Home
        Pressure Interface
         Contents
1.      14  TW LN         E                              SP RED EE O DR 14  o                        ens E       i ii 16  ARA                                                      RR RED RAE E              17  CAI UE Sa 17  O ELITES                 e E               18  26 1 Force calibration      dina it 19  3 6 2 Pressure Calibration PROCE QUITE sida Sd 22  3 6 3 pH Calibration Procedure  optional                                      rna aaia a 25  3 6 4 Analog Record Out Select             o       26            e                 E E          27  SC    TEMPERATURE DIFFERENCE ORE ira 27  PA        EE EEE    ani 28  305 3 FACTORY DIAGNOS TIC            ia ATE east EEEE                  28  Appendix 1   System Specifications              nssnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn mennan nn nnnm             nnmnnn a na 29    4 PRESSURE INTERFACE   USER MANUAL    SAFETY    The Pressure Interface has been designed for use only in teaching and research applications  It is not intended for clinical or criti   cal life care use and should never be used for these purposes  Nor for the prevention  diagnosis  curing  treatment  or alleviation  of disease  injury  or handicap    e Do not open the unit  the internal electronics pose a risk of electric shock     e Do not use this apparatus near water     e To reduce the risk of fire or electric shock  do not expose this apparatus to rain or moisture  Objects filled with liquids should  not be placed on
2.  Myograph  see figure 2 3  to the P1 outlet on the cap of the buffer bottle   see fig  2 3      e Connectthe    perfusion outlet    at P2 on the Pressure Myograph Unit  see figure 2 3  to the P2 inlet on the Pressure Interface   see fig  2 3         Figure 2 3 Pressure Interface tube connection    CHAPTER2 11    NOTE     TO FILL THE TUBING AND AVOID AIR BUBBLES  MOUNT THE    CALIBRATION SHUNT TUBE    AS DESCRIBED IN CHAPTER 5   FILL THE P1 SCHOTT BOTTLE WITH A MAXIMUM OF 400 ML OF BUFFER  IN THE    PRESSURE    MENU SET THE PRESSURE  P1 AND P2 AS FOLLOW     P1   150mm Hg   P2   7Omm Hg   Set the Pressure ON  Set Flow ON    Let the pressure regulator run until no air bubbles are visible in the perfusion circuit     IMPORTANT  NEVER RUN AIR BUBBLES THROUGH A VESSEL AS THIS MAY DAMAGE IT     NOTE II  NEVER FILL THE BUFFER BOTTLE WITH MORE THAN 400 ML OF BUFFER AS HIGHER VOLUMES MAKES PRECISE CONTROL  OF THE PRESSURE MORE DIFFICULT     2 6 Extra perfusion inlet outlet    The extra perfusion inlet outlet 3 way valves on the Pressure Myograph allow the introduction of special agents or reagents to  the perfusion circuit during an experiment  They are situated at P1 and P2 on the Myograph Chamber and the small handle on  top of the valves controls the flow direction     2 7 Connecting superfusion flow  optional     Creating a superfusion flow in the Pressure Myograph Chamber requires an external peristaltic pump  which is not part of the basic  Pressure Myograph System  Contact DMT f
3.  NEXT STEP    pH1 CALIBRATION    Step no    1   2     Apply buffer with high pH value  to the pH and Temp  Probe   When the pH reading   is stable  go to next step     pH1 readings  1863    NEXT STEP    pH1 CALIBRATION    Step no    2       4     Apply buffer with the Low pH value  to the pH and Temp  Probe    When the pH reading   is stable  go to next step     pH1 readings  883    pH1 CALIBRATION    Step no    4     The pH1 sensor is now calibrated  if OK  go to next step  else  repeat the calibration    pH1 value  4 00    Probe Temp   25 0    NEXT STEP    NEXT STEP       CHAPTER3 25    3 6 4 Analog Record Out Select  optional     The RECORD OUT SELECT sub menu determines how data is sent to the  BNC analog outputs at the back of the Pressure Interface  There are 4  analog outputs  Each output is individually programmable  Any change  made to RECORD OUT will only affect the data collected by the external data  acquisition system  such as an AD Instruments PowerLab   So itis important  to check the data acquisition setup when a change is made to the analog  output with this setting  The analog output works in range of  2 5V to  2 5V     Use    SELECT    to select the OUT channel number to be changed  Press     ENTER    to go to the output setup     Use    SELECT    to select the line to be changed     Use the arrow keys or    DEFAULT    to change select a new setup     26 PRESSURE INTERFACE   USER MANUAL    SELECT REC  BNC ANALOG OUT    Select Record Out1   Select Record Out
4.  the apparatus     e Do not block any ventilation openings  Install in accordance with the manufacturer   s instructions     e Only use secure industry standard connectors and tubing for pressure connections  Faults  defects and mistakes due to  wrong connections void warranty  We are not accountable for results and mistakes due to inappropriate pressure hookup     e Do not install near any heat sources such as radiators  heat registers  stoves  or other apparatus that produce heat   e Only use attachments and accessories specified by the manufacturer   e Unplug this apparatus during lightning storms or when unused for long periods of time    The Pressure Interface is delivered with an external 100 240VAC to 24VDC ADAPTER     Protect the power ADAPTER and cord from being walked on or pinched  Particularly at power plugs and the point where they con   nect to the apparatus     Refer all servicing to qualified service personnel  Servicing is required when the apparatus has been damaged in any way  such    as  the power supply cord or plug is damaged  liquid has spilled onto or objects have fallen into the apparatus  the apparatus has  been exposed to rain or moisture  does not operate normally  or has been dropped     SAFETY 5    UNPACKING        PRESSURE INTERFACE AND MYOGRAPH SYSTEM    Please take a few minutes to carefully inspect your new Pressure Myograph System for damage  which may have occurred during  handling and shipping  If you suspect any kind of damage  please con
5.  the oxygen supply pipe on the Pressure Myograph Chamber cover to an adjustable oxygen supply  See figure 2 2    Pipe  for oxygen supply        12 PRESSURE INTERFACE   USER MANUAL    2 10 Connecting a suction device  Connect the    large    pipe on the Pressure Myograph Chamber cover to a vacuum pump via a suction bottle and a vacuum valve    as illustrated in fig  2 1  The internal diameter of the silicone tube used for this connection must be 2 mm   A 2 mm tube is not  included with the Myograph System      2 11 3 Way valve adjustments    The 3 way valves on each side of the Pressure Myograph Unit  at P1  amp  P2  have three different settings to control the inlet   outlet  flow to the Pressure Myograph Chamber  The flow settings are illustrated in fig  2 4 and fig  2 5        Figure 2 4 Three way valve flow settings at P1       Figure 2 5 Three way valve flow settings at P2    CHAPTER 2 13    CHAPTER 3   THE PRESSURE INTERFACE MENUS    This chapter contains a detailed description of how to navigate the touch screen menus and how to use the special features of  the Pressure Interface     3 1 General description of how to navigate the touch screen    Menus on the Pressure Interface are all accessible by a touch screen  To access a menu  simply touch the corresponding button  on the screen  When a setting needs to be changed  press the relevant    SELECT    button or in the display below  touching the     ON    or    OFF    buttons     The line selected will turn blue  indica
6. 2     Select Record Out3     Select Record Out4     SELECT    SELECT    SELECT    SELECT       SELECT REC  BNC ANALOG OUT    Select Record Out1   Select Record Out2   Select Record Out3     Select Record Out4     RECORD TOUTPUT SET    Record 1  Force  Rec  1 out  2 5V   100mN  Rec  1 out  2 5V  100mN    DEFAULT    RECORD 1 OUTPUT SET  Record 1  Force   Rec  1 out  2 5V   100mN  Rec  1 out  2 5V  100mN    DEFAULT    SELECT  SELECT  SELECT    SELECT    SELECT    SELECT    SELECT    SELECT  SELECT    SELECT       The following parameter in the Pressure Interface can be selected as output  on the Analog channel  Force  Probe temperature  Chamber temperature   pH1  Pressure regulation ON OFF  Flow ON OFF   Optional  Digital output 1   Digital output 2  Digital input 1  Digital input 2   Pressure   1  Pressure   2     When the parameter is selected press    ENTER    to save the value     The output range is  2 5V to  2 5V  The user can select the parameter value  that gives     2 5V and the value that gives  2 5V  In this example  100mN is  2 5V and   100mN is  2 5V     Selecting asymmetrical values is also valid  They can be  2 5V   OmN and   2 5V   100mN     NOTE  REMEMBER TO PRESS THE    ENTER    KEY TO SAVE THE LINE CHANGES  TO MEMORY     3 6 5 Interface settings    The INTERFACE SETTINGS sub menu in SETTINGS has an additional 3 sub   menus  These 3 additional sub menus are     TEMPERATURE DIFFERENCE  pH SETUP  FACTORY DIAGNOSTICS    3 6 5 1 TEMPERATURE DIFFERENCE  OFFSET     T
7. ACE SETTINGS       18 PRESSURE INTERFACE   USER MANUAL    3 6 1 Force calibration Procedure       Figure 3 1   Weight Calibration Set up    gs       As      Yi       Figure 3 2   Illustration of how to fit the balance just behind the glass cannula fixation plate    CHAPTER3 19    Enter the FORCE CALIBRATION sub menu to begin the transducer  calibration  The display will show the force calibration procedure     Step 1   The calibration procedure is listed in 6 individual steps and needs  to be performed one at a time to calibrate the system  Step 1 involves  setting up the chamber for calibration  Fill the cnamber to a normal level  with double distilled water  Press    NEXT STEP        Step 2   Involves setting up the calibration kit for the actual weight  calibration     Make sure that the tip of the calibration bridge arm is placed behind the  glass cannula fixation plate as illustrated in fig  3 71  Carefully move the  calibration bridge until the tip of the arm is placed freely behind the glass  cannula fixation plate  It should not be touching the fixation plate     Press    NEXT STEP    when the calibration kit has been properly placed     Step 3   This step initiates the heating process for the chambers  In order for  the calibration to be accurate  the transducers must be heated to the same  temperature to be used in your experiment  This allows for heat induced  expansion of the electronic parts in the transducer  Otherwise  inaccurate  readings and transducer drift 
8. ALLED WITH DRIVERS FOR THE DIGITAL CCD  USB CAMERA  FOLLOW THE PROCEDURES IN SECTION 2 2 TO SETUP THE PRESSURE MYOGRAPH SYSTEM     If you have not purchased a computer from DMT follow the MyoVIEW Il Quick Installation Guide to install MyoVIEW Il on your own  computer     2 2 Setting up the complete Pressure Myograph System    This section describes how to connect the cables in the Pressure Myograph System as illustrated in figure 2 1  Before connecting  any of the myograph equipment ensure that for the 110P Pressure Myograph system that the USB camera is connected to your  PC  For the 111P Pressure Myograph system please connect both USB cable of the DMT microscope to the computer  Install the  MyoVIEW Il Data Acquisition Software on your PC as described in the MyoVIEW II Quick Installation Guide     NOTE  BEFORE PROCEEDING WITH THE CONNECTION PROCEDURE  MAKE SURE THAT THE PRESSURE INTERFACE  THE  MICROSCOPE AND THE PC ARE SWITCHED OFF     1  Pressure Myograph Unit to Pressure Interface connection   Connect the Pressure Myograph Unit to the Pressure Interface by using the grey 25 44 pin connection cable  The end of the  cable with the temperature probe is connected to the Pressure Myograph Unit     2  Pressure Interface to PC connection   Connect the Pressure Interface to the computer with the USB cable  from the back panel of the Pressure Interface to the  USB port on the computer     3  Pressure Interface and oxygen supply connection   Connect the    Oxygen    inlet on the 
9. DMI    DANISH MYO TECHNOL Ost    PRESSURE INTERFACE    FOR PRESSURE MYOGRAPH SYSTEMS    ae    kre  1 Ro         qu   un i 2 3 1   LL         o        gt  Eq        User Manual    Version 1 1       2 PRESSURE INTERFACE   USER MANUAL    PRESSURE INTERFACE    FOR PRESSURE MYOGRAPH SYSTEMS    TRADEMARKS    PowerLab   and LabChart   are registered trademarks of ADInstruments Pty Ltd     The names of specific recording units  such as PowerLab 4 35  are trademarks of ADInstruments Pty Ltd     Danish Myo Technology A S reserves the right to alter specifications as required   This document was  as far as possible  accurate at the time of printing   Changes may have been made to the software and hardware it describes since then   New information may be supplied separately     This documentation is provided with the  Pressure Interface    No part of this document may be reproduced by any means without  the prior written permission of Danish Myo Technology A S     Copyright O Danish Myo Technology A S       DANISH MYO TECHNO Low    TRADEMARKS    3    CONTENTS           MN AS assassina i e se ce aa wc les E ee re A ce i ice a EA 3  A                                   RE REP SE E 5  Unpacking the Pressure Interface and Myograph Systeim          c   sssccssssseccsensseeccnnssecennnseeconsseeeeonssseeeeasseenenssseeeenssseeeenssssenennsssens 6  Chapter 1                        cca 7  od PRC SUE                                                                           7  Lo Pressure Mio sra o
10. ER MANUAL    Icon for Pressure ON OFF                 Icon for Flow ON OFF                                        Icon for Heater ON OFF              OA CT     Icon for Buzzer ON OFF                                                                                   MAIN MENU     00 00 00     g  1  Pressure P1   122 0 mm Hg  Pressure P2   72 0 mm Hg    Force chamber   20 2 mN e ZERO key    This key is used to zero the  force output      If ON the icon is GREEN    pH1 value  6 81  Probe temp   36 8   C       Pressure setup menu   rst      Heat setup menu      OE ONE          Timer alarm            SUE A einnsean cinched arctic eee ann            Buzzer icon     This icon indicates the status on the Buzzer   If the icon is grey the Buzzer is OFF   If the icon is green the Buzzer is ON  active      The Buzzer will make a sound when the Timer reaches zero     OFF ON  Timer   The Timer is a countdown watch that can be set to a maximum of 24 hours  00 00 00  If the buzzer is activated it will sounds when the timer reached zero   Heat icon   This icon indicates the status of the chamber heat   If the icon is grey Heat is OFF   If the icon is green the Heat is ON  active    OFF ON    Flow valve icon   This icon indicate the status on the Flow valve   If the icon is grey the Flow valve is OFF  NO Flow    If the icon is green the Flow valve is ON  active    OFF ON    Pressure icon     This icon indicates the status of the Pressure regulation   If the icon is grey the Pressure regulation 
11. ansducer P1    Pole    DI Sakae Micropositioner for longitudinal adjustment of glass cannulas    Figure 1 3 The Pressure Myograph Unit    Fixation plate for left glass cannula     e Fixation plate for right glass cannula  9        Outlet connecting pipe  e    perio e Perfusion inlet pipe       99 999    0000000000                  Silicone tube connecting outlet connecting       Silicone tube connecting perfusion inlet pipe and right  pipe with left glass cannula       glass cannula  Left glass cannula e  i A        e Right glass cannula              O Myo chamber window    Figure 1 4 The Pressure Myograph chamber   magnified    8 PRESSURE INTERFACE   USER MANUAL    CHAPTER 2   SETTING UP        PRESSURE MYOGRAPH SYSTEM    FlowMeter 161FM    USB cable  Waste Bottle       USB cable       Data Acquisition  Software and PC       dm a              Peristaltic Pump    USB cable  optional           Pressure Myograph   111P         ss 4       incl  Micr  a Pressure Interface  incl  Microscope     pH sensor          optiona  E  C  o  Pressure and      O   Oxygen Supply RE SEIA pls   optional     Oxygen Supply  Superfusion  Sucction Bottle Buffer       Outlet Superfusion buffer  Inlet Superfusion buffer  Inflow Perfusion buffer  Outflow Perfusion buffer  Cables   Tubing    Figure 2 1 Example of the Complete Pressure Myograph System   Model 111P    Fig  2 1 is a diagram of a complete setup for the Pressure Myograph System   111P  The diagrams include optional equipment  such as a peris
12. back panel of the Pressure Interface to an adjustable oxygen supply using the    Atlas  Copco Cablair hose    delivered with the Pressure Myograph System     IMPORTANT MAKE SURE THAT THE HOSE IS SECURELY FASTENED TO THE    OXYGEN    INLET USING THE HOSE CLIP  THE  PRESSURE FROM THE OXYGEN SUPPLY MUST NOT EXCEED 1 0 BAR     4  Microscope camera to PC connection   Inside the DMT Microscope there is a USB camera wired to a USB cable and one USB cable to power the microscope light   Connect the two USB cable from the DMT microscope to the computer USB ports  For the 110P system with the external  USB camera connect the USB cable from the USB Camera to your PC for the 110P system     5  Pressure Interface to external data acquisition system connection  optional    If you have not obtained the MyoVIEW Il software the four BNC ports on the rear of the Pressure Interface enable connection  of an external data acquisition system for recording inlet pressure P1  outlet pressure P2  longitudinal force  probe  temperature  chamber temperature  flow ON OFF  pressure ON OFF  pH1 value and  Optional Dig out1  Dig out2  Dig in1 dig  in2      6  Turn on the power   Turn the main power to the Pressure Interface ON at the power switch  Turn ON the computer  The Pressure Myograph  System is now ready to be used with MyoVIEW Il Data Acquisition Software   NOTE   A DETAILED GUIDE TO CONNECTING THE PRESSURE MYOGRAPH SYSTEM FOR AN EXPERIMENT IS DESCRIBED IN DETAIL  IN CHAPTER 5     For installation 
13. d between 3900 to 4500  Apply 125 mmHg pressure on the  Myo  When the pressure reading  is stable  go to next step   Pressure P1  4293  Pressure P2  4203  Pressure P3  4211                         Step 4   Apply a pressure of 200mm Hg to the Pressure Myograph Chamber    using the    Big Ben    Pressure Manometer  When the 3 blue lines of relative PRESSURE CALIBRATION  values stabilize  go to    NEXT STEP       ser  DO OB    The relative values should between 4600 to 5200  Apply 200 mmHg pressure on the    Myo  When the pressure reading  is stable  go to next step   Pressure P1  4905    Pressure P2  4788  Pressure P3  4 784    NEXT STEP    lines are the actual pressure on all three pressure sensors  The values must PRESSURE CALIBRATION    Step 5   The calibration is now finished  The readings in the three bottom    be 200 0    1 0mm Hg     Step no   Ove    If the readout is 200 0   1 0 mm Hg  Carefully remove the    Calibration Shunt Tube    and reconnect the tubing to the calibration is OK   the glass cannulas  Otherwise repeat the calibration   Pressure P1  200 0 mm Hg  Pressure P2  200 0 mm Hg  Pressure P3  200 0 mm Hg       24 PRESSURE INTERFACE   USER MANUAL    3 6 3 pH Calibration Procedure  optional     Before the pH calibration is performed be sure to select the way the pH  sensor is to be used  See the sub menu    pH SETUP    under INTERFACE  SETTINGS     The temperature is an important parameter in the calibration formula and  is obtained automatically if AUTO is sel
14. ected in Temperature compensation  function  shown below     If MANUAL is selected  the Manual temperature is used in the pH calibration  formula  and the temperature probe is de activated    Inthe MANUAL mode  the temperature of the calibration buffers is measured  with a thermometer and entered manually in the    manual temperature    line     Step 1   The pH calibration procedure is listed in 4 individual steps and  needs to be performed one at a time  Step 1 involves cleaning the pH probe  and the temperature sensor with double distilled water  When ready Press     NEXT STEP        Step 2   Place the pH meter electrode and temperature probe in the HIGH  buffer solution  here pH 7  and turn on stirring of the HIGH buffer solution   When the relative pH output in the blue line is stable  go to    NEXT STEP        Step 3   Place the pH meter electrode and temperature probe in the LOW  buffer solution  here pH 4  and turn on stirring of the LOW buffer solution   When the relative pH output in the blue line is stable  go to    NEXT STEP        Step 4   The calibration is now finished  The values in the two bottom lines  are the actual pH reading  and the temperature reading     pH SETUP MENU    Low buffer pH  4 0 SELECT  High buffer pH  7 0 SELECT    Temperature comp   AUTO SELECT    Manual temp  Value  20   C SELECT    Step no   ME         Follow the pH calibration  procedure in the User Manual   Clean the pH probe and the    temperature probe   When ready go to next step    
15. en the relative force reading  is stable  go to next step     Force Chamber  3261    BACK NEXT STEP    Step no   GUI  5     Carefully place the 2 9 weight  On the pan    When the relative force reading  is stable  go to next step     Force Chamber  3346    NEXT STEP    FORCE CALIBRATION    Step no   UE  Dw    The transducer is now calibrated   Force read out should be 9 81 mN    0 1 mN  If OK go to next step     Otherwise  repeat the calibration     NEXT STEP    Force Chamber   9 81 mN    CHAPTER 3 21    3 6 2 Pressure Calibration Procedure    The setup to perform the pressure transducer calibration procedure is illustrated in Fig  3 3  Carefully follow the procedure  described below to calibrate the Pressure Interface System  The calibration procedure is listed in 5 individual steps and needs to  be performed one at a time to calibrate the system        Figure 3 3 Illustration of setup for pressure transducer calibration    Step 1   involves setting up the Myograph Chamber for pressure calibration     Step 2   Connect the    Big Ben    Pressure Manometer to the 3 way valve at P1 on the Pressure Myograph Chamber as illustrated  in fig  3 3     Step 3   Inside the Myograph Chamber  on the right hand side  is a small stainless steel pipe  which is connected to the right glass  cannula via a small silicone tube  Fig  3 4 B   Carefully disconnect the silicone tube from the stainless steel pipe     Step 4   Carefully disconnect the silicone tube between the    outlet connecting pi
16. he TEMPERATURE DIFFERENCE function allows the user to fine tune the  temperature setpoint of the system  Although the temperature setpoint for  the system can be set in the HEAT MENU  the actual temperature for the  system may not heat to the exact setpoint  This function allows the user to  adjust the temperature of the chamber and bottle individually so the EXACT  temperatures can be reached  This is referred to as a temperature offset   TEMP OFFSET on chamber   The    SELECT    and    ALL    functions are the  same in this menu as previously described  Pressing    ENTER    will save the  values for future experiments     RECORD 1 OUTPUT SET    Record 1  Force    Rec  1 out  2 5V   100mN SELECT    Rec  1 out  2 5V  100mN SELECT    INTERFACE SETTINGS    TEMPERATURE  DIFFERENCE    FACTORY  DIAGNOSTICS    TEMP OFFSET    Chamber  1 20  Bottle  1 1  C       TEMP OFFSET    Chamber  1 26  Bottle  WASG       CHAPTER 3 27    3 6 5 2 pH SETUP MENU    Before the pH calibration is performed  be sure to have selected the way the  pH sensor is intended to be used  This is done in the sub menu    pH SETUP        The temperature is an important parameter in the calibration formula  and is obtained automatically if AUTO is selected in the Temperature  compensation  as shown below    If MANUAL is chosen  the Manual temperature is used in the pH calibration  formula  and the probe temperature is deactivated    Inthe MANUAL mode  the temperature of the calibration buffers is measured  with a t
17. hermometer and entered manually in the    manual temperature    line     3 6 5 3 FACTORY DIAGNOSTICS    Entering FACTORY DIAGNOSTICS will display the LOGIN CODE TO  DIAGNOSTICS window  This window is for trained technicians and is used for  diagnostics and troubleshooting purposes  The general user will not have  access to this window  Entering the proper 5 digit pin number  however  will  give the trained technician access to Diagnostics panels that will provide  information during a malfunction  or mechanisms to change other settings  controlled by the onboard computer     28 PRESSURE INTERFACE   USER MANUAL    pH SETUP MENU  Low buffer pH  4 0  High buffer pH  7 0    Temperature comp   AUTO  Manual temp  Value  20   C    Type login code to get acces     EM O       5 o       Code init value  51761 BBBBB    SELECT    SELECT    SELECT    SELECT       APPENDIX 1   SYSTEM SPECIFICATIONS    TECHNICAL SPECIFICATIONS    External 100 240VAC to 24VDC adapter     Current max    Dimensions LxDxH in cm   Net weight   Environmental humidity     Environmental operating temperature     Environmental storage temperature     Fluid capacity     One pH input range   pH calibration    4 analog outputs   Analog output range   Output impedance   Air connection     USB connector     TRANSDUCER DATA    Maximum range   Operating temperature   Frequency range      10    3 3Amps at 24VDC    34 x25 x 15   5kg    20  to 80  RH  non condensing     15 to  40  C     4 to  70  C    500 ml bottles  We recom
18. is OFF     If the icon is green the Pressure regulation is ON  active    OFF ON    Zero key   This key is used to zero the output from the force transducer  Pressing ZERO  this key will reset the baseline of the chart traces without affecting the    calibrations     CHAPTER3 15    3 3 Pressure menu    The Pressure is controlled from this menu  Use the    SELECT    key to select  the line to be changed  When selected the line change color is blue  edit  mode   Use the    DEFAULT    button or the arrow keys to change the setpoint  value  Pressing    ENTER    will save the new setpoint     P1 Inlet setpoint is the pressure value that the regulator is setting on the  inlet side of the Myograph  Pressure range is 0 250mmHg     P2 Outlet setpoint is the pressure value that the regulator is setting on the  outlet side of the Myograph  Pressure range is 0 250mmHg     Pressure gradient This sets the speed at which the regulator modifies the  pressure  The pressure change range is 1 10mmHg sec     NOTE  P1 MUST NEVER BE LOWER THAN P2 TO PREVENT BACKFLOW IN THE  SYSTEM     The Pressure regulator and the Flow are also controlled from this menu   To turn the Pressure regulator ON press the Pressure ON button  When  activated the button changes to Green     In the Main Menu the pressure icon also turns to green when pressure  regulation is on              ON    To turn the Flow ON press the Flow ON button  When activated the key  changes to Green     In the Main Menu the Flow icon also tur
19. may occur     To start heating  press    HEAT ON        Covering the chambers with the chamber covers will expedite the heating   Place the temperature probe into the chamber for the first calibration to  monitor when the chamber has reached the target temperature  With the  covers heating will take about 20 to 30 minutes for the chambers and  transducers to reach 37  C  Once the chamber s  are heated and have  reached the target temperature  press    NEXT STEP        20 PRESSURE INTERFACE   USER MANUAL    FORCE CALIBRATION    Step no    8   2         5        Follow the weight calibration  procedure in the User Manual   Fill the chamber with double  distilled water    When ready go to next step     NEXT STEP    Step no    JBC  to  5     Place the calibration bridge on  the myograph  Place the tip of the  Bridge behind the glass cannula  Fixation plate    When ready go to next step     BACK NEXT STEP    Step no         68  5   6     Turn the heat on  Wait until  the temperature is stable     Temperature set point  37 0   C  Probe temperature  36 8   C    Step no   OBULU    Turn the heat on  Wait until  the temperature is stable     Temperature set point  37 0   C  Probe temperature  36 8   C    BACK               HEAT OFF    NEXT STEP       Step 4   This is the first step in the actual weight calibration process  A  4 digit number will be displayed in blue at the bottom of the screen  If  nothing has been disturbed during the heating process  the zero   O gram  calibration sh
20. mend a volume of 400ml to ensure optimal pressure  regulation    pH 0 14   manual with instructions on the display    BNC connectors at rear to connect to external data recorder    2 5Volt     lt 200 Ohm    maximum pressure 1 Bar     For downloading firmware and connection to PC      200MN    15 to  50  C   O   20Hz     APPENDIX 1 29    DMT A S   Skejby Science Center  Skejbyparken 152  DK 8200 Aarhus N  Denmark    Tel    45 87 41 11 00  Fax   45 87 41 11 01    www dmt dk  sales dmt dk  support dmt dk    DMT Asia Ltd    Rm 2402B  Great Eagle Centre  23 Harbour Road   Wanchai  Hong Kong S A R   P R  China    Tel    852 6621 8337  Fax   852 3020 7554    www dmt asia com  sales dmt asia com  support dmt asia com    DMT Asia  China office    Rm 28C  No  8 Dong Fang Road  Lu Jia Zui Financial District  Shanghai 200120   P R  China                  21 5425 1330  Fax   86  0  21 5877 0063    www dmt asia com  sales dmt asia com  Support dmt asia com    DMT USA  Inc    201 East Liberty Street  Suite 6   Ann Arbor  MI 48104  USA    Tel    1 770 612 8014  Fax   1 678 302 7013    www dmt usa com  sales dmt usa com  support dmt usa com       
21. n a 8  Chapter 2   Setting up the Pressure Myograph System                      1  211x 2ssssrsreaaeee ai eee iii 9  AL De Lume UPS Ed 10  2 2 Setting up the complete Pressure Myograph SYyStemM          erre eee ereee rene eee re rene ee eee rene ene sas ererecoca nana rerecoca saca rerecoca saca rerenaca 10  2 9 The first w  ight and pressure calibrations axis ga                                                                                                                                              10  2 4 External Pressure Myograph Unit connections                cccccccccssesseeceecccesseecenccecesseeneuccecesseesenececessserenececeseeerenececcsssarenececesseasenesees 11  ZO OM Me    P                       E a                11  2O EXTA      Ret OCIS nn 12  2 1 Connecting superfusion flow  optional  siga iii Sa 12  2 8 Connecting the Flowmeter   DMT 161FM  optional                    ccsccccccssesseeeeecccesseesenccecesseesenececesseerenececesseerenececesssarenececesssetesesees 12  2 9 Connecting the oxygen SUPDPIY ai Sa 12                 SUC ON        E           arene 13  2 11 3 Way Valve AUSENTE  ais da 13  Chapter 3   The Pressure Interface menus                       211ss   arrese eran iii eee iii iii 14  3 1 General description of how to navigate the touch screen san icnicestinccasdzasessdnsvmecnccunerocceccedeostussaseasdenennciadcudcansuavaunumeraettentuctuaduseuative 14       ON      SOFC    accent                                                             
22. ns to green when Flow is on     z    ON    16 PRESSURE INTERFACE   USER MANUAL    PRESSURE SET  P1 Inlet setpoint   120 0mmHg SELECT  P2 Outlet setpoint   85 0mmHg SELECT    Pressure gradient  5 0mmHg sec  SELECT    Pressure   Corr      DEFAULT    PRESSURE SET    P1 Inlet setpoint   120 0mmHg  P2 Outlet setpoint   85 0mmHg  5 0mmHg sec     Pressure   Corr   o  3     DEFAULT    Pressure gradient     PRESSURE SET   P1 Inlet setpoint   120 0mmHg SELECT  P2 Outlet setpoint   85 0mmHg SELECT  Pressure gradient     Pressure     DEFAULT    PRESSURE SET    P1 Inlet setpoint   120 0mmHg  P2 Outlet setpoint   85 0mmHg  5 0mmHg sec     Pressure gradient     Pressure     DEFAULT       3 4 Heat Menu    The Chamber and Bottle heating temperature are controlled from this menu  SET CHAMBER TEMPERATURE    To turn the heat on  or to change the temperature for the system  press the     HEAT    button in the Main Menu  The display will enter the Heat menu and  allow the user to change the system temperature  as well as turn heat on or Probe temperature  36 6   C  off  Pressing    DEFAULT    will automatically reset the temperature setpoint to  37  C  Manually change the temperature by pressing the up or down arrows   Pressing    ENTER    will save the new temperature setpoint     Temperature setpoint  37 0   C    LIMES  To turn the heat on  touch    ON    and the    ON    button will turn green  The  system will heat to the designated temperature setpoint  SET CHAMBER TEMPERATURE    Temperature 
23. oVIEW Il software program  optional   e Computer  optional     FlowMeter   Model 161FM  optional    e User manual    FlowMeter   Model DMT 161FM     e MyoFlow    FlowMeter software including drivers  CD ROM     Peristaltic pump for superfusion flow  optional    e Watson Marlow peristaltic pump    6 PRESSURE INTERFACE   USER MANUAL    CHAPTER 1   SYSTEM OVERVIEW    1 1 Pressure Interface    i gt  aaa i  LLELE  L  A  gt            SS        eas e 2x 500 ml Schott Duran bottles    Pressure Interface    touch display       Power indicator    Figure 1 1 The Pressure Interface   front panel       ON OFF switch e    Earth point        Inlet for oxygen car   bogen to pressurize  bottles  max 1 bar     24V DC port 9  for power       Connection port for RR EENE TA  Pressure Myograph unit           USB connection port    for computer   E       e Connection port     for pH meter  optional   Analogue recorder output   connection to  external data acquisition software  optional     Figure 1 2 The Pressure Interface   rear panel    CHAPTER 1    7    1 2 Pressure Myograph Unit    Horizontal adjustment screw                                     9 Micropositioner for vertical adjustment of glass  cannulas    Force transducer                Outlet pressure transducer 25 pin serial port for communication with  connection pipe                           Pressure Interface        Inlet 3 way valve  6    Outlet 3 way valve  e                 Outlet pressure transducer P2             Inlet pressure tr
24. of MyoVIEW Il refer to the MyoVIEW II Quick Installation Guide     2 3 The first weight and pressure calibrations  Prior to shipment the Pressure Myograph System has gone through two days of continuous testing  including final weight and  pressure calibrations  However in order to ensure that the Myograph is working at highest performance  DMT recommends that    new weight and pressure calibrations are performed before the first use     The weight and pressure calibration procedures are described in detail in chapter 3 6 1 and 3 6 2     10 PRESSURE INTERFACE   USER MANUAL    2 4 External Pressure Myograph Unit connections    This section illustrate how to connect the perfusion circuit  the superfusion circuit  oxygen supply  pressure supply and suction  device on to the Pressure Myograph unit     Suction pipe to vacuum         Perfusion inlet   1        Perfusion outlet P2              Extra perfusion outlet e     e  Superfusion inlet silas    Pipe for oxygen supply     eee       Access hole for reagents  buffer   temperature or pH probe    Figure 2 2 External Pressure Myograph connection    2 5 Connecting perfusion flow    The perfusion flow is controlled by the Pressure Interface  Instructions on how to control the perfusion flow are provided in  chapter 3     To connect the Pressure Interface to the Pressure Myograph unit  use the silicone tube  internal diameter 1mm  delivered with  the Pressure Myograph System     e Connect the    perfusion inlet    at P1 on the Pressure
25. or further information about recommended perfusion pump models  specifications  and prices     The superfusion circuit is connected to the Pressure Myograph Unit as illustrated in fig  2 1     2 8 Connecting the Flowmeter   DMT 161FM  optional     The DMT Flowmeter   Model 161FM is optional for flow measurements in the range of 15   4000ul min  The Flowmeter is easily  combined with any DMT Pressure Myograph System as illustrated in fig  2 1     Use of the Flowmeter permits control of intravascular pressure  pressure gradient and flow conditions with high accuracy  This  enables in vitro studies on vessel segments under conditions very close to those found in vivo  Adding the DMT FlowMeter DMT  161FM to the Pressure System also enables measurement of flow rate as a parameter  The Flowmeter data will make it possible  to record live traces of the flow dependent parameters as shear stress  Reynolds number and vascular resistance due to changes  in vessel structure and function in various physiological and pharmacological studies     To add the FlowMeter DMT161FM to the Pressure System the cable from the flow is attached to the    Flowmeter    port marked on  the backside of the Pressure Interface  The Software MyoVIEW Il will recognize the Flowmeter DMT161 as soon as it is connected    to the Pressure Interface and will be ready for use immediately  Please consult the MyoVIEW II video manuals linked in the help  menu of MyoVIEW II     2 9 Connecting the oxygen supply    Connect
26. ould be stable  The 4 digit number will not fluctuate and after  30 45 seconds    NEXT STEP    can be pressed  If the 4 digit number is not  stable  wait before pressing    NEXT STEP        Step 5   This step is the 2 gram weight calibration  At this step  place the  2 gram weight in the pan closest to the transducer  over the transducer   to simulate the stretch created by the contraction of a mounted muscle  preparation     Remember  a 2 gram weight in 8 90   vector is divided  and the transducer  will only detect 1 gram or 9 81 mN of force  The weight placement should  cause a positive increase in the 4 digit number  Wait at least 30 to 45  seconds for the applied force to stabilize before pressing    NEXT STEP     Only  when the 4 digit number has stabilized should you press    NEXT STEP        Step 6   This step verifies that the calibration was performed correctly  The     Force Chamber    reading should be 9 81   0 1 mN  If the    Force Chamber     reading is off by more than 0 1 mN  then remove the weight  press    BACK    to  return to Step 4  and repeat the calibration process  If the    Force Chamber     reading is satisfactory  then press    NEXT STEP    to end the calibration   After calibrating the force transducer  carefully remove weight  balance  and calibration bridge  The Pressure Myograph System is now ready for  longitudinal force measurements           FORCE CALIBRATION    sere  OVO0BOO    Make sure that the transducer is  not subjected to any force   Wh
27. pe    and the    outlet pressure transducer  connecting pipe     fig  3 4 A     Step 5   Close the inlet from the Pressure Interface to P1 by turning the handle on the 3 way valve towards the normal inlet way   The inlet to P1 now comes from the pressure manometer  See fig  3 3     22 PRESSURE INTERFACE   USER MANUAL       Figure 3 4 Disconnection of silicone tubes       Figure 3 5 Mounting of    Resistance tube shunt       CHAPTER3 23    3 6 2 1 Pressure calibration   PRESSURE CALIBRATION  Press    setting    in the    Main Menu    and then press    Pressure Calibration      sen  DODDO    Step 1   Press    NEXT STEP    when the system is ready  Follow the Pressure calibration  Procedure in the User Manual   Prepare the Myo Chamber for  Calibration   When ready go to next step     NEXT STEP    Step 2   Apply a pressure of 50mm Hg on the Pressure Myograph using the  PRESSURE CALIBRATION       Big Ben    Pressure Manometer  When the 3 blue lines of relative values    stabilize  go to    NEXT STEP      se 09000    The relative values should between 3300 to 3900  Apply 50 mmHg pressure on the  Myo  When the pressure reading  is stable  go to next step   Pressure P1  3672  Pressure P2  3615  Pressure P3  3676    BACK NEXT STEP    Step 3   Apply a pressure of 125mm Hg to the Pressure Myograph Chamber  using the    Big Ben    Pressure Manometer  When the 3 blue lines of relative PRESSURE CALIBRATION  values stabilize  go to    NEXT STEP       Step no  BO    The relative values shoul
28. setpoint  37 0   C  In the Main Menu the thermometer icon turns to green when the heat is on     Probe temperature  36 6   C              IMM    3 5 Timer Alarm Menu    SET TIMER  The Timer and Buzzer Alarm are controlled from this menu   Set Timer  00 00 00  Use  SELECT    to program the timer  The timer can be programmed in  intervals from 0 to 23 59 59  Hours   minutes   seconds   When Set Timer Timer   o    is selected use the up and down arrows to program the timer  The hours are    programmed first  and then use the    RIGHT    button to get to the minutes Buzz Alarm   o   and then the seconds    WIE   Pressing    ENTER    will save the programmed time     SET TIMER    Set Timer  00 00 00 Set Timer  00 00 00                   Set Timer  00 00 00       To start the    Timer    press the Timer    ON    button  When the Timer is on  the  button turns green  When the timer reached zero it switches automatically SetTimer 00 00 00        to    OFF       gt      If the Buzzer is set ON it will make a tone when the timer reaches zero     Buzz Alarm     CHAPTERS 17            activate the    Buzzer Alarm    press the Buzzer Alarm    ON    button  When  ON the button turns green and in the Main Menu the bell icon also turns SET TIMER  green     Set Timer  00 00 00       O    ON Buzz Alarm   on     3 6 Settings Menu    The    Settings Menu    contains several sub menus  These sub menus SETTINGS  include     PRESSURE CALIBRATION  pH CALIBRATION  optional   RECORD OUT SELECT  INTERF
29. tact us immediately and the matter will be pursued as soon  as possible  If the packing material appears damaged  please retain it until a possible claim has been settled     We recommend that you store the packing material for any possible future transport of the Pressure Myograph System  In case  of transport  and the original packing material is unavailable  please contact the DMT Sales Department for advice and packing  instructions  After unpacking your new Pressure Myograph System  please use the following list to check that the system is  complete     Pressure Myograph Unit   e Force calibration kit  including    bridge        balance    and 2 gram weight   e Resistance tube shunt  calibration tube   e Two glass cannulas  e Chamber cover    Pressure Interface   e Pressure Interface to chamber cable with a temperature probe  e Power adapter  the shape of the AC plug varies by country  be sure that the plug has the right shape for your location   e USB cables for PC connection   e Atlas Copco Cablair hose for main pressure supply  e Two 500m  Schott Duran Bottles with specially designed screw caps    Riester    Big Ben    Pressure Manometer    Accessories   e Allen key  e Spare screws  e Small screwdriver  e Suture wire  e Tube of high vacuum grease  e Tube of grease for linear slides  e Silicone tubing  internal diameter Imm    Computer  hardware  software  amp  manuals   e USB digital camera  optional   e CD with DMT Pressure Myograph Systems user manuals  e CD with My
30. taltic pump to enable superfusion flow and pH probes for measuring pH in the superfusion buffer  Also the diagram  includes the DMT Flowmeter DMT 161FM  The Flowmeter enables flow measurements in the range of 15 4000ul min     Fig  2 1 is intended as an overview of the setup of the Pressure Myograph Systems  A detailed step by step description is provided  in the remaining sections of Chapter 2     The Pressure Myograph System will work with a range of different microscopes  the DMT Microscope  Zeiss Axio Vert A1  Nikon  TS 100F  Motic AE2000 or your own inverted microscope of choice     NOTE  THE MICROSCOPE IS ESSENTIAL FOR THE DIMENSIONAL ANALYSIS OF THE MOUNTED VESSELS     The DMT Microscope is a standard inverted microscope with a built in digital CCD USB camera  The DMT Microscope is very  effective and recommended in setups requiring standard image analysis     The Zeiss Axio vert A1 is an inverted microscope that provides the highest degree of flexibility in image analysis  The Zeiss  Axio vert A1  a Nikon TS 100F and Motic AE 2000 enable the use of a wide range of objectives and facilitate sophisticated    fluorescence techniques     Contact the DMT Sales Department for further product information and prices on any optional equipment  including microscopes     CHAPTER2 9    2 1 Setting up step by step    NOTE  IF YOU HAVE PURCHASED A COMPUTER FROM DMT IN CONJUNCTION WITH YOUR PRESSURE MYOGRAPH SYSTEM  THEN THE MYOVIEW Il DATA ACQUISITION SOFTWARE HAS ALREADY BEEN INST
31. ting that the interface is in edit mode  and waiting for input  When    DEFAULT    is chosen  a default value will be PRESSURE SET  displayed  To change the numeric value for the selected parameter touch    the up or down arrow keys  P1 Inlet setpoint   120 0mmHg SELECT    P2 Outlet setpoint   85 0mmHg SELECT  Pressure gradient  5 0mmHg sec  SELECT    Pressure     FIOW     DEFAULT    Once the desired setting has been chosen  pressing    ENTER    will save the  selection  and the new value will be stored in memory  The selected line will PRESSURE SET  revert to black  Pressing the white    X    in the red box will exit the menu and    return you to the Main Menu  P1 Inlet setpoint   120 0mmHg SELECT    P2 Outlet setpoint   85 0mmHg SELECT  Pressure gradient  5 0mmHg sec  SELECT    Pressure   Flow     MULTI INTERFACE    Multi Myograph System  Model 110P 111P  Software Revision 01 02  Date  Nov  27 2011    3 1 Power up screen    After turning on the Pressure Interface  an    Introduction    screen appears   It displays the software versions number  While this screen is displayed the  system is auto initialising     After a few seconds  the    MAIN MENU    display will appear        3 2 Main Menu    The Main Menu gives a good overall picture of how the Pressure Interface is working  It displays values for the input pressure   output pressure  force  pH  probe temperature and time  and the status of systems including the time  heat  flow and pressure     14 PRESSURE INTERFACE   US
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Motori Diesel OEM 6081 da 8,1 l ( —199,999)    iOptron SmartStar -PR EQ User's Manual  オットーボック製品 取扱説明書 ② ( 製品篇 ) オットーボック カスタム装具  Service Manual Emperor Ninja (ENG)  Axor 3-jet User's Manual  Confort AT-L  LG VS880 Brochure (Spanish)  Samsung HMX-S10BP kamera cyfrowa Instrukcja obsługi    Copyright © All rights reserved. 
   Failed to retrieve file