Home

Quick Immunity Sequencer - Kikusui Electronics Corp.

image

Contents

1. 7 2 7 2 Preparing the Test Conditions 7 3 7 3 Results of Sequence Running rrrrree 7 6 Chapter 8 IEC 61000 4 29 8 1 8 1 Outline of the Test 8 2 8 2 Preparing the Test Conditions 8 2 8 2 1 DC Voltage Dips and Short Interruptions 8 2 8 2 2 DC Voltage Variations 8 5 8 3 Results of Sequence Running 8 7 Appendix A 1 A 1 Test Standard Selection Guide A 1 Index l 1 SD003 PCR LA Safety Precautions Before performing an immunity test using this application software carefully read the operation manuals for the following products and the specific hardware descriptions found therein to avoid improper connections or incorrect handling Incorrect connection or handling of any equipment device in the following configuration may result in serious accidents involving EUT damage or fire AC Power Supply PCR LA Series AC Power Supply PCR L Series Conventional GPIB interfaces IBO3 PCR LA and IB11 PCR L Three phase drivers 3P03 PCR LA and 3P02 PCR L Immunity tester IT01 PCR L Definition of Symbols XCAUTION Indicates the possibility of damage to the test system or DUT if you neglect the CAUTION statement and perform improper operation Make sure to follow the operating
2. Mseq1 R E q2 Seq3 Seq4 Seqs Seas WA ia T ClsSeq3 f a 61000 4 11 Ciseq4 O Over swing a Sweep in frequency Odd non multiple of 3 O Odd multiple of 3 NO Even harmonics C 5eq5 CSeq 6 C 5eq7 C 5eq8 seq9 Seq 10 61000 4 13 6100 4 27 K al 6100 4 28 DC 6100 4 29 7 Trigger 13 Repeat all 6 Repetition count 05V 00a 9 Seq end Output off 11 UT Fig 4 3 Sweep in Frequency Setting Window Angle Duration Interval Frequency sweep method This test uses the AC Power Supply PCR LA s waveform banks The waveform banks are memory for storing special waveforms Up to 14 types of waveforms can be used Quick Immunity Sequencer goes from frequency to frequency among the harmonic frequencies stored in the waveform banks to perform a stepwise sweep Determining sweep frequencies Specify the sweep frequencies by harmonic orders that is integral multiples of the fundamental frequency This means that the sweep frequency step must be an integral multiple of the fundamental frequency SD003 PCR LA IEC 61000 4 13 4 9 NOTE e As shown in Fig 4 4 the sweep frequency step is an integral multiple of the fundamental frequency Because the frequency changes significantly compared with the standard s requirements this feature can be used only in preliminary tests gt S5min step 1 x fl Un
3. Select one of 1 3 5 and 10 times or enter a value ranging from 1 to Input range 9999 8 4 IEC 61000 4 29 SD003 PCR LA 8 2 2 DC Voltage Variations 2 Test type 4 Test level 1 Seq tabs 3 UT 8 Repetition interval a Test1 qco Quick Im nunity Sequen er 100 a iA DER Eile Edit Instrument Opti E y o T 2 Open Cond Out On To 200 Expand Contract Abort 11 Sequences to run CDC Voltage dips DC Short interruptions gt DC Voltage variations ma GEL Seq2 Seq3_ Seq seas f sege Seq7 seas sibo sea10 61000 4 11 61000 4 13 haf eq end derr PO output off H7 i i i i v i j 5 m 0 4 28 DC ae 0 A E e e a E A e ee dho 4 29 1 0 Trigger 9 Repetition count 6 Reduced 7 Voltage voltage time increasing time 13 Repeat all 5 Voltage decreasing time all 12 Seq end Output off Fig 8 2 DC Voltage Variations Setting Window iv The following items other than 2 4 5 6 and 7 are the same as for DC Voltage dips and DC Short interruptions Please refer to 8 2 1 DC Voltage Dips and Short Interruptions 1 Seq tabs 2 Test type Select DC Voltage variations 3 UT 4 Level test level For DC Voltage variations enter the rate of change of DC voltage with respect to UT Enter one of 80 85 and 120 or enter a value rangin
4. cre errr ere rere reer eens 1 2 1 2 1 Preparing for Installation 1 2 1 2 2 Installing the KI VISA Driver 1 3 1 2 3 Installing Quick Immunity Sequencer 1 6 Chapter 2 Overall Picture of Test Sequences 2 1 2 1 Outline of Immunity Test 2 2 2 2 Test Configuration 2 3 2 3 Launching Quick Immunity Sequencer 2 4 2 4 Setting up I O Configuration 2 5 2 5 Troubleshooting at Startup 2 6 2 5 1 Setting up KI VISA 2 7 2 6 Overall Picture of Test Sequences 2 10 2 7 Menus and Toolbar 22 c ect e cece rer c eres 2 12 2 7 1 Menus cer cere rer eer reece reer reece 2 12 2 7 2 Toolbar 22 c cere r err ee ee ee eee eeee 2 15 2 8 RunningaTest 0c e crete errr eer etre eee 2 17 2 8 1 Turning Power on off in the Test Configuration 2 17 2 8 2 Running Procedure ere rcer cee ee 2 18 2 8 3 Troubleshooting During Running 2 19 2 8 4 Corrective Measures in the Event of Problems 2 19 2 9 Observing a Test Signal 2 19 Chapter 3 IEC 61000 4 11 3 1 3 1 Outline of the Test
5. 2xfl 10x fl 20x fl 40xfi f Hz Fig 4 4 Example of Sweep in Frequency A frequency sweep test switches the waveform banks total 14 to change the frequency stepwise The contents of the waveform banks are harmonic orders the ratio to the fundamental voltage and durations The following items other than 2 3 4 5 and 7 are the same as for the combined harmonics flat curve Please refer to 4 2 1 Combined Harmonics Flat Curve 1 Seq tabs 2 Test type Select Sweep in frequency from the following e Flat curve e Over swing e Sweep in frequency e Odd non multiple of 3 e Odd multiple of 3 e Even harmonics 4 10 IEC 61000 4 13 SD003 PCR LA NOTE SD003 PCR LA 3 Harm harmonic order Enter the order of the harmonics superimposed on the fundamental wave A sweep step progresses downward from the highest row Input range Enter a value ranging from to 40 4 Level Enter the ratio of the harmonic to the fundamental voltage superimposed on the fundamental wave Input range Enter a value ranging from 0 0 to 100 0 5 Duration Enter the duration of the harmonic superimposed on the fundamental wave Enter a value ranging from 0 001 seconds to 999 hours 999 minutes Input range 909 999 seconds e Waveform switching timing is at the 0 phase angle of the fundamental wave That is the current waveform is not switched at the end of th
6. 10 000 v Wimis ANIA i Interval 6 Repetition count 4 Duration 13 Repeat all 7 Trigger AA 9 Seq end Output off Fig 4 2 Over Swing Setting Window The following items other than 2 and 3 are the same as for the flat curve Please refer to 4 2 1 Combined Harmonics Flat Curve 1 Seq tabs 2 Test type Select Over swing from the following e Flat curve e Over swing e Sweep in frequency e Odd non multiple of 3 e Odd multiple of 3 e Even harmonics SD003 PCR LA IEC 61000 4 13 4 7 3 Harm Level Angle test level Enter the ratio of the 3rd and 5th harmonics to UT superimposed on UT Likewise enter the phase angle Voltage ratio Enter a value ranging from 0 0 to 100 0 input range Phase angle input range Enter a value ranging from 0 deg to 360 deg Even when the test level is changed the rms value remains at the UT value 4 Duration 5 Interval repetition interval 6 Repeat repetition count 7 Trigger 8 Sequences to run 9 Seq end Output off 10 Bank 11 UT 12 Freq 13 Repeat all 4 8 IEC 61000 4 13 SD003 PCR LA 4 2 3 Sweep in Frequency 10 Bank 2 Test type 8 Sequences to run 1 Seq tabs 3 Harmonic order 4 Level 5 Duration DoR il 100 an Mild f Out On To 200 Expand Contrac Ea Testl qco Quick Im nunity Sequencer Eile Edit Instrument Optidh p Open Cond E
7. 13 Repeat all 7 Trigger 9 Seq end Output off Fig 4 1 Flat Curve Setting Window Set the following items Select a value provided in the dropdown list or enter a value within the input range into the Edit box After entering a value press the Enter key to accept it 1 Seq tabs Allows you to set the test conditions in the panel of each seq tab Seq 1 to Seq 10 These tabs show the heading numbers of the test conditions to be set in the following Test conditions can be copied or pasted between seq tabs For more information please refer toEdit Menu in 2 7 1 Menus SD003 PCR LA IEC 61000 4 13 4 3 NOTE 2 3 4 4 4 IEC 61000 4 13 Test type Select Flat curve from the following e Flat curve e Over swing e Sweep in frequency e Odd non multiple of 3 e Odd multiple of 3 e Even harmonics x UT x sqrt2 test level Enter the ratio of the clipped section to the UT s peak value UT x 2 If no clipping is made the value should be 1 Select one of 0 80 0 90 and 0 95 or enter a value ranging from 0 78 to Input range 0 99 Input range with Select either 0 90 or 0 95 or enter a value ranging from 0 89 to 0 99 ITO1 PCR Even when the test level is changed the rms value remains at the UT value Duration Enter the duration of the flat curve waveform Enter the value in seconds or in hours minutes seconds format Select 120 s or enter a value ranging
8. 3 2 3 2 Preparing the Test Conditions 3 3 3 2 1 Voltage Dips and Short Interruptions 3 3 3 2 2 Voltage Variations ec rere rete cere 3 7 3 3 Results of Sequence Running rr e 3 9 Chapter 4 IEC 61000 4 13 4 1 4 1 Outline ofthe Test 4 2 4 2 Preparing the Test Conditions 4 3 SD003 PCR LA Contents 3 4 Contents 4 2 1 Combined Harmonics Flat Curve 4 3 4 2 2 Combined Harmonics Over Swing 4 7 4 2 3 Sweep in Frequency r crc eect eee 4 9 4 2 4 Individual Harmonics 4 12 4 3 Results of Sequence Running 4 15 Chapter 5 IEC 61000 4 14 5 1 5 1 Outline ofthe Test 5 2 5 2 Preparing the Test Conditions 5 3 5 2 1 Voltage Fluctuation cee c errr ere ree 5 3 5 2 2 Interval c errr reer errr reer eee 5 6 5 3 Results of Sequence Running r rrr rrr eee 5 8 Chapter 6 IEC 61000 4 27 6 1 6 1 Outline ofthe Test eee eee 6 2 6 2 Preparing the Test Conditions 6 3 6 3 Results of Sequence Running rrr rrr rrr eee 6 6 Chapter 7 IEC 61000 4 28 7 1 7 1 Outline of the Test
9. 8 Interval repetition interval 10 000 seconds 9 Repeat repetition count The repetition count in the Seq 1 tab is not indicated No 2 DC Voltage dips 70 1 Seq tab The Seq 2 tab is used 2 Test type DC Voltage dips 3 UT 230 V 4 Level test level 70 5 Duration 0 100 seconds 7 Interval repetition interval 10 000 seconds 8 Repeat repetition count The repetition count in the Seq 2 tab is not indicated No 3 DC Vhort interruptions 1 Seq tab The Seq 3 tab is used 2 Test type DC Short interruptions 3 UT 230 V 4 Level impedance Not indicated 5 Duration 0 100 seconds IEC 61000 4 29 8 7 e 7 Interval repetition interval 10 000 seconds e 8 Repeat repetition count The repetition count in the Seq 3 tab is not indicated Items common to No 1 No 2 and No 3 e Trigger Not indicated e Sequences to run Sequences are run in the order of Seq 1 Seq 2 and Seq 3 e Seq end Output off Not indicated e Repeat all The number of repetitions for the sequences to be run in the order of Seq 1 Seq 2 and Seq 3 is 1 8 8 IEC 61000 4 29 SD003 PCR LA Appendix Guides users to the appropriate standard depending on the nature of the testing desired A 1 Test Standard Selection Guide Slow DC voltage changes Single or Items you wish to test Generic term three phase or Standard Refer t
10. Lets you assign a desired name to a test condition file and save it in any folder This menu item is used to save all the settings related to the standard E Save As Test Result Allows you to assign a name to a running result file and save it to a selected folder E Exit Exits the Quick Immunity Sequencer E Seq tab Copy Copies all the items of a selected seq tab into memory This function is slightly similar to copying to the clipboard E Seq tab Paste Copies the contents of the memory copied using Seq tab Copy to a selected seq tab 2 12 Overall Picture of Test Sequences SD003 PCR LA E Copy to next Seq tab This function is a combination of Seq tab Copy and Seq tab Paste It copies all items of the selected seq tab to memory and pastes them to the next tab E Default data for Class1 2 3 X Each of classes 1 to 3 and X has submenus for test levels 1 to 6 Inapplicable classes and test levels are grayed out and cannot be selected Clicking a test level makes it possible to detect the selected standard buttons and option buttons and to set default values conforming to these buttons Instrument Menu A CAUTION SD003 PCR LA E Send Sends the set test conditions to the PCR LA When transmission is completed the Run button is enabled This menu item has the same function as the Send button on the toolbar E Run Starts the test This menu item has the same function as the Run button on the toolbar E Sto
11. Quick In munity 5equence DER a i 9 100 a 7 j Out On To 200 Expand Corgiract Abort File Edit Instrument Opt we Open Cond s Seq2 Seq3_ Seq4 Seg5 J a6 Seq Seg8_ Seq9 sea 10 7 Seqi Ag ic Seq2 gt ENEN Cl5eq3 voltage fluctuation s Delta2 Delta 61000 4 11 seq 4 O Interval 2 0 m i 8 0 Ml a0 mly Jseq5 RT Cl5eq6 Cl5eq7 61000 4 13 L seq8 1seq9 mgt Bist eq end ELFI SPAL Output off 6100 4 27 e F rigger 6140 4 28 DC ica a j 6100 4 29 lt i 12 Freq 6 Repetition count 13 Repeat all 11 UT 7 Trigger 9 Seq end Output off DEV DOA Fig 5 1 Voltage Fluctuation Setting Window Re Delta2 Re Delta3 Re Set the following items Select a value provided in the dropdown list or enter a value within the input range into the Edit box After entering a value press the Enter key to accept it 1 Seq tabs Allows you to set the test conditions in the panel of each seq tab Seq 1 to Seq 10 These tabs show the heading numbers of the test conditions to be set in the following Test conditions can be copied or pasted between seq tabs For more information please refer to Edit Menu in 2 7 1 Menus SD003 PCR LA IEC 61000 4 14 5 3 2 Test condition type Select Voltage fluctuation from the following e Voltage fluctuation e Int
12. Select one of 1 3 5 and 10 times or enter a value ranging from to Input range 9999 7 Trigger Select trigger signal output The red symbol W_ indicates a location selected using the checkbox 8 Sequences to run Check the corresponding checkboxes of the seq tabs whose test conditions are used to run a test sequence Test sequences are run in turn starting at Seq 1 Seq tabs whose corresponding checkboxes have not been checked are not used to run a test 9 Seq end Output off When this checkbox is checked a test signal is turned off at the end of sequence running the AC Power Supply PCR LA output is turned off 10 Bank This test uses the AC Power Supply PCR LA s waveform banks The waveform banks are memory for storing special waveforms Up to 15 types of waveforms can be stored One of the stored waveforms sine wave is a reference waveform for the AC power supply and cannot be overwritten Thus the actual number of waveform types that can be stored is 14 For the Bank indication the fluctuating indicated bar is displayed in green If this bar is displayed in red the number of waveforms to be stored has exceeded 14 types In this case the Send button remains disabled to prevent further testing e Waveform bank data is stored temporarily not backed up Waveform bank data is lost when the AC Power Supply PCR LA POWER switch is turned off IEC 61000 4 13 4 5 11 UT Enter the initial voltage equal
13. e 1 Seq tab The Seq 4 tab is used e 2 Test type Odd multiple of 3 e 3 Harm harmonic order 3rd e 4 Level 1 0 e 5 Angle 0 e 6 Duration 5 000 seconds e 7 Interval 1 000 second No 1 2nd row Odd harmonics multiples of 3 e 1 Seq tab The Seq 4 tab is used e 2 Test type Odd multiple of 3 e 3 Harm harmonic order 9th e 4 Level 1 0 e 5 Angle 0 e 6 Duration 5 000 seconds e 7 Interval 1 000 second No 1 3rd row Odd harmonics multiples of 3 e 1 Seq tab The Seq 4 tab is used e 2 Test type Odd multiple of 3 e 3 Harm harmonic order 15th e 4 Level 1 0 e 5 Angle 0 e 6 Duration 5 000 seconds e 7 Interval 1 000 second IEC 61000 4 13 4 17 Items common to No 1 1st to 3rd rows e Repeat repetition count The repetition count in a seq tab is not indicated e Trigger Not indicated e Sequences to run Sequences are run in the Seq 4 tab e Seq end Output off Not indicated e Bank Not indicated e UT 230V e Freq Not indicated e Repeat all The number of repetitions for the sequences to be run in Seq 4 is 1 4 18 IEC 61000 4 13 SD003 PCR LA Pt Chapter 5 IEC 61000 4 14 Describes the Voltage fluctuation immunity test according to the IEC 61000 4 14 standard SD003 PCR LA IEC 61000 4 14 5 1 5 1 Outline of the Test The Voltage fluctuation immunity test determines the equipment s immunity against Voltage fluctua
14. Segs Seq9 Seqt Seq 2 M Seq 3 O voltage dips 7 61000 4 11 d Seq Short interruptions L seqs Voltage variations A oO Seq 6 F Seq 61000 4 13 L seq8 ect Ces ees M tease ot cr ae L seq9 100 0 0 ivr 4 wm Pak Jik Seq end 61000 4 14 CI Output off R t 1 me he i DefFeasing Reduced Indjeasing Interval 61000 4 27 Jf fFrea 50 00 sH trigger oo w o y joo vy oto AMI epeat all 1 omogo himis himis himis himis m7 A ANY AAA No Repeat all Seq Item UT Level Angle Decrea Reduced Increasing Interval il E i Voltage Variation 100 0 40 0 100 0 100 0 100 0 100 Repeat ll 1 Seq Repeat 0 00 00 89 6 0 004 Fig 2 4 Window Outline Example of IEC 61000 4 11 E Window Outline 1 Selection of the standard Select one of the following six standards e TEC61000 4 11 Voltage dips Short interruptions and Voltage variations e TEC61000 4 13 Harmonics and Interharmonics e TEC61000 4 14 Voltage fluctuation e TEC61000 4 27 Unbalance IEC61000 4 28 Power frequency variations IEC61000 4 29 Voltage dips Short interruptions and Voltage variations for DC power inputs 2 Test condition settings This part consists of the seq tabs Seq 1 to Seq 10 checkboxes and edit boxes 2 10 Overall Picture of Test Sequences SD003 PCR LA Seq tabs Seq 1 to Seq 10 Each of the seq tabs Seq 1 to Seq 10 allows you to set test conditions corresponding to the standards The color of each tab corr
15. Unbalance e TEC61000 4 28 2002 07 Edition1 1 Power frequency variations e TEC61000 4 29 2000 08 Ist Edition Voltage dips Short interruptions and Voltage variations for DC power inputs The following discussions give only the standard number The issue date and standard version are omitted The equipment and devices to which each standard is applied are those whose rated current per phase is 16 A or less with the exception of IEC 61000 4 29 Note that Quick Immunity Sequencer was not designed with above maximum current limitation in mind Thus the product is used at the user s discretion Parts of Quick Immunity Sequencer do not conform to certain standards For more information please refer to Hl Items not conforming to the standard s requirements in each chapter Preface P 1 P 3 Required Operating Conditions NOTE P 2 Preface Quick Immunity Sequencer requires the following PC environments OS version Windows 98 SE Windows Me Windows 2000 or Windows XP VISA driver supporting the VISA COM software One of the following is required e KI VISA 2 2 x or later latest version may be downloaded from the Kikusui website e NI VISA 2 6 or later e Agilent VISA 2 x separate Agilent VISA COM setup is required ActiveX instrument driver perldrv Ver 2 91 or later latest version may be downloaded from the Kikusui website 1 0 interface One of the following is required e RS 232C e GPIB
16. at 60 Hz E Three phase operations The test conditions are all set in relation to phase U Test signals change simultaneously in three phases on the basis of phase U SD003 PCR LA 3 2 Preparing the Test Conditions 3 2 1 Voltage Dips and Short Interruptions 2 Test type 6 Duration 10 Sequences to run 1 Seq tabs 3 UT 4 Test level 7 Repetition interval fk Test1 qco Quick In munity Seq encer AER Eile Edit Instrument Opti O g S 9 101 gt Md 7 Open Cond Out O To 20V Expand Contract Abort Ie Voltage dips 61000 4 11 O Short interruptions i yr F Interval J ae O Voltage variations 1000 4 13 ty 61000 4 14 V seq2 s5eq3 5eq 4 C seq5 seq6 seq7 seq8 seq9 5eq 10 oO o q 100 0 v Vrms m y hange angle 4 F gt 0 v de potz frea 50 00 rigger T 4 gt Qa D0 4 28 Ei YV p0 4 29 J a mie 12 Freq 8 Repetition count 5 Starting phase ange gs Interval 13 Repeat all 9 Trigger aay 11 Seq end Output off f 000A Fig 3 1 Voltage Dips and Short Interruptions Setting Window Set the following items Select a value provided in the dropdown list or enter a value within the input range into the Edit box After entering a value press the Enter key to accept it 1 Seq tabs Allows you to set the test conditions in the panel of
17. 000 seconds e 8 Repeat repetition count The repetition count in the Seq 3 tab is not indicated Items common to No 1 No 2 and No 3 e Trigger Not indicated e Sequences to run Sequences are run in the order of Seq 1 Seq 2 and Seq 3 e Seq end Output off Not indicated e Freq Not indicated e Repeat all The number of repetitions for the sequences to be run in the order of Seq 1 Seq 2 and Seq 3 is 1 3 10 IEC 61000 4 11 SD003 PCR LA Po Chapter 4 IEC 61000 4 13 Describes Harmonics and interharmonics immunity tests according to the IEC 61000 4 13 standard SD003 PCR LA IEC 61000 4 13 4 1 4 1 Outline of the Test The Harmonics and interharmonics immunity tests determine the equipment s immunity against distortions in supply voltage waveforms Shown below are the specific tests that comprise these immunity tests e Flat curve e Over swing e Sweep in Frequency e Odd harmonics non multiples of 3 e Odd harmonics multiples of 3 e Even harmonics e Interharmonics e Meister curve E Test sequence The tests that determine compliance with the standard provide the recommended test sequences in flowchart form for each test To enable greater flexibility in preliminary tests and other tests Quick Immunity Sequencer does not adopt these sequences Instead Quick Immunity Sequencer is designed to allow each test to be performed independently E Electromagnetic environment classes Quick Immunity Sequencer
18. 110 V 120 V 200 V 220 V 230 V and 240 V or Input range enter a value ranging from 0 0 V to 305 0 V NOTE e An appropriate voltage range must be selected Use the toolbar or the Instrument Menu to set the voltage range For more information please refer to E To 100 V 200 V button in 2 7 2 Toolbar 13 Freq Enter the rated frequency of the EUT Select either 50 Hz or 60 Hz or enter a value ranging from 1 Hz to Input range 999 90 Hz 14 Repeat all Enter the repetition count of all the selected seq tabs Seq 1 to Seq 10 Select one of 1 3 5 and 10 times or enter a value ranging from to Input range 9999 SD003 PCR LA IEC 61000 4 27 6 5 6 3 Results of Sequence Running The running results are indicated every time the sequence for one seq tab finishes The starting date and time are shown in the row above the No 1 row Shown below is a typical test example An indicated value becomes a blank or 0 if it is not related to the test The Repeat set value in the panel for each seq tab is not indicated Table 6 1 Example of Indication of Running Results Repeat all Ua Ub Uc Ub Uc Dura Inter em Ur mea Angle Angle tion val 1 1 1 Unbalance 230 0 50 0 100 0 95 2 90 0 125 240 30 60 2 1 2 Unbalance 230 0 50 0 100 0 90 0 80 0 131 239 15 60 3 1 3 Unbalance 230 0 50 0 110 0 66 0 71 0 139 2
19. 2 000 10 000 2 1 2 Voltage dips 230 0 70 0 0 0 0 200 0 0 10 000 3 1 3 Short interruption 230 0 0 0 0 0 0 0 100 0 0 10 000 SD003 PCR LA E Description of Table 3 1 Number in the following items corresponds to the numbered steps in Figs 3 1 and 3 2 No 1 Voltage variations 40 e 1 Seq tab The Seq 1 tab is used e 2 Test type Voltage variations e 3 UT 230 V e 4 Level test level 40 e 5 Decreasing time voltage decreasing time 2 000 seconds e 6 Reduced reduced voltage time 1 000 second e 7 Increasing time voltage increasing time 2 000 seconds e 8 Interval repetition interval 10 000 seconds e 9 Repeat repetition count The repetition count in the Seq 1 tab is not indicated No 2 Voltage dips 70 e 1 Seq tab The Seq 2 tab is used e 2 Test type Voltage dips e 3 UT 230 V e 4 Level test level 70 e 5 Change angle starting phase angle 0 e 6 Dips Short duration 0 200 seconds e 7 Interval repetition interval 10 000 seconds e 8 Repeat repetition count The repetition count in the Seq 2 tab is not indicated No 3 Short interruptions e 1 Seq tab The Seq 3 tab is used e 2 Test type Short interruptions e 3 UT 230 V IEC 61000 4 11 3 9 e 4 Level test level 0 e 5 Change angle starting phase angle 0 e 6 Dips Short duration 0 100 seconds e 7 Interval repetition interval 10
20. Cancel 7 Check the Use GPIB instruments NI 488 2M compatible API checkbox 8 Enter for example GPIBO in the Index box and Interface PCI 4301 in the Model box Check the Use this board checkbox 9 Click the Enable VISA C API button 10 Follow the prompts displayed in the window The KI VISA Setup window shown in step 6 is displayed once again 11 Click the OK button This completes the setup process 1 2 3 Installing Quick Immunity Sequencer gt Click on Installation of Quick Immunity Sequencer in the Setup Start Window in Fig 1 1 on page 1 3 This launches the Installation window The installer program installs Quick Immunity Sequencer and the ActiveX driver for the PCR LA series jie Kikusui Quick Immunity Sequencer BETA_2004 01 22 InstallShi__ R Welcome to the InstallShield Wizard for Kikusui Quick Immunity Sequencer KIKUSUI BETA_2004_01_22 ELECTRONICS The InstallShield R Wizard will install Kikusui Quick Immunity Sequencer BET4_2004_01_22 on your computer To continue click Next M mi Instruments and Power Supplies WARNING This program is protected by copyright law and international treaties Cancel 1 6 Setup SD003 PCR LA 2 Follow the prompts displayed in the window BETA 2004 01 22 Installshi R ja Kikusui Quick Immunity Sequencer i License Agreement at i Please read the following license agreement carefully SR EEEESTION 1 SOFTWARE LICENSE AGREE
21. Certain Quick Immunity Sequencer items do not conform to the standard s requirements in the test configuration including the hardware The following table shows these items and the performance achieved in the test configuration Performance achieved Item Standard s requirements in the test configuration A test is conducted with settings made for only one Phase basis setting phase Only phase U can be for three phase operations Ideally any of phases U specified V and W can be specified as a phase to be tested 18 at 50 Hz or 21 6 1 Within 10 at 60 Hz with the exception of zero cross Phase setting for three phase operations 1 In single phase operations the standard s requirements are met if the ITO1 PCR L immunity tester is used e We recommend using the ITO1 PCR L immunity tester for single phase operations e The standard s performance specifications relating to voltage rise and fall time of a test voltage generator 1 us to 5 us are met The performance achieved without the ITO1 PCR L is approximately 30 us e The standard s performance specifications relating to the starting phase angle of Voltage dips or Short interruptions setting accuracy within 10 are met Without the ITO1 PCR L the standard s performance specifications will be met at an angle of multiples of 18 such as 90 180 etc for 50 Hz The setting accuracy at other phase angles is 18 at 50 Hz or 21 6
22. Example of Indication of Running Results UT ni a Item wr ur mee Dani REPS Dans REPO pans t all ition ition2 ition3 val2 1 1 5 Voltage 2300 100 0 8 0 5 000 8 0 5 000 8 0 5 000 fluctuation 2 1 6 Interval 230 0 100 0 90 0 160 0001 60 000 5 8 IEC 61000 4 14 E Description of Table 5 1 Number in the following items corresponds to the numbered steps in Figs 5 1 and 5 2 No 1 Voltage fluctuation 1 Seq tab The Seq 5 tab is used 2 Test condition type Voltage fluctuation 3 UT test level 100 0 4 Deltal Voltage fluctuation 8 0 5 Repetion1 repetition interval 5 000 seconds 4 Delta2 Voltage fluctuation 8 0 5 Repetion2 repetition interval 5 000 seconds 4 Delta3 Voltage fluctuation 8 0 5 Repetion3 repetition interval 5 000 seconds No 2 Interval 1 Seq tab The Seq 6 tab is used 2 Test condition type Interval 3 UT test level 1 100 0 4 UT Interval2 test level 2 90 0 5 Interval repetition interval 60 000 seconds SD003 PCR LA Items common to No 1 and No 2 e Repeat repetition count The repetition count in a seq tab is not indicated e Trigger Not indicated e Sequences to run Sequences are run in the order of Seq 5 and Seq 6 e Seq end Output off Not indicated e Adrs Not indicated e UT 230V e Freq Not indicated e Repeat all The number of repetitions for sequences t
23. On To 200 Expand Contraci Seq7_ Seq8 Seqo sell 10 61000 4 11 UTx 110 0 0 100 himis 0 100 v himis 6100 4 27 61dh0 4 28 DC 6100 4 29 6 Repetition count 1 Re Delta2 Re Deltas Re 13 Repeat all 11 UT 7 Trigger 9 Seq end Ouitput off JSV 0 004 Fig 5 2 Interval Setting Window Set the following items Select a value provided in the dropdown list or enter a value within the input range into the Edit box After entering a value press the Enter key to accept it The following items other than 2 3 4 and 5 are the same as for Voltage fluctuation Please refer to 5 2 1 Voltage Fluctuation 5 6 IEC 61000 4 14 SD003 PCR LA 1 Seq tabs 2 Test condition type Select Interval from the following e Voltage fluctuation e Interval 3 UT Interval1 test level 1 Enter the center value of Voltage fluctuation This should be a ratio to the rated voltage of the EUT For most cases set it to the value of the UT of the previous seq tab in running sequences Select one of 90 100 and 110 or enter a value ranging from Input range 1 0 to 150 0 Z CAUTION Check the set value of UT taken between the seq tabs used to run a test If the UT set value differs between the seq tabs unintended voltage change may occur depending on selection of seq tabs whose sequences are run 4 UT Interval2
24. Serial instruments KI IS4 automatically searches for ASRL Serial I O Resources when FindRsrc method is invoked You do not have to configure the resources manually This option enables multiple sessions for the same ASRL resource to be opened from different processes Using this option makes I O operations bit slower Enable VISA C API Cancel 2 Check the Use ASRL Serial instruments checkbox in the ASRL panel 3 Click the Enable VISA C API button 4 Follow the prompts displayed in the window The KI VISA Setup window shown in step 2 is displayed once again 5 Click the OK button This completes the setup process Overall Picture of Test Sequences 2 7 E Measure 2 Using the GPIB interface for KI VISA When using National Instruments GPIB or Agilent GPIB use the VISA driver from the respective manufacturers Users using the CONTEC or Interface GPIB should use the KI VISA 1 Click the Windows Start button point to Program Kikusui IO Software and VISA then click on IO Config E When using the CONTEC GPIB board 2 Click on the GPIB CONTEC tab KI VISA COM I O Confie useTmc ASRL c pte ut 488 2m I Use GPIB instruments CONTEC 4PI GPIB ver4 01 required GPIB Board Index r Model 5 3 Check the Use GPIB instruments CONTEC API GPIB Ver 4 01 required checkbox 4 Enter for example GPIBO in the Index box and GP IB PCI F in the Model box Check the Use this board checkb
25. drivers may prevent any of the VISA drivers from operating properly VISA VISA Virtual Instrument Software Architecture is a standard specification for instrument connecting software established by the VXIplug amp play Systems Alliance VISA driver Driver software implemented in accordance with VISA specifications KI VISA VISA driver originally developed by Kikusui Electronics in compliance with VXIplug amp play VISA specification 2 2 Table 1 1 Required VISA Drivers and How to Obtain Them I O interface VISA driver required How to Obtain RS 232C Program CD ROM or 3 2 Interface GPIB KI VISA Ver 2 2 x or later KIKUSUI website CONTEC GPIB Continues on the next page SD003 PCR LA National Instruments NLVISA Ver 2 6 or later CD ROM supplied with GPIB or National Instru 4 GPIB ments website ae ae CD ROM supplied with Agilent GPIB HERETO Tr rarios KOLUR AE ar Agilent Technolo or later gies website 1 For use with the RS 232C interface there is no need to install KI VISA if a VISA driver from another manufacturer is already installed 2 The RS 232C interface supports a standard PC communications port provided and virtual serial communications port via USB RS 232C converters However for virtual communications port provided by a converter or similar configuration the VISA driver may not run properly depending on the specific device driver pro
26. either 40 or 70 ITO1 PCR L 5 Change angle starting phase angle Enter the starting phase angle of Voltage dips or Short interruptions Select one of 0 45 90 135 180 225 270 and 315 or enter Input range a value ranging from 0 to 359 6 Dips Short duration Enter the duration of Voltage dips or Short interruptions Enter the value in seconds or in hours minutes seconds format The number of periods corresponding to time entered is then indicated Select one of 0 01 s 0 02 s 0 1 s 0 2 s 0 5 s 0 6 s and 1 s or enter a Input range value ranging from 0 001 seconds to 999 hours 999 minutes 999 999 seconds 3 4 IEC 61000 4 11 SD003 PCR LA The ITO1 PCR L allows the setting of a 0 5 period or value ranging Input range from 1 to 3000 periods Since this item is set as a time value you with must convert the number of periods into a time value to enter ITO1 PCR L 50 Hz 0 010 s to 60 000 s 60 Hz 0 008 s to 50 000 s NOTE e How to enter in hours minutes seconds format A time value of up to 999 999 999 999 may be set by delimiting each unit with a colon If only one colon is used the value is treated as if in minutes seconds format Without a colon it is regarded as a second value Second values can be entered up to the third decimal place 7 Interval repetition interval Enter the interval between the poi
27. from 0 020 seconds to 999 hours Input range 999 minutes 999 999 seconds How to enter in hours minutes seconds format A time value of up to 999 999 999 999 may be set by delimiting each unit with a colon If only one colon is used the value is treated as if in minutes seconds format Without a colon it is regarded as a second value Second values can be entered up to the third decimal place Waveform switching timing is at the O phase angle of the fundamental wave That is the current waveform is not switched at the end of the set time but at the 0 phase angle of the next fundamental wave Interval repetition interval Enter the interval between the point at which the return voltage UT is reached and the next clipped voltage At the end of repetition intervals this is the time taken to move to the next seq tab Enter the value in seconds or in hours minutes seconds format For most cases select 10 000 s Alternatively enter a value ranging from Input range 0 020 seconds to 999 hours 999 minutes 999 999 seconds SD003 PCR LA NOTE NOTE SD003 PCR LA e Waveform switching timing is at the 0 phase angle of the fundamental wave That is the current waveform is not switched at the end of the set time but at the 0 phase angle of the next fundamental wave 6 Repeat repetition count Enter the number of repetitions for the test conditions in the seq tab in question
28. place 3 tp1 transition time Enter the transition time Enter the value in seconds or in hours minutes seconds format Select either 1 s or 10 s or enter a value ranging from 0 001 seconds to Input range 999 hours 999 minutes 999 999 seconds 4 f1 Af variation ratio and duration Enter a variation ratio Select one of 15 6 3 3 4 and 15 or enter a value Maputa ranging from 50 0 to 50 0 Enter the duration of f1 Af Enter the value in seconds or in hours minutes seconds format For most cases select 120 s Alternatively enter a value ranging from Input range 0 001 seconds to 999 hours 999 minutes 999 999 seconds 5 tp2 transition time Enter the return time Enter the value in seconds or in hours minutes seconds format Select either 1 s or 10 s or enter a value ranging from 0 001 seconds to Input range 999 hours 999 minutes 999 999 seconds 6 f2 2nd frequency and duration Enter the power frequency of the EUT Select either 50 Hz or 60 Hz or enter a value ranging from 1 Hz to Input range 999 99 Hz 7 4 IEC 61000 4 28 SD003 PCR LA Enter the duration of f2 Enter the value in seconds or in hours minutes seconds format When the test sequence moves to the next tab it takes approx 1 3 seconds for its preparation For most cases select 60 s Alternatively enter a value ranging from Input
29. procedure SD003 PCR LA This Manual is made up of the following sections Preface Provides an outline of the product and the required operating conditions Chapter1 Setup Describes how to install the Quick Immunity Sequencer application software Chapter 2 Overall Picture of Test Sequences Describes how to launch Quick Immunity Sequencer and provides an overall picture of test sequences Chapter3 IEC 61000 4 11 Describes Voltage dip Short interruption and Voltage variation immunity tests according to the IEC 61000 4 11 standard Chapter 4 IEC 61000 4 13 Describes Harmonics and interharmonics immunity tests according to the IEC 61000 4 13 standard Chapter5 IEC 61000 4 14 Describes the Voltage fluctuation immunity test according to the IEC 61000 4 14 standard Chapter6 IEC 61000 4 27 Describes the Unbalance immunity test according to the IEC 61000 4 27 standard Chapter 7 IEC 61000 4 28 Describes the Power frequency variation immunity test according to the IEC 61000 4 28 standard Chapter 8 IEC 61000 4 29 Describes Voltage dip Short interruption and Voltage variation immunity tests for DC power inputs according to the IEC 61000 4 29 standard Appendix Guides users to the appropriate standard depending on the nature of the testing desired Il Arrangement of this manual SD003 PCR LA Preface P 1 About This Manual This Manual describes the Quick Immunity Sequencer application software for the AC Power S
30. range 0 001 seconds to 999 hours 999 minutes 999 999 seconds 7 Repeat repetition count Enter the number of repetitions for the test conditions in the seq tab in question Select one of 1 3 5 and 10 times or enter a value ranging from 1 to Input range 9999 8 Trigger Select trigger signal output The red symbol W_ indicates a location selected using the checkbox 9 Sequences to run Check the corresponding checkboxes of the seq tabs whose test conditions are used to run a test sequence Test sequences are run in turn starting at Seq 1 Seq tabs whose corresponding checkboxes have not been checked are not used to run a test 10 Seq end Output off When this checkbox is checked a test signal is turned off at the end of sequence running the AC Power Supply PCR LA output is turned off 11 UT Enter the rated voltage of the EUT Select one of 100 V 110 V 120 V 200 V 220 V 230 V and 240 V or Input range enter a value ranging from 0 0 V to 305 0 V NOTE e An appropriate voltage range must be selected Use the toolbar or the Instrument Menu to set the voltage range For more information please refer to E To 100 V 200 V button in 2 7 2 Toolbar 12 Repeat all Enter the repetition count of all the selected seq tabs Seq 1 to Seq 10 Select one of 1 3 5 and 10 times or enter a value ranging from to Input range 9999 SD003 PC
31. run in turn starting at Seq 1 Seq tabs whose corresponding checkboxes have not been checked are not used to run a test SD003 PCR LA IEC 61000 4 11 3 5 11 Seq end Output off When this checkbox is checked a test signal is turned off at the end of sequence running the AC Power Supply PCR LA output is turned off 12 Freq Enter the rated frequency of the EUT Select either 50 Hz or 60 Hz or enter a value ranging from 1 Hz to Input range 999 90 Hz Input range with Select either 50 Hz or 60 Hz ITO1 PCR L 13 Repeat all Enter the repetition count of all the selected seq tabs Seq 1 to Seq 10 Select one of 1 3 5 and 10 times or enter a value ranging from to Input range 9999 3 6 IEC 61000 4 11 SD003 PCR LA 3 2 2 Voltage Variations 2 Test type 11 Sequences to run 1 Seq tab 3 UT 4 Test level 8 Repetition interval COR f amp Testl qco Quick Im unity Seque ncer Eile Edit Instrument Optio E 11 9 Gy Mid g Open Cond f Stop Pause Out Off Expand pntract Abort eel d heg 1 Seq2 W5e93_ Seg4 Segi Seq Seg7 Seg8 sedh sea 10 Seq X Seq 3 voltage dips 61000 4 11 Fi Seq 4 O Short interruptions Ll5eg5 Voltage variations Ne d Seq 6 61000 4 13 UT o Seq _ seq8 seq9 wf Ye 6100 4 27 req P N a 6100 4 28 DC 7 6100 4 29 Sle 6 Reduced 13 Freq 9 Repetition count voltage time 7 Vol
32. specified are selected in the running sequence the number of sequence addresses will exceed 100 If this Adrs indicator bar is displayed in red the number of the sequence addresses has exceeded 100 disabling transmission of the test condition data In this case reduce the number of the seq tabs selected for sequences to run until the indicator bar changes to green so that the number of sequence addresses becomes 100 or less 11 UT Enter the rated voltage of the EUT Select one of 100 V 110 V 120 V 200 V 220 V 230 V and 240 V or Input range enter a value ranging from 0 0 V to 305 0 V e An appropriate voltage range must be selected Use the toolbar or the Instrument Menu to set the voltage range For more information please refer to E To 100 V 200 V button in 2 7 2 Toolbar 12 Freq Enter the rated frequency of the EUT Select either 50 Hz or 60 Hz or enter a value ranging from 1 Hz to Input range 999 90 Hz IEC 61000 4 14 5 5 13 Repeat all Enter the repetition count of all the selected seq tabs Seq 1 to Seq 10 Select one of 1 3 5 and 10 times or enter a value ranging from to Input range 9999 5 2 2 Interval 10 Aars 2 Test condition type 8 Sequences to run 1 Seq tabs 3 Test level 1 3 Test level 2 5 Repetition interval T Test1 qco Quick In munity Sequence DER File Edit Instrument Opti n Help i oy 100 g W Open a ae
33. the PCR LA must be set as follows in advance Model Baud rate Parity Data length Stop bit Panel display PCR LA 19200bps None 8 2 0822 PCR L 9 600bps None 8 2 0834 3 When using the IT01 PCR L immunity tester check the Use IT01 PCR L checkbox Select or enter DEVx into the ITO1 Device Name dropdown list box The ITO1 PCR L immunity tester supports only the GPIB interface If the ITO1 PCR L is also used the AC Power Supply PCR LA is controlled as well via the GPIB The GPIB interface cannot be used with the RS 232C INTERFACE E Occurrence of a communication error e If the As an interface error has occurred Quick Immunity Sequencer is terminated Turn off the power switch of the PCR L A message appears the cable to the ITO1 PCR L may not be connected or the ITO1 PCR L s GPIB address DEV x may be incorrect Make sure the GPIB cable is connected properly 4 For three phase operations check the 3 phase system checkbox If Use ITO1 PCR L checkbox is checked uncheck it e The ITO1 PCR L immunity tester cannot be used in three phase operations with Quick Immunity Sequencer Overall Picture of Test Sequences 2 5 2 5 Troubleshooting at Startup E it takes a long time to display the startup window If the PCR LA series device driver is used via the RS 232C interface the timeout is set to 30 seconds This is because it takes around 30 seconds to detect a communication error caused by an incorrect device name or communi
34. 00 4 28 DC TERT N N 6100 4 29 xs mm E ne 13 Repeat all 11 UT 10 Seq end Output off E Fig 7 1 Power Frequency Variation Setting Window Set the following items Select a value provided in the dropdown list or enter a value within the input range into the Edit box After entering a value press the Enter key to accept it 1 Seq tabs Allows you to set the test conditions in the panel of each seq tab Seq 1 to Seq 10 These tabs show the heading numbers of the test conditions to be set in the following Test conditions can be copied or pasted between seq tabs For more information please refer to Edit Menu in 2 7 1 Menus SD003 PCR LA IEC 61000 4 28 7 3 2 f1 1st frequency and duration Enter the power frequency of the EUT Select either 50 Hz or 60 Hz or enter a value ranging from 1 Hz to Input range 999 90 Hz Enter the duration of f1 Enter the value in seconds or in hours minutes seconds format For most cases select 60 s Alternatively enter a value ranging from Input range 0 001 seconds to 999 hours 999 minutes 999 999 seconds NOTE e How to enter in hours minutes seconds format A time value of up to 999 999 999 999 may be set by delimiting each unit with a colon If only one colon is used the value is treated as if in minutes seconds format Without a colon it is regarded as a second value Second values can be entered up to the third decimal
35. 3 Flat x UT x sqrt2 test level 0 95 4 Duration 120 000 seconds 5 Interval repetition interval 10 000 seconds No 2 Over swing 1 Seq tab The Seq 2 tab is used 2 Test type Over swing 3 Harm test level 3rd and 5th orders 3 Level test level 3rd 1 0 Sth 1 0 3 Angle test level 3rd 0 Sth 0 4 Duration 1 000 second 5 Interval repetition interval 10 000 seconds Items common to No 1 and No 2 Repeat repetition count The repetition count in a seq tab is not indicated Trigger Not indicated Sequences to run Sequences are run in the order of Seq 1 and Seq 2 Seq end Output off Not indicated Bank Not indicated UT 230 V Freq Not indicated Repeat all The number of repetitions for the sequences run in the order of Seq 1 and Seq 2 is 1 IEC 61000 4 13 4 15 Frequency Sweep Table 4 2 Example of Indication of Running Results Frequency Sweep No Roe Seq Item UT Flatt Harm Level Angle pute Interval all tion 1 1 3 Sweep frequency 230 0 2 1 0 1 000 1 1 3 Sweep frequency 230 0 6 1 0 1 000 1 1 3 Sweep frequency 230 0 8 1 0 1 000 4 16 IEC 61000 4 13 E Description of Table 4 2 Number in the following items corresponds to the numbered steps in Fig 4 3 No 1 1st row Frequency sweep e 1 Seq tab The Seq 3 tab is used e 2 Test type Sweep in frequency e 3 Harm har
36. 35 1 60 6 6 IEC 61000 4 27 E Description of Table 6 1 Number in the following items corresponds to the numbered steps in Fig 6 1 No 1 1st row Unbalance 1 2 3 4 5 6 7 8 Seq tab The Seq 1 tab is used Ua 100 0 Ub 95 2 Uc 90 0 Ub Angle 125 Uc Angle 240 Duration 30 000 seconds Interval repetition interval 60 000 seconds a_i a i i No 2 2nd row Unbalance 1 Seq tab The Seq 2 tab is used 2 Ua 100 0 3 Ub 90 0 4 Uc 80 0 5 Ub Angle 131 6 Uc Angle 239 7 Duration 15 000 seconds 8 Interval repetition interval 60 000 seconds No 3 3rd row Unbalance 1 2 3 4 5 6 7 8 Seq tab The Seq 3 tab is used Ua 110 0 Ub 66 0 Uc 71 0 Ub Angle 139 Uc Angle 235 Duration 1 000 second Interval repetition interval 60 000 seconds a a a a i SD003 PCR LA Items common to No 1 No 2 and No 3 e Repeat repetition count The repetition count in a seq tab is not indicated e Sequences to run Sequences are run in the order of Seq 1 Seq 2 and Seq 3 e Seq end Output off Not indicated e UT 230V e Freq 50 0 Hz e Repeat all The number of repetitions for the sequences to be run in the order of Seq 1 Seq 2 and Seq 3 is 1 SD003 PCR LA IEC 61000 4 27 6 7 6 8 IEC 61000 4 27 SD003 PCR LA E Chapter 7 IEC 61000 4 28 Describes the Power frequency variation immu
37. K INIT IN SEQ TRIG OUT and SEQ STAT OUT terminals BNC connectors is shared in the AC Power Supply PCR LA and is not isolated The BNC connectors are also not isolated from the SLOTs internal circuit Note the slight time lag approx 100 us between the trigger signal output and actual output change Overall Picture of Test Sequences 2 19 Upper part of the rear of AC Power Supply the AC Power Supply PCR LA PCR LA s internal circuit SEG TRIG OUT 5V T 5Va o V PEAK IN 1k Max SEQ TRIG OUT SEQ TRIG OUT terminal ky A Z A SLOT 4 SLOT 3 SLOT 2 GPIB SLOT 1 Fig 2 5 SEQ TRIG OUT Terminal and Internal Circuit E IT01 PCR L trigger signal output With the ITO1 PCR L immunity tester a trigger signal is output from the TRIG OUT terminal BNC connector at the front of the IT01 PCR L immunity tester For more information please refer to the ITO1 PCR L Immunity Tester Operation Manual 2 20 Overall Picture of Test Sequences SD003 PCR LA Cn Chapter 3 IEC 61000 4 11 Describes Voltage dip Short interruption and Voltage variation immunity tests according to the IEC 61000 4 11 standard SD003 PCR LA IEC 61000 4 11 3 1 3 1 Outline of the Test NOTE 3 2 IEC 61000 4 11 The Voltage dip Short interruption and Voltage variation immunity tests determine the equipment s immunity against an abrupt drop or gentle decline in supply voltage E Items not conforming to the standard s requirements
38. MENT 1 Grant of License KIKUSUI Electronics Corp hereinafter referred to as KIKUSUI grants to you the following rights concerning this software which can be downloaded on the condition that you agree to the terms of this Agreement and related written materials collectively referred to as the SOFTWARE a To use of the SOFTWARE for the purpose of using our products that accept the terms in the license agreement OI do not accept the terms in the license agreement InstallShield ja Kikusui Quick Immunity Sequencer BETA 2004 01 22 InstallShi J Setup Type Choose the setup type that best suits your needs SEEGERS TONNE CORE Please select a setup type Complete All program Features will be installed Requires the most disk space Choose which program features you want installed and where they will be installed Recommended for advanced users InstallShield ia Kikusui Quick Immunity Sequencer BETA 2004 01 22 InstallShi Eg InstallShield Wizard Completed KIKUSUI ELECIRO N ICS The InstallShield Wizard has successfully installed Kikusui Quick Immunity Sequencer BETA_2004_01_22 Click Finish to exit Test amp Measurement the wizard Instruments an Power Supplies Shortcut Kikusui Quick Immunity Se SD003 PCR LA Setup 1 7 Click the Finish button This completes installation of Quick Immunity Sequencer When installation is complete a Quick Immunity Sequencer shortcut
39. National Instruments Agilent Technologies CONTEC or Interface e Please disable any PC system features such as Advanced Power Management APM or Suspend modes Interrupts known as SMIs that are periodically issued to the CPU in these modes will prevent normal operation of the product Personal computer CPU Pentium 233 MHz or better OS Windows 98 SE Windows Me Windows 2000 or Windows XP Memory Minimum of 128 MB HDD Free space of at least 100 MB is recommended for installation and additional free space of 10 GB or more for data CD ROM drive Required for product installation Mouse Required Display 1024 x 768 or larger Color 16 bit High Color or better SD003 PCR LA P 4 Required Hardware SD003 PCR LA AC power supplies PCR LA and PCR L series of the following versions PCR LA series Ver 3 10 or later PCR L series Ver 2 04 or later IT01 PCR L immunity tester The ITO1 PCR L immunity tester is recommended for IEC 61000 4 11 Voltage dips Short interruptions and Voltage variations in single phase The ITO1 PCR L immunity tester enables the following e The standard s performance specifications relating to voltage rise and fall time of a test voltage generator 1 us to 5 us are met The performance achieved without the ITO1 PCR L is approximately 30 us e The standard s performance specifications relating to the starting phase angle of Voltage dips or Short interrup
40. On Off button When output is off clicking the Out On button turns output on When the output is on the button indication changes to Out Off The initial voltage of the valid seq tab is output enabling a test based on the output ON state H To 100 V 200 V button This is a voltage range selector button If the 100 V range is active it displays To 200 V If the 200 V range is active it displays To 100 V In the 100 V range output of up to 152 5 V is available in the 200 V range output of up to 305 0 V is available Note that in the 200 V range the maximum output current is half that of the 100 V range e For IEC 61000 4 29 Voltage dips Short interruptions and Voltage variations for DC power inputs DC voltage applies In the 100 V range output of up to 215 5 V is available in the 200 V range output of up to 431 0 V is available E Expand button Horizontally expands the preview waveform in the middle of the window This enlarges the beginning of the waveform for a valid seq tab e Quick Immunity Sequencer was designed so that the preview waveform and actual output will generally be in agreement However in certain rare cases the preview waveform and the actual output will be off by a single period due to an error in zero cross detection 2 16 Overall Picture of Test Sequences SD003 PCR LA e The time axis direction of the preview waveform area is finite and extends approx 7 times the window size In certain case
41. Part No Z1 003 152 IB006433 Nov 2006 USER S MANUAL PCR LA Series Application Software SD003 PCR LA E Test1 qco Quick Immunity Sequencer File Edit Instrument Option Help 5 E 1 an MM Open Cond Stop Pause Out Off Expand Contract Abort I Seq 1 Segi Seg2 5e93_ Seq4 Seq5 Seg6 Seq7_ Seq8 Seq9_ Seq 10 X Seq2 Seq3 O Voltage dips 61000 4 11 Seq4 hort interruptions e Seq5 Voltage variations ae Seq6 Seq 61000 4 13 Seq8 ur n 5eq9 100 0 v vrms h Seq 10 Seq end 61 000 4 14 C Output off Repeat T e Decreasing Reduced Increasing Interval r bes 1000 427d Frag 50 00 v H2 trigger 0 100 lo100 0100 o100 v Repeat alll A him s him s him s him s 61000 4 29 m nns AA gt No Repeat all Seq Item Level Angle Decrea Reduced Increasing Interval Variation 100 0 40 0 100 0 100 0 100 0 100 Repeat All1 Seq Repeat 0 0 00 00 896V KIKUSUI Use of This Manual Please read through and understand this User s Manual before operating the product After reading always keep the manual nearby so that you may refer to it as needed When moving the product to another loca tion be sure to bring the manual as well If you find any incorrectly arranged or missing pages in this manual they will be replaced If the manual it gets lo
42. R LA IEC 61000 4 28 7 5 7 3 Results of Sequence Running The running results are indicated every time the sequence for one seq tab finishes The starting date and time are shown in the row above the No 1 row Shown below is a typical test example An indicated value becomes a blank or 0 if it is not related to the test The Repeat set value in the panel for each seq tab is not indicated Table 7 1 Example of Indication of Running Results Repeat all Item ur gp Pte A gime PL Pua pmo 2 i f2 time taf time tion time Variation of frequency 230 0 50 00 3 0 60 000 10 000 120 000 50 00 10 000 60 000 Variation of 230 0 50 00 3 0 60 000 10 000 120 000 50 00 10 000 60 000 frequency 7 6 IEC 61000 4 28 E Description of Table 7 1 Number in the following items corresponds to the numbered steps in Fig 7 1 No 1 1st row Power frequency variations 1 Seq tab The Seq 1 tab is used 2 f1 50 00 Hz 2 f1 time 60 000 seconds 3 tp1 time 10 000 seconds 4 tpl Deltaf 3 0 4 Duration 120 000 seconds 5 tp2 time 10 000 seconds 6 2 50 00 Hz 6 f2 time 60 000 seconds No 2 2nd row Power frequency variations 1 Seq tab The Seq 2 tab is used 2 f1 50 00 Hz 2 f1 time 60 000 seconds 3 tp1 time 10 000 seconds 4 tpl Deltaf 3 0 4 Duration 120 000 seconds 5 tp2 time 10 000 se
43. Send button is not available unless output is turned off using the Out Off button because the waveform banks have been overwritten E Run button Starts the test This button is enabled if the set values are sent to the PCR LA using the Send button You can also turn on output beforehand using the Out On button then start a test using the Run button Overall Picture of Test Sequences 2 15 A CAUTION NOTE E Stop button Stops the test mid course Note that output remains enabled If you also wish to switch off output click the Abort button instead This button is enabled only while a test is running E Pause button This button pauses a test immediately before moving to the next tab after completing a sequence in a given seq tab This function can be used for example to check or replace the EUT e Before connecting or disconnecting the EUT always confirm that the AC Power Supply PCR LA output is OFF If you click the Pause button while a test is running the button indication changes to Cancel until the sequence of the seq tab running at that time is complete Checking the Cancel button cancels the pause waiting status and returns the button displayed to the original Pause button When the sequence of the relevant seq tab is finished and the test is paused the button indication changes to Continue enabling the Out On Off button Click the Continue button to resume running the sequence E Out
44. Voltage dip 3 2 Voltage Dips 3 3 Voltage fluctuation immunity test 5 2 Voltage Variations 3 7 W waveform banks 4 5 4 10 Waveform preview 2 11 l 2 Index SD003 PCR LA SD003 PCR LA QUICK IMMUNITY SEQUENCER Application Software User s Manual
45. abs 4 Uc File Edit Instrument Opti Help 6 Phase angle Ua Uc 7 Duration 5 Pip angle 8 Repetition interval Cdhtr t Abert 6 Seq Seq8 Seq 9 Hea 10 Ua 150 0 vii himis a lt uration interval himis E 125 ideg 240 v deg Ub 95 2 wi o 1 ii Open Cond d E ok o To 2004 M Seq 1 Seq 1 Seq2 Seg3 Seq4 e wi e eee 5eq 3 61000 4 11 Clseq4 s5eq5 ar _ Seq6 seq7 61000 4 13 seq8 seq9 h _ seq 10 eq end 61090 4 C output off wo UT 100 0 Repeat 1 GN 6100 4 27 req 50 00 z oF peat alll Uc 90 0 vi 6100 4 28 Dc 610 0 4 29 lt m aam 13 Freq 9 Repetition count 14 Repeat all 12 UT 11 Seq end Output off Fig 6 1 Unbalance Setting Window t SSS Uc UbA UcA Dura Inte os 0 004 Set the following items Select a value provided in the dropdown list or enter a value within the input range into the Edit box After entering a value press the Enter key to accept it 1 Seq tabs Allows you to set the test conditions in the panel of each seq tab Seq 1 to Seq 10 These tabs show the heading numbers of the test conditions to be set in the following Test conditions can be copied or pasted between seq tabs For more information please refer to Edit Menu in 2 7 1 Menus SD003 PCR LA IEC 61000 4 27 6 3 2 Ua Enter a ratio of a quantity t
46. ary tests conducted before actual testing for the standard immunity allowance tests or stress tests E Items not conforming to the standard s requirements Certain Quick Immunity Sequencer items do not conform to the standard s requirements in the test configuration including the hardware For more information please refer to the chapter relating to the standard in question E Three phase operations The test conditions are all set in relation to phase U Test signals change simultaneously in three phases on the basis of phase U Note that for IEC 61000 4 27 unbalance although a test signal can be set on a phase basis there is a time lag in the test signal between phases For IEC 61000 4 29 Voltage dips Short interruptions and Voltage variations for DC power inputs there are no settings for the three phases NOTE e The ITO1 PCR L immunity tester cannot be used in three phase operations with Quick Immunity Sequencer 2 2 Overall Picture of Test Sequences SD003 PCR LA 2 2 Test Configuration Z CAUTION SD003 PCR LA This section provides a brief discussion of single phase and three phase operations of AC Power Supply PCR LA Series For more information on connecting please refer to the operation manuals for the following products e Before performing an immunity test using this application software carefully read the operation manuals for the following products and the specific hardware descriptions found the
47. cation cable connection or other errors Note that the initial setting provided on installation of Quick Immunity Sequencer is the RS 232C interface If a communication error is detected when you launch Quick Immunity Sequencer the I O Configuration dialog box please refer to 2 4 Setting up I O Configuration will appear Check the driver settings E Error message This application cannot be launched because VISA32 DLL cannot be found Reinstalling the application may solve this problem This may indicate that no VISA driver is installed Install an appropriate VISA driver For more information please refer to 1 2 Installation This may indicate that use of KI VISA as the VISA driver has not been set up To use the RS 232C interface set the driver according to the procedure given in Measure 1 below To use the GPIB interface set according to the procedure given in Measure 2 below For more information please refer to 2 5 1 Setting up KI VISA E Error message Runtime error XXX at XXXXXXXX The PCR LA series device driver may be installed incorrectly Re install Quick Immunity Sequencer or download and install the device driver for the PCR LA series from the download page at http www kikusui co jp E Error message Unable to open instrument in device DEVx This may indicate that no GPIB driver has been installed or that the GPIB driver has not been correctly set Check the GPIB driver setting E Error mes
48. ck the 3 phase system checkbox In this case as ITO1 PCR L cannot be used uncheck the Use ITO1 PCR L checkbox 2 14 Overall Picture of Test Sequences SD003 PCR LA Option Menu E Default voltage and frequency Lets you select the voltage and frequency defaults for the set values entered when creating a new file The settings configured for the Quick Immunity Sequencer on shipment from the factory are 100 V 50 Hz E Color Lets you modify the color for background progress bars characters and preview waveforms for each seq tab Help Menu E Quick Immunity Sequencer Help Opens the Help file E About Quick Immunity Sequencer Displays the Quick Immunity Sequencer version information 2 2 Toolbar l gt ow Open Cond SD003 PCR LA ad ai L 8 380 a M 7 Send Run Stop Pause Out On To 2004 Expand Contract Abort E Open button Opens the dialog box used to select a test condition file This button has the same function as Open on the File Menu E Cond button Saves the test condition settings under the same name as the currently open test condition file replacing the earlier version This button has the same function as Save Test Conditions on the File Menu E Send button After setting the test conditions click the Send button to send set data to the PCR LA When the data has been sent the Run button is enabled and no longer grayed out For IEC 61000 4 13 harmonics immunity test the
49. conds 6 2 50 00 Hz 6 f2 time 60 000 seconds Items common to No 1 and No 2 Repeat repetition count The repetition count in a seq tab is not indicated Trigger Not indicated Sequences to run Sequences are run in the order of Seq 1 and Seq 2 Seq end Output off Not indicated UT 230 V Repeat all The number of repetitions for the sequences to be run in the order of Seq 1 and Seq 2 is 1 SD003 PCR LA Ce Chapter 8 IEC 61000 4 29 Describes Voltage dip Short interruption and Voltage variation immunity tests for DC power inputs according to the IEC 61000 4 29 standard SD003 PCR LA IEC 61000 4 29 8 1 8 1 Outline of the Test The Voltage dip Short interruption and Voltage variation immunity tests for DC power inputs determine the equipment s immunity against an abrupt drop or gradual decline in the supply voltage 8 2 Preparing the Test Conditions 8 2 1 DC Voltage Dips and Short Interruptions 2 Test type 4 Test level 1 Seq tabs 3 UT 6 Repetition interval DOR 9 Sequences to run Ea Test1 qco Quick Im munity Sequer cer File Edit Instrument Opti a a Ii 3 100 p Wid 7 Open Cond z F Out On To 200 Exgipnd Contract Abort Bom seq 1 Sq1_ seq2 seq3 seq segl sege Seq Sells Seg9 Seq 10 Seq 2 y C seq3 DC Voltage dips L__61000 4 11 J _ seq 4 O DC Short interruptions seq5 O DC voltage variations ae 5
50. dard a transition between seq tabs will take several seconds For more information please refer to Interval repetition interval in 6 2 Preparing the Test Conditions 2 8 1 Turning Power on off in the Test Configuration A CAUTION Follow the procedure given below to turn power on off Departing from this SD003 PCR LA procedure may generate an error or cause equipment device failure E Turning power on Turn on the IT01 PCR L immunity tester POWER switch 2 Turn onthe AC Power Supply PCR LA POWER switch z gt Need Launch Quick Immunity Sequencer Overall Picture of Test Sequences 2 17 E Turning power off 2 3 Exit Quick Immunity Sequencer Turn off the AC Power Supply PCR LA POWER switch Turn off the IT01 PCR L immunity tester POWER switch 2 8 2 Running Procedure A 5 Select single phase or three phase Click on I O Configuration from the Instrument Menu and check the 3 phase system checkbox in the dialog displayed Set the test conditions Send the set test conditions to the PCR LA Click the Send button on the toolbar When transmission is completed the Run button on the toolbar is enabled Run the sequences Click the Run button on the toolbar While the sequences are running the Sequence Running indication appears in a progress bar that graphically represents the status of the sequence in question When execution of the set test conditions has been completed the test auto
51. does not specify an electromagnetic environment class or assume it to be a setting condition An electromagnetic environment class for test conditions should be set at the user s discretion E Items not conforming to the standard s requirements Certain Quick Immunity Sequencer items do not conform to the standard s requirements in the test configuration including the hardware The following table shows these items and the performance achieved in the test configuration Performance achieved 4 2 IEC 61000 4 13 Item Standard s requirements in the test configuration Frequency sweep The frequency step of a sweep is 0 1 x fl minimum The frequency step of a sweep is x f1 minimum integral multiple Interharmonics Meister curve Required Not provided E Three phase operations The test conditions are all set in relation to phase U Test signals change simultaneously in three phases on the basis of phase U SD003 PCR LA 4 2 4 2 1 Preparing the Test Conditions Combined Harmonics Flat Curve 10 Bank 2 Test type 8 Sequences to run 1 Seq tabs 3 Test level 5 Repetition interval Ea Test1 qco Quick Imm inity Sequencer DER File Edit Instrument Optio p 5 2 100 a Open Cond d Out On To 200 61000 4 11 61000 4 13 610 4 0 200 himis 10 000 w Jims 11 UT 6 Repetition count 4 Duration H
52. e 2 6 Even harmonics 4 12 F File Menu 2 12 Flat Curve 4 3 H Harmonics and interharmonics immunity tests 4 2 Help Menu 2 15 High impedance 8 3 T O interface P 2 Individual Harmonics 4 12 Installation 1 2 Instrument Menu 2 13 interface error 2 7 ITO1 PCR L immunity tester P 3 Items not conforming to the standard s requirements 2 2 3 2 4 2 5 2 6 2 SD003 PCR LA K KI VISA 1 2 L Low impedance 8 3 O Occurrence of a communication error 2 5 Odd harmonics multiples of 3 4 12 Odd harmonics non multiples of 3 4 12 Option Menu 2 15 OS version P 2 Over Swing 4 7 Overall Picture of Test Sequences 2 10 P Power frequency variation immunity test 7 2 R Required Hardware P 3 Required Operating Conditions P 2 S Save Test Conditions 2 12 Save As Test Conditions 2 12 Save As Test Result 2 12 Selection of the standard 2 10 Setting up I O Configuation 2 5 Short Interruptions 3 3 Sweep in Frequency 4 9 T Test Configuration 2 3 Test Standard Selection Guide A 1 Three phase operations P 3 2 2 3 2 4 2 5 2 7 2 Toolbar 2 15 Trigger signal output 2 19 Troubleshooting at Startup 2 6 Troubleshooting During Running 2 19 Index l 1 Turning power off 2 18 Turning power on 2 17 U Unbalance immunity test 6 2 Using the product with the conventional PCR L series P 1 V Versions of the AC power supplies for which this manual is valid P 1 VISA 1 2 VISA driver P 2
53. e set time but at the 0 phase angle of the next fundamental wave 6 Repeat repetition count 7 Trigger Select trigger signal output A trigger signal is output at the start of the superimposed waveform output of harmonics selected using the checkboxes 8 Sequences to run 9 Seq end Output off 10 Bank For the Bank indication the fluctuating indicated values are displayed in green If this value is displayed in red the number of waveforms to be stored has exceeded 14 types In this case the Send button remains disabled to prevent further testing Paring down the number of waveforms stored e Method 1 Reduce the number of checked checkboxes in the seq tabs e Method 2 Set the Duration setting to 0 zero 11 UT 12 Freq 13 Repeat all IEC 61000 4 13 4 11 4 2 4 Individual Harmonics Three harmonics are given below as individual harmonics Each harmonic is set in the same way for all tests e Odd harmonics non multiples of 3 e Odd harmonics multiples of 3 e Even harmonics 12 Bank 2 Test type 3 Harmonic order 5 Angle 7 Repetition interval 1 Seq tabs 4 Level a Test1 qco Quick I munity 5equencer File Edit Instrument Opi n Helpi Ed wi 100 Willd Out On To 2004 ee Contract RA SAPS NETS ME AEN NEST Flat curve el Over swing EN REET 61000 4 13 10 Sequences to run 6 Duration 4 o EJ oa mi oO Sweep in frequ
54. each seq tab Seq 1 to Seq 10 These tabs show the heading numbers of the test conditions to be set in the following Test conditions can be copied or pasted between seq tabs For more information please refer to Edit Menu in 2 7 1 Menus SD003 PCR LA IEC 61000 4 11 3 3 2 Test type Select either Voltage dips Short interruptions or Voltage variation The diagram displayed on the right in the window changes depending on the test type button selected 3 UT Enter the initial voltage equal to voltage at return For most cases enter the rated voltage of the EUT Select one of 100 V 110 V 120 V 200 V 220 V 230 V and 240 V or Input range enter a value ranging from 0 0 V to 305 0 V NOTE e An appropriate voltage range must be selected Use the toolbar or the Instrument Menu to set the voltage range For more information please refer to E To 100 V 200 V button in 2 7 2 Toolbar A CAUTION Check the set value of UT taken between the seq tabs used to run a test If the UT set value differs between the seq tabs unintended voltage change may occur depending on selection of the seq tabs whose sequences are run 4 Level test level For Voltage dips enter a residual voltage in with respect to UT This level is set to 0 no input in the case of Short interruptions Select either 40 or 70 or enter a value ranging from 0 to Input range 200 Input range with Select
55. eated as if in minutes seconds format Without a colon it is regarded as a second value Second values can be entered up to the third decimal place 6 Interval repetition interval Enter the interval between the point at which the return voltage UT is reached and the next voltage drop At the end of repetition intervals this is the time taken to move to the next seq tab Enter the value in seconds or in hours minutes seconds format For most cases select 10 000 s Alternatively enter a value ranging from Input range 0 001 seconds to 999 hours 999 minutes 999 999 seconds 7 Repeat repetition count Enter the number of repetitions for the test conditions in the seq tab in question Select one of 1 3 5 and 10 times or enter a value ranging from 1 to Input range 9999 8 Trigger Select trigger signal output The red symbol W indicates a location selected using the checkbox 9 Sequences to run Check the corresponding checkboxes of the seq tabs whose test conditions are used to run a test sequence Test sequences are run in turn starting at Seq 1 Seq tabs whose corresponding checkboxes have not been checked are not used to run a test 10 Seq end Output off When this checkbox is checked a test signal is turned off at the end of sequence running the AC Power Supply PCR LA output is turned off 11 Repeat all Enter the repetition count of all the selected seq tabs Seq 1 to Seq 10
56. ency Odd non multiple of 3 Odd multiple of 3 Even harmonics __ wl nN oO RAR AAA a Angle Duration Interval 9 Trigger 15 Repeat all 8 Repetition count 05V 000A 11 Seq end Output off 13 UT Fig 4 5 Individual Harmonics Setting Window Odd Harmonics Non multiples of 3 The individual harmonics test switches the waveform banks total 14 to change harmonics stepwise The following items other than 2 3 4 5 6 7 and 9 are the same as for the flat curve Please refer to 4 2 1 Combined Harmonics Flat Curve 1 Seq tabs 2 Test type Select Odd non multiple of 3 Odd multiple of 3 or Even harmonics from the following 4 12 IEC 61000 4 13 SD003 PCR LA 3 4 5 6 NOTE 7 SD003 PCR LA e Flat curve e Over swing e Sweep in frequency e Odd non multiple of 3 e Odd multiple of 3 e Even harmonics Harm harmonic order Enter the order of the harmonic superimposed on the fundamental wave Changes in individual harmonics progress downward from the highest row Odd harmonics The values 5 5 7 7 11 13 17 19 23 25 29 31 35 and 37 are fixed non multiples and cannot be modified of 3 Odd The values 3 3 9 9 15 21 27 33 and 39 are fixed and cannot be harmonics multiples of 3 modified Even harmonics Enter an even number ranging from 2 to 40 Level Ent
57. eq6 seq7 61000 4 13 Cl5eq8 Cl 5eq9 Cl5eq 10 eq end j Cloutputoff ff nNOS bo i 6p D Eg 6109 0 4 27 61090 4 28 DC eM a gyo 4 29 8 Trigger 7 Repetition count 5 Duration ing Interval 11 Repeat all 10 Seq end Output off w ODoA Fig 8 1 DC Voltage Dips and Short Interruptions Setting Window Set the following items Select a value provided in the dropdown list or enter a value within the input range into the Edit box After entering a value press the Enter key to accept it 8 2 IEC 61000 4 29 SD003 PCR LA 1 Seq tabs Allows you to set the test conditions in the panel of each seq tab Seq 1 to Seq 10 These tabs show the heading numbers of the test conditions to be set in the following Test conditions can be copied or pasted between seq tabs For more information please refer to Edit Menu in 2 7 1 Menus 2 Test type Select either DC Voltage dips DC Short interruptions or DC Voltage variations as an immunity test The diagram displayed on the right in the window changes depending on the test type button selected 3 UT Enter the initial voltage equal to voltage at return Input range Select one of 100 V 110 V 120 V 200 V 220 V 230 V and 240 V or enter a value ranging from 0 0 V to 431 0 V NOTE e An appropriate voltage range must be selected Use the toolbar or the Instrument Me
58. er the ratio of the harmonic to the fundamental voltage superimposed on the fundamental wave Input range Enter a value ranging from 0 0 to 100 0 Angle phase angle Enter the phase angle of the harmonic superimposed on the fundamental wave Input range Enter a value ranging from 0 to 360 Duration Enter the duration of the harmonic superimposed on the fundamental wave Input range Enter a value ranging from 0 000 seconds to 999 hours 999 minutes 999 999 seconds e Waveform switching timing is at the 0 phase angle of the fundamental wave That is the current waveform is not switched at the end of the set time but at the 0 phase angle of the next fundamental wave Interval repetition interval Enter the rest time until the next harmonic is superimposed During this period only the fandamental wave is present Input range Enter a value ranging from 0 000 seconds to 999 hours 999 minutes 999 999 seconds IEC 61000 4 13 4 13 NOTE e Waveform switching timing is at the 0 phase angle of the fundamental wave That is the current waveform is not switched at the end of the set time but at the 0 phase angle of the next fundamental wave 8 Repeat repetition count 9 Trigger Select trigger signal output A trigger signal is output at the start of the superimposed waveform output of harmonics selected using the chec
59. erface Installation of KI VISA Installation of Quick Immunity Sequencer f4 Fal Ie B My Computer Ui Fig 1 1 Setup Start Window SD003 PCR LA Setup 1 3 2 Click on Installation of KI VISA The KI VISA Installation window is displayed i Kikusui KI VISA COM 2 2 5 beta 3 InstallShield Wizard Welcome to the InstallShield Wizard for Kikusui KI VISA COM 2 2 5 beta 3 KIKUSUI ELECT I P R ON cs The InstallShield R Wizard will install Kikusui KI YISA COM e 2 5 beta 3 on your computer To continue click Next 2 2 5 beta 3 ter T it lick Next Test amp Measurement WARNING This program is protected by copyright law and international treaties 3 InstallShield Wizard Completed KIKUSUI ELECTRONICS The InstallShield Wizard has successfully installed Kikusui CORP KI VISA COM 2 2 5 beta 3 Click Finish to exit the wizard Test amp Measurement Launch IO Config 4 Check the Launch IO Config checkbox 5 Click the Finish button This completes installation of the KI VISA driver KI VISA Setup window then launches 1 4 Setup SD003 PCR LA E When using the RS 232C Interface 6 Click on the ASRL tab KI VISA I O Config GPIB CONTEC GPIB NI 488 2M JV Use ASRL Serial instruments I Enable dynamic search for available serial ports JV Enable Multiple Session This option enables multiple sessions for the same ASRL resource to be opened from different processes Using this option makes I O operati
60. erval 3 UT test level Enter the center value of Voltage fluctuation This should be a ratio to the rated voltage of the EUT Select one of 90 100 and 110 or enter a value ranging from Input range 1 0 to 150 0 Z CAUTION Check the set value of UT taken between the seq tabs used to run a test If the UT set value differs between the seq tabs unintended voltage change may occur depending on selection of seq tabs whose sequences are run 4 Delta1 2 3 Voltage fluctuation Enter the Voltage fluctuation range This should be a ratio to the rated voltage of the EUT Select one of 12 8 8 and 12 or enter a value ranging from 50 to 50 Input range The duration of Voltage fluctuation is 2 seconds fixed 5 Repetition1 2 3 repetition interval Enter the repetition interval of Voltage fluctuation Enter the value in seconds or in hours minutes seconds format For most cases select 5 s Alternatively enter a value ranging from 3 Input range seconds to 999 hours 999 minutes 999 999 seconds NOTE e How to enter in hours minutes seconds format A time value of up to 999 999 999 999 may be set by delimiting each unit with a colon If only one colon is used the value is treated as if in minutes seconds format Without a colon it is regarded as a second value Second values can be entered up to the third decimal place e A transi
61. esponds to the color of the waveform displayed in the waveform preview To change colors use Color in the Option Menu Edit boxes The edit boxes used to enter a value are of the dropdown type Click the down arrow head control of an edit box to select a typical value To enter a value type the desired value directly into the box then accept the typed value by pressing the Enter key Edit boxes not in the panel area of a seq tab Seq 1 to Seq 10 are common to all seq tabs They are used to set voltage frequency or repetition count for all sequences The voltage or frequency edit box is displayed depending on the specific standard in question Checkboxes Seq 1 to Seq 10 Check the corresponding checkboxes of the seq tabs whose test conditions are used to run a test sequence Test sequences are run in turn starting at Seq 1 seq tabs whose corresponding checkboxes have not been checked are not used to run a test Checkboxes Trigger For trigger output selection the red symbol W indicates at a location selected using a checkbox Use these checkboxes when observing test signals on an oscilloscope For more information on trigger signals please refer to 2 9 Observing a Test Signal Checkbox Seq end Output off When this checkbox is checked a test signal is turned off at the end of sequence running the AC Power Supply PCR LA output is turned off 3 Waveform preview This part displays the waveform selected using seq
62. format For most cases select 180 s Alternatively enter a value ranging from 0 Input range seconds to 999 hours 999 minutes 1000 seconds 6 4 IEC 61000 4 27 SD003 PCR LA NOTE e How to enter in hours minutes seconds format A time value of up to 999 999 1000 may be set by delimiting each unit with a colon If only one colon is used the value is treated as if in minutes seconds format Without a colon it is regarded as a second value Second values after decimal places cannot be entered e A transition between the seq tabs takes several seconds Accordingly Interval time repetition time at the end requiring a transition between the seq tabs is equal to the set value plus several seconds 9 Repeat repetition count Enter the number of repetitions for the test conditions in the seq tab in question Select one of 1 3 5 and 10 times or enter a value ranging from to Input range 9999 10 Sequences to run Check the corresponding checkboxes of the seq tabs whose test conditions are used to run a test sequence Test sequences are run in turn starting at Seq 1 Seq tabs whose corresponding checkboxes have not been checked are not used to run a test 11 Seq end Output off When this checkbox is checked a test signal is turned off at the end of sequence running the AC Power Supply PCR LA output is turned off 12 UT Enter the rated voltage of the EUT Select one of 100 V
63. from Input range 0 001 seconds to 999 hours 999 minutes 999 999 seconds 8 Interval repetition interval 9 Repeat repetition count 10 Trigger 11 Sequences to run 12 Seq end Output off 13 Repeat all 8 6 IEC 61000 4 29 SD003 PCR LA 8 3 Results of Sequence Running The running results are indicated every time the sequence for one seq tab finishes The starting date and time are shown in the row above the No 1 row Shown below is a typical test example An indicated value becomes a blank or 0 if it is not related to the test The Repeat set value in the panel for each seq tab is not indicated Table 8 1 Example of Indication of Running Results No Repeat Seq Item UT Level Hopr Duration nee Interval all asing asing 1 1 1 Pe Voltan 230 0 85 2 000 1 000 2 000 10 000 variation 2 1 2 DC Voltage dips 230 0 70 0 100 10 000 3 1 3 DG ahont 230 0 70 0 100 10 000 interruption SD003 PCR LA E Description of Table 8 1 Number in the following items corresponds to the numbered steps in Figs 8 1 and 8 2 No 1 DC Voltage variations 40 1 Seq tab The Seq 1 tab is used 2 Test type DC Voltage variations 3 UT 230 V 4 Level test level 85 5 Decreasing time voltage decreasing time 2 000 seconds 6 Reduced reduced voltage time 1 000 second 7 Increasing time voltage increasing time 2 000 seconds
64. g from 0 Input range to 200 SD003 PCR LA IEC 61000 4 29 8 5 5 Decreasing voltage decreasing time Enter the time taken to reduce voltage from the initial voltage UT to the reduced voltage Enter the value in seconds or in hours minutes seconds format For most cases select 2 000 s Alternatively enter a value ranging from Input range 0 001 seconds to 999 hours 999 minutes 999 999 seconds NOTE Voltage rise and drop are approximated stepwise via the PCR LA sequence function The sequence function allows up to 100 steps to be set For example if there is only one valid seq tab and DC voltage variations have been set to it both voltage rise and drop are approximated in 48 steps If all seq tabs are selected made valid and configured for DC voltage variations both voltage rise and drop are approximated in three steps 6 Reduced reduced voltage time Enter the reduced voltage duration Enter the value in seconds or in hours minutes seconds format Select one of 0 01 s 0 03 s 0 1 s 0 3 s and 1 s or enter a value ranging Input range from 0 001 seconds to 999 hours 999 minutes 999 999 seconds 7 Increasing voltage increasing time Enter the time taken to return from the reduced voltage to the initial voltage UT Enter the value in seconds or in hours minutes seconds format For most cases select 10 000 s Alternatively enter a value ranging
65. is created on the desktop jo 1 8 Setup SD003 PCR LA Chapter 2 Overall Picture of Test Sequences Describes how to launch Quick Immunity Sequencer and provides an overall picture of test sequences SD003 PCR LA Overall Picture of Test Sequences 2 1 2 1 Outline of Immunity Test An immunity test generates a test signal using a voltage generator applying it to the equipment under test EUT EUT operations are classified into the following four performance levels in terms of functional loss or performance degradation in order to assess the test results The classification criteria are the performance levels specified by the agreement between the equipment manufacturer and a test requester or between the equipment manufacturer and a purchaser a Normal performance within the specification limits specified between the manufacturer and a test requester or purchaser b Temporary functional loss or performance degradation resolved after interference has been removed EUT automatically regains normal performance without operator intervention c Temporary functional loss or performance degradation that requires adjustment with operator intervention d Unrecoverable functional loss performance degradation or data loss caused by hardware failure or software corruption Quick Immunity Sequencer can be used for testing according to the standard Since it allows test conditions to be extended and set it can also be used for prelimin
66. kboxes 10 Sequences to run 11 Seq end Output off 12 Bank For the Bank indication the fluctuating indicated values are displayed in green If this value is displayed in red the number of waveforms to be stored has exceeded 14 types In this case the Send button remains disabled to prevent further testing Paring down the number of waveforms stored e Method 1 Reduce the number of checked checkboxes in the seq tabs e Method 2 Set the settings for Duration and Interval repetition interval to 0 zero 13 UT 14 Freq 15 Repeat all 4 14 IEC 61000 4 13 SD003 PCR LA 4 3 Results of Sequence Running The running results are indicated every time the sequence for one seq tab finishes The starting date and time are shown in the row above the No 1 row Shown below are some typical test examples An indicated value becomes a blank or 0 if it is not related to the test Combined Harmonics Table 4 1 Example of Indication of Running Results Flat Curve and Over Swing No Renee Seq Item UT Flatt Harm Level Angle Pus Interval all tion 1 1 1 Flat curve 230 0 0 95 120 000 10 000 2 1 2 Over swing 230 0 3 5 1 0 1 0 0 0 1 000 10 000 SD003 PCR LA E Description of Table 4 1 Number in the following items corresponds to the numbered steps in Figs 4 1 and 4 2 No 1 Flat curve 1 Seq tab The Seq 1 tab is used 2 Test type Flat curve
67. matically stops Stop There are two ways to stop a test the Stop button and the Pause button Stop button Stops the test mid course Note that output remains enabled If you also wish to switch off output click the Abort button instead Pause button This button pauses a test immediately before moving to the next tab after completing a sequence in a given seq tab This function can be used for example to check or replace the EUT If you click the Pause button while a test is running the button indication changes to Cancel until the sequence of the seq tab running at that time is complete Clicking the Cancel button cancels the pause waiting status and returns the button displayed to the original Pause button When the sequence of the relevant seq tab is finished and the test is paused the button indication changes to Continue Click the Continue button to resume running the sequence 2 18 Overall Picture of Test Sequences SD003 PCR LA 2 8 3 Troubleshooting During Running E The Run button remains disabled although the Send menu button was clicked With the ITO1 PCR L immunity tester you must turn on the PCR LA output before running a test Click the Out On button to turn on the PCR LA output E The Send button remains disabled although no test is being run If IEC 61000 4 13 has been selected you must turn off the PCR LA output This is because bank calculations were made while the test condition data was bei
68. monic order 2nd e 4 Level 1 0 e 5 Duration 1 000 second No 1 2nd row Frequency sweep e 1 Seq tab The Seq 3 tab is used e 2 Test type Sweep in frequency e 3 Harm harmonic order 6th e 4 Level 1 0 e 5 Duration 1 000 second No 1 3rd row Frequency sweep e 1 Seq tab The Seq 3 tab is used e 2 Test type Sweep in frequency e 3 Harm harmonic order 8th e 4 Level 1 0 e 5 Duration 1 000 second Items common to No 1 1st to 3rd rows e Repeat repetition count The repetition count in a seq tab is not indicated e Trigger Not indicated e Sequences to run Sequences are run in the Seq 3 tab e Seq end Output off Not indicated e Bank Not indicated e UT 230V e Freq Not indicated e Repeat all The number of repetitions for the sequences to be run in Seq 3 is 1 SD003 PCR LA Individual Harmonics Table 4 3 Example of Indication of Running Results Odd Harmonics Multiples of 3 No Repeat Seq Item UT Flatt Harm Level Angle Dum Interval all tion i 1 4 Odd harmonics 230 0 3 1 0 0 5 000 1 000 multiple of 3 i 1 4 Odd harmonics 230 0 9 1 0 0 5 000 1 000 multiple of 3 i i 4 Odd harmonics 230 0 15 1 0 0 5 000 1 000 multiple of 3 SD003 PCR LA E Description of Table 4 3 Number in the following items corresponds to the numbered steps in Fig 4 5 No 1 1st row Odd harmonics multiples of 3
69. ng sent 2 8 4 Corrective Measures in the Event of Problems If a problem occurs in the equipment under test EUT operate Quick Immunity Sequencer and the hardware to abort testing For better protection of the EUT we recommend aborting via hardware E Operating Quick Immunity Sequencer Click the F9 function key or the Abort button on the toolbar to abort testing The AC Power Supply PCR LA output is also turned off E Operating the hardware Turn off the AC Power Supply PCR LA POWER switch 2 9 Observing a Test Signal SD003 PCR LA When observing a test signal using an oscilloscope use the trigger signal output described below For trigger signal output from the AC Power Supply PCR LA a trigger signal selected under the test conditions for each standard is output The ITO1 PCR L s trigger signal output is not related to a trigger signal selected in the test condition setting for each standard E Trigger signal output from the AC Power Supply PCR LA A trigger signal is output from the SEQ TRIG OUT terminal BNC connector at the rear of the AC Power Supply PCR LA A low level signal is output for several 10 us at a location on the waveform where trigger signal output has been specified The signal level in cases other than signal output is approx 5 V The low level is approx 0 V The BNC connectors are isolated from the INPUT and OUTPUT terminal boards of the AC Power Supply PCR LA However the line for each signal of the PEA
70. nity test according to the IEC 61000 4 28 standard SD003 PCR LA IEC 61000 4 28 7 1 7 1 Outline of the Test The Power frequency variation immunity test determines the equipment s immunity against Power frequency variations E Electromagnetic environment classes Quick Immunity Sequencer does not specify an electromagnetic environment class or assume it to be a setting condition An electromagnetic environment class for test conditions should be set at the user s discretion E Three phase operations The test conditions are all set in relation to phase U Test signals change simultaneously in three phases on the basis of phase U 7 2 IEC 61000 4 28 SD003 PCR LA 7 2 Preparing the Test Conditions 3 Transition 5 Transition 9 Sequences to run 1 Seq tabs 2 f1 3 time 4 f1 Af 5 time 6 f2 qco Q eque e LJ File Edit Instrument Opti Help a S Pi 100 a A Open Cond nd Out OF To 200 Expand Cdhtract Abo v se Seq Seq2_ Seg3 Seq4 Seqis Seq Seg Seqs Seq Seq 10 Ip seq 2 7 7 _ 5eq 3 61000 4 11 Clseq4 _ Seq5 ae seq6 i Ll5eq7 i 61000 4 13 Ll5eq8 mrtg a ie R AR md tees ee Renee L seq9 re o r Llseg 10 a i i i Seq end 6100 4 14 Output off i p ATtDeta f N fe vAN UT 100 0 v i 50 0 S tp 50 00 Hz a ee i jo 100 jo to0 jooo yt Repeat all 1 him s himis ims7 him sA 61
71. nt at which the return voltage UT is reached and the next Voltage dip At the end of repetition intervals this is the time taken to move to the next seq tab Enter the value in seconds or in hours minutes seconds format For most cases select 10 000 s Alternatively enter a value ranging Input range from 0 020 seconds to 999 hours 999 minutes 999 999 seconds Input range with Enter a value ranging from 10 000 s to 999 000 s increment 1 s ITO1 PCR L NOTE e If the interval is set to 0 02 seconds at 50 Hz frequency or to 0 18 seconds at 60 Hz frequency it consists of two periods of the interval The interval is set as the number of periods in which 1 is added to the number of periods for a test signal contained in the set time A 1 is added because one period is required to detect a Zero Cross 8 Repeat repetition count Enter the number of repetitions for the test conditions in the seq tab in question Select one of 1 3 5 and 10 times or enter a value ranging from 1 to Input range 9999 Input range Select one of 3 6 9 and 12 times or enter a multiple of 3 within the with range of 3 to 9999 mice e 9 Trigger Select trigger signal output The red symbol W_ indicates at a location selected using a checkbox 10 Sequences to run Check the corresponding checkboxes of the seq tabs whose test conditions are used to run a test sequence Test sequences are
72. nu item has the same function as the To 100 V button on the toolbar E 200 V range Sets the PCR LA voltage range to 200 V When the voltage range is set to 200 V a check mark is displayed to the left of this item This menu item has the same function as the To 200 V button on the toolbar E AC mode Selects AC mode from AC AC S modes This function is valid only in the PCR LA series In the PCR L series AC mode is the only valid mode E AC S mode Selects AC S mode from AC AC S modes This function is valid only in the PCR LA series In the PCR L series AC mode is the only valid mode E I O Configuration Configures settings related to communications with the PCR LA series and other devices When controlling the test system using the RS 232C interface type COMI or COM2 into the Edit box To control the test system using the GPIB interface type DEV1 DEV2 and so on into the Edit box The number following DEV is a GPIB address This menu item also lets you select whether to use not use the ITO1 PCR L immunity tester If this is set for use the AC Power Supply PCR LA is also controlled via the GPIB interface only In this case type DEV1 DEV2 and so on in the same way into the ITO1 Device Name dropdown list box e If the Use ITO1 PCR L checkbox is checked Voltage variation tests according to IEC 61000 4 11 are not possible Only Voltage dips and Short interruptions may be selected For three phase systems che
73. nu to set the voltage range For more information please refer to E To 100 V 200 V button in 2 7 2 Toolbar CAUTION Check the set value of UT taken between the seq tabs used to run a test If the UT set value differs between the seq tabs unintended voltage change may occur depending on selection of seq tabs whose sequences are run 4 Level test level For DC Voltage dips enter a residual voltage in with respect to UT This level is set to 0 no input in the case of DC Short interruptions Input range Select either 40 or 70 or enter a value ranging from 0 to 200 For DC Short interruptions enter the impedance condition Input range Select either High impedance or Low impedance When High impedance is selected an inverse current from the EUT is blocked When Low impedance is selected an inrush current from the EUT is absorbed SD003 PCR LA IEC 61000 4 29 8 3 NOTE 5 Duration Enter the duration of the DC Voltage dip or DC Short interruption Enter the value in seconds or in hours minutes seconds format Select one of 0 01 s 0 03 s 0 1 s 0 3 s and 1 s or enter a value ranging Input range from 0 001 seconds to 999 hours 999 minutes 999 999 seconds e How to enter in hours minutes seconds format A time value of up to 999 999 999 999 may be set by delimiting each unit with a colon If only one colon is used the value is tr
74. o DC Abrupt voltage changes Abrupt voltage changes and short interruptions falling to OV 3 2 1 Voltage Dips and Abrupt voltage changes and Voltage IEC61000 4 11 Short Interruptions specifying the phase angle at the change start of change 3 2 2 Voltage Slow voltage changes Variations 4 2 1 Combined Peak of waveform clipped Single or Harmonics Flat Curve ae 4 2 2 Combined Waveform with sharp peaks Harmonics Over ONARE IEC61000 4 13 Swing distortion 4 2 3 Sweep in Superimposing noise on power Frequency supply 4 2 4 Individual Harmonics 5 2 1 Voltage oe a ey paa IEC61000 4 14 Fluctuation i e nad 5 2 2 Interval Voltage Unbalance in three phases nabalance c Three phase IEC61000 4 27 OF Pep arig the Test phase angle Conditions change Power Single or 7 2 Preparing the Test Changes in power frequency feopieney hese phase TEC61000 4 28 Conditions Abrupt DC voltage changes 8 2 1 DC Voltage Dips Abrupt DC voltage changes and Voltage and Short Interruptions short interruptions falling to OV chais DC IEC61000 4 29 8 2 2 DC Voltage Variations SD003 PCR LA Appendix A 1 A 2 Appendix SD003 PCR LA A ActiveX instrument driver P 2 Adrs 5 5 Bank 4 5 C Corrective Measures in the Event of Problems 2 19 D DC Voltage Dips and Short Interruptions 8 2 DC Voltage Variations 8 5 Default voltage and frequency 2 15 Determining sweep frequency 4 9 E Edit Menu 2 12 Error messag
75. o be run in the order of Seq 5 and Seq 6is 1 SD003 PCR LA IEC 61000 4 14 5 9 5 10 IEC 61000 4 14 SD003 PCR LA Cn Chapter 6 IEC 61000 4 27 Describes the Unbalance immunity test according to the IEC 61000 4 27 standard SD003 PCR LA IEC 61000 4 27 6 1 6 1 6 2 IEC 61000 4 27 Outline of the Test E Electromagnetic environment classes The Unbalance immunity test determines the equipment s immunity against unbalance in three phases Quick Immunity Sequencer does not specify an electromagnetic environment class or assume it to be a setting condition An electromagnetic environment class for test conditions should be set at the user s discretion E Items not conforming to the standard s requirements Certain Quick Immunity Sequencer items do not conform to the standard s requirements in the test configuration including the hardware The following table shows these items and the performance achieved in the test configuration Item Standard s requirements Performance achieved in the test configuration Voltage and phase changes in three phase operations Concurrent changes between phases Presence of a lag of several seconds between phases Minimum duration of unbalance 0 1 seconds 1 to several seconds Responsiveness to voltage changes 1 us to 5 us Approx 30 us SD003 PCR LA 6 2 Preparing the Test Conditions 2 Ua 3 Ub 10 Sequences to run 1 Seq t
76. o the rated voltage of the EUT Select either 100 or 110 or enter a value ranging from 0 0 to Input range 150 0 3 Ub Enter a ratio of a quantity to the rated voltage of the EUT Select one of 66 87 90 93 5 and 95 2 or enter a value fnuputealiee ranging from 0 0 to 150 0 4 Uc Enter a ratio of a quantity to the rated voltage of the EUT Select one of 71 74 80 87 and 90 or enter a value Iepure eange ranging from 0 0 to 150 0 5 Phase angle Ua Ub Enter the Ua Ub phase angle Select one of 125 127 131 134 and 139 or enter a value ranging Input range from 1 to 360 6 Phase angle Ua Uc Enter the Ua Uc phase angle Select one of 235 238 239 and 240 or enter a value ranging from Input range 1 to 360 7 Duration Enter the duration of unbalance Enter the value in seconds or in hours minutes seconds format Select one of 2 s 15 s 30 s and 60 s or enter a value ranging from 1 000 Input range second to 999 hours 999 minutes 1000 seconds 8 Interval repetition interval Enter the repetition interval of unbalance In this period each phase voltage is set to 100 the Ua Ub phase angle is set to 120 and the Ua Uc phase angle is set to 240 The increment is 1 second Enter the value in seconds or in hours minutes seconds
77. ons bit slower Mind that multiple open is always possible from the same process even if this option is unchecked Enable VISA C API OK Cancel 7 Check the Use ASRL Serial instruments checkbox 8 Click the Enable VISA C API button 9 Follow the prompts displayed in the window The KI VISA Setup window shown in step 6 is displayed once again 10 Click the OK button This completes the setup process E When using a CONTEC GPIB 6 Click on the GPIB CONTEC tab KI VISA COM I O Config useTwc ASRL Use GPIB instruments CONTEC API GPIB ver4 01 required GPIB Board Index Model ch dd Remove Enable VISA C API OK Cancel 7 Check the Use GPIB instruments CONTEC API GPIB Ver 4 01 required checkbox 8 Enter for example GPIBO in the Index box and GP IB PCI F in the Model box Check the Use this board checkbox 9 Click the Enable VISA C API button 10 Follow the prompts displayed in the window The KI VISA Setup window shown in step 6 is displayed once again 11 Click the OK button This completes the setup process SD003 PCR LA Setup 1 5 E When using an Interface GPIB 6 Click on the GPIB NI 488 2M tab KI VISA COM I O Confie USBTMC asRL GPIB CONTEC JV Use GPIB instruments NI 488 2M compatible APT m GPIB Board Index epteo x Model Interface PCI 4301 z JV Use this board Auto Search Add Remove Enable VISA C API OK
78. ox 5 Ensure to click the Enable VISA C API button 6 Follow the prompts displayed in the window The KI VISA Setup window shown in step 2 is displayed once again 7 Click the OK button This completes the setup process 2 8 Overall Picture of Test Sequences SD003 PCR LA E When using the Interface GPIB board 2 Click on the GPIB NI 488 2M tab KI VISA COM I O Config IV Use GPIB instruments NI 488 2M compatible API GPIB Board Index ePteo Model Interface PCI 4301 IV Use this board Auto Search Add Remove Enable VISA C API OK Cancel Check the Use GPIB instruments NI 488 2M compatible API checkbox Enter for example GPIBO in the Index box and Interface PCI 4301 in the Model box Check the Use this board checkbox Ensure to click the Enable VISA C API button Follow the prompts displayed in the window The KI VISA Setup window shown in step 2 is displayed once again jo lee iP IN Click the OK button This completes the setup process SD003 PCR LA Overall Picture of Test Sequences 2 9 2 6 Overall Picture of Test Sequences 1 Selection of 2 Test condition 3 Waveform 4 Running the standard settings preview results Q eque e Eile Egt Instrument Option Help g 2 11 9 gt WA 7 Op Cond top Pause Out OFf T Expand Contract Abort gei 5eg iV Seq2 Seq3_ seq4 Seq5 Seg Mseg7
79. p Stops the test Note that output remains enabled This menu item has the same function as the Stop button on the toolbar E Pause Entering step toward pause A test is paused immediately before moving to the next seq tab Until the sequence of the seq tab running at that time is complete the menu indication changes to Cancel When the test is paused output can be turned on off allowing the EUT to be checked or replaced While the test is paused the menu indication changes to Continue This menu item has the same function as the Pause button on the toolbar e Before connecting or disconnecting the EUT always confirm that the AC Power Supply PCR LA output is OFF E Output On Turns on the PCR LA output This allows you to start a test with the PCR LA output set to ON This menu item has the same function as the Out On button on the toolbar E Output Off Turns off the PCR LA output This menu item has the same function as the Out Off button on the toolbar Overall Picture of Test Sequences 2 13 NOTE E Abort This function is assigned to the F9 function key It disables output change and turns output off In the event of an emergency you must also turn off the PCR LA POWER switch This menu item has the same function as the Abort button on the toolbar E 100 V range Sets the PCR LA voltage range to 100 V When the voltage range is set to 100 V a check mark is displayed to the left of this item This me
80. rein to avoid improper connections or incorrect handling Incorrect connection or handling of any equipment device in the following configuration may result in serious accidents involving EUT damage or fire e AC Power Supply PCR LA Series e AC Power Supply PCR L Series e GPIB interfaces IB03 PCR LA and IB11 PCR L e Three phase drivers 3P03 PCR LA and 3P02 PCR L e Immunity tester ITO1 PCR L E Single phase operations 1 Equipment under test EUT Use either Trigger signal output Voltmeter or oscilloscope Fig 2 1 Test Configuration for Single phase Operations Example Using AC Power Supply PCR LA E Single phase operations 2 AC Power Supply PCR LA Series Equipment Immunity tester under test IT01 PCR L EUT Trigger signal output Voltmeter or oscilloscope Fig 2 2 Test Configuration for Single phase Operations Example Using AC Power Supply PCR LA and IT01 PCR L Immunity Tester Overall Picture of Test Sequences 2 3 ANCAUTION Before turning on off the IT01 PCR L immunity tester always confirm that the AC Power Supply PCR LA output is OFF to avoid damaging the equipment For operating procedures please refer to 2 8 1 Turning Power on off in the Test Configuration E Three phase operations RS 232C Use either GPIB AG Power Supply PC 1BO3 PCR LA PCR LA Series 3P03 PCR LA L___ in phase U Trigger signal output AC Power Supply 7 Eq
81. s depending on settings the entire waveform cannot be displayed E Contract button Horizontally contracts the preview waveform in the middle of the window allowing you to view the entire condition However in certain cases if a long duration setting has been made the entire waveform cannot be displayed E Abort button Disables output change and turns output off In the event of an emergency you must also turn off the AC Power Supply PCR LA POWER switch 2 8 Running a Test b Q A Seq 1 Seq2 Seq 3 Seq 4 Seq5 Seq6 Seq CJseq8 Ciseq9 seq 10 Seq end C output off CITT UT 100 0 y Freq 50 0 Hz Repeat all 1 NOTE A test is conducted by running sequences of the seq tabs selected using the checkboxes in turn starting with the smallest numbered seq tab The test disregards seq tabs that have not been selected In an example of the figure shown on the left Seq 1 to Seq 7 are run in turn Because the repetition count for all sequences Repeal all is 1 the test halts when the sequences are run up to Seq 7 Since the Seq end Output off checkbox has not been checked output remains enabled at the end of the test e A transition between the seq tabs takes approx 1 3 seconds Accordingly the repetition interval at the end requiring a transition between the seq tabs is equal to the set value plus approx 1 3 seconds e For testing according to the IEC 61000 4 27 stan
82. sage Error write to instrument in device DEVx The GPIB cable may be disconnected Check that the GPIB cable is properly connected The device name may be incorrect Check the device name on the dialog box shown in 2 4 Setting up I O Configuration 2 6 Overall Picture of Test Sequences SD003 PCR LA This may indicate that the IT01 PCR L immunity tester has not been connected in a configuration in which the IT01 PCR L is expected Uncheck the Use ITO01 PCR L checkbox in the dialog box shown in 2 4 Setting up I O Configuration or connect the ITO01 PCR L to the PCR LA E Error message Different model access in device DEVx The setting for single phase or three phase system for the connected PCR LA may be incorrect Check the setting for the 3 phase system checkbox in the dialog box shown in 2 4 Setting up I O Configuration E Error message As an interface error has occurred Quick Immunity Sequencer is terminated Turn off the power switch of the PCR L A This may indicate that the communication cable is disconnected Turn off the PCR LA POWER switch Reconnect the communication cable then re start Quick Immunity Sequencer 2 5 1 Setting up KI VISA SD003 PCR LA E Measure 1 Using the RS 232C interface for KI VISA 1 Click the Windows Start button point to Program Kikusui IO Software and VISA then click on IO Config KI VISA COM I O Confie USBTMC ASRL GPIB CONTEC GPIB NI 488 2M JV Use ASRL
83. st or soiled a new copy can be provided for a fee In either case please contact Kikusui distributor agent and provide the Kikusui Part No given on the cover This manual has been prepared with the utmost care however if you have any questions or note any errors or omissions please contact Kikusui distributor agent Microsoft and Windows are registered trademarks of Microsoft Corporation USA Pentium is a trademark of Intel Corporation USA National Instruments is the registered trademark of National Instruments Corporation USA NI 488 and NI 488 2 are registered trademarks of National Instruments Corporation USA Other company brand and product names provided in this manual are trademarks or registered trademarks of their respective holders Reproduction and reprinting of this manual whole or partially without our permission is prohibited Both unit specifications and manual contents are subject to change without notice Copyright 2005 2006 Kikusui Electronics Corporation Safety Precautions e cere reece eeee l Preface P 1 P 1 About This Manual 52 eee errr rrr ce eee P 1 P 2 Product Outline P 1 P 3 Required Operating Conditions P 2 P 4 Required Hardware rrr errr etter eee eee P 3 Chapter 1 Setup 1 1 1 1 Package Contents eer r cere e erence eee eens 1 2 1 2 Installation
84. tabs Seq 1 to Seq 10 For voltage this shows relative values with the initial value set to 1 The Expand and Contract buttons on the toolbar allow the waveform to be horizontally expanded or contracted in ten steps Note that there may be one period of a deviation from the actual output The color of the waveform corresponds to the color of the seq tab set Use Color in the Option Menu to change colors for a seq tab 4 Running results Progress bar Displays the status of a test sequence run in bar graph form To change colors use Color in the Option Menu List of the running results Displays the set values for the seq tab whose sequence has been run The specific items indicated depend on the standard in question Please refer to the corresponding chapter SD003 PCR LA Overall Picture of Test Sequences 2 11 2 Menus and Toolbar 2 7 1 Menus File Menu Edit Menu E New Sets the default values and creates a new test condition file under an untitled file name E Open Opens the dialog box used to select a test condition file This menu item has the same function as the Open button on the toolbar The extension for a test condition filename is qco E Save Test Conditions Saves the test condition settings under the same name as the currently open test condition file replacing the earlier version This menu item is used to save all the settings related to the standard E Save As Test Conditions
85. tage increasing time 14 Repeat all 10 Trigger 5 Voltage decreasing time 12 Seq end Output off Fig 3 2 Voltage Variations Setting Window The following items other than 2 5 6 7 and 11 are the same as for Voltage dips and Short interruptions Please refer to 3 2 1 Voltage Dips and Short Interruptions 1 Seq tabs 2 Test type Select the Voltage variations NOTE e If the Use ITO1 PCR L checkbox is checked in the dialog box displayed by clicking on I O Configuration from the Instrument Menu Voltage variations are not possible SD003 PCR LA IEC 61000 4 11 3 7 3 UT 4 Level test level 5 Decreasing voltage decreasing time Enter the time taken to reduce voltage from the initial voltage UT to the reduced voltage Enter the value in seconds or in hours minutes seconds format For most cases select 2 000 s Alternatively enter a value ranging Input range from 0 001 seconds to 999 hours 999 minutes 999 999 seconds 6 Reduced reduced voltage time Enter the reduced voltage duration Enter the value in seconds or in hours minutes seconds format For most cases select 2 000 s Alternatively enter a value ranging Input range from 0 001 seconds to 999 hours 999 minutes 999 999 seconds 7 Increasing voltage increasing time Enter the time taken to return from the reduced voltage to
86. test level 2 Enter the center value of Voltage fluctuation This should be a ratio to the rated voltage of the EUT For most cases set it to the value of the UT of the next seq tab in running sequences Select one of 90 100 and 110 or enter a value ranging from Input range 1 0 to 150 0 5 Interval1 2 repetition interval Enter the repetition interval of Voltage fluctuation Enter the value in seconds or in hours minutes seconds format Select either 60 s or 120 s or enter a value ranging from 0 001 seconds Input range to 999 hours 999 minutes 999 999 seconds NOTE e A transition between the seq tabs takes approx 1 3 seconds Accordingly Interval2 repetition interval at the end requiring a transition between the seq tabs is equal to the set value plus approx 1 3 seconds 6 Repeat repetition count 7 Trigger 8 Sequences to run 9 Seq end Output off 10 Adrs SD003 PCR LA IEC 61000 4 14 5 7 5 3 11 UT 12 Freq 13 Repeat all Results of Sequence Running The running results are indicated every time the sequence for one seq tab finishes The starting date and time are shown in the row above the No 1 row Shown below is a typical test example An indicated value becomes a blank or 0 if it is not related to the test The Repeat set value in the panel for each seq tab is not indicated Table 5 1
87. the initial voltage UT Enter the value in seconds or in hours minutes seconds format For most cases select 10 000 s Alternatively enter a value ranging Input range from 0 001 seconds to 999 hours 999 minutes 999 999 seconds 8 Interval repetition interval 9 Repeat repetition count 10 Trigger 11 Sequences to run Check the corresponding checkboxes of the seq tabs whose test conditions are used to run a test sequence Test sequences are run in turn starting at Seq 1 seq tabs whose corresponding checkboxes have not been checked are not used to run a test NOTE e If the Use IT01 PCR L checkbox is checked in the dialog box displayed by clicking on I O Configuration from the Instrument Menu the seq tabs to which Voltage variation has been set are not used in the test 12 Seq end Output off 13 Freq 14 Repeat all 3 8 IEC 61000 4 11 SD003 PCR LA 3 3 Results of Sequence Running The running results are indicated every time the sequence for one seq tab finishes The starting date and time are shown in the row above the No 1 row Shown below is a typical test example An indicated value becomes a blank or 0 if it is not related to the test Voltage dips Short interruptions and Voltage variations Table 3 1 Example of Indication of Running Results No m Seq Item UT Level Angle pa Reduced ue Interval 1 1 1 Voltage variation 230 0 40 2 000 1 000
88. tion between the seq tabs takes approx 1 3 seconds Accordingly Repetition 3 repetition interval at the end requiring a transition between the seq tabs is equal to the set value plus approx 1 3 seconds 6 Repeat repetition count Enter the number of repetitions for the test conditions in the seq tab in question Select one of 1 3 5 and 10 times or enter a value ranging from 1 to Input range 9999 5 4 IEC 61000 4 14 SD003 PCR LA NOTE SD003 PCR LA 7 Trigger Select trigger signal output The red symbol W indicates a location selected using the checkbox 8 Sequences to run Check the corresponding checkboxes of the seq tabs whose test conditions are used to run a test sequence Test sequences are run in turn starting at Seq 1 Seq tabs whose corresponding checkboxes have not been checked are not used to run a test 9 Seq end Output off When this checkbox is checked a test signal is turned off at the end of sequence running the AC Power Supply PCR LA output is turned off 10 Adrs To implement the Voltage fluctuation immunity test Quick Immunity Sequencer uses the AC Power Supply PCR LA sequence function This sequence function can accommodate up to 100 addresses adrs For Voltage fluctuation 18 sequence addresses are used for one seq tab while two sequence addresses are used for one seq tab for Interval If six or more of the seq tabs for which Voltage fluctuation is
89. tion in the supply voltage E Electromagnetic environment classes 5 2 IEC 61000 4 14 Quick Immunity Sequencer does not specify an electromagnetic environment class or assume it to be a setting condition An electromagnetic environment class for test conditions should be set at the user s discretion E Items not conforming to the standard s requirements Certain Quick Immunity Sequencer items do not conform to the standard s requirements in the test configuration including the hardware The following table shows these items and the performance achieved in the test configuration Item Standard s requirements Performance achieved in the test configuration tf tr of Voltage fluctuation The timing for five periodic transitions must be specified The timing for five periodic transitions is not specified Start phase of three phase operations 270 for each phase 270 only at the first period in phase U E Three phase operations The test conditions are all set in relation to phase U Test signals change simultaneously in three phases on the basis of phase U SD003 PCR LA 5 2 Preparing the Test Conditions The test conditions are prepared using two separate windows the Voltage fluctuation and interval setting windows 5 2 1 Voltage Fluctuation 10 Aars 2 Test condition type 4 Voltage 8 Sequences to run 1 Seq tabs 3 Test level l eae 5 Repetition interval E Testl qco
90. tions setting accuracy within 10 are met Without the ITO1 PCR L the standard s performance specifications will be met at an angle of multiples of 18 such as 90 180 etc for 50 Hz The setting accuracy at other phase angles is 18 at 50 Hz or 21 6 at 60 Hz GPIB For the GPIB interface the IBO3 PCR LA GPIB interface is required for AC Power Supply PCR LA Series or the IB11 PCR L GPIB interface for AC Power Supply PCR L Series Three phase operations In three phase operations the 3P03 PCR LA three phase driver is required for AC Power Supply PCR LA Series or the 3P02 PCR L three phase driver for AC Power Supply PCR L Series Preface P 3 P 4 Preface SD003 PCR LA Chapter 1 Setup Describes how to install the Quick Immunity Sequencer application software SD003 PCR LA Setup 1 1 1 1 Package Contents The Quick Immunity Sequencer package contains the following Item name Quantity Program CD ROM 1 User s Manual this manual Z1 003 152 1 1 2 Installation 1 2 1 Preparing for Installation NOTE 1 2 Setup The ActiveX instrument driver specially designed for the PCR LA series requires a VISA driver You must first install a VISA driver The VISA driver required depends on the I O interface Refer to Table 1 1 below to choose the necessary VISA driver e Use of the wrong VISA driver with the GPIB interface may disable the GPIB interface Installing multiple VISA
91. to voltage at return For most cases enter the rated voltage of the EUT Select one of 100 V 110 V 120 V 200 V 220 V 230 V and 240 V or enter a value ranging from 0 0 V to 305 0 V Input range NOTE e An appropriate voltage range must be selected Use the toolbar or the Instrument Menu to set the voltage range For more information please refer to E To 100 V 200 V button in 2 7 2 Toolbar 12 Freq Enter the rated frequency of the EUT Select either 50 Hz or 60 Hz or enter a value ranging from 1 Hz to Input range 699 50 Hz 13 Repeat all Enter the repetition count of all the selected seq tabs Seq 1 to Seq 10 Select one of 1 3 5 and 10 times or enter a value ranging from 1 to Input range 9999 4 6 IEC 61000 4 13 SD003 PCR LA 4 2 2 Combined Harmonics Over Swing 10 Bank 2 Test type 8 Sequences to run 1 Seq tabs 3 Test level 5 Repetition interval Ea Test1_qco Quick In munity Sequencer DER Eile Edit Instrument Opt p A Open Cond e f If Msi 5 E q2 Seg3 Seq4 Seq5 Seq6 Seq Segs Seq9 Seq 10 Grace Harm Level oj 61000 4 11 D oror sana Harm Level f Angle deg O Sweep in Frequency MA O Odd non multiple of 3 61000 4 13 9 100 an Wild Out On To 2004 Expand Contract 10 0 180 soo fo O Odd multiple of 3 OEvenharmonics J erent fee eet ee nee se TET
92. uipment PCR LA under test Series 3P03 PCR LA in phase V EUT AC Power Supply PCR LA Series 3P03 PCR LA in phase W Voltmeter or oscilloscope Fig 2 3 Test Configuration for Three phase Operations Example Using AC Power Supply PCR LAs NOTE e The IT01 PCR L immunity tester cannot be used in three phase operations with Quick Immunity Sequencer 2 3 Launching Quick Immunity Sequencer To launch Quick Immunity Sequencer double click the Quick Immunity Sequencer shortcut on the desktop P Kikusui Quick Immunity Se Launching Quick Immunity Sequencer displays the window shown in Fig 2 4 on page 2 10 2 4 Overall Picture of Test Sequences SD003 PCR LA 2 4 Setting up I O Configuration NOTE NOTE SD003 PCR LA Set up the I O configuration using I O Configuration from the Instrument Menu For more information on this menu please refer to 2 7 1 Menus I O Configuration PCR Device Name COM2 Select or type a VISA Resource of the instrument When using PCR LA series in R5232C set 19200 N 8 2 0822 When using PCR L series in RS232C set 9600 N 8 2 0834 use 1701 PCR L 13 phase system O 1 Select DEV1 COM1 COM2 etc in the PCR Device Name dropdown list box 1 in DEV 1 stands for a GPIB address COM1 refers to COM communications port 1 Select COM1 or COM2 when using the RS 232C interface 2 When using the RS 232C interface
93. upply PCR LA Series For information on the PCR LA series hardware please refer to the AC Power Supply PCR LA Series Operation Manual Using the product with the conventional PCR L series Quick Immunity Sequencer can also be used for the conventional AC Power Supply PCR L Series In this case read PCR LA as PCR L For information on the PCR L series hardware please refer to the AC Power Supply PCR L Series Operation Manual Versions of the AC power supplies for which this manual is valid PCR LA series Ver 3 10 or later PCR L series Ver 2 04 or later P 2 Product Outline SD003 PCR LA The Quick Immunity Sequencer application software for the AC Power Supply PCR LA Series allows you to simulate various phenomena in the AC power supply environment It can be used to perform immunity tests for electric or electronic equipment connected to low voltage power distribution systems under the conditions stipulated in the following standards Since test conditions may be set to exceed the ranges specified by the standard the product can be used in preliminary tests conducted before actual testing for the standard immunity allowance tests and stress tests e TEC61000 4 11 2001 03 Edition1 1 Voltage dips Short interruptions and Voltage variations e JEC61000 4 13 2002 03 1st Edition Harmonics and Interharmonics e JEC61000 4 14 2002 07 Edition1 1 Voltage fluctuation e JEC61000 4 27 2000 08 Ist Edition
94. vided 3 KI VISA supports the following GPIB models CONTEC GP IB PCDL GP IB PM GP IB PCI F and GP IB CB F API GPIB driver Ver 4 01 or later recommended Interface PCI 4301 LabVIEW compliant GPC 4301N driver Ver 1 21 or later recommended When using a CONTEC GPIB board install an ordinary API GPIB driver rather than API GPLV driver Lab VIEW compliant NI 488 2M API compatible When using an Interface GPIB board install GPC 4301N LabVIEW compliant NI 488 2M API compatible instead of an ordinary GPC 4301 driver 4 We recommend that users holding a National Instruments license use NI VISA 1 2 2 Installing the KI VISA Driver 1 Load the Quick Immunity Sequencer setup CD into the CD ROM drive The following Setup window is displayed lol xi File Edit View Favorites Tools Help Bak gt gt amp A Qsearch Favorites Smeda lt 4 G GD Address E H PCR_LA_ApH Setup_Image Qis_V1_01 5A601702 Readme ht Go Links Installation of Quick Immunity Sequencer Quick Immunity Sequencer Setup Readme Microsoft Internet Explorer Quick Immunity Sequencer needs a VISA library which is based on the VISA specification 2 2 or later i e KI VISA 2 2 x NI VISA 2 6 or later Agilent I O Libraries K01 00 or later etc Regarding the VISA library of National Instruments or Agilent Technologies ask each vender about it KIKUSUI KI VISA also supports GPIB of Contec and Int

Download Pdf Manuals

image

Related Search

Related Contents

Manuel d`utilisation Édition 1 / 04_2014  AUTOMARK/SLIDEMATIC SYSTEM  User Manual - VIAIR Corporation  HRSA HSMES Database User Manual 5.0  ドメイン指定受信設定方法【docomo】  

Copyright © All rights reserved.
Failed to retrieve file