ACD-14 FX and ACD-14 TRMS FX Digital Clamp-On
Contents
1. ACD 14 FX ACD 14 TRMS FX User Manual Bedienungshandbuch Mode d emploi e Manuale d Uso e Manual de uso Bruksanvisning ACD 14 FX ACD 14 TRMS FX Owner s Manual Limited Warranty and Limitation of Liability Your Amprobe product will be free from defects in material and workmanship for 1 year from the date of purchase This warranty does not cover fuses disposable batteries or damage from accident neglect misuse alteration contamination or abnormal conditions of operation or handling Resellers not authorized to extend any other warranty on Amprobe s behalf To obtain service during the warranty period return the product with proof of purchase to an authorized Amprobe Test Tools Service Center or to an Amprobe dealer or distributor See Repair Section for details THIS WARRANTY IS YOUR ONLY REMEDY ALL OTHER WARRANTIES WHETHER EXPRESS IMPLIED OR STAUTORY INCLUDING IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY ARE HEREBY DISCLAIMED MANUFACTURER SHALL NOT BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES ARISING FROM ANY CAUSE OR THEORY Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages this limitation of liability may not apply to you Repair All test tools returned for warranty or non warranty repair or for calib
2. resolution is useful for identifying the minute current changes flame detector applications Flame signal current check should indicate steady flame signal of at least 2pA for a rectification type 1 5yA for an ultraviolet type or 8 for self checking systems If a flame signal current with inadequate strength or fluctuation beyond 10 check the following to avoid the risk of unwanted flame relay dropout For gas or oil flames Minipeeper e Low supply voltage e Detector location e Defective detector wiring Dirty viewing windows e Faulty Minipeeper For oil flames Photocell e Detector location amp wiring Smoky flame or poorly adjusted air shutter e Faulty Photocell e Temperature over 165 F 74 C at photocell For gas flames Flame Rod e Ignition interference A flame signal current difference with the ignition both on and off greater than 0 5pA indicates the presence of ignition interference e Insufficient ground must be at least 4 times the detector area Flame lifting off burner head ground or not continuously in contact with the flame rod e Temperature in excess of 600 F 316 C at the flame electrode insulator causing short to ground HOLD The hold feature freezes the upper ACA display for later viewing Press the HOLD button momentarily to activate or exit this feature MAX The Max feature compares and displays the measured maximum value on the upper ACA display as fast as
3. TERMS IN THIS MANUAL A WARNING Identifies conditions and actions that could result in serious injury or even death to the user A CAUTION Identifies conditions and actions that could cause damage or malfunction in the instrument WARNING To reduce the risk of fire or electric shock do not expose this product to rain or moisture The meter is intended only for indoor use To avoid electrical shock hazard observe the proper safety precautions when working with voltages above 60 VDC or 30 VAC rms These voltage levels pose a potential shock hazard to the user Inspect test leads connectors and probes for damaged insulation or exposed metal before using the instrument If any defects are found replace them immediately Do not touch test lead tips or the circuit being tested while power is applied to the circuit being measured To avoid accidental short circuit of bare un insulated hazardous live conductors or bus bars switch them off before insertion and removal of the current clamp jaws Contact with the conductor could result in electric shock Keep your hands fingers behind the hand finger barriers that indicate the limits of safe access of the meter and the test leads during measurement A CAUTION Disconnect the test leads from the test points before changing meter functions INTERNATIONAL ELECTRICAL SYMBOLS A Caution Refer to the explanation in this Manual Caution Risk of electric shock Earth Ground
4. AC voltage To activate the Hz Frequency function press the Hz button momentarily Note 400 0 range is designed with 1000M high input impedance for least current drain in measuring small signals and can cope better with most commercially available voltage output transducers adapters The non zero display reading is normal when the meter inputs are open circuit which will not affect actual measurement accuracy The meter will show close to zero reading when the inputs are shorted Open input is actually a floating condition which is not a zero volt input condition Note Hz input sensitivity varies automatically with function range selected while activating the Hz function Lowest range has the highest sensitivity and the highest range has the lowest sensitivity Activating the Hz function WHILE measuring the specific function signal auto ranging can automatically select the best sensitivity range to avoid electrical noise that may cause unstable Hz reading However if the Hz reading shows zero due to insufficient sensitivity activate the Hz function BEFORE measuring the specific function signal The meter is then in the lowest range and can select the highest sensitivity range Q Resistance and al Continuity functions Inputs are made through the test leads terminals Move the slide switch to the position The unit defaults at Q Press SELECT button momentarily to select Continuity function which is convenient
5. E 2 6 600 unspecified Type K thermocouple range amp accuracy Accuracy specified to 350A continuous not included and 400A for 60 second max with 5 minutes cool down interval Added range to indicate instantaneous over range current values TRMS model ACD 14 TRMS FX accuracy is specified from 10 to 100 of range 15 DC pA Burden Accuracy Voltage 400 04 1 5 4d 2 8 20004A 1 2 3d 2 8 AC pA Burden Accuracy Voltage 50Hz 500 2 400 0uA 2 0 5d 2 8 20004A 1 5 5d 28 Ohms Lee 400 092 0 8 6d 4 000kQ 40 00kQ 0 6 4d 400 0kQ Open Circuit Voltage 0 4VDC typical Audible Continuity Tester Audible threshold between 10Q and 120Q Diode Tester Open Circuit Test Current Voltage Typical lt 1 6 VDC 0 25 Capacitance 500 0nF 5 000uF 3 5 50 00uF 500 0uF 3000uF Additional 50 00nF range accuracy is specified Accuracies with film capacitor or better Specified with battery voltage above 2 8V approximately half full battery Accuracy decreases gradually to 12 at low battery warning voltage of approximately 2 4V Hz Frequency Sensitivity 400 0mV 300mV 10Hz 2kHz 4 000V AN 5Hz 40kHz 40 00V 28V 5Hz 100kHz 400 0V 100V 5Hz 100kHz 600V 500V 5Hz 5kHz 400 0uA 500uA 10Hz 30kHz 2000uA 500uA 10Hz 30kHz Display counts 5000 Best resolutio
6. GD D Double Insulation or Reinforced insulation E Fuse AC Alternating Current DC Direct Current Conformity symbol the instrument complies with the valid directives It complies with the EMC Directive 89 336 EEC and the Low Voltage Directive 73 23 EEC with their valid standards Underwriters Laboratories Inc Note Canadian US Product Description Note Please refer to Specification Section for function details 1 2 3 4 5 6 7 8 9 Clamp Jaw for AC current magnetic field pick up Hand Finger Barrier to indicate the limits of safe access of the meter during measurement Power ON OFF buttons for ACA current function and MAX Hold features Push button for frequency selection Input Jack for all functions EXCEPT clamp on ACA current function Common Ground reference Input Jack for all functions EXCEPT clamp on ACA current function Slide switch Selector to turn the lower display ON OFF and to Select functions Lower Display for Input Jack functions 3 3 4 digits 4000 counts LCD Upper Display for ACA current Clamp on 3 3 4 digits 4000 counts LCD 10 Jaw trigger for opening the ACA current clamp jaw Operation DC Voltage AC Voltage Hz Frequency functions Inputs are made through the test lead terminals Move the slide switch to the vn position The unit defaults at DC voltage Press SELECT button momentarily to select
7. for checking wiring connections and operation of switches A continuous beep tone indicates a complete wire circuit Diode test function Inputs are made through the test lead terminals Move the slide switch to the Q position The unit defaults at Q Press SELECT button momentarily 2 times to select Diode test function Normal forward voltage drop forward biased for a good silicon diode is between 0 400V to 0 900V A reading higher than that indicates a leaky diode defective A zero reading indicates a shorted diode defective An OL indicates an open diode defective Reverse the test leads connections reverse biased across the diode The digital display shows OL if the diode is good Any other readings indicate the diode is resistive or shorted defective Capacitance function Inputs are made through the test leads terminals Move the slide switch to the position The unit defaults at Q Press SELECT button momentarily 3 times to select F capacitance function CAUTION Measuring Resistance Continuity Diode or Capacitance function on a live circuit will produce false results and may damage the instrument In many cases the suspected component must be disconnected from the circuit to obtain an accurate measurement reading ACA Current clamp on function Inputs are made through the clamp on jaws for non invasive ACA current measurements Press the OFF push button momentarily to pow
8. 25ms with auto ranging capability Press the MAX button for 1 second or more to activate or exit this feature 11 Auto ranging When there is more than one measuring range under a selected meter function the LED annunciator a turns on in the upper left corner The meter will automatically switch to the best resolution range when making measurements No manual ranging selection is required Auto Power Off APO When the meter is on the Auto Power Off APO feature will switch the meter into a sleep mode automatically to extend battery life after approximately 30 minutes of the last front panel activity To wake up the meter from APO press the buttons momentarily or move the slide switch to the OFF position and then back on again Always set the slide switch to the OFF position manually when the meter is not in use Maintenance A WARNING To avoid electrical shock disconnect the meter from any circuit remove the test leads from the input jacks and turn OFF the meter BEFORE opening the case Do NOT operate with open case Trouble Shooting If the instrument fails to operate first check the batteries and test leads etc and replace as necessary Double check the operating procedures as described in this User s Manual If the instrument voltage resistance input terminal is subjected to a high voltage transient e g lightning or switching surge by accident or abnormal conditions of operation the series fusible resistor
9. V 1 2 50us surge for all models Pollution degree 2 Overload Protections ACA Clamp on jaws AC 400A RMS continuous amp terminals all functions 600VDC VAC RMS 14 Standard Accessories Test leads pair banana plug style bead Type K Thermocouple batteries user s manual amp soft carrying case Replacement Part Numbers Test Leads pair MTL 90B Bead Type K Thermocouple TPK 59 Carrying Case SV U Electrical Specifications Accuracy is reading digits number of digits or otherwise specified at 73 F 9 F 23 C 5 C amp less than 75 R H True RMS Model ACD 14 TRMS FX ACV AcuA amp ACA clamp on accuracies are specified from 5 to 100 of range or otherwise specified Maximum Crest Factor are lt 1 96 at full scale amp lt 3 92 at half scale and with frequency spectrums besides fundamentals fall within the meter specified AC bandwidth for non sinusoidal waveforms DC Voltage AC Voltage aer 400 0 3 4d 50Hz 500 2 0 400 0V NMRR gt 50dB_ 50 60Hz CMRR gt 120dB DC 50 60Hz Rs 1kQ CMRR gt 60dB DC to 60Hz Rs 1kQ Input Impedance 10MQ 30pF nominal 1000 0 for 400 0 range Max Hold where applicable Specified accuracy 50 digits for changes gt 25ms in duration ACA Current Clamp on 50Hz 60Hz 40 00A 1 9 8d Input Impedance 10MQ 30pF nominal Type K Temperature Le 20 300 2 3 PETO
10. er on and off the separate ACA function upper display The MAX HOLD push button can also be used to power on this function This twin ACA function display can be used simultaneously with voltage measurements or any of the other slide switch functions lower display or it can be used alone A CAUTION Clamp the jaws around only one single conductor of a circuit for load current measurement Enclosing more than one conductor of a circuit will result in differential current like identifying leakage current measurement Make sure the jaws completely closed or else it will introduce measurement errors Strong Electro magnetic field environments such as high current transformers motors and conductors may affect measurement accuracy EAS AES Temperature function Insert the banana plug type K temperature bead probe TPK 59 observing the proper polarities Move the slide switch to the position The unit defaults at degree Celsius Press SELECT button momentarily to select degree F Fahrenheit HA Current function Inputs are made through the test leads terminals Move the slide switch to the yA position The unit defaults at DC Press SELECT button momentarily to select AC You can select the Hz function by momentarily pressing the Hz button 10 Application notes The DC pA function is designed especially for HVAC R flame sensor applications The
11. n 0 001Hz Accuracy 0 5 4d 16 Visit www Amprobe com for e Catalog e Application notes e Product specifications e User manuals
12. ration should be accom panied by the following your name company s name address telephone number and proof of purchase Additionally please include a brief description of the problem or the service request ed and include the test leads with the meter Non warranty repair or replacement charges should be remitted in the form of a check a money order credit card with expiration date or a purchase order made payable to Amprobe Test Tools In Warranty Repairs and Replacement All Countries Please read the warranty statement and check your battery before requesting repair During the warranty period any defective test tool can be returned to your Amprobe Test Tools distribu tor for an exchange for the same or like product Please check the Where to Buy section on www amprobe com for a list of distributors near you Additionally in the United States and Canada In Warranty repair and replacement units can also be sent to a Amprobe Test Tools Service Center see address below Non Warranty Repairs and Replacement US and Canada Non warranty repairs in the United States and Canada should be sent to a Amprobe Test Tools Service Center Call Amprobe Test Tools or inquire at your point of purchase for current repair and replacement rates In USA In Canada Amprobe Test Tools Amprobe Test Tools Everett WA 98203 Mississauga ON L4Z 1X9 Tel 877 AMPROBE 267 7623 Tel 905 890 7600 Non Warranty Repairs and Replacement E
13. s will be blown become high impedance like fuses to protect the user and the instrument Most measuring functions through this terminal will then be an open circuit A qualified technician should then replace the series fusible resistors and the spark gaps Refer to the LIMITED WARRANTY section for obtaining warranty or repair service Cleaning and Storage Periodically wipe the case with a damp cloth and mild detergent do not use abrasives or solvents If the meter is not to be used for periods of longer than 60 days remove the batteries and store them separately 12 Battery replacement The meter uses 2 standard 3V alkaline button batteries ANSI NEDA 5004LC IEC CR2030 The lower battery is used for the slide switch functions and the upper battery is used for the upper display ACA function separately Low Battery Indication When the battery voltage drops below approx 2 4V a battery symbol will appear in the upper right corner of the LCD display indicating that the battery needs replacing To replace the batteries loosen the two screws from the case bottom and remove the bottom case Slide the battery out the side of the holder see example below and replace with a new battery observe polarity Replace the bottom case Note Take care to insure that the two 2 springs located near the lower battery line up with the speaker in the rear cover Re fasten the screws 13 Specifications General Specifications Di
14. splay 3 3 4 digits 4000 counts dual LCD displays Update Rate 3 per second nominal Polarity Automatic Low Battery Indication Below approx 2 4V Operating Temperature 0 C to 40 C 32 F to 104 F Relative Humidity Maximum relative humidity 80 for temperatures to 31 C 88 F decreasing linearly to 50 relative humidity at 40 C 104 F Altitude Operating below 2000m 6 562 ft Storage Temperature 20 C to 60 C 4 F to 140 F lt 80 R H with battery removed Temperature Coefficient nominal 0 15 x specified accuracy 0 C 18 C or 28 C 40 C F 32 F 64 F or 82 F 104 F or otherwise specified Sensing Average sensing for ACD 14 FX True RMS for ACD 14 TRMS FX Power Supply Two 3V alkaline button batteries ANSI NEDA 5004LC IEC CR2032 Battery Life 95 hours approx in DCV mode Power Consumption 2 5 mA typical for ACD 14 2 9mA typical for ACD 14 TRMS FX Low Battery Indication Below approx 2 4V battery symbol appears in upper right Auto Power Off After being idle for 30 minutes APO Consumption ACD 14 FX 0 8pA typical except 25 typical for temperature function ACD 14 TRMS FX 10pA typical except 25 typical for temperature function Dimension L 190mm X W 63mm X H 32mm L 7 48 X W 2 48 X H 1 25 Weight 204 g 0 450 Lb approx Jaw opening amp Conductor diameter 26mm 1 02 max Safety Category III 600 Volts AC amp DC CE Approved UL Listed Transient protection 6 5k
15. urope European non warranty units can be replaced by your Amprobe Test Tools distributor for a nominal charge Please check the Where to Buy section on www amprobe com for a list of distributors near you European Correspondence Address Amprobe Test Tools Europe P O Box 1186 5602 BD Eindhoven The Netherlands Correspondence only no repair or replacement available from this address European customers please contact your distributor ACD 14 FX ACD 14 TRMS FX Contents Safety EE 3 Product Description Operation Maintenance le MEET 14 Congratulations Your new instrument has been crafted according to quality standards and contains quality components and workmanship It has been inspected for proper operation of all of its functions and tested by qualified factory technicians according to the long established standards of Amprobe Safety Regulations This manual contains information and warnings that must be followed for operating the instrument safely and maintaining the instrument in a safe operating condition If the instrument is used in a manner not specified by the manufacturer the protection provided by the instrument may be impaired OVERVOLTAGE CATEGORY Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations Note Examples include switches in the fixed installation and some equipment for industrial use with permanent connection to the fixed installation