TMP-REG with TMP-SET (optional) Temperature
Contents
1. 90 ag 80 70 T 60 5 y Hysteresis 1000 50 SE fo OMS bo O P 5f _setpoint 50 ________ 40 EEEE oe og ARI ee ees 40 i g On No j orio 20 10 SE O S 10 t 0 10 20 30 40 50 60 70 80 S0 100 0 10 20 30 40 50 60 70 8 amp 0 50 100 Relay action Increasing Relay action Decreasing Example how to keep temperature low despite warming effects Set up the TMP REG system in the following way to keep temperature at say 15 C in aquaria due to warming effects from sun shining and heating from pumps 1 Connect the submersible Eheim pump to relay 1 on the back side of the TMP REG instrument relay 2 is not used in this example 2 Connect the coil to the output of the submersible Eheim pump via the tubing Use so much tube that the coil can be placed in the cooling bath 3 Place the coil into the cooling bath Make sure that the temperature of the cooling bath is cooler than the wanted Setpoint 15 C e g water with ice cubes 4 Connect to the coil the remaining tubing and place the tubing into the aquaria Now when the relay turns on the submersible pump will suck water from the aquaria and pump it through the coil While water is pumped through the coil the water will be cooled down So the water that is pumped back into the aquaria is cooled Instrument settings 5 Press the OK button several times until the display reads REL1 Wait for one second and the display now reads SET or toggle using arrow buttons u
2. SE gt SE VALJ SVENSK HJALPTEXT UK gt UK SELECT ENGLISH HELPTEXT CZ gt CZ VYBER CESKOU NAPOVEDU xxxx gt SET CORRECT PASSWORD C LIN gt TEXT ENTERED BY USER IN PRESET CURR gt CURRENT INPUT VOLT gt VOLTAGE INPUT POTM gt POTENTIOMETER INPUT TEMP gt TEMPERATURE SENSOR INPUT When current selected 0 20 gt INPUT RANGE IN mA 4 20 gt INPUT RANGE IN mA When voltage selected 0 10 gt INPUT RANGE IN VOLT 2 10 gt INPUT RANGE IN VOLT 0 0 1 gt INPUT RANGE IN VOLT 0 2 4 gt INPUT RANGE IN VOLT YES gt CALIBRATE POTENTIOMETER LOW NO gt CALIBRATE POTENTIOMETER LOW YES gt CALIBRATE POTENTIOMETER HIGH NO gt CALIBRATE POTENTIOMETER HIGH DEC P 1111 111 4 11 11 1 111 DI LO XXXX DI HI XXAX REL U PERC DISP DEC P 1111 111 1 UNIT C F gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt DECIMAL POINT POSITION DECIMAL POINT POSITION DECIMAL POINT POSITION DECIMAL POINT POSITION DISPLAY READOUT LOW DISPLAY READOUT HIGH SET RELAY IN PERCENTAGE SET RELAY IN DISPLAY UNITS SELECT PT SENSOR TYPE SELECT NI SENSOR
3. TYPE SELECT TC SENSOR TYPE SELECT PT SENSOR TYPE SELECT PT SENSOR TYPE SELECT PT SENSOR TYPE SELECT PT SENSOR TYPE SELECT PT SENSOR TYPE SELECT PT SENSOR TYPE SELECT PT SENSOR TYPE SELECT PT SENSOR TYPE SELECT PT SENSOR TYPE SELECT PT SENSOR TYPE SELECT NI SENSOR TYPE SELECT NI SENSOR TYPE SELECT NI SENSOR TYPE SELECT NI SENSOR TYPE When Pt and Ni sensor selected SELECT 2 WIRE SENSOR CONNECTION SELECT 3 WIRE SENSOR CONNECTION SELECT 4 WIRE SENSOR CONNECTION SELECT TC SENSOR TYPE SELECT TC SENSOR TYPE SELECT TC SENSOR TYPE SELECT TC SENSOR TYPE SELECT TC SENSOR TYPE SELECT TC SENSOR TYPE SELECT TC SENSOR TYPE SELECT TC SENSOR TYPE SELECT TC SENSOR TYPE SELECT TC SENSOR TYPE SELECT TC SENSOR TYPE SELECT TC SENSOR TYPE SELECT TC SENSOR TYPE When temperature selected DECIMAL POINT POSITION DECIMAL POINT POSITION DISPLAY AND RELAY SETUP IN CELSIUS DISPLAY AND RELAY SETUP IN FAHRENHEIT gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt gt ENTER RELAY 1 SETUP SKIP RELAY 1 SETUP RELAY 1 DISABLED RELAY SETPOINT ACTIVATE AT INCREASING SIGNAL ACTIVATE AT DECREASING SIGNALL RELAY HYSTERESIS HOLD RELAY AT ERROR ACTIVATE RELAY AT ERROR DEACTIVATE RELAY AT ERROR UNDEFINED STATUS AT ERROR RELAY ON DELAY IN SECONDS RELAY OFF DELAY IN SECONDS ENTER R
4. the last option available Once the entire configuration has been entered the display will show The following two pages show the complete menu routing diagram and scrolling help texts NB Please note that the TMP REG is password protected The Password is 1234 The RE1 is set to act on a decreasing signal the RE2 is set to act on an increasing signal Use the fast set point adjustment to adjust the set point FAST SET POINT ADJUSTMENT It is possible to quickly change set point values without having to scroll through the whole instrument menu With the instrument turned on press the A button once Now choose relay 1 or 2 and press the OK button Then change the set point value using the arrow buttons Finish by pressing the OK button Test relay functions by pressing both arrow buttons simultaneously The LED diode will toggle its state When done press the OK button several times until the display reads Page 3 of 9 ROUTING DIAGRAM If no keys are activated for 2 minutes the display returns to default state 1 0 without saving configuration changes A Increase value choose next parameter Decrease value choose previous parameter 5 Save the chosen parameter and go to the next menu Hold Back to previous menu return to menu 1 0 without saving Power up LOLIGOSYSTEMS COM Fast setpoint adjustment and relay test Increase setpoint v Decrease setpoint Save and exit th
5. C EN 61010 1 UL Standard for Safety UL 508 Page 9 of 9
6. ELAY 2 SETUP SKIP RELAY 2 SETUP RELAY 2 DISABLED RELAY SETPOINT ACTIVATE AT INCREASING SIGNAL ACTIVATE AT DECREASING SIGNAL RELAY HYSTERESIS HOLD RELAY AT ERROR ACTIVATE RELAY AT ERROR DEACTIVATE RELAY AT ERROR UNDEFINED STATUS AT ERROR RELAY ON DELAY IN SECONDS RELAY OFF DELAY IN SECONDS OUTPUT RANGE IN mA OUTPUT RANGE IN mA OUTPUT RANGE IN mA OUTPUT RANGE IN mA DISPLAY VALUE FOR OUTPUT LOW DISPLAY VALUE FOR OUTPUT HIGH O ERR 23 mA gt 3 5 mA gt OmA gt NONE gt RESP AXX K gt E PAS NO gt YES gt N PAS NAMUR NE43 UPSCALE AT ERROR NAMUR NE43 DOWNSCALE AT ERROR DOWNSCALE AT ERROR UNDEFINED OUTPUT AT ERROR ANALOGUE OUTPUT RESPONSE TIME IN SECONDS ENABLE PASSWORD PROTECTION ENABLE PASSWORD PROTECTION SELECT NEW PASSWORD Page 5 of 9 L oLico USING THE RELAYS LOLIGOSYSTEMS COM The TMP REG instrument has two independent relays to control the activity of submersible pumps For each relay choose a set point and a hysteresis value Now decide if the relay should be activated when the signal drops below a set point DECR or if it should be activated when it rises above INCR e g for cooling control the relay should act on a increasing INCR signal since temperature will increase above the set point due to heating effects in the chamber e g sun shining Graphic depiction of the relay function setpoint t Relay units A Relay units 100 100 90
7. e menu A and v simultaneously change relay state HYSe t PN T INCR NONE Y 1 EER PEPE SEIE LE slab ee areata a ee as a eS SiK roo o ew eS ew en Swe eee me T NONE 1 0 Default state 1 1 Only if password protected 1 2 Only is FastSet is enabled 1 3 FastSet and relay test disabled at password 5000 9999 1 4 No relay outputs 1 5 Not valid for these input signals Potentiometer 0 20 mA and voltage 1 6 No analogue output 1 7 Only if input signal is temperature To default state 1 0 Page 4 of 9 LOLIGOSYSTEMS COM SCROLLING HELP TEXT Display in default state xxxx hardware error SE BR gt SENSOR WIRE BREAKAGE SE SH gt SENSOR SHORT CIRCUIT IN HE gt INPUT OVERRANGE IN LO gt INPUT UNDERRANGE 9 99 9 gt DISPLAY OVERRANGE 1 9 9 9 gt DISPLAY UNDERRANGE HW ER gt HARDWARE ERROR EE ER gt EEPROM ERROR CHECK CONFIGURATION RA ER gt RAM MEMORY ERROR CJ ER gt CJC SENSOR ERROR Fastset Enabled F SET REL1 gt FAST SET MENU SELECT RELAY REL2 gt SETP xxxXKX gt RELAY SETPOINT PRESS OK TO SAVE Fastset Disabled SETP xxxx gt RELAY SETPOINT READ ONLY Open menus DE gt DE WAEHLE DEUTSCHEN HILFETEXT DK gt DK VAELG DANSK HJAELPETEKST ES gt ES SELECCIONAR TEXTO DE AYUDA EN ESPANOL FR gt FR SELECTION TEXTE D AIDE EN FRANCAIS IT gt IT SELEZIONARE TESTI DI AIUTO ITALIANI
8. l oLico LOLIGOSYSTEMS COM TMP REG with TMP SET optional Temperature Analyzer and Regulator System User manual version 2 4 The TMP REG system is used for monitoring and regulating water temperature of sea or fresh water in fish tanks respirometers aquaria etc 1 Controller instrument 2 Data Cable 3 Pt100 temperature probe 4 Power cord 5 Converter piece 6 User manual 7 TMP SET OPTIONAL Submersible Eheim pump Adapter cable for pump Soft PVC tubing Stainless steel cooling coil LIST OF PARTS Ng CONTENTS BIST OP PARTS A E A E A E A dass E E ie suaeate ers 1 CONTENT orire ride N E E NEEE E EEE E EEEE 1 CONFIGURATION OPERATING THE FUNCTION KEYS aasssusssssnssnnnrnnnnrnnnnrnrnnrrnnrnnnrnrnnnrrerrnnns 3 USING THE RELAT siirsin rinine E n E E EEE ae ERA 6 FAST SET POINT ADIUS IMENT owstedeecoxscmase deen ESE EETAS eas dete senrdieeksenseie des ctesemaeeiewis 3 PA SWORD PROTECTION oeras cent einnar EEE EEEE E santas ana snexaandvextaexaxoh anaes clas pemuswansednesaee 7 DEFAULT SETTING ern inean e E E A E eds TEE E E AE nae 8 USING THE CONTROLLER FOR DATA ACQUISITION caceircscnoevnesnanessendvediacss seems caseseseeeanenven tes 8 CRANGE TEMPERATURE UNITS popretie nrin noran ine A OAE A E AA EE an EE EE ON 8 TMPHREG OPE CIRICATION Sarea EE E O E EE E 9 Page 1 of 9 l oLico LOLIGOSYSTEMS COM SETUP Connect the Pt100 temperature probe to the input labelled IN on the front side of the TMP REG in
9. n set 0 LO to 15 and O HI to 25 Now the 0 5 Volt output voltage will be O V at 15 C and 5 V at 25 C The gain will be 2 C V CHANGE TEMPERATURE UNITS It is possible to measure and control temperature in other units than C Press OK several times until UNIT appears Now use the A or V buttons to toggle between C or F When done press OK several times until appears Page 8 of 9 L oLico LOLIGOSYSTEMS COM TMP REG SPECIFICATIONS Supply voltage universal 21 6 253 VAC 50 60 Hz or 19 2 300 VDC Internal consumption 3 2 W Max consumption 3 5 W Isolation voltage test operation 2 3 kVAC 250 VAC Signal noise ratio Min 60 dB 0 100 kHz Response time programmable 0 4 60 s Calibration temperature 20 28 C Accuracy lt 0 1 of reading Temperature Coefficient lt 0 01 of reading C EMC immunity influence lt 0 5 of reading Potentiometer input min 10 Q Potentiometer input max 100 kQ Relay function Setpoint Hysteresis in display counts 0 1 25 1 2999 On and Off delay 0 3600 s Sensor error detection Make Break Hold Max voltage 250 VRMS Max current 2A AC Max AC power 500 VA Max current at 24 VDC 1A IMPORTANT DO NOT connect relays to gt 500W equipment max 2 A 250 V Marine approval Det Norske Veritas Ships amp Offshore Standard for Certification No 2 4 Observed authority requirements Standard EMC 2004 108 EC Emission and immunity EN 61326 LVD 73 23 EE
10. ntil it reads SET and then press OK Page 6 of 9 L oLico LOLIGOSYSTEMS COM 6 Now use the A or V buttons to increase or decrease the set point value Set it to 15 C and then press OK 7 Set the action of the relay to INCR and then press OK 8 Now enter the hysteresis value Use arrow buttons to set the value to 0 5 C and then press OK 9 Press OK several times until the display reads N to finish Relay 1 will be activated every time temperature rises above 15 C pumping water through the coil This will cause the temperature of the water to decrease and as it reaches 14 5 C the relay de activates After some time temperature will rise above 15 C again due to heating effects and relay 1 is activated once more In this way temperature in the aquarium is automatically kept at 14 5 15 C at all times PASSWORD PROTECTION Using a password will stop access to some of the menu and parameters There are two levels of password protection Passwords between 0000 4999 will allow access to the fast set point adjustment and relay test Using this password stops access to all other parts of the menu Passwords between 5000 9999 stop access to all parts of the menu fast set point and relay test Current set point is still shown By using the master password 2008 all configuration menus are available If you want to enable password protection press the OK button several times until the display read
11. s E PAS Use an arrow button to choose YES and then press OK Now set the password using the arrow buttons and press OK when done If you want to disable password protection go to the menu option E PAS again and set it to NO Finish by pressing the OK button Page 7 of 9 L oLico LOLIGOSYSTEMS COM DEFAULT SETTINGS The TMP REG instrument is delivered with the following default settings IN TEMP TYPE PT PT TY 100 CONN 4W DEC P 111 1 UNIT ke REL1 SETP 25 0 ACT DECR HYS 0 5 ERR HOLD ON DE O OF DE 0 REL2 SETP 25 0 ACT INCR HYS 0 5 ERR HOLD ON DE O OF DE 0 ANALOG OUTPUT 4 20 converted into a 0 5V instrument output 0 LO 0 O HI 100 O ERR 3 5mA RESP 1 0 E PASS YES N PASS 1234 USING THE CONTROLLER FOR DATA ACQUISITION The instrument produces a 0 5 Volts analog output signal for data acquisition purposes Connect the data cable to the socket marked Out on the backside of the TMP REG instrument Connector pin 1 is positive connector pin 4 is O zero The low instrument value 0 LO corresponds to a 0 V output and the high value 0 HI corresponds to a 5 V output This means that the gain becomes HI value 5 V With the default values the gain is then 20 C V e g if the output voltage is 4 V the calculated temperature value is 80 C If the 0 5 Volt analog signal is noisy try to shorten the range of interest of the temperature output E g the temperature signal will always be within the range of 15 25 C The
12. strument Now connect the power cord to the input labelled POWER on the backside of the TMP REG instrument The TMP REG will now turn on and start reading temperature values There are two relays on the backside of the instrument which can be used to control the activity of the submersible Eheim pump to regulate temperature in the water Temperature analyzer TMP REG rs Tubing Cooling heating bath Pump cable Adapter cable Pt100 temperature probe Da _ Eheim pump Water tank Page 2 of 9 L oLico LOLIGOSYSTEMS COM CONFIGURATION OPERATING THE FUNCTION KEYS The TMP REG is a menu driven instrument The front panel has three buttons for operation e g two arrow buttons A and V and one OK button Use these three buttons to scroll through the menu and accept instrument settings For each menu there is a scrolling help text which is automatically shown in the display this starts after five seconds if no key has been activated Use the menu to calibrate the temperature probe and set relay action and level of control A will increase the numerical value or choose the next parameter V will decrease the numerical value or choose the previous parameter OK will accept the chosen value and end the menu To summarize the A and V buttons are used to toggle betweens options The OK button is used to accept settings and go to the next option Press and hold the OK button for 0 5 seconds to get