        MRI Guidelines for InterStim Therapy neurostimulation systems
         Contents
1.  Clinician Programmer cannot be used to change or verify neurostimulator  settings       Use the Model 3037 Patient Programmer to turn the neurostimulator off and reduce  the amplitude to 0 0 V      If the magnet switch is enabled  the patient may experience uncomfortable or  unintended stimulation during the MRI scan      If an active program with bipolar electrode settings is used  bipolar mode   the  likelihood of uncomfortable stimulation during an MRI scan will be reduced      The patient should be instructed to see the implanting physician or managing  physician to verify neurostimulator function after the MRI head scan     9  Ensure that the head coil is a transmit receive head coil    Note  If you are unsure if your MRI system has RF transmit receive head coil capability   check with the MRI equipment manufacturer  Do not use a receive only head coil or  an RF transmit body coil because of the risk of heating and potential nerve and tissue  damage during MRI scans    10  Instruct the patient to immediately inform the MRI operator if any discomfort   unexpected stimulation  shocking  or heating occurs during the examination    11  Verify that all proposed MRI examination parameters comply with the MR scanning  conditions on page 5 of this manual  If these conditions cannot be met  do not proceed  with the MRI examination     During an MRI head scan    Conduct the MRI head scan in Normal operating mode only  Do not scan in First level  controlled operating mode because 
2.  amp  Medtronic  MRI Guidelines for InterStim  Therapy neurostimulation systems    Instructions for use                   1USA  Rx only    Explanation of symbols on product or package labeling  Refer to the appropriate product for symbols that apply     Conformit   Europ  enne  European Conformity   This  symbol means that the device fully complies with AIMD  Directive 90 385 EEC  NB 0123  and R amp TTE Directive    oN  MA                   0123 1999 5 EC   EC  REP Authorized representative in the European community  Manufacturer       For USA audiences only    e gE             Magnetic Resonance  MR  Conditional    Magnetic Resonance  MR  Unsafe    2012 02 MRI Guidelines for InterStim Therapy neurostimulation systems English 1    Medtronic    InterStim    InterStim iCon    N Vision    and SoftStart Stop   are trademarks of  Medtronic  Inc   registered in the U S  and other countries     2 English MRI Guidelines for InterStim Therapy neurostimulation systems 2012 02    Table of contents    Description 5  MR scanning conditions 5  MRI and InterStim Therapy 5  Information for prescribers 5  Warnings 5  Precautions 7  MRI head scan using an RF transmit receive head coil 8  Preparing for an MRI head scan 8  During an MRI head scan 10  After an MRI head scan 11  Patient programmer instructions for MRI technicians 11  InterStim iCon Model 3037 Patient Programmer instructions 11          Refer to the Indications Insert for indications and related information    Refer to the appropri
3.  and the risk  of uncomfortable  unintended stimulation  often described as shocking or jolting sensation   is justified   Ensure that the electrodes remain off during the MRI scan by turning the neurostimulator  off  If the neurostimulator cannot be turned off  eg  because the battery is depleted  which  may occur in an abandoned or unused system  and the neurostimulator programmed in a  unipolar configuration  ie  the case is programmed as a positive electrode  is on during the  MRI examination  the likelihood of undesired MRI induced effects  eg  shocking and jolting   increases dramatically   Patient interaction during MRI     If possible  do not sedate the patient so that the patient  can provide feedback of any problems during the examination  Monitor the patient during  the MRI examination  Verify that the patient is feeling normal and is responsive between  each individual scan sequence of the MRI examination  Discontinue the MRI immediately  if the patient experiences any heating  pain  shocking sensations uncomfortable  stimulation  or unusual sensations   Serial numbers ineligible for MRI scans  Model 3023 Neurostimulator only      Due to  increased risk of neurostimulator damage  patients implanted with InterStim Model 3023  Neurostimulators with the following serial numbers should not have MRI scans      Less than NBV132955H     Between NBV133037H and NBV133063H     Between NBV628045S and NBV628263S    Control magnet function after MRI  Model 3023 Neurostimulator
4.  only      Do not conduct  an MRI scan if the patient can only use a control magnet to turn the neurostimulator on or  off  An MRI scan may permanently damage the magnet switch in the neurostimulator  If the  magnet switch in the neurostimulator is damaged  the patient will require a patient  programmer to turn the neurostimulator on or off     MRI head scan using an RF transmit receive head coil    Ae all of the instructions stated in this manual are followed  MRI examinations of  the head only using a 1 5 T horizontal closed bore MRI system with an RF transmit receive  head coil may be safely performed     Preparing for an MRI head scan  Prior to performing an MRI head scan on a Medtronic InterStim Therapy patient    1  Determine whether the patient has any other implants or conditions that would prohibit  or contraindicate an MRI examination  The most restrictive MRI exposure requirements  must be used if the patient has multiple active medical device implants  Contact the  appropriate device manufacturers if you have questions    If the following requirements cannot be met  do not proceed with the MRI  examination     8 English MRI Guidelines for InterStim Therapy neurostimulation systems 2012 02    2  Verify the implant location of the InterStim Therapy system and ensure that the  InterStim Therapy system will not be inside the transmit receive head coil  Do not use  the RF transmit body coil    Note  When implanted per approved indications  InterStim Therapy components 
5. 00       Germany    Medtronic GmbH  Tel   02159  81490  Fax  02159  8149100    Greece    Medtronic Hellas S A   Tel  210 67 79 099  Fax 210 67 79 399    Hungary    Medtronic Hungaria Kft   Tel  1 889 06 00   Fax 1 889 06 99    Ireland    Medtronic Ireland Ltd   Tel   01  890 6522   Fax  01  890 7220  Italy    Medtronic Italia SpA  Tel  02 241371   Fax 02 241381   Tel  06 328141   Fax 06 3215812    Japan    Medtronic Japan  Tel  03 6430 2016  Fax 03 6430 7110    Latin America   Medtronic  Inc   Tel   1305  500 9328  Fax  1786  709 4244    Norway    Medtronic Norge AS   Tel  067 10 32 00   Fax 067 10 32 10  Poland    Medtronic Poland Sp  z o o   Tel   022  465 69 00   Fax  022  465 69 17  Portugal    Medtronic Portugal  Lda   Tel  21 724 5100   Fax 21 724 5199                      Russia    Medtronic Russia  Tel   8495  580 7377  Fax  8495  580 7378    Slovakia   Medtronic Slovakia  o z   Tel  0268 206 911   Fax 0268 206 999    Spain    Medtronic Ib  rica  S A   Tel  91 625 0400   Fax 91 650 7410    Sweden   Medtronic AB   Tel  08 568 585 00  Fax 08 568 585 01    Switzerland    Medtronic  Schweiz  AG  Tel  031 868 0100   Fax 031 868 0199       The Netherlands   Medtronic B V    Tel   045  566 8000  Fax  045  566 8668  U K     Medtronic U K  Ltd   Tel  01923 212213  Fax 01923 241004    USA    Medtronic  Inc    Tel   1763  505 5000   Fax  1763  505 1000  Toll free   1 800  328 0810          Medtronic    Alleviating Pain   Restoring Health   Extending Life    Manufacturer asal  Me
6. 012 02 MRI Guidelines for InterStim Therapy neurostimulation systems English 5    contains ferromagnetic material  which can be affected by the MRI magnet and is unsafe  in the MRI environment     MRI RF transmit body coil     Do not use an RF transmit body coil or a receive only head  coil  Serious patient injury could result     MRI transmit receive head coil     An MRI examination of the head only  no other  part of the body has been tested  can be conducted safely using an RF transmit receive  head coil when all instructions in this manual are followed   Limitations        These instructions apply only to Medtronic InterStim Therapy neurostimulation  systems for approved indications listed as eligible for an MRI head scan     Prior to an MRI head scan  determine whether the patient has multiple active  medical device implants  such as deep brain stimulation systems  implantable  cardiac defibrillators  and others   The most restrictive MRI exposure requirements  must be used if the patient has multiple active medical device implants  Contact  the appropriate device manufacturers if you have questions  If you are unclear what  implants may be present  perform an x ray to determine implant type and  location     Scans must be limited to 1 5 T horizontal closed bore MRI systems  Use of different  MRI systems  such as open bore or 3 0 T MRI  is not permitted because it may  expose the patient to unacceptable or unknown risks     Do not place any part of the implanted system w
7. are  outside of the head coil    3  Inform the patient of the risks of undergoing an MRI    Note  Patients should always inform healthcare personnel that they have an implanted  neurostimulation system before any medical procedure begins    4  If possible  do not sedate the patient so that the patient can provide feedback of any  problems during the examination    5  If the patient does not use a Model 3037 Patient Programmer  instruct the patient to  see the implanting physician or managing physician to confirm MRI head scan eligibility   refer to Table 1  and neurostimulator settings for MRI  refer to Table 2  using the Model  8840 Clinician Programmer    6  Determine the implanted neurostimulator model  Check the current Patient ID card or  use the InterStim iCon Model 3037 Patient Programmer or N Vision Model 8840  Clinician Programmer    Note  Refer to the  InterStim iCon Model 3037 Patient Programmer instructions  on  page 11 or the appropriate programmer manual for additional instructions for use    7  Review Table 1 and use the implanted neurostimulator model to determine MRI head  scan eligibility  Eligibility for an MRI head scan is limited to specific neurostimulator  models and serial number series     Table 1  MRI head scan eligibility by neurostimulator model             Model Description Eligibility details   3058 InterStim II All Model 3058 Neurostimulators  serial numbers be   ginning with NJY  are eligible for head coil only MRI  scans    3023 InterStim Chec
8. ate Information for Prescribers booklet for contraindications   warnings  precautions  adverse events summary  individualization of treatment  patient  selection  use in specific populations  resterilization  and component disposal    Refer to System Eligibility  Battery Longevity  Specifications reference manual for  neurostimulator selection  battery longevity calculations and specific neurostimulator  specifications           2012 02 MRI Guidelines for InterStim Therapy neurostimulation systems English 3    4 English MRI Guidelines for InterStim Therapy neurostimulation systems 2012 02    Description    MR scanning conditions    MR Conditional  Non clinical testing has demonstrated that InterStim Therapy  systems have been found to be MR Conditional  If this patient is implanted with an InterStim  Il Model 3058 Neurostimulator or an eligible serial number of an InterStim Model 3023  Neurostimulator  when implanted as a system including a neurostimulator  lead  and  extension as applicable   MRI examinations of the head only may be safely performed  under the following conditions      1 5 Tesla  T  horizontal closed bore   a Maximum spatial gradient of 19 T m  1900 gauss cm      RF transmit receive head coil only  no RF transmit body coil      Gradient slew rate limited to 200 T m s     Normal operating mode  Scanning frequency of approximately 64 MHz only      If possible  do not sedate the patient     Model 3058 and eligible Model 3023 Neurostimulators  Turn the neurostim
9. conducting  an MRI head scan  Scans conducted using other modes  eg  First level controlled operating  mode  may increase the risk of shocking sensations uncomfortable stimulation or unusual  sensations during MRI scans    Gradient limitation     Limit the gradient slew rate to 200 T m s or less  Exposure to gradient  systems with a gradient slew rate exceeding 200 T m s may result in overstimulation or  shocking  particularly for unipolar capable devices    Scanning frequency of approximately 64 MHz only     Do not conduct MRI scans with  nonproton scanning frequencies  such as  13C  23Na  or 31P   Frequencies other than 64  MHz have not been tested    RF transmit receive head coil only     Use a transmit receive head coil to limit RF exposure  to the head only and ensure that the neurostimulator is well outside of the RF field  Do not  use a receive only head coil because it increases the risk of heating and potential nerve  and tissue damage during MRI scans    Neurostimulator settings during the MRI examination     Before conducting an MRI scan   turn the neurostimulator off  For the Model 3023 Neurostimulator  disable the Magnet    2012 02 MRI Guidelines for InterStim Therapy neurostimulation systems English 7    Switch  Failure to use these settings could result in uncomfortable  unintended stimulation  during the MRI examination  If the neurostimulator is not turned off or cannot be confirmed  to be off  the scan should not be performed unless it is a medical emergency
10. dtronic  Inc    710 Medtronic Parkway  Minneapolis  MN 55432 5604  USA   www medtronic com   Tel  1 763 505 5000   Fax 1 763 505 1000       EC   REP             Authorized Representative  in the European Community  Medtronic B V    Earl Bakkenstraat 10   6422 PJ Heerlen   The Netherlands   Tel  31 45 566 8000   Fax 31 45 566 8668    Europe Africa Middle East  Headquarters   Medtronic International Trading S  rl  Route du Molliau 31   Case Postale 84   CH 1131 Tolochenaz   Switzerland   www medtronic eu   Tel  41 21 802 7000   Fax 41 21 802 7900    Asia Pacific   Medtronic International Ltd    Suite 1106 11  11 F  Tower 1  The Gateway  25 Canton Road  Tsimshatsui   Kowloon   Hong Kong   Tel  852 2919 1362   Fax 852 2907 3998       Contacts for specific countries are listed inside this cover     9 4 002     M947945A    All Rights Reserved  M947945A002                                                                                 Medtronic  Inc  2012    0         
11. higher SAR and gradient levels are permitted     10 English MRI Guidelines for InterStim Therapy neurostimulation systems 2012 02      Monitor the patient both visually and audibly  Check the patient between each imaging  sequence  Discontinue the MRI examination immediately if the patient reports any  problems or has any discomfort      During the MRI head scan  the patient may feel tugging and or vibration of the  neurostimulator  If the tugging or vibration causes the patient considerable discomfort   stop the MRI head scan     After an MRI head scan     Verify that the patient has not experienced adverse effects as a result of the MRI head  scan  Contact Medtronic to report any adverse effects      Restore the neurostimulator settings to pre MRI head scan values      Instruct the patient to see the implanting physician or managing physician       if the patient has any questions about neurostimulator function      if assistance is required to return program parameters to pre MRI head scan settings      if the patient programmer displays a power on reset  POR  screen    Patient programmer instructions for MRI technicians  These instructions are for the use of the InterStim iCon Model 3037 Patient Programmer  at the patient s MRI appointment  For additional instructions  refer to the Model 3037 Patient  Programmer user manual     InterStim iCon Model 3037 Patient Programmer instructions       Decrease Increase       Neurostimulator on  Neurostimulator off    Sync    Navigat
12. ithin the RF transmit receive head  coil  Excessive heating can cause tissue damage and result in serious or  permanent patient injury     Ifthe system is removed  remove all portions of the neurostimulation system to  allow for future MRI scans  Even partial systems can have MRI interactions such  as RF heating  Excessive heating can cause tissue damage and result in serious  or permanent patient injury   Risks associated with MRI examination     Exposing a patient with an implanted  neurostimulation system or component to MRI settings other than those listed in this manual  may potentially injure the patient or damage the neurostimulator  The known potential risks  are as follows     Induced electrical currents from the MRI to the neurostimulation system or  component may cause heating  especially at the lead electrode site  resulting in  tissue damage   Note  This warning applies even if only a lead or extension is implanted   Factors that increase the risks of heating and tissue damage include  but are not limited  to  the following       Higher MRI Specific Absorption Rate  SAR  RF power levels       MRI RF transmit receive coil that is near or extends over the implanted lead  including  abandoned leads or portions of leads     6 English MRI Guidelines for InterStim Therapy neurostimulation systems 2012 02      Induced electrical currents may also stimulate or shock the patient  depending on  the potential circuit paths  the exposure to the pulsed gradient magnetic field
13. k the serial number  No MRI scans if the serial  number is        Less than NBV132955H    Between NBV133037H and NBV133063H    Between NBV628045S and NBV628263S    7427T InterStim Twin No MRI scans     a If a programmer is used to check the neurostimulator serial number  the letter suffix  H or S  may not  appear     8  Using the eligible neurostimulator model  review Table 2 and use the appropriate  programmer to change or verify the program settings to comply with neurostimulator  settings for an MRI head scan     2012 02 MRI Guidelines for InterStim Therapy neurostimulation systems English 9    Note  Before changing a neurostimulator program setting  record the current setting  so that the neurostimulator can be programmed to pre MRI program settings following  the MRI examination     Table 2  Recommended neurostimulator settings for an MRI head scan             Model Programmer Settings  3058 Model 3037 Patient Programmer Turn the neurostimulator off   3023 Model 8840 Clinician Programmer Turn the neurostimulator off and disable    the magnet switch         a For the Model 3023 Neurostimulator  the Model 8840 Clinician Programmer is required to disable the  magnet switch  If you have verified that the magnet switch has been disabled  the Model 3037 Patient  Programmer may be used to turn the neurostimulator off     Note  For an emergency MRI head scan on a patient implanted with the Model  3023 Neurostimulator  If an emergency MRI head scan is required and the Model  8840
14. lect a program   3  Synchronize the programmer and neurostimulator   4  Verify that the program is active M           Figure 4  Programmer screen showing an active program     Increasing or decreasing amplitude  1  Synchronize the programmer and neurostimulator   2  While holding the programmer over the neurostimulator      Press the Decrease     key to decrease amplitude       Press the Neurostimulator on    key and then press the Increase key to  increase amplitude     all Amplitude       Figure 5  Programmer screen showing an amplitude of 2 0 V     2012 02 MRI Guidelines for InterStim Therapy neurostimulation systems English 13    Contacts   Asia   Medtronic International Ltd     Tel  02919 1362  Fax 02907 3998    Medtronic Asia Ltd   Tel   02  548 1148  Fax  02  518 4786    Australia     Medtronic Australasia Pty  Ltd     97 Waterloo Road   North Ryde  NSW 2113  Australia   Tel   61 2 9857 9000  Fax  61 2 9878 5100  Toll free 1 800 668 6700    Austria    Medtronic Osterreich GmbH  Tel  01 240440   Fax 01 24044 100    Belgium    Medtronic Belgium S A   Tel  02 456 0900   Fax 02 460 2667    Canada    Medtronic of Canada Ltd   Tel   1 905  460 3800   Fax  1905  826 6620    Czech Republic   Medtronic Czechia s r o   Tel  2 965 795 80  Fax 2 965 795 89    Denmark    Medtronic Danmark A S  Tel  45 32 48 18 00   Fax 45 32 48 18 01  Finland    Medtronic Finland Oy LTD  Tel   09  755 2500   Fax  09  755 25018    France    Medtronic France S A S   Tel  01 5538 1700   Fax 01 5538 18
15. or    Figure 1  Model 3037 Patient Programmer keys     2012 02 MRI Guidelines for InterStim Therapy neurostimulation systems English 11    Synchronizing the programmer and neurostimulator   Hold the programmer over the neurostimulator and press the Sync    key    Viewing the neurostimulator model and serial number   To view the neurostimulator model  IM  and serial number  IS  information    screens   1  Synchronize the programmer and neurostimulator     2  If using the patient s Model 3037 Patient Programmer  press the Up arrow on the  Navigator key   3  Press the Left arrow on the Navigator key once to select information screens     4  Press the Down arrow on the Navigator key     5  Press the Left or Right arrows on the Navigator key to scroll through each information  screen until the neurostimulator model  IM  or serial number  IS  screens appear   Figure 2         Figure 2  Neurostimulator model and serial number information screens     Note  Refer to Table 1 for information about using the neurostimulator model and serial  number when determining MRI head scan eligibility     Turning the neurostimulator on or off  To turn the neurostimulator on or off  hold the programmer over the neurostimulator and  press the Neurostimulator on    or off    key  Figure 3         Figure 3  Neurostimulator on and off keys     12 English MRI Guidelines for InterStim Therapy neurostimulation systems 2012 02    Changing a program  41  Synchronize the programmer and neurostimulator   2  Se
16. s  and  the pulsed RF field      The static magnetic field  pulsed gradient magnetic field  and or the pulsed RF field  generated by MRI may permanently damage the neurostimulator  requiring explant  or replacement      MRI may affect the operation of the neurostimulator  MRI may also reset the  parameters to power on reset  POR  settings  requiring reprogramming with the  clinician programmer      The neurostimulator may move within the implant pocket and align itself with the  MRI field  which may cause patient discomfort     Precautions   Obtain the latest MRI guidelines     Always obtain the latest MRI guidelines  Refer to the  contact information on the last page of this manual  or go to www medtronic com mri   Enter the neurostimulator model on the website  www medtronic com mri  to obtain the  latest version of the manual containing these guidelines  Copies of these MRI guidelines  may not be the most up to date version if not received directly from the website or in another  manner from Medtronic at the time of the patient   s MRI appointment     N External devices are MR Unsafe in the MRI environment     Do not bring any of  the following Medtronic external control devices into the MRI scanner room      Patient programmer     Clinician programmer     Control magnet    These devices contain ferromagnetic material  which can be affected by the MRI magnet  and are unsafe in the MRI environment    Normal operating mode  MRI scans      Use only Normal operating mode when 
17. ulator   off     Eligible Model 3023 Neurostimulators only  Disable the magnet switch  Scanning under different conditions may result in severe patient injury or device  malfunction     MRI and InterStim Therapy   It is important to read this information in its entirety before conducting a magnetic resonance  imaging  MRI  head scan on a patient with any implanted component of a Medtronic  InterStim Therapy neurostimulation system  These instructions do not apply to other  implantable products  or other devices  products  or items  Contact Medtronic at the  appropriate address or phone number listed at the back of this manual if you have any  questions    MRI systems generate powerful electromagnetic fields that can produce a number of  interactions with implanted components of the neurostimulation system  Some of these  interactions  especially heating  are potentially hazardous and can lead to serious or  permanent patient injury  The following information describes the potential interactions and  control measures that should be taken to minimize the risks from these interactions     Information for prescribers    Warnings   MRI during test stimulation     Explant all test stimulation components if an MRI scan is  required  Physicians should not prescribe MRI for patients undergoing test stimulation or  who have any neurostimulation system components that are not fully implanted  MRI has  not been evaluated with test stimulation components  The external neurostimulator    2
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
990 - Austromex  User`s manual MEMS OCT PROBE  TruPortal™ Release Notes  GAMME D`OUTILS RAMSET T3  MANUEL DE L`UTILISATION  使用上のご注意  Festschrift Bonin - Leuphana Universität Lüneburg  be.ez LA robe One Blue  Mode d`eMploi  Terminale-S_files/DS3 TS2 2012 Titrages et Newton    Copyright © All rights reserved. 