1. Codonics Infinity 1 4 1    Release Notes    Operating Software    Version 1 4 1 Release Notes    Summary    Version 1 4 1 software provides new features and corrects  defects for the Codonics Infinity   C and LX Medical Image  Servers     This software is compatible with all Infinity C and LX  systems     NOTE  This version of software cannot be downgraded  to a previous released version after being upgraded to  version 1 4 1     The Infinity Medical Image Server is an expandable DICOM  storage appliance that provides a solution for storing and  retrieving medical images  Applications include temporary  storage of large DICOM data sets directly from any  modality  and storage of digital images in an environment  without a centralized  digital image archive     This document details the changes from version 1 4 0  software  Infinity User   s Manuals provide further  information on the use of the Infinity systems  If you are  upgrading from an older version  previous Infinity Release  Notes can be downloaded from the Codonics web site     New Features      French Study Manager  The Study Manager is now  available in French along with the previously existing  support for English and Japanese  The French Study  Manager will allow French speaking users to connect to  the Infinity Server   s database and search for objects  related to patients  studies  series  images  and reports   Once a search is performed  a user can select and view an  object  edit its database information  ex
2. ages from one patient could be retrieved with the  other patient   s images  This situation could also occur if  the Administration tool is incorrectly configured to use a  Study Matching type that is not Study UID  such as  Accession Number or Study ID  and the chosen  identifier is not unique within the facility     A hard power loss could result in data corruption  To  prevent this condition  connect Infinity to a Codonics   recommended Uninterruptible Power Supply  UPS  and  minimize the number of times Infinity is hard powered  down     When rebuilding the RAID on an Infinity LX Server  only  the amount of time will take longer than previous  releases  In the unlikely event that a RAID is required to  be re built  the amount of time required to complete the  operation has increased from an average of eight hours  to ten hours     When exporting studies from Infinity  Storage Commit  is not supported  If the Storage Commit box is selected in  the Export Tab of the Administration tool  the Infinity  will retry the storage commit request  but ultimately will  not succeed     Infinity does not accept certain transfer syntaxes  A  transfer syntax of 1 2 840 10008 1 2 2   Explicit VR Big  Endian cannot be stored to Infinity     Infinity beeping and muting capability       If there is an issue with the RAID card  the system  will beep  The beep can only be muted after Infinity  software has started       Ifthe system was improperly shut down and the  RAID card is in a failure o
3. ahoo    Mail also  treats Infinity e mail as spam and puts it in the spam  folder  A user has to go to the spam folder and click the  e mail to indicate to Yahoo  that it is not spam     S     O CODONICS    Automatic system shutdown will not occur if DICOM  settings are not saved and the Infinity Administration  tool remains open  A user should save their settings and  close the Administration tool before shutting down the  system     Refreshing the System Status user interface page  navigates a user back to the main page  When refreshing  the System Status user interface  e g   by pressing the F5  key or using a browser button   the main page always  displays  For example  if the Drives page is currently  displayed  refreshing the page does not refresh the  Drives page but instead displays the main page     Selecting Search without any criteria may return too  many studies in the Infinity Web Viewer  A message  will alert a user that there are too many studies  Use a  criteria in one of the searchable fields  e g   Smith in the  Patient Name Field  to reduce the number of studies  returned     Sorting on columns in the browser tab of the Infinity    Web Viewer may display a message that too many   studies exist if the current search criteria returns too   many studies     Using the Infinity Web Viewer browser s Back button   will cause the Login prompt to display  It is not   recommended to select the Back button when using the   Infinity Web Viewer       When using the Infi
4. arks when hovering over Windows   system  time  Manually setting the format to English and then  back to Japanese refreshes Infinity and displays the  appropriate kanji characters     1  Open the Start menu and click Control Panel       Click Regional and Language Options       Opening two instances of the Administration tool can  cause errors  Opening two Administration tools and then  closing the second one can cause the first tool to go into a  state in which making changes and saving them causes  errors       Itmay take up to two minutes after resetting minimum  temporary storage before DICOM services will start       When performing an upgrade from 1 2 0 software  if the  system name is not set or the system name is changed  when performing the upgrade  DICOM stations  configured in the Study Manager will not be preserved       Infinity cannot compress certain study types  Infinity  will not compress Enhanced CT or Enhanced MR studies  using lossy compression when the system is configured  to compress incoming studies       Upgrading from 1 1 0 or 1 2 0 software to 1 4 1 software  does not create JPEG images for studies already stored  on the Infinity system  Contact Codonics Technical  Support to generate JPEG images for web viewing of  studies already stored with 1 1 0 or 1 2 0 software     Technical Support    If problems occur during software installation  contact  Codonics Technical Support between the hours of 8 30AM  and 5 30PM EST  weekends and U S  holidays excl
5. lection window and perform a search using  desired filter s         AJ Selection  1 Study       Fie View Options               I8 Patient list Study list    0  Desktop    Patient Name  Doe  John    Q Search    DCM4CHEE Patient ID    Patient ID  Database          Patient Name       psi see ciate     0  Send to Desktop  Display    Retrieve       Export study    TE Patient    EH Properties             In order to initiate a retrieve  right click the desired study  from the search results and select Export study  NOTE   The Display and Retrieve options are not supported    When prompted  select Database as the Connection  Selection      Upgrading from Version 1 3 1 or earlier no longer  removes Administrative DICOM stations  The upgrade  process was corrected to preserve DICOM stations in  when upgrading from version 1 3 1 or earlier          Rebooting the Infinity between 3 30 AM and 4 30 AM   system clock  during a database backup no longer  causes Infinity to stop working     Patient ID field is no longer truncated if greater than 32  characters  The patient identification field length was  increased from 32 to 64 characters  which meets the  DICOM standard and prevents the possibility of  truncation     Known Common Issues    This section details common issues with Infinity 1 4 1  software that are likely to be experienced by most users          The locale setting in the Configurator utility on the  SmartDrive will permanently set the database for a  specific locale and canno
6. nity Web Viewer  pressing the  Enter key rather than clicking on the Search button  when searching may cause the user to be logged out     If a user makes changes to their user preference   settings in the Infinity Web Viewer and then exits the    browser without logging out  the changes are not saved   for future sessions  Save preference settings before   logging out     When storing studies in the Japanese locale  Latin 1   e g   English  names that include special characters   e g   umlauts  will not display the special characters     The Help About screen on the Infinity Web Viewer  incorrectly states that the version is Beta 1     Structured report instances do not appear in query   results  Upon querying a study from the Study Manager   structured reports  if present  are not displayed  If the   retrieve function is selected to move the study to the   Infinity database  the structured reports  if present  for    the study will be imported successfully and displayed for  selection from the study level     When permanent storage is full  to Query Retrieve  studies stored on Infinity  use the AE title temp set up  on the DICOM Store SCU  e g   workstation      Known Uncommon Issues    This section details uncommon issues with Infinity 1 4 1  software that are unlikely to be experienced by most users     WARNING  Images for one patient could be included  in a study for another patient  If a patient   s images have  the same Study UID as another patient   s images  then  im
7. port it to another  DICOM device  or delete it     Product Improvements      Study Manager importing of studies only requires  matching criteria of the study UID  When importing  studies from a USB HDD via the Study Manager  the  default matching criteria is now only study UID       Infinity Basic Web Viewer has expanded selections for  modality and body part  The Infinity Basic Web Viewer  was updated to include additional selections for  modality  AS  BI  CD  CF  CP  CS  DD  DF  DG  DM  DR   DS  DX  EC  ES  FA  FS  HC  KO  LP  LS  MA  MG  MS   NM  PR  PT  RG  ST  TG  VF  XA  and body parts   Abdomen  Ankle  Arm  Breast  Clavicle  Coccyx  Elbow   Extremity  Hand  Heart  Hip  Jaw  Knee  Neck  Pelvis   Shoulder  Skull  C Spine  L Spine  S Spine  T Spine        User interface text elements are fully available for  display  Minor modifications were made for the  correction of text elements that were truncated when  using Japanese and French translated modes     User interface elements are all fully translated  Minor  adjustments were made for the correction of text  elements that were not translated when using Japanese  and French translated modes       Non DICOM objects are supported  Non DICOM  objects in notepad   txt  format are now supported for  association to studies in the Infinity database     Defects Corrected      The Infinity Viewer no longer displays incorrect slice  location information  As described in Infinity Product  Bulletin 301  part number 903 300 001   inco
8. r critical state  the system  will beep for two to three minutes during software  start up     Studies may be sent to the trash if the system is  rebooted during study transfer  It is possible that patient  images will be placed in the trash if the system is  rebooted during a study transfer  Do not reboot or shut  down Infinity until all images from a study are present in  the Study Manager     Q    O CODONICS    In the Infinity Configurator  attempting to cut  copy  or  paste text in the fields produces incorrect outputs  A  user should properly type in any letters or numbers and  not use shortcuts to copy text     If the SmartDrive is inserted or removed from the  Infinity  the tree on the left side of the System Status  user interface will become out of sync with the panes  on the right  A user should either refresh the browser  using CTRL F5  or close the System Status user interface  and reopen it to correct the issue     Deleting a study using the Infinity Study Manager  when it is still receiving studies causes orphan files  If a  user deletes a study while it is still being received  the  remaining incoming files will still be processed  but will  not belong to any study     Reboot time may take longer when Infinity is in a  critical state  e g   if the RAID array is rebuilding   The  time it takes for Infinity to reboot is longer when it is ina  critical state than when it is in an optimal state     When Infinity is set to Japanese locale  it may display  question m
9. rrect slice  location information may have been displayed by the  Infinity Study Manager and DICOM Viewer  In 1 4 0  software for Infinity C and 1 3 1 update software for  Infinity LX  the viewer did not display the slice location  information  With 1 4 1 software  the data is now  correctly displayed in the Information Text Overlay and  Localizer       Query Retrieve from Study Manager now works  correctly  To query retrieve studies using the Study  Manager  the Default AE title must be set to either temp  or perm in Administration   To set the default AE title   simply go to the File   gt  System   s info    drop down menu  in Administration     S    O CODONICS       a a T    Setup          System s Information       Server s name  Infinity G6X01    Database s name   0292    Database release  784    Default AE Title   perm  cree               Control     Editing the default AE title will modify Infinity   s calling  AE title to temp or perm when exporting studies through  the Study Manager  In previous software versions  the  calling AE title was DxDCMserver when exporting  studies through the Study Manager  The calling AE title  on exports will remain DxDCMserver in 1 4 1 software  unless the default AE title is modified in Administration   In order to initiate a query retrieve from the Study  Manager  the DICOM station must be added in the Study  Manager  refer to Infinity User   s Manual for details      In order to initiate a query  select the DICOM station  from the Se
10. t be changed  Once Infinity is  configured to a specific locale  Japanese or English   any  changes to the locale setting will cause studies stored  before the locale change to be unsearchable and display  incorrectly     Disconnecting the SAS cable or power cable to the  Infinity C Expansion Module during data transfer can  lead to loss of data  Refer to Infinity Medical Image  Expansion Module User   s Manual for further details     The LEDs on the Infinity C hard drives are red until the  Infinity software is running  The LEDs on the front of  each hard drive are red when the Infinity is turned on  and stay red until the Infinity software is running  Do not  pull a drive during this time     The Login dialog box is delayed for Remote Desktop  Connection  When a user first opens Remote Desktop  Connection for Infinity  there is at least a 30 second delay  before the Login dialog box opens  During this time  a  blank blue background is displayed  This condition only  occurs for the first time after a reboot  Subsequent logins  occur more quickly     Automatic system shutdown will not occur if studies  are opened for viewing in the Infinity Study Manager     Audit logs are not fully enabled as a default setting   Audit logging must be enabled from the Infinity  Administration tool     E mail status notifications might be handled  incorrectly by different e mail clients  Microsoft  Hotmail treats Infinity e mail as spam and requires that a  user click to display the message  Y
11. uded      2  3  Select English Format   4  Select Japanese Format   5      Click the Apply button     Get it all with just one call  1 800 444 1198    All registered and unregistered trademarks are the property of their respective owners     Specifications subject to change without notice  Patents pending     Copyright    2010 Codonics  Inc  Printed in the U S A  Part No  901 409 001 02     Phone   1 440 243 1198  Email  support codonics com  Website    www codonics com    R  CODONICS    We bring the future into focus    17991 Englewood Drive Codonics Limited KK  Middleburg Heights  OH 44130 USA New Shibaura Bldg  F1   440  243 1198 1 3 11  Shibaura    440  243 1334 Fax   Email info codonics com  www codonics com    Phone  81 3 5730 2297  Fax  81 3 5730 2295    Minato ku  Tokyo  105 0023 JAPAN    
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
iogear IOG-GFR204SD card reader  http://utomir.lib.u-toyama.ac.jp/dspace/ Title 看護フィジカル  Installation - Cedgreentecheast.com  Bürgerinitiative will gegen Funkturm klagen Geplante        CURASA CPAP EUT USER MANUAL  CMX32M Hardware Manual - RTD Embedded Technologies, Inc.  Einbau- und Bedienungsanleitung Rollotron® Pro standard    Copyright © All rights reserved. 