Home
Barco Coronis 5MP
Contents
1. Keep your original packaging It is designed for this display and is the ideal protection during transport and storage B4100592 CORONIS 5MP 05 08 2014 _ S 1 7 Welcome 2 OB 4100592 CORONIS SMP 05 08 2014 2 Parts controls and connectors 2 PARTS CONTROLS AND CONNECTORS 2 1 Display front view Overview 0006060 Image 2 1 Front view Luminate Left key Right key Menu key Power on off key Ambient light sensor Power LEDs Side downstream USB 2 amp 3 B4100592 CORONIS 5MP 05 08 2014 2 Parts controls and connectors 2 2 Display rear view Connector compartment cover closed Image 2 2 Rear view cover closed Connector compartment cover Display release button for portrait position Mounting anchor Display release button for landscape position Connector compartment cover open Image 2 3 Rear view cover open 24 Vdc power input 5 Vdc 0 5 A power output Display Port video input DVI video input Ethernet connection see note below USB upstream USB downstream 1 4 D B4 100592 CORONIS 5MP 05 08 2014 2 Parts controls and connectors lt The Ethernet connection is used for maintenance purposes and is not supported for user application 2 3 Foot Single foot Image 2 4 Tilt amp Swivel neck cover Cable duct covers Dual foot Image 2 5 Dual tilt and swivel foot Cable cover Left amp right arm cover Foot lock cl
2. 30 dip in Ur 70 U 30 dip in Ut 5MP requires continued for 25 cycles for 25 cycles operation during power A ban ait 6 East mains interruptions it is eae ae GER reine aaa dip in recommended that the T t Coronis 5MP be powered from an uninterruptible power supply or a battery Power frequency 50 60 Not applicable Power frequency Hz magnetic field magnetic fields should IEC 61000 4 8 be at levels characteristic of a typical location in a typical commercial or hospital environment Conducted RF 3 Vrms 3 V Portable and mobile IEC 61000 4 6 150 kHz to 80 MHz RF communications 40 Ur 60 dip in Ur 40 Ur 60 dip in Ur for 5 cycles for 5 cycles 1 is the a c mains voltage prior to application of the test level 2 Coronis 5MP doesn t contain susceptible components to magnetic fields 40 WO B44100592 CORONIS SMP 05 08 2014 Immunity test IEC 60601 Compliance level Test levels Radiated RF 3 V m IEC 61000 4 3 80 MHz to 2 5 GHz At 80 MHz and 800 MHz the higher frequency range applies Important information Electromagnetic environment guidance equipment should be used no closer to any part of the Coronis 5MP including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P d 1 2VP 80 MHz to 800 MHz d 2 3VP 800 MHz to 2 5 Ghz Where P is the maximum
3. About Continuous ALC X Continuous ALC can only be selected when the DICOM display function is selected Therefore please refer to Display functions page 26 to correctly set the display func tion Enabling continuous ALC will continuously recalculate the DICOM display function taking the averaged ambient light into account To select continuous ALC 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Calibration gt Ambient Light menu 3 Enter the Continuous ALC submenu 4 Select Enabled Disabled as desired and confirm 5 18 Image scaling About image scaling Enabling image scaling will multiply each individual pixel to one or more adjacent pixels so that the size of the displayed image will be a multiple of the original image source video input signal Image scaling is only possible when the resolution of your display s video input signal is less than or equal to half the maximum resolution of the display To enable disable image scaling 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Image Source menu 3 Enter the Scaling submenu 4 Select Enabled Disabled as desired and confirm 28 B4100592 CORONIS 5MP 05 08 2014 5 Advanced operation 5 19 Display orientation About Display orientation Your display automatically detects its physical orientation portrait or landscape and by default automat ically adjusts the image orientation to this This means that
4. 2014 7 Important information 7 7 Technical specifications Overview Screen technology TFT AM LCD Dual Domain IPS Active screen size diagonal 540 9 mm 21 3 Active screen size Hx V 422 4 mm x 377 9 mm 16 5 x 13 3 Aspect ratio H V 5 4 Resolution 5 MP 2560 x 2048 Pixel pitch 0 165 mm H x 0 165 mm Color mans Gray mage Number of grayscales LUT 1024 gray levels 10 12 in LUT out Viewing angle H V 170 Uniform Luminance Yes Technology ULT Per Pixel Uniformity TPPU Ambient Light Compensation Yes ALC Backlight Output Stabilization Yes BLOS Maximum luminance 1100 cd m panel typical DICOM calibrated luminance 500 cd m ULT off Contrast ratio ULT off 850 1 typical Response time Tr Tf 36 ms typical Scanning frequency H V 30 150 kHz 15 80 Hz Rousing color fow Video input signals DVI D Dual Link DisplayPort per display Video inout terminals USB ports 1 upstream endpoint 3 downstream USB standard 2 0 Power requirements nominal 100 240V Power consumption nominal ower management DVI DMPM DP DMPM 280 MHz English German French Dutch Spanish Italian Portuguese Polish Russian Swedish Chinese simplified Japanese Korean Arabic Dimensions with stand W x H Portrait w x hmax x d 392 x 658 2 x 110 2 mm w x hmin x d 392 x x D 558 2 x 110 2 mm Landscape w x hmax x d 484 7 x 565 5 x 110 2 mm w x hmin x d 484 7 x 465 5
5. Assurance providing maximum diagnostic confidence and uptime Barco recommends to install MediCal QAWeb Agent and apply the default QAWeb policy ee at least This policy includes calibration on regular intervals Connecting to MediCal QAWeb Server offers even more possibilities Learn more and sign up for the free MediCal QAWeb Essential level at www barco com healthcare qa 4 2 Key indicator lights About the key indicator lights By default the indicator lights of the keys will be dimmed which makes the keys unavailable at that mo ment To make the keys illuminate and available for further actions touch one of the keys As a result all keys will be illuminated and are now available for further actions However if no further actions are taken within the following 5 seconds the keys will dim again The key auto dim function can be disabled in the OSD menus Please refer to Key in dicator lights page 22 for detailed instructions on how to do this 4 3 Standby switching 18 About standby switching The connected power supply also provides a switch that can be used to turn the power completely off To use the display please make sure to switch on this power supply This can be done by pushing the on off switch on the power supply into the position Switching on the display while it is in standby mode or vice versa can be done by 1 Illuminate the keys as previously described 2 While the keys are illumin
6. Installation Place the device on a flat solid and stable surface that can support the weight of at least 3 devices If you use an unstable cart or stand the device may fall causing serious injury to a child or adult and serious damage to the device This apparatus conforms to CE0120 MDD 93 42 EEC A1 2007 47 EC class llb product CE 2004 108 EC CE 93 42 EEC A1 2007 47 EC class Il b IEC 60950 1 2005 A1 2009 2ND EDITION IEC 60601 1 2ND ED 1988 A1 1991 A2 1995 UL 60601 1 1ST EDITION CAN CSA C22 2 NO 601 1 M90 2005 IEC 60601 1 2005 A1 2012 ANSI AAMI ES 60601 1 2005 C1 2009 A1 2012 CAN CSAC22 2 No 60601 1 08 2008 DEMKO EN 60601 1 2006 EN 60601 1 2 2007 CCC GB9254 2008 GB4943 2001 GB17625 1 2003 KC VCC FCC class B ICES 001 Level B FDA 510K RoHS National Scandinavian Deviations for CL 1 7 2 Finland Laite on liitettava suojamaadoituskoskettimilla varustettuun pistorasiaan Norway Apparatet ma tilkoples jordet stikkontakt Sweden Apparaten skall anslutas till jordat uttag 7 2 Environmental information 36 Disposal Information Waste Electrical and Electronic Equipment H E This symbol on the product indicates that under the European Directive 2012 19 EU governing waste from electrical and electronic equipment this product must not be disposed of with other municipal waste Please dispose of your waste equipment by handing it over to a designated collection point for the recycli
7. The RoHS of Chinese Mainland is included in the MCV standard of the Ministry of Information Industry of China in the section Limit Requirements of toxic substances in Electronic Information Products 34 8 4 F ABAS MARTH AE ee eee nn Printed Circuit Assemblies FR oe HA Mik LCD panel i e FB Be Bt Plastic enclosure BY FA H 3 Heatsinks B4100592 CORONIS 5MP 05 08 2014 37 7 Important information 3 4 8 4 ABBAS MARKER Component name Hazardous substances and elements Gs a gz A ZRKA SR IX Pb Hg Cd Cr6 PBB ik PBDE Pp of a EB JR HE AY as Power Supply Unit i ie ae X iF iH BA Paper Manuals HA BAB CD manual O ATRAASASMREZREM AYRE PHS SWE SIT 11363 2006 THEMEN RSERUAE O Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement in SJ T11363 2006 X RRBASASOREVEBZRAMKR WRMA PHS SMM SIT 11363 2006 EM EH REZX X Indicates that this toxic or hazardous substance contained in at least one of the homogeneous materials used for this part is above the limit requirement in SJ T11363 2006 fp BA Bi A BS A oh EIP AD AR AR BEA ma RE IRE et Wh E Ep GR se FA ABR EFUP R Barco ATR A JEFUPRE SERB HN A AB AY S OE FA Fil EF m SEF Op BA EA MARIE SRE DR HE All Electronic Information Products EIP that are sold within Chinese Mainland must comply with the Electronic Information Products P
8. is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer of the user of the Coronis 5MP can help prevent electromagnetic interfer ence by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Coronis 5MP as recommended below according to the maximum output power of the communications equipment Rated maximum output Separation distance according to frequency of transmitter power of transmitter 150kHz to 80MHz 80MHz to 800MHz 800MHz to 2 5GHz d 2 3 P 0 12 0 23 0 73 2 3 23 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies x These guidelines may not apply in all situations Electromagnetic propagation is af fected by absorption and reflection form structures object and people 7 5 Explanation of symbols Symbols on the device On the device or power supply you may find the following symbols nonrestrictive list C Indicates compliance with the Directive 93 42 EEC as Class device C Indicates compliance with the Directive 93 42 EEC as Class II device 0120 FE Indicates compliance with Part 15 of the FCC rules Class A or Class B 5 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter Where P is th
9. it from the AC inlet to avoid damage by transient over voltage To fully disengage the power to the device please disconnect the power cord from the AC inlet B4100592 CORONIS SMP 05 08 2014 __ 35 7 Important information Power cords e Utilize a UL listed detachable power cord 3 wire type SJ or equivalent 18 AWG min rated 250 V min provided with a hospital grade type plug 5 15P configuration for 120V application or 6 15P for 240V application Do not overload wall outlets and extension cords as this may result in fire or electric shock e Mains lead protection U S Power cord Power cords should be routed so that they are not likely to be walked upon or pinched by items placed upon or against them paying particular attention to cords at plugs and receptacles e The power supply cord should be replaced by the designated operator only at all time e Use a power cord that matches the voltage of the power outlet which has been approved and complies with the safety standard of your particular country Water and moisture Never expose the display to rain or moisture Never use the display near water e g near a bathtub washbasin swimming pool kitchen sink laundry tub or in a wet basement Ventilation Do not cover or block any ventilation openings in the cover of the set When installing the device in a cupboard or another closed location heed the necessary space between the set and the sides of the cupboard
10. light allowed in this type of room e the preset ambient light correction value for this reading room These parameters are stored in your display and determine the preset ambient light correction value to take into account to recalculate the DICOM display function when Ambient Light Compensation ALC is enabled Please refer to Ambient Light Compensation ALC page 27 to enable ALC The available reading rooms for your Coronis 5MP are CR DR MAMMO Corresponds to light conditions in diagnostic reading rooms for computed radiology digital radiology or mammography This setting has the lowest maximum ambient light CT MR NM Corresponds to light conditions in diagnostic reading rooms for computed tomography magnetic resonance or nuclear medicine scans Staff Office Corresponds to light conditions in office rooms Clinical Viewing Room Corresponds to light conditions in diagnostic reading rooms for clinical viewing Emergency Room Corresponds to light conditions in emergency rooms Operating Room Corresponds to light conditions in operating rooms This setting has the highest maximum ambient light B4100592 CORONIS 5MP 05 08 2014 27 5 Advanced operation To select a reading room 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Calibration gt Ambient Light menu 3 Enter the Reading Room submenu 4 Select one of the available reading rooms and confirm 5 17 Continuous ALC
11. output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range 4 Interference may occur in the vicinity of equipment marked with symbol Cp 3 Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Coronis 5MP is used exceeds the applicable RF compliance level above the Coronis 5MP should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the Coronis 5MP 4 Over the frequency range 150 kHz to 80 MHZ field strengths should be less than 3 V m B4100592 CORONIS 5MP 05 08 2014 41 7 Important information amp These guidelines may not apply in all situations Electromagnetic propagation is af fected by absorption and reflection from structures objects and people Recommended separation distance The Coronis 5MP
12. the top of the cover slightly away from the display and lift the cover upwards piin Removal of the cable covers of a single foot To get access to the cable routing channel of a single foot the cable covers should be removed To remove them proceed as follows 1 Remove the tilt amp swivel neck cover 1 Therefore first tilt the display panel in upright position Then lift and pull backward the neck cover 2 Also to facilitate cable routing afterwards open both cable duct covers 2 B4100592 CORONIS 5MP 05 08 2014 3 Display installation Removal of the cable cover of a dual foot To get access to the cable routing channel of a dual foot the display stand cable cover should be removed To remove this cover proceed as follows 1 Gently press and hold the lock clips at the inside of the top of the cover 2 Slide the cover upwards on the shaft while holding the clips pressed It may be needed to tilt the displays if the arm is blocking the movement of the cover yo 3 4 Connecting the signal cables CAUTION Only connect one of the two video links Connecting both inputs simultane ously will result in driver errors B4100592 CORONIS 5MP 05 08 2014 _ 13 14 3 Display installation To connect the signal cables to the display You may connect the display to a display controller by using the DVI or the DP connection The active input can be selected in the OSD menu 1 Connect the DVI or DP
13. to comply with the guidelines set forward by the latest medical guidelines The available video encoding modes for your display are listed below x Please note that the display controller connected to your display might not always sup port all these video encoding modes B4100592 CORONIS 5MP 05 08 2014 __ _ SSF 29 5 Advanced operation Standard This mode uses the standard 8 bit encoding mechanism HDR1 This mode features 10 bit video encoding following the High Dynamic Range HDR standard usually used on color displays HDR2 HDR2 is a Barco specific mode featuring 10 bit video encoding following the High Dynamic Range HDR standard but allowing full refresh rate which is not always possible with HDR1 This mode is usually used on color displays Dual 10 only with Barco display controller Dual 10 is a Barco specific 10 bit video encoding mode which is usually used on grayscale displays and which allows full refresh rate Automatic When selecting this option your Coronis SMP will automatically assign the correct video encoding settings to the connected video input signals To manually select a video encoding mode 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Image Source menu 3 Enter the Video Encoding submenu 4 Select one of the available video encoding modes and confirm 5 22 Grayscale conversion modes 30 About grayscale conversion modes Grayscale conversion modes specif
14. x 110 2 mm Dimensions w o stand W x H 392 x 484 7 x 110 2 mm x D Dimensions packaged W x H 779 x 542 x 255 mm x D Net weight with stand 15 3 kg B4100592 CORONIS 5MP 05 08 2014 D 45 7 Important information Net weight w o stand 9 3 kg stand TT Certifications CE0120 MDD 93 42 EEC A1 2007 47 EC class Ilb product CE 2004 108 EC CE 93 42 EEC A1 2007 47 EC class II b IEC 60950 1 2005 A1 2009 2ND EDITION IEC 60601 1 2ND ED 1988 A1 1991 A2 1995 UL 60601 1 1ST EDITION CAN CSA C22 2 NO 601 1 M90 2005 IEC 60601 1 2005 A1 2012 ANSI AAMI ES 60601 1 2005 C1 2009 A1 2012 CAN CSAC22 2 No 60601 1 08 2008 DEMKO EN 60601 1 2006 EN 60601 1 2 2007 CCC GB9254 2008 GB4943 2001 GB17625 1 2003 KC VCC FCC class B ICES 001 Level B FDA 510K RoHS Quick Installation Sheet Video cables 1 x DVI Dual Link 1 x DisplayPort Main cables UK European CEBEC KEMA USA UL CSA adaptor plug NEMA 5 15P Chinese CCC Warrant Operating humidity 8 70 non condensing Storage humidity 5 95 non condensing Operation altitude 3000 m torage altitude 7500 m 7 8 Open source license information 46 Open source license information Open source license usage This product contains software components released under an Open Source license A copy of the source code is available on request by contacting your Barco customer support representative EACH SEPARATE OPEN SOURCE SO
15. 5MP should assure that it is used in such an environment Re ene Electromagnetic environment Guidance RF emissions Group 1 The Coronis 5MP uses RF energy only for its internal function erie Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment CISPR 11 ee je as Harmonic emissions Class D PEO ARMES ee AISNE and those directly connected to IEC 61000 3 2 _ the public low voltage power Voltage fluctuations flicker Complies supply network that supplies emissions buildings used for domestic IEC 61000 3 3 purposes B4100592 CORONIS 5MP 05 08 2014 39 7 Important information This Coronis 5MP complies with appropriate medical EMC standards on emissions to and interference from surrounding equipment Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received including interfer ence that may cause undesired operation Interference can be determined by turning the equipment off and on If this equipment does cause harmful interference to or suffer from harmful interference of surrounding equipment the user is encouraged to try to correct the interference by one or more of the following mea sures e Reorient or relocate the receiving antenna or equipment e Increase the separation between the equipment and receiver Connect the equipment into an outle
16. Coronis SMP User Guide MDCG 5121 BARCO B4100592 03 05 08 2014 visibly yours Barco nv President Kennedypark 35 8500 Kortrijk Belgium Phone 32 56 23 32 11 Fax 32 56 26 22 62 Support www barco com esupport Visit us at the web www barco com Printed in Belgium 7 Welcome 1 WELCOME 1 1 About the product Overview The high grade Coronis 5MP display system offers unsurpassed film like images with ultra high resolution luminance and contrast The 21 3 LCD provides unequaled image quality without pixelization a wide viewing angle and perfect geometry Building on proven technology Coronis 5MP provides ultimate precision and performance while main taining the highest brightness and contrast The advanced 10 bit architecture ensures accurate image representations with 1024 simultaneous shades of gray out of a palette of 4096 CAUTION Read all the important safety information before installing and operating your Coronis 5MP Please refer to the dedicated chapter in this user guide 1 2 What s in the box Overview Your Coronis 5MP comes with e the display e a system tilt and swivel foot e this Coronis 5MP user guide e asystem CD e a DisplayPort cable e a DVI cable e an USB cable e an AC power cord an external power supply If you ordered a Barco display controller it s also in the box together with its accessories A dedicated user guide is available on the system CD
17. D main menu 2 Navigate to the Configuration gt User Interface gt Controls menu 3 Enter the Sound submenu 4 Select Enabled Disabled as desired and confirm 5 8 Ethernet over USB About Eth over USB By default the Ethernet over USB will be disabled This does not disable the USB hub and still allows to make use of the display s USB downstream connectors and any external devices connected to it key board mouse Please note that a connection from a PC USB downstream connector to the display s USB upstream connector is required to be able to use this functionality By enabling the Ethernet over USB you allow the internal Coronis 5MP processor to communicate directly over USB with the connected PC Enabling the Ethernet over USB is only recommended when you are using a non Barco display controller If you are using a Barco display controller this communication will automatically be done over the connected video cable s To enable disable the eth over USB 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Connectivity menu 3 Enter the Eth over USB submenu 4 Select Enabled Disabled as desired and confirm 5 9 USB About USB By default USB will be enabled This will allow the connected PC to communicate directly over USB with the internal Coronis 5MP processor By disabling USB communication between the internal Coronis 5MP processor and the connected PC will not be possibl
18. FTWARE COMPONENT AND ANY RELATED DOCUMENTA TION ARE PROVIDED AS IS WITHOUT EXPRESS OR IMPLIED WARRANTY INCLUDING BUT NOT B4100592 CORONIS 5MP 05 08 2014 7 Important information LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICU LAR PURPOSE IN NO EVENT SHALL THE COPYRIGHTHOLDER OR ANY OTHER CONTRIBUTOR BE LIABLE FOR DIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGES MORE INFORMATION DETAILS IS TO BE FOUND IN EACH SPE CIFIC OPEN SOURCE LICENSE Copyright on each Open Source Software component belongs to the respective initial copyright holder each additional contributor and or their respective assignee s as may be identified in the respective doc umentation source code README file or otherwise You shall not remove or obscure or otherwise alter the respective copyrights You acknowledge living up to the conditions of each separate Open Source Software license A list of the Open Source Software components used is available in the applicable EULA through the customer section of the Barco website or through other online means B4100592 CORONIS SMP 05 08 2014 ___ 47
19. ated touch the standby key for approximately 2 seconds As a result the display will switch on or will switch to standby mode When the display is switched on the power LED is blue When the display is in standby mode the power LED is orange The power LED can be disabled in the OSD menus Please refer to Power status indi cator light page 22 for detailed instructions on how to do this B4100592 CORONIS 5MP 05 08 2014 4 Daily operation x In case of a power outage recovery your display will always start up in the power mode it was in before the power interruption i e standby or on This protects your display against inadvertent image retention problems 4 4 Bringing up the OSD menus How to bring up the OSD menus The OSD menu allows you to configure different settings to make your Coronis 5MP fit your needs within your working environment Also you can retrieve general information about your display and its current configuration settings through the OSD menu Bringing up the OSD menus can be done by 1 If not already done so switch on the display as previously described 2 Illuminate the keys as previously described 3 While the keys are illuminated touch the menu key As aresult the OSD main menu comes up in the center of the screen However if no further actions are taken within the following 90 seconds the OSD will disappear again The OSD menu auto exit function can be disabled in the OSD menu Please ref
20. e This does not disable the USB hub and still allows to make use of the display s USB downstream connectors and any external devices connected to it keyboard mouse Please note that a connection from a PC USB downstream connector to the display s USB upstream connector is required to be able to use this functionality B4100592 CORONIS SMP 05 08 2014 23 5 Advanced operation amp Enabling the USB endpoint is only recommended when you are using a non Barco dis play controller If you are using a Barco display controller this communication will au tomatically be done over the connected video cable s To enable disable USB 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Connectivity menu 3 Enter the USB submenu 4 Select Enabled Disabled as desired and confirm 5 10 DPMS mode About DPMS mode Enabling the Display Power Management System DPMS mode on your display will optimize its diagnos tic lifetime by automatically switching off the backlight when the display is not used for a specified period of time By default DPMS mode is enabled on your display but it also needs to be activated on your workstation To do this go to the Power options properties window of your workstation Barco recommends setting DPMS activation after 20 minutes of non usage Ver When DPMS mode is enabled on your display an additional OSD power saving function becomes available hibernate Please refer t
21. e maximum output power rating of the transmitter in watts W according to the transmitter manufacturer 42 DD B44100592 CORONIS SMP 05 08 2014 7 Important information oil Indicates the device is approved according to the UL regulations il Indicates the device is approved according to the UL regulations for Canada and US a Indicates the device is approved according to the UL Demko regulations Indicates the device is approved according to the CCC regulations Indicates the device is approved according to the VCCI regulations Indicates the device is approved according to the KC regulations Indicates the device is approved according to the BSMI regulations Indicates the USB connectors on the device Indicates the DisplayPort connectors on the device Indicates the manufacturing date BWAASVDAGeOCS Indicates the temperature limitations for the device to safely operate within specs x x lt a Indicates the device serial no Warning dangerous voltage Caution Consult the operating instructions BE gt Eg 6 Values for xx and yy can be found in the technical specifications paragraph B4100592 CORONIS 5MP 05 08 2014 _ 43 7 Important information Indicates this device must not be thrown in the trash but must be recycled according to the European WEEE Waste Electrical and Electronic Equipment directive ee Indicates Direct Current DC Indicates Alternating Current AC Sta
22. e the color depth to 8 or to 10 bit To manually set EDID timings 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Image Source menu 3 Enter the Timings submenu 4 Select Refresh Rate Preferred Orientation or Color Depth 5 Select one of the available settings and confirm 5 24 Display info About display info Your display serial number color type native resolution firmware versions etc are available in a dedi cated submenu of the OSD menu To retrieve info about your display 1 Bring up the OSD main menu 2 Navigate to the About this Display menu to make the information visible on the screen 5 25 Display status About display status The Status submenu of the OSD menu provides info on the current status of your display runtimes tem peratures etc the status of the connected image sources video encoding mode timings etc the current calibration status of your display display function luminance ALC etc and the status about activated connections To retrieve the status of your display 1 Bring up the OSD main menu 2 Navigate to the Status menu 3 Enter the Display Image Sources Calibration or Connectivity submenu as desired B4100592 CORONIS 5MP 05 08 2014 31 5 Advanced operation 32 B4100592 CORONIS 5MP 05 08 2014 6 Cleaning your display 6 CLEANING YOUR DISPLAY 6 1 Cleaning instructions To clean the display Clean the display using a
23. e to the Configuration gt Calibration menu 3 Enter the Display Function submenu 4 Select one of the available display functions and confirm B4100592 CORONIS 5MP 05 08 2014 5 Advanced operation 5 15 Ambient Light Compensation ALC About ALC COM display function is selected Therefore please refer to Display functions page Ambient Light Compensation ALC can only be enabled on your display when the DI 26 to correctly set the display function When ALC is enabled the DICOM display function will be recalculated taking a preset ambient light cor rection value into account This value is determined by the selected reading room Therefore it is also important to select a realistic reading room when enabling ALC This can be done by following the instruc tions in Reading rooms page 27 To enable disable ALC 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Calibration gt Ambient Light menu 3 Enter the Ambient Light Compensation submenu 4 Select Enabled Disabled as desired and confirm 5 16 Reading rooms About reading rooms Reading rooms can only be selected when the DICOM display function is selected Therefore please refer to Display functions page 26 to correctly set the display function The American Association of Physicists in Medicine AAPM composed a list of pre defined reading rooms Each of these reading rooms are defined by following parameters e the maximum
24. er to OSD menu automatic close function page 21 for detailed instructions on how to do this The OSD menu position can be modified in the OSD menu Please refer to OSD menu position page 21 for detailed instructions on how to do this 018 e 4 5 Navigating through the OSD menus How to navigate through the OSD menus Navigating through the OSD menus can be done by e Use the up down keys to move through the sub menus change values or make selections Togo into a submenu or confirm adjustments and selections use the menu key e Us the standby key to cancel adjustments or exit a sub menu Exit all OSD menus at once by touching the standby key for approximately 2 seconds B4100592 CORONIS SMP 05 08 2014 19 4 Daily operation 20 B4100592 CORONIS 5MP 05 08 2014 5 Advanced operation 5 ADVANCED OPERATION 5 1 OSD menu position About the OSD menu position By default the OSD menu comes up in the bottom center position of the screen This position can be modified however There s choice from a number of predefined positions To change the position of the OSD menu 1 Bring up the OSD main menu 2 Navigate to the Configuration gt User Interface gt Menu menu 3 Enter the Position submenu 4 Select one of the available OSD menu positions and confirm 5 2 OSD menu language About the OSD menu language By default the OSD menu comes up in English However there s a wide range of o
25. ill adjust its backlight to reach the target To set the luminance target 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Calibration menu 3 Enter the Luminance Target submenu 4 Set a luminance target value as desired and confirm The default factory DICOM calibrated luminance value is available in the technical spec ifications table The guaranteed backlight lifetime is valid for this setting 5 13 Viewing modes About viewing modes The Coronis 5MP can be used in two viewing modes Diagnostic This mode provides the full calibrated luminance and is intended for using the display for diagnostic purposes Text In this mode the luminance is reduced to approximately half of the luminance This is intended for using the display with office applications such as word processing Please note that text mode is not persistent once powered off the unit will restart in diagnostic mode B4100592 CORONIS 5MP 05 08 2014 __ _ SSF 25 5 Advanced operation amp As the Coronis 5MP is intended to be used in a diagnostic environment the diagnostic mode should always be selected To select a viewing mode 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Calibration menu 3 Enter the Viewing Mode submenu 4 Select Diagnostic Text as desired and confirm 5 14 Display functions 26 About display functions Native uncorrected panels will display all grayscale colo
26. ip behind cable cover Display positioning sliders B4100592 CORONIS SMP 05 08 2014 5 2 Parts controls and connectors 6 n B4100592 CORONIS SMP 05 08 2014 3 Display installation 3 DISPLAY INSTALLATION Prior to installing your Coronis 5MP and connecting all necessary cables make sure to have a suitable display controller physically installed in your computer If you are using a Barco display controller please consult the user guide delivered with it to do this 3 1 Mounting the displays WARNING Do not change the display orientation by turning the panel when it is mounted on its stand To change from landscape to portrait or the other way around remove the panel from its stand and use the appropriate mounting hooks as described above Using the single tilt amp swivel foot portrait orientation 1 Unpack the single tilt amp swivel foot and place it on a solid surface 2 Unpack the display 3 Hook the top gaps in the rear of the panel over the top anchors of the stand attachment plate 1 4 Push the bottom of the panel firmly towards the stand so that the bottom anchors clip in the bottom gaps 2 5 Check that the panel is firmly attached to the stand 6 Tilt and swivel the panel to the desired position and angle B4100592 CORONIS SMP 05 08 2014 7 3 Display installation SE Using the single tilt amp swivel foot landscape orientation 1 Follow the same steps as for the portrait instal
27. isplay If you have properly installed your display and display controller the Windows start up messages will appear once the boot procedure is finished Your Coronis 5MP will be running in a basic video mode at a default refresh rate when first time starting up If you are using a Barco display controller please consult the user guide delivered with it to install the drivers software and documentation When this is done your display will automatically detect the connected video input signal s and apply the correct video mode and refresh rate B4100592 CORONIS 5MP 05 08 2014 4 Daily operation 4 DAILY OPERATION 4 1 Recommendations for daily operation Optimize the lifetime of your display Enabling the Display Power Management System DPMS of your display will optimize its diagnostic life time by automatically switching off the backlight when the display is not used for a specified period of time By default DPMS is enabled on your display but it also needs to be activated on your workstation To do this go to Power Options Properties in the Control Panel Barco recommends setting DPMS activation after 20 minutes of non usage ce Use a screen saver to avoid image retention Prolonged operation of an LCD with the same content on the same screen area may result in a form of image retention You can avoid or significantly reduce the occurrence of this phenomenon by using a screen saver You can activate a
28. lation but hold the display in landscape position to hook the panel on the foot To release the panel from the display foot 1 Push the push button 3 at the rear of the panel 8 B4100592 CORONIS 5MP 05 08 2014 3 Display installation 2 Pull the bottom of the panel towards you and at the same time lift the panel from the stand attachment plate Using the dual tilt amp swivel foot 1 Unpack the dual tilt amp swivel foot and place it on a solid surface 2 Unpack the first display B4100592 CORONIS 5MP 05 08 2014 3 Display installation 10 Hold the display slightly angled with the logo downwards and position the top of the mounting anchor at the back of the display perfectly aligned on the top of the mounting clamp of the foot Push the downside of the display gently backwards until the mounting clamp locks into the mounting anchor Verify that mounting clamp and mounting anchor are perfectly fitting and that the display is firmly mounted Repeat step 2 to 5 for the second display Caution Limit transportation distances of the mounted displays and foot system 7 Pull out the foot lock clip 8 Retain the clip for possible further transportations 9 Now that the height positioning is unlocked you can adjust the position of the displays B4100592 CORONIS 5MP 05 08 2014 3 Display installation Note When both displays are not correctly aligned minor corrections are po
29. nce with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this device does cause harmful interference to radio or television reception which can be determined by turning the device off and on the user is encouraged to try to correct the interference by one or more of the following measures e Reorient or relocate the receiving antenna e Increase the separation between the device and receiver e Connect the device into an outlet on a circuit different from that to which the receiver is connected e Consult the dealer or an experienced radio TV technician for help Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment Canadian notice This ISM device complies with Canadian ICES 001 Cet appareil ISM est conforme a la norme NMB 001 du Canada 7 4 EMC notice General information No specific requirement on the use of external cables or other accessories except power supply With the installation of the device use only the delivered power supply or a spare part provided by the legal manufacturer Using another can result in a decrease of the immunity level of the device Electromagnetic emissions The Coronis 5MP is intended for use in the electromagnetic environment specified below The customer or the user of the Coronis
30. nd by a Equipotentiality 7 6 Legal disclaimer 44 Disclaimer notice Although every attempt has been made to achieve technical accuracy in this document we assume no responsibility for errors that may be found Our goal is to provide you with the most accurate and usable documentation possible if you discover errors please let us know Barco software products are the property of Barco They are distributed under copyright by Barco N V or Barco Inc for use only under the specific terms of a software license agreement between Barco N V or Barco Inc and the licensee No other use duplication or disclosure of a Barco software product in any form is authorized The specifications of Barco products are subject to change without notice Trademarks All trademarks and registered trademarks are property of their respective owners Copyright notice This document is copyrighted All rights are reserved Neither this document nor any part of it may be reproduced or copied in any form or by any means graphical electronic or mechanical including photocopying taping or information storage and retrieval systems without written permission of Barco 2014 Barco N V All rights reserved Patent information This product is covered under the following intellectual property rights US Patent RE43 707 US Patent 7 038 186 US Patent 7 166 829 US Patent 6 950 098 European Patent 1 274 066 B4100592 CORONIS 5MP 05 08
31. ng of waste electrical and electronic equipment To prevent possible harm to the environment or B4100592 CORONIS 5MP 05 08 2014 7 Important information human health from uncontrolled waste disposal please separate these items from other types of waste and recycle them responsibly to promote the sustainable reuse of material resources For more information about recycling of this product please contact your local city office or your municipal waste disposal service For details please visit the Barco website at http www barco com en AboutBarco weee Mercury Notice LAMP CONTAINS MERCURY DISPOSE ACCORDING TO STATEILOCAL LAW This Barco product consists of materials that may contain mercury which must be recycled or disposed of in accordance with local state or country laws e Within this system the lamps in the device contain mercury Turkey RoHS compliance x EE T rkiye Cumhuriyeti AEEE Y netmeli ine Uygundur Republic of Turkey In conformity with the WEEE Regulation H EAk RoHS Chinese Mainland RoHS AR HE PEAR E FHE AN m R EEA EAA PEAMROHS ATAA T Barco ih PARISH A SA RASMRHZMAS Ss HEA RHS HES a A E p EE AP ML MOV SHEAS mPa SORMReEk h According to the China Administration on Control of Pollution Caused by Electronic Information Products Also called RoHS of Chinese Mainland the table below lists the names and contents of toxic and or hazardous substances that Barco s product may contain
32. not remove cover No serviceable parts inside Refer servicing to qualified personnel Do not expose this apparatus to rain or moisture Type of protection electrical Display with external power supply Class equipment Degree of safety flammable anesthetic mixture Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide Non patient care equipment Equipment primarily for use in a health care facility that is intended for use where contact with a patient is unlikely no applied part The equipment may not be used with life support equipment The user should not touch the equipment nor its signal input ports SIP signal output ports SOP and the patient at the same time Power connection Equipment with external 24 VDC power supply e Power requirements The equipment must be powered using the delivered medical approved 24 VDC SELV power supply The medical approved DC power supply must be powered by the AC mains voltage The power supply is specified as a part of the ME equipment or combination is specified as a ME system To avoid the risk of electric shock this equipment must only be connected to a supply mains with protective earth The equipment should be installed near an easily accessible outlet The equipment is intended for continuous operation Transient over voltage If the device is not used for a long time disconnect
33. o Hibernate page 24 for more information on hibernation and how to enable this function To enable disable DPMS mode on your display 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Power Management menu 3 Enter the DPMS Mode submenu 4 Select Enabled Disabled as desired and confirm 5 11 Hibernate 24 About hibernate Enabling hibernation will not only switch off the backlight but will also force the display to disable other functionalities so that power consumption is further reduced to a minimum This happens after a specific period of time which can be manually adjusted Hibernate can only be enabled on your display when the DPMS mode is enabled first Therefore please refer to DPMS mode page 24 to do this SSS B44100592 CORONIS 5MP 05 08 2014 5 Advanced operation To enable disable hibernation on your display 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Power Management menu 3 Enter the Hibernate submenu 4 Select Enabled Disabled as desired and confirm To specify the hibernate timeout 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Power Management menu 3 Enter the Hibernate Timeout submenu 4 Set the timeout value as desired and confirm 5 12 Luminance target About the luminance target The luminance target of your Coronis 5MP is adjustable over a predefined range When you change the luminance target the display w
34. of the display controller to the DVI or DP connector of the display 2 If you want to make use of the display s USB downstream connectors connect a PC USB downstream connector to the display s USB upstream connector 3 If you have chosen to use the display s USB downstream connectors connect a USB device to one of the downstream connectors 4 Connect the supplied external DC power supply to the 24 VDC power input of the display B4100592 CORONIS 5MP 05 08 2014 3 Display installation 5 Plug the mains connector of the external DC power supply to a grounded power outlet by means of the supplied power cable 3 5 Routing the cables To route the cables 1 Route the cables alongside the neck and in the cable ducts in the tilt amp swivel foot 2 Re attach the connector compartment cover and neck cover Close the cable duct covers Push the cable duct covers upwards to lock them To route the cables on a dual foot If the display is mounted on the system dual tilt amp swivel foot cables are best routed behind the arms and through the cable channel in the foot shaft as shown in the pictures below B4100592 CORONIS SMP 05 08 2014 15 3 Display installation 3 6 First time starting up 16 Overview You are now ready to start up your Coronis 5MP for the first time 1 Switch on your Coronis 5MP as described in Standby switching page 18 2 Turn on the computer connected to your d
35. ollution Control Labeling Standard of Chinese Mainland marked with the Environmental Friendly Use Period EFUP logo The number inside the EFUP logo that Barco uses please refer to the photo is based on the Standard of Electronic Information Products Environmental Friendly Use Period of Chinese Mainland O 7 3 Regulatory compliance information 38 Indications for use The Coronis 5MP MDCG 5121 is intended to be used in displaying and viewing digital images including digital mammography for review and analysis by trained medical practitioners Caution USA Federal law restricts this device to sale by or on the order of a physician Details amp exemptions are in the Code of Federal Regulations Title 21 801 Part D B4100592 CORONIS 5MP 05 08 2014 7 Important information FCC class B This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference re ceived including interference that may cause undesired operation This device has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This device generates uses and can radiate radio frequency energy and if not installed and used in accorda
36. r levels with luminance increments that are not optimal for crucial diagnostic information Studies have shown however that in medical images certain grayscale color parts contain more diagnostic information then others To respond to these conclusions display functions have been defined These functions emphasize on these parts containing crucial diag nostic information by correcting the native panel behavior ee S you select Native the native panel behavior will not be corrected Dynamic Gamma 1 8 These are gamma functions that are shifted to take into account the non zero luminance of an LCD panel when driven with a black signal They are especially useful in CT applications to improve the perception of low Hounsfield values DICOM Digital Imaging and Communications in Medicine is an international standard that was developed to improve the quality and communication of digital images in radiology In short the DICOM display function results in more visible grayscales in the images Barco recommends selecting the DICOM display function for most medical viewing applications User This display function will be automatically selected when display functions are defined by MediCal QAWeb Gamma 1 8 Select one of these display functions in case the display is to replace a CRT display with a gamma PAM of 1 8 or 2 2 respectively To select a display function Dynamic Gamma 2 2 1 Bring up the OSD main menu 2 Navigat
37. rface gt Indicator Lights menu 3 Enter the Power Status submenu 4 Select Enabled Disabled as desired and confirm 5 5 Key indicator lights About the key indicator lights By default after lighting up the key indicator lights will dim again if no further actions are taken within the following 5 seconds However this behavior can be changed so that the key indicator lights are always on or always off To configure the key indicator lights 1 Bring up the OSD main menu 2 Navigate to the Configuration gt User Interface gt Indicator Lights menu 3 Enter the Keys submenu 4 Select Automatic Always On Always Off as desired and confirm 5 6 Power lock function 22 About the power lock function By enabling the power lock function the Coronis 5MP is forced to remain switched on This means that it can t be switched to stand by mode until the power lock function is disabled again To enable disable the power lock function 1 Bring up the OSD main menu 2 Navigate to the Configuration gt User Interface gt Controls menu 3 Enter the Power Lock submenu 4 Select Enabled Disabled as desired and confirm B4100592 CORONIS 5MP 05 08 2014 5 Advanced operation 5 7 Keys sound About keys sound By default the keys will make a beep sound on every touch However this sound can be disabled so that no sound will be heard when using the keys To enable disable the keys sound 1 Bring up the OS
38. screen saver in the Display properties window of your workstation Barco recommends setting screen saver activation after 5 minutes of non usage A good screen saver displays moving content In case you are working with the same image or an application with static image elements for several hours continuously so that the screen saver is not activated change the image content regularly to avoid image retention of the static elements Understand pixel technology LCD displays use technology based on pixels As a normal tolerance in the manufacturing of the LCD a limited number of these pixels may remain either dark or permanently lit without affecting the diagnostic performance of the product To ensure optimal product quality Barco applies strict selection criteria for its LCD panels To learn more about LCD technology and missing pixels consult the dedicated white papers available at www barco com healthcare Enhance user comfort Every Barco multi head display system is color matched with the highest specifications in the market tant to use all displays of a multi head configuration at the same rate to preserve color Barco recommends keeping color matched displays together Furthermore it is impor E matching throughout the economic lifetime of the system B4100592 CORONIS 5MP 05 08 2014 _ S 1 4 Daily operation Maximize quality assurance The MediCal QAWeb system offers online service for high grade Quality
39. sponge cleaning cloth or soft tissue lightly moistened with a recognized clean ing product for medical equipment Read and follow all label instructions on the cleaning product In case of doubt about a certain cleaning product use plain water Do not use following products e Alcohol solvents at higher concentration gt 5 e Strong alkalis lye strong solvents Acid e Detergents with fluoride e Detergents with ammonia e Detergents with abrasives e Steel wool Sponge with abrasives Steel blades e Cloth with steel thread CAUTION Take care not to damage or scratch the front glass or LCD Be careful with rings or other jewelry and do not apply excessive pressure on the front glass or LCD CAUTION Do not apply or spray liquid directly to the display as excess liquid may cause damage to internal electronics Instead apply the liquid to a cleaning cloth B4100592 CORONIS 5MP 05 08 2014 33 6 Cleaning your display 34 B4100592 CORONIS 5MP 05 08 2014 7 Important information 7 IMPORTANT INFORMATION 7 1 Safety information General recommendations Read the safety and operating instructions before operating the device Retain safety and operating instructions for future reference Adhere to all warnings on the device and in the operating instructions manual Follow all instructions for operation and use Electrical Shock or Fire Hazard To prevent electric shock or fire hazard do
40. ssible in relative height and relative angle by loosening and tightening the opposite pairs of allen screws opposite sides or opposite diagonals of the positioning sliders 3 2 VESA mount installation The display mounting anchor is compatible with the VESA 100 mm standard WARNING Never move a display attached to an arm by pulling or pushing the display itself Instead make sure that the arm is equipped with a VESA approved handle and use this to move the display Please refer to the instruction manual of the arm for more information and instructions WARNING Use an arm that is approved by VESA according to the VESA 100 mm stan dard Use an arm that can support the weight of the display Refer to the technical specifica tions of this display for the applicable weight Overview The display mounting anchor is compatible with the VESA 100 mm standard 1 Unscrew the four fixation screws in the mounting anchor B4100592 CORONIS SMP 05 08 2014 1 3 Display installation 2 Attach the display firmly to the VESA mounting bracket with 4 screws M4 x 10 mm You should mount the display in portrait position Landscape position is not supported 3 3 Removing the covers 12 Removal of the connector compartment cover The connector compartment cover should be removed to get access to the connectors To remove the connector compartment cover proceed as follows 1 Gently push the two lips on top of the cover 2 Pull
41. t on a circuit different from that to which the receiver is connected e Consult the dealer or an experienced technician for help Electromagnetic immunity The Coronis 5MP is intended for use in the electromagnetic environment specified below The customer or the user of the Coronis 5MP should assure that it is used in such an environment Immunity test IEC 60601 Compliance level Electromagnetic environment guidance Electrostatic discharge 6kV contact 6kV contact Floors should be wood ESD 8kV air 8kV air concrete or ceramic tile IEC 61000 4 2 g in If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast 2kV for power supply 2kV for power supply Mains power quality transient burst lines lines should be that of a typical IEC 61000 4 4 1kV for input output 1kV for input output CMMercial or hospital lines lines environment Surge 1 KV line s to line s 1 kV line s to line s Mains power quality IEC61000 4 5 2 kV line s to earth 2 kV line s to earth SPOVUId be that of a typical commercial or hospital environment Test levels Voltage dips short lt 5 Ur 1 gt 95 dip in lt 5 Uz gt 95 dip in Mains power quality interruptions and voltage Ur for 0 5 cycle Ur for 0 5 cycle should by that of a variations on power supply input lines IEC 61000 4 11 typical commercial or hospital environment If the user of the Coronis 70 Ur
42. ther languages avail able for the OSD menu of your Coronis 5MP To change the language of the OSD menu 1 Bring up the OSD main menu 2 Navigate to the Configuration gt User Interface menu 3 Enter the Language submenu 4 Select one of the available languages and confirm 5 3 OSD menu automatic close function About the OSD menu automatic close function By default the OSD menu will disappear automatically after approximately 90 seconds of inactivity How ever this function can be disabled so that the OSD menu remains on the screen until manually closed To enable disable the OSD menu automatic close function 1 Bring up the OSD main menu 2 Navigate to the Configuration gt User Interface gt Menu menu 3 Enter the Automatic Close submenu 4 Select Enabled Disabled as desired and confirm B4100592 CORONIS 5MP 05 08 2014 _ 21 5 Advanced operation 5 4 Power status indicator light About the power status indicator light By default when the display is switched on the power status indicator light is dimmed This behavior can be changed so that the power status indicator light will be blue when the display is switched on When the display is in stand by mode the power status indicator light will always turn orange even when the power status indicator light is disabled To enable disable the power status indicator light 1 Bring up the OSD main menu 2 Navigate to the Configuration gt User Inte
43. when your display is physically rotated the image will rotate along The OSD menu however allows to overrule this behavior and force the image orientation to portrait or landscape regardless of the physical orientation of the display This may be especially useful when oper ating your display with the screen facing upwards To set the Display orientation 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Image Sources menu 3 Enter the Display Orientation submenu 4 Select Landscape Portrait Automatic as desired and confirm 5 20 Video input signals About input signals The available input signals for your display are DisplayPort 1 The input corresponding to the DisplayPort connector DVI 1 The input corresponding to the DVI connector Automatic Selection The input is automatically selected To manually select a video input signal 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Image Source menu 3 Enter the Input Signal submenu 4 Select one of the available input signals and confirm 5 21 Video encoding modes About video encoding modes Video encoding modes specify how the color and luminance information is encoded in the video signal In consumer displays this is usually done by a limited 8 bit encoding mechanism Your Coronis 5MP however features 10 bit encoding enabling it to better match the DICOM defined grayscale range Such an extensive range is necessary
44. y how color generated on the display controller is converted to grayscale in your display The available grayscale conversion modes are No Conversion Use Red Channel This mode is intended for grayscale displays where gray is sent over the red channel Use Green Channel This mode is intended for grayscale displays where gray is sent over the green channel gray is sent over the blue channel Use All Channels This mode is intended for grayscale displays where gray is sent over the red green and blue channel This is done by means of a standard conversion mechanism where 30 red 59 green and 11 blue are used to generate gray To manually select a grayscale conversion mode 1 Bring up the OSD main menu 2 Navigate to the Configuration gt Image Source menu 3 Enter the Grayscale Conversion submenu 4 Select one of the available color conversion modes and confirm B4100592 CORONIS 5MP 05 08 2014 5 Advanced operation 5 23 EDID timings About EDID timings Following EDID timings are available for your Coronis SMP Refresh Rate Allows to manually select the refresh rate of the image source video input signal depending on the maximum refresh rate of the display controller connected to your display Preferred Orientation Allows to change the orientation of the image source video input signal to landscape portrait or to let the display automatically assign the correct orientation Color Depth Allows to chang
Download Pdf Manuals