1. Lexan 500R Red UL94V0 Dimensions 4 37 x 8 50 x 1 77 111 x 216 x 45 mm Weight 1 21 Ibs 550 g Output Two standard safety banana jacks 4 mm SAFETY SPECIFICATIONS CE Wu Electrical Double insulation or reinforced insulation between the primary or secondary and the outer case of the handle conforms to IEC 1010 2 032 Common Mode Voltage 600 V Category III Pollution Degree 2 Dielectric Strength 5550 V 50 60 Hz between primary secondary and the outer case of the handle Electromagnetic Compatibility EN 50081 1 Class B EN 50082 2 Electrostatic discharge IEC 1000 4 2 Radiated field IEC 1000 4 3 Fast transients IEC 1000 4 4 Magnetic field at 50 60 Hz IEC 1000 4 8 ORDERING INFORMATION AC Current Probe SR634 Cat 2113 48 Includes a user manual and a product warranty and registration card Accessories Leads set of 2 5 ft safety 1000V Cat 2111 29 Banana plug adapter to nonrecessed plug 65 Cat 1017 45 OPERATION Please make sure that you have already read and fully understand the WARNING section on page 1 Making Measurements with the AC Current Probe Model SR634 Connect the current probe to an AC current range on your DMM or other current measuring instrument The AC Current Probe SR634 is a multi ratio unit The outputs are as follows Current Flow Output of Probe in Conductor Ratio 250 5 Ratio 500 5 Ratio 1000 5 100A 2000 mA 1000 mA 500 mA 250A 5 A 500
2. Use extreme caution when clamping around bare conductors or bus bars INTERNATIONAL ELECTRICAL SYMBOLS o This symbol signifies that the current probe is protected by double or reinforced insulation Use only factory specified replacement parts when servicing the instrument A This symbol signifies CAUTION and requests that the user refer to the user manual before using the instrument This is a type A current sensor This symbol signifies that application around and removal from HAZARDOUS LIVE conductors is permitted DEFINITION OF MEASUREMENT CATEGORIES Cat l For measurements on circuits not directly connected to the AC supply wall outlet such as protected secondaries signal level and limited energy circuits Cat Il For measurements performed on circuits directly connected to the electrical distribution system Examples are measurements on household appliances or portable tools Cat Ill For measurements performed in the building installation at the distribution level such as on hardwired equipment in fixed installation and circuit breakers Cat IV For measurements performed at the primary electrical supply lt 1000V such as on primary overcurrent protection devices ripple control units or meters RECEIVING YOUR SHIPMENT Upon receiving your shipment make sure that the contents are consistent with the packing list Notify your distributor of any missing items If the equipment appears to be damaged file a claim immediate
3. 0 mA 2500 mA 1250 mA Select the 10 A range on your DMM or instrument Use the switch on the inside handle to select the 1000 5 range Clamp the probe around the conductor and work down until the appropriate range and resolution in reached Take the reading on the meter and multiply it by the ratio selected to obtain the measured current e g 500 mA reading 500 x 200 100 000 mA or 100 A with the 1000 5 ratio e For best accuracy avoid if possible the proximity of other conductors which may create noise Tips for Making Precise Measurements e When using a current probe with a meter it is important to select the range that provides the best resolution Failure to do this may result in measurement errors e Make sure that probe jaw mating surfaces are free of dust and contamination Contaminants cause air gaps between the jaws increasing the phase shift between primary and secondary It is very critical for power measurement MAINTENANCE Warning e For maintenance use only original replacement parts e To avoid electrical shock do not attempt to perform any servicing unless you are qualified to do so e To avoid electrical shock and or damage to the instrument do not get water or other foreign agents into the probe Cleaning To ensure optimum performance it is important to keep the probe jaw mating surfaces clean at all times Failure to do so may result in error in readings To clean the probe jaws use very fine sand paper fi
4. AC Current Probe Model SR634 User Manual DESCRIPTION The SR634 Catalog 2113 48 is designed for use in industrial environments The ergonomic design allows it to easily attach to cables or small bus bars The circular jaws guarantee a very good accuracy and low phase shift The probe has a measurement range up to 1000 Arms continuous and is compatible with any AC Ammeter multimeter or other current measurement instrument with an input impedance lower than 0 6Q To achieve the stated accuracy use the SR634 with an ammeter having an accuracy of 0 75 or better WARNING These safety warnings are provided to ensure the safety of personnel and proper operation of the instrument Read the instruction manual completely and follow all the safety information before attempting to use or service this instrument Use caution on any circuit Potentially high voltages and currents may be present and may pose a shock hazard Read the Safety Specifications section prior to using the current probe Never exceed the maximum voltage ratings given Safety is the responsibility of the operator ALWAYS connect the current probe to the display device before clamping the probe onto the sample being tested ALWAYS inspect the instrument probe probe cable and output terminals prior to use Replace any defective parts immediately NEVER use the current probe on electrical conductors rated above 600V in overvoltage category Ill CAT III
5. current carrying conductor tesi sample centered Load impedance 0 20 lead lt 40 mQ 0 10 for 250 A range Accuracy Per IEC 185 Frequency Range 30Hz 5kHz current derating above 1 kHz using the formula 41000A x 1 F in kHz Load Impedance 0 40 max Working Voltage 600V Cat III Common Mode Voltage 600V Cat III Open Secondary Voltage lt 25V by limiting circuit Influence of Adjacent Conductor lt 1mA A AC Influence of Conductor in Jaw Opening 0 02 of reading Influence of Frequency Range 250 5 From 65 to 100 Hz lt 1 of R From 100 to 1000 Hz lt 5 of R Range 500 5 From 65 to 100 Hz lt 1 of R From 100 to 1000 Hz lt 1 of R Range 1000 5 From 65 to 100 Hz lt 0 5 of R From 100 to 1000 Hz lt 1 of R MECHANICAL SPECIFICATIONS Operating Temperature 14 to 122 F 10 to 50 C Storage Temperature 4 to 158 F 20 to 70 C Influence of Temperature lt 0 13 per 10 K Influence of Humidity Range 250 5 lt 0 6 amp lt 2 Range 500 5 lt 0 4 amp lt 0 6 Range 1000 5 lt 0 2 amp lt 0 2 Jaw Opening 2 25 57 mm max Maximum Conductor Size 2 05 52 mm Envelope Protection IP 40 IEC 529 Drop Test 1 m IEC 68 2 32 Mechanical Shock 100 g IEC 68 2 27 Vibration 5 to 15 Hz 0 15 mm IEC 68 2 6 15 to 25 Hz 1 mm 25 to 55 Hz 0 25 mm Polycarbonate Material Handles ABS Grey and Lexan 500R Red UL94V0 Jaws
6. ly with the carrier and notify your distributor at once giving a detailed description of any damage AEMC INSTRUMENTS 4 37 111 mm 3 98 101 mm 2 25 57 mm 1 71 43 5 mm t 3 90 99 mm ELECTRICAL SPECIFICATIONS Current Range 1 to 1000A AC continuous lt 1 kHz Transformation Ratio 250 5 500 5 1000 5 Output Signal 5mA 10mA 20mA AC A AC 5A AC at 250 500 or 1000A Accuracy and Phase Shift Range 250 5 20 mA A Primary 1 to 5A 5A 12 5A 50A 250A 300A Accuracy S 10 0 1A 10 5 2 5 2 2 Phase N A N A 10 10 10 10 Overload 300 A for 30 mn on 15 mn OFF 1 22 31 mm _ aS 1 38 35 mm 1 77 45 mm Range 500 5 10 mA A Primary 1 to 10A 10A 25A 100A 500A 600A Accuracy lt 6 0 1A 6 3 2 1 1 Phase N A 6 4 3 2 5 2 5 Overloa Range 1 d 600 A for 30 mn on 15 mn OFF 000 5 5 mA A Primary 1 to 20A 20A 50A 200A 1000A 1200A Accuracy lt 6 0 1A 5 3 1 5 1 1 Phase N A 5 3 1 5 ly 1 Overloa Hz externa d 1200 A for 30 on Reference conditions 23 C 3 K 20 15 mn OFF to 75 RH 48 to 65 magnetic field lt 40 A m no DC component no external
7. ne 600 to avoid scratching the jaw then gently clean with a soft oiled cloth REPAIR AND CALIBRATION You must contact our Service Center for a Customer Service Authorization number CSA This will ensure that when your instrument arrives it will be tracked and processed promptly Please write the CSA on the outside of the shipping container If the instrument is returned for calibration we need to know if you want a standard calibration or a calibration traceable to N I S T includes calibration certificate plus recorded calibration data Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive e Dover NH 03820 USA Tel 800 945 2362 Ext 360 603 749 6434 Ext 360 Fax 603 742 2346 or 603 749 6309 repair aemc com Or contact your authorized distributor Costs for repair standard calibration and calibration traceable to N I S T are available NOTE All customers must obtain a CSA before returning any instrument TECHNICAL AND SALES ASSISTANCE If you are experiencing any technical problems or require any assistance with the proper use or application of this instrument please call our technical hotline 800 343 1391 508 698 2115 Fax 508 698 2118 Chauvin Arnoux Inc d b a AEMC Instruments techsupport aemc com www aemc com En 99 VMAN 100134 v5 04 09