1. 2400 LV 2410 2410 C 200V 1A 20W SourceMeter 200V 1A 20W SourceMeter with Contact Check 20V 1A 20W SourceMeter 1100V 1A 20W SourceMeter 1100V 1A 20W SourceMeter with Contact Check 60V 3A 60W SourceMeter 60V 3A 60W SourceMeter with Contact Check 100V 3A 100W SourceMeter 100V 3A 100W SourceMeter with Contact Check 100V 10A 1000W Pulse Mode SourceMeter SourceMeter Line I V Characteristics All SourceMeter instruments provide four quadrant operation In the first and third quadrants they operate as a source delivering power to a load In the second and fourth quadrants they operate as a sink dissipating power internally Voltage current and resistance can be measured during source or sink operation Local IN OUT HI I Remote Vimeter Compliance Local SENSE HI Remote SENSE LO IN OUT LO X Source I Measure V I or Q configuration A p 100mA 20V 200V 100mA B 14 Duty cycle limite Local IN OUT HI Imeter Compliance Vsource Feedback to Adjust Vsource Remote SENSE HI Remote SENSE LO Local IN OUT LO Source V Measure I V or Q configuration 1100V 200V 20V 1A 200mA 20mA 20V 200V 1100V 20mA 100mA Lal aind 1A Model 2400 and 2400 LV SourceMeter Model 2410 High Voltage SourceMeter 100V 10A 1000W Pulse Mode
2. 15Q and 50Q e No energy passes through the device under test during the operation Enabled either from the front panel or remotely over the GPIB e 3 fault notification methods GUARD SENSE IN OUT HI 350us Contact Check gt optional SENSE HI SENSE LO IN OUT LO Contact check option for 4 wire or 6 wire applications 1 888 KEITHLEY us only www keithley com SourceMeter Line Unique 6 Wire Ohms Technique SourceMeter instruments can make standard 4 wire split Kelvin and 6 wire guarded ohms measurements and can be configured for either the constant current or constant voltage method The 6 wire ohms technique Uses guard and guard sense leads in addition to the 4 wire sense and source leads e Locks out parallel current paths when measuring resistor networks or hybrid circuits to isolate the component under test e Allows users to configure and plot data easily from Series 2400 SourceMeter instruments making characterization of two three and four terminal devices a snap GUARD GUARD SENSE meter IN OUT HI SENSE HI Vmeter DUT R2 RI R3 SENSE LO E3 Hares ia IN OUT LO v 6 Wire Ohms Circuit All test current flows through R1 because the high current guard drives the voltage across R2 to OV sea e Ea g g 2 p TN modiy iktik
3. 1A 10A q No contact check failure lt 1 00 Q lt B5Q lt 475 Q ranges add 0 05 For 0 01 PLC add 0 05 of range to offset specifications except 200mV 1A 10A ranges Always contact check failure gt 3 00 Q gt 16 5 Q gt 525 Q ada nag o x ie 7a 1 888 KEITHLEY uss only KEITHLEY www keithley com A GREATER MEASURE OF CONFIDENCE Series 2400 SourceMeter Line Resistance Measurement Accuracy Local or Remote Sense gt 5 SOURCE V MODE MANUAL OHMS Total uncertainty V source accuracy I meas 3 Manual ohms only except 2420 2425 2430 2440 for 2Q range and 2410 or 2400 for 200MQ range 4 Source readback enabled offset compensation ON Also available on 2410 2420 2425 2430 and 2440 with similar accuracy enhancement 6 WIRE OHMS MODE Available using active ohms guard and guard sense Max 5 In pulse mode limited to 0 1 PLC measurement Guard Output Current 50mA except 1A range Accuracy is load dependent 6 Except 2440 default test current is 5A Refer to White Paper no 2033 for calculation formula Except 2440 default test current is 0 5uA GUARD OUTPUT IMPEDANCE lt 0 1Q in ohms mode ure accuracy 4 wire remote sense Default Test Enhanced Accuracy Normal Accuracy 23 C 5 C 23 C 5 C Default Test Current 1 Year rd h 1 Year 4 ed h Default Current 2420 2425 ear rdg ohms ear rdg ohms Range Resolution 2400 2410 2430 2440 2400 2410 2420 2425 24
4. Accuracy Measurement gt 6 7 1 Year Accuracy 1 Year Programming 23 C 5 C Default Measurement 23 C 5 C Model Range Resolution rdg amps Resolution rdg amps Source Sink Limit 1 00000 uA 50 pA 0 035 600 pA 10 pA 0 029 300 pA 10 0000 uA 500 pA 0 033 2nA 100 pA 0 027 700 pA 100 000 uA 5nA 0 031 20 nA 1nA 0 025 6nA eh de fone 1 00000 ma 50 nA 0 034 200 nA 10nA 0 027 60 nA Err a ae P 10 0000 mA 500 nA 0 045 2uA 100 nA 0 035 600 nA S 100 000 mA 5puA 0 066 20 uA 1 uA 0 055 6uA Riss 1 00000 A 50 uA 0 27 900 uA 10 uA 0 22 570 uA S 1 00000 uA 50 pA 0 035 600 pA 10 pA 0 029 300 pA 10 0000 uA 500 pA 0 033 2nA 100 pA 0 027 700 pA D 100 000 uA 5nA 0 031 20 nA Ind 0 025 6nA 1 05A 21 V o 2410 2410 C 1 00000 mA 50 nA 0 034 200 nA 10 nA 0 027 60 nA 21 aK 1100 y oO 20 0000 mA 500 nA 0 045 4A 100 nA 0 035 1 2 uA D 100 000 mA 5 uA 0 066 20 uA 1pA 0 055 6uA 1 00000 A 50 uA 0 27 900 uA 10 uA 0 22 570 uA 10 0000 uA 500 pA 0 033 2nA 100 pA 0 027 700 pA 5 100 000 uA 5nA 0 031 20nA Ind 0 025 6nA 0 1 00000 mA 50 nA 0 034 200 nA 10 nA 0 027 60nA 43 154 21 V j 2420 2420 C 10 0000 mA 500 nA 0 045 2A 100 nA 0 035 600 nA 05 A 63 v Q 100 000 mA 5 uA 0 066 20 uA 1 uA 0 055 6uA _ N 1 00000 A 50 uA 0 067 900 uA 10 uA 0 066 570 uA D 3 00000 A 50 uA 0 059 2 7 mA 10 uA 0 052 1 71 mA 5 10 0000 uA 500 pA 0
5. 3 ISO 17025 accredited calibrations within 3 years of purchase for Models 2425 2425 C C 2430 3Y ISO 3 ISO 17025 accredited calibrations within 3 years of purchase for Models 2430 2430 C C 2440 3Y ISO 3 ISO 17025 accredited calibrations within 3 years of purchase for Models 2440 2440 C TRN 2400 1 C Course Unleashing the Power of Your SourceMeter Instrument Not available in all countries Ww a x wi Z x m 8 oO a 1 888 KEITHLEY v s only KEITHLEY www keithley com A GREATER MEASURE OF CONFIDENCE Series 2400 SourceMeter Line System Speeds MEASUREMENT MAXIMUM RANGE CHANGE RATE 75 second MAXIMUM MEASURE AUTORANGE TIME 40ms fixed source Sweep Operation Reading Rates rdg second for 60Hz 50Hz Source Measure gt Measure Source Measure Pass Fail Test 5 Source Memory Speed NPLC Trigger Origin To Mem To GPIB To Mem To GPIB To Mem To GPIB To Mem To GPIB Fast 0 01 internal 2081 2030 1754 1551 1515 1369 902 900 981 165 162 165 IEEE 488 1 Mode 0 01 external 1239 1200 1254 1018 990 1035 830 830 886 163 160 163 Fast 0 01 internal 2081 2030 1198 1210 1551 1515 1000 900 902 900 809 840 165 162 164 162 IEEE 488 2 Mode 0 01 external 1239 1200 1079 1050 1018 990 916 835 830 830 756 780 163 160 162 160 Medium 0 10 internal 510 433 509 433 470 405 470 410 389 343 388 343 133 126 132 126 IEEE 488 2 Mod
6. SourceMeter with Contact Check 40V 5A 50W SourceMeter 40V 5A 50W SourceMeter with Contact Check Accessories Supplied Model 8605 Test Leads LabVIEW Software Driver downloadable LabTracer Software downloadable 1 888 KEITHLEY us only 7 60V 20V 1A 100mA 20V 60V a Duty cycle 3A Model 2420 3A SourceMeter 100V 60 20 1A 100mA 20 60 100V 100mA 1A Duty cycle af imite Pulse mode only 10A Model 2430 1kW Pulse Mode SourceMeter A GREATER MEASURE OF p p 100V 60 20 100mA 20 60 100V 100mA wall Model 2425 100W So 1A E Duty cycle limite 3A urceMeter 10V 40v Model 2440 5A Sourc CONFIDEN A 100mA 1A 3A Duty cycle limite eMeter u v q lt u Z q wu S x e n CE Series 2400 SourceMeter Line Automation for Speed A SourceMeter instrument streamlines production testing It sources voltage or current while making Cec onoN gt measurements without needing to change connections It is designed for reliable operation in non Devices stop production environments To provide the throughput demanded by production applications the SourceMeter instrument offers many built in features that allow it to run complex test sequences A without computer control or GPIB communications slowing things down e Passive devices e Transient suppression devi
7. 0000 V 500 uV 0 02 2 4 mV 100 uV 0 015 1ImV 0 15 V jus 1100 V 21 mA 1000 00 V 50 mV 0 02 100 mV 10 mV 0 015 50 mV 0 5 Vius 200 000 mV 5 uV 0 02 600 uV 1 uv 0 012 300 uV 2420 2420 C 2 00000 V 50 wv 0 02 600 uV 10 uV 0 012 300 uV 21 V 3 15 A 20 0000 V 500 uV 0 02 2 4 mV 100 uV 0 015 1 mV 0 08 Vus 63 V 1 05 A 60 0000 V 15 mV 0 02 7 2 mV 1 mV 0 015 3 mV 0 14 Vus 200 000 mV 5 wv 0 02 600 LV 1 pv 0 012 300 uV 2425 2425 C 2 00000 V 50 uV 0 02 600 uV 10 uV 0 012 300 uV 21 V 3 15 A i 20 0000 V 500 uV 0 02 2 4 mV 100 uV 0 015 1mV 0 08 V us 105 V 1 05 A 100 0000 V 2 5 mV 0 02 12 mV 1 mV 0 015 5 mV 0 25 Vius 200 000 mV 5 uv 0 02 600 uV 1 uv 0 012 300 uV 105 V 1 05A 2430 2430 C 2 00000 V 50 wv 0 02 600 uV 10 uV 0 012 300 uV nea aE 20 0000 V 500 uV 0 02 2 4 mV 100 pV 0 015 1mv 0 08 Vus 105 V 10 5 A 100 0000 V 25 mV 0 02 12 mV mV 0 015 5 mV 0 25 Vus pulse mode only 200 000 mV 5 wv 0 02 600 uV 1 pv 0 012 300 uV 2440 2440 C 2 00000 V 50 uV 0 02 600 uV 10 wv 0 012 300 uV 10 5 V 5 25 A 10 0000 V 500 uV 0 02 1 2 mV 100 uV 0 015 750 uV 0 08 V us 42 V 1 05 A 40 0000 V 5 mV 0 02 4 8 mV 1mV 0 015 3 mV 0 25 Vius TEMPERATURE COEFFICIENT 0 18 C and 28 50 C 0 15 x accuracy specification C VOLTAGE REGULATION Line 0 01 of range Load 0 01 of range 100 OVER VOLTAGE PROTECTION User selectable values 5 tol
8. 033 2nA 100 pA 0 027 700 pA 7 100 000 uA 5nA 0 031 20nA 1nA 0 025 6nA 1 00000 mA 50 nA 0 034 200 nA 10 nA 0 027 60nA 3154 21 V 2425 2425 C 10 0000 mA 500 nA 0 045 2A 100 nA 0 035 600 nA 108 A 105 y 100 000 mA 5 uA 0 066 20 uA luA 0 055 6uA i 1 00000 A 50 uA 0 067 900 uA 10 uA 0 060 570 uA 3 00000 A 50 uA 0 059 2 8 mA 10 uA 0 052 1 71 mA 10 0000 uA 500 pA 0 033 2nA 100 pA 0 027 700 pA 100 000 uA 5nA 0 031 20nA Ind 0 025 6nA 1 00000 mA 50 nA 0 034 200 nA 10 nA 0 027 60nA 1 05A 105 V 2430 2430 C 10 0000 mA 500 nA 0 045 2A 100 nA 0 035 600 nA i 100 000 mA 5 uA 0 066 20 pA luA 0 055 GUA 10 5 A 105 V 1 00000 A 50 uA 0 067 900 uA 10 uA 0 060 570 uA pulse mode only 3 00000 A 500 uA 0 059 2 8mA 10 uA 0 052 1 71 mA 10 00000 A 500 uA 0 089 5 9 mA 10 uA 0 082 1 71 mA 10 0000 uA 500 pA 0 033 2nA 100 pA 0 027 700 pA 100 000 uA 5nA 0 031 20nA 1nA 0 025 6 nA 1 00000 mA 50 nA 0 034 200 nA 10 nA 0 027 60 nA 5254 10 5 V 2440 2440 C 10 0000 mA 500 nA 0 045 2puA 100 nA 0 035 600 nA 05A 42 y 100 000 mA 5 uA 0 066 20 uA 1A 0 055 6pA i 1 00000 A 50 uA 0 067 900 uA 10 uA 0 060 570 uA 5 00000 A 50 uA 0 10 5 4 mA 10 uA 0 10 3 42 mA TEMPERATURE COEFFICIENT 0 18 C and 28 50 C 0 15 x accuracy specification C NOTES CURRENT REGULATION Line 0 01 of range Load 0 01 of range except Model 2440 1
9. 2400 2410 Only Specifications valid for continuous output currents below 105mA For operation above 105mA 5A range 0 05 100pA continuous for gt 1 minute derate accuracy 10 35mA above 105mA A ENEE i 2 Full operation 1A regardless of load to 30 C 50 C for Model 2420 and 2440 Above 30 C 50 C for Model Lu VOLTAGE LIMIT Bipolar voltage limit compliance set with single value Min 0 1 of range 2420 and 2440 ambient derate 35mA C and prorate 35mA Q load 4 wire mode For current sink operation 4 OVERSHOOT lt 0 1 typical 1mA step RL 10kQ 20V range for Model 2400 2410 2420 2425 on 1A 3A or 5A ranges maximum continuous power is limited to approximately 1 2 rated power or less 2430 10V range for Model 2440 pisces on current up to 30 C ambient See power equations in the User s Manual to calculate allowable Wn luty cycle for specific conditions x 3 For sink mode 14A to 100mA range accuracy is Ww F x Model 2400 0 15 offset 4 Models 2410 2420 2425 2430 2440 0 5 offset 3 s CONTACT CHECK SPECIFICATIONS requires C version For 1A range accuracy is P PRN Model 2400 1 5 offset 8 Models 2410 2420 2425 2430 2440 1 5 offset 3 SPEED 9504s for verification and notification 4 10A range only ss pulse mode vied to 2 5ms pulse hers oe ae cycle lt a CONTACT CHECK 2Q 15Q 509 5 Speed Normal 1 PLC For 0 1 PLC add 0 005 of range to offset specifications except 200mv
10. 30 2440 2400 lt 0 20000 93 Source Icc Meas Vicc Source Ico Meas Vicc Source Icc Meas Vicc Source Icc Meas Vicc 2 00000 Q 10 uQ 1A Source Iioc Meas Vico Source Iioc Meas Vice 0 17 0 00032 Source Tiec Meas Vcc 20 0000 Q 100 uQ 100 mA 100 mA 0 10 0 003 Q 0 11 0 006 Q 0 10 0 003 Q 0 07 0 001 Q 200 000 Q 1 mQ 10 mA 10 mA 0 08 0 03 Q 009 O01 Q 0 08 0 03 Q 0 05 0 01 Q 2 00000 kQ 10 mQ 1mA 1mA 0 07 03 Q 0 08 06 Q 0 07 03 Q 0 05 0 1 Q 20 0000 kQ 100 mQ 100 uA 100 uA 0 06 3 Q 0 07 6 Q 0 06 3 Q 0 04 1 Q E 200 000 KQ 1 Q 10 uA 10 uA 0 07 30 Q 0 07 60 Q 0 07 30 Q 0 05 10 Q Fr 2 00000 MQS 10 Q 1 uA 1 uA 0 11 300 Q 0 12 600 Q 0 11 300 Q 0 05 100 Q Go 20 0000 MQ 100 Q 1 uA 1 pA 0 11 1 KQ 0 12 24 kQ 0 11 1 kQ 0 05 500 Q g 200 000 MQ 1 ko 100 nA 0 66 10 KQ 0 66 24 kQ Source Icg Meas Vicc 0 35 5 kQ 5 gt 200 000 MQ Source Icc Meas Vicc Source Icc Meas Vice Source Icc Meas Vicc Source Icc Meas Vicc 5 a 2190 _ 6 97 9 Ley are sala 0 18 C and 28 50 C 0 15 x accuracy NOTES 1 Speed Normal 1 PLC For 0 1 PLC add 0 005 of range to offset specifications except 200mV 1A 10A ranges add 0 05 For SOURCE I MODE MANUAL OHMS Total uncertainty I source accuracy V meas 0 01 PLC add 0 05 of range to offset specifications except 200mV 1A 10A ranges add 0 5 z ure accuracy 4 wire remote sen
11. 3V 500mA diode clamped POWER SUPPLY 100V to 240V rms 50 60Hz automatically detected at power up Model 2400 190VA Model 2410 210VA Model 2420 220VA Model 2425 2430 250VA Model 2440 240VA COOLING Model 2410 2420 2425 2430 2440 Forced air variable speed EMC Conforms to European Union Directive 89 336 EEC EN 61326 1 SAFETY Conforms to European Union Directive 73 23 EEC EN61010 1 VIBRATION MIL PRF 28800F Class 3 Random WARM UP 1 hour to rated accuracies DIMENSIONS 89mm high x 213mm wide x 370mm deep 342 in x 8 in x 14 in Bench Configuration with handle and feet 104mm high x 238mm wide x 370mm deep 4 in x 9 in x 14 6 in WEIGHT 3 21kg 7 08 lbs Model 2425 2430 2440 4 1kg 9 0 lbs ENVIRONMENT Operating 0 50 C 70 R H up to 35 C Derate 3 R H C 35 50 C Storage 25 C to 65 C KEITHLEY GREATER MEASURE OF CO INF I DE NCE el S 1S VU oO a wn wT lt b wn Oo wT 2 O o o N wn oO k oO YN SOURCE AND MEASURE
12. All other ranges 100 MAXIMUM PULSE WIDTH 5ms from 90 rising to 90 falling edge 2 5ms 10A range MINIMUM PULSE WIDTH 150ys MINIMUM PULSE RESOLUTION 50us typical 70s max limited by system jitter SOURCE ACCURACY Determined by settling time and source range specifications OUTPUT SETTLING TIME 0 1 800us typ source I 10A into 10Q limited by voltage slew rate 500us typ source I 10A into 10 limited by voltage slew rate OUTPUT SLEW RATE Voltage 10Q load 0 25V us 30 on 100V range 0 08V us 30 on 20V range 10A range Current 0Q load 0 25A us 30 on 100V range 0 08A us 30 on 20V range 10A range NOTES 1 2400 2410 Only Specifications valid for continuous output currents below 105mA For operation above 105mA continuous for gt 1 minute derate accuracy 10 35mA above 105mA 2 Speed Normal 1 PLC For 0 1 PLC add 0 005 of range to offset specifications except 200mV 1A 10A ranges add 0 05 For 0 01 PLC add 0 05 of range to offset specifications except 200mV 1A 10A ranges add 0 5 3 Accuracies apply to 2 or 4 wire mode when properly zeroed In pulse mode limited to 0 1 PLC measurement A KEITHLEY GREATER MEASURE OF CONFIDENCE Series 2400 SourceMeter Line Current Accuracy Local or Remote Sense 6 Accuracies apply to 2 or 4 wire mode when properly zeroed 7 In pulse mode limited to 0 1 PLC measurement 8 Model 2400 and 2400 C only Source 3
13. Ww 7 lt x wi A Z x m 8 4 oO a Series 2400 Five instruments in one IV Source IVR Measure Six models 20 100W DC 1000W pulsed 1100V to 1pV 10A to 10pA Source and sink 4 quadrant operation 0 012 basic measure accuracy with 5 2 digit resolution 2 4 and 6 wire remote V source and measure sensing 1700 readings second at 412 digits via GPIB Pass Fail comparator for fast sorting binning Available high speed sense lead contact check function Programmable DIO port for automation handler prober control Standard SCPI GPIB RS 232 and Keithley Trigger Link interfaces Keithley LabTracer 2 0 I V curve tracing application software download 1 888 KEITHLEY us only SourceMeter Line Keithley s SourceMeter family is designed specifically for test applications that demand tightly coupled sourcing and measure ment All SourceMeter models provide precision voltage and current sourcing as well as measurement capabilities Each SourceMeter instrument is both a highly stable DC power source and a true instrument grade 5 2 digit multimeter The power source characteristics include low noise precision and readback The multimeter capabilities include high repeatability and low noise The result is a compact single channel DC parametric tester In operation these instruments can act as a voltage source a current source a voltage meter a current meter and an ohm mete
14. ad bii ii ii Vibe Voltage 1 Ming 1 s Ties A W a zza Wiel maol ono Mies Free LabTracer 2 0 device characterization software downloadable KEITHLEY GREATER MEASURE OF CO INF I DE NCE 2 oO wo D wn oO a eTe O 5 O n a VU oO D wT 2 a O O gt 2 pis oD kK SOURCE AND MEASURE wn S 1S VU oO for 2 wow lt b wn wo wT je O l o N i oO i oO Yn SOURCE AND MEASURE Series 2400 Voltage Accuracy Local or Remote Sense SourceMeter Line Source Measurement 5 4 Accuracy 1 Year Default Accuracy 1 Year Programming 23 C 5 C Measurement 23 C 5 C Output Slew Rate Model Range Resolution rdg volts Resolution rdg volts 30 Source Sink Limit 200 000 mV 5 uV 0 02 600 uV 1 wv 0 012 300 uV 2400 2400 C 2 00000 V 50 uV 0 02 600 uV 10 aV 0 012 300 uV 21 V 1 05A i 20 0000 V 500 uV 0 02 2 4 mV 100 uV 0 015 1 5 mV 0 08 V us 210 V 105 mA 200 000 V 5 mV 0 02 24 mV ImV 0 015 10 mV 0 5 Vius 200 000 mV 5 uV 0 02 600 uV 1 wv 0 012 300 uV 2400 LV 2 00000 V 50 uV 0 02 600 uV 10 WV 0 012 300 uV 21 V 1 05 A 20 0000 V 500 uV 0 02 2 4 mV 100 uV 0 015 1 5 mV 0 08 V us 200 000 mV 5 uv 0 02 600 uV 1 wv 0 012 300 uV 2410 2410 C 2 00000 V 50 uV 0 02 600 uV 10 uV 0 012 300 uV 21 V 1 05 A 20
15. ces e ICs RFICs MMICs Discrete semiconductor devices Standard and Custom Sweeps e 5000 5 2 digit readings can be stored in the e Temperature coefficient non volatile buffer memory Forward voltage reverse breakdown leakage current e DC parametric test e DC power source Custom sweep e HIPOT e Dielectric withstanding my o 5 Sweep solutions greatly accelerate testing with Sep p 3 5 e m automation hooks Three basic sweep waveforms cart ode aa ee are provided that can be programmed for single Start modules LEDs photodetectors mm event or continuous operation They are ideal for e Circuit protection devices TVS 1 V Y R V I and V R characterization Bias Bias MOV Fuses etc J Linear Staircase Sweep Moves from the start 5 e Airbags 5 level to the stop level in equal linear steps Linear staircase sweep e Connectors switches relays Logarithmic Staircase Sweep Done on a log Tests scale with a specified number of steps per lt Stop T decade eakage wn e H S54 Custom Sweep Allows construction of special Start Low voltage resistances sweeps by specifying the number of measure e LIV g ment points and the source level at each point Blas Bias e IDDQ JE Up to 1700 readings second at 4 digits to the Logarithmic staircase sweep e l V characterization 9 GPIB bus e Isolation and trace resistance gt 100 a Built In Test Se
16. d at the limit providing fault protection ACCESSORIES AVAILABLE TEST LEADS AND PROBES COMMUNICATION INTERFACE 1754 2 Wire Universal 10 Piece Test Lead Kit KPCI 488LPA IEEE 488 Interface Controller for the PCI Bus 5804 Kelvin 4 Wire Universal 10 Piece Test Lead Kit KUSB 488B IEEE 488 USB to GPIB Interface Adapter 5805 Kelvin 4 Wire Spring Loaded Probes TRIGGERING AND CONTROL 5808 Low Cost Single pin Kelvin Probe Set 2499 DIGIO Digital I O Expander Assembly 3809 Rone Cosh Kelvin Clip Lead Set 8501 1 Trigger Link Cable DIN to DIN 1m 3 3 fi 8607 2 Wire 1000V Banana Cables 1m 3 3 ft 8501 2 Trigger Link Cable DIN to DIN 2m 6 6 ft CA 18 1 Shielded Dual Banana Cable 1 2m 4 ft 8502 Trigger Link to BNC Breakout Box SWITCHING HARDWARE 8503 Trigger Link Cable DIN to Dual BNC 1m 3 3 ft 7001 Two Slot Switch System 8505 Male to 2 Female Y DIN Cable for Trigger Link 7002 Ten Slot Switch System RACK MOUNT KITS 7019 C 6 Wire Ohms Switch Card 4288 1 Single Fixed Rack Mount Kit 7053 High Current Switch Card 4288 2 Dual Fixed Rack Mount Kit CABLES ADAPTERS 4288 4 Dual Fixed Rack Mount Kit 7007 1 Shielded GPIB Cable 1m 3 3 ft 4288 5 Shelf Type Side by Side Rack Mounting Kit 7007 2 Shielded GPIB Cable 2m 6 6 ft 4288 9 Dual Fixed Rack Mounting Kit 7009 5 RS 232 Cable SOFTWARE 8620 Shorting Plug LabTracer 2 0 Curve Tracing Software downloadable A GREATER MEASURE OF CONFIDENCE Series 2400 Ordering Info 2400 2400 C
17. e 0 10 external 438 380 438 380 409 360 409 365 374 333 374 333 131 125 131 125 Normal 1 00 internal 59 49 59 49 58 48 58 48 56 47 56 47 44 38 44 38 IEEE 488 2 Mode 1 00 external 57 48 57__ 48 57 48 57__ E 56 47 56 47 44 68 44 688 Single Reading Operation Reading Rates rdg second for 60Hz 50Hz Measure Source Measure5 Source Measure Pass Fail Test gt Speed NPLC Trigger Origin To GPIB To GPIB To GPIB Fast 488 1 0 01 internal 537 140 135 Fast 488 2 0 01 internal 256 256 79 83 79 83 Medium 488 2 0 10 internal 167 166 72 70 69 70 Normal 488 2 1 00 internal 49 42 34 Gl 35 80 Component for 60Hz 50Hz 6 Source Measure Pass Fail Test5 7 Speed NPLC Trigger Origin Measure To GPIB Source Pass Fail Test To GPIB Fast 0 01 external 1 04 ms 1 08 ms 0 5 ms 0 5 ms 4 82 ms 5 3 ms Medium 0 10 external 2 55 ms 2 9 ms 0 5 ms 0 5 ms 6 27 ms 7 1 ms Normal 1 00 external 17 53 ms 20 9 ms 0 5 ms 0 5 ms 21 31 ms 25 0 ms NOTES 1 Reading rates applicable for voltage or current measurements Auto zero off autorange off filter off display off trigger delay 0 and binary reading format 2 Purely resistive lead 14A and 10uA ranges lt 65ms 3 1000 point sweep was characterized with the source on a fixed range 4 Pass Fail test performed using one high limit and one low math limit 5 Includes time to re program source to a new level befo
18. erance Factory default none CURRENT LIMIT Bipolar current limit compliance set with single value Min 0 1 of range OVERSHOOT lt 0 1 typical full scale step resistive load 10mA range ADDITIONAL SOURCE SPECIFICATIONS All Models TRANSIENT RESPONSE TIME 30s minimum for the output to recover to its spec following a step change in load COMMAND PROCESSING TIME Maximum time required for the output to begin to change following the receipt of SOURce VOLTage CURRent lt nrf gt command Autorange On 10ms Autorange Off 7ms OUTPUT SETTLING TIME Time required to reach 0 1 of final value after command is pro cessed 100s typical Resistive load 104A to 100mA range DC FLOATING VOLTAGE Output can be floated up to 250VDC Model 2440 40VDC from chassis ground REMOTE SENSE Up to 1V drop per load lead COMPLIANCE ACCURACY Add 0 3 of range and 0 02 of reading to base specification OVER TEMPERATURE PROTECTION Internally sensed temperature overload puts unit in standby mode RANGE CHANGE OVERSHOOT Overshoot into a fully resistive 100kQ load 10Hz to IMHz BW adjacent ranges 100mV typical except 20V 200V 20V 60V on Model 2420 20V 100V on Model 2425 and 2430 range boundary and Model 2440 MINIMUM COMPLIANCE VALUE 0 1 of range 1 888 KEITHLEY us only www keithley com ADDITIONAL PULSE MODE SOURCE SPECIFICATIONS 2430 and 2430 C only MAXIMUM DUTY CYCLE 8 hardware limited 10A range only
19. ght systems integration for applications such as binning and sorting Built in component handler interface Start of test and end of test signals e 5V 300mA power supply e Optional expander accessory Model 2499 DIGIO adds 16 digital I O lines Trigger Link Interface All SourceMeter instruments include Keithley s unique Trigger Link inter face which provides high speed seamless communications with many of Keithley s other instruments For example use the Trigger Link interface to connect a SourceMeter instrument with a Series 7000 Switching System for a complete multi point test solution With Trigger Link the 7000 Series Switching Systems can be controlled by a SourceMeter during a high speed test sequence independent of a computer and GPIB Optional Contact Check Function The Contact Check function makes it simple to verify good connections quickly and easily before an automated test sequence begins This elimi nates measurement errors and false product failures associated with con tact fatigue breakage contamination loose or broken connection relay failures etc Some capabilities of this function are e 350us verification and notification process time The output of the SourceMeter instrument is automatically shut off after a fault and is not re activated until good contact is verified protecting the device under test from damage and the operator from potential safety hazards e 3 pass fail threshold values 2Q
20. quencer Source Memory List The Source Memory list provides faster and eas ier testing by allowing you to setup and execute up to 100 different tests that run without PC intervention Bias Bias Stores up to 100 instrument configurations each containing source settings measurement settings pass fail criteria etc e Pass fail limit test as fast as 500s per point e Onboard comparator eliminates the delay caused when sending data to the computer for analysis e Built in user definable math functions to calculate derived parameters Example Test Sequence Test Pass Fail Test If Passes Test If Fails Test Test 1 Check V at Go to Test 2 100mA against pass fail limits Test 2 CheckV at 1A Go to Test 3 1 Bin part to bad bin against pass fail 2 Transmit data to limits computer while Test 3 Check leakage 1 Bin part to good bin ae placing current at 500V 2 Transmit readings to 3 Return to Testi and test against computer while handler pass fail limits is placing new part 3 Return to Test 1 m 4 7 lt x wi Z x m 8 4 oO a 1 888 KEITHLEY us only KEITHLEY www keithley com A GREATER MEASURE OF CONFIDENCE Series 2400 Digital 1 0 Interface The digital I O interface can link the SourceMeter instrument to many popular component handlers including Aetrium Aeco and Robotronics Other capabilities of the interface include e Ti
21. r Manufacturers of components and modules for the com munications semiconductor computer automotive and medical industries will find the SourceMeter instruments invaluable for a wide range of characterization and production test applications Advantages of a Tightly Integrated Instrument By linking source and measurement circuitry in a single unit these instruments offer a variety of advantages over systems configured with separate source and measurement instruments For example they minimize the time required for test station development setup and maintenance while lowering the overall cost of system ownership They simplify the test process itself by eliminating many of the complex synchronization and connection issues associated with using multiple instruments And their compact half rack size conserves pre cious real estate in the test rack or bench Power of Five Instruments in One IV Source IVR Measure The tightly coupled nature of a SourceMeter instrument provides many advantages over separate instruments For example it provides faster test times by reducing GPIB traffic and simplifies the remote programming interface It also protects the device under test from damage due to accidental overloads thermal runaway etc Both the current and voltage source are programmable with read back to help maximize device measurement integrity If the readback reaches a programmed compli ance limit then the source is clampe
22. re making measurement Time from falling edge of START OF TEST signal to falling edge of END OF TEST signal 7 Command processing time of SOURce VOLTage CURRent TRIGgered lt nrf gt command not included GENERAL Noise Rejection NPLC NMRR 0 01 0 1 60 dB CMRR 80 dB 80 dB 100 dB Fast Medium Slow 1 1 Except lowest 2 current ranges 90dB LOAD IMPEDANCE Stable into 20 000pF typical COMMON MODE VOLTAGE 250V DC 40V DC for Model 2440 COMMON MODE ISOLATION gt 10 Q2 lt 1000pF OVERRANGE 105 of range source and measure MAX VOLTAGE DROP BETWEEN INPUT OUTPUT AND SENSE TERMINALS 5V MAX SENSE LEAD RESISTANCE 1MQ for rated accuracy SENSE INPUT IMPEDANCE gt 10 Q GUARD OFFSET VOLTAGE lt 150 V typical 300uV for Models 2430 2440 SOURCE OUTPUT MODES Pulse Model 2430 only Fixed DC level Memory List mixed function Stair linear and log MEMORY BUFFER 5 000 readings 5 digits two 2 500 point buffers Includes selected measured value s and time stamp Lithium battery backup 3 yr battery life 1 888 KEITHLEY us only www keithley com A SOURCE MEMORY LIST 100 points max PROGRAMMABILITY IEEE 488 SCPI 1995 0 RS 232 5 user definable power up states plus factory default and RST DIGITAL INTERFACE Interlock Active low input Handler Interface Start of test end of test 3 category bits 5V 300mA supply Digital I O 1 trigger input 4 TTL Relay Drive outputs 3
23. se 2 Accuracies apply to 2 or 4 wire mode when properly zeroed wn 2 o Wn xN SERVICES AVAILABLE 2400 3Y EW 1 year factory warranty extended to 3 years from date of shipment 2400 C 3Y EW __ 1 year factory warranty extended to 3 years from date of shipment 2400 LV 3Y EW 1 year factory warranty extended to 3 years from date of shipment 2410 3Y EW 1 year factory warranty extended to 3 years from date of shipment 2410 C 3Y EW 1 year factory warranty extended to 3 years from date of shipment 2420 3Y EW 1 year factory warranty extended to 3 years from date of shipment 2420 C 3Y EW _ 1 year factory warranty extended to 3 years from date of shipment 2425 3Y EW 1 year factory warranty extended to 3 years from date of shipment 2425 C 3Y EW 1 year factory warranty extended to 3 years from date of shipment 2430 3Y EW 1 year factory warranty extended to 3 years from date of shipment 2430 C 3Y EW 1 year factory warranty extended to 3 years from date of shipment 2440 3Y EW 1 year factory warranty extended to 3 years from date of shipment 2440 C 3Y EW __ 1 year factory warranty extended to 3 years from date of shipment 2400 3Y ISO 3 ISO 17025 accredited calibrations within 3 years of purchase for Models 2400 2400 C 2400 LV C 2410 3Y ISO 3 ISO 17025 accredited calibrations within 3 years of purchase for Models 2410 2410 C 2420 3Y ISO 3 ISO 17025 accredited calibrations within 3 years of purchase for Models 2420 2420 C C 2425 3Y ISO