1. connection between electrical equipment and the potential equalization busbar of the electrical installation Please consult IEC 6060 1 2005 for requirements This system is subject to European WEEE Directive 2002 96 EC This product contains electrical and electronic components and must not be disposed of using standard refuse collection Please consult local directives for disposal of electrical and electronic equipment bt lt gt P lJ O Check the 3M Ranger system website to ensure you have the most recent version of this document www rangerfluidwarming com reorder 202456B 3M Ranger Irrigation Fluid Warming System English Protective earth ground Consult instructions for use Follow instructions for use 5 Recycle to avoid environmental contamination This product contains recyclable parts For information on recycling please contact your nearest 3M Service Center for advice v Date of manufacture E Manufacturer e m m LE EO Sterile ethylene oxide Storeatroom temperature 8 C 33 Store at room temperature DEHP free A Quantity Single use Latex free b Keep dry Temperature limits 20 C Explanation of Signal Word Consequences WARNING Indicates a hazardous situation which if not avoided could result in death or serious injury CAUTION Indicates a hazardous situation which if not avoided could result in minor or m
2. cross contamination The cleaning tool provides only superficial cleaning it does not disinfect or sterilize the interior of the unit NOTICE 1 To avoid device damage Do not immerse the warming unit in cleaning or sterilizing solutions The unit is not liquid proof Do not clean the warming unit with solvents Damage to the case label and internal components may result Do not insert metallic instruments in the warming unit Do not use abrasive materials or solutions to clean the heater plates Do not allow spills to dry inside the unit as this may make it more difficult to clean the unit TO CLEAN THE EXTERIOR OF THE WARMING UNIT 1 Disconnect the Ranger irrigation fluid warming unit from the power source 2 Wipe the outside of the unit with warm soapy water nonabrasive cleaning solutions dilute bleach or cold sterilants Do not use abrasive materials 3 Wipe with a dry soft cloth TO CLEAN THE HEATING PLATES The Ranger hardware cleaning tool is intended to clean both heating plates of the warming unit It is not necessary to disassemble the warming unit to use the tool 12 Check the 3M Ranger system website to ensure you have the most recent version of this document www rangerfluidwarming com reorder 202456B 3M Ranger Irrigation Fluid Warming System English METHOD 1 Unplug the irrigation fluid warming unit 2 Unfold the cleaning tool Wet the foam pads with a n
3. failure Temperature over temperature alert temperature or has gone will rise to 50 C within a few system cut off dark alert sounds seconds The safety backup system is activated at 50 C and power is immediately shut off to the heating plates Under temperature alert 33 C LED displays LO and Heater temperature has fallen temperature alert sounds to 33 C Alerts will stop alert light flashes when temperature rises above 33 C Continue use of unit NOTE The difference in over temperature alert system cut off points does not indicate any difference in safety and is simply a result of tighter controller tolerances The Ranger irrigation fluid warming system is designed with very tight controls to regulate its temperature to the 41 C set point The microprocessor controller measures the temperature at the fluid pathway inside the warming plate four times every second making it very responsive to changes in flow rate Check the 3M Ranger system website to ensure you have the most recent version of this document 3M Ranger Irrigation Fluid Warming System Instructions For Use www rangerfluidwarming com reorder 202456B Preparation and Setup of the Ranger Irrigation Fluid Warming Unit 1 Attach the Ranger irrigation fluid warming unit to the pole Tighten the pole clamp securely CAUTION To prevent tipping clamp the Ranger irrigation fluid warming unit to an LV pole with a minimum 14 35 6 c
4. 1 1 CAN CSA C22 2 No 601 1 CLASSIFICATION MEDICAL GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601 1 CAN CSA C22 2 No 601 1 ANSI AAMI ES60601 1 2005 CSA C22 2 No 60601 1 08 Control No 4HZ8 Classified under IEC 60601 1 Guidelines and other national versions of the Guidelines as Class I Type B Ordinary equipment Continuous operation Classified by Underwriters Laboratories Inc with respect to electric shock fire and mechanical hazards only in accordance with IEC EN 60601 1 and in accordance with Canadian CSA C22 2 No 601 1 Classified under the Medical Device Directive as a Class IIb device Electrical characteristics Temperature characteristics INPUT VOLTAGE SET POINT TEMPERATURE 100 120 or 220 240 VAC 41 C OPERATING FREQUENCY OVER TEMPERATURE ALERT 100 120 VAC 50 60 Hz 48 C 220 240 VAC 50 60 Hz MAXIMUM HEATING POWER UNDER TEMPERATURE ALERT 900 W 33 C FUSE OVER TEMPERATURE CUTOFF 2x T10A H 250V for 100 120 VAC 50 C 2 x T6 3A H 250V for 220 240 VAC FUSE TYPE Time delay high breaking Leakage Current Meets leakage current requirements in accordance with UL IEC 60601 1 Environmental Conditions OPERATING TEMPERATURE RANGE 15 to 40 C 59 to 104 F STORAGE TEMPERATURE RANGE 20 to 45 C 4 to 113 F OPERATING HUMIDITY 10 to 85 RH noncondensing ATMOSPHERIC PRESSURE RANGE 50 kPa to 106 kPa C 0086 3M is a tra
5. Check the 3M Ranger system website to ensure you have the most recent version of this document Irrigation fluid warming system Operator s manual Manuel de l utilisateur du syst me de r chauffement pour liquide d irrigation Sp lfl ssigkeitserw rmungssystem Benutzerhandbuch Manuale per l operatore del sistema di riscaldamento dei fluidi di irrigazione 5 Sistema de calentamiento de fluidos de irrigaci n Manual del operador amp D Gebruikershandleiding verwarmingssysteem voor irrigatievloeistoffen Anv ndarhandbok for spolv tskeuppv rmningssystem Varmesystem til udskylningsv ske Brugervejledning Oppvarmingssystem for irrigasjonsv ske brukerveiledning Huuhtelunesteen l mmitysj rjestelm K ytt opas Manual do utilizador do Sistema de www rangerfluidwarming com reorder 202456B 3M Ranger Irrigation Fluid Warming System Model 247 Operator s Manual Eyxetpidto yprjonc 0 ppavonc vypwv Karai vronc Instrukcja obstugi systemu do podgrzewania plyn w stosowanych przy irygacji 9 Irrig ci sfolyad k meleg t rendszer Kezel i k zikonyv 2 Ohrivaci syst m pro iriga n tekutiny N vod k obsluze GD Irigacinio skys io ildymo sistema Operatoriaus vadovas irigasyon s v s s tma sistemi Kullanim k
6. Indications for Use en IQ 4 Definition of Sei Gl 8s ccssisesiciaiccdiscddsusvonveonsnsnienstonsapearieanseuntanaisuacesdssvvieinspunsausdguaveuavnnine 4 Explanation of Signal Word Consequences 5 bro M 6 Lenin T T 6 pj 6 Product Description aula Sua DU MM RI MN M f DN E IUE 7 Ranger Irrigation Fluid Warming Unit e erri alten bye 7 Ranger Irrigation Fluid Warming Disposable Set aaa 7 Model 247 Product Safety Features aaa assssssasssssssasssassssssssssssssws 8 Instructions For Use 9 Preparation and Setup of the Ranger Irrigation Fluid Warming Unit 9 Removing the Irrigation Fluid Warming Set from the Ranger Irrigation Fluid Verdad 9 Troubleshooting 10 Maintenance and Storage e 12 14 Check the 3M Ranger system website to ensure you have the most recent v
7. aged Use only 3M specified replacement parts Keep power cord visible and accessible at all times The plug on the power cord serves as the disconnect device The wall socket outlet shall be as close as practical and shall be easily accessible 3 Toreduce the risks associated with air embolism Never use fluids if air bubbles are present in the fluid line CAUTION 1 Not for LV use 2 To reduce the risks associated with cross contamination The cleaning tool provides only superficial cleaning it does not disinfect or sterilize the interior of the unit 3 To reduce the risks associated with impact and facility medical device damage Clamp the Ranger irrigation fluid warming unit to an LV pole with a minimum 14 35 6 cm radius wheelbase and at a height no higher than 44 112 cm 4 To reduce risks associated with environmental contamination Follow applicable regulations when disposing of this device or any of its electronic components NOTICE 1 Federal law USA restricts this device to sale by or on the order of a licensed healthcare professional 2 To avoid device damage Do not clean the warming unit with solvents Damage to the case label and internal components may result Do not immerse the warming unit in cleaning or sterilizing solutions The unit is not liquid proof Do not insert metallic instruments in the warming unit Do not use abrasive materials or solutions to clean the heater
8. demark of 3M Company RANGER BAIR HUGGER and the BAIR HUGGER logo are trademarks of Arizant Healthcare Inc used under license in Canada 2013 Arizant Healthcare Inc a 3M company All rights reserved 3M Deutschland GmbH Health Care Business Carl Schurz Str 1 41453 Neuss Germany 3M Health Care 2510 Conway Ave St Paul MN 55144 USA TEL 800 228 3957 www rangerfluidwarming com 202456B 08 13
9. ersion of this document www rangerfluidwarming com reorder 202456B Check the 3M Ranger system website to ensure you have the most recent version of this document www rangerfluidwarming com reorder 202456B 3M Ranger Irrigation Fluid Warming System English Technical Service and Order Placement Technical Service USA TEL 1 952 947 1200 1 800 733 7775 FAX 1 952 947 1400 Outside of the USA Contact your local 3M Patient Warming representative In warranty Repair and Exchange To return a device to Patient Warming for service first obtain a Return Authorization RA number from a customer service representative Please use the RA number on all correspondence when returning a device for service A shipping carton will be delivered to you at no charge if needed We will service and ship your device within five 5 working days of our receipt Call your local supplier or sales representative to inquire about a loaner device while your device is being serviced When You Call for Technical Support Remember we will need to know the serial number of your unit when you call us The serial number label is located on the bottom of the warming unit 202456B English 3 Check the 3M Ranger system website to ensure you have the most recent version of this document www rangerfluidwarming com reorder 202456B English 3M Ranger Irrigation Fluid Warming System Introduction The 3M Ranger irrigation fl
10. g infused or clamp is open proximal to the warming cassette Warming unit is below patient level creating excessive back pressure Cause Under temperature condition caused by very high flow using very cold fluid or defective heater relay Cause Open wire on temperature sensor Cause Electrical interference www rangerfluidwarming com reorder 202456B English Solution Unplug unit Call 3M Patient Warming technical support Solution Make sure fluid is drained from warming cassette before sliding out the cassette that fluids are no longer infusing and that clamp is closed proximal to the warming cassette Raise unit above patient level Solution Alerts should stop when temperature rises above 33 C If alerts continue turn off unit unplug unit and discontinue use Call 3M Patient Warming technical support Solution Do not use unit Call 3M Patient Warming technical support Solution Remove the unit Refer to biomedical technician or call 3M Patient Warming technical support 11 Check the 3M Ranger system website to ensure you have the most recent version of this document www rangerfluidwarming com reorder 202456B English 3M Ranger Irrigation Fluid Warming System Maintenance and Storage Cleaning the Ranger Irrigation Fluid Warming Unit Clean the Ranger irrigation fluid warming unit on an as needed basis CAUTION 1 To reduce the risks associated with
11. ilavuzu 4U7 v BRE EMIT RR ERA disali cg oll ils 2545 As paita aquecimento de fluidos de irrigac o Ke m ligation Fluid Warming System i a Check the 3M Ranger system website to ensure you have the most recent version of this document www rangerfluidwarming com reorder 202456B Check the 3M Ranger system website to ensure you have the most recent version of this document www rangerfluidwarming com reorder 202456B Ranger Irrigation Fluid Warming System Model 247 Operator s Manual English 1 Francais 15 Deutsch 33 Italiano 49 Espanol 65 Nederland 81 Svenska 97 Dansk 113 Norsk 129 Suomi 145 Portugu s 161 EAAnviKa 177 Polski 195 Magyar 211 Cesky 227 Lietuviy Kalba 243 259 T rk e 275 Baa 291 Xj 307 As pl 323 Check the 3M Ranger system website to ensure you have the most recent version of this document www rangerfluidwarming com reorder 202456B Check the 3M Ranger system website to ensure you have the most recent version of this document www rangerfluidwarming com reorder 202456B Table of Contents Technical Service and Order Placement 3 In warranty Repair and Exchange 3 When You Call for Technical Support 3 Introduction a P 4
12. lternately flashes a temperature of 33 C or lower and the word LO Alert indicator light that comes on when either an over or under temperature condition occurs Ranger Irrigation Fluid Warming Disposable Set Irrigation fluid warming sets include a Cassett cassette tubing connectors flow chamber and clamps Fluid source Inlet tube p White clamp Outlet tube Blue clamp Exit Flow Rate Priming Volume Patient Line Length Irrigation set 0 867 mL min 308 mL 213 cm 1 Refer to the instructions provided with each disposable set for information on use 2 With the bag hung 100 cm above the scope under gravity flow Check the 3M Ranger system website to ensure you have the most recent version of this document www rangerfluidwarming com reorder 202456B English 3M Ranger Irrigation Fluid Warming System Model 247 Product Safety Features The following chart describes the safety alarm features of the Ranger irrigation fluid warming unit Alert Type Set Point Indication Description Primary over temperature 48 C LED alternately displays Primary Alert Alert sounds alert system and temperature controller cuts power to the alert sounds alert heaters until temperature falls light flashes below 48 C When heater temperature drops below 48 C audible and visual alerts stop Independent back up 50 C LED displays the Primary
13. m radius wheelbase and at a height no higher than 44 112 cm Failure to do so may result in damage to the product see Figure 1 2 Slide the irrigation cassette into the slot in the warming unit The cassette can only fit into the device one way see Figure 2 3 Prime the warming set For more information about priming the set refer to instructions provided with the warming sets 4 Plug the power cord into an appropriate outlet Turn the unit ON In a few seconds the alpha numeric display will illuminate It takes less than 2 minutes to warm up to the 41 C set point temperature 5 Begin flow When flow is completed remove the warming set and discard according to institutional protocol Removing the Irrigation Fluid Warming Set from the Ranger Irrigation Fluid Warming Unit Close inlet clamp proximal to the fluid warming cassette and open all clamps distal to the cassette 2 Disconnect the warming set from the fluid source if applicable 3 Allow a small amount of fluid to drain from the end of the patient line to reduce pressure in the cassette 4 Removethe warming set from the warming unit and discard according to institutional protocol Figure 1 44 14 35 6 cm 112 cm Figure 2 English Check the 3M Ranger system website to ensure you have the most recent version of this document English Troubleshooting Condition Nothing illuminates on the
14. n Open flow to reduce temperature Alerts will stop when the temperature drops below 48 C The unit is ready to use Alerts will stop when the temperature drops below 48 C The unit is ready to use Turn off unit and unplug it Discontinue infusion of fluids Do not warm fluids before infusing them through the Ranger irrigation fluid warming unit Solution Power to heating plates will shut off if temperature rises to 50 C Turn unit off and unplug it Discontinue use of unit Discard disposable set Alert will continue to sound if you do not unplug unit Call 3M Patient Warming technical support Solution Make sure test screw is completely tightened If it is missing turn unit off and unplug it Call 3M Patient Warming technical support Check the 3M Ranger system website to ensure you have the most recent version of this document Condition Alert sounds but warming unit has been turned off Condition Cannot remove cassette from unit Condition Alerts indicator light illuminates and alert sounds alpha numeric display alternately flashes a temperature of 33 C or below and the word LO Condition Alpha numeric display reads Er 4 or Open Condition Alpha numeric display reads Er 5 or Open 3M Ranger Irrigation Fluid Warming System Cause Independent backup safety system has been activated Cause e Cassette is too full fluids are still bein
15. oderate injury NOTICE Indicates a situation which if not avoided could result in property damage only p 202456B English Check the 3M Ranger system website to ensure you have the most recent version of this document www rangerfluidwarming com reorder 202456B English 3M Ranger Irrigation Fluid Warming System WARNING reduce the risks associated with hazardous voltage and fire and thermal energy hazards Do not substitute other devices i e Model 245 for the Ranger irrigation fluid warming unit or Ranger irrigation disposable sets Do not continue use of the unit if the over temperature alert continues to sound and the temperature does not return to the set point temperature Immediately stop fluid flow and discard the disposable set Have the warming unit tested by a biomedical technician or call 3M Patient Warming technical service 2 reduce the risks associated with hazardous voltage and fire Do not modify or service this device and do not open the warming unit case there are no user serviceable parts in the Ranger warming unit Connect power cord to receptacles marked Hospital Only outlet Hospital Grade or a reliably grounded Use only the power cord specified for this product and certified for the country of use Do not allow the power cord to get wet Do not use the Ranger warming system when it appears the warming unit power cord or disposable is dam
16. onabrasive solution such as Alconox brand detergent 3 Insertthe tool from the back of the unit and pull the tool all the way out from the front 4 Rinse the tool with water and repeat 3 times Discard the tool according to institutional protocol 5 Wipe off the unit to remove excess fluid TO CLEAN RESISTANT DRIED ON FLUIDS 1 Spray a nonabrasive solution inside the slot of the warming unit and let sit for 15 20 minutes 2 Clean the unit by using the cleaning tool NOTE You may use a nonmetal instrument such as a cotton swab to clean the upper channels If you are unable to adequately clean the unit call 3M Patient Warming technical service Storage Store all components in a cool dry place when not in use Servicing There are no user serviceable parts in the Ranger irrigation fluid warming unit All service must be performed by 3M Patient Warming or an authorized service technician Call 3M Patient Warming technical service at 800 733 7775 or 952 947 1200 for service information Alconox is a trademark of Alconox Inc 13 Check the 3M Ranger system website to ensure you have the most recent version of this document www rangerfluidwarming com reorder 202456B English 3M Ranger Irrigation Fluid Warming System Specifications Physical Characteristics WARMING UNIT 4 5 in 11 cm high x 7 5 in 19 cm wide x 10 in 25 cm long wt 7 lb 7 oz 3 4 kg CERTIFICATIONS IEC EN 6060
17. plates Do not allow spills to dry inside the unit as this may make it more difficult to clean the unit 3 The Ranger irrigation fluid warming unit meets medical electronic interference requirements If radio frequency interference with other equipment should occur connect the unit to a different power source Check the 3M Ranger system website to ensure you have the most recent version of this document Product Description www rangerfluidwarming com reorder 202456B 3M Ranger Irrigation Fluid Warming System English The Ranger irrigation fluid warming system consists of a Model 247 warming unit and a sterile disposable fluid warming set Alert indicator light Alphanumeric display Ranger Irrigation Fluid Warming Unit Slot for warming cassette switch I V pole clamp The warming unit is a compact lightweight liquid resistant device with a clamp located on the side for attachment to an LV pole When mounted to LV pole the unit fits easily above the 3M Bair Hugger 500 or 700 series warming unit A carrying handle on the top of the unit makes it easy to transport Located on the front panel you will find Alphanumeric display that indicates the heater temperature during normal operation In an over temperature condition the display alternately flashes a temperature of 48 C or higher and the word HI An audible alert also sounds In an under temperature condition the display a
18. uid warming system is designed to warm irrigation fluids and deliver these at flow rates under gravity up to 867 mL min when the bag is hung 100 cm above the scope At these flow rates the device maintains fluid output temperatures ranging from 33 C to 41 C Note This is for room temperature fluids only It takes less than 2 minutes to warm up to the 41 C set point temperature The alert points on the Model 247 are 48 C and 50 C The fluid warming disposable set is sterile latex free single use only and is designed to be used with the reusable Model 247 warming unit For additional information on Ranger disposable sets visit us online at rangerfluidwarming com This manual includes operating instructions and unit specifications for the Model 247 irrigation fluid warming system For information about using Ranger irrigation fluid warming disposable sets with the Model 247 irrigation fluid warming unit refer to the Instructions for Use included with each disposable component The Ranger irrigation fluid warming system should only be used in healthcare facilities by trained medical professionals Indications for Use The Ranger irrigation fluid warming system is intended to warm irrigation fluids Definition of Symbols Fuse CAUTION Type B Applied Part Voltage Alternating Current AC An equipotentiality plug grounded conductor other than a protective earth conductor or a neutral conductor providing a direct
19. warming unit panel Condition Alert indicator illuminates and alert sounds alpha numeric display alternately flashes a temperature of 48 C or higher and the word HI Condition Alert sounds alpha numeric display and alert indicator light go dark Condition Unit alerts soon after plugging it in unit does not have to be turned on for this condition to occur Heater temperature rises to 50 C and unit shuts down soon after plugging itin unit does not have to be turned on for this condition to occur 10 3M Ranger Irrigation Fluid Warming System Cause Unit is not turned on plugged in or power cord is not plugged into an appropriate outlet Unit failure Cause Temporary over temperature condition because An extreme change in flow rates occurred e g from 500 ml min to stop flow Unit was turned on and reached set point temperature before warming cassette was inserted Fluids were prewarmed to above 42 C before being run through the warming unit Cause Primary controller failure Unit will no longer operate Cause Test screw on bottom of unit is loose or missing www rangerfluidwarming com reorder 202456B Solution Turn unit on Make sure the power cord is plugged into the power entry module of the warming unit Make sure the warming unit is plugged into a properly grounded outlet Call 3M Patient Warming technical support Solutio