Mosaic User`s Manual - Keck 3D Fusion Microscope
Contents
1. 4 72 Goto Step sje 0 104908 0 The Positions Control P4 ie 71 0 1 nee Move Up P5 51364 104908 0 Move Down This is the window s primary control It R emove contains the list of stage positions that will be visited and imaged during the selected __Remove all time lapse and a set of buttons for working with them Stage positions are m Mosaic Pattern always visited starting from the top of the xo list and going to the bottom the Magnification 4 Overlap Percentage i P Yjo numbering of the positions is irrelevant i e if PS is higher in the list than PO Generate Raster Generate Snake P5 will be visited first not P0 JV Enable Mosaic Patterns Stage Controls Add Position adds the current X Y stage a coordinates to the end of the list Scales Logarithmic Z ese 9 mm sec For convenience this action can also be invoked by pressing the ENTER RETURN key Acceleration mm sec Goto Step moves the stage to the F Tipwhs currently highlighted position in the i m positions list and then highlights the next position in the list After the stage starts moving this button changes to Stop to Accept Cancel allow the user to stop the stage movement For convenience this action can also be invoked by pressing the SPACE key Pressing SPACE starts the movement pressing it again while the stage is moving stops the stage
2. Move Up and Move Down move the selected position up or down in the list Allows one to re arrange the order of stage positions Remove deletes the selected position from the list This can also be invoked by pressing the DELETE key Remove All deletes all positions from the list Additionally the Positions Control allows the stage to be moved using the dedicated arrow keys on the keyboard i e not the ones on the numerical keypad The stage s speed is controlled via the Max Speed slider in the Stage Controls section o To use the arrow keys to move the stage keyboard and mouse focus must be set to the Positions control This simply means either clicking on a position in the list or just clicking anywhere in the positions list if there are no positions The Mosaic Pattern Control e This control is used to generate the raster or snake mosaic patterns Refer to the section Creating a Pattern Mosaic for details The Stage Controls e Max Speed is the speed in millimeters per second the stage will move at This affects the stage s speed when moving using the keyboard arrow keys and during the time lapse The Acceleration parameter controls how quickly the stage accelerates in millimeters per second squared to and from its target speed Max Speed When the stage is instructed to move to an XY position it will accelerate at the specified rate until it reaches the target speed It will th
3. will need to see a live real time display of your sample in order to work with it o If you are acquiring brightfield DIC fluorescence or epi fluorescence images start the Spot Advanced program from its shortcut on the Keck desktop Then press the Live button on the program s toolbar on the right side of the program window If you are acquiring confocal images i e single photon two photon reflectance start the Confocal Subsystem program from its shortcut on the Keck desktop Then press Grab and the live display of your sample will start momentarily After you have established a real time display of your sample using either the Spot Advanced or the Confocal Subsystem programs click back to the Mosaic Configuration window and move it such that both it and the live image display are visible You will also need to move the Time Lapse Series window so that it doesn t obstruct your view of the real time display You may now begin working with your sample Use the arrow keys on the keyboard to move the XY stage You will see movement in the image display as you re moving the stage You may also use the joystick located to the right side of the microscope to move the XY stage Creating a Manual Mosaic Click Enable Mosaic if you haven t already done so Prepare a real time display of your sample as outlined in the section Preparing to View Your Sample in Real Time
4. Leave the Enable Mosaic Patterns box unchecked Now use the arrow keys and or joystick to move the stage Find your objects of interest using the live image display then click the Add Position button or press ENTER to add its stage coordinates to the list Remember to test your mosaic setup before starting the time lapse Use the Goto Step button to step through your positions to ensure the stage moves using reasonable motions for what you will be imaging During the time lapse the images are NOT assembled or stitched together since this mode of operation is not created to image contiguous regions of a sample Creating a Pattern Mosaic Click Enable Mosaic if you haven t already done so Check the Enable Mosaic Patterns box This will enable the Mosaic Pattern control Prepare a real time display of your sample as outlined in the section Preparing to View Your Sample in Real Time With pattern mosaics the user does not specify the actual stage positions they are computed by the software In order to compute a snake or raster pattern however you must provide two stage positions the upper left and the lower right boundaries of the region of interest Move the stage using the keyboard arrow keys and or the joystick and add those positions to the list Enter the magnification you are using and press Generate Raster or Generate Snake depending on which pattern of movement
5. Using Mosaics in the Keck 3DFM Time Lapse Software Rafal Norton Keck Software Developer rnorton ece neu edu Optical Science Laboratory Center for Subsurface Sensing and Imaging Systems CenSSIS Northeastern University Date Created January 17 2008 Overview of the Mosaic Capabilities For the purposes of this document a mosaic is essentially a sequence of X Y stage coordinates that have been either specified by the user or computed by the software The 3DFM Keck microscope is currently equipped with a Prior Pro Scan II motorized translation stage which is what performs the actual mosaics The mosaic features are available for all of the current time lapse modes The mosaic configuration window is invoked by clicking on the Mosaic menu in the Time Lapse Series application and selecting Mosaic Configuration During the selected time lapse the stage visits each of these positions in the order that they appear in the Positions list from top to bottom At each stage position the appropriate imaging operations for the selected time lapse are performed and after all the positions in the list have been visited this constitutes one round e When using mosaics with any of the slice time lapse modes the Number of images parameter is the number of mosaics to be taken so instead of a single image a group of images representing a mosaic is taken o For example if 10 images are entered in t
6. en decelerate from the target speed when it is halfway to its destination coordinates o To disable acceleration set the Acceleration slider to 100 When acceleration is disabled the stage will immediately start moving at its target speed Max Speed When moving the stage using the arrow keys only the Max Speed parameter is in effect since it is not known at what stage coordinates movement will stop The S Curve parameter provides fine tuning of the stage acceleration It controls the rate of change in millimeters per second squared from the time the stage is stationary to the time the stage reaches the acceleration rate specified by the Acceleration slider Important Use these parameters carefully and reasonably especially when imaging aqueous samples Sudden stops and motions can cause the sample or parts of it to move or drift thus degrading the quality and accuracy of the mosaic Using acceleration and a low target speed can minimize the drifting of aqueous material The actual stage motions can always be observed by having a real time display of your sample and stepping through each individual position If the stage motions are too great one can make adjustments and test them before starting the actual time lapse When viewing a live display at higher magnifications the P minimum stage speed of 1 mm sec may still create too large a movement In that case use the joystick located to the righ
7. he Number of images entry box this will be 10 mosaic rounds The time lapse will visit and image each position in the list from top to bottom wait the specified amount of time then will visit and image the positions again from the top of the list down This will repeat until 10 rounds have been completed When using mosaics with any of the z stack time lapse modes the Number of steps which is normally the number of images to take for each z stack becomes the number of mosaics to be taken for each z stack The software currently provides two mosaic types manual and pattern mosaics e A manual mosaic is really a collection of arbitrary user defined positions that will be visited and imaged in the order in which they appear in the list These are useful when one is only interested in imaging discreet arbitrary regions of a sample Refer to the section Creating a Manual Mosaic for instructions A pattern mosaic will image a contiguous region of a sample in either a raster or a snake pattern Refer to the section Creating a Pattern Mosaic for instructions Overview of the Mosaic Controls Mosaic Configuration i x Select Enable Mosaic to begin using the Disable Mosaic controls of the Mosaic Configuration Qe window The current X Y stage Beers coordinates will be inserted into the ee i 934s PO 51364 106572 j Add Position Positions list P1 49171 0 106572 P2
8. t of the Keck microscope to move the stage instead of the keyboard arrow keys Preparing to View Your Sample in Real Time e Make sure the appropriate shutters are open otherwise you will not see your sample Start the Shutter Control 1 1 program from its shortcut on the Keck desktop o For one photon two photon or reflectance imaging i j ae sen te 55 CONFOCAL SPOT open the Confocal shutter by clicking on its Open button as shown to the right Open peh Closed Closed For brightfield DIC fluorescence or epi fluorescence imaging open the Spot shutter by clicking on its Open button as shown to the right 137 99 212 iss Position Position You will also need to move a mirror into or out of the optical path depending on what kind of imaging you will be doing You will see a rack with two mirrors directly below the microscope objectives Start the Zaber Positioner program from its shortcut on the Keck desktop o For one photon or reflectance imaging click on the Hot Mirror button You will hear the mirror rack moving Wait for it to move into place For two photon imaging click on the Cold Mirror and wait for it to move into place For brightfield DIC fluorescence or epi fluorescence no mirror must be in the optical path Click on the Open button and wait for the rack to move into place Start the appropriate live viewing application You
9. using reasonable motions for what you will be imaging Each mosaic will be assembled and placed into the folder named stitched located in the same folder as your image files It is highly recommended to also perform a test run first before starting the time lapse o For slice time lapse operations enter I for the Number of images and 0 for the wait time parameters For z stack time lapse operations specify I for the Number of z stacks I for the Number of steps 1 for the Step size and 0 for the wait time parameters Start the time lapse After it has finished examine the assembled mosaic image located in the stitched folder to be sure the result is acceptable before starting the time lapse
10. you prefer o It is generally recommended to use the Snake pattern especially when imaging aqueous samples The snake pattern uses smaller movements which reduce the drifting of aqueous liquid material while the mosaic is being performed The stage motions of the snake and raster patterns are illustrated below using a 3x3 mosaic EAE HAE 456 65 4 7 819 7 819 Raster Pattern Motion Snake Pattern Motion You may optionally specify overlap percentages in the X and or Y directions Overlap controls how much of an image appears in the next one in the mosaic sequence and is used to compensate for movements of the sample that may have occurred while the mosaic was being performed The software will now generate the stage positions that will image the region of interest using the selected pattern and will place them in the Positions list Important Do not edit the generated position list in any way Doing so will cause incorrect assembling of the images in the mosaic sequence Do not delete add any new positions or rearrange the order of the generated list To obtain a new list of positions using different parameters and or a different region of interest you must first delete the existing ones via Remove All then repeat all the steps in this section Remember to test your mosaic setup before starting the time lapse Use the Goto Step button to step through your positions to ensure the stage moves
Download Pdf Manuals
Related Search
Related Contents
KYOCERA FS-1128MFP Plexor(TM) Systems: Instrument Setup and Data Analysis for the Bio 1982 , Volume , Issue Dec-1982 Multitech Modem MT9234ZBA-USB User's Manual Martin Steinmann. S. Giedion. la mécanisation de la maison et "la "user manual" Graco XTR Airless Spray Guns User's Manual 53HSR (10-13) Mode d`emploi Instrukcja obsługi ECS-7700/7100 Copyright © All rights reserved.