Home
Evolution Technologies OM-900M Oxygen Equipment User Manual
Contents
1. Factory preset at 2 0 0 5 Ipm bypass setting CLASSIFICATIONS AND SPECIFICATIONS SPECIFICATIONS Cont Oxygen delivery at level Liter Flow Equivalency 6 Continuous flow 2 LPM 0 5 LPM Maximum breathing rate 40 breaths per minute Power supply Blinking red light Low battery energy level replace batteries CLASSIFICATIONS AND SPECIFICATIONS SPECIFICATIONS Cont Classification according to EN 60601 1 Degree of protection against electric shock Type BF Degree of protection against harmful ingress of water IPX1 as per IEC 60529 Electromagnetic compatibility EMC according to EN 60601 1 2 Radio interference suppression EN 55011 Radio interference immunity EN 61000 4 2 to 6 8411 Vibrations Within IEC 60068 2 64 CLASSIFICATIONS AND SPECIFICATIONS SEPARATION DISTANCES Recommended separation distances between portable and mobile RF Communications equipment and the EVOLUTION MOTION The EVOLUTION MOTION is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled The customer or user of the EVOLUTION MOTION can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications transmitters and the EVOLUTION MOTION as recommended below According to the maximum output power of the communication equipment Separation distance according to transmission frequency in meters Max Ou
2. MOTION conserver is designed for a long and accurate life however as with any electronic device prudent care is required The unit should be kept clean and free from moisture and dust as well as extreme temperature Do not expose the unit to water such as when bathing or swimming It is advisable to keep the device in a carrying bag to afford a degree of protection HYGIENIC PREPARATION The unit and its accessories must be hygienically prepared at regular intervals Also carry out a functional check after the hygienic preparation see Functional Check on page 19 INTERVALS The unit and its accessories must be cleaned at the intervals listed below We also recommend carrying out disinfection at these intervals Please refer to the instructions supplied with the disinfectant used You are advised to use suitable gloves for disinfection work e g household or disposable gloves INTERVAL COMPONENT CLEANING DISINFECTION As required Case Fittings Wipe down Wipe disinfection with a lint free cloth As required Carrying Bag Hand wipe using warm Wipe disinfection water and mild soap CARE AND MAINTENANCE PROCEDURE Carry out hygienic preparation of the unit and accessories as described on page 16 We recommend using a 0 5 TERRALIN solution for wipe disinfection Follow the instructions enclosed with the disinfectant WARNING Take special care that no liquids enter the unit as this may cause damage You should under
3. VOLUTION OTION ane TABLE OF CONTENTS General InformallOl 3 06 o oum eto P ae Iob 28g ERa oe Sees seeders 3 Important Safety Rules and Precautions Operating The DIE 4 lt 5 cse e Ead o EASA EAE EEEE EERE 4 6 ACCESSON S ACDAIS cick as ca a aw wanes See hae us bed ERG 6 Introduction PUIPOSE ossi sao m ue e ee ae RAKES OS WR A Ee um a wae 7 User Qualificati eua xa re REM EE EEG EE ELE OE ath eases 7 FUNCTION Ln 8 Description of Parts and Controls sso bre Ree 9 10 Assembly and Use Non Portable Use 0 00 es 11 Installing Alkaline Batteries 0 00006 00 ee 12 Monitoring Battery Energy Level ics cssc idee ax x n 13 Installing ING System sss ku RR See XE RE a see ee eee een 13 Operating SUUCIONS 3 4 20 93 50020 te nonren ta MORIA ERU S S 14 15 Oxygen Cylinder Duration llle 16 TABLE OF CONTENTS Care and Maintenance Hygienic Preparation eseecastkER PRSE et beVaci ber VER RR x3 bs 17 ItitetVals aris ae are cC A E Ret e SR he ae eA dra 17 Procedure esses nn 18 Patie t Grange 2 cios 4454550426 braaa KISE P Bri Fb cex Peoria 18 Functional Check 2 ca cc ccna nee rr hh hm RR RR en ee 19 Disposal esee OR CER e EO RC ae C ede e 19 Products Spare Parts Accessories 000 cece eee eee 20 Troubleshooting Guide L9 ice Eu REX devi ibersbeetideaageds 21 22 Classifications and Specifications Technical Data sss es 23 25 Separation Distances nia als godere RES hg R
4. Your EVOLUTION MOTION conserver must not be immersed in liquid or cleaned with liquid agents Prevent water or other liquid substances from entering the unit Protect your EVOLUTION MOTION conserver from the cold and from continued exposure to water such as rain Please observe the section Hygienic Preparation on page 17 in order to avoid infection or bacterial contamination Prevent dust or any small particles from entering the unit Take care not to get entangled in the nasal cannula tube which could impede movement and lead to discomfort around the throat e Your EVOLUTION MOTION conserver is protected against interference To ensure safety of use however do not use a mobile phone in the immediate vicinity of the unit IMPORTANT SAFETY RULES amp PRECAUTIONS CAUTION Cont Oxygen conserving systems only work reliably upon sufficiently strong inhalation Therefore please observe the following Do not use the EVOLUTION MOTION at night or while sleeping Do not use the EVOLUTION MOTION for babies or children Do not use the EVOLUTION MOTION if you only breathe through your mouth Do not use the EVOLUTION MOTION if you breathe more than 40 times per minute Closely observe the permissible ambient conditions listed in the Technical Data section on pages 23 26 Failure to observe them may lead to a fire risk or damage to the unit Tighten all screwed unions by hand only Do not use a tool Do not
5. no circumstances use a cleaning agent Cleaning agents containing alcohol or grease pose a fire risk in combination with compressed oxygen Pay special attention to the oxygen inlet and outlet to make sure they remain free of dust etc If the oxygen inlet connection becomes contaminated with dirt oil or grease DO NOT USE OR ATTEMPT TO CLEAN Contact your supplier for service or repair CAUTION The carrying bag must never be washed in a washing machine spin dried or dried in a laundry drier W NOTE You are advised to use suitable gloves e g household or disposable gloves for disinfection work When cleaning your carrying bag be careful not to scrub the plastic window and do not roughen the seams Repeat cleaning if necessary Hang the bag in a well ventilated area and allow to air dry Do not hang in direct sunlight as this may cause its external fabric to fade PATIENT CHANGE Carry out a wipe disinfection on the unit s surfaces before you hand the unit over to a new patient CARE AND MAINTENANCE FUNCTIONAL CHECK CHECKING FOR LEAKS 1 Close the valve on the oxygen cylinder 2 Depressurize the EVOLUTION MOTION by inhaling several times using the nasal cannula The gauge indicator should drop to zero 3 Check that all screwed unions and tube connections are tight If necessary tighten them by hand W NOTE Do not use a tool 4 Ensure that the unit is set to the PULSE position 5 Slowly open the
6. use the EVOLUTION MOTION with a humidifier Do not use if leaking or damaged Always open the cylinder valve slowly Be sure to carry extra AA size alkaline batteries in the event they are needed WM NOTE Always ensure that your oxygen oylinder is sufficiently full We recommend always keeping a full spare cylinder in reserve Oxygen supplied by this equipment is supplemental only and is not intended for life support applications ACCESSORIES REPAIRS CAUTION Malfunctions and a lack of biocompatibility may result if third party articles are used Please bear in mind that in these cases any guarantee entitlement and liability shall lapse where accessories recommended in the instruction manual or original spare parts are not used e Servicing and repair work must only be carried out by the manufacturer CHAD Therapeutics or by trained personnel Refer repairs to authorized personnel Please contact your Home Care Provider if you have any questions INTRODUCTION PURPOSE The EVOLUTION MOTION electronic conserver is designed for use as part of a portable ambulatory oxygen system for the purpose of facilitating long term oxygen therapy LTOT It provides mobile LTOT patients with an extended operating time of their mobile system The EVOLUTION MOTION conserver includes a combination of a low pressure regulator and an oxygen conserver and is capable of delivering a precise amount of supplemental oxygen at the optimal p
7. valve on the oxygen cylinder until the needle in the gauge indicator no longer moves 6 Close the cylinder valve again 7 Observe the needle in the gauge indicator for approximately one minute If the needle remains in its position everything is OK However if the contents indicator shows a continuous decrease in pressure there is a leak in the system In this event contact your Home Care Provider DISPOSAL X The Unit Do not dispose of the unit in the household waste Consult an authorized electronic waste eCyCling company for the proper disposal of the unit Disposal of Batteries Do not dispose of used batteries in the household waste Contact a public waste ax disposal authority for proper disposal instructions PRODUCTS SPARE PARTS ACCESSORIES STANDARD PRODUCT ORDER NUMBER DESCRIPTION OM 900M EVOLUTION MOTION with CGA 870 Connection SPARE PARTS ORDER NUMBER DESCRIPTION RP 3040 Black S Cylinder Adjustment Handle FR 870G Seal Washers U S Only bag of 10 ACCESSORIES ORDER NUMBER DESCRIPTION OP 150 800 3 in 1 carry bag fits M4 M6 M7 and M9 cylinders OP 150T Horizontal carrying tote fits M4 M6 M7 and M9 cylinders WF NOTE Only manufacturer specified seal washers may be used with the EVOLUTION MOTION conserver These accessories are available from your Home Care Provider TROUBLESHOOTING GUIDE PROBLEM PROBABLE CAUSE SOLUTION Software needs to be reset O
8. which may lead to respiratory distress Disposable tubing should be disposed of in accordance with local ordinances and or regulations for disposal Replacements are available through your homecare provider OC 401S case of fifty 50 4 ft 1 52m cannulas or equivalent IMPORTANT SAFETY RULES amp PRECAUTIONS ZN WARNING Cont Please remember that this is for your own safety Keep all parts free of oil and grease Hydrocarbon compounds such as oil grease petroleum based products cleaning agents containing alcohol hand cream or adhesive bandages can cause explosive reactions if they come into contact with highly compressed oxygen Please wash and dry your hands properly prior to operating your oxygen equipment Never use aerosol sprays near the oxygen equipment Do not use in the presence of flammable anesthetic mixture Keep your oxygen equipment at least 5 ft 1 5 m away from any electrical appliance Be sure to turn off the oxygen supply by closing the cylinder valve when not in use Do not use cannula tubing that is longer than 7 ft 2 13 m Do not use a mask or pediatric or other low flow cannula tubing when operating the unit Do not use the EVOLUTION MOTION conserver as a handle for carrying your oxygen cylinder CAUTION To prevent the unit from overheating do not place it near any heating devices and do not expose it to direct sunlight Do not expose the unit to extreme temperatures
9. your first use or after you have made a change to your normal settings Place the nasal cannula into position with the prongs in the nostrils and begin breathing Fig G The EVOLUTION MOTION conserver will now start to deliver oxygen The amount of oxygen delivered is determined by the setting Adequate oxygen delivery will be Fig G achieved because of the precise time in the breathing cycle that the br dice 8 pulse of oxygen is delivered ASSEMBLY AND USE OPERATING INSTRUCTIONS Cont The Active setting should be used during exercise or activity and the Rest setting should be used when inactive or at rest The Rest and Active settings should be selected in accordance with a physician s prescription or with a physician s approval The motion sensor will select Rest or Active mode in accordance with the below table ACTIVITY RESULT Motion occurs for less than 2 seconds Remain in Rest mode Motion occurs for more than 2 seconds Switch to Active mode If motion occurs between 2 and 4 seconds Revert to Rest mode 10 seconds after motion ceases If motion occurs for more than 4 seconds Revert to Rest mode 50 seconds after motion ceases We NOTE To help prevent possible damage to the unit keep the EVOLUTION MOTION conserver in a carrying bag Several bags are available for use with different cylinder sizes and configurations STEP 6 When finished using the system turn off the oxygen supply cylinder valve and continue br
10. G INSTRUCTIONS STEP 1 STEP 2 STEP 3 STEP 4 Make sure the EVOLUTION MOTION conserver is set to the PULSE position Battery saving technology has eliminated the need for an OFF position The unit will go to Sleep Mode after three minutes of non use The unit will turn on again when an inspiration is detected or the Selector Button is pressed To reduce the risk of rapid oxygen recompression and fire OPEN THE CYLINDER VALVE SLOWLY and completely so that the pressure gauge moves slowly as it indicates the cylinder pressure Listen for leaks If a leak is present close the cylinder valve check the seal washer and reinstall If the leak persists DO NOT USE THE EQUIPMENT Contact your supplier for repair To select the Rest and Active settings on the EVOLUTION MOTION OM 900M conserver 1 6 that are equivalent to your oxygen prescription press and hold the REST Selector Button on the Selector Knob until the appropriate LED display light is illuminated Next press and hold the Active Selector Button Q on the Selector Knob until the appropriate LED display light is illuminated WT NOTE To check your current settings press and release the appropriate Rest or Active Selector Button WM NOTE The Active setting can only be set equal to or greater than that of the Rest setting STEP 5 Because the EVOLUTION MOTION stores the settings from the last time it was used you only have to complete this step before
11. R I ACE Eu ide 26 OVMNOOS KEY uorursibesRTISEESPERURERRUSS eee os Foe e hie 27 Limited Warranty issus RERO E Rd REX RES ESSERE ede ented eee 28 Important Information to Record 20 4264 29 nbrbhRERRRX RS RR X ex REREERG 29 GENERAL INFORMATION This manual provides the information necessary to operate the EVOLUTION MOTION electronic oxygen conserver with built in regulator in accordance with a physician s prescription The EVOLUTION MOTION conserver can be used with any CGA 870 post valve cylinder see Fig A at home E EH or away from home to provide your specific oxygen ogg e requirements It requires two 2 1 5 volt AA alkaline Parallel batteries for operation Pin Index Statements in this manual preceded by the following words are of special significance LN WARNING Indicates there is a possibility of injury to you or others E FIGURE A CAUTION CGA 870 Indicates there is a possibility of damage to the device or to other property Post Valve Cylinder W NOTE Indicates points of particular interest or emphasis that allow for more efficient and convenient operation of the equipment The CHAD Therapeutics EVOLUTION MOTION Model OM 900M is intended for prescription use only to be used as part of a portable oxygen delivery system for patients that require supplemental oxygen up to 6 liters per minute in their home and for ambulatory use Federal law restricts this device to sale by or on the ord
12. e setting CF is designed for emergency use only The amount of oxygen delivered when using the EVOLUTION MOTION in continuous flow mode is preset at 2 LPM liters per minute When not in use the cylinder should be turned off by turning the cylinder valve toggle counter clockwise The CF mode continuous flow is a mechanical override designed as a safety feature in case battery power fails It is set at 2 LPM CF When in the CF mode the user can still change the LED display lights by pressing the Selector Button but it has no effect on oxygen output CAUTION In the event that it is necessary to operate the unit in the continuous flow mode DO NOT obstruct the flow of oxygen from the Oxygen Supply Outlet by placing your finger over the outlet or blocking the flow through the oxygen tubing in any way Doing so may render the unit inoperable and or damage the sensor in the unit WE NOTE Remember that in continuous flow mode the oxygen will be consumed at a much faster rate Return to another source before depleting the oxygen cylinder Battery Holder This compartment holds two 2 AA size alkaline batteries Oxygen Pressure Gauge This enables the user to monitor the contents of the compressed oxygen cylinder Oxygen Supply Outlet Use this fitting to attach a standard cannula Cylinder Alignment Pins When assembling the unit these parallel pins must go into the holes on the post valve Seal Washer This creates the interface between the
13. eathing through the nasal cannula until no further oxygen is detected STEP 7 Remove the nasal cannula STEP 8 When not in use store in a clean dry location OXYGEN CYLINDER DURATION Because the total delivery of oxygen via the EVOLUTION MOTION conserver is related to breathing rates and activity level it is user adaptive in that the total oxygen delivered per minute will automatically adjust with user need These cylinder durations can vary widely depending on whether more time is spent in Rest or Active mode and depending on breath rate To use the duration chart below you can use an average of your Rest setting and your Active setting if you are using your device equally between the two setting options For example if your Rest setting is 2 and your Active setting is 4 then your tank duration on an M6 B cylinder at 20 breaths a minute would be approximately 4 6 hours If you spend more time in rest mode your duration would be higher and if you breathe faster than 20 breaths per minute while active the duration may be lower This chart is intended to be used as a guide only and is an estimation of use times Actual results may vary SETTING Continuous Flow 2 LPM CYLINDER TYPE CYLINDER VOLUME Estimated Cylinder Duration in Hours Based on 20 breaths per minute 36 liters 113 liters 164 liters 171 liters 198 liters 246 liters 425 liters 680 liters CARE AND MAINTENANCE The EVOLUTION
14. er of a physician IMPORTANT SAFETY RULES amp PRECAUTIONS Read this instruction manual carefully to ensure your complete understanding before operating your EVOLUTION MOTION electronic conserver This manual is part of the unit and must be available at all times Use the unit for the designated purpose only see Purpose on page 7 For your own safety and that of your patients please observe the following points OPERATING THE UNIT N WARNING Failure to observe the following warnings may result in damage to the unit or injury to life or limb Smoking near oxygen equipment is strictly prohibited While using your EVOLUTION MOTION conserver your clothes may come into contact with oxygen enriched air making your clothes more flammable This also applies for a time after use until the increased oxygen concentration has escaped from your clothing For this reason you must keep cigarettes matches burning tobacco and open flames such as lighted candles or fireplaces away from the area where the system is being stored or operated Avoid the creation of any spark such as static electricity caused by any type of friction near the oxygen equipment F NOTE Oxygen will not burn however it does vigorously accelerate the burning of any flammable material e Cannula tubing is a disposable accessory that should be replaced periodically following normal usage Continue reuse of cannula tubing may cause contamination
15. ith stationary oxygen cylinders it should only be used while awake and reasonably attentive The EVOLUTION MOTION conserver is not intended for use during sleep because in the unlikely event of operational malfunction or dislodging of the cannula the user could be unaware and not make the necessary corrections ASSEMBLY AND USE INSTALLING ALKALINE BATTERIES The EVOLUTION MOTION conserver uses advanced technology that prolongs battery life through efficient power use With normal use of four 4 hours per day at any setting your batteries should last a minimum of one 1 year Two 2 AA alkaline batteries are packaged in the box with your EVOLUTION MOTION conserver Follow these easy steps to install them STEP 1 While holding the unit in one hand gently press down on the battery door with your thumb and open the hinge door away from the EVOLUTION MOTION label Fig E1 and E2 STEP 2 Drop the batteries into the slot making sure they are inserted in the proper direction as indicated by the and symbols Figs E3 STEP 3 Rotate the door cover back in place until a click is heard Changing FIGURE E1 FIGURE E2 FIGURE E3 Alkaline Batteries NOTE The EVOLUTION MOTION conserver is packaged with batteries inside A special seal is used at the positive terminal to prevent battery oxidation If it has not already been removed by your Home Care Provider be sure to remove the seal before using the unit f
16. level the OM 900M stores two selectable settings one for rest and one for activity By sensing the user s activity the device can automatically switch from a lower setting in Rest mode to a higher setting in Active mode to meet the user s increased oxygen needs For these reasons the EVOLUTION MOTION OM 900M is designed to be an integral component of a lightweight long lasting ambulatory system DESCRIPTION OF PARTS amp CONTROLS aoe me Low Battery A FIGURE B Indicator Light EVOLUTION MOTION Conserver Rest L Selector Cylinder Top View Button ar Adjustment Handle Active Selector Button OXYGEN INLET CONNECTION Oxygen A Cylinder Pressure Seal Washer Alignment FIGURE C EVOLUTION MOTION Conserver Side View Battery Holder FIGURE D EVOLUTION MOTION Conserver Bottom View DESCRIPTION OF PARTS amp CONTROLS Low Battery Indicator Light The EVOLUTION MOTION conserver incorporates a visual indicator light that alerts you when a battery change is needed The following signal will be displayed as your batteries become depleted Blinking red light Low battery energy level Replace batteries Cylinder Adjustment Handle This is used to attach the unit to any CGA 870 post valve cylinder Selector Knob This enables the user to select their prescribed oxygen flow setting It also enables s he to select the continuous flow mode setting The continuous flow mod
17. o such action The foregoing warranty is exclusive and in lieu of all other express warranties Implied warranties if any including but not limited to the implied warranties of merchantability and fitness for a particular purpose shall not extend beyond the duration of the express warranty provided herein In no event shall CHAD Therapeutics be liable for loss of use or profit or other collateral special or consequential damages The expected service life of this device is 5 years IMPORTANT INFORMATION TO RECORD Your Name Date You Received Your Unit Prescribed Oxygen Flow Setting At Rest During Exercise Home Care Provider s Name Home Care Provider s Phone Number Physician s Name Physician s Phone Number Notes NOTES Parent Company Drive Medical Design amp Manufacturing 99 Seaview Boulevard Port Washington NY 11050 Toll free 877 224 0946 www drivemedical com LX 2 Li ISO 13485 Registered driv CHAD Manufactured by Inovo Inc 401 Leonard Blvd N Lehigh Acres FL 33971 Toll free 888 446 6862 www chadtherapeutics com EC REP MDSS Schiffgraben 41 30175 Hanover Germany PM 00042 13 E Printed in the U S A CHAD is a registered trademark of Inovo Inc EVOLUTION is a trademark of Inovo Inc
18. oint in the breathing cycle Operationally the EVOLUTION MOTION conserver greatly increases efficiency in the delivery of oxygen maximizing the beneficial effects and eliminating unnecessary oxygen waste The EVOLUTION MOTION is not suitable for use during sleep The EVOLUTION MOTION is not suitable for children Use the unit exclusively for the purpose described above USER QUALIFICATION Prior to beginning therapy patients must be given instruction by qualified personnel on how to use the unit INTRODUCTION FUNCTION The EVOLUTION MOTION OM 900M oxygen therapeutic device which combines a regulator motion sensor and an oxygen conserver is designed for use as part of an ambulatory oxygen system When we breathe approximately one third of the time is spent inhaling and two third exhaling As a result oxygen delivered by continuous flow is wasted during exhalation By eliminating oxygen flow during exhalation a two thirds savings is possible Additionally the application of the EVOLUTION MOTION is based on the fact that only oxygen inhaled at the beginning of the breath actually reaches the alveoli and is absorbed by the body The oxygen inhaled during the remainder of the breath is not used and is exhaled again However it is the addition of the patented motion sensing technology which makes the OM 900M a unique therapeutic device For those patients who have increased oxygen requirements based on their activity
19. or the first time ASSEMBLY AND USE MONITORING BATTERY ENERGY LEVEL The EVOLUTION MOTION conserver is equipped with a low battery indicator light that alerts you when a battery change is needed The following signal will be displayed as your batteries become depleted Blinking red light Low battery energy level Replace batteries ef Disposal of Batteries Do not dispose of used batteries in the household waste Contact a public ma Waste disposal authority for proper disposal instructions INSTALLING THE SYSTEM STEP 1 Loosen the cylinder adjustment handle STEP 2 Lower the EVOLUTION MOTION conserver over any CGA 870 post valve cylinder with the alignment pins toward the holes on the cylinder neck Fig F STEP 3 Align the two 2 pins and seal washer with the corresponding holes on the cylinder post valve STEP 4 While holding the unit in place tighten the cylinder adjustment handle by turning clockwise Fig F oe UR ONT W NOTE Tighten only by hand The use of a tool to tighten the handle MOTION CN to the cylinder may damage the unit MOTION CAUTION If you are unable to eliminate leaks by manually tightening the cylinder adjustment handle replace the seal washer If leaks persist the unit must be returned for service STEP 5 Attach a standard cannula to the oxygen supply outlet See page 14 for an illustration demonstrating the proper positioning of the nasal cannula ASSEMBLY AND USE OPERATIN
20. pen the battery door and remove the batteries Wait 10 seconds Replace the batteries and close the battery door pulse incorrectly reversed contacts and a cotton swab to clean contacts cylinder if empty Oxygen cannula is blocked Remove kinks Clean or replace if Short battery used alkaline life Replace batteries See Installing Alkaline Batteries section on page 12 Non functioning units are subject to warranty provisions and the manufacturer repair return policy If necessary call your Home Care Provider NOTE Do not attempt to open the electronic compartment of the unit If the case is opened or tampered with the warranty is void TROUBLESHOOTING GUIDE PROBLEM Increasing Rest Mode setting changes Active Mode setting Stays in Active Mode while riding in an automobile The Activity light does not illuminate during motion Sometimes the Activity light or Activity Mode stays on longer PROBABLE CAUSE The Active Mode setting must be equal to or higher than the Rest Mode setting The device may be interpreting the car s motion as activity The battery may be dead or depleted The activity time was not long enough to activate the activity sensor Switching times between Rest and Active Modes are variable SOLUTION Select the Rest Mode setting first then set the Active Mode according to your prescription The device has been designed to be very responsive to a u
21. ser s activity As a result the device may interpret a person sitting in a car bus train elevator or other forms of mobile transportation as activity If you are a passenger you may choose to adjust the Active Mode temporarily to equal your Rest Mode If you are the driver consult your physician to see if the Active flow setting is appropriate NEVER adjust your settings while driving Replace or recharge the batteries Continue activity and monitor situation The Activity light will illuminate after 3 4 seconds of activity The time in which the device switches from Active Mode back to Rest Mode varies depending upon the length of time the user remains active The longer the user remains active the longer the device takes to switch back to Rest Mode once the activity has stopped This is proper operation of the unit CLASSIFICATIONS AND SPECIFICATIONS TECHNICAL DATA SPECIFICATIONS EAS Product class according to 99 42 EEC b 6 1 L 15 5 cm x 2 5 H 6 4 cm x 3 1 W 7 9 cm Approximately 14 9 ounces 422 grams with batteries Temperature range Operation 14 F to 104 F 10 C to 40 C Storage 40 F to 158 F 40 C to 70 C Humidity range Operation 15 95 non condensing Storage Up to 95 non condensing Operating Altitude 1 000 to 10 000 feet 304 8 to 3 048 meters 700hPa 1060hPa Standard nasal cannula up to 7 ft 2 13 mj Built in 25 5 PSI 17 3 bar Continuous flow emergency
22. tput Power Watts 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 3 5 V YP D 3 5 V v P Dz 7N Y P CLASSIFICATIONS AND SPECIFICATIONS SYMBOLS KEY Warning consult accompanying documents Refer to the instruction manual memm EE ea O O O oe orvosin saara seare saper OOOO R oraaa nen voras OOO Serial number of the unit its bag or carrying it under your jacket SN IPX1 HON No smoking or open flames The EVOLUTION MOTION is protected against dripping A few drops of rain will not damage the unit but you should protect it from continued exposure to water by keeping it in LIMITED WARRANTY The EVOLUTION MOTION oxygen conserver has been carefully manufactured and inspected and is warranted to be free from defects in workmanship and materials Under this warranty CHAD Therapeutics obligation shall be limited to the replacement or repair of any such units or parts that prove by CHAD s inspection to be defective within two years from the date of purchase Any abuse operation other than the intended use of the product as outlined in this manual negligence accident or repair by other than authorized service professionals shall immediately void this warranty This warranty does not extend to the battery or cannula CHAD Therapeutics will not accept damages or charges for labor parts or expenses incurred in making field repairs except upon written authorization prior t
23. valve and the EVOLUTION MOTION conserver Besides offering a rugged interface it also surrounds the oxygen path in a ring of stainless steel or brass Oxygen Inlet Connection Interface between cylinder valve and conserver that allows oxygen to flow into the regulator LN WARNING Use only a manufacturer specified seal washer Other seal washers may not be oxygen compatible and may cause an oxygen leak creating an increased fire risk ASSEMBLY AND USE Make certain that your hands are free of oil grease and other contaminants Inspect the unit to insure that it has a manufacturer specified seal washer in good working condition attached to the inlet nozzle Secure the cylinder in an upright position Inspect the post valve of the cylinder and the EVOLUTION MOTION conserver to ensure they are free of contaminants If any indication of damage or contamination is detected DO NOT use the equipment and contact your Home Care Provider A WARNING Use ONLY a manufacturer specified seal washer Other seal washers may not be oxygen compatible and may cause an oxygen leak creating an increased fire risk Do not use the device if the manufacturer specified seal washer is missing Contact your Home Care Provider for assistance NON PORTABLE USE The EVOLUTION MOTION conserver is designed to extend the life of portable oxygen supplies when away from the primary source While the EVOLUTION MOTION conserver may be used w
Download Pdf Manuals
Related Search
Related Contents
田中光栄 八木のぶお 桜井和寿 Doctrine - Open Source PHP ORM Solar Power Controller (SDPC-SF) Canada - Buyandsell.gc.ca No.294-1104 Electric Self-Cleaning EvenAir™ Convection Plus Wall Oven FieldServer FS-8700-19 Computer Drive User Manual User Manual Betriebsanleitung Operating instructions Manometer 一一一ーと竺 Copyright © All rights reserved.
Failed to retrieve file