Home

TMO2D Oxygen Analyzer - GE Measurement & Control

image

Contents

1. E205 l TA A x Zu R23 2 5 L STD ALARM QAU DIVe ALARM LAB GP 01 STANDARD VOLTAGE FUSE TABLE VIL TAG FUSE JUMP 100 126 172A 220 240 1 445 ro 00 Figure B 7 Oxygen Analyzer Display PCB Assembly Dwg 4703 1226 E 20 DRTN ES BRTN SSEMBLE PER IPC A 160 AND OR PANAMETRICS STANDARD 957 002 NOT INSTALL CR4 CRS RT3 RT4 R18 R13 AND TB4B DR WEATHERPROOF UNITS INSTALL JUMPER WS B 7 1 20 97 So AOSTO DV Co RES 00v 5 SH 3 TALL RES A R2 Y1 ue EN 1 ay 4 DV47 1 8W oo end NET IB TALe PSEN E2 P80C32 Ls x es U eADC ENS 54 H 2 gt El Po
2. Ues U30 B 8 1 20 97 TURN ARE AR U14 npe co XXX MO DD PAIN DOUT MAXIS9AC Z SHDN REF SH _1ADC_ENS ZICS U14 ES SH TSLLKO SCLK 10h Why AV AV f 5 RIM 6 4 7 ASIN 819 EJASAD N AY AOSTO CR4 AN mE 2 A NODO W 1 DOPO L rast el DIO Ole Amun aznmo Et VIN TEMP wvpgur esv REF STRIMULe GH D SEL 497803 4 Va 1 00 1 03 Be A VIOH U17 1 3 acl 6 FUT U20 zw mo Sr 78L SAC HC14 ar ADO 19 7 Di 45V _ ER m ue L ases at AO idooor 8 1 we 39 le EDI 5 T AR 5 22 A K Veg so Svr 3Ue4 x sa SH 1Ense amp lce BE VB 7 Poen det D8224 499K 1X
3. Oxygen Analyzer Display PCB Schematic Dwg 4700 1226 sh 2 lt lt lt lt xXXxxX WME ADS NECO QJ OKON OOPF 10 100V of tay BE ola lalol Noise GE Al 00T p J25 i 609617 MC7 C r af B a IN RTN Figure B 9 Ue HPR L20 24 15V e VIN VOUT VIN5 OUT C64 9 1000PF 1K V B 9 1 20 97 Fl red BL 1 1 SEE TABLE 4 WHT De SI Inx gt LO_ALRM HI_ALRM 705 517 JOR ASS PWR UN SI e Lo C33 4 7 100V HH E z 599 GM AO0T OT 44001 U30 Fl AAA L1 LM340 5 BL DIV 2 220UH HON m aes VIN VOUT K5 f SEE TABLE 4 Wi R42 86 6K 17 By k 3 We 4 d v S E d 15521 CHESS IS SEE TABLE 4 VD
4. ER PART IF I COV ET HEAD Cy sE CART 8 250014 HEATSINK I USED M WITH APPR P LUGS SR para CIEN 4 REQD 6 OOO mR 3 Enron HERREN INTERNAL GROUND SCREW EXTERNAL GROUND SCREW 6 N PREN 30 2 PLCS MINIMUM DF 5 THREADS ENGAGED THESE THREADS PERMANENTLY CEMENTED TO PREVENT DISASSEMBLY SEE NOTE 1 AS a R VD KO TAG DATA BE EF D MJ Kal Analyzer PANAMETRICS Model Na E Serial No Voltage 24V Watts epw J VIO MIN II PATENT NO 5012669 y E OR SIMILAR P D F HEATSINK WH ASSEMBLY RE TEST 714 426 AF HANGES TD THIS DRAWING REQUIRE PRIOR
5. zz nd OHJA SLZI Set Settling Time Enter the amount of time the process gas should remain on before the unit resumes taking measurements Enter the time in minutes and seconds Note A arms and recorders are not updated during the settling period Settling Time Press YES to enter the Set Settling MM SS XX XX Time option The X s represent the previously entered value Settling Time Use the numeric Keys to enter the MM SS 14 30 desired settling time then press YES to confirm the entry AUTO CALIBRATION MENU Press YES again to exit Set Settling Time AUTO CALIBRATION MENU Press YES again to exit Set Error Handling 4 24 Programming the TMO2D 9 30 93 Set Error Handling Use the Set Error Handling option to program the display screen the alarms and the recorders to respond to any invalid measurement that occurs during Auto Cal Follow the steps below to enable Error Handling set the display screen and select the alarm and recorder responses AUTO CALIBRATION MENU Set Error Handling Enable Error Handling YES no Auto Cal Error Effects Set Display Response Display AutoCal Error YES no Auto Cal Error Effects Set Alarm Response Select Alarm Effects No Effect Select Alarm Effects Trip High Select Alarm Effects Trip Low Select Alarm Effects Trip Both At this prompt pres
6. Recorder A 0 mA Value 960 X XX Recorder A 0 mA Value 960 0 00 At this prompt press NO until the Recorder Menu displays Press YES to enter the Recorder Menu Use the NO or arrow keys to move the brackets to A and press YES to confirm the selection Use the NO or arrow keys to select the desired recorder range Press YES to set the low end value The X s represent the previously entered value for 0 mA Use the numeric keys to enter a 0 O and O then press YES to confirm the entry Programming the TMO2D 9 30 93 The Recorder Menu cont Recorder A 20 mA Value 960 X XX Recorder A 20 mA Value 0 100 0 Select Recorder to set A B done Press YES again to proceed to the high end value The X s represent the previously entered value for 20 mA Use the numeric keys to enter a 1 0 0 and 0 then press YES to confirm the entry Press YES to exit Recorder B is programmed the same way After entering the necessary values Select Recorder to set A B DONE MAIN MENU Recorders Use the NO or arrow keys to select DONE Press YES to exit This completes the Recorder Menu Press NO to proceed to the next Main Menu title 4 10 Programming the TMO2D 9 30 93 The Alarm Menu The third Main Menu is the Alarm Menu The TMO2D is provided with two four pol
7. option then press YES Use the NO or arrow keys to select HOURS Press YES to enter the number of hours The X s represent the previously entered hours Use the numeric keys to enter a 1 and 2 then press YES to confirm the entry To set the time interval for 3 days AUTO CALIBRATION MENU Set Time Interval Select AutoCal Interval hours DAYS Press YES twice to re enter the Set Time Interval submenu Use the NO or arrow keys to select DAYS Programming the TMO2D 4 19 9 30 93 Set Time Interval cont Auto Cal Interval Days X Auto Cal Interval Days 3 Press YES to enter the number of days The X s represent the previously entered days Use the numeric keys to enter a 3 then press YES twice to confirm the entry and proceed to the next option Now enter a time of day at which Auto Cal will occur Enter the time in 24 hour format For example set the Auto Cal to occur at 1 45 P M 13 45 in 24 hour time Auto Cal at Time HH MM 13 45 AUTO CALIBRATION MENU Set Time Interval AUTO CALIBRATION MENU Set Zero Gas Use the numeric keys to enter a 1 3 4 and 5 1 45 pm then press YES to confirm the entry Press YES again to exit Press NO to proceed to the next option 4 20 Programming the TMO2D 9 30 93 ese Set Zero Gas The Set Zero Gas option is used to enter the o
8. AUTO CALIBRATION MENU Perform AutoCal Auto Cal in Progress Span Gas is ON XX XX Auto Cal in Progress Measuring XX Auto Cal in Progress Zero Gas is ON XX XX Auto Cal in Progress Measuring XX Auto Cal in Progress Settling XX XX Working AUTO CALIBRATION MENU Perform AutoCal AUTO CALIBRATION MENU Zero AutoCal Press YES to activate Auto Cal The X s represent the countdown for gas equilibration time When the span gas equilibration time reaches 0 the following prompt occurs When the span gas measuring countdown reaches 0 the following prompt occurs When the zero gas equilibration time reaches 0 the following prompts occur When the span gas measuring countdown reaches 0 the following prompt occurs When the process gas settling time reaches zero the unit immediately applies the Auto Cal results to the calibration data before returning to the Auto Cal menu Press NO to proceed to the next option Programming the TMO2D 9 30 93 Zero AutoCal Selecting this option causes the Display to reset the Auto Cal drift curve to the original calibration data AUTO CALIBRATION MENU Zero AutoCal Zero AutoCal YES no Working AUTO CALIBRATION MENU Zero AutoCal View Drift Curve Press YES to enter the Zero Auto Cal option Use the NO or arrow
9. mae ar n m n SINIKONES booa 0 00 jaa co OSOS topo O JO jn w EN Coro UUUUUUUU Polo no TYPO CO S COT oo O C 4 CO Ne Lee UU c 2 no al To NU D CX op UUUUUU DIDI TO NOUR OOM UU UU LL MOK NOUR WRF O CREF ONLY VCC 8 SN S1S5 Lk SH amp 3 8 5 RESP P3 3 UDU oo Mime RTN RETURN RTN 4 VD Ou UUUUUUUU DUNN MONUI Js COP O U CX G2 CIC u cee ce lee um ONO CO Pol O UUUU POs N MAXS15C REF IN 704 442 RS 232 CABLE ASSY The g DISPLAY DSi aa a em ea NIKAN OUOUUUUUOULU OOO0O000o00 MIOSULHSCOTO ES S SCLK p M Ba BOTONES UG He VOUT DS1202 PFI jg PFU 2 A al XL wDI wDo H RES E MAX6910 R CEIN CEQUT OSCIN RES U3 OSCSEL RES RES AM27C010 ded es 841 7 150NS PL6 RiR NE CE RTN SCLKSH 2 i 16 PD AnI OOO 4100s D c ooo OO OMNA IC a a a m e NID OESS C O O u ROO B0 fas ee os oe oo oe oo MOS UI 4 COPO r7 O roro os zs m 00 ES OONDUBRWMTHO O14 COO CO 7
10. 83 GAS INLET DUTLET 1 4 NPTF ALL DIMENSIONS ARE REF ui ESSE THIS DRAWING WEIGHT 29 LBS IUR AUTRURIZATION F 4 3 Kg RI MUTUAL LISSER amp ENSIONS INCH mm Fi WIRING DIAGRAM 701 015 701 016 AND 701 017 BE INSTALLE CURDANCE WI ICAL SAFETY RE Figure B 1 TMO2 Outline and Installation Dwg 2712 211 LOW 9 Cees 1 CLE 5 ls SERVICE SHOWN WITH gt ALLED ALTERNATE DI SI W F CWPF gt UNIT B 1 264 271 5 50 DIA 4 HOLES SEE NOTE 4 amp 5 NOTES 1 MATERIAL 12 THK 6061 T6 ALUMINUM ALLOY 2 FINISH CLEAR IRIDITE 3 REMOVE ALL BURRS amp SHARP EDGES 4 FRONT SURFACE TO BE FREE OF ALL SCRATCHES BLEMISHES AND TOOL MARKS 5 PAINT FRONT SURFACE amp ALL EDGES DARK GRAY COLOR 26152 PER FED STD 595 6 10 SEMIGLOSS CLEAR EPOXY TOP COAT 125R TYP 4 PLACES 43 TYP BOTH ENDS 1 20 97 Figure B 2 TMO2D Rack Model Outline and Installation Dwg 421 255 B 2 G NC NIMUM OF 8 EADS ENGAG
11. N FROM FACTORY MUTUAL AUTHORIZATI AND ISSEP 1 20 97 Figure B 3 TMO2 Oxygen Analyzer Assembly Dwg 705 468 B 3 n N a LENZ N 92191 190195 a 9 4 DO NOT TRIM LEADS INSTALL ON LO POSITION 1 GAIN SWITCH R41 O2 GAIN ADJ TB1 4 COMP TB1 3 O2 TB1 2 RTN TB1 1 VIN 703 1036 E13 RTN E12 15V m RIO RI3CI9 E5 COMP INPUT R27 ES 10M E14 OFFSET COMP GAI R29 R44 R20 ZERO ADJ 10V REF 10V REF INVERSE CO THIS DRAWING IS A FACTORY MUTUAL RESEARCH APPROVED DESIGN amp NO CHANGE IS PERMITTED UNLESS AUTH
12. SETUP MENU Press YES to set the display Set Display Display Compensation Then use the NO or arrow keys to YES no move the brackets to the desired entry SETUP MENU Press YES to confirm the entry and Set Display exit SETUP MENU Press NO to proceed to the next Set Communications submenu The Set Communications submenu has two choices 1 Baud Rate and 2 Update Rate Baud Rate The Display contains a bi directional industry standard RS 232C serial port that allows operation of the instrument remotely with a terminal or computer that supports the RS 232C protocol All keypad operations and most display operations can be performed remotely via this port The Display supports the communication rates of 9600 4800 2400 1200 and 300 baud The default setting is 9600 baud Word size is fixed at 8 bits 1 stop bit no parity Note 300 baud is provided for compatibility with older equipment however the use of 300 baud greatly limits the computation speed of the TMO2D and its use is not recommended 4 6 Programming the TMO2D 9 30 93 Set Communications The Set Communications submenu is used to set a baud rate and an cont update rate SETUP MENU Press YES to enter the Set Set Communications Communications submenu SET COMMUNICATIONS Press YES to enter the baud rate Set Baud Rate SEL
13. US TLE2064C El z 1 OVREF 10VREF over o Lo o 31 TABLE 2 ATIENUATE REF SELECT R27 COMP ors ne 14 NOTE 4 SAI SAI 6 7 i y 1 1 R28 COMP K GAIN SEE TABLE 3 LF444 D 10K Qs 181 TORE BIA TBIB ajo 02 R12 CRN SPECIALS 0 4 20MA Qs VPO109N3 COMP 0 4 20MA THIS IS A FACTORY MUTUAL RESEARCH APPROVED DESIGN AND NO CHANGE IS PERMITTED UNLESS AUTHORIZED BY FACTORY MUTUAL 1 20 97 Analyzer PCB Schematic Dwg 2700 1036 Figure B 6 REF DES UNUSED NOTES 1 AXIAL LEAD RESISTOR VALUES ARE EXPRESSED IN OHMS 1 RN55 UNLESS OTHERWISE SPECIFIED UNLESS OTHERWISE SPECIFIED HIGHEST EVER USED mE 2 CAPACITOR VALUES ARE EXPRESSED IN MICROFARADS 50V UNLESS OTHERWISE SPECIFIED C24 3 INDICATES MARKING Q5 SANA 4 R46 MAY BE REMOVED DURING CALIBRATION PROCEDURE NOT USED TABLE 1 IC POWER VRA REF DES TYPE 25V U9 Ul LM340 u2 LT1054 U3 LM3524 w2 LF444 SAI OP200 MTI QUA U5 U6 U7 U8 Uo TABLE 4 OPTION GROUPS XMIR OUTPUT 020 420 ORION 020 420 GAS COMP 02 03 v Me E 3 R35 PRESS COMP 04 05 R36 01 STANDARD U5 TLE2064C R37 E18
14. No 251 Min Hwa Road Ku Shan District Kaohsiung Taiwan Telephone 886 7 552 5498 Fax 886 7 552 3596 E mail panaexim ms 11 hinet net United Kingdom Panametrics UK Limited Unit 2 Villiers Court 40 Upper Mulgrave Road Cheam Surrey SM2 7AJ England Telephone 020 8643 5150 Fax 020 8643 4225 E mail uksales panametrics ie
15. X XX corresponding comp input milliamp value for the given point Repeat the procedure for every point in the background grid The unit will then display the Working message before returning to the Calibrate Background option Working Calibrate Background At this prompt press NO to Enter Grid Data proceed to the next option Calibrate Background Press YES to return to the Done Calibration Menu Programming the TMO2D 4 35 9 30 93 No Compensation The second option in the Edit Calibration Data submenu is No Compensation for Normal or Damped Response Refer to page 4 30 Proceed to the following sections for NORMAL or DAMPED response CALIBRATION MENU Edit Calibration Data Enable Compensation yes NO Set System Response normal DAMPED At this prompt press YES to enter the Enable Compensation option Use the NO or arrow keys to select NO Press YES to confirm the entry and enter the Set System Response option Note Normal response may NOT be used in conjunction with either pressure or background compensation Refer to page 4 39 for Damped Response Normal Response Set System Response NORMAL damped Calibrate Oxygen Enter Grid Data O Grid Entry of Points 2 At this prompt use the NO or arrow keys to select NORMAL then press YES to confirm the entry Press YES to ente
16. prompt NO returns to the AUTO CAL Error Effects Set Display Response prompt AUTO CAL MENU Perform AutoCal YES or NO AutoCal in Progress Span Gas is ON XX XX TMO2D automatically calibrates Press NO to interrupt otherwise the unit calibrates and returns to the Perform AutoCal prompt NO Abort AutoCal Span Gas is ON YES or NO Abort AutoCal Settling TMO2D automatically aborts AutoCal Press NO to interrupt otherwise the unit settles and returns to the Perform AutoCal prompt NO Abort Settling Settling YES or NO YES returns to the AUTOCAL MENU Perform AutoCal prompt NO causes the TMO2D to resume settling then return to the Perform AutoCal prompt AUTO CAL MENU Zero AutoCal YES or NO Zero AutoCal YES no Use the NO or arrow keys to move the brackets to the desired selection YES pom Resets calibration AUTO CAL MENU View Drift Curve YES or NO Drift Values X XX 9602 X XX mA Press YES to view drift values YES NO AUTO CAL MENU Done YES or NO YES returns to the CALIBRATION MENU Auto Cal Parameters prompt NO returns to the AUTO CAL MENU Set Time Interval prompt CALIBRATION MENU Edit Calibration Data YES or NO Enable Compensation YES no Use the NO or arrow keys to move the brackets
17. provides increased gain Note Do not use the 0 position on Sl R27 and R28 provide the offset and gain adjustments for the compensation circuits High or low bridge voltage is selected via jumper plug J3 Low voltage should be used for most applications High voltage is used with hydrogen and helium background gasses Most analyzers will be magnetized and marked as high H medium MD or low L assemblies The oxygen ranges that can be calibrated for each type of magnetization are Magnetization O Range High H 0 1 2 5 10 21 Medium M 0 2596 95 100 Low L 0 100 D 4 The TMO2 Transmitter Calibration Procedure 11 30 92 Internal Background Gas Compensation Zero Internal Background gas compensation is used to provide a single 0 4 20 milliamp oxygen signal that is constant despite changes in the gas composition Note In the following procedure the gas mixture having the largest oxygen sensitivity should be used as Gas A Set SA1 switches as follows 1 3 4 amp 6 OFF up position 2 amp 5 ON Set J3 to the LO position HI if using hydrogen helium 1 Set S1 to position 9 and adjust R41 fully clockwise 2 Apply background Gas A ex 100 CO into the oxygen transmitter 3 Adjust R27 until the voltage at E14 equals 2 0 volts 4 Adjust R28 until the voltage at E10 equals 2 0 volts 5 Turn R20 fully clockwise 6 Adjust R29 until TB1 3 is 0 4 milliamps 7 Tu
18. 2 1 1 20 97 Mounting the TMO2 The TMO2 Transmitter is shown in Figure 2 1 below For optimum Transmitter response mount it as close as possible to the process being monitored preferably within 5 feet 1 52 meters Be sure to keep the Transmitter level to within 15 of vertical and provide at least 9 inches of clearance above the top cover of the Transmitter to allow access to the printed circuit board PC Board for calibration and maintenance Note Be sure to have the correct certified configuration if the Transmitter is mounted in a hazardous area Ambient temperature must be 20 C to 40 C 4 to 104 F Figure 2 1 The TMO2 Transmitter in Mounted Position 2 2 Installation 1 20 97 Mounting the TMO2 A complete Sample System includes the Transmitter and associated Sample System sample tubing and components already mounted to a metal panel Figure 2 2 below shows a Sample System which includes calibration gas inlet valves filter coalescer with drain valve pressure regulator and bypass flowmeter in addition to the mandatory sample flow meter and flow regulating valve Mount the TMO2 Sample System as close to the process as possible Once the Sample System is mounted connect the oxygen sample input and output lines via the 1 4 tube fittings at the bottom of the Transmitter The sample line is generally recommended to be 1 4 stainless steel tube and 1 4 stainless steel tube fittings Attempt to minimize the samp
19. 24VDC Red Display Terminal Block Power Return Black Oxygen White Compensation Blue or Green IWARNING BE SURE TO PLUG UP THE UNUSED CONDUIT HOLE ON THE SIDE OF THE TRANSMITTER TO MAINTAIN THE APPROPRIATE WEATHERPROOF OR EXPLOSION PROOF RATING Note Refer to Figure 2 4 on the following page for a detailed description of all wiring connections Installation 2 7 1 20 97 Tree aaa l Wiring Recorders Optionally the TMO2D can be wired to Recorders AutoCal and AutoCal System Alarms Alarms Figure 2 4 below diagrams all possible wiring connections etc Table 2 2 Maximum Cable T N4 Lengths and Gauges OXYGEN SPARE RECORDERS Cell INPUTS 0 4 20 mA AUTOCAL ALARMS di REC A REC BPROCESS CAL LOW HIGH 53088 ass IT Y Y Y 1 22 z2 CNC NOC NC NOC NC NCC NC NO SOLENOID VALVE POWER OPTIONAL RECORDER B PROCESS GAS INLET OUTLET AUTO CALIBRATION SYSTEM Figure 2 4 All Transmitter and Display Wiring Options 2 8 Installation 1 20 97 RS 232C Serial Port The TMO2D also contains a bidirectional industry standard RS 232C serial port J8 which can be connected to a terminal or computer that supports the RS 232C protocol See Figure 2 5 below for RS 232 wiring pin connections see Appendix C for a corresponding key chart Pin 2 Transmit from Pin 3 Receive to Pin 7 Return ground L Figure 2 5 Display RS 232C Serial Port Con
20. 9 30 93 The Setup Menu The first Main Menu is the Setup Menu The Setup Menu contains the following six submenus Set Time Set Date e Set Backlight e Set Display Set Communications Done These submenus allow the user to alter operating parameters Once entered these values will remain in the Display memory until they are changed Refer to Appendix E for a flow diagram of the Setup Menu Set Time The first display in the Setup Menu is Set Time which sets the current time in 24 hour format For example to enter 1 15 pm 13 15 in 24 hour time MAIN MENU Setup SETUP MENU Set Time Enter 24 hour time HH MM XX XX 13 15 Enter 24 hour time HH MM 13 15 SETUP MENU Set Time Press YES to enter the Setup Menu Press YES to set the time Then use the numeric keys to enter al 3 1 and 5 The X s represent the previous time entered Press YES to confirm the entry Press YES again to exit Press NO to proceed to the next submenu Programming the TMO2D 9 30 93 Set Date Set Backlight The Set Date submenu is used to set the current date in USA month day year format For example to enter February 24 1991 SETUP MENU Set Date XX XX XX 2 24 91 Enter Date MM DD YY 2 24 91 Enter Date MM DD YY SETUP MENU Set Date SETUP MENU Set Backlight Press YES
21. Electronic Display The Transmitter has no power switch and will begin operating once it has been connected to the Display and the power has been turned on Allow 90 minutes after power is turned on for the Transmitter to warm up and stabilize During this time establish gas flow through the system Open the necessary valves to establish the flow at 1 0 0 2 SCFH 500 100 cc min Make sure that nothing impedes the flow of gas thereby causing a pressure or vacuum buildup within the Transmitter For proper operation the Transmitter should be vented to the atmosphere Note The Transmitter has been calibrated at atmospheric pressure Unless pressure compensation is being used operating the Transmitter at other pressures will cause erroneous readings The Electronic Display unit contains a 2 line by 24 character backlit Liquid Crystal Display screen LCD On power up the Display unit tests its memory RAM then searches for valid calibration data from the Display as well as input from the TMO2 Transmitter If calibration data has already been entered into the Display the unit immediately begins taking measurements from the Transmitter and the LCD begins displaying percent oxygen concentration If valid calibration data has not been entered and stored in the Display or if the the Transmitter is not hooked up to the Display the LCD will display erroneous readings Proceed to Chapter 4 page 4 1 to enter data into the Display O
22. TABLE 3 VALUE OPTIONS BASE DIAGRAMS BOTTOM VIEWS U6 REF DES EMTER U7 1 20 97 D CESCK3 4 PON 22 N S PON POPO PMY VA UN Y EF ON A d Ci o19 v2 KLK2KS Kg STD E22 2 5V J dup 8 EI d TF 88 UMBER TABLE REC STD UTPUTS STD ALARM 08 DIV2 ALARM 03 2 a C39 ml 6 f STD ALARM GAIO4 N C E l DIV2 ALARM CHS 2 C STD ALARM 06 cag DIV2 ALARM 07 ceX Jg 7 N STD ALARM QAU8 ues m E e DIV2 ALARM EAIS DISPLA 5 SE STD ALARM 10 83 o DRINS e d E DIV2 ALARM 41
23. TMO2 Equipment Completely assembled TMO2 oxygen analyzer Sample Gas Equipment Cylinder of 10096 oxygen with regulator Cylinder of primary background gas Gas A Cylinder of secondary background gas Gas B e Sample system to control combine the above gasses Drawings 705 468 TMO2 Top Assembly e 703 1036 Assembly Analyzer PCB 700 1036 Schematic Analyzer PCB 712 211 Outline and Installation D 2 The TMO2 Transmitter Calibration Procedure 11 30 92 Setup Preliminary Explanation 1 Connect gas inlet of the TMO2 transmitter to the sample system 2 Connect 0 20 milliamp meter to TB1 of transmitter pin 3 and pin 2 return 3 Connect 24 VDC power supply to TB1 of the transmitter PCB Connect the positive terminal to pin 1 and the negative terminal to pin 2 of the TB1 Turn on power and let transmitter warm up for at least 30 minutes The following procedures describe how to calibrate various TMO2 oxygen analyzers Most analyzers have internal background gas compensation to correct the output for changes in the gas composition To begin the compensation will be fixed to 2 00v while the oxygen circuit is calibrated using zero and span ex 0 and 10 oxygen The primary background gas referred to as Gas A will compose the remainder of the mixture Next the compensation circuit is calibrated In this procedure the TMO2 transmitter is exposed to full scale oxygen ex 10 ina back
24. Up 0 2 0005 3 1 AutoCal K Entering Parameters 4 17 Wiring onen 2 8 Key Functions 0 05 4 2 Keypad B Description 0 0 0 0 ee ee eee 3 2 Key Functions 004 4 2 Backlight DELUD EN aes a ke eril us cuir 4 5 M Baud Rate EAEE EPIRA eens 4 7 Mounting C Display unless 2 4 Sample System 00 2 3 Calibration Menu 4 17 Transmitter 0 2 ee eee eee eee 2 2 Auto Cal Parameters 4 17 Edit Calibration Data 4 30 O Communications f E11 SER GENE aa adonde eni 4 6 Operation E E as EE EE EIE S 3 1 Connections Display ZINC LLLI 3 1 dn c ms 2 5 Establishing Gas Flow 4 56 3 1 Powering Up is tet RR 3 1 x P Date Setup ae adas adds 4 5 Perform AutoCal Option ps pyr ee SENG RS a Description Performance Specifications Display ds iem eee eee 3 1 Display 00 sees reer nn 5 2 Keypad es nee see eee ee bee eric 3 2 Power Pro Brea ase ccena cai cate cab iie 4 2 MAINE ond beeper aaa 2 10 Display Program l Description of 008 3 1 Description AE UM ANULUM a E i Menu Navigation 4 3 Display Navigation 43 Mounting een 2 4 Recorder Menu 0 0005 4 9 Performance Specifications 5 2 Physical Specifications 5 3 SEUD vie eo rer 4 6 Specifications 0 00 00 00 5 2 Ind
25. electronic measuring bridge circuit The circuit is unbalanced as the electrical resistance of the thermistors change with temperature The circuit imbalance causes a voltage to appear across the bridge which is proportional to the oxygen concentration in the gas being measured See Figure A 1 on the following page Thermoparamagnetic Magnetic Wind Theory A 1 4 8 92 TMO 1000 OXYGEN SENSOR WIND GENERATING COOLED THERMISTORS WIND Qm RECEIVING WARME THERMISTORS Figure A 1 The TMO2 Oxygen Transmitter The TMO2 Transmitter Compensation Bridge TMO2D Electronics Microprocessor based Compensation for Variations in Background Gas Composition As the background gases which make up the balance of an oxygen containing mixture change the thermal and magnetic properties change and affect heat transfer to and from the thermistors This is ultimately reflected in thermistor temperature and will affect oxygen measurement if left uncompensated For this reason the TMO2 transmitter features a unique bridge within a bridge circuit The oxygen measuring bridge described above is one arm of yet another compensation bridge which maintains the oxygen bridge at a constant temperature regardless of variations in background gas compensation By sensing the compensation bridge current necessary to restore oxygen bridge temperature a signal is obtained which is transmitted to the TMO2D electronics where microprocessor
26. enable compensation or select no compensation To Enable Compensation Enable Compensation Use the NO or arrow keys to select YES no YES Press YES to confirm the entry and enter the Select Compensation option The Select Compensation option allows the user to calibrate for either pressure or background compensation proceed to the following sections for a description of each choice 4 30 Programming the TMO2D 9 30 93 Edit Calibration Data To Calibrate for Pressure Compensation the user must enter both cont transducer data and pressure grid data Select Compensation Use the NO or arrow keys to select PRES bkgd PRES then press YES to confirm the entry Calibrate Pressure Press YES again to enter Transducer Data transducer data Pressure Curve Entry Use the numeric keys to enter the of Points 2 number of points in the pressure transducer curve then press YES to confirm the entry Point 1 Press YES again to enter the mmHg XXX X pressure in mmHg for the given point in the transducer curve Point 1 Use the numeric keys to enter the mmHg 767 3 value then press YES to confirm the entry Point 1 Press YES again to enter the P mA XXX corresponding mA value for the given point in the transducer curve Point 1 Use the numeric keys to enter the PmA 9 17 value then press YES to confirm the entry Repeat the procedure for each poin
27. keys to select the desired entry Select YES to zero the Auto Cal drift curve Select NO to cancel the process and return to the Auto Cal Menu Press NO to proceed to the next option Selecting this option allows the user to view the Auto Cal drift curve AUTO CALIBRATION MENU View Drift Curve Drift Values X XX 9602 X XX mA Press YES to scroll through the drift curve Press YES to exit and return to the Auto Cal Menu Note f the unit is uncalibrated and or has not undergone the auto cal procedure no calibration data has been previously entered there will be no points in the drift curve AUTO CALIBRATION MENU View Drift Curve AUTO CALIBRATION MENU Done CALIBRATION MENU Auto Cal Parameters CALIBRATION MENU Edit Calibration Data Press NO to proceed to the next option Press YES to exit Press NO to proceed to the next submenu Programming the TMO2D 4 29 9 30 93 Edit Calibration Data The Edit Calibration submenu contains the following two options Enable Compensation for Pressure and Background No Compensation for Normal or Damped Response Note Normal response may NOT be used in conjunction with either pressure or background compensation At the Edit Calibration Data prompt press YES to enter the Enable Compensation option CALIBRATION MENU Edit Calibration Data Proceed to the appropriate section to either
28. prompt select YES to enter the calibration data for the given point or select NO to have the unit determine the value for the given point Note Each curve must have AT LEAST TWO entered data points XX X 02 XXX X mmHg Enter Point YES no XXX 0 XXX X mmHg O2 mA X XX To enter the calibration data use the NO or arrow keys to select YES then press YES to confirm the entry Use the numeric keys to enter the corresponding milliamp value for the given point Repeat the procedure for every point in the pressure grid The unit will then display the Working message before returning to the Calibrate Pressure option Working Calibrate Pressure Enter Grid Data Calibrate Pressure Done CALIBRATION MENU Edit Calibration Data At this prompt press NO to proceed to the next option Then press YES to return to the Calibration Menu At the Edit Calibration Data prompt press YES to enter the Enable Compensation option Programming the TMO2D 4 33 9 30 93 Edit Calibration Data To Calibrate for Background Compensation the user must enter cont background grid data Enable Compensation YES no Select Compensation pres BKGD Calibrate Background Enter Grid Data O Grid Entry of Curves 2 O Grid Entry of Points 2 Point 1 O XX X At
29. 10 100V TMOZD p u 3 ae a P_ ASSEMBLY 3 12V0 DARDO WIRING et 09 7 D zo E WE OD W Os N Wal e V a PR_ALRM Oxygen Analyzer Display PCB Schematic Dwg 700 1226 sh 3 K6 CAL_ALRI Figure B 10 W7 AUTOCAL TABLE 4 uL TAGH FUSE JUMPER 100 180 1 24 WLW3 P20 240 1 44 We B 10 Appendix C RS 232C Serial Port 4 8 92 RS 232C Serial The TMO2D contains a bidirectional industry standard RS 232C serial Port port which can be connected to a terminal or computer that supports the RS 232C protocol see Chapter 4 page 4 6 All keypad and most display operations can be performed remotely through this serial port Table C 1 below provides an illustration of the terminal computer keys and how they correspond to the keys on the display keypad Table C 1 TMO2D RS 232C Seri
30. 34 Fax 43 1 602 25 34 11 E mail panametrics netway at Benelux Panametrics B V Postbus 111 3870 CC Hoevelaken The Netherlands Telephone 31 0 33 253 64 44 Fax 31 0 33 253 72 69 E mail info panametrics nl France Panametrics S A BP 106 11 Rue du Renard 92253 La Garenne Colombes Cedex France Telephone 33 0 1 47 82 42 81 Fax 33 0 1 47 86 74 90 E mail panametrics panametrics fr Germany Panametrics GmbH Mess und Pruftechnik Robert Bosch Stra e 20a 65719 Hofheim Germany Telephone 49 6122 8090 Fax 49 6122 8147 E mail panametrics t online de Italy Panametrics S r l Via Feltre 19 A 20132 Milano Italy Telephone 02 2642131 Fax 02 26414454 E mail info panametrics it Japan Panametrics Japan Co Ltd 2F Sumitomo Bldg 5 41 10 Koishikawa Bunkyo Ku Tokyo 112 0002 Japan Telephone 81 03 5802 8701 Fax 81 03 5802 8706 E mail pci panametrics co jp Korea Panametrics Korea Ltd Kangnam P O Box 1902 Seoul Korea Telephone 82 2 555 4611 Fax 82 2 556 4351 E mail panakor soback kornet nm kr Spain Panametrics Instumentaci n S L Diamante 42 28224 Pozuelo de Alarcon Madrid Spain Telephone 34 91 351 82 60 Fax 34 91 351 13 70 E mail info panametrics infonegocio com Sweden Panametrics AB Box 160 147 23 Tumba Sweden Telephone 46 0 8 530 685 00 Fax 46 0 8 530 357 57 E mail pana O panametrics se Taiwan Panametrics Exim Ltd 8F
31. ECT BAUD RATE Use the NO or arrow keys to scroll 9600 baud through the choices SELECT BAUD RATE 4800 baud SELECT BAUD RATE 2400 baud SELECT BAUD RATE 1200 baud SELECT BAUD RATE Press YES to select the desired 300 baud baud rate and exit SET COMMUNICATIONS Press NO to proceed to the next Set Baud Rate submenu Note The baud rate can be changed via a terminal connected to the RS 232C port However this is not recommended as the TMO2D will immediately change to the new baud rate Display and keyboard operation will not be correct until the baud rate of the terminal is changed to match the new baud rate set in the TMO2D SET COMMUNICATIONS Press YES to set the update rate Set Update Rate Programming the TMO2D 4 7 9 30 93 Set Communications cont Update Rate This option is used to change the interval at which data goes to the serial port The current data interval will appear in brackets and users can enter a new data interval Data intervals range from 0 to 300 seconds A data rate of zero will prevent data from passing to the serial port SET COMMUNICATIONS Set Update Rate SET COMMUNICATIONS Data Interval X 180 SET COMMUNICATIONS Data Interval 180 SET COMMUNICATIONS Set Update Rate SET COMMUNICATIONS Done SETUP MENU Set Communications SETUP MENU Done MAIN MENU Setup Pre
32. EMENTS YOU MUST SHIELD AND GROUND ELECTRICAL CABLES AS DESCRIBED IN APPENDIX F Installation 2 5 1 20 97 X AA Wiring the Transmitter to Remove the Transmitter cover by loosening the set screw and the Display unscrewing the cover The Transmitter PC board is located directly beneath see Figure 2 3 below Note There is a 1 printed on the PC Board next to the 1 pin connection on TBl 1 Route the cable into the Transmitter through one of the 3 4 conduit holes on the side of the Transmitter 2 Unplug TB1 on the Transmitter PC Board by carefully pulling it directly up without bending the pins attached to the board 3 Loosen the TB1 sidescrews and insert the colored wires into the corresponding openings on top of TB1 see Table 2 1 on page 2 7 for color coded pin connections 4 Tighten the sidescrews and carefully plug TB1 back onto the PC Board 5 Similarly connect the cable to the OXYGEN CELL Terminal Block located on the rear panel of the Display see Table 2 1 for color coded pin connections Set Screw Figure 2 3 Top View of Transmitter PCB and TB1 amp Protective Conductor Terminal 2 6 Installation 1 20 97 Wiring the Transmitter to the Display cont Table 2 1 Transmitter to Display Wiring Pin Connections Transmitter Wire Connections Terminal Block Power
33. J Y U3 gt ES o C Ull AL2 A14 JLAL2 VIN VBUTS EF DES TYPE 5 ARTN E RIN E U9 U17 74HC14 TC J5 5 C e U4 74HC00 u9 1 15 0 D DV Uli Lale R DT A hos 705 515 U19 78L054 K c E Pl PS 713 2 LCD DISPLAY ASSY u20 78L05A SAIS U21 6N137 3 7 R9RIS RI ge 5 Gt e 7 E Bs VD u22 6N137 E U23 AD822A RAM e U 1822 S LC E P pon p gt BASE DIAGRAMS eu4 26 Ue HPR 120 CETTTUWUEEPAESS va U29 MC34063 o E D weed lo 4 Oxygen Analyzer Display PCB Schematic Dwg 700 1226 sh 1 Figure B 8 U30 Lw340 5 3 8 AEN or E en us 1cL76604 gt ae T MS ce MAXSLSC 4 9 5 4 S K1 K2 K5 K6 K3 K4 K7 K8 01 02 03 04 06 C INPUT RTN lt CASE OUT
34. LLCEICEXkiCEGEXC X XNE Z JDC LG TlHILL LL LL The Alarms Menu cont Enter Alarm Setpoint Use the numeric keys to enter 2 Low 0 25 0 and 5 then press YES twice to confirm the entry and proceed to the next prompt Select Alarm to set Use the NO or arrow keys to select high low DONE DONE MAIN MENU Press YES to exit Alarms This completes the Alarms Menu Press NO to proceed to the next Main Menu title 4 12 Programming the TMO2D 9 30 93 The Tests Menu DVM Test The fourth Main Menu is the Tests Menu The Tests Menu provides assistance in installing and troubleshooting the TMO2D The Tests Menu contains the following four submenus DVM Test Recorder Calibrate Test Alarms Test Done Refer to Appendix E for a flow diagram of the Tests Menu The DVM test allows the Display to operate as a simple digital voltmeter to measure in milliamps the Transmitter s oxygen and compensation signals The Display is updated approximately twenty times per second facilitating connection and calibration of the Transmitter Follow the steps below to test the oxygen signal input from the Transmitter MAIN MENU At this prompt press NO until the Setup Test Menu displays MAIN MENU Press YES to enter the Test Menu Tests Tests Press YES to select the DVM DVM Test submenu Select DVM Input Use the NO or arrow keys to select GA
35. MENU Then use the NO or arrow keys to Set Time Interval view the eight options within this submenu AUTO CALIBRATION MENU Set Zero Gas AUTO CALIBRATION MENU Set Span Gas AUTO CALIBRATION MENU Set Settling Time AUTO CALIBRATION MENU Perform AutoCal AUTO CALIBRATION MENU Zero AutoCal AUTO CALIBRATION MENU View Drift Curve AUTO CALIBRATION MENU Done 4 18 Programming the TMO2D 9 30 93 Auto Cal Parameters cont The Auto Cal options are described individually in the following sections Set Time Interval The Set Time Interval option is used to set the time interval at which an Auto Cal will occur Pressing YES at the Set Time Interval prompt provides the option of selecting an interval of either hours or days Hours can be entered fractionally i e 90 minutes 1 5 hours up to a maximum of 24 An interval of zero hours prevents Auto Cal from occurring Days can range from 0 to 99 Fractional days are not permitted An interval of zero days prevents Auto Cal from occurring If a non zero number of days is entered the Display will prompt for the time of day when the Auto Cal should occur For example to set the time interval for 12 hours Scroll to the Set Time Interval AUTO CALIBRATION MENU Set Time Interval Select AutoCal Interval HOURS days Auto Cal Interval Hours XX XX Auto Cal Interval Hours 12 00
36. O 25 tess sists ehe ys e s pete or iE Ie wem ed a e perte ted D 9 span Compensation en rien E wa leis ER EID bled ra D 9 Appendix E Main Menu Flow Diagrams Appendix F Installation Instructions for CE Mark Compliance Installation Instructions for CE Mark Compliance 0 cece eee eee eee F 1 vii Chapter 1 Features and Capabilities Introduction to the TMO2D ssllsleeee System Description 0 0 0 cee eee eee System Operation is eis eevee ae crete eee Wee rr ees 11 30 92 Introduction to the In this section you will find the following TMO2D Anintroduction to the TMO2D Oxygen Analyzer and its components A description of the TMO2D configurations and accessories A brief discussion of the TMO2D operating system The TMO2D Oxygen Analyzer is a complete system that measures oxygen concentration from 0 100 in gaseous mixtures The main components of the Analyzer are the TMO2 Transmitter a measuring cell mounted within an aluminum housing and the TMO2D Electronic Display The TMO2D Electronic Display is connected to the TMO2 Transmitter via a 3 or 4 wire non shielded cable The Electronic Display provides user friendly software that allows you to enter parameters specific to your process application i e your O range recorder output range alarm settings etc System Descrip
37. ORIZED BY FACTORY MUTUAL E6 18 OXYGEN SIGNAL El 10VREF COMP OFFSET ADJ REF VOLTAGE ADJ P COMP ATTENUATE MULTIPLICATIV E COMP ADDITIVE COMP 1 20 97 Figure B 4 PCB Analyzer Assembly Dwg 2703 1036 B 4 HY 70 4 REQ D b J AMETRICS Oxy An A a PANAMETRICS VENEN OD TYP 4 PLCS O SS AMETRICS Xy Anal a PANAMETRICS TH na U AU ANG UAL IS DRAWING IS FACTORY MUT CONTROLLED NO ES PERMITTED UNLESS HORI ED BY FACTORY MUTUAL APPROX AS SHOWN 1 20 97 Figure B 5 TMO2D Top Assembly Dwg 705 517 sh 2 A4 TEMP PROBE ASSEMBLY 705 455 eos wore Ji 703 930 ES P3 BRIDGE VOL
38. October 2002 j M M HM B M 1 Process Control Instrument Division TMO2D Oxygen Analyzer User s Manual 910 084A5 IATTENTION This manual contains instructions for TMO2D units that use software versions STD 004 E or earlier Units manufactured in Waltham with serial numbers below 1350 or in Shannon with serial numbers below 300E were supplied with these software versions r PANAMETRICS 4 8 92 TMO2D Oxygen Analyzer lii 1 20 97 Warranty Maintenance Policy The TMO2D Oxygen Analyzer is warranted by PANAMETRICS to be free from defects in material and workmanship Liability under this warranty is limited to servicing calibrating and replacing any defective parts of the instrument Fuses and batteries are specifically excluded from any liability This warranty is effective from the date of delivery to the original purchaser For the warranty to be valid the equipment must be determined by Panametrics to have been defective This warranty is effective with respect to the following one year for electronic failures one year for mechanical failures shorts or opens to the transmitter If damage is determined to have been caused by misuse or abnormal conditions of operation the owner will be notified and repairs will be billed at standard rates after approval If any problems develop take the following step
39. PTF sample inlet amp outlet Area Classification Explosion proof weatherproof INIEX Cert 90C 103 882 to CENELEC EEx d IIC T6 Accuracy 0 1 of span electronics only Ambient Temperature Effect 0 01 of full scale per C Analog Output Standard Single isolated 0 4 20 mA 500 ohm maximum field programmable Optional Dual isolated 0 4 20 mA 500 ohm maximum field programmable Alarms Standard Two Form C SPDT 2 A max 117 VAC field programmable 0 01 set point resolution 0 05 of span dead band Optional Two hermetically sealed relays for Class I Div 2 hazardous areas 2 standard or hermetically sealed relays for automatic calibration Auto Cal Digital Output RS 232C serial port Display 2 line x 24 character backlit LCD Analog Input 4 to 20 mA from TMO2 TC Transmitter Power 100 120 220 240 VAC 10 50 60 Hz 35 watts max provides 24 VDC 1 A max to TMO2 TC Transmitter Fuses 110 120 VAC 0 5 A Slo Blo 220 240 VAC 0 25 A Slo Blo Temperature Operating 0 to 50 C 32 to 122 F Storage 20 to 70 C 4 to 158 F 5 2 Specifications 1 20 97 Physical Dimensions Rack Mount 5 25 H x 19 W x 9 25 D 133 x 483 x 235 mm Bench Mount 5 25 H x 9 W x 9 25 D 133 x 229 x 235 mm Panel Mount 5 25 H x 9 W x 9 25 D 133 x 229 x 235 mm Weatherproof fiberglass 11 25 H x 9 38 W x 4 38 D 286 x 238 x 111 mm Weatherproof
40. S comp done the GAS input O2 DVM TEST Press YES to confirm the entry X XX mA The X s represent the milliamp signal which will update continuously until any key is pressed Programming the TMO2D 4 13 9 30 93 DVM Test cont Recorder Calibrate Test The compensation signal input is displayed the same way Select DVM Input GAS comp done TESTS DVM Test TESTS Recorder Calibrate Use the NO or arrow keys to select DONE Press YES to exit Press NO to proceed to the next submenu The Recorder Calibrate Test allows a O to be output to the recorder to facilitate adjustment of the recording device s zero and span For example to enter a O of 36 39 to be output to recorder A TESTS Recorder Calibrate A B done Select Recorder to test Set Recorder A to 960 XX XX Set Recorder A to 960 36 39 A B DONE Select Recorder to test TESTS Recorder Calibrate TESTS Alarms Test Press YES to select the Recorder Calibrate Test Use the NO or arrow keys to select A Press YES to confirm your selection The X s represent the previously entered setpoint Use the numeric keys to enter 3 6 3 and 9 then press YES twice to confirm the entry and proceed to the next prompt Use the NO or arrow keys to select DONE Press YES to exit Press NO to procee
41. TAGE SELECT Lo A2 VO FCB 703 930 H w he MN VREF gg INV N ug Se cs cA CLS La RT osc Telefe lee fot ICT SPP imssza RIN 2 5 p 4700PF 100V el MJEIZ1 dy VR4 Ny 1N5250A 20v Q2 MJET81 RAL 4700 U4 VRI LM3858 2 5 2 5V N VR3 LM385BZ 1 2 1 2V LM3858Z 1 2 12V R21 40 2K0 196 R23 1000 0 1 1O PCB 703 930 A2 VO PCB 703 930 BLK A OXYGEN SENSOR ASSEMBLY 705 447 w2 o 7e US TLE2064C RAS 7 100K y 10M R24 1000 0 1 REF 40 2k0 1 10 SAI RI 40 2K 0 196 R46 2 00K O2 GAIN R38 200K0 196 R39 200K0 196 U4 ARIN RAO R22 40 2K0 196 PRESS Var x COMP SWITCH US TLE2064C PRESSURE SENSOR 4 RA6 220K E Varn PRESSURE COMPENSATION RIT 40 2K 0 1 1 TLE2064C US
42. TMO2 Display and Keypad While displaying oxygen concentration the Display will ignore all keys except the NO key On receipt of a NO the LCD will begin displaying Enter Code and await the program entry code 1 2 3 During code entry the Display continues to update the data display alarm status and recorder ouput 4 2 Programming the TMO2D 9 30 93 Menu Navigation After the passcode 1 23 is entered the LCD switches to Menu Mode which allows programming of the Display i e setting parameters and calibration data performing alarm and recorder tests While in Menu Mode data collection is suspended and alarm status and recorder outputs hold their current values In Menu Mode the first line of the LCD shows the title of the current menu in capital letters The second line displays the current menu options The two types of response keys available in the Menu Mode are YES NO and Selector keys Pressing YES selects the displayed option Pressing NO skips that option and displays the next option in the list The Selector option is used to choose between two or three possible choices in the menu Note The menu lists are circular i e skipping over the last option in the list returns to the first option in the list The following sections describe the programming procedure and menu navigation in detail one menu at a time Also refer to Appendix E for flow diagrams of each menu Programming the TMO2D 4 3
43. TO CALIBRATION MENU Set Span Gas Enter the equilibration time in this case a 3 for 3 minutes then press YES to confirm the entry Press YES to exit Press NO to proceed to the next option 4 22 Programming the TMO2D Set Span Gas 9 30 93 The Set Span Gas option is used to enter the oxygen concentration and the equilibration time for the span gas The default concentration for the Span Gas is 20 93 the concentration of oxygen in air The procedure to set the span gas is identical to setting the zero gas For example set the span gas for a gas that is 50 O and 50 N for an equilibration time of 4 minutes AUTO CALIBRATION MENU Set Span Gas Span Gas 960 XX X Span Gas Ov 50 00 Span Gas ON for MM SS X XX Span Gas ON for MM SS 4 00 AUTO CALIBRATION MENU Set Span Gas AUTO CALIBRATION MENU Set Settling Time Press YES to enter the Set Span Gas option The X s represent the previously entered value Use numeric keys to enter a 5 and 0 for 5096 oxygen then press YES twice to confirm the entry and proceed to the next prompt The X s represent the previously entered equilibration time Enter the equilibration time in this case a 4 for 4 minutes then press YES to confirm the entry Press YES to exit Press NO to proceed to the next option Programming the TMO2D 4 23 9 30 93 nn zz
44. Therefore by carefully combining a magnetic field gradient and a temperature gradient with the TMO2 transmitter measuring cell an oxygen containing gas can be made to flow creating magnetic wind The intensity of this induced flow depends on the concentrations of oxygen in the sample The TMO2 Transmitter In the TMO2 transmitter measuring cell permanent magnets are used Oxygen Measurement to create a magnetic field The temperature within the measuring cell Bridge is controlled to a constant 113 45 C to maintain thermal equilibrium The cell also contains two pairs of thermistors with one thermistor of each pair located within the magnetic field Since the thermistors are electrically heated a temperature gradient is created within the magnetic field When a small sample of oxygen containing gas diffuses into the measuring cell it is attracted to the field causing the sample pressure to become locally higher The sample pressure is slightly lower in the vicinity of the thermistors because the gas is less strongly attracted at higher termperatures This difference in sample pressure causes the gas to flow out from the center of the magnetic field and over the thermistors The inner wind generating thermistors decrease in temperature as they lose heat to the magnetic wind while the outer wind receiving thermistors increase in temperature as they pick up heat from the magnetic wind The thermistor pairs are additively connected in an
45. al Port and Corresponding Keys 6 9 036 039 6 o IN 01B 008 ESCAPE BACKSPACE RS 232C Serial Port C 1 Appendix D The TMO 2 Transmitter Calibration Procedure Introduction ur 0 x one o awe ws ee CR e D 1 Equipment and Drawings eese D 2 Setup eo cy ra D 3 Preliminary Explanation eee D 3 Internal Background Gas Compensation D 5 External Compensati0N o oooocoooococnnn o D 6 Internal Pressure Compensation oooooooooommmo D 7 External Pressure Signal Calibration D 8 Internal Hydrogen Gas Compensati0M D 9 External Hydrogen Gas Compensation D 9 Internal Methane Gas Compensation Reverse Compensation D 9 1 20 97 ees Introduction Factory calibration procedures are included here for reference only Do not attempt factory calibration in the field For field calibration refer to the TMO2D calibration sheet that was shipped with your unit IWARNING TO ENSURE THE SAFE OPERATION OF THE TMO2D YOU MUST INSTALL AND OPERATE IT AS DESCRIBED IN THIS MANUAL IN ADDITION BE SURE TO FOLLOW ALL APPLICABLE SAFETY CODES AND REGULATIONS FOR INSTALLING ELECTRICAL EQUIPMENT IN YOUR AREA The TMO2 Transmitter Calibration Procedure D 1 11 30 92 Equipment and Drawings Test Equipment e Two 3 1 2 digit DVMs HP 34384 or equivalent Pressure generator 0 1 PSI
46. an accommodate three types of pressure compensation Additive Multiplicative and External Additive and multiplicative pressure compensation are performed on the analyzer circuit board whereas external pressure compensation produces an auxiliary 4 20 TB1 4 milliamp signal for use in TMO2D compensation calculations This section describes additive and multiplying pressure compensation Additive pressure compensation is used to make corrections on 10046 oxygen analyzers Multiplying should be used on the lower oxygen ranges Calibration of the auxiliary external pressure signal is covered in the next section Additive Comp Set SA1 switches as follows 1 4 amp 5 OFF up 2 3 amp 6 ON Multiplying Comp Set SA1 switches as follows 2 3 amp 6 OFF 1 4 amp 5z ON Set J3 to the LO position HI if using hydrogen helium The TMO2 Transmitter Calibration Procedure D 7 11 30 92 Zero Span Compensation External Pressure Signal Calibration Zero Oxygen Span Refer to page D 5 Zero and follow steps 1 through 12 setting E14 to 2 94 volts additive comp or 0 80 volts multiplying comp in step 3 With single background gas systems use only steps 1 through 6 1 Apply span gas ex 100 oxygen into the oxygen transmitter 2 Adjust S1 and R41 until the output TB1 3 is 20 milliamps 3 Apply span pressure to the pressure sensor ex 1 0 PSI 4 Adjust R28 until TB1 3 is 20 milliamps 5 Apply min
47. ay seeleeeeeeee e 2 4 Making Wiring Connections 0 cece cece E NE E E EENE EEO eens 2 5 Wiring the Transmitter to the Display lseeeeeee IA 2 6 Wiring Recorders AutoCal System Alarms etc 0 000 cece cece eee 2 8 RS 232C Senal Port nen 22 2 ie Oa ea Ri AR eee OE de 2 9 Connecting Power Lu eive eb idv xg eb rk aa Ei 2 10 Chapter 3 Operation IntroductiOm ns en RE ROCA ab tte es s 3 1 Powering Up the TMO2D 5r ue Red eie ee ne ER RUE A ud e 3 1 Establishing Gas Flow 20 5 e he ed he eee ehe s herein d 3 1 Operating the Display en rey b ener ber Hr ere bx ah ag 3 1 The LCD Display 2 22 22 u A a EVEN ah Metus 3 2 The Keypad cu 2 nn AA EUM ey ae E s 3 2 1 20 97 es Table of Contents cont Chapter 4 Programming the TMO2D Introductions a p eto tote Passed Tey Td nee aad ok sdb oed rA Vere Er 4 1 Entering Data into the User Program 4 1 Key Punctions 2 22 04 2 EE a BR a RI Bee Res DE a 4 2 Programming the TMO2D via the Display lseeeeeee I 4 2 Display Navigation an ae ERE EDI RE AUS E 4 2 Menu Navigation zi uon dun sn era pao dual us UO gan VER sa A beard au Sgt 4 3 The Setup Menu nen 20 5 eh REESE eese Pene ee Go ES 4 4 NIME 4 4 NIS DI ANN 4 5 Set Backlight nn eee ache ae A ELE eL EOS EU te ess 4 5 setDbisplay 2 5 oy ser er sen Ela Bis Sela ees See bre Pe SE E IER SE 4 6 Set Communications near Rn ERR ION SALON GRR ON TR TL in 4 6 The Rec
48. based background gas compensation is performed Thermoparamagnetic Magnetic Wind Theory Appendix B Outline and Installation Drawings TMO2 Outline and Installation Dwg 4712 211 B 1 TMO2D Rack Model Outline and Installation Dwg 4421 255 B 2 TMO2 Oxygen Analyzer Assembly Dwg 705 468 B 3 PCB Analyzer Assembly Dwg 2703 1036 B 4 TMO2D Top Assembly Dwg 705 517 sh 2 B 5 Analyzer PCB Schematic Dwg 4700 1036 B 6 Oxygen Analyzer Display PCB Assembly Dwg 703 1226 B 7 Oxygen Analyzer Display PCB Schematic Dwg 700 1226 sh 1 B 8 Oxygen Analyzer Display PCB Schematic Dwg 700 1226 sh 2 B 9 Oxygen Analyzer Display PCB Schematic Dwg 700 1226 sh 3 B 10 1 20 97 6 10 DIA 155 5 71 DIA 145 ER LUCK SCREW gt THREAD GROUND LUG 2 mm gt 3 4 NPT TYP Y 10 47 266 EXP E NOTE ciel Samal 39 DIA TYP 210 203 242 WPF SHOWN WITH OPTIONAL FLAME ARRESTOR FOR EXPLLISIUN PR LHF UNIT O SIGNAL CONNEC e 16 eoo 3 27 DIA u E
49. ct Recorder to set A B done Use the NO or arrow keys to move the brackets to the desired recorder YES Rcd A Output mA 0 20 4 20 Use the NO or arrow keys to move the brackets to the desired recorder output range YES Recorder A 4 mA Value 02 XX XX Enter the numeric value to represent the low end value for the recorder output range YES YES Recorder A 20 mA Value 3602 X XX Enter the numeric value to represent the high end value for the recorder output range YES Select Recorder to set A B DONE Note Repeat the same procedure for recorder B After setting the recorders use the NO or arrow keys to move the brackets ta DONE to exit YES Figure E 1 Setup and Recorder Menu Flow Diagrams Main Menu Flow Diagrams MAIN MENU Alarms YES or NO Select Alarm to set HIGH low done Use the NO or arrow keys to move the brackets to the Desired alanm to set YES Enter Alarm Setpoint High 02 XX X Enter the numeric value for the alarm value YES YES Select Alarm to set high low DONE Note Repeat the same procedure for the low alarm After setting the alarms use the NO or arrow keys to move the brackets to DONE to exit YES TESTS MAIN MENU Tests YES or INO DVM Tests vESI or NO Select DVM Input GAS comp done Use the NO or arrow ke
50. d to the next submenu 4 14 Programming the TMO2D 9 30 93 Alarms Test The Alarms Test trips and resets the alarm relays via the keypad in order to test the operation of external alarm devices For example to test the high alarm relay TESTS Alarms Test Select Alarm to test HIGH low done Turn High Alarm ON off done Turn High Alarm on OFF done Press YES to select the Alarms Test Use the NO or arrow keys to select HIGH Press YES to confirm your selection then use the NO or arrow keys to select ON Press YES to confirm the ON selection Note Upon selecting the ON option and pressing YES the High Alarm relay will turn on and the selection brackets will skip to OFF Turn High Alarm ON off done Press YES to turn the high alarm off Note Upon selecting the OFF option and pressing YES the High Alarm relay will turn off and the selection brackets will skip to ON Turn High Alarm on off DONE Select Alarm to test high low DONE Use the NO or arrow keys to select DONE Press YES to confirm the selection then use the NO or arrow keys to select DONE again Programming the TMO2D 4 15 9 30 93 Alarms Test cont TESTS Alarms Test TESTS Done MAIN MENU Done This completes the Tests Menu Menu title Press YES to confirm the selection and exit Pres
51. d zero gas measurements the Display reconnects the Transmitter to the process stream Then a final settling time takes place while the process gas replaces the zero gas Once settling has been completed if no error has occurred the Display calculates the amount of drift and applies the calculated drift to the factory oxygen calibration data The factory calibration data is not changed The corrections to the calibration data are stored as the Drift Curve If the Auto Cal measurements were out of range the data is disregarded and the Display uses the drift curve already stored in memory If Auto Cal Error Handling is enabled the screen alarms and recorders will respond as programmed Programming the TMO2D 4 17 9 30 93 Auto Cal Parameters Note Auto Calibration requires the Auto Cal option to be ordered cont with the TMO2D Auto Cal also requires a sampling system with solenoid valves and two distinct calibration gases or one calibration gas and air Using the Auto Cal procedure without these options will produce an erroneous calibration and prevent proper operation of the TMO2D Follow the steps below to enter the Auto Cal Parameters submenu MAIN MENU At this prompt press NO until the Setup Calibration Menu displays MAIN MENU Press YES to enter the Calibration Calibration Menu CALIBRATION MENU Press YES to select the Auto Cal Auto Cal Parameters Parameters submenu AUTO CALIBRATION
52. der our sole responsibility that the TMO2D Display and Control Module to which this declaration relates is in conformity with the following standards EN 61326 1998 Class A Annex A Continuous Unmonitored Operation EN 61010 1 1993 A2 1995 Overvoltage Category II Pollution Degree 2 following the provisions of 89 336 EEC EMC Directive and 73 23 EEC Low Voltage Directive Shannon August 1 2001 p um T Ap A Mr James Gibson GENERAL MANAGER 1 LS ENISO9002 T V ESSEN Shannon ISO 9001 U S CERT DOC Rev F 7 19 01 DECLARATION DE CONFORMITE AS a PANAMETRICS Panametrics Limited Shannon Industrial Estate Shannon Co Clare Ireland d clarons sous notre propre responsabilit que le TMO2D Display and Control Module r latif cette d claration est en conformit avec les documents suivants EN 61326 1998 Class A Annex A Continuous Unmonitored Operation EN 61010 1 1993 A2 1995 Overvoltage Category II Pollution Degree 2 suivant les r gles de la Directive de Compatibilit Electromagn tique 89 336 EEC et de la Directive Basse Tension 73 23 EEC Shannon August 1 2001 p um T Ap A Mr James Gibson GENERAL MANAGER 1 LS ENISO9002 T V ESSEN Shannon ISO 9001 U S CERT DOC Rev F 7 19 01 KONFORMITATSERKLARUNG re u PANAMETRICS Panametrics Limited Shannon Industrial Estate Shannon Co Clare Ireland erkl ren in alleiniger Verantwortung da das Produkt TMO2D Display and Contr
53. e de 4 9 The Alarm Men isa uer 4 aaa ee eee 4 11 The Tests Menu 2 2 eG baie a EORR OR Rn 4 13 The Calibration Menu eee nun nun nun nn 4 17 9 30 93 ese Introduction The TMO2D contains an interactive user friendly program that allows the user to change operating parameters as desired The user program has six main menus Use the front panel keypad and display to check or change the current operating parameter settings Data is stored in memory and is retained for several years even if the main power is lost New data overrides any previously entered data The user program consists of the following six main menus e Setup Recorders Alarms Test Calibration Resume Entering Data into the To enter data into the user program or to check previously entered User Program values enter the Menu Mode When power is turned on the TMO2D is in the Operate Mode To enter the Menu Mode 1 Press the NO Key 2 Key in the code 1 23 The TMO2D will display an asterisk after each digit is entered The LCD screen now displays the Setup Menu the first of the six main menus At this display press YES to enter this menu or NO to go on Pressing NO continuously scrolls through all six main menus Note The first five options are main menus the RESUME prompt is used to exit the Menu Mode and return to Operate Mode If an incorrect code is entered or a non numeric YES NO or arrow key is pres
54. e double throw relays for use in activating alarm devices Both relays provide a normally open and a normally closed set of contacts The alarms are addressed as HIGH and LOW The HIGH alarm relay will trip when the current reading is greater than or equal to the high alarm setpoint The LOW alarm relay will trip when the current reading is less than or equal to the low alarm setpoint Use the steps in the following example to set the high alarm setpoint to 75 O and the low alarm setpoint to 25 O Refer to Appendix E for a flow diagram of the Alarm Menu MAIN MENU Setup MAIN MENU Alarms Select Alarm to set HIGH low done Enter Alarm Setpoint High 0 X XX Enter Alarm Setpoint High 0 75 0 Select Alarm to set high LOW done Enter Alarm Setpoint Low 950 X XX At this prompt press NO until the Alarm Menu displays Press YES to enter the Alarm Menu Use the NO or arrow keys to select HIGH Press YES to enter the high alarm setpoint The X s represent the previously entered setpoint Use the numeric keys to enter 7 and 5 then press YES twice to confirm the entry and proceed to the next prompt Use the NO or arrow keys to select LOW Press YES to enter the low alarm setpoint The X s represent the previously entered setpoint Programming the TMO2D 4 11 9 30 93 ESQ CEE SxCIK
55. ermoconductivity TB1 pin 4 These signals may then be used in a TMO2D algorithm to provide an improved oxygen calculation Note In the following procedure the gas mixture having the largest oxygen sensitivity should be used as Gas A Set SA1 switches as follows 1 3 5 amp 6 OFF up position 2 amp 4 ON Set J3 to the LO position HI if using hydrogen helium Zero Refer to page D 5 Zero and follow steps 1 through 12 but set E14 to 0 30 volts in step 3 and disregard step 4 Oxygen Span 1 Apply span background Gas A ex 10 oxygen remainder CO into the oxygen transmitter 2 Adjust S1 and R41 until the TB1 3 is about 20 milliamps D 6 The TMO2 Transmitter Calibration Procedure 11 30 92 External Compensation Internal Pressure Compensation 1 Apply background Gas B ex 100 Hydrogen into the oxygen transmitter 2 Adjust R28 until TB1 4 is about 20 milliamps 3 Apply background Gas A ex 100 CO into the oxygen transmitter 4 Adjust R27 until the TB1 4 is about 4 milliamps 5 Repeat steps 1 through 4 until the desired outputs are acquired These values are arbitrary and may vary somewhat since they will be entered into TMO2D or System 1 calibration tables The levels should extend from 1 to 35 milliamps taking all expected gas mixtures into account Note R27 and R26 are interdependent adjustment of one pot will affect the setpoint of the other pot The analyzer circuit board c
56. ero calibration gas YES YES AUTO CAL MENU Set Span Gas YES or NO Span Gas 9602 X XX Enter the percentage of oxygen in the span calibration gas YES YES Span Gas ON for MM SS X XX Enter the equilibration time for the span gas YES YES AUTO CAL MENU Set Settling Time YES or NO Settling Time MM SS X XX Enter equilibration time for process gas YES YES AUTO CAL MENU Set Error Handling ne 3 y YES or NO YES YES AUTOCAL Error Effects Done ES Enable Error Handling YES no Use the NO or arrow keys to move the brackets to the desired selection Select YES or select NO AUTOCAL Error Effects Set Display Response YES or NO Display AutoCal Error YES no Use the NO or arrow keys to move the brackets to the desired selection Select YES or select NO AUTOCAL Error Effects Set Alarm Response or NO Select Alarm Effect No Effect Trip High Trip Low Trip Both Use the NO or arrow keys to scroll to the desired selection AUTOCAL Error Effects Set Recorder Response YES or NO Select Recorder Effect No Effect Force High Force Low Hold Last Value Use the NO or arrow keys to scroll to the desired selection YES returns to the AUTO CAL MENU Set Error Handling
57. et old a eal eus C 1 vi 1 20 97 D ADAAXALA EAELLLLLLLELCELpECCCC ku IEEDDD2 EALEABdOoOOO SSA ADA1A1b3AI AX3 21oQ j3z ECLLD Table of Contents cont Appendix D The TMO2 Transmitter Calibration Procedure Introduction op ra a lei D 1 Equipment and Drawings o oo oocoococo I el D 2 SEP ne cee et eae Br eel eT BE AA obige D 3 Preliminary Explanation 0 0 0 cece cee eee ene e eens D 3 Internal Background Gas Compensation D 5 EEUU D 5 Span Compensation es l6sonerserenersehweergBenenuer a eqergecew x gh ene erent D 6 External C omp nsatiOn independent gota Asean br gta queda D 6 LO E EE D 6 Oxy Gen Span oa eI MER St QE M IA EC HER EE eb ea D 6 External Compensation 0 0 0 cece eee s D 7 Internal Pressure Compensation 20 0 c eee teen nent tenes D 7 Zelt enero Moe Acie ets tanh dda ceria Wet bea ca AS Se ee ee das D 8 Span Compensation sac cc wes a cee awa a RAE Ra RURALES SNC See RR Ga RN ues D 8 External Pressure Signal Calibration 20 0 0 eee eee eee nee D 8 Zoo KM dd dal Dl a o D 8 Oxygen SPAN 2m ae a la ad D 8 External Pressure Signal 5 tia a ee Ebene uera D 9 Internal Hydrogen Gas Compensation sseleeeeeeeee ett eens D 9 yo RM D x D 9 spat Compensation es ce Devi aa Ale eka nei epu RIEN ie D 9 External Hydrogen Gas Compensation D 9 Internal Methane Gas Compensation Reverse Compensation ote re tiaa oa eoe dq Pent handed quu lb qu eid D 9 LT
58. ex 1 20 97 Index cont R Recorder Calibration Test o o 4 14 Recorder Menu 000000e 4 0 Recorders Wigs eee hass 2 8 RS 232C Wide are RE Hee ese 2 9 Sample System Mounting u ca pi retra il ies 2 3 Set Communications 0 4 8 Settling Tim eus eesi e a ea 4 24 Setup Menu Backligbt o ooooocococooomo o 4 5 Communications lees 4 6 Dates ccm a MOERS RARUS 4 5 Display iious nien transan sey hese oe 4 6 TIME DEM 4 4 Span Gas Option 000 4 23 Specifications llle 5 1 Display oies so beard eed bees eee bees oe 5 2 Functional ser mes 5 2 T Test Menu Recorder Calibrate 4 14 Tests Menu 2222222220 4 13 Alarms Test 12e nee 4 15 DVM Test ex 3 2 4 3 is a a e 4 13 Time SUD evita a teri ett 4 4 Time Interval for Auto Cal oooooo o 4 19 Transmitter Mounting the 20000 2 2 WIS 2 0 ee wet ned sse Bat n 2 6 U Update Rate n ovn 4 8 in Setup Menu 0 4 8 W Wiring Alarms cuore hoe 2 8 AutoCal System 2 222202020 2 8 Power IE Nee 2 10 R cordets ner ees 2 8 RS 232 nn iw mv usus 2 9 Transmitter to Display 2 6 Z Zero Gas Option 000 4 21 Index DECLARATION OF CONFORMITY AS a PANAMETRICS Panametrics Limited Shannon Industrial Estate Shannon Co Clare Ireland declare un
59. ground of Gas A followed by full scale oxygen in Gas B the secondary background gas The type of gases used for Gas A and Gas B will vary depending on the application As an example the calibration for a customer who uses the TMO2 to measure oxygen in a hydrogen tank would use hydrogen as Gas A Nitrogen might be used for Gas B since a system leak would inject air nitrogen into the mixture In addition circuits can be added to the analyzer board to provide pressure compensation reverse compensation and compensation attenuation These options are switch selectable at calibration to provide proper compensation for hydrogen helium methane and pressure sensitive applications Swich SA1 on the circuit board is used to select the following six operating functions SAI I Selects negative offset voltage e SA1 2 Selects positive offset voltage e SAI 3 Selects inverse compensation e SAI 4 Selects compensation attenuation e SA1 5 Selects multiplying compensation e SA1 6 Selects additive compensation The TMO2 Transmitter Calibration Procedure D 3 11 30 92 Preliminary Explanation cont Zero adjustments are made via R20 R29 and R44 When properly adjusted zero shift errors due to background gas changes can be reduced Oxygen signal gain is controlled by rotary switch S1 and R41 S1 serves as a course gain conrol while R41 provides fine gain adjustment Rotating either control in the clockwise direction
60. ing YES at Perform AutoCal or when the specified time arrives the following occurs The top line of the display changes to Auto Cal in Progress At this point the Process Cal relay will trip to the Cal position and the Zero Span relay will trip to the Span position The second line of the display reads Span Gas is ON along with the equilibration time The equilibration time will begin counting down to zero The Auto Cal can be interrupted at this point by pressing NO The Display will prompt Abort AutoCal Two things can be done Press NO to resume the countdown Press YES to return the Display to the state it was in previous to the Auto Cal When the equilibration time reaches zero the Display will read Measuring and count down the measurement cycles At the end of the measurement the Zero Span relay will reset to the Zero position and the second line of the display will change to Zero Gas is ON The equilibration countdown and the measurement countdown are performed as for the span gas At the end of the Auto Cal procedure the Process Cal relay will reset to the Process position and the settling time will count down to zero The Display then reads Working as it performs the curve fitting functions Note To set Auto Cal to begin automatically refer to page 4 17 Programming the TMO2D 4 27 9 30 93 Perform AutoCal cont To manually activate the Auto Cal procedure
61. ins turned cn After completing this option use the NO er arrow keys to move the brackets to DONE to exit YES Select Alarm to test high low DONE Note Repeat the same procedure for the low alarm YES TESTS Done YES or NO YES returns to the MAIN MENU Tests prompt NO returns to TESTS DVM Tests prompt Figure E 2 Alarm and Test Menu Flow Diagrams Main Menu Flow Diagrams Goes to MAIN MENU Resume Calibration MAIN MENU YES or NO Goes to MAIN MENU Edit Calibration Data ALIBRATION MENU uto Cal Parameters YES or NO Enable AutoCal YES no Use the NO or arrow keys to move the brackets to the desired selection Select NO or select YES AUTO CAL MENU Set Time Interval YES or NO Select AutoCal interval hours DAYS Use the NO or arrow keys to move the brackets to the desired selection YES AutoCal Interval Days X Enter the time interval YES YES AutoCal at Time HH MM XX XX Note This prompt will not appear if interval is being entered in hours Enter the time AutoCal is to begin in 24 hour format YES YES AUTO CAL MENU Set Zero Gas YES or NO Zero Gas 9502 X XX Enter the percentage of oxygen in the zero calibration gas YES YES Zero Gas ON for MM SS X XX Enter the percentage of oxygen in the z
62. ion and displays it as a percentage of the sample see Figure 1 3 below Note The TMO2D milliamp output range can be independent of the Display range Transmitter Display Figure 1 3 TMO2 Transmitter and TMO2D Display 1 4 Features and Capabilities Chapter 2 Installation Introduction mnn 2 1 Mounting the TMO2 Transmitter lt lt lt 2 2 Mounting the TMO2 Sample System 2 3 Mounting the TMO2D Electronic Display 2 4 Making Wiring Connections 1 20 97 ees Introduction This section provides information on how to mount and wire the Transmitter and Display Read this section if you are installing the TMO2D for the first time A typical TMO2D installation consists of the following Mounting the TMO2 Transmitter Mounting the TMO2 Sample System Mounting the TMO2D Electronic Display Wiring the Transmitter to the Display Wiring Recorders AutoCal System Alarms etc Connecting Power Caution The interconnecting wiring between the transmitter and display must be completed before powering up IWARNING TO ENSURE THE SAFE OPERATION OF THE TMO2D YOU MUST INSTALL AND OPERATE IT AS DESCRIBED IN THIS MANUAL IN ADDITION BE SURE TO FOLLOW ALL APPLICABLE SAFETY CODES AND REGULATIONS FOR INSTALLING ELECTRICAL EQUIPMENT IN YOUR AREA ALL INSTALLATION PROCEDURES SHOULD BE PERFORMED BY TRAINED SERVICE PERSONNEL Installation
63. l the desired outputs are obtained These values are arbitrary and may vary somewhat since they will be entered into TMO2D or System 1 calibration tables The levels should extend from 1 to 35 milliamps taking all expected gas mixtures into account Note R27 and R26 are interdependent adjustment of one pot will affect the setpoint of the other pot Set SAT switches as follows 1 3 4 amp 6 OFF up 2 amp 5 ON Set J3 to the HI position HI if using hydrogen helium Refer to page D 5 Zero and follow steps 1 through 12 but set E14 to 0 92 volts in step 3 and set E10 to 4 0 volts in step 4 Refer to page D 6 Span Compensation and follow steps 1 through 7 Refer to page D 6 External Compensation but set J3 to the HI position Set SAI switches as follows 2 4 amp 6 OFF up 1 3 amp 5 ON Set J3 to the LO position HI if using hydrogen helium Refer to page D 5 Zero and follow steps 1 through 12 setting E14 to 3 15 volts in step 3 Refer to page D 6 Span Compensation and follow steps 1 through 7 The TMO2 Transmitter Calibration Procedure D 9 Appendix E Main Menu Flow Diagrams Setup and Recorder Menu Flow Diagrams Alarm and Test Menu Flow Diagrams Calibration Menu Flow Diagram 9 30 93 To enter PROGRAM press the NO key Key in 1 2 3 MAIN MENU Setup SETUP MENU Set Time YES or NO ye or NO E
64. lay features a two line by 24 character backlit liquid crystal display LCD and comes in either rack bench panel or weatherproof configurations A color coded cable is used to connect the Display to the Transmitter A Sample System is required to bring a gas sample flow to the Transmitter at a well defined pressure The Sample System design will depend on the application Generally it consists of the mounted Transmitter a sample gas inlet flow regulating valve flow meter and calibration gas inlet valves The Sample System may also include a pressure gauge regulator pump aspirator coalescer filter or other components Figure 1 2 below is an illustration of a Sample System which includes most of the components described above Sample Inlet E Y h 4 Figure 1 2 A TMO2 Sample System Features and Capabilities 1 3 11 30 92 System Operation The TMO2 Transmitter uses the thermoparamagnetic magnetic wind principle to measure oxygen concentration in gases see Appendix A for magnetic wind theory During operation the TMO2 Transmitter generates an electrical signal proportional to oxygen concentration and optionally one additional signal for either background gas or pressure compensation The TMO2D Hlectronics Display accepts the oxygen and compensation signals from the Transmitter and using transmitter calibration background gas compensation and pressure compensation data calculates oxygen concentrat
65. le line length Similarly connect calibration gas inlets sample outlets bypass outlet and filter coalescer drain as applicable Figure 2 2 A Mounted TMO2 Sample System Installation 2 3 1 20 97 Mounting the TMO2D Electronic Display The Electronic Display comes in four mounting configurations bench rack panel and weatherproof mount see Appendix B for mounting dimensions No special mounting requirements are needed for the Display If you have a bench mount simply put the Display in a convenient location connect the wires from the Transmitter and connect the power If you have a rack or panel mount insert the Display into the rack or panel connect the wires from the Transmitter and connect the power The weatherproof model can be similarly mounted Note The power cord is the main disconnect device IMPORTANT 7o comply with the European Low Voltage Directive you must install a switch or circuit breaker on the input power line For greatest safety locate the circuit breaker or switch near the unit that the line serves Installation 1 20 97 ees Making Wiring Normally a 3 or 4 wire color coded cable is supplied to connect the Connections Transmitter to the Display First make wiring connections from Terminal Block 1 TB1 pins 1 4 on the Transmitter PC Board to the OXYGEN CELL Terminal Block located on the rear panel of the Display IATTENTION EUROPEAN CUSTOMERS IN ORDER TO MEET CE MARK REQUIR
66. nections Installation 2 9 1 20 97 i Connecting Power Once you have mounted and connected the Transmitter and Display and if applicable wired recorders alarms etc connect the Display to one of the following power supplies Japan 100 VAC U S 110 120 VAC Europe 220 VAC Australia 240 VAC Caution The interconnecting wiring between the Transmitter and Display must be completed before powering up Note The power cord is the main disconnect device IMPORTANT To comply with the European Low Voltage Directive you must install a switch or circuit breaker on the input power line For greatest safety locate the circuit breaker or switch near the unit that the line serves D Protective Conductor Terminal 2 10 Installation Chapter 3 Operation Introduction esos ko nu xn hee a nO e ees 3 1 Powering Up the TMO2D serere 3 1 Establishing Gas FloW 42 v vee a YR Ea 3 1 Operating the Display oooooooocorrnnnnonnmmo 3 1 The LCD Display i xx aan aaa RR E 3 2 The Keypad c i sn ey xh hne ne 3 2 11 30 92 Introduction Powering Up the TMO2D Establishing Gas Flow Operating the Display Once the TMO2D has been installed the analyzer must be powered up and a gas flow must be established before operating the analyzer via the Electronics Display This section covers the following Powering Up The Transmitter Establishing Gas Flow Operating The
67. nter 24 hour time HH MM X XX Enter the time in 24 hour format YES YES SETUP MENU Set Date YES or NO Enter Date MM DD YY XX XX XX Enter the date in MM DD YY ormat YES YES SETUP MENU Set Backlight YES or NO SETUP MENU Remain ON min X Enter how many minutes the backlight is to stay on before automatically turning off YES YES SETUP MENU Set Display real or NO Display Compensation YES no Use the NO or arrow keys to move the brackets to the desired choice Note If compensation is not enabled no option is given YES SETUP MENU Set Communications YES or NO et Baud Rate ET COMMUNICATIONS um q PESI or NO 9600 baud 4800 baud 2400 baud 1200 baud 300 baud Use the NO or arrow keys to scroll to the desired selection YES SET COMMUNICATIONS a Set Update Rate YES or NO SET COMMUNICATIONS Data Interval X Enter the desired rate at which data is sent to the serial port YES YES SET COMMUNICATIONS I Done YES or NO SETUP MENU Done YES or NO YES returns to the MAIN MENU Setup prompt NO returns to the SETUP MENU Set Time prompt MAIN MENU Goes to Ri 2 ecorders MAIN MENU YES or NO Alarms Sele
68. nter Point YES no Use the NO or arrow keys to move the brackets to the desired selection YES proceeds to the following prompt NO causes the calibration to skip this point and go to the next point X XX 02 Curve 1 O2 mA X XX Enter the milliamp signal for the point YES YES X XX 02 Curve 1 Comp mA X XX Note This prompt displays for background compensation only Enter the milliamp signal for the point Note The last two prompts appear for each point YES YES CALIBRATE BKGD Done YES or NO YES returns to the CALIBRATION MENU Edit Calibration Data prompt NO returns to the CAL PRESSURE Transducer Data prompt for pressure and the CALIBRATE BKGD Enter Grid Data prompt for all other compensation CALIBRATION MENU Done YES or NO YES returns to MAIN MENU Calibration prompt NO returns to the CALIBRATION MENU Auto Cal prompt MAIN MENU Resume YES or NO Returns to MAIN MENU O2 Pres xX XX XX Setup x xx xxx x Alarm L Notes 1 X s represent the previously entered value 2 Pressing YES the first time after entering a value replaces the X value 3 Bracketed choices are CAPPED when chosen 4 Use NO to erase mistakes 9 30 93 iagram Menu Flow D Calibration Figure E 3 E 5 Appendix F Installation Instructions for CE Mark Compliance 7 1 96 Installation NOTICE In
69. ol Module folgende Normen erf llt EN 61326 1998 Class A Annex A Continuous Unmonitored Operation EN 61010 1 1993 A2 1995 Overvoltage Category II Pollution Degree 2 gem den Europ ischen Richtlinien Niederspannungsrichtlinie Nr 73 23 EWG und EMV Richtlinie Nr 89 336 EWG Shannon August 1 2001 p um T Ap A Mr James Gibson GENERAL MANAGER 1 LS ENISO9002 T V ESSEN Shannon ISO 9001 U S CERT DOC Rev F 7 19 01 PANAMETRICS WORLDWIDE OFFICES a PCI Division 221 Crescent Street Suite 1 Waltham MA 02453 3497 USA Telephone 781 899 2746 Toll Free 800 833 9438 Fax 781 894 8582 ISO 9001 PANAMETRICS E mail pci panametrics com Web Site http www panametrics com CERTIFIED MAIN OFFICES USA Panametrics Inc 221 Crescent St Suite 1 Waltham MA 02453 3497 USA Telephone 781 899 2719 Toll Free 800 833 9438 Fax 781 894 8582 E mail pci panametrics com Web Site www panametrics com ISO 9001 Certified Ireland Panametrics Limited Shannon Industrial Estate Shannon Co Clare Ireland Telephone 353 61 470200 Fax 353 61 471359 E mail info panametrics ie ISO 9002 Certified 1 29 02 INTERNATIONAL OFFICES Australia Panametrics Pty Ltd P O Box Gymea N S W 2227 Australia Telephone 61 02 9525 4055 Fax 61 02 9526 2776 E mail panametrics panametrics com au Austria Panametrics Messtechnik GmbH Waldgasse 39 A 1100 Wien Austria Telephone 43 1 602 25
70. order Menu 2 2 La SE tak ee e o po AED e T edd 4 9 The Alar Menus pre ane ea dew wes A Yew ead Hom wera ren 4 11 The Iests Menu eme anm RC ER oye NS Whe ese x ara d xr rae edere drehte eps 4 13 DVM Test hen cese Trovit Tate Vu gii Va le gee PUDE he ad UE yh 4 13 Recorder Calibrate Test 4 2 2 2 3 san ee E TRAN AES DEEST bs 4 14 Alarms Test e nn Lem becca takes a Wai u eter be 4 15 The Calibration Men c 55 5 4 Ss 808 jc 5222 8 e PECES Ta sk un Ed OPE 4 17 Auto Gal Parametros Ea it ue QUA E E 4 17 Edit Calibration Data 22 2 2 022 02 vei eh a a RA S Res a ee 4 30 Chapter 5 Specifications TMO2 Oxygen Transmitter 2 ce cece nent n eee 5 1 TMO2D Display Specifications teen een eens 5 2 Performance seis oa ara ie as weal CPG PR ROGO X eu a Gees Ban 5 2 Eunct onal 4 epe sk ue T urea erue bevel Gia ge moe ee a 5 2 Physical iier e eret aaa ois ton anata ice gob m Rodas d outs Hog aging 5 3 Appendix A Thermoparamagnetic Magnetic Wind Theory Basic Principle uL sede LI nna edi epi eset d quei a age A 1 The TMO2 Transmitter Oxygen Measurement Bridge 0 0 ee eee eee ee A 1 The TMO2 Transmitter Compensation Bridge 0 0 cee ee eee ee eee A 2 TMO2D Electronics Microprocessor based Compensation for Variations in Background Gas Compositions 3 2 35 eed er Ek SY SE Den seht sr irn ema A 2 Appendix B Outline and Installation Drawings Appendix C RS 232C Serial Port RS 232C Serial Port ius c nee
71. peration 3 1 11 30 92 The LCD Display The TMO2D is operated via the Display keypad In order to facilitate operation it is important to familiarize yourself with the Display and keypad functions The first line of the LCD screen displays the current measurement source or menu title and a real time clock The second line of the LCD screen displays the measured data on the left and the current alarm condition on the right see Figure 3 1 below Note The LCD contains an electroluminescent EL panel to enhance readability of the screen during operation To activate the EL panel press any key except the NO key The Keypad For operational purposes the keypad Figure 3 1 below contains the digits 0 9 a minus sign a decimal point and four special operations keys e YES NO eft Arrow Right Arrow Note For details on how to use the keypad to program the Display refer to Chapter 4 fc d g ge Anc aly yzer PANAMETRICS HEBES 2 d Figure 3 1 TMO2D Display and Keypad 3 2 Operation Chapter 4 Programming the TMO2D Introduction oss vex nuo xn y ee slew n e a 4 1 Entering Data into the User Program 4 1 Programming the TMO2D via the Display 4 2 The Setup Men 42 sun e o cec nk e DO Rea ce E aa 4 4 The Recorder Menu uk o IEEE ER RESI AE o
72. r grid data Enter Grid Data O Grid Entry Use the numeric keys to enter the of Points 3 number of points in the gas calibration curve then press YES twice to confirm the entry and proceed to the next prompt Point 1 Use the numeric keys to enter the 02 XX XX percent gas concentrations for each of the points in the curve as prompted then press YES to confirm each entry Note Percent gas concentrations MUST be in ascending order XX XX 9605 Use the numeric keys to enter the Oz mA XX XX corresponding mA values for the given percent gas concentration The unit will then display the Working message before returning to the Calibrate Oxygen option Working Calibrate Oxygen At this prompt press NO to Enter Grid Data proceed to the next option Calibrate Oxygen Then press YES to exit to the Done Calibration Menu 4 38 Programming the TMO2D 9 30 93 No Compensation cont CALIBRATION MENU Press NO to return to the main Edit Calibration Data Calibration Menu CALIBRATION MENU Press YES to return to the Main Done Menu Programming the TMO2D 4 39 Chapter 5 Specifications TMO2 Oxygen Transmitter 000 eee ee TMO2D Display Specifications sL 1 20 97 TMO2 Oxygen Measuring Ranges Transmitter 0 196 O5 0 296 O 0 596 O 0 10 O 0 25 O 0 100 O Mea
73. r grid data Use the numeric Keys to enter the number of points in the gas calibration curve then press YES twice to confirm the entry and proceed to the next prompt 4 36 Programming the TMO2D 9 30 93 ese No Compensation cont Normal Response cont Point 1 Use the numeric keys to enter the 02 XX XX percent gas concentrations for each of the points in the curve as prompted then press YES to confirm each entry Note Percent gas concentrations MUST be in ascending order XX XX 02 Use the numeric keys to enter the Oz mA XX XX corresponding mA values for the given percent gas concentration The unit will then display the Working message before returning to the Calibrate Oxygen option Working Calibrate Oxygen At this prompt press NO to Enter Grid Data proceed to the next option Calibrate Oxygen Then press YES to exit to the Done Calibration Menu At the Edit Calibration Data prompt press NO to return to the main Calibration Menu CALIBRATION MENU Edit Calibration Data CALIBRATION MENU Press YES to return to the Main Done Menu Programming the TMO2D 4 37 9 30 93 No Compensation cont Damped Response Set System Response At this prompt use the NO or normal DAMPED arrow keys to select DAMPED then press YES to confirm the entry Calibrate Oxygen Press YES to ente
74. rn R20 fully counter clockwise 8 Adjust R44 until TB1 3 is 0 4 milliamps 9 Repeat steps 5 through 8 as required 10 Apply background Gas B ex 10096 N5 into the oxygen transmitter 11 Adjust R20 until the output is 0 4 milliamps 12 Reapply Gas A and fine tune R29 for 0 4 milliamps output The TMO2 Transmitter Calibration Procedure D 5 11 30 92 Span Compensation 1 Apply span background Gas A ex 10 oxygen remainder CO into the oxygen transmitter 2 Adjust S1 and R41 until the output is 20 milliamps 3 Apply span background Gas B ex 10 oxygen remainder N3 into the oxygen transmitter 4 Adjust R28 until the output is 20 milliamps 5 Reapply span background Gas A ex 10 oxygen remainder CO into the oxygen transmitter 6 Adjust R27 until the output is 20 milliamps 7 Repeat steps 3 through 6 until a consistent output is obtained Note R27 and R28 are interdependent adjustment of one pot will affect the setpoint of the other pot External A TMO2 transmitter with external compensation provides a separate Compensation 0 20 mA output to indicate the relative thermoconductivity of the gas mixture In normally compensated transmitters this signal is internally multiplied by the internal oxygen signal to give a single compensated oxygen 0 4 20 mA output In externally compensated transmitters however one output is an uncompensated oxygen signal TB1 pin 3 and the second output represents th
75. s Notify PANAMETRICS giving full details of the problem Be sure to include the model and serial number of your O2 Analyzer PANAMETRICS will then give you a RETURN AUTHORIZATION NUMBER if the analyzer has to be returned to the factory If PANAMETRICS instructs you to send your O2 Analyzer back to the factory please send it prepaid to the authorized repair station as indicated in the shipping instructions If damage has been caused by misuse abnormal conditions or if the warranty has expired an estimate will be made and provided upon request before repairs are started The PANAMETRICS TMO2D Oxygen Analyzer warranty is limited to that which is stated above PANAMETRICS will not be liable for anything beyond this Copyright Panametrics Inc 1997 1 20 97 Table of Contents Chapter 1 Features and Capabilities Introduction to the TMO2D 2 00 22 a a ner 1 1 System Description ero reet ruere eee en ba Etpe Ha rend 1 1 The TMO2 Transmitter llle mh 1 2 The TMO2D Electronics Display sseeeeeee III 1 3 The Sample System 2 ot neta ee Kev PbEMRU ERE ERG RE UE ei a 1 3 System Operation vx s ex o e RR s po us RR RUE EO AEN dee nue epi ces E AA E e 1 4 Chapter 2 Installation Introduction re eie wv vob oe Gee ea a x er e eR Sb he i 2 1 Mounting the TMO2 Transmitter l l eee eee eae 2 2 Mounting the TMO2 Sample System 0 0 cee ees 2 3 Mounting the TMO2D Electronic Displ
76. s NO to exit the submenu Press YES to exit the Tests Menu Press NO to proceed to the next Main 4 16 Programming the TMO2D 9 30 93 The Calibration Menu Auto Cal Parameters The fifth Main Menu is the Calibration Menu The Calibration Menu is used to enter measurement parameters and calibration data into the TMO2D The Calibration Menu contains the following three submenus Auto Cal Parameters Edit Calibration Data Done Refer to Appendix E for a flow diagram of the Calibration Menu The Display can be programmed to perform an automatic calibration procedure Auto Cal at specified time intervals ranging from minutes to months By performing measurements on two calibration gases zero and span the Display can correct for changes in the response of the Transmitter without operator intervention At the specified time interval the Display activates a solenoid controlled valve on an optional sampling system This isolates the Transmitter from the process stream and connects the Transmitter to the span calibration gas After a programmable settling time during which the span gas replaces the process gas measurements are taken of the span calibration gas The Display then activates a second solenoid valve to connect the Transmitter to the zero calibration gas The Transmitter then settles as the zero gas replaces the span gas and measurements are taken of the Zero calibration gas After taking span an
77. s YES to set the Error Handling option Use the NO or arrow keys to make the desired selection then press YES to confirm the entry Press YES to set the display response Use the NO or arrow keys to make the desired selection then press YES to confirm the entry Press YES to set the alarm response Use the NO or arrow keys to make the desired selection then press YES at the desired alarm response Programming the TMO2D 4 25 9 30 93 qxE __ o AEC EGGSLLCLU LEZGJA EECCCLEOEEC LEEE EEDDS9DEZ A ndl Set Error Handling cont Auto Cal Error Effects Press YES to set the recorder Set Recorder Response response Select Recorder Effect Use the NO or arrow keys to make No Effect the desired selection then press YES at the desired recorder response Select Recorder Effect Force High Select Recorder Effect Force Low Select Recorder Effect Hold Last Value Auto Cal Error Effects Press NO until the Done Done prompt appears AUTO CALIBRATION MENU Press YES to return to the Set Set Error Handling Error Handling option AUTO CALIBRATION MENU Press NO to proceed to the next Perform AutoCal option 4 26 Programming the TMO2D 9 30 93 ese Perform AutoCal The Perform AutoCal option is used to activate the Auto Cal procedure without waiting for the specified Auto Cal interval When an Auto Cal is performed either by press
78. sed rather than 1 2 3 the LCD will automatically resume displaying data and the user must press NO to re attempt the code Once all three digits have been entered correctly the Display will cease collecting data and the LCD will switch to Menu Mode Programming the TMO2D 4 1 9 30 93 Key Functions Programming the TMO2D via the Display Display Navigation The YES key is used to confirm numeric entries or to select a displayed menu option The NO key is used to clear a numeric entry or to scroll forward through the menu options The Left Arrow key has two functions tis used as a backspace key during numeric entry The right most digit of the entry will be erased for each press of the left arrow tis used to step backward through a list of menu options The Right Arrow key is used to scroll forward through the menu options It is equivalent to pressing the NO key in the Menu Mode This section gives a brief explanation of Display and menu navigation and then takes you through the programming procedure On power up the first line of the LCD screen contains the current measurement source and a real time clock The second line of the LCD contains the measured data on the left and the current alarm condition on the right see Figure 4 1 below Oxygen Analyzer a PANAMETRICS Figure 4 1
79. span pressure to the pressure sensor ex 0 PSI 6 Adjust R27 until TB1 3 is 20 milliamps 7 Repeat steps 3 through 6 until a consistent output is obtained Note R27 and R26 are interdependent adjustment of one pot will affect the setpoint of the other pot This system uses an analyzer with separate oxygen and pressure signal outputs These signals would normally be used by TMO2D programs to produce a pressure corrected oxygen value Set SAT switches as follows 1 4 5 amp 6 ZOFF up 2 amp 3 ON Set J3 to the LO position HI if using hydrogen helium Refer to page D 5 Zero and follow steps 1 through 12 setting E14 to 2 0 volts in step 3 With single background gas systems use only steps 1 through 6 1 Apply span background gas ex 10 oxygen remainder CO into the oxygen transmitter 2 Adjust S1 and R41 until the TB1 3 is about 15 milliamps D 8 The TMO2 Transmitter Calibration Procedure 11 30 92 External Pressure Signal Internal Hydrogen Gas Compensation Zero Span Compensation External Hydrogen Gas Compensation Internal Methane Gas Compensation Reverse Compensation Zero Span Compensation 1 Apply span pressure to the pressure sensor ex 1 0 PSI 2 Adjust R28 until TB1 4 is approximately 20 milliamps 3 Apply minspan pressure to the pressure sensor ex 0 0 PSI 4 Adjust R27 until the output is approximately 10 milliamps 5 Repeat steps 1 through 4 unti
80. ss YES to set the update rate For example enter 1 8 and O for a 180 sec 3 min interval The X represents the previously entered data interval Press YES to confirm the entry Press YES again to exit Press NO to proceed to the next submenu Press YES to exit Press NO to proceed to the next submenu Press YES to exit This completes the Setup Menu Press NO to proceed to the next Main Menu title Programming the TMO2D 9 30 93 The Recorder Menu The second Main Menu is the Recorder Menu The Recorder Menu is used to select which recorder A or B to adjust and allows entry of all necessary information for that recorder The TMO2D provides a choice of one non isolated 0 4 20 mA recorder output or two isolated 0 4 20 mA recorder outputs Both recorder options can be set for a 0 20 mA or a 4 20 mA response Output can be scaled anywhere within the range of the transmitter Note The Display is programmed to accept settings for two recorders however if only a single non isolated recorder output is used only recorder A is effective Use the steps in the following example to set up recorders We will set recorder A for a 0 20 mA output with 0 mA equal to 0 oxygen and 20 mA equal to 100 oxygen Refer to Appendix E for a flow diagram of the Recorder Menu MAIN MENU Setup MAIN MENU Recorders Select Recorder to set A B done Rcd A Output mA 0 20 4 20
81. stainless steel Consult factory Explosion proof Consult factory Weight Rack Mount 5 4 lbs 2 4 kg Bench Mount 7 4 lbs 3 4 kg Panel Mount 4 7 lbs 2 1 kg Weatherproof fiberglass 6 5 lbs 3 0 kg Weatherproof stainless steel Consult factory Explosion proof Consult factory Environmental Rack Bench Panel Mount General purpose Weatherproof fiberglass NEMA 4X IP65 Weatherproof stainless steel Consult factory Explosion proof Consult factory European Compliance This unit complies with EMC Directive 89 336 EEC and 73 23 EEC Low Voltage Directive Installation Category II Pollution Degree IL Specifications 5 3 Appendix A Thermoparamagnetic Magnetic Wind Theory Basic Principle icc uisu rh 0 mm ner A 1 The TMO2 Transmitter Oxygen Measurement Bridge A 1 The TMO2 Transmitter Compensation Bridge A 2 TMO2D Electronics Microprocessor based Compensation for Variations in Background Gas Composition A 2 4 8 92 M Basic Principle Since its magnetic susceptibility is approximately 100 times greater than most common gases oxygen can be easily distinguished from other gases based on its attraction into a magnetic field The magnetic susceptibility of oxygen varies with temperature more strongly attracted at low temperatures and less at high temperatures
82. structions for CE CE MARK COMPLIANCE IS REQUIRED ONLY FOR Mark Compliance UNITS USED IN EEC COUNTRIES For CE Mark compliance the TMO2D must be enclosed in a NEMA 4 stainless steel case In addition you must shield and ground the electrical connections as shown in Table F 1 below Note f you make the modifications as discussed in this appendix your unit will comply with the EMC Directive 89 336 EEC Table F 1 Wiring Modifications for CE Compliance Termination Modification 1 When connecting power select the cable entry closest to the chassis ground 2 Connect the power ground to the nearest chassis ground using the shortest run of wire possible Grounding lugs are provided in the TMO2D enclosure Input Output Use shielded cable to interconnect system enclosures Connect the shields to the nearest chassis ground using the shortest run of wire possible Grounding lugs are provided in the TMO2D enclosure After you make all the necessary electrical connections seal unused cable entry holes with standard conduit plugs or equivalent Installation Instructions for CE Mark Compliance F 1 1 20 97 Index A E Alarm Menu sslsleeeeeeeeee 4 11 Edit Calibration Data Alarms Background Compensation 4 30 Tests ee UNS PEEL EP 4 15 Pressure Compensation 4 30 Wire eg mai E ES 2 8 Response Parameters 4 30 Analyzer Error Handling in Auto Cal 4 25 Powering
83. surement Resolution 0 01 O Accuracy 1 of span Linearity 0 1 of span Repeatability 0 2 of span Drift Zero and span 0 8 of span month 0 2 O range and higher 1 5 of span month 0 1 O range Speed of Response 4 seconds for 63 of a step change in O with Display Ambient Temperature Influence Negligible if transmitter is heated at least 5 C 9 F above highest ambient temperature Transmitter Temperature Controlled to 45 C 113 F Atmospheric Pressure Influence 0 2 of span per mm Hg optional pressure compensation available Sample Flow Rate Influence Less then 1 of span with 50 to 1000 cc mm no flame arrestors 0 1 to 2 SCFH Line Voltage Influence Negligible Ambient Requirements Relative Humidity 0 100 non condensing Temperature 20 to 40 C 4 to 104 F higher temperatures available Sample Requirements Flow 50 to 1000 cc min 0 1 to 2 SCFH Pressure 20 psig maximum Power Requirements 24 VDC 1A max Powered by TMO2D electronics Electrical Output 0 4 20 mA to TMO2D electronics Transmitter Electronics Separation 300 m 1000 ft maximum shielded cable not required Physical Characteristics Dimensions H x W x D 143 x 152 x 235 mm 5 625 x 6 x 9 25 inches Weight 4 3 Kg 9 5 Ibs Specifications 5 1 1 20 97 TMO2 Oxygen Transmitter cont TMO 2D Display Specifications Performance Functional Connections 3 4 in NPTF electrical conduit 1 4 in N
84. t in the transducer curve The unit will then display the Working message before returning to the Calibrate Pressure option Programming the TMO2D 4 31 9 30 93 Edit Calibration Data cont Calibrate Pressure Transducer Data Calibrate Pressure Enter Grid Data Pressure Grid Entry of Curves 2 Curve 1 mmHg 760 0 Press NO to proceed to the Enter Grid Data option Press YES to enter grid data Use the numeric keys to enter the number of curves in the pressure calibration grid then press YES twice to confirm the entry and proceed to the next prompt Use the numeric Keys to enter the pressure in mmHg for each of the curves in the pressure grid then press YES twice to confirm the entry and proceed to the next prompt Note Be sure to repeat the procedure for each curve number Pressure Grid Entry of Points 2 Point 1 O XX X Use the numeric Keys to enter the number of points in the pressure grid then press YES twice to confirm the entry and proceed to the next prompt Use the numeric keys to enter the percent gas concentrations for each of the points in each curve as prompted then press YES to confirm the entry Note Percent gas concentrations MUST be in ascending order 4 32 Programming the TMO2D 9 30 93 Edit Calibration Data cont XX X 02 XXX X mmHg Enter Point yes NO At this
85. this prompt use the NO or arrow keys to select YES then press YES again to confirm the entry Use the NO or arrow keys to select BKGD then press YES to confirm the entry Press YES again to enter grid data Use the numeric Keys to enter the number of curves in the background grid then press YES twice to confirm the entry and proceed to the next prompt Use the numeric keys to enter the number of points then press YES twice to confirm the entry and proceed to the next prompt Use the numeric keys to enter the percent gas concentrations for each of the points in each curve as prompted then press YES to confirm the entry Note Percent gas concentrations MUST be in ascending order X XX 96O Curve 1 Enter Point yes NO At this prompt select YES to enter the calibration data for the given point or select NO to have the unit determine the value for the given point Note Each curve must have AT LEAST TWO entered data points 4 34 Programming the TMO2D 9 30 93 Edit Calibration Data cont X XX 96O Curve 1 To enter calibration data use the Enter Point YES no NO or arrow keys to select YES then press YES to confirm the entry XX XX Oo Curve 1 Use the numeric keys to enter the Oz mA X XX corresponding gas input milliamp value for the given point then press YES to confirm the entry XX XX 02 Curve 1 Use the numeric keys to enter the Comp mA
86. tion The TMO2D Oxygen Analyzer is configured with the following components and accessories The TMO2 Transmitter The TMO2D Hlectronic Display The Sample System specific to your application on which the Transmitter is mounted e A three or four wire color coded cable T N4 Note All applications require a sample system consult Panametrics whether you are ordering a sample system or planning to build one Features and Capabilities 1 1 11 30 92 The TMO2 Transmitter The TMO2 Transmitter is self contained consisting of the oxygen measuring cell and associated electronics enclosed in a cast aluminum housing see Appendix B for outline and installation dimension drawings The Transmitter has no moving parts It requires 24 VDC power and outputs a 4 20 or 0 20 mA signal proportional to oxygen concentration It can also output an optional pressure or background gas compensation signal The Transmitter housing is weatherproof and can be made explosion proof with inlet and outlet flame arrestors see Figure 1 1 below Flame Arrestors Figure 1 1 TMO2 Transmitter with Inlet and Outlet Piping 1 2 Features and Capabilities 4 8 92 The TMO2D Electronics Display The Sample System The TMO2D Hlectronics Display is an electronic module containing user friendly software that allows you to enter specific parameters for O range recorder output range alarm settings etc The Disp
87. to set the date Then use the numeric keys to enter a 2 24 and 91 The X s represent the previous date entered Press YES to confirm the entry Press YES again to exit Press NO to proceed to the next submenu The LCD contains an electroluminescent EL panel to enhance the readability of the screen in dim light EL panels have a finite life span dimming with use To maintain the life of the EL backlight the Display will automatically turn the backlight off after a predetermined time period The Backlight time out period can be set from 0 never on to 60 minutes The default timeout is 3 minutes For example to set this time to 10 minutes SETUP MENU Set Backlight SETUP MENU Remain ON min X 10 SETUP MENU Remain ON min 10 SETUP MENU Set Backlight SETUP MENU Set Display Press YES to set the backlight Then use the numeric keys to enter 10 The X s represent the previous time entered Press YES to confirm the entry Press YES again to exit Press NO to proceed to the next submenu Programming the TMO2D 9 30 93 Set Display Set Communications In order for a compensation value to be displayed either background or pressure compensation must be enabled through the Calibration Edit Calibration Data Menu see page 4 30 If neither pressure nor background compensation is enabled a Comp not enabled message will be displayed
88. to the desired selection Select YES or select NO Set System Response normal DAMPED Use the NO or arrow keys to move the brackets to the desired selection I YES E Select Compensation pres BKGD CAL PRESSURE Transducer Data Use the NO or arrow keys to move the brackets to the desired selection YES If you select PRES If you select BKGD CALIBRATE BKGD Enter Grid Data YES or NO PRES CURVE ENTRY of Points X Enter the number of points YES YES Point 1 mmHg X XX Enter the signal value in millimeters of mercury YES YES Point 1 P mA X XX Enter a value for pressure in milliamps Note These two prompts repeat for each point YES YES Working Bkgd represents the compensation you chose earlier YES or NO BKGD GRID ENTRY of Curves X Enter the number of curves in the pressure gas grid YES YES Curve 1 mmHg X X Note This prompt only appears if you selected pressure compensation Enter the signal for the curve in millimeters of mercury This prompt repeats for each curve YES YES BKGD GRID ENTRY of Points X Enter the number of points in the pressure gas curve YES YES Point 1 9602 X XX Enter the percentage of oxygen for point 1 This prompt repeats for each point YES YES X XX 9602 Curve 1 E
89. xygen concentration and the equilibration time for the zero gas If 100 nitrogen is used for the zero gas the concentration would be zero If a mixture is used the oxygen concentration of the mixture should be entered The default concentration is 0 oxygen After entering the O concentration the display will automatically prompt for the equilibration time for the zero gas The equilibration time should allow for the distance the calibration gas must travel and for the settling time of the Transmitter cell Note For a successful auto calibration to be performed the equilibration time should be at least 3 minutes The default is 5 minutes For testing purposes the equilibration time can be as low as 10 seconds The maximum equilibration time is 30 minutes For example set the zero gas for 100 nitrogen for an equilibration time of 3 minutes AUTO CALIBRATION MENU Press YES to enter the Set Zero Set Zero Gas Gas option Zero Gas The X s represent the previously 9602 X XX entered oxygen concentration Zero Gas Use numeric keys to enter a 0 then 02 0 00 press YES twice to confirm the entry and proceed to the next option Zero Gas ON for The X s represent the previously MM SS X XX entered equilibration time Programming the TMO2D 4 21 9 30 93 Set Zero Gas cont Zero Gas ON for MM SS 3 00 AUTO CALIBRATION MENU Set Zero Gas AU
90. ys to move the brackets to the desired DVM input YES 02 DVM TEST X XX mA The display shows the DVM input for oxygen ANY KEY Select DVM Input gas comp DONE Note Repeat the same procedure for comp After completing the DVM tests use the NO or arrow keys to move the brackets to DONE to exit YES TESTS Recorder Calibrate YES or NO Select Recorder to test A B done Use the NO or arrow keys to move the brackets to the desired recorder to be calibrated YES Set Recorder A to 02 X XX Enter the numeric value to represent the percent oxygen value to output to the recorder YES YES Select Recorder to test A B DONE Note Repeat the same procedure for Recorder B After completing the recorder calibration use the NO or arrow keys to move the brackets to DONE to exit YES 9 30 93 Goes to MAIN MENU Calibration PB TESTS Alarms Test MES or NO Select Alarm to test HIGH low done Use the NO or arrow keys to move the brackets to the desired alarm to test YES I High Alarm ON of done Use the NO or arrow keys to turn the alarm on or off YES Turn High Alarm on off DONE Note After making a selection the brackets jump to the opposite selection For example if YES is pressed at ON the brackets jump to OFF but the alarm rema

Download Pdf Manuals

image

Related Search

Related Contents

Exit - Canvys  RCA DRC275A DVD Player User Manual  Cartilha de Orçamento Familiar  HD15K HD17K  Insta-Fluor Plus - 株式会社パーキンエルマージャパン  i.Sound ISOUND-1596  SOMMAIRE  Table des matières    IC1005 USER MANUAL MANUEL DE L'UTILISATEUR  

Copyright © All rights reserved.
Failed to retrieve file