1.    This unit is marked according to the Medical Device  0537 Directive 93 42 EEC  if the unit contains the CE   mark     This symbol indicates that the waste of electrical  and electronic equipment must not be disposed as  unsorted municipal waste and must be collected  separately  Please contact an authorized represent   ative of the manufacturer for information concerning  the decommissioning of your equipment     30 User   s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    7  Warnings and precations Imaging plate unit    7T  Warnings and precations    THE UNIT IS A CLASS 1 LASER PRODUCT    Note  When covers are removed the unit is a class 3B laser product     avoid expo   sure to the laser beam     CAUTION   Use of controls or adjustments or performance of procedures other  than those specified herein may result in hazardous laser radiation exposure    e When handling imaging plates  protective covers and hygiene bags always  take the appropriate hygiene measures and precautions to prevent cross con   tamination     e The imaging plates are harmful if swallowed     Do not move or knock the unit when it is reading an imaging plate       This unit must only be used to read image plates supplied by the manufacturer  and must not be used for any other purpose  NEVER use imaging plates  pro   tective 
2.   NOTE   Start image capture from the dental imaging soft   ware you are using     CAUTION    For optimum image quality the exposed IP must be  read as soon as possible after exposure  If the ex   posed IP is not read immediately after exposure but  left unread for a period of time it must be protected  from ambient light and X rays as both may degrade  the image quality     1  Pull the flap to open the hygiene bag     2  Keep the IP in the protective cover so that you  do not touch the IP or allow it to be exposed to  ambient light  and then slide the protective cover  and IP out of the hygiene bag     3  Hold the protective cover and IP so that the  white side of the protective cover is on the right   The metal disk on the IP will also be on the right   Insert the protective cover and IP into the unit  door    A magnet will hold the IP in the correct position              16 User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    2  Intraoral imaging plate unit Imaging plate unit    Slide the protective cover off of the IP  and leave  the IP in position half way in the unit door     4 4    7 A As soon as the protective cover is removed the    unit detects that the IP is in the unit door and will  switch from the standby state to the ready state     ready to use  and automatically slide the IP into
3.   degrade the diagnostic value of the image  Dust wear on the sensitive side can appear as  white grey spots on the image     User s manual 206749 25    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 4  Handling and care of imaging plates    e Unprotected IPs must not come in contact with  the patient  the patient s saliva or any other bod   ily fluids    Always use hygiene bags and protective covers  with IPs when taking exposures to      eliminate cross contamination     maintain image quality     maximize IP service life     4 3 Cleaning  If IPs are handled and stored correctly cleaning  should not be necessary or can be kept to an abso   lute minimum   Clean the IP if there you see any visible marks or  stains  which are not obvious scratches  on the im   age or if there is a reason to believe that the IP is  contaminated       Use the microfibre cloth supplied       First very gently wipe the IP in a backwards  forwards movement in both the widthwise and  lengthwise directions and then finish with a cir   cular wiping movement     e For stubborn marks and stains that cannot be be  removed using the microfibre cloth  use 70     96  ethanol  70   96 EtOH   30   4 DI WATER   or ethanol anhydride  Wipe the IP as described  above and then wipe dry     NEVER USE abrasive chemicals or materials t
4.   the black side of the protective cover is on the  same side as the black side of the hygiene bag     En  wm    5  Peel off the cover paper from the sealing tape  and then fold the flap  along the pre formed line   over and onto the sealing tape        6  Press and slide your thumb along the tape to  ensure that the flap is properly sealed        7  Turn the sealed hygiene bag over and check  that the IP and protective cover are in the cor   rect position    You must be able to see the light side of the pro   tective cover and the metal disk on the IP        12 User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    2  Intraoral imaging plate unit    Imaging plate unit    Imaging plate holders  It is recommended that imaging plate holders be  used to ensure accurate IP positing and consistently  good images quality     MAY           y    Using imaging plate holders improves image quality  because     the IP is positioned correctly in relation to the  tooth    there is no positioning guesswork    the IP is not bent and thus distortion is  eliminate    the IP cannot move in relation to the X ray unit  images are standardized and reproducible  there is no overlapping nor cone cut off   IP wear and tear is minimized     image quality can be maintained irrespective of  who takes the image    time is sav
5.   the unit     CAUTION    If the IP does not slide into the unit after the  protective cover has been removed  the IP is  misaligned or has been placed in the unit the  wrong way round  Reposition the IP and insert it  into the unit again     If after repositioning the IP it still does not slide  into the unit press the Start key to manually  switch the unit from the standby to the ready  state and the IP will then slide into the unit when  the protective cover is removed     The busy animation will appear on the display  which indicates that the imaging plate is being  read        After few seconds a preview image will appear    Progress on the unit display  animation    N PC  A read out progress window will appear on  the PC display  After a few seconds the image  will appear in the dental imaging software    The image can now be saved  Refer to the  documentation supplied with the dental imaging  software you are using        User   s manual 206749 17    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 2  Intraoral imaging plate unit    CAUTION    If the metal disk on the rear of the IP can be  seen on the digital image  it indicates that the IP  was exposed from the wrong side        CAUTION   RETRIEVING IMAGES   If the image is not transferred to the PC because  of a network  PC or software 
6.  04301 Tuusula   FINLAND   Tel   358 10 270 2000   Fax   358 10 270 2230    See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit    IV User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit    Contents   1  The Scan eXam   a centetecttccact cece a intra muuoc eem uel EN R D cMEIREM E  1  TOT MOT debis oe tOd oa ftia eed ada cun pe dU sat PALM URDU aR RM Du RUE d o RO rnb Ro E 1  1 2 System installation  ee aoi doe t cse bulo to e Co dote Mo aaa ESE 2  Positioning the WIN e ENT IE 2  Positioning the PC seese e e E Leine Eaa 3   Other devices piprita E ee e S 3   2  Intraoral imaging plate IDE  uuc ioc ce saa rev eco paa bat auic wide vlt nzo dear add 4  2 1 Main parts and controls   usos esr oriatur theta k o Ro RA dep Quies pa Fa RE k pus 4  2 2 PCCCS SOMES P                                      5  2 3 Display symbols and what they mean                          ssee 7  2 4 Using the SYStemM                ccccceceecceceeeeeeeesseeesseeeeeeeeeeeeeeeeceeaeeaaeeseeeeesseeeeees 10  Preparing the System e ems 10  Preparing an IP for expoOSULe      nonecoi tese noie weoip orta Font t teal etat etus nnna 11  Huizienatefelsitzdsrelelc S T ereere ermmm 13   Taking an exposes ditatus anion EE es Pa se eee
7.  1 5 m  4 5 ft     The minimum vertical distance between the patient  and the PC is 2 5 m  6 5 ft      Other devices    DO NOT connect any other devices to the unit or  the PC connected to the unit that are       not part of the supplied system    not supplied by the manufacturer of the unit      not recommended by the manufacturer of the  unit     User s manual 206749 3    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 2  Intraoral imaging plate unit  2  Intraoral imaging plate unit  2 1 Main parts and controls  Power on   off key and status light   O  green   on      flashing 7 entering the standby mode       yellow   in the standby mode        di    V Proximity sensor    P   a   Displa H Z  eae d  LN            rUhit door and  plate carrier           Imaging plate collector    Rear of unit Power supply connector for the  power supply unit  PSU      A    i  CAUTION     Only use the PSU supplied with the   unit or an approved spare PSU   supplied by an authorized distributor      Refer to technical specifications     RJ 45 connector for Ethernet cable     direct connection to PC     network  LAN  connection  Ethernet cable  not included    Refer to technical specifications        PC  not included  with digital imaging  software that conforms to the MDD    4 User   s manual 206749    Reviewed  
8.  4   B2 3 Express Share ecciesie quat tuce eg qurace taco vdu e RES cU acea a ides eee 6   B3 Setup Options aae doe etra aS cene tenuti e debi c Ste mY eee ee Uno asc es  8  B31 Scanner sorro ae a DR 8  EU ae mani aaa                            er 9   Image Scannihg RERO 9   MISHA NS SIAL Chart siirius eas SS sal ede ee aa a 9   cz                            10   mage POSS SUG soisissa ES AMAN  10   Last image eret i euis puente ome tides eT ee 10   Scanner Unt Serial IUD uec qsa ique pegulb genit buszcc iiam septa nni 10   Be Acro m                             11   B3  3 VV OMIM                                                           11  Readout SIAM sab sacsaneansarasimmacniunponmmaiuanei naeaiian D s a RR R SIN 12   TOUENIESS ODGIRUOI   ocean paese np porap tre postres 3H OR HU ibusk ir MEET GER pa IM Qra ES 13   Plate GIS MO T EC                         13   B3 4 Power ODDS ieiuna a beant absol bau cuc EaR 14  Power Options ND LEUR 14   B4 Tro  bleshootihg os aen sauer qv qadcuis pacti cuta paca usur NE REM e EEEa e a esas 15  User s manual 206749 VII    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit    VIII User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this
9.  48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit B  Installation and setup    Key the serial number of the unit into the Scan   ner serial number field  The serial number of  the unit will appear on the unit display when the  unit is switched on  It can also be found on the  type label on the back of the unit    Make sure that the Computer network connec   tion that provides the LAN network connection  is selected     B2 2 IP method  using the unit IP address     If your system does not allow the direct connection  method to be used to connect the PC s   connection  can be done using an IP address    1  Follow steps 1 to 5 from the previous section   Direct connection method  uses the unit s n      2  PC  From the Settings tab select IP based and  then select the Enable changing IP address  box     Scanner Settings   Drvar Irfa           Scanner Connection     Direct Connection Direct Connection settings         Scanner sera number  ETZE    Computer newak connection         Phased Scanner IP ecdcress    18  168 02 Send to Scanner     Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    B  Installation and setup Imaging plate unit    Obtain an IP address for the unit from your net   work administrator and key it into the IP 
10.  P is the maximum output power rating of the transmitter  in watts  W  according to the transmitter manufacturer     NOTE 1  At 80 MHz and 800 MHz  the separation distance for the higher frequency range    NOTE 2  These guidelines may not apply in all situations  Electromagnetic propagation is  affected by absorption and reflection from structures  objects and people           User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28    D506723    Imaging plate unit A  Technical Specifications    A 8 User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    B  Installation and setup Imaging plate unit    Appendix B  Installation and setup  Installation and setup must only be carried out  by service personnel trained and approved by  the manufacturer of the unit     B1  System installation    B1 1 Positioning the unit  Position the unit on a stable flat surface so that vi   brations will not degrade the image quality   The unit can also be attached to a wall  under or on  a shelf using the optional scanner mounting kit   pt  no  204138      Do not position the unit in direct sunlight or near  bright light  Sunlight or bright light must no
11.  and Dental Chart is se   lected a preview image with a dental chart for tooth  numbering will appear before the image is saved     Using the dental chart    1  After an IP has been read a window will open  that shows a preview image and a dental chart     2  Click the tooth   teeth on the chart that cor   respond to the tooth   teeth in the image  Tooth  numbers will be assigned to the selected teeth     The tools at the top of the window allow the im   age to be manipulated     3  Click OK to save the image and tooth numbers     Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit B  Installation and setup    Resolution    Super gives a pixel size of 35y  This results in im   ages with better resolution  but more memory is  required as the image files are larger     High  default  gives a pixel size of 64u  This results    in images with less noise especially if short expo   sure times are used     Image Processing    Image Processing   v  Noise Filtering  Noise Filtering  normally selected  makes images  smoother when they are taken at low radiation  doses   Last image  Last Image    Retrieve latest image from the scanner unit memory     Retrieve now      Refer to section Retrieve last image           Scanner Unit Serial number  Scanner Unit Serial Number      Add serial number to new images    Add
12.  determine the status of this document  Printed out  12 07 12 12 51 28 D506723    B  Installation and setup Imaging plate unit    B4 Troubleshooting    PROBLEM  The unit does not come on  The unit s power on   off  status light and display are off     CAUSE   SOLUTION  The main power supply to the unit is off or the unit is  not switched on     i  Ifthe LED on the PSU is not on it is not receiv   ing power from the mains  Switch the mains  power on     ii  Ifthe LED on the PSU is on switch the unit on     iii  If status light and display still do not come on  check the unit s membrane control panel and  cabling  Replace if faulty     PROBLEM   Animation displaying either Unit connection or  Dental imaging software  flashing yellow  appear  on the unit display and the unit will not work        CAUSE   SOLUTION    Defective RJ45 cable  Replace     Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit B  Installation and setup    PROBLEM  The IP connection between the unit and the PC  does not work     CAUSE   SOLUTION    i  Check that the Direct connection method was  configured correctly  If all the setting are correct  but the connection still does not work  use the IP  connection method     ii  The unit has been configured to work with one  PC only and does not recognize any other PCs   Reconfigure t
13.  document  Printed out  12 07 12 12 51 28 D506723    Introduction Imaging plate unit    1  The Scan eXam     1 1 Introduction    This manual describes how to use the Scan eXam   digital imaging plate scanner  the unit  which is part  of the digital intraoral imaging plate system  the sys   tem   The complete system comprises the following       the Scan eXam  digital imaging plate scanner   the unit        Dental Imaging Plates  IPs   protective covers   hygiene bags and other related imaging plate  accessories       APC  not supplied  with suitable dental imaging  software       Alocal area network  LAN  cable  not supplied   will be required if the system is to be used in a  network     The unit is a laser scanning device designed to  automatically read reusable dental intraoral imaging  plate sizes 0  1  2  and 3  After reading the images  can be viewed on the PC using the dental imaging  sofware     The unit can either be connection directly to the PC  or to network via the LAN cable     The unit can be set up to work with a single PC  the  single user configuration  or with several PCs  the  Express Share  multiple user  configuration    With the Express Share configuration up to eight  PCs can be used  one at a time  with the unit     User s manual 206749 1    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    
14.  ener ey 14  Reading an imaging plate a   ss coeno a ieu epe eaaet me toni ee deii cin  e 16  Removing IPs from the plate collector                                     eessssesssss 18   MS   eg  0   Q0 10   TNR NER NE EE Dn RR 19  Retrieve last image m                       19   Shutting down the Unit een eee rele en per eee een 20   IP erasing mode  Initial erasing of the IPS                                      sss 20   ISI TOES cues cecidi tend niae cw ve AR Deae AK a SA lng diu Rd 22   SEE ULI OOM NE I m OT 23  eme  istae o ooo 23  Selecting the tooth numbering feature                                  ssesesuuusssss 23   Using the tooth numbering feature                                 sesseeessessees 23   3 2 FSS SO PO E o eiue idola portet e aaa 24  3 3 Image PROCESSING sie abiit E o Coa Mod eti imt tie i dua 24  User s manual 206749 V    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit    4  Handling and care of imaging plates                                   ssssssseee  25  a Generalene i a ee er ee Pe eea 25  A2 WING esee E a er a E e 25  43 Cleaning e LL LE EIL EE E E 26  ce Ec oS ETE 27  4 5 Replacement ids iio casa wig a e Duero dud 27  4 0 DISPOSA ie e DOS A Au uu i 27   5  Unit care and MAIMGTIANCE io cicasssucnnecszesnnneteosousenocercemdteancnnioceraecennes 28  5 1 Cleaning the U
15.  the eras   ing mode than the normal read and erase mode   This is to ensure that IPs that have not been  used recently are erased properly     After the IP has been erased it will be ejected  from the unit  The IP can now be used to take  an exposure     To exit the erasing mode  either wait 15 sec   onds for the unit to automatically exit the mode   or press and hold down the Start key until the  erase animation disappears     21    See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 2  Intraoral imaging plate unit    Errors    e If there is a problem with the unit the error state    symbol and an error number will appear on the   N unit display     Error    Restart the unit  The error should clear  If not  contact your authorized distributor for assist   ance     e PC  If the message   Degraded image quality    appears on the PC display during or after an IP  is read it indicates that the image may not be  within the factory set quality limits    If the image looks okay no action is necessary   If the message appears frequently  contact your  authorized distributor for assistance     22 User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    3  Setup options Imaging plate unit    3  Setup options    There are 
16.  values in the  following table     14 User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    2  Intraoral imaging plate unit Imaging plate unit    If the exposure time is too short images will be  noisy  Such images may still usable for some  diagnostic purposes     If the exposure time is too long images will be  too dark or will show patient movement  These  images will not be good enough for accurate  diagnostic examination     The exposure values table below should only be  used as a guide     Recommended exposure values  s  for DC x ray units     60 kV  7 mA 70 kV  7 mA  shortcone   longcone  Adult     Child   Adult Child   Adult Child   Adult Child    Bitewing  Maxillary incisor    Maxillary molar  Occlusal          3  Protect yourself from radiation and take the  exposure           4  After the exposure remove the sealed hygiene  bag from the patient s mouth   CAUTION   If there is a risk of cross contamination  wash   disinfect and dry the hygiene bag before open   ing it     User s manual 206749 15    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 2  Intraoral imaging plate unit    Reading an imaging plate
17. 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit    A  Technical Specifications       Guidance and manufacturer s declaration     electromagnetic emissions       The eXam5 is intended for use in the electromagnetic environment specified below  The  customer or the user of the eXam5 should assure that it is used in such an environment                                   Emissions test Compliance   Electromagnetic environment   guidance   RF emissions Group 1 The eXam5 uses RF energy only for its internal   CISPR 11 function  Therefore  its RF emissions are very low and  are not likely to cause any interference in nearby  electronic equipment    RF emissions Class B The eXam5 is suitable for use in all establishments    CISPR 11 including domestic establishments and those directly  connected to the public low voltage power supply   Harmonic Class A network that supplies buildings used for domestic   emissions purposes    IEC 61000 3 2   Voltage Complies   fluctuations  flicker   emissions   IEC 61000 3 3   A 4 User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22    See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28    D506723      Technical Specifications    Imaging plate unit       Guidance and manufacturer s declara
18. Addition workstation in   formation  for example  user name  location etc   can be entered into the field next to the work  station identifier number     IMPORTANT NOTE   If only one PC is connected to the unit do not  select the Use Express Share check box         Express Share        v Use Express Share    Workstation identifier           Fiv 2251 616  jao         Scanner Autorelease timeout  seconds         diii      Cancel      The Scanner Autorelease timeout is the length  of time that the unit will remain reserved and  unused by a PC before the PC automatically re   leased the unit so that it can be used by another  PC in the system  the scanner can be reserved  in advance from another PC     The default setting is 40 seconds  This can be  changed by keying in a new value    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    B  Installation and setup Imaging plate unit    2  Click OK to connect the PC to the unit   NOTE     An automatic technique will automatically locate  the unit within the local area network and con   nect the PC     3  Repeat the above process for all the other PCs  in the network  Make sure that you give each PC  a different Workstation identifier     4  Checkthe installation by starting image capture  using the imaging software   If the Use Express Share was selected the  Workstation ident
19. ENGLISH    Scan eXam   Digital imaging plate scanner    User s Manual    0 805 0329    206749 rev  3  2012 05     Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Copyright    Manufactured by    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22    Document code  206749 rev 3  2012 05   eXam  is a registered trademark of Kaltenbach  amp  Voigt  GmbH    All rights reserved     Documentation  trademark and the software are copyrighted  with all rights reserved  Under the copyright laws the  documentation may not be copied  photocopied  reproduced   translated  or reduced to any electronic medium or machine  readable form in whole or part  without the prior written  permission of PaloDEx Group OY     The original language of this manual is English     PaloDEx Group OY reserves the right to make changes in  specification and features shown herein  or discontinue the  product described at any time without notice or obligation   Contact your PaloDEx Group OY representative for the most  current information     CE    0537    PaloDEx Group OY  Nahkelantie 160  P O  Box 64   Fl
20. Imaging plate unit Introduction    NOTES    Only personnel trained and authorized by the manu   facturer of the unit are allowed to install and config   ure the unit     Only use the imaging plates  protective covers and  hygiene bags supplied by the manufacturer of the  unit     Please read the section 7  Warnings and preca   tions before using the unit     1 2 System installation    Positioning the unit    Do not position the unit in direct sunlight or near  bright light  Sunlight or bright light must not be al   lowed to shine directly on the unit door into which  the IPs are inserted     Position the unit on a stable flat surface so that vi   brations will not degrade the image quality    The unit can also be attached to a wall  under or on  a shelf with the optional mounting kit     The unit must not be positioned so that it it touching  other equipment  It must not be placed on top of or  under other equipment     The unit can be positioned within the environment in  which the patient is examined and treated  patient  environment      2 User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Introduction Imaging plate unit    Positioning the PC    The PC connected to the unit should not be used in  the patient environment    The minimum horizontal distance between the pa   tient and the PC is
21. Mi EHE 28  5 2 Disinfecting DEDI iaaeaie a i aa 28  5 3 MING VANCE eR EET EET 29  BE REDON reren T e aaa e pL 29  WIS DOS eT pP 29   6  Symbols that appear on or in the unit                               seeeeusssssse 30   7  Warnings and DIecallOfiss cnc vesutVuenda uatinedesuuitced ladanddduebsbelerencents 31   Appendix A  Technical Specifications                             ssseseeeeeeeesssss 1  NIU e REDE 1  A 2 Imaging Plates and hygiene bags                      c  seeeseeseeeeeeeeeeeeeeeeeeeeeeeeeeeentes 2  A 3 Malri dimiernst  riBasuicecdep ei cin CIS dud tondadis ase cE edad aac dedpdN uU ANDER REEL CE  3   VI User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit    Appendix B  Installation and setup                                       eeeseeeessssss 1  HT  SysiemnsiallallOi iocos ode e i oet ofeca tono doce us encre a cete Ln e doa dede Less 1  BT  T Positioning Ine OPI  ac io Pero a reo Le Lee acea eb E EIE IOS EE RR do eed 1  B1 2 Positioning ihe PC s   not supplied  1 2  erre te ee Stone deesset 1   B1 3 Other devices mee eC EUN 2   B2  Connecting the unit to a PC   LAN    etate enne tuu iub 2  B2 1 Direct connection method  uses the unit s n                                      2   B2 2 IP method  using the unit IP address                                    suus
22. Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    2  Intraoral imaging plate unit Imaging plate unit    2 2 Accessories  For additional information about the accessories  listed below contact your authorized dealer  Not all  accessories are available for all units     Imaging plates  IPs   Equivalent to film sizes O  1     IDOT version  An identification code  IDOT  is print   ed on the IP and will appear on the intraoral image   The IDOT identification mark allows the IP used for  an exposure to be easily identified and removed if it  damaged     L3    EN Standard  STD  version  The standard version has  no identification mark               Protective covers  For IPs 0  1  2 and 3     S  j    Hygiene bags  For IPs 0  1  2 and 3        User s manual 206749 5    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 2  Intraoral imaging plate unit    CAUTION    For optimum performance only use IPs  protective  covers and hygiene bags supplied by the manufac   turer of the unit or the manufacturer s authorized  distributors    The manufacturer will not be held responsible for  problems caused by using accessories from other  manufacturers     IP storage b
23. at the     30   dip in U7   for 25 cycles     30   dip in Ur   for 25 cycles    eXam5 be powered  from an uninterruptible  power supply or a         5 96 Ur   5   Ur battery     gt 95   dip in Ur    gt 95 96 dip in Ur   for 5 sec for 5 sec  Power frequency   3 A m 3 A m Power frequency     50 60 Hz   magnetic field  IEC 61000 4 8             magnetic field should be  at levels characteristic  of a typical location in a  typical commercial or  hospital environment        NOTE Uris the AC mains voltage prior to application of the test level              User   s manual 206749 A 5  Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit A  Technical Specifications       Guidance and manufacturer s declaration     electromagnetic immunity       The eXam5 is intended for use in the electromagnetic environment specified below  The  customer or the user of the eXam5 should assure that it is used in such an environment        Immunity IEC 60601 test   Compliance Electromagnetic environment   guidance  test level level       Portable and mobile RF communications  equipment should be used no closer to any  part of the eXam6  including cables  than the  recommended separation distance calculated  from the equation applicable to the frequency  of the transmitter    Recommended separation distance    Conducted   3 Vr
24. covers or hygiene bags from other manufacturers     e This unit  or its accessories  must not be modified  altered or remanufactured  in any way     e Only the manufacturer s authorized service personnel are authorized to carry  out maintenance and repair of the unit  There are no user serviceable parts  inside the unit     e Infection control procedures must be observed when using accessories  such  as film holders  x ray tube guides and imaging plates  When using accessories  always follow the manufacturer s instructions on how to use the accessory and  prevent cross contamination from one patient to another     e This unit can interfere with other devices due to its EMC characteristics   e Other devices can interfere with this unit due to their EMC characteristics       This unit complies with IEC 60601 1 standard  Accessory equipment connect   ed to this device must be in compliance with the related nationally harmonized  IEC standards       Unit not suitable for use in the presence of flammable anaesthetic mixture with  air or with oxygen or nitrous oxide     User s manual 206749 31    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 7  Warnings and precations    e For ethernet connections  use an unshielded CAT6 LAN cable  so that multiple  chassis must not be connected  The PC   Ethernet s
25. e PCs Ether   net connector s  on the NIC  If the LED is not on  replace the NIC     PROBLEM  The unit display is not active but the power on   off  status light is on  green or yellow     CAUSE   SOLUTION  e Green LED   defective display or its cabling   Replace    e Yellow LED   unit is in standby  energy save   mode     Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    
26. ed and profitability increased    Problems caused by manually positioning the IP  include     incorrect vertical alignment  distortion  cone cut off    poor projection standardization    inferior image quality     For more information on imaging plate holders and  systems contact you dealer     User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22    13    See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 2  Intraoral imaging plate unit    Taking an exposure    1  Place the IP  in its sealed hygiene bag  into the  appropriate imaging plate holder and then insert  it into the patient s mouth in the position for the  image you wish to take    Note that the back of the sealed hygiene bag   the black side  must face the X ray source        2  Select exposure values appropriate for the ex   posure you are taking     The system will produce excellent images even  if the exposure values differ considerably from  the optimum values  In most cases the same  exposure values can be used for virtually all  imaging purposes     For normal everyday use select the Adult  Bitewing exposure time from the following  table    If required the exposure time can be increased  for very large patients and reduced for children     The optimum exposure values also depends on  the performance of the X ray unit being used  and may vary by  1 step from the
27. er options allow to select when the  scanner will start       after Plate insert  the scanner will automatically  start when it detects and IP in the plate carrier       after Cover removal  after the IP and protective  cover have been inserted into the plate carrier   the unit will automatically start when the protec   tive cover is removed     The Start delay options allow the start delay time to  be selected       Short   approximately 0 2 seconds    Medium   approximately 0 4 seconds      Long   approximately 0 6 seconds    Select Manual if you want the unit to start reading  an IP only when the Start key is pressed     Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28    D506723    B  Installation and setup Imaging plate unit    Touchless operation    Touchless operation    Scanner will wake up automatically from standby mode if touchless operation  sensor is in use and activated      v Use touchless operation sensor          If Use touchless operation sensor is selected the  unit will automatically  wake up  from the standby  mode as soon as the sensor detects a movement  within a distance of 20 cm     If itis NOT selected the unit will  wake up  when the  start is pressed     Plate eject mode    Plate eject mode     Drop in plate collector       The options are       Drop in plate collector  the IP is ejected  into the p
28. failure  the image  can be retrieved from the unit s memory as long  as the unit is NOT switched off    For information on how to do this see section  Retrieve last image     5  After the IP has been read it will be automati   cally erased and then ejected from the unit into  the plate collector     After the IP has been ejected the ready anima   tion will reappear on the display  The unit is now  ready to read the next IP     N    Removing IPs from the plate collector    When removing IPs from the plate collector hold  them by their edges  Alternatively  pull the plate col   lector out of the unit and tip the IPs out onto a flat  clean surface  If they are not to be reused immedi   ately or within a short period of time  store them in  their storage box        18 User   s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    2  Intraoral imaging plate unit Imaging plate unit    Standby mode    If no IP is inserted into the unit within a certain pe   riod of time  the unit will beep several times and the  status light will start to flash  When it turns yellow  the unit is in the standby mode     The unit will automatically exit the standby mode as  soon as a new imaging plate is inserted into the unit  or when the Start key is pressed     After four hours the unit will exit the standby mode  and automat
29. field in   v Enable changing IP addi the Scanner IP address area      192 168 2 1  NOTE  The PC and the unit must be in the    same subnet when setting the IP address of the  unit        Scanner IP address    on the unit and then click the Send to Scanner       3  PC   Unit  Press and hold down the Start key  button on the settings window     Send to Scanner         You will hear a beep which indicates that the  PC is now sending the IP address the unit  The  Startup animation will appear on the unit display    Startup     gt   animation    Name Serial number FW x xx  CORE xxx IP XXX XXX XXX XXX       4  PC  When the Startup animation clears click  OK to connect the PC to the unit     5  Now connect the other PCs in the network to the  unit  Just enter the IP address into the IP field  and then click OK to connect the PC to the unit   itis not necessary to hold down the Start key  and click the Send to Scanner button with the  other PCs once the unit has already got an IP  address      Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit B  Installation and setup    B2 3 Express Share    1  PC  If the unit is to be used with several PCs  select the Use Express Share check box and  select a unique Workstation identifier number   between 1 and 8   for the PC being configured   from the drop down list  
30. he system for multiple users     PROBLEM    The unit is connected to the PC but cannot scan  images     CAUSE   SOLUTION    i  Reserve the unit and check that it is configured  for multiple users  If not reconfigure the system     ii  Ifthe Dental imaging software animation   flashing yellow  appears it indicates that a  patient has not been selected for the scanned  image s   Select a patient     iii  If the Insert animation appears on the unit dis   play while inserting the IP but the unit does not  insert the IP into the unit nor start scanning  the  reflective sensor in the unit door is faulty  refer  to the service manual         Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    B  Installation and setup Imaging plate unit    PROBLEM  Local area network  LAN    subnet configuration  problem    CAUSE   SOLUTION    i  Ifthe Dental imaging software animation   flashing yellow  appears and cable s    router s   are known to be okay ask your network adminis   trator for assistance        ii  Conflict   mismatch IP addresses and   or sub   net mask of the unit   PC   Use command   ipconfig   from the command  prompt  Start      Run      cmd in Windows  to  get information about the active network set   tings   Change either the IP address of the PC or the  unit so that they are both in the same subnet   If the proble
31. ibutors    The manufacturer of this unit will not be held re   sponsible for any problems caused by using acces   sories from other manufacturers    Proper handling  cleaning and storage of the IPs  ensures the best image quality and maximum serv   ice life of the IPs  Refer to section 4  Handling and  care of imaging plates     Preparing the system  1  PC  Switch on the PC connected to the unit     2  PC  Open the dental imaging software and a  new or existing patient card where you wish to  store the intraoral images    NOTE    If you are using the system for the first time you  may wish to check and   or change the setup op   tions  See section 3  Setup options     O 3  Press the on off key to switch the unit on     The startup animation will appear on the display  and the unit will carry out self test during which  the IP carrier will slide out of the unit door     Startup     gt   animation    Name Serial number FW x xx  CORE xxx IP XXX XXX XXX XXX       10 User   s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    2  Intraoral imaging plate unit Imaging plate unit    When the status light turns green and ready ani     mation  indicating IP insertion  appears on the  unit display  the unit is ready to use  in the ready  state     NOTE     If the ready animation does not appear  check  the syste
32. ically shut down     Retrieve last image    If the last image read is not transferred to the PC  because of a network  communication  PC or soft   ware failure  the last image read can be retrieved     IMPORTANT NOTE   The LAST read image can only be retrieved if the  unit is left on  If the unit is switched off the image  will be lost     To retrieve the last scanned image     1  Correct the problem that caused the communi   cation failure  When the connection between the  unit and the PC is re established the last read  image will automatically be transferred to the  PC     2  PC  Ifthe image is not automatically transferred  to the PC  select the Setup  gt  Scanner page  from the dental imaging software your are using     3  PC  In the Last Image field  click the Retrieve  now button to retrieve the last read image     User s manual 206749 19    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 2  Intraoral imaging plate unit    NOTE    If required you can select different parameters   e g  resolution  show image preview etc   for  the image to be retrieved     4  PC  Click OK to close the Setup window   The last read image will be transferred to the  PC     Shutting down the unit  O 1  Press and hold power on   off key until the indi        cator light goes off     NOTE    If there is an untra
33. ifier of the PC  1   8  being  used will appear on the unit display        ES    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit B  Installation and setup    B3 Setup options    The scanner setup options allow you to configure  the scanner to the user s clinical requirements     B3 1 Scanner    1  From the dental imaging software you are using  select Setup   Scanner page     KaVo Scan eXam Setup       m Status    Scanner  KaVo Scan eXam    Version  6 11          m Image Scanning    Show Image Preview and Dental Chart    Resolution  C Super   High          m Image Processing   v Noise Filtering                   m Last Image  Retrieve latest image from the scanner unit memory             Add serial number to new images                Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    B  Installation and setup Imaging plate unit    Status      Settings   Workflow   Power options   Driver Info    Status  Scanner  KaVo Scan eXam    Version  6 11       Shows the scanner type  version and serial number     Image Scanning    Image Scanning    Show Image Preview and Dental Chart    Resoktin    Super 6 figh    poer mw       If Show Image Preview
34. images  with less noise especially if short exposure times  are used     2  Click OK to close the window and accept the  selected resolution     3 3 Image Processing    1  In the Image processing click the Noise Filter   ing check box   Noise filtering is normally always selected  It  makes images smoother when they are taken at  lower radiation doses     2  Click OK to close the window and select the  noise filtering     24 User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    4  Handling and care of imaging plates Imaging plate unit    4  Handling and care of imaging plates    The correct use  handling  cleaning and storage  of imaging plates guarantees the best image  quality and maximum service life of the imaging  plates     4 1 General    e X ray exposures DO NOT cause IPs to age       The light coloured side of the IP is the side  which  stores  the x ray image  It is the SENSI   TIVE side       Image information  stored  on an IP after an  exposure is sensitive to light     4 2 Handling      Handle IPs in the same way as you would han   dle CDs or DVDs     e Hold IPs by their edges using your finger tips       Protect the sensitive side of the IP from     scratches  wear and mechanical damage    stains  dirt  dust and fingerprints    NOTE    Stains   fingerprints on the sensitive side can
35. ion     28 User    manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    5  Unit care and maintenance Imaging plate unit    5 3 Maintenance  The unit does not require any maintenance     5 4 Repair    If the unit is damaged or malfunctions in any way it  must only be repaired by service personnel author   ized by the manufacturer of the unit     5 5 Disposal    At the end of the useful working life of the unit and   or its accessories make sure that you follow national  and local regulations regarding the disposal of the  unit  its accessories  parts and materials  The unit  includes some or all of the following parts that are  made of or include materials that are non environ   mentally friendly or hazardous      electronic circuit boards     electronic components    User s manual 206749 29    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 6  Symbols that appear on or in the unit  6  Symbols that appear on or in the unit    DANGEROUS VOLTAGE  LASER RADIATION    IMPORTANT INFORMATION  Refer to user s man   ual     Ll      gt  e e    Direct current    a Eternet connector RJ45 straight cable    CE  0537  Symbol MDD 93 42 EEC
36. lassification   CLASS 1 LASER PRODUCT  EN 60825 1  2007    Dimensions  H x W x D  196 mm x 215 mm x 382mm  7 7in x 8 5in x 15 0in     9 8     10 3 kg  21 6     22 7 Ib     24 VDC  PSU  100     240 VAC  50 60 Hz   Less than 1 5 A   35 um  Super   64 um  High    14 bits grayscale    Interface cable For Ethernet connections  use an unshielded CAT6  LAN cable  so that multiple chassis must not be  connected  The PC   Ethernet switch to which unit is  connected to should be approved appropriately  e g   EN 60950  IEC 60950  UL 60950      Operating environment  10  C    40  C  30     90 RH   700     1060 mbar  Storage   transportation  10  C      50  C  0     90 RH   500     1080 mbar  environment       User   s manual 206749 A 1    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit A  Technical Specifications  A 2 Imaging Plates and hygiene bags    Imaging plates  IP     Dimensions  mm  27 x54  Image size  pixels   35 um 771 x 1542   mage sie  CS   em  us     jus ooo fes    Storage environment Imaging plates must be stored in their protective box below  33  C  The box must be kept closed to remain dust free     Photo stimulable phosphorous material uniformly coated  on a support plastic material  Shielded with a top coat layer  on the active surface and encapsulated with lacquer  around edges    Imaging 
37. late collector after the IP has been  scanned       Leave in plate carrier  the IP remains in the  plate carrier after the IP has been scanned   The Leave in plate carrier option is recom   mended for users who want to handle the IPs  with more care and reduce wear and tear on  them  This option extends IP service life and al   lows greater hygiene standards to be observed     Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit B  Installation and setup    B3 4 Power options    1  From the dental imaging software you are using  select Setup   Power options page        Power options  Standby after  seconds    600 Y       v Beep when entering standby mode    Shutdown after  minutes  240 v          Power options    Standby after  seconds   Allows you to select the  length of time the scanner will remain unused be   fore it enters the standby mode  Display switches off  and power LED turns yellow      Beep when entering standby mode  Select this  and audible signal will be heard when the scanner  enters the standby mode     Shutdown after  minutes   Allows you to select the  length of time the scanner will remain unused be   fore the unit automatically switches itself off     Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to
38. m persists ask your network admin   istrator for assistance     iii  Network connection not functioning or not stable  due to LAN signal quality problems   Install a network switch between the unit and PC  or network socket     PROBLEM   Ethernet link not active    The Ethernet connection consists of correctly in   stalled cabling and any switches and or hubs used   basically the complete Ethernet link  If the link is  active it does not necessarily mean that the unit is  physically connected to PC  It only means that the  unit is physically connected to something that is  Ethernet compatible  hub  switch  another PC etc      Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit B  Installation and setup    CAUSE   SOLUTION   First remove and then reconnect the Ethernet cable  between the unit and PC to see if this solves the  problem     i  Physical connection is okay  link active   but the  Dental imaging software animation  flashing  yellow  appears on the unit display  The unit is  not configured correctly  Reconfigure     ii  Cabling not correct  link not active   and the  Unit connection animation appears on the unit  display  The unit is not physically connected to  the PC  Connect the unit to the PC        iii  The PCs Ethernet connector s  is faulty  There  is usually a green link LED near th
39. m setup described in the installation  instructions     Preparing an IP for exposure  IMPORTANT NOTE  If the IP is being used for the very first time or if it  has not been used within the last 24 hours  it must  be erased before use to remove any fogging caused  by background radiation  See section  Erasing an  imaging plate     CAUTION    When handling IPs  protective covers and hygiene  bags take all appropriate measures and precautions  to prevent cross contamination     1  Place the IP you wish to use onto a protective  T ad EPIS   cover  The light blue side  sensitive  of the IP  i      en must face and be placed on the half of the pro   M     1               4 4 tective cover that is the same shape as the IP     Fold the half of the protective cover with the  semicircular cutout over onto the IP  The metal  disk on the back  black side  of the IP must ap   pear in the semicircular cut out        User s manual 206749 11    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 2  Intraoral imaging plate unit    3  Turn the protective cover and IP over so that the  black side of the protective cover is uppermost   This makes it easier to slide the protective cover  and IP into the hygiene bag        4  Slide the protective cover and IP into the hy   giene bag as far as they will go  Make sure that
40. ms 3V d  12NP  RF  Eesi  a d 1 24 P 80 MHz to 800 MHz  4 6  d 2 34 P 800 MHz to 2 5 GHz  Radiated where P is the maximum output power rating  RF 3 V m 3 V m of the transmitter in watts  W  according to the  IEC 61000    80 MHz to transmitter manufacturer and d is the  4 3 2 5 GHz recommended separation distance in    metres  m      Field strengths from fixed RF transmitters  as  determined by an electromagnetic site survey     should be less than the compliance level in  each frequency range    Interference may occur in the vicinity of  equipment marked with the following symbol      t     NOTE 1 At 80 MHz and 800 MHz  the higher frequency range applies     NOTE 2 These guidelines may not apply in all situations  Electromagnetic propagation is  affected by absorption and reflection from structures  objects and people                      Field strengths from fixed transmitters  such as base stations for radio  cellular cordless   telephones and land mobile radios  amateur radio  AM and FM radio broadcast and TV  broadcast cannot be predicated theoretically with accuracy  To assess the electromagnetic  environment due to fixed RF transmitters  an electromagnetic site survey should be considered   If the measured field strength in the location in which the eXam5 is used exceeds the applicable  RF compliance level above  the eXam5 should be observed to verify normal operation  If  abnormal performance is observed  additional measures may be necessary  such as reorienting  of rel
41. nsferred image in the unit s  memory the unit cannot be shut down  The im   age must be transferred first  Refer to section  Retrieve last image     IP erasing mode  Initial erasing of the IPs   If an IP is being used for the very first time or has  not been used for 24 hours and has not been ex   posed to ambient light the erasing procedure must  be carried out before the IP is used  The erasing  procedure removes any fogging due to background  radiation   The dedicated erasing mode erases the IP but does  not send the image  blank  to the dental imaging  software     20 User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    2  Intraoral imaging plate unit       User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22    Imaging plate unit    Press and hold down the Start key for several  seconds until the erase mode animation ap   pears on the unit display  The unit is now in the  erase mode     Hold the IP by its edges  or use a protective  cover  and position it so that the side with the  metal disk  the back  is on the right    Insert the IP into the unit door  remove the pro   tective cover if used   A magnet will hold the IP  in position and then automatically slide it into the  unit     NOTE    It may take longer to erase IPs using
42. o  clean IPs    Unsuitable cleaning solutions methods may  damage or destroy the IPs or leave residues on  the sensitive surface that may appear on the  images     26 User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    4  Handling and care of imaging plates Imaging plate unit    4 4 Storage    e Unpacked  exposed to ambient light in the dedi   cated storage box    e Below 33  C   80  RH and shielded from X rays  and ultraviolet radiation       Ifan IP is stored for over 24 hours in a hygiene  bag or in a location that is shielded from ambi   ent light  the IP must be erased  to remove any  potential fogging  before being used to take an  exposure     4 5 Replacement  Replace an IP if     e the image shows marks  spots or dots  which  still appear on images even after the IP has  been properly cleaned     e the IP is mechanically damaged  scratched  or  badly bent     4 6 Disposal    IPs must be disposed of in accordance with all local   national and international regulations regarding the  disposal of non environmentally friendly or hazard   ous materials     Phosphor substance  under the top coat  on the  sensitive side of IPs must not be swallowed     User s manual 206749 27    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group O
43. o connect the unit to the PC  using the direct connection method if another device  is already connected to the PC using direct con   nection  If the direct connection field is not active   greyed out  or the system does not work correctly  after the unit has been connected  reconnect the  unit using the IP connection method     1  After positioning the unit connect it to the PC s   in the local area network using the Ethernet  cable s   not supplied     2  Switch the unit on   The dental imaging software animation will ap   pear  This indicates that the unit is not communi   cating with the PC s  in the network        3  PC  Install the dental imaging software to be  used in the PC s      Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    B  Installation and setup Imaging plate unit    4  PC  Open the dental imaging software and se   lect the scanner setup window        Scanner   Settings   Driver Info         Status  Scanner  KaVo Scan eXam  Version  6 11 Serial No  KJ0009001             5  PC  From the scanner setup window select the  Settings tab to open the Scanner Connection    page        Intel R    amp 2557LM Gigabt Network Connection    6  PC  Select Direct Connection     Intel FE   amp 2557LM Gigebt Network Connection       Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14
44. oDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 2  Intraoral imaging plate unit    rJ     3  Imaging plate  l  Yellow  wrong way round  rotate    Protective cover        Remove   disconnect    Insert   connect    Busy  Unit in operation        Progress   animation   QT Unit in erasing mode  Check  2  Something wrong or take alternative action  ri  8 User s manual 206749   Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21   D506723    Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28    2  Intraoral imaging plate unit Imaging plate unit    Dental imaging software  Software not open  not ready or waiting for user ac   tion     Unit connection    Not connected or connection not working     Rotate    Error state and error number    SEDE    Error       Check documentation supplied with the unit         Unit in service mode    E   Service technicians only     i    User   s manual 206749 9    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 2  Intraoral imaging plate unit    2 4 Using the system    For optimum performance only use IPs  protective  covers and hygiene bags designed for this unit and  supplied by authorized distr
45. ocating the eXam5       Over the frequency range 150 kHz to 80 MHz  field strengths should be less than 3 V m              A 6 User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    A  Technical Specifications Imaging plate unit       Recommended separation distances between portable and mobile RF communications  equipment and the eXam5        The eXam5 is intended for use in an electromagnetic environment in which radiated RF  disturbances are controlled  The customer or the user of the eXam5 can help prevent  electromagnetic interference by maintaining a minimum distance between portable and  mobile RF communications equipment  transmitters  and the eXam5 as recommended  below  according to the maximum output power of the communications equipment        Rated maximum    Separation distance according to frequency of transmitter m       output power of  transmitter W    150 kHz to 80 MHz    d 124P    80 MHz to 800 MHz    d 1 2  P    800 MHz to 2 5 GHz    d 234P                            0 01 0 12 0 12 0 23  0 1 0 38 0 38 0 73  1 1 2 1 2 2 3  10 3 8 3 8 7 3  100 12 12 23       applies        For transmitters rated at a maximum output power not listed above  the recommended  separation distance d in meters  m  can be estimated using the equation applicable to the  frequency of the transmitter  where
46. ox  For storing IPs safely and conveniently       Microfiber cloth  For cleaning IPs    IP holders   For bitewing  periapical and endodontic exposures   optional  not included    See section Imaging plate holders for more infor   mation        6 User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    2  Intraoral imaging plate unit Imaging plate unit    2 3 Display symbols and what they mean    During use symbols and animations will appear on  the unit display  These   indicate the status of the unit    help you to operate the unit correctly    Show user mistakes and corrective actions    display error codes    display a preview image    The main symbols are   Startup    During startup the unit serial number  IP address  Startup and other information will appear on the unit display      gt   animation    Name Serial number FW x xx  CORE xxx IP XXX XXX XXX XXX       Express Share wait  Express Share configuration  The unit is not re   served by any PC in the system     Express Share reservation  Express Share configuration  The unit has been  reserved by a PC  e g  PC number 2         Unit door  Protective cover and imaging plate    Yellow  remove protective cover    User   s manual 206749 7       Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the Pal
47. plates do not include phosphor     The typical service life for an imaging plate is several  hundreds of cycles provided that the imaging plate is  handled with care and according to the supplied   instructions  The use of genuine hygiene accessories     protective covers and hygiene bags  will extend the  service life of the imaging plates     Disposal Imaging plates are industrial waste and must be disposed  of in accordance with local and national regulations  concerning the disposal of such material    Never use damaged imaging plates     Hygiene bags     Material       Food grade polyethylene   Latex free   Packaging Supplied in boxes     For the best performance it is recommended the hygiene bags are used  within two years from the date of manufacture  The date of manufacture  is printed on the bottom of the box containing the hygiene bags   DDMMYYXX      Extended storage time or exceeding the specified storage conditions  may compromise the performance of the adhesive tape and or the  plastic material from which the hygiene bags are made     Disposal Observe relevant national requirements        A 2 User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    A  Technical Specifications Imaging plate unit    A 3 Main dimensions       User s manual 206749 A 3    Reviewed  Taskinen Jari Pekka 29 05 
48. s the unit serial number to all new images     Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    B  Installation and setup Imaging plate unit    B3 2 Settings    See B2 Connecting the unit to a PC   LAN for  information on Settings     B3 3 Workflow    1  From the dental imaging software you are using  select Setup   Workflow page     Scanner   Settings Workflow   Power options   Driver Info         Start delay       Manual  Start readout by pressing unit s START key        m Touchless operation    Scanner will wake up     standby mode if touchless operation  sensor is in use and activated    Iv eases es ee          m Plate eject mode     Drop in plate collector                m Restore factory settings  Resets the Workflow  Power options and Hygiene values to their factory  settings     Restore factory settings                  Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  Aarynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit B  Installation and setup          Readout start                  Start after  Start delay     Plate insert v     Medium v      C Manual  Start readout by pressing unit s START key     Select Automatic if you want the scanner to start  automatically    The Start aft
49. several setup options in the dental imag   ing software that allow you to set the image quality  to your requirements    To select the setup options     1  From the dental imaging software you are using  select Setup    Scanner page     3 1 Tooth numbering    Numbers can be assigned to the teeth that appear  on the image     Selecting the tooth numbering feature    1  In the Image Scanning section  click the Show  Image Preview and Dental Chart check box     2  Click OK to close the window and activate tooth  numbering     Using the tooth numbering feature    After an IP has been read a window will open that  shows the image and a tooth location map     1  Click the tooth teeth on the map that corre   spond to the tooth teeth in the image  Tooth  numbers will be assigned to the selected teeth     The tools at the top of the window allow the im   age to be manipulated     2  Click OK to save the tooth numbers with the im   age     User s manual 206749 23    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 3  Setup options    3 2 Resolution    1  In the Image Scanning section  select either  Super or High  default    Super gives a pixel size of 35u  This results in  images with better resolution  but more memory  is required as the image files are larger   High a pixel size of 64u  This results in 
50. t be al   lowed to shine directly on the unit door into which  the IPs are inserted     The unit must not be positioned so that it touching  other equipment  The unit must not be placed on top  of or under other equipment     The unit can be positioned within the environment in  which the patient is examined and treated  patient  environment      B1 2 Positioning the PC s   not supplied     The PC s  connected to the unit should not be used  in the patient environment    The minimum horizontal distance between the pa   tient and the PC s  is 1 5 m  4 5 ft     The minimum vertical distance between the patient  and the PC s  is 2 5 m  6 5 ft      Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit B  Installation and setup    B1 3 Other devices    DO NOT connect any other devices to the unit or  the PC s  connected to the unit that are       not part of the supplied system    not supplied by the manufacturer of the unit      not recommended by the manufacturer of the  unit     B2  Connecting the unit to a PC   LAN    The procedure for connecting the unit to a single  PC or several PCs in a local area network  LAN  is  exactly the same except that every PC in the LAN  needs to be given a unique ID number     B2 1 Direct connection method  uses the unit s n     IMPORTANT NOTE    It may not be possible t
51. tion     electromagnetic immunity       The eXam6 is intended for use in the electromagnetic environment specified below  The  customer or the user of the eXam5 should assure that it is used in such an environment        Immunity test    IEC 60601 test level    Compliance level    Electromagnetic  environment   guidance       Electrostatic  discharge  ESD   IEC 61000 4 2     6 kV contact     8 kV air     6 kV contact     8 kV air    Floors should be wood   concrete or ceramic tile   If floors are covered with  synthetic material  the  relative humidity should  be at least 30          Electrical fast     2 kV for power supply     2 kV for power    Mains power quality          transients bursts   lines supply lines should be that of a  IEC 61000 4 4  1 kV for input output  1 kV for typical commercial or  lines input output lines hospital environment   Surge  1 kV differential mode    1 kV differential Mains power quality  IEC 61000 4 5 mode should be that of a  typical commercial or  hospital environment   Voltage dips   lt 5   Ur  lt 5   Ur Mains power quality  short   gt 95 96 dip in U7    gt 95   dip in U7  should be that of a  interruptions and   for 0 5 cycle for 0 5 cycle typical commercial or  voltage variations hospital environment  If  on power supply   40   Ur 40   Ur user of the eXam5  lines  60   dip in Ur   60   dip in Ur  requires continued  IEC 61000 4 11 for 5 cycles for 5 cycles operation during power  mains interruptions  it is  70   Ur 70 96 Ur recommended th
52. witch to which unit is con   nected to  should be approved appropriately  e g  EN 60950  IEC 60950  UL  60950   After installation check that the IEC 60601 1 leakage current levels are  not exceeded     e In order to maintain safe and correct functioning of the unit  only the power  supply unit  PSU  delivered with the unit or distributed by authorized dealers   Please refer to the unit technical specifications for a list of the PSUs     e If this device will be used with 3rd party imaging application software not sup   plied by the manufacturer  the 3rd party imaging application software must  comply with all local laws on patient information software  This includes  for  example  the Medical Device Directive 93 42 EEC and or FDA if applicable       Medical electrical equipment needs special precautions regarding EMC and  needs to be installed according to EMC information     32 User s manual 206749    Reviewed  Taskinen Jari Pekka 29 05 12 11 01 21  Approved  A  rynen Teemu 29 05 12 14 48 22  See the PaloDEx Group Oy PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    A  Technical Specifications Imaging plate unit  Appendix A  Technical Specifications    A 1 Unit    Intraoral imaging plate unit    eXam5   Class 1 or 2 equipment depending on the classification  Classification IEC60601 1   Of the PSU  No applied part    Continuous operation      IPXO  enclosed equipment without protection against  ingress of liquids    Laser Safety C
53. y PDM system to determine the status of this document  Printed out  12 07 12 12 51 28 D506723    Imaging plate unit 5  Unit care and maintenance    5  Unit care and maintenance    WARNING   Switch the unit off and disconnect it from the main  power supply before cleaning or disinfecting the  unit    Do not allow liquids to enter the unit     5 1 Cleaning the unit    Use a non abrasive cloth moistened with either      cool or lukewarm water      soapy water      mild detergent      butylalcohol      or ethanol  ethyl alcohol  70   9696   to clean the unit  After cleaning wipe the unit with a  non abrasive cloth moistened with water     Never use solvents or abrasive cleaners to clean  the unit  Never use unfamiliar or untested cleaning  agents  If you are not sure what the cleaning agent  contains  DO NOT use it    If you use a spray cleaning agent DO NOT spray it  directly into the unit door     5 2 Disinfecting unit    CAUTION  Wear gloves and other protective clothing when  disinfecting the unit     Wipe the unit with a cloth dampened with a suitable  disinfectant solution such as ethanol 9696  Never  use abrasive  corrosive or solvent disinfectants  All  surfaces must be dried before the unit is used     WARNING  Do not use any disinfecting sprays as the vapor  could ignite and cause injury     Disinfecting techniques for both the unit and the  room where the unit is used must comply with all  local and national regulations and laws concerning  such equipment and its locat
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
User manual for the measuring program ic.exe (ISee!)  info document  MA2012-6  Blue Microphones Spark Digital Studio Microphone for Mac/PC + iPad 830 User's Manual    GBC LEATHERGRAIN COVERS IVORY (100)  SAM 9.1 – MANUAL NOVAS FUNCIONALIDADES  取扱説明書 (3.99 MB/PDF)  DN482/483 - Pdfstream.manualsonline.com  Traitement des ulcères veineux – La thérapie de compression    Copyright © All rights reserved. 