        HumidAire 2i™ User Manual
         Contents
1.        Note  Some models are not available in all regions        f the product fails under conditions of normal use  ResMed will repair or replace   at its option  the defective product or any of its components  This Limited  Warranty does not cover    a  any damage caused as a result of improper use  abuse  modification or  alteration of the product    b  repairs carried out by any service organization that has not been expressly  authorized by ResMed to perform such repairs    c  any damage or contamination due to cigarette  pipe  cigar or other smoke    d  any damage caused by water being spilled on or into a flow generator   Warranty is void on product sold  or resold  outside the region of original  purchase  Warranty claims on defective product must be made by the initial  consumer at the point of purchase    This warranty is in lieu of all other express or implied warranties  including any  implied warranty of merchantability or fitness for a particular purpose  Some  regions or states do not allow limitations on how long an implied warranty lasts   so the above limitation may not apply to you    ResMed shall not be responsible for any incidental or consequential damages  claimed to have occurred as a result of the sale  installation or use of any ResMed  product  Some regions or states do not allow the exclusion or limitation of  incidental or consequential damages  so the above limitation may not apply to  you  This warranty gives you specific legal rights  and you m
2.  Make sure that the CPAP  APAP or bilevel device is switched off at the  power outlet before connecting the HumidAire 2i    e Make sure the power cord and plug are in good condition and the  equipment is not damaged     Place the device on a firm  flat surface near your bed  Make sure that it is placed  at a level lower than your sleeping position so that any excess condensation  drains back into the water chamber     CAUTION    Be careful not to place the device where it can be bumped into and knocked  over     WARNING  Do not fill the water chamber while it is attached to the docking station and  the CPAP  APAP or bilevel device     Turn the water chamber so that the air outlet points up  Over a sink  carefully fill  the water chamber through the air outlet up to the maximum water level mark  shown on the side of the water chamber base     Air outlet    Maximum  water level  mark       The HumidAire 2i should be filled with fresh water before use  Always ensure   that the water chamber is clean  See    Cleaning and Maintenance    on page 13   WARNING   Do not overfill the water chamber  as water may enter the CPAP  APAP  or  bilevel device and air tubing    CAUTION   Do not use additives  in particular aromatic additives  eg  scented oils and    perfumes   They may reduce the humidification output of the device and or  cause deterioration of the water chamber materials     Preparing For Use    English    7 Carefully  slide and push the filled water chamber onto the humidifi
3.  active  The HumidAire 2i has a rapid heating element  which allows very short heating times  Do not be concerned if the light only  flashes on very briefly     Lie down and arrange the air tubing so that it is free to move if you turn in your  sleep     WARNING  Do not leave long lengths of air tubing around the top of your bed  It could  twist around your head or neck while you are sleeping     Stopping Treatment    To stop treatment at any time  remove your mask and press the Front or the  Enter key  depending on your model   Alternately  if the SmartStart function is  enabled  simply remove your mask and treatment will end    You may notice that the device continues to blow air gently after you have  pressed the Front or the Enter key  depending on your model   This continues  for up to one hour to assist cooling of the heater plate  However  you can switch  off the power at any time and allow the heater plate to cool without airflow     Operating Instructions 11    12    Helpful Hints    Condensation in the Mask or Air Tubing   You may experience excessive condensation in the mask or air tubing  This may  occur seasonally  particularly when the weather is cool  Turn the humidity control  dial down until the condensation decreases    Another cause of condensation is air blowing directly on the air tubing  for  example from fans and air conditioning units  This causes the humidified air to  quickly lose temperature  resulting in increased condensation  Avoid this  situatio
4.  reliable operation provided that it is operated and  maintained in accordance with the instructions provided by ResMed  Applicable  ResMed warranty details are provided with the device at the time of original  supply  Of course  as with all electrical devices  if any irregularity becomes  apparent  you should exercise caution and have the device inspected by an  Authorized ResMed Service Center     Cleaning and Maintenance    17    English    18    Troubleshooting    If there is a problem  try the following suggestions  If the problem cannot be  solved  contact your equipment supplier or ResMed  Do not attempt to open the    unit     WARNING    Service and repair should only be performed by an authorised service    agent  Under no circumstances should you attempt to service or repair the    unit yourself     Problem    Device not working   no display on LCD  screen     Possible Cause    Power cord is not plugged  in properly     Device not switched on     Faulty power outlet     Solution    Check that the power cord is plugged  firmly into the power outlet     Check that the switch at the back of  the device is in the ON position     Try another power outlet        Heater plate light not  lighting up    The Front key on the  device has not been  pressed     Control dial is in the OFF  position     Press the Front key on the  device to activate the heater plate     Turn the control dial up from the OFF  position        Leaking water  chamber    The water chamber may  be damag
5.  with condensate  the mask    might fill with water  The  in patients who are unab  mask off  for example fol    HumidAire 2i should therefore be used with caution  e to guard their airway  or are unable to pull the  owing a stroke  Attention should be given to the       relative risks and benefits    Do not operate the HumidAire 2i if it is not working properly or if any part of it  has been dropped or damaged    Keep the power cord away from heated surfaces     Do not drop or insert any  Do not use the HumidAir  materials     object into any opening or tubes   e 2i in the presence of flammable or explosive    Medical Information    English    3    The HumidAtre 2i    HumidAire 2i Components    Control dial          Docking  station    Heater plate    Air outlet    Water chamber lid    Seal    Cross member    Water chamber base  Side catch    Tub plate    Figure 2  Water chamber components    Preparing For Use    Setting Up the HumidAire 2i    WARNING    Make sure that the CPAP  APAP  or bilevel device is switched off at the  power outlet before connecting the HumidAire 2i           1 Remove the front cap from the device  To do this    1  Press the latch underneath the front cap    2  While pressing the latch  lift and pull the front cap away from the device  The  front cap should detach from the bottom and right side  The front cap should  still be attached to the left side    Left side       N f Right side               Qi Front cap      s   Latch  3  Hold the device down 
6. RESMED    HumidAire 2i     User Manual  English    ResMed Ltd  Manufacturer  1 Elizabeth MacArthur Drive Bella Vista NSW 2153 Australia   ResMed Corp  US Designated Agent  14040 Danielson Street Poway CA 92064 6857 USA   ResMed  UK  Ltd  EU Authorized Representative  96 Milton Park Abingdon Oxfordshire OX14 4RY UK   ResMed Offices Australia  Austria  Belgium  Brazil  China  Finland  France  Germany  Greece  Hong Kong  India  Ireland  Italy   Japan  Malaysia  Mexico  Netherlands  New Zealand  Norway  Portugal  Singapore  South Africa  Spain  Sweden  Switzerland   UK  USA  see www resmed com for contact details      Patents pending    Protected by design registrations  AU 147335  CH 128 711  CH 128 712  DE 40202008  DE 40201723  ES 153516  ES 153518   FR 021407  GB 3001820  GB 3001821  JP 1164266  JP 1164267  SE 75598  SE 75715  US D476077    US D477868  Other designs pending   VPAP and HumidAire are trademarks of ResMed Ltd and are registered in U S  Patent and Trademark Office        2008 ResMed Ltd     English    Contents    Introduction           0 0    ete e tenet ens    Definitions  User Owner Responsibility    Medical Information                0 00  ccc ccc teen    Intended Use  Contraindications  Warnings and Cautions    The HumidAire 2i 3 oi c5ek etek eee eS YR a UE Se Ss    HumidAire 2i Components    Preparing For Use          0   0000 0 eee    Setting Up the HumidAire 2i    Operating Instructions         Starting Treatment  Stopping Treatment  Helpful Hints    Clean
7. amber with demineralised  water     Seasonal Changes  You may have to adjust your humidity setting from season to season  or if you  travel to a different climate     Cleaning and Maintenance    You should regularly carry out the cleaning and maintenance described in this  section     Cleaning the HumidAire 2i between patients  The HumidAire 2i should be disinfected when used between patients  Detailed  disinfection instructions are available from the ResMed website  www resmed com and must be consulted before re use     Note  The HumidAire 2i is available in both a clear polycarbonate  PC  version as  well as a smoky gray polysulphone  UDEL  version  Only the clear polycarbonate   PC  version is able to be disinfected     Cleaning the HumidAire 2i for home users  WARNING  Beware of electrocution  Do not immerse the device or power cord in water     Always unplug the device before cleaning and be sure that it is dry before  plugging back in     1 Turn the main power switch at the back of the device to OFF  and unplug the  power cord from the power outlet     2 Allow 10 minutes for the heater plate and any excess water to cool     3 Remove the water chamber  To do this  lift the latch at the bottom of the water  chamber and pull the water chamber away from the docking station        WARNING     Take care when handling the HumidAire 2i  The water in the water  chamber may be hot      Do not touch the tub plate at the base of the water chamber while  removing the water chamber f
8. ay also have other  rights which vary from region to region    For further information on your warranty rights  contact your local ResMed dealer  or ResMed office        RO01 307 2 05 06    Limited Warranty    English    23    24    
9. e HumidAire 2i  These  solutions may cause damage and reduce the life of the product    e bleach  chlorine   alcohol   or ammonia based solutions     moisturizing  antibacterial  or glycerine based soaps   e water softening and descaling agents   If using a dishwasher  avoid heavy duty or high temperature cycles  greater  than 158  F   Do not use strongly alkaline detergents and rinse aids     DOCKING STATION  Wipe the outside of the docking station with a damp cloth  Do this periodically     Reassembling the Water Chamber  When all the water chamber parts are clean and dry  reassemble the water  chamber     1  Place the cross member on its supports inside the water chamber base     Notches       Supports    Supports       Cleaning and Maintenance 15    16    2  Fit the seal onto the rim of the water chamber base  Press along the edge to  ensure that the seal is fitted properly           3  Fit the lid onto the water chamber base  Press the lid and base firmly  together to close        Note  If the side catches become detached from the water chamber base   reattach them as shown        The side catches consist of two parts  If they become separated clip them back  together as shown  Insert the smaller end  a  of the hinge  b  first  hold it flat   and press firmly into place        Servicing    This product  HumidAire 2i  should be inspected by an authorized ResMed  Service Center 5 years from the date of manufacture  Prior to this  the device is  intended to provide safe and
10. e you start treatment     Note  The device automatically detects the presence or absence of the  HumidAire 2i  There is no need to enter the menus to change the humidifier  setting     2 Pre heat the water before starting treatment by activating the Warm Up Feature   Skip this step if you do not wish to use the Warm Up feature     Warm Up Feature  not available in all models    The Warm Up feature allows you to pre heat the water in the humidifier prior to  starting treatment  The result is that the air is comfortably humidified at the start  of treatment    You can begin using the device at any time during the humidifier  Warm Up   period  The  Warm Up  feature will automatically turn off after 30 minutes if  treatment has not started     Operating Instructions 9    10    To activate the Warm Up Feature     1  Depending on your model  press the Right  warmup  key or the Enter key on  the control panel         Right key   warmup     VPAP III ST A    Enter key       2  The Warming screen  which may vary in appearance according to your  model  will be displayed to indicate that the Warm Up feature is activated   ST a    WARMING  menu exit           Humidifier Warming               E                        3  While the HumidAire 2i is in Warm Up mode you can      Start treatment at any time by pressing the Front or the Enter key   depending on your model     e Access the menus by pressing the Left  menu  key  For the  VPAP    Il STA with QuickNav     use the Up Down and Enter ke
11. ed or cracked        The lid may not be on  correctly     Water chamber seal may  be twisted     Side catches are not  engaged properly     The tub plate is incorrectly  fitted     Contact your supplier for a  replacement     Press the lid down firmly   Make sure the seal is seated properly  inside the water chamber     Make sure the two side catches are  engaged properly     Make sure the tub plate is fitted  correctly        Problem    No air flow    Possible Cause    The air tubing is not  connected properly     The air tubing is blocked   pinched or punctured     The CPAP APAP or bilevel  device is not working  correctly     The water chamber is  overfilled     Solution    Connect the air tubing as described in       Preparing For Use    on page 5     Unblock or free the air tubing  Check    the air tubing for puncture    See the Troubleshooting s    English    S     ection in the    CPAP APAP or bilevel device manual     Empty some water from t    he water    chamber  Make sure that the water    level is not above the maxi  level mark     mum water       Water splashing on  your face    The water chamber is  overfilled     The control dial is set too   high and condensation is   forming in the air tube and  mask     Air blowing on the air  tubing  from fans and air  conditioning units  is  causing condensation     Empty some water from t  chamber        he water    Turn the control dial down     Insulate the air tubing  Pla    cing the air    tubing under the bed covers 
12. er docking  station  already attached to the CPAP  APAP  or bilevel device  until it clicks down  into place  Ensure that the water chamber is fully engaged     Docking  station       WARNING   J   Always connect the HumidAire 2i docking station before connecting the  water chamber     To prevent water from entering the CPAP  APAP  or bilevel device  only  move or transport it with the HumidAire 2i attached if the water chamber  is empty and thoroughly dry     8 Connect one end of the air tubing firmly onto the air outlet on the water chamber  and the other end to the mask system     EEG   ask system        Ae tZ    L7                  Air outlet    Air tubing    The HumidAire 2i is now ready for use  To start treatment  see    Operating  Instructions    on page 9     Operating Instructions    Starting Treatment    The HumidAire 2i should be filled with fresh water and attached to the CPAP   APAP  or bilevel device  with the air tubing and mask system connected  See     Setting Up the HumidAire 2i    on page 5     English    1 Switch on the power at the power outlet  Switch the main power switch at the  back of the device ON     Main  power  switch          When the device is turned on a prompt for the Warm Up feature will appear  not  available on all models      AUTOSET SPIRIT  menu warmup    The Warm Up Feature is automatically displayed when the HumidAire 2i is  connected to the CPAP  APAP  or bilevel device  When activated  this feature  allows the water to warm up befor
13. ied in this manual  Connection of other delivery tubes or accessories  could result in injury  or damage to the device    The HumidAire 2i tubing is for single patient use and must not be re used on  another person  This is to avoid the risk of cross infection    Always place the HumidAire 2i device lower than the level of the patient  If it  is placed at or above the level of the patient  heated water may flow down  the air tubing and into the mask    Do not attempt to dismantle the HumidAire 2i docking station  There are no  userserviceable parts inside  Repairs and internal servicing should only be  performed by an authorised service agent    Take care when handling the HumidAire 2i  The water in the water chamber  may be hot    If liquids are inadvertently spilled on the HumidAire 2i docking station  unplug  the CPAP APAP or bilevel device from the power outlet and allow the docking  station to drain and dry before using     The water chamber shou  transporting the HumidA     d not be connected to the docking station when  re 2i in the carry bag  Place the water chamber in a    separate compartment in the carry bag   Make sure that the water chamber is empty and thoroughly dried before    transporting the HumidAi    Do not operate the Humi    re 2i   dAire 2i if the CPAP APAP or bilevel device has a    damaged power cord or plug     There exists the possibili    y that if the device is placed above the level of the    patient and is tipped over  or if the tubing is filled
14. ing and Maintenance  Servicing  Troubleshooting             Technical Specifications      Humidifier Performance  Limited Warranty              Introduction    The ResMed HumidAire 2i    is designed to humidify the air delivered to the  airway during continuous positive airway pressure  CPAP   automatic positive  airway pressure  APAP   and bilevel therapy  Nasal congestion or dryness of the  nose  mouth or throat may be experienced during therapy  These symptoms are  caused by dry climates  cold air temperatures  the airflow delivered from the  CPAP  APAP  or bilevel system and mouth leak  Moist air helps relieve  symptoms of dryness and can make therapy more comfortable  The  HumidAire 2i is designed to form a single heated humidification unit when  attached to a CPAP  APAP  or bilevel unit     Definitions    This manual contains special terms and icons that appear in the margins to draw  your attention to specific and important information     WARNING    Alerts you to possible injury     CAUTION    Explains special measures for the safe and effective use of the device     Note  Is an informative or helpful note     User Owner Responsibility    The user or owner of this system shall have sole responsibility and liability for any  injury to persons or damage to property resulting from       operation which is not in accordance with the operating instructions supplied      maintenance or modifications carried out unless in accordance with  authorised instructions and by autho
15. may help    prevent condensation        Air does not feel  moist    The control dial is set too  low     HumidAire 2i heating  element is not working     Turn the control dial up     Contact your equipment s    upplier        Air too moist    Control dial is set too high     High ambient relative  humidity     Turn the control dial down    Adjust your humidity setting        Troubleshooting 19    20    Technical Specifications    Performance   Maximum heater plate temperature  167  F   Temperature cut out  199  F   Maximum output humidity  95  relative humidity  control dial set to maximum  setting  flow rate 50 L min  ambient temperature of 68  F  ambient relative  humidity of 40       Dimensions  H x W x D    4 8 in x 7 2 in x 8 6 in  docking station and water chamber   Weight   2 Ib  docking station and unfilled water chamber    Water Capacity   To maximum fill line  400 mL    Power Supply   Input range for AutoSet Spirit Unit with HumidAire 2i  100 240 V  50 60 Hz    lt 528 VA  maximum power consumption    Actual power consumption will vary  depending on factors such as temperature  setting  set pressure  the use of accessories  height above sea level and ambient  temperature     Housing Construction  Docking station  Flame retardant engineering thermoplastic    Water chamber  Injection molded plastic    Environmental conditions   Operating temperature   41  F to  104  F   Operating humidity  10 95  non condensing   Storage and transport temperature   4  F to  140  F   St
16. n when possible  Placing the air tubing under the bed covers with you  may help prevent condensation    We recommend that you arrange the device so that any excess condensation  drains back into the water chamber     Symptoms of Dryness Still Present   If you still experience symptoms of dryness  then your humidity setting may be  set too low to be effective  Gradually turn the humidity control dial up over a few  nights until the symptoms disappear    If you are still experiencing symptoms of dryness  even after increasing the  humidity setting  you may be breathing through your mouth  Mouth breathing or  mouth leaks cause drying and irritation of the upper airway  If this is a problem   a full face mask or chin strap may help  Consult your clinician or equipment  supplier for details     Nasal Infection or Illness  If you have a cold or nasal infection  be sure to clean the HumidAire 2i thoroughly  each day  Do not put aromatic oils or decongestants into the water chamber     CAUTION   Do not use additives  in particular aromatic additives  eg  scented oils and  perfumes   They may reduce the humidification output of the device and or  cause deterioration of the water chamber materials     Scaly Mineral Deposits   You may notice a white powder or deposits in the water chamber base  These  are mineral deposits from the tap water  To remove these deposits  rub the  water chamber base gently using a mild washing detergent    To prevent these deposits forming  fill the water ch
17. orage and transport humidity  10 95  non condensing   IEC 60601 1 Classification   Class II  double insulation    Type CF   Electromagnetic Compatibility   Product complies with all applicable electromagnetic compatibility requirements   EMC  according to IEC60601 1 2  for residential  commercial and light industry  environments     Humidifier Performance  The following settings have been tested at 23  C  40  RH ambient conditions                    4  gt 95 18  10  gt 95 14  15  gt 95 13  20  gt 95 12  25  gt 95 12       1  AH  Absolute Humidity in mg L   2  BTPS  Body Temperature Pressure Saturated     Note  The manufacturer reserves the right to change these specifications  without notice     Technical Specifications 21    22    Symbols Which Appear on the Device    A  v                                  IPX        X    ANNA    Attention  consult accompanying documents    Type CF applied part    Class I  equipment    Drip Proof    Hot Surface    Do not touch    Maximum water level    Limited Warranty    ResMed warrants that your ResMed product shall be free from defects in  material and workmanship for the period specified below from the date of  purchase by the initial consumer  This warranty is not transferable                 Product Warranty Period  ResMed humidifiers  ResControl     ResLink     ResTraxx    1 Year   ResMed flow generators 2 Years  Accessories  mask systems  including mask frame  cushion  90 Days   headgear and tubing   Excludes single use devices       
18. rised persons     Please read this manual carefully before use     Introduction    English    Medical Information    Intended Use    The HumidAire 2i is indicated for the humidification of the air delivered from a  ResMed compatible CPAP therapy device  Refer to the relevant device manual  for information regarding its compatibility with the HumidAire 2i    The HumidAire 2i is intended for single patient re use in the home environment  and multi patient re use in a hospital institutional environment  Contact your  equipment supplier or ResMed office if you require further information    The HumidAire 2i is for use only as recommended by a physician     CAUTION  USA ONLY        Federal law restricts this device to sale by or on the order of a physician        Contraindications    The HumidAire 2i is contraindicated for use with patients whose supraglottic  airways have been bypassed  Please also refer to the relevant system manual for  contraindications associated with CPAP  APAP  and bilevel therapy     Warnings and Cautions    The following are general warnings and cautions  Specific warnings and cautions  appear next to the relevant instructions in the manual     WARNINGS    Use the HumidAire 2i only for its intended use as described in this manual   The HumidAire 2i is designed only for use with specific ResMed CPAP APAP  or bilevel devices  Do not use the HumidAire 2i with any other devices    The HumidAire 2i should only be used with delivery tubes or accessories  specif
19. rom the docking station  It may be hot     4 Discard any excess water by turning the water chamber upside down  allowing  the water to drain out of the air outlet     Cleaning and Maintenance    English    13    5 Disassembling the Water Chamber    1  To open the side catches pull out from the lower edge     Side catch       me  Lo    Front tab _       a   AN    2 FN  Ls p RE             3  Separate the water chamber by pulling the two halves apart  Use your thumb  to pull down on the front tab     Front tab    Seal       Base       6 Discard any remaining water     The tub plate does not need to be removed from the water chamber base  If  removed  replace the tub plate by     1  Turning the water chamber base upside down     2  As shown below  position the tub plate so that the open side is facing down   Then firmly press the tub plate down into the water chamber     CD    w plate  4    English       Note  The tub plate will not fit into the water chamber if it is facing up  Before  fitting the tub plate  make sure the O ring is fitted snugly inside the groove of  the tub plate     Cleaning   WATER CHAMBER   Wash the lid  seal  cross member and base in warm water using a mild detergent  or in a dishwasher  Rinse them thoroughly with clean water and allow them to   dry away from direct sunlight  Do this daily or after each use  Inspect the water  chamber for wear and deterioration  Replace any damaged parts     CAUTION   Do not use any of the following to clean any part of th
20. with your left hand  With your right hand  firmly pull  the right side of the front cap out and then up until the whole front cap  comes off   Notes      To replace the front cap     a  Align the air outlet connector inside the front cap to the air outlet on the  device     b  Push the front cap firmly onto the device until it clicks into place     Remember to take the front cap with you when travelling  in case you  need to use your CPAP APAP or bilevel device without a humidifier     Preparing For Use    5    English    2 Separate the HumidAire 2i water chamber from the docking station  To do this   lift the latch at the bottom of the water chamber and pull the water chamber  away from the docking station        3 Slide the humidifier docking station  with the water chamber detached  onto the  device until it clicks into place     Docking station       Notes     The images in this manual are for illustrative purposes only and may look  different from your model    e Under normal operation  there is no need to remove the HumidAire 2i  docking station once it is attached to the CPAP APAP or bilevel device  If you  need to remove it  press the latch underneath the docking station and pull it  away from the device       When removing the docking station  ensure that the water chamber is  removed first     4 Insert the power cord into the socket at the rear of the device  Plug the other end  of the power cord into a power outlet        Power To  cord power  outlet    WARNING    
21. ys to  access the menus      Stop the Warm Up feature at any time by pressing the Right  exit  key   this does not apply to the VPAP III ST A with QuickNav      3 Turn the humidity control dial to the desired setting  The control dial is labelled    as shown in the diagram below  A setting of 1 provides minimum humidification   and a setting of 6 delivers maximum humidification  The setting marked with a  O switches heated humidification off    Initially  set the control dial to 3  If necessary adjust the control dial up or down  to find the most comfortable and effective setting  You can adjust the humidity  setting at any time        Your aim is to have moist air delivered at your mask while avoiding condensation  in the air tubing  Your ideal setting will depend on the temperature and humidity  of the surrounding air    See    Helpful Hints    on page 12 for further tips on the correct humidity setting     Fit your mask as described in the mask user instructions     WARNING    Before starting therapy ensure that you are wearing the mask and that the  mask and air tubing are fitted correctly     English    Press the Front key or the Enter key  depending on your model  on the control  panel  Alternately  if the SmartStart function is enabled  simply breathe into the  mask  Air should begin to flow from the device          VPAP III ST 4    Heater  plate light       Note  During operation  the heater plate light illuminates periodically to indicate  that the heating element is
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
General International 30-460HC M2 Use and Care Manual  Nobo Diamond  LE MANUEL D`UTILISATION  Panasonic WV-SW155 CAD File  MANUAL DE INSTRUCCIONES / INSTRUCTIONS MANUAL  取扱説明書〈詳細版〉 [F-01B]  GPC® 184    02-Italie mode d`emploi 17-05-11  Polycom CX300 R2 Guia do Usuário    Copyright © All rights reserved. 