Home

Hanna Instruments Thermometer HI 9063 User's Manual

image

Contents

1. Connect a K type thermocouple probe to the meter Vu To switch on press the ON OFF key on the front of the meter LS If the meter does not come on make NOFE sure that the batteries are correctly installed The thermometer will carry out a self diagnostic test routine the LCD will show all segments for a few seconds or as long as ON OFF is held followed by the per centage indication of the remaining e battery life The thermometer then enters normal measurement mode If a temperature probe is plugged in the meter displays the measured temperature XU r L If no probe is plugged in or if reading deg is over range the display shows flash ing dashes If a measurement is slightly over the range of the meter specifications the display flashes the closest full scale value To switch the thermometer OFF press the ON OFF key MEASURING SCALE C F Theinstrumentis factory settothe Cscale but measurements can be performed in either the Celsius or Fahrenheit scale Press the C F button toselectthe desired scale CIF HOLD MODE The HOLD function is activated by press ing the HOLD key heed The measured temperature is held on the display until HOLD is pressed again The H tag blinks on the display to indicate the HOLD mode Note Although the display is frozen ale the meter continues measuring and updating Hi and Lo values HIGH LOW TEMPERATURES The maximum and minimum temp
2. HI 766B3 Spring loaded small surface probe with insulated shaft max 200 C 390 F HI 766C Penetration probe max 9007 1650 F HI 766C1 Ultra fast Penetration probe max 300 C 570 F HI 766D Air probe max 3007C 570 F HI 766E1 General purpose probe max 900 C 1 650 F HI 766E2 General purpose probe max 9007 1650 F HI 766F High temperature flexible wire probe without handle max 1100 C 2000 F HI 766F1 Flexible wire probe without handle max 480 900 F HI 766TR1 Penetration probe max 250 C 482 F HI 766TR2 Penetration long probe max 2507 482 F HI 766TV1 Pipe clamp probe max 2007 390 F with detachable handle amp mini connector to be used in conjunction with the HI 766HD probe handle HI 766PA Roller surface probe max 320 C 600 F HI 766PB Surface probe max 650 C 1200 F HI 766PC Penetration probe max 900 C 1650 F HI 766PD Air probe max 3007C 570 F HI 766PE1 General purpose probe max 900 C 1 650 F HI 766PE2 General purpose probe max 900 C 1 650 F grill surface probe HI 766B4 Grill surface probe with 70 cm 27 6 cable protected with stainless steel jacket max 250 C 482 F HI 7664B4S Spare stainless steel sensor for HI766B4 probe OTHER ACCESSORIES HI 710021 Spare protective case HI 721317 Rugged carrying case HI 766EX Extension cable for K type probes HI 766HD Rugged thermocouple probe handle with 1m 3 3 cable fitted with mini connector CE DECLARATION OF
3. meter If you need additional technical information do not hesitate to e mail us at tech hannainst com This instrument is in compliance with the directives PRELIMINARY EXAMINATION Remove the instrument from the packing material and exam ine it carefully to make sure that no damage has occurred during shipping If there is any noticeable damage notify your Dealer or the nearest Hanna Office Note Save all packing materials until you are sure that the instrument functions correctly Any defective item must be returned in the original packaging together with the supplied accessories GENERAL DESCRIPTION HI 9063 is a waterproof microprocessor based K type ther mocouple thermometer which provides very accurate measurements in a wide range of temperatures The meter is also provided with low battery detection and BEPS Battery Error Preventing System which turns the unit off when the batteries are discharged avoiding erroneous readings caused by low battery level HI 9063 features include autoranging capability dual level LCD for simultaneously displaying of maximum and minimum measured temperatures C F selection button and hold function The meter is supplied complete with four 1 5V AA batteries and instruction manual FUNCTIONAL DESCRIPTION 2 HI 9063 K thermocouple Thermometer 3 Autoranging 2 CLEAR C F HOLD 1 Temperature probe connector 2 Liquid Crystal
4. CONFORMITY BIEN instruments ce DECLARATION OF CONFORMITY pu struments Italia Srl 35030 8 eola di Rubano PD ITALY herewith certify that the K type thermometer HI 9063 has been tested and found to be in ci oral nis e with EMC Directive s B AAGIHEC and Low Voltage Directive 73 23 EEC ble normatives EN 50082 1 Electroma TEC 61000 IEC 61000 4 3 RF Radiated EN 50081 1 n ctromagnetic Compatibility Generic Emission Standard N 55022 Radiat ENGI010 1 Safety requi control and labora tory se mity Standard electrical equipment for measurement Date of Issue 24 10 2002 LL A Marsilio Technical Director of nts S r l Recommendations for Users Before using this product make sure that it is entirely suitable for the environment in which it is used Operation of this instrument in residential areas could cause unacceptable interference to radio and TV equipment requiring the operator to take all necessary steps to correct interference Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid electrical shock do not use these instruments when voltage at the measurement surface exceeds 24 Vac or 60 Vdc To avoid damage or bums do not perform any measurement in microwave ovens IST9063R1 06 05
5. Display 3 ON OFF key 4 Measuring unit selection key 5 HOLD key 6 HI LO values reset key Ern D eo LE LOLCL 1 Current temperature value 2 HOLD indicator 3 Low battery indicator 4 Minimum temperature value 5 Maximum temperature value 6 K type probe indicator 7 Measuring unit C or F SPECIFICATIONS Range 50 0 to 1350 C 58 0 to 2462 F Resolution 0 1 C up to 199 9 C 1 C outside 0 1 F up to 399 9 F 1 F outside Accuracy 0 2 F S for 1 year excluding probe error 3 C 6 F Typical EMC Deviation with HI 766 probes Battery Type 4 x 1 5V AA IEC LR6 Life approx 2000 hours of continuous use Probe K type thermocouple see Accessories Environment 10 to 50 C 14 to 122 F RH 100 Dimensions 196 x 80 x 60 mm 7 7x3 1x2 4 Weight 500 g 1 1 Ib Range may be limited by probe FACTORY RECALIBRATION All Hanna Instruments thermometers have been accurately pre calibrated at the factory It is generally recommended to have all thermometers recalibrated at least once a year For an accurate annual recalibration contact your nearest Hanna Service Center OPERATIONAL GUIDE INITIAL PREPARATION f Each meter is supplied complete with batteries gt Remove the back cover unwrap the 9 batteries and install them while pay ing attention to their polarity
6. Instruction Manual HI 9063 Portable Waterproof Microprocessor K Type Thermocouple Thermometer JR www hannainst com WARRANTY All Hanna Instruments meters are warranted for two years against de fects in workmanship and materials when used for their intended purpose and maintained according to instructions Sensors and probes are war ranted for a period of six months This warranty is limited to repair or replacement free of charge Damages due to accidents misuse tampering or lack of prescribed mainte nance are not covered If service is required contact the dealer from whom you purchased the instru ment If under warranty report the model number date of purchase serial number and the nature of the foilure First obtain a Returned Goods Authorization number from the Customer Ser ice department then return the instrument with the Authorization included along with shipment costs prepaid If the repair is not covered by the war ranty you will be notified of the charges When shipping any instrument make sure it is properly packaged for com plete protection 3 5 Hanna Instruments reserves the ight to modify the design construction and appearance of its products without advance notice Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for correct operation Please read it carefully before using the
7. era tures are continuously monitored and displayed in the lower portion of the LCD Note When reading goes over range the Hi and Lo values display dashes until cleared CLEAR FUNCTION Upon pressing the CLEAR key the current reading is assigned to the highest and lowest temperature val ues The High Low values may be cleared at any time during measure ment BATTERY REPLACEMENT The instrument is powered by four 1 5 V alkaline batteries and is provided with the Battery Error Prevention System BEPS which turns the unit off when a low battery signal is detected When the remaining battery level is less than 10 a warning symbol blinks on the display to indicate a low battery condition It is recommended to replace the batteries as soon as the low battery condition is detected Battery replacement must only take place in a nonhazardous area using four 1 5V alkaline batteries In order to replace rundown batteries simply remove the two screws on the rear cover of the instrument and replace the four batteries with new ones paying attention to the correct polarity Reattach the cover and tighten the two screws ACCESSORIES K TYPE THERMOCOUPLE PROBES with integral handle 1 m 3 3 cable amp mini connector HI 766A Roller surface probe max 320 C 600 F HI 766B Surface probe max 650 C 1200 F HI 766B1 90 Surface probe max 450 040 F HI 766B2 Spring loaded surface probe max 900 1 650 F