         Contents
1.                                                                                                                                             Figure 7       Technical manual Planmeca Intra x ray unit 33    PARTS REPLACEMENT  amp  REPAIR       e  Remove the eight M4 DIN 934 attachment nuts and 94 3 DIN 6798 washers from the  Generator PCB  Fig  8  1   Unscrew the two M4x8 DIN 84 screws  Fig  8  2   Lift the  generator assembly away from the wall adapter and open the two M3x6 DIN 912  screws located behind the generator assembly frame  Fig  8  3      f  Remove the Generator PCB        M4 DIN 934 nut  and M4 washer                                           ps                                                                                                                                                                       _tech_4_3APSU2 e     JE  if                   Figure 8  g  Install the new Generator PCB in reverse order  Note  that the new Generator PCB is  not attached to the frame with the two M3x6 DIN 912 screws located behind the  generator assembly frame        34 Planmeca Intra x ray unit Technical manual    PARTS REPLACEMENT  amp  REPAIR       7 2 Replacing the tube head PCB    a  Unscrew the two M4x20 DIN 7984 fastening screws  Fig  9  1  of the cone  Fig  9  2    inside the cone      b  Remove the tube head front cover  Fig  9  3  and the cone    c  Disconnect the two connectors from the tube head PCB  Fig  9  4    d  Pull the tube head PCB  Fig  9  5  from the tube
2.       Figure 11    j  Join the old and new cables together as shown in Fig  12 below  Tape the joining point  firmly  Do not tape the old Dixi interconnection cable  it is pulled out from the opening    on the bracket arm  see figure Fig  13     Old Dixi interconnection cable       Grounding lead of the new cable    ps    Grounding lead of the old cable    Tech_2A e          Figure 12       38 Planmeca Intra x ray unit Technical manual    PARTS REPLACEMENT  amp  REPAIR       k  Coat the new arm cable with silicone  grease or similar  Pull from the tube head end of  the arm cable and simultaneously move the arm up and down        Old Dixi interconnection cable                New arm cable    Figure 13    NOTE Make sure that the arm cable and the Dixi interconnection cable are  not twisted when you are pulling them into the arm     l  Stop pulling when the new cable is 390 400 mm out from the opening of the tube head  support  Fig  14  1   Route the cable through the opening on the support frame  Fig  14   2         390 400 mm          Figure 14       Technical manual Planmeca Intra x ray unit 39    PARTS REPLACEMENT  amp  REPAIR       m  Strip approx  11 mm of the outer jacket of the cable  Slide the shield onto the  interconnection cable  Align the leads and insert them into the RJ45 connector in the  order shown on the Fig  15 below     Colour according EIA TIA 586A                1 Green   White Green   White 1  2 Green Green 2   gt  3 Orange   White Orange   White 3    4 B
3.      AN CAUTION    ELECTRICAL SAFETY RULES   The unit contains hazardous voltages  While servicing internal parts   always turn off externally the power to the unit  and wait for 2 minutes  before touching any electrical parts    Always replace the fuses with ones of the same type and rating   Otherwise patient  operator or equipment safety cannot be  guaranteed    The circuit boards can be damaged due to static discharges and  require careful handling     IN CAUTION    GENERAL SAFETY RULES   The unit must be serviced only by qualified personnel  trained by  PLANMECA  Repairs and parts replaced by unqualified personnel  carry no warranty    Periodical maintenance as described in this manual must be  performed on a regular basis  to ensure the safety and image quality  of the unit    Some procedures described in the unit could be dangerous  if not  followed as stated        Technical manual Planmeca Intra x ray unit 1    GENERAL  amp  TECHNICAL DATA    AN CAUTION    Check that the X ray unit is installed properly and no mechanical play  caused by wear  corrosion  metal fatigue or ageing can be found  between the wall bracket and horzontal arm        1 2 Manual versions    Planmeca pursues a policy of continual product development  Although every effort is made  to produce up to date product documentation this publication should not be regarded as an  infallible guide to current specifications  We reserve the right to make changes without prior  notice     NOTE This manual is o
4.    PLANMECA    technical manual    TABLE OF CONTENTS       1 GENERAL  amp  TECHNICAL DATA       ssssssssssssss eee ennenen eee 1  Ti Warnings  aind    Cautions  eria e a dea deaa bi E E 1  1 2    Manual Versions  ida ci 2  1 3 Technical Specifications T 3  1 4 User s statement for Planmeca Intra esse sees sese ee esse ee eee ee eree eee neee enner 5  1 5  SEMG information acusa A vie daa Abe adie iene  8  2    SERVICE MODE HHH 12  2 1  Control panel neces  ciseccsetet evens peeves cece A e bt E aaaea 12  2 2 Howto enter exit the Service mode sese sees eee eee eee 12  2 3 X ray tube filament preheating voltage Calibration eee eee eee eee eee eee eee eee 13  2 4 KV FANGS Selection eedonia tt ld dd eed vaca eves seid ave doe 13  2 5 Operation of the preprogrammed settings keys  KV hold  sese eee eee 14  2 6    Dimming he displays  ordinario AE A A tay eda haca ca cada aaa edad 15  2 7 Duration of the displays dimming time out o oo    eee eee eee 16  2 8 Disabling the exposure Key    ooooooccconnoccccconononcccnnonanncnnnnnnencnnnnnnn cnn rre name rene recen erre nena rr nene nn 17  2 9 Automatic identification of the cascade Card sss sees ee sese e sees e seene 18  2 10 Specifying the PCB versions sse eee eee eee 18  2 11    Ready State seting unica ind ptos 19  3 RECALLING THE FACTORY PREPROGRAMMED EXPOSURE  VALUES A as 20  TROUBLESHOOTING     ninas 21  ERROR  MESSAGES   cui essen 22  5 1 Error message shortform table T 22  5 2 Detailed error message Explanations esse sese sese
5.   generator PCB  replace in this order   The possible reason is also faulty feedback cable  9  pole  in the tube head     X ray tube anode voltage  kV  dropped suddenly    The x ray tube voltage suddenly drops  and a knocking sound is heard from the tube head at  the same time  The exposure is aborted and this error indicated  This phenomenon should be  considered quite normal if it doesn   t occur frequently  If the generator is damaged during the  exposure  or the arm extension cable is broken  this error is indicated and during next expo   sure the error E 30 occurs     If this error occurs constantly it is probably caused by a faulty tube head or tube head PCB or  generator PCB  replace in this order   The possible reason is also faulty feedback cable in the  tube head     E 12 X ray tube cathode filament preheating voltages are not calibrated    The tube head PCB has been replaced  but the X ray tube cathode filament preheating volt   ages are not calibrated     See paragraph 2 3    X ray tube filament preheating voltage calibration    on page 13 for details  how to perform the calibration        Technical manual    Planmeca Intra x ray unit 23    ERROR MESSAGES       E 13 Filament preheating voltage calibration failed    During preheat calibration the filament voltage is measured to be in specified limits  The fila   ment circuit in the tube head or in the tube head PCB is faulty or the tube head is faulty     Perform the calibration procedure again  If it fails  replac
6.   s 12V voltage feed or the communication between the control panel and  tube head CPU is failed  The arm extension cable or control panel telephone cable is faulty     Check the cables  Measure the 12V voltage at the tube head PCB P2 connectors pins 1 and  4  Check if the red LED D7 on the tube head PCB is on  Replace the tube head PCB  control  panel or the generator PCB  in this order         26 Planmeca Intra x ray unit Technical manual    ERROR MESSAGES       E 71 FLASH memory check sum error  tube head CPU     Tube head CPU internal error   Replace the tube head PCB     E 81 EEPROM memory defective  tube head CPU     Tube head CPU internal error   Replace the tube head PCB     E 83 Config register error  tube head CPU     Tube head CPU internal error   Replace the tube head PCB        Technical manual Planmeca Intra x ray unit 27    MECHANICAL ADJUSTMENTS       6 MECHANICAL ADJUSTMENTS    6 1 Adjusting the balance of the arm    Adjust the balance of the arm by turning the adjustment nuts with a screwdriver  The adjust   ment nuts are located inside the bracket arm and can be reached through the openings at the  under side of the bracket arm                    T    Openings Tor  adjustments                                  Arm in adjustment position    NOTE Adjust the arm from the lower part of the  adjustment nut        WRONG    Figure 1       28 Planmeca Intra x ray unit Technical manual    MECHANICAL ADJUSTMENTS       6 2 Adjusting the bracket arm angles    In case th
7.  Do not connect any other equipment to the terminal of the generator             box   Figure 22       44 Planmeca intra x ray unit Technical manual    DIAGRAMS       8 DIAGRAMS       Technical manual Planmeca Intra x ray unit 45       Tube head assembly                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        1  1  1  PLANMECA  gt   113 10 13 1  1  K 4800 Z AD LAL   5 Z a aN   xx N  Z2362a  NQryorssceg 1  aim cette Re pZ head dl ve    YELL 1  Dx x H 1  asar RED    RED 1  RED 1  RED    RED    RED    RED 1  RED    RED i       Remote  control      gt  A fuss  Arm cable  Feedback cable  Tubehead grounding o         Code 10006682 Code 6320001 wire  Code 6320013 g 2  323  Es  i 5  gt   DD  PLANMECA E   113 10 11 Oe     S I  Generator assembly B 1  ET  P1  Main switch Main switch 3 5  cable  brown cable  blue PLANMECA  lt   Code 6320005 Code 6320004 Mains voltage selector P14  113 10 17 cable 220 240 V  See  P5 Code 6113005 S  BLU NI BLU p3 P4 a6  BAN LD  Tea   Me k  IAINS 8 E  INPUT E 8  Z  BLU Mains voltage sel
8.  and connect the feedback cable to the tube  head PCB terminal P1     s  Connect the arm cable leads to the connectors as described in Fig  20 below     Red  Black  thin          White       Black     S   Blue G   D   Yellow                gt         AAA                       OOOO             100                   HHIH    aan eal                 IN          Figure 20       42 Planmeca Intra x ray unit Technical manual    PARTS REPLACEMENT  amp  REPAIR       t    Connect the arm cable to the tube head PCB and connect the grounding lead to the  grounding point near the support frame opening     Arm cable                Grounding lead of arm cable       Tube head grounding lead Feedback cable  Figure 21    u  Check that the arm cable remains in the groove of the tube head support when moving  the arm from one extreme position to another  Secure the arm cable to the support  frame with a cable tie     v  Pull the cables through the extension arm and connect the cables to the generator  assembly connectors     w  Replace the removed covers and plugs        Technical manual Planmeca Intra x ray unit 43    PARTS REPLACEMENT  amp  REPAIR       7 5 Replacing the control panel cable    a  Press the clip of the connector  Fig  22  1  and pull the socket from the terminal of the  generator box  Fig  22  2     b  Detach the connector from the control panel in the same way as from the generator  box    c  Connect the new cable to the generator box and to the control panel     AN CAUTION   
9.  ee eeee esse ee eee renee 23  6 MECHANICAL ADJUSTMENTS    sss sese essen eenn 28  6 1 Adjusting the balance Of the arm ooo    cece eee eee eee 28  6 2 Adjusting the bracket arm angles esse sese essere essere essere enen 29  6 3 Adjusting the stiffness of the tube head   s horizontal axle sees 30  6 4 Vertical tube head stiffness adjustment SCreW sees eee eee eee eee eee 31  7 PARTS REPLACEMENT     REPAIR         cccccccsssssssssseseeeeeeeeeesseeeeeeseeeess 32  7 1 Replacing the Generator PCB see eee eee eee 32  7 2 Replacing the tube head PCB sse eee eee eree 35  7 3 Replacing the tube head asses sese sees sees sees ereenn 36  7 4 Replacing the arm e 37  7 5 Replacing the control panel Cable oe e essere essere enserre ennenen 44  3 DACARA MS ar aaa naaar eaa aa ramaan Aaaa AAA KA ain aaa 45    COPYRIGHT PLANMECA 2007 10  PUBLICATION NUMBER 10006491 revision 12       Technical manual Planmeca Intra x ray unit TOC 1    GENERAL  amp  TECHNICAL DATA       1 GENERAL     TECHNICAL DATA    1 1 Warnings and cautions    AN WARNING    THE FOLLOWING WARNINGS  CAUTIONS AND NOTES MUST  ALWAYS BE CONSIDERED WHILE SERVICING THE UNIT  IN ORDER  TO AVOID EITHER PERSONAL INJURY OR DAMAGE TO THE UNIT     AN CAUTION    RADIATION SAFETY RULES   Some procedures described in this manual produces X ray radiation   Always follow the rules for radiation protection    Never attempt to open the tube head  It does not contain any  serviceable parts  and radiation safety could not be guaranteed any  more
10.  head    e  Install the new PCB in reverse order     Perform the x ray tube filament preheating voltage calibration  see section 2 3    X ray  tube filament preheating voltage calibration    on page 13                             Figure 9       Technical manual Planmeca Intra x ray unit 35    PARTS REPLACEMENT  amp  REPAIR       7 3 Replacing the tube head    a  Remove the tube head PCB as described in section 7 2    Replacing the tube head  PCB    on page 35     b  Unscrew the six fastening screws of the tube head cover  Fig  10  1  and remove the  cover  Fig  10  2     c  Unscrew the five screws from the support frame  Fig  10  3  and remove the tube head   Fig  10  4   Note  that there is a grounding plate under the lower left attachment screw    d  Install the new tube head in reverse order  Make sure that the grounding plate touches  the tube head PCB    NOTE Do not overtighten the tube head cover fastening screws to avoid  damaging the cover     e  Perform the x ray tube filament preheating voltage calibration  see section 2 3    X ray  tube filament preheating voltage calibration    on page 13           Grounding plate    Figure 10       36 Planmeca Intra x ray unit Technical manual    PARTS REPLACEMENT  amp  REPAIR       7 4 Replacing the arm cables    NOTE The Dixi interconnection cable is attached to the arm cable and you    have to replace both cables  The Dixi interconnection cable is  attached to the new arm cable at the factory     NOTE The x ray units with seri
11.  mm     Rated line voltage  100  110 117  220 240 V   10   Line voltage regulation 10      Maximum line current  6 1 A at 230 V   12 2 A at 115 V     Technique factors that constitute the maximum line current condition  70 kV  8 mA    Generator rating and duty cycle    1 4 kW  duty cycle 1 15  The wait period is controlled automatically by calculating it according  to the formula tw   15 x texp     Maximum deviation of peak tube potential from indicated value   2 0kV    Maximum deviation of tube current from indicated value   10     Maximum deviation of exposure time from indicated value   10     DEFINITION OF MEASUREMENT CRITERIA    Exposure time  The beginning and end points of the exposure time are defined at 70  of the peak radiation  waveform measured with a calibrated x ray monitor    Peak tube potential  Is defined as the high voltage mean value measured with a calibrated non invasive kVp  meter    Tube current    Is defined using the voltage over the feedback resistor measured with a calibrated multimeter   The mA value is calculated by dividing the voltage by the resistance value        Technical manual Planmeca Intra x ray unit 5    GENERAL  amp  TECHNICAL DATA       The nominal x ray voltage together with the highest x ray tube current  obtainable from the high voltage generator when operated at it   s highest x   ray tube voltage    70 kV  8mA  The nominal x ray tube current when operated at the highest x ray tube  voltage   8 mA  70 kV  The x ray tube voltage and
12.  number starts to blink  and the number can now  be changed with the parameter adjustment keys        00    Select key Parameter adjustment keys    C Accept the new setting by pressing the Select key   O  Press the Mode key briefly to exit the service mode                 14 Planmeca Intra x ray unit Technical manual    SERVICE MODE       2 6 Dimming the displays  Enter the service mode according to the instruction given in section 2 2    How to enter exit the    service mode    on page 12   Press the parameter adjustment up key until the parameter number 16 appears on the kV dis     play   16 Y  O  Parameter adjustment up  key  kV    The number indicating the brightness of the displays  from 1 to 5  factory default is 5  is  0 s  shown on the time display                 Press the Select key until the number starts to blink  and the value can now be changed with  the parameter adjustment keys     D O O    Select key Parameter adjustment keys    e amp  Accept the new setting by pressing the Select key   f mone d Press the Mode key briefly to exit the service mode           Technical manual Planmeca Intra x ray unit 15    SERVICE MODE       2 7 Duration of the displays dimming time out    Enter the service mode according to the instruction given in section 2 2    How to enter exit the  service mode    on page 12     Press the parameter adjustment up key until the parameter number 17 appears on the kV dis     play   17 C2 L A Parameter adjustment up  key  kV    The number indic
13.  ray unit is intended for use in an electromagnetic environment in which radiated RF dis   turbances are controlled  The customer or the user of Planmeca Intra X ray unit can help prevent electro   magnetic interference by maintaining a minimum distance between portable and mobile RF  communications equipment  transmitters  and the Planmeca Intra X ray unit as recommended below   according to the maximum output power of the communications equipment                             Separation distance according to frequency of transmitter  Rated maximum output power m  of transmitter  150 kHz to 80 MHz   80 MHz to 800 MHz   800 MHz to 2 5 GHz   W d   124P d   12  P d   23 P  0 01 0 2 0 2 0 3   0 1 0 4 0 4 0 7   1 1 2 1 2 2 4   10 4 0 4 0 8 0  100 12 0 12 0 24 0                      Technical manual Planmeca Intra x ray unit 11    SERVICE MODE    2 SERVICE MODE       2 1 Control panel    PLANMECA    Preprogrammed setting  keys and indicator lights    Adult child selection key  and indicator light       Exposure warning    indicator light READY                     Ready indicator light    PRET       Exposure key          kV display 66 3           HA display  kV mA  Time display 0 250  9 s       Mode key  amp  gt  O                  Select key  Cw  O ________ Parameter adjustment keys    2 2 How to enter exit the service mode     amp  Press and hold down the select key for 4 seconds     Press and hold down the Mode key for more than 2 seconds  until the four uppermost prepro   grammed
14.  setting indicator lights come on     To exit from the service mode    f yon d Press the Mode key briefly        12 Planmeca intra x ray unit Technical manual    SERVICE MODE       2 3 X ray tube filament preheating voltage calibration       Enter the service mode according to the instruction given in section 2 2    How to enter exit the  service mode    on page 12     Press and hold down the adult child mode selection key for 2 seconds or until the indicator  light starts to blink  The indicator lights will start to blink indicating that you are in the preheat   ing voltage calibration mode     Move as far away from the x ray tube as the length of the cable from the control panel per   mits     Press and hold the exposure key on the control panel until 16 exposures are performed   approx  30 seconds      Press the Mode key briefly to exit the service mode     2 4 kV range selection                Enter the service mode according to the instruction given in section 2 2    How to enter exit the  service mode    on page 12     Press the parameter adjustment up key until the parameter number 14 appears on the kV dis     play   14 L   O  Parameter adjustment up  key  kV    The code of the kV range  0   9  is shown on the time display  The kV ranges are  0   50 70   1   55 70  2   60 70  3   66 70  4   70  5   50 68  6   55 68  7   60 68  8   66 68 and 9   68  kV     Press the Select key until the kV range code starts to blink  and the range can now be  changed with the parameter 
15.  temperature will appear on the time  display  The control panel does not operate  Wait until the temperature drops              Technical manual Planmeca Intra x ray unit 21    ERROR MESSAGES       5 ERROR MESSAGES          E 11 The error code is displayed on the time display     S          CS Press the select key to clear the error from the display     5 1 Error message shortform table                                                          ERROR CODE ERROR MESSAGE EXPLANATION   E DO Exposure key was released too early during the exposure    E 10 X ray tube Anode voltage  kV  overshoot    E 1l X ray tube Anode voltage  kV  dropped suddenly    E 12 X ray tube cathode filament preheating voltages are not calibrated    E 13 Filament preheating voltage calibration failed    E 29 Membrane keyboard key short circuited pressed during the self test or faulty  display board    E 30 kV value does not reach or it exceeds the given value  difference more than  5      E 3  X ray tube Anode current  mA  missing  or not in specified limits    E 33 X ray tube Filament voltage  V  missing  or outside the range  too low or too  high     E 3U X ray tube Anode voltage  kV  missing  or below the specified limit    E 36 Too long exposure    E 37 kV feedback signal open circuit or short circuit    E 38 mA feedback signal open circuit or short circuit    E 50 Tube head temperature sensor short circuit    E 51 Tube head temperature sensor open circuit    E 52 Filament voltage feedback not in speci
16.  tube current which result in the highest electric  output power   70 kV  8 mA  The nominal electric power for a load time of 0 1 sec and at the nominal x   ray tube voltage   1 4 kW at 70 kV  8mA    Anode heating cooling curve of the X ray tube            y PENNEN  K  Ese   PST    TIME  min     HEAT STORAGE  kJ        6 Planmeca Intra x ray unit Technical manual    GENERAL  amp  TECHNICAL DATA       X ray tube assembly heating cooling curve       X ray tube assembly heating cooling curve    140    120    100    Heat Storage  kJ   o     o Oo    E  o    N  o       0 30 60 90 120 150 180 210 240 270 300    Time  min                 Technical manual Planmeca Intra x ray unit 7    GENERAL  amp  TECHNICAL DATA       1 5 EMC information    AN WARNING    Use of any accessories and cables other than those specified in  Planmeca Intra X ray unit   s documentation  with exception of cables  sold by Planmeca as replacement parts for internal components  may  result in increased emission or decreased immunity of the X ray unit     AN WARNING    Planmeca Intra X ray unit should not be used adjacent to or stacked  with other equipment  If adjacent or stacked use is necessary  the  Planmeca Intra X ray unit should be observed to verify normal  operation in configuration which it will be used        Guidance and manufacturer   s declaration   electromagnetic emissions       Planmeca Intra X ray unit is intended for use in the electromagnetic environment specified  below  The customer or the 
17.  tube head feedback cable  Replace the generator PCB  tube head PCB or tube head   replace in this order      E 31 X ray tube anode current  mA  missing  or not in specified limits    The tube current is sampled periodically  by the tube head CPU  and if the actual measured  mA value differs more that   2 mA from the specified value this error is displayed  The tube  head  tube head PCB or tube head feedback cable can be faulty     Proceed with the filament definition  see paragraph 2 3    X ray tube filament preheating volt   age calibration    on page 13 for details  If this does not help and the error occurs constantly  check the tube head feedback cable  replace the tube head PCB  or tube head  in this order      E 33 X ray tube filament voltage  V  missing  or outside the range  too low or  too high   The tube filament voltage is sampled periodically  by the tube head CPU  and if the actual    measured filament voltage is not in the specified limits  1 0   4 5 V  then this error is dis   played  The filament circuit in the tube head or on the tube head PCB can be faulty     Check the tube head feedback cable  Replace the tube head PCB or the tube head  in this  order         24 Planmeca Intra x ray unit Technical manual    ERROR MESSAGES       E 34 X ray tube anode voltage  kV  missing  or below the specified limit    This error occurs in the beginning of the exposure  when the tube anode voltage does not  rise  The tube head  tube head PCB  generator PCB or arm extensio
18. 40 V     50 60 Hz    8 AT 250 VAC at 220 240V   15 AT 250 VAC at 100 V  110 115 V     1 15  automatic control    Class    Type B       Technical manual    Planmeca Intra x ray unit 3    GENERAL  amp  TECHNICAL DATA       Mechanical data    Weight total 23 kg  51 Ibs   tube head 4 2 kg  9 3 Ibs  with standard cone  4 5 kg  10 lbs  with long cone    Color RAL 9016    Environmental requirements    Ambient temperature operating  5  C    40  C storage  10  C    50  C  Humidity 25    75     External mains fuse recommendation  The recommendation for the external mains fuses are   e units with 100V  or 115V  voltage setting  16A  time lag  e units with 220 240V  voltage setting  10A  time lag    No other equipment should be connected to the same fused mains line as the x ray unit  In  some countries an additional external fault current guard is also required     Original manufacturer    PLANMECA Oy  Asentajankatu 6  FIN 00880  Helsinki  FINLAND  phone   358 20 7795 500       4 Planmeca Intra x ray unit Technical manual    GENERAL  amp  TECHNICAL DATA       1 4 User s statement for Planmeca Intra    Radiation leakage technique factors    The maximum rated peak tube potential is 70 kV and the maximum rated continuous tube  current is 0 53 mA for the maximum rated peak tube potential     Minimum filtration    The radiation port contains an added 1 0 mm aluminium filtration  The measured half value is  0 50   0 55 at 70 kV  The measured value corresponds to an aluminium equivalent of 2 0
19. ALUES        Press and hold down any of the preprogrammed setting keys when switching the unit on  The  error code E 29 will appear on the time display     Intra       e ez    0 O  E 29               a o     NO       Do not clear the error code by pressing the select key  but press the Occlusal exposure key  for 6 seconds  The factory preprogrammed values and density value O will be stored into  memory     PLANMECA    Press this key          20 Planmeca Intra x ray unit Technical manual    TROUBLESHOOTING       4    TROUBLESHOOTING    The control panel displays do not come on and the indicator light of  generator box is not on  Problems in mains voltage  or a fuse is blown     Check the mains voltage  Check the mains cable  Check and replace  if necessary  the fuses  located on the lower left side of the generator box  Open the generator box and check and  replace  if necessary  the generator fuse F2  Replace the generator   Perform the checks and  parts replacements in this order      The control panel displays do not come on and the indicator light of  generator box is on    Control panel power failure  or the control panel is defective  The control panel operates with  12 V produced on the generator low voltage supply     Check the control panel cable and the telephone cable  Replace the control panel   Perform  the checks and parts replacements in this order      Temperature of the tube head too high       bl E       If the temperature of the tube head exceeds 60   C  the
20. adjustment keys     GOO       Select key Parameter adjustment keys       Accept the new kV range by pressing the Select key     Press the Mode key briefly to exit the service mode        Technical manual    Planmeca Intra x ray unit 13    SERVICE MODE       2 5 Operation of the preprogrammed settings keys  kV hold     Enter the service mode according to the instruction given in section 2 2    How to enter exit the  service mode    on page 12     Press the parameter adjustment up key until the parameter number 15 appears on the kV dis   play     15 Cw    de Parameter adjustment up  key  kV       The number indicating the operation of preprogrammed settings keys  0 or 1  is shown on the  U s  time display              0  Normal operation  factory setting   The preprogrammed settings keys operate as described  in the Planmeca Intra user   s manual     1  kV hold  The selection of a preprogrammed setting does not affect the prior manually  inserted kV value  but the exposure time is recalculated to achieve constant optical density on  the film  In the case the kV value is now manually altered  the exposure time will also be  changed automatically  If the exposure time is manually altered  the unit will return into man   ual mode  At any time  the preprogrammed setting for the kV value  mA value and exposure  time for any tooth position is called by selecting the preprogrammed setting and pressing the  child mode selection key twice     Press and hold down the Select key until the
21. al number IXRF41386 or smaller need old    7mm  0 28 in   thick arm cable  part number 06320002  Do not use the  new  9mm  0 35 in   thick arm cable  part number 10006682  in these  X ray units     Remove the tube head as described in section 7 3    Replacing the tube head    on page  36   Unscrew the three M4x6 ULS screws and remove the generator cover     Disconnect the arm cable connectors from the Generator PCB  terminals P1 and P8    Detach the grounding lead of the arm cable   Fig  11  1      Remove the extension arm cover plugs  Fig  11  2    Pull the arm cable and the Dixi interconnection cable through the extension arm   Detach the plug of the tube head support  Fig  11  3      Remove the cover plug from the bracket arm opening  Detach the Dixi interconnection  cable from the Dixi control box OR pull the cable out from the bracket arm  Fig  11  4    Unscrew the M3x6 ISO 7380 screw of the tube head support  Fig  11  5  and remove  the cover plate of the tube head support  Fig  11  6   The plate is attached with double   sided adhesive tape    Cut off the connectors from the tube head end of the arm cable push the cable through  the support frame  Fig  11  7         Technical manual    Planmeca Intra x ray unit 37    PARTS REPLACEMENT  amp  REPAIR       i    Cutoff the connectors from the generator end of the arm cable and Dixi interconnection  cable  The grounding lead should be left longer that the other leads  Fig  11  8                        Tech_1APSU2 eps       
22. ating the time out value  0  1  2  3  4  5  is shown on the time display  The  U s  time out values are  0   no dimming  1   1 min   2   2 min   3   5 min   4   20 min  and 5   60  min  Factory default is 1                 Press the Select key until the number starts to blink  and the value can now be changed with  the parameter adjustment keys     E O O    Select key Parameter adjustment keys     amp  Accept the new setting by pressing the Select key   O Press the Mode key briefly to exit the service mode                 16 Planmeca Intra x ray unit Technical manual    SERVICE MODE       2 8 Disabling the exposure key  Enter the service mode according to the instruction given in section 2 2    How to enter exit the    service mode    on page 12   Press the parameter adjustment up key until the parameter number 18 appears on the kV dis     play   18 Y  O  Parameter adjustment up  key  kV    The number indicating the mode of exposure key operation is shown on the time display  0    U s  exposure key normal operation  1   exposure key operation disabled  The factory default is O                 Press the Select key until the number starts to blink  and the value can now be changed with  the parameter adjustment keys     D O O    Select key Parameter adjustment keys    e amp  Accept the new setting by pressing the Select key   f mone d Press the Mode key briefly to exit the service mode           Technical manual Planmeca Intra x ray unit 17    SERVICE MODE       2 9 Automatic i
23. chnical manual                            SERVICE MODE       2 11 Ready state setting                The Planmeca Intra X ray unit can be set so that the Ready indicator light will only come on  when the Dimaxis program is ready for the exposure  i e  the Waiting for exposure message is  on the computer screen     Enter the service mode according to the instruction given in section 2 2    How to enter exit the  service mode    on page 12     Press the parameter adjustment up key until the parameter number 21 appears on the kV dis     play   cl Y  O  Parameter adjustment up  key  kV    The number indicating the ready state setting  0 or 1  is shown on the time display     0  Normal operation  factory setting   The Planmeca Intra X ray unit will go into Ready state  regardless of PC operation     1  The Ready indicator light will only come on when the Dimaxis program is ready for the  exposure    Press and hold down the Select key until the number starts to blink  and the number can now  be changed with the parameter adjustment keys        oso    Select key Parameter adjustment keys       CS Accept the new setting by pressing the Select key        Press the Mode key briefly to exit the service mode        Technical manual    Planmeca Intra x ray unit 19    RECALLING THE FACTORY PREPROGRAMMED EXPOSURE VALUES    3 RECALLING THE FACTORY PREPROGRAMMED  EXPOSURE VALUES       The factory preprogrammed exposure values are in the Planmeca Intra user s manual  sec   tion 9    EXPOSURE V
24. dentification of the cascade card  Enter the service mode according to the instruction given in section 2 2    How to enter exit the    service mode    on page 12   Press the parameter adjustment up key until the parameter number 19 appears on the kV dis     play   19 Y  O  Parameter adjustment up  key  kV       The value indicating the type of the cascade card is shown on the time display  Value  lt  108  108 s  equals an old card and value  gt  108 equals a new card  The type of the cascade card affects  the adjustment functionality of the preheat and mA              2 10 Specifying the PCB versions    Enter the service mode according to the instruction given in section 2 2    How to enter exit the  service mode    on page 12     Press the parameter adjustment up key until the parameter number 20 appears on the kV dis     play   20 C2 L A Parameter adjustment up  key  kV    The value indicating the type of the PCB versions is shown on the time display  1   CPU PCB  3 s  Version C3  or older  with older cascade PCB  2   CPU PCB version C5 with newer cascade  PCB and 3   CPU PCB version E1  works with both old and new cascade PCB  The factory  default is 3     Press the Select key until the number starts to blink  and the value can now be changed with  the parameter adjustment keys       OO    Select key Parameter adjustment keys    CS Accept the new setting by pressing the Select key   O  Press the Mode key briefly to exit the service mode     18 Planmeca Intra x ray unit Te
25. e  the generator 12V power supply is faulty  or the  mains filtering capacitors charging circuit is faulty  The extension cable can be too long and or  the wire cross sections too small     If the error occurs immediately after switching the unit on  replace the tube head PCB  If the  error occurs only after starting the exposure  measure the 12 V voltage at the generator PCB  P13 connector  If the voltage drops at the beginning of the exposure  generator   s green indi   cator light dims   replace the generator PCB  Check the connectors of the arm cable and the  length of the extension cable  as well as the wire cross sections  The properties of extension  cable are given in the figure below                                                                                                                                                                              Ferrite    Generator PCB   Extension cable max 12 m   Extension   Arm cable  Std    l   cable Con       nectors      i    G       1 Yel Grn  i    a o       Molex 3    i 7 Yel HV1 Ki   Molex 3  PA  3 Wht HV2 i p lt     IS o    Molex 6   HV2 T   0    i Gry  12V ETA pa    S BIK EXP   dexa o  i S ped  KVC Hane SIH   P8   4 Bm GND  e Talele      Blu SCL    ser Iela OE  1 Pink ELMP  ME  I0    N C  6 151 Barca HA    1  sR i       o  lo    Screw  gt  Lo    terminal       i y 1 12V Gry j  GND Brn  P14      ELMP Pink    3                      E 61 Communication error between control panel and tube head CPU    The tube head PCB 
26. e bracket arm angles need to be adjusted it can be done with the limiting plates     Remove the cover plug from the end of the bracket arm  Attach the limiting plate to the arm  with two M3 Allen screws  Adjust the angle with the limiting plate adjustment screws        Intradj1 eps        Limiting plate                                                       Figure 2       Technical manual Planmeca Intra x ray unit 29    MECHANICAL ADJUSTMENTS       6 3 Adjusting the stiffness of the tube head   s horizontal axle    Remove the plug from the tube head   s axle and adjust the tightness of the two adjusting  screws evenly  arrows on the figure below         Adj_stiffness eps    WARNING   Do not touch the painted screws  or the earth spring attachment screw                    2 5 Allen screw    Figure 3       30 Planmeca Intra x ray unit Technical manual    MECHANICAL ADJUSTMENTS       6 4 Vertical tube head stiffness adjustment screw    Adjust the stiffness of the vertical tube head by turning the adjustment screw on the support  axle manually or with a wrench tool     The stiffness of the vertical tube head has been preadjusted at the factory  and can be  changed  if necessary     Turn the adjustment screw 0 5   1 rounds clockwise if you want to tighten the tube head and  0 5   1 rounds counterclockwise to loosen it     NOTE Do not turn the adjustment screw too much counterclockwise to  avoid the screw to come loose        intra_stiffness eps                S  Vertical tube head 
27. e the tube head or tube head PCB  in  this order   The possible reason is also faulty feedback cable in the tube head     E 29 Control panel key short circuited pressed during the self test    This error can occur only during the self test  During the self test the unit checks that all keys  are open  normal state if not pressed   If a key is found to be in short circuit  this error is dis   played  Because the control panel keys are arranged in a matrix  one key   s short could cause  the whole keyboard to operate erroneously  therefore this check is important     Replace the control panel     The factory predetermined settings are recalled and the density value is set to zero by press   ing any of the keys during the self test  The error code E 29 appears on the time display  after  which the right hand preprogrammed setting key must be pressed and held down for 6 sec   onds  See section 3    RECALLING THE FACTORY PREPROGRAMMED EXPOSURE VAL   UES    on page 20     E 30 kV value does not reach or it exceeds the given value  difference more  than 5      The tube voltage is sampled periodically  by the tube head CPU  and if the actual measured  kV value differs more than   3 kV from the specified value this error is displayed  The tube  head  tube head PCB  generator PCB or arm extension cable signals HV1  HV2 or KVC  see  wiring diagram  can be faulty  See also error E 11 and E 34     Check the incoming mains voltage during the exposure  Check the arm extension cable and  the
28. ector    P GND ania 100 115 V  P2  BRN SWITCH Code 6113006 Z exe   2  P11   Deadman P12  Remote spiral  P43  Remotekey   app 3  SW conn keyboard conn  board conn  z  ON OFF EPEEE  MainSwitch E a EEE    m   gt  o  Planet Connector  YELL GRN Code 6320054  YELL GRN  YELL  YELL  RED  BLK  WHT  BLK  BLU  Power supply cable 115 V     A    Code 10006683 Display spiral cable  _ Code 10001193  Power supply cable 220 V   Code 10006584    Date  Mains PLANMECA O Y 26 11 2004 Prostyle intra X ray unit  Asentajankatu 6  Designed  100 V   1 ne 00880 Helsinki Finland E Wiring diagram  standard wall mounting installation  110 115 V  210  cheer De  220 240 V   10  S Jankavaara  Approved  J Kolehmainen Rev G                         fulfis the highest acknowledged fulfils the highest acknowledged  international standards international standards                MATION TECHNOL   ye aron  Ter oy       www planmeca com    PA  PLANMECA OY Asentajankatu 6  00880 Helsinki  Finland  AIN tel   358 20 7795 500  fax  358 20 7795 555  sales planmeca com    
29. etic site  survey  should be less than the compliance  level in each frequency range     Interference may occur in the vicinity of  equipment marked with the following sym   bol       gt      NOTE 1  At 80 MHz and 800 MHz  the higher frequency range applies     NOTE 2  These guidelines may not apply in all situations  Electromagnetic propagation is affected by  absorption and reflection from structures  objects and people                    a Field strengths from fixed transmitters  such as base stations for radio  cellular cordless   telephones and land mobile radios  amateur radio  AM and FM radio broadcast and TV broadcast  cannot be predicted theoretically with accuracy  To assess the electromagnetic environment due  to fixed RF transmitters  an electromagnetic site survey should be considered  If the measured  field strength in the location in which Planmeca Intra X ray unit is used exceeds the  applicable RF compliance level above  Planmeca Intra X ray unit should be observed to  verify normal operation  If abnormal performance is observed  additional measures may be  necessary  such as re orienting or relocating Planmeca Intra X ray unit     b Over the frequency range 150 kHz to 80 MHz  field strengths should be less than 3 V m                 10 Planmeca Intra x ray unit Technical manual    GENERAL  amp  TECHNICAL DATA          Recommended separation distances between  portable and mobile RF communications equipment and Planmeca Intra X ray unit       Planmeca Intra X
30. fied limits   E 57 Exposure key pressed during self test    E 60   15VDC voltage is out of limits   E 61 Communication error between control panel and tube head CPU                    22 Planmeca Intra x ray unit Technical manual    ERROR MESSAGES                ERROR CODE ERROR MESSAGE EXPLANATION  E  71 FLASH memory check sum error  tube head CPU     E 8l EEPROM memory defective  tube head CPU     E 83 Config register error  tube head CPU               5 2 Detailed error message explanations    E 00 Exposure key was released too early during exposure    The most probable cause  if the key really was pressed firmly during the whole exposure  is  faulty control panel or faulty separate exposure switch     Replace the control panel or the separate exposure switch  Check the control panel cable  tel   ephone cable and the arm cable     E 10 X ray tube anode voltage  kV  overshoot    E 11    This condition is monitored by the watch dog circuit on the tube head PCB during the whole  exposure and if the anode voltage rises above 95 kV the exposure is immediately aborted  and this error indicated  A knocking sound may be heard from the tube head at the same  time  This kind of arcing can occur now and then without any special reason  and should be  considered a normal phenomenon  If however the occurrence frequency becomes too high  it  could be an indication of a degrading tube head     If this error occurs constantly it is probably caused by a faulty tube head or tube head PCB or
31. is  short circuited  the tube head PCB or the tube head feedback cable is faulty     Check the feedback cable  Replace the temperature sensor  beside the signal connectors of  the tube head   or the tube head PCB  in this order      E 51 Tube head temperature sensor open circuit    The tube head temperature sensor signal is measured by the tube head CPU  The tempera   ture sensor is damaged  the tube head PCB or the tube head feedback cable is faulty     Check the feedback cable  Replace the tube head PCB     E 52 Filament voltage feedback not in specified limits    The filament voltage is monitored by the tube head CPU  The amplifier is faulty   Replace the tube head PCB        Technical manual Planmeca Intra x ray unit 25    ERROR MESSAGES       E 57 Exposure key pressed failure during self test    The tube head CPU checks the state of the exp signal when the unit is switched on  The  exposure key or the separate exposure switch can be short circuited  The arm  extension   control panel or telephone cable can be faulty  The tube head PCB or the generator PCB can  be faulty     Check the cables and the separate exposure switch  Replace the control panel  tube head  PCB or the generator PCB  in this order      E 60   15VDC voltage is out of limits    The tube head CPU measures the internal voltages generated by tube head PCB power sup   ply from the 12V operating voltage  If this error occurs before exposure  the tube head PCB is  faulty  If the error occurs after the exposur
32. lative humidity should be at  least 30         Electrical fast  transient burst     2 kV for power  supply lines     2 kV for power  supply lines    Mains power quality should be  that of a typical commercial or  hospital environment       IEC 61000 4 4  1 kV for input output    1 kV for input output   lines lines  Surge  1 kV line to line  1 kV line to line Mains power quality should be  IEC 61000 4 5  2 kV line to earth   2kV line to earth   that of a typical commercial or    hospital environment        Voltage dips  short  interruptions and  voltage variations  on power supply  input lines    IEC 61000 4 11     lt 5   Ur     gt 95   dip in Ur   for 0 5 cycle   40   Ur    60   dip in Ur   for 5 cycles   70   Ur    30   dip in Ur   for 25 cycles    lt 5   Ur     gt 95   dip in Ur   for5s     lt 5   Ur     gt 95   dip in U7   for 0 5 cycle   40   Ur    60   dip in Ur   for 5 cycles   70   Ur    30   dip in Ur   for 25 cycles    lt 5   Ur     gt 95   dip in Ur   for 5 s    Mains power quality should be  that of a typical commercial or  hospital environment  If the user  of Planmeca Intra X ray unit  requires continued operation dur   ing power mains interruptions  it  is recommended that Planmeca  Intra X ray unit be powered from  an uninterruptible power supply        Power frequency   50 60 Hz   magnetic field    IEC 61000 4 8       3 A m       3 A m       Power frequency magnetic fields  should be at levels characteristic  of a typical location in a typical  commercial or h
33. lue Blue 4  5    Blue   White T SE 5   gt  6 Orange Orange 6    Brown   White Brown   White 7  8 Brown Brown 8  NOTE     The colours marked to the figure are only examples  The lead colours of the cable  you are using can differ from the colours marked to the figure     Figure 15       Shield  RJ45 connector    Figure 16       40 Planmeca Intra x ray unit Technical manual    PARTS REPLACEMENT  amp  REPAIR          n  Insert the RJ45 connector into the cable crimp tool  Press the handles of the crimp tool    Figure 17          as far as they will go  Check with the cable tester that the cable is functioning     DSC03823 jpg    o  Check the RJ45 connector to make sure that the cable is properly attached to the  connector     RJ45 connector before crimp    RJ45 connector after crimp       I                      Wires             UTP Cable             Figure 18             A 11mm        gt                                            Na UTP Cable             Technical manual    Planmeca Intra x ray unit 41    PARTS REPLACEMENT  amp  REPAIR       p  Pull the Dixi interconnection cable out from the bracket arm opening  Connect the Dixi  interconnection cable to the Dixi control box or push the cable into the bracket arm   Attach the cover plug to the bracket arm opening        Dixi interconnection  cable                         Tech_4APSU2 eps    Figure 19  q  Attach the tube head to the support frame with the five M4x8 DIN 7984 screws     r  Slide the tube head PCB into its position
34. n cable signals HV1   HV2 or KVC  see wiring diagram  can be faulty  See also error E 11 and E 30     Check the incoming mains voltage during the exposure  Check the arm extension cable and  the tube head feedback cable  Replace the generator PCB  tube head PCB or tube head   replace in this order      E 36 Too long exposure    The control panel CPU monitors the exposure time by measuring the state of the exp signal   If  however  the tube head CPU continues the exposure more than the maximum exposure  time  then the control panel CPU terminates the exposure and this error occurs  This is a  safety procedure  that guarantees that the exposure is terminated under all conditions     Replace the tube head PCB if this error occurs repeatedly     E 37 kV feedback signal open circuit or short circuit    The kV feedback signal is monitored by the tube head CPU  The internal connection of the  tube head or tube head feedback cable is faulty     Check the condition of the feedback cable  replace the tube head PCB or the tube head  in  this order      E 38 mA feedback signal open circuit or short circuit    The mA feedback signal is monitored by the tube head CPU  The internal connection of the  tube head or tube head feedback cable is faulty     Check the condition of the feedback cable  replace the tube head PCB or the tube head  in  this order      E 50 Tube head temperature sensor short circuit    The tube head temperature sensor signal is measured by the tube head CPU  The sensor 
35. nly valid for the software versions from 3 01 and  later        2 Planmeca intra x ray unit Technical manual    GENERAL  amp  TECHNICAL DATA       1 3 Technical specifications    Generator    X ray tube  Focal spot size  Cone diameter    Max  symmetrical radiation field    Total filtration  Inherent filtration    Anode voltage   Anode current   Target angle   Exposure times   Reference current time product  Lowest current time product  Max  nominal anode voltage  Max  electrical output  Electrical output at 0 1 sec   Max  loading energy    SSD  Source Skin Distance   Standard Long  Long with rectangular collimator    Mains voltage  Apparent resistance    Mains frequency  Fusing    Duty cycle  Electrical classification    Constant potential  microprocessor controlled   operating frequency 66 kHz    Toshiba D 0711SB  0 7 mm according to IEC 60336    60 mm  2 36 in    Rectangular 33 x 43 mm  1 30 x 1 69 in      260mm at SSD 200 mm     60 mm at SSD 300 mm  according to IEC 806    min  2 mm Al equivalent at 70 kV  according to IEC 60522    1 mm Al equivalent at 70 kV  according to IEC 60522    50  52  55  57  60  63  66  70 kV   2 kV   8  7  6  5  4  3  2 mA   5    0 2 mA    16     0 01 3 2 sec    5    0 001 sec    26 steps  8 mAs at 70 kV  8 mA  1 sec    0 02 mAs at 2 mA  0 01 sec    70 kV   560 W at 70 kV  8 mA   560 W at 70 kV  8mA   1800 mAs h at 70 kV    200 mm  8 in   300 mm  12 in    306 mm  12 04 in      100 V  110 115 V  220 240 V     0 3 ohms 110 115 V     0 8 ohms 220 2
36. ospital environ   ment  The power frequency mag   netic field should be measured in  the intended installation location  to assure that it is sufficiently low           NOTE Ur is the a c  mains voltage prior to application of the test level              Technical manual    Planmeca Intra x ray unit 9    GENERAL  amp  TECHNICAL DATA          Guidance and manufacturer   s declaration   electromagnetic immunity       Planmeca Intra X ray unit is intended for use in the electromagnetic environment specified below  The  customer or the user of Planmeca Intra X ray unit should assure that it is used in such an environment        IEC 60601 Compliance Electromagnetic environment     Immunity test test level level guidance       Portable and mobile RF communications  equipment should be used no closer to any  part of the Planmeca Intra X ray unit  includ   ing cables  than the recommended separa   tion distance calculated from the equation  applicable to the frequency of the transmit   ter     Recommended separation distance    Conducted RF   3 Vrms 3 Vrms d   124P    IEC 61000 4 6   150 kHz to 80 MHz  d   12  P 80 MHz to 800 MHz    Radiated RF 3 V m 3 V m d   23  P 800 MHz to 2 5 GHz    IEC 61000 4 3   80 MHz to 2 5 GHz l f  where P is the maximum output power rat   ing of the transmitter in watts  W  according  to the transmitter manufacturer and d is the  recommended separation distance in  metres  m      Field strengths from fixed RF transmitters   as determined by an electromagn
37. stiffness adjustment screw   _      Figure 4       Technical manual Planmeca Intra x ray unit 31    PARTS REPLACEMENT  amp  REPAIR       7 PARTS REPLACEMENT  amp  REPAIR    AN WARNING    Make sure that the power supply is switched off before starting parts  replacement     7 1 Replacing the Generator PCB    a  Unscrew the three M4x6 ULS screws and remove the generator cover                 Generator cover                                  Figure 5       32 Planmeca intra x ray unit Technical manual    PARTS REPLACEMENT  amp  REPAIR       b  Measure that the mains voltage is not present at the mains input terminals  P5  marked  N and L   c  Disconnect the connectors from terminals P1  P3  P5  P6 P7  P8 and P9     iit  oo                         II                                                                                                                                                                                                                                                                              Gf  N      EN       P5   it Z  bee x 5    E l  Y   Ogle     Soe  Figure 6    d  Loosen the four M4x8 ISO 7380 screws of the generator assembly frame  Fig  7  1   Lift  the generator assembly upwards  Fig  7  2   The generator assembly can now be lifted  away from the wall adapter  Fig  7  3      ps                          _tech_4_2APSU2 e                                                                                                                                      
38. user of the Planmeca Intra X ray unit should assure that it is used in  such an environment                    Emissions test Compliance Electromagnetic environment     guidance  RF emissions Planmeca Intra X ray unit uses RF energy only  Group 1 for its internal function  Therefore  its RF emis   sions are very low and are not likely to cause any  CISPR 11 interference in nearby electronic equipment   RF emissions Planmeca Intra X ray unit is suitable for use in all  Class B establishments  including domestic establish   ments and those directly connected to the public  CISPR 11 low voltage power supply network that supplies  Harmonic emissions buildings used for domestic purposes   Class A  IEC 61000 3 2  Voltage fluctuations   flicker emissions Complies  IEC 61000 3 3                      8 Planmeca intra x ray unit Technical manual    GENERAL  amp  TECHNICAL DATA          Guidance and manufacturer   s declaration   electromagnetic immunity       Planmeca Intra X ray unit is intended for use in the electromagnetic environment specified below  The cus   tomer or the user of Planmeca Intra X ray unit should assure that it is used in such an environment        Immunity test    IEC 60601  test level    Compliance level    Electromagnetic environment   guidance       Electrostatic  discharge  ESD     IEC 61000 4 2     6 kV contact   8 kV air     6 kV contact   8 kV air    Floors should be wood  concrete  or ceramic tile  If floors are cov   ered with synthetic material  the  re
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Service Manual  Vespa – Simulation User Manual and Reference - VeSPA  Metra Electronics 99-7008 User's Manual    Tripp Lite B021-000-19-SH rack console  Beewi Scara Bee  Manual de instrucciones  Caldera de condensación a gas WTC 15 … 32-A  SeHAT User Manual - SOCSO Health Screening Programme  製品安全データシート (MSDS)    Copyright © All rights reserved. 