1. Pelstar LLC has been officially certified as the manufacturer IS O1 3485 2003 of Health o meter Professional medical devices The Pelstar CERTIFIED LLC quality assurance system covers the development production sales and service of medical scales and measuring systems This product is tested to and meets the requirements of 1 UL60601 1 First Edition Safety for Medical Electrical Equipment 2 CSA C22 2 No 601 1 M90 Medical Electrical Equipment 2650KL 2700KL Scale Connectivity Information To enable reliable transmission of weight height and Body Mass Index BMI data these scales are designed to connect to a computer monitor or other electronic device via a built in USB port This connectivity helps minimize a source of potential medical errors caused by incorrectly copying and then recording patient data These scales are EMR EHR compatible For more information regarding how to connect these scales to other electronic devices please contact Health o meter Professional Scales Technical Support available during normal business hours at 1 800 683 3722 2650KL 2700KL Scale Disposal These Health o meter Professional scales must be disposed of properly as electronic waste Follow the national regional or local regulations which apply to you for disposal of electronic waste or batteries Do not dispose of this device in the domestic waste stream 2650KL ASSEMBLY INSTRUCTIONS Before Assembly Each
2. Symptom Possible Cause Corrective Action Scale does not turn on 1 Dead Battery 1 Replace batteries 2 Faulty electrical outlet 2 Use a different outlet 3 Bad power supply 3 Replace AC adapter Questionable weight or 1 External object interfering 1 Remove interfering object the scale does not zero with the scale from the scale 2 The display did not show 2 Ask the patient to get off the 0 0 before weighing scale zero the scale and begin weighing process again 3 Scale is not placed on a 3 Place the scale on a level level floor floor and begin weighing process again 4 Scale is out of calibration 4 Check weight with known weight value 5 Improper Tare 5 Place the item to be tared on the scale Press REWEIGH Once the weight of the item is displayed press TARE Place the patient and the tared item back on the scale Press REWEIGH again Weighing is performed but The patient is not sitting or Ask the patient to stand or sit the weighing process takes standing still still too long and the weight does not lock on the displayed The display shows Over The load on the scale Remove the excess weight and message exceeds the capacity use the scale according to its 1000 Ib 454 kg limits The display shows The batteries are discharged Replace batteries according to Lo bAtt message instructions 24 WARRANTY Limited Warranty What does the Warranty Cover Health o meter Professi
3. 1 1 ER R Note In order to mount the scale s display assembly onto the wall a set of wall anchors and screws will be required not included Refer to the User Instructions for secure display assembly mounting Tools Required Phillips head screwdriver not included Allen wrenches included 11 2700KL ASSEMBLY INSTRUCTIONS CONTINUED Assembling the 2700KL Scale Refer to Figures 4 6 Caution It is recommended that the assembly be performed by two persons 1 Place scale platform on a flat level and dry surface and carefully turn the platform upside down Caution Ensure that the wheelchair ramps are held in place when moving the scale ramps can flip open 2 Thread the 4 feet into the underside of the platform Each foot should be completely hand tightened into the platform Back out 1 2 full turns This will allow for proper load cell operation and scale accuracy 3 Carefully flip the scale over to its upright position on the floor Warning Do not drop scale as this can permanently damage the load cells 4 Adjust the feet in order to level the scale Ensure that the feet are not completely tightened into the platform for proper load cell operation 5 Open the scale to its operational position by folding open the wheelchair ramps 6 Connect the load cell cable to the scale platform by inserting the load cell cable connector into the load cell port See Figure 5 7 Ensure that the other load
4. 2013 15013485 2003 CERTIFIED P N UM2650KL 2700KL Rev 2 031913 25
5. 8 17 QUICK START OPERATING INSTRUCTIONS Keypad Capacity 1000lbX 21b 454kgX 1kg ER RECALL ON OFF TARE PRETARE REWEIGH ENTER DEPRESS AND HOLD FOR PRETARE Figure 10 Function Description Q ON OFF Turns scale on and off ZERO Zeros the scale prior to a weighing Holds the weight value of the patient on the display until the button is PE LEASE pressed again to clear the value Also moves the cursor from left to right on the display for entering height and pretare value Prompts entry of height data to calculate patients Body Mass Index BMI RECALL Recalls the previous weight TARE Tares out weight of an object on the platform TARE PRETARE PRETARE Prompts entry of a tare value that will be deducted from the weight on the platform Also releases tare weight returns display to zero Also used to increment height value and pretare value REWEIGH __ Allows repeated weighing of the patient without stepping off the scale Used to accept height value in BMI mode and tare value in PRETARE ENTER mode Also releases the recall value and BMI value 18 QUICK START OPERATING INSTRUCTIONS CONT Tare Function When using these scales the weight of an object such as a wheelchair can be subtracted from the total weight in order to determine the weight of the patient alone The Tare Function automatically performs this subtraction The 2650KL and 2
6. into the calibration path The following calibration assumes the scale is set to pounds and 400 Ib the default weight is employed as the calibration weight Scale Calibration Operator Action Display 1 Make sure the display is turned off then press and hold the U 01 followed by a 5 digit HOLD RELEASE button while pressing the ON OFF button idle scale count number 2 Press the ENTER button 3 Make sure there is nothing on the scale then press the ENTER Clear followed by button again 0400 0LB 4 Use the gt and A buttons to set the calibration weight to a Shows 400 0 or entered value other than the default weight The calibration weight should calibration weight be as high as possible with 400 Ib as a minimum calibration weight 5 Press the ENTER button Alternating Load followed by 400 0LB 6 Load the scale with the required weight Display increments showing raw weight data 7 Without touching or shaking the scale allow the display to SUCC or Err1 stabilize then press the ENTER button 8 If the calibration is successful the display will show SUCC and turn off If the calibration is not successful the display will show Err1 and turn off Contact customer service for an Err1 display 1 800 638 3722 23 TROUBLESHOOTING Troubleshooting Before contacting service personnel refer to the following instructions to check and to correct any failures
7. pages provide instructions for maintenance cleaning calibrating and troubleshooting for the 2650KL or 2700KL scale Maintenance operations other than those described in this manual should be performed by qualified service personnel Caution Before first use or after long periods of non use check the scale for proper operation and function If the scale does not operate correctly refer to qualified service personnel 1 Check overall appearance of the total scale for any obvious damage wear and tear 2 Inspect the AC adapter for cord cracking or fraying or for broken bent prongs Cleaning Proper care and cleaning is essential to ensure a long life of accurate and effective operation Caution Disconnect the scale from the AC adapter power source prior to cleaning the unit 1 Clean all external surfaces with a clean damp cloth or tissue Mild antimicrobial or antibacterial soap and water solution may be used Dry with a clean soft cloth 2 Do not immerse the scale into cleaning or other liquid solutions 3 Do not use Isopropyl Alcohol or other solutions to clean the display surface 4 Do not use abrasive cleaners 22 CALIBRATION Calibration These scales have been factory calibrated and does not require calibration prior to use For maintenance the scale can be calibrated Calibration of your 2650KL or 2700KL scale is performed using kilograms kg or pounds Ib according to the units of measure used upon entering
8. 2650KL scale is shipped disassembled in a single carton Carefully inspect the carton for shipping damage before unpacking If damage is found contact your shipper or a Health o meter Professional representative immediately at 1 800 815 6615 Claims must be filed with the shipper as soon as possible after receipt of the package The following information details what you will find inside the main carton as you unpack the parts for assembly To prevent scratching any components carefully remove each assembly from the carton and unwrap the packing materials Set the carton aside for storage To avoid damaging the scale parts when unpacking do not use a box cutter knife scissors or any sharp object to open the protective inner packaging Parts List 1 Scale Platform 1 User Instructions 1 Box Containing 1 Display Assembly Load Cell Connector Cable AC Adapter 4 Feet 1 1 AR R Note In order to mount the scale s display assembly onto the wall a set of wall anchors and screws will be required not included Refer to the user instructions for secure display assembly mounting Tools Required Phillips head screwdriver not included Allen wrenches included 2650KL ASSEMBLY INSTRUCTIONS CONTINUED Assembling the 2650KL Scale Refer to Figures 1 3 Caution It is recommended that the assembly be performed by two persons 1 Place scale platform on a flat level and dry surface and carefully turn the p
9. 700KL scales allow tare weight to be entered manually by using the keypad or automatically as outlined in this section TARE 1 With all weight off the scale press the ON OFF button to turn on the scale 2 Wait until the O OLB or O OKG appears on the display 3 Place the wheelchair on the scale The display will show a value for the wheelchair s weight 4 Press the TARE button The scale will zero out and the word TARE will appear at the bottom center of the display indicating that there is a tare value for this patient s wheelchair 5 Remove the weight from the scale Once it is removed from the scale the display will show a negative weight equal to the weight of the wheelchair 6 Place the patient and wheelchair on the scale The scale will automatically deduct the tare weight of the wheelchair from the gross weight of the patient and wheelchair together 7 The weight of the patient will appear on the scale 8 With no weight on the scale the tare value will remain on the display until changed cleared or until the scale is turned off 9 To remove the Tare press the TARE button again The word TARE will disappear from the bottom of the display and the tare value will be cleared from the display The scale will resume normal operation PRETARE Limit 200 0 Ib 1 Make sure that there is no object on the weighing platform 2 Press the ON OFF button to turn on the scale 3 Wait until the O
10. Health o meter FP fr OT s8 Bt 60 ra i Model 2650KL 2700KL Patient Wheelchair Scales d zah User Instructions P N UM2650KL 2700KL Rev 2 031913 Health o meter Model 2650KL 2700KL Thank you for your purchase of this Health o meter Professional product Please read this manual carefully and keep it for easy reference or training TABLE OF CONTENTS Caution ANG 00010 STEC eare 3 SPECIICALIONS Tor These Scales akad 4 Certifications Connectivity Disposal ssaa aaa aaa aaa aaa aaa a aaaa 6 Assembly Instructions 2650KL Scale lt lt lt lt lt lt 7 Assembly Instructions 2700KL Scale ooocccnnccnncccncocncococononcncononoconononanonancnnnnncnnns 11 Display Assembly Wall Mounting Instructions cooccocncoccccnconcocncnncononnncanincnnanons 15 SGUDZBAUENSSY MANSO emite 16 Q ick Start 10 1 6 11 611 010 S ercer ordered rro 18 Mate AE AA 22 CANO 9 PA 23 6 16 POONG i ea LEE RO HAZE ee ee 24 CA AP E E 29 Note These scales have been factory calibrated and does not require calibration prior to use Please register your scale for warranty coverage at http www homscales com customer registration aspx For User Instructions updates and revisions please go to http www homscales com CAUTION AND WARNING e To prevent injury and damage to you
11. LY Figure 7 14 DISPLAY ASSEMBLY WALL MOUNT INSTRUCTIONS Display Assembly Wall Mounting Instructions Caution When wall mounting the display assembly ensure that the mounting location is free of electrical lines gas pipes and other safety hazards Ensure that wall is of construction that can support the display assembly 1 Position the display assembly on the wall and carefully trace around the display outline 2 Mark a level horizontal line 2 4 70mm below the top of the display outline 3 Mark a vertical line through the center of the display outline The center line is 4 72 114mm from each side of the display outline 4 On the level horizontal line mark two drill holes 2 74 57mm away from the center line 5 Use the appropriate plastic or metal anchors for the type of wall construction and carefully drill the appropriate sized holes 6 Insert the anchors into the wall and screw in the two mounting screws into the anchors Adjust the screw head depth such that the bracket can be mounted over the two screw heads 7 Position the display assembly mounting holes onto the mounting screws and pull down on the display assembly to secure the bracket over the screws Required Parts 1 Display Assembly 2 8 Anchoring Screws 4 2mm nominal OD 2 Mounting Anchors Compatible with Wall Construction Note Anchors and Anchoring Screws are not included with these scales 15 SETUP BATTERIES TRANSPO
12. Note The scale will not calculate BMI for a patient that is 24 pounds 12 kg or less If a height is not entered for the patient within 30 seconds of pressing the BMI button the scale will return to the normal weighing mode 1 Make sure there is no object on the weighing platform 2 Press the ON OFF button to turn the scale on 3 Wait until the 0 0 LB or 0 0 KG appears on the display 4 Ask the patient to step on the scale The scale will begin to display weight information and will lock on the patient s weight when the stable weight of the patient is determined Depending on the movement of the patient on the scale it may take several seconds for the scale to lock onto the weight 5 Once a weight has been displayed while the patient is still on the scale Health o meter recommends performing a second weighing by pressing the REWEIGH button 6 After obtaining the patient s weight on the scale press the BMI button while the scale is on 7 The scale will prompt you to enter the patient s height Use the keypad A and buttons to enter the patients height Note When weighing in pounds the height is displayed in inches When weighing in kilograms the height is displayed in centimeter 8 Once height is keyed in press the ENTER button 9 The display will show the patients BMI 10 Press the ENTER button to return to the normal weighing mode of the scale 21 MAINTENANCE AA Maintenance The following
13. OLB or 0 0KG and ZERO appear on the display 4 Hold down the TARE button for 3 to 5 seconds The word TARE will appear at the bottom center of the display and the user will be prompted to enter the tare value 5 Use the keypad gt and A buttons to enter the weight of the wheelchair or other object Press the ENTER button 6 The value entered will show as a negative value on the display 19 QUICK START OPERATING INSTRUCTIONS CONT 7 Place the patient in the wheelchair and roll the wheelchair on the scale The scale will automatically deduct the tare weight as entered from the gross weight of the patient and the wheelchair 8 The weight of the patient will appear on the scale 9 With no weight on the scale the tare value will remain on the display until changed cleared or until the scale is turned off 10 To remove the Tare press the TARE button again The word TARE will disappear from the display and the tare value will be cleared from the display The scale will resume normal operation Weighing a Patient Note When these scales are turned on they will always default to the settings and units last used lb or kg 1 Make sure there is no object on the weighing platform 2 Press the ON OFF button to turn the scale on 3 Wait until the O OLB or 0 0KG appears on the display 4 Tare out the weight of the wheelchair and roll the patient in the wheelchair onto the scale The
14. RT Preparing the 2650KL 2700KL Scale for Use 1 Remove the protective plastic film covering the display the film says FACTORY CALIBRATED 2 Place batteries into the battery holder see Figure 7 3 Plug the scale s AC adapter into the power source 4 Ensure all objects are removed from the scale s platform Press the ON OFF button to turn the scale on The display will show Start then 0 0LB or 0 0KG 5 Press the UNIT button to select the weight mode Ib or kg Note These scales will always default to the settings and units last used lb or kg 6 When the AC adapter is disconnected from the scale the scale will switch to battery power if batteries are installed Note If the set up procedure failed refer to the troubleshooting instructions If the problem is not corrected refer to qualified service personnel Replacing Inserting Batteries 1 Unplug the AC adapter from the power source 2 Remove the battery cover on the side of the display assembly by unscrewing the 2 battery cover screws see Figure 7 3 Replace or install the batteries with new units Health o meter recommends using an alkaline or lithium battery rated for 130 F 54 C operation Do not use Zinc Carbon batteries 4 Re attach the battery cover to the display assembly Warning If the scale will not be used for some time remove the batteries to avoid a safety hazard Disposal of batteries must be performed
15. The scale is now assembled Plug the AC adapter into a power source and or insert the batteries 11 Carefully position the display assembly on the desk table or other flat surface Caution Ensure that the load cell cable is adequately secured to prevent a tripping hazard to the patient or user Note Refer to Display Assembly Wall Mounting Instructions located in this User Instruction Manual 2650KL ASSEMBLY INSTRUCTIONS CONTINUED _ GUAR Load Cell Port Figure 3 10 2700KL ASSEMBLY INSTRUCTIONS Each 2700KL scale is shipped disassembled in a single carton Carefully inspect the carton for shipping damage before unpacking If damage is found contact your shipper or a Health o meter Professional representative immediately at 1 800 815 6615 Claims must be filed with the shipper as soon as possible after receipt of the package The following information details what you will find inside the main carton as you unpack the parts for assembly To prevent scratching any components carefully remove each assembly from the carton and unwrap the packing materials Set the carton aside for storage To avoid damaging the scale parts when unpacking do not use a box cutter knife scissors or any sharp object to open the protective inner packaging Parts List 1 Scale Platform 1 User Instructions 1 Box Containing 1 Display Assembly Load Cell Connector Cable AC Adapter 4 Feet
16. cell cable connector is securely plugged into the display assembly load cell jack located on the rear panel of the display assembly The load cell jack is labeled LOAD CELL see Figure 3 8 If the AC adapter will be used connect the AC adapter cord plug into the AC power jack located on the back of the display assembly The AC power jack is labeled POWER see Figure 3 Caution Do not plug the scale into a power source until the scale is fully assembled 9 The scale is now assembled Plug the AC adapter into a power source and or insert the batteries 10 Carefully position the display assembly on the desk table or other flat surface Caution Ensure that the load cell cable is adequately secured to prevent a tripping hazard to the patient or user Note Refer to Display Assembly Wall Mounting Instructions located in this user instruction Manual 12 O LLI H a O gt Y y O H O a m H Y gt aul 00 LLI U Y lt _ x N N DISPLAY ASSEMBLY BRACKET SCREWS DISPLAY ASSEMBLY LOAD CELL CABLE CONNECTOR FOOT Figure 4 A O a Ko go GU E A Figure 5 13 y aii zz a ae sm HU Umm ma AZER ZZ a A D Y f Pa 4 Wa N AY 14 AN a a A ch ey BY a al FA 4 lt A W D g Ja i N N W j 1 Y tw Y w Zw Y Figure 6 BATTERIES DISPLAY ASSEMB
17. esigned to provide the user with features that make the weighing process simple fast and convenient These scales are set up to use motion sensing weighing technology in order to determine the actual weight of a moving patient Weight can be displayed in pounds or in kilograms For privacy and convenience in use the scale features a remote display head which may be held by the operator placed on a table or other flat surface or mounted on a wall Both scale models have 2 wheels built in for easy mobility Each scale can be powered using an AC adapter included or by 6 AA batteries not included 2650KL 2700KL Specifications Capacity and Resolution 2650KL 2700KL 1000 Ib x 0 2 Ib 454 kg x 0 1 kg Power Requirements 120V 240V AC 50 60Hz 9V DC adapter or 6 AA batteries Environmental Operating temperatures 50 F to 95 F 10 C to 35 C Storage temperatures 30 F to 125 F 0 C to 50 C Maximum Humidity 85 RH Physical Dimensions Platform 2650KL Assembled Size and Weight Length 31 72 80 0 cm Length 34 7 87 cm Width 31 79 4 cm Width 33 83 8 cm Height 2 72 6 2 cm Height 6 15 2 cm Weight 80 5 Ib 36 5 kg Physical Dimensions Platform 2700KL Assembled Size and Weight Length 32 72 82 6 cm Length 41 7 105 4 cm Width 36 91 4 cm Width 56 143 cm Height 2 72 6 2 cm Height 3 8 cm Weight 80 5 Ib 36 5 kg Certification Descriptions manana aana
18. in accordance with the prevailing national regional or local regulations which apply to you 16 SETUP BATTERIES TRANSPORT CONTINUED Preparing a 2650KL 2700KL Scale for Transport The ramps on these scales can be easily folded for movement from one room to another Both scales include mounting pins to secure the display assembly during transport the pins are located on the right wheelchair ramp To prepare the scale for transport follow these steps 1 Remove the display assembly from the wall by lifting up the display assembly off of the mounting screws 2 Unplug the AC adapter from the display assembly and remove the adapter from the outlet Carefully store the AC adapter and cord to prevent damage during transport 3 With the scale on the floor fold both ramps onto the platform 4 Position the display assembly onto the mounting pins located on the right folded ramp The display assembly is mounted onto the ramp by sliding the display assembly brackets onto the raised pins see Figures 8 amp 9 5 For the 2650KL grasp the guardrail located on the opposite side the wheels and lift the scale onto the two wheels The scale can now be transported by rolling the unit on its wheels 6 For the 2700KL grasp the handle located on the opposite side of the wheels and lift the scale onto the two wheels The scale can now be transported by rolling the unit on its wheels Mounting pins on right folded ramp Figure
19. ivates replace the batteries or switch to the AC adapter as soon as possible e When operating the scale with the display assembly on a flat surface ensure that the load cell cable is routed to prevent a tripping hazard to the patient or user e When operating the scale with the display assembly on the wall ensure that the display is securely mounted with the appropriate anchors and Screws e o prevent scale damage ensure that the wall mounted display is removed off the wall prior to scale transport e When wall mounting the display assembly ensure that the mounting location is free of electrical lines gas pipes and other safety hazards e When replacing inserting batteries be sure to use all new batteries 3 CAUTION AND WARNING e To prevent patient injury the patient must be attended throughout the entire weighing event Exercise caution when rolling patient onto and off of the scale e To prevent injury excise care when adjusting wheelchair ramps e To prevent injury and damage to your scale do not drop or mishandle scale while handling or transporting scale As this can permanently damage the load cells and may also cause injury to one self SPECIFICATIONS FOR THESE SCALES General The Model 2650KL and 2700KL Patient Scales use highly sophisticated microprocessor technology Each precision scale is designed to provide accurate reliable and repeatable weight measurements In addition each scale is d
20. latform upside down Caution Ensure that the wheelchair ramps are held in place when moving the scale ramps can flip open 2 Thread the 4 feet into the underside of the platform Each foot should be completely hand tightened into the platform Back feet out 1 full turns This will allow for proper load cell operation and scale accuracy 3 Carefully flip the scale over to its upright position on the floor Warning Do not drop scale as this can permanently damage the load cells 4 Adjust the feet to level the scale Be certain that the feet are not completely tightened into the platform for proper load cell operation 5 Open the scale to its operational position by folding open the wheelchair ramps 6 Connect the load cell cable to the scale platform by inserting the load cell cable connector into the load cell port see Figure 2 7 Ensure that the other load cell cable connector is securely plugged into the display assembly load cell jack located on the rear panel of the display assembly The load cell jack is labeled LOAD CELL see Figure 3 8 If the AC adapter will be used connect the AC adapter cord plug into the AC power jack located on the back of the display assembly The AC power jack is labeled POWER see Figure 3 9 Connect the AC adapter to the power port located on the back of the display assembly Caution Do not plug the scale into a power source until the scale is fully assembled 10
21. onal scales are warranted from date of purchase against defects of materials or in workmanship for a period of two 2 years If product fails to function properly return the product freight prepaid and properly packed to Pelstar LLC see To Get Warranty Service below for instructions If the manufacturer determines that a defect of material or in workmanship exists the customer s sole remedy will be repair or replacement of the scale at no charge Replacement will be made with a new or remanufactured product or component If the product is no longer available replacement may be made with a similar product of equal or greater value All parts including repaired and replaced parts are covered only for the original warranty period Who is Covered The original purchaser of the product must have proof of purchase to receive warranty service Please save your invoice or receipt Pelstar dealers or retail stores selling Pelstar products do not have the right to alter or modify or any way change the terms and conditions of this warranty What is Excluded Your warranty does not cover normal wear of parts or damage resulting from any of the following negligent use or misuse of the product use on improper voltage or current use contrary to the operating instructions abuse including tampering damage in transit or unauthorized repair or alternations Further the warranty does not cover natural disasters such as fire flood hurricanes and t
22. ornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To get Warranty Service make sure you keep your sales receipt or document showing proof of purchase Call 1 800 638 3722 or 1 708 377 0600 to receive a return authorization RA number which must be included on the return label Attach your proof of purchase to your defective product along with your name address daytime telephone number and description of the problem Carefully package the product and send with shipping and insurance prepaid to Pelstar LLC Attention R A Return Department 9500 West 55 Street McCook IL 60525 If your scale is not covered by warranty or has been damaged an estimate of repair costs or replacement costs will be provided to you for approval prior to servicing or replacing PELSTAR LLC 9500 West 55 Street McCook IL 60525 USA 1 800 638 3722 or 1 708 377 0600 PLEASE REGISTER YOUR SCALE FOR WARRANTY COVERAGE AT htto www homscales com customers registration aspx Health o meter is a registered trademark of Sunbeam Products Inc used under license Health o meter Professional products are manufactured designed and owned by Pelstar LLC We reserve the right to improve enhance or modify Health o meter Professional product features or specifications without notice Pelstar LLC
23. r scale please follow these instructions very carefully e Assemble and operate the scale per the enclosed user instructions e Operate these scales exclusively with the AC adapter provided by Health o meter Use of an unspecified adapter will void the warranty and can pose a serious safety hazard e Prior to using these scales inspect the AC adapter cord for cracking fraying or for broken bent plug prongs e Prior to using these scales ensure that the AC adapter is plugged into an outlet with the rated voltage appropriate for operation e Ensure that the AC adapter outlet is wired to a circuit breaker or other protected power source e Unplug the AC adapter and carefully store both the adapter and adapter cord before moving the scale e Ensure that the AC adapter and scale do not come in contact with liquids excessive temperature or excessive humidity e Operating these scales at voltages and frequencies other than specified can damage the equipment and will void the warranty e Do not transport the scale with any weight on the platform e Do not exceed recommended weight limit of 1000 Ib 454 kg for these scales e For accurate weighing these scales must be placed on a flat stable surface Also verify before each use the proper operation according to the procedure described in this manual e Do not use in the presence of flammable or explosive materials e For accurate weighing if the Lo bAtt indicator act
24. scale will begin to display weight information and will lock on the patient s weight when the stable weight of the patient is determined Depending on the movement of the patient on the scale it may take several seconds for the scale to lock onto the weight Caution To prevent patient injury the patient must be attended throughout the entire weighing event Exercise caution when rolling patient onto and off of the scale 5 Once a weight has been displayed while the patient is still on the scale Health o meter recommends performing a second weighing by pressing the REWEIGH button 6 The weight may be held on the display by pressing the HOLD RELEASE button To release press the HOLD RELEASE button again 7 After the weight measurement has been taken and is recorded roll the patient off the scale 8 Press the ON OFF button to turn the scale off Weight Recall The last weight reading on the scale can be recalled by pressing the RECALL button To clear the display of the recalled weight press the ENTER button Note Clearing the recalled weight from the display will not clear the recalled weight from memory The recalled weight will remain in memory until another weight is obtained 20 QUICK START OPERATING INSTRUCTIONS CONT BMI Function Body Mass Index is another name for Quetelet Index a formula that is a convenient indicator of BMI with various exemptions such as athletes children the elderly and the infirm

Download Pdf Manuals

image

Related Search

Related Contents

Moxa Managed Ethernet Switches User`s Manual  取扱説明書 ケーブル通信ユニット  The Redbox Range of Products  Neumann.Berlin SM 2 User's Manual  manuale di istruzione e manutenzione bwt duna ½”-¾  Boiler Gas BG 10 / Boiler Gas/Elektro BGE 10    Manual de instrucciones (PDF - 3,5Mb)  仕 様 書  LevelOne 8-Port Combo KVM Switch with Expansion Slot  