QUICK START GUIDE Picus® & Picus® NxT Electronic Pipettes

image

Contents

1. wheel 3 Confirm the setting by pressing a the operating button or the right softkey for OK which will exit the editing mode OR b the middle softkey for NEXT to move on to edit the next highlighted setting 4 Repeat steps 2 and 3 for all the settings you wish to edit To activate the advanced functions 1 Press the right softkey for ADV to choose the desired advanced function e g mixing tracker counter 2 Scroll through the list of advanced functions using the softkey for NEXT 3 Turn the selected advanced function on by turning the adjustment wheel 4 Press the operating button or the softkey for OK to accept the activation To use the pipette with the selected mode 1 Attach the pipette tip s 2 Press the operating button as many times as needed to complete the pipetting sequence until the liquid is dispensed 3 Eject the tip by pressing the tip ejector button Setup It is possible to personalise settings in Picus Picus NxT Press the left softkey to open the menu and scroll down using the adjustment wheel to choose SETUP Setup Options Sound turns sounds on or off for example for the adjustment wheel messages and battery status indicator Adjustment lets the user to adjust the pipette at 1 2 or 3 calibration points 1 point adjustment calibration at 1 point set the calibration volume in ul 2 point adjustment calibration at 10 and 100 of the nominal volume 3 point adjust
2. QUICK START GUIDE Picus amp Picus NxT Electronic Pipettes Getting Started 1 SE Sartorius Picus Picus NxT pipettes are delivered ready for use with the battery partially charged However for the best results we recommend fully charging the pipette for approximately one hour before first use Using the USB Charger connect the charger to the pipette s USB socket and plug the charger adapter into a power outlet mains Using the charging stand or carousel make sure the charging stand is connected to a power outlet mains and insert the pipette into the grooves on the charging stand so that the fi charging contacts meet Insert the replaceable Safe Cone Filter s into the tip cone s before use to reduce the possible risk of the pipette becoming contaminated Alternatively you may use filter tips Press the ON OFF button on top of the pipette to turn the power on F Press the tip ejector when indicated by the display The pipette is now ready for programming and use Product Overview Operating Buttons USB charging socket ON OFF button Operating button Adjustment wheel Electronic tip ejector Hotkey for memory places Softkeys for programming ee ee ee ee i Display 1 Softkey functions Current mode Pipetting volume and aliquots in multi dispensing mode Advanced Function ADV if selected Speed Battery indicator Arrow indicating pipetting direction won JJO Noo 0
3. e Pipetting Modes Advanced functions can be used in conjunction with the main pipetting modes as shown below Pipetting Mode Advanced Function Tracker Mixing Counter Excess Auto Repeated Volume dispensing Blow out Adjustment Pipetting v y v y Reverse Pipetting v y v Manual Pipetting y Multi dispensing v y v Diluting y v Sequential Dispensing y Multi aspiration v Titrate Protocols V v v v Only in Picus NxT models The availability of advanced functions depends on the pipetting mode selected for the protocol Press softkey for MENU Choose mode Press softkey for EDIT Change the setting and press softkey for OK or NEXT Start pipetting by pressing operating button Operation Button Functions Operating button Activates aspiration and dispensing Confirms the selected settings Adjustment wheel Enables fast and comfortable navigation in menu and volume setting Controls pipetting aspiration and dispensing operations in manual pipetting mode Softkeys Activate the programming functions To select the mode 1 Press the left softkey to open the menu then turn the adjustment wheel to scroll between modes 2 Press the operating button or the right softkey for OK to confirm the mode selection To edit the settings 1 Turn the adjustment wheel or press the middle softkey for EDIT to unlock the mode for editing 2 Change the settings by turning the adjustment
4. ment calibration at 10 50 and 100 of the nominal volume Adjustment is always done in Pipetting mode and the adjustment is valid in all modes when activated AG Sartorius Biohit Liquid Handling Oy Backlight changes the display backlight colour H 2 r ah TEUA User ID allows the user to personalise the pipette with an ID RELLE GLP Info only in Picus for setting and displaying the previous and next HA service and maintenance dates BEE MeN Reset resets pipette to factory settings Information displays the software version and battery charging level C Languages displays available user interface languages English French German Chinese and Russian Pipette lock only in Picus NxT locks the pipette e g in case of contamination If password protection is enabled the admin password is required to unlock the pipette Passwords only in Picus NxT 2 level passwords user level to limit the use of the pipette to a specific user and admin level to protect stored programs and other settings Date amp Time only in Picus NxT for setting date and time Reminders only in Picus NxT for setting service related reminders This Quick Start Guide is for basic operation of the Picus Picus NxT electronic pipettes For further information and more detailed instructions please refer to the user manual available at www sartorius com A hardcopy can be requested from lhinfo finland sartorius c
5. om The status of the information specifications and illustrations in this quick guide are subject to change Sartorius reserves the right to make changes to the technology features specifications and design of the equipment without notice All trademarks are Sartorius property unless otherwise stated Patents granted or pending Patents 109882 120861 4451311 10 0937762 255595 PL202091 006476 Sartorius Biohit Liquid Handling Oy Sartorius Lab Instruments GmbH amp Co KG Laippatie 1 Weender Landstrasse 94 108 00880 Helsinki Finland 37075 Goettingen Germany Phone 358 9 755 951 Phone 49 551 308 0 lhinfo finland sartorius com Fax 49 551 308 3289 www sartorius com Publication No SUL6017 e150804 Order No 85037 545 23 Ver 08 2015