Stryker Sdc Hd Digital Image Capture System
Contents
1. loss of definition Refocus the coupler e aniy Clean and dry both the scope and the coupler windows Remove the coupler from the camera head and remove any moisture that has built up between the two components Optics are dirty Rotate the scope If dust particles in the picture rotate the dust is located on the scope itself Follow the manufacturer s instructions for cleaning the eyepiece and negative lens If particles in the picture do not move when you rotate the scope the particles are located behind the scope Remove the scope and clean the window on the front of the coupler with a dry or alcohol tipped cotton swab If dust particles lie between the coupler and camera remove the coupler and clean the coupler and camera windows Note Ensure all components are completely dry before reassembling them or fogging may result Blurry picture Ensure the coupler is in focus Increase the enhancement Ensure the specialty switch is not set to FLEXSCOPE unless you are using a flexible scope 31 Possible Solution White balance See the solution for Picture is too dark WB not good See the solution for Picture is too bright Perform the white balance procedure with the light source connected to the scope Use metal halide or xenon lighting no fluorescent lighting Note If this troubleshooting guide does not solve the problems that you are having with your camera call Stryker Technical Support at 1
2. 1088 Medical Video Camera Operating and Maintenance Manual stryker ENDOSCOPY Table of Contents Table of Contents Table of Contents Le 1 Warnings and Cautions 3 Symbol Definitions eee nn 5 Product Description and Intended Use 7 The Camera Console ee 7 Th Camera Head ur aea xc RO a wap De ca ae 10 The Coupler ick Rove NP CAU DURO RARO COR DECOR CR Ken 11 Setting Up the 1088 Camera 12 Setting Up the Console csse RII 12 Setting Up the Camera Head serene 17 Setting Up the Coupler lesen IRI 18 Operating the 1088 Camera 20 Using the Console Controls ccce 20 Powering On the Camera sehn 20 Using the White Balance Switch een 20 Using the Gain Button _ hh nn 21 Using the Enhance Up and Down Buttons 21 Selecting a Surgical Specialty s serrer 21 Using the Light Zoom Switch ee 21 Using the Shutter Switch nnn 22 Using the NTSC PAL Switch COO OCCUR 22 Controlling Remote Video Accessories _ 22 Using the Camera Head Buttons 0 0000 eee eens 23 Using the P BUftoDra aca deir a ra D SHE ce ed 23 Using the W Button RR 24 Using the Up and Down Buttons leen 24 Cleaning and Maintaining the 1088 Camera 25 Cl
3. 22 23 Product Description and Intended Use Power Switch Powers the camera on and off White Balance Button Starts the white balance operation Gain Button Selects four levels of light sensitivity settings Enhance Buttons Increase or decrease image sharpness Specialty Screen Displays which surgical preset has been selected Specialty Button Selects different surgical presets Camera Connector Connects to the 1088 Camera Head Sidne Port Connects to the Sidne Console to enable voice operation Ethernet Port Provides an ethernet connection for remote diagnoses and future software upgrades RGB Out S Video 1 Out S Video 2 Out Composite Out AC Power Inlet Equipotential Ground Plug Light Zoom Switch Selects either the LIGHT or ZOOM function for the arrow buttons on the camera head NTSC PAL Switch Selects NTSC or PAL video signals for analog outputs Shutter Switch Enables disables the auto shutter Remote Out 1 Connects to a video accessory remote switch Remote Out 2 Connects to a video accessory remote switch DVI Out 1 DVI Out 2 Fuse Panel Contains two 0 63A fuses The Camera Head The camera head connects to the camera console and captures video and photographic images which it relays to the camera console It features several controls that are accessible through a button control pad located on the top of the camera head see the Operating the Camera Head section of this manual Figure 2 b
4. Camera Head Buttons section of this manual Place the Light Zoom switch in the Zoom position to make the arrow buttons on the camera head control zoom Place the Light Zoom switch in the Light position to make the arrow buttons on the camera head control the light level 21 Using the Shutter Switch The SHUTTER switch on the rear console panel activates or deactivates the automatic shutter e Select ON to enable the automatic shutter The camera will automatically adjust picture brightness in response to varying light levels e Select OFF to disable the automatic shutter You will adjust picture brightness by using an automatic light source Note You may use an automatic light source and the camera auto shutter at the same time but the camera auto shutter should normally be used alone Using the NTSC PAL Switch The NTSC PAL switch selects which type of analog output the camera Will use Choose NTSC for NTSC analog outputs Composite S VHS RGB Choose PAL for PAL analog outputs Controlling Remote Video Accessories The 1088 Camera can remotely control two video accessories such as the SDC Pro a VCR or a photo printer enabling the user to capture images or start and stop video recording by pressing the buttons on the camera head See also the Using the Camera Head Buttons section of this manual 1 Connect the video accessory to one of the remote outputs on the rear console panel Use the provid
5. ose CA 95138 ut i LTD KADAN l N inii l ADUIIUUULDZ d HI 15 Wiring Diagram 3 Camera Light Source Flat Panel Monitor and CRT Monitor StrykerVision 1 Monitor DVI I VGA Adapter X7000 Light Source CRT Monitor Note If you are using any device with unterminated analog video inputs you must connect a cable from the VIDEO OUT of that device to the VIDEO IN on the monitor Note An additional monitor may be connected using any open camera output 3 Power on the monitor 4 Power on the camera Note A color bar pattern will appear on the monitor when the camera head is not connected to the camera console Follow the instructions in the Setting Up the Camera Head section of this manual to connect the camera head to the console 16 Setting Up the 1088 Camera Setting Up the Camera Head 1 Connect the camera head to the console Unscrew the soaking cap from the cable connector if necessary Align the red dot on the cable connector with the red dot on the camera connector port on the front console panel see Figure 4 Push in the connector until it locks in place Figure 4 Connecting the camer
6. 877 478 7953 inside the U S or refer to the Service and Claims section of this manual 32 Electromagnetic Compatibility Tables Electromagnetic Compatibility Like other electrical medical equipment the 1088 Camera requires special precautions to ensure electromagnetic compatibility with other electrical medical devices To ensure electromagnetic compatibility EMC the 1088 Camera must be installed and operated according to the EMC information provided in this manual Note The 1088 Camera has been designed and tested to comply with IEC 60601 1 2 2001 requirements for EMC with other devices Caution Portable and mobile RF communications equipment may il affect the normal function of the 1088 Camera Warning Do not use cables or accessories other than those A provided with the 1088 Camera as this may result in increased electromagnetic emissions or decreased immunity to such emissions Warning If the 1088 Camera is used adjacent to or stacked with A other equipment observe and verify normal operation of the 1088 Camera in the configuration in which it will be used prior to using it in a surgical procedure Consult the tables below for guidance in placing the 1088 Camera Guidance and Manufacturer s Declaration Electromagnetic Emissions 1088 Camera is intended for use in the electromagnetic environment specified below The customer or the user of 1088 Camera should ensure that it is used in such an environment Emiss
7. Department 5900 Optical Court oan Jose CA 95138 Stryker and Stryker Endoscopy are registered trademarks of Stryker Corporation 40 International Service International Service For service in the USA call your Stryker Endoscopy representative or Stryker Endoscopy Customer Service at 1 800 624 4422 Outside of the USA please contact your Stryker Endoscopy distributor at one of the following locations Stryker Corporation 2725 Fairfield Road Kalamazoo MI 49002 USA Phone 1 269 385 2600 Telex 224464 STRYKER KMZ Fax 1 269 385 1996 Stryker Canada 45 Innovation Drive Hamilton Ontario Canada L9H 7L8 Phone 905 690 5700 800 668 8323 toll free Fax 1 905 690 5698 Stryker Deutschland GmbH Gewerbeallee 18 D 45478 Mulheim an der Ruhr GERMANY Phone 49 208 999 060 Fax 49 208 999 0666 Stryker Latin America 15100 N W 67th Ave Suite 210 Miami Florida 33014 USA Phone 1 305 821 1888 Fax 1 305 826 0067 Stryker B V Marinus van Meelweg 17 P O Box 8747 5657 En Eindhoven THE NETHERLANDS Phone 31 40 2922522 Fax 31 40 2922555 Stryker Osteonics SA 5 Chemin des Aulx 5 1228 Plan les Ouates Case Postale 725 1212 Grand Lancy 1 Geneve SWITZERLAND Phone 41 22 884 0111 Fax 41 22 884 0199 Stryker European Rep RA QA Manager ZAC Satolas Green Pusignan Av De Satolas Green 69881 MEYZIEU Cedex France Phone 33 1 48175000 Fax 33 1 48632175 Stryker India Private Limited First Floor C 5 S
8. Road Newbury Berkshire RG14 5 EP United Kingdom Phone 44 1635 262400 Fax 44 1635 262464 42 Stryker Middle East Africa Via Della Posta 6934 Bioggio Switzerland Phone 4021 212 1122 Fax 4021 212 1133 NV Stryker SA Belgium Ikaros Business Park Fase III Ikaroslaan 12 1930 Zaventem Brussels BELGIUM Phone 32 2 717 92 10 Fax 32 2 717 92 49 Stryker Chile Avenida Nueva Tajamar 481 Oficina 805 Piso 8 Torre Norte santiago CHILE Phone 562 244 3600 Fax 562 244 3696 Stryker Spain Manuel Tovar 35 28034 Madrid SPAIN Phone 34 91 7283500 Fax 34 91 3580748 Stryker AB Scandinavia Krossverksgatan 3 9 216 10 Malmo SWEDEN Phone 46 40 69 18 100 Fax 46 40 69 18 190 Stryker AB Denmark Sankt Annae Plads 9 1021 Copenhagen Denmark Phone 45 33 9360 99 Fax 45 33 9320 69 MANUFACTURER Stryker Endoscopy Inc 5900 Optical Court San Jose CA 95138 USA Phone 408 754 2000 Fax 408 754 2505 stryker ENDOSCOPY 5900 Optical Court San Jose CA 95138 1 800 624 4422 1 800 729 2917 1000 400 675 Rev A
9. dip in Ut for 25 cycles lt 5 Ut gt 95 dip in Ut for 5 sec Compliance Level 2 4 6kV contact 2 4 8kV air 2kV line to ground 1kV line to line 0 5 1kV differential mode x 0 5 1 2kV common mode 596 Ut gt 95 dip in Ut for 0 5 cycle 4096 Ut 6096 dip in Ut for 5 cycles 7096 Ut 30 dip in Ut for 25 cycles lt 5 Ut 29595 dip in Ut for 5 sec NOTE Ut is the a c mains voltage prior to application of the test level 34 Electromagnetic Environment Guidance Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment If the user of 1088 Camera requires continued operation during power mains interruptions it is recommended that 1088 Camera be powered from an uninterruptible power supply or a battery Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Electromagnetic Compatibility Tables Guidance and Manufacturer s Declaration Electromagnetic Immunity 1088 Camera is intended for use in the electromagnetic environment specified b
10. from the date of purchase This warranty does not apply to any unit which has been subject to misuse neglect improper installation or that which has been altered adjusted or tampered with by any person other than Stryker Endoscopy authorized personnel If upon examination by authorized service personnel it is determined that the malfunction is due to misuse or abuse warranty provisions will not apply An estimate of the cost of repair work will be given to the customer prior to servicing and repairing the unit The customer is responsible for returning the defective equipment to the factory at his or her own expense Stryker Endoscopy or its representative will service the unit repair or replace any defective parts thereof and return the unit If upon examination it is determined that the fault has been caused by misuse or abnormal conditions of operation the repairs will be billed to the customer as out of warranty repairs Instruments repaired under Stryker Endoscopy s standard repair program will be issued a thirty day warranty against defects in both materials and workmanship provided the original warranty period has passed Instruments submitted due to defects in materials and workmanship during the warranty period will be repaired at no charge to the customer The warranty as set forth herein is exclusive and in lieu of all other warranties remedies obligations and liabilities of Stryker Endoscopy Inc expressed or imp
11. level has reached its maximum level pressing the W button again will cycle the level back to the lowest setting Note Whether the button selects light or zoom depends on which function has been selected with the LIGHT ZOOM switch on the rear console panel See the Using the LIGHT ZOOM Switch section of this manual for details Using the Up and Down Buttons The up and down buttons work together to increase or decrease the light level or zoom settings Whether the button selects light or zoom depends on which function has been selected with the LIGHT ZOOM switch on the rear console panel See the Using the LIGHT ZOOM Switch section of this manual for details e ZOOM When ZOOM is selected on the rear console panel pressing the arrow buttons will raise or lower the magnification setting between 100 percent and 200 percent in 64 steps Press the button once for individual steps or hold down the button for a quicker smoother transition LIGHT When LIGHT is selected on the rear console panel pressing the arrow buttons will raise or lower the automatic shutter light level setting in 64 steps Press the button once for individual steps or hold down the button for a quicker smoother transition 24 Cleaning and Maintaining the 1088 Camera Cleaning and Maintaining the 1088 Camera Cleaning the Camera Console Warning Disconnect the console from the AC power source N before cleaning Should the camera console need clean
12. operating the 1088 Camera ensure all components have been set up according to the instructions in the Setting Up the 1088 Camera section of this manual Warning Before using the 1088 Camera in a surgical procedure N test all components to ensure proper function Ensure that a video image appears on all video monitors before beginning any procedure Using the Console Controls Powering On the Camera Press the power switch to power the camera on or off Using the White Balance Switch The WHITE BALANCE switch is used to correct slight color differences that exist between different light sources or endoscopes Perform the white balance procedure before every surgical procedure Note Ensure a scope and light source are attached to the camera and that the camera light source and monitor are powered on before adjusting the white balance 1 Pointthe scope at several stacked 4x4 white gauze pads a white laparoscopic sponge or any clean white surface 2 Lookat the monitor and make sure that no glare is visible off of the white surface 3 Press and hold the WHITE BALANCE button until WB begins flashing on the video monitor 4 Continue pointing the scope at the white surface until the video monitor indicates that white balance is OK The video picture Will change color f the video monitor indicates that white balance is NOT GOOD repeat the process f you cannot achieve a white balance that is OK r
13. scope to completely dry before reassembly Note Be sure the threads on the coupler are completely dry before reattaching the coupler to the camera head Any moisture on the threads will cause the camera and coupler windows to fog during use Sterrad System 28 Caution Always install the soaking cap prior to sterilization Clean and prepare the camera head and cable as recommended in the Cleaning and Disinfection sections Ensure the soaking cap is installed 2 Sterilize the camera head and cable using the Sterrad oterilization System 3 Allow the camera head cable coupler and scope to completely dry before reassembly Note Be sure the threads on the coupler are completely dry before reattaching the coupler to the camera head Any moisture on the threads will cause the camera and coupler windows to fog during use Cleaning and Maintaining the 1088 Camera Using Sterile Drapes Using sterile drapes will ensure maximum longevity of your 1088 Camera Head For best results follow the instructions provided by the drape manufacturer Sterile camera drapes for use with the 1088 Camera Head can be ordered under Stryker product number 240 010 200 Maintaining the 1088 Camera Replacing the Fuses 1 Unplug the power cord from the wall outlet and remove the cord from the camera console 2 Unlatch the fuse holder above the AC inlet and remove it You may need to press the tab on the fuse holder with a slender screwdriv
14. DA Commercial Complex New Delhi 110 017 INDIA Phone 91 11 686 6740 Fax 91 11 696 6020 Stryker Australia No 50 Broughton Road Artarmon NSW 2064 AUSTRALIA Phone 61 2 9415 5100 Fax 61 29 4294127 Stryker Singapore PTE LTD 70 Bendemeer Road 4403 32 Hiap Huat House SINGAPORE 339940 Phone 65 293 0119 Fax 65 293 7028 Stryker Pacific Ltd Suite 2501 Citibank Tower Citibank Plaza 3 Garden Road Central HONG KONG Phone 61 2 9415 5100 Fax 61 29 4294127 Stryker Mexico S A de C V Calle Sacramento 410 Col Insurgentes San Borja C P 03100 Mexico D F MEXICO Phone 525 488 0890 Fax 525 488 0891 41 Stryker Finland PL 80 Makelankatuz FIN 00501 Helsinki FINLAND Phone 358 0 9 7744 680 Fax 358 0 9 7744 6820 Stryker Korea 11F Dong Sung Bldg 154 24 Samsung dong Kangnam ku Seoul KOREA 135 090 Phone 82 2 34517572 Fax 82 2 552 4156 Stryker China Limited Room 903 905 Office Tower 2 Beijing Sun Dong An Plaza 138 Wang Fu Jing Da Jie Beijing 100006 P R China Phone 86 10 65136183 Fax 86 10 83913571 Stryker Japan Dai Tokyo Kasai Shinjuku Bldg 3 25 3 Yoyogi Shibuya ku Tokyo 151 0053 Phone 813 535 29106 Fax 813 535 21789 Stryker Europe Headquarters Cite Centre Grand Rue 92 CH 1820 Montreux SWITZERLAND Phone 41 21 966 1201 Fax 41 21 966 1200 Stryker Taiwan 5F 1 23 Pa Te Road Section 1 Taipei TAIWAN R O C Phone 886 2 2322 2895 Fax 886 2 2357 8543 Stryker U K Ltd Hambridge
15. a head to the console Note To unplug the camera from the control unit grasp the knobbed portion of the connector and pull straight out This releases the connector latching mechanism Caution Do not severely bend the camera cable or damage may AN result 17 Setting Up the Coupler 1 Attach the coupler to the camera head Twist the camera end of the coupler into the camera head clockwise until it forms a tight seal see Figure 5 Figure 5 Attaching a coupler to the camera head Caution Do not overtighten the coupler as this may damage the front window of the camera Note For direct coupled scopes scopes that require no coupler thread the endoscope directly into the camera head until it forms a tight seal Caution Do not overtighten a direct coupled scope as this may N damage the front window of the camera 18 Setting Up the 1088 Camera 2 Attach an endoscope to the coupler e Remove the red dust cap if it is present Push down on the endobody clamp a and insert the scope into the scope end of the coupler b See Figure 6 Figure 6 Attaching an endoscope to the camera head Release the endobody clamp 3 Attach a light cable from the light source to the light post on the endoscope c see Figure 6 Warning Before each use check the outer surface of the N endoscope to ensure there are no rough surfaces sharp edges or protrusions 19 Operating the 1088 Camera Note Before
16. ance procedure See the Using the White Balance Switch section of this manual Check the color settings on the monitor Adjust the light level with the camera head Increase the light source output Check the fiberoptic light cable for excessive broken fibers Press the GAIN switch to increase illumination Adjust the light level with the camera head Decrease the light source output Ensure GAIN is OFF Plug the electrocautery generator into a separate electrical outlet and separate the 1088 power cord from the electrocautery power cord Separate the camera cable from the electrocautery cable Reposition the electrocautery grounding pad on the patient Ensure GAIN is OFF Reduce Enhancement Check for and replace faulty video cables Check to ensure that all devices in the video system are plugged in and powered on Check the connector on the camera head cable for broken pins Troubleshooting Possible Solution Image is not Release the scope from the coupler and then centered reconnect it Make sure the scope is seated correctly in the coupler Variability in Perform the white balance procedure See the color Using the White Balance Switch section of this reproduction manual between different Check the settings on video peripherals light sources or peripherals Ensure the light source has a proper infrared filter check with manufacturer specifications Foggy picture Refocus the camera
17. atings Denotes compliance to CAN CSA C22 2 No 601 1 M90 Yus and UL2601 1 Ame Type BF applied part Equipotentiality Protective earth ground Product Description and Intended Use Product Description and Intended Use The Stryker Endoscopy Model 1088 Medical Video Camera is a high definition camera used to capture still and video images of endoscopic surgical applications The 1088 Medical Video Camera consists of three main components the camera console the camera head e the focusing coupler The 1088 also comes with various cables which like the other components can be purchased together or separately The Camera Console The camera console is the control center for the 1088 Medical Video Camera and processes the video and photographic images captured during the surgical procedure The front console panel features controls for adjusting the enhance gain and white balance and for selecting surgical specialty settings that optimize camera performance for specific surgical procedures The rear console panel provides ports for connecting the 1088 Camera to viewing and recording equipment such as video monitors the SDC HD or photo printers The features of the front and rear console panels are listed in Figure 1 Figure 1a The 1088 Camera console front panel Figure 1b The 1088 Camera Console rear panel SH ue Seo dor x o 10 11 12 ire 14 15 16 17 18 19 20 21
18. e used with coupler C mount camera head used with C mount scopes C mount 1 32UN 2A Auto Shutter Range 1 60 to 1 50 000 second Gain OFF 0 dB LOW 5 dB MEDIUM 10 dB HIGH 15 dB 37 Operating Conditions Temperature 0 to 40 C Relative Humidity 10 to 75 Transport amp Storage Conditions Temperature 20 to 60 C Relative Humidity 1096 to 7596 Atmospheric Pressure 700 hPa to 1060 hPa Input Electrical Ratings 100 240V 50 60 Hz 0 6A Total Shipping Weight 11 Ibs 5 0 kg Dimensions Camera Control Unit 12 5 w x 3 8 h x 15 25 d 31 8cm w X 9 7cm h X 38 7cm d Camera Head Cable to Camera Control Unit 10 3 feet 3 15m sealed cable or 20 7 foot 6 30m sealed cable sensed by CCU Enhancement 16 levels switchable Classification amp Approvals Complies with medical safety standards IEC 60601 1 1988 A1 1991 A2 1995 IEC 60601 2 18 1996 A1 2000 CAN CSA C22 2 No 601 1 M90 UL2601 1 2nd Edition AS NZS 3200 1 0 1998 Complies with medical EMC standard IEC 60601 1 2 2001 Class Equipment Type BF Applied Part Water Ingress Protection IPXO Ordinary Equipment Continuous Operation 38 Warranty Stryker Endoscopy warrants the Model 1088 Medical Video Camera against defects in both materials and workmanship to the registered owner at the time of purchase All components including the charge coupled device CCD located in the camera head are covered by the warranty for a period of one year
19. eaning the Camera Console sss 25 Cleaning the CameraHead ss wean eee 25 Disinfecting the Camera Head ccce 26 Sterilizing the Camera Head senes 27 100 Ethylene Oxide Gas _ hmm 27 SIGs SySteM ania aid DD re RC aoc Ova o aO ied ae nee 28 Stertad System ew aC wwe EH TR 28 Using Sterile Drapes 0 0 eee ee ee ee RR hn 29 Maintaining the 1088 Camera 00 00 cee eee 29 Replacing the Fuses _ eee ee es 29 Periodic Maintenance Schedule 29 Disposing ofthe 1088 nn 29 Troubleshooting 30 Electromagnetic Compatibility Tables 33 Technical Specifications 37 Waan 2227s des i Sade ied r 39 Service and Claims 40 International Service 41 Warnings and Cautions Warnings and Cautions Please read this manual and follow its instructions carefully The words warning caution and note carry special meanings and should be carefully reviewed Warning The personal safety of the patient or physician may be involved Disregarding this information could result in injury to the patient or physician Caution Special service procedures or precautions must be followed to avoid damaging the instrument Note Special information to make maintenance easier or important information more clear An exclamation mark wi
20. ed remote cables See Diagram 2 in the Setting Up the Console section of this manual 2 Pressthe P button on the camera head for less than one second to activate Remote 1 One beep will sound 3 Press the P button on the camera head for more than one second to activate Remote 2 Two beeps will sound 22 Operating the 1088 Camera Using the Camera Head Buttons The camera head contains a diamond shaped four button pad that offers hand control of the 1088 Camera see Figure 7 The functions of each button are described in the following pages Figure 7 The camera head buttons Using the P Button The P button controls up to two remote video accessories Press the P button on the camera head for less than one second to activate Remote 1 One beep will sound Press the P button on the camera head for more than one second to activate Remote 2 Two beeps will sound 23 Using the W Button The W button activates the white balance function or the light zoom function Press the W button for more than one second to activate the white balance function The white balance function may also be initiated using the WHITE BALANCE switch on the front console panel See the Using the White Balance Switch section of this manual for details Press the W button for less than one second to increase the light or zoom Each press will raise the light or zoom level one out of four total increments When the light or zoom
21. efer to the Troubleshooting section of this manual 20 Operating the 1088 Camera Using the Gain Button The GAIN button increases the brightness of the video picture If the picture is too dark even when the light source is at full brightness and the camera light level is maximized you can brighten the picture by increasing the gain When you increase the gain the picture will get brighter but may also be more grainy There are four levels of gain OFF LOW MEDIUM and HIGH Press the GAIN button repeatedly to scroll through the four levels The current gain level will be displayed on the monitor OFF is the setting you should normally use Using the Enhance Up and Down Buttons The Enhance buttons increase or decrease the picture sharpness There are 16 levels of enhancement The current enhancement level will be displayed on the monitor Selecting a Surgical Specialty The Surgical Specialty switch selects one of eight pre established camera settings Each camera setting optimizes camera performance for a specific surgical application The eight settings are Arthroscopy Cystoscopy e Standard e Thoracoscopy Laparoscopy ENT Microscope Flex Scope The selected type will display on the specialty screen on the 1088 front console panel Using the Light Zoom Switch The Light Zoom switch on the rear console panel determines the function of the arrow buttons on the camera head See also the Using the
22. elow The customer or the user of 1088 Camera should ensure that it is used in such an environment Immunity IEC 60601 Test Compliance Electromagnetic Environment Test Level Level Guidance Portable and mobile RF communications equipment should be used no closer to any part of the 1088 Camera system including its cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended Separation Distance Conducted RF 3 Vrms d 147 P IEC 61000 4 6 150 kHz to 80 MHz Radiated RF 3 V m d 1 17 P 80 MHz to 800 MHz IEC 61000 4 3 80MHz to 2 5 GHz d 2 33 JP 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey 9 should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 35 Guidance and Manufacturer s Declaration Electromagnetic Immunity 1088 Camera is intended for use in the electromagnetic en
23. elow lists the features of the camera head Figure 2 The 1088 HD Camera Head 1 Soaking Cap Protects the cable connector during cleaning and sterilization 2 Camera Head Captures photographic and video images provides camera controls and connects to a focusing coupler 3 Camera Cable Cable Connector Connects the camera head to the camera console 10 Product Description and Intended Use The Coupler The coupler threads on to the face of the camera head enabling a scope to be attached to the camera It provides a focusing ring to adjust image sharpness The features of the coupler are listed in Figure 3 below Figure 3 The coupler Camera End Threads onto the camera head Focusing Ring Adjusts the coupler focus Ventilation Port Provides ventilation for the coupler lens Endobody Clamp Secures the scope to the coupler ae SN a Scope End Receives the endoscope 11 Setting Up the 1088 Camera Note Stryker Endoscopy considers instructional training or in service as an integral part of the 1088 Medical Video Camera Your local Stryker Endoscopy sales representative will perform at least one in service at your convenience to help set up your equipment and instruct you and your staff on its operation and maintenance To schedule an in service contact your local Stryker Endoscopy representative after your equipment has arrived Setting Up the 1088 Camera involves three steps 1 Setting up the con
24. epartment for service Never soak the camera and cable in the same tray with sharp instruments Caution Separate the camera head coupler and scope prior to N disinfection Refer to the coupler and scope product manuals for disinfection instructions Caution Always install the soaking cap prior to disinfection AN Caution To minimize galvanic corrosion avoid soaking N dissimilar metals in close proximity Caution Avoid leaving the camera in sterilization solutions N longer than necessary as this will accelerate normal product aging Camera life is extended by the consistent use of a single sterilization method 1 Clean and prepare the camera head and cable as recommended in the Cleaning section Ensure the soaking cap is installed 2 Immerse the camera head and cable in Cidex Plus glutaraldehyde solution or Cidex OPA solution for disinfecting Specifications for each are as follows Cidex Plus disinfection Soak time 10 minutes minimum as recommended by the sterilant manufacturer 24 hours maximum Temperature 25 C 77 F Cidex OPA disinfection Soak time 12 minutes minimum as recommended by the sterilant manufacturer 24 hours maximum Temperature 20 C 68 F 3 After soaking rinse the camera head and cable in sterile water agitating them to remove all traces of glutaraldehyde solution 4 Dry all parts thoroughly with a sterile towel 26 Cleaning and Maintaining the 1088 Camera Sterilizi
25. er to release the latch 3 Replace the fuse with the same value and rating Warning To avoid the risk of fire only use fuses of the value N specified on the fuse label located on the rear panel of the camera control unit 4 Reinstall the fuse holder until the tab snaps in place Periodic Maintenance Schedule Warning To ensure safe operation of the Model 1088 Medical Video Camera you should perform the following procedure periodically Every 12 months check the earth leakage current to 500pA 300pA in U S A ground impedance to 0 1 ohms and power consumption less than or equal to rated power Use a true RMS digital multimeter and safety analyzer to perform this test Note Refer calibration and operating difficulties not detailed in this manual to your Stryker Endoscopy sales representative Disposing of the 1088 The 1088 must be disposed of according to local laws and hospital practices The device does not contain any hazardous materials 29 Troubleshooting No color bar Incorrect picture color Picture is too dark Picture is too bright Noise or snow on picture when using electrocautery probes Noise or snow on picture when not using electrocautery probes No video picture when the camera head is plugged in 30 Possible Solution Ensure the video out from the camera is connected to the video in on the monitor Ensure all video systems are powered on Perform the white bal
26. face of the endoscope to ensure that there are no rough surfaces sharp edges or protrusions The warranty is void if any of these warnings is disregarded otryker Endoscopy accepts full responsibility for the effects on safety reliability and performance of the equipment only if Readjustments modifications and or repairs are carried out exclusively by Stryker Endoscopy e The electrical installation of the relevant operating room complies with the applicable IEC CEC and NEC requirements Stryker Endoscopy reserves the right to make improvements to the product s described herein Product s therefore may not agree in detail to the published design or specifications All specifications are subject to change without notice Please contact the local Stryker Endoscopy Distributor listed in the Other Service section or phone your local Stryker Endoscopy sales representative or agent for information on changes and new products Warning Federal law United States of America restricts this N device to use by or on order of a physician Warnings and Cautions Symbol Definitions In addition to the cautionary symbols already listed other symbols found on the 1088 Camera and in this manual have specific meanings that clarify the proper use and storage of the 1088 Camera The following list defines the symbols associated with this product 4 Operating humidity ratings o Operating pressure ratings Operating temperature r
27. ing wipe it down with a sterile cloth and mild cleaning solution Caution Never immerse or sterilize the camera console as this N will damage the camera and void the warranty Cleaning the Camera Head Warning The camera head and cable must be cleaned and N sterilized prior to the first use and after every subsequent use 1 Immediately after use unplug the camera head connector from the console and protect it with the soaking cap Place the cap over the connector and turn it clockwise until secure Caution Always install the soaking cap before rinsing the camera Failure to properly tighten the soaking cap will corrode the connector pins and void the warranty 2 Rinse the camera head and cable in lukewarm tap water Use a mild enzymatic detergent and a soft brush to remove resistant debris and bioburden from all surfaces Rinse off all soap residue with water 4 Clean glass surfaces with alcohol Dry the camera head and cable thoroughly with a soft towel or gauze surgical sponge 6 Prepare the camera head and cable for disinfection by coiling the camera cable into a loop about ten inches in diameter avoiding any kinks or twists in the cable 25 Disinfecting the Camera Head Caution Before soaking inspect the camera cable for cuts and N breaks Never soak a cut or a broken cable because moisture will seep into the camera head and cause damage Return any camera with a damaged cable to the Stryker Endoscopy Repair D
28. ions test Compliance Electromagnetic Environment guidance 1088 Camera uses HF energy only for its RF emissions CISPR internal function therefore its RF emissions are Group 1 I 11 very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B 1088 Camera is suitable for use in all Fama eri sa ons establishments including domestic Class A establishments and those directly connected to IEC61000 3 2 i the public low voltage power supply network that supplies buildings used for domestic Voltage Fluctuations I dir purposes flicker emissions Complies IEC61000 3 3 33 Guidance and Manufacturer s Declaration Electromagnetic Immunity 1088 Camera is intended for use in the electromagnetic environment specified below The customer or the user of 1088 Camera should ensure that it is used in such an environment Immunity Test Electrostatic Discharge ESD IEC61000 4 2 Electrical fast transient burst IEC61000 4 4 Surge IEC61000 4 5 Voltage dips short interruptions and voltage variations on power supply input lines IEC61000 4 11 Power frequency 50 60Hz magnetic field IEC 61000 4 8 IEC 60601 Test Level 6kV contact 8kV air 2kV for power supply lines 1kV for input output lines 1kV differential mode 2kV common mode lt 5 Ut gt 95 dip in Ut for 0 5 cycle 40 Ut 60 dip in Ut for 5 cycles 70 Ut 30
29. lied including the implied warranties of merchantability and fitness for use and of consequential damages These products are being sold only for the purpose described herein and such warranty only runs to the purchaser In no event shall Stryker Endoscopy be liable for any breach of warranty in any amount exceeding the purchase price of the product No agent employee or representative of Stryker Endoscopy has the authority to bind the Company to any other warranty affirmation or representation concerning this instrument This warranty is valid only to the original purchaser of Stryker Endoscopy products directly from Stryker Endoscopy or from a Stryker Endoscopy authorized agent The warranty cannot be transferred or assigned by the original purchaser 39 Service and Claims Caution Never open the camera head oor its cable or attempt any N service not described in this manual These units have been factory sealed to prevent moisture from entering the electronic components If the camera head or the cable seal is intentionally broken the equipment warranty will be void If service is needed either during or after the warranty period 1 Contact Stryker Endoscopy at 1 800 624 4422 or phone your local Stryker Endoscopy sales representative 2 Package all the components carefully in the original shipping container if possible 3 Ship the camera pre paid and insured to otryker Endoscopy Customer Service Attention Repair
30. ng the Camera Head Caution f the coupler and camera head are sterilized as a N single unit disconnecting the coupler will compromise the sterility of the two products 100 Ethylene Oxide Gas Caution Always install the soaking cap prior to sterilization 1 Clean and prepare the camera head and cable as recommended in the Cleaning and Disinfection sections Ensure the soaking cap is installed 2 Sterilize the camera head and cable with an ethylene oxide sterilization unit according to the following parameters Temperature 55 C 131 F Humidity 70 RH 5 Vacuum 53 34 cmHg 21 inHg Preconditioning time 60 minutes Sterilant mixture 10096 ethylene oxide Sterilant concentration 600 5 mg L ethylene oxide Exposure time 120 minutes maximum 180 min Aeration 12 hours 55 C 3 Allow the camera head cable coupler and scope to completely dry before reassembly Note Be sure the threads on the coupler are completely dry before reattaching the coupler to the camera head Any moisture on the threads will cause the camera and coupler windows to fog during use 27 Steris System Caution Always install the soaking cap prior to sterilization Clean and prepare the camera head and cable as recommended in the Cleaning and Disinfection sections Ensure the soaking cap Is installed 2 Sterilize the camera head and cable using Steris System 1 with Steris Sterilant 20 3 Allow the camera head cable coupler and
31. ration distance m according to frequency of transmitter Rated maximum output power W of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz d 1 17 P d 1 17 P d 2 33 P For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 36 Technical Specifications Technical Specifications Imaging System 1 3 Progressive Scan CCDs High Definition Scanning System Horizontal 47 536 kHz Vertical 60 02 Hz Video Outputs Digital Analog Two Digital Video Interface DVI RGBHV 1280 x 1024 SXGA resolution Connector 29 pin DVI I Composite One NTSC or PAL standard 1 0V P P 75 Ohms Connector BNC coaxial Y C Two S VHS Y 1 0V P P C 0 29V P P 75 ohms Connector 4 Pin mini DIN Component One RGB R 0 7V p p 75 ohms G 1 0 V p p w SYNC 75 ohms B 0 7 V p p 75 ohms Connector 8 pin DIN Resolution Horizontal 1100 lines Vertical 600 lines Mounting Endoscope Eyepiec
32. s e f desired connect any remote outputs using the remote cables supplied with the 1088 Camera See Diagram 2 Devices connected to the remote outputs of the 1088 Camera can be operated by using the buttons on the camera head See the Operating the Camera Head section of this manual for details f desired connect the Sidne interface as well See Diagram 2 Warning When the 1088 Camera is used with other equipment leakage currents may be additive Ensure that all systems are installed according to the requirements of IEC 60601 1 1 Warning Do not touch the internal pin of the VIDEO OUT BNC N jack and the patient simultaneously 13 Wiring Diagram 1 Camera Light Source and Flat Panel Monitor Flat Panel Monitor DVI I VGA Adapter 1088 Camera IAG ANALOG OUTPUTS O A MADE IN U S A DIGITAL OUTPUTS San Jose CA 95138 USA X7000 Light Source ENDOSUITE stryker idt TECHNOLOGY ENDOSCOPY ho er patana PUN c eRe i OT E MADE IN U S A Setting Up the 1088 Camera Wiring Diagram 2 Camera Light Source SDC Pro HD Sidne and Flat Panel Monitor StrykerVision 1 Monitor UDUDUU DEVCE PORTS oo ri LIE 9050000 eee eee e e r 9 1088 Camera X7000 Light Source stryker EN
33. sole 2 Setting up the camera head 3 Setting up the coupler Setting Up the Console Caution Portable and mobile RF communications equipment may A affect the normal function of the 1088 Camera When choosing a location for the 1088 Camera consult the Electromagnetic Compatibility section of this manual to ensure proper function To set up the console make the following connections 1 Connect the AC power Connect the AC power cord to the AC inlet on the rear console panel Connect the other end to a hospital grade outlet Warning Always use the hospital grade power cord supplied N with this camera 12 Setting Up the 1088 Camera 2 Connect the video output he rear panel provides four analog and two digital video outputs which can be used together or independently 8 pin DIN push only connectors Analog Composite Composite BNC push and turn connectors S VHS 1 S VHS 4 pin Mini Din push only connectors S VHS 2 S VHS 4 pin Mini Din push only On some monitors S connectors VHS inputs may be labeled Y C ETUDES pin push only connectors with two tightening knobs DVI I 2 29 pin push only The DVI connectors can connectors with two also output analog SXGA tightening knobs signals through a DVI I to VGA adapter Use the cables and outputs described above to connect the 1088 to other operating room equipment Diagrams 1 3 on the following pages describe typical set up
34. thin a triangle is intended to alert the user to the presence of important operating and maintenance instructions in the literature accompanying the product presence of hazardous voltage Refer all service to authorized personnel To avoid potential serious injury to the user and patient and or damage to this device the user must A A lightning bolt within a triangle is intended to warn of the Warning 1 Read this operating manual thoroughly and be familiar with its contents prior to using this equipment 2 Carefully unpack the unit and check if any damage occurred during shipment If damage is detected please refer to the Service and Claims section in this manual 3 Be a qualified physician having complete knowledge of the use of this equipment 4 Test this equipment prior to a surgical procedure This unit was fully tested at the factory before shipment 5 Avoid removing covers on the control unit as doing so may break the camera head seals or cause electric shock 6 Attempt no internal repairs or adjustments not specifically detailed in this operating manual T Pay close attention to the care and cleaning instructions in this manual A deviation may cause damage 8 Never sterilize the camera console 9 Disconnect the control unit from the electrical outlet when inspecting the fuses 10 Read the entire instruction manual before assembling or connecting the unit 11 Before each use check the outer sur
35. vironment specified below The customer or the user of 1088 Camera should ensure that it is used in such an environment a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the 1088 Camera system is used exceeds the applicable RF compliance level above the 1088 Camera system should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the 1088 Camera unit b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the 1088 Camera System The 1088 Camera system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The user of the 1088 Camera system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the 1088 Camera system as recommended below according to the maximum output power of the communications equipment Sepa