USER`s MANUAL
Contents
1. But it will skip the self diagnose and self calibrating Battery installing can also turn on the meter directly 11 MEASUREMENT Measuring and measurement interface Remove the holder and lay the Measurement Aperture just against the position that you want to measure Then push the Measuring Button Within 2 seconds a measurement will be completed Meanwhile the result will be shown in the Display Window as below and stored into the memory automatically 7 Notes Fis kii k the eighth group Angle mode is shown in the first column the values of a current measurement in the second and the averages in the third express that the current measuement is the third of five measurements in 12 MEASUREMENT E Deleting measurements To delete a current measurement please push down operation button for more than 2 seconds Note Only the measurements in a current group can be deleted one by one E Turn off the meter The meter will shut off automatically according to the preset time See page 25 Note After turning off the meter will be in a sleeping state which consume little power Even so it is strongly recommended to remove the battery when no use for a long time interval 13 MEASUREMENT Notice The holder removed from the main device should be keptin a safe and clean place to prevent from damage or pollution While measuring please keep environment light from directly ir
2. then click the Query button to get a result 29 USING SOFTWARE Meter Setup This group of commands is used to set the measuring conditions of MG268 F2 in a communication state AAngle Mode There are six options to be select including single angle double angle and treble angle AGroup Presetting To preset a number of measurements in a group you can choose from 1 to 99 A Verified Values This is for modifying the verified values of the Calibrating Standard in the Holder Note It is very important to keep the verified values correct for a calibrating standard Therefor don t modify those values at will 30 USING SOFTWARE A Shutoff Time This command is for setting a no operation time interval for which the meter will be shut down automatically There are three options to be selected AClear Memory This is for cllearing all the data stored in the memory of MG268 F2 permanently And so It is better to reconsider before clearing Help AContents The contents of help AAbout KSJ Data Operator About the copyright and version of KSJ Data Operator and other informations 31 USING SOFTWARE Tool bar Current Data Equivalent to the corresponding menu command Historical Data 3 Equivalent to the corresponding menu command Print amp Equivalent to the function button of print Print Equivalent to the corresponding menu command Function Button Delet Button To de
3. into the holder A When moisture condensation occurs you may be unable to make a correct measurement A To avoid damage of the meter due to leak of corrosive liquid we strongly recommend you use a high quality battery A After turning off the meter will be in a sleeping state which consume little power Even so it is strongly recommended to remove the battery when no use for a long time interval A When the battery run out Low Battery will appear the Display Window replace the old battery with a new one at once A For environment dispose the wasted batteries properly A The meter should not be taken down at will in order to avoid permanent damage 16 MENU OPERATION E Menu interface On the measurement Interface push the Menu Wheel to enter the menu interface as below E Option selecting Turn up or down the Menu Wheel to select the options Push down the Menu Wheel for entering your option selected The following describes what the symbols mean Return to the prior menu Select to return Enter the next submenu 4 Select to enter Anytime press the Measuring Button to terminate menu operation and return to the measuring interface 17 MENU OPERATION Option Select this option and push down the Menu Wheel to return to measurement interface i menu and its submenu 18 MENU OPERATION inair submenu In data menu select inquiry
4. off by itself in a preset time interval CONSTRUCTION Whole set of MG268 F2 Display Window Menu Wheel Front Shell Rear Shell Bat Chamber Cover CONSTRUCTION Main Device Measuring Button Rear Shell RS232 Port Mark For Location Incidence Direction Front Shell Measurement Aperture CONSTRUCTION Holder With Calibrating Standard Calibrating Standard Gloss Value Label Holder CONSTRUCTION Battery in Position Battery Chamber Cover Battery With Positive Pole Up AA or LR6 Alkaline or Rechargeable Rear Shell Front Shell 10 MEASUREMENT E Turn on the meter Lay the main unit into the holder then push down the Menu Wheel to turn on the meter The infos of model and series number will be shown in the Display Window meanwhile the meter performs self diagnose and self calibrating After passing this procedure the display should turn to measuring mode the meter would be ready for operation Or else it shows as below The messages prompt you to confirm that 1 The calibrating Standard STD is clean 2 The main device is set into the holder properly 3 Recalibrating is required when environment temperature change is great In this case the meter needs some time to establish a balance of temperature before calibrating Notes If the main unit don t lay into the holder it is possible to turn on the meter by pushing down the Menu wheel
5. ATIONS Specifications Unit MG268 F2 Glossmeter Gs 209 0 0 119 9 Gs 20 120 2000 Measuring Range GU Gs 609 0 0 119 9 Gs 609 120 1000 Gs 859 0 0 119 9 Gs 85 120 160 Resolution GU 0 1 Repeatability GU 0 2 0 2 Reproducibility GU 0 5 0 5 Deviation GU 1 5 1 5 Zero point Deviation GU 0 2 Operating Temperature 10 C 40 SPECIFICATIONS MG268 F2 Specifications Unit Storage Temperature 10 C 60 C Relative Humidity Up to 85 no moisture condensation for operation Gs 20 9x10 Measuring Spot Gs 60 9x16 Gs 85 5x39 Power Supply One AA sized battery 1 2V 1 5V Rechargeable or alkaline Consumption It s different to battery types An alkaline battery may support 20000 measurements or more Volume 155x48x75 without a holder Weight 9 400 with a battery and a holder MAIN FUNCTIONS MG268 F2 has the main functions as below key to get gloss values at a measurement The result will be automatically stored into memory e Optional angle groups You may complete a measurement at all three angles You can also choose to do a measurement at one or two angles e Self diagnosis and self calibration This function help to find some faults and so you can deal with them in time eFriendly display interface In measurement mode current values averages and other informati
6. KSJ 268 2 GLOSSMETER USER s MANUAL Please read this manual seriously before using Made in China KSJ Photoelectrical Instruments Co Ltd Add Middle of Binnan Street Dongbin Industrial Zone Donghai Bincheng Quanzhou 362000 China Tel 86 595 22391040 Fax 86 595 22391048 Website www ksj cn www glossmeter cn E mail sales ksj cn keshijia public qz fj cn CONTENTS Brief Introduction eee ee eee Specifications Functions Construction Measurement Warning And Notice Seeds ee ete RENAA Menu Operation Using Software Trouble shooting Warranty VARAN Packing List aw BRIEF INTRODUCTION Thanks for your choice to purchase MG268 F2 MG268 F2 is designed and manufactured according to 1502813 and GB9754 It is also compatible with the standards of ASTM D523 ASTM D2457 GB T13891 GB7706 and GB8807 The technical parameters conform with JJG696 2002 The gloss values can be traced to the those of the National Primary Standards in SIMT the P R China Main Applications e All kinds of coating and finishing surfaces such as paints varnishes printing ink etc Decorative materials such as marble granite polishing brick ceramic tile and so on The other kinds of materials and objects such as plastic woodenware paper etc SPECIFIC
7. e CD coming with MG268 F2 The procedure of intallation and uninstall is same as a normal software and it is unnecessary to be described here 25 USING SOFTWARE How to connect between MG268 F2 and PC There are two connecting cables for MG268 F2 to connect with a PC One is a RS232 Cable 9 pins and the other is a USB Adapter optional extra Using a COM port Simply onnect one terminal to MG268 F2 and the other to a COM port on a PC Using a USB port Connect a RS232 Cable and a USB Adapter and then connect separately the other terminal of the RS232 Cable to MG268 F2 and the USB plug of the USB Adapter to a USB port on a PC Note It is necessary to preinstall a driver for the USB Adapter Please refer to the document in the CD coming with MG268 F2 26 USING SOFTWARE E Software interface Interface of Current Data default Title Bar 27 USING SOFTWARE Interface of Historical Data Title Bar 28 USING SOFTWARE E Menu commands Data Operation A Current Data Connect 268 2 and a PC turn MG268 F2 run KSJ Data Operator set the upload date then click Upload button to upload data Notes The data is uploaded one by one group with sounds and flash indicating Uploading troubles can usually be settled according to the hints of a pop up window A Historical Data Turn to the interface of Historical Data set a time interval for data inquiring
8. e conden sation occurs easily If so don t use the meter any more and wait for the moisture to evaporate See above item also Corrosion caused by a bad battery Abad or over discharged battery will produce corrosive liquid to damage the electric poles and elements When it happens please remove the battery in time and clean the poles with a cotton bud dipped in absolute ethanol If the meter don t work well please contact us for repair 34 TROUBLE SHOOTING E Problems with communication AFailing to communicate through a Series Port on a PC It is all right to use a Series Port to carry out a communication Problems usually occur because MG268 F2 don t connect with a PC properly MG268 F2 is not in a turn on state No data is in meter s memory A Failing to communicate through a USB Port on a PC Please be sure that driver of the USB adapter has been installed properly All connecting cables are connected correctly USB adapter works in order X Notes A USB adapter i e USB Serial Port Convertor be easily purchased in market 35 TROUBLE SHOOTING How to verify your glossmeter A glossmeter is a kind of metrical instruments It is important and necessary to keep its accuracy For this reason the meter must be verified in a period of time usually a year Then we recommend you to verify the glossmeter as follow Send it to a public organization which is authorised by t
9. eter with Checking Standard that is available for purchase 21 MENU OPERATION gt menu and its submenu 22 MENU OPERATION submenu and its next menu Enter either menu select an option and push down the Menu Wheel to confirm Then it will be changed to the measurng interface automatically submenu This menu is for setting a no operation time interval for which the meter will be shut down automatically In the menu select an option and push down the Menu Wheel to confirm And then it will return to the measuring interface 23 MENU OPERATION Enter the menu turn the Menu Wheel to set a number of measurements in a group and confirm Then it will return to the measuring interface automatically Note The range is between 1 to 99 24 USING SOFTWARE Brief introduction KSJ Data Operator is a simple software for data processing Using it we can perform a communication between MG268 F2 and a PC and upload the data of measurements Then we can process the data uploaded and export them to an Excel file or as a printout We can also operate on a PC to set the measuring conditions for MG268 F2 and clear its memory Software installation and uninstall To use KSJ Data Operator you need the following an IBM compatible PC with a CD ROM Microsoft Windows 2000 Windows NT 4 0 Windows XP or later version The software is on th
10. he government and able to perform the gloss verifying Send it to a manufacturer that produces glossmeters Purchase certified standards to verify the glossmeter by yourself Note We also provide some models of Standards for gloss verifying as below MG268S H high gloss MG268S M medium gloss MG268S L low gloss 36 WARRANTY Limited Warranty And Technical Support The limited warranty period is one year commencing on the date of purchase If your meter needs service carry it to the local reseller or contact us at http www ksj cn for repair e We provide spare parts anytime for distributors and customers The inquiries on the gloss measurement may be directed to support ksj cn We also provide glossmeter customizing for a special application Ww Notes The meter should not be taken down at will in order to avoid precision destroying and maintaining difficulty f the breakdown is caused by taking down or inappropriate use we reserve the rights not to guarantee it free in repair 37 PACKING LIST Main Device Model MG268 F2 Holder With Calibrating Standard RS232 Cable KSJ Data Operator Carrying case User s Manual Tracable Certificate Note 1 Unit 1 Unit 1 Piece 1 2 1 1 USB Adapter is provided an optional accessory 38
11. lete selected data shown in the Data Area permanently Delet All Button To delete all data shown in the Data Area permanently Excel Button To export data to an Excel document Print Button To make a printout 32 TROUBLE SHOOTING E Errors in self diagnose and self calibration When the meter is turned on with the Holder the warning message right hand side may come out It says something would be wrong Please check and settle the troubles in turn 1 1 Clean the Calibrating Standard Observe the surface of the Calibrating Standard against light If any dirt such as dust fingermarks steam and filth is found it should be cleaned gently with lens paper or a piece of soft cloth dipped in absolute ethanol Some abrasion and damage on the surface will change its verified values In this case please contact the manufacturer to deal with Set the main unit into the Holder Only when the main unit lay in the Holder properly the meter can pass the procedure of self diagnose ang self calibration 33 TROUBLE SHOOTING Recalibate the meter A big temperature difference between environment and the meter would badly affect measuring and the warning message will be shown In this case please wait for a period of time till the temperature get to a balance and then calibrate the meter Moisture condensation When the meter is taken from cold location to a hot environment moistur
12. on are synchronously shown on the Display Window The clear characters and simple menus make operation more easily e Presettable measurement number a group According a requirement you can choose a proper number between 1 to 99 for a group A correct choice will make measurement and data processing easier eLarge memory for measurement data In 3 angle mode more than10000 measurements or 999 groups can be allowed to stored in MG268 F2 enquiry for data in memory Through menu operation you can browse the data in memory MAIN FUNCTIONS e Deleting defective measurements By pressing Measuring Button for two seconds more the measurement displaying on the Display Window will be deleted This function is only effective in current group RS232 port 9 pins Through this series port MG268 F2 can communicate with a PC ePractical and friendly software KSJ Data Operator is not only friendly but practical It can help you to upload the data from MG268 F2 preset the measuring conditions make a printout perform data processing and export the data to an Excel document power consumption MG268 F2 needs only a 1 2v to 1 5v DC power supply You can choose a piece of AA battery which may be alkaline or rechargeable An alkaline battery will support more than 20000 measurements The power for communicating is taken from a PC for low battery and full memory e Automatic shutoff If no operation MG268 F2 will be turned
13. option and push down the Menu Wheel to enter the inquiry submenu Turn up and down to set the group number that you want to inquire and then push down the Menu Wheel to enter the inquiry interface Here you can browse all the measurements in this group by turning the Menu Wheel Notes 15 represents the first measurement a current group that has been preset to involve five measurements represents the averages of this group 2 Turn up and down the Menu Wheel to browse data Push down the Menu Wheel or press the Measuring Button to exit inquiry and return to the measuring interface 19 MENU OPERATION submenu Enter the submenu and turn the Menu Wheel to select to clear all data in memory Or else select Menu and its submenu 20 MENU OPERATION submenu Enter the submenu turn the Menu Wheel to select a value and push down the Menu Wheel to confirm then turn the Menu Wheel to modify the value push down the Menu Wheel to confirm again It is the same operation for other values Calibrate option In menu turn the Menu Wheel to select option and push down the Menu Wheel to confirm The calibrating will be carry out automatically and meanwhile the result will be shown in the Display Window Finally it will returnto the measurement interface Notes 1 Before calibrating be sure that the Calibrating Standard is clean 2 You can also calibrate the m
14. radiating into the Measurement Aperture If necessary especially under strong light shade it with a piece of light tight cloth A big temperature difference between environment and the meter would badly affect measuring In this case please wait for a period of time till the temperature get toa balance and then calibrate the meter again f measurement operation last a long time for example an hour or more it s necessary to recalibrate the meter Pay attention to keep Calibrating Standard and Measurement Aperture clean Any dirt such as fingermarks dust steam and filth should be cleaned gently with lens paper or a piece of soft cloth dipped in absolute ethanol 14 MEASUREMENT If memory is full of data a prompt would be shown as below in Display Window In this case measuring operation will be in order but the results can t be stored into memory any more It is better to upload the data in memory to PC and clear the memory When a battery runs out the following message will be shown Please replace the old battery with a new one 15 WARNING AND NOTICE A The whole set of the meter should be kept in dry and non corrosive environment under the temperature between 10 C to 60 C A Although the exterior parts of the meter are made of good quality plastic they may not resist some chemicals expecially some strong solvents such as acetone and so on A meter is not in use please lay it