Masterdock™ - Scott Safety
Contents
1. 11 3 4 1 Formatting SD ne UT 12 3 5 ANRE ME REN 12 3 6 Installing Software esses nnne 12 4 Ub deri nde der ETE 13 4 1 Bump Testing and 13 4 2 LET AINS and DISplay Seenaa 13 4 3 WASTCKOOCK WUY tinci deii oc eiusd 14 4 3 1 Connecting Masterdock to 14 4 3 2 Masterdock Utility Modes ui EE beasts 14 4 3 3 Masterdock Utility Main Display 15 4 3 4 Masterdock MENUS st isti rei ditiones 15 3A i Opara eoe 16 4 342 OHIO MENU onssa a 17 ASAS 17 zem m 18 5 Pans EIST and ovid da ud a e d adu 19 5 1 Cleaning and Routine Maintenance 19 5 2 20 5 3 21 li P N 087 0045 Rev E 01 10 CG T Masterdock Docking Station Instruction Manual This Page Left Intentionally Blank P N 087 0045 Rev E 01 10 CG T Masterdock Docking Station Instruction Manual List of Figures Figure 2 1 Masterdock
2. 1 Insert cylinder 2 Thread cylinder into Masterdock into place 3 Verify pressure gauge indicates greater than 50 psi Figure 3 2 Installing Calibration Gas 10 of 20 P N 087 0045 Rev E 01 10 Masterdock Docking Station Instruction Manual 3 4 Secure Digital Memory Card The Masterdock docking station comes with an SD memory card for electronic storage of all bump tests and calibrations for transferring of data to a PC The Masterdock will not perform a bump test or calibration with out the SD card installed and sufficient memory available Do not use the SD card for saving any other file types or data To install the SD card open the cover found on the bottom of the Masterdock next to the USB connection Verify the SD card is unlocked prior to installation Slide the SD card into the slot fully until it clicks and close the cover To remove the SD card push the card in and allow it to release then pull the card out Keep the SD port cover closed at all times when not removing or installing an SD card This will prevent dust and contaminants out of the data drive Figure 3 3 Installing SD Memory Card The SD memory card that comes with the Masterdock has been formatted for use New SD cards will need to be formatted prior to use with the Masterdock Replacement SD cards must not exceed 2GB NOTE CUSTOMERS MUST SUPPLY AN EXTERNAL SD MEDIA CARD READER TO ACCESS TH
3. 8 Figure 3 1 Masterdock Wall Mounting 0 9 Figure 3 2 Installing Calibration 10 Figure 3 3 Installing SD Memory 11 Figure 4 1 Masterdock Utility Main Display 15 Figure 4 2 Masterdock Utility Operation 16 Figure 4 3 Masterdock Utility Option 17 Figure 4 4 Masterdock Utility View 17 Figure 4 5 Masterdock Utility Help menu 18 P N 087 0045 Rev E 01 10 CG T Masterdock Docking Station Instruction Manual This Page Left Intentionally Blank VI P N 087 0045 Rev E 01 10 CG T Masterdock Docking Station Instruction Manual List of Tables Table 1 1 Acronym Quick Reference List 5 Table 2 1 Masterdock COMpOnents s oet ot oe rei aeta E 8 Table4 1 Status LED IndiGalloris o oi NE bed 13 Dir ME Pans 5 19 Table 5 2 Troubleshooting 20 VII P N 087 0045 Rev E 01 10 CG T Masterdock Docking Station Instruction Manual Record of Revisions 04 09 Initial release 08 09 Updated LED indications and troubleshoot
4. 8 Calibration Gas Dock 7 Protege Dock 10 Air Filter Battery Cover On Side of Unit 11 USB Port 6 Activation LEDs SD Card Slot IE AN 3 StatusLEDs 10 8 of 20 P N 087 0045 Rev E 01 10 Masterdock Docking Station Instruction Manual 3 Setup and Mounting 3 1 Mounting the Masterdock The Masterdock does not need to be mounted on a wall to operate It may be placed on any desk top table top or other horizontal surface and will operate However users may choose a more permanent location such as mounted on a wall To mount the Masterdock on the wall drill four holes per the dimensions shown in Figure 3 1 Mounting hardware must not exceed 14 5 mm in diameter and be capable of supporting the weight of the Masterdock NOTE THE MASTERDOCK WEIGHS APPROXIMATELY 8 5 LBS WITHOUT A CYLINDER OR PROTEGE DOCKED Allow for sufficient spacing around the Masterdock to utilize all the features such as installing calibration cylinders and accessing the battery compartment Tue 310 mm zz 4 5 115 Power Connection Ind Allow additional 15 380 mm for Rear Side Displayed cylinder clearance not drawn to scale Figure 3 1 Masterdock Wall Mounting Dimensions 9 of 20 P N 087 0045 Rev E 01 10 Masterdock Docking Station Instruction Manual 3 2 Powering the Masterdock The master do
5. P N 087 0045 Rev E 01 10 SCOIT Masterdock Docking Station Instruction Manual 5 3 Scott Repairs For quick and effective service of your instrumentation and to reduce time spent on repairs the SCOTT Instruments Service Department requires a Return Maintenance Authorization number be issued prior to any product being shipped for service repairs Contact the Service Department at Phone 1 800 247 7257 Fax 1 704 291 8330 Monday Friday 8 30 AM to 5 00 PM EST e mail techsupport scotths us tycoint com www scotthealthsafety com 21 of 20 P N 087 0045 Rev E 01 10 SCOIT Health amp Safety Monroe Corporate Center PO Box 569 Monroe NC 28111 Telephone 1 800 247 7257 FAX 704 291 8330 www scotthealthsafety com Printed in the USA Copyright 2010 SCOTT All Rights Reserved
6. Filter Verify the Air Inlet port on the bottom of the Masterdock is not operated without the filter in place The filter prevents dust and other contaminants from entering the Masterdock unit CAUTION OPERATING THE MASTERDOCK WITHOUT THE AIR FILTER IN PLACE MAY CAUSE CALIBRATIONS OR BUMP TESTS TO FAIL Periodically inspect and replace the air filter as necessary 3 6 Installing Software The Masterdock Utility software comes on the CD packaged with the docking station The utility is used to upgrade firmware set the date and time and to input calibration gas lot numbers for data recording Installation of the software onto a PC occurs in two stages First a driver must be loaded for the USB then the utility program is installed Insert the CD into a PC Once the start up screen appears click on the appropriate buttons and follow the on screen prompts 12 of 20 P N 087 0045 Rev E 01 10 Masterdock Docking Station Instruction Manual 4 Operation 4 1 Bump Testing and Calibration Prior to performing any bump test or calibration the Protege unit and the Masterdock must be powered on Verify the Battery LED is green on the Masterdock and the pressure gauge indicates greater than 50 psi Ensure the air inlet is not obstructed and the filter is clean 1 Open the door to the docking station and insert the powered on Prot g unit face down so that that charging contacts of the Prot g align with contacts i
7. any malfunction of or damage to THE PRODUCTS resulting from accident alteration misuse or abuse THIS WARRANTY IS MADE IN LIEU OF ALL OTHER WARRANTIES EXPRESSED OR IMPLIEDINCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE IN ADDITION SCOTT EXPRESSLY DISCLAIMS ANY LIABILITY FOR SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES IN ANY WAY CONNECTED WITH THE SALE OR USE OF SCOTT PRODUCTS AND NO OTHER FIRM OR PERSON IS AUTHORIZED TO ASSUME ANY SUCH LIABILITY 1 3 1 Contacting Scott Health amp Safety To contact Scott Health amp Safety call fax email or write to Scott Health amp Safety 4320 Goldmine Road Monroe NC 28110 Website www scotthealthsafety com Phone 800 247 7257 FAX 704 291 8330 4 of 20 P N 087 0045 Rev E 01 10 Masterdock Docking Station Instruction Manual 1 4 Acronyms Quick Reference Throughout the duration of this manual several acronyms are used Provided in Table 1 1 is a quick reference chart to quickly identify any acronym that may be unfamiliar to users Table 1 1 Acronym Quick Reference List MF uM Alternating Current Direct Current File Allocation Table Pounds b Pounds Liquid Crystal Display Sb oecure Digital USB Universal Serial Bus 5 of 20 P N 087 0045 Rev E 01 10 T Masterdock Docking Station Instruction Manual This Page Left Intentionally Blank 6 o
8. E DATA ON THE CARD SCOTT HEALTH amp SAFETY DOES NOT SUPPLY A READER Each record contains roughly 200 bytes of data allowing a 2GB memory card to hold thousands of records 11 of 20 P N 087 0045 Rev E 01 10 Masterdock Docking Station Instruction Manual 3 4 1 Formatting an SD Card If the Service Requested LED blinks orange on the Masterdock or if you are using a new SD card for the first time the SD card may need to be formatted The Masterdock does not automatically format a card it must be formatted using a PC SD cards typically come formatted by the manufacturer but for an SD card to operate properly in the Masterdock it must be formatted for FAT32 a 32 bit file allocation system If the card is formatted for FAT or FAT16 or unformatted it must be formatted for FAT32 Before using an SD card or checking whether it needs formatting verify the sliding locking tab is in the unlocked position as shown in Figure 3 3 NOTE FORMATTING AN SD MEMORY CARD WILL ERASE ALL DATA ON THE CARD BACKUP ANY DATA PRIOR TO FORMATTING Verify the lock tab on the memory card is unlocked and connect the SD memory card to a PC Open the My Computer icon on the desk top Locate the SD card it is typically identified as Removable Disk Right click and select Format When the Format Disk window appears set or verify the File System is set to FAT32 and select Start The PC will notify when the format is complete 3 5 Air
9. PORS THE MASTERDOCK IS NOT FOR USE WITH ANY OTHER PRODUCT THAN THE PROTEGE PORTABLE GAS MOINITOR CAUTIONS VERIFY THE GAS INLET PORT HAS A CLEAN FILTER ATTCHED AND AN UNOBSTRUCTED FLOW PATH PRIOR TO USE DO NOT USE A MAGNET NEAR THE SD MEMORY CARD TO PREVENT LOSING OR CORRUPTING DATA DO NOT USE ANY SOLVENTS CHEMICALS OR POLISHES CONTAINING SILICON COMPOUNDS TO CLEAN THE MASTERDOCK AS THESE CAN CAUSE DAMAGE TO THE SENSORS OF THE PROTEGE 1 2 2 General Rules for Use of the Masterdock It is critical that this docking station be used and maintained properly Failure to do so could jeopardize the people whose lives depend on its operation and cause injury or death USE THE DOCKING STATION CORRECTLY Only use the Masterdock to bump test and calibrate Scott s Prot g portable gas monitor POWER MUST BE ON TO WORK The Masterdock will only function while powered on VERIFY CALIBRATION CYLINDERS ARE SUFFICIENTLY CHARGED Calibration cylinders must have at least 50 psi of pressure to accurately calibrate or bump test Prot g units DO NOT USE CALIBRATION GAS BEYOND THE EXPIRATION DATE Verify Calibration gas has not exceeded its expiration date prior to use DO NOT USE IN UNKNOWN OR POTENTIALLY HAZARDOUS ATMOSPHERES Always verify an atmosphere is free of toxic or combustible gases prior to powering up the Masterdock Keep the air filter clean to prevent contaminants from entering the units air system 3 of 20 P N 087 0045
10. Rev E 01 10 Masterdock Docking Station Instruction Manual 1 3 Warranty Statement SCOTT HEALTH amp SAFETY LIMITED WARRANTY ON Masterdock Docking Station scott Health amp Safety SCOTT a division of Scott Technologies Inc warrants it s MASTERDOCK DOCKING STATION THE PRODUCTS to be free from defects in workmanship and materials for a period of one 1 year from the date of original manufacture by SCOTT This warranty applies to all components of THE PRODUCTS supplied at the time of original sale of THE PRODUCTS SCOTT s obligation under this warranty is limited to replacing or repairing at SCOTT s option THE PRODUCTS or components shown to be defective in either workmanship or materials Only personnel of SCOTT or when directed by SCOTT authorized SCOTT agents are permitted to perform warranty obligations This warranty does not apply to defects or damage caused by any repairs of or alterations to THE PRODUCTS made by owner or any third party unless expressly permitted by SCOTT product manuals or by written authorization from SCOTT To obtain performance under this warranty and as a condition precedent to any duty of SCOTT the purchaser must return such products to SCOTT a SCOTT authorized distributor or a SCOTT authorized service center Any product returned to SCOTT shall be sent to SCOTT HEALTH amp SAFETY Attn Warranty Claim Dept P O Box 569 Monroe NC 28111 This warranty does not apply to
11. ck must be powered from an external power supply plugged into the rear side receptacle or with 4 C cell batteries External power supplies may be plugged into the wall or to a 12V cigarette lighter socket Refer to Figure 3 1 for the location of the AC power connection The battery compartment is located on the side of the Masterdock Push up on the tab and then pull up on the door to remove the cover Rechargeable batteries may be used however the Masterdock will not recharge those batteries when connected to an external source Rechargeable batteries must be removed and charged in an external charger 3 3 Connecting Calibration Gas Insert either 58L or 34L cylinder of calibration gas into the Calibration Gas Dock and thread into place NOTE IF USING THE 58L CYLINDER THE RUBBER SLEEVE MUST BE REMOVED FROM THE CALIBRATION GAS DOCK The Gas Gauge will indicate cylinder pressure Cylinder pressure must be above 50 PSI to perform a calibration or bump test There is no isolation valve The Masterdock controls flow automatically through a solenoid valve The cylinder does not need to be removed between bump tests or calibrations When cylinder contents fall below 50 psi remove from dock simply by unthreading the cylinder Pull UI SAL O Power loe YY O Seven p X MASTERDOCK lt _ MASTERDOCK NI LJ Lu _ lt MASTERDOCK J
12. f 20 P N 087 0045 Rev E 01 10 T Masterdock Docking Station Instruction Manual 2 Introduction The Masterdock is an automated bump testing and calibration station designed exclusively for performing bump tests and calibration of Scott s Prot g portable gas monitors The Masterdock allows users to quickly and effectively verify the proper operation of the Prot g with the touch of a button Removable memory in the Masterdock logs all calibrations and testing for accurate record keeping The Masterdock can be powered using the included AC adapter or by installing four alkaline C cell batteries The Masterdock eliminates the need to connect a calibration adapter and a remote calibration cylinder to the Prot g A calibration cylinder is installed into the docking station allowing for multiple bump tests and calibrations of Prot g units simply by docking a Prot g into the Masterdock and pressing a button The Masterdock can be used on any flat surface such as a desk top or can be mounted on a wall Wall mounting instructions are included in this guide Refer to Figure 2 1 7 of 20 P N 087 0045 Rev E 01 10 Masterdock Docking Station Instruction Manual 1 3 4 9 xX Q lt 7 a O 9 gt 11 12 Kc 8 10 Figure 2 1 Masterdock Components Table 2 1 Components Component Air Outlet Bump and Calibration Activation Buttons Calibration Gas Gauge
13. ger Click on the icon to the left of Ports Identify the COM number associated CP210x to UART Bridge Controller This COM port number is necessary for configuring the Masterdock Utility 6 Open the Masterdock Utility software 7 Select Operation gt Setup Port Select the COM number previously identified 4 3 2 Masterdock Utility Modes The Masterdock utility can be used in User Mode or Administrator Mode All options available in User Mode are available in Administrator Mode However Administrator Mode also allows users to set the Gas Lot Number set the Location synchronize the Masterdock to the time on the PC and configure the Masterdock to perform a calibration in the event of a bump test failure The Administrator mode is password protected To select Administrator Mode Select Operation gt Operation Mode gt Administrator Mode At the password prompt screen a Change Password option is available NOTE THE DEFAULT PASSWORD FOR THE ADMINISTRATOR MODE IS SCOTT ALL CAPITAL LETTERS 14 of 20 P N 087 0045 Rev E 01 10 SCOIT Masterdock Docking Station Instruction Manual 4 3 3 Masterdock Utility Main Display The Masterdock Utility Main Display shows Firmware Revision the date and time Gas Lot Number serial Number and Location of the Masterdock When in Administrator Mode the options to sync time with the PC clock and Calibrate if a Bump test fails become available A toolbar located at the top of the displa
14. ication Reason Recommended Action s All LEDs extinguished Service Request LED is red unable to perform a Bump Test or Calibration LEDs indicate normal operation Green however unable to perform a Bump Test or Calibration Pump does not operate Masterdock Utility fails to open COM port Masterdock is not receiving power SD Card problem OR Date Time Synchronization is required Empty calibration cylinder OR Defective gauge R Obstruction in inlet line Faulty pump OR Obstructed line Port has not been USB driver has not been installed correctly Verify power source If battery operated verify batteries have sufficient charge If powered externally verify power source is connected Verify AC adapter is not damaged Check for loose or broken wires Verify SD card is installed Connect SD card to PC and remove data to clear memory on card Replace SD memory card synchronize the Masterdock date time using the Masterdock Utility Replace cylinder Verify gauge indicates properly with a full tank Verify inlet line and air filter are clear of obstructions Check pump wiring clear obstructions from air inlet air outlet and port inside docking station door to Protege or replace pump Refer to Paragraph 3 6 and Paragraph 4 3 Replace Calibration Gas bottle Refer to Pump is stalled Cal Bump and Cal LED flash red Gas bottle is empty Paragraph 3 3 20 of 20
15. ing table 09 09 Updated P N s 10 09 Updated P N s 01 10 Updated Table 4 1 and 5 2 vill P N 087 0045 Rev E 01 10 Masterdock Docking Station Instruction Manual 1 Safety and General Information WARNING ALL INDIVIDUALS WHO HAVE OR WILL HAVE RESPONSIBILITY FOR USING OR TESTING THIS PRODUCT MUST READ AND UNDERSTAND THE CONTENTS OF THIS MANUAL THE PRODUCT WILL PERFORM AS DESIGNED ONLY IF USED AND TESTED IN ACCORDANCE WITH THE MANUFACTURER S INSTRUCTIONS FAILURE TO FOLLOW MANUFACTURER S INSTRUCTIONS WILL RENDER THE WARRANTY AND APPROVALS NULL AND VOID FAILURE TO FOLLOW THESE INSTRUCTIONS MAY ALSO RESULT IN SERIOUS INJURY OR DEATH 1 1 Important Notices scott Health and Safety can take no responsibility for use of its equipment if it is not used in accordance with the instructions If further operational or maintenance details are required but not provided in this manual contact Scott Health and Safety or their agent Scott Health and Safety shall not be liable for any incidental or consequential damages in connection with any modifications errors or omissions in this manual While every effort has been made to ensure accuracy in this owner s manual no responsibility can be accepted for errors or omissions This publication is not intended to form the basis of a contract and the company reserves the right to amend the design content and specifications of the detector without notice Additionally indu
16. n the docking station door 2 Close the door to the docking station 3 Verify the Bump and Cal LEDs are lit green If not open door and verify the Prot g is firmly seated in docking station 4 Push Bump to perform a bump test or Cal to perform a calibration The LED above the selected button will turn off and the status LED will begin blinking orange 5 When the status LED lights green the Prot g unit has completed testing If the LED lights red instead of green the test has failed 4 2 LED Alerts and Displays Refer to Table 4 1 for Status LED indications and their meanings Table 4 1 Status LED Indications Status LED Green 9 Orange Blinking Low Battery external power not A Dowel p external power is supplied sufficient SD card is locked OR SD cards needs formatting OR SD memory is SD card is not installed OR SD card memory is full OR a communication failure has occurred OR The date time needs Service Request synchronization with the OR synchronize the P date time using the Masterdock Utility Masterdock Utility 13 of 20 P N 087 0045 Rev E 01 10 Masterdock Docking Station Instruction Manual 4 3 Masterdock Utility 4 3 1 Connecting Masterdock to a PC Use the supplied USB cable to connect the Masterdock to a PC Right click on the My Computer icon found on the desktop and select properties 3 Select the Hardware Tab Select Device Mana
17. nie COT Instruction Manual Masterdock Docking Station Service Request i ae WARNING READ AND FOLLOW THE ENTIRE CONTENT OF THIS MANUAL PRIOR TO USE FAILURE TO DO SO MAY RESULT IN SERIOUS INJURY OR DEATH COPYRIGHT 2009 SCOTT ALL RIGHTS RESERVED P N 087 0045 Rev E 01 10 CG T Masterdock Docking Station Instruction Manual This Page Left Intentionally Blank P N 087 0045 Rev E 01 10 CG T Masterdock Docking Station Instruction Manual Table of Contents ts satety and General INIOFMAUNOM 1 1 1 S 1 1 2 Warnings Cautions and cies re rx rra n ER XR re XE Pa PER ERR EE 2 1 2 1 General Warnings and Cautions 3 1 2 2 General Rules for Use of the Masterdock 3 1 3 Waramy Statemen 4 1 3 1 Contacting Scott Health amp Safety eese 4 1 4 Acronyms Quick Referente s Deu don 5 25 7 9 SeTUD ANG MOUNN 9 3 1 Mounting the 9 3 2 Powering the 10 3 3 Connecting Calibration 10 3 4 Secure Digital Memory Card
18. on Clock from PE Yes Ho Calli 5 Ho Ready NUM Figure 4 2 Masterdock Utility Operation Menu 16 of 20 P N 087 0045 Rev E 01 10 T Masterdock Docking Station Instruction Manual 4 3 4 2 Option Menu Language Used to Select which language for the Masterdock utility to display Masterdock Utility 1 0 Bl xl Operation Option View Help bm Chinese Simplified w English Gas Lot Setting Gas Lot No Information Firmware Ver Date Time Setting SH Location ClockfromPC Yes No Cal lf Bump tails C Yes No Figure 4 3 Masterdock Utility Option Menu 4 3 4 3 View Menu Toolbar Used to display or hide the Toolbar Status Bar Used to display or hide the Status Bar Masterdock Utility 1 0 Oj x Operation Option view Help amp P Toolbar Status Bar Gas Lot Setting Gas Lot Information Firmware Wer Date Time Setting SH Location Clock fom PE ves No Calli Bump fails C Yes No Ready m Figure 4 4 Masterdock Utility View Menu 17 of 20 P N 087 0045 Rev E 01 10 CG T Masterdock Docking Station Instruction Manual 4 3 4 4 Help Menu About Masterdock Utility Used to display the current software revision level of the Masterdock Utility Masterdock Utility 1 0 nml x Ope
19. ration Option view Help About Masterdack Utility 1 0 Information Firmware Date Time Setting SM Location Gas Lot Setting Gas Lot No Clock fom PE ves Calli ane C Yes Ho Ready NUM Figure 4 5 Masterdock Utility Help menu 18 of 20 P N 087 0045 Rev E 01 10 Masterdock Docking Station Instruction Manual 5 Parts List and Maintenance Table 5 1 Parts List Scott Portable gas Detection CD Masterdock Software and Instruction Manual Included Calibration Gas 34L 077 0351 Calibration Gas 58L 077 0352 5 1 Cleaning and Routine Maintenance The Masterdock is designed to be free of routine maintenance unless operated with batteries As needed replace batteries with C Cell batteries Rechargeable batteries may be used in the Masterdock however batteries will not recharge when installed in the Masterdock Batteries must be charged with an external charger CAUTION NEVER LEAVE DEAD BATTERIES INSTALLED IN THE MASTERDOCK AS THEY MAY LEAK OR CORRODE AND CAUSE DAMAGE TO COMPONENTS Clean the Masterdock as needed using only compressed air or a clean dry cloth If needed a dampened cloth may be used for external cleaning however do not use any cleaners or solvents 19 of 20 P N 087 0045 Rev E 01 10 SCOIT Masterdock Docking Station Instruction Manual 5 2 Troubleshooting Table 5 2 Troubleshooting Chart ind
20. stry standards codes and legislation are subject to change Updated copies should be obtained by users to ensure the most recently issued regulations standards and guidelines are available 1 of 20 P N 087 0045 Rev E 01 10 T Masterdock Docking Station Instruction Manual 1 2 Warnings Cautions and Notes Throughout this document warnings cautions and notes have been interspersed to draw attention to potentially unsafe hazardous or unique situations that require user attention Each warning caution or note is labeled and quickly identified using an icon WARNING INDICATES A POTENTIALLY HAZARDOUS SITUATION WHICH IF NOT AVOIDED COULD RESULT IN DEATH OR SERIOUS INJURY CAUTION INDICATES A POTENTIALLY HAZARDOUS SITUATION THAT NOT AVOIDED MAY RESULT IN MINOR OR MODERATE INJURY IT MAY ALSO BE USED TO ALERT AGAINST UNSAFE PRACTICES NOTE HIGHLIGHTS VARIOUS INSTANCES WHERE AN ATTENTION TO DETAIL IS CRITICAL TO PRODUCT PERFORMANCE 2 of 20 P N 087 0045 Rev E 01 10 Masterdock Docking Station Instruction Manual 1 2 1 General Warnings and Cautions The following list of warnings and cautions pertain to the general use and care of the Masterdock Failure to follow these warnings and cautions may result in death injury or poor equipment performance WARNINGS USE THE MASTERDOCK ONLY IN AN ATMOSPHERE THAT HAS BEEN VERIFIED TO BE FREE OF TOXIC OR COMBUSTIBLE GASES OR VA
21. y has three quick use buttons Get Configurations Upgrade Firmware and Help Masterdock Utility 1 0 loj x Operation Option View Help ka PU Information Firmware Ver Date Time Setting SH Location Gas Lot Setting Gas Lot Clock from PE Yes No Cal If Set Ready Figure 4 1 Masterdock Utility Main Display 4 3 4 Masterdock Utility Menus Four Menus are available for use in the Masterdock Operation Option View and Help Not all selections will be available depending on which mode is selected 15 of 20 P N 087 0045 Rev E 01 10 T Masterdock Docking Station Instruction Manual 4 3 4 1 Operation Menu Setup Port Used to change the COM port Refer to Paragraph 4 3 1 for more information Get Configurations Used to poll the Masterdock about current information for display including the Firmware Revision Gas Lot Number Serial Number and Location of the Masterdock Operation Mode Used to select either user mode or Administrator Mode Refer to Paragraph 4 3 2 for more information Upgrade Firmware Used to Upgrade the Firmware installed in the Masterdock Masterdock Utility 1 0 E iol xl Operation Option View Help Setup Pork Get Configurations Fe tion Mad k BM ecd poc ARN Gas Lot Setting Gas Lot No Upgrade Firmware Date Time Setting ch Locati
Download Pdf Manuals
Related Search
Related Contents
Lightolier IS:2002 User's Manual Hotpoint WMUD 10637P washing machine Nikon Zoom Wide Angle-Telephoto AF Zoom Nikkor 28-80mm f/3.3-5.6G Autofocus Lens - AF28-80_3.3 Rev. AG+ System Board User's Manual パーフェクトトレジャー取扱説明書 Mode d`emploi et instructions de montage Four avec micro Acer 08P4100 Network Card User Manual Manuel d`installation et d`utilisation Copyright © All rights reserved.