Home
        Multi Wire Myograph System
         Contents
1.    Probe temperature     At any given time  if the force applied on any channel    is out of range  the force reading for the overloaded ran ae  channel will turn yellow as a warning   Force chamber 2   0 26 mN    Force chamber 3   1 06 mN  Force chamber 4   0 76 mN    Probe temperature  37 0   C    ACTUAL FORCE       Three menus are accessible from the default    Actual  Force    screen or display  These menus are  Zero   Heat  and Settings        Force chamber 1   2 36 mN          Force chamber 2   Force chamber 3     Force chamber 4        Probe temperature  37 0   C    p             9     nm    Ld b 3 x  C Hm        iib   b p     9                nU  00     ee       16 Multi wire myograph system Model 620M   User manual    Zero Menu    This menu is used to zero the output of the trans   ducers  When using a data acquisition program like  LabChart by AD Instruments     using this feature will  reset the baseline of the chart traces without affect   ing the calibrations or physically changing any pre   load tensions placed on the mounted vessels  The  channels can be changed individually by pressing     SELECT    or all at once by pressing    ALL     Pressing     ENTER    will execute the zero function and return  the user to the ACTUAL FORCE display     Heat Menu    The heating unit and temperature are controlled from  this menu  To turn the heat on or change the preset  temperature for the system  access the temperature  control menu  Pressing the    HEAT    key will ente
2.   1 0 M solution     1  Make a 1 0M solution of CaCl   110 99  in double distilled H O  Filter sterilize the calcium    solution through a 0 22 um filter  The sterilized solution can be stored in the refrigerator  for up to 3 months       Dissolve all the chemicals except the CaCl  in approximately 80  of the desired final    volume of double distilled H O while being constantly stirred  For example  if 1 litre of  PSS is to be made  then dissolve all the chemicals in 800mL of double distilled H5O       Add the appropriate volume of 1 0M CaCl  for the total volume of PSS being made  for    example  1 6mL of 1 0M CaCl  for 1 litre of buffer   Continue to stir the PSS while the  CaCl  is being added       Bring the solution up to the final volume with double distilled H5O  Continue to stir    the solution until the EDTA is fully dissolved  This takes about 15 minutes at room  temperature     Chapter6 49    Before use   5  Dilute the 25 x PSS stock solution 1 25 using double distilled H5O   6  Add   e 091 g L Glucose  e 100 g L NaHCO     f  Aerate the solution with carbogen  95  O    5  CO   for at least 20 minutes  If necessary  wait further for the pH of the buffer to reach pH 7 4     High potassium Physiological Saline Solution  KPSS     1x 60mM KPSS   Chemical Mol Wt  mM g 0 5L g L g 2L g 4L  NaCl  58 45  74 7 2 18 4 37 8 73 17 46  KCI  74 557  60 2 24 4 47 8 95 17 89  KH5PO   136 09  1 18 0 08 0 16 0 32 0 64  MgSO  7H50  246 498  1 17 0 145 0 29 0 58 1 16  NaHCO   84 01  14
3.   Step no   og  5     Make sure that the transducer is  not subjected to any force   When the relative force reading  is stable  go to next step     Force Chamber 1  3261    NEXT STEP    CHAMBER 1 CALIBRATION    Step no   JIB    Carefully place the 2 g weight  On the pan    When the relative force reading  is stable  go to next step     Force Chamber 1  3346       BACK NEXT STEP       Chapter3 19    Step 6 is to verify that the calibration was performed  correctly  The    Force Chamber 1    reading should be  9 81   0 1 mN  If the    Force Chamber 1    reading is  off by more than 0 1 mN  then remove the weight   press    BACK    to return to Step 4  and repeat the cali   bration process  If the  Force Chamber 1  reading is  satisfactory  then press  NEXT STEP   Calibrate the  other chambers in the same manner     To calibrate the transducers to a data acquisition sys   tem such as the AD Instruments PowerLab and Lab   Chart system  see Quick Start Guide for the 620M     2  VALVE DELAY     Pressing  VALVE DELAY  in the SETTINGS menu will  allow the user to modify the time duration that the  vacuum valves stay open for washes  Factory default  is set at 1 second  but 1 second is not enough time  to completely empty a chamber with even as small a  volume of 5 mL     Pressing  SELECT  next to any given channel will  cause the line selected to turn blue  The up and down  arrow keys can then be used to modify the length of  time the vacuum valves stay open after the valves  have b
4.   The purpose of checking endothelial function  therefore  is     1  To check whether the endothelium is intact  The procedure is performed to verify that the  endothelium was not damaged during dissection or mounting     2  To verify that endothelium was sufficiently removed  denuded  if an experiment requires  removal of the endothelium     These 2 functional parameters can be used as inclusion exclusion criteria  If the vessel does  not contract or relax according to the end user s acceptable criteria  new vessels can be  mounted if desired  Again  depending on the vessel and the conditions  these inclusion ex   clusion criteria will need to be decided upon and assessed by the end user     Concentration dependent responses  the actual experiment    After the wake up protocol is completed  the actual experiment can be conducted  A typical  experiment would consist of concentration response curves to one or several agonists in the  presence or absence of antagonists  Depending on the agonists  multiple concentration re   sponse curve protocols can be performed in the same vessel segment  If the agonists used  are water soluble and their effects are easy to wash out  then animals can be preserved and  data collection can be maximized by performing several concentration responses to different  agonists     42 Multi wire myograph system Model 620M   User manual    6 2 Example of step by step experimental procedure  Dissection and mounting    1  Isolate organ  vascular bed  or i
5.   This allows the vessels to slowly  heat up     4  Perform your normalization using the DMT normalization module on LabChart for small  vessels to determine the optimal passive tension  If using larger arteries such as mouse  aortic rings  a length tension relationship will need to be performed to determine optimal  passive tension to conduct the reactivity at     5  Make a note of what the passive tension is     6  Wash the vessels with bubbled  warmed  37  C  PSS by pressing the    All Channels    but   ton on the front of the interface to evacuate the chambers of buffer     f  Immediately replace the buffer  bmL  with fresh  bubbled  warmed  37  C  PSS     8  Allow the artery to continue equilibrating for another 20 minutes     Chapter6 43    9  During the next period of equilibration  if the tension slips or changes  continue to ad   just the passive tension so that it remains at the noted passive tension in step 5     10  After 20 minutes from the first wash  now 40 minutes since the start of equilibration    do another single wash with warmed  bubbled PSS     11  Once a total of 60 minutes has passed  the Wake up Protocol can be initiated     The wake up protocol    1  Reset the baseline values to  Zero  by using the  Zero  function menu of the inter   face     2  Use a combination of PSS containing 60mM K   60mM KPSS  and a contractile agonist   either norepinephrine for mesenterics or some other agonist for other types of resist     ance arteries  for the wake up pro
6.   press    NEXT STEP      Step 4 is the first step in the actual weight calibra   tion process  A 4 digit number will be displayed in  blue at the bottom of the screen  If nothing has been  perturbed during the heating process  the zero  O  gram  or 0 00 mN calibration should be stable as in   dicated by the 4 digit number and  NEXT STEP  can  be pressed at this time  If the 4 digit number is not  stable  then wait until the number has stopped fluc   tuating before pressing  NEXT STEP      Step 5 is the 2 gram weight calibration  At this step   place the 2 gram weight in the pan closest to the  transducer so as to simulate a vessel pulling on the  jaw or pin attached to the transducer  Remember   a 2 gram weight in a 90   vector is cut in half  and  the transducer will only detect 1 gram or 9 81 mN of  force  The weight placement should cause a positive  increase in the 4 digit number  Wait at least 10 to  15 seconds for the applied force to stabilize before  pressing  NEXT STEP   Once the 4 digit number has  stabilized  press  NEXT STEP      CHAMBER 1 CALIBRATION    step vo  DD C  9  GJ    Turn the heat on  Wait until  the temperature is stable     Temperature set point  37 0  C  Probe temperature  36 8   C    BACK HEAT ON  imo  NEXT STEP    CHAMBER 1 CALIBRATION       Step no   OBUCE    Turn the heat on  Wait until  the temperature is stable     Temperature set point  37 0   C  36 8   C    BACK   Haron   HEAT OFF NEXT STEP    Probe temperature     CHAMBER 1 CALIBRATION     
7.  10  M stock Ach or carbachol to 5mL bath   Allow to relax for 2 minutes     4  Wash the contractile agonist and Ach or carbachol from the vessel preparation  Wash at  least 5 to 6 times over 15 to 20 minutes to completely eliminate the effects of the vaso   pressor and Ach or carbachol     5  Preparation is now ready for experiments   Concentration responses  Example of agonist concentration response experiment     5 0 uL of 10  M to get 10  M in the 5 mL bath  1 0 uL of 10   M to get 3x10  M in the 5 mL bath  3 5 uL of 10   M to get 108 M in the 5 mL bath  1 0 uL of 10  M to get 3x108 M in the 5 mL bath  3 5 uL of 10  M to get 10  M in the 5 mL bath  1 0 uL of 10  M to get 3x10  M in the 5 mL bath  3 5 uL of 10   M to get 10   M in the 5 mL bath  1 0 uL of 10  M to get 3x10 M in the 5 mL bath  3 5 uL of 10  M to get 10   M in the 5 mL bath    Each concentration should be added cumulatively without washing the previous addition out   Once the concentration response is complete  the vessels can be washed completely  5 to  6 washes over 20 minutes   and another concentration response can be performed with a  different agonist     For example  the first concentration response can be norepinephrine  NE   This can be washed  out  the ECs can be determined from the concentration response  The NE EC     can then be  used to contract the vessel again  An acetylcholine  Ach  or carbachol concentration response  can be performed to assess endothelial function once a stable contraction
8.  9 0 625 1 00 2 00 5 00  Glucose  180 16  DD 0 5 1 00 2 00 4 00  EDTA  380  0 026 0 005 0 01 0 02 0 04  CaCl   110 99  1 6 O 8mL 1 6mL 3 2mL G 4mL    1 0 M solution     1  Make a 1 0M solution of CaCl   110 99  in double distilled H O  Filter sterilize the calcium  solution through a 0 22 um filter  The sterilized solution can be stored in the refrigerator  for up to 3 months     2  Dissolve all the chemicals except the CaCl  in approximately 80  of the desired final  volume of double distilled H2O while being constantly stirred  For example  if 1 litre of PSS  is to be made  then dissolve all the chemicals in 800mL of double distilled H5O     3  Add the appropriate volume of 1 0M CaCl  for the total volume of PSS being made  for  example  1 6mL of 1 0M CaCl  for 1 litre of buffer   Continue to stir the PSS while the  CaCl  is being added    4  Bring the solution up to the final volume with double distilled H5O  Continue to stir  the solution until the EDTA is fully dissolved  This takes about 15 minutes at room    temperature     5  Aerate the solution with carbogen  95  O    5  CO2  for about 20 minutes     50 Multi wire myograph system Model 620M   User manual    Appendix 1     Terms of warranty    Warranty   DMT A S warrants to the original user that myograph systems manufactured by DMT A S will  be free from defects in materials and workmanship for a period of three years after the date  of delivery  DMT A S will repair or replace any defective part  subject to the condi
9.  SELECT    The line to be modified will turn blue  indicating that  the interface is waiting for input  When  ALL  is cho   sen  all lines corresponding to all 4 channels will turn  blue  Changing the numeric value for the chosen pa   rameter can be done by touching the up or down ar   row keys     ALL  Once the desired setting has been chosen  press   ing  ENTER  will lock the selection and be stored in    memory  Pressing the white  X  in the red box will  exit that menu and take you automatically to the Ac     tual Force Display     SET FORCE TO ZERO       Force chamber 1   2 36 mN    Force chamber 2   0 26 mN    Force chamber 3   1 06 mN    Force chamber 4   0 76 mN       SET FORCE TO ZERO    SELECT    SELECT    SELECT    SELECT       Force chamber 1   2 36 MN    Force chamber 2   0 26 MN    Force chamber 3   1 06 mN    Force chamber 4   0 76 mN    MEASUREMENT RANGE    Range Chamber 1  200 mN    Range Chamber 1  800 mN    Range Chamber 1  400 mN    200 mN    Range Chamber 1        SELECT  SELECT  SELECT    SELECT    Chapter 3 15    Power Up Screen   After turning on the 620M Interface  an    Introduc   tion    screen appears  The system is auto calibrating  the A D converters while this screen is displayed     DMT620 MYOGRAPH    Multi Myograph System    Model DMT 620  Software Revision 03 00 09  Date  Nov  27 2009       After a few seconds  the    ACTUAL FORCE    display will  appear     ACTUAL FORCE  Force chamber 1   Force chamber 2   Force chamber 3     Force chamber 4
10.  TEMP OFFSET  ON CHAMBER         The  SELECT  and  ALL  functions are the same  in this menu as previously described for the   VALVE DELAY  menu  Pressing  ENTER  will t 4 2  C uice  store the numbers in memory for future experi    44  C SELECT  ments     TEMP OFFSET ON CHAMBER       1 2  C SELECT    1 4  C SELECT    Il  FACTORY DIAGNOSTICS   Entering FACTORY DIAGNOSTICS will display  the LOGIN CODE TO DIAGNOSTICS window  This  window is for trained technicians and used for    diagnostics and troubleshooting purposes  The  general user will not have access to this window     B o    Entering the proper 5 digit pin number  however   will allow the trained technician access to Di  fode init value 51761 OCOL  agnostics panels that will provide information  during a malfunction or mechanisms to change  other settings controlled by the onboard compu   ter     LOGIN CODE TO DIAGNOSTICS    Type login code to get acces          22 Multi wire myograph system Model 620M   User manual    Chapter 4     The Multi Wire Myograph Unit    This chapter contains a complete explanation of how to adjust  calibrate and maintain the  myograph 620M system so that the myograph is always performing at peak performance     4 1 Changing and adjusting the mounting supports    Each chamber can accommodate mounting supports for either small vessels   gt 50um  or  larger segments   gt 500um   Because the mounting supports can be changed easily  experi   ments can be performed with different vessels of varyin
11.  and will appear as a  flat line trace at 20 mN  even though the force read   ings on the interface may exceed 20 mN  Therefore   change the FORCE REC  OUT settings to an appropri   ate setting so as to capture any maximal response  from the vessel of interest  This value should not  exceed the settings for the transducer range  which  is defined by the sub menu  MEASUREMENT RANGE  and is explained in the next section     The  SELECT  and  ALL  functions are the same in  this menu as previously described for the  VALVE DE   LAY  menu  Pressing  ENTER  will store the numbers  in memory for future experiments     Anytime this function is changed  a new weight cali   bration on the transducers should be performed  en   tering the new voltage values into the data acquisi   tion system being used     4  MEASUREMENT RANGE     The MEASUREMENT RANGE sub menu in SETTINGS  determines the maximum force capacity of the trans   ducer  The factory setting is 200 mN  but the trans   ducer capacity can be changed to 400 mN  800 mN  or a maximum of 1600 mN of force detection  de   pending on the size of the vessel used     The  SELECT  and  ALL  functions are the same in  this menu as previously described for the  VALVE DE   LAY    menu  Pressing    ENTER    will store the numbers  in memory for future experiments     Whenever the force recording output range is  changed  a new weight calibration should be per   formed on the transducers  and the new voltages  that correspond to the new weigh
12.  calibration chamber 4     FORCE CALIBRATION    Force calibration chamber 1   Force calibration chamber 2     Force calibration chamber 3     Force calibration chamber 4     CHAMBER 1 CALIBRATION    SELECT  SELECT  SELECT    SELECT    SELECT  SELECT  SELECT    SELECT    xX       step o  89  3  6  C  9   9J    Follow the Weight calibration    procedure in the User Manual   Prepare the jaws and chamber for    calibration   When ready go to next step     NEXT STEP    Step no   OBO  s     Place the calibration bridge on  the myograph  Be careful when  placing the bridge  The pin  must not touch the wire jaw   When ready go to next step     NEXT STEP       Step 3 initiates the heating process for the cham   bers  In order for the calibration to be accurate  the  transducers must be heated to the experimental tem   perature to be used to accommodate heat induced  expansion of the electronic parts in the transducer   Otherwise  inaccurate readings and transducer drift  may occur  introducing large errors into the experi   ment  To start heating  press    HEAT ON      Covering the chambers with the chamber covers will  expedite the chamber heating  Place the tempera   ture probe into the chamber for the first calibration  to monitor when the chamber has reached the tar   get temperature  Heating will take about 20 to 30  minutes for the chambers and transducers to come  to 37  C with the chamber covers in place  Once the  chamber s  are heated and have reached the target  temperature
13.  ence corresponding to 100 mmHg  ICioo  by a factor k  The factor is for rat mesenteric arteries   0 9   Again  this value should be optimized for the particular tissue preparation being used  by a length tension curve     IC   k e IC     The normalized internal  lumen  diameter is then calculated by     The micrometer reading X  at which the normalised internal circumference is attained is  calculated by      IC    IC      2    Appendix5 59    Appendix 6     Service Check    For successful studies utilizing small vessels or other small tubular tissues  the myograph  needs to be performing optimally  To make sure that our customers always are dealing with  first class myographs  DMT offers a Myograph Service Check at a very favourable price     The Myograph Service Check includes a complete disassembly of the system for inspection  of all mechanical and electronic parts  The myograph is then reassembled  adjusted and all    electronic and mechanical parts are tested     Please note that the service does not include replacement of transducers or any other need   ed spare parts  Please contact DMT for information about prices     60 Multi wire myograph system Model 620M   User manual    Appendix 7     Shipping Instructions    If the myograph system needs to be sent back for service or repair  please read the following  shipping instructions very carefully     Before packing the myograph system  please remember that the myograph is a very delicate  piece of equipment  and theref
14.  gently  slide the calibration kit forward toward the micrometer so that the transducer arm rests  on the wire or pin  creating a force reading on that channel  Carefully slide the calibration  kit back toward the transducer slowly until the force reads zero or very close to zero  At  this point  as soon as the force reads zero  the transducer arm will be properly placed for  weight calibration     4  Go to the FORCE CALIBRATION sub menu of the SETTINGS menu on the Interface to be   gin the actual transducer calibration  The process that is described above is reiterated in  6 steps once the FORCE CALIBRATION sub menu is initiated  which is described in detail  in Chapter 3        Transducer arm    Figure 4 7   Illustration of the proper placement for the balance transducer arm for calibration    Chapter 4 27    4 3 Checking the force transducer    The myograph force transducer is a strain gauge connected to a Wheatstone bridge  The  force transducers for each chamber are housed in a separate  protective compartment  Fig   ure 4 8   While the protective cover offers some mechanical protection for the force trans   ducers  they are still very vulnerable to applied forces exceeding 1 Newton  100 grams  or  fluid running into the transducer compartment due to insufficient greasing of the transducer  pinhole             e Transducer house    Figure 4 8   Illustration of the proper transducer house    If the force readings on the Interface appear unstable or noisy  then first check
15.  instructions are followed at all times     Myograph chamber tubing   To prevent the tubing from becoming blocked with buffer salt deposits after an experiment   remove the chamber cover from the Myograph Chamber and turn on the vacuum and press  the vacuum valve for about 10 seconds by holding down the valve button s  down  Turn off  the vacuum and gas supply  Remove any water or buffer remaining in the chamber or on the  tubing using absorbent paper     Force transducer  The force transducer is the most delicate and fragile component of the myograph system   Extreme care must be used when handling or touching the force transducers     As a part of daily maintenance  inspect the grease around the transducer pin extending from  the transducer housing pinhole before starting any experiment  Insufficient grease in this  area will allow buffer and water to enter the transducer housing and cause damage to the  force transducer     IMPORTANT NOTES     e DMT recommends that the high vacuum grease sealing the transducer pinhole is checked  and sealed at least once a week  especially if the myograph is used frequently    e DMT takes no responsibilities for the use of any other kinds of high vacuum grease other  than the one available from DMT    e DMT takes no responsibilities for any kind of damage applied to the force transducers     Linear slides  Check the linear slides  under the black covers  for grease at least once a week  In case of  insufficient lubrication  grease the slid
16.  is reached  a  plateau   A 3  concentration response can be performed similar to the 2   only using sodium  nitroprusside or nitroglycerin for endothelium independent relaxation     These final concentrations will need to be modified depending on the agonist s  used  the  range of efficacy for that agonist  and the volume of buffer used in the chambers     6 3 Practice exercises    The exercises listed in this section are short experiments to try for tractice to gain a familiarity  with how a typical experiment should be performed  These are only suggestions for practice     Experiment 1  Noradrenaline concentration response curve  The purpose of this exercise is to determine the sensitivity of rat mesenteric small arteries  to the vasoconstrictor  noradrenaline in the presence and absence of compounds that would    affect the pharmacology of noradrenaline in a mounted vessel  Noradrenaline is the same as  norepinephrine     Chapter 6 45    Background    Noradrenaline causes contraction of mesenteric small arteries through activation of a   adrenoceptors  whereas noradrenaline activation of B adrenoceptors causes vasodilatation   The purpose of this exercise is to determine the contraction sensitivity to noradrenaline  The  vasodilatory effect of noradrenaline is eliminated throughout the experiment by the constant  presence of the B adrenoceptor antagonist  propranolol     Rat mesenteric arteries are densely innervated by sympathetic nerves  which have a highly  efficient r
17.  of the Interface and  the Multi Wire Myograph Unit     Chapter 2 describes step by step instructions to set up a complete 620M Wire Myograph  System  including accessories     Chapter 3 is a complete manual to the 620M Interface  This chapter describes  in detail  how  to navigate the menus and how to use the special features of the 620M Myograph System     Chapter 4 contains procedures describing general and daily maintenance of the myograph  unit  e g  adjustment of supports  weight calibration of the force transducer and cleaning  instructions     Chapter 5 describes how to mount vessels in the wire myograph system  This includes a  complete dissection and mounting procedure for small mesenteric arteries  This chapter also  contains information on the normalization procedure     Chapter 6 contains information on how to perform actual experiments on the myograph  system   his includes protocols and step by step instructions on how to conduct a typical  vascular reactivity experiment  as well provide a couple of practice experiments to get familiar  with the system as well as typical responses from the vessel used for experiments  This  chapter also provides buffer recipes that can be used for any vascular reactivity experiment     Appendices contain additional information about terms of warranty  ocular calibration  procedures  reading a micrometer  principles of weight calibration  normalization theory  fuse  replacement  myograph service  shipping instructions  accessori
18.  operation with Professor M  J  Mulvany and The University of Aarhus     During the late 1980s and through the 1990s  several improvements were applied to the  myograph systems  such as a new mechanical design  a more robust transducer  and a new  electronic system  New systems also were introduced  such as the automatic dual myograph  510A  the multi myograph 610M and the confocal myograph 120CW  In 2000  J  P  Trading  changed its company structure and became known as DMT     Today  DMT is one of the world s leading designers and manufacturers of wire myographs   pressure myographs  culture myographs and organ tissue baths  Driven by our global cus   tomer base  our singular goal is to develop and manufacture first class research equipment  within the fields of physiology and pharmacology     Introduction    5    Safety    The 620M Multi Wire Myograph System has been designed for use only in teaching and re   search applications  It is not intended for clinical or critical life care use and should never be  used for these purposes  or for the prevention  diagnosis  curing  treatment  or alleviation of  disease  injury  or handicap     e Do not open the unit  the internal electronics pose a risk of electric shock   e Do not use this apparatus near water     e To reduce the risk of fire or electric shock  do not expose this apparatus to rain or mois   ture  Objects filled with liquids should not be placed on the apparatus       Do not block any ventilation openings  Install i
19.  small vessels  Figure 4 1   6  Loosen screw A to move the micrometer side jaw toward or away from the micrometer  f  Loosen screw B to move transducer side jaw toward or away from the transducer    8  Loosen screw C to vertically alight the transducer side jaw  Screw C is the screw on the  transducer side support that is furthest away from the transducer    Fine adjusting the jaws for small vessels  Figure 4 2 and Figure 4 3   9 Tightening Screw   will move the micrometer side jaw downward and to the left  10  Tightening both screws   and III will move the micrometer side jaw straight down    11  Tightening both screws Il and IV will move the micrometer side jaw straight up    Micrometer e       Figure 4 2  Fine adjustments of the jaws in the myograph chamber     wn rw  D  V x x    Figure 4 3   lllustrations of properly aligned jaws  depicted on the far left  and incorrectly aligned  jaws  depicted in the middle and far right             24 Multi wire myograph system Model 620M   User manual    Fine adjusting the pins for larger vessels  Figure 4 4 and Figure 4 5     12  Loosen screw A to move the micrometer side arm holder sideways   13  Loosen screw B to move the micrometer side pin toward or away from the transducer   14  Loosen screw C to align the transducer side pin horizontally     15  Loosen screws D and E to align the heights of the pins vertically                 1       5  Zr  4               Figure 4 4   Fine adjustments of the pins in the myograph chamber    Pin
20.  that the cham   bers are connected properly to the Interface and that the chambers are plugged all the way  into the interface     If the force reading s  are still unstable or noisy  then perform a new calibration of the force  transducer as described in Chapter 3 and Chapter 4 2     During the new calibration  monitor the relative force reading values in the FORCE CALIBRA   TION sub menu on the Interface  Steps 4 and 5 of the calibration procedure   The normal  operating values for the force transducer during calibration should be between 3000 and  3500     e  fthe value is O  a single digit  or a three digit number  the force transducer is broken and  needs to be replaced    e  f the value is less than 2000 or greater than 4500  the force transducer is broken and  needs to be replaced    e  f the message    OFF    is displayed on the main page of the Interface  even though the  chamber is plugged in at the rear of the interface  the force transducer is broken and  needs to be replaced  In addition  if the force reading s  appear yellow in color  cannot be  reset to zero  AND the transducer cannot be recalibrated  the force transducer is broken  and needs to be replaced     If any other problems related to the force transducer are encountered  please contact DMT  for advice or further instructions     28 Multi wire myograph system Model 620M   User manual    4 4 Force Transducer Replacement    If the force transducer breaks and needs to be replaced  follow this step by ste
21.  to be optimized  depending on the vessel chosen for observation   The step by step protocol at the end of this chapter lists a wake up protocol that is ideal for  rat mesenteric resistance arteries and may prove to be a good starting point for other vessels  of interest     Final vascular function assessment    One of the last components of the wake up protocol is an agonist mediated contraction and  test for endothelial function  The contraction should be a contraction elicited by a submaxi   mal concentration of a vasopressor of choice  For example  in mouse aorta  10   mol L phe   nylephrine in an endothelium intact vessel segment would be an appropriate stimulus  This  contraction can be used as a final stimulus in the wake up protocol  In addition  this stimulus  can be used in two ways     1  To assess whether the vascular smooth muscle is contracting properly     2  It can be used as the contraction needed to test an endothelium dependent relaxation   if intact endothelial experiments are to be performed     It is recommended that endothelial function  or lack of  is evaluated  Stimulating a vessel  segment with acetylcholine or carbachol causes the release of nitric oxide  NO  from the en   dothelial cells  If the endothelium is undamaged after dissection and mounting  a substantial  relaxation will occur  If the endothelium is removed or damaged during dissection and mount   ing  a partial relaxation or no relaxation to acetylcholine or carbachol will be observed   
22.  were used  which precluded the use of small  vessels  In addition  the vessel segment had to be directly manipulated with dissecting in   struments  causing mechanical trauma  Investigations of smaller vessels  therefore  were  limited to in vivo perfusion experiments and histological examination     In 1976 Professor M  J  Mulvany and Professor W  Halpern described  for the first time  a new  technique that made it possible to investigate highly isometric responses from vessels with  internal diameters as small as 100 um  The mounting procedure was refined in 2 ways  1   both ends of each mounting wire were secured under tension without any direct manipula   tion of the vessel  and 2  segments of small vessels could not be atraumatically mounted as  ring preparations in a myograph for recording of highly isometric force measurements     During the late 1970s  some improvements were made to the myograph  and in 1981  a new  dual myograph that allowed simultaneous testing of two vessels was introduced  In parallel   the technique became widely acknowledged  resulting in a growing interest in the myograph  systems  In 1986  the growing demand resulted in the foundation of the private company  J   P  Trading  with the purpose of making the myograph systems commercially available world   wide  At the same time  J  P  Trading initiated a comprehensive improvement programme for  the existing myograph systems as well as a development programme of new myograph sys   tems in close co
23.  will cause corrosion of the stainless steel jaws and pins     IMPORTANT NOTES     Be very careful using HCl or HNOs because these acids may cause extreme damage to  the stainless steel chambers and supports  DO NOT USE bleach to clean the chambers   Repeated use of chlorinated solutions such as bleach and HCI will cause damage to the  stainless steel parts of your myograph system  Avoid using them if at all possible    After cleaning  ALWAYS check that the grease around the transducer pin is sufficient to  keep the buffer and water from entering the transducer housing     If red or brown discolorations appear on the chamber sides or on the supports  the following  cleaning procedure will work in most cases     1     Incubate the myograph chamber and supports for 30 minutes with 2mM T 1210 Tetrakis    2 pyridylmethyl  ethylenediamine solution dissolved in double distilled water       Use acotton tip applicator to mechanically clean all the affected surfaces during the last    15 minutes of the incubation period       Wash the myograph chamber and supports several times with double distilled water       Incubate the myograph chamber with 96  ethanol for 10 minutes while continuing the    mechanical cleaning with a cotton tip applicator       Remove the ethanol solution and wash a few times with double distilled water  Incubate    the myograph chamber and supports with an 8  acetic acid solution for 10 minutes and  continue the mechanical cleaning with a swab stick       Wash t
24. 00  0 5 1 00 2 00 4 00  0 005 0 01 0 02 0 04    O 8mL 1 6mL 3 2mL 6 4mL    48 Multi wire myograph system Model 620M   User manual    5       Make a 1 0M solution of CaCl   110 99  in double distilled H5O  Filter sterilize the calcium    solution through a 0 22 um filter  The sterilized solution can be stored in the refrigerator  for up to 3 months       Dissolve all the chemicals except the CaCl  in approximately 80  of the desired final    volume of double distilled H O while being constantly stirred  For example  if 1 litre of  PSS is to be made  then dissolve all the chemicals in 800mL of double distilled H5O       Add the appropriate volume of 1 0M CaCl  for the total volume of PSS being made  for    example  1 6mL of 1 0M CaCl  for 1 litre of buffer   Continue to stir the PSS while the  CaCl  is being added       Bring the solution up to the final volume with double distilled H5O  Continue to stir    the solution until the EDTA is fully dissolved  This takes about 15 minutes at room  temperature     Aerate the solution with carbogen  95  O    5  CO   for about 20 minutes     25x Concentrated PSS     Chemical Mol Wt  mM g 0 5L g L g 2L g 4L  NaCl  58 45  3250 94 98 189 96 379 92 159 84  KCl  74 557  117 5 4 375 8 75 17 5 35 0  KH5PO   136 09  29 5 2 0 4 0 8 0 16 0  MgSO  7H50  246 498  29 25 3 625 1 25 14 5 29 0  NaHCO   84 01  14 9 0 625 1 25 2 50 5 00  Glucose  180 16  5 5 0 5 1 00 2 00 4 00  EDTA  380  0 65 0 125 0 25 0 50 1 0  CaCl   110 99  40 20mL 40mL 80mL 160mL   
25. 16780 um    Appendix 3 55    Appendix 4     Principles of weight calibration    Calibrating the force transducer with a standard weight is based on simple physics where a  known amount of weight acting on a balance produces a net torque  The magnitude of the  torque T about a point of rotation P is defined by              aor    T r F sing    The distance from the point of rotation to the point on the object is denoted as    r     where the  force    F    is acting at the angle  8  The physics of the weight calibration setup is illustrated in  Figure A4 1       Transducer Arm  TA        Figure A4 1 Theoretical principle of the weight calibration    Applying the weight on the pan arm creates net torque acting at the    centre of gravity     result   ing in a force    F    acting on the force transducer  The following two equations describe the  forces working during the weight calibration     d  pps Poll  REI g   sind     weigt    2  T  o F_  Sing     56 Multi wire myograph system Model 620M   User manual    Where    ra is the length of the pan arm     F     is the force acting on the pan arm when apply   ing the weight     F    is equal to the acceleration of gravity times the mass of the weight  The  length of the transducer arm is    r     and the force acting on the force transducer is    F       The net torque acting at    center of gravity    is constant for the weight calibration setup  There   fore  equation 1 and 2 are equal  making it possible to calculate the force 
26. DANISH MYO TECHNOLOGY    Multi Wire Myograph System  Model 620M       User Manual  Version 1 0    Multi Wire Myograph System   Model 620M    User Manual    Trademarks    Pentium is a registered trademark of the Intel Corporation  Windows  Windows 95  Windows  98  Windows ME  Windows NT  Windows 2000  Windows XP  Windows Vista  and Windows 7  are registered trademarks of Microsoft Corporation  All other trademarks are the properties    of their respective owners     DMT reserves the right to alter specifications as required   This document was  as far as possible  accurate at the time of printing  Changes may have  been made to the software and hardware it describes since then  New information may be    supplied separately     This documentation is provided with the DMT Multi Wire Myograph System      Model 620M   Version 1 0    No part of this document may be reproduced by any means without the prior written    DMT A S   Skejby Science Center  Skejbyparken 152  DK 8200 Aarhus N  Denmark    Tel    45 87 41 11 00  Fax   45 87 41 11 01    www dmt dk  sales dmt dk  support amp dmt dk    permission of DMT     Copyright    2010 DMT A S    DANISH MeO TH CRN OL _ OG    DMT Asia   Rm 28C Liang Feng Building  No  8 Dong Fang Road   Lu Jia Zui Financial District  Shanghai 200120   P R  China    Tel   86  0  21 6486 9685  Fax   86  0  21 5877 0063    www dmt asia com  sales dmt asia com  Support dmt asia com    DMT USA  Inc    1201 Peachtree Street  400 Colony Square  Suite 200 630  Atla
27. Perform the normalization as described in Chapter 5 3 and Appendix 5     3  Perform the wake up protocol as described in Chapter 6 1     4  Add aconcentration of noradrenaline to the vessel that corresponds to the concentration  that causes 70  of the maximum contraction  When the contraction reaches a stable  plateau  add increasing cumulative concentrations of acetylcholine to the chamber   using Table 6 2 below  Wait for a stable response before adding the next concentration    of acetylcholine     Acetylcholine final  concentration  bmL volume   solution to the bath     1x1019 M  3 x 10109 M  1x 109 M  3 x 109 M  1x 108 M  3 x 108 M  1x 107 M  3 x 10    M  1x 108 M  3 x 108 M  1x 10   M    Add the following stock    5 0uL of 107 M  1 0uL of 10   M  3 5uL of 10   M  1 0uL of 10   M  3 5uL of 10   M  1 0uL of 10  M  3 5uL of 10  M  1 0uL of 103 M  3 5uL of 103 M  1 0uL of 102 M  3 5uL of 102 M    Table 6 2 The order of acetylcholine stock solutions to be added to the myograph chamber and the  final concentration of the added acetylcholine in a 5mL bath     6 4 Buffer Recipes    Physiological Saline Solution  PSS     1x PSS   Chemical Mol Wt   NaCl  58 45   KCl  74 557   KH5PO   136 09   MgSO  7H50  246 498   NaHCO   84 01   Glucose  180 16   EDTA  380   CaCl   110 99      1 0 M solution     mM  130    4 7  118  1 17    14 9  D D  0 026  1 6    g 0 5L g L g 2L g 4L  3 799 7 598 15 20 30 39    0 175 0 35 0 70 1 40  0 08 0 16 0 92 0 64  0 145 0 29 0 58 1 16  0 625 1 25 2 50 5 
28. Servicing is required when the apparatus  has been damaged in any way  such as  the power supply cord or plug is damaged  liquid has  spilled onto or objects have fallen into the apparatus  the apparatus has been exposed to rain  or moisture  does not operate normally  or has been dropped     Multi wire myograph system Model 620M   User manual    EMC EMI    This equipment has been tested and complies with the limits for a Class B Digital device   pursuant to part 15 of the FCC rules  These limits are designed to provide reasonable protec   tion against harmful interference in residential installations  This equipment generates  uses   and can radiate radio frequency energy and  if not installed and used in accordance with  the instructions  may cause harmful interference to radio communications  However  there  is no guarantee that interference will not occur in a particular installation  If this equipment  does cause harmful interference to radio or television reception  which can be determined by  monitoring the interference while turning the equipment off and on   the user is encouraged  to correct the interference by one or more of the following measures     e Reorient or relocate the receiving antenna   e Increase the separation between the equipment and receiver     e Connect the equipment into an outlet on a circuit different to that which the receiver is  connected to     e Consult the dealer or an experienced radio TV technician for help     Approvals    Complies wit
29. a aaa A A D 14  Chapter 3     The Interface Menus              ccssscsssccnsccssecnseenseenseenseeceeenseenssensseensenssenees 15  Chapter 4     The Multi Wire Myograph Unit                                           eere eene 23  4 1 Changing and adjusting the mounting supports                                  anna anna naaa 23  4 2 Calibration of the force transducer scavshecarecesancenasneascassennen TIR SUFRE ERE SOM REd POI REDIERE  26  4 3 Checking the force WS CCE ercsi ceirar a 28  4 4 Force Transducer ICD lac ele  Mes acaceuculetuecentaceasvessaunssanntenadebeneneensensvbocneneseesaaneubet 29  45 Myograph MaintenaN E ieee ane crceeene nde E ET 30  Chapter 5     Dissection  mounting and normalization                                             32  5 1 Dissection Protocol for Small Mesenteric Arteries                                     ereees 32  5 2 Mounting protocol TOF small GELT IOS        aina ada tUm E REPRE RE RERO d RS  36  NOrmal CIO INN gerne errr tome E anne na ate in dex US S IRE RM ETIN DIE 40  Chapter 6   Experimental protocols                                     eere eren e ener nnn nnn nnn 41  6 1 Summary of procedures   start to TIFIell                       x2 sd dorus rn balcatt anle iie muc diues 41   6 2 Example of step by step experimental procedure                              eene 43  OPERA A AA A o NNI FII IRR UM IMEEM IMMER M 45  DA BUHer SCI Sas cee nn E EA A A A 48  Appendix 1     Terms of warranty                         leere eere nenne ne
30. acting on the force    transducer     7    tate SI  cT etm g   sind     weigt  FL   fy  m  ias E Sind   PRESTA    The length of the pan arm  r    is 2 cm  the length of the transducer arm  r    is 4 cm  the  weight is 2 g  both angles  6  and 6    are 900 and the acceleration of gravity is 9 81m s    Therefore  the force acting on the force transducer is     p CIO Baud messsindo  4 cm sin9O0  F  oe claire Ts  Since   N is equal to 1kg m s   F  is equal to     F    9 81 mN   9 81 mN   1 0 g     Appendix 4 57    Appendix 5     Normalization theory    The importance of making a normalisation before initiating an experiment with any tubular  tissue segment is described in Chapter 6 3  In this appendix the mathematical rationale and  calculations underlying the normalisation procedure are described in detail     Mathematical Calculations     Let  X  Y   be the pair of values representing the micrometer reading and force reading re   spectively characterising each step in the normalisation procedure  YO is the force reading at  the start position of the normalisation procedure where the wires are just Separated and the  force reading is approximately zero  Then given that tension on the vessel is equal to force  divided by wall length  the wall tension at the i th micrometer reading is calculated by    T   Y  Y       28    a   a      where 0 is the microscope eyepiece reticule calibration factor in mm per division and a  and  a  are the vessel end points when measuring the length o
31. ales dmt dk sales dmt asia com sales dmt usa com    support dmt dk support dmt asia com support dmt usa com    
32. aplace relationship  Figure 5 14   The IC  is calculated from the IC455  by multiplying a factor to give an internal circumference at which the active force production  and optimal sensitivity to agonists of the vessel segment is maximal  For rat mesenteric ar   teries  this factor is 0 9  Both this factor and the transmural pressure have to be optimized  when other small resistance arteries are used  regardless of species  The normalized internal  diameter is calculated by dividing IC  with 7     Appendix 5 contains a complete description of the mathematical rationale and calculations  of the normalization procedure     Force  mNimm   Wal Tension T  mN mm   C        100 200 300 ao 500 600 700  00 900  Internal Circumference  gum     Time  min     Figure 5 13 Illustration of the step  Figure 5 14 Illustration of the exponential  wise normalization procedure curve fitting and determination of IC       40 Multi wire myograph system Model 620M   User manual    Chapter 6   Experimental protocols    This chapter contains detailed information regarding a standard reactivity experimental pro   tocol  A typical experimental protocol for vascular reactivity experiments involves equilibration  of the mounted vessel segment  a wake up protocol  a functional test of the mounted vessel  segment  and the actual experiment  which may involve a concentration response curve to  one or several vasopressors and or vasodilators  First  each component of an entire experi   mental protocol is disc
33. assive tension for the vessel segment  a vessel segment would  have to develop a stimulated tension equivalent to an effective active pressure above  13 3 kPa or 100 mm Hg  This is only an example  and it is up to the end user to deter   mine appropriate inclusion exclusion criteria for the vessels being used for their particu   lar experiments     The wake up protocol is performed after the vessel segment has been heated  equilibrated  and stretched with the appropriate passive tension  The present procedure is to be used on  rat mesenteric arteries  Another procedure may be needed for other animal species and tis   sue or vessel types     The wake up protocol consists of a series of stimuli and washout periods  The stimuli chosen  will depend on the mounted vessel  For example  noradrenaline  also known as norepine   phrine  is a good agonist to use to stimulate contractions in mesenteric resistance arter   ies  Methoxamine and cirazoline  a1 receptor agonists  are also good agonists to use in the  mesenteric resistance arteries  If using mouse or rat aorta for reactivity experiments  the o    Chapter6 441    receptor agonist  phenylephrine  would be a good choice     If other resistance arteries are used  these agonists may not stimulate for optimal contrac   tions  KPSS  high potassium physiological salt solution  should always be a part of the wake   up protocol  see Chapter 6 3 for recipe   The number of stimuli  the order  and length of time  for each stimulus may need
34. b data acquisition    system  optional     a Connection to  oxygen supply    BNC Cables    Vacuum pump   optional  IN                   eSeeeee  eR ARA eee ee Ba       Suction bottle  optional     Figure 2 1 The complete Multi Wire Myograph System   Model 620M  2 2 Setting up step by step    This chapter contains a complete step by step description of how to set up a complete myo   graph 620M system as illustrated in Figure 2 1    1  Interface   PC Connection   Data acquisition is possible either by connecting the Interface directly to a PC or through  a PowerLab data acquisition and analysis system  optional      l  Direct PC Connection   Connect the Interface to one of the COM ports on the PC using a serial cable  cable  not included      Il  PowerLab  Optional    Connect the Interface to the PowerLab unit using BNC cables  Connect Rec 1 on the  Interface to Input 1 on the PowerLab  Rec 2 to Input 2 etc   Connect the PowerLab unit to one of the USB ports on the PC using the USB cable  delivered with the PowerLab system  Follow the  DMT Quick Start Guide  instructions  delivered with the PowerLab system to install the PowerLab driver and Chart software  on the PC     Chapter 2 13    2  Oxygen Supply    Connect the gas supply  95  O5  5  CO  or 21  O5  5  CO   balance N   with tubing run   ning from the gas supply to the gas inlet on the back of the Interface  Oxygen is supplied  to the chambers by tubing attached to the stainless steel vacuum pipe  The oxygen and  vacuum tub
35. back quickly  whereas in the  vein  the blood will run back very slowly  if at all  It is important that you perform this  on vessels other than those you will use  as this procedure can damage the vessels     8  Dissect away the vein using scissors to cut the adipose and connective tissue between  the artery and vein  One method is to cut the vein in one position and afterwards  gently  pull the vein away from the artery  In this way  a fine membrane of connective tissue  becomes visible between the adipose tissue and the artery  Carefully cut the fine mem   brane to remove the vein and adipose tissue while avoiding any direct contact between  the scissor and artery  Figure 5 4A B         Figure 5 4  A B Removal of vein    34 Multi wire myograph system Model 620M   User manual       Figure 5 5 A  B and C Cutting free the artery to be studied    9  Clean the artery by removing any remaining adipose or connective tissue  Gently pull  away adipose or connective tissue to make the connective tissue membrane more vis   ible  Cut the membrane to remove the tissue visible  Cut the membrane to remove the  tissue     10  Cut the distal end of the artery section to be investigated  Afterwards cut the proximal  end while ensuring that the vessel segment has the correct length  Figure 5 5 A C      Chapter5 35    5 2 Mounting protocol for small arteries    Mounting small arteries involves threading wires through the lumen of the vessel and secur   ing the wires in the lumen to the jaws s
36. een activated with the push buttons on the  front panel of the interface     Pressing  ALL  will cause all the lines to turn blue   meaning all chambers can be modified at the same  time  Again  the up and down arrow keys can be used  to modify the length of time the vacuum valves stay  open     Pressing  ENTER  after modifying the value s  for  valve delay will lock in the number s  and be retained  in memory every time the system is turned on    20 Multi wire myograph system Model 620M   User manual    CHAMBER 1 CALIBRATION    Step no   Ovog    The transducer is now calibrated    Force read out should be 9 81 mN    0 1 mN  If OK go to next step   Otherwise  repeat the calibration     Force Chamber 1   9 81 mN    NEXT STEP    EMPTY VALVES DELAY          EMPTY VALVES DELAY    Chamber 1  5 Sec     Chamber 2  6 Sec     Chamber 3  6 Sec     Chamber 4  5 Sec        EMPTY VALVES DELAY    Chamber 1  5 Sec     Chamber 2  6 Sec     Chamber 3  6 Sec     Chamber 4     5 Sec     3  FORCE REC  OUT     The FORCE RECORDING OUTPUT  or FORCE REC   OUT  sub menu determines the upper limit for force  sent from the BNC analogue output connectors  This  will only affect the data collected from the interface  to a data acquisition system such as AD Instruments  PowerLab and LabChart software  The factory default  setting for FORCE REC  OUT is 20 mN  meaning that  if the force of the mounted vessel exceeds 20 mN   the force recorded in the data acquisition software  will not record more than 20 mN
37. es and fill in the values in the Ocular Calibration Sheet  Repeat the procedure  twice     52 Multi wire myograph system Model 620M   User manual    Ocular Calibration Sheet     Date   Operators    Microscope  type and       Ocular type        Stage Micrometer Reticule Scale    Lengan Longin    1    3       1      oso w   m KI RI  ea   mm E vf as  om  3            s      p ee ee Lm    mm  mm    Calculations  1  Calculate the length between the two positions on the stage micrometer by subtracting  the value of position 1 from the value of position 2  Multiply the length in divisions with    the length of each division to get the length in mm     2  Calculate the length between the two positions on the reticule scale by subtracting the  value of position 1 from the value of position 2     3  Calculate the mean length value of both the stage micrometer and the reticular scale   4  Calculate the conversion factor   Mean Stage Micrometer Length  mm     Mean Reticule Scale  div     Appendix2 53    Appendix 3     Reading a millimetre micrometer    Sleeve scale Thimble scale  e  e      Lini beni ee       Figure A3 1 Overview of the micrometer parts  actual reading 20000 um   20 mm     Sleeve scale    The micrometer sleeve scale has a total length of 25 mm divided into 50 equal parts  Each  part of a division above the horizontal line represents 1 mm  where each 5  line is marked by  a longer line and a number designating the length in mm  Each division below the horizontal  line is p
38. es and spare parts  and  system specifications     About this manual    9    Unpacking the myograph system    Take a few minutes to carefully inspect your new 620M multi channel wire myograph system  for damage which may have occurred during handling and shipping  If you suspect any kind of  damage  please contact DMT immediately and the matter will be pursued soon as possible   If the packing material appears damaged  please retain it until a possible claim has been  settled     We recommend that you store the packing material for any possible future transport of the  Wire Myograph System  In case of transport and the original packing material is unavailable   please contact DMT Sales Department for advice and packing instructions     After unpacking your new Wire Myograph System  please use the following list to check that  the system is complete     e 1 interface unit   e 4 chamber units with mounted stainless steel jaws   e 4 set mounting support pins  200um    e 4 chamber covers   e 1 external temperature probe   e    1 power cord    e 1 calibration kit  including  bridge    balance  and 2 gram weight    e 4 plastic funnels   e 1 roll of 40 um stainless steel wire   e 1tube of high vacuum grease   e 1tube of grease for linear slides   e 5 spare screws for mounting of jaws   e 3 Allen keys   e 1 small screwdriver   e 1 Multi Wire Myograph System 620M User Manual   e 1 manual by Professor M  J  Mulvany      Procedures for Investigation of Small  Vessels using small vessel 
39. es with the  Grease for Linear Slides  included with  your system     Cleaning the myograph  DMT strongly recommends that the myograph chambers and surrounding areas are cleaned  after each experiment     At the end of each experiment  use the following procedure to clean the myograph chambers  and supports     1  Fill the myograph chamber to the edge with an 8  acetic acid solution and allow it to  work for a few minutes to dissolve calcium deposits and other salt build up  Use a cotton   tipped applicator to mechanically clean all chamber surfaces     2  Remove the acetic acid and wash the myograph chamber and supports several times  with double distilled water     3  If any kind of hydrophobic reagents have been used which might be difficult to remove    using steps 1  and 2   then try incubating the chamber and supports with 96  ethanol  or a weak detergent solution  i e  0 1  triton 100      30 Multi wire myograph system Model 620M   User manual    4  To remove more resistant or toxic chemicals  incubate the myograph chamber and sup     9     6     ports with 1M HCI for up to 1 hour  In exceptional cases  incubate the chamber and sup   ports with no stronger than a 3M HNOs5 solution for about 15 minutes     Wash the myograph chamber and supports several times with double distilled water   If acids such as 1M HCI and 3M HNOs are used to clean the chambers  make sure ALL    surfaces are thoroughly dried after copious washes with double distilled water  Any re   sidual acid
40. euptake mechanism that removes noradrenaline from the neuromuscular junction   The reuptake mechanism will create a concentration gradient between the solution around  the vessel segment and the receptors on the smooth muscle  To correctly determine the  sensitivity of noradrenaline in this vascular type  it is necessary to eliminate this concentration  gradient by performing the experiment in the presence of cocaine to block the noradrenaline  reuptake     To determine the sensitivity of noradrenaline  cumulative concentrations of noradrenaline  are added to the the vessel segment  Each concentration is applied until a steady response  has been reached  a plateau   The next concentration is added once the plateau is reached     When the vessel segment is fully contracted and does not continue to respond after further  increases in the noradrenaline concentration  the experiment is over     Protocol  Prepare the following stock solutions   e Noradrenaline  109 105  104  103  107 M  e Propranolol  103 M  e Cocaine  103 M  1  Mount the vessels as described in Chapter 5 2   2  Perform the normalization as described in Chapter 5 3 and Appendix 5     3  Perform the wake up protocol as described in Chapter 6 1     4  After the final wash  incubate the 1 or 2 vessel segments in 1uM propranolol and 3uM  cocaine for at least 10 minutes  Incubate the other remaining vessels in PSS alone     5  Following Table 6 1 below  add increasing concentrations of noradrenaline into the bath   Wai
41. f the mounted vessel segment     The internal circumference of the mounted vessel at the ith reading is calculated by   C  IG ALIA em     where IC  is the internal circumference of the mounted vessel when the wires are just sepa   rated and is given by     I    2 tn    d  where d is the wire diameter  For 40 um wires  IC    205 6 um     Using the Laplace relation  the effective pressure Pi is calculated for each pair of readings   The effective pressure is an estimate of the internal pressure  which is necessary to extend  the vessel to the measured internal circumference     T     ce     The stepwise distension is continued until the calculated effective pressure exceeds the tar   get transmural pressure  The target value needs to be optimized for the individual tissue  preparation  optimal active force as determined by the length tension relationship for that  tissue   For rat mesenteric arteries  the target transmural pressure is normally 100 mmHg   13 3 kPa            IC  T   100 mmg      27    100 mmHg    58 Multi wire myograph system Model 620M   User manual    An exponential curve is fitted to the internal circumference pressure data as illustrated in  Figure 5 14 in Chapter 5  Now the isobar corresponding to 100 mmHg is used to calculate  the IC100 value from the point of interception between the function of the exponential curve  and the function of the 100 mmHg isobar     The normalized internal circumference IC  is calculated by multiplying the internal circumfer  
42. fully pull the vessel along the wire   Figure 5 7B C   Be careful not to stretch the vessel seg   ment if the end of the wire catches the vessel wall     Step 3  Securing the first wire and mounted vessel    Once the vessel segment is threaded onto the wire  grab  the free end of the wire  nearest you  with the forceps and  move the jaws apart     While controlling the movement of the wire with one set  of forceps  use another pair of forceps to gently pull the  vessel segment along the wire until the area of interest  IS positioned in the gap between the jaws  The near end  of the vessel segment should sit about 0 1 mm inside  the jaw gap so that the vessel is not pinched by the wire  against the jaw face on the side closest to the user  Fig   ure 5 8 A      While still holding the free end of the wire with the forceps   move the jaws together to clamp the wire  Secure the wire  under the near fixing screw on the micrometer side jaw by  bending the wire around the screw in a clockwise direc   tion  Tightening the screw will also tighten the wire if the  wire is mounted in this way  Figure 5 8B      dh d dh    Figure 5 7 A  Band C  Mounting step 2       Figure 5 8 Aand B  Mounting step 3    Chapter5 37    Step 4  Removing excess vessel    e Using forceps  gently rub the vessel segment on the far  side of the jaw to separate any excess vessel segment  from the area of interest clamped in the gap between the  jaws  Do this in the area indicated by the tip of the forceps  in Fig
43. g internal diameter  Continuous use  and repeated greasing of the transducer arm holes will cause some misalignment of the  mounting supports  The mounting supports  therefore  whether they are the jaws for wires or  the pins  will need occasional adjustments     Changing and adjustment of the supports is performed using the following step by step pro   cedure     NOTE  THE TRANSDUCERS ARE FRAGILE AND SENSITIVE TO MECHANICAL STRAIN  BE VERY  CAREFUL WHEN CHANGING OR ADJUSTING THE MOUNTING SUPPORTS     Changing the supports  Figure 4 1     1  Use the micrometer to separate the supports as far apart as possible    2  Use the small screwdriver provided to gently loosen screw D on the support attached on  the transducer side using the small screwdriver  Screw D is the screw on the transducer   side support closest to the transducer    3  Gently pull the support away from the transducer pin    4  Loosen screw B on the micrometer side with the appropriate fitting allen key    5  Pull the support away  Note  Number the supports with the chamber number they were  removed from using some kind of permanent marker  Store the supports in the provided    plastic case  Numbering the supports will save time when the supports are changed  again  limiting the amount of adjustments needed after each change              o        o      ee      ee    AB CD    Figure 4 1 Myograph unit   screws for changing supports and coarse adjustment of the jaws    Chapter 4 23    Course adjusting the jaws for
44. h the EMC standards  EN 61326 2 6 2005  EMC 89 336 EEC  EN 61000 3 2    Certified with the safety standards  EN 61010 1 2001  Directive 2006 95 EC  EN 61010 1 Corr 1 2003  EN 61010 1 Corr 1 2003  EN 61010 2 101 2003    EMC EMI 7    Certificate of Conformity    DMT A S  Skejbyparken 152  8200 Aarhus N   Denmark   hereby declares its responsibility that the following product     Multi Wire Myograph System  Model 620M    is covered by this certificate and marked with CE label conforms  with the following standards     EN 61010 1 2001 Safety requirements for electrical equipment for meas   EN 61010 1 Corr 1 2003 urement  control  and laboratory use      EN 61010 1 Corr 1 2003 Part 1  General requirements     EN 61010 2 101 2003 safety requirements for electrical equipment for meas   urement  control and laboratory use   Part 2   101  Par   ticular requirements for in vitro diagnostic  IVD  medical  equipment     EN 61326 2 6 2005 Electrical equipment for measurement  control and labo   ratory use   EMC  Requirements   Part 2 6  Particular requirements In vitro  diagnostic  IVD  medical equipment     With reference to regulations in the following directives   2006 95 EC  39  S356  EEC        8 Multi wire myograph system Model 620M   User manual    About this manual    This manual contains a complete list of procedures that describe how to install  maintain and  use the Multi Wire Myograph System   Model 620M     Chapter 1 provides an overview of the construction and basic features
45. hapter5 39    5 3 Normalization  Normalizing the vascular preparation is important for 3 reasons     1  Experiments with elastic preparations like vessels can only have meaning if they are per   formed under conditions where the size is clearly defined     2  Clearly defined conditions are required in pharmacological experiments  as the sensitiv   ity of preparations to agonists and antagonists is dependent on the amount of stretch     3  The active response of a preparation is dependent on the extent of stretch  which makes it  important to set the preparation to an internal circumference giving maximal response     The aim of the normalization procedure is to stretch the segment to a normalized internal  circumference  IC4  defined as a set fraction of the internal circumference  IC455  that a fully  relaxed segment would have at a specified transmural pressure  For small rat arteries  the  target transmural pressure is typically 100 mm Hg  13 3 kPa      Principles of the normalization procedure    In practice  the normalization is performed by stretching the segment stepwise and measur   ing the micrometer and force readings  Figure 5 13  steps 1 4   These data are converted into  values of internal circumference  um  and wall tension T  mN mm   respectively     Plotting wall tension against internal circumference  results in an exponential curve  By ap   plying the isobar curve corresponding to 100 mm Hg  the IC4o5 is calculated from the point of  intersection using the L
46. he myograph chamber and supports several times with double distilled water       Dry the surfaces using absorbent paper  i e  Kim Wipes  or cotton tip applicators     IMPORTANT NOTES     In exceptional cases  the supports  jaws or pins  may need to be removed from the myo   graph chamber and cleaned individually to assure proper cleaning of all support sur   faces  NEVER SOAK THE SUPPORTS IN ANYTHING STRONGER THAN 8  ACETIC ACID FOR  EXTENDED PERIODS OF TIME  i e  several hours or overnight      Chapter 4    31    Chapter 5     Dissection  mounting and normalization    This chapter contains some practical protocols and procedures that describe how to quickly  get started using the Multi Wire Myograph 620M System  This chapter is aimed toward the  novice myograph user  however it may contain some hints that also will be useful for the ex   perienced myograph user     5 1 Dissection Protocol for Small Mesenteric Arteries    The Multi Wire Myograph 620M system is versatile in that a large variety of physiological and  pharmacological studies using ring preparations of varying sizes from different species can  be performed  This section describes the dissection of mesenteric resistance arteries as an  example of how to get started with mounting and measuring function from small resistance  arteries     1     Euthanize the laboratory animal in accordance to the local and national laws and regula   tions  A midline laparotomy is performed to expose the mesenteric bed       Use scis
47. he wire  pointing towards the user  Figure 5 6 B C         UJ    e Fill Myograph Chamber with room temperature PSS  see  recipe in Chapter 6 3      Step 2  Threading the vessel onto the wire    e Grab the vessel with forceps in a way as not to damage it   The best way to do this is to grab either the very end of the  vessel where it was cut  either end  or grab a very small  part of the adventitia on the outside of the vessel with   out pulling the vessel too hard  Pulling too roughly on the  adventitia will cause damage to the vessel  While holding C  the vessel  try to slide the vessel onto the wire through  the lumen  If the lumen is closed  try one of the following  techniques        Figure 5 6 A  Band C  Mounting step 1    36 Multi wire myograph system Model 620M   User manual    1  Use the wire to gently push the lumen open  blood  streaming out is a very good sign     2  Hold the vessel on the very outside edge of the closed  lumen of the vessel  With another pair of forceps  grab  the very outside edge of the vessel directly across  from the first pair of forceps and gently pull to open  the lumen     3  Hold the vessel about 3 mm from the closed end with  a set of forceps and use another pair of forceps to  squeeze the blood remaining in the lumen out through  the closed end     CAUTION  these techniques may increase the risk of  damaging the vessel and should be done very care   fully     Pull one  opened end of the vessel segment so the wire  enters the lumen  Care
48. ing need to be inserted into the chamber in order to aerate the heated buffer   Needle valves on the back of the interface can be adjusted to regulate the amount of  bubbling that occurs  Turning the regulator clockwise increases the bubbling while turn   ing it counter clockwise decreases the bubbling  Each regulator has a lock device at   tached that can be used when the desired bubbling is achieved  Figure 2 2     NOTE  The needle valves need to be greased  using the grease for the linear slides  and  turned at regular intervals to prevent them from sticking or permanently freezing     3  Vacuum Connection    The system has a built in manifold with separate valves that allows each chamber to be  drained individually  After connecting the vacuum source at the back of the Interface   the vacuum pipes need to be inserted into the chambers in order for this feature to work  properly  The pipes are inserted into the chamber by gently pulling up on the curved part  of the pipe  turning it 90   counter  clockwise and gently lowering it into the chamber  A  chamber can then be emptied by pressing the corresponding numbered button  Pressing  the    all    button will empty all the chambers at the same time  Figure 2 2     NOTE  When draining the chambers using the automatic vacuum function  press the  appropriate button for an additional 3 5 seconds after the initial emptying  This will help  drain residual buffer and solutions retained in the tubing and valves     e Funnel    e Te
49. laced between each 1 mm mark  scale above the horizontal line  and represents 0 5  mm     Thimble scale    The thimble is divided into 5O equal parts  and one complete rotation of the thimble is in   dicated by the smallest division on the sleeve  which equals 0 5 mm  Each division on the  thimble scale is 10 um  If the thimble scale falls between two lines  then a number between  O and 10 um must be approximated     Example 1    1  Note that the thimble has stopped at a point beyond    10    on the sleeve indicating 10000  um  10 mm      2  Note that there is no mark completely visible between the 10 mm mark and the thim   ble     3  Read the value on the thimble corresponding to the intersection with the horizontal line  on the sleeve     A  Reading on sleeve  10000 um   B  No additional mark visible  O um     C  Thimble reading  380 um hs iR one  Total reading   10380 um       Figure A3 2 Example 1   reading   10380 um    54 Multi wire myograph system Model 620M   User manual    Example 2    1  Note that the thimble has stopped at a point beyond    16    on the sleeve indicating 16000  um  16 mm      2  Note that this time a mark is visible between the 16 mm mark and the thimble indication  500 um     3  Read the value on the thimble corresponding to the intersection with the horizontal line  on the sleeve     A  Reading on sleeve  16000 um  B  One additional mark visible  500 um  C  Thimble reading  280 um   Total reading  16780 um       Figure A3 3 Example 2   reading   
50. le is to use the eyepiece reticule  typically consisting of 50 100 divisions  to  measure the length of an object in terms of reticular divisions spanned by the object  Having  the conversion factor specific for the eyepiece reticule and used magnification  the length of  the object in mm is easily calculated     All reticules need to be calibrated in order to determine the specific conversion factor for a  particular eyepiece reticule and the used magnification  A stage micrometer will be needed  to do this  A stage micrometer is simply a microscope glass slide having a scale engraved on  the surface  A typical micrometer scale is 2 00 mm long engraved with divisions of 0 01 mm   equalling 10um per division  However  micrometer glass slides with finer divisions can also  useful for calibrating a stereomicroscope to be used with the wire myograph     Ocular Calibration Procedure    1  Decide which microscope magnification is to be used for the segment length measure   ments  Use the largest possible fixed magnification where the eyepiece reticule scale still  covers the whole gap of the myograph jaws     2  Place the stage micrometer on the microscope stage and focus on it  Fit one of the divi   sion lines on the stage micrometer to one of the division lines of the reticule scale  While  keeping the stage micrometer fixed on the microscope stage  find another position on  both scales where the division lines are aligned  Read the position of the two  fit points   on both scal
51. list  all 4 chambers for calibration     To begin the calibration  press    SELECT    for the  chamber which calibration will be performed on  The  text for the chamber to be calibrated will turn blue   Pressing  ENTER  will enter the 6 step procedure  for calibrating the force transducer on the desired  chamber     The calibration procedure is listed in 6 individual  steps and needs to be performed for each channel  or transducer when calibrating the system  Step 1 in   volves setting up the chamber for calibration  Make  sure the chamber contains the pins or jaws  depend   ing on the type of vessel being studied  If jaws are  being used for smaller vessels  a wire needs to be  strung on the transducer side jaw for the calibra   tion  Fill the chamber with double distilled water for  the volume to be used experimentally  Press  NEXT  STEP        Step 2 involves setting up the calibration kit ap   propriately for the actual weight calibration  Verify  that the transducer arm pin does not touch the  mounting wire on the jaw or the mounting pin for  larger ves sels  as instructed  The pin should be as  close as possible to the mounting wire or mounting  pin without touching in order to get the most accu   rate calibration  Press    NEXT STEP    when the cali   bration kit has been properly placed     18 Multi wire myograph system Model 620M   User manual    FORCE CALIBRATION    Force calibration chamber 1     Force calibration chamber 2     Force calibration chamber 3     Force
52. mber cover DMT Item   100064       40 mm funnels DMT Item   100082    Appendix9 63    Appendix 9     System Specifications    Tissue size   Chamber     Chamber material     Chamber volume   Chamber suction   Chamber gassing   Chamber cover   Force range   Force resolution   Micropositioners     Weight calibration     Heating    Temp  range   Temp  resolution   Temp  probe   Output reading   Analogue output   Digital output   Voltage      gt 60um    gt 450um up to 10 mm   Four individual chambers   Acid resistant stainless steel   Max  8ml   Manual or automatic  time controlled  user defined  Individually controlled per chamber by needle valves  Supplied with connections for gassing   User selectable at     200  400  800  1600mN  O 01mN   Manually operated   Semi Automatic   Built in   Ambient temp    50  C   0 1   C   External   Force  mN    2 5V F S    Serial interface   RS232   RS485   100 to 240VAC  auto  50 60Hz    64 Multi wire myograph system Model 620M   User manual    Notes    65    DANISH MYO TECHNOL OG    DMT A S DMT Asia DMT USA  Inc   Skejby Science Center Rm 28C Liang Feng Building 1201 Peachtree Street  Skejbyparken 152 No  8 Dong Fang Road 400 Colony Square  DK 8200 Aarhus N Lu Jia Zui Financial District Suite 200 630  Denmark Shanghai 200120 Atlanta  GA 30361   P R  China USA  Tel    45 87 41 11 00 Tel   86  0  21 6486 9685 Tel    1 770 612 8014  Fax   45 87 41 11 01 Fax   86  0  21 5877 0063 Fax   1 678 302 7013  www dmt dk www dmt asia com www dmt usa com  s
53. mperature probe    e For drug application       Figure 2 2 Suction connection Figure 2 3 Chamber cover    4  Chamber Covers  The chamber covers will help maintain the temperature and other buffer conditions  gas  tension  pH  fairly constant  Holes in the chamber covers serve different purposes  and  they are illustrated in Figure 2 3  The slots allow the covers to be placed over the cham   ber around the support arms and gas vacuum tubes     2 3 The first weight calibration    Prior to the shipment of the Multi Wire Myograph 620M System  it has gone through two days  of continuous testing  including a final weight calibration  However  DMT recommends that  a new weight calibration is performed before using the myograph system for the first time   The weight calibration procedure is described in detail in the FORCE CALIBRATION sub menu  under SETTINGS  as explained in Chapter 3     14 Multi wire myograph system Model 620M   User manual    Chapter 3     The Interface Menus    Chapter 3 is a complete manual for the 620M Interface  The chapter contains a detailed  description of how to navigate the touch screen menus and how to use the special features    of the 620M myograph     Menus on the 620M interface are all accessible by  a touch screen  To access a menu  simply touch the  screen to access a menu  When a setting needs to  be changed  the setting can be changed by pressing  the    SELECT    icon on the touch screen correspond   ing to the desired channel to be changed    
54. myograph    e 1 CD  Dissection and mounting of small vessels in wire myographs      The shape of the AC plug varies by country  be sure that the plug fits the outlets for your  location     10 Multi wire myograph system Model 620M   User manual    Chapter 1     System overview    1 1 Interface Front Panel    Y x      Xx    Um      iX                           Myo Interface display e      Heat indicator   Valve buttons      Power indicator    Figure 1 1 Interface Front Panel    1 2 Interface Rear Panel    4 Transducer connections       Gas regulator  needle valve  3 RS 232 Port for serial connection to PC          Gas input 3   Temperature probe            ij Oo C O       e  Power connector         e 4 Recorder outputs            Vacuum input ON OFF switch USB output    Figure 1 2 Interface Rear Panel    Chapter 1 11    1 3 Multi wire myograph unit    Allen screws for fine alignment e   Connection to Myo Interface      9    of the myograph jaws    Micropositioner eco eR         Myograph jaw connected to micropositioner e    Myograph jaw connected to force transducer e       Figure 1 4 Close up of myograph jaws             Figure 1 5 Mounting jaws for small vessels    Figure 1 6 Mounting pins for larger vessels    12 Multi wire myograph system Model 620M   User manual    Chapter 2     Setting up    2 1 The Complete Myograph 620M System    DMT CS200 Pulse Train PC data acquisition and  Stimulator  optional  analysis software  optional     Myo Interface front panel        PowerLa
55. n accordance with the manufacturer s  instructions     e Do not install near any heat sources such as radiators  heat registers  stoves  or other  equipment or devices that produce heat     e Only use attachments and accessories specified by the manufacturer    e Unplug this apparatus during lightning storms or when unused for long periods of time    e This apparatus must be grounded      Usea three wire grounding type cord similar to the one supplied with the product    e Do not modify the polarized or grounding type plug  A polarized plug has two flat blades   one being wider than the other  A grounding type plug has two blades and a third  round   grounding pin  The wide blade or the third prong is provided for your safety  If the provid   ed plug does not fit into your outlet  consult an electrician for replacement of the obsolete  outlet    e Be advised that different operating voltages require the use of different types of line cord    and attachment plugs  Check the voltage in your area and use the correct type  See the  table below     Voltage Line plug according to standard  110 125 V UL81 and CSA C22 2 No  42    220 230 V CEE 7 page VII  SR section 107 2 D1 IEC 83  page C4  DAN BS 1363 of 1984  Specification for 13A fused plugs and  switched and unswitched socket outlets   Protect the power cord from being walked on or pinched  particularly at power outlets and the  point where they connect to the apparatus        Refer all servicing to qualified service personnel  
56. na n unu a nea au aaa nana nna unas 51  Appendix 2     Calibration of Eyepiece Reticule                                         eus 52  Appendix 3     Reading a millimetre micrometer                                           eese 54  Appendix 4     Principles of weight calibration                                           eeu 56  Appendix 5     Normalization theory                               eee aaa aaa aaa nana naaa naaa 58  Appendix 6     Service Check                        1  11e  e eere een ee eene nen anna nun aa aaa nana 60  Appendix 7     Shipping Instructions                                      eere eren enn nnn nnn 61  Appendix 8     Myograph accessories and spare parts                                             62  Appendix 9     System Specifications                                1   ee cee rere rene nennen nnn nnns 64  NS M                                               65    Multi wire myograph system Model 620M   User manual    Introduction    Until the mid 1970s most of the information about the mechanical  morphological and phar   macological properties of vascular smooth muscle were only obtainable from studies on rela   tively large vessels  At that time rat tail arteries were the smallest vessels to be investigated in  detail due to limitations in the available in vitro techniques  For example  studies measuring  the contraction force were routinely performed with only one of the mounting wires secured   Futhermore  relatively large wires  100 200 um 
57. nta  GA 30361  USA    Tel    1 770 612 8014  Fax   1 678 302 7013    www dmt usa com    sales dmt usa com  Support dmt usa com    Trademarks    3    4    Contents    ATQOE SS ire EXRMInS NMESESEExEFREMIMUNAMENMCC DUCK ENMENUNNENNIEEENNENNSNENINEEEVNEENNEAKENENNANNNNNFEKUENI NER 3  Toitgsre lU a UO A                      n 5  cria e                                                  M                            6  3 142  M T                                                      use 7  PAD FOV IS RR c                                        EEE 4  Certificate of ConforMity             ssessessuasuas ann un nana naaa anna n nena nenau man naaa naaa naaa aua mauu mua 8  About this malludluissssiwsskexeua 2E ENNN M FENN newt MA SV ENNHMEKMNNM EXE YIN EE EN NENEE  EUMEEKYS TEE ux KNEES RN RAN DUE 9  Unpacking the myograph system                             eere eere eene ener n nura nana naaa 10  Chapter 1     System overview                           eeeeee aaa eene nnn nne nn unnn nnn nana ana naaa 11  ELME CE FONT PIN A trt A MIETEN onu En A OI UE 11  1 2 Interaee Rear Pane acusa E ront RISQUE RI Sn ITANE REN UNE 11  Lo  NMUR wire mMyocraph UN en EA A en dA sua E Des 12  Chapter 2     Setting UD viscceennnnesciawecsevnnensenathuswenawusenawawersenwesnbiasusemsdensnnensennenaensnswamneteans 13  2 1 The Complete Myograph 620M System              ssseeeeee naaa naaa manamana nana Raman aan naaa aa 13  Fh DONS UD SOP Dy en ete acs se vate incre ets nce nce Ra T REOS 13  ANN a Oe O aa aa
58. o that isometric contractions can be measured one  the wires are secured to the jaws  The force transducers of the Multi Wire Myograph 620M  system have a sensitivity of 0 01 mN   1 mg  and can be damaged if the applied force ex   ceeds around 1 N   100 g      IMPORTANT  DO NOT PRESS THE JAWS TO HARD AGAINST EACH OTHER  A DISTANCE OF  20  um AFTER THE JAWS JUST TOUCH PRODUCES SUFFICIENT FORCE TO CLAMP THE WIRES  BETWEEN THE JAWS  AS DESCRIBED BELOW  IN ADDITION  DO NOT USE TOO MUCH TORQUE  FORCE TO TIGHTEN THE SCREWS  WHEN TIGHTENING THE SCREWS ON THE WINGS OF THE  JAWS TO SECURE THE WIRE  STOP TURNING WHEN THE SCREW STOPS MOVING  THIS CAN  BE DONE EASILY BY ONLY USING THE THUMB AND INDEX FINGER ON THE SUPPLIED SCREW   DRIVER TO TIGHTEN THE SCREWS     Step 1  Securing the first wire    e Cut lengths of wire  40 um stainless steel or 25 um tung   sten  about 2 2 cm long     e Holding the wire at the far end  place the centre of the  wire between the jaws and bring the jaws together so that  the wire is clamped  Figure 5 6 A   Again  do not press the  jaws against each other too hard  After the jaws just touch  each other  the micrometer should not be turned more  than 20 um to hold the wire     e Secure the wire on the micrometer side jaw first by bend   ing the far end of the wire towards the left  and wrap it  around the screw in a clockwise direction  Tightening the  screw will the aid in tightening the wire  The wire will be  clamped between the jaws with the near end of t
59. ore  care must be taken when preparing the equipment for  shipping  DMT recommends that each part of the myograph system be wrapped individually   i e  with bubble wrap  and placed together in a large box  preferably the box the myograph  system was originally shipped in   Place the wrapped items in the middle of the box and fill  the box with packing peanuts or packing chips     IMPORTANT   Before closing the box  make sure that none of the enclosed items are loose  as transport by  road or air can be quite rough     Address the box to  Attn  Henrik Pedersen  DMT A S  Skejby Science Center  Skejbyparken 152  DK 8200 Aarhus N  Denmark    Phone   45 87 41 11 O0    Make sure that all four sides of the box are marked  fragile   Indicate on the top of the box  that it contains goods returned for repair service     Customers outside the EC must also enclose a pro forma invoice stating that the box contains  goods being returned for repair or service     If arranging transportation through a courier  please keep in mind the high value of the myo   graph system and that standard insurance provided by the courier  in most cases  is insuf   ficient to cover damage or loss of the myograph system  In most cases  additional insurance  coverage will be needed     Appendix 7 61    Appendix 8     Myograph accessories and spare parts    This appendix contains a complete list of equipment needed to set up a Multi Wire Myograph  System  In addition  the appendix contains a list of special Myog
60. ounting with vessel segments from the first or  second branch from the superior mesenteric artery  approximate internal diameter 200   300um in the rat        Use high quality forceps and dissection scissors to dissect the vessel segment of in     terest  Start cutting through the mesenteric membrane along both sides of the vessel   about 1 2 mm from the vessel  Avoid accidentally cutting the artery by cutting along the  length of the vessels and never perpendicular to them  Figure 5 2 A B        Cut away as much adipose tissue as needed around the vessels to distinguish between    the artery and vein  The artery can easily be identified by the following characteristics   Figure 5 3      32 Multi wire myograph system Model 620M   User manual          Second Branch    First Branch    Figure 5 1 Branching of the mesenteric arteries       Figure 5 3 Distinguishing between artery and vein    Chapter5 33    e The branch points of arteries are V shaped whereas those of veins are more U   shaped    e The arterial wall contains a thick layer of smooth muscle cells compared to the vein  wall  which only contains a single or a few layers of smooth muscle cells  The morpho   logical differences should be clearly visible under the stereomicroscope     e If you still have difficulty identifying the artery from the vein  and the vein and artery  still contain some blood  then try moving the blood forward by very gently squeezing  the vessels with forceps  In the artery  the blood will run 
61. p replacement  procedure carefully     1  Remove the pin or jaw from the transducer pin coming out of the transducer house   2  Disconnect the Myograph Chamber from the Interface     3  Turn the Myograph Chamber upside down and remove the transducer housing by loosen   ing the two screws  A B  as illustrated in Figure 4 9        Figure 4 9   The 2 screws that secure the transducer house to the chamber    4  The replacement transducer will be shipped with the new transducer inside a new trans   ducer house     5  Place a small amount of vacuum grease  clear or whitish grease  around the bottom of  the transducer housing to seal the transducer housing when put back in place     6  Carefully realign the transducer housing with the new transducer on the Myograph Cham   ber and reinsert the allen screws through the bottom of the Myograph Chamber     f  Tighten the screws and place some vacuum grease around the transducer pin that pro   trudes from the transducer housing  Make sure that the hole is completely sealed to  prevent buffer solution or water from entering the transducer housing and damaging the  new force transducer     IMPORTANT NOTE     Calibrate the new force transducer before performing a new experiment  as described in  Chapter 3 and 4 2     Chapter 4 29    4 5 Myograph Maintenance    The Multi Wire Myograph System Model 620M is a very delicate and sophisticated piece of  research equipment  DMT recommends that the following sections are read carefully and  that the
62. r  the menu and allow the user to change the default  system temperature  as well as turn the heat on or  off  Pressing    DEFAULT    will automatically reset the  temperature setpoint to 37  C  Manually change the  temperature by pressing the up or down arrows     To turn the heat on  touch    ON    and the    ON    icon  will turn green  indicating the heat has been turned  on  The system will heat to the designated tempera   ture setpoint  Pressing  ENTER  will send the user  back to the    ACTUAL FORCE    display     Settings Menu    The    Settings Menu    contains several sub menus  that can be accessed to change functional aspects  of the interface  These sub menus include       FORCE CALIBRATION     VALVE DELAY     FORCE REC  OUTPUT     MEASUREMENT RANGE    INTERFACE SETTINGS    O1    DN P    SET FORCE TO ZERO    Force chamber 1   2 36 mN    Force chamber 2   0 26 mN    Force chamber 3   1 06 mN    Force chamber 4   0 76 mN    SET CHAMBER TEMPERATURE    Temperature setpoint  27 0   C    Probe temperature  36 6   C       cue CY I  SET CHAMBER TEMPERATURE       Temperature setpoint  37 0   C    Probe temperature  36 6   C    SETTINGS       FORCE VALVE FORCE REC  MEASUREMENT  CALIBRATION DELAY OUTPUT RANGE    INTERFACE  SETTINGS       Chapter 3 17    1  FORCE CALIBRATION     Entering the FORCE CALIBRATION sub menu begins  the transducer calibration procedure  Begin the  calibration procedure by pressing    FORCE CALIBRA   TION    to enter the sub menu  The sub menu will 
63. raph 620M accessories and  Spare parts     Other than focusing on development and manufacturing  DMT has specialized in offering our  customers first class laboratory equipment needed for a Multi Wire Myograph set up at very  competitive prices  Please contact the DMT Sales Department for further product informa   tion and prices     General myograph equipment    This section contains a complete list of laboratory equipment needed when setting up a Multi  Wire Myograph system     e Dissection stereo microscope DMT Item   100072  DMT recommends the Zeiss Stemi 2000 Stereo Microscope which includes ocular mi   crometer and stage micrometer      Mounting forceps DMT Item   100081  DMT recommends Dumont Medical No 5 tip  0 10 mm x 0 06 mm        Dissection scissors DMT Item   100073  DMT recommends Geuder G 19745 8 cm straight trabeculum       Light source DMT Item   100070    DMT recommends Schott Cold Light Source  either Model KL 200 or Model KL 1500  LCD     e Water bath including heater DMT Item   100133  DMT recommends Julabo 5 L open bath circulator with plexiglass bath tank     e Vacuum pump DMT ltem   100130  DMT recommends a membrane vacuum pump having a volume of at least 6 L minute       Dissection petri dish DMT Item   100103  DMT recommends a  9 cm glass Petri dish coated with a 5 mm Sylgard polymer layer     Myograph 620M system accessories    This section contains a list of special accessories available for the multi wire myograph sys   tem   Model 620M     e ADIn
64. renaline needs to be optimized since too low a concentration makes it impossible to  evaluate the relaxation  It is  however  difficult to relax super maximally contracted arteries   which may lead to an underestimation of the sensitivity to acetylcholine  It is recommended   therefore  to apply a concentration of noradrenaline that induces 70 80  of the maximal  contraction response  In practice  this concentration is found by performing a noradrena   line concentration response curve as described in Experiment 1 of the Practice Exercises   Chapter 6 2   This value is known  pharmacologically  as the EC75 or ECas  This value can be  calculated based on the following equation if the EC  and Hill slope are obtained from a full  concentration response     log ECs    log EC     1 Hill Slope  x log F  100   FJ      F   70 or 80  depending on the desired value   The vessel segment is contracted with the noradrenaline  When the contraction has stabi   lized and reached a plateau  cumulative concentrations of acetylcholine are added to relax  the vessel  Each concentration is added after a steady response has been achieved  When    the vessel segment is either fully relaxed or does not relax any further to increasing concen   trations of acetylcholine  the experiment is over     Chapter 6    47    Protocol    Prepare the following stock solutions     e Acetylcholine  10   105  10    10   10    102M    e Noradrenaline  102 M    1  Mount the vessels as described in Chapter 5 2     2  
65. s from top view                            X       Pins from side view    L    J       X X    Figure 4 5   Illustrations of properly aligned pins  depicted on the far left  and incor   rectly aligned pins  depicted in the middle and far right                                 Chapter4 25    4 2 Calibration of the force transducer    As a part of the general maintenance of the myograph  DMT recommends that the myograph  is weight calibrated at least once a month  The myograph should also be weight calibrated  every time the interface has been moved  Although lab benches are all Supposedly perfectly  horizontal  small differences in lab bench pitch can affect the calibration of the system  The  myograph also should be calibrated if the system has been idle for longer than a month  A  step by step procedure is included in the FORCE CALIBRATION sub menu under SETTINGS  as  explained in Chapter 3  See Appendix for Principles of the Weight Calibration     Force transducer calibration procedure   This section contains step by step instructions to calibrate the force transducer and should  be used in conjunction with the steps described in Chapter 3  FORCE CALIBRATION sub menu  under SETTINGS      1  Move the jaws pins apart  If calibrating the transducer with the jaws in place  make sure  a wire is mounted on the transducer side jaw  If pins are being used  the wire does not  have to be put in place  Fill the chamber with distilled water or buffer  Use the same vol   ume that will be 
66. solated artery  arteries  i e  aorta or carotid arteries  of  choice from species of interest     2  Place the organ  vascular bed  or isolated artery into a beaker containing cold or room  temperature PSS     3  Transfer the organ  vascular bed  or isolated artery in a dissection dish for cleaning   Cleaning is best done under a dissection microscope in a petri dish coated with some  kind of material such as Sylgard so as to be able to pin the organ or vessel down without  causing major damage to the vessels in general  The petri dish should contain room  temperature PSS     4  Cut the vessel to an appropriate length   2 mm   Use a cut up plastic metric ruler under  the dissecting microscope as a reference for length     5  Fill the chambers of the 620M with 5 mL of room temperature PSS in each    6  Transfer a single ring into a chamber of the 620M    f  Mount the artery on the wire  25 um tungsten or 40 um stainless steel  or the pins  de   pending on the internal diameter of the vessel being studied  The mounting procedure  on wires can be found in Chapter 5 2  In addition  watch the accompanying video sent    with your system for step by step visual instructions     8  Place the chambers with the mounted arteries on the interface and plug the transducer  cables into the back of the interface     Equilibration    1  Turn on the heat  which should be preset to 37   C   2  Turn on your gas to bubble your chambers     3  Wait 20 minutes without touching the vessels or system
67. sors to remove a section of intestine together with its feeding vasculature  in     cluding the superior mesenteric artery  Be careful not to damage the vasculature during  this procedure  As a precaution to prevent the mesenteric arcade from drying  wrap the  arcade in a paper towel or gauze soaked in Physiological Saline Solution  PSS  see Chap   ter 6 3 for recipe   The proximal end of the intestinal section must be about 10 cm from  pylorus in rats  Make a cut in the proximal end of the intestine for later identification       Place the excised intestinal section in a petri dish  about 9 cm in diameter  coated with a    5 mm thick layer of Sylgard at the bottom to hold the fixing pins  Immediately fill the petri  dish with cold PSS that has been well aerated with your gas of choice  carbogen   95   O    5  CO  or 21  O5  5  CO   balance N    The dissection is performed without further  oxygenation of the PSS       Pin down the proximal end of the intestinal section on the left hand side of the petri dish    without stretching the vessels  Pin down the remaining intestinal section in a counter   clockwise direction  In this configuration  proximal end at the left side  distal end at the  right side and running counter clockwise from proximal to distal side   the arteries should  be underneath the arcade while the veins will be on the top       Select the vessel segment to be investigated  Figure 5 1   Novice myograph users should    start practicing their dissections and m
68. struments PowerLab data acquisition system  Including LabChart data acquisition and analysis software    e DMT Normalisation module DMT ltem   100044  LabChart add on module software     62 Multi wire myograph system Model 620M   User manual      Plastic mounting jaws DMT Item   100092  With built in platinum electrodes for point stimulation during electrophysiological experi   ments    e Chamber covers for NFS DMT Item   100065  With built in platinum electrodes for field stimulation during electrophysiological experi   ments     e 25 um tungsten wire DMT Item   100128  For mounting of ring preparations with internal diameter as small as 60 um     e Standard PC system DMT Item   100032    e CS 200  4 channel  current stimulator DMT Item      100052  Combined pulse and train generator     Myograph 620M system spare parts    This section contains a complete list of standard spare parts available for the multi wire myo   graph system   Model 620M  For parts not listed in this appendix or for special parts which  may need to be custom made  please contact DMT for further information     Force transducer DMT Item   100078  e Roll of 40 um stainless steel wire DMT Item   100112    High vacuum grease DMT Item   100129    Grease for linear slides DMT Item   100087  e Stainless steel jaws DMT Item   100091  e Mounting support pins  e 200 um DMT Item   100094  e 250 um DMT Item   100095  e 300 um DMT Item   100096  e 400 um DMT Item   100097  e Calibration kit DMT Item   100055  e Cha
69. t calibration should  be entered in the data acquisition being used  such  as AD Instruments PowerLab and LabChart     FORCE REC  OUTPUT RANGE    Force1 Rec  Force2 Rec    Force3 Rec    Force4 Rec      Range       Range       Range       Range     200 mN  1100 mN  20 mN    1200 mN          FORCE REC  OUTPUT RANGE      Range     Range     Range       Range       Range     Range     Range       Range     200 mN  1100 mN  20 mN    1200 mN    200 mN  1100 mN  20 mN    1200 mN         Range       Range       Range       Range     200 mN  1100 mN  20 mN    1200 mN       Chapter3 21    5  INTERFACE SETTINGS     The INTERFACE SETTINGS sub menu in SETTINGS  has an additional 2 sub menus  These 2 additional  Sub menus are     INTERFACE SETTINGS       TEMPERATURE  DIFFERENCE       TEMPERATURE DIFFERENCE    Il  FACTORY DIAGNOSTICS DIAGNOSTICS    l  TEMPERATURE DIFFERENCE   The TEMPERATURE DIFFERENCE function allows       the user to fine tune the temperature setpoint of UES spied a UNA een  the system  Although the temperature setpoint    Chamber t  1 2  C  secr    for the system can be set in the HEAT MENU  Chamber 2  1 1  C   ser   the actual temperature for the system may not Chsmberde 3598 en   heat to the exact defined setpoint  Therefore    the user can adjust the temperature of each us MA   sese J       chamber individually to fine tune the tempera   ture setting so that EXACT temperatures can be  achieved for any particular chamber  This is re   ferred to as a temperature offset 
70. t for a stable contraction or plateau between each concentration     46 Multi wire myograph system Model 620M   User manual    Noradrenaline final Add the following stock  concentration  5mL volume   solution to the bath     1x 10 9 M 5 0uL of 10  M  3 x 10 9 M 1 0uL of 10   M  1 x 10 8 M 3 5uL of 10   M  3 x 10 8 M 1 0uL of 104 M  1 x 10 7 M 3 5uL of 104M  3 x 10 7 M 1 0uL of 103 M  1x 10 6 M 3 5uL of 103 M  3 x 10 6 M 1 0uL of 102 M  1x 10 5 M 3 5uL of 102 M    Table 6 1 The order of noradrenaline stock solutions to be added to the myograph chamber and the  final concentration of the added noradrenaline in a 5mL bath     Experiment 2  Acetylcholine concentration response curve    The purpose of this exercise is to determine the sensitivity of the endothelium dependent  vasodilator  acetylcholine  and get a sense of how well acetylcholine causes endothelium de   pendent relaxation in normal  noradrenaline pre contracted  rat mesenteric small arteries     Background    Acetylcholine causes relaxation of rat mesenteric small arteries by activating muscarinic M3  receptors at the endothelial cell layer  This receptor activation causes the release of endothe   lium derived relaxing factors     Rat mesenteric arteries do not show spontaneous tone in the wire myograph  which is why it  is necessary to first induce a contraction to be able to observe the relaxation to acetylcholine   In this exercise  the contraction is induced by noradrenaline  The required concentration of  norad
71. tioned under the first    secured wire  Carefully move the jaws together  Figure  5 114   The procedure clamps the second wire to prevent  it from sliding off the jaw and damaging the vessel seg   ment when securing the wire to the transducer side jaw   Secure one end of the wire in a clockwise direction un   der one of the screws on the transducer side jaw  Figure  5 11B      Step 7  Securing the mounted vessel    Secure the other  free end of the wire under the remain   ing screw on the transducer side jaw  Again  the wire is  passed clockwise around the screw  The wire will tighten  as the screw is tightened  Figure 5 12A   Move the jaws  apart to slightly stretch the vessel segment  Make sure  that the vessel on the far side of the jaws does not extend  beyond the face of the jaws  Even a small extension will af   fect the normalization procedure  In case of excess vessel  on the far side of the jaws  move the jaws together again  and remove excessive tissue using forceps as described  in Step 4  A better method for the skilled operator is to  move the jaws slightly apart and use scissors to make a  small slit in the vessel wall where the vessel is clamped   Figure 5 12B   Once the wires are secured and the vessel  is cleaned of any residual  overhanging vessel  position  the wires so that they are just touching  but no force is    registered on the force transducers  Figure 5 120         Figure 5 11 A and B  Mounting step 6       Figure 5 12 A and B  Mounting step 7    C
72. tions  limita   tions and exclusions     Exclusions    Force transducers  separately or part of myograph systems manufactured by DMT A S  are  disclaimed from any warranty     Limitations    This warranty shall not apply to equipment subjected to accidental damage  improper use   alteration  or deterioration     Warranty on third party products will be as determined by their respective manufacturer   DMT A S shall not be liable for consequential  incidental  special  or other direct or indirect  damages resulting from economic loss or property damage sustained by you or any end user  from the use of the products sold or services rendered hereunder     Warranty Returns    A    Return Material Authorization     RMA  number is required for all returns  This number  Should be clearly indicated on all returned myograph systems     Products damaged due to improper or inadequate packaging when returned for RMA pur   poses are not granted warranty coverage     Appendix 1 51    Appendix 2     Calibration of Eyepiece Reticule    Principles of Ocular Calibration  The purpose of calibrating the eyepiece reticule is to determine a conversion factor      so  that the microscope can be used for measuring vessel segment lengths mounted in the wire    myograph     Several types of eyepiece reticules are available  The most simple is a horizontal scale as  illustrated in Figure A2 1     X  46 5o 60 70    90 190    0 d y       Figure A2 1  Horizontal eyepiece reticule scale    The basic princip
73. tocol  See page 47 49 for PSS and KPSS recipes     3  Start the wake up protocol by stimulating the vessel with KPSS   10uM noradrena   line   Stimulate for 3 minutes     4  Wash out 4 times with regular PSS over 5 minutes and wait 5 minutes   5  Repeat steps 3 and 4     6  While the vessel is in regular PSS  stimulate the vessel with 10uM noradrenaline  and  wait 3 minutes     f  Wash out 4 times with regular PSS over 5 minutes and wait 5 minutes   8  Stimulate the vessel with KPSS  Wait until the contraction is stable at a plateau  This  will take more than 5 minutes  and can take up to 10 or 15 minutes  maybe longer de     pending on the vessel examined     9  Wash out 4 times with regular PSS over 5 minutes or until the vessel has returned to  baseline  Wait 5 minutes     10  Stimulate the vessel with KPSS   10uM noradrenaline   Stimulate for 3 minutes     11  Wash out 5 or 6 times over 20 minutes with regular PSS  Continue washes if the vessel  hasn t reached baseline      NOTE  use the appropriate agonist for your vessel that causes a robust contraction   Final vascular function assessment  1  Add contractile agonist  used during the standard start so that the final concentration  in the bath is 24M  5uL of 103 M stock to the 5mL bath   Allow contraction to occur for    3 to 5 minutes     2  DO NOT WASH THE VESSEL     44 Multi wire myograph system Model 620M   User manual    3  Add acetylcholine  Ach  or carbachol at a final concentration of LOM in the bath  5uL  of 1 x
74. ure 5 9A  Dissect free and remove the excess vessel  from the chamber  A small amount of vessel should be A  clamped in the area indicated by the circle in Figure 5 9B   By having a small amount of vessel secured against the  jaw face will prevent the vessel from moving and subse   quently being damaged during the introduction of the 2 4  wire   B          C  Figure 5 9 A  B and C  Mounting step 4    Step 5  Introducing the 2    wire    e Move the jaws apart using the micrometer  Figure 5 10A    Take a second wire  and hold it about one third down from  the far end using a pair of forceps  Align the wire parallel  with the vessel segment so the wire can be passed into  the far end of the lumen  Gently feed the wire through the  lumen of the vessel in one movement using the already  mounted wire as guide  Figure 5 10B   Hold the wire at  a point at least 10 mm from the vessel to prevent the  vessel from being stretched during the manoeuvre  Be  careful not to touch the lumen of the vessel with the end  of the wire when pushing the wire end through the near  end of the lumen  Once the wire has successfully passed  through the lumen of the vessel  place the wire in a posi     tion so that sufficient length of the wire can be secured at  both the near and far screws on the transducer side jaw   Figure 5 10C      Qm 5 10 A  Band C  Mounting step 5       38 Multi wire myograph system Model 620M   User manual    Step 6  Securing the 2   wire    e Align the 2  wire so that it is posi
75. used during the experiments     2  Set up the calibration kit  bridge and balance  on one of the myograph chambers as il   lustrated in Figure 4 6  Also place the weight on one of the chambers  Turn the heat on  as discussed in Chapter 3  The system takes about 20 to 30 minutes to reach 37  C   Obviously  lower temperatures take less time and higher temperatures take more time to  reach  Make sure adequate time is allowed so that calibration can be performed at the  temperature at which the experiments will be performed  Placing the calibration kit and  weight on the chamber allows them to warm up to the experimental target temperature   No need to bubble the chambers while waiting for the system to heat up        Figure 4 6   Weight calibration kit shown in place on a single myograph chamber    26 Multi wire myograph system Model 620M   User manual    3  When the system reaches target temperature  adjust the calibration kit so that the tip of  the transducer arm is as close to the wire  if jaws are being used  or pin on the transduc   er side as possible without touching  as illustrated in Figure 4 9  One way to do this is to  use the following technique  Start with the calibration kit in place so that the transducer  arm of the bridge with the pans is not touching any part of the jaw or wire  if the jaws  are being used  or not touching any part of the pins  Go to the main menu displaying the  forces  and zero the channel being calibrated so the force reads zero  Slowly and
76. ussed  and then step by step instructions are listed on how to perform  a full vascular reactivity protocol     6 1 Summary of procedures   start to finish    Equilibration    Once a vessel has been mounted  regardless of whether it is a mouse aortic ring on pins or  rat mesenteric artery on wires   the vessel needs to equilibrate in the Myograph Chamber be   fore reliable  reproducible results can be obtained during the experiments  The equilibration  period allows the vascular ring preparation to heat up to experimental temperature slowly  while giving the preparation time to reset ion gradients that may have been disturbed during  dissection and cleaning  The equilibration period also allows the preparation time to achieve  a Stable level of passive tension  regardless of whether the normalization for small vessels  or a pre determined passive tension was used  Typically  equilibration takes no more than 1  hour     The wake up protocol  The purpose of performing a wake up protocol is to   1  Re activate the mechanical  functional  and signalling properties of the vessel segment    2  Check that responses to different types of stimuli are normal and verify that the function   ality of the vascular ring preparation was not damaged during the dissection or mounting  procedures     3  Check whether the vessel segment meets inclusion exclusion criteria to be used for ex   perimentation  For example  if the DMT normalization module from ADInstruments is  used to determine the p
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Samsung HG32ED590HB Felhasználói kézikönyv  Gateway GT150 F1 User's Manual    SANYO DENTETSU SA NY O DE NT ET SU SANYO DENTET  Tassenwärmer / Cup warmer - Kafi  Invacare® Cadiz General Cleaning After Use    Mode d`emploi Fours multifonctions H 330 BP, H 330  Delta T2761 Installation Guide  SAILOR 150 FleetBroadband    Copyright © All rights reserved. 
   Failed to retrieve file