TURBINA 405 inglese
Contents
1. 34 INSTALLATION Check to make sure following items are received Dental turbine Universal wrench Pins Operating manual e DEVICE CAN BE INSTALLED BY USER e Make sure that compressed air used is dry and oilfree filtered e Screw S405 turbine to cable or to Quick connector Insert Quick Connector in turbine lengthwise by pressing the two push buttons IMPORTANT e The instrument must not be started without a tool inserted into the chuck 35 OPERATING INSTRUCTIONS Try turbine to make sure it rotates and that air and water come out of holes spray To prevent exhausted air from striking hand S405 turbines with Borden coupling are made to O exhaust into cable FARO prepared cables If cable doesn t have this exhaust A remove O ring B using small tool fig 1 2er oe fig 1 36 Replacing burr SP405 turbine push button version To replace burr just press push button on turbine head fig 2 pull out old burr and fit new one The push button burr locking system is EJ designed to be fully effective when rotating To avoid damaging locking system do not attempt to remove burr with turbine stopped without pressing push button fig 2 37 MAINTENANCE Cleaning e This device can be sterilized in conformity with the parameters of the Official Pharmacopoeia Y for example 121 C 249 8 F to
2. 40 Cartridge replacement For easy cartridge replacement it is recommended C to leave burr in place and recover it later S Proceed as follows e Using universal wrench A supplied unscrew cover fig 5 then draw out cartridge by pressing from burr side e Insert new cartridge being careful to position reference pin in its seat fig 6 e Screw down cover using the same universal wrench fig 5 41 Changing light bulb To change light bulb L proceed as follows fig 7 L e Disconnect turbine from cable by pressing two small push buttons on quick connector e Unscrew sleeve M of turbine body counter clockwise and withdraw it up to stop M e Use tool as a lever under lamp disconnect lead and withdraw lamp e Fit new bulb being careful to snap it correctly into position in its seat e Screw down sleeve and connect cable fig 7 Cleaning nozzles If small holes from which emerge water and air are clogged free passage using pin S provided fig 8 It is recommended to carry out this operation regularly once a week lS fig 8 43 TROUBLESHOOTING Turbine sounds Check cartridge condition and change it if necessary see page 41 different it vibrates In case of difficulty see our Service Department too much and cutting power is reduced Optic
3. DECLARATION OF CONFORMITY The manufacturer FARO S p A Via Faro 15 20060 Ornago Milano ITALY declares under its own responsability that following product Product name Dental Office Turbine It fully conforms to the following European directives and national acceptances 93 42 CEE medical devices directive Class which the equipment belongs to Ila It conforms to the following standards CEI EN 60601 1 CEI EN 50081 1 CEI EN 50082 1 CEI EN 55022 Model S405 IEC 801 2 It is factory new in accordance with the provisions of art IEC 801 3 Dott Angelo Favonio 6 para 2 of D L 626 94 IEC 801 4 Managing Director 26 TABLE OF CONTENTS e Symbols used in this manual page 28 e Safety rules page 29 30 31 e Features page 32 33 e Description of parts page 34 e Installation pag 35 e Operating instructions page 36 37 e Maintenance pag 38 39 40 41 42 43 e Troubleshooting page 44 45 e Data sheets page 46 47 48 e Guarantee certificate page 49 27 SYMBOLS USED IN THIS MANUAL Danger The paragraphs marked with this symbol contain instructions to be followed carefully to avoid damage to the device the operator and the patient Warning These instructions warn to be very careful to avoid situations which could damage the device Prohibition This icon identifies what to prevent doing to avoid damage to the device Suggestions With this icon information is
4. Power input W a gt AT 47 CLASSIFICATION OF SL405 turbine with light DEVICE e Device of Class II e Device of Type B e Device not protected against liquid penetration e Device not suited for use in the presence of anasthetic mixtures inflammable with air oxygen or nitrous oxide Q To obtain a safe system the use of a FARO electronic circuit is recommended 48 GUARANTEE CERTIFICATE FARO offers the final customer a 12 months guarantee for the turbine and 6 months guarantee for the rotor starting from the date of purchase Repairs under guarantee must be performed at FARO expenses and transportrisks are at the risk of the purchaser Repair under guarantee is considered valid only when all sections of the certificate have been filled in and sent in advance to FARO by Fax 039 6010540 The guarantee covers faults due to the bad quality of the material or manufacturing defects in the case of valid claims the guarantee covers free repair or replacement Claims for damages and or interest are excluded The guarantee is not considered valid at the sole discretion of FARO if the fault is due to tampering damage incorrect use improper maintenance and normal wear and tear 49
5. supplied permitting more effective use of the device SAFETY RULES Follow all instructions set forth in this manual dental operations e Device is designed for use only in a dental office by medical personnel to perform specific e Device must be sterilized before use each time e Device is not delivered sterilized therefore it must be sterilized before use 121 C 249 8 F to 135 C 275 F e Sterilization parameters must be in conformity with Official Pharmacopoeia for example e Do not immerse turbine in any disinfecting liquid e Device must always be washed before sterilization see page 38 29 Device must be installed by user in a specific power supply system dental unit and be connected to a cable Never press button when turbine is still rotating Systems of fastening to cables meet EN standards 29168 Fit in the turbine only burrs with stems meeting ISO standards 1797 in turbine Make sure burr is perfectly seated Avoid extended positionning of light in patient s eyes turbine with light S405 version with light is not protected against penetration of liquids Do not attempt to repair turbine yourself Only perform operations set forth in this manual For any work not described in this manual apply to our Service Department Turbines of series S405 must be used only for purpose described 30 When the equipment is packed
6. 135 C 275 F e Office turbine must be sterilized after each patient e Before sterilization device must always be cleaned externally with a brush under running water O or with an alcohol based solution e S405 turbines can be disinfected externally with a chemical disinfectant disinfection with spray or with damp cloth e Do not leave S405 turbines immersed in a disinfection liquid e This device cannot be cleaned in thermodisinfector 38 Lubrication S405 turbines must be lubricated either daily and before and after each sterilization by spraying special FARO lubricant into rear other lubricants should not be used To lubricate turbines with Borden A or 4 way B connection proceed as follows fig 3 e Unscrew the ringnut connecting the cable to the turbine e Insert aerosol lubricant injector in hole 1 or 2 and spray in lubricant e Reconnect cable to handpiece and run device without load for a few seconds 39 To lubricate handpieces designed for a QUICK CONNECTION proceed as follows fig 4 e Disconnect turbine by pressing two push buttons on the connection e Fit lubricant aerosol bottle injector in rear of turbine and spray in lubricant e Remove aerosol injector from turbine and reconnect handpiece to quick connector Run turbine without load for a few seconds ALWAYS LUBRICATE TURBINE AFTER STERILIZATION fig 4
7. al fiber turbine Check condition of light bulb and change it if necessary see page 42 does not light In case of difficulty see our Service Department 44 No water or spray Disconnect turbine from cable and check if water come out If not see comes out of turbine our Service Department Check to see if hole in Turbine is clogged and if necessary clean it see page 43 In case of difficulty see our Service Department Burr locking system See our Service Department or rotor are damaged 45 DATA SHEETS Direct coupling SP405 turbine Borden 3 way 4 way Rotation Speed rpm 395 000 390 000 Max power output W gt 12 5 12 Operating pressure bar 2 2 0 2 2 2 0 2 Max air consumption l min 45 35 Max water consumption ml min 80 80 Noise level dB A lt 60 lt 60 Vibrations m s7 lt 25 lt 2 5 Head dimensions mm 12 4 H 15 12 4 H 15 Weight g en 74 46 Quick connection SP405 turbine SLP405 light turbine Borden 3 way 4 way lux Rotation Speed rpm 395 000 380 000 Max power output W gt 103 10 8 Operating pressure bar 2 4 0 2 2 4 0 2 Max air consumption l min 42 39 Max water consumption ml min 80 80 Noise level dB A lt 60 lt 60 Vibrations m s7 2 25 lt 2 5 Head dimensions mm 12 4 H 15 12 4 H 15 Weight g 68 68 78 Light bulb SLP405 turbine data sheets Power supply V 355 Current input mA 740
8. for shipment or stock it can lie for maximum 15 weeks to environment conditions that do not exceed the following ranges gt Sorrounding temperature of 20 C 4 F or 70 C 158 F gt Relative humidity from 10 to 90 Atmospheric pressure from 500 to 1060 mBar This turbine in version with light must be powered in conformity with CEI EN standard 60601 1 31 FEATURES Purpose of device is to provide movement to a rotating body i e a burr for performance of specific operations in dentistry The S405 turbine is produced in following models SP405 with push button Borden 4 way Quick Connection SLP405 with push button and light electrified Quick Connection Q ALL S405 TURBINES HAVE A CONTAMINATED LIQUID HOLDING SYSTEM e Turbines with burr replacement system by pressing push button provided on turbine head e Direct turbine versions with Borden and 4 way systems for connection to cable e Turbine version with Quick cable connection system using a Quick connector Borden 4 way electrified e Optical fiber version providing illumination of operating area e All S405 models can be sterilized in an autoclave in conformity with Official Pharmacopoeia for example 121 C 249 8 F to 135 C 275 F 33 Push button Optical fiber Borden Connection 4 way Connection Quick Connection DESCRIPTION OF PARTS nABWN re