Home

MiniBump - Advanced Calibration Designs, Inc.

image

Contents

1. addition a 3 Norprene hose with a quick connect is available as an optional item The MiniBump can also be removed from the Bump Station and used as a portable gas generator Mounting the Bump Station Housing To mount the case remove the four mounting screws from inside the unit Drill four mounting holes in the wall as per the location of the mounting brackets and mount the case with the gas outlet pointing toward the ground The front of the case has been labeled This Side Down Mounting in this configuration helps prevent dust dirt and water from entering the instrument through the gas outlet port Removing or Inserting the MiniBump Unit into the Wall Mount Housing If the MiniBump unit is ordered with the Bump Station the instrument is shipped inside the wall mount housing The instrument may also be used as a portable gas generator or it may need to be removed from the housing to change the gas cell or the batteries To remove the MiniBump follow these steps 1 Unscrew the 4 Phillips screws that hold the front of the housing to the back Note that the screws do not come completely out of the front cover Remove the front of the housing from the back Remove the one Phillips screw that holds the clear plastic retaining plate located over the instrument Pull out the clear retaining plate by lifting up and out Remove the instrument from the housing To insert the MiniBump back into the housing follow the directions i
2. battery or exhausted cell condition occurs The circuit board is attached directly to the bottom case via solder joints to the battery terminals If damaged or inoperable it should be replaced as part of the case rather than removed from these terminals Delivery cup The instrument comes standard with a small flexible rubber cup designed for delivering the gas to the sensor head The distance between the generator and the sensing electrodes should be kept to a minimum for fast response Page 3 ll START UP To start the generator press and hold the ON OFF Switch located in the middle front of the instrument until the GREEN LED turns on approximately 0 3 seconds Release the switch immediately thereafter Please note that the label on the front of the MiniBump instrument says to press and hold for 3 seconds This is no longer necessary but holding the button down for 3 seconds will still turn on the unit with no adverse effects The GREEN LED will remain lit for approximately 120 seconds Gas will begin to be generated immediately and should reach the sensor within a few seconds Gas will continue to be generated for 30 seconds followed by a 90 second purge of air through the sensor Note Some sensors have faster response times than others If the system does not respond to the gas within 30 seconds continue to hold the MiniBump test cap on the sensor as some gas is still being released during the purge cycle If the RED LED begin
3. RED LED will light and the instrument will continue to run for approximately 30 seconds before power down If the cell has been depleted it should be removed and replaced with a new generating cell Please refer to the section on cell life for proper instructions on replacing the generating cells If the cell has not been depleted there may be a faulty electrical connection between the cell and the instrument Remove the cell and inspect the pins protruding from the cell and the sockets in the cell chamber Remove any debris and or corrosion that may be present and retest the cell If either the pins or sockets become damaged or non repairable replace the cell chamber and or the cell Page 6 V Cell Life The generating cell is consumed during normal operation All cells are rated for up to 1000 test maximum 250 minimum Once a cell has been depleted it may be replaced with a new cell to provide another 1000 tests Replacement cells may be stored in their containers in a cool humid area for several years under most conditions Different types of generating cells can be used with each instrument The microprocessor in the MiniBump reads the type of gas of the cell from a memory chip embedded in the cell body It then adjusts itself for each type of gas output For a complete list of correct cell part numbers please refer to the parts list to follow To replace the generating cell in the instrument first remove the instrument cap locat
4. ZACD Advanced Calibration Designs Inc MiniBump Calibration Gas Instrument INSTRUC TION MANUAL 11O Press and Hold 3 Seconds Release Il www GoACD com WARNING This instrument generates test gas for toxic gas detectors The instruction manual should be read and understood prior to operation of the instrument Failure to operate the instrument correctly can lead to improper testing of your detection system This instrument conforms to the protection requirements of the EC DIRECTIVE 89 336 EEC on Electromagnetic Compatibility EMC in accordance with the provisions of Statutory Instrument 2372 The following standards have been applied EN 50081 1 Emissions Standard Residential Commercial and Light Industry EN 50082 1 Immunity Standard Residential Commercial and Light Industry Il III IV Vi VII VIII Instruction Manual Calibration Gas Instrument TABLE OF CONTENTS General Description Start Up Normal Operation Release Il Non Normal Operation Cell Life Maintenance Accessory Items Parts List Optional Bump Station Accessory Warranty Information 910 0550 00 Page 2 Page 4 Page 5 Page 6 Page 7 Page 8 Page 9 Page 10 Page 11 l GENERAL DESCRIPTION The MiniBump is a small battery powered portable electrochemical gas generator designed to functionally test toxic gas sensors and alarm systems The MiniBump is designed to bump a gas detection system into alarm within
5. approximately 30 seconds The MiniBump uses the following components to produce the calibration gas air mixture Internal Micro Pump A small air pump draws in ambient air to blend with the electrochemically generated gas Electrochemical Generating Cell The electrochemical generating cell contains an electrolyte solution and either inert or consumable electrodes depending upon the gas being generated Gas is generated in the solution and readily permeates out of the cell and to the detection system 1 0 Press and Hold 3 Seconds G Page 2 Alkaline AA Batteries A set of two fully charged alkaline AA batteries provides approximately 10 12 hours of operation To replace the batteries open the battery cover with either a fingernail or small flat object It is important that when changing the batteries in the instrument to not use a metal object such as a screw driver Using a metal object to remove the batteries will frequently blow the pico fuse in the unit preventing the unit from powering up NOTE Rechargeable baiteries will provide substantially fewer hours of operation and will self discharge while the instrument is in storage Users should also be aware that use of rechargeable batteries will void the CSA approval of the MiniBump for intrinsic safety Microprocessor Based Circuitry When powered up the microprocessor checks for remaining cell life It also monitors battery life and indicates when either a low
6. d in lieu of any and all other warranties and representations expressed or implied and all other obligations or liabilities on the part of Advanced Calibration Designs Inc including but not limited to the warranty of fitness for a particular purpose In no event shall Advanced Calibration Designs Inc be liable for direct incidental or consequential loss or damage of any kind connected with the use of its products or failure of its products to function or operate properly The following is a listing of the available electrochemical MiniCal cells and their standard warranty when installed in equipment manufactured and supplied by Advanced Calibration Designs Inc Chlorine Hydrogen Hydrogen Sulfide Hydrogen Cyanide One year or 250 tests Page 12 Manufactured by Advanced Calibration Designs Inc 2024 W McMillan Street Tucson Arizona 85705 USA Phone 520 290 2855 Fax 520 290 2860 Int l Phone 001 520 290 2855 Fax 001 520 290 2860 web site www GoACD com Page 13 ZACD Advanced Calibration Designs Inc 2024 W McMillan St Tucson Arizona 85705 USA Telephone 520 290 2855 Fax 520 290 2860 Email acd goACD com www goACD com Revision 11 20 13
7. ed on the top of the instrument The generating cell pulls directly up out of the cell chamber Do not twist or rotate the generating cell while removing it from the cell chamber or damage may occur to either the cell pins or the cell chamber sockets The replacement cell is replaced in reverse fashion Care must be taken when replacing the cell to correctly line up the electrode pins with the corresponding sockets Do not force the cell into the cell chamber if it is not properly aligned otherwise the pins may become damaged rendering the cell and or instrument useless NOTE Never attach a sample draw instrument directly to the end of the MiniBump delivery cup with no Y or T connector in between the two instruments This may cause damage to both the MiniBump and or sample draw instrument Page 7 VI MAINTENANCE ON OFF Switch The ON OFF Switch is identified by the universal symbols of 1 and 0 and is activated through the front membrane panel Itis a physical switch mounted on the circuit board If the membrane area gets damaged it is replaced as part of the entire front label If the switch becomes nonfunctional it must be repaired or replaced as part of the circuit board Caution Both the electrochemical cell and the alkaline batteries used in the MiniBump instrument contain corrosive chemicals While it is not expected that the chemicals in the cell or alkaline batteries will leak during normal operation it is recommend
8. ed that both the cell and the batteries be removed from the instrument if it is being stored for periods longer than one week between use Note The generating cells contain very small amounts of acidic solution and should be disposed of as per local or federal regulations For more information please contact the factory Page 8 ACCESSORY ITEMS PARTS LIST P N 115 0400 00 210 0400 00 250 0401 00 400 0500 00 510 0500 01 510 0550 02 510 0570 02 510 0590 02 715 0100 00 715 0403 0X 730 0400 00 730 0415 00 910 0550 00 The following items and spare parts are available for the MiniBump Description Screw Cap with hose fitting Wall mount case for MiniBump Battery Cover MiniGen MiniBump main board Chlorine generation cell Hydrogen sulfide generation cell Hydrogen cyanide generation cell Hydrogen generation cell Rubber cal cup adapter Hose with connector Norprene 3 Nylon carrying case Hard body water resistant padded accessory case Instruction manual MiniBump Page 9 Vill OPTIONAL BUMP STATION ACCESSORY The MiniBump can also be used in the optional wall mount Bump Station This weather and corrosion resistant housing is designed to place the MiniBump unit in an easily accessible area for frequent functional testing of a variety of gas detection units When the MiniBump unit is ordered with the Bump Station the instrument is shipped with the rubber cal cup adapter installed on the instrument In
9. f sample draw instrument is high the MiniBump may need to operate several times to provide a sufficient gas sample Be sure to disconnect the sample draw unit before the bag is completely emptied of gas NOTE Never attach a sample draw instrument directly to the end of the MiniBump delivery cup with no Y or T connector in between the two instruments This may cause damage to both the MiniBump and or sample draw instrument Page 5 IV Non Normal Operation If either the batteries or the generating cell become depleted or the generating cell is removed from the MiniBump the instrument enters a non normal operating condition and the RED LED will turn on Under a low battery condition the RED LED will begin to flash If the low battery condition occurs during power up the RED LED will flash for approximately 30 seconds and then the instrument will power down If the low battery condition occurs after start up the RED LED will begin flashing and the instrument will continue to run for approximately 30 seconds before power down It is recommended to replace depleted batteries with two AA heavy duty Alkaline batteries for maximum operating life If the cell becomes depleted or is removed the RED LED will light continuously If the depleted cell condition occurs during power up the RED LED will flash for approximately 30 seconds and then the instrument will power down If the depleted cell condition occurs after start up the
10. n reverse order Po oust Page 10 IX STANDARD WARRANTY We warrant gas calibration equipment manufactured and sold by us to be free from defects in materials workmanship and performance for a period of one year from date of shipment Any parts found defective within that period will be repaired or replaced at our option free of charge F O B factory This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal service and which must be cleaned repaired or replaced on a routine basis Such items may include a Electrochemical type generating cells b Batteries Warranty is voided by abuse including rough handling mechanical damage alteration or repair procedures not in accordance with the instruction manual This warranty indicates the full extent of our liability and we are not responsible for removal or replacement cost local repair costs transportation costs or contingent expenses incurred without our prior approval Page 11 Advanced Calibration Designs Inc s obligation under this warranty shall be limited to repairing or replacing and returning any product which shall be returned to Advanced Calibration Designs Inc at its manufacturing facilities with transportation charges prepaid and which Advanced Calibration Designs Inc s Material Review Board examination shall disclose to its satisfaction to have been defective This warranty is expresse
11. s flashing or remains lit continuously it indicates the instrument is in either low battery condition or cell failure mode Please refer to the section on Non Normal Operation The instrument will shut off automatically after the 120 seconds expires and the GREEN LED turns off The instrument may be turned on again immediately after the GREEN LED turns off Page 4 lll NORMAL OPERATION The instrument is designed to generate gas for approximately 120 seconds per bump test Each cell will provide up to 1000 tests The MiniBump will generate gas until either the battery becomes discharged or the cell is consumed If either condition occurs a RED LED will light indicating a non normal condition refer to the Non Normal Operation section that follows The calibration gas is delivered through a small flexible rubber cup attached to the cell manifold cap The rubber cup is placed over around or in front of the gas sensor Be sure to make a tight seal to the sensor face to prevent the generated gas from escaping prior to reacting on the sensor This method is primarily used for calibrating diffusion type sensors Alternatively the output gas may be collected into a gas collection bag for calibrating sample draw instruments If this method is used it is recommended that the generated gas is allowed to diffuse into a calibration bag for several minutes several consecutive tests to partially fill the bag with generated gas If the flow rate o

Download Pdf Manuals

image

Related Search

Related Contents

MODE D`EMPLOI 2014  Programming Features - Earth People Technology  Biostar TA760G-M2+ motherboard  RIVELA DT-E TX_RX _Hw10 Fw10_ manuale rev0 aggiornato  Brugermanual - WhiteAway.com  manual NEBNext DNA Library Prep Master Mix Set for Illumina  グリーン調達基準  Clamp Meter - PCE Instruments  1- Pair Port PoE (IEEE 802.3af) Mid-Span Power Supply  Port Port.fh11  

Copyright © All rights reserved.
Failed to retrieve file