Home
Laboratory Centrifuge Model GF614B User Manual
Contents
1. a Bio Seal should be employed contact your Graham Field authorized distributor to purchase this Bio Seal In the event that materials of a higher risk group are being used more than one level of protection must be provided The use of flammable or explosive materials as well as those materials which have a vigorous chemical reaction is prohibited A WARNING For your safety and the durability of your machine never transport or store centrifuge with tube holders inside machine A WARNING If components are damaged or missing contact your Graham Field authorized distributor immediately DO NOT use substitute parts A WARNING GF Health Products Inc assumes no responsibility for any damage or injury caused by improper installation or use of this product INTENDED USE The Grafco Model GF614B is a continuous duty centrifuge designed for the small lab or doctor s office for the purpose of separating laboratory fluid any other use is explicitly prohibited GF614B Centrifuge 4 SUPPLIED EQUIPMENT The following items come standard with each Model GF614B centrifuge Equipment Quantity Description supplied with Lo Rotor six place fixed angle 6B Tube holder 75 mm ES Tube holder 100 mm fF 6 o Tube holder 125 mm OPTIONAL ACCESSORIES The following items are optional accessories for the Model GF614B centrifuge Please contact your Graham Field authorized distributor to purchase these accessories Optional tem
2. it should slide freely 3 Once a proper fit has been achieved replace the rotor knob or nut and turn it until it is hand tight 4 Replace the tube holders and ensure that they are seated properly 5 Perform the SETUP Initial setup procedure to ensure that the rotor has been installed correctly and that the centrifuge has not been damaged Tube Holder Replacement Replace the tube holders after 24 months of use or if damaged Calibration to be performed only by a qualified service technician For continued proper operation of the centrifuge test every two years to ensure that the top speed is within specification for top speed see SPECIFICATIONS section Testing ground continuity and leakage current to be performed only by a qualified service technician A WARNING To reduce the risk of electric shock test the ground continuity and line leakage every two years GF614B Centrifuge 12 TROUBLESHOOTING Problem Solution SSCS The rotor does not spin freely Excessive noise when the device is running The centrifuge does not run GF614B Centrifuge Ensure that nothing has fallen into the rotor chamber If there is nothing obstructing the rotor contact your Graham Field authorized distributor for further assistance Ensure that the load is balanced Ensure that nothing has fallen into the rotor chamber Ensure that the nut in the center of the rotor is tight Possible faulty motor contact your Graham Fie
3. Number Description OO O accessories GF614B 75 75 mm tube holder package of 6 GF614B 100 100 mm tube holder package of 6 GF614B 125 125 mm tube holder package of 6 SETUP Unpacking 1 Check for obvious damage to the carton or its contents If damage is evident please notify the carrier and GF Health Products Inc 2 Unpack the centrifuge and verify that all of the supplied equipment is present see Equipment supplied with GF614B list above GF614B Centrifuge 5 Main features Picture of GF614B with main features labeled follows circuit breaker in back of unit g 7 power cord and plug N in back of unit a rotor a beneath lid tube holders for 75 mm tubes raf lid latch GF614B 75 _ lp knob tube holders timer for 100 mm bee nn j GF614B 100 tube holders ay for 125 mm tubes GF614B 125 GF614B Laboratory Centrifuge Fixed angle rotor for spinning 75 mm 100 mm and 125 mm test tubes in specially designed tube holders Cool Flow air flow design that prevents overheating of samples Heavy gauge durable steel construction Lid safety switch that prevents the centrifuge from operating unless the lid is closed and latched Removable rotor for easy cleaning Brushless A C motor Transparent lid for observation of samples and optical speed calibration Setup location Choose a setup location which meets the following criteria a A bench top clearance height of 20 is required i
4. e seen eneeseeeeeneesseeeseeseaeesneevsaeesseeseneevenessneeseesesneneaneas 10 PROVIDE ADEQUATE VENTILATION c cccssssssssesseeeeseeeseeesseeneneeeseenes 10 ALWAYS SPIN BALANCED LOADS ccec ccs cece eeneeeneeveneeesteneneneeenee 10 KEEP THE TUBE HOLDERS CLEAN insnrrrrrrsnnessnensne 10 KEEP THE ROTOR CHAMBER CLEAN cccesscssessecsessteeeseeeseeeeseeseeeees 11 TUBE HOLDER REPLACEMENT cccscesseesseseseesseesseeeveneeseneseeeneneeeseess 12 CALIBRATION 00 cccecccteeseneeeseeseneeeseeseeeessneseeeesneeveneesseeseeeveaessnneesesesenenes 12 TESTING GROUND CONTINUITY AND LEAKAGE CURRENT 12 TROUBLESHOOTING ti sscscucescceseatavauracstwassiawacisetacnauvenacnsncvesuccuesaivenedaseivasconbeanaten 13 SPECIFICATIONS rt tas sd near ise tea da re een EAEan KaKa EE KKK 14 WARRANTY nn anaua Eana ss uruauuanasvasuasancscanatuesiontinesiadendavases 15 GF Health Products Inc is not responsible for typographical errors All illustrations specifications packaging and warranties contained in this literature are based on the latest product information available at the time of printing The most current product information including the most current version of these instructions can be found online at www grahamfield com Graham Field and Grafco are registered trademarks of GF Health Products Inc GF614B Centrifuge 2 IMPORTANT SAFETY GUIDELINES PLEASE READ BEFORE USE The safety statements presented in this
5. Graico A Graham Field Brand Laboratory Centrifuge Model GF614B User Manual Read this manual before operating the GF614B Laboratory Centrifuge Save this manual for future reference GF614B INS LAB RevB13 Info The Grafco GF614B Laboratory Centrifuge complies with all requirements of UL standard 61010A 1 61010A 2 20 Can CSA C22 2 No s 1010 1 1010 2 20 CONTENTS IMPORTANT SAFETY GUIDELINES PLEASE READ BEFORE USE 3 INTENDED USE enr rnnnesnnnsnneensnnesnneesnee sen eesnne seen snrennenne 4 SUPPLIED EQUIPMENT cccccceceeesseseeeesseeveneesseeveseessaeseneessneesseessonewsesennenaes 5 OPTIONAL ACCESSORIES cccceecccceseneeeceveneeseeeveneesenessenessneessnessonessnesseneese 5 SETUP E EA E AE A A E T A A AEA 5 UNPACKING wiccsscessaxcbaavceusrusdaissavenssssiuecsuainetsnssdaventebasiqlaucalasasddsausesanazexivaauan sne 5 MAIN FEATURES nnnsansnnnnnennsennsnnrenrrnnrnnrnnnnunrnnrnnnnnnnnnnennnnnnnnnennnnnnnnn ennnen ennenen 6 SETUP LOCATION iscscdisscrctesvesdpsucnssvevaassestatanedssucabanidaessubasesnatveasdaviewan cauansaet 6 INITIAL SETUP Se 7 BALANCED LOADS wisiccsscicssitrsctessacess cascesesedaasavasedsatetasisdaaantdanssenacasiacseaueneate 8 SPECIAL FEATURES 0 cccccescccsecseeseneeseneeeceeveneeeneessaessneesseeseseesseveneenseenees 8 OPERATION cccccececsesseeseceeeseeeeeeeseeesseesseesseeseseeseeesseessneeseeevegeevsaessneeveeeenneneas 9 MAINTENANCE Loo cece ccc c ete
6. bient air through the air intake cover on the top of the lid and exhausts this air in the rear of the base ALWAYS Spin Balanced Loads Ensure that you always spin a balanced load The Model GF614B has a unique counter balanced motor mount design which along with its rubber suction feet produces excellent vibration dampening However out of balance loads may break glass test tubes and or produce unsatisfactory separation results Proper load balancing will help to improve sample separation and extend the life of the centrifuge Refer to earlier SETUP Balanced loads section for additional information on balancing the load Keep the Tube Holders Clean A WARNING Always follow the safety guidelines of materials in the event that a substance known to be potentially toxic radioactive or contaminated with a pathogenetic microorganism is spilled in or on the centrifuge Small glass fragments left in the tube holder after tube breakage may adhere to the next test tube inserted in that holder When this tube is handled these fragments may puncture protective gloves and lacerate the operator s fingers or hand Remaining fragments may provide stress points on subsequent tubes and result in additional breakage GF614B Centrifuge 10 If tube breakage occurs carefully remove the tube holder Properly dispose of the sample and tube fragments and thoroughly clean both the inside and outside of the tube holder Replace the tube holder in the ro
7. chapter refer to the basic safety information that the operator of the centrifuge shall pay attention to and abide by There are additional safety statements in other chapters or sections which may be the same as or similar to the following or specific to the operations Z WARNING Indicates a potential hazard situation or unsafe practice that if not avoided could result in death or serious personal injury A CAUTION Indicates a potential hazard or unsafe practice that if not avoided could result in minor or moderate personal injury A NOTICE Indicates a potential hazard or unsafe practice that if not avoided could result in product property damage Info Provides application recommendations or other useful information to ensure that you get the most from your product A WARNING Important Read and understand these instructions before installing or using the Grafco GF614B Laboratory Centrifuge If you do not understand any part of these instructions contact your medical professional or Graham Field authorized distributor for direction in the use of this product GF614B Centrifuge 3 A WARNING For the safety of both the operator and service personnel take care when handling substances that are known to be toxic radioactive or contaminated with pathogenic microorganisms when using this centrifuge When Risk Group Il materials are used as identified in the World Health Organization Laboratory Bio Safety Manual
8. f an overload preventing the device from operating GF614B Centrifuge OPERATION Info Follow the previous Initial setup procedure before initial operation l Ensure that the centrifuge s power cord is plugged into a properly grounded 115 Volt AC 60 Hz electrical wall outlet Turn the lid latch knob counter clockwise and open the lid Place the test tube samples into the tube holders Be sure to follow the rules for balanced loads Close the lid and turn the lid knob clockwise to its complete stop position Turn on the device by turning the timer to the desired run time The centrifuge should begin to spin Once the timer reaches zero 0 the motor will stop and the rotor will coast to a stop Do not open the lid until the rotor has come to a complete stop Turn the lid latch knob counter clockwise and open the lid 10 Remove the samples 11 The centrifuge is immediately ready for operation GF614B Centrifuge MAINTENANCE With proper care and maintenance your centrifuge will provide years of laboratory service For proper care always follow the maintenance instructions in this section A WARNING Electrical shock hazard Do not open the case Any disassembly of the device must be performed by a qualified service technician Provide Adequate Ventilation Place the centrifuge on a hard smooth surface for good air circulation For cooling purposes the centrifuge draws in am
9. ing the timer to 10 minutes 6 Listen to the centrifuge You should hear a smooth whirring sound Info After the centrifuge has passed this procedure it is ready for operation A WARNING Ensure that Centrifuge is installed as described before use GF614B Centrifuge 7 Balanced Loads A NOTICE Your centrifuge must contain a balanced load in order to work properly Spinning unbalanced loads may shorten the life of the device and produce unpredictable results Observe the following rules when loading the rotor 1 Opposing tube holders must be identical 2 Opposing tube holders must be empty or loaded with equally weighted samples 3 If an odd number of samples is to be spun fill a tube with water to match the weight of the unpaired sample and place it across from the unpaired sample Special features Lid Safety Switch The lid is secured to the top of the cabinet by a latching knob and pawl system When the knob is rotated clockwise the pawl grips the underside of the cabinet opening and prevents the lid from opening A mechanical stop positions the pawl and prevents it from rotating completely When rotated to the stop position the pawl makes contact with a microswitch mounted underneath the cabinet top The lid safety switch prevents the centrifuge from operating while the lid is open Circuit Breaker The Model GF614B has a 4 Amp circuit breaker at the rear of the base that disconnects power in the event o
10. ld authorized distributor for further assistance Ensure that the centrifuge is plugged into a properly grounded 115 Volt AC 60 Hz electrical wall outlet Ensure that the lid latch knob is turned completely clockwise to its stop position If not the centrifuge will not operate Ensure that the circuit breaker at the rear of the base that disconnects power in the event of an overload and prevents the device from operating is not tripped If the switch is white the breaker has tripped Contact your Graham Field authorized distributor for further assistance 13 SPECIFICATIONS General Specifications for the Grafco Model GF614B Centrifuge Rotor Rotor Accessories Rated for a rotation frequency of 3 500 RPM 17 mm x 125 mm Maximum capacity 90 ml 6 x 15 ml Maximum sample density is 1 15 grams ml water density 1 0 grams ml Overall Dimensions H x W x D 8 75 x 11 75 x 14 222 3 mm x 298 5 mm x 355 6 mm Centrifuge Motor 1 30 HP A C Protection Breaker 4 Amp re settable Timer mechanical 1 to 30 minutes accuracy 10 Voltage Requirement 115 10 Volts frequency i GF614B Centrifuge 14 WARRANTY GF Health Products Inc warrants the Grafco Laboratory Centrifuge Model GF614B for a period of one year for defects in workmanship and materials If within the warranty period the product or component part is proven to GF Health Products Inc s satisfaction to be defective GF Health Products Inc
11. n order to open the lid b A clearance envelope the space around the centrifuge required for safety of at least 24 x 24 is required with the centrifuge at the center No person or hazardous material shall be permitted in the clearance envelope during operation The operator time within the envelope shall be limited to the time necessary for loading unloading and centrifuge operation only GF614B Centrifuge 6 c Unencumbered air flow is required Proper ventilation is necessary to prevent the overheating of samples as well as premature failure of the centrifuge d A flat level surface is required on which to place the centrifuge Secure the centrifuge to the operating surface by its four suction feet No level adjustment is necessary e The power outlet must always be within reach as the power cord is the means of emergency disconnection Initial Setup Info If any problems are found during the initial setup procedure refer to the Troubleshooting section If you are still unable to solve your problem contact your Graham Field authorized distributor for further assistance 1 Plug the centrifuge s power cord into a properly grounded 115 Volt AC 60 Hz electrical wall outlet 2 Turn the lid latch knob counter clockwise and open the lid 3 Spin the rotor by hand check for free and level rotation 4 Close the lid Rotate the lid latch knob clockwise to its complete stop position 5 Turn the centrifuge on by turn
12. shall provide at its option one of the following 1 repair or replacement of any defective or nonconforming part or product or 2 a credit and or refund of the original selling price This warranty does not include any labor charges incurred in replacement part s installation or any associated freight or shipping charges to the manufacturer The warranties contained herein contain all the representations and warranties with respect to the subject matter of this document and supersede all prior negotiations agreements and understandings with respect thereto The recipient of this document hereby acknowledges and represents that it has not relied on any representation assertion guarantee warranty collateral contract or other assurance except those set out in this document GF614B Centrifuge 15 GF Health Products Inc Graham Field 2935 Northeast Parkway Atlanta Georgia 30360 telephone 770 368 4700 fax 770 368 2386 GRAHAM FIELD www grahamfield com 2013 GF Health Products Inc
13. tor Motor and Electrical Maintenance The Model GF614B uses a brushless A C motor The motor and electrical components should not need servicing for the life of the centrifuge Keep the Rotor Chamber Clean A NOTICE Do not submerge the centrifuge in water or other cleaning solutions as this will cause damage and void your warranty Clean the rotor chamber every six months or whenever there is a tube breakage as described above Instructions follow to remove the rotor clean and re install the rotor To remove the rotor A WARNING Unplug the centrifuge from the electrical outlet at this time to reduce the risk of electric shock 1 Open the lid 2 Remove the test tube holders 3 Remove the knob or nut in the center of the rotor by turning it counter clockwise A nut driver may be required 4 The rotor sits on a cone shaped adapter Pull the rotor up and off of this adapter To clean Clean the rotor chamber rotor and accessories thoroughly using either isopropyl alcohol soap and water or bleach with a clean towel or cloth A NOTICE Do not use fully or partially halogenated hydrocarbons ketones esters or any other chemicals not listed in the previous paragraph they could damage the rotor and tube holders GF614B Centrifuge 11 To re install the rotor 1 Place the rotor back onto the cone shaped adapter You may need to turn the rotor slightly to align it properly 2 Slide the rotor onto the rotor cone
Download Pdf Manuals
Related Search
Related Contents
109 Farmacêutico 5180 Owner`s Manual AMX UDM-0404 User's Manual HUMATROPEN® 6 mg - Eli Lilly Canada User Manual - Oregon Scientific CFW700 Variable Speed Drives Slant/Fin INTREPID Oil-Fired Boilers User's Manual sfra45 user manual - Caltest Instruments 取扱説明書 Copyright © All rights reserved.
Failed to retrieve file