        USER`S AND M AINTENANCE M ANUAL
         Contents
1.       Serial number    e Use by  lt date gt     ro 8 Ble D    www icaretonometer com       LOT    STERILE R    x   G    e Keep dry    e Manufacturing date       Lot number    e Sterilized using irradiation    e Stand by    e Do not discard this product with other household type waste    e Send to appropriate facility for recovery and recycling    e EU WEEE  European Union Directive for Waste of Electronic  and Electrical Equipment     14         
2.  case you end up taking  multiple measurements for some reason  the eye details can be  incorrect and must be fixed  See Icare LINK user manual      www icaretonometer com       Instruction for the patient    Switch on the unit  load the probe   Go in front of a wall mirror   Take the tonometer in your hand with the white  main button pointing upwards   Move the tonometer sideways toward your eye so  that you see in the mirror the white tip of the probe  in front of the black pupil   Move the tonometer closer so that the forehead  and cheek supports touch the skin   Check that the white tip of the probe is seen in  front of the center of the cornea   Turn the tonometer from the sideway position so  that the tonometer with the loaded probe points  straight to the center of the cornea  pupil    Open your eyes wide and try not to blink your eye  while measuring   Press the measurement button once  You will get  the final result after the probe touched you eye for  six times and after one long beep  In automatic  mode the six interim measurements are taken in  sequence     A new measurement needs to be taken if     REPEAT ERROR LED is displayed  for example  if the  probe made contact with the eyelid        10    Icare tonometer ONE manual English       CLEANING AND DISINFECTION    The forehead and cheek supports and the eye support must be cleaned for each new patient with disinfectant for example 70  alcohol solution  using wipe  Do not immerse the tonometer in water or other l
3.  inch      Press the measurement button lightly to perform the    measurement  taking care not to shake the tonometer     The tip of the probe should make contact with the  central cornea  pupil   The way you take  measurements depends on the chosen mode    In normal mode a value is shown briefly after each  successful measurement  When the six measurements  in sequence are completed and the measurement  valid  the final value of IOP is displayed  If there is too  much variation between the measurements  the  REPEAT ERROR LED is displayed and the measurement  needs to be performed again  The device stores the  quality of the measurement in its internal memory   which is only accessible by downloading to a PC using  the Icare LINK software    A new measurement needs to be taken if     REPEAT ERROR LED is displayed  for example  if the  probe made contact with the eyelid    Encountering unusual values  for example  over   22 mmHg or below 8 mmHg     The tonometer   s internal memory stores the time and  date of a measurement  the measurement result and  the quality level of the measurement  This information  can be uploaded to a PC through a USB cable using  Icare   LINK software  If a patient does not have the  software  the tonometer is taken to a health care  professional who can transfer the data   NOTE  When using Icare LINK software  it assumes by default that the  first accepted measurement is for the right eye and the second  accepted measurement is for the left eye  In
4. 1 ranges   5 7 mmHg  7 10 mmHg  10 14 mmHg  14 18 mmHg  18 21 mmHg  21 24 mmHg  24 27 mmHg  27 30 mmHg  30 35 mmHg  35 40 mmHg  40 50 mmHg    SETTING UP THE TONOMETER BEFORE TAKING MEASUREMENTS    Before taking measurements your tonometer  the measurement position must be adjusted correctly by health care professional  Set up includes   e Choosing a measurement mode  e Adjustment of the measurement position  e Loading the probe before each measurement    CHOOSING A MEASUREMENT MODE    The device can operate in two modes    Normal   e Normal mode is used to take individual measurements one at a time  The measurement button must be pressed once to initiate each of  the six measurements to obtain final IOP reading  Normal mode is generally used when measurements are taken by someone other than  the patient    Automatic   e Automatic mode is especially useful in self tonometry  In automatic mode a single press of the measurement button initiates the  measurement function and the tonometer takes six measurements automatically to obtain final IOP reading     A NOTE     Device is designed to perform six subsequent movements of probe to cornea  measurements after which the device calculates the final IOP range   The final IOP range is displayed on device LED panel     To change the measurement mode     1      Turn on the tonometer   2  Inserta straight pin into the mode switch hole on the side of the tonometer     Mode switch hole       Measurement mode switch    3  Push the switch w
5. A CAUTION     Federal  US  law restricts this device to sale by or on the order of a physician or properly licensed practitioner     e When you have opened the package  check for any external damage or faults  particularly for damage to the case  If you suspect  that there is something wrong with the tonometer  contact the dealer    e Use the tonometer only for measuring intraocular pressure  Any other use is improper and the manufacturer cannot be held liable  for any damage arising from improper use  or for the consequences thereof    e Never open the casing of the tonometer  except for the battery compartment or to change the probe base    e Never use the tonometer in wet or damp conditions nor allow the tonometer to get wet    e The probe base  battery compartment cover  screws  collar and probes are so small that a child could swallow them  Keep the  tonometer out of the reach of children    e Do not use the device near inflammable substances  including flammable anesthetic agents       Certain microbiological agents  for example  bacteria  can be transmitted from the forehead  eye or cheek support  To avoid this   the forehead and cheek support and eye piece must be cleaned for each new patient with disinfectant  for example an alcohol  solution    e The tonometer conforms to EMC requirements  IEC 60601 1 2   but interference may occur in it if used near   lt 1m  a device  such as  a cellular phone  causing high intensity electromagnetic emissions  Although the tonomete
6. ENGLISH     care    EEEETEIEIE ONE    USER S AND MAINTENANCE MANUAL          Icare tonometer ONE manual English       Icare   ONE  Model  TA02  INSTRUCTION MANUAL v1 12 01 14 EN    A        0044    This device complies with   Medical Device Directive 93 42 EEC  Canadian Medical Device Regulations       Copyright    2012 Icare Finland Oy  Made in Finland    Icare Finland Oy     yritie 22  FI 01510 Vantaa  Finland   Tel   358 9 8775 1150  Fax  358 9 728 6670  www icarefinland com  info icarefinland com    TABLE OF CONTESTS   Table of CONTESTS IRI DRE A ee nen Cen RR IE RR II RI RR ee re 2  Safety WN SUF UNG CIN morire poimaine dia p  n rata ina cabine dia auf neant aie  e nai e vse nr i illa uta lieu simi i  i nu ist 2  Indications TDF  IS Eaton ema lia mata da mata mata aa tata atata ma ta aa aaa tau aaa 3  E je je Bote aha gel  a ce   iale ae   NS IRENE RER RR RR PPR RR E IRI RR ANR RR PI RARI RE IRI NE IRINA PERIE TI INT INN OR INI PN RI E NR IVI 4   ip FURS 5 Lo RIP IRI PORN IRINA ON OI PI a pele ae E PI e e ae aia ep ilps na pe E e ae ei a aly ee i a e daly eae ete E Poeta 4  Package COnNtents ennen TRI CE RR ERORI REC RI 0 RR RR EN RP MIRI IN RR RER RI E RR MRI RDI RECE PIE SPRE ECE NI RISC MIINE AER RR N a INN RER RR 4  Turnin Ne device ON ie cama pi alai i aa ri a af ar dn sa i a ara ri ia i ia dat 5  Setting up the tonometer before taking measurements sea esuate aaea oaia tame aaa ie mea ea adie a dle dala iai ea aa 80 ai pl ea dar ta aa ai aia aie sara da 6  Choosing 
7. a measurement ModE RR RR RR RI RR RR RO RI RR N RR RI II N NINO RR RE III RR RR RR RR RI NR PRR RI IRI IP IRI RR 6  Adjusting the measurement DOSIRIIOFI ceasca als do limes a i iun aul fa este ionii a dala iai i cd all aa 6  Loading the OS i miar ca acea mac aaa nica a aa a i ca ua a a usa nu at a dau ini na nn at at bg daia e ta 7  Measunne intraocular Pressure  ri aranera E PI RI RR RR ROI e RI RR EP RR NE O RR N PRR RR RIN A NE IN RANI NR RAE RONI 8  Patient Use MMStrUCtIONS vis coaie tite da bca i n i ra dn n 5 a nl ia aa 80 a a   n al Sa E atac 9  Eirormessages SMA Ne AOS cier ERROR ORE RER ERORI RER RE RD RR PR RN RI RR RR IE E RER IER RR N N 10  Differences between professional user and the patient using the Icare ONE Tonometer             eee nana 10  Cleaning and disinfection PRPRREDE PRR PRIME RE RR RI RR RIN e PRR EP OR PR IE ROI PAR OR PR Ie n RU PR EI PRR IA 11  Changing the Dei EREI IRIS eta ital a li al lt tt i a aa aa d ca a tc aaa aaa aa tat a it la et a 11  Service O PO GEICIRI   E  6 ee ee im ea ni mea dala i i aa ci N i n ata a a i te lit ii ia ec a tal 12  Spare pan Sio oaste to maus asta E E E ea pia a atat ma dat pia atat pa a i pal 12  TecHnicakand performance data merer aoaea EEA EE EEEE E E A EA 12  Glinicalpertormance CIEL ie es eines ae A E EAE EEE E E A EEE EEE EEE EERE E E AREE AEE 13  SYMBOLS 99 O RI RR NR RR RR A ERE RN NI EEE E get E N RR E EEE EE EE E E 14  SAFETY INSTRUCTIONS    N saune    The tonometer must not come into contact with the pa
8. button once  You will get the final result after the probe touched you eye for six times and after one long beep     In automatic mode the six interim measurements are taken in sequence     A CAUTION     Always perform measurement on RIGHT eye first followed by LEFT eye  Complete the RIGHT eye measurement sequence before changing to the  LEFT eye     www icaretonometer com    English       Icare tonometer ONE manual English       A new measurement needs to be taken if     e REPEAT ERROR LED is displayed  for example  if the probe made contact with the eyelid     e Encountering unusual values  for example  over 22 mmHg or below 8 mmHg     ERROR MESSAGES AND INDICATORS    Icare Tonometer ONE display panel shows the following LEDs     LED Cs State   Description and Action    PROBE BLUE blinking Load probe  oE  eo OOO       CLEAN RED lit Clean or change the probe base    DIFFERENCES BETWEEN PROFESSIONAL USER AND THE PATIENT USING THE ICARE ONE    TONOMETER    Instructions for the health care professional    Check that the tonometer is set up correctly by a  doctor or a nurse    The patient should relax  and look straight ahead at a  specific point while keeping eyes wide open    Bring the tonometer near the eye  Move the  tonometer closer so that the forehead and cheek  support touch the skin  Check that the white tip of the  probe is seen in front of the center of the cornea   pupil   The distance from the tip of the probe to the    center of the cornea must be 4 8 mm  5 32 5 16
9. ent   Temperature   10   C to  35   C   Relative humidity  30   to 90     Atmospheric pressure  800 hPa     1060 hPa     Storage environment   Temperature   10   C to  55   C   Relative humidity  10   to 95     Atmospheric pressure  700 hPa     1060 hPa  e Transport environment   Temperature   40   C to  70   C   Relative humidity  10   to 95     Atmospheric pressure  500 hPa     1060 hPa    CLINICAL PERFORMANCE DATA    Performance data is obtained from a clinical study  performed according to ANSI Z80 and ISO 8612 for tonometers   The mean of paired difference  Goldmann lcare ONE tonometer  were 0 6   lt 16 mmHg 0 1   gt 16 lt 23 1 0   gt 23 0 8  and and standard deviation is 3 1    Bland   Altman plot for IOP values of Goldmann   tonometer vs Icare One   tonometer    AQ      4040404 E 4O  Coon  0 004440     9 44  4040     mee  a    e o o   e     e eee    Difference  IOP  One      IOP  GAT         Mean IOP  GAT   IOP  ONE    Frequency Count 999   AAA  EENG   0604 ttt 5 FHKE    Scatterplot of IOP values of test tonometer against the IOP values of Goldmann reference tonometer  With regression line and 95   confidence intervals    IOP lcare    One    IOP_ONE   0 00077190   1  0290 I0P GAT       0 0 20 30 40 50 60 70  IOP Goldmann  Frequency Cont       AAA  EEE    6064 8     5 HH e    13  www icaretonometer com         Icare tonometer ONE manual English       SYMBOLS    e Caution       See operating instructions  for more information    BF type device    e Single use disposable 
10. iquid     If the tonometer displays a CLEAN error  CLEAN LED   the probe base  see Figure below  may be dirty or dusty  To clean the probe base first unscrew    the probe base off the tonometer  Carefully inject isopropyl alcohol  rubbing alcohol  through the top of the probe base  Dry the probe base by  injecting air into the probe base and drying the part gently  If it is not possible to clean the probe base  then replace it with a new one     Probe base          le      Tonometer probe base          Probe base             Tonometer probe base unscrewed    CHANGING THE BATTERIES    Lift the silicone lid that protects the USB port and keeps the battery compartment cover in place  Open the battery compartment cover by pressing  the silicone lid slightly and sliding the battery compartment cover as shown in figure        Battery compartment  cover          Silicone lid          Opening the silicone lid and battery cover    Insert two CR123A lithium batteries in the correct order      end upwards as shown in figure  Close the cover firmly and press the silicone lid in place  to cover the USB port     A NOTE    Make sure you use batteries with built in PTC protection  For Example Energizer Lithium Photo 123 3V CR123A     11  www icaretonometer com       Icare tonometer ONE manual English          Inserting new batteries    SERVICE PROCEDURES    Replace the batteries when the device indicates low battery voltage     Clean or change the probe base if the probe does not move smooth
11. ith mode switch key provided with the tonometer on the bottom of the mode switch hole of the tonometer until the  mode change signal is heard  The switch must be pressed for approximately three seconds for the mode to change     The audible sound indicates the measurement mode        one beep  normal mode  e three beeps  automatic mode    ADJUSTING THE MEASUREMENT POSITION    The tonometer has two adjustable supports  one for the forehead and one for the cheek as shown in figure  These supports are adjusted to ensure  accurate measurement distance and alignment  Adjust the supports using the adjustment wheels  so the distance from the tip of the probe to the  surface of the cornea is 4 8 mm  5 32 5 16 inch  and the probe is horizontal     www icaretonometer com    English       Icare tonometer ONE manual English            lt  Forehead support    Adjustment wheels    _  lt    Cheek support    Tonometer forehead and cheek supports with adjustment wheels    An optional eye support can be used for easy alignment and additional stability during measurement  When the eye support is used  the cheek  support is not used and needs to be adjusted to minimum setting  The eye support can be used for both eyes by rotating it from one side         lt  Eye support  Forehead and eye support  with the cheek support in minimum position    LOADING THE PROBE    The Icare ONE tonometer TA02 uses disposable tonometer probes as shown in figure  Each probe is meant for single use only  The probes a
12. ly  CLEAN LED is lit      A NOTE    No other service procedures can be carried out by the user  All other service and repair must be carried out by the manufacturer or a certified  service center     SPARE PARTS    Single use probes  10 included with the tonometer when purchased   Probe base kit   Eye supports   Mode switch key   USB cable   Batteries   carry on case    These can be purchased from a local distributor of Icare products     TECHNICAL AND PERFORMANCE DATA    Type TA02   The device conforms to CE regulations   Dimensions  approximately 11 cm x 8 cm x 3 cm   Weight  approximately 150 g   Power supply  2eCR123 batteries  make sure you use batteries with built in PTC protection  For Example Energizer Lithium Photo   123 3V CR123A    Measurement range  5 50 mmHg   Display range  5 50 mmHg   Accuracy  95  tolerance interval relatively to manometry     1 8 mmHg   Repeatability  coefficient of variation    lt  9    IOP  lt 20 mmHg   lt  3    IOP  gt 20 mmHg   e Accuracy of display  display is divided into 11 ranges    5 7 mmHg   7 10 mmHg   10 14 mmHg   14 18 mmHg   18 21 mmHg   21 24 mmHg   24 27 mmHg   27 30 mmHg   30 35 mmHg   35 40 mmHg   40 50 mmHg   Display unit  mmHg   The serial number is located on the inside of the battery compartment cover   There are no electrical connections from the tonometer to the patient   The device has BF type electric shock protection    12  www icaretonometer com         Icare tonometer ONE manual English          Operation environm
13. nd by patients under supervision and control of the health care professional  Patient use is by prescription only     www icaretonometer com    Icare tonometer ONE manual English       USAGE CONTRAINDICATIONS    The TAO2 tonometer is contraindicated in the presence of the following conditions   Corneal Scarring  Microphtalmos  Buphthalmos  Nystagmus  Keratoconus  abnormal central corneal thickness  and topical anesthetics     INTRODUCTION    Icare ONE tonometer is intended to monitor intra ocular Pressure  IOP  of the human eye  It is indicated to be used by health care professionals and  by patient under prescription use only     It is based on a patented  induction based rebound method  which allows intraocular pressure  IOP  to be measured accurately  rapidly and without  an anesthetic     Icare One uses sterilized disposable probes in order to reduce the risk of microbiological contamination to a negligible level    Intraocular pressure changes are due to the effects of pulse  breathing  eye movements  and body position  Several measurements are taken using  the handheld device within fractions of a second to average the micro changes of IOP that occur during measurement  In the automated mode  six  pre programmed measurements are needed to obtain an accurate reading     Icare ONE tonometer can be used by glaucoma patients to monitor their intraocular pressure  Results are stored in the internal memory of the  tonometer and can be transferred to a PC through a USB cable u
14. r   s own electromagnetic emissions are  well below the levels permitted by the relevant standards  they may cause interference in other  nearby devices  for example   sensitive sensors       If the device is not to be used for a long time  remove its batteries  as they may leak  Removing the batteries restores the device to  normal measurement mode    e For every patient to be measured take a new probe to avoid cross contamination of bacteria or viruses and infection of the eye   e Use probes taken only from original intact packaging  Sterility of the probe cannot be guaranteed once the seal is compromised    e The tonometer probes are labeled single use only  Re sterilization or re use of the probe can result in incorrect measurement or  damage to the probe head    e Be sure to dispose of the single use probes properly  for example  in a container for disposable needles   Probes may become  contaminated with micro organisms after one time use    e Do not connect the USB cable during measurements  When connected it is impossible to perform measurement    e Do not change the batteries or probe base when the USB cable is connected    e Batteries  packaging materials and probe bases must be disposed of according to local regulations  Use only original probes  see  Figure 5  and accessories     INDICATIONS FOR USE    Icare ONE tonometer  TA02  is indicated for the monitoring of intraocular pressure  IOP  of the human eye  It is intended to be used by health care  professionals a
15. re packed  individually in blister     ee    Disposable tonometer probe and probe container    To load the probe   1  Open the lid of the probe container as shown in figure     A CAUTION     To prevent contamination do not touch the probe tip with fingers        Open the probe container    2  Insert the probe into the probe base by turning the probe container upside down as shown in figure     www icaretonometer com         Icare tonometer ONE manual English       Inserting probe into probe base    3  After the probe has been inserted  be careful not to point the tonometer downward to prevent the probe from falling out     4  Press the measurement button once to activate the inserted probe  During activation the device magnetizes the probe  probe  moves rapidly back and forth      Once the probe is activated  the PROBE LED remains lit  The tonometer is now ready to take a measurement     MEASURING INTRAOCULAR PRESSURE       A measurement sequence is a set of six measurements  The measurement value  shown after all six measurements are taken  is  displayed on LED display  The measurement value is an average of four successful interim measurements of those six interim  measurements performed  The highest and the lowest value is left out from final value displayed  If there is too much variation in  average  SD  error code REPEAT is lit on LED display  See error messages and indicators     e In the automatic mode the six measurements in a sequence are taken automatically and in 
16. sing the Icare LINK software  Consult your doctor to interpret the results of the  self measurement     PACKAGE CONTENTS    The package contains     Icare ONE tonometer   10 sterilized single use probes   2 eye supports  small and large    2 Batteries   USB cable   User manual   Instructions for downloading Icare LINK software  Warranty card   Carry on case   Mode switch key    www icaretonometer com         Icare tonometer ONE manual English       1  Eye support   2  Support position indicator  3  Measurement button   4  Forehead support   5  Probe   6  Cheek support   7  Probe base      IOP indicator LEDs     Mode switch     Forehead cheek support adjustment wheel    PROBE ERROR LED     POSITION ERROR LED     REPEAT ERROR LED     CLEAN ERROR LED    NOW PUNR    TURNING THE DEVICE ON    Press the measurement button to turn the tonometer on  All LEDs are displayed briefly    One of the lower IOP indicator LEDs remains lit for 2 seconds while other LEDs switch off  The LED indicates the battery voltage      GREEN   OK     YELLOW   REDUCED     RED   LOW  change battery    IOP indicator LEDs    One indicator LED remains lit for 2 seconds to show battery voltage    PROBE LED    POSITION LED  REPEAT LED  CLEAN LED    Tonometer LED panel    Following a brief pause  the PROBE LED blinks to remind the user to load the single use probe into the tonometer prior to measurement     www icaretonometer com       icare tonometer ONE manual       Accuracy of display  display is divided into 1
17. that the first accepted measurement is for the right eye and the second accepted  measurement is for the left eye     MEASUREMENT STEP       Measurement position    PATIENT USE INSTRUCTIONS    e When taking measurements     A NOTE    Before performing self tonometry the patient must have the tonometer set up by a health care professional  The doctor or nurse has adjusted the  distance between the tip of the probe and the center of the cornea with the forehead and cheek support to be 4 8 mm   5 32 5 16 inch      A NOTE  The tonometer must be held in a vertical position  Angle of inclination has limited to   22   to ensure reliable outcome     e Doctor or nurse has adjusted the measurement position      Doctor or nurse has demonstrated to the patient how to use the tonometer    1  Switch on the unit  load the probe   2  Goin front of a wall mirror   3  Take the tonometer in your hand with the white main button pointing upwards   4  Move the tonometer sideways toward your eye so that you see in the mirror the white tip of the probe in front of the black pupil   5  Move the tonometer closer so that the forehead and cheek supports touch the skin   6  Check that the white tip of the probe is seen in front of the center of the cornea  pupil    7  Turn the tonometer from the sideway position so that the tonometer with the loaded probe points straight to the center of the cornea   pupil    8  Open your eyes wide and try not to blink your eye while measuring   9  Press the measurement 
18. the normal mode the user must take  each measurement individually until all six measurements are taken     To measure intraocular pressure     1  Check that the tonometer is set up correctly and the measurement position is adjusted by a doctor or a nurse   2  The patient should relax  and look straight ahead at a specific point while keeping eyes wide open      1 6 1 3         Patient with the tonometer correctly positioned    3  Bring the tonometer near the eye in front of a wall mirror as shown in Figure  Measurement step   Move the tonometer closer so  that the forehead and cheek support touch the skin  The distance from the tip of the probe to the center of the cornea must be  4 8 mm  5 32 5 16 inch      4  Press the measurement button lightly to perform the measurement  taking care not to shake the tonometer  The tip of the probe  should make contact with the central cornea  The way you take measurements depends on the chosen mode        In Automatic mode  press the measurement button once to perform the measurement  A long beep indicates the end of one  measurement sequence  six measurements   In automatic mode single measurement values are not shown   e In Normal mode  press the measurement button to take one individual measurement at a time  A short beep sounds after each    measurement  A long beep sounds when the six measurement sequence is complete   In addition to showing the IOP range  the accurate IOP value is stored in the internal memory   Icare ONE is constan
19. tient s eyes  except for the probes  which may do so for a fraction of a second during  measurement  Do not bring the tonometer into contact with the eye or push it into the eye  the tip of the probe should be 4 8mm  or 5 32 5 16  inch  from the eye      A WARNING     The device probe tips have not been evaluated for the presence of endotoxins  The probe tips are for single use only  and are packaged sterile     A WARNING     The Icare ONE Tonometer  TA02  is intended to be used only under supervision and control of the health care professional only     www icaretonometer com       Icare tonometer ONE manual English       A WARNING     Health care professionals must inform patients not to modify or discontinue their treatment plan without receiving instructions from the health care  professional     N anni    Do not change your medication or dosage without consulting your doctor     Misses    IOP accuracy and repeatability are unknown for patients with corneal astigmatism  gt  3 00 D of astigmatism  irregular astigmatism and keratoconus     A WARNING     Patients with hearing loss may not detect signals that indicating correct measurements     N sunau    Patients with physical challenges may experience difficulty in using the hand held tonometer       N wanna    No modification of this equipment is allowed    A CAUTION     Before using the tonometer read this manual carefully  Keep it for future use  It contains important information on using and servicing the  tonometer     
20. tly controlling the measurement position  When too far or near from eye or if the measurement angle is too tilted  the    tonometer beeps twice and displays a red LED indicating an error  Position   Press the measurement button again to clear the error message and  follow the corrective actions according to indicators described on page 10     www icaretonometer com         icare tonometer ONE manual       In normal mode a value is shown briefly after each successful measurement  When the six measurements in sequence are completed and the  measurement result is valid  the final value range is displayed  If there is too much variation between the measurements  the REPEAT ERROR LED is  displayed and the measurement needs to be performed again  The device stores the quality of the measurement in its internal memory  which is  only accessible by downloading to a PC using the Icare   LINK software     A new measurement needs to be taken if     e REPEAT ERROR LED is displayed  for example  if the probe made contact with the eyelid   e Encountering unusual values  for example  over 22 mmHg or below 8 mmHg     The tonometer   s internal memory stores the time and date of a measurement  the measurement result and the quality level of the measurement   This information can be uploaded to a PC through a USB cable using Icare   LINK software  The tonometer is taken to a health care professional who  can transfer the data     A NOTE    When using Icare LINK software  it assumes by default 
    