        User`s Manual - Test Equipment Depot
         Contents
1.    CA11E  HANDY CAL   Voltage Current Calibrator   2     User s  Manual       In addition to this Users Manual  there is another manual for the instrument   IM CA11E 01 E  which provides instructions for its safe use and explains its  functions  Refer to it as needed     YOKOGAWA       Yokogawa Meters  amp  Instruments Corporation    IM CA11E 02E  1st Edition Apr  2006  KP     8  Calibration Procedure  E Calibration Procedure  To maintain a high level of accuracy  it is recommended that the CA11E HANDY CAL be calibrated  annually  The    Selecting the Standards    section below presents calibration methods using the  recommended standards     E Selecting the Standards  Source feature    Items to be Names of Ranae Measuring   calibrated standards g range  Max 110 mV  Max 1 1 V    Max 11 V  Max 33 V    Digital multimeter 20 mA 4 20 mA    Max 24 mA    Digital multimeter   20 mASINK   standard DC 20 mASINK 0 01 to 7 mA  voltage generator 5 to 28    Measurement feature    Items to be Names of Generated  0 01  1 00 mV  M d l 2552  0 01  1 V  a    0 01  10 V   0 02  30 V     Standard DC a    E Environmental Conditions for Calibration  Ambient temperature  230 21  C  Relative humidity  45 to 75  R H  Warm up  Warm up time as specified for the standard    Accuracy    0 005  5 uV   0 005  20 uV   0 005  200 uV   0 005  2 mV     Model 7561   Yokogawa     Digital multimeter  or equivalent    t  m      2      0 001  0 8 uV     Standard DC     Yokogawa   voltage generator    or equivalent 
2.    E Calibration Points  e The calibration points are as specified in the following tables   e It is possible to independently select the necessary range to be recalibrated   e Always calibrate the zero  0  point and full scale  FS  as a pair for generation     Measurement     Calibration points Standard Value 2  100 mV FS  FS    Source     Calibration points Standard Value     aw Cf om S   jo fow  1V   1000 mV   O e  10 V   oo Y  30 V   ooo y  20 mA  20 mA  Sins     1  Make adjustments to CA11E according to the  reading of the standard  CA11E output Value   as specified in the table         2  Set the Value to the standard as specified in  the table        E Calibrating the Generation Feature  Operation procedure     lt 1 gt  Warm up the standard     lt 2 gt  Before turning on the power of the CA11E calibrator  connect it to the standard     lt 3 gt  Turn on the power of the instrument     lt 4 gt  Simultaneously press and hold the  A1  and  V4  keys  shown in the figure in the    Assign   ment of Keys for Calibration    section below  for about 2 seconds to enter the calibration  mode     lt 5 gt  Select the generation range to calibrate using the MEASURE SOURCE selection switch  and range selection rotary switch  The display unit shows    CAL        SOURCE        ON        0     and  the lower limit     lt 6 gt  Read the output value of the CA11E using the standard  digital multimeter   and using the   A  and  V  adjustment keys adjust the CA11E so that the output value is
3.  set to the offset  value  Then press the  W1  input determination  ENTER  key to fix the setting   After fixing the setting  the display unit reading changes to    CAL        SOURCE        ON        FS      and a full scale of the range     lt 7 gt  Read the output value of the CA11E using the standard  digital multimeter   and using the   A  and  V  adjustment keys adjust the CA11E so that the output value is set to the full  scale value  Then press and hold the  W1  input determination  ENTER  key for about 1  second to fix the setting   After fixing the setting  the display unit shows    0 FS    blinking  Re press and hold the  V1   input determination  ENTER  key for about 1 second to write the calibration coefficients to  the EEPROM of the instrument   This overwrites the previous calibration coefficients    When this is complete  the instrument returns to the status in Step 5     lt 8 gt  Repeat Steps 5 to 7 for each range to be calibrated     To return to the previous step    lt 9 gt  To return to the previous step without fixing the setting  press the  A1  input cancellation  key  Note that this is possible only before writing to the EEPROM     To return to the normal operation mode    lt 10 gt Simultaneously press and hold the  A1  and  W4  keys  shown in the figure in the    Assign   ment of Keys for Calibration    section below  for about 2 seconds  or press the  POWER  key  to turn off the power once and then press it again to turn it back on     Test Equipment D
4. P are excluded    EN61326 ClassB  EN55011 ClassB Group1  EN61000 3 2  EN61000 3 3  Immunity  EN61326  Performance criterion under immunity test environments  B   self returnable performance deterioration   Test conditions of EMC and Immunity standards   AC adapter  B9108WB  and Lead cables  B9108MS  are used     Battery life     Safety standards     EMC standards     Test Equipment Depot   800 517 8431   99 Washington Street Melrose  MA 02176       10  How to Use the Carrying Case and Rubber Boot  E The carrying case  B9108NK  may be used as follows         1  Undo the fasteners on the top  2  With the fastener on the logo   and sides of the case cover to side of the case centered  lift the  open it  cover and pivot it to the side and   under the case itself      3  Re do the fasteners at the top  and sides of the cover   Note  The fastener on the logo side cannot be undone     E Rubber Boot  93038  E Anti skid pad    Anti skid pad   supplied        The optional rubber boot provides  shock protection and can be used  with a strap     Note  When using the instrument with a rubber boot  the anti skid pad at the bottom is not  needed  When used without a rubber boot in a leaning position  the supplied anti skid  pad should be used     IM CA11E 02E  lt 2 gt     FAX 781 665 0780   TestEquipmentDepot com    
5. epot   800 517 8431   99 Washington Street Melrose  MA 02176       99 Washington Street  Melrose  MA 02176   Fax 781 665 0780  TestEquipmentDepot com    Test Equipment  Depot    1 800 517 8431    E Notice regarding the Manual   lt 1 gt  The information contained in this User   s Manual is subject to change without notice    lt 2 gt  Every effort has been made to ensure that the information contained herein is accurate   However  should any concerns  errors  or omissions come to your attention  or if you  have any comments  please contact us     E Calibration Precautions  e Connection for 20 mA SINK calibration  Connect the CA11E calibrator to the standard as shown below     DMM  L current range    O  Constant voltage  power supply   Vo  5 28V        E Assignment of Keys for Calibration  When the CA11E calibrator is in the calibration mode  keys are assigned as specified here              Input cancellation key             Input determination Tr   ENTER  key 2 fut rrree E ENE EEEE             Mode switching  calibration or normal    Press and hold two keys  simultaneously for 2 seconds                     Adjustment keys  six keys  B l             Digit to be adjusted by the  A4 V4  adjustment keys         i             E Calibrating the Measurement Feature  Operation procedure     lt 1 gt  Warm up the standard     lt 2 gt  Before turning on the power of the CA11E calibrator  connect it to the standard     lt 3 gt  Turn on the power of the instrument     lt 4 gt  Simultaneousl
6. r about 2 seconds  or press the  POWER  key  to turn off the power once and then press it again to turn it back on     IM CA11E 02E  lt 1 gt     FAX 781 665 0780   TestEquipmentDepot  com    9  Specifications  E Source Functions  Accuracy      of set value   uV  mV or WA   at 23  5  C    Range Range of A Setting  Selection Generated Signal eure    Resolution  0 to 30 00 V 0 05    20 mV Maximum output current  1 mA  0 to 11 000 V Maximum output current  10 mA  0 05    2 mV When the load is 1k or greater  and   1 2 3 4 5 V 1 Vstep greater   the error of the lead cables is also  excluded     0 05   4 uA Maximum load  12 V  4 20 mA  4 8 12 16 20 mA 4 mAstep  20 mA SINK    0 1 to 24 000mA   0 1  4uA External power supply  5 to 28 V    Temperature coefficient  1 10 of accuracy   C  but  0 005    10 uV    C in the 100 mV range     The resolution can be changed between 5 digits and 4 digits using a DIP switch     0 to 1100 0 mV 0 05    0 2 mV       E Measurement Functions  Accuracy      of reading   least significant digits   at 23   5  C    Range  0 to  30 00 V 0 05    2 dgt 10 mV  Oto 11 000V   0 05    2 dgt  mV    Input impedance Approximately 1 MQ  Oto  1100 0mV   0 05   2 dat  Oto  110 00 mV   0 05   7 dgt  0 to  24 000 mA 0 05    4 dgt Input impedance  45 Q    Temperature coefficient  1 10 of accuracy   C  but  0 005    10 uV    C in the 100 mV range     The resolution can be changed between 5 digits and 4 digits using a DIP switch        Power supply  Four 1 5 V alkaline batte
7. ries  LR6  AA size  or dedicated AC adapter   sold separately   Approximately 50 hours for 5 V DC output with a load of 10 kQ or greater   when running on alkaline batteries   Approximately 25 hours for 20 mA output with a load of 5 V   when running on alkaline batteries   Automatic Power Off  After a period of approximately 10 minutes with no operations  Generation Signal Level Setting   By four sets of up and down keys  Response of generator  Approximately 1 second  between the time for an output signal to start  changing and the time when it enters the accuracy range   Loading conditions  Less than 0 1 uF  DC V   Measured value indication updating intervals  Approximately 1 second  Display  7 segments LCD  Maximum allowable applied voltage   30 V DC between each terminal and ground  Operating temperature and humidity range   0 to 50  C  20 to 80  R H   no condensation   Storage temperature and humidity range    20 to 50  C  90  R H  or less  no condensation   Dimension  Approximately 192  H  x 92  W  x 42  D  mm  excluding protrusions   Weight  Approximately 440 g  Accessories  Lead cables  B9108MS  for measurement and generation   one pair  consisting of one black cable and one red cable   4 alkaline batteries  2 User   s manuals  Optional accessories  Dedicated AC adapter   A1020UP  AC100V  A1022UP  AC120V  B9108WB  AC 220 240V   Carrying case  B9108Nk   Rubber boot  93038   Strap  97040   Accessory case  B9108XA   EN61010 1  only AC adapter B9108WB    A1020UP and A1022U
8. y press and hold the  A1  and  V4  keys  shown in the figure in the    Assign   ment of Keys for Calibration    section below  for about 2 seconds to enter the calibration  mode     lt 5 gt  Select the measurement range to calibrate using the MEASURE SOURCE selection switch  and range selection rotary switch    CAL    and    MEASURE    appear and    FS    blinks on the display unit    If a value nearly equivalent to full scale has been input  a measured value and    FS     appear      lt 6 gt  Set up the standard in order to input the full scale value to the instrument  Wait until the  reading stabilizes  then press and hold the  V1  input determination  ENTER  key for about  1 second to fix the setting     lt 7 gt  After fixing the setting     0    and    FS    indications on the display unit start blinking  Re press  and hold the  V1  input determination  ENTER  key for about 1 second to write the calibra   tion coefficients to the EEPROM of the instrument   This overwrites the previous calibration  coefficients      lt 8 gt  Repeat Steps 5 to 7 for each range to be calibrated     To return to the previous step    lt 9 gt  To return to the previous step without fixing the setting  press the  A1  input cancellation  key  Note that this is possible only before writing to the EEPROM     To return to the normal operation mode    lt 10 gt Simultaneously press and hold the  A1  and  V4  keys  shown in the figure in the    Assign   ment of Keys for Calibration    section below  fo
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
蘇生バッグの組立て間違いについて  User`s Manual  L`ACTU DES TRAVAUX DANS VOTRE QUARTIER - Issy  中国の民事裁判における当事者の地位に関する考察 (2): 処分権への  Philips Respironics Auto Bi - Flex (español) - TR-KinE  "user manual"  Zebra Printhead 2746e 203 dpi  Built-in Oven BlMFBSSOOXPS User Manual    Owner's manual & Installation manual Mode d'emploi et    Copyright © All rights reserved. 