Home
        XDP Explosion-Proof Display Package
         Contents
1.                                                    4 99  127     SIDE VIEW    9 14  232                                                                                                                                                                    Figure 2 1  XDP Mounting Dimensions       Installation    2 9    July 2003                                                                                        NOTE  This unit requires an external power    disconnect device such as a switch or  circuit breaker  The disconnect device  must be marked as such  clearly visible   directly accessible and located within 1 8  m  6 ft  ofthe XDP                 LEFT SIDE CONNECTIONS       0 987    654    QUE  v3       C NICNC    NU C NC    NO C NC   RTN RX T                                                 00000000000                                                                An                                     ory Jumper                                                                                                                                                                            AC POWER INPUT   TB1       Pin No     Description       L    Line Input  100 240 VAC        N    Line Neutral          G    Line Ground                                                                                                      FUSE       Rating 1 25 Amp       Voltage   250 V       Size 5 x 20 mm          Type T             Ground Terminal    RIGHT SIDE CONNECTIONS          Q  Mug  Gra 
2.                                          4 17  LOCK Menus A en e ede tes do Ge 4 17  Lock NA per eem Uu I Eis 4 18  GN 3 Now das ana nra aa ea eia Kup Cr dag Rete REN a ae Nuus dte RUNS 4 19  Calibrating the System    ouem nen te rere ed e Ed ne 4 20  Initial Preparation  zie es er ge Ugh dd ba qur nc Rp ER angle RE 4 20  Performing a Zero Calibration                                                 4 21  Performing a Span Calibration                                                4 23  Viewing and Clearing the Cal Drift                                           4 24    Chapter 5  Advanced Programming    Introduction nA A Bess ER 5 1  Programming Levels                                                            5 1  Setup Level Prosrammins        Ie a I Re ARS ae nahe ob kl 5 2  Configuring the Display                                                      5 2  Accessing the Setup Menu                                                    5 4  Configuring the 4 20 mA XDP Analog Output                                     5 5  Configuring the Low and Low Low Alarms                                      5 8  Testing the XDP Alarm Relays                                               5 12  Factory Level Programming                                                     5 14  Entering the Sensor Serial Number                                            5 14  Configuring the 4 20 mA Analog Input                                         5 15  Entering Factory Calibration Data                    
3.                           5 16  Loading Software  sus exe sr GN Na KG Hal ai isch 5 18  Exiting the User Program    22238 a nenn speed one DAA 5 19       July 2003    Table of Contents  cont      Chapter 6  Specifications    Performance Specifications   Lai du Baas Rn ee  Physical Specifications                                                           Functional Specifications                                                         Environmental Specifications                                                      Transmitter Specifications  dat ee dee eee en EUER EO    Appendix A  CE Mark Compliance    Introduction  Fuss Se dss Sr ri u an baban dads aa  EMC Compliance    rss Den ne Dae ns ER  EVD  Compliance vies savers ar re iD II REG a a    Appendix B  Additional Wiring Connections    Introduce a t atur c    Connecting the System Outputs                                                      Wiring the System Analog Output   J5                                           Wiring The System Alarm Relay   J2                                            XDP  Alar mR  lays    ot dt eee cs e E eere ee s    Appendix C  Menu Maps    Operator Level  Cal  and  Disp  Menu Map                                          Operator Level  Opt  Menu Map                                                   Setup Level Menu Map    eel su ce eee I Rx E ee du eher n  Factory Level Menu Mapua ha Gan LG med ee dept letter PA       July 2003       Warranty    Return Policy    Each instrument manufactured by 
4.                         1 3 Namur Analog Output                   4 9  Gas Ranges                            1 3  P  H Padlock Symbol                        3 3  Hazardous Location Rating                1 1 Password  Factory Level                       5 14  l Operator Level                       3 4  Infrared Keypad   i  Lente dues ts 3 2 pda M D ET 24  Installation Site User Program                        5 1  Ms a a a NG AN 2 2 Photograph  XDP                       1 2  prae  Requirements                         2 2 Powenng E RAN IM ML    Installing the Sample System               2 3 Frevenune TINO PAO IEIR eiir E    Programming  K Factory Level                       5 14  Keypad  i  tees Mat an SMS 3 2  6 2 R  L Rating  Environmental                    1 1  Return Policy                         1 vii  Line Power Requirements                 6 1  Loading New Software                  5 18  Locking the User Program                4 17  LVD Compliance                2 1 2 3  A 1       2 Index    July 2003    Index  cont         S T  Sample System Teminal Block J7                       B 2  Installing                             2 3 Temperature  Ambient                    2 1  Mounting                            2 3 Testing  Typical oi iis  tt a 2 3 Analog Input                        5 15  Serial Number  Entering                 5 14 Low Alarm      2  saaan nono 5 8  Serial Port Low Low Alarm                      5 10  Baud Rate ss trust nf 4 16 System Analog Output        
5.                        Onfiguring                          5 2  Generator Coolant                     1 3 Contrast Adjustment                    4 4  D a seen nern 13 Desaster 3 2  B Output Devices                        5 3  Run Mode Screen                      3 3  Backlight  Setting Up                    4 5 Switching Windows                    4 3  Basic Features                          1 1 Unis 104 2  a Bere  5 3  Disconnect Device                   2 1  2 3  Display  See Digital Display  Display Windows  Switching               4 3  E  EMC Compliance                       A 1  Environmental Rating                    1 1  Error Messages                          3 5  Exiting the User Program                 5 19       Index    July 2003  zZ  Je    Index  cont      F M  Factory Menu Menu Map  4 20 mA Analog Input                 5 15 Factory Level                        C 4  Calibration 32 2 0 ya 5 16 Operator Level                   C 1  C 2  Loading Software                     5 18 Set  p Level    ri vv C 3  Serial Number                       5 14 Messages  Error                        3 5  Factory Level Programming               5 14 Mounting  Fail Safe Alarm Mode                    4 7 The Sample System                   2 3  Fault Alarm  System   Setting              4 6 The Transmitter                       2 3  Flow Rate  Calibration Gases             4 20 THeE DR See ren 2 3  Busen ein ger aa ud 6 1 Mounting Dimensions  XDP              2 9  G N  Gas Mixtures   
6.                   Use  i   and       to position the  cursor under the desired char   acter  Use fi   and  E  to enter  a time between 0 and 99 min    When done  press  ENTER      Press  ESC      You are now back at the User  Menu     Press  ESC  as many times as necessary to return to normal RUN  mode  or proceed to the appropriate section of this chapter to continue    programming the XDP     The system fault alarm is not normally used for XDP applications                                            Fault Alarm However  if you do intend to use this alarm relay  the fault alarm type  may be specified by entering the User Program  as described on page  3 4  and proceeding as follows   Note  See Appendix B  Additional Wiring Connections  for  instructions on wiring the system fault alarm   Cal   Opt   Disp Use the  i   and       keys to  select  Opt  and press   ENTER    Disp Use the  i   and  E  keys to  select  User  and press   ENTER    Lock Menus  Note     Lock All    and    Versions    menu options also appear on the  above list   4 6 Setup and Calibration    July 2003    Setting the System  Fault Alarm  cont            User       Analog Output    Contrast                If necessary  use the  I   and   E  keys to select  Fault Alarm   and press  ENTER      Note  The    Backlight    and    Comm    options are also available in    the above list box     The XDP fault alarm has two possible operating modes       Non Fail Safe  The alarm is wired to the normally open  NO  and
7.              Now  the XDP analog output is driven to exactly 20 mA     Press the  I   or     key until  the meter reads exactly 20 mA   When done  press  ENTER      Press  ESC  to return to the  setup submenu     To continue programming the setup menu  proceed to the appropriate  section for instructions  Otherwise  press  ESC  four times to return to    live measurement mode     To enter the factory calibration data  access the setup menu as  described on page 5 4 and complete the following steps           Range Cal        Serial Number  4 20mA Input  Factory Calibration                 Use the  i   and  E  keys to  select  Factory Calibration  and  press  ENTER      At the next prompt  select the range to be calibrated  H2 AIR is used    as an example in this manual            Gas Curve        H2 CO2  AIR CO2                Use the  i   and  E  keys to  select  H2 AIR  and press   ENTER      At the next prompt  enter the number of data points available for    calibrating the chosen gas curve           Gas Curve  H2 AIR    Enter   of Points                         Use  i   and       to position the  cursor under the desired char   acter  Then  use  i   and  E  to  increment the value  Press   ENTER  when done        Advanced Programming    July 2003    Entering Factory  Calibration Data  cont      Caution   Incorrect entry of the calibration data points will cause  the instrument to give unreliable results     At the following prompt  enter the  mA  pent  values for each of 
8.            Note  The Aout Type  Zero Setpoint  and Span Setpoint options are  also available in the above list box     At the following prompt  the system analog output of the meter is  driven at exactly the low end of the output range              User Aout Zero  Trim Press the       or  E  key until  Zero Trim your output device reads the  correct value  When done   press  ENTER      A    xx V                      User  Aout You are now back at the sys   Test tem analog output setup menu   Span Trim                      Either press  ESC  until you return the User Menu or proceed to the  appropriate section to continue setting up the system analog output        Setup and Calibration    4 13    July 2003    Setting the Span Trim    Enter the system analog output menu  as described on page 4 8  and    proceed as follows     IMPORTANT  Before proceeding with this section  make sure that  the output percentage being trimmed has been  entered in the Test menu on page 4 12           User  Aout  Test       Zero Trim             Use the  i   and  E  keys to  select  Span Trim   and press   ENTER      Note  The Aout Type  Zero Setpoint  and Span Setpoint options are  also available in the above list box     At the following prompt  the system analog output of the meter is  driven at exactly the high end of the output range              User Aout Span Trim                   User  Aout  Test       Zero Trim                Press the  i   or     key until  your output device reads the  
9.          4 12  Cable u    u san 2 6 XDP Alarm Relays                    5 12  Data Length  oe a Uni unes 4 16 XDP Analog Output                    5 6  Node ID     set eae crinii e 2 40 use ue 4 15 Tool  Wiring                            2 4  Parity  nn 4 16 Transmitter  Setting UD  c nee 4 15 Documentation                        1 1  SLOP  Bits    au au are nr 4 16 Mounting                            2 3  Type  an 4 16 Requirements                         1 2  WIEINS ap cue tot and Gam sans EG 2 6 Specifications                         6 2  Setup Menu WINS ica Dr et da ess 2 5  4 20 mA Analog Output                5 5 Troubleshooting                         3 5  ACCESSING 24 u a chine iln Pp pea 5 4  Low and Low Low Alarms              5 8 U  Site Preparation                         2 3 User Program  Spa CASE  O 4 20 Acces na AT te de 3 4  Speuifications ae E mei Vat Ed    A er GO SCR 5 19  Environmental  PE 6 2 Factory Menu Map                    C 4  Functional    sa po ee ne NG 6 2 a AA 51  evels  Perfomance ET 6 1 Locking Unlocking                     4 17  Physical MET ee 6 1 Moving Back One Level                 4 1  Transmitter           ES 6 2 Operator Menu Map               C 1  C2  O ARNG SUC jupe so p P  ssword8s        rl Ir ee 5 1  System Functions  Wiring                 B 1 Setup Menu Map                      C3  System Overview                       1 2 Done sa nes ren 5 18  W  Warm Up Time                          3 2  Watranty   i  cede c tem p eR  1 vii  W
10.        see text for available options           Stop Bits     ENT    Comm Type           Figure C 2  Operator Level  Opt  Menu Map       Menu Maps    C 2    July 2003          ENT                            Cfg                                                    H2 AIR      may also be H2 CO2 or AIR CO2                                Trip Point trip percent                      0 Fault Alarm t 0 Analog Output        system functions     Rome am  range zero H2 AIR XDP Fault Alarm  Gas Curve Relay1                                                    Figure C 3  Setup Level Menu Map                                                       AY  AY  Process Ca Ra    Process Cal Relay                                     Zero Span Relay          Menu Maps C 3    July 2003       password              gt   cx           20 mA trim value    4 mA trim value    mA Input Test                                           sensor number                            BS  cx    4 20mA Out         of points ENT    ENT     mA  Pent  data    ENT       H2 AIR                   AIR CO2       ENT                                  Load Slot                               91 52 pcnt       Figure C 4  Factory Level Menu Map       Menu Maps    C4    July 2003    Index  A C  Accessing the Setup Menu                5 4 Cable  Serial Port                        2 6  Accessing the User Program               3 4 Calibration  Active Curve  Selecting                   4 2 DIE  ER 4 24  Adjusting Display Contrast             
11.     UP Arrow DOWN Arrow  Figure 3 1  The Digital Display Window       The XDP is factory configured to display H2 AIR as the active  display range  However  changing the active gas curve automatically  updates the display to show the new curve  See Chapter 5  Advanced  Programming  to reconfigure the display  if necessary     Note  The XDP 5 infrared keypad permits programming of the  instrument through the glass faceplate without removing the  cover  Thus  all programming procedures may be performed  while the unit is installed in a hazardous environment        3 2    Operation    July 2003              _ _ _ _ __ _ __ _ _ UU    Initial Screen Displays Immediately upon power up  the digital display shows the following  sequence of informational screens  as the meter performs its internal  systems checks           BSeries Loader v1 03 This is a typical example of  Boot is Flash  the first initialization screen     Program CRC valid   Booting from Flash                            Scanning Hardware This is a typical example of  Found Image Slot 1 the second initialization  screen     Downloading Slot 1                      Scanning Hardware This is a typical example of  Initializing Machine the final initialization screen   Initializing Ul                      If the meter passes all of its internal system checks  the following data  screens appear           1 H2 AIR This message appears for a  few seconds  until the sensor  begins outputting a valid gas    mA Input Under Range
12.     Use the  i   and  E  keys to  select  Trim  and press   ENTER      At the following prompt  the XDP analog output of the meter is    driven at exactly 4 mA           4 20mA Out Trim                Press the  i   or  E  key until  your output device reads  exactly 4 mA  When done   press  ENTER      At the following prompt  the XDP analog output of the meter is    driven at exactly 20 mA        4 20mA Out Tr  Trim                      4 20mA Out       4 20 Out Range  Test                      Press the  i   or  E  key until  your output device reads  exactly 20 mA  When done   press  ENTER      Press  ESC  to return to the  setup menu     To continue programming the setup menu  proceed to the appropriate  section for instructions  Otherwise  press  ESC  four times to return to    live measurement mode        Advanced Programming    5 7    July 2003    Configuring the Low and To configure the XDP Low and Low Low alarms  access the setup  menu as described on page 5 4 and complete the following steps     Low Low Alarms             Range Cal  4 20mA Out             Use the  i   and  E  keys to  select  Alarms  and press   ENTER      Note  The other six alarm relays listed in the alarm setup menu are    discussed in the next section                                                                                                              The Low Alarm 5   Alarms Use the       and  E  keys to  select  Low Alarm  and press   Low Low Alarm  ENTER    Normal Alarm   Alarms Low Al
13.     content signal                 1 H2 AIR 8 The temporary absence of  valid signal triggers the Low  Low alarm  and this screen    Low Low H2 pcnt  XX X  appears                       Note  The padlock icon at the upper right hand corner of the above  displays indicates that the User Program is locked with  password protection        Operation 3 3    July 2003       Starting the Sample Gas    Accessing the User  Program    Initiate the flow of sample gas to the transmitter and allow a few  minutes for the system to reach equilibrium  If the sample gas has a  hydrogen concentration higher than the Low Low alarm setting  the  Low Low alarm resets  and the following data screen appears              1 H2 AIR 8    91 52    This is a typical display  The  actual concentration of your    pent  sample gas is shown here                    Note  Allow approximately 30 minutes for the sensor to reach  temperature equilibrium  to ensure accurate readings    Notice that the display shows the slot number and the input range on  the top line  and the current live value along with the unit type are  displayed in the center of the screen     Note  The    Slot Number    refers to the location of the applicable  circuitry in the digital controller module  Slot O is the main  circuit board  slot 1 holds the XDP circuit board  and slots 2   4 may be used for optional circuit boards     The XDP software permits the operator to configure the meter for his  specific requirements  To accomplish thi
14.    4 4 Factory Setup                        5 16  Alarms Fail  ie   uei ara 4 24  Fail Safe Mode                       4 7 Flow Rate  Gas                       4 20  Fault  Setting                         4 6  A Guts eaba unl erie 4 20  Low Low Setup                      5 10 A rss 3 1  4 20  LOW  Setup  src Pas 5 8 Preparation                          4 20  Setpoints                            1 4 Span Point                          4 23  SEIS naa ne BS 1 3 AN PAA 4 20  Specifications                         6 2 Zero PORT  Les hada re 4 21  System Fault                         B 1 CE Mark Compliance                 2 4  A 1  MY  DOS  PP T 1 3 Cleaning the XDP                       3 1  YA a eee oris edere 2 5 Common Problems  Preventing             3 1  XDP Relays                         5 12 Configuring  Digital Display               5 2  Ambient Temperature                    2 1 Connecting  Analog Output  System  See Part Name  Setting Up                           4 8   Contrast  Adjusting                      4 4  Span Setpoint                       4 11  span Tamm    4 14 D  D Rl Ie Diagnostics sos eser Le eai fu ES  3 5  Types erue UV ERR LER e DUT 4 9 Digital Display  WAITING es Kaban tot aed a as Pe bs det Pa 2 6 Aonya RG 53  ge      ooo Sean  D NE ae ee T Backlight Setup  eei a 4 5  Zero Eis Lan dresse reat d a Da neat 4 13 Boot  x  oot Up Screens                      3 3  Analog Output  XDP                     5 5 CD NES ata ata P er 3 2  Application    ponents    
15.    and        keys to select  Range  and  press  ENTER      Press  ENTER      Use the  i   and  E  keys to  select the desired gas mixture  to be monitored and press   ENTER    H2 AIR is used as  an example in this manual      Press  ESC  three times to  return to the main menu     Press  ESC  as many times as necessary to return to normal RUN  mode  or proceed to the appropriate section of this chapter to continue    programming the XDP        4 2    Setup and Calibration    July 2003    Switching Display If multiple XDP display windows have been configured  see Chapter   Windows 5  Advanced Programming  for instructions   the display may easily  be switched between these windows from the within the User  Program  To accomplish this  access the User Program as described  on page 3 4 and proceed as follows     Note  As a shortcut  the display windows may be switched from  normal run mode by using the        and  E  keys                                                                             Cal Opt   Disp   Use the  i   and       keys to  select  Disp  and press   ENTER     Cal Opt All available display windows   1  H2 AIR pent are listed  Use the  1   and  E     keys to select the desired win   POI ES  dow and press  ENTER     Cal Opt   Disp   Press  ESC  to exit the User  Program and return to mea   surement mode    1 H2 AIR After pressing  ESC  at the  pent previous prompt  the chosen  9 1   52 display appears                       Note  Although the digital display ca
16.   1  Turn the system power ON and allow at least 30 minutes for the  sensor to reach temperature equilibrium     2  Begin the flow of the zero gas to the inlet port of the transmitter  at  about 250 cc min  0 5 SCFH  and atmospheric pressure        4 20    Setup and Calibration    July 2003       Performing a Zero  Calibration    To perform a field calibration of the meter  access the main menu of  the User Program  see page 3 4  and proceed as follows             EJ Opt Disp                         Range   Cal         Slct Cal Menu Item                   Range Cal  Manual Cal             View Cal  Drift          If necessary  use the  1   and        keys to select  Cal  and  press  ENTER      Press  ENTER      Use the  i   and       keys to  select  Cal  and press   ENTER      If necessary  use the  I   and   E  keys to select  Manual Cal   and press  ENTER      Note  A    Clear Cal Drift    option also appears on the above list              Gas Curve   H2 AIR    H2 CO2                         Use the  i   and  E  keys to  select the calibration gas curve  and press  ENTER      Note  The H2 CO2 calibration gas is used as an example here           Gas Cur  Manual Cal    Span                Use the  i   and  E  keys to  select  Zero  and press the   ENTER  key        Setup and Calibration    4 21    July 2003  EEE EE An zz    Performing a Zero  Calibration  cont      At the following prompt  enter the percentage of target gas  H2 in this  example  in the zero gas mixture  usua
17.   Advanced Programming       5 9    July 2003                                                                                                                                                          The Low Low Alarm E  Alarms Use the       and  E  keys to  Low Alarm select  Low Low Alarm  and  Low Low Alarm PE  Normal Alarm  Ala  Low Low Alarm Use the  i   and  E  keys to  select  Trip Point  and press  Test  ENTER    Fail Safe  At the next prompt  enter the percentage of hydrogen in air that  should trip the low low alarm  typically 80 00     Alarms  Trip Point Use  i   and       to position the  cursor under the desired char   acter  Then  use  i   and  E  to  XXX XX    increment the value  Press     ENTER  when done   Ala  Low Low Alarm Use the  i   and  E  keys to  Trip Point select  Test  and press  TES  Fail Safe  Alarms Low L  Test Use the  i   and  E  keys to  select  Trip  and press  REA  ENTER    After verifying that the alarm has actually tripped  continue to the  next prompt   Alarms Low L  Test If necessary  use the  i   and  Trip  E  keys to select  Reset  and  use   5 10 Advanced Programming    July 2003        The Low Low Alarm  cont   The program alternates between the two previous prompts  each time   ENTER  is pressed  To leave the test loop  press  ESC      Note  Upon leaving the test loop  the alarm is automatically reset to  its normal state              Ala  Low Low Alarm Use the  i   and     keys to   Trip Point select  Fail Safe  and press   ENTER   
18.   Lock All    and    Versions    menu options also appear on the  above list              User Use the       and  E  keys to  Fault Alarm select  Analog Output  and  Analog Output press  ENTER     Contrast                   Note  The    Backlight    and    Comm    options are also available in  the above list box     The following analog output features may be set in this menu       Analog Output Type  configure the system analog output for a  0 20 mA  4 20 mA or 0 2 V output device      Zero Setpoint  enter the live measurement value that corresponds  to the low end of the system analog output range       Span Setpoint  enter the live measurement value that corresponds  to the high end of the system analog output range minus the zero  setpoint       Test  manually generate a system analog output with a known value       Zero Trim  adjust the system analog output to equal the low end  of the recorder range        4 8    Setup and Calibration    July 2003    Setting Up the System  Analog Output  cont      Choosing the System  Analog Output Type      Span Trim  adjust the system analog output to equal the high end  of the recorder range minus the zero setpoint     Proceed to the appropriate sub section to program the desired system  analog output feature     Enter the system analog output menu  as described on page 4 8  and  proceed as follows              Use the  i   and  E  keys to  select  Aout Type   and press   ENTER      User Aout       Zero Setpoint             Span Set
19.   Low Low Alarm select the  Normal Alarm  for    NormakAlarm testing  and press  ENTER      XDP Fault Alarm                         Alar  Normal Alarm If necessary  use the  i   and   E  keys to select  Test  and    Poe                                  Alarms Norma  Test Use the  i   and     keys to  select  Trip  and press  Reret  ENTER                        5 12 Advanced Programming    July 2003    Testing the XDP Alarm After verifying that the alarm has actually tripped  continue to the  Relays  cont   next prompt           Alarms Norma  Test If necessary  use the  i   and    Trip      keys to select  Reset  and  press  ENTER                        The program alternates between the two previous prompts  each time   ENTER  is pressed  To leave the test loop  press  ESC      Note  Upon leaving the test loop  the alarm is automatically reset to  its normal state                             Alar  Normal Alarm Press  ESC     Alarms If desired  select another alarm   Low Low Alarm relay for testing  and repeat the  above instructions    Normal Alarm   XDP Fault Alarm                      When the testing has been completed  press  ESC  to return to the  setup menu  To continue programming the setup menu  proceed to the  appropriate section for instructions  Otherwise  press  ESC  four  times to return to live measurement mode        Advanced Programming 5 13    July 2003    Factory Level  Programming    Entering the Sensor  Serial Number    To program any of the features descr
20.   Test                                                                Alarms  Fail Safe Use the       and     keys to   Non Fail Safe select desired alarm type and   Fail Safe press  ENTER     Ala  Low Low Alarm Press the  ESC  key to leave   Trip Point the    Low Low Alarm    sub   menu    Test   Alarms Press  ESC  to return to the  setup menu     Low Alarm    Low Low Alarm    Normal Alarm                         To continue programming the setup menu  proceed to the appropriate  section for instructions  Otherwise  press  ESC  four times to return to  live measurement mode        Advanced Programming 5 11    July 2003    Testing the XDP Alarm To test the remaining six XDP alarm relays  access the setup menu as  Relays described on page 5 4 and complete the following steps           Range Cal  4 20mA Out    Use the  i   and  E  keys to  select  Alarms  and press   ENTER                  Note  To configure the low and low low alarms  see the previous  section     The following XDP alarm relays may be tested in this submenu       Normal Alarm   e XDP Fault Alarm     Gas Curve Relay 1  see Appendix B for wiring instructions     Gas Curve Relay 2  see Appendix B for wiring instructions     Process Cal Relay  see Appendix B for wiring instructions       Zero Span Relay  see Appendix B for wiring instructions     Testing of the  Normal Alarm  relay is used here as an example  Use  the same procedures to test any of the other alarm relays        Alarms Use the       and  E  keys to
21.   common  C  contacts  and the alarm 1s  Off  until a fault condition  energizes the alarm relay to close these contacts and trigger the    alarm       Fail Safe  The alarm is wired to the normally closed  NC  and  common  C  contacts  and the alarm relay is energized to keep the  alarm    Off    during normal operation  A fault condition de   energizes the alarm relay and triggers the alarm by allowing these    contacts to close           User Fault  Alarm  Non Fail Safe    Fail Safe                      User       Analog Output       Contrast                Use the  i   and  E  keys to  select the desired alarm mode  and press  ENTER      You are now back at the User  Menu     Press  ESC  as many times as necessary to return to normal RUN  mode  or proceed to the appropriate section of this chapter to continue    programming the XDP        Setup and Calibration    4 7    July 2003    Setting Up the System  Analog Output    The system analog output is not normally used for XDP applications   However  if you do intend to use this output  it may be configured by  entering the User Program  as described on page 3 4  and proceeding  as follows     Note  See Appendix B  Additional Wiring Connections  for  instructions on wiring the system analog output                                      Cal   Opt   Disp Use the  i   and       keys to  select  Opt  and press   ENTER    Cal Disp Use the       and  E  keys to  select  User  and press   ENTER    Lock Menus                   Note   
22.  19    Chapter 6    Specifications    Performance Specifications                                6 1  Physical Specifications                                    6 1  Functional Specifications                                  6 2  Environmental Specifications                              6 2    Transmitter Specifications    July 2003    Performance  Specifications    Physical Specifications    Accuracy    0 5  of full scale    Ambient Temperature Range    10 to 65  C  14 to 149  F     Ambient Relative Humidity  maximum    35   65  C  50   40  C  65   20  C  75   0  C    Linearity    0 5  of full scale    Reliability   50 000 hours MTBF minimum    Resolution    0 1  of full scale    Response Time     90  of reading in 50 seconds    Stability   zero drift less than 0 5  of full scale in 24 hours    Temperature Drift   less than 0 5  of full scale per 100  F    Enclosure     Panel Mount  Explosion Proof   Class I  Division 1  Groups B  C and D     Fuse   1 25 A  250 V  5x20 mm  IEC Type T    Power Input   100 240 VAC  50 60 Hz  40 W max     Power Output to Transmitter   24 0   2 0 VDC  1 A max     Note  The power supply meets CISPR 55022 and CISPR 55014   Level B EMI requirements and IEC1010 1 safety standards        Specifications    6 1    July 2003    Functional  Specifications    Environmental  Specifications    Transmitter  Specifications    Alarms  terminal blocks J2  J3      Dual general purpose contacts   Contact rating  2 A  28 V  SPDT    XDP Fault  J3   fail safe type  X
23.  363      Draw    ions    XDP Wiring Connecti    Figure 2 2       Installation    2 10    Chapter 3    Operation    Introduction  ss ss kaum xanax nn 3 1  Preventing Common Problems                             3 1  Cleaning the Enclosure                                   3 1  Powering Up the System                                  3 2  Accessing the User Program                               3 4  Exiting the User Program                                  3 5    Error Messages    July 2003    Introduction    Preventing Common  Problems    Cleaning the Enclosure    The XDP is an easily operated monitoring device  Make sure that the  system has been installed in accordance with the instructions given in  Chapter 2  Installation  before applying power  This chapter includes  discussions of the following topics       preventing common problems    powering up the system    programming the meter      taking measurements    IWARNING    To ensure safe operation of the XDP  it must be  installed and operated as described in this manual   In addition  be sure to follow all applicable local  safety codes and regulations for installing electrical  equipment     The complexity of accurately measuring the target gas mixtures  dictates that some basic precautions be observed in using the XDP   Failure to observe these simple procedures is often the cause of some  common problems  Compliance with the following points will help to  eliminate such problems       Calibration of the analyzer shoul
24.  E  to  increment the value  Press   ENTER  when done     At the next prompt  enter the percentage of hydrogen in air that  should generate a mA output of 20 00  typically 100 0040         4 20mA Out 4   Span    XXX XX                      Use  i   and       to position the  cursor under the desired char   acter  Then  use  i   and  E  to  increment the value  Press   ENTER  when done        Advanced Programming    5 5    July 2003    The Range Submenu  cont               4 2  4 20_Out_Range Press  ESC  to leave the  4 20  Out_Range  submenu    H2 CO2   AIR CO2                The Test Submenu          4 20mA Out  4 20_Out_Range         Use the  i   and  E  keys to  select  Test  and press   ENTER                     At the next prompt  enter the mA value to be output as a test signal           4 2  Enter mA Value Use  i   and       to position the  Enter mA Value cursor under the desired char   acter  Then  use  i   and  E  to                         XX XX     increment the value and press  a  ENTER  when done   4 20mA Out En  Test Measure the XDP analog out     Enter mA Value put test signal and compare it  to the displayed value  When    done  press  ENTER         XX XX                      The above two prompts repeat so that more than one test value may  be tried  When the testing is complete  press  ESC  from the first  prompt to leave the Test submenu        5 6 Advanced Programming    July 2003    The Trim Submenu          4 20mA Out       4 20_Out_Range  Test         
25.  ENTER                        Setup and Calibration    4 15    July 2003    Setting Up the Serial  Port  cont      At the following prompt  choose one of the following baud rates  300   1200  2400  9600  19200  38400  57600  or 115200           User Node  Baud Rate                9600  38400             User No  Data Length    7                            User Node ID  Parity    Odd    Even                      User Node  Stop Bits    zZ             RS 485    User Node  Comm Type                User       Contrast  Backlight                      Use the  i   and  E  keys to  select the desired baud rate  and press  ENTER      Use the  i   and  E  keys to  select the desired data length  and press  ENTER      Use the  1   and  E  keys to  select the desired parity setting  and press  ENTER      Use the  i   and  E  keys to  select desired number of stop  bits and press  ENTER      Use the       and  E  keys to  select the type of serial port  desired  Then  press  ESC  six  times     You are now back at the User  Menu     Press  ESC  as many times as necessary to return to normal RUN  mode  or proceed to the appropriate section of this chapter to continue    programming the XDP        Setup and Calibration    July 2003    Other Opt Menu  Options    Lock Menus    In addition to the User Menu described in the previous section  the  Opt Menu includes the following options     e Lock Menus  e Lock All    e Versions    Proceed to the appropriate section for a discussion of these op
26.  Setting the Display After entering the User Program  as described on page 3 4  the main  Backlight menu appears  To set the display backlight  proceed as follows   Cal Disp Use the       and       keys to                   Lock Menus                         select  Opt  and press   ENTER      Use the  i   and  E  keys to  select  User  and press   ENTER      Note     Lock All    and    Versions    menu options also appear on the    above list           User       Contrast    Backlight    Comm                         Use the  i   and  E  keys to  select  Backlight  and press   ENTER      Note  The    Fault Alarm  and    Analog Output  options are also  available in the above list box     The display backlight has three possible settings     e Off   the backlight is permanently turned off      On   the backlight is continuously on      Timed   the backlight comes on whenever a key is pressed and  remains on until a specified time interval has elapsed without any    keypad activity          User Backlight  Off  On                        Use the  i   and  E  keys to  select the desired option and  press  ENTER      Note  If    Off    or    On    was chosen above  the following two prompts  do not appear and you are returned to the User Menu        Setup and Calibration    4 5    July 2003    Setting the Display  Backlight  cont      Setting the System          User Backlight Timed                User Backlight  Off  On                      User       Contrast  Backlight    Comm 
27.  alleiniger Verantwortung  da   die Produkte    APX Advanced Paramagnetic Oxygen Transmitter  XDP Explosion Proof Display Package    folgende Normen erf  llen       EN 61326 1998  Class A  Annex A  Continuous Unmonitored Operation    EN 61010 1 1993   A2 1995  Overvoltage Category II  Pollution Degree 2    gem     den Europ  ischen Richtlinien  Niederspannungsrichtlinie Nr   73 23 EWG und EMV Richtlinie  Nr   89 336 EWG     Die oben aufgef  hrten Ger  te und zugeh  rige  mitgelieferte Handhabungssysteme tragen keine CE   Kennzeichnung gem     der Druckger  te Richtlinie  da sie in   bereinstimmung mit Artikel 3  Absatz 3   gute Ingenieurpraxis  der Druckger  te Richtlinie 97 23 EG f  r DN lt 25 geliefert werden     Shannon   June 1  2002 po Et i  Mr  James Gibson  GENERALDIREKTOR               LS  EN ISO 9002 T  V ESSEN    Shannon ISO 9001  U S        CERT DOC Rev G 5 28 02    GE Panametrics ATEX COMPLIANCE    GE Panametrics  221 Crescent Street  Suite 1  Waltham  MA 02453  U S A     as the manufacturer  declare under our sole responsibility that the product  XDP Explosion Proof Display Package    to which this document relates  in accordance with the provisions of ATEX Directive 94 9 EC Annex Il   meets the following specifications     C     112 GD EEx d IIC T6  1180 KEMA01 ATEX2128 T85   C   20  C to  65  C  Furthermore  the following additional requirements and specifications apply to the product     e Having been designed in accordance with EN 50014  EN 50018  and EN 5028
28.  no connection     Wiring The System Terminal block J2 contains connections for the system fault alarm  Alarm Relay   J2 relay  To wire this alarm  complete the following steps     Note  The system fault alarm on terminal block J2 is usually factory   configured as a    fail safe    alarm  That is  it is energized  during normal operation and de energized when tripped    1  Connect the System Fault alarm as follows    a  Connect pin  1 to the alarm device input     b  Connect pin  2 to the alarm device return     c  Pin  3 is not used for fail safe operation       Additional Wiring Connections B 1    July 2003    XDP Alarm Relays       In addition to the alarm relays discussed in Chapter 2  Installation   there are four supplemental alarm relays at terminal block J7       Gas Curve Relay 1    Gas Curve Relay 2    Process Cal Relay      Zero Span Relay    Note  Contact GE Panametrics for information on the use and  functions of these alarms     Table B 1  Terminal Block J7 Connections    Alarm Terminal    Gas Curve Relay 1          Gas Curve Relay 2          Process Cal Relay          Zero Span Relay                   To wire the supplemental alarms  refer to Figure 2 3 on page 2 10 and  Table B 1 above  Then  complete the following steps     Note  A fail safe alarm is wired to the normally closed  NC   contacts  while a non fail safe alarm is wired to the normally   open  NO  contacts    1  For non fail safe operation  connect pin  1 to the alarm input     2  Connect pin  2 t
29.  the unit will  comply with the Low Voltage Directive  73 23 EEC         CE Mark Compliance A 1    Appendix B    Additional Wiring Connections    Introduction    KEN dieu an a dans  Connecting the System Outputs                              ADP Alarm Relays  aa arar ae    July 2003       Introduction In addition to the features included specifically for typical XDP  applications  the general purpose controller used in the XDP has its  own built in features  These include the following       system analog output      system fault alarm    Also  there are four supplemental XDP alarm relays on terminal block  J7 that can be wired and utilized     Connectingthe System Although the XDP software is not intended to interface with the   Outputs general controller system functions  those functions are active and  may be accessed  To connect either or both of these outputs  refer to  Figure 2 3 on page 2 10 and proceed to one of the following sections     Wiring the System Terminal block J5 contains connections for the system analog output    Analog Output   J5 The connections on terminal block J5 are optional  in that they are not  required for proper operation of the analyzer  To wire the system  analog output  complete the following steps     IMPORTANT  The system analog output range does not  automatically switch to the currently active curve     1  Connect pin  1 to positive input on the output device   2  Connect pin  2 to the negative input on the output device     3  Pin  3 has
30.  unlocked   you may reenter the User Program by simply pressing the  ESC  key  only  To re lock the menus  see the instructions in Chapter 4  Setup  and Calibration     Note  If the unit is powered down  the User Program will be locked  when the unit is powered up again     To program a specific feature of the XDP  proceed to the appropriate  section of Chapter 4  Setup and Calibration     As an aid in diagnosing operational problems with the XDP  the built   in software has been designed to report several error conditions  directly on the display screen  These error messages  along with the  causes and suggested remedies are listed in Table 3 1 on page 3 6     IMPORTANT  Table 3 1 on page 3 6 lists the error messages in  order of priority  If multiple errors occur  simultaneously  only the highest priority error is  displayed  errors with the same priority level cannot  occur simultaneously         Operation    3 5    July 2003    Error  Message    ADC In Out  of Range     Table 3 1  Screen Error Messages    The analog to   digital converter  has returned an  invalid value     Suggested Remedy    A low level hardware failure is likely   Contact the factory     XDP Analog  Output  Response    Holds last  value       mA Input  Under  Range     The analog signal  from the  transmitter is less  than 2 4 mA       Could be caused by a loose or incorrect    electrical connection  Check all  electrical connections       The sensor calibration may have  drifted too far  Contact the f
31. 1  the product meets the  fault tolerance requirements of electrical apparatus for category    d        e The product is an electrical apparatus and must be installed in the hazardous area in accordance with  the requirements of the EC Type Examination Certificate  The installation must be carried out in  accordance with all appropriate international  national and local standard codes and practices and site  regulations for flameproof apparatus and in accordance with the instructions contained in the manual   Access to the circuitry must not be made during operation     Only trained  competent personnel may install  operate and maintain the equipment     The product has been designed so that the protection afforded will not be reduced due to the effects of  corrosion of materials  electrical conductivity  impact strength  aging resistance or the effects of  temperature variations     The product cannot be repaired by the user  it must be replaced by an equivalent certified product   Repairs should only be carried out by the manufacturer or by an approved repairer     The product must not be subjected to mechanical or thermal stresses in excess of those permitted in  the certification documentation and the instruction manual     The product contains no exposed parts which produce surface temperature infrared  electromagnetic  ionizing  or non electrical dangers     Cf  6     CERT ATEX C 7 16 03       GE Panametrics    WORLDWIDE    OFFICES          MAIN OFFICES  GE PANAMETRICS 
32. 4    Sg d    3 8          C NICNC NOCNC NOC             RIN  24                                           00000000000                      A                                                                                                                   XDP ALARMS   J8       Pin No     Description       1    ormal  NC              ormal  C        ormal  NO        Fault  NC              Fault  C   Fault  NO        wire colors for Panametrics       Low  NC      704 668 serial cable        SYSTEM ANALOG OUTPUT   J5    SYSTEM FAULT ALARM   J2    Low  C     RS232 485 COMM   J4       Pin No     Description    Pin No     Description    OI BR cnm    Low  NO     Pin No    Description Color                    No connections  required here      See Appendix B     1       2             3       No connections  required here    See Appendix B        Low Low  NC     1 Transmit  TX  Red       Low Low  C     2 Receive  RX  White                Low Low  NO     Green                   3 Return  RTN           OTHER CONNECTIONS   J7       Pin No    Description       1                No connections       required here    See Appendix B              OI III  BB        ND       a  e                    TRANSMITTER CONNECTIONS   J6    XDP ANALOG OUTPUT   J8       Pin No     Description    Pin No     Description       1    Analog Input  IN     1    Analog Output  AOUT        2    Power Output   24     2    Return  RTN              3       Return  RTN           3       No Connection             ing 4702
33. DP Normal  J3   fail safe type  XDP Low  J3   fail safe or non fail safe mode   trips if H2  lt 85   setpoint is programmable   XDP Low Low  J3   fail safe or non fail safe mode   trips if H2  lt 80   setpoint is programmable     Analog Input  terminal block J6    4 20 mA  28 V    Analog Output  terminal block J8    4 20 mA  28 V    Communications  terminal block J4    Standard RS232 RS485 Serial Port    Display   4 line  backlit liquid crystal display  LCD     Keypad   6 key  infrared    Measurement Ranges  for sample application        70 100  hydrogen in air  e 0 100  hydrogen in carbon dioxide      0 100  air in carbon dioxide    The design  manufacture  and certification of the XDP have been  conducted in a manner that allows its operation in Zone 1    Gas Group IIC  flameproof protection method in an ambient  operating temperature of  10 to 65 C  14 to 149 F      This unit complies with EMC Directive 89 336 EEC and the  73 23 EEC Low Voltage Directive  Installation Category II  Polution  Degree 2   It is in compliance with ATEX directive 94 9 EC Annex  IL  per the document at the back of this manual     For complete transmitter specifications  refer to the documentation  provided with the transmitter        6 2    Specifications    Appendix A    CE Mark Compliance    Introduction AA A TUE e OS  EMG Complianee     mk AA ae UC CR Ge NG    LVD Compliance  see ea aan    July 2003    Introduction For CE Mark compliance  the XDP must meet both the EMC and  LVD directives     
34. GE Panametrics is warranted to be  free from defects in material and workmanship  Liability under this  warranty is limited to restoring the instrument to normal operation or  replacing the instrument  at the sole discretion of GE Panametrics  Fuses  and batteries are specifically excluded from any liability  This warranty  is effective from the date of delivery to the original purchaser  If GE  Panametrics determines that the equipment was defective  the warranty  period is       one year for general electronic failures of the instrument      one year for mechanical failures of the sensor    If GE Panametrics determines that the equipment was damaged by  misuse  improper installation  the use of unauthorized replacement parts   or operating conditions outside the guidelines specified by GE  Panametrics  the repairs are not covered under this warranty     The warranties set forth herein are exclusive and are in lieu of  all other warranties whether statutory  express or implied   including warranties or merchantability and fitness for a  particular purpose  and warranties arising from course of  dealing or usage or trade      If a GE Panametrics instrument malfunctions within the warranty period   the following procedure must be completed     1  Notify GE Panametrics  giving full details of the problem  and  provide the model number and serial number of the instrument  If the  nature of the problem indicates the need for factory service  GE  Panametrics will issue a RETURN AU
35. IMPORTANT  CE Mark compliance is required only for units used  in EEC countries     EMC Compliance For EMC compliance  the electrical connections must be shielded and  grounded as shown in Table A 1 below  After all the necessary  electrical connections have been made  seal any unused cable entry  holes with standard conduit plugs or equivalent     Note  If the instructions in this appendix are followed  the unit will  comply with the EMC Directive 89 336 EEC     Table A 1  Wiring Modifications for EMC Compliance    Wiring Modification    1  When connecting the power  select the cable  entry closest to the chassis ground    2  Use shielded cable  to connect the power to the  XDP enclosure  Connect the shield to the  nearest chassis ground terminal      Connect the power line ground wire to the    nearest chassis ground terminal        Input Output   1  Use shielded cable    to interconnect the Model  XDP enclosure with any external I O devices     Connect the shields to the nearest chassis  ground terminal      Wires enclosed in a properly grounded metal conduit do not  require additional shielding     LVD Compliance For compliance with the European Union s Low Voltage Directive   73 23 EEC   the analyzer requires an external power disconnect  device such as a switch or circuit breaker  The disconnect device must  be marked as such  clearly visible  directly accessible  and located  within 1 8 m  6 ft  of the XDP        Note  If the instructions in this appendix are followed 
36. INTERNATIONAL OFFICES   USA Australia Japan   GE Panametrics P O  Box 234 2F  Sumitomo Bldg    221 Crescent St   Suite 1 Gymea N S W  2227 5 41 10  Koishikawa  Bunkyo Ku  Waltham  MA 02453 3497 Australia Tokyo 112 0002    USA   Telephone  781 899 2719   Toll Free  800 833 9438   Fax  781 894 8582   E mail  panametrics  ps ge com  Web  www  gepower com panametrics  ISO 9001 Certified    Ireland   GE Panametrics   Shannon Industrial Estate  Shannon  Co  Clare   Ireland   Telephone 353 61 470200  Fax 353 61 471359   E mail info  panametrics ie  ISO 9002 Certified    July 2003       Telephone 61  02  9525 4055  Fax 61  02  9526 2776  E mail panametrics   panametrics com au    Austria   Waldgasse 39   A 1100 Wien   Austria   Telephone  43 1 602 25 34  Fax  43 1 602 25 34 11  E mail panametrics netway at    Benelux   Postbus 111   3870 CC Hoevelaken   The Netherlands   Telephone  31  0  33 253 64 44  Fax  31  0  33 253 72 69  E mail info   panametrics nl    France   BP 106   11 Rue du Renard   92253 La Garenne Colombes Cedex  France   Telephone 33  0  1 47 82 42 81  Fax 33  0  1 47 86 74 90   E mail panametrics   panametrics fr    Germany   Mess und Pruftechnik  Robert Bosch Stra  e 20a   65719 Hofheim   Germany   Telephone  49 6122 8090   Fax  49 6122 8147   E mail panametrics   t online de    Italy   Via Feltre  19 A   20132 Milano   Italy   Telephone 02 2642131   Fax 02 26414454   E mail info   panametrics it    Japan   Telephone 81  03  5802 8701  Fax 81  03  5802 8706  E mail 
37. July 2003    Process Control Instruments    XDP Explosion Proof  Display Package       User   s Manual 910 204B    GE Panametrics    July 2003    Table of Contents    Chapter 1  General Information    Introduction  4  ote ne ein E epe er rU NA 1 1  Basic AA AT 1 1  System Overview    ns GNG nn en PNG nen pee nen ARA 1 2  A Typical Application    4 4 4  ies ana ues ee a a es De eee NG 1 3   Alarm Settings  cused  nadie dee ia A wre 1 3    Chapter 2  Installation    Introduction Jub i aros a un had an 2 1  Choosing an Installation Site                                                      2 2  Mounting the XDP System    au name name 2 3  Mounting the XDP Enclosure                                                  2 3  Mounting the Remote Transmitter Sample System                                  2 3  Wiring the XDDB AA ATO Ae 2 4  Wiring The Remote Transmitter   J6                                             2 5  Wiring The XDP Alarm Relays   J3                                             2 5  Wiring the XDP Analog Output   J8                                             2 6  Wiring the Optional Serial Port   J4                                             2 6  Wiring The Line Power as te ner en ee Ak aed terre 2 7    Chapter 3  Operation    Introduction ec ee ee PURI DE EE TUE tos 3 1  Preventing Common Problems                                                     3 1  Cleaning the Enclosur           bes ole ne er m ERR a bre eid S 3 1  Powering Up the System    ra RE na bs 3 2   T
38. NTER  when done   User  Aout You are now back at the sys   Aout Type tem analog output setup menu     Zero Setpoint    Span Setpoint                      Either press  ESC  until you return the User Menu or proceed to the  appropriate section to continue setting up the system analog output        4 10 Setup and Calibration    July 2003          Setting the Span Enter the system analog output menu  as described on page 4 8  and  Setpoint proceed as follows   User  Aout Use the  1   and  E  keys to            select  Span Setpoint   and  press  ENTER         Zero Setpoint  Span Setpoint                Note  The Aout Type  Zero Trim  and Span Trim options are also  available in the above list box     At the next prompt  enter the desired span setpoint  the percentage of  target gas that corresponds to the upper end of the specified system  analog output range minus the zero setpoint                                          User Aout Span Set Use  i   and       to position the  cursor under the desired char   acter  Then  use  i   and  E  to   XXXX XXXXX    increment the value  Press   A  ENTER  when done    User   Aout You are now back at the sys    Zero Setpoint tem analog output setup menu    Span Setpoint   Test    Either press  ESC  until you return the User Menu or proceed to the  appropriate section to continue setting up the system analog output        Setup and Calibration 4 11    July 2003    Testing the System  Analog Output    Enter the system analog output menu  as d
39. R  key     Note  If the calibration    Failed     repeat the calibration procedure   If another failure occurs  contact the factory for assistance        Zero Drift                      4 22    Setup and Calibration    July 2003       Performing a Span  Calibration    Stop the flow of the zero calibration gas  and begin the flow of the  span gas to the inlet port of the transmitter  at about 250 cc min  0 5    SCFH  and atmospheric pressure           Gas Curve  Manual         Zero                         Use the  i   and  E  keys to  select  Span  and press the   ENTER  key     At the following prompt  enter the percentage of target gas  H2 in this  example  in the span gas mixture  usually 100 0096            Gas Curve  Ma  Span       Enter Span Gas Pent    XXX XX                      Use  i   and       to position the  cursor under the desired char   acter  Then  use  i   and  E  to  increment the value  Press   ENTER  when done     A  YES  response at the next prompt sends the last live value to the  analog output  while a  NO  response sends the calibration gas value    to the analog output           Gas Curve  Man  Cal  Hold 4 20 mA Output    During Calibration     ms                   Gas Curve  Man  Cal  Introduce Cal Gas  x x H2 CO2                    lt Back   Next gt    Cancel       Use  i   and       highlight the  desired answer  YES or NO   to the question shown  Then   press the  ENTER  key     Make sure the span gas is  flowing  and use      and       to  highl
40. THORIZATION NUMBER   RAN   and shipping instructions for the return of the instrument to a  service center will be provided     2  If GE Panametrics instructs you to send your instrument to a service  center  it must be shipped prepaid to the authorized repair station  indicated in the shipping instructions     3  Upon receipt  GE Panametrics will evaluate the instrument to  determine the cause of the malfunction     Then  one of the following courses of action will then be taken     e  f the damage is covered under the terms of the warranty  the  instrument will be repaired at no cost to the owner and returned        f GE Panametrics determines that the damage is not covered under  the terms of the warranty  or if the warranty has expired  an estimate  for the cost of the repairs at standard rates will be provided  Upon  receipt of the owner s approval to proceed  the instrument will be  repaired and returned        vii    Chapter 1    General Information    Introductorio ee 1 1  Basic Features                                           1 1  System OVeIrVIeW   ad be a a n 1 2    A Typical Application                                     1 3    July 2003    Introduction    Basic Features    The XDP Explosion proof Display Package has been designed as one  component of a system that monitors the composition of a binary gas   The second component is a remotely mounted sensor transmitter   such as the GE Panametrics Model TMO2 TC or similar device     Note  Refer to the documentat
41. Use the  i   and       keys to select    Add    or    Remove    and press  the  ENTER  key to change the status of the selected output  An  output is selected if an asterisk     appears to its left     3  When all of the desired outputs have been selected  use the      and        keys to select    OK    and press the  ENTER  key           From 1 Cfg H2 AIR    pent                            1 H2 AIR    18 73 mA                      Press  ESC  three times to  return to the main menu  Then   press  ESC  again to return to  run mode     The selected display is shown    Notice that the padlock no  longer appears  as the menus  are now unlocked         Advanced Programming    5 3    July 2003    Configuring the Display   cont      Accessing the Setup  Menu    The newly configured display shows the slot number and the active  curve on the top line  and the current live value along with the unit  type are displayed in the center of the screen     Note  With the menus unlocked  you may reenter the User Program  by simply pressing the  ESC  key only  To re lock the menus   see the instructions in Chapter 4  Setup and Calibration     Upon reentering the User Program  the    Disp    main menu option  will now be available  and the available display windows may be  switched as described in Chapter 4  Setup and Calibration  Also  the  display may be reconfigured at any time by repeating the above  procedure     To access the setup menu from the main menu of the User Program   proceed as 
42. actory calibrated  for the range s  and gas mixture s  specified at the time of purchase   However  to ensure the accuracy of the readings  the system should  be recalibrated every 2 6 months  The optimum calibration interval  depends on the specific details of the application and is best  determined by periodic calibration checks after the initial installation     Caution   The calibration procedures described below should only  be performed by trained service personnel     In order to complete the calibration procedure  the following items  are required       zero gas  usually either 100  air or 100  carbon dioxide     span gas  usually 100  hydrogen       a suitable sample system  see Figure 2 1 on page 2 4     IMPORTANT  Each gas cylinder must be certified as to the exact  composition of the gas  The accuracy of the  calibration is only as good as the accuracy of the  calibration gases used     Proceed with the instructions in this section to calibrate the system     A temporary connection between the transmitter and the gas cylinder  may be made for calibration purposes  However  if a permanent  connection is preferred  it should be as short as possible  with an  isolation valve right at the gas inlet port on the transmitter  see the  transmitter and or sample manual for details      IWARNING   When working in a hazardous area  carefully follow  all appropriate safety guidelines during the  calibration process     To set up for calibration complete the following steps   
43. actory for  sensor recalibration or replacement       May be caused by a failure in the ADC    circuitry  Contact the factory for  assistance      No transmitter signal  Check the  transmitter wiring     Holds last  value       mA Input  Over Range     Gas Pcnt  Under  Range     Gas Pent  Over Range     The analog signal  from the transmit   ter is more than  21 6 mA     The target gas per   centage is below  the preset range     The target gas per   centage is above  the preset range     The sensor calibration has drifted  Contact  the factory for assistance with field recali     bration     The sensor calibration has drifted  Per   form a field calibration  see    Calibrating  the System    on page 4 20     The sensor calibration has drifted  Per   form a field calibration  see    Calibrating  the System    on page 4 20      Holds last  value    Outputs zero  value    Outputs span  value       Low H2 pet     Occurs in H2 AIR  measurement  mode  when the  H2 reading is  below the Low  Alarm setpoint   usually 85       This indicates air contamination  and the  Low Alarm will trip  Remove the source  of the air contamination     no response          Low Low  H2 pet        Occurs in H2 AIR  measurement  mode  when the  H2 reading is  below the Low   Low Alarm set   point  usually  80          This indicates air contamination  and the  Low Low Alarm will trip  Remove the  source of the air contamination           no response       3 6    Operation    Chapter 4    Setup and Cali
44. alog output    setting up the serial port      calibrating the system    Note  While in the User Program  press  ESC  at any time to abort  the current operation and move back up one menu level     Access the User Program as described on page 3 4  and refer to the  menu maps in Figures C 1 and C 2 on pages C 1 and C 2  At the  following screen  enter the operator level password   The default  password is 2719            Password Use  i   and       to position the  cursor under the desired char   acter  Then  use  i   and  E  to  increment the value  Press   e  ENTER  when done     Enter Password       XXXX                      You will now be at the main menu screen  Proceed directly to the  appropriate section to perform the desired programming task        Setup and Calibration 4 1    July 2003       Selecting the Active  Curve    At times it will be desirable to view a curve other than H2 Air  For  example  during the first stage of a shutdown  it will be necessary to  view the H2 CO2 curve  To select the XDP active curve  access the  User Program as described on page 3 4  and proceed as follows              Opt Disp                      Opt Disp                Range Cal    Slct Cal Menu Item                Cal  Select Active Curve                            Select Active Curve       H2 CO2  AIR CO2                         Cal  Select Active Curve                   Use the  i   and       keys to  select  Cal  and press   ENTER      Press  ENTER      If necessary  use the  1
45. arm Use the       and     keys to  select  Trip Point  and press   Test  ENTER     Fail Safe   At the next prompt  enter the percentage of hydrogen in air that  should trip the low alarm  typically 85 00      Alarms  Trip Point Use  i   and       to position the  cursor under the desired char   acter  Then  use       and  E  to   XXX XX B  increment the value  Press   s  ENTER  when done    Alarms Low Alarm Use the       and  E  keys to   Trip Point select  Test  and press   en   Fail Safe   5 8 Advanced Programming       July 2003  The Low Alarm  cont            Alarms Low A  Test Use the       and  E  keys to    select  Trip  and press  TE  ENTER                  After verifying that the alarm has actually tripped  continue to the  next prompt           Alarms Low A  Test If necessary  use the       and  Trip  E  keys to select  Reset  and    Bee i                      The program alternates between the two previous prompts  each time   ENTER  is pressed  To leave the test loop  press  ESC      Note  Upon leaving the test loop  the alarm is automatically reset to  its normal state              Alarms Low Alarm Use the  1   and  E  keys to   Trip Point select  Fail Safe  and press   ENTER     Test                   Alarms  Fail Safe  Non Fail Safe    Use the  1   and  E  keys to  select desired alarm type and  press  ENTER         Fail Safe                   Alarms Low Alarm       Press the  ESC  key to leave    Trip Point the  Low Alarm  submenu     Test                    
46. ation    July 2003    Mounting the XDP  System    Mounting the XDP  Enclosure    Mounting the Remote  Transmitter Sample  System    This section explains how to mount the XDP system components and  the transmitter sample system at the installation site     IMPORTANT  The XDP requires an external power disconnect  device such as a switch or circuit breaker  The  disconnect device must be marked as such  clearly  visible  directly accessible  and located within 1 8 m   6 ft  of the unit     IMPORTANT  This symbol indicates Caution   risk of electric    shock   AN    Refer to Figure 2 1 on page 2 9  and complete the following steps to  mount the XDP enclosure     1  Select a flat surface on a vertical wall or instrument panel with  sufficient space to hold the XDP enclosure     2  Using the mounting dimensions shown in Figure 2 1 on page 2 9   locate and prepare four  4  mounting holes or two  2  horizontal  mounting rails on the vertical surface     3  Secure the XDP enclosure to the mounting surface with four  4   sets of mounting hardware  Be sure that the enclosure is in a  vertical position  with the display window at the top     Mount the remote transmitter in accordance with the instructions in  its accompanying documentation     Note  Most GE Panametrics transmitters may be located up to  2 800 ft  850 m  from the XDP enclosure     To provide reliable and accurate results  the transmitter is often  installed in a suitable sample system  The sample system is designed  to 
47. bration    Introduction    Gaan RAE BER ANAN GT eee NAE PERA RR 4 1  Selecting the Active Curve                                 4 2  Switching Display Windows                                4 3  Adjusting the Display Contrast                             4 4  Setting the Display Backlight                               4 5  Setting the System Fault Alarm                             4 6  Setting Up the System Analog Output                       4 8  Setting Up the Serial Port                                 4 15  Other Opt Menu Options                                  4 17    Calibrating the System                                   4 20    July 2003  Aa      Introduction Although the XDP is set up at the factory with default values that are  suitable for the intended application  the User Program provides a  means for customizing many of the meter parameters  Proper setup of  the system is very important to ensure accurate data readings  The  following procedures are described in this chapter     IWARNING    To ensure safe operation of the XDP  it must be  installed and operated as described in this manual   In addition  be sure to follow all applicable local  safety codes and regulations for installing electrical  equipment  In addition  all procedures should be  performed by trained service personnel       selecting the active curve     switching display windows     adjusting the display contrast    setting the display backlight    setting the fault alarm     setting up the an
48. correct value  When done   press  ENTER      You are now back at the sys   tem analog output setup menu     Either press  ESC  until you return the User Menu or proceed to the  appropriate section to continue setting up the system analog output        Setup and Calibration    July 2003    Setting Up the Serial  Port    After entering the User Program  as described on page 3 4  the main  menu appears  To set up the serial port  proceed as follows     Note  For the XDP the serial port is normally used only for  downloading software upgrades to the meter        Cal Disp Use the       and       keys to  select  Opt  and press     ENTER                     Cal MOS Disp Use the  1   and  E  keys to  User select  User  and press     ENTER    m              Note     Lock All    and    Versions    menu options also appear on the                above list   User Use the  1   and  E  keys to  Contrast select  Comm  and press  ENTER    Backlight                     Note  The    Fault Alarm    and    Analog Output    options are also  available in the above list box     IMPORTANT  Each of the following prompts shows the usual  default value for that parameter     At the following prompt  choose one of the following numbers as the  Node ID for the meter  16  32  48  64  80  96  112  128  144  160  176   192  208  224  or 240              User Node ID Use the  1   and  E  keys to  Node ID scroll the list of available node  pa ID numbers  When the desired  number is highlighted  press  16 v 
49. d be checked approximately every  2 6 months  depending upon the specific application  see     Calibrating the System    on page 4 20        Never use thread sealant on any connection in the sample gas flow  path  Thread sealant emits combustible vapors that can cause  reading errors       The infrared keypad may not function properly if direct sunlight  strikes the face of the XDP or if the XDP is installed in an  environment that has a high infrared content  To minimize such  factors  provide suitable shading during installation     If the glass window or the case of the XDP becomes soiled  use a soft  cloth dampened with water for cleaning  Never use solvents or  detergents to clean the XDP        Operation    3 1    July 2003    Powering Up the  System    The Digital Display  Window    Check the wiring connections and close both the XDP and transmitter  enclosures before applying power  Energize the external disconnect  device to power up the XDP  Allow the transmitter to warm up for at  least thirty minutes  before taking measurements     IMPORTANT  The XDP requires an external power disconnect  device such as a switch or circuit breaker  The  disconnect device must be marked as such  clearly  visible  directly accessible  and located within 1 8 m   6 ft  of the unit     The digital display window at the top of the XDP enclosure includes  the components shown in Figure 3 1 below     Display Window    ENTER Key        GE Panametrics    o    LEFT Arrow N        RIGHT Arrow
50. deliver a clean  representative gas sample to the inlet of the  transmitter at the proper temperature  pressure and flow rate     Note  A suitable sample system may be ordered directly from GE  Panametrics  The sample system would be supplied fully  assembled on a flat plate with four  4  mounting holes  See the  sample system and or transmitter manuals for more details        Installation    July 2003    Wiring the XDP To completely wire the XDP  connect the following items     analog output from transmitter  J6   e  24 VDC input to transmitter  J6   e XDP alarm relays  J3     XDP analog output  J8     serial port  J4    optional    line power  J1     Note  See Appendix B  Additional Wiring Connections  for a  discussion of connections to terminal blocks J2  J5  and J7     Refer to the wiring diagram shown in Figure 2 2 on page 2 10  while  completing the instructions presented in this section      Attention European Customers   To meet CE Mark requirements  install all cables as  described in Appendix A  CE Mark Compliance     IMPORTANT  The XDP requires an external power disconnect  device such as a switch or circuit breaker  The  disconnect device must be marked as such  clearly    visible  directly accessible  and located within 1 8 m   6 ft  of the unit     IWARNING    To ensure safe operation of the XDP  the unit must  be installed and operated as described in this  manual  Also  be sure to follow all applicable local  safety codes and regulations for installing electr
51. escribed on page 4 8  and  proceed as follows              User  Aout Use the  1   and  E  keys to  select  Test   and press     ENTER              Span Setpoint       Zero Trim          Note  The Aout Type  Zero Setpoint  and Span Trim options are  also available in the above list box     At the next prompt  enter the desired    Test Percent  value  which is  the system analog output value  expressed as a percentage of the  system analog output span  used by the meter as a TEST signal           User Aout Aout Test Use  I   and  E  to increment  Test Percent the test percent to the desired    Fi value           xx V       Enter as many values as desired at the above prompt to verify that the  analog output device is accurately recording the specified test values   When done  press  ENTER               User  Aout You are now back at the sys   tem analog output setup menu        Span Setpoint    Zero Trim                      Either press  ESC  until you return the User Menu or proceed to the  appropriate section to continue setting up the system analog output        Setup and Calibration    July 2003    Setting the Zero Trim    Enter the system analog output menu  as described on page 4 8  and  proceed as follows     IMPORTANT  Before proceeding with this section  make sure that  the output percentage being trimmed has been  entered in the Test menu on page 4 12                       User  Aout Use the       and  E  keys to  Test select  Zero Trim   and press  FM   Span Trim  
52. follows            era Cal Opt Disp Use the  i   and       keys to    select  Cal                        Opt Disp Press  ENTER               Use the  i   and       keys to  select  Setup  and press   ENTER      Range Cal          Slct Cal Menu Item                          Range Cal ea  At this programming level   4 20mA Out only the two options shown  are available   Alarms                Proceed to the appropriate section to configure the desired option        5 4    Advanced Programming    July 2003       Configuring the 4 20 mA To configure the XDP 4 20 mA analog output  access the setup menu  XDP Analog Output as described on page 5 4 and complete the following steps           4 20mA Out                The Range Submenu          4 20mA Out  4 20 Out Range  Test                Trim                   4 2  4 20 Out Range       H2 CO2  AIR CO2                Use the  i   and  E  keys to  select  4 20mA Out  and press   ENTER      Use the  i   and  E  keys to  select  4 20 Out Range  and  press  ENTER      Use the  i   and  E  keys to  select the desired output range  and press  ENTER      Note  The H2 AIR range is used in this manual as an example  To  program the other ranges  use similar procedures     At the next prompt  enter the percentage of hydrogen in air that  should generate a mA output of 4 00  typically 70 00             4 20mA Out 4   Zero    XX XX                      Use  i   and       to position the  cursor under the desired char   acter  Then  use  i   and 
53. g Cal   Opt   Disp                         Lock Menus                      Use the  i   and       keys to  select  Opt  and press   ENTER      Use the  i   and  E  keys to  select  Setup   Then  press   ENTER   and proceed to the  appropriate section     Note     Lock All    and    Versions    menu options also appear on the    above list                                                                    XDP Software T  Setup Use the  i   and  E  keys to  Load Slot select  Load Slot  and press  Load Main EIER  Setup Load Slot Press  ENTER    ID  When the download is com   Size  plete  the meter automatically  Received Block  reboots into run mode   Write to Slot 1   5 18 Advanced Programming    July 2003       System Software    Exiting the User  Program          Setup  Load Slot                   BSeries Loader v1 03  Boot is Flash    Load by Software   Load via Comm Y N                           ID   Size     Received Block   Write to Flash             Use the  i   and  E  keys to  select  Load Main  and press   ENTER      To respond to the question   press  ESC  for    Y    or press   ENTER  for  N      When the download is com   plete  the meter automatically  reboots into run mode     When you have finished programming all of the required functions   press  ESC  as many times as necessary to return to normal run mode   The number of key presses required depends on which feature was  programmed last  The XDP may then be placed into service        Advanced Programming    5
54. he Digital Display Window                                                  3 2   Initial Screen  Displays    edi das in sep teh a Haren Re et E 3 3   Starting the Sample Gas                                                       3 4  Accessing the User Program             Sa DAG esse san ae nenne neh 3 4  Exiting  the User Program    2 a a RR VER DEAE GNG 3 5  Error Messages    nn  nen asian DAL Helle BE er 3 5       iii    July 2003  es    Table of Contents  cont      Chapter 4  Setup and Calibration    AA Eh eS ttt ra 4 1  Selecting the Active Curve    0 0 0 0    ccc ccc eee hme 4 2  Switching Display Windows                                                      4 3  Adjusting the Display Contrast                                                    4 4  Setting the Display Backlight                                                      4 5  Setting the System Fault Alarm                                                2  4 6  Setting Up the System Analog Output                                              4 8  Choosing the System Analog Output Type                                       4 9  Setting the Zero Setpoint  ici As 4 10  Setting the Span Setpoint                                                    4 11  Testing the System Analog Output                                             4 12  Setting the  Zero Tim  Xa nd nuls ede Mar hehe qu 4 13  setting the Span Trim ss sas a he Der en a 4 14  setting Up the  Serial Port    bl rn una 4 15  Other Opt Menu Options               
55. he XDP must be configured for the  desired display parameter  To set the active curve that is output to the  digital display  complete the following steps                          Password Use  i   and       to position the  Enter Password cursor under the desired char   Er acter  Then  use  i   and  E  to  increment the value  Press  A  ENTER  when done   Cal Opt Disp If necessary  use the  1   and          keys to select  Cfg  and  press  ENTER    Notice Disp is  unavailable at this time                      GES Cal Opt Disp Press  ENTER                        5 2    Advanced Programming    July 2003       Configuring the Display   cont            From 1 Cfg             From 1 Cfg H2 AIR    pent                         From 1 Cfg H2  pent   0 Display   0 Fault Alarm  Add Remove MOS                Cancel                      The currently active range is  listed  Press  ENTER  to select  this range   H2 AIR is used  here as an example      A list of available units is   shown  Use the  i   and  E   keys to select the desired units  and press  ENTER      A list of the available output  devices is shown  Follow the  instructions below to select the  desired outputs     Note     0  Analog Output  is also available in the above list box     To select the desired output s   complete the following steps     1  Use the       and    l keys to highlight a specific system output  For  this example  make sure that  0 Display  is selected as one of the    outputs to be displayed     2  
56. he time the XDP was ordered  Thus  the equipment should be suited  to the application and the planned installation site  However  before  installing the unit  read these guidelines to verify that the best  installation site has been chosen for optimum instrument accuracy  and reliability     Make sure that the XDP and transmitter enclosures are suitable for  the environmental conditions at the installation site     Install the system in a location with little or no vibration     Mount the transmitter in accordance with the instructions provided  in its accompanying documentation     Make sure the ambient temperature at the XDP installation site is   10 to 65  C  14 to 149  F      Use a suitable electrical conduit to interconnect the remote  transmitter and the XDP enclosure     Use a suitable sample system between the gas source and the  sample inlet of the transmitter     Observe all normal safety precautions  Specifically  do not exceed  the maximum pressure and temperature ratings of the transmitter  or the sample system     Install the XDP system in a location that provides ready access for  programming  testing  and servicing the unit     Protect all cables from excessive physical strain  bending  pulling     twisting  etc    Do not subject the cables to temperatures above  65  C  149  F  or below  50  C   58  F      Be sure that the line voltage used at the planned installation site  corresponds to the factory preset line voltage rating for the XDP        2 2    Install
57. he transmitter for a discussion of the specific technology employed  by the sensor mechanism   To be suitable for use in an XDP system   the transmitter must meet the following requirements       powered by the  24 VDC output from the XDP    output a 4 20 mA signal to the XDP analog input    meet all performance specifications for the system      suitable for use in the specified environment    The XDP accepts the 4 20 mA analog input from the transmitter and  processes the information  Then  the results are displayed on the  built in digital display window  In addition  the data may be output as  a linear analog signal to drive a data recorder and or controller device        General Information    July 2003    A Typical Application    Alarm Settings    Heavy duty electricity generators are typically cooled with hydrogen  gas to maximize generator efficiency  The hydrogen gas level is  continuously monitored  at both ends of the generator  An XDP  system with a remote transmitter is ideally suited for this service     During normal operation  the hydrogen gas must be monitored for the  presence of air  which would create an explosion hazard  Then  in  preparation for a maintenance shutdown  the hydrogen gas is purged  with carbon dioxide gas before the chamber is opened up and exposed  to air  At the conclusion of the maintenance work  the chamber is  resealed and the air is purged with carbon dioxide  Finally  the carbon  dioxide is purged with hydrogen prior to resuming ope
58. ibed in this section  access the  User Program as described on page 3 4 and enter the correct value   Consult Factory  at the password prompt  Then  refer to Figure C 4  on page C 4 and go to one of the following sections     entering the sensor serial number    configuring the 4 20 mA input    entering factory calibration data    loading software    To enter the XDP system transmitter serial number  access the setup  menu as described on page 5 4 and complete the following steps              Range Cal   Setup   Use the  1   and  E  keys to    Alarms select  Serial Number  and    ENTER    Serial Number pes      4 20mA Input                   Note     4 20mA Out    and    Factory Calibration  menu options also  appear on the above list     At the next prompt  enter serial number of the transmitter  as provided  by GE Panametrics              Sensor Serial Num Use  i   and       to position the  cursor under the desired char   acter  Then  use  i   and  E  to  increment the value  Press   a  ENTER  when done     Sensor Serial Num       XXXXX                            Range Cal   Setup You are now back at the setup    Alarms menu     Serial Number    4 20mA Input                   To continue programming the setup menu  proceed to the appropriate  section for instructions  Otherwise  press  ESC  four times to return to  live measurement mode        Advanced Programming    July 2003    Configuring the 4 20 mA To configure the 4 20 mA analog input  access the setup menu as  descr
59. ibed on page 5 4 and complete the following steps     Analog Input    The Test Submenu    The Trim Submenu          Range Cal        Serial Number  4 20mA Input    Factory Calibration                 Use the  i   and  E  keys to  select  4 20mA Input  and  press  ENTER      To test the analog input  program a test value into the XDP analog  output  see page 5 6  and connect the analog output signal to the  analog input terminals  Then  proceed as follows           4 20mA Input    Trim                      Use the  i   and  E  keys to  select  Test  and press   ENTER      At the next prompt  the mA value of the test signal is shown           4 20mA Input Test  mA Input Test       X XXX             After noting the mA value of  the analog input signal  press   ENTER  to continue     To trim the analog input  make sure the XDP analog output has been  calibrated  see page 5 5  and connect the XDP analog output signal to    the analog input terminals           4 20mA Input  Test                      Use the  i   and  E  keys to  select  Trim  and press   ENTER      The XDP analog output is now driven to exactly 4 mA           4 20mA In  4mA Trim                      Press the  I   or    l key until  the meter reads exactly 4 mA   When done  press  ENTER         Advanced Programming    5 15    July 2003    Configuring the 4 20 mA  Analog Input  cont      Entering Factory  Calibration Data          4 20mA I  20mA Trim    xx V                         4 20mA Input  Test               
60. ibration    4 25    Chapter 5    Advanced Programming    Introduction       ss sus 20a AAA ann 5 1  Programming Levels                                      5 1  Setup Level Programming                                  5 2  Factory Level Programming                               5 14    Exiting the User Program                                 5 19    July 2003    Introduction    Programming Levels    The XDP is factory configured to provide optimum performance in  the specified application  Because many of the advanced settings  programmed into the meter during this procedure are essential for  reliable and accurate operation  the operator does not see these menu  items during routine programming of the XDP  However  in order to  address operational problems or to fine tune the performance of the  system for a specific application  these additional programming  features are made available to authorized service personnel     Caution    Always consult GE Panametrics before performing any  of these advanced programming functions  Improper  setup may impair the performance of the XDP system   and or prevent it from meeting its design specifications     Advanced programming of the XDP is divided into two categories   which require passwords that are different from each other and from  the operator level password discussed on page 3 4  All three  programming levels and the required passwords are listed below       Operator  Basic  Level  password   2719    Setup  General  Level  pas
61. ical  equipment  In addition  all procedures should be  performed by trained service personnel     Remove the two side covers on the XDP enclosure  This is  accomplished by loosening the set screw in the cover and using a long  screwdriver or rod in the slots provided to unscrew the cover  Proceed  to the appropriate sub section to wire each of the terminal blocks     Note  A long  pin like  plastic wiring tool is stored inside the XDP  enclosure  Insert the tool into the connector pin being wired to  open the hole  and then insert the wire  When the tool is  removed  the wire is clamped into the connector        2 4 Installation    July 2003    Wiring The Remote  Transmitter   J6    Wiring The XDP Alarm  Relays   J3    Terminal block J6 contains the analog signal and loop power  connections for the remote transmitter  To wire the transmitter   complete the following steps     1  Connect the XDP to the transmitter as follows     Note  Refer to the transmitter manual for details on wiring the  transmitter  s terminal block s      a  Connect pin  1 to the positive terminal of the analog output  from the transmitter     b  Connect pin  2 to the  24 VDC loop power input on the  transmitter     c  Connect pin  3 to the return connection of the analog output  from the transmitter     Proceed to the next section to continue wiring the XDP     Terminal block J3 contains connections for the XDP alarm relays  To  wire these alarms  complete the following steps     Note  A failsafe alar
62. ight  Next   Then  press  the  ENTER  key     Note  The value  x x  shown at the above prompt is the current  measured value for the active gas curve     The meter now recalibrates its span point           Gas Curve  Man  Cal  Hit Next When Stable  x x H2 CO2                    lt Back   Next gt    Cancel       When the displayed reading  has stabilized  use  i   and        to highlight  Next   Then  press  the  ENTER  key        Setup and Calibration    4 23    July 2003    Performing a Span  Calibration  cont      Viewing and Clearing the  Cal Drift    At the next prompt  the new span drift value and an indication that the  meter passed or failed the calibration attempt are displayed           Gas Curve  Man  Cal  Span Drift  x xx mA Passed     lt Back   Finish      If the calibration    Passed      use  i   and       to highlight   Finish   Then  press the   ENTER  key                          Note   f the calibration  Failed   repeat the calibration procedure   If another failure occurs  contact the factory for assistance           Gas Curve  Manual      Press  ESC  three times to  return to the Cal Menu        Zero                Press  ESC  twice to return to the main menu  Then  press  ESC   again to exit the User Program  or proceed to the appropriate section  of this chapter to continue programming the meter     The XDP internally monitors the amount of drift from the factory  calibration  and the operator may view the drift values and clear them   if desired  To acc
63. ion supplied with the specific  transmitter used for complete information on the transmitter     The XDP performs the following basic functions       supplies the 24 VDC required to power the transmitter     accepts and processes the 4 20 mA output from the transmitter    shows a selected process parameter on a digital display     provides a user interface for programming the system      provides user configurable alarm relays and analog outputs    The XDP is wall mounted via its integral mounting plate and it is  connected to the transmitter with an electrical conduit  Both  components can be rated for use in Class I  Division 1  Groups B  C  amp   D hazardous environments     The XDP incorporates the following basic features       wall mountable  explosion proof enclosure     40 W universal VAC power input   e 24 VDC power output     4 20 mA analog input     high resolution  backlit LCD digital display     infrared     through the glass     programming keypad  e a Windows like graphical user interface     three built in gas measurement ranges     two isolated 4 20 mA analog outputs     four general purpose alarm relays      built in error reporting diagnostics    See Figure 1 1 on page 1 2 for a picture of the XDP        General Information    1 1    July 2003       Basic Features  cont      System Overview       Figure 1 1  The XDP    In a binary gas mixture  the transmitter measures the concentration of  a target gas in a known background gas  see the documentation for  t
64. iring  See Part Name  Wiring Diagram                        2 10  Wiring Insertion Tool                     2 4  Index 3    July 2003  ees    Index  cont      X  XDP  Calibrating                          4 20  Cleaning  u    ae 3 1  Mounting                            2 3  Mounting Dimensions                  2 9  Photograph                           1 2  Powering Up                         3 2  Removing Covers                     2 4  System Functions                     B 1  Wining a TT 2 4  Wiring Diagram                      2 10  Z  Zero GAS  paa ne PBA me Mey NDN 4 20       4 Index    GE Panametrics D  CLARATION    CONFORMITY    GE Panametrics  Shannon Industrial Estate  Shannon  Co  Clare  Ireland    declare under our sole responsibility that the    APX Advanced Paramagnetic Oxygen Transmitter  XDP Explosion Proof Display Package    to which this declaration relates  are in conformity with the following standards       EN 61326 1998  Class A  Annex A  Continuous Unmonitored Operation    EN 61010 1 1993   A2 1995  Overvoltage Category II  Pollution Degree 2    following the provisions of the 89 336 EEC EMC Directive and the 73 23 EEC Low Voltage Directive     The units listed above and any ancillary sample handling systems supplied with them do not bear CE  marking for the Pressure Equipment Directive  as they are supplied in accordance with Article 3  Section 3   sound engineering practices and codes of good workmanship  of the Pressure Equipment Directive 97 23   EC f
65. lly 0 00             Gas Curve    H2 CO2 Use  i   and       to position the  cursor under the desired char     acter  Then  use  i   and  E  to  XX XX     increment the value  Press    Enter Zero Gas Pcnt                       ENTER  when done           A  YES  response at the next prompt sends the last live value to the  analog output  while a  NO  response sends the calibration gas value  to the analog output        Gas Curve  Man  Cal Use  i   and       highlight the  Hold 4 20 mA Output desired answer  YES or NO   to the question shown  Then                 SR   During Calibration  press the  ENTER  key     YES   NO   Gas Curve  Man  Cal Make sure the zero gas is       flowing  and use  l   and       to  highlight  Next   Then  press  xx H2 CO2 the  ENTER  key       Back   Next gt    Cancel    Note  The value  x x  shown at the above prompt is the current  measured value for the active gas curve     Introduce Cal Gas                   The meter now recalibrates its zero point           Gas Curve  Man  Cal When the displayed reading  Hit Next When Stable has stabilized  use  i   and        to highlight  Next   Then  press  xx H2 CO2 the  ENTER  key       Back   Next gt    Cancel    At the next prompt  the new zero drift value and an indication that the  meter passed or failed the calibration attempt are displayed                             Gas Curve  Man  Cal If the calibration    Passed      use  i   and       to highlight   Finish   Then  press the  x xx mA Passed  ENTE
66. m is wired to the normally closed  NC   contacts  while a non failsafe alarm is wired to the normally     open  NO  contacts     1  Connect the Normal alarm as follows  this alarm is factory   configured for failsafe operation      a  Connect pin  1  NC  to the alarm device input   b  Connect pin  2  C  to the alarm device return   c  Pin  3  NO  is not used for fail safe operation    2  Connect the XDP Fault alarm as follows  this alarm is factory   configured for failsafe operation      a  Connect pin  4  NC  to the alarm device input   b  Connect pin  5  C  to the alarm device return     c  Pin  6  NO  is not used for fail safe operation       Installation    2 5    July 2003    Wiring The XDP Alarm  Relays   J3  cont      Wiring the XDP Analog  Output   J8    Wiring the Optional  Serial Port   J4    3  Connect the XDP Low alarm as follows  this alarm is factory   configured in failsafe mode  but may be reconfigured      a  Connect pin  7  NC  to the alarm device input   b  Connect pin  8  C  to the alarm device return   c  Pin  9  NO  is not used for fail safe operation     4  Connect the XDP Low Low alarm as follows  this alarm is factory   configured in failsafe mode  but may be reconfigured      a  Connect pin  10  NC  to the alarm device input   b  Connect pin  11  C  to the alarm device return   c  Pin  12  NO  is not used for fail safe operation     Proceed to the next section to continue wiring the XDP     Terminal block J8 contains connections for the XDP   s anal
67. n be switched at any time  between alternate inputs  an alarm or a recorder cannot be  switched on the fly to respond to alternate inputs  They must  be specifically configured to accept one input at a time        Setup and Calibration 4 3    July 2003       Adjusting the Display  Contrast    After entering the User Program  as described on page 3 4  the main  menu appears  To adjust the contrast of the LCD display  proceed as  follows              Cal   Opt   Disp                   User             Use the       and       keys to  select  Opt  and press   ENTER      Use the  i   and  E  keys to  select  User  and press   ENTER      Note     Lock All    and    Versions    menu options also appear on the    above list              User          Analog Output    Backlight                   Use the  i   and  E  keys to  select  Contrast  and press   ENTER      Note  The    Fault Alarm  and    Comm    options are also available in    the above list box                 User  Adjust Contrast    Adjust Contrast                            User           Analog Output    Backlight                Use the  i   and  E  keys to  increment the value to a num   ber between 0  min  contrast   and 99  max  contrast   When  done  press  ENTER      You are now back at the User  Menu     Press  ESC  as many times as necessary to return to normal RUN  mode  or proceed to the appropriate section of this chapter to continue  programming the XDP        4 4    Setup and Calibration    July 2003      
68. o the alarm device return     3  For fail safe operation  connect pin  3 to the alarm input       B 2    Additional Wiring Connections    Appendix C    Menu Maps    Operator Level  Cal  and  Disp  Menu Map                   C 1  Operator Level  Opt  Menu Map                            C 2  Setup Level Menu Map                                   C 3    Factory Level Menu Map                                 C 4    July 2003       password       MAIN MENU                 gt                                                                                   CAL MENU                   AIR CO2          Select Active Curve             Clear Cal Drift       H2 CO2             H2 AIR mA             Zero Span Drift       Clear Drift        Span Gas Pcnt Zero Gas Pcnt    Hold O    Introduce    utput  Hold Output     Cal Gas Introduce Cal Gas    Figure C 1  Operator Level  Cal  and  Disp  Menu Map    Next When Stable Next When Stable                      Menu Maps    C 1    July 2003       password       MAIN MENU               gt                           E                                               Versions    USER MENU             system functions     91 52 pcnt          Software Versions ENT                      Fault Alarm                       Non Fail Safe       Fail Safe                   ENT    Adjust Contrast                                                          Zero Set Value       Span Set Value       Test Percent                Zero Trim Value       ENT Span Trim Value   
69. og output   To wire this analog output  complete the following steps     1  Connect pin  1 to positive input on the analog output device   2  Connect pin  2 to the negative input on the analog output device   3  Pin  3 has no connection     Proceed to the next section to continue wiring the XDP     Terminal block J4 contains connections for the XDP   s serial port  The  connections on terminal block J4 are optional  in that they are not  required for proper operation of the analyzer  To wire the serial port   complete the following steps    Note  This connection may be made with a GE Panametrics   704 668 cable  If this cable is used  pin  1 is the red wire   pin  2 is the white wire  and pin  3 is the green wire    1  Connect pin  1 to the transmit pin on the computer    2  Connect pin  2 to the receive pin on the computer     3  Connect pin  3 to the return pin on the computer     Proceed to the next section to continue wiring the XDP        2 6    Installation    July 2003    Wiring The Line Power    The terminal block on the power supply board contains connections  for the line power to the XDP  Make these connections only with wire  that meets the following specifications       individual conductor gauge of 18 AWG minimum  maximum  current capacity of 12 Amps       voltage rating of 600 Volts minimum      insulation temperature rating of 105  C minimum    IWARNING   The rating of the wire insulation must be at least  15 C above the expected ambient temperature     Refer 
70. omplish this  access the main menu of the User  Program  see page 3 4  and proceed as follows           Opt Disp If necessary  use the  1   and        keys to select  Cal  and    press  ENTER                        Press  ENTER                           Range Cal Use the  i   and       keys to  select  Cal  and press   ENTER    Slct Cal Menu Item                4 24    Setup and Calibration    July 2003    Viewing and Clearing the  Cal Drift  cont         Range   Cal      View Cal Drift  Clear Cal Drift                Use the  i   and  E  keys to  select  View Cal Drift  and  press  ENTER      Note  The    Manual Cal  menu option also appears on the above list          ox         View Cal Drift    Zero Drift x xx       Span Drift x xx             View Cal Drift  Clear Cal Drift                Clear Cal Drift  Clear Drift  Warning  Data Will Be Reset                    ms i             Range   Cal      View Cal Drift                Clear Cal Drift       Both the zero and span drift  values  in mA  are shown   Press the  ENTER  key when  you are ready to continue     Use the  i   and  E  keys to  select  Clear Cal Drift  and  press  ENTER      Use  i   and       highlight the  desired answer  YES or NO   to the question shown  Then   press the  ENTER  key     Press  ESC  to return to the  Cal Menu     Press  ESC  as many times as necessary to return to normal RUN  mode  or proceed to the appropriate section of this chapter to continue    programming the XDP        Setup and Cal
71. or DN lt 25     Shannon   June 1  2002 p Ct i  Mr  James Gibson  GENERAL MANAGER          LS  EN ISO 9002 T  V ESSEN    Shannon ISO 9001  U S        CERT DOC Rev G 5 28 02    GE Panametrics D  CLARATION    CONFORMITE    GE Panametrics  Shannon Industrial Estate  Shannon  Co  Clare  Ireland    d  clarons sous notre propre responsabilit   que les    APX Advanced Paramagnetic Oxygen Transmitter  XDP Explosion Proof Display Package    relatif 4 cette d  claration  sont en conformit   avec les documents suivants       EN 61326 1998  Class A  Annex A  Continuous Unmonitored Operation    EN 61010 1 1993   A2 1995  Overvoltage Category II  Pollution Degree 2    suivant les r  gles de la Directive de Compatibilit   Electromagn  tique 89 336 EEC et de la Directive  Basse Tension 73 23 EEC     Les mat  riels list  s ci dessus ainsi que les syst  mes d   chantillonnages pouvant   tre livr  s avec  ne portent  pas le marquage CE de la directive des   quipements sous pression  car ils sont fournis en accord avec la  directive 97 23 EC des   quipements sous pression pour les DN lt 23  Article 3  section 3 qui concerne les  pratiques et les codes de bonne fabrication pour l ing  nierie du son     Shannon   June 1  2002 pa En  Mr  James Gibson    DIRECTEUR GENERAL                 LS  EN ISO 9002 T  V ESSEN    Shannon ISO 9001  U S        CERT DOC Rev G 5 28 02    GE Panametrics KONFORMIT  TS   ERKLARUNG    GE Panametrics  Shannon Industrial Estate  Shannon  Co  Clare  Ireland    erkl  ren  in
72. pci  panametrics co jp    Korea   Kwanghee Bldg   201  644 2  Ilwon dong  Kangnam Ku  Seoul 135 945   Korea   Telephone 82 2 445 9512  Fax 82 2 445 9540   E mail jkpark  panaeng co kr    Spain   Diamante 42   28224 Pozuelo de Alarcon   Madrid   Spain   Telephone 34  91  351 82 60   Fax 34  91  351 13 70   E mail info  panametrics infonegocio com    Sweden   Box 160    147 23 Tumba   Sweden   Telephone  46  0 8 530 685 00  Fax  46  0 8 530 357 57  E mail pana  panametrics se    Taiwan   7th Fl 52  Sec 3 Nan Kang Road  Taipei  Taiwan   ROC   Telephone 02 2788 3656   Fax 02 2782 7369   E mail rogerlin lumax com tw    United Kingdom   Unit 2  Villiers Court   40 Upper Mulgrave Road  Cheam   Surrey SM2 7AJ   England   Telephone 020 8643 5150   Fax 020 8643 4225   E mail uksales     panametrics ie       
73. point          Note  The Test  Zero Trim  and Span Trim options are also  available in the above list box           User Aout Aout Type Use the       and    l keys to   0 20mA select the desired system ana   log output type  and press   ENTER     0 2V                Note  The    Namur    option is also available in the above list box   The Namur NE43 specification requires a 4 20 mA output to  be clamped in specific bands to indicate out of range errors           User Aout    You are now back at the sys   tem analog output setup menu        Zero Setpoint  Span Setpoint             Either press  ESC  until you return the User Menu or proceed to the  appropriate section to continue setting up the system analog output        Setup and Calibration    4 9    July 2003    Setting the Zero Setpoint Enter the system analog output menu  as described on page 4 8  and  proceed as follows           User  Aout Use the       and  E  keys to  select  Zero Setpoint   and    press  ENTER             Aout Type        Zero Setpoint       Span Setpoint             Note  The Test  Zero Trim  and Span Trim options are also  available in the above list box     At the next prompt  enter the desired zero setpoint  the percentage of  target gas that corresponds to the lower end of the specified system  analog output range         User Aout Zero Set Use  i   and       to position the  cursor under the desired char   acter  Then  use  i   and  E  to                   He increment the value  Press   A  E
74. ral Information    Chapter 2    Installation    Introduction 2 os eke cee PER arr aan den 2 1  Choosing an Installation Site                               2 2  Mounting the XDP System                                 2 3    Wiring the XDP   ia cip  re KARA et PAG 2 4    July 2003    Introduction    This chapter provides a general description of the XDP Explosion   proof Display Package and gives directions on how to install and wire  the unit for proper operation     IMPORTANT  The XDP requires an external power disconnect  device such as a switch or circuit breaker  The  disconnect device must be marked as such  clearly  visible  directly accessible  and located within 1 8 m   6 ft  of the unit     Be sure to observe all installation limits and precautions described in  this chapter  Pay particular attention to the ambient temperature range  of    10 to  65  C  14 to  149  F  specified for the instrument     IWARNING    To ensure safe operation of the XDP  the unit must  be installed and operated as described in this  manual  Also  be sure to follow all applicable local  safety codes and regulations for installing electrical  equipment  In addition  all procedures should be  performed by trained service personnel     Proceed to the next page to begin the installation        Installation    July 2003    Choosing an  Installation Site    All environmental and installation factors should have been discussed  with a GE Panametrics applications engineer or field sales person at  t
75. ram is now locked   To program another meter function  proceed to the appropriate  section of this chapter   4 18 Setup and Calibration    July 2003    Versions    To view the current software versions installed in your meter  access  the main menu as described on page 3 4  and proceed as follows        Cal Disp Use the  I   and       keys to  select  Opt  and press   ENTER                        Use the  i   and  E  keys to  select  Versions  and press   ENTER           Lock Menus  Lock All    Versions                   Note  The    User    menu option is also available above              Versions After noting the software ver     0 BSERIES STD 001 0 sion data  press  ENTER      1 XDP H2G 001 B  2                    The following software information is listed in the above display     e Slot 0  controller main circuit board software  e Slot 1  XDP circuit board software      Slots 2 4  not normally used for the XDP    Note  The versions of the software installed in Slots 1 and 2 at the  above prompt are used only as an example  The actual  versions of your software will appear instead           Press  ESC  to return to the  main menu          Lock Menus  Lock All    Versions                   Press  ESC  until you exit the User Program  or proceed to the  appropriate section of this chapter to continue programming the  meter        Setup and Calibration    4 19    July 2003    Calibrating the System    Initial Preparation    Prior to shipment  the XDP transmitter system is f
76. ration  Thus  at  various times  the following gas mixtures must be monitored     e 70 100  hydrogen in air    0 100  hydrogen in carbon dioxide    e 0 100  air in carbon dioxide    GE Panametrics offers several transmitters that are well suited for  operation in all of these environments  and the XDP has been  designed to permit quick and easy switching between the analysis and  display of these three gas mixture inputs     When factory configured for this typical application  the XDP s built   in alarms are set up as follows     XDP Normal Alarm  failsafe  tripped if      active curve is not H2 AIR    a  b  field calibration is in progress    e      any built in error condition occurs    d  power to the XDP is interrupted      XDP Fault Alarm  failsafe  tripped if   while in measurement mode   a  any error condition  except Low H2 or Low Low H2  occurs  b  power to the XDP is interrupted  Or  while in calibration mode   a  ADC is out of range  mA input under over range error occurs    b  power to the XDP is interrupted       General Information    1 3    July 2003  yq _ _   _   _          _ __    _ 2 _  _  _ _______ O LLLLLLLLLL LCD AAZ    Alarm Settings  cont     Low Alarm  field programmable  failsafe or non failsafe  tripped  in H2 AIR measurement mode by     a  H2 level below 85  setpoint  programmable       Low Low Alarm  field programmable  failsafe or non failsafe   tripped in H2 AIR measurement mode by     a  H2 level below 80  setpoint  programmable        1 4 Gene
77. s  it is necessary to leave the  measurement mode and enter the User Program as follows  see the  menu maps in Figures C 1 and C 2 on pages C 1 and C 2            While in measurement mode   press the  ESC    ENTER   and   ESC  keys in sequence     1 H2 AIR 8    91 52    pent                Enter the operator level password  2719  at the next screen  This  password grants access only to those menus needed by the operator of  the meter  A different password is required to access the more  advanced menus  see Chapter 5  Advanced Programming  for  details            Password    Use  i   and       to position the  cursor under the desired char   acter  Then  use       and  E  to  increment the value  Press      ENTER  when done     Enter Password  XXXX                            3 4    Operation    July 2003       Accessing the User  Program  cont      Exiting the User  Program    Error Messages               KU Opt Disp This screen shows the highest    level of the User Program   which is referred to as the  main menu                    To leave the User Program  proceed as follows from the main menu                       Opt Disp Press the  ESC  key   1 H2 AIR You are now back in normal  9 1 5 2 pnt run mode                    Notice that the padlock is no longer visible at the upper right corner  of the display window  Once the User Program has been accessed  with a valid password  the menus at that programming level remain  unlocked when you return to run mode  With the menus
78. sword   Consult Factory      Factory  Advanced  Level  password   Consult Factory    Access the User Program as described on page 3 4  the menu  structure that becomes available depends on which of the above three  passwords is entered at the following prompt                       Password Use  i   and       to position the   Enter Password cursor under the desired char    XXXX acter  Then  use  I   and  E  to  increment the value  Press   s  ENTER  when done    Cfg Opt Disp This screen shows the initial    level of the User Program  the  main menu                        To program any features of the XDP not covered in Chapters 3 or 4   proceed to the appropriate section of this chapter        Advanced Programming    July 2003    Setup Level  Programming    Configuring the Display    To program any of the features described in this section  access the  User Program as described on page 3 4 and enter the correct value   Consult Factory  at the password prompt  Then  refer to Figure C 3  on page C 3 and go to one of the following sections       configuring the display    configuring the 4 20 mA output    configuring the Low and Low Low alarms      testing all XDP alarms    After powering up  the XDP performs a series of internal tests  If the  display has not yet been configured  the following screen appears              x Unassigned 8 The padlock at the upper right  means that the menus are cur   rently password protected                    Before any data can be displayed  t
79. the  known data points for the selected gas curve  To enter the data     complete the following steps     1  Use  i   and       to position the cursor at the desired position in the    mA value  and use       and  E  to increment the number     2  Use  i   and       to position the cursor at the desired position in the       pent value  and use  1   and  E  to increment the number     3  Use  i   and       to position the cursor at the double arrows at the  far right  and use  i   and  E  to increment the number and switch    to a different calibration point     4  Repeat steps 1 3 until all of the data points have been entered                                         H2 CO2       AIR CO2    Gas Curve  H2  Data   mA pent A   Bixx xxxxx XXX XXXX v  1  6   Gas Curve                    Press  ENTER  when all of the  data points have been entered     Press  ESC  to return to the  setup submenu     To continue programming the setup menu  proceed to the appropriate  section for instructions  Otherwise  press  ESC  four times to return to    live measurement mode        Advanced Programming    5 17    July 2003    Loading Software To download a new software program into the XDP  access the User  Program as described on page 3 4 and proceed as follows     IMPORTANT  The download takes place via the XDP  s serial port   Before proceeding  make sure that the serial port is  properly connected and configured  Also  the new  software file must be available on the connected PC           Cf
80. tions     To protect the User Program from unauthorized modifications   password protection may be reinstated  To accomplish this  access the  main menu as described on page 3 4  and proceed as follows     Note   f the meter is powered down  the password protection is in  effect by default when the meter is powered up           Cal Disp Use the       and       keys to  select  Opt  and press   ENTER                           Use the  i   and  E  keys to  select  Lock Menus  and press   ENTER          Lock Menus  Lock All                   Note     User    and    Versions    menu options are also available  above           1 H2 AIR 8 The meter returns directly to  pent run mode  with the padlock  9 1 E 52 icon indicating that the User    Program is now locked                 To program another meter function  proceed to the appropriate  section of this chapter        Setup and Calibration    July 2003                                                             Lock All To lock all meter functions  such as display window selection  in  addition to the User Program  access the main menu as described on  page 3 4  and proceed as follows    Cal   Opt   Disp Use the  I   and       keys to  select  Opt  and press   ENTER    UM Disp Use the  1   and  E  keys to  select  Lock All  and press  Lock All IENTEBL  Note  The    User    menu option is also available above   1 H2 AIR 8 The meter returns directly to  pent run mode  with the padlock  9 1   52 icon indicating that the User  Prog
81. to Figure 2 2 on page 2 10 and complete the following steps     IWARNING   Before proceeding with the section  verify that the  line power has been turned off at the external  disconnect device     1  Attach a cable or conduit with the three line power conductors to a  cable entry port on the electronics enclosure     Caution   Be sure that the input voltage to the XDP complies with  the value specified at the time of purchase     2  Connect the line power leads to terminal block J1 as follows     a  Connect the  ine power lead  black  to pin L on the power  connector     b  Connect the neutral power lead  white  to pin N on the power  connector     c  Connect the ground power lead  green  to the chassis ground  screw shown in Figure 2 2 on page 2 10     3  Make sure that the factory installed jumper wire at pin G on the  power connector is secure     This completes the wiring of the XDP  Proceed to Chapter 3   Operation  for instructions on using the meter        Installation    2 7    July 2003       8 23  209                                                                                                                                                                                                                                                                                                                                                127                 6 00  152     8 00  203                          FRONT VIEW                                                          
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
図書館情報・公民館情報・文化財巡り旅    EINFÜHRUNG  Panels 170 series  microUSBケーブル01取扱説明書(共通)  NL Gebruiksaanwijzing voor afneembare kogel  "user manual"  Einbauhinweise und Betriebsanleitung  operating manual bedienungsanleitung mode d`emploi  Bedienungsanleitung    Copyright © All rights reserved. 
   Failed to retrieve file