Home

Dräger Multi-PID 2 - Ribble Enviro Ltd

image

Contents

1. The microprocessor controls the components of the instrument and interprets and records the signal generated by the photoionization detector PID Recorded data and setup information entered into the microprocessor s memory are retained when the Multi PID 2 is turned off A pump continuously pulls the air under test through the Multi PID 2 s PID The Multi PID 2 converts the concentration of photoionizable compounds in the sample into an electrical signal The microproces sor subtracts any background from the signal and divides this signal by a sensitivity obtained by calibrating with a standard gas of known concentration This concentration appears on the Multi PID 2 s dis play and depending on the values entered through the Multi PID 2 s keypad an alarm status may be displayed and an audio signal may be heard The Multi PID 2 can detect thousands of different types of airborne gases and vapors and its response depends on the characteristics as well as the concentration of each compound The Multi PID 2 does not distinguish one type of compound from another but displays a number indicating the total concentration of all photoionizable compounds in the sample 20 00623767 04 eps Introduction A standard of isobutylene at a known concentration may be used for setting the sensitivity If the Multi PID 2 is calibrated with isobutylene it displays concentrations in units equivalent to ppm of isobutylene If isobutylen
2. Appendices Materials in sample stream Stainless steel Teflon Viton polypropy lene nitrile chorobutadiene rubber nickel Inlet filter Replaceable Teflon Polypropylene 1 um Inlet flow rate Greater than 300 mL min Operating temperature range 0 to 50 C 32 to 122 F I S Certified Temperature Range 0 to 40 C 32 to 105 F Operating humidity range O to 95 relative humidity non condensing Operating humidity range with optional humidity pre filter tube O to 1006 relative humidity non condensing Operating altitude range Up to 10 000 ft 3000 m Storage temperature range 10 to 65 C 14 to 149 F Storage humidity range O to 100 relative humidity non condensing Storage altitude range Up to 15 000 ft 4570 m Operating concentration range 0 1 to 10 000 ppm isobutylene Accuracy 10 or 2 ppm whichever is greater Precision 1 of calibration calibrated with 100 ppm isobutylene Response time Less than 3 seconds to 90 Detection limit 0 1 ppm isobutylene Dilution probe NOTE Normal calibration concentrations bet ween 100 to 20 000 ppm 20 High accuracy calibration concentrations between 100 to 1000 ppm 15 Concentrations between 1000 to 20 000 ppm 20 Specifications subject to change without notice 95 Appendices Installing Alternate AC
3. Electronic problem Contact Dr ger Safety for assistance 89 Troubleshooting Situation Cause Action Pump Error If the pump sounds labored the pump is operating out side its normal operating range Check for obstructions in the sample line Verify that the sample line sample probe and inlet filter are not blocked Ensure that the sample outlet is not blocked The Multi PID 2 has aspirated aliguid Contact Dr ger Safety The pump has failed Contact Dr ger Safety Blocked Filter Replace the inlet filter Dr ger Safety Part No 83 17 681 Instrument over range High concentrations of gases will saturate the detector and electronics Wait until the display returns to nor mal PIDs are designed to detect rela tively low levels of gases The detector has become saturated Move the Multi PID 2 to a location where it can sample clean air Sample clean air until the readings stabilize at a low level Detector has been short cir cuited by foreign matter in the cell Contact Dr ger Safety 90 General Ouestions Troubleshooting Fault Very low or no instrument response detected yet com pounds are known to be pre sent Possible Cause The Multi PID 2 has not been properl Remedy Verify the concentration of the span gas Calibrate the Multi PID 2 See page 73 Memory slots are not set up correctl
4. Calibration complete NOTE If you are not using the Dr ger Safety Flow Match Regulator ensure that the inlet pressure is less than 0 35 bar 5 psi when calibrating NOTE While the Cal status is active all alarms are deactivated 77 Routine Maintenance Battery Charging Afully charged battery pack powers the Multi PID 2 for approximately 8 hours If the instrument is to be used for more than 8 hours carry a spare battery pack Part No 83 17 670 When the first one has been discharged replace it with the spare NOTE If you do not turn the Multi PID 2 off before removing the battery pack you will reset the instrument and you will lose all logged data and setup parameters When the Multi PID 2 displays the low battery symbol the battery pack reguires charging When the low battery symbol is displayed you have 30 minutes of operation left The Multi PID 2 will turn itself off before the battery pack becomes critically low WARNING Do not remove or recharge the battery pack in a hazardous location Upon return from field work charge the battery packs as outlined in Battery Charging on page 26 Use only the AC adapter specified for use with the Multi PID 2 Optionally you can use the off line charger to charge the battery pack independently of the instrument See Off Line Charger on page 50 NOTE You must use the 220 V battery charger Part No 83 17 661 in order to comply with the requirements of the
5. 2 Usethe V key to select Switch to GAS mode resp Switch to TVOC mode then press select Unit Setup Menu Unit setup functions are used to select the Multi PID 2 features There are nine functions which can be set on the Multi PID 2 Pump Backlight User Mode Clock Date format Passcode Units and Sample Collection Figure 14 shows a menu detailing the User Setup functions Press the menu key in any operating mode to access Unit Setup Menu Functions 55 Menu Functions GAS mode Unit Setup User Mode Logging Off Units Sample Clock Date format PPM Backlight PPB Passcode Mg M3 Enter current time Enter current date MM DD YYYY DD MM YYYY on off Change Passcode or lock unlock instrument TVOC mode Unit Setup Pump off Backlight on User Mode Clock on Date format jof Passcode Units Logging off Sample Collection Tag Interval Enter current time Enter current date MM DD YYYY DD MM YYYY Change Passcode or lock unlock instrument PPM PPB Mg M3 Enter sample liters Figure 14 Unit Setup Map 56 _04 eps 01323792_04 Menu Functions Pump Only available in TVOC mode The function Pump turns the pump on and off The detector is turned off wh
6. The Sample Collection function is used to suck a defined air vol ume through the device e g to fill a preceding sampling tube sam pling tube with activated carbon 67 33 011 or a gas sampling tube with a defined air volume The sampling tube or the air collected using the gas sampling tube can be later analysed in a lab This process is menu driven The desired air volume in litres can be adjusted using the arrow keys It takes three minutes to take a sample of one litre In TVOC mode the Sample Collection function is in the Unit Setup menu In Gas mode this function is in the main menu 1 Press the menu key Highlight Unit Setup then press select Use the W key to highlight Sample Collection then press select 2 Use the arrow keys A and V to set the value 3 Press done to accept the set value Memory slots Menu Only available in TVOC mode General Information The response of a PID to a compound is dependent on its ionization potential In certain situations when the user is confident that only a single specific compound exists they can set up the Multi PID 2 using an isobutylene calibration to mimic the Multi PID 2 response as if the Multi PID 2 was calibrated using that specific single compound In situations where only a single pure compound is present in air the Multi PID 2 should be calibrated with a standard of that specific com pound as span gas The Multi PID 2 s 15 memo
7. The response factors were determined over the range 5 500 ppm based on a 100 ppm isobutylene calibration Isobutylene RF 1 0 The following formula is used for calculation of response factors Actual Concentration Multi PID 2 Response Response Factor A response factor less than 1 0 indicates a compound response bet ter than that of isobutylene A response factor greater than 1 0 indica tes a lower response than that of isobutylene NOTE It does not matter which response factor is entered The Multi PID 2 s response is not specific to any one compound The dis played reading represents the total concentration of all ioniz able compounds in the sample Table 5 Response Factors Compound Response Factor Acetaldehyde 10 5 Acetone 1 2 Acetophenone 2 0 Acrolein 2 Propenal 4 0 Acrylic Acid 10 9 Acrylonitrile ND 9 Allyl Chloride 3 Chloro 1 Propene 3 9 Benzene 0 5 Bromoform Tribromomethane 2 0 1 3 Butadiene 0 7 n Butane 0 2 n Butanol 3 4 97 Appendices Compound Response Factor n Butyl Acetate Butyl Acrylate Butyl Cellosolve n Butyl Acrylate n Butyl Mercaptan Butanthiol Carbon Disulfide Chlorobenzene Crotonaldehyde 2 Butenal Cumene Isopropylbenzene Cyclohexane Cyclohexanol Cyclohexanone Diacetone alcohol 1 2 Dichlorobenzene ortho cis 1 2 Dichloroethylene trans 1 2 Dichloroethylene N N Dimethylformamide DMF 1 4 Dioxane Epichlorohydrin Ethanol Ethyl
8. Humidity Tube function can be used in combination with a PID pre filter tube humidity Order No 81 03 531 to enhance the measuring performance by reducing the humidity This tube is used to measure the Total Volatile Organic Compounds value TVOC for 15 minutes filter capacity of the tube at 40 C and 100 96 relative humidity Then you are asked to check the tube and to replace it if necessary 1 Press menu then use the W key to highlight Humidity Tube then press select 70 Routine Maintenance Routine Maintenance 71 Routine Maintenance Maintenance Intervals The device should be calibrated before measuring especially if it was not used for a longer period of time If used the 11 7 eV lamp should be replaced approx every six months Calibration Technical Description Periodic calibration is required to compensate for the Multi PID 2 out put changes due to inlet filter restriction ionization chamber contami nation lamp cleanliness pump wear and other factors During calibration the Multi PID 2 is first exposed to zero air A small zero signal is generated This zero signal is stored by the micropro cessor The Multi PID 2 is next exposed to span gas This span gas signal is stored by the microprocessor The microprocessor subtracts the zero signal from the span gas signal and divides the difference by the user entered span gas concentration The resulting sensitivity is stored in
9. Charger eret tesi nie PARU o i Hx ns ii Eee 50 Menu FKE LOS eee i pp ii el le Erler 53 Usernterface Basic Menu sica casa einasi ka ai a eaa ia Pa vues a 54 Operation Modes Unit Setup Menu Memory slots MENU ee bean 65 Data Eog Optioris Menu iaia id 68 je NEU XUE 69 Edit Alarms Menu es 70 Humidity Tube Menus TM 70 Routine Maintenance dee i ene eie dec tne tn eed nene e pea iai 71 Maintenance Intervals ener arar reisen 72 Calibration Technical Description 72 Galibratiom Introduction sanpaolo 73 Calibration Using the Flow Match Regulator see en 76 Battery Charging cirie od ii init 78 Maintenance of the UV Lamp iui 80 Replacing the Sample Inlet Filter nnns 84 Waste electrical and electronic equipment nennen 85 Meets 3p ks 87 General Information ai citri ER i oed En EE ERE REOR ENTRA eR ER rn 88 Troubleshooting General Questions ABPENGICSS Lure esee ctp a EUR ORAE nice i ORE EUR ME SKIES a i a o ees anne S 93 Specification EE PE 94 Installing Alternate AC Plug on the Battery Charger u 96 Galibration Gas Supplier iC re i 96 Presets and Response Factors ree arc narco rana nnne nnne nennen 97 Reference T As 103 Figures Figure T M lt PID 2 2 Xe eben Ene I 18 Figure 2 Multi PID 2 Block Diagram oooco
10. Graphic Display The Multi PID 2 uses an 8 line graphic display The display will always be used for reporting detected concentration In order to accommo date the range of concentrations the Multi PID 2 can detect the meter reading will be reported using one of two resolutions A resolu tion of 0 1 ppm will be used for concentrations below 100 ppm and a resolution of 1 ppm will be used for concentrations above 100 ppm ALARM e e CHARGE Dr ger Multi PID 2 9 3 2006 view clear menu 01123792 04 eps Figure 5 Multi PID 2 Display The display reports instantaneous concentration at all times when the pump is on In Interval mode the display will report instantaneous concentrations as well as PEAK STEL or TWA The Multi PID 2 is designed for ease of use with a logically organized internal menu structure user interface The Multi PID 2 User Menu is shown in Figures 6 and 7 The Multi PID 2 has three soft keys under the graphic LCD display which always show the available functions of the soft keys in any Screen 29 Using the Multi PID 2 Multi PID 2 User Menu Unit Setup Switch to TUOC mode Data Log Options D select Unit Setup Switch to GAS mode Memory slots v select GAS mode Unit Setup User Mode Logging Off Switch to TVOC mode Units Sample Data Log Options Clock PPM Clear values Date format PPB Edit Alarms Backlight Sample Collectio
11. In this mode the soft keys view sample and menu are active Pressing the view key shows the latest measured values stored in the Datalogger Pressing the sample key starts a new benzene measurement menu selects the Multi PID 2 s internal menu for the Multi PID 2 setup by the user 58 Menu Functions TAG Tag mode is identified by the word TAG in the upper right hand corner of the Multi PID 2 display Tag mode will continuously display the instantaneous concentration of total volatile compounds Tag mode also allows the user to manually tag and datalog readings Tag mode allows the user to datalog a background reading a sample rea ding and assign Site Codes to readings Site Codes are set using the Dr ger GasVision software See Connecting Accessories on page 43 for more information on how to use the Dr ger GasVision software In Tag mode the soft keys are view tag and menu Activating tag mode 1 Press the menu key and select Unit Setup then press select Select User Mode then press select 2 Select tag then press select In this mode the soft keys view tag and menu are active view selects datalogger review tag selects No Label Background or Sample menu selects the Multi PID 2 s internal menu for the Multi PID 2 setup by the user Interval Interval displays the instantaneous rea
12. MSDS enn 24 Maxim maccuracy cxtat uhstieplca en ERICH DII E E e TRIS DA 33 Memory Slots doit a nei e bae reete ev nor eel 66 Numerio Value oce esee a aa aka d Dee Ris 36 Operation viii dai lu Raw ew ISI PUE ne DU Mur ELE Pel x eO a asi RN EA 33 Operation Modes ita iio eee a e RA Re Bun eR rx npa 54 Overview Lii iai asi de See SUO T PII edes FPES 18 Passcode o vos asas out O Aa x 64 PEAK MO0G 5 ses v Pub oe nb RA AD a ET i du EE TA 39 PEAK modernas Reip ee a IU OPER EY Dade DUM a 60 Photoionization Detector sz Liss aaa ai na ne baa ET DIE 21 Photoionization detector PID u seen I ERU OPEP RE E IE Ha ree uu 20 Preparing for Field Operation LL 40 Pump Lk esie mc he PP WU or anb E Eq Hd pater wee pee I 57 Regulators for Compressed Gases eee eee 74 Removing Battery Pack siii arte 3 al ea C Re x ia 26 Replacing the Battery Pack 14 esos ass Hr 26 Response Factors tisi rur RENS EL ae as E RO ERROR REN E Pris E RR EO REL 66 Response time iuis iini REA E USUS eua edu bed ie a eidem a 95 Safety Practices wi es ane ni en ka ARI a Siri ern 17 Set the Clock zur copa id dad trae ded an da bapti d 62 Set the time and date nenne 63 Setting the Time Cocco co platani ee ee LOVER aaa 62 106 Short term exposure limit STEL ssessssssssse Ie 19 Short Term Exposure Limit STEL Mode sese 39 Soft Keys MC 32 STEL caloulation ra er Rn a ak Ba A 60 STEL MOJ
13. The following values are displayed benzene concentration in xx ppm STEL value the highest benzene measurement value recorded so far SMAX the amount of measurements already carried out SAMP corresponds with the amount of PID pre filter tube benzene already used and the sampling interval period of 90 seconds SECS Pressing the view key shows the Datalogger measurement values Pressing the menu key takes you back to the selection menu Pressing the sample key starts the measurement the pump is acti vated and the measured air is drawn through the PID pre filter tube benzene for 90 seconds The measurement status is indicated by the flashing concentration value and by the flashing word sampling The number 90 displayed under SECS counts down to 0 So the dis played number shows the remaining measuring time SAMPLE DONE signalled in the display indicates that the measurement is fin ished The PID pre filter tube benzene is a one shot device and cannot be used for a second measurement Repeatedly pressing the sample key will call up the information that the PID pre filter tube benzene in use is no good and that a new PID pre filter tube benzene must be put into the pre filter tube holder Only after inserting a new tube will pressing the sample key start a new measurement 54 TVOC mode In TVOC mode the Total Volatile Organic Compounds are detected Changing the Operating Mode 1 Press the menu key
14. 4 7 Hydrogen Sulfide H2S 3 3 Isoamyl Acetate IAMYACET 1 8 Isobutyl Acetate IBUTACET 2 6 Isobutyraldehyde 9 IBUTALDE 14 Isopentane IPENTANE 8 2 Isoprene 2 Methyl 1 3 Butadiene ISOPRENE 0 6 Isopropanol IPA 4 4 Appendices 101 Appendices Compound Code RF Isopropyl Acetate IPACETAT 2 6 Isopropyl Ether IPROPETH 0 8 Methyl Bromide Bromomethane MeBROM 1 6 Methyl Ethyl Ketone MEK 0 8 Methyl Isobutyl Ketone MIBK 1 0 Methyl Mercaptan Methanethiol METHMERC 0 5 Methyl Methacrylate MeMeACRY 1 4 Methyl tert Butyl Ether MTBE MTBE 0 8 Monomethylamine MMeAMINE 1 3 n Nonane nNONANE 1 4 Iso Octane 2 2 4 Trimethylpentane IOCTANE 1 2 n Pentane nPENTANE 10 4 n Propanol nPA 51 Propionaldehyde Propanal 5 PROPANAL 14 8 n Propyl Acetate nPROACET 3 1 Propylene PROPYLEN 1 2 Propylene Oxide PROPOXID 5 8 Styrene STYRENE 0 4 Tetrachloroethylene PCE PCE 0 5 Tetrahydrofuran THF 1 5 Toluene TOLUENE 0 5 Trichloroethylene TCE 0 5 Trimethylamine TRMeAMIN 0 9 Vinyl Acetate VINACET 1 2 Vinyl Bromide VINBROM 0 4 Vinyl Chloride Chloroethylene VINCHLOR 1 7 Vinylidene Chloride 1 1 DCE 1 1 DCE 0 8 meta Xylene mXYLENE 0 5 ortho Xylene oXYLENE 0 5 para Xylene pXYLENE 0 5 102 NOTES 1 Peak alarm levels have been established as the TLV Ceiling concentration or the TLV STEL concentration in those cases where no TLV Ceiling value exists In those cases where no STEL exists for a compound the
15. 4th edition July 2006 Subject to alteration
16. Acetate Ethyl Acrylate Ethylbenzene Ethyl Cellosolve 2 Ethoxyethanol Ethylene Glykol Ethyl Ether Diethyl Ether Ethyl Mercaptan Ethanethiol Ethylene n Heptane n Hexane 2 Heptanone Hydrogen Sulfide 2 3 1 8 3 1 1 8 0 6 1 3 0 4 1 2 0 6 1 3 3 4 0 9 1 8 0 5 0 8 0 4 0 8 1 3 6 5 8 8 3 8 2 3 0 5 1 3 0 1 1 2 0 6 10 1 2 4 4 7 24 3 8 98 Compound Response Factor Isoamyl Acetate Isobutyl Acetate Isobutyl Alcohol Isobutyraldehyde Isopentane Isoprene 2 Methyl 1 3 Butadiene Isopropanol Isopropyl Acetate Isopropyl Ether Mercaptopropionic Acid Methacrylic Acid Methyl n Amyl Ketone 2 heptanone Methyl Bromide Bromomethane Methyl Chloroform 1 1 1 TCA Methylene Chloride Methyl Ethyl Ketone 2 Butanone Methyl Isobutyl Ketone Methyl Mercaptan Methyl Methacrylate Methyl tert Butyl Ether MTBE Monomethylamine n Nonane iso Octane 2 2 4 Trimethylpentane n Pentane Polypropylene Propane n Propanol Propionaldehyde Propanal n Propyl Acetate Propylene Propylene Oxide Styrene 1 8 2 6 0 3 14 8 2 0 6 44 2 6 0 8 04 000 21 1 6 000 4 000 4 0 8 1 0 0 5 14 0 8 1 8 14 12 10 4 0 8 000 5 1 14 8 3 1 1 2 5 8 0 4 Appendices 99 Appendices Compound Response Factor Syltherm XLT 9 8 Tetrachloroethylene Perchloroethylene 0 5 Tetrahydrofuran 1 5 Therminol 000 4 Toluene 0 5 1 1 1 Trichloro
17. Lamp Driver Circuit UV Lamp High Voltage Plates Light Sensor Electrometer 125 Volt Detector Cell Sample In Collector Electrode M hn gt Mt e Repeller Electrode lonized Molecule z x9 ee gt Sample M Figure 3 Photoionization Detector The ionized molecules in the detector cell are subjected to a conti nuous electric field between the repeller electrode and the collector electrode The ions move in the electric field generating a current which is proportional to the concentration of the ionized molecules in the detector cell An electrometer circuit converts the current to a voltage which is then fed to the microprocessor The UV lamp is operated by a high voltage lamp driver circuit which delivers high voltage energy to the lamp through plates in the lamp holder The lamp driver power supply is controlled by the micropro cessor based on a feedback signal from a light sensor on the driver circuit board 22 00723767_04 eps Introduction Unpacking Instrument Remove the Multi PID 2 from its shipping box The following accesso ries are included with the Multi PID 2 Sample Probe Instrument Manual Multi Tool AC Adapter or AC Adapter with AC Line Cord Wrist Strap Replacement Sample Inlet Filters 25 pieces Reference Card NOORON gt Ensure that all of these accessories have been included with the instrument If any items are missing or damaged
18. STEL value has been established as eguivalent to the TWA value Compound is Not Detectable ND by the Multi PID 2 Please contact Dr ger Safety Applications Department for further details Response factor for this compound using the 10 6 eV lamp is equal to 0 00 because the 11 7 eV lamp is required to io nize and detect this compound Please contact Dr ger Safety Applications Department for further details In cases where recommended exposure limits are below the detection limit for the compound in question the estimated lower limit of detection has been substituted for those valu es The Multi PID 2 is not suitable for monitoring these com pounds at ACGIH recommended levels A 1000 ppm TWA has been established for those com pounds which are Simple Asphyxiants and for which no ex posure value exists For those compounds which lack established exposure le vels an arbitrary value of 200 ppm has been established for the TWA STEL and PEAK alarms Reference 1 2 Maslansky Carol J and Steven P Maslansky Air Monitoring In strumentation New York Van Nostrand Reinhold 1993 American Conference of Governmental Industrial Hygienists Threshold Limit Values for Chemical Substances and Physical Agents and Biological Exposure Indices 1994 1995 Reprinted with permission NOTE The TLV BEI publication is revised annually 3 National Institute for Occupational Safety and Health NIOSH Recommendations for O
19. Securite Intrins que Class I Division 1 Groups A B C and D Maximum surface temperature 135 C T4 Model Multi PID 2 d 1 0158 x Il 2G TRL OSATEX21031X EEx ib IIC T4 Ta 0 C to 40 C Dr ger Safety D23560 L beck Germany WARNING Substitution of components may impair intrinsic safe Avertissement La substitution de composants peut compromettre la securite intrinsique To prevent ignition of flammable or combustible atmospheres disconnect power before servicing CAUTION To reduce the risk of explosion only use AC Adapter and Serial Port or recharge the batteries outside of the hazardous location Avertissement Afin de prevenir l inflammation d atmosph res dangereuses ne charger les batteries que dans des emplacements 3 designes non dangereux Use only Dr ger battery pack g Part No 8317670 3 11 Notices and Warnings WARNING 1 All calibration maintenance and servicing of this device including battery charging must be performed in a safe area away from hazar dous locations 2 Disconnect all power before servicing 3 Do not open UV Lamp Cap when unit is energized 4 Only use the AC Adapter in a safe area away from hazardous loca tions 5 Only use the Serial Port in a safe area away from hazardous locati ons CAUTION To reduce the risk of fire or injury to persons read and follow these instructions 1 There are no operator replaceable parts inside the Multi
20. and running on the PC prior to download To download data 1 Press the menu key 2 Usethe V key to choose Data Logger Options the press the select key 3 Use the V key to choose Download to PC then press the select key 4 Press the next key at the Connect Instrument to PC prompt 5 The display will show Downloading Data and data will now download to the PC Press the done key to stop the download at any time Clear values Menu Only available in GAS mode The Clear values function can be used to clear the latest measure ment values for the highest measured concentration SMAX the lat est STEL value or all values ALL which are shown in the display 1 Press menu then use the W key to highlight Clear values then press select 2 Selecting the values and pressing the clear key clear the high lighted values 69 Menu Functions Edit Alarms Menu Only available in GAS mode The Edit Alarms function can be used to edit the latest set alarm values for the STEL and the REAL alarm 1 Press menu then use the W key to highlight Edit Alarms then press select 2 Usethe W key to highlight STEL alarm or REAL alarm then press select 3 Usethe A W and B keys to enter the desired value 5 Highlight the checkmark then press done Humidity Tube Menu Only available in TVOC mode The
21. applicable Council Directives If you do not require portable operation you can use the Multi PID 2 while it is connected to the AC adapter WARNING The Multi PID 2 is not intrinsically safe when con nected to an AC adapter The AC adapter automatically charges at a high charge rate until the battery pack is fully charged It then maintains the full charge with a low continuous charge rate indefinitely so there is no danger of over charging When the Charge LED on the top of the Multi PID 2 is red the battery is charging When the Charge LED turns green the battery is fully charged 78 Routine Maintenance Battery Pack Care Leaving the Multi PID 2 for more than 3 months without charging the batteries may result in the loss of recorded data and setup parame ters If the Multi PID 2 is not used for long periods of time recharge the battery for a few hours every 30 days to avoid loss of data See Battery Charging on page 26 for instructions on charging the battery Please observe the following For replacement battery pack use only Part No 83 17 670 Do not dispose of the battery pack in a fire The cell may explode The battery pack is 24 cadmium Cd by weight This battery pack must be disposed of properly Check with local codes for special disposal instructions Do not open or mutilate the battery pack Exercise care in handling battery packs in order not to short the terminals with conducting materials such a
22. contact Drager Safety immediately 23 Introduction Support Eguipment and Consumables Calibration For normal operation these items are required 1 Calibration Gas Regulator Part No 68 10 688 2 Calibration gas containing 100 ppm isobutylene Part No 68 10 687 Other concentrations of the calibration gas may be re quired This will depend on your application 3 Zero air clean dry air without any organic impurities There are several alternatives for clean or zero air calibration you can use a bottle of clean air certified as having not more than 0 1 ppm total hydrocarbons connected directly to the instrument the clean air can be transferred to a Tedlar bag which can then be connected to the instrument clean ambient air without detectable contamination or ambient air run through the charcoal filter Part No CH 24 101 4 If compound threshold limit values TLVs are exceeded you should use a gas bag for sampling and calibration To determine the TLV of the compounds contained in the calibra tion gas refer to the Material Safety Data Sheet MSDS supplied with your calibration gas cylinder Field Operation For field operation the Multi PID 2 Field Kit Part No 83 17 663 is available The field kit includes a cable kit a carrying case and a calibration regulator and a spare battery pack Refer to the check list in on page 41 to ensure you have all the neces sary accessories and equipment before b
23. correct concentra tion and calibrate the Multi PID 2 as described on page 74 Memory Slots have not been properly programmed Program the memory slots See page 66 Detector has been short cir cuited by foreign material Contact Dr ger Safety Date and time setting are not retained The battery pack was remo ved before the Multi PID 2 was turned off Replace the battery pack and reset the time and date See page 26 The Multi PID 2 has not been used for 3 months or more Connect the Multi PID 2 to the battery charger and recharge the battery Reset the date and time Display is blank Battery pack is critically low Recharge or replace the bat tery pack The battery pack is not con nected to the Multi PID 2 pro perly Ensure the battery pack con nector is securely attached to the Multi PID 2 Sample flow is less than 300 ml min Inlet filter is plugged Replace the inlet filter See page 84 Inlet filter is incorrectly instal led Ensure the inlet filter is instal led correctly See page 84 Sample outlet is blocked Clear blockage from sample outlet Pump has been damaged Contact Dr ger Safety Liquid has been aspirated The Multi PID 2 has been exposed to a solvent that can pass through the inlet filter Contact Dr ger Safety Very corrosive gases have been sampled Contact Dr ger Safety Cannot download
24. loca tion 88 Troubleshooting Troubleshooting Situation Cause Action Zero Air Error Contamination of sample lines sample probe inlet fil ter or fittings before the detector Clean or replace contaminated item Span gas and zero air are mixed up Ensure that clean air is used to zero the Multi PID 2s Ambient air is contaminated If you are not sure of the cleanliness of the ambient air use a commercial zero grade air to zero the Multi PID 2 Span Gas Error Span gas and zero air are mixed up Ensure that clean air is used to zero the Multi PID 2 Span gas concentration is too low Span gas concentration should be higher than 25 ppm UV lamp window is dirty Clean the lamp window See page 82 UV lamp is failing Install a new lamp See Removing and Replacing the UV Lamp on page 80 The concentration and sam ple gas are not com patible with the Multi PID 2 Contact Dr ger Safety UV Lamp Error UV lamp has not started immediately This fault may occur momentarily when the Multi PID 2 is first turned on Allow 60 seconds for UV lamp to start and fault to clear Turn the Multi PID 2 off then on again Wait 60 seconds for UV lamp to start UV lamp is not installed Install UV lamp See Removing and Replacing the UV Lamp on page 80 Incorrect lamp is installed Use only a getterless UV lamp 83 18 307 or 83 18 317
25. the selected Memory Slots with the zero signal and the alarm levels This number is then multiplied by the response factor and displayed The Multi PID 2 readings are always relative to the calibration gas After calibration with isobutylene the Multi PID 2 will respond directly in units equivalent to isobutylene Most volatile organic compounds will be detected by the Multi PID 2 It cannot distinguish between isobutylene and other ionizable com pounds A reading of 10 ppm indicates all ionizable compounds that are present have generated an ion current proportional to 10 ppm of isobutylene The reading is actually 10 ppm isobutylene equivalent units The Multi PID 2 readings give an indication of the total ionizab les present and their concentration relative to the calibration gas 72 Routine Maintenance Calibration Introduction The Multi PID 2 must be calibrated in order to display concentration in ppm units eguivalent to the calibration gas First a supply of zero air which contains no ionizable gases or vapors is used to set the Multi PID 2 s zero point Then a calibration gas containing a known concentration of a photoionizable gas or vapor is used to set the sen sitivity Clean indoor ambient air may be substituted for a cylinder of zero air Due to the Multi PID 2 s sensitivity outdoor air is usually unsuitable for calibration unless the activated charcoal filter Part No CH 24 101 and filter tube holder Part No 83
26. unsure about the status of the battery replace the battery pack with one that is fully charged See battery charging on page 26 2 After calibration is complete sample the calibration gas and the bag of zero air to ensure the Multi PID 2 has been calibrated cor rectly 3 Selectthe correct operating mode See Section regarding Opera tion Modes page 54 Using the Multi PID 2 41 Using the Multi PID 2 42 Connecting Accessories Connecting Accessories 43 Connecting Accessories Computer The Multi PID 2 will download information stored in its datalogger to a Windows based computer This option may be used if you need to prepare reports based on the Multi PID 2 s recorded data This fea ture may also be used if you need the recorded data in a format that can be imported into a spreadsheet or database for further calculati ons NOTE The Multi PID 2 is not classified for use in hazardous locations when connected to any computer The instructions below will provide you with the basic information for using the Multi PID 2 with the Multi PID 2 communications software the Multi PID 2 Comm In order to use these instructions you must be familiar with Microsoft Windows and it must be installed and running on your computer To initiate communications between the Multi PID 2 and a PC 1 Turn the Multi PID 2 off NOTE You must turn the instrument off before connecting or discon necting the computer cable 2 Th
27. 00 For Your Safety For Your Safety Strictly follow the Instructions for Use Any use of this instrument reguires a full understanding and strict adherence to these instructions This instrument is only to be used for the purposes specified here see Intended Use on page 18 Maintenance The Multi PID 2 must be inspected and serviced by trained service personnel at regular intervals Repair of the Multi PID 2 may only be carried out by trained service personnel Only authentic Dr ger Safety spare parts may be used for mainte nance Observe chapter Maintenance Intervals on page 72 We recommend that a service contract be obtained with the service of Dr ger Safety and that all repairs are also carried out by them Use in areas subject to explosion hazards This instrument and its components have been tested and approved according to the European Directives 94 9 EEC ATEX Directive and 89 336 EEC The electromagnetic Compatibility Directive and may be used only under the conditions explicitly specified in the EC Declaration of Conformity of August 28 2003 see page 11 Modifications of components or the use of faulty or incomplete parts is not permitted When making repairs to equipment or components of this type the relevant national regulations must be adhered to Liability for proper functioning or damage The liability for the proper functioning of the instrument is irrevocably transferred to the owner or operator t
28. 19 093 is used Commercially sourced zero air should not have more than 0 1 ppm total hydrocarbons THC To determine the TLV of the compounds contained in the calibration gas refer to the Material Safety Data Sheet MSDS supplied with your calibration gas cylinder NOTE Disconnect the Multi PID 2 from the AC adapter before begin ning calibration 738 Routine Maintenance Compressed Gases Cylinders of compressed gas such as calibration gas must be hand led with extreme care For safety the calibration gas cylinders must be secured before use Please observe the following handling procedures Mark each new regulator with its intended gas service and never use a regulator for more than one service To ensure safety and avoid contamination regulators should be dedicated to one ser vice Do not change gas service or adapt equipment without con sulting your gas supplier Do not heat or expose cylinders or regulators to temperatures abo ve 52 C 125 F The cylinders may rupture at high temperatures Use only the specified regulator for the calibration gas Confirm re gulator type and material with your specialty gas supplier Always secure oylinders before removing the cylinder valve protec tion cap Do not drag or roll cylinders Large cylinders should only be mo ved on carts designed for compressed gas cylinders Do not move cylinders without the valve protection cap in place Wear safety glasses when working wi
29. Dr ger Dr ger Multi PID 2 Portable Photoionization Monitor Instrument User manual Software Version 3 nn Contents For Your Safety 3e ii Rois 5 Notices and Warnings endete idad 7 EGG Warning a ras EE 8 The Multi PID 2 Intrinsic Safety I S Notice iii 8 ATEX Directive and EMC Directive i 11 WARNING i E A i ia aaa ias a a S a Ai Aa saaa 12 o AAA na eene a E RIO AA 15 Aboutthis Manuals tiziana een ANAIS 16 Warnings and Safety Practices Approved Models of the Multi PID 2 e Excessive Heat and Cold a 17 Intended Use 18 Multi PID 2 Overview 18 IC CHROM iais ao sia a A Ks i a i sa asa 20 Photoionization Detector umi is 21 Unpacking Instrument Liss siais ere eode a a aaa 23 Support Equipment and Consumables e 24 Using the Multi PID anei 25 s uc dire 26 DisSpla tas ect P EC BR ERA EIE Lr 29 Multi PID 2 User Menu 30 Keys ian 32 Beginning Operation 33 Preparing for Field Operation ii 40 Connecting Accessories ii ee re eet Aiden ee dei i i a TE 43 COMPU ot S DIE UO De DE S as 44 Pre filter Tube Holder rnnt nent ttt nnnc nens 45 Pre filters and Sample Collection Tubes i 46 aH M EE 46 Wrist MP iii nre HT nant REOR Ad RR FEE 47 DC Power Cord 47 Belt Clip Holster 47 11 7 eV UV Lamp 48 Off Line
30. L mode the STEL PEAK and TWA are calculated If you select INTERVAL mode these values are automatically recorded in the Multi PID 2 s memory The Multi PID 2 can log up to 12 000 ent ries In TAG mode operation the Multi PID 2 prompts you to locate a site and then to record a background and sample readings for the site You can record up to 12 000 manual entries There is no averaging of data in TAG mode Tags are set via the PC using the Dr ger GasVi sion software Recorded data can be reviewed on the display or downloaded to a computer Data are recorded by date and time The Multi PID 2 has 5 keys for instrument operation The keys are used to set up and calibrate the Multi PID 2 All information entered with the keys and stored in the Multi PID 2 s memory is retained when the instrument is switched off The clock and calendar continue to operate and do not need to be reset when the Multi PID 2 is turned on 19 Introduction General Operation The Multi PID 2 is a microprocessor controlled air monitor for measu ring the presence of photoionizable compounds in air at parts per mil lion levels The block diagram in Figure 2 shows the main components of the Multi PID 2 Sample In Sample Out Inlet Photoionization gt Filter Detector Pump Power Display and Supply Microprocessor I O Connector Keypad Electrocal Connections gt gt Gas Connections Figure 2 Multi PID 2 Block Diagram
31. O iii A e ode LIRA 60 Support EguipM nt liti EE ri a AR LEE EE DELE 24 SystemiAlents I oes demise aa uei uude A i ea 37 TAG Mode Display us s a Ra et vila k age a at 35 Threshold limit values TLVS LL 24 Uu stess dl Aaa ARA di Anksti aa bak ai ieee aa i ea 36 Time weighted average TWA 0 0 cece ccc eee eee eee eee eee eens 19 Time Weighted Average TWA Mode 0 cc cece He 39 Turning the M lt PID 2 On stash deus ia a oa Cob Eb a ed Feat t ae 33 TWA accumulator 1 32 22 genie ep EDO aie eR eor buscas ee er teen es 39 TWA mode xni battle GERBERA bad NUI eae Cr eR E due Fette de 60 Unit Setup Fungtions u u xe er ae ai EI eere dri eee iais 55 Unit Set p Map ois sa ita Nas A EA m idt O ardere 56 Units ya setae eer esee ene la dou tree te bee DAR verdes de el DA 64 Unpacking Instrument 32 cose EIL fe Re a 23 User Interface o2 n ra ia ei DE MUR Ue der d eee 54 User Menu Limit ere e ole Aon A RR Qe eee a e a eee Roe EUR dn 29 30 User Menu GAS mode esee 30 User Menu TVOC mode hae 31 User Mode citi La er 58 Using tree S hi Incolla dona NN ais 25 Warning Practices essc sa 17 Zero all licia troie eppure E bee er Kai ia ie 24 73 107 108 109 110 111 Drager Safety AG amp Co KGaA RevalstraBe 1 D 23560 L beck Germany Tel 49 451882 0 Fax 49 4518 82 20 80 www draeger com 90 23 767 GA 4635 301 en Dr ger Safety AG amp Co KGaA
32. PID 2 ex cept the battery pack UV lamp and sample inlet filter 2 Forreplacement battery pack use only Dr ger Part No 83 17 670 3 There are no operator serviceable parts inside the Multi PID 2 4 Do not dispose of the battery pack in a fire The cells may explode The battery pack must be disposed of properly Check with local co des for possible special disposal instructions 5 Do not open or mutilate the battery pack If the Multi PID 2 is used in a manner not specified the protection provided by the Multi PID 2 may be impaired 6 Exercise care in handling battery packs in order not to short the ter minals with conducting materials such as rings bracelets and keys The battery or conductor may overheat and cause burns 7 Do not defeat proper polarity orientation between the battery pack and battery charger 8 Charge the battery pack using the AC adapter provided with or iden tified for use with this product only in accordance with the instruc tions and limitations specified in this manual For AC adapter use only Part No 64 05 428 115 Volt AC Part No 83 17 661 220 Volt AC When using the AC adapter do not block access to AC outlet in use with adapter AC adapter is not to be used in a ha zardous area 12 These optional accessories may be used with the Multi PID 2 in a hazardous location Name and Discription Calibration Regulator Wrist Strap Belt Clip Holster Carrying Case User s Reference Ca
33. Plug on the Battery Charger In most cases the Multi PID 2 will be shipped with an AC line cord that will fit the AC wall outlet in your area If this cannot be done you may need to obtain an AC line cord suitable for the AC receptacle in your area The AC line cord attached plug and receptacle must be marked with your country s certification mark and the cord must have a Harmoniza tion HAR mark The line cord must be rated for either 100 to 120 VAC at 60 Hz or 220 to 240 VAC at 50 Hz The voltage rating will depend on the voltage in your area Contact your Dr ger Safety representative to obtain more information Calibration Gas Supplier The recommended span gas is isobutylene in air 100 ppm isobuty lene in air may be obtained from Dr ger Safety The exact concentration will be determined by your application Other concentrations and other gases may be obtained from your local gas supplier If you cannot locate a local gas supplier contact Dr ger Safety s Technical Support Department 96 Appendices Presets and Response Factors Presets simplify Memory Slot programming and provide standard response factors and alarm levels for approximately 100 compounds The name response factor and three alarm levels are all set from the preset You can change any of the values entered in the Memory Slots Chan ges made to the library information that has been loaded into a Preset will have no effect on the original library entry
34. adings 17 Introduction Intended Use The Multi PID 2 measures the concentration of airborne gases and vapors that can be ionized by a photoionization detector The Multi PID 2 automatically displays and can record these concen trations In the TVOC mode the Multi PID 2 does not distinguish between individual compounds The reading displayed represents the total concentration of all pho toionizable chemicals present in the sample The Multi PID 2 is fac tory set to display concentration in units of ppm Benzene is selectively measured in GAS mode This measurement requires the use of the PID pre filter tube benzene The pre filter tube adsorbs all substances with the exception of ben zene Only benzene passes into the detector Multi PID 2 Overview Figure 1 Multi PID 2 18 04 ops 00523792 04 6 Introduction The Multi PID 2 operates automatically The Multi PID 2 display updates itself once per second You can read concentrations directly from the display The Multi PID 2 will perform short term exposure limit STEL time weighted average TWA and PEAK calculations when it is in INTER VAL mode You can view any of these results in INTERVAL mode The Multi PID 2 has two datalogging options TAG and INTERVAL mode TAG mode allows the user to manually tag and store readings during a walkthrough Interval mode allows the user to datalog at sel ectable intervals of 1 second to 999 seconds In INTERVA
35. alarm press the accept key Once acknowledged the alarm indi cators are cleared The alarm status will remain until the alarm condi tion clears The Multi PID 2 updates the instantaneous concentration once every second Following every update the instantaneous concentration is compared to the peak alarm level and if exceeded an alarm is triggered In Interval mode if a 15 minute average exceeds the STEL a STEL alarm is generated The TWA alarm is generated when the current average of concentration since the TWA was last cleared has excee ded the TWA exposure limit During calibration all alarms are disabled Once the calibration is complete the alarms are re enabled 38 Using the Multi PID 2 Short Term Exposure Limit STEL Mode The Short Term Exposure Limit STEL mode displays the concentra tion as a 15 minute moving average The Multi PID 2 maintains 15 samples each representing a one minute averaging interval Once every minute the oldest of the 15 samples is replaced with a new one minute average This moving average provides a 15 minute average of the last 15 minutes with a one minute update rate Since the average is calculated using 15 one minute averages the meter display will only update once every minute STEL is set to zero each time the instrument is turned on Since STEL is a 15 minute moving average there is no need to clear or reset the STEL STEL calculations are always being performed by
36. allel port Do not connect the serial port of the converter to a parallel port Pre filter Tube Holder The optional pre filter tube holder Order No 83 19 093 is shown in figure 12 The pre filter tube holder is designed to hold the optional charcoal pre filter tube Order No CH 24 101 for calibration in clean air the optional PID pre filter tube humidity Order No 81 03 531 to enhance the measuring performance in TVOC mode by reducing the humidity the optional sampling tube with activated carbon Order No 67 33 011 and the PID pre filter tube benzene Order No 81 03 511 for selective benzene measurement in GAS mode The tube holder attaches to the Multi PID 2 on the inlet nozzle repla cing the standard inlet nozzle cap It is a simple twist lock application po Do SL ce DR k ES NS Figure 12 Tube Holder To insert a pre filter tube into the tube holder 1 Unscrew the top part of the tube holder immediately above the tube window 2 Break off each end of the glass tube by inserting the end into the opening as shown Insert tube into holder 4 Re screw on the top part of the tube holder note the tube holder is designed to compensate for any minor irregularities in the tube that result from the breaking off of each end Connecting Accessories 02523792 04 eps 45 Connecting Accessories To remove a pre filter tube from the tube holder 1 Unscrew the top part of t
37. and either connect the Multi PID 2 to zero air connect the optional activated charcoal filter Part No CH 24 101 and filter tube holder Part No 83 19 093 or leave the instrument unconnected so it can sample clean ambi ent air Then press the next key The Multi PID 2 will take 60 90 seconds to set the zero point for calibration 4 The Multi PID 2 display will show Zero air calibrated Continue calibration Press the next key 5 Enterthe span gas concentration if the concentration displayed on the Multi PID 2 is different than the concentration of the span gas cylinder Press the new key and follow the display prompts For further information see page 36 for numeric value duration time and date entry If the span concentration matches the span gas cy linder concentration press the next key To exit calibration press the cancel key 6 Ensurethe calibration gas cylinder is upright and open the regula tor by turning the valve counter clockwise Open the regulator until the ball is 1 8 from its rest position 7 Connectthe span gas to the Multi PID 2 if the zero air calibration was done with either a Zero Air cylinder or the optional activated charcoal tube and tube holder either should be disconnected be fore connecting the span gas fitting to the Multi PID 2 8 The Multi PID 2 will take 60 90 seconds to set the span set point When finished with the calibration the Multi PID 2 will display
38. ations by selecting STEL as the Display mode In PEAK mode the reading is updated once a second In the back ground the Multi PID 2 datalogger is sampling the concentration and measuring min max and average concentrations for the selected averaging interval At the end of every interval one entry is placed in the datalogger until the datalogger is full TWA mode sums concentrations every second until 8 hours of data have been accumulated Once 8 hours of data have been summed the accumulation stops This sum will only be complete after 8 hours so the meter displays the current sum divided by 8 hours While you are in TWA mode the time on the status display will show the number of minutes and hours of data that TWA has accumulated When this reaches 8 hours the Multi PID 2 stops accumulating data and the TWA is complete TWA calculations are always being performed by the Multi PID 2 You can display the results of the calculations by selecting TWA as the Display mode Interval mode logs readings at user selected intervals of 1 second to 999 seconds 60 Menu Functions ALARM e e CHARGE Urdger Multi PID 2 Intervals 00323792_04 eps Figure 15 Intervall time adjustment Interval simultaneously displays the STEL TWA and PEAK reading along with the instantaneous reading The Multi PID 2 s Datalogger can store up to 15000 entries The interval you select will determine the perio
39. ccncncccononccononccononananonnnnnnonnncnnnnnn cnn nan nnnnnnn cn nnnnnnannnns 20 Figure 3 Photoionization Detector ii 22 Figure 4 Battery Pack Removal and Replacement i 27 Figure 5 Multi PID 2 Display iii 29 Figure 6 User Menu GAS mode 30 Figure 7 User Menu TVOC mode 31 Figure 8 GAS mode Display 33 Figure 9 Logging Off Mode Display ie 34 Figure 10 TAG Mode Display u ie 35 Figure 11 Interval Mode Display ii 35 Figure 12 Tube Holder iii a en 45 Figure 13 Dr ger Pre filter Tubes i 46 Figure 14 Unit Setup Map eere rri arat de tr As spada ee tema Eu 56 Figure 15 Intervall time adjustment iii 61 Figure 16 Setting the Time iii 62 Figure 17 Function Memory Slots Settings i 66 Figure 18 Calibration with Flow Math seeeeeennnnn nennen 76 Figure 19 Removing the UV Lamp eene sasas 81 Figure 20 Replacing the Inlet Filter i 85 Tables Table 1 Table 2 Table 3 Table 4 Table5 Table 6 SM AIS CE M Check List for Field Items Additional Field Items Averaging Intervals and Period STEL TWA and PEAK Operation u 61 Response ACTOS o eter ias a ii Ss riii ui fia eere aaa 97 Library Entries e 1
40. ccupational Safety and Health Compen dium of Policy Documents and Statements January 1992 Appendices 103 Appendices 104 Index AG adapter 2 2 2 u iais ia a BOUM d LI o S lhi badd 27 73 Dro D n 9 Alda A A a oie 37 Approved Models tacita dd quee dati 17 ATEX DIFeGtV aos ia a a Rena pi den 11 Averaging Interval hen nenne ere xh KS eben nee nee 61 B cklight L a iO ENS RA S aa Aa REI eve hate tartrate P 57 Basic Menu susce aaa dete Area nn ER Ruten learn nn ala pra a 54 Battery Charging ui iii s aaa ade ERE NER HORE a d a ale WG aso en 26 Beginning Operation Li 33 Cable kit iii er Re ans ee 24 Calibrating with the Flow Match Regulator 0 cece cece 77 Calibration i melon oi debit died lasted a a is es 24 Calibration Gas seb feat dk A ede a ew ee lke eee 75 Calibration Dr RD 24 Calibration Gas Regulator nnn 24 Calibration Introduction LL 73 Calibration Technical Description LL 72 Calibration Using the Flow Match Regulator Li 76 Charging the Battery Pack LL 27 Check List for Field Items Li 40 Clock T 62 Compressed Gases iii i Rex err e 74 Computer siusis a Sua as m e e de a PE ORE a oi 24 Connecting the Flow Match Regulator seesssss nn 76 GCons mabl s Lis csse gs dete ete e A Nu rS ee 24 Date cuu ne E 36 Date Format san RE DAA eS odes 63 Daten l schen ede ius ais ed Ca
41. conds The measurement status is indicated by the flashing concentration value and by the flashing word sampling The number 90 displayed under SECS counts down to 0 So the dis played number shows the remaining measuring time SAMPLE DONE signalled in the display indicates that the measurement is fin ished The PID pre filter tube benzene is a one shot device and cannot be used for a second measurement Repeatedly pressing the sample key will call up the information that the PID pre filter tube benzene in use is no good and that a new PID pre filter tube benzene must be put into the pre filter tube holder Only after inserting a new tube will pressing the sample key start a new measurement TVOC mode The Multi PID 2 can power up in Logging Off Tag or Interval mode depending on the mode that was set by the previous user The current mode is shown in the upper right hand corner of the display e e CHARGE Multi PID 2 9 3 2006 Figure 9 Logging Off Mode Display 34 _04 eps 00923792 04 e 9 3 2006 tag 9 3 2006 clear view CHARGE Multi PID 2 menu Using the Multi PID 2 01023792 04 eps 02723792 04 eps Figure 11 Interval Mode Display 35 Using the Multi PID 2 Numeric Value Duration Time and Date Entry In cases where the system reguires the user to enter a number dura tion time
42. d P soft keys enter the passcode Press the B soft key until the Check mark is highlighted and select the done soft key NOTE If you change the passcode make sure you record the setting If you cannot remember your passcode once the instrument is lok ked there is no way to unlock it without losing all your data log contents Unlocking the instrument without a password is accomplished by doing a reset or cold start To do the reset hold the middle upper key while turning the instrument on with the on off key Units The Units function can be used to select the unit of the displayed measurement value PPM parts per million PPB parts per billion ana Mg M3 milligrams per cubic meter 1 Press the menu key Highlight Unit Setup then press select Use the W key to highlight Units then press select 2 Select either PPM PPB or Mg M3 then press select 64 Menu Functions NOTE To utilize the Mg M3 unit of measure the instrument must be configured and operated with a compound chosen from the Multi PID 2 s Preset RF library because a molecular weight value is needed for the Mg M3 calculation If you wish to use the Mg M3 unit of measure while doing TVOC work you should use a memory slot with Isobutylene selected from the Preset RF library Refer to Section Response Factors and Memory Slots for further information Sample Collection
43. d of operation Atthe end of the period the datalogger will be full and you will see the Data Log Full status Table 4 Averaging Intervals and Period STEL TWA and PEAK Operation Averaging Interval Hours of Operation to example Fill the Datalogger 15s 50 1m 200 2m 400 5m 1000 15m 3000 The Multi PID 2 s display can show four values in Interval mode STEL TWA and PEAK plus the instantaneous reading Activating Interval mode 1 Press the menu key and select Unit Setup then press select Select User Mode then press select 2 Select Interval then press select 61 Menu Functions In Interval mode the soft keys are view clear and menu view selects datalogger review clear shows the selection options TWA PEAK or All Use the V key to select one of the selection options then press clear to delete the stored data of the corresponding option menu selects the Multi PID 2 s internal menu for the Multi PID 2 setup by the user Clock The function Clock is used to set both the current date and time Entering Numbers with the Soft Keys to Set the Clock For all information that you must enter the left center and right soft keys correspond to the W A und P arrow Figure 16 Set ting the Time on page 62 The W und A keys are used to change the character highlighted by the cursor The P key is u
44. data or communicate with PC Cable is not connected Connect the Multi PID 2 to the PC See page 24 92 Appendices Appendices 93 Appendices Specification Size 9 230 mm long x 3 80 mm deep x 4 25 110 mm wide at display tapering to 2 6 67 mm at handle Weight 1 9 pounds 0 86 kg Detector Instant on photoionization detector with standard 10 6eV UV lamp optional 11 7 eV lamp available Keypad 5 keys with tactile feedback Display 8 line LCD display for alphanumeric rea douts and soft key display Datalogger memory 200 hours at 1 minute sample interval Serial output RS 232 9600 baud 8 data bits with no parity for connection to a Windows M com patible computer Audio output 85 decibels on Alarm Inlet connection 1 8 8 175 mm fitting Battery type Nickel cadmium rechargeable cell with intel ligent charger The battery pack is 2496 cad mium Cd by weight Charge discharge time 4 hr 8 hr Input Power 12 VDC 1 0A Battery charger Automatically charges and maintains full charge in battery pack U S and Canada Part No 64 05 428 Input 120 VAC 60 Hz 24 W Output 12 VDC 1000 mA Regulatory Approvals CSA UL All other Countries Part No 83 17 660 Input 100 250 VAC 50 60 Hz 0 5 0 3 A Output 12 VDC 2 08A Regulatory Approvals ATEX UL 94
45. dings as well as STEL TWA and PEAK readings Interval mode is identified by the letters INT in the upper right hand corner of the Multi PID 2 display Interval auto matically calculates and updates STEL TWA and PEAK readings Interval mode also automatically stores these readings in the Multi PID 2 s memory at a preset interval selected by the user In inter val mode data is logged at a user selected interval value between 1 second and 999 seconds The microprocessor accumulates all readings in an averaging interval that you select and determines the minimum average and maximum readings It stores these numbers along with the highest priority instrument status and the most recent time and date These recorded data can now be reviewed and edited Recorded data can also be printed using the PC For each averaging interval the Multi PID 2 prints the minimum of all the minima the average of all the readings for the interval and the maximum of all the maxima 59 Menu Functions In STEL mode 15 samples are combined to form a 15 minute aver age Once every minute the oldest of the 15 samples is replaced with a new one minute average This moving average provides a 15 minute average with a one minute update rate so the meter display will only update once every minute STEL is set to zero each time the instru ment is turned on STEL calculations are always being performed by the Multi PID 2 You can display the results of the calcul
46. e Maintenance 81 Routine Maintenance Cleaning the UV Lamp Window During the course of normal operation a film builds up on the window of the UV lamp The rate at which the film develops depends on the type and concentration of the gases and vapors being sampled and results from the UV light interacting with them Hot gases and vapors may contribute to a decrease in sensitivity because they may condense on the lamp window Condensation may eventually evaporate off the window but it will usually leave a residue that must be removed by cleaning the lamp window Since some UV inhibiting deposits are invisible to the eye regular cle aning of the lamp window is strongly recommended Clean the lamp frequently to prevent a heavy buildup of deposits that may be difficult to remove WARNING Do not remove the UV lamp in a hazardous location The UV lamp cleaning kit Part No 83 18 318 includes aluminum oxide powder 3 0 micron powder and cotton swabs Additionally you will require approximately 10 mL of methanol When ordering methanol specify purge and trap grade The instructions provided below are for 10 6 eV lamps Refer to Cle aning the Lithium Fluoride Window for special instructions to clean the 11 7 eV lamp window WARNING Never touch the UV lamp window with your fingers 82 To clean the UV lamp 1 2 3 Remove the lamp as outlined in Removing and Replacing the UV Lamp on page 80 Allow the lamp to co
47. e Multi PID 2 must be connected to a serial port Use the com puter cable Part No 83 17 667 to connect the Multi PID 2 to one of the computer s serial ports Remember which serial port you are using as you will need this information when you set up the Dr ger GasVision software Normally you will use Com1 or Com2 3 If the Multi PID 2 printer cable plugs directly into the port on the computer without the gender changer or the 9 to 25 pin adapter you are most probably connected to a parallel port You will need at least one of the adapter cables to connect the Multi PID 2 to a serial port Once all connections have been made turn on the Multi PID 2 instrument 4 Startthe Dr ger GasVision software as you would any Windows program The Dr ger GasVision splash screen will appear on the PC Once the Dr ger GasVision software is running the PC is rea dy to receive data from the Multi PID 2 5 Onthe Multi PID 2 the number of data bits has been fixed at 8 stop bits has been fixed at 1 Parity has been set at None and the Flow control is Xon Xoff These values are set automatically by the Dr ger GasVision software e 44 6 On the Multi PID 2 press the menu key choose Data Log Options then press the select key 7 Choose the Download to PC option using the W key and press the select key 8 Follow the Multi PID 2 prompts to start the data download NOTE A DB 25 female connector is sometimes a par
48. e socket on the off line charger Replace the battery pack in the Multi PID 2 as outlined in Battery Charging on page 26 You can keep the battery pack fully charged indefinitely without over charging it by leaving it connected to the off line charger while the charger is operating Connecting Accessories 51 Connecting Accessories 52 Menu Functions Menu Functions 53 Menu Functions User Interface Basic Menu The Multi PID 2 is designed for ease of use with a logically organized internal menu structure user interface The Multi PID 2 User Menu is shown in Figures 6 and 7 on page 30 and page 31 The Multi PID 2 has three soft keys under the graphic LCD display which always show the available functions of the soft keys in any screen Operation Modes GAS mode The GAS mode is a filter tube mode This means that all measure ments reguire the installation of the pre filter tube holder and of the pre filter tubes In this selective measuring mode the PID pre filter tube benzene which is mounted at the head end of the pre filter tube holder and adsorbs all substances with the exception of benzene is used to selectively measure benzene The entire process is menu driven The selective benzene measurement value is displayed after 90 seconds Activating the GAS mode opens a dialog which queries a device cali bration The benzene measurement screen is displayed after installa tion of the PID pre filter tube benzene
49. e were the only photoionizable chemical in the sample then the Multi PID 2 would display its concentration directly The Multi PID 2 responds more or less readily to other chemicals than it does to isobutylene Because it has a medium sensitivity to isobuty lene this gas has been chosen as a reliable means of reporting an average concentration of total ionizable compounds present Gases other than isobutylene can be used to calibrate the Multi PID 2 However all response factors are based on an isobutylene calibration Photoionization Detector The photoionization detector PID is shown in Figure 3 The PID mea sures the concentration of photoionizable chemicals in the gas stream from the sample inlet and produces an electrical signal for the microprocessor A UV lamp generates photons which ionize specific molecules in the gas stream The permanent air gases argon carbon dioxide nitro gen oxygen water vapor etc require a relatively high energy for ioni zation and are not ionized by the UV photons Many of the compounds considered pollutants including most hydrocarbons are ionized The gas stream is directed into the PID through a small port at the center of the UV lamp window and through a series of larger ports around the perimeter of the lamp window This arrangement permits a high sample flow rate and short response time 21 Introduction Lamp Power Supply Sample Out Lampholder
50. eginning field operation Computer The Multi PID 2 may also be connected to a computer The computer must be a Windows based PC Use the cable kit Part No 83 17 667 to connect the Multi PID 2 to the computer Stored datas are evaluated using the Dr ger GasVision software Part No 83 14 034 WARNING The Multi PID 2 is not classified for use in hazar dous locations when connected to a computer danger of an explosion 24 Using the Multi PID 2 Using the Multi PID 2 25 Using the Multi PID 2 Battery Charging Before beginning operation of the Multi PID 2 the battery pack must be charged You can also remove the battery pack and replace it with a fully char ged spare battery pack Part No 83 17 670 NOTE You must use the 220 V battery charger Part No 83 17 661 in order to comply with the requirements of the applicable Council Directives Removing and Replacing the Battery Pack WARNING Do not remove or recharge the battery pack in a hazardous location danger of an explosion To remove the battery pack 1 If the Multi PID 2 has been turned on turn it off by pressing the ON OFF key for five seconds and then releasing it NOTE If you do not turn the Multi PID 2 off before removing the battery pack you will reset the instrument and you will lose all logged data and setup parameters 2 Locate the battery hatch on the back of the instrument See Figure 4 3 Loosen the two Phillips screws
51. en you turn the pump off This prevents the detector from being damaged when there is no sample flowing through the detector When the pump and the detector are off the meter display will conti nue to read normally but the instantaneous reading is at 0 0 Turn the pump and detector off when concentration measurements are not necessary and the Multi PID 2 will only be used for setup or reviewing data By operating the instrument with the pump and detec tor off when you do not need them you will conserve the battery and ultraviolet UV lamp To turn the pump on 1 Press the menu soft key Unit Setup press select Pump press select on then press select To turn the pump off 2 Press the menu soft key Unit Setup press select Pump press select off then press select Backlight The function Backlight is used to switch the backlighting on and off when there is insufficient light to read the display Switching on the backlight 1 Press the menu key Select Unit Setup then press select Select Backlight then press select Select on then press select Switching off the backlight 2 Press the menu key Select Unit Setup then press select Select Backlight then press select select off then press select To extend the operating life of the battery pack turn the backlighting off when it is not requ
52. er on to the Multi PID 2 as well as the turn the power off To turn on the Multi PID 2 press the key To turn the power off press the key and hold it down for 5 seconds and then release it This is done to prevent accidental power off Soft Keys The three soft keys on the Multi PID 2 are located directly below the display Each key has varying functions for configuring the Multi PID 2 editing the data and controlling the display Since only three soft keys are available each function is broken down into a path 32 Beginning Operation Turning the Multi PID 2 On 1 Turn the Multi PID 2 on by pressing the key 2 The Multi PID 2 will display the instrument s software version num ber Next the Multi PID 2 will proceed to the mode display 3 The Multi PID 2 has an instant on lamp For maximum accuracy and stability allow the Multi PID 2 to warm up for 10 minutes prior to calibration Default Display The display shows the last mode the Multi PID 2 used The resolution of the display changes with the magnitude of the reading A reading of 0 to 99 9 will be displayed with a resolution of 0 1 ppm A reading greater than 99 9 will be shown with a resolution of 1 ppm The meter will display concentrations up to 10 000 ppm GAS mode In GAS mode the PID pre filter tube benzene Order No 81 03 511 can be used to measure benzene or a sample collection tube Order No 67 33 011 can be filled with a defined amount of a
53. et A B 2 C D 3 select E G edit H I or J N select 15 O S T Z edit Name I Enter new Name RRF Enter new RRF STEL alarm Enter new STEL TWA alarm Enter new TWA Display selected slot REAL alarm Enter new REAL settings Formula Wt Enter new FW I done Figure 17 Function Memory Slots Settings 66 _04 eps 01523767 04 e Menu Functions Response factors built into the Multi PID 2 are used to correct for the response of the PID to a specific compound These correction factors are often called response factors RF The Multi PID 2 has 15 memory slots Each memory slot can store one preset RF Library RF selections simplify Memory Slot programming and provide standard response factors for approximately 100 compounds The selection allows you to select an entry from a pre programmed library The name Name response factor RRF the three alert levels STEL TWA and REAL and the formula FWt are all set from the library To select a preset to program the selected Memory Slot 1 Press menu then use the W key to highlight Memory Slots then press select To select a memory slot Press the W key until the memory slot you wish to use is high lighted Press the select key The name of the compound RRF STEL TWA REAL and FWt are displayed for a few seconds for this memory slot Press done to return t
54. ethane 000 4 Triethanolamine ND 9 Trichloroethylene TCE 0 5 Trimethylamine 0 9 1 2 4 Trimethyl Benzene 2 3 1 3 5 Trimethyl Benzen 1 7 Vinyl Acetate 1 2 Vinyl Bromide 0 4 Vinyl Chloride Chloroethylene 1 7 Vinylidene Chloride 1 1 DCE 0 8 meta Xylene 0 5 ortho Xylene 0 5 para Xylene 0 5 Table 6 Library Entries Compound Code RF Acetaldehyde ACETAL 10 5 Acetone ACETONE 1 2 Acrolein 2 Propenal 5 ACROLEIN 4 0 Allyl Chloride 3 Chloro 1 Propene ALLCHLOR 3 9 Benzene BENZENE 0 5 Bromoform Tribromomethane 5 BROMFORM 2 0 1 3 Butadiene 13BUTADI 0 7 n Butanol nBUTANOL 3 4 100 Compound Code RF n Butyl Acetate nBUTACET 2 3 n Butyl Acrylate nBUTACRY 1 8 n Butyl Mercaptan Butanethiol 9 nBUTMERC 0 6 Carbon Disulfide CS2 1 3 Chlorobenzene CHLOBENZ 0 4 Crotonaldehyde 2 Butenal CROTONAL 1 2 Cumene Isopropylbenzene CUMENE 0 6 Cyclohexane CYCHEXAN 1 3 Cyclohexanone CYCHEXON 0 9 1 2 Dichlorobenzene ortho 12DCBENZ 0 5 cis 1 2 Dichloroethylene cis12DCE 0 8 trans 1 2 Dichloroethylene trn12DCE 0 4 N N Dimethylformamide DMF N N DMF 0 8 1 4 Dioxane DIOXANE 1 3 Epichlorohydrin 5 EPICLHYD 6 5 Ethanol ETHANOL 8 8 Ethyl Acetate ETHYACET 3 8 Ethyl Acrylate ETHYACRY 2 3 Ethylbenzene ETBENZEN 0 5 Ethyl Cellosolve 2 Ethoxyethanol ETHCELLO 1 3 Ethyl Ether Diethyl Ether ETHETHER 1 2 Ethyl Mercaptan Ethanethiol 9 ETHMERC 0 6 Ethylene ETHYLENE 10 1 n Heptane nHEPTANE 2 4 n Hexane nHEXANE
55. field to reduce or eliminate down time of the instrument If you are going to be in the field for a single 8 10 hour day then you should include the following accessories Table 2 Check List for Field Items y Spare battery pack Part No 83 17 670 Y Spare 10 6eV lamp Part No 83 18 307 Y Multi PID 2 multi tool Part No 64 05 415 y Short sample probe Part No 64 05 412 y Calibration regulator Part No 68 10 688 Y Tank s of calibration gas Part No 68 10 687 Y Spare inlet filters Part No 83 17 681 y Carrying case Part No 83 17 664 y DC power cord Part No 64 05 421 y Tube Holder Part No 83 19 093 y Dr ger Charcoal Tube Carbon filter Part No CH 24 101 for zero point calibration y Drager Tube PID Humidity Filter Part No 81 03 531 Sampling Tube Activated Charcoal Type BIA Part No 67 33 011 PID Pre Filter Tube Benzene Part No 81 03 511 y Instrument manual Part No 90 23 767 If you will be in the field for more than one day you should include the following additional items Table 3 Additional Field Items y AC adapter Part No 83 17 661 y Cable kit Part No 83 17 667 y Computer and associated cables 40 Operational Check List Before beginning field work set up and calibrate the Multi PID 2 for your particular application To ensure the instrument is in working order before heading into the field 1 Ensure the battery pack is fully charged If you are
56. he tube holder immediately above the tube window 2 Liftthe pre filter tube from the tube holder 3 Either insert another tube or re attach the top part of the tube hol der Pre filters and Sample Collection Tubes Dr ger Safety provides a variety of pre filter tubes for the Multi PID 2 The tubes are custom developed for Dr ger Safety AG amp Co and address clean air calibration and humidity reduction to facilitate more accurate sampling The optional sample collection tubes are standard Dr ger tubes used for collecting VOC samples for analysis later in a lab or other analytic facility Figure 13 Drager Pre filter Tubes Sample Line A 3 meter 9 sample line Part No 64 05 411 may be connected to the Multi PID 2 for remote sampling Connect the sample line to the Multi PID 2 inlet using the fittings supplied with the sample line NOTE When using the sample line be especially careful not to aspirate liquids or solids as they will damage the Multi PID 2 46 04 eps 02623792 04 e Connecting Accessories Wrist Strap To use the wrist strap Turn off the Multi PID 2 and then remove the battery cover See Removing and Replacing the Battery Pack on page 26 e Place the metal ring of the wrist strap in the circular notch located on the battery door Replace the battery cover Adjust the strap length as necessary DC Power Cord The Multi PID 2 can be connected to a car ba
57. in the top of the battery hatch 4 The battery hatch can now be removed b Liftthe battery pack out of the case and carefully disconnect the battery pack connector from the Multi PID 2 6 Attach the connector from the charged battery pack to the Multi PID 2 NOTE The connector is polarized It will only fit one way Do not force the connection 7 Place the battery pack in the Multi PID 2 case Ensure the battery wires are not pinched or strained 8 Ensurethe wrist strap ring is in proper position This ring holds the wrist strap in place 9 Replacethe battery hatch and then replace the two screws Do not overtighten the screws as you will damage the case 26 Using the Multi PID 2 Charging the Battery Pack NOTE Only use the AC adapter specified for use with the Multi PID 2 Using another AC adapter will result in damage to the battery pack the Multi PID 2 or the adapter itself To charge the battery pack 1 2 Plug the AC adapter into the jack located on the bottom of the Multi PID 2 Plug the AC adapter into an AC outlet If you are using the Euro pean AC adapter ensure the correct plug is installed on the line cord If it is not correct for the wall outlet in your area then it must be replaced The Charge LED on the Multi PID 2 indicates the charge state Red indicates the battery is being charged Green indicates the battery is fully charged and ready for use Itis normal for a fully charged batte
58. ired 57 Menu Functions User Mode The User Mode function selects one of the offered logging modes The logging modes are In GAS mode Logging off and sample In TVOC mode Logging off Tag and Interval Logging off Logging Off mode is identified by the word LOG with a diagonal line through the word LOG in the upper right corner of the Multi PID 2 display Logging Off will continuously display the concen tration of total volatile compounds present that the Multi PID 2 can ionize The reading is updated approximately once per second In Logging Off mode the only soft key displayed is menu menu selects the Multi PID 2 s internal menu for the instrument setup by the user Activating Logging off 1 Press the menu key and select Unit Setup then press select Select User Mode then press select 2 Select Logging Off then press select In this mode the only soft key displayed is the menu key menu selects the Multi PID 2 s internal menu for the Multi PID 2 setup by the user Sample Only available in GAS mode This function starts the measurement with the PID pre filter tube ben zene benzene measurement See also page 33 and page 54 Activating sample mode 1 Press the menu key and select Unit Setup then press select Select User Mode then press select 2 Select sample then press select
59. ith your calibration gas cylinder 75 Routine Maintenance Calibration Using the Flow Match Regulator Connecting the Flow Match Regulator to the Cylinder WARNING Observe proper handling procedure for all gases To connect the flow match regulator to the gas cylinder 1 Connect the regulator to the calibration gas cylinder 2 If you are using a portable tank of calibration gas Part No 68 10 687 connect the regulator Part No 68 10 688 directly to the tank 3 When the regulator is connected properly you can read the cylin der contents from the regulator gauge Connect the adapter tubing to the regulator Check if the Multi PID 2 is sucking in the gas pressure free e g by means of a T piece paying attention the fact that the flow from the gas oylinder is higher than the inlet flow of the Multi PID 2 ss Calibration cancel new next Figure 18 Calibration with Flow Math 76 04 eps 01623792 04 ej Routine Maintenance Calibrating the Multi PID 2 with the Flow Match Regulator 1 Ensure that the short sample probe is connected to the Multi PID 2 inlet If you are using the long probe for sampling then en sure the long probe is connected to the Multi PID 2 NOTE Ensure the sample probe is free of any contamination as this will affect the calibration 2 Press the CA key 3 Follow the instructions on the display
60. ith the dilution probe Limitations of Lithium Fluoride Lamp Window This lamp is intended for special applications only It is not suitable for normal operation due to limitations of the lamp window material The 11 7 eV lamp window material is lithium fluoride LiF LiF is com posed of two light elements which are easily disrupted within the cry stal lattice by the UV light generated by the lamp Disruption of the lattice causes the crystal to turn a yellowish color and again perfor mance declines Unlike other lamp windows LiF readily absorbs water from atmosphe ric humidity When contaminated by moisture the window loses its ability to transmit UV light NOTE Do not touch the lamp window or handle it near water Using the 11 7 eV UV Lamp With an 11 7 eV lamp installed your instrument functions as a detec tor responding to compounds which ionize at 11 7 eV or less The 11 7 eV lamp may be useful for detecting compounds not ionized by the standard 10 6 eV lamp Due to the lamp window limitations the lifetime of the 11 7 eV lamp is restricted and it must be used sparingly according to the following instructions 48 Connecting Accessories To use the 11 7 eV lamp in your the Multi PID 2 1 Remove the 11 7 eV lamp from the supplied desiccant bottle and install the lamp as outlined in Removing and Replacing the UV Lamp on page 80 NOTE Do not remove or replace any detector lamp in a hazardous location 2 T
61. ll result in damage to the battery pack the off line charger or the adapter Charging from an AC Source NOTE Do not charge the battery pack in a hazardous location To re charge your battery pack 1 Remove the battery pack as outlined in Battery Charging on page 26 2 Attach the connector from the battery pack to the socket on the off line charger NOTE The connector is polarized It will only fit one way Do not force the connection 3 Plug the AC adapter into the jack on the opposite face of the off line charger 4 Plug the AC adapter into an AC outlet If you are using the Euro pean AC adapter ensure the correct plug is installed on the line cord If it is not correct for the wall outlet in your area then it must be replaced 5 The LED on the upper face of the off line charger indicates the charge state of the battery pack Red indicates the battery is being charged Green indicates the battery is fully charged 6 Charging a fully discharged battery pack will take approximately four hours 50 9 10 It is normal for a fully charged battery pack to indicate it is charging red light when first plugged in The LED will turn green within a few minutes to indicate the battery is fully charged When the battery pack is fully charged remove the AC adapter first from the wall outlet then from the off line charger Remove the battery pack connector from the socket on the off line charger Replace the ba
62. n Passcode Clear data log Download to PC SMAX STEL ALL Enter current time Enter current date MM DD YYYY DD MM YYYY on off Change Passcode or lock unlock instrument STEL alarm REAL alarm Enter new STEL Enter new REAL Enter sample liters Figure 6 User Menu GAS mode 30 en eps 02823792 04 en e Using the Multi PID 2 Unit Setup C Switch to GAS mode Memory slots bs duc Unit Setup Switch to GAS mode Memory slots Data Log Options Humidity Tube Pump Backlight User Mode Clock Date format Passcode Units Sample Collection Unit Setup Switch to TUOC mode Data Log Options v select Logging off Tag Interval Enter current time Enter current date MM DD YYYY DD MM YYYY Change Passcode or lock unlock instrument PPM PPB Mg M3 Clear data log Edit interval Download to PC Attach Tube holder and insert new tube Enter sample liters Figure 7 User Menu TVOC mode 31 02923792 04 eneps Using the Multi PID 2 Keys Fixed Keys The Multi PID 2 has two fixed keys The first fixed key is the key on off The second fixed key is the dedicated CAD key calibration key Pressing the CAD key will start calibration in almost any mode The O key is used to both turn pow
63. nter the correct date 7 Press the done soft key to confirm the date and the display will return to the main screen Date Format Use the Date format function to select the displayed date format MM DD YYYY or DD MM YYYY Selecting a format 1 Press the menu key and select Unit Setup then press select and select Date format then press select and select date format then press select 63 Menu Functions Passcode The Passcode function can be used to change the passcode as well as to activate or de activate the menu function block The lock function is used to prevent access to the Multi PID 2 s sensitive options Sensitive options are those which can affect the Multi PID 2 s readings If any of these functions is selected when the instrument is locked the Multi PID 2 will prompt you to unlock the instrument before you can access the functions The unlock function is used to obtain access to the Multi PID 2 s sensitive options The change function can be used to change the set passcode The unit always has a passcode in force When shipped from the fac tory the code is 0000 1 Press the menu soft key Highlight Unit Setup and press the select soft key Use the W soft key to highlight Passcode the press the select soft key 2 Choose either unlock change or lock soft keys You will be prompted for a passcode Using the A V an
64. nts may affect safety rating CAUTION To reduce the risk of fire or injury to persons read and follow these instructions 1 All calibration maintenance and servicing of this device Part No 68 10 688 83 17 673 83 17 677 83 17 664 90 23 770 64 05 411 64 05 412 83 19 093 CH 24 101 67 33 011 81 03 511 81 03 531 including battery charging must be performed in a safe area away from hazardous locations Disconnect all power before servicing 2 There are no operator replaceable parts inside the Multi PID 2 except the battery pack UV lamp and sample inlet filter 3 There are no operator serviceable parts inside the Multi PID 2 Notices and Warnings Notices and Warnings WARNING 1 2 10 For replacement battery pack use only Part No 83 17 670 Do not dispose of the battery pack in a fire The cells may explode The battery pack must be disposed of properly Check with local codes for possible special disposal instruc tions Do not open or mutilate the battery pack If the Multi PID 2 is used in a manner not specified the protection provided by the Multi PID 2 may be impaired Exercise care in handling battery packs in order not to short the terminals with conducting materials such as rings bra celets and keys The battery or conductor may overheat and cause burns Do not defeat proper polarity orientation between the bat tery pack and battery charger Charging the batte
65. o the extent that the instrument is improperly serviced or repaired by personnel not employed or authorized by Dr ger Safety or if the instrument is used in a manner not conforming to its intended use Dr ger Safety cannot be held responsible for damage caused by non compliance with the recommendations given above The warranty and liability provisions of the terms of sale and delivery of Dr ger Safety are likewise not modified by the recommendations given above Dr ger Safety AG amp Co KGaA For Your Safety Copyright Information This document contains proprietary information that is protected by copyright All rights are reserved No part of this publication may be reproduced in any form whatsoever or translated into any language without the prior written permission of Dr ger Safety Trademarks Registered names trademarks etc used in this document even when not specifically marked as such are protected by law Notices and Warnings Notices and Warnings Notices and Warnings FCC Warning This instrument has been tested and found to comply with the limits for a Class B Digital Device pursuant to Subpart B Class B of Part 15 of the FCC rules These limits are designed to provide reasonable protection against harmful interference when the eguipment is opera ted in a commercial environment This eguipment generates uses and can radiate radio freguency energy and if not installed and used in accordance with the instr
66. o the main menu To edit a memory slot 1 Press menu then use the W key to highlight Memory Slots then press select 2 Use the W key to select the memory slot then press select to display the memory slot contents Press done to return to the main menu or 3 Use the W key to select the memory slot then press edit to change the values for RRF STEL TWA REAL and FWt 4 Press edit then highlight either Name RRF STEL TWA REAL or FWt using the W key then press select 5 Usethe A W and B keys to enter the desired number 6 Highlight the checkmark then press done 7 Pressthe done key to return to the main display or press select to edit another value 8 Toreturn to the main menu press select then done 67 Menu Functions To select a pre programmed compound stored in the Multi PID 2 s Preset RRF library 1 Press menu then use the W key to highlight Memory Slots then press select 2 Usethe V key to select the memory slot then press edit to display the memory slot contents 3 Press preset 4 Usethe V key to choose the alpha range of the compound of interest then press the select key 6 Choose the compound of interest using the W key and press the select key 7 Press the back key to accept the choice or press edit to modify the preset values for RRF STEL TWA REAL and FWt as sh
67. ol to room temperature to avoid thermal shock which could crack the window Dampen a cotton swab in methanol and dab it into the cleaning compound Allow a small amount of the cleaning compound to ad here to the cotton tip WARNING Do not use methanol to clean the 11 7 eV lamp win dow Using small circular motions rub the cleaning compound onto the lamp window Do not exert excessive force onto the window since the window can be permanently damaged by misuse Continue cleaning for approximately one minute Dampen a new cotton swab with methanol and remove any trace amounts of the cleaning compound from the lamp window Wipe the window and the lamp with a dry lint free tissue to remove any trace amounts of the cleaning compound and methanol All tra ces of cleaning compound must be removed If you removed an O ring from the lamp replace it now Ensure your hands are clean and dry Without touching the lamp window replace the lamp as outlined in Removing and Replacing the UV Lamp on page 80 Cleaning the Lithium Fluoride Window To clean the 11 7 eV lamp window 1 Clean the lamp window with dry aluminum oxide powder on a dry cotton swab Do not use methanol or water 2 Wipethe window and the lamp with a dry lint free tissue to remove any trace amounts of the cleaning compound and methanol All tra ces of cleaning compound must be removed 3 The lamp window can also be regenerated by storing the lamp in a desiccant for a
68. or date the following mechanism is used The number of digits to be entered depends on the type of value being entered Upon entering a value entry screen a bar icon below the left most digit highlights it as the active digit The arrows A middle soft but ton and V left soft button are used to increase decrease the digit The bar icon is moved to the next digit to the right using the B soft button Once the user has finished entering the value they use the right soft button to move the bar icon to the far right to highlight the OK check mark y When the bar is under the OK check mark V the left soft button changes to done The user is then able to input the value Pressing the right soft button while the OK check mark is highlighted will return the bar icon to the left most digit In some cases units may be specified e g ppm or hh mm in others there may be no units No cancel option is available to the user They must input a value Enter new PEAK Enter new PEAK Enter new PEAK Enter new PEAK y 36 04 eps 01223792 04 ej Instrument Status Using the Multi PID 2 The instrument status is shown on the left of the first line of the status display Each status has a priority assigned to it If more than one sta tus is in effect then the status with the highest priority is displayed until the condition is corrected or until the option is turned off Table 1 is a list of the possible sy
69. own above Data Log Options Menu The Data Log Options function can be used for the following func tions Clear data log Edit interval only in TVOC mode and Download to PC Clear Data Log The Clear data log function clears all measurement values in the Multi PID 2 memory NOTE Deleted information cannot be recovered You should play back and print or download the contents of the datalogger before deleting any information ihe empty the data logger i in the Multi PID 2 Press the menu key gt Use the V key to choose Data Logger Options and then press the select key 3 Choose Clear data log and then press the select key 4 Presseitherthe yes or no key in response to Are you sure you want to clear all data Edit interval 68 Menu Functions Only available in TVOC mode 1 Press the menu key Use the V key to highlight Data Logger Options then press select 3 Highlight Clear data log then press select 4 Usethe A W and B keys to enter the desired sampling interval 5 Highlight the checkmark then press done Download to PC The Download to PC function starts the download of data from the Multi PID 2 onto the PC The Multi PID 2 computer cable Part No 83 17 667 must be connected between the Multi PID 2 and the PC before beginning the data download The Dr ger GasVision software must be installed
70. rd Long Sample Probe Short Sample Probe Tube Holder Drager Tube Carbon filter Sampling Tube Activated Charcoal Type BIA PID Pre Filter Tube Benzene Dr ger Tube Humidity filter Do not use any other accessories with the Multi PID 2 in a hazardous location Part No 68 10 688 83 17 673 83 17 677 83 17 664 90 23 770 64 05 411 64 05 412 83 19 093 CH 24 101 67 33 011 81 03 511 81 03 531 Notices and Warnings 18 Notices and Warnings 14 Introduction Introduction Introduction About this Manual This manual provides detailed instructions for setup operation and maintenance of the Multi PID 2 Portable Photoionization Monitor Before unpacking the instrument please read Warnings and Safety Practices This section describes possible hazards that might injure the user damage the instrument or compromise its operation Some general safety information is also provided To help you learn to use the Multi PID 2 guickly this manual is organi zed by tasks beginning with Using the Multi PID 2 in Using the Multi PID 2 on page 25 Accessories are covered in Connecting Accessories on page 43 Routine maintenance is covered in Routine Maintenance on page 71 Troubleshooting techniques are covered in Troubleshooting on page 87 The Multi PID 2 manual uses a few conventions for key names on the keypad and for text that is shown on the display Display Text Te
71. ry is only to be done in a nonhazardous area Charge the battery pack using the AC adapter provided with or identified for use with this product only in accordance with the instructions and limitations specified in this manual For AC adapter use only Part No 64 05 428 115 Volt AC Part No 83 17 661 220 Volt AC When using the AC adap ter do not block access to AC outlet in use with adapter AC adapter is not to be used in a hazardous area Notices and Warnings ATEX Directive and EMC Directive rdger SAFETY EC Declaration of Conformity We Dr ger Safety AG amp Co KGaA Revalstra e 1 23560 L beck Germany declare that Equipment Photoionization Monitor Model name Multi PID 2 in accordance with the following Directives 949 EEC ATEX Diroctive 89 336 EEC The Electromagnetic Compatibility Directive has been designed and manufactured lo the following standards Safety EN61010 1 2001 EMC EN61326 1997 A1 1998 and A2 2001 Explosive Atmosphere ENS0014 1998 and ENS0020 2002 hereby deckare that the equipment named above has been designed to comply with the relevant sections of the above referenced standards and all essential requirements of the Directives Signed by AO Alena Name Dr Heinz Dicter Sluma a Title President Business Unit Gas Detection Technology 3 Germany 23560 L bcck Revalstrabe I s On August 28 2003 8 3 Model Multi PID 2 Photoionization Monitor Certified Intrinsically Safe
72. ry slots can be used to store calibration information for 15 different span gases 65 Menu Functions The Multi PID 2 will always detect all ionizable compounds present in a sample regardless of the response factor RF selected NOTE Even if the Multi PID 2 has been calibrated with a specific com pound its response is not specific and the presence of another ionizable impurity may render the numerical result invalid lt is often impractical to carry a range of different standards into the field Approximate results can be obtained by calibrating the Multi PID 2 with the recommended span gas and entering the appropriate response factor The response factor is based on the ratio of the response of the spe cific compound to the response of the span gas The response factor multiplies the Multi PID 2 s reading then displays and records it See table 5 on page 97 for response factors This list should be used only for concentrations up to 500 ppm of the specific compound as response factors change with concentration WARNING The Multi PID 2 is a total VOC instrument and will ALWAYS respond to all ionizable compounds pre sent in a sample The Multi PID 2 will NEVER respond only to a specific compound if an RF is pro grammed It is the responsibility of the user to be aware of the limitations of PID instruments Response Factors and Memory Slots Memory slots 1 edit pres
73. ry to indicate it is charging red light when first plugged in The Charge LED will turn green within a few minutes to indicate the battery is fully charged When the battery pack is fully charged remove the AC adapter first from the wall outlet and then from the Multi PID 2 00823792 04 eps Figure 4 Battery Pack Removal and Replacement 27 Using the Multi PID 2 Charging a fully discharged battery pack will take approximately 4 hours Optionally you can use the off line charger Part No 64 05 404 to charge the battery pack when not installed in the Multi PID 2 If you are charging the battery pack in the instrument you can use all the features of the Multi PID 2 while the battery pack is being char ged Leaving the AC adapter connected to the Multi PID 2 will not harm the battery or the AC adapter in any way If the Multi PID 2 is to be left unused for an extended period of time leave it connected to the AC adapter so that the battery will be fully charged and ready for opera tion On average a fully charged battery pack will provide 8 hours of conti nuous operation Battery life is shorter if the instrument is turned off and then on again repeatedly or if the backlighting is turned on 28 Using the Multi PID 2 Display The Multi PID 2 has a graphic display for reporting detected concen tration and to guide you through configuration options All functions of the Multi PID 2 will be reported on the display
74. s rings bracelets and keys The battery or conductor may overheat and cause burns Charge the battery pack using the AC adapter provided with or identified for use with this product only in accordance with the in structions and limitations specified in this manual For AC adapter use only Part No 64 05 428 North America or Part No 83 17 661 Europe 79 Routine Maintenance Maintenance of the UV Lamp Removing and Replacing the UV Lamp NOTE Do not remove or replace the UV lamp in a hazardous location To remove the UV lamp from your instrument 1 Ensure the instrument is turned off WARNING You must turn the instrument off before removing the lamp cover 2 Remove the lamp housing cover 3 Tilt the Multi PID 2 slightly and remove the UV lamp WARNING Do not touch the wire grid inside the detector cell Any dust or dirt in the detector cell can be blown out with a gentle jet of compressed air Do not insert any object other than the UV lamp into the lamp holder 4 Without touching the lamp window place the new lamp into the Multi PID 2 lamp holder window first See Figure 19 80 Aum cuce Drager Multi PID 2 Figure 19 Removing the UV Lamp NOTE Do not force the lamp into the lamp holder 5 Replace the lamp housing cover Tighten the cover down Do not over tighten 6 Calibrate the Multi PID 2 and then continue normal operation 01823792 04 eps Routin
75. sample The GAS mode is a filter tube mode This means that all measure ments require the installation of the pre filter tube holder 83 10 093 and of the pre filter tubes The PID pre filter tube benzene which is mounted at the head end of the tube holder and adsorbs all sub stances with the exception of benzene is used to selectively measure benzene The entire process is menu driven The selective benzene measurement value is displayed after 90 seconds Activating the GAS mode opens a dialog which queries a device cali bration The benzene measurement screen is displayed after installa tion of the PID pre filter tube benzene ALARM e e CHARGE Urdger L Multi PID 2 1 4 2006 2 view sample menu Figure 8 GAS mode Display 03023792 04 eps Using the Multi PID 2 33 Using the Multi PID 2 The following values are displayed benzene concentration in ppm STEL value the highest benzene measurement value recorded so far SMAX the amount of measurements already carried out SAMP corresponds with the amount of PID pre filter tube benzene already used and the sampling interval period of 90 seconds SECS Pressing the view key shows the Datalogger measurement values Pressing the menu key takes you back to the selection menu Pressing the sample key starts the measurement the pump is acti vated and the measured air is drawn through the PID pre filter tube benzene for 90 se
76. sed to advance the cursor to the next character on the right When the cursor is advanced past the right most character or symbol it wraps around to the first character again To accept the changes press the done soft key when the checkmark is highlighted 4 ALARM e e CHARGE Urdger L Multi PID 2 01423792 04 eps Figure 16 Setting the Time 62 Menu Functions To set the time and date 1 Press the menu soft key 2 With Unit Setup highlighted press the select soft key 3 Using the W soft key highlight Clock Press the select soft key The A and V soft keys are used to change the character un derlined by the cursor The B key is used to advance the cursor to the next character on the right When the cursor is advanced pastthe right most character it wraps around to the first character again Formatting characters such as the colon in the time the deci mal in a concentration and the slash in date are skipped when advancing the cursor 4 Usethe A and V keys to enter the correct time The time is formatted as Hour Minute 5 Use the B soft key until the checkmark Y on the display is highlighted Press the done soft key to confirm the time and move to the date option 6 When setting the date the Multi PID 2 prompts you for the current date formatted as Month Day Year Use the A W and D to e
77. ses QUU RIPE IERI deed e PRI ee 68 Declaration of Conformity x oer e p eem qur drm aa xe ug 11 Default Display xa O ae ee n mU ET DIE Ls 33 Display cas cise re ea TEC UT S Wr are E Sor OC PES 29 Drager GasVision Software u O ER dia UH a P Wer a E x 24 Durationi saro nere eere eV V TU MEER UD QU Fa PR EO rae dra 36 EMG Directive e RR EP VETE DE sion eae ais MR ER NE IDEE 11 Entering Numbers ici sir RARI RE nennen RE Biken sada RIE PR 62 Excessive Heat and Cold iii 17 105 FCG Warning cet er a ae A a ia a a aca ta ds 8 Field Check List vvv e A S a E eo Pane 40 Field Kit usi as d Dee eme a A E Han a deca le 24 Field Operation Aarau ET 24 EixediKeys oa leer riga v e su a ee 32 GAS mode DiSplay ie 44i dd 222 pli Re a BRAUN BRE TIN TAS en 33 General Information sr aa EA ET Rn iE 65 General Operation iurirrra u kinas ola es edet uA Saou a om lala es 20 Graphic Display siue ct len sa aa RUDI na RIO ERES qd mias rd 29 Instrument Status a sei eet v eI p es ol LO a a ie o AE 37 Intended Use sc ia ee odas Eo e eu eee S drum 18 Interval Mode Display 2s 2 04 verse PII ae gow are ae ass xa qup Es 35 Intervall time adjustment LL 61 Intrinsic Safety i uen mie erus es PU PUR pala de ated wee i gs 8 Introduction rire help oe AS E a Tal M FU a 15 Kayser ced Liusi a aa lea 32 Logging Off Mode Display sssssssssse II nnn 34 Main componenis L ii idea E TET Hals Pare s 20 Material Safety Data Sheet
78. stem alerts Table 1 System Alerts Multi PID 2 Display Description Zero Air Error Span Gas Error Pump Error UV Lamp Error Blocked Filter Low Battery Icon Data Log Full Unit Locked Unit Communicating with PC Instrument Over Range Alarm R Icon Alarm T Icon Alarm S Icon Zero gas too high Span gas too low Pump fault UV lamp fault Blocked sample inlet filter Low battery Data memory full Unit locked by passcode Instrument communicating with PC Sample concentration over instrument operating range Real time concentration alarm TWA concentration alarm STEL concentration alarm System Alerts and Alarms While operating the instrument system alerts can occur To accura tely identify the source of the alarm each type of alarm has been given a unigue status In addition to the status the Multi PID 2 also has an audio alarm and an alarm LED To conserve power the Multi PID 2 alternates between the LED and audio Different alarms are identified by the freguency at which the Multi PID 2 alternates between the audio and LED Peak alarm is 5 times per second STEL alarm is 2 5 times per second and TWA alarm is 1 25 times per second 37 Using the Multi PID 2 A soft key is used for acknowledging alarms and is named accept If no alarm exists then the accept key is not shown To clear the
79. t least 5 days NOTE You may not be able to fully clean the 11 7 eV lamp window Routine Maintenance 83 Routine Maintenance Replacing the Sample Inlet Filter The Multi PID 2 is eguipped with a combined dust and water filter to reduce detector contamination As the filter collects dust the Multi PID 2 s inlet flow rate and sensitivity decrease The filter will not allow water to pass through but the filter will not stop all solvents NOTE Do not aspirate liquid samples with the Multi PID 2 Replace the filter on a weekly basis or more frequently if the Multi PID 2 is used in a dusty or wet environment You must replace the fil ter if the Multi PID 2 has been exposed to water If you are sampling hot gases or vapors condensation in the sample line may also affect the filter WARNING Do not replace the inlet filter in a hazardous loca tion To replace the inlet filter 1 Turn the instrument off Unscrew the filter housing from the detec tor housing Be careful not to lose the O ring seal 2 Remove the Teflon Polypropylene filter and install the new filter Part No 83 17 681 Place the filter so that the Teflon side is fa cing down in the filter housing and the mesh side is facing the Mul ti PID 2 Handle the filter disk only by the edges The mesh may be dama ged or contaminated by excessive handling Use forceps if possi ble 3 Replace the filter housing 4 Recalibrate and then continue normal opera
80. th compressed gases Do not store cylinders in a hazardous location Store cylinders in an upright position away from possible sources of heat or sparks Never plug obstruct or tamper with safety relief devices Regulators for Compressed Gases When connecting a regulator to a large cylinder 74 Ensure cylinder valve and regulator connection match Ensure regulator construction materials are compatible with the gas and that the cylinder pressure gauge will withstand the cylin der pressure Never use the regulator as a shut off valve Close the cylinder when it is not in use Do not subject the regulator to an inlet pressure greater than re commended Do not move or detach the regulator when it is pressurized or when it is in use Routine Maintenance e Beforeconnection ensure the gas cylinder valve and the regulator CGA connection are clean e Turnthe pressure control valve on the cylinder all the way off clo se the cylinder Turn the regulator outlet to off Open the gas cy linder valve slowly and check for leaks Adjust the delivery pressure and then open the regulator outlet valve Calibration Gas Adequate ventilation must be provided when the Multi PID 2 is being calibrated If compound threshold limit values TLV are exceeded you should use a gas bag for sampling and calibration To determine the TLV of the compounds contained in the calibration gas refer to the Material Safety Data Sheet MSDS supplied w
81. the Multi PID 2 You can display the results of the calculations by selecting Interval Mode Time Weighted Average TWA Mode The TWA accumulator sums concentrations every second until 8 hours of data have been combined If this value exceeds the TWA alarm setting a TWA alarm is generated The TWA is not calculated using a moving average Once 8 hours of data have been summed the accumulation stops In order to reset the TWA accumulator press the clear key then select TWA using the W key then press clear This sum will only be complete after 8 hours so the meter displays the current sum divided by 8 hours TWA calculations are always being performed by the Multi PID 2 You can display the results of the calculations by selecting Interval mode PEAK Mode The PEAK mode displays the current detected concentration The reading is updated once a second In the background the Multi PID 2 datalogger is sampling the concentration and measuring minimum maximum and average concentrations for the selected averaging interval At the end of every interval one entry is placed in the datalog ger until the datalogger is full In order to reset the PEAK reading press the clear key then select PEAK using the W key then press clear 39 Using the Multi PID 2 Preparing for Field Operation Field Check List When using the Multi PID 2 for field operation the following items should be carried into the
82. tion 84 ALARM CHARGE Drager Multi PID 2 Figure 20 Replacing the Inlet Filter Waste electrical and electronic equipment EU wide regulations for the disposal of electric and electronic appliances which have been defined in the EU Directive 2002 96 EC and in national laws are effective from August 2005 and apply to this device Common household appliances can be disposed of using special col lecting and recycling facilities However as this device has not been registered for household usage it must not be disposed of through these means The device can be returned to your national Dr ger Safety Sales Organization for disposal Please do not hesitate to contact the above if you have any further questions on this issue 01923792 04 eps Routine Maintenance 85 Routine Maintenance 86 Troubleshooting Troubleshooting 87 Troubleshooting General Information If you have a service related guestion about the Multi PID 2 consult this manual first If you cannot find the answer in this documentation contact Dr ger Safety s Technical Support Department When you call please have the following information ready 1 Adescription of what happened and what you were doing when the problem occurred 2 Any corrective action that you have tried 3 The exact wording of any messages that appeared on the display WARNING Do not service the Multi PID 2 in a hazardous
83. ttery pack in the Multi PID 2 as outlined in Remo ving and Replacing the Battery Pack on page 26 You can keep the battery pack fully charged indefinitely without over charging it by leaving it connected to the off line charger while the charger is operating Charging from a DC Source To re charge your battery pack 1 Remove the battery pack as outlined in Removing and Replacing the Battery Pack on page 26 2 Attach the connector from the battery pack to the socket on the off line charger NOTE The connector is polarized It will only fit one way Do not force the connection 3 4 9 10 Plug the DC power cord into the jack on the opposite face of the off line charger Plug the DC power cord into a vehicle auxiliary 12 VDC or cigaret te lighter socket The LED on the upper face of the off line charger indicates the charge state of the battery pack Red indicates the battery is being charged Green indicates the battery is fully charged Charging a fully discharged battery pack will take approximately 4 hours It is normal for a fully charged battery pack to indicate it is charging red light when first plugged in The LED will turn green within a few minutes to indicate the battery is fully charged When the battery pack is fully charged remove the DC power cord first from the vehicle auxiliary 12 VDC or cigarette lighter socket then from the off line charger Remove the battery pack connector from th
84. ttery through the ciga rette lighter with the DC power cord While the Multi PID 2 is connec ted to the car battery the Multi PID 2 battery is being charged NOTE The 2Multi PID 2 is not classified for use in hazardous locations with a DC power cord To connect the Multi PID 2 to a DC power supply 1 Turn the instrument off by pressing the ON OFF key for five secon ds 2 Connectthe DC power cord Part No 64 05 421 to the Multi PID 2 AC adapter jack on the rear of the instrument 3 Connect the other end of the DC power cord to the cigarette ligh ter in the car 4 Turn the instrument on again by pressing the ON OFF key If the vehicle is running ensure the engine exhaust does not contami nate your samples Belt Clip Holster Use the belt clip holster Part No 83 17 677 to protect the instru ment and to mount the instrument to a belt 47 Connecting Accessories 11 7 eV UV Lamp General Information The Multi PID 2 is eguipped with a standard 10 6 eV UV lamp An 11 7 eV UV lamp Part No 83 18 317 is available for special applica tions Install this lamp as outlined in Removing and Replacing the UV Lamp on page 80 When you install a UV lamp other than the standard 10 6 eV lamp all the response factors become invalid You must set the response factor for the current cal memory to 1 Contact Dr ger Safety s Applications Department for further information NOTE Do not use the 11 7 eV lamp in conjunction w
85. uction manual may cause harmful interfe rence to radio communications Operation of this eguipment in a resi dential area is likely to cause harmful interference in which case the user will be reguired to correct the interference at their expense The Multi PID 2 Intrinsic Safety I S Notice wath LISTED CME To The Multi PID 2 is classified for use in class I division 1 groups A B C D hazardous locations T4 135 C rating It has been listed by MET Laboratories Inc to comply with Underwri ters Laboratories Inc UL 913 Standard for Intrinsically Safe Appa ratus and Associated Apparatus for use in Class I Division 1 Groups A B C D Hazardous Classified Locations Sixth Edition when powered by MX700010 Battery Pack The Multi PID 2 is not intended to detect combustible levels of gases The Multi PID 2 is classified for use in atmospheres containig combu stible levels of gases 04 eps 02323792 04 e These optional accessories may be used with the Multi PID 2 in a hazardous location Name and Discription Calibration Regulator Wrist Strap Belt Clip Holster Carrying Case User s Reference Card Long Sample Probe Short Sample Probe Tube Holder Drager Tube Carbon filter Sampling Tube Activated Charcoal Type BIA PID Pre Filter Tube Benzene Dr ger Tube Humidity filter Do not use any other accessories with the Multi PID 2 in a hazardous location Substitution of compone
86. urn the instrument on and wait for the ready status If the lamp cannot be started contact Dr ger Safety s Technical Support De partment 3 Calibrate the Multi PID 2 See Calibration Using the Flow Match Regulator on page 76 4 Allresponse factors are invalid when an 11 7 eV lamp is installed The response factor must be set to 1 5 Recalibrate the instrument every 15 minutes of operation 6 Every hour of operation switch off the instrument remove the lamp and examine the window for yellowing If the window is yel low then regenerate the window according to the procedure in Cleaning the Lithium Fluoride Window on page 83 7 After use remove the lamp and store it in the supplied desiccant bottle Do not leave the 11 7 eV lamp in the instrument when you turn it off Always remove the 11 7 eV lamp and store it in the supplied desiccant bottle 49 Connecting Accessories Off Line Charger General Information The off line charger Part No 64 05 404 allows you to charge a bat tery pack independently of the Multi PID 2 To charge a battery pack without removing it from the Multi PID 2 see Battery Charging on page 26 The off line charger requires one of the following the Multi PID 2 accessories for operation AC Adapter Part No 64 05 428 North America AC Adapter Part No 83 17 661 Europe DC Power Cord Part No 64 05 421 Use only the AC adapter specified for use with the Multi PID 2 Using another AC adapter wi
87. xts shown on the Multi PID 2 display are written with quotation marks e g Enter current time The names of soft keys are also written with quotation marks e g select In the text you will find various warnings and notes WARNING A warning indicates an operation that could cause personal injury if precautions are not followed CAUTION A caution indicates an operation that could cause instrument damage if precautions are not followed NOTE A note indicates significant information 16 Introduction Warnings and Safety Practices Please read the Notices and Warnings section of this user s manual before operating the Multi PID 2 Approved Models of the Multi PID 2 This manual provides operational information for all models of the Multi PID 2 The Multi PID 2 is intrinsically safe and approved for use in hazardous locations Refer to the Notices and Warnings section of this manual for details of each approval Throughout the manual notes are provided to inform you of any limita tions of usage for the Multi PID 2 models WARNING If the Multi PID 2 you are using is not specifically identified as intrinsically safe with a label on the Multi PID 2 do not use it in a location where flammable concentrations of gases and vapors may exist Excessive Heat and Cold WARNING Do not expose the instrument to intense sunlight for prolonged periods Exposure to excessive heat or cold may result in erroneous re
88. y Program the memory slots See page 66 Response Factor has been set to zero Set Response Factor to 1 0 or refer to Table 6 for correct response factor Detector is leaking UV lamp is the incorrect size Ensure the UV lamp is cor rectly installed Ensure the lamp cover is tight Do not over tighten Ensure the O ring seal on the lamp cover is positioned cor rectly Contact Dr ger Safety for correct lamp Sampling environment is extremely humid or it is rai ning Water vapor is not ionized by the PID Water vapor does scatter and absorb light that causes a higher or lower rea ding The Multi PID 2 has been designed to work in a humid environment but you may notice a decreased response UV lamp is failing Install a new UV lamp See page 80 High concentrations of non ionizable compounds Chemical compounds with an ionization potential higher than 10 6 eV or 11 7 eV scat ter and absorb UV light Sen sitivity may be decreased Contact Dr ger Safety for more information 91 Troubleshooting Fault Possible Cause Remedy Erroneous high readings Sampling environment is very humid Water vapor may contain mineral salts which carry a charge The water vapor becomes an electrolytic solu tion which becomes ionized when it enters the detector The Multi PID 2 is not pro perly calibrated Ensure that the calibration gas is the

Download Pdf Manuals

image

Related Search

Related Contents

  Epson StylusRIP Adobe PostScript Software for Windows and Macintosh Quick Start Guide  "取扱説明書"  Samsung DVD-R6000 User Manual  Model 8174 HF Monitor Receiver Operating and Service Manual  XZR400 Analyseurs d`oxygène  HP Color LaserJet CM6030 CM6040 MFP Service Manual    BETRIEBSANLEITUNG USER MANUAL CD 1230 R  Storage Options 55662 User Guide  

Copyright © All rights reserved.
Failed to retrieve file