Cryo 5 - User`s Manual
Contents
1. Lightly typing AY results in reducing the air current flow to Level 1 Lightly typing Am results in increasing the air current flow to Level 6 Holding the key pressed modifies the air current flow each second During the Minimal Load or the Defrosting Phase this display indicates When the Minimal Load has been reached it indicates the preset Level 3 4 Start Stop Key Start Stop 5a Defrosting Key r 4 3 5b Defrosting Indication Display 6 After Sales Service Indication Display This key has the following functions n Stand by mode or during additional loading the treatment begins by pressing this key The treatment time counts down by the second During the treatment the compressor functions permanently The treatment can be interrupted by pressing this key The air current flow and the therapy time stop By pressing this key again the interrupted treatment can be continued Therapy time starts again at the previous value Atthe end of the treatment Cryo 5 returns to Stand by Mode All parameters return to their default value The device can be used again for a treatment Defrosting is made in Stand by Mode by pressing both keys simultaneously for a few seconds Defrosting should be started in 2 cases When the air current flow of the Cryo 5 is lower than normal the cold reserve is blocked with ice and the air cannot circulate freely The cause of this is a very humid
2. the Cryo 5 What does it produce What are the advantages of the Cryo 5 How can you reach the desired cold concentration The Running of the Cryo 5 3 Device Description A compact device which generates cold air for skin cooling The Cryo 5 produces cold air at that can be applied with variable air current flows The preset parameters are user friendly A large cold air production capacity is provided Only ambient room air and electricity are required No consumables Depending upon the surface and the access of the area to be treated the device offers 6 air current flow levels level 1 300 l min level 6 700 I min In order to employ Cryo 5 in its entire capacity the daily running stages and principal elements should be understood The Cryo 5 is basically composed of the following elements a cold circuit with a cold reserve a compressor a cooling fan Fan C situated in the low part of the unit a therapy fan Fan T a defrosting system of the cold reserve a micro processor controlled driving system The daily running of the Cryo 5 consists of the following stages The Minimal Load The Total Load Additional The Stand by Important Advice for Optimal Use With Load we mean the time the device needs to produce a certain quantity of cold that it will store in its reserve The Minimal Load of the Cryo 5 begins to be created as soon as the device is switched on Thi
3. tube and nozzels are easy to be cleaned with a non abrasive glass cleaning product The device must be given to specifically authorized personnel or organization for disposal 12 UL Classification The device bears the following UL Mark yy C US CLASSIFIED BY UNDERWRITERS LABORATORIES INC WITH RESPECT TO ELECTRIC SHOCK FIRE MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 2601 1 AND CAN CSA C22 2 NO 601 1 SIE o 1KD1
4. 11 Base 80400751 Castor 75 mm 65370620 mounting bracket for air hose 95372210 Air Hose Changes possible Safety Functional and Safety Test 11 Safety and Maintenance Cryo 5 is produced in accordance with Safety Standard VDE 0750 Part 1 IEC 601 1 UL 2601 Good Manufacturing Practices and Safe Medical Devices Act and Amendments As the manufacturer Zimmer Elektromedizin GmbH can only be held responsible for safety and reliability of the device when the device is connected to the correct hospital grade electrical outlet the device is used in accordance with the User Manual repairs modifications or any other changes are made only by those authorized to do so by Zimmer Elektromedizin The device consists of no parts which must be user supplied The user must only vacuum the filter 11 when the indication light is lit and empty the Water Container 12 when full If any functional and safety test is required proceed as follows These control checks should only be made by knowledgeable qualified personnel with the practical necessities which enable these technical checks to be independently performed The following controls must be made Control of the outside function and safety state of the device and accessories Control of readability of descriptions and displays Control of functioning according User Manual Chapter 4 On Off Switch Control of the protective earth resistance ac
5. Cryo 5 Users Manual Front Panel N q 16 Cryo 5 Back Panel Description of the Front Side 2a 2b 3a 3b 4 5a 5b 10 Therapy Time Adjustment Digital Therapy Time Display Air Current Flow Regulation Key Air Current Flow Indication Display Start Stop Key Defrosting Key Defrosting Indication Display After Sales Service Indication Display Indication Display for Emptying the Defrosted Water Container Indication Display for Cleaning the Filter On Off Indication Display Stand by Indication Display Description of the Back Side 11 12 13 14 15 16 17 18 On Off Switch Filter for Room Air Access Door to the Container for Defrosted Water Power Fuses Identification Plate Power Cable Holder for Treatment Tube Treatment Tube Base and Castors 10 11 12 Cryo 5 Table of Contents Assembly Notice Short Start Up Procedure Device Description Description of the Front Panel Description of the Back Panel Modifications of the Treatment Parameters Indications Contraindications Advice Technical Information Accessories Safety and Maintenance Safety Functional and Safety Test General Comment Error Indications Casing Care Disposal UL Classification Page 12 13 14 15 16 17 18 20 1 Assembly Notice 1 Castors If the desired castors are not mounted on the Cryo 5 proceed as 18 follows Lay the Cryo 5 on its left side opposite to t
6. cording IEC 601 1 maximal value 0 2 Q Control of ground leakage current according UL 2601 maximal value 300 pA General Comment Error Indications Casing Care Disposal The controls checked as well as the results must be mentioned in the device book If any control is not operating correctly the device must be repaired or the user must be informed that it is dangerous This User Manual is to be considered as part of the device itself and is therefore always to be kept in close vicinity of the device The User Manual is the basis for correct usage and handling of the device as well as for the patients and operators safety If an error is detected an error indication is displayed on the screen as follows DE HB In some cases the error is self correctable after the device has been switched off In this case wait a few seconds and switch the device back on If the error indication is still displayed after switching the device back on contact customer service In the event of an Error F22 switch off the device and wait 30 minutes before switching on again Certified Service Personnel can be reached by contacting the sales representative or by contacting Zimmer Elektromedizin directly When the expulsed air or air volume temperature is decreasing defrosting is necessary Cryo 5 casing should be cleaned with commercial plastic cleaner Glass front panel treatment
7. environment When the Cryo 5 will not be used for an extended period i e holidays it is suggested to defrost the unit and to empty the Water Container prior to leaving in order to prevent water accumulation in the device During defrosting the Stand By Indication Display stops blinking and the Defrosting Indication Display is lit The device draws in the ambient air which passes through the cold circuit without being cooled and draws it out through the treatment tube The defrosted water flows into the container at the back of the device The defrosting procedure stops automatically once the temperature of the cold circuit has reached ambient room temperature or manually by pressing both keys again After defrosting the device proceeds to a Minimal Load It indicates that the defrosting procedure is running Once completed the display stops blinking This indication display is lit when the Cryo 5 presents a defect in the functioning It is indicated by an acoustic signal and an error number in the Digital Therapy Time Display F It is then better to switch off the device and call your distributor With the error number he can give instructions on how to proceed 7 Indication Display for Emptying the Defrosted Water Container NN ey V 8 Indication Display for Cleaning the Filter 9 On Off Indica tion Display ER 10 Stand By Indi cation Display This display indicates that the con
8. he one with the holder for the treatment tube Pay attention not to damage the filter at the backside Castor axle can be plugged in the 4 special holes on the base of the unit 2 Holder for Holder is affixed with the 2 corresponding screws on the right Treatment tube side of the unit 16 3 Treatment The tube can be screwed on below the front panel of the Cryo 5 Tube The cold air exit is situated there 17 4 Handpiece To be inserted into the distal end of the treatment hose to provide a smaller more easily directed flow of air 1 Switch on the Device 2 Utilization 3 Beginning of the Treatment 4 Endofthe Treatment 2 Short Start Up Procedure Press the On Off Key 1 The On Off Indication Display 9 is then lit When the Cryo 5 has been switched on it begins to store a reserve of cold air The unit needs 10 minutes to reach the minimal cold temperature The loading time is indicated by the blinking of the Stand By Indication Display 10 during which time the device cannot be used When the unit has reached its minimal cold temperature it is ready for use and the Stand By Indication Display 10 stops blinking It is then constantly lit The unit is preset with an air current flow level of 3 3b and a treatment time of 15 minutes 2b Select the desired air output Press the Start Stop Key The end of the treatment is indicated by the stop of the air current flow and an acoustic signal What is
9. mal Magnetic Circuit Breaker Power Input Power Input in Stand by Volume Level Treatment Temperature Air Volume Protection Class MDD MPG Treatment Tube Length Housing Dimensions Dimensions of the Base Height with Castors Weight Environmental Conditions Storage Conditions Packaging 3 Inches 75 mm 120V 60 Hz 12A or 14A 1 4 KVA about 260 W 70 dBa max 29 F max 35 C 1000l min max Class Type B according IEC 601 1 Class lla 71 inches 180 cm H 38 inches 970 mm W 13 inches 330 mm L 18 inches 460 mm W 470 mm L 510 mm 43 inches 1080 mm 121 Ibs 55 kg Temperature from 50 F to 95 F Steady humidity level from 30 to 95 Pressure from 700 hPa to 1060 hPa Temperature from 38 F to 158 F Relative humidity from 10 to 100 Pressure from 500 hPa to 1060 hPa The device should only be shipped in the original packaging Storage and shipment only with an empty water container Explanation of Labels 9 Device Type B f Protection level against electric shock Technical changes possible Refer to user manual Accessories delivered with unit Available Accessories 10 Accessories Only authorized accessories should be used 1 User Manual 1 air hose 1 mounting bracket for air hose 1 Container for defrosted water Accessory Number 95370000 Air Hose 80025520 Container for defrosted Water 654406
10. n these parameters are modified in Stand by Mode before pressing Start Stop to begin a treatment they are stored until the device is switched off After switching off and on the preset parameters are available again Indications Contra Indications 7 Indications Contra Indications The Cryo 5 Cold Air Device is intended to minimize discomfort associated with laser treatments Alternative uses include skin cooling as a local anesthetic for procedures that induce minor discomfort Hyper sensitivity to cold Areas of impaired sensation Areas of impaired circulation Open wounds Do not use in the presence of explosive gases Caution 8 Advice After transporting Cryo 5 especially sideways wait at least 30 minutes before switching on the unit Do not place the Cryo 5 near a heating device heating system fango unit sauna etc Insure a minimum of 24 free space behind Cryo 5 for air circulation Under extreme environmental conditions over 86 F and high humidity Cryo 5 s cooling capacity may be reduced Electrical and magnetic fields may interfere with the normal functioning of the device The use of Cryo 5 close to any device generating electromagnetic fields radiology diathermy is impermissible Cryo 5 packaging is to be disposed of and kept out of the reach of children Cryo 5 should not be used the presence of explosive or flammable gases 9 Technical Information Operating Voltage Ther
11. ng of keys and indication displays in 2 ways Graphic elements check Typical display check with treatment time and parameters Afterwards Cryo 5 begins to store its cold reserve and all displays indicate until the device is ready for use In order to clean the treatment air of larger impurities the Cryo 5 is equipped with a metallic filter at the back of the device Its maintenance is very simple it should be cleaned regularly with a vacuum cleaner The operation must only be made when the indication display for cleaning the filter is lit By switching on the Cryo 5 or by starting defrosting the cooling system warms and produces defrosted water The Water Container holding approx 2 liters can be reached through the door at the rear of the device It can be opened by pushing the black slide downwards It is advisable to disinfect the Water Container after emptying The level sensor hanging ball should also be cleaned Check before closing the door to be sure that the ball is placed in the Water Container The device should only be used when the Water Container is placed in the unit Air Current Flow Treatment Time 6 Modification of Preset Parameters Cryo 5 provides the possibility to individualize the treatment according to your needs Air Current Flow is preset at Level 3 and can be modified by pressing the keys A Y Treatment Time is preset at 15 minutes it may be modified to 30 00 and 60 00 minutes Whe
12. s phase is characterized by the simultaneous working of the compressor and the Fan C These 2 elements are always functioning together During the Minimal Load no functions are available and the indication displays only bring a horizontal slash no therapy is possible When the unit has reached its minimal cold temperature the Stand By Indication Display 10 stops blinking The compressor and Fan C still function and now produce the Additional or Total Load The Total Load refers to the steady maintenance of the maximum cold temperature in the cold reserve For example During a treatment Starting of Fan T with the Start Key the compressor and the Fan C run automatically in order to compensate for the loss of cold capacity When the Fan T stops end of treatment the compressor and Fan C will continue running in order to complete the cold reserve During this loading period treatment is possible Afterwards the unit remains in Stand By Mode If the device is not used for approximately 30 minutes in Stand by Mode the cold reserve will naturally lose part of its capacity The compressor and Fan C will then start running in order to complete the cold reserve During the Stand by Mode treatment is possible Cold reserve additions are made automatically The frequency required to maintain adequate cold reserve may change according to the climatic conditions around the Cryo 5 The first treatment should be made after the cold re
13. serve is complete or Total Load and not after the Minimal Load Otherwise the possibility exists that the unit may not have enough cold reserve for subsequent treatments it would then automatically proceed to the Minimal Load and treatment would not be possible for a few minutes Displays and Keys 2a Therapy Time Adjustment Ov O Ec E 2b Digital Therapy Time Display B Be 3a Air Current Flow Regula tion Key Av w C E J 3b Air Current Flow Indication Display a E 4 Description of the Front Panel Function The therapy time can be adjusted with these 2 keys Any modification of the treatment time in Stand By Mode will be stored for the following treatments Lightly typing Ow results in decreasing the treatment time from 15 to 0 minutes Lightly typing CD results in increasing the treatment time from 15 to 30 to 60 minutes During the Minimal Load or the Defrosting Phase this display does not indicate a treatment time rather The therapy time is pre set at 15 minutes but is modifiable to suit the user s needs During treatment the time counts down by the second At the end of treatment 00 00 is indicated in the Digital Therapy Time Display The end of the treatment is indicated by the stop of the air current flow and an acoustic signal The unit is preset with an air current flow 3 But it may be modified before as well as during the treatment
14. tainer for defrosted water should be emptied The container is accessible through the door at the back of the device When the indication display is lit no treatment is possible In order to protect the cold circuit and to control the purity of the air drawn out of the Cryo 5 a filter has been installed 11 at the back of the device The indication display is lit when the filter should be cleaned Switch off the device and clean the filter with a vacuum After cleaning the filter display must be switched off manually Proceed as follows Hold the Start Stop key pressed and switch on the device This display is lit as long as the device is switched on If it should go out along with the other indication displays it is possible that a malfunction of the Cryo 5 has occurred which burned out the primary fuses of the device In this case please contact your After Sales Service Representative who will then check the cause of the break down and change the primary fuses if needed When this display is lit the Cryo 5 is ready for use it has enough cold capacity to proceed to treatments When this display blinks the Cryo 5 is proceeding to a Minimal Load or defrosting No treatment is then possible 1 On Off Switch 11 Filter for Room Air 12 Access Door to the Container for Defrosted Water 5 Description of the Back Panel Upon switching on the Cryo 5 an acoustic signal sounds The device checks the correct functioni