User Manual for Body Sensor „ARES“
Contents
1. AthenaDiax User Manual for Body Sensor ARES 9 2015 version 3 english CE 0633 AthenaDiax General Information Prior to using the body sensor the user must become familiar with the function of the body sensor through reading this manual carefully Intended Purpose of the Body Sensor The AthenaDiaX body sensor is manufactured to record and store cardiac signals over long periods up to 7 days The body sensor consists of ECG recorder ARES ECG patch ZEUS e ZEUS 7D application up to 7 days e ZEUS 2D application up to 2 days ZEUS XDA application up to 7 days with monitoring extension Cardiac signals are detected and recorded in the following cases e Paroxysmal atrial fibrillation to detect strokes or for monitoring after ablation e Assessment of cardiac arrhythmia e Diagnosis of cardiac genesis especially for patients at risk e g after a stroke syncope Safety and Patient Information for Use e The body sensor should not be stored or used nearby strong electronic and electromagnetic fields e g X ray This can cause errors in the recording e The body sensor should not be used when computer tomography or magnetic resonance systems MR or MRI are performed e Only original accessories should be used e The body sensor should not be used while swimming and in the sauna e The body sensor casing should not be exposed to strong hits After a strong hit to the casing the product should not be used any
2. electrode is a single use product and it should be only used for a maximum recording of 24 hours Make sure that the ECG electrodes are correctly installed A clicking sound is heard After sticking the body sensor to the chest the signal recording and storage will start automatically Function Control With proper signal recording and storage the green LED blinks every 2s for a short time Signal Control It is possible to activate this status indicator while recording Therefore the body sensor needs to be firmly padded twice End of the Recording Pull off the body sensor from the skin slowly and carefully Removing rapidly can cause skin irritation First loosen gently an easy accessible point of the body sensor from the skin then remove the device from the thorax evenly Disconnect the ECG recorder ARES from ECG patch ZEUS Send the ECG recorder ARES to your doctor for analysis of the recorded ECG Screening and Analysis of the ECG data Connect the reader IRIS to the PC via the USB cable Place the ECG recorder ARES into the reader Make sure that ARES is oriented correctly An evaluation of the recorded signals will to be done by trained personnel AthenaDiaX GmbH Fehrbelliner Str 38 16816 Neuruppin Germany 49 0 3391 739 27 77 www athenadiax com E Mail info athenadiax com AthenaDiax Care and Maintenance e The body sensor should only be cleaned with a slightly damp cloth Use a non agg
3. Notice the technical data limits for the ambient conditions temperature pressure and humidity for storage transportation and operation of the body sensor Use the body sensor only within these limits e Portable and mobile RF communications equipment can affect the function of the body sensor The body sensor should be installed and used according to the information on electromagnetic compatibility e Magnetic and electric fields or ionizing radiation can affect the function of the device Therefore do not use the body sensor nearby equipment with high electromagnetic fields or ionizing radiation such as X ray apparatus or diathermy e Repairs should be carried out only by authorized personnel under the responsibility of AthenaDiax e lf the device is maintained and handled by the user improperly AthenaDiaX has no liability e Showering is with ZEUS XDA not allowed It is only allowed with ZEUS 2D and 7D e ZEUS 2D with monitoring extension must be used only for one patient e Maximum time of monitoring with ZEUS 7D ZEUS XDA ZEUS 2D with monitoring extension is 7 days 168h and for ZEUS 2D without monitoring extension 2 days e After start of recording the body sensor must not be without skin contact for more than 1 hour The ARES stops recording after 1 hour without skin contact Classification The body sensor is a medical device class lla Type BF according to EN 60601 1 and EN 60601 2 47 It is a portable device with inte
4. more e The body sensor should not be used on open heart e The evaluation of the results should be made by qualified personnel e The ECG recorder ARES should be checked prior to each use for visible damage Defective devices or accessories should not be used on patients AthenaDiaX GmbH Fehrbelliner Str 38 16816 Neuruppin Germany 49 0 3391 739 27 77 www athenadiax com E Mail info athenadiax com A AthenaDiaX e Avoid direct sunlight on the body sensor during the wearing time e The compliance of the safety requirements may be guaranteed only for AthenaDiaX products and product configurations e The body sensor can be operated together with pacemakers or other stimulators when all participating devices are used according to their intended purpose However automatic ECG analysis may not be possible in this case e The device can be damaged by using external defibrillators A risk to the patient does not exist e Incase of malfunction send the device to an authorized AthenaDiaX representative e The patient should take off the device if a physical problem e g severe skin irritation occurs during the recording e Extreme cold and heat should be avoided e To obtain correct measurement results the patient should pay attention to an intact adhesive joint between patch and body e The body sensor is not suitable for direct monitoring the clinical status of patients because it has no signal display or alarm functions e
5. ressive solution for disinfecting the body sensor e Do not use abrasive materials to clean the device e The body sensor should not be cleaned with organic solvents such as benzene or ether e Prevent penetration of liquids do not spray e Never insert the device into liquids disinfectants Service and Repair Unauthorized repairs or modifications of the body sensor s parts or accessories can cause impairment of the function or hazards to the user or patient Therefore repairs should be carried out only by the manufacturer or by an authorized person Illegal opening causes loss of warrant For service or repair contact your distributor Disposal e The ECG recorder ARES and the reader IRIS should be properly disposed as old electrical equipment and the ECG recorder should never be disposed to the normal household waste e The ECG patch contains a lithium battery Improper disposal can cause environmental hazards Please act in accordance with local or national regulations for disposal of batteries Self tests After connecting the ECG recorder to the patch and during operation of the body sensor a comprehensive functional test is performed automatically All necessary functions are checked If a result of a self test is negative the ECG recording will not start and the body sensor displays the error status This condition is indicated by a red LED for a period of about 1 hour Subsequenily the body sensor switches off it
6. rnal power supply for continuous operation Product Liability The manufacturer of the body sensor accepts product liability under the following conditions e lf the body sensor was operated only with original accessories AthenaDiaX GmbH Fehrbelliner Str 38 16816 Neuruppin Germany 49 0 3391 739 27 77 www athenadiax com E Mail info athenadiax com AthenaDiaX lf repairs of the body sensor and its accessories are performed exclusively by the manufacturer or by persons trained and authorized by the manufacturer lf this manual was considered in using the device If the safety checks were carried out Start of Signal Recording If necessary shave the respective chest area Clean and dry the skin For this purpose alcohol pads can be used Assemble the ECG recorder ARES and ECG patch ZEUS A clicking sound is heard Check to see whether all sides of the ECG recorder are snapped into the retaining rails of the ECG patch ZEUS To achieve proper assembly notice the mark on the ECG recorder and patch Both lines 1 and 2 must be properly aligned to each other Take off the liner from the ECG patch ZEUS and stick the pre assembled body sensor on the chest Place the body sensor to the highlighted chest area in the figure below optimal place Only for ZEUS 7D After affixing the body sensor remove the protective foil from the top of the ECG patch Only for ZEUS with monitoring extension The monitoring extension ECG
7. self Internal Monitoring and Protective Measures The body sensor includes internal protection measures e ESD protection according to EN 60601 2 47 6 kV for contact discharge 8 kV air discharge e No defibrillation strength according to DIN EN 60601 1 e Dielectric strength test voltages applied part according to DIN EN 60601 1 and DIN EN 60601 2 47 e Leakage currents applied part according to DIN EN 60601 1 and DIN EN 60601 2 47 AthenaDiaX GmbH Fehrbelliner Str 38 16816 Neuruppin Germany 49 0 3391 739 27 77 www athenadiax com E Mail info athenadiax com AthenaDiaxX e EMC according to EN 60601 1 2 Support For questions and problems with the body sensor and accessories please contact us at AthenaDiaX GmbH AthenaDiaX GmbH Fehrbelliner Str 38 SaalmannstraBe 9 16816 Neuruppin 13403 Berlin Germany Germany E Mail Info athenadiax de E Mail Info athenadiax de www athenadiax de www athenadiax de 49 3391 739 27 77 49 30 692 067 950 AthenaDiaX GmbH Fehrbelliner Str 38 16816 Neuruppin Germany 49 0 3391 739 27 77 www athenadiax com E Mail info athenadiax com
Download Pdf Manuals
Related Search
Related Contents
Impressora Zebra Série XiIII Guia de Segurança e Referência Rápida Vitalvision Istruzioni seratec ® baby-time Fisher SRC-850A User's Manual Samsung RT29FBRNDSP User Manual User Manual • • Lana • • • Cotone • • wool • • • Cotton • • • Laine • • • Coton Téléchargez le carnet de stage Copyright © All rights reserved.