Home

Sample Kit manual.book

image

Contents

1. The radius of the tip can be calculated from three pieces of information MEASUREMENTS Topography Scan forward a e Line fit 33 2nm Topography range xt Topography CrossSection 15 1nm Rawdata Topography range 9 72nm Som Section mm gt Bead Section Analysis A section analysis ofa zoomed in area of the bead sample The radius R of the beads The radius r of the tip e the height between the tops of the beads the lowest point that the tip reaches between them R R r h R r 2 Rd 2h Using h 15 nm and R 60 nm the bead radius we find that the tip has an approximate radius of 68 nm The manufacturer specifications for the tip radius is 10 nm so this is rather large It is possible that one or more beads are clinging to the end of the tip 53 GLASS BEADS Bead tip geometry The tip may not fit all the way down between two beads The geometry of the beads and the tip allow the determination of the tip radius This is an easy way to estimate the actual tip tadius On the other hand we realize that the tip radius influences the image feature size especially when the two have comparable dimensions Relevance of the Glass Beads The glass beads sample demonstrates the small scale interactions of sub micron bodies This alone is an interesting use for the beads but many other uses have been discovered Different sized beads have been shown to disrupt
2. AFM Setup The five components of an AFM setup regulator Even if this parts are present in every AFM their implementation can differ substantially However a common point to all AFM is the force sensor also called AFM probe It is plausible that the results strongly depend on the sharpness of the tip and the spring constant of the cantilever This will be the subject of The Force Sensor The deflection detection system needs to be very sensitive and can be implemented in different ways which will be discussed in The Position Detector The feedback system will be described in The PID Feedback System The AFM can be operated in different modes which will be discussed in AFM Operation Modes Finally The Scanning System and Data Collection will deal with the positioning or scanning system which needs to provide nanometer resolution The Force Sensor In AFM the force sensor needs to meet the two following requirements e Contact Mode see AFM Operation Modes The spring constant of the cantilever needs to be small such that the cantilever can be sufficiently deflected and the deflection can be detected Ideally the spring constant should be smaller than the interatomic spring constant which is about 10 N m Dynamic mode see AFM Operation Modes The portion of perturbation 10 A BRIEF INTRODUCTION TO ATOMIC FORCE MICROSCOPY E F RE 2 transmitted to the cantilever is given by aans aofo fres gt Where fy is the excitor
3. duction continues likely leading to an even greater prevalence in their use and a further expansion in the variety their applications Glass Beads The glass beads sample is a piece of glass covered with a thin layer of very small colloidal silica i e glass particles The beads will group in clusters some of which exhibit crystalline structures Colloidal particles are often used to make small scale regular structures such as calibration patterns hard disk recording media and photonic crystals AFM can be used to character ize defects in such structures but here the structure is used to characterize the 3D shape of the AFM tip as opposed to the 2D characterization done with nanotubes in the previous chapter Measurements Sample Preparation The glass bead sample must be prepared The kit contains an empty glass slide and a vial of diluted bead solution which will be used to create the sample In this case the beads have a diameter of 120nm may vary in future Once the sample is prepared it can be used for measurements as long as it remains clean To prepare the sample Assemble the following materials e glass slide e diluted bead solution ethanol clean tissue To clean the glass slide and remove any accumulated dirt Drip some ethanol on the slide Wipe the ethanol off with the clean tissue When the sample is clean Place the vial in a beaker filled part way with water Place the beaker in an ult
4. Image Analysis Reflective Side Figure Reflective Side Overview is an image taken of the reflective side of the foil seen in both height and surface views The most noticeable feature in this figure is the set of parallel lines of varying height running across the image These lines extend for very long stretches along the foil Try scanning at the maximum scan range of your microscope You should find that the lines extend for this entire scan range Another feature of the sample topog raphy is the set of pits that are lower down that the average height level These pits are visible as dark spots that in the bright lines running from the bottom to the top of the color map chart Topography Scan backward Topography Scan backward 60pm Line ft 850nm Line fit 1um 510n Topography range Opm Opm xt 60pm Reflective Side Overview Measurement with in a scan range of 60 pm left color map chart with a height scale of 850 nm right 3D view projection The two charts in figure Reflective Side Overview complement each other by highlighting slightly different information The height image points out the well defined lines of different height while the three dimensional surface view gives a good idea of the transition between the higher and lower regions For example the height image makes it clear how deep the pits are but not how steep the drop from the surrounding area into the pit is The color map better illustra
5. RAM sample ZOutput ForwardScan 4 07 um Rawdata ZOutput range 1 05pm Opm x 39 9um Optimized Gain Optimized gain setting for RAM chip structure MEASUREMENTS Lower the gain Lower the integral gain well below the optimal setting As you lower the gain the feedback loop will not work quickly enough to provide high resolution Note the poorly defined edges in figure Low Gain At a lowered gain the feedback loop is not responding quickly enough to respond to changes in height ZOutput ForwardScan 4 07 um Rawdata ZOutput range 1 05pm Opm PA 39 39pm Low Gain The feedback loop is not responding quickly enough Raise the gain Gradually raise the gain to well above the optimal settings At some point the z controller will start to overcompensate for feedback errors when the tip encounters steps in the sample This overcompensation is also called over shoot Overshoot and Undershoot When the gain settings are increased further the controller will react to this overshoot by undershooting the undershoot will be less than the overshoot These overreactions initiate an oscillation which eventually subsides The fre quency of this oscillation is either the mechanical resonance frequency of the scanner or the resonance frequency of the cantilever itself At even higher gains the oscillation will not subside Instead it will steadily increase most likely resulting in da
6. different cells Cell disrup tion is the process wherein the cell wall growth is disturbed for the purpose of extracting products out of the cells in which they are produced This dis ruption can be used for example to obtain DNA from within a cell nucleus Extensive research has been performed and now sold in different sizes according to their use in specific cells Beads have also been very useful in the study of chromatography the separa tion of a mixture of substances in a phase separated medium There is a sta tionary phase held in some sort of container and a moving phase flowing through it The different substances in a given mixture are drawn to either the stationary or mobile phases based on some property size charge etc that they exhibit Silica beads coated with a substance similar to that of a cell membrane have been used to simulate cellular interactions This provides a simple way to reproduce crucial biological phenomena that are not fully understood In RELEVANCE OF THE GLASS BEADS other experiments researchers have suggested that silica beads coated with gold may kill cancerous cells when exposed to near infrared light 55 56 GLASS BEADS Staphylococcus Aureus AFM is used for various biological applications such as determining the structure of proteins and the binding force between antibodies and antigens and other key lock pairs While most of these measurements must be per formed in a physiologi
7. follow the topography Investigate this by measuring the same area with different scan rates Fabrication of the Gold Clusters The gold clusters sample consists of a thin layer of gold clusters on a piece of a silicon wafer To help the gold stick to the surface a 5 nm layer of titanium was applied to the surface before adding the gold The gold layer is 200 nm thick As the gold atoms were deposited on the titanium they joined into clusters on the surface To prepare the sample small pieces of the silicon wafer were cut to fit onto the sample disc The pieces were then cleaned in a sonic bath for a few min utes each in water acetone ethanol and finally propanol The cleaning ensures that the silicon has no pieces of dirt on its surface which would result in an uneven coating of titanium and gold Both the titanium and the gold were applied to the silicon with the use of a sputtering machine Samples of titanium and gold are at the bottom of the chamber with the silicon samples at the top as shown in figure Sputtering machine p 41 The titanium and gold are bombarded with high energy argon ions to release metallic atoms The atoms that are released are actually ions meaning they have electric charge A strong magnetic field is applied in the chamber and the charged FABRICATION OF THE GOLD CLUSTERS Sputtering machine The vacuum pump controls the pressure inside the chamber The boxes directly above the gold and titanium
8. force is set by the user set point The feedback regulator main tains this set point by moving the scanner in the direction vertical to the sample This movement generated by the regulation is then plotted as topog raphy of the sample The major parameter to set in this mode is the interac tion force This must be set to a minimum value such that the tip is just in contact with the surface The inconvenience of this method is that the tip might easily be contaminated or broken and that sticky samples can not be imaged correctly Non Contact Mode In this mode the cantilever is oscillated at high frequencies 100kHz 1MHz and small amplitude lt 1nm The tip is then scanned at some nanometers above the surface and ideally never touching the surface This oscillation increases the sensitivity in such a way that even the marginally attractive Van 15 A BRIEF INTRODUCTION TO ATOMIC FORCE MICROSCOPY de Waals force can be detected The feedback regulator reacts to changes of oscillation amplitude phase or frequency This mode is very difficult to implement as the measurement is very sensitive to long range and short range attractive forces and thus the equilibrium sit uation is very unstable and drift dependent Dynamic Mode Dynamic mode is probably the most used mode nowadays As in the non contact mode the cantilever is oscillated but with a much greater amplitude Hence the tip is touching the surface periodically The contact w
9. from the se miconductor This electrode is called the gate The gate structure can be formed using a metal electrode on top of an insu lating oxide layer on top of the semiconductor giving rise to the term MOS Metal Oxide Semiconductor When a positive voltage is applied to the gate of an n channel transistor additional electrons move to the surface of the p region These electrons then form a conductive channel between the two n regions hence the name n channel Most digital ICs microproces sors memory chips use a technology known as CMOS Combined MOS that uses both n and p channel MOS transistors INTEGRATED CIRCUIT TECHNOLOGY Integrated circuit production The production process of integrated circuit involves many steps which can be grouped into three main steps Define the structure of the circuit s various components Deposit the material of which the component consists Remove the spurious material The structure of an integrated circuit is defined using photolithographic processes First a layer of light sensitive resin is deposited on the wafer Next the resin is exposed to a light dark pattern by projecting a mask onto the resin Finally depending on the process either the exposed or non exposed areas of the resin are removed and the areas that will not be processed are protected by the remaining resin The production process begins with a Silicon disc called a wafer The wafer is already lightly infuse
10. is now the dominant transistor in digital circuits Bipolar transistors The first transistor type to be made it was discovered in 1948 in the attempt to build a FET This work earned William Brad ford Shockley John Bardeen and Walter Houser Brattain the 1956 Nobel Prize in Physics Bipolar transistors are the predominant transistors in ana log circuits This discussion will only deal with the field effect transistor FET because it is easier to understand which is probably why it was invented earlier In a FET two regions of the same type are separated by a region of the other type When the like regions are n type this is called an n channel FET when 23 24 CHIP STRUCTURE IN SILICON the like regions are p type this is called a p channel FET These two arrange ments are illustrated in figure PNP and NPN Transistors Doped with an element having 5 valence electrons EE Doped with an element having 3 valence electrons PNP and NPN Transistors p and n type regions arranged into p and n channel field effect transistors If two electrodes are attached to two regions of the same type and a voltage is applied this arrangement will not allow current to flow between the two regions because either one of the diodes will block the current flow However this situation can be changed by applying an electrical field hence the term field effect to the center region using a third electrode that is electrically isolated
11. of thickness less than or equal to 100 um and most foils are somewhere in the range of 5 to 75 um The foil in this sample is approximately 30 um thick Generally two sheets of a thicker aluminium foil are rolled together at the same time to save time and energy The two sheets are rolled one on top of the other with oils or other lubricant between them as well as between the foil and the rollers as shown in figure Foil Production Thus the two sides of the foil are exposed to different conditions during the rolling process The reflective side of the foil was in contact with a harder surface the roller it is flatter and shinier The non reflective side is only in contact with another sheet of aluminium foil another non reflective side and so is less flat and reflective The lines of different height on the shiny side of the foil are a result 79 ALUMINIUM FOIL of the rollers pushing into the foil so the lines run in the same direction as the length of the roller as depicted in figure Foil Production Reflective a Non reflective er Reflective Foil Production The process that leaves one side of the aluminium foil shinier than the other side The pits seen in the various measurements are caused by the oil placed between the rollers and the foil to prevent overheating The oil gets trapped between the roller and foil and is forced down since it has no other place to go Sample Maintenance Of all the samples in t
12. on or off Transistors are generally made from a semiconducting material which con sists almost exclusively of atoms of an element with four valence electrons usually Silicon The transistors consist of an arrangement of regions of the semiconductor that differ in the amount and type of impurity elements they contain These regions are referred to as either n type or p type In n neg ative regions the semiconductor is infused with an element that has an extra valence electron The extra electrons improve the conductivity of the semi conductor by making the motion of the negatively charged electrons the dominant conduction mechanism In p type positive regions the semicon ductor is infused with an element with less electrons This lack of electrons causes the creation of extra holes a hole being a positively charged electron deficiency The transition between a p and an n region will normally block the flow of current from the n to the p region but allow current flow from the p to the n region Such a transition is called a diode The n and p regions can be made in various arrangements to make different types of transistors The main transistor types are bipolar and field effect transistors Field Effect Transistors FETs Although invented first the first working FET was built later than the bipolar transistor A particular type of field effect transistor the Metal Oxide Semiconductor Field Effect Transistor MOSFET
13. quickly wear the tip Set a small force in the setpoint 10nN in static mode 60 in dynamic mode PS PMMA THIN FILM Approach the sample For more information on approaching the sample see the AFM user manual Select a region with well defined bumps and little or no dirt Zoom in on Phase Contrast If necessary consult your user documentation for how to set up your AFM to take phase contrast images Approach the sample as described above Take a phase contrast image Image Analysis Topography of PS PMMA The polymer film should appear as a collection of randomly shaped ran domly spaced spots The spots which are the hydrophilic PMMA in a sea of PS are approximately 20 nm high and different regions of the sample may have different sized spots Close to the centre of the sample the spots have an approximate diameter of 0 5um Farther to the outside of the sample the spots are larger The size of the spots has in part to do with the amount of time that the mixed solution is allowed to rest on the silicon before it is spin coated The outer parts of the sample may have had more liquid solution on their surface for a slightly longer period of time than the inner regions due to the spin coating process Figure PMMA Bumps Surrounded by PS shows two topography frames of the polymer film The left image is in a range of 10pm In a scan range of this size many of the bumps are visible and it is possible to
14. sec PI Gain A EE 0 8 E 0 6 S E S 3 4 04 a 0 2 0 0 1 2 Time sec PID Gain slope an overshoot peak and a settling time wobbles As next the differ ential gain D will be turned up in addition to P It can be seen in figure PD THE AFM SETUP Gain that the derivative gain reduces both the overshoot and the settling time and had little effect on the rise time In order to see the influence of the Integral gain I the D gain is turned down and the I gain up As can be observed in figure P Gain the I controller further reduced the overshoot and decreased the settling time The response is much smoother now albeit with an increased rising time When the P I and D gains are combined in an appropriate way it is possible to obtain the response shown in figure PID Gain with no overshoot short rise time and short settling time The correct PID settings are sample dependent and have to be determined for each measurement AFM Operation Modes The AFM can be operated in different modes This depends on the sample and on the information one would like to acquire Among several modes here only the most common ones are discussed contact non contact and dynamic mode Contact Mode This mode is the most basic mode which was also the first real mode in which AFMs were operated The tip is always in contact with the sample while probing the surface Thereby the deflection of the cantilever and thus the interaction
15. that not only the very apex of the tip but also its side walls interact with the sample during scanning the measured image is always a convolution between the tip shape and the sample Therefore it is important that the feature size of the sample and their aspect ratios are some orders bigger than the radius of cur vature and aspect ratio of the tip respectively The Position Detector Another critical part of the AFM is the deflection measurement system Ide ally the sensing system must be able to measure the deflection of the canti lever with angstrom resolution and must not perturb the cantilever in any way The most used detection system is therefore an optical technique based on the reflection of a laser beam on the cantilever The idea of the technique is shown in figure AFM Setup A laser beam is focused on the very end of the cantilever which reflects it back on a segmented photo diode The deflection angle of the cantilever is thereby enhanced i e a small displacement of the cantilever results in a bigger displacement of the reflected laser beam on the photo diode The further away the diode the bigger this mechanical amplifi cation However the photo diode cant be placed to far away because of exter nal perturbation One reason for that is that the laser deflection method is sensitive to the ambient light the light reflected by the sample or the canti lever and other possible sources of light The optical detection system allows
16. vibration frequency with amplitude a and f is the resonance frequency It is therefore usual to use cantilevers with high resonance fre quency in order to avoid low frequency acoustic or mechanic perturbation such as building vibrations AFM probes are typically micro fabricated The single leg or V shaped can tilevers are usually made out of silicon silicon dioxide or silicon nitride Typical cantilevers are several hundred micrometers long several tens of micrometers large and on the order of one micrometer thick For silicon these dimensions will result in spring constants between 0 1 and 1N m and resonance frequencies between 10 and 100kHz To further avoid such perturbation the AFM can be placed on a vibration isolation table or under a acoustic hood AEM Probe Tip and cantilever of an AFM probe Image courtesy of IBM Thanks to recent developments in microtechnology it is possible to fabricate cantilevers with integrated sharp tips It is important to keep in mind that THE AFM SETUP the quality of the tip i e the shape of the tip determines the quality of the measurement The critical dimensions of an AFM tip are its aspect ratio height width the radius of curvature sharpness and its material The ideal tip has a high aspect ratio a small radius of curvature and is made of an extremely hard material The shape of the tip is of great importance when it comes to the interpretation of the measurement Due to the fact
17. Alu minium foil is only one common household product that can be imaged with AFM to determine its microscopic structure Other items from around the house to try imaging include regular vs glossy paper hair the Teflon coating on a razor blade rubber plastic glass metals or anything else that you can find Not all of the items that you try will be flat enough to scan but the ones that are may yield interesting results Measurements The aluminium foil sample consists of two pieces of household aluminium foil A sheet of aluminium foil has two sides one more reflective than the other One of the pieces of foil has the shiny side up the other has the dull side up Both pieces are glued to the sample disc Image Acquisition The reflective side is the easier side to measure so begin with this piece It should be possible to see the cantilever s reflected image in the shiny piece of aluminium The diffuse reflecting sample has no usable mirror image of the cantilever Instead you can approach on its shadow which is not as easily visible Approach the cantilever to its reflected image or shadow Continue until the two are close but without touching See user manual for more information on how to approach samples Start automatic approach Measure the reflective and diffusive sample at the same scan In order to later compare the measurements Set the color map height scale identical in both measures ALUMINIUM FOIL
18. Yet User Manual nanoSurf Extended Sample Kit NANOSURF AND THE NANOSURF LOGO ARE TRADEMARKS OF NANOSURF AG REGIS TERED AND OR OTHERWISE PROTECTED IN VARIOUS COUNTRIES 29 8 07 BY NANOSURF AG SWITZERLAND PROD BT0 RO 2 Table of contents A Brief Introduction to Atomic Force Microscopy Introduction coi rad caidas tara E SE recites The AFM St muito apani The Force Sensor The Position Detector The PID Feedback System AFM Operation Modes The Scanning System and Data Collection Chip Structure in Silicon WMEASUEMEMES bre Image Acquisition Image Analysis Integrated Circuit Technology ccceeeeseseeeeeeeteneetseeeeeeeaees Transistors Integrated circuit production Sample Maintenance rrnnrnnnnnnnnrnnrvnnnnvnnnrnnsnrnnnvnenrnssnnnnnrensnne CD Stamper Mea SEE ado Image Acquisition Image Analysis Optical Data Storage rrranrnonvnnnvvnernnnvrnnnrnrsvnnenrnenrnrsnrnennnesnne Sample Maintenance rnrnrnnnnvnnnrnnrvnnnnrnnnrnnsnnnnnnnnnvnnsnnnnnven nn Gold Clusters MeasUre Me tiiniician iaa do Image Acquisition Image Analysis Fabrication of the Gold Clusters cccccccccnnnnanananancncnnnononanos Nanotubes Me surementis tt lis Image Acquisition Image Analysis Carbon Nanotubes oooococococoncccccccnncnnncnnnononininnnnano nono nonnonnnnnnnnns Glass Beads Measurement lada Sample Preparation Image Acquisition Image Analysis Relevance of th
19. aph ite These bonds are stronger than the chemical bonds of diamonds making nanotubes very durable Nanotubes naturally align themselves into bundles held together by Van der Waals forces 47 48 NANOTUBES Nanotube Molecular Structure Japanese physicist Sumio lijima discovered the hollow cylindrical nanotubes while studying fullerene synthesis in 1991 Today nanotubes are used in a range of applications that is remarkable considering their short history A nanotube may or may not conduct electricity depending on its structure This opens the possibility of very tiny electrical circuit elements particularly transistors see the description in the Chip Structure chapter Nanotubes have also been shown to conduct heat very effectively a property that could be applied to the cooling of tiny mechanical elements It has also been shown that by removing the cap at one end of a nanotube it can be used as a nano test tube Under the proper conditions nanotubes can be made sufficiently long to serve as carbon fibres Researchers have found that newly synthesized nano tubes exhibit strong absorption of outside molecules and this may have direct applications in medicine A company called Babolat uses nanotubes in their tennis rackets to keep them stiffer and controlled electron emission from nanotubes have been used in television like displays Building nano tubes will become simpler and more cost efficient as research into their pro
20. are sensors that measure the rate at which the metals are applied to the silicon above the rectangles above the sensors The lines that run through the sensors are the shutters that open or close depending on which metal is being applied to the silicon titanium and gold atoms are directed upwards in the field onto the silicon By controlling the pressure in the chamber and the exposure time of the silicon the exact thickness of titanium and gold layers can be controlled 41 42 GOLD CLUSTERS Nanotubes The carbon nanotube sample represents the group of materials with macro molecules that are used for molecular nanotechnology Other well known examples are self assembled particles DNA and nanotubes made of other materials AFMs can be used to characterize and manipulate such molecules The well defined structure of nanotubes makes them ideal for demonstrating the influence the structure of the end of the AFM tip has on the measured image While manipulation is not demonstrated here that does not mean it is impossible to do Measurement The carbon nanotube sample consists of a piece of silicon wafer on which carbon nanotubes are deposited Nanotubes are less than 10 nm in diameter and can reach lengths of several 100 micrometers Image Acquisition With this sample if the scan parameters are not well optimised the tip is likely to be damaged Therefore if you start with a relatively large range 15 um and successively z
21. ateral scale long variations are wavy short variations are rough The distinction between roughness and waviness is made because roughness gen erally has a different effect on the function of a device than waviness For example in optical devices waviness causes optical aberrations roughness causes light dispersion and absorption MEASUREMENT In the easyScan software the surface roughness is calculated as follows Before doing any evaluation on the data shown in chart the software proc esses the height data according data filter that was set View Panel or chart Bar depending on the software that you are using Before calculating rough ness subtract the average slope of the data by selecting the Line fit Plane fit data filter In figure Image of the Gold Clusters Sample p 38 the number of points is expressed as the number of data samples taken 256 This means that the microscope will record 256 lines of data in the y direction with each line consisting of 256 equally spaced data points in the x direction The Root Mean Square RMS roughness Sq is defined to be the sum of the absolute value of the difference of the height at each point and the mean height It is therefore a measure of how different the height of the sample is from its average height which is the value calculated by Since height is measured in nm the difference will also be in nm The square of the difference will be in nm 2 and Sq will be measured in
22. ation area of thin film coat ings Although the techniques used for producing coatings are in many respects similar to the techniques used for chip production they differ in both the surface area that is coated and in the greater variety of materials used for coatings Moreover the thin film coating processes are meant to change the function of the entire surface of the sample as opposed to producing discrete structures Therefore AFM applications concentrate on characteris ing statistical properties of these coatings rather than the size of specific structures The most commonly used statistical property of a surface is roughness of which there are many different parameters This chapter shows how AFM measurements are used to determine the surface root mean squared rough ness parameter Sq of the gold cluster sample Measurement Image Acquisition The maximum height value attained in the topography of the gold clusters sample is close to 10 nm not taking into account any dirt that may appear on the surface The image in figure Image of the Gold Clusters Sample appears to be free of any dirt Starting with a large scan range 1 5 um allows you to identify any potential dirt so you can locate a smaller area free of dirt and zoom in on it This sample is quite smooth i e has a very low roughness Therefore the measurements are very sensitive to disturbances meaning there is more noise apparent in the height signal The horizontal li
23. by 6 others 51 GLASS BEADS In figure Well Ordered Beads for example the top right corner of the image is far better ordered than the top left Notice also that there are some beads slightly higher up than others they are not all perfectly coplanar pography Scan forward Line ft 52 3nm Topography range Opm xt 3 99pm Well Ordered Beads The center part of a spot of bead solution Some sections have a crystalline structure while others are less ordered Image Analysis Section analysis can be used to determine the distance between desired mark ers Figure Bead Section Analysis p 53 illustrates how to find the vertical distance from the point from where two beads touch to the top of one of the two beads Since the beads are approximately spherical and have a 120 nm diameter the vertical distance between one peak and the adjacent valley should be half the diameter 60 nm However in this section analysis the vertical distance is only about 15 nm The discrepancy is due to the fact that scanning is limited by the shape of the tip which cannot fully extend down between beads Figure Bead tip geometry p 54 illustrates how this occurs with a tip that is approximately spherical at the end The tip may track some of the height drop between two adjacent beads but it will not track the full extent of the drop While this effect is negligible for larger beads it becomes more signif icant the smaller the beads are
24. cal solution some samples such as the Staphylococcus Aureus sample can be measured under normal AFM scanning conditions The staphylococcus staph aureus bacteria sample is a glass slide covered with millions of bacteria The bacteria have been killed and fixed to the slide so the sample is safe to touch and scan The name staphylococcus comes from the Greek word staphylos meaning bunch of grapes and the Latin word coccus for the round shape of the individual bacterium Aureus is a strain of the staphylococcus bacteria one which is commonly contracted by humans The round bacteria group together in random clusters which makes them look like grapes in bunches Measurement Image Acquisition The glass slide is only slightly reflective so it can be difficult to judge the tip sample distance for the approach If you can see the cantilever s shadow or reflection you can use it to judge the distance You can try to make the reflection more visible by moving the sample holder slightly If you cannot see the cantilever s reflection perform a very slow coarse approach while judging the distance on the focal plane of the side view as follows e When the tip is on the sample the focal plane crosses the sample at the tip position e When the tip is further away the focal plane crosses the sample more behind the cantilever The bacteria have been fixed to the glass slide with a burning process The process leaves a mark where the bacte
25. characteristics of the oxide layer the tip quality decreases quite fast compared to usual tip wear Thanks to the sharp steps this sample is ideal to learn and train how to optimise the PID feedback settings and vibration amplitude in dynamic mode The later will be the subject of the analysis section Image Acquisition The general guide how to start a measurement is described in the user man ual In this particular case Find a clean spot on the sample Approach the sample Start the measurement Adjust the slope The procedure for adjusting the slope is described in the user manual Adjust PID gains Find the optimum vibration amplitude see next section Image Analysis Setting the vibration amplitude is crucial for achieving the optimum resolu tion In Static mode the main parameter to regulate the image quality are PID feedback settings and Set point In dynamic mode additionally the set 62 MICROSTRUCTURE ting of the vibration amplitude plays a great role In general the vibration amplitude must correspond to the size of the sample features small structures require a small amplitude big structures requite a big amplitude However if you want to look at small structures on top of big structures you need a small amplitude and slow scan speed Topography Scan forward xt Topography Scan forward 124nm Mean fit 125nm Opm xe SOpm Raw data 250nm Amplit
26. d with either n type or p type impurities After some intermediary steps the impurities are infused into the wafer in a process called doping Photolithography processes are again used to define which part of the wafer will be doped If the original Silicon was slightly doped with n type material a p type region must be made see figure PNP and NPN Transistors p 24 Creation of a p type region can be accomplished by bombarding the desired area with high energy p type atoms which become embedded just beneath the surface When the Silicon is heated to a very high temperature the p Lightly doped p type EEG pay n type Yo Sa Doping Process 26 CHIP STRUCTURE IN SILICON type atoms will diffuse further down into the Silicon By controlling the temperature and exposure time manufacturers can set the depth of the p type region This p type region is the site of one transistor which is finally created by adding two small n type regions The n type regions are created with the same method used to create the p type region Figure Doping Process p 25 shows the steps for implanting the p and n type regions into the Silicon wafer indicating the active area s at each stage of the process To construct the gate first the whole piece of Silicon is covered in a thin layer of insulating gate oxide and then covered with a thicker layer of con ductive highly doped polycrystalline Silicon polysilicon which behaves like a metal strict
27. e Glass BeadS occcccccccnnnnnonananannncncncnnnonanos TABLE OF CONTENTS Staphylococcus Aureus 57 Measurement 2 s s0iescct O 57 Image Acquisition 57 Image Analysis 58 Microstructure 61 Measurement sclcceaiae enki al ee 61 Image Acquisition 61 Image Analysis 61 Micro fabrication colacao an tives ram 64 PS PMMA Thin Film 65 Measurement 25 nc avira an eaa adia nite eed 65 Image Acquisition 65 Image Analysis 66 Phase Contrast Imaging rrronvnnnrnnrnrnnnvnnrnrnrnnnnnnnnnnrennnnnrnrnssrnnsnnnensene 68 Skin Cross Section 69 Measurement el dd 69 Image Acquisition 69 Image Analysis 71 Biological Samples botica ti e 72 Aluminium Foil 75 MeasurenentS sica ns Ae in 75 Image Acquisition 75 Image Analysis 76 Bulk Foil Production 3 4s00 aston SA eee 79 Sample Maintenance cccesccseseeseeeeeeeeeeeseeeeeaeeseaeeeaeeseeeseaeeeeeeaas 80 TABLE OF CONTENTS About this Manual The sample kit contains ten samples chosen from among various disci plines tools with which to handle them and a manual to guide the student through the investigation of each one Each individual sample was dually designed to emphasize interesting features of the sample itself as well as to highlight diverse elements of the AFM In some cases the samples have the additional quality of representing actual current uses of AFM in their respective fields of science The manual is intended to serve as a collection of suggesti
28. ection of the images above at the position indicated by the arrow The topography image is smeared out and the topography line graph shows a too small slope The reason therefore can be found in the amplitude signal The peaks are larger this means that the correction to the amplitude to the Set point value is not as quick as in figure Microstructure Due to the small vibration amplitude when the tip needs more time from the moment where it lost the contact to the surface to the moment it gains con tact again During this time the topography is uncertain and the tip is vibrat ing at the free vibration amplitude Increasing the vibration amplitude or decreasing the scan speed will increase the quality again 63 64 MICROSTRUCTURE Micro fabrication HE so Si E o photoresist SiO Si SiO photoresist SiO Si SiO SiO Si Process Flow Fabrication process flow for the Silicon dioxide microstructures This sample can be fabricated with only a few simple steps If you can access a clean room facility you can easily reproduce the fabrication procedure shown in figure Process Flow Start with a plain Si wafer Make a thermal oxidation for 200nm Spin coat the photo resist on the wafer Transfer the pattern to the photo resist by photolithography Develop the resist Transfer the pattern from the resist into the silicon dioxide by wet etching in buffered hydrofluoric acid PS PMMA Thin Film The PS PMMA
29. es scan rotation raw data The orientation of the scan direction with respect to the features in the sample can make a huge difference when data processing is used on the image An interesting analysis to perform on this sample is a section analysis A sec tion running perpendicular to the lines of different height can help to deter mine the height differences between adjacent lines as well as the maximum height difference between the highest and lowest points in the line selected Figure Reflective Side Cross Section shows the section analysis of a line running through the lines of different height on the shiny side of the aluminium foil The line selected runs through one pit which is visible as a dip in the sec tional cross section chart The Measure Length tool is used to measure the half width 705nm and depth 50 nm of the pit 77 78 ALUMINIUM FOIL Topography Scan backward 60m Line fit 850nm Oum x 60m Tool status ES 180nm Length 707 5nm Deltaz 53 72nm Width 705 4nm Height 53 72nm Mean fit 191nm Opm Section 20 84m Reflective Side Cross Section Analysis of the overview measurement Diffuse Reflective Side On inspection of the measured surfaces such as shown in figure Comparison it immediately becomes clear that the diffuse reflection of the surface is due to its much rougher and more irregular surface There appear to be lines running f
30. he Sample kit the aluminium foil sample is the only one that can hardly be damaged by cleaning To clean the sample drop some ethanol onto the aluminium foil and wipe it dry with a clean paper tissue 80 81 82
31. his kit OPTICAL DATA STORAGE The CD manufacturer uses the stamper to press the shape made by the bumps of the stamper into a piece of polycarbonate plastic The indented side of the plastic is then sputtered with aluminium and finally covered by a protective layer of acrylic When a CD player reads the CD it focuses a laser light beam through the clear polycarbonate plastic layer and onto the alumi num Figure CD Cross Section illustrates the order in which these layers appear The laser beam reflects off the aluminum and hits a detector which simply reads the difference between 1s the edges of the bumps and Os thus decoding the stored information What is more complicated is the way the CD stores the information in a manner that reduces the chances of errors This method called EFM eight to fourteen modulation involves converting the original 8 bit data to 14 bit Adding the extra bits guarantees that there are always between 2 and 10 Os in a row and never consecutive 1s Label Acrylic ng Plastic AT Laser CD Cross Section To read the CD a laser is focused on the aluminum through the thick plastic layer below the label and aluminum sputter Furthermore CDs use a process called interleaving whereby information is not stored or retrieved sequentially The bumps and pits in one track do not make up one long chain of information but rather scrambled information that the CD reader p
32. igh right lower force set point 2 nN note dirt moved around by the previous measurement with too high force set point MEASUREMENT Image Analysis Length Use the length tool to measure the length of various nanotubes For best results measure the straighter nanotubes as the curved or bent ones may be difficult to measure Compare the lengths of the longer nanotubes to the shortest ones and note the wide variations Height and Width Use the section tool to measure the height and the width of a nanotube First use the line section tool to draw a line perpendicular to the axis of the nano tube to be measured Then measure the nanotube height using the distance tool and measure the width using the length tool Figure Nanotube Cross Section Measurements shows how tip sharpness affects the measurements Tip Geometry In theory nanotubes are cylindrical The above measurements indicate that the vertical distance between the top of one nanotube and the silicon sub strate is around 3 6 nm which is the nominal diameter The width measure ments however clearly do not represent a cylinder with this diameter The reason for this discrepancy is that the horizontal measurements are lim ited by the shape of the end of the tip A simple model of the end of the tip is a sphere When a sphere scans over a cylinder that is smaller than the sphere radius the measured topography will be a cut off cylinder with a radius equal to the sum of
33. in the smaller scan range are different Rather zooming in allows higher resolution of smaller detail The second graph on the right in figure Zoomed Image shows the height difference between consecutively measured points This image displays the slope of the topography and gives better contrast for finer details For exam ple the textured surface of the bacteria is enhanced in the deflection image as is the shape of the cell walls separating the bacteria 59 60 STAPHYLOCOCCUS AUREUS ZOutput ForwardScan E A Raw data 450nm ZOutput rang ZOutput ForwardScan 4 97 um Derived data S0nm ZOutput range Zoomed Image Left color map chart of raw data Right color map chart of derived data The maximum height attained in the image on the left is 450 nm The height scale for the image on the right is arbitrary in the sense that it has nothing to do with the height of the bacteria Microstructure The microstructure sample consists of a structured silicon dioxide layer on silicon The structures are regularly distributed squares Depending on the stock the sample has square silicon dioxide island or square holes in the silicon dioxide layer This kind of sample is often used to calibrate the orthogonality of a microscope imaging device Measurement This sample is in general quite easy to measure and there are not any special settings to be considered However due to the abrasive
34. ince the speed is equal to the radius times the rotational speed As the reader moves outward the CD player has to slowly decrease its speed to make sure that the information is always read at the same rate Sample Maintenance CAUTION Cleaning the sample is always accompanied by the risk of permanently dam aging it To reduce the risk of damaging the sample it is best to reduce the need for cleaning it by always storing it in its container when not in use Only clean the sample when it has become too dirty to allow good measure ments Clean the sample using only highly pure solvents which prevents the depo sition of residues To remove any large dust that may have accumulated on the surface leave the chip and the sample disk in an ultrasonic bath as follows SAMPLE MAINTENANCE 10 minutes in distilled water Sonicating first in water removes any salt which would not be removed by the alcohol 10 minutes in ethanol Removes finer dirt 10 minutes in propanol Removes even finer dirt Using the propanol last also leaves the sample cleaner as it evaporates leaving fewer spots To ensure that the sample dries as cleanly as possible dab it off with a clean P y as p paper tissue to remove excess propanol when removing it from the propanol 35 36 CD STAMPER Gold Clusters The gold clusters sample consists of a thin layer of gold clusters on a piece of a silicon wafer This sample represents the applic
35. ins the regular structures of the chip ZOutput ForwardScan 40um Line fit 1 35pm ZOutput range Oum Opm xe 39 9pm RAM Chip Structure 40 um image taken at center of chip CHIP STRUCTURE IN SILICON Note the well ordered repeating structure Figure RAM Chip Structure p 19 shows the repeating structure at the center of the chip The height of the structure is approximately 1pm Each individual element in the image consists of two rounded squares each with a hole at the center separated by a groove Together these two rounded squares and the groove form a memory cell The holes connect to the tran sistors located below them The gate of the transistors is located below the groove There are also lines running perpendicular to the gates Image Analysis Since the chip structures are so well defined this sample is conducive to test ing the effects of your instrument s integral gain setting Note that some systems have more than one type of gain setting If your system has just one gain setting it is the integral gain CAUTION Excessively high and low gains can result in a damaged tip Monitor your system carefully when adjusting the gains Optimize your gain settings If you have not already done so make sure your gains are set to levels that produce reasonable images The line trace in figure Optimized Gain repre sents well optimized gain settings the tip is accurately tracking the topogra phy of the
36. ith the sur face attenuates the oscillation amplitude The feedback regulates this atten uation compared to the desired set point Ideally the damping of the amplitude is related to the tip sample interaction force which is therefore defined with the set point The set point of this mode is given by the per centage of damped amplitude compared to the undamped amplitude i e a set point of 100 gives no interaction and a set point of 60 means that the 40 of the vibration energy is lost in the interaction between tip and sample As in contact mode the goal is to keep the interaction as small as possible in order to avoid damage or contamination of the tip In this case this means that the set point needs to be as near to 100 as possible The oscillation amplitude is also an important parameter Generally the oscillation amplitude has to be in the order of the features that have to be observed i e large features need large amplitudes and tiny features need a small amplitude In order to measure tiny features on large features small amplitude and slow scan speeds are recommended The achievable resolution of the dynamic mode is comparable to the contact mode However due to the fact that the tip is only periodically in contact with the sample the tip is less damaged and the lateral sticky forces are neg ligible The Scanning System and Data Collection The scanning system of the AFM must be capable of placing the tip with a subatomic reso
37. k drives high resolution z height measurements and high resolution magnetic field images using Magnetic Force Microscopy In the field of optical data storage and other optical devices holograms interference grids optical surface characterisation tools can not be used because size of the structures is comparable to the wavelength of the light This makes the AFM the most cost effective tool for surface topography measurement The size of CD and DVD structures must be very well defined and this requirement is well served by the measurement evaluation tools in AFM soft ware which is demonstrated in this chapter Measurement The CD stamper sample contains a piece of the master copy of aCD This is the original that creates the imprint in the pressed CD that you listen to ACD has small indentations called pits whereas the stamper has bumps in the corresponding places Image Acquisition Set a large scan range approximately 50um At this size you can see many bumps and it is even possible to make out the curvature of the rows tracks Each bump is approximately 200 nm high Practice zooming in on individual bumps This sample is good for practicing zooming in on individual surface features as bumps are visible at a variety of scan sizes Take an image of well ordered bumps at least 5 or 6 tracks wide CD STAMPER Try to get an image similar to figure 20 um Image of CD Stamper which is suitable for measuring the t
38. le the topography of the surface could be plotted by measuring the distance dependent tunnelling current The STM was a revolution in the field of high resolution microscopy however this technique could only be used to image conducting samples First STM First STM left invented by Binnig top and Rohrer bottom in 1981 Image courtesy of IBM New scanning probe microscopes SPM based on the STM principle have therefore been invented Among those the most promising was the atomic force microscope AFM The success of AFM is due to its capability to achieve atomic resolution and to simultaneously measure topography and other force related material properties AFM could also be used to image a huge variety of samples which of course did not need to be electrical conduc A BRIEF INTRODUCTION TO ATOMIC FORCE MICROSCOPY tors and could also be used in different environments like gas vacuum and liquid The principle of AFM is very easy and straightforward The AFM detects the force interaction between a sample and a very tiny tip lt 10nm radius mounted on a cantilever with a low spring constant lt 10N m The force interaction between sample and tip is related to the deflection of the canti lever i e the more the tip presses into the sample the greater the deflection of the cantilever and the greater the force exercised on the sample A regulat ing feedback system tries to keep the deflection of the cantilever and thus the force i
39. locate areas with large pieces of dirt The image on the right is a zoomed in region Phase Contrast of PS PMMA The phase contrast image gives information about the hardness of the observed sample The PS can be discerned from the PMMA not only by topography but also by their different density Figure Phase Contrast of PS MEASUREMENT Topography Scan forward Line fit 14 6nm Line fit 14 6nm Perg TR a o pe ae 5 PK Min p gt y a TES cece oon id E ee A PO ARAN t r xt PMMA Bumps Surrounded by PS Lefi seen in a 10um scan range right seen in a scan range of Sum The image were taken at centre of the sample Phase Scan forward Line fit 14 7nm Line ft 12 2 2 52pm Phase Contrast of PS PMMA Left Topography of the PS PMMA thin film Right Phase contrast of the PS PMMA thin film The small dirt prickles result in a third phase contrast PMMA p 67 shows a 2 5m topography and phase contrast image The two blends are clearly discernable in both images The phase contrast mode could thus be used to discern materials with locally different hardness which are not visible in topography PS PMMA THIN FILM Phase Contrast Imaging The phase contrast image provides significantly more contrast than the top ographic image and is sensitive to material and surface properties such as stiffness and chemical composition In general changes in phase angle du
40. lution which is needed in order to image the sample with atomic resolution THE AFM SETUP The movement of the tip or sample in the three axes can be realised in several ways There are different implementations e g piezoelectric electromag netic etc As described in The PID Feedback System the topography image is generated by the feedback system which moves the scanner This motion data is sent to the PC software through the AFM controller usually line by line The software combines the lines to a three dimensional image where the height is usually represented with a colour code 18 A BRIEF INTRODUCTION TO ATOMIC FORCE MICROSCOPY Chip Structure in Silicon The characterization of chips also known as Integrated Circuits ICs is an important application of AFM technology The dimensions of the structures in these circuits are decreasing rapidly and no other tool is able to character ise these dimensions without destroying the sample This particular chip is a Random Access Memory RAM chip The struc tures on a RAM chip are very regular and rather large which make it ideal for looking at how the height measurement in the AFM works In particular we will investigate the function of the z controller Measurements Image Acquisition Set a large scan range between 10 and 80pm The chip structure can be clearly seen at this size Approach the reflective part at the center of the sample This is the section that conta
41. ly speaking the term Metal Oxide Semiconductor is not correct for currently produced MOS FETs Next to ensure that only the gate regions are left covered with polysilicon the photolithography process is used again The areas of the oxide and poly silicon that are left exposed can then be removed with suitable chemicals Figure MOS Transistor shows the transistor after the unwanted polysilicon has been removed Conductive gate polysilicon lt Insulator oxide Source Drain p type active area Lightly doped silicon wafer MOS Transistor Cross section of an n p n MOS transistor after the gate has been added Finally the source and drain are coated with metal for the wiring process which is accomplished using photolithography Usually several layers of metal and insulator must be used to make a functional IC SAMPLE MAINTENANCE The benefits of reducing the dimensions of a transistor are clear the smaller the transistor the less space required to run a device The higher the density of transistors on a single chip the greater the processing power Decreasing the dimensions is not straightforward however and requires intensive research into all stages of the production process the production of the sem iconducting wafer the light sensitive resins the process of the optical projec tion of the masks the doping process the materials and methods used in the production of the insulating oxide layers a
42. m decaying or getting dirty Measurement The skin cross section is a narrow specimen which has many distinct layers and distinct structures within these layers The specimen being narrow makes it important to line up the tip properly to touch down on the skin as opposed to the glass slide around it You will know if you are scanning the glass because it will appear very flat in a height scale of 100 nm while all of the layers of skin have noticeable features in this z range Figure Skin Over view gives an idea of the structures that can be found within the skin Image Acquisition In contrast to the other samples in the sample kit the macroscopic position of the AFM tip on the sample determines what kind of structures you will see Thus this sample is a good sample to practice coarse positioning of the sample Moreover the skin is hardly visible see Skin Overview To prepare the measurement Look at the sample under an optical microscope Search some distinct features particles or dirt Memorize the position of the skin Put the sample under the AFM microscope 70 SKIN CROSS SECTION ET Bici Skin Overview The image is an overview of the skin cross section sample The different layers are visible as well as the AFM cantilevers indicates The image was taken with the easyScope camera microscope Align the tip to the memorized position The fact that the specimen has many different layers is also important in scan
43. mage to the probe tip The oscillation should be visible in both the topography and the error signal deflection or amplitude depending on the measurement mode images Be sure to monitor your system for indications that the controller is becom ing unstable First it will overshoot and then it will ring which is repre 21 22 CHIP STRUCTURE IN SILICON ZOutput ForwardScan 4 07pm Rawdata ZOutput range 1 05um Opm x 39 9um High Gain Oscillating signal when gain is set too high sented by a vibration with decreasing amplitude at the step edges Also the error signal in this case the cantilever deflection will start to increase again Line fit 1 35um Output range Opm xt 39 9um RAM chip structure imaged at different integral gain settings In the lowest region ofthe image the gain is too low at the center it is optimized at the top it is too high Integrated Circuit Technology An integrated circuit IC is a single chip of semiconducting material usually Silicon on which all the functional elements transistors wiring and con tacts of an electronic circuit are integrated Transistors The transistors are the enabling part of the integrated circuit In analog cir cuits transistors are used as amplifiers in which a small current or voltage INTEGRATED CIRCUIT TECHNOLOGY controls a much larger current or voltage In digital circuits a voltage is used to switch a current
44. n of the analog music This is accomplished by measuring the intensity of the 31 32 CD STAMPER ZOutput CrossSection 153nm Rawdata ZOutput range TT Opm Section 20 9um Bump length Using the Measure Length tool in the track direction wave at regular intervals and categorizing these measurements into discrete intensity values The greater the frequency of the intervals the closer the approximation of the original wave The rate at which these measurements are taken is called the sampling rate A CD uses a sampling rate of approx imately 44 000 samples per second so a CD can store just over 780 million bytes of digital music closely approximating the original analog signal The actual CD stamper is created through a process involving photolithog raphy An original CD stamper is made on a piece of glass which is polished to ensure flatness and then covered with photoresist The CD is then exposed to UV light through a mask which has clear parts where the CD will have pits and opaque parts everywhere else When the CD is then placed in devel oper solution only the sections of the photoresist that have been exposed to the radiation will dissolve Next the CD master is sputtered with a very thin layer of metal usually silver Then it is played to make sure that the process was successful Finally the master is covered with a thicker layer of metal This negative CD is the CD stamper represented by the sample in t
45. nd the production of the metal connections Sample Maintenance CAUTION Cleaning the sample is always accompanied by the risk of permanently dam aging it To reduce the risk of damaging the sample it is best to reduce the need for cleaning it by always storing it in its container when not in use Only clean the sample when it has become too dirty to allow good measure ments Clean the sample using only highly pure solvents which prevents the depo sition of residues To remove any large dust that may have accumulated on the surface leave the chip and the sample disk in an ultrasonic bath as follows 10 minutes in distilled water Sonicating first in water removes any salt which would not be removed by the alcohol 10 minutes in ethanol Removes finer dirt 10 minutes in propanol 27 28 CHIP STRUCTURE IN SILICON Removes even finer dirt Using the propanol last also leaves the sample cleaner as it evaporates leaving fewer spots To ensure that the sample dries as cleanly as possible dab it off with a clean paper tissue to remove excess propanol when removing it from the propanol CD Stamper This sample represents two application areas Data storage and optical devices The data storage industry uses the AFM for quality control in the production of various parts of hard disk drives and the masters used for pro ducing CDs and DVDs Two key capabilities make the AFM the tool of choice with hard dis
46. nes across the images are noise that is inherent to images taken with any scanning instrument such as an AFM The noise is not an important factor in height signals above 30 or 40 nm since the signal to noise ratio is very high In very flat samples however the lower ratio results in an image with lower resolution To reduce the effects GOLD CLUSTERS ZOutput ForwardScan 1 5pm E G o E v a Oum Opm Ke 1 5pm Image of the Gold Clusters Sample of the noise cover the microscope with a padded cover and or keep it on a vibration isolated table In a scan range of about 10pm many clusters are present but not always apparent The clusters vary in size but they are typically round with a diam eter of approximately 100nm In a scan range of 10um there will be close to 100 clusters across the image which makes it difficult to resolve individual structures Figure Image of the Gold Clusters Sample shows a scan range of 1 5um and even in this range it is difficult to resolve individual clusters with much certainty Image Analysis The gold cluster sample is useful for introducing the concept of roughness Roughness is one type of deviation of the sample surface from an ideal sur face For plane surfaces the only other other type of deviation is called wav iness The opposite of roughness is smoothness the opposite of waviness is flatness The separation between roughness and waviness is determined by the l
47. ning since not all of the layers will be visible in one scan range The best strategy to see all of the structures within the skin sample is Choose a big scan range 50um to100um Begin at one side of the skin cross section Take an image Retract to a safe position Move slowly across the sample Only move the sample a little bit so as not to skip over any interesting struc tures When doing this in a controlled way it is possible to assemble a mosaic of several images to a bigger one as shown in figure Skin Mosaic MEASUREMENT Skin Mosaic 4x2 images at 100um scan rage each The mosaic image shows a nice overview of the skin section The images were assembles with a common image treatment software Image Analysis Figure Skin Images shows the skin images taken while moving along the cross section with the coarse adjustment Figure Skin Images A shows the outer layers of dead epithelial skin cells The outermost layer on the right side in the image is already beginning to flake off and is much less dense than the inner layers depicted on the left Contin uing to move in the same direction across the skin will unveil deeper and deeper layers of skin Figure Skin Images B contains multiple layers of skin The image shows the living epithelial layer of skin The structures at the bottom right hand cor ner of the image are the beginning of yet another layer of skin This region is the one which contains collagen the
48. nm For a sample with perfectly uniform height the height of each point will be equal to the mean height The difference between the mean and the height at each point will be zero so the RMS roughness will also be zero ZOutput ForwardScan ZOutput ForwardScan ZOutput ForwardScan Line fit 9 89nm Line fit 10 5nm Mey ae om ATT Line fit 10 5nm ZOutput range E amp Onm y 402nm Scan Speed and Rougness Gold clusters sample scanned at 400 nm 800 nm and 1500 nm at the same scan rate time scan line 39 GOLD CLUSTERS In figure Scan Speed and Rougness p 39 shows scans of the gold clusters sample taken at three different scan ranges all centred on the same point and scanned at the same rate Roughness analysis was performed on each The roughness analysis on the smallest scan region 400 nm image yielded a Sq of 1 6 nm The next larger scan region 800 nm had a slightly smaller Sq of 1 58 nm and the largest region 1500 nm yields a still smaller value of 1 3 nm One possible explanation for this change in roughness is that the measure ment speed is too high at the larger scan sizes so the tip does not follow the surface as accurately As the scan sizes increase the scan rate lines of data recorded per second remains the same so the probe tip encounters variances in the surface topography at a faster rate This makes it harder for the feed back loop to
49. nteraction constant Therefore the cantilever must be moved away from the surface or towards the surface depending on how the force changes This movement is then recorded as topography signal when the tip is scanned over a sample The topography can thus also be interpreted as a map of equal forces In this way it is possible to detect any kind of force as long as the tip is sensitive enough i e as long as the force interaction induces a meas urable deflection of the cantilever Hence not only interatomic forces but also long range forces like magnetic force and electrostatic force can be detected The AFM Setup Independently of the type of tip sample interaction an AFM basically consist of five major parts shown in figure AFM Setup and described in the following paragraphs 1 A force sensor which is basically a sharp tip lt 10nm mounted on a sen sitive cantilever 2 A system which is moving the sample or the sensor in order to probe the sample surface 3 A sensor which detects the cantilever deflection for example a laser deflection system or piezoresistive system 4 A feed back system which regulates the force interaction 5 Controller electronics which records movements controls the feedback loop and sends the measured data to ta personal computer software THE AFM SETUP 5 controller and PC laser 3 deflection detection system Y ren 4 feed back 1 force sensor 2 scanning system
50. ons rather than a how to manual By exploring the samples there is freedom to make mistakes and to learn through those mistakes which conditions produce desired results and which do not It is not neces sary to read this manual in the predicted order The manual starts with a brief introduction into atomic force microscopy AFM followed by a description of the samples of the extended sample kit The samples introduce the following topics e Chip Structure in Silicon Setting the optimum PID feedback gains e CD Stamper Using the distance and length tool Gold clusters Doing roughness analysis Nanotubes Choosing the optimum Set point e Glass beads Estimating the tip radius Staphylococcus Aureus Approaching a non reflective sample Microstructure Setting the correct vibration amplitude e PS PMMA Thin Film Performing phase contrast Skin Cross Section Positioning the tip at specific positions Aluminium Foil Estimating data processing artefacts A Brief Introduction to Atomic Force Microscopy Introduction In 1986 Gerd Binnig and Heinrich Rohrer won the Nobel Prize in Physics for the invention of the scanning tunnelling microscope STM and the fact that it could achieve atomic resolution They observed that if they held a metallic tip of 10 angstrom above a conductive surface they could measure a tunnelling current in the order of a nanoampere When the tip was then scanned over the conducting samp
51. onsidering that the free bacteria are spherical with an approximate diameter of 0 7um It is possible that the proc ess which fixes the bacteria to the slide results in flattening them as well Figure 3D View is a parallel perspective projection of the 3D data of the same measurement The height scale is exaggerated with respect to the x and y MEASUREMENT Topography Scan forward 118p Raw data 610nm 1 79 3D View A parallel perspective projection of the same 3D data in the 10x10um image in figure Staphylococcus Bacteria Images right image shows that some bacteria are stacked on top of the others scale the bacteria appear to be much higher than they actually are Each view illustrates a different feature of the sample The color map image emphasizes the round shape of the bacteria as well as their random horizon tal clusterings and bare spots The color map image provides color contrast which makes it possible to determine that some bacteria are higher than oth ers through the color scale but the 3D view directly shows that some bacteria are clearly stacked on top of several others Figure Zoomed Image shows a 5um scan range of the bacteria Zooming in on the bacteria highlights the topographical features of the individual bacte ria The little dots that are visible on the surface of the bacteria in figure Zoomed Image were not visible in figure Staphylococcus Bacteria Images This is not because the bacteria
52. oom in on an area of interest it may not be possible to measure the nanotubes at high resolution Figure Nanotube Images shows images of nanotubes taken at optimal and less than optimal settings Set a small scan range 2 um or less Take a scan Optimize scanning parameters Zoom out by taking a scan at a relatively large scan range 15 um Identify an area of interest Zoom back in Figure Optimising the Set Point illustrates an optimization sequence At first the set point is too high 10nN so the nanotube gets pushed around This 44 NANOTUBES Topography Scan forward Topography Scan forward E z E co co E o o SG A z gt gt O I o D D 2 2 o o E gt gt Ra S a o oO S z D D o o 8 8 e OF ne 15pm i 15pm Opm Opm x x Nanotube Images Left blunt tip and set point too high 19 nN Right sharp tip and better set point 2nN makes it appear streaky and not as wide as it should be With the Set Point lowered the nanotube is imaged more stably Note that the dirt that was pushed to the side in the first scan is visible on the side of the second scan Topography Scan forward Topography Scan forward 506nm 506nm Mean fit 5 77nm Mean fit 5 77nm Topography range Topography range Onm Onm xe Optimising the Set Point Lefi force set point 10 nN is too h
53. possible to treat the sample without damaging it these microscopes do not provide specific height information about the sample The skin cross section in this sample was treated and is obviously not still living material The sample shows however that even the soft tissue of skin can be scanned in contact mode Atomic force microscopy gives information about the height of the structures that make up mammalian skin even at very high resolution Dynamic force microscopy provides a method to explore the topography of soft tissue at a very high magnification without damaging the sample since there is much less applied force The fact that dynamic force microscopy can be performed with the sample submerged in solution means that skin sam ples or other living samples may be scanned without treating or damaging them AFM is currently being used to study DNA RNA proteins and indi vidual living cells One of the keys to studying biology is understanding that structure is related to function The functions of certain proteins cells etc are unknown but the opportunity to explore the microscopic structure of living material offers many clues as to specific cellular functions 73 74 SKIN CROSS SECTION Aluminium Foil Aluminium foil is a common household product exhibiting interesting fea tures on a height scale under lum This sample illustrates a different way to view something that people are used to seeing on a relatively large scale
54. primary protein responsible for bind ing tissues within the skin The bottom figure Skin Images C shows a hair follicle and the structure within it This is the hole through which actual hair would have grown Hair follicles are quite large compared to the other structures in skin so an entire follicle may not fit into one scan range Figure Skin Images D shows the collagen layer of the skin section The widely varying height of the collagen bundles makes it difficult to resolve detail on the structure of the collagen The collagen cross sections appear circular 71 72 SKIN CROSS SECTION E E m E v 3 Line fit 362nm Line fit 317nm Line fit 273nm Oum ye 95 Sum Skin Images A Outer layers of dead epithelial skin B Epithelial layer of skin C Hair follicle D collagen layer when they run through the skin cross section but they may not always be exactly perpendicular to the cross section of skin All the parts just discussed can also be seen in figure Skin Mosaic p 71 The epithelial skin cells are in the bottom left corner the hair follicle at the top left corner and the collagen cell on the right side of the image Biological Samples Soft tissue samples can be imaged with optical microscopes but only up to a certain resolution and electron microscopes involve treating the sample BIOLOGICAL SAMPLES Even if the tissue sample is in the resolution range for optical microscopy or if it is
55. r ing scanning are related to energy dissipation during tip sample interaction and can be due to changes in topography tip sample molecular interactions deformation at the tip sample contact Depending on the operating conditions different interaction forces might be required to produce accurate images These changes will often affect the phase image In practice the material difference between different regions usually has to be substantial for contrast to be realized e g rubber particles on a glass substrate carbon fibres on an epoxy layer Thus phase contrast images are purely qualitative due to inaccurate or unknown spring constants unknown contact geometry and contributions from different types of tip sample interactions Skin Cross Section This sample is typical for medical microscopy The main application is the study the cross section of cells with high resolution The skin cross section sample is a thin cross section of skin protected by plastic and fixed to a glass slide Move the sample back and forth in the light You can see that at the centre of the slide there is a thin curve The entire length of the curve is the skin consecution made up various layers The skin sample was first fixed to the glass slide and then dehydrated through immersion in a series of water ethonol mixtures with increasing eth anol concentration Next the skin was infiltrated with and embedded in a plastic which protects the skin surface fro
56. rack distance ZOutput ForwardScan Raw data 221nm ZOutput range 20 ym Image of CD Stamper Note that the curvature of the tracks is not discernible at this scan range Image Analysis There are two different methods for measuring the track distance The Measure Length tool measures a cross section as shown in figure Track Dis tance with Lenght Tool p 31 and gives a measurement of 1 63 um The Measure Distance tool shown in figure Track Distance with Distance Tool p 31 measures the perpendicular distance between two parallel lines It gives a measurement of 1 598 um basically identical to the nominal dis tance of 1 6 um You can also measure the length of individual bumps in the image The shortest possible bumps in a CD stamper are approximately 0 8 um long while the longest possible bumps are close to 3 um long Figure Bump length p 32 shows a section of the CD stamper along one track image taken with 20 um scan range The length of one of the bumps was measured using the Measure Length tool on a cross section OPTICAL DATA STORAGE 158nm Rawdata 79 5nm Opm Section 20 79m Track Distance with Lenght Tool Using the Measure Length tool Track Distance with Distance Tool Using the Measure Distance tool Optical Data Storage CDs continue to be the most common optical storage devices The original music stored on a CD is an analog wave and the CD stores a digital versio
57. rasonic bath for approximately 20 minutes GLASS BEADS This process will break up any groupings of beads in the solution The beads will form aggregates over time due to simple attractive interactions between them such as Van der Waals forces but the goal is to have individual beads come together on the glass slide and eventually form crystalline structures Use an eyedropper to place a drop of bead solution onto the cleaned glass slide Try to form the largest drop possible without spilling over the sides of the glass Dry the sample in air at room temperature This may take several hours When the bead solution is dry Bake the sample disc in an oven at 250 for 2 hours and 30 minutes Baking the sample helps the beads remain fixed Allow the sample to cool completely before imaging Image Acquisition Approaching the Sample This sample is one of the more difficult to approach as it is non metallic and not very reflective If you can see the cantilever s shadow or reflection you can use it to judge the distance If you find it difficult to recognise the cantilever s reflection then slightly move the sample holder the strucures on the sample will move but the reflection will stay in the same place If you cannot see the cantilever s reflection perform a very slow coarse approach while judging the distance on the focal plane of the side view as follows e When the tip is on the sample the focal plane crosse
58. ria have been burned which makes it possible to locate the parts of the slide that are covered with bacteria STAPHYLOCOCCUS AUREUS Image Analysis The individual bacteria are approximately 0 7 um in diameter so it is possi ble to make out several bacteria in a relatively large scan range This sample has regions with a very high concentration of bacteria as well as some with lower concentrations and some bare spots The left image in figure Staphylo coccus Bacteria Images shows a 20x20um scan region densely packed with bacteria It should be easy to zoom in on a much smaller scan region where the bacteria are still very concentrated Topography Scan forward E E 5 D E v 3 Raw data 450nm Topography range Opm xt 19 9um Staphylococcus Bacteria Images Left 20x20um scan range Some regions appear to have crystalline structure while other regions contain unordered gaps The variations in color brightness of the individual bacteria correspond to the variations in height of the bacteria Right 10x10um scan range Both images have a height scale of 450 nm The image of the 10x10um scan on the right shows a less dense region In a scan range of this size and where the bacteria are less dense it is clear why the bacteria were named after grapes The round bacteria appear like individ ual grapes and clusters like the one in the top right corner appear like bunches of grapes The height scale of 450nm is small c
59. rom the top of these images to the bottom but the lines are not as well defined or as thin as those on the reflective side of the foil There are pits in the diffuse reflective side as well but they appear less deep and more elongated than on the reflective side Additionally whereas the pits on the reflective side are mainly located on bright i e high lines here they seem to run parallel to the smudged lines To further analyse the dull side of the foil attempt a section analysis You should find that the shiny side is between two and three times as flat as its non reflective counterpart BULK FOIL PRODUCTION Topography Scan backward Topography Scan backward 60pm 60pm Mean fit 2um Mean fit 2um Topography range Topography range Opm Opm Opm xt 60pm Opm xt 60pm Comparison Lefi reflective side Right diffuse reflective side Bulk Foil Production Commercially available aluminium foil generally has one side different from the other The two pieces of aluminium foil in this sample are representative of these different sides The lines across the foil and the craters spread throughout it are conse quences of the rolling process that renders the foil flat and shiny Aluminium is mined refined and processed into large sheets Initially the sheets are quite thick close to 1 cm and they must go through several rolling procedures until they are flat enough for use as foil Foil is loosely defined as a sheet
60. s the sample at the tip position e When the tip is further away the focal plane crosses the sample more behind the cantilever MEASUREMENTS Scanning Start with a low force set point for best results Applying too much force may move some of the beads around and create wide horizontal stripes across the image If you get stripes in your image Lift the tip and then Bring it back into contact If the tip is simply dirty you can remove the dirt by Retracting the tip Re extending it again If there are still stripes in your image Move to another region on the sample The problem may be that the region where you are scanning does not have perfectly fixed beads In a region of more ordered beads the beads will stay in place Set the scan range to 1pm Since the beads are approximately 120 nm in diameter you should be able to see about 10 of them across the image If your image shows islands of beads surrounded by very flat areas move to a region of better ordered beads In general the region with the best ordering is close to the center of the spot on the slide Figure Well ordered beads shows a well ordered region near the center of the spot Identifying the Bead Structure In a well ordered region each bead will be surrounded by 6 others Identify a single bead and count the beads around it to see if it is surrounded by 6 others Then check to see if each of these 6 beads is in turn surrounded
61. sample is a thin layer of a blend of two polymer solutions spread onto a piece of silicon wafer PS polystyrene and PMMA poly methyl methacrylate when mixed together separate into well defined phases on the silicon The silicon substrate is first cleaned to make sure that it was free of large dirt It is submerged in an ultrasonic bath in distilled water then ethanol then propanol to perform the cleaning Since the idea is to have an even coating of the PS PMMA film dirt on the silicon risks ruining the film To create the PS PMMA mixture first individual solutions of PS and PMMA were created Each polymer was dissolved in toluene a potent sol vent then parts of each polymer solution were mixed together after allowing an entire day for the polymers to dissolve A few drops of the mixture were placed on a piece of silicon and the substrate was spin coated dry Spin coat ing leaves the silicon surface with only a very thin layer film of the PS PMMA blend Measurement Image Acquisition Topography The tip can be positioned on the entire sample However the structure vary depending upon region For acquiring topographical images the AFM can be operated in static or dynamic mode For recording a phase contrast see Phase Contrast p 66 the AFM needs to be operates in phase contrast mode which is a extended dynamic mode Set a scan size to maximum 10pm Due to the adhesive character of the sample a bigger scan sizes will
62. senta tion of the feed back system is shown in figure PID feed back set point Se SEE I S canti lever measure difference deflection deflection A AS error y paa I ASreg I scanner move Ms PID topography PID feed back As the name suggests the PID amplifier has three domains of amplification Proportional Gain Integral Gain THE AFM SETUP Differential Gain These three gains can be set individually and define how fast and in which manner the error is minimised and the therefore how good the topography of the sample is reproduced in the measurement In most cases an appropri ate combination of these three gains leads to the desired result Thus it is important to understand its characteristics To illustrate the effect of the PID gains consider the following experiment A step signal from 0 to 1 will be measured see figure Step The goal is to Ste Displacement m Step reproduce the rectangular step as precisely as possible Hence the PID gains must be adjusted Figure P Gain shows the result when only the proportional gain P is turned up We note that the topography shows a long rise time Displacement m E P Gain 13 14 A BRIEF INTRODUCTION TO ATOMIC FORCE MICROSCOPY Displacement m 1 Time sec PD Gain 14 12 gt aiai E E Eos E 206 2 504 02 0 0 0 5 1 5 2 Time
63. tes this MEASUREMENTS Figure Detail of Reflective Side shows a zoomed measurement on the area indicated in figure Reflective Side Overview All images show the same area the image on the left was measured in the same orientation as the overview the other images are measured in a 90 degrees rotated direction In the image on the left the lines of different height are well defined and quite large The image in the centre shows the same area measured at 90 degrees rotation In both the left and centre image the mean height of each line is subtracted before displaying the image Due to this data processing the centre image clearly shows that the larger lines are composed of smaller lines with less height contrast on a smaller scale and the holes in the surface become very conspicuous To show that both the left and the centre data actually display the same structure the image on the right shows the same measurement without data processing Topography Scan backward Topography Scan forward Topography Scan forward 10Aym Rawdata 600nm E E E E E 2 Topography range Topography range m Topography range Op Oum 5 Opm 101pm Opm xe 101pm x 101pm x Detail of Reflective Side Aluminium foil in a scan range of 10 um Left image 0 degrees scan rotation mean fit processed data Centre images 90 degrees scan rotation mean fit processed data Right image 90 degre
64. the radii of the tip and the cylinder The top of the cut off cylinder will have the same height as the actual cylinder i e the nanotube as illustrated in figure Nanotube tip Geometry Notice that the shape of the tip does not permit it to touch the silicon surface at the edge of the nanotube Since the tip is approximately spherical there is a significant area of the silicon surface on either side of the nanotube that it cannot touch At the point where it does touch the silicon surface a right triangle can be constructed to approximate the actual tip radius as shown in figure Nanotube tip Geometry 45 NANOTUBES Topography Scan forward Topography Scan forward 506nm 1 06pm Mean fit 5 77nm 3 E K P w E c G o Topography range Onm Opm Opm xt 1 05pm xe Topography CrossSection E E c c im im a a gt Y 2 gt LI awa SVAN CRRA E a ira a a O e E E c c a T T E T T Onm Section 164nm Onm Section 328nm Nanotube Cross Section Measurements Left with sharp tip height is 3 7 nm width is 20 nm right with blunt tip height is 3 6 mm width is 47 nm Nanotube tip Geometry The larger circle represents the end of the tip where it is approximately spherical The smaller circle is the cross sectional view of a nanotube By setting up a right triangle whose hypotenuse is the line joining the centers of the two circle it is possible
65. to calculate the tip radius CARBON NANOTUBES One leg of the triangle has length d which is the horizontal distance meas ured from the highest point on the nanotube to the point where the tip touches the silicon The other leg is R r which is the radius of the tip minus the radius of the nanotube The hypotenuse is the sum of the two radii Using the Pythagorean relationship we find that R r d R r R 2rR 1 d R 2rR r d 4rR d2 R C 4r Entering the values determined from the analysis in figure Nanotube Cross Section Measurements p 46 d 10 nm and 23 5 nm and r 1 8 nm gives a tip radius R of approximately 14 nm for the sharp tip This is reasonable considering that the nominal tip radius given by the manufacturer is 10 nm For the blunt tip d 10 nm and 23 5 nm and r 1 8 nm the tip radius becomes 77 nm Carbon Nanotubes A carbon nanotube is as the name suggests a tiny cylinder composed of carbon atoms More specifically it is a lattice of graphitic carbon rolled into a tube Figure Nanotube Molecular Structure p 48 shows an example of the molecular structure of a carbon nanotube The ends of the tube in figure Nanotube Molecular Structure p 48 are not capped but it is possible to seal a nanotube at both ends with a fullerene A fullerene is similar to a nanotube in molecular structure but it is spherical rather than cylindrical The bonds that hold nanotubes together are entirely sp2 bonds as in gr
66. to measure deflections below one angstrom There are other detection techniques used for detecting the deflection of the cantilever like Interferometric optical system Piezoresistive detection which will not be further discussed here 11 A BRIEF INTRODUCTION TO ATOMIC FORCE MICROSCOPY The PID Feedback System Before starting any AFM measurement it is necessary to understand how the feedback regulation system works This regulation enables the acquisition of an AFM image As described previously the cantilever deflection is detected by a sensor This position is then compared to a set point i e a constant value of cantilever deflection chosen by the user As the deflection of the cantilever is directly related to the tip sample interaction force the set point is usually given in Newton N Typical forces are in the nN range The dif ference between the actual interaction force and the desired force is called the error signal AS This error signal is then used to move the tip or sample to a distance where the cantilever has the desired deflection This movement is then plotted in function of the lateral position of the tip and is the so called topography The goal of the feedback system is to minimize the error in a very fast manner so that the measured topography corresponds to the real topography of the sample Therefore the error signal must be amplified by a PID amplifier Proportional Integral Differential A schematic repre
67. ude Scan forward Raw data 449m 49 3mY Opm xt Line ft 14 3m Y 50pm Microstructure Topography and amplitude image of the microstructure sample The line graphs show a cross section of the images above at the position indicated by the arrow The vibration amplitude was set to 00mV Figure Microstructure shows the topography and amplitude image of the microstructure sample The line graphs show a cross section of the images above at the position indicated by the arrow It is clearly visible that the in the topography the slopes are steep After each perturbation the amplitude signal is also corrected to the Set point value very quickly MEASUREMENT Topography Scan forward Amplitude Scan forward 50um Mean fit 150nm Line ft 43 68m Topography range Amplitude range Opm Topography Scan forward Amplitude Scan forward 99 4nm 74 5mY 3 a E a 5 E gt amp EG Ge a Ge ada baie ci E 2 g E Sa amp I 100nm 25 2mY el Oum mu 50m Opm yt 50pm Vibration Amplitude Settings Topography and amplitude image of the microstructure sample The line graphs show a cross section of the images above at the position indicated by the arrow The vibration amplitude was set to 0mV FigureVibration Amplitude Settings shows the topography and amplitude image of the microstructure sample with too low vibration amplitude The line graphs show a cross s
68. uts back together This way there will never be consec utive numbers one ending with a 1 and the other beginning with a 1 This process also ensures that if a part of the CD is damaged there will not be entire sections of a song missing but rather only split seconds from several songs Since there are always between 2 and 10 Os together a CD has between 3 and 11 spaces between 1s and these 9 lengths are represented by different bump lengths on the CD stamper 33 CD STAMPER In an image of a large scan range attempt to locate the smallest bump and see how many of the 9 lengths you can find in the image They should approximately correspond to 4 3 5 3 6 3 7 3 11 3 the length of the smallest bump Within this information is encoded data that tells the CD player where to read so that you hear the information in the correct order CDs also use parity check bits which count the total number of 1s in a string of characters If the number is even the check bit would be a 0 if odd it would be 1 Having many check bits increases the chances of detecting errors in reading the signal The final step in being able to hear the encrypted music is the CD player reading the CD The laser beam that reads the CD has to stay focused on the track at all times so the beam and the detector slowly move along the CD in the radial direction as the information is read At the same time as it moves outward though the speed of the bumps increases s

Download Pdf Manuals

image

Related Search

Related Contents

Americas: TH  取扱説明書/2.6MB  おかしなアイスキャンディー チュッパチャプス  Instruction Manual  Remeha Avanta Plus Gas 360 Technical Instruction  Bulletin 0020 - Mr.Cappuccino  King Canada BB-4022 User's Manual  

Copyright © All rights reserved.
DMCA: DMCA_mwitty#outlook.com.