Choicemed Pulse Oximeter User Manual
Contents
1. SpO Heart rate BPM v BPM BPM x Low power indication SN Serial No
2. Indication 1 The or 1 Normal lamps are is product 2 Replace the suddenly off automatically batteries powered off when no signal is detected longer than 8 seconds 2 Power quantity of the batteries is started being inadequate Error3 or 1 Low power 1 Change Error4 2 Receiving new battery Displayed on tube being 2 Please screen shielded or contact with damaged local together with customer broken service connector center 3 Mechanical 3 Please center 13 Detailed descriptions of product functions 13 1Display mode OLED 13 2 SPO2 Measurement range 70 99 Accuracy 2 on the stage of 80 99 2 on the stage of 70 80 13 3 Pulse Measure range 30 235 BPM Accuracy 2 BPM or 2 larger 13 4 Pulse Intensity Bargraph Indication 13 5 Power Requirements Two AAA alkaline Batteries 13 6 Battery life Two AAA 1 5V 600mAh alkaline batteries could be continuously operated as long as 30 hours 13 7 Outline dimension Length 58mm Width 32mm Height 34mm Weight 50g batteries 13 8 Environment requirements Operation Temperature 5 40 C Storage Temperature 10 40 C including two AAA Ambient Temperature 15 80 in operation 10 80 in storage Declaration EMC of this product comply with IEC60601 1 1 2 standard 14 Symbol Definitions Symbol Definition The equipment type is BF Refer to user manual A before application Hemoglobin saturation
3. Instruction Manual Fingertip Pulse Oximeter 1 General Description Haemoglobin Saturation is percentage of Oxyhemoglobin HbO2 capacity compounded with oxygen by all combinativable haemoglobin Hb obin HbO2 capacity in blood In other words it is consistence of Oxyhemoglobin in blood It is a very important ecological parameter for Respiratory circulation System Many respiratory diseases can result in haemoglobin Saturation being lowered in human blood Moreover the following factors can also lead to problems in oxygen supply so that human haemoglobin saturation might be reduced Automatic Organic Regulation Malfunction caused by Anesthesia Intensive Postoperative Trauma hurts resulted in by some medical examination and etc In the situation illnesses such as light head asthenia vomitory and etc might happen to patients and even endanger the patient s life Therefore it is very important to know Hemoglobin saturation of patient timely in clinical medical aspects So that doctors can find problems in time The Fingertip Pulse Oximeter features in small volume low power consumption convenient operation and being portable It is only necessary for patient to put one of his fingers into a _ fingertip photoelectric sensor for diagnosis and a display screen will directly show measured value of hemoglobin Saturation It has been proved in clinical experiments that it features in rather high precise and repeatabi
4. graph Switch PR Low power indicator SPO2 waveform It is apparent the heart rate bargraph display corresponds with pulse rate 8 Product Accessories 8 1 One hang lace 8 2 Two batteries 8 3 One User manual 9 Battery Installations 9 1 Put the two AAA batteries into battery cassette in correct polarities 9 2 Push the battery cover horizontally along the arrow shown as below Notes must be Battery polarities correctly installed Otherwise damage might be caused to device Please put or remove batteries in right order or is likely to damage the device bracket Please remove the battery if the Oximeter will not be used for long time 10 Hang Lace Installations 10 1 Thread thinner end of the hang lace through the hanging hole 10 2 thread thicker and of the lace through the threaded end before pulling it tightly 11 Maintenance and Storage 11 1 Replace the batteries timely when low voltage lamp is lighted 11 2 Clean surface of the fingertip Oximeter before it is used in diagnosis for patients 11 3 Remove the batteries inside the battery cassette if the Oximeter will not be operated for a long time 11 4 It is best to preserve the product in a place where ambient temperatures 10 40 C 14 104 F and humidity is 10 80 It is recommended that the product should be kept in a dry environment anytime A wet ambient might affect its lifetime and even might damage the Misplace for c
5. lity 2 Measurement principle Principle of the Oximeter is as follows An experience formula of data process is established taking use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin R Hb and Oxyhemoglobin O2 Hb in glow and near infrared zones Operation principle of the instrument is Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology so that two beams of different wavelength of lights 660nm glow and 940nm near infrared light can be focused onto human nail tip through perspective clamp finger type sensor Then measured signal can be obtained by a photosensitive element information acquired through which will be shown on two groups of LEDs through process in electronic circuits and microprocessor Diagram of Operation Principle 1 Glow and Infrared ray Emission Tube 2 Glow and Infrared ray Receipt Tube 3 Technical Specifications 3 1 Display type OLED display SPO2 display range 0 99 PR display range 30 254 BPM PR display mode bargraph Low power indication k 3 2 Power requirements Two AAA 1 5V Alkaline Battery 3 3 Power consumption less than 40mA 3 4 Resolution 1 for SPO2 and 1BPM for Pulse rate 3 5 Measurement Accuracy SPO2 80 99 2 70 80 3 lt 70 no definition PR 30 235 BPM 2 or 2 BPM 3 6 Measurement Performance in Weak Filling Condition requi
6. n in sport procedure is not recommended and etc The product is not suitable to monitor patient continuously 6 Operation Instructions 6 1 Installing two AAA batteries into battery cassette before covering its cover 6 2 Nip the clamp as diagram 6 3 Plug one of fingers into rubber hole of the Oximeter it is best to plug the finger thoroughly before releasing the clamp 6 4 Press the switch button once on front panel 6 5 Your finger do not tremble during the Oximeter is working Your body is not recommended in moving status 6 6 Read correspondent datum from display screen 6 7 Six display modes After turn on the oximeter each time you press the power switch the oximeter will switch to another display mode there are 6 display modes shown as follows Pulse Oximeter When you press the power switch for a long time more than one second the brightness of the oximeter will be changed by degrees there are 10 levels on brightness the default level is level four Declaration Please use the medical alcohol to clean the rubber touching the finger inside of Oximeter and clean the test finger using alcohol before and after each test The rubber inside of the Oximeter belongs medical rubber which has no toxin and no harmful to the skin of human being When your finger is plugged into the Oximeter your nail surface must be upward S 7 Brief Description of Front Panel SPO2 Power PR Bar
7. ontact with receive emissi local on tube customer 4 Amp circuit service malfunction center 4 Please contact with local customer service center 1 Please change battery Error displayed on screen 1 Low power 2 Emission tube damaged 3 Current control circuit malfunction 2 Please contact with local customer service center 3 Please contact with local customer service product Please follow the law of the government to deal with used battery 12 Possible Problems and resolutions local Possible Problems reason Solution Oxyhemoglobi 1 Finger is not 1 Retry by nor heart rate plugged plugging the can not be correctly finger shown 2 Patient s 2 Try some normally Oxyhemoglobi more times If n value is too you can make low to be sure about no measured problem existing in the product Please go to a hospital timely for exact diagnosis Oxyhemoglobi 1 Finger might 1 Retry by nor heart rate not be plugged plugging the is shown deep enough finger unstably 2 Finger is 2 Try not to trembling or move patient s body is in movement status The Oximeter 1 Power of 1 Please can not be batteries might replace powered on be inadequate batteries or not be there 2 Please at all reinstall the 2 Batteries batteries might be 3 Please installed contact with incorrectly local customer service centre 3 The Oximeter might be damaged
8. red the test equipment BIO TEK INDEX Pulse Oximeter tester the pulse wave is available without failure when the simulation pulse wave amplitude is at 6 3 7 Interference Resistance Capacity against Ambient Light Device work normally when mixed noise produced by BIO TEK INDEX Pulse Oximeter tester 3 8 It is equipped with a function switch through which the Oximeter can be powered off in case no finger is the Oximeter longer than 8 seconds 4 Product Properties 4 1 Operation of the product is simple and convenient 4 2 The product is small in volume light in weight total weight is about 50g including batteries and convenient in carrying 4 3 Power consumption of the product is low and the two originally equipped two AAA batteries can be operated continuously for 30 hours 4 4 Low voltage warning will be indicated in visual window when battery voltage is so low that normal operation of the Oximeter might be influenced 4 5 The product will automatically be powered off when no signal is in the product for longer than 8 seconds 5 Product Operation Scope The fingertip Oximeter can be used to measure human Haemoglobin Saturation and heart rate through finger The product is suitable for use in family hospital including clinical use in internist surgery Anaesthesia paediatrics intensive care and etc Oxygen Club social medical organizations physical care in sports It can be used before or after sports Operatio