Contents

1. 10 50 60 Hz Rated Voltage Current 115V 12A max for 750 watts 8A max for 500 watts 230V 7A max for 750 watts 5A max for 500 watts Fuse Rating 115V F15A 250V for 750 watts F10A 250V for 500 watts 230V F8A 250V for 750 watts F6 3A 250V for 500 watts Weight 15 lbs 6 8 Kg Dimensions 9 25 H x 7 5 W x 13 5 D 235 mm x 190 mm x 340 mm Output Voltage 1000 V rms max Output Frequency 20 KHz nom or 40 KHz Converter Weight 2 lbs 900 g Dimensions 7 25 L x 2 5 Dia 183 mm x 63 5 mm Standard 12 Horn Weight 0 75 Ibs 340 g Dimensions 5 375 Lx 5 Dia 136 mm x 13 mm Environmental Pollution Degree Operating Limits Temperature 41 104 F 5 40 C Relative Humidity 20 90 Non Condensing Altitude 6 651 ft 2000 m Shipping Storage Temperature 35 120 F 2 49 C Relative Humidity 10 95 Non Condensing Ambient Pressure Extremes 40 000 ft 12 192 m Restriction of RoHS Compliant Substances ROHS Relative humidity Maximum relative humidity 80 for temperatures up to 31 C decreasing linearly to 50 relative humidity to 40 C Only use IEC approved Fast acting fuses Cooper Bussman series S500 CAUTION LOW SURFACE TENSION LIQUIDS ORGANIC SOLVENTS The probes solid or with a replaceable tip are tuned elements that resonate at a specific frequency If the replaceable tip is removed or isolated from the rest of the probe t
2. The screen will display REVIEW The REVIEW function provides a window on the process by displaying various operating parameters without process interruption Pressing the ENTER REVIEW key repeatedly during processing will consecutively display the following information a Selected amplitude e g Amplitude 40 b Set processing time and elapsed processing time e g Set 5 30 25 Time 0 57 03 c Selected pulsing cycle and actual pulsing cycle e g Pulse 01 01 d Amount of power in watts and accumulated amount of energy in JOULES delivered to the probe e g 20 watts 0000000 Joules e Elapsed time since processing was initiated e g Elapsed time 1 27 33 The number of Joules displayed is dependent on the watts being drawn and the processing time 14 ENERGY SETPOINT The energy setpoint continuously monitors the amount of energy in Joules watts seconds that is being delivered to the probe and terminates the ultrasonics when the energy level reaches a predetermined setpoint To set the energy setpoint press the ENERGY key The screen will display Using the numeric keys set the energy setpoint The screen will display Press the ENTER REVIEW key The screen will display SAVE RECALL SAVE The save function retains in memory up to 9 1 9 control parameters under a storage identification ID number To store the parameters under an ID number press the SAVE RECALL key The s
3. Sonics requires that each customer certify one of the statements on page 19 Prior to returning any equipment please make a copy of the form on page 19 fill in the form and send it back with the equipment being returned This form must accompany any equipment that is being returned for repair SONICS SAFETY CERTIFICATION FORM Items being returned Please check only one item below The equipment was never used or exposed to any radiological biological or chemical agents and is safe to handle use or dispose of The equipment was used but not in conjunction with or exposed to any radiological geological or chemical agents and is safe to handle use or dispose of ___ The equipment was used in conjunction with or exposed to radiological biological or chemical agents and has been decontaminated rendering it safer for handling use or disposal Authorization By accepting authorization to return the equipment listed above the undersigned assumes all responsibility and liability for radiological biological and chemical decontamination Sonics reserves the right to refuse delivery of the equipment without necessary documentation or where we determine they have not been properly decontaminated Sonics reserves the right to bill the customer for any and all costs associated with the decontamination and or disposal of the equipment we determine was not properly decontaminated In the event the equipment has been exposed to r
4. as nr D Puna o 0 Og oe E or seo YN o x O O Oo COD O S 2 Ge lt x oo 03 22 OF CL os Y n5 x Les O 5S ES 595 12 oo 220 ey ae Susa 05 ES ES DEE TEw se ESTOS 2085 Lote Ep a eS m2 pa E a w 6S Lele 0 y BXL328 05 72 Bye DAS Y0 D S ES ES 03023 28 S2889E328 83205 5352 32 eS Gs D 0 O ces 5 Se SS aeeooy a2 S0 4 5 a b 2552 Ese 25 oN 2 oo 3 ob 2802E d gt S Enay BOLAGFG A ES ES SPSREL ef 25058285 gretos Sha5 Be SS R ZOU S o o 2 u2 0 0 ESTLD VC OS ES CER ES S niet osgahs S802 Bge E y f Pass oO 0O z Ss pda ow ony fD Oo D Pu Y ES Gs Q 0 E O 7 lt OO E ea SES5P0R ES pee Legs Eybi 2 503 sF eS Ae o D is ra a GO 4 des ES SS N 0USO0 Eo or sx cs o 0D Lo UE 00 EO or po uSo2sto Se E LCSL 732302 ENS L2EESGQ 02 ES ES EDO Ba SFotEN on oc Vosso 5339 6S ES DoeSS o yg 0ES2 20pOS uo 8 S22E79 g2 aS ES o Zes 38 o23D22ES E uvn2 303 as Ste SS us SaBQseeS8 go 2eRocsa ss ETLE Slos Se aS os ore ES gt 053E 00 SG FRHGAEGBEBELTE Esozo woco FG se ESO ee amp SS come Bis ONE PRE PN E NT GIN Ne DOPE O NE POPE INE NE POPE PN NE SINE LE PN NL SON LE OPN NE SIN PLE NL NON DLE PN PEN PRE NIE GIN NE PROPEL PN PROPEL PDE PED 2 ONO ERE ERE GENER MELEE REDE PCR SAR 2 2 2 2 2 2 2 ROA NOR NOR NEA RA CA ORME REA NOR MOR NOA NOR ER MOLES CLR ER EAR EERE NEE ER MEER AER ER ME EEE AE EERE RAE EER IMPORTANT SAFEGUARDS READ BEFORE INSTALLING OR USING THE EQUIPMENT
5. required Protease inhibitors are biological or chemical compounds that functions by reversibly or irreversibly binding to the protease Proteases generally belong to one of four evolutionarily distinct enzyme families based on the functional groups involved in cleavage of the peptide bond Therefore several different types of inhibitors are generally required to protect proteins from proteolysis during extraction and purification Cellular disruption is the first step in RNA isolation and one of the most critical steps affecting yield and quality of the isolated RNA Typically cell disruption needs to be fast and thorough Slow disruption for example placing cells or tissue in guanidinium isothiocyanate GITC lysis solution without any additional physical shearing may result in RNA degradation by endogenous RNase released internally yet still inaccessible to the protein denaturant GITC This is especially a concern when working with tissues high in endogenous RNase such as spleen and pancreas Incomplete disruption may also result in decreased yield because some of the RNA in the sample remains trapped in intact cells and therefore is unavailable for subsequent purification For most samples thorough disruption can be monitored by close inspection of lysate after disruption There should be no visible particulates except when disrupting materials containing hard non cellular components such as connective tissue or bone When processing
6. so xg ee E ve Se e ES 2 e R E ee e ee E Ze lt 3 ve lt 9 e ee E ve Se e eS so o ee ve OI III E E E EEEE IIS SOF EA E A SSSI SI III AS OA SA A SAE OSS Pea ots Fetes wets Pes LES ES rane SL L LL a ous 6 2u28585 2 offo F2 ae 255459 223 3 5 ES ES FERRER GeseEesBes sede ESEE pe f DA oroo 8 D DEORE 0 D 2 o 20530 O 0 EX da SL Lgo 813 8255 575S gg oi F 55588 3 238552232 2559 2947 F SO Nn 4 aN ES cB zos 5 Ese 8552 CESE se 28 eS SS S gt u o0 gt LPEPL ESO c 20 ce 5 8 ce ES 6 L EQoF o a8 2 00595S Sec 20 o eS A gt o ec O 0 O DICO 50 Su RY eS ozott 2 9 8 F855 D 55 2 o 85 S E eS na oo 552 D azau n ow PHS gt ES 8 Eco lt 3 x N Ye OCB Oo 55 G6 Ses lt Me E O E E oo n raf S 523 g SS asg 255 28852 ES gt E Eo oO SCE gt z c50 cCc E gt gt 5S 6 Enap S 3585228 2588 azon 27 S as OT O OG 0 oe ZoasSos oO gt Co i UD aks CS O zD0 lt x O O Cc o E O S E O Q S A eae E a o we ES 2ecess SGM Ez foe 2835 o 2 SS mie DOE DW HEOEHDDE NS OL Oo v ose foe NI Se asas oe ES D com Cg OF 052 Q de oe StS SE ESSGSLe e CS 5 Bose BEF e C Z 585985 FS SESLIFESS 507 252 ES ES 55550 E FSE3O0 o2ee 20530 o52 gt 2 LS SB lt EBLO aS SRR 35853 sack 2228 SB amp SS 2 gt OF 0 Do E L es a Sesto so 0 poel 852 Szab Ba a 5scpg3a Se Byres S of 8 Sia 5 amp 38
7. From this point cells can be lysed as described above for bacteria or 2 isolating RNA from the soil or sediment directly For example adding soil to a diatomaceous earth and lysis buffer and then sonicating The sample is then centrifuged to remove solid debris Cultured cells are relatively easy to disrupt Cells grown in suspension are collected by centrifugation rinsed to remove culture medium and then lysed by sonicating in GITC lysis buffer Placing the flask or plate on ice while washing and lysing the cells will further protect the RNA from endogenous Rnases released during the disruption process Detergent cell lysis is sometimes used in conjunction with ultrasonic processing to facilitate disruption Detergents break the lipid barrier surrounding cells by solubilizing proteins and disrupting lipid lipid protein protein and protein lipid interactions Detergents like lipids self associate and bind to hydrophobic surfaces They are composed of a polar hydrophilic head group and a nonpolar hydrophobic tail and are categorized by the nature of the head group as either ionic cationic or anionic nonionic or zwitterionic Their behavior depends on the properties of the head group and tail NOTE When working with detergents do not use a probe with a replaceable tip Use a solid probe instead Unfortunately there is no standard protocol available for selecting a detergent to use for membrane lysis the ideal detergent will depen
8. MICROTIPS STEPPED MICROTIP TIP DIAMETER 1 8 3mm 3 16 5mm 1 4 6 5mm 1 8 3mm INTENSITY ultra high very high very high VOLUME batch 1 10ml 3 20ml 5 50ml 250ul 10ml STANDARD PROBES TIP DIAMETER 1 2 13mm 3 4 19mm 1 25mm VOLUME batch 10 250ml 25 500ml 500 1000ml 25 500mI 500 1000mI FUNCTIONS OF KEYS CONTROLS INDICATORS AND CONNECTORS FRONT PANEL Screen 059 key CLEAR key ENTER REVIEW key TIMER TEMP PULSER ENERGY key START STOP key PAUSE key SAVE RECALL key AMPL key I key O key Displays prompts and the following control parameters e Amplitude selected e Output power delivered to the probe in watts and as percentage of the total power e Selected duration of processing e Actual processing time e Elapsed time e Set and read temperature e Pulse duration e Accumulated amount of energy in Joules delivered to the probe Input digits Clears preceding entry Enters data into the program and selects various parameters for display on the LCD screen Used with the numeric keys to set the duration of ultrasonic application from 1 second to 9 hours 59 minutes 59 seconds Used with the numeric keys to set the high temperature limit from 1 C to 99 C Red indicator lights when the temperature limit has been reached Used with the numeric keys to set the pulse mode The ON cycle and OFF cycle can be set independently from 1 second to 59 secon
9. Your Ultrasonic Processor has been designed with safety in mind However no design can completely protect against improper usage which may result in bodily injury and or property damage For your protection and equipment safeguard observe the following warnings at all times read the operating instructions carefully before operating the equipment and retain this instruction manual for future reference If the Ultrasonic Processor is used in a manner contrary to that specified in this instruction manual the protection features designed into the unit may be impaired WARNINGS e Make sure the Ultrasonic Processor is properly grounded via a 3 prong outlet e High voltage is present in the power supply DO NOT REMOVE THE COVER Refer all servicing to qualified service personnel e Never operate the power supply unless it is connected to the converter e Never secure anything to the probe except at the nodal point point of no activity e Never touch a vibrating probe e Air cool the converter with dry compressed air when operating at high amplitude for prolonged duration e Hearing protection is highly recommended It is recommended that a sound abating enclosure or ear protection be used when operating the Ultrasonic Processor SYMBOLS Caution Risk of electric shock Hazardous voltage Caution Risk of danger Refer to User Manual SPECIFICATIONS Power Supply Operational Input Voltage 115V 10 50 60 Hz 230V
10. cause the maximum power that the power supply is capable of delivering to be delivered to the sample For example with the 500 watt Ultrasonic Processor the maximum power that the power supply is capable of delivering will only be delivered when the resistance to the movement of the probe is high enough to draw 500 watts Press the I key The screen will display the power rating and the frequency of the Ultrasonic Processor and the following control parameters AMPLITUDE The amplitude is the only parameter that must be set in order for the Ultrasonic Processor to be operational The other control parameters Time and Pulse do not have to be set for continuous operation AMPL displays the percentage of amplitude that was previously selected Press the AMPL key and the A or W key for a reading of 40 then depress the ENTER REVIEW key The screen will display The Ultrasonic Processor is now ready for continuous operation To energize the equipment press the START key or the footswitch To de energize the equipment press the STOP key or release the footswitch If the Time or Pulse functions must be used refer to the appropriate paragraph s below To increase or decrease the amplitude in small increments when the equipment is on depress the AMPL key to display the percentage of amplitude that was previously selected then depress the A or W key as required NOTE If the START key is pressed and the time limit has not bee
11. difficult cells pretreatment with an enzyme to weaken the cell walls is beneficial Lysozyme byaluronidase glycosidase glucalase lyticase zymolase and lysostaphin digestion are among the enzymatic methods frequently used with yeast Lysozyme zymolase and lysostaphin digestion are among the enzymatic methods frequently used with bacteria and yeast to dissolve a coat capsule capsid or other structure not easily sheared by ultrasonics Enzymatic treatment is usually followed by sonication in a GITC lysis buffer Collogenase may be used with collogen lysostaphin with staphylococcus and trypsin hyaluronidase with liver and kidney Yeast can be extremely difficult to disrupt To process yeast sonicate in a tube containing the sample guanidinium based lysis buffer and small glass beads 0 5 1 mm Enzymatic pretreatment as outlined above is strongly recommended To disrupt filamentous fungi scrape the mycelial mat into a cold mortar add liquid nitrogen grind to a fine powder with a pestle then sonicate in lysis buffer to solubilize completely As fungi may also be rich in polysaccharides pretreatment with polyvinylpyrrolidone PVP may be beneficial Disruption of cells found in soil and sediments is accomplished by 1 isolating the bacterial cells from the material prior to the RNA isolation procedure This is accomplished by homogenization of wet soil in a mechanical blender followed by a slow speed to pellet the bacteria cells
12. mortar and pestle During this process it is important that the equipment and tissue remain at cryogenic temperatures The tissue should be dry and powdery after grinding Grinding should be followed by sonication in a GITC lysis buffer Processing frozen tissue in this way is cumbersome and time consuming but effective Ultrasonic processing will typically cause the temperature of the sample to increase especially with small volumes Since high temperatures inhibit cavitation the sample temperature should be kept as low as possible preferably just above its freezing point This can be accomplished by pulsing the ultrasonics on and off while keeping the sample vessel immersed in an ice bath While processing the sample occasionally touch the vessel to ensure that the sample is relatively cool Increasing hydrostatic pressure typically 15 60 psi and viscosity can enhance cell disruption For microorganisms the addition of glass beads in the 0 5 to 1mm size range promotes cell disruption Beads are almost a prerequisite when working with spores and yeast A good ratio is one volume of beads to two volumes of liquid Glass beads are available from Cole Parmer Instruments Fisher Bioblock Scientific 625 East Bunker Court Parc d innovation BP 50111 Vernon Hills Illinois 60061 F 67403 illkirch cedex USA France Phone 1 800 323 4340 Phone 03 88 67 14 14 Fax 847 247 2929 Fax 03 88 67 11 68 Email info coleparmer com Email infosObiob
13. In animal cells the plasma membrane is the only barrier separating cell contents from the environment but in plants the plasma membrane is also surrounded by a rigid cell wall Plant cell walls consist of multiple layers of cellulose These types of extracellular barrier confer shape and rigidity to the cells Plant cell walls are particularly tough making them very difficult to disrupt mechanically or chemically The lack of an extracellular wall in animal cells make them relatively easy to lyse Soft fresh plant tissue can often be disrupted by sonicating in a lysis buffer Other plant tissues like pine needles need to be ground dry without liquid nitrogen Some hard woody plant materials require freezing and grinding in liquid nitrogen prior to being ultrasonically processed Plant cell suspension cultures and calluses can be lysed by sonication in a lysis buffer for 30 seconds to 2 minutes The diversity of plants and plant tissue make it impossible to give a single recommendation for all samples However one should be aware that most plant tissues typically contain polysaccharides and polyphenols that can coprecipitate with RNA and inhibit downstream assays Treating a plant tissue lysate with polyvinylpyrrolidone PVP will precipitate such problematic components from the lysate before the actual RNA isolation is carried out Single cell organisms micro organisms consist of a semipermeable tough rigid outer cell wall surrounding the pr
14. USER S GUIDE AUTOTUNE SERIES HIGH INTENSITY ULTRASONIC PROCESSOR WITH TEMPERATURE CONTROLLER MICROPROCESSOR CONTROLLED SERIAL NO AC OR HIGHER 500 WATT MODEL 750 WATT MODEL TABLE OF CONTENTS Warranty Important Safeguards and Warnings Specifications Low Surface Tension Liquids Organic Solvents SECTION INSTALLATION IIS SCHOM secs cteavocet fee a E e O E 1 Electrical Requirements ce ceeeeeeeeeeeeee eee r 1 Installing the Ultrasonic Processor sisscioniansasiia aio 1 SECTION II OPERATION Principles of Ultrasonic DiSrUptiON oooococcccccconoconoconaconnonnnnnnnnnnonnnos 2 Functions of Controls Indicators and Connectors c0 ee 3 Preparationtor USC tii diia 5 Using the Ultrasonic Processor unsvicciinian daria 8 SECTION lll SERVICE INFORMATION Return of Equipment ssrin iii 17 SECTION IV OPERATING SUGGESTIONS AND TECHNIQUES cocoa ia 20 The Ultrasonic Processor supplied with this instruction manual is constructed of the finest material and the workmanship meets the highest standards It has been thoroughly tested and inspected before leaving the factory and when used in accordance with the procedures outlined in this manual will provide you with many years of safe and dependable service SONICS 8 MATERIALS INC 53 CHURCH HILL ROAD NEWTOWN CT 06470 PHONE 203 270 4600 REV gt 10 2 A ee o eS 6 lt 9 e pd lt 9 E Me Se ve Se es ee R E ve E ae
15. aced If the wattmeter reads above 20 watts either the converter booster or power supply has failed and the Ultrasonic Processor should be returned for repair 3 If the Ultrasonic Processor stops working due to an overload condition as indicated on the display press the O key investigate and remedy the problem then press the I key to restart the equipment 17 CLEANING INSTRUCTIONS The generator and converter may be cleaned using a solvent free cleaning solution i e glass cleaner Horns and probes should be cleaned using isopropyl alcohol Horns are made from titanium and can be autoclaved RETURN OF EQUIPMENT It is suggested that if a unit is in need of repair it should be sent back to the factory In order to receive prompt service always contact the factory before returning any equipment Include date of purchase model number and serial number For equipment not covered by the warranty a purchase order should be forwarded to avoid unnecessary delay Care should be exercised to provide adequate packing to insure against possible damage in shipment The equipment should be sent to the Service Department with all transportation charges prepaid and return of shipment indicated Please obtain a Return Authorization Number RA prior to returning the instrument SAFETY CERTIFICATION FORM Federal law prohibits the transfer of equipment or products contaminated with radiological biological or chemical waste residue
16. adiological contamination the signature of the Radioactive Safety Officer is required Print name RA Signature Date SECTION IV OPERATING SUGGESTIONS AND TECHNIQUES DISRUPTING CELLS The disruption of cells is an important method in the field of proteomics and in the isolation and preparation of intracellular products Isolation of subcellular fractions and concentration of proteins allow for more efficient identification and study of proteins of interest From research levels through to production many areas of biotechnology particularly recombinant technology necessitate the use of ultrasonics for cell disruption Cell disruption focuses on obtaining the desired product from within the cell and it is the cell wall that must be disrupted to allow access to the contents of the cell All cells have a plasma membrane a protein lipid bilayer that forms a barrier separating cell contents from the extracellular environment Lipids constituting the plasma membrane are amphipathic having hydrophilic and hydrophobic moieties that associates spontaneously to form a closed bimolecular sheet Membrane proteins are embedded in the lipid bilayer held in place by one or more domains spanning the hydrophobic core In addition peripheral proteins bind the inner or outer surface of the bilayer through interactions with integral membrane protein or with polar lipid head groups The nature of the lipid and protein content varies with cell type
17. and automatically adjusts the amount of power delivered by the engine in order to compensate for these ever changing conditions The greater the terrain rate of incline and greater the resistance to the movement of the vehicle the greater the amount of power that will be delivered by the engine to overcome that resistance The Ultrasonic Processor is designed to deliver constant amplitude As the resistance to the movement of the probe increases additional power will be delivered by the power supply to ensure that the excursion at the probe tip remains constant Using a more powerful power supply will not deliver more power into the liquid Rather it is the resistance to the movement of the probe that determines how much power will be delivered into the sample The AMPLITUDE control allows the ultrasonic vibrations at the probe tip to be set to any desired level Although the degree of cavitation required to process the sample can readily be determined by visual observation the amount of power required cannot be predetermined A sensing network continuously monitors the output requirements and automatically adjusts the power to maintain the amplitude at the preselected level The greater the resistance to the movement of the probe due to higher viscosity deeper immersion of the probe into the sample larger probe diameter or higher pressure the greater the amount of power that will be delivered to the probe Setting the amplitude to 100 will not
18. arious methods can be implemented to measure the efficiency of ultrasonic disruption Typically counting the cells using a microscope is a satisfactory method However for greater accuracy a protein assay should be used This procedure is widely recognized as a good method for measuring cell disruption by taking into account the amount of protein released after disruption The disrupted cells are then tested and checked against this number for percentage breakage There are several types of protein assays The most common is the Folin Reaction Lowry Assay method as it is comparatively simple and provides consistent results This colorimetric method has a sensitivity to protein of around 8 ug mL in the assay solution The assay turns blue in the presence of proteins due to the reaction of copper ions in the alkaline solution with protein and the reduction of phosphomolybdate phosphotungstic acid in the Folin reagent by aromatic amino acids in the treated protein 27 Fractional protein release Rp is calculated using the following equation and multiplying the result by 100 Rp Cf Cb Ct Cb Cf Free protein Ct total protein Cb Background protein This gives the actual disruption percentage taking into account the background levels of protein before disruption
19. creen will display Press the key one more time The screen will display Press the ENTER REVIEW key The screen will display Using the 1 9 numeric keys enter the ID number The screen will display Press the ENTER REVIEW key The screen will display RECALL The recall function can retrieve from memory any of 9 stored control parameters for verification or usage To retrieve any parameters press the SAVE RECALL key The screen will display nor Press the ENTER REVIEW key The screen will display The screen will display E Press the ENTER REVIEW key R4 TIME 5 30 25 TEMP TTC The sober wih alsplay PULSE 2 5 1 0 AMPL 40 Note To review all the information that has been stored press all the numeric keys consecutively IMPORTANT Proper care of the probe is essential for dependable operation The intense cavitation will after a long period of time cause the tip to erode and the power output to decrease without showing up on the wattmeter The smoother and shinier the tip the more power will be transmitted into the sample Any erosion of the probe tip will increase the rate of future erosion For that reason it is recommended that after every 5 or 6 hours of use the tip be examined and if necessary polished with emery cloth or an abrasive wheel or machined in a lathe Since the probe is tuned to vibrate at a specific frequency it is most important that only the contaminated surface be removed This proced
20. d on the intended application In general nonionic and zwitterionic detergents are milder and less denaturing than ionic detergents and are used to solubilize membrane proteins when it is critical to maintain protein function and or retain native protein protein interactions for enzyme assays or immunoassays Zwitterionic detergents and nonionic detergents are commonly used for these purposes In contrast ionic detergents are strong solubilizing agents and tend to denature proteins thereby destroying protein activity and function There are a few commonly used ionic detergents that are only mildly denaturing including sodium cholate and sodium deoxycholate The choice of detergent for cell lysis also depends on sample type Animal cells bacteria and yeast all have differing requirements for optimal lysis due to the presence of absence of a cell wall Because of the dense and complex nature of animal tissues they require both detergent and sonication In addition to the choice of detergent other important considerations for optimal cell lysis include the buffer pH salt concentration and temperature Consideration should be given to the compatibility of the chosen detergent with downstream applications For example if the detergent used for lysis must be removed then a dialyzable detergent should be selected If pretreatment with enzymes or detergents cannot be used the freeze thaw method should be considered The freeze thaw method is comm
21. diverse therefore it is difficult to make one recommendation for all bacteria Ultrasonic processing will lyse most Gram positive and Gram negative bacteria including mycobacteria Typically glass beads and lysis solutions are added to a bacterial cell pellet and the sample is sonicated for a few minutes It is possible to lyse some Gram negative bacteria by sonication in lysis solution alone Bacteria cell walls can be digested with lysozyme to form spheroplasts Gram positive bacteria usually require more rigorous digestion increased incubation time increased incubation temperature etc than Gram negative organisms The spheroplasts are then easily lysed in a GITC lysis buffer with ultrasonics Gram negative bacteria typically require 10 to 15 minutes of processing while staphylococcus requires 20 to 30 minutes The level of intensity that should be used is application dependent For example high intensity might be recommended for the break up of cells but should never be used when the release of intracellular components might be objectionable e g Organelle isolation The ability to control the amplitude at the probe tip is a prerequisite for process optimization And because each application requires its own set of processing parameters due to variation in volume and composition the optimum amplitude can only be determined empirically When processing a new sample it is recommended that the amplitude be set first at 50 80 with a m
22. ds Red indicator lights when pulser is in the OFF portion of the cycle Used to set the ENERGY SETPOINT Starts or stops the ultrasonics In the STOP mode the red indicator goes off Suspends operation Red indicator lights when the processing cycle is interrupted Used with the numeric keys to assign a number to a program and store that program in memory Up to 9 programs 1 9 can be stored Also used to recall any of 9 stored programs Controls the amplitude of vibration at the probe tip Switches the main power on Switches the main power off Used with the AMPL key when the unit is on stand by to set the amplitude of vibration at the probe tip Also used to increase or decrease the amplitude in small increments while the unit is running To accomplish this task depress the AMPL key to display the percentage of amplitude previously selected then depress the A or Y key as required FUNCTIONS OF KEYS CONTROLS INDICATORS AND CONNECTORS cont REAR PANEL 9 pin D sub connector Connects to external actuation device and enable power and frequency monitoring 9 PIN D SUB CONNECTOR e ewan ICO CTN 8 and 9 Enables the intensity to be remotely adjusted using an external 10k potentiometer See below to pin 9 10K lt to pin 8 to pin 6 PREPARATION FOR USE Connect the Power Cord to the receptacle on the rear of the processor Plug the Power Cord into a grounded electrical outlet If the unit is already o
23. e and corrosive fumes and extremes of temperature and humidity If processing flammable liquids use an approved fume hood and do not place the power supply in the fume hood When positioning the unit be sure to leave adequate space behind the unit so that all connections can be easily disconnected SECTION II OPERATION PRINCIPLES OF ULTRASONIC DISRUPTION The ultrasonic power supply converts 50 60 Hz line voltage to high frequency electrical energy This high frequency electrical energy is transmitted to the piezoelectric transducer within the converter where it is changed to mechanical vibrations The vibrations from the converter are intensified by the probe creating pressure waves in the liquid This action forms millions of microscopic bubbles cavities which expand during the negative pressure excursion and implode violently during the positive excursion This phenomenon referred to as cavitation creates millions of shock waves in the liquid as well as elevated pressures and temperatures at the implosion sites Although the cavitational collapse lasts but a few microseconds and the amount of energy released by each individual bubble is minimal the cumulative effect causes extremely high levels of energy to be released into the liquid The larger the probe tip the larger the volume that can be processed but at a lesser intensity For information regarding the processing capability of each probe consult the tables below TAPERED
24. ed microtip 1 8 8mm 630 0422 Probe Y 18mm with threaded end and replaceable tip 630 0220 Tapered microtip 1 8 8mm 630 0418 Tapered microtip 3 16 5mm 630 0419 Tapered microtip 1 4 6mm 630 0420 Probe solid or with threaded end and replaceable tip same as 5 Replaceable tip same as 6 Extender Y 13mm 630 0410 Extender 19mm 630 0409 Extender 1 25mm 630 0444 Full wave extender 34 19mm 10 254mm long 630 0518 Full wave extender 1 25mm 10 254mm long 630 0519 High gain probe 4 19mm solid 630 0306 High gain probe 34 19mm with threaded and replaceable tip 630 0305 High gain probe 1 25mm solid 630 0310 High gain probe 1 25mm with threaded and replaceable tip 630 0311 Replaceable tip 34 19mm or 1 25mm same as 6 Full wave probe 12 13mm solid 10 254mm long 630 0217 Full wave probe Y 13mm 10 254mm long with threaded and replaceable tip 630 0218 Aluminum coupler 630 0562 34 19mm solid probe 630 0208 Cup horn 1 Y 88mm 630 0503 Cup horn 2 Y 64mm 630 0431 Cup horn 3 76mm 630 0496 USING THE ULTRASONIC PROCESSOR The speed control on an automobile can to a certain extent be compared to an Ultrasonic Processor The speed control is designed to maintain the vehicles rate of travel constant As the terrain changes so do the power requirements The speed control senses these requirements
25. he element will no longer resonate at that frequency and the power supply will fail Unlike aqueous water based solutions which rarely cause problems solvents and low surface tension liquids are problematic These liquids penetrate the probe replaceable tip interface and force the particulates into the threaded section isolating the tip from the probe When processing low surface tension liquids ALWAYS use a solid probe SECTION INSTALLATION INSPECTION Prior to installing the Ultrasonic Processor perform a visual inspection to detect any evidence of damage which might have occurred during shipment Before disposing of any packaging material check it carefully for small items The equipment was thoroughly inspected and carefully packed before leaving our factory The carrier upon acceptance of the shipment assumed responsibility for its safe delivery Claims for loss or damage sustained in transit must be submitted to the carrier If damage has occurred contact your carrier within 48 hours of the delivery date DO NOT OPERATE DAMAGED EQUIPMENT Retain all packing materials for future shipment ELECTRICAL REQUIREMENTS The Ultrasonic Processor requires a fused single phase 3 terminal grounding type electrical outlet For power requirements check the label on the back of the unit INSTALLING THE ULTRASONIC PROCESSOR The Ultrasonic Processor should be installed in an area that is free from excessive dust dirt explosiv
26. icrotip and then increased of decreased as required Prior to sonication cells can be treated with various agents to aid the disruption process Lysis can be promoted by suspending cells in a hypotonic buffer which causes them to swell and burst more readily by physical shearing Lysozyme can be used to digest the polysaccharide component of yeast and bacterial cell walls Alternatively processing can be expedited by treating tough cells with glass beads to facilitate the crushing of cell walls This treatment is commonly used with yeast cells Viscosity of a sample typically increases during lysis due to the release of nucleic acid material DNase can be added to samples 25 50 ug ml to reduce this problem Nuclease treatment is not required for sonicated material because sonication shears chromosomes Because proteolysis can be a problem whenever cells are manipulated it is advisable to add protease inhibitors to samples undergoing lysis NOTE All living organisms contain proteolytic enzymes proteases and peptidases Proteases are required for a variety of cellular functions such as cellular repair or the digestion of extracellular material In whole cells protease activity is tightly regulated by compartmentalization or inhibitors to prevent damage to cellular proteins Cell lysis disturbs this regulation and proteolytic degradation of the sample becomes a concern Therefore addition of protease inhibitors to cell lysis buffers is often
27. lock fr When processing a sample with ultrasonics always immerse the probe deep enough below the surface of the sample to inhibit aerosoling or foaming foaming substantially reduces cavitation Processing at a lower power setting without foam is much more effective than processing at a higher power setting with foam Decreasing the power increasing processing time and lowering the temperature of the sample will usually prevent aerosoling and foaming Do not use any antifoaming agents or surfactants During cavitation free radicals are formed which if they are allowed to accumulate can greatly affect the biological integrity of the sample by reacting with proteins polysaccharides or nucleic acids Although during short periods of processing their formation is not normally considered a problem for longer durations the addition of free radical scavengers such as carbon dioxide NO cysteine reduced glutahione dithiothreitol or other SH compounds might be beneficial Saturating the sample with a protective atmosphere of helium or nitrogen gas or dropping a small pellet of dry ice in the sample will also inhibit free radical formation Whereas it is true that gas is required for effective cellular disruption it is not necessary that the vapor phase be oxygen or air since any gas except carbon dioxide will work just as well Following ultrasonic processing the cell debris can be centrifuged at 15 000 rpm for 10 minutes Since the greates
28. n switch the unit off by depressing the O key If the optional footswitch is used insert the footswitch plug into the jack located on the rear panel Make sure that the plug is inserted all the way in Screw the booster into the converter and secure forcefully with the spanner wrenches provided Using the spanner wrenches provided forcefully secure the booster to the probe Mount the converter probe assembly in a stand Secure the clamp to the converter housing only Do not secure the clamp to the probe Connect the converter cable to the power supply Connect either one of the air fittings on top of the converter to a source of dry compressed air TIF REGAL TP TIGHTENING a 13 3 5 8110 A 19 19 gt s 2 j pa 12 15 y I Y bl mr bd P K gt Cx gt A h a an 4 49 0 1 1 dpe 2 LA 7 EE 3 kS Le VITI Y YLL y LL DESCRIPTION Converter Model CV33 CV00033 Four element coupler 630 0558 Stepped top s 1 8 3mm 630 0535 Booster BHNVCGD Probe Y 13mm solid 630 0219 Probe 12 18mm with threaded end and replaceable tip 630 0220 Probe 3 19mm solid 630 0208 Probe 3 19mm with threaded end and replaceable tip 630 0207 Probe 1 25mm solid 630 0209 Probe 1 25mm with threaded end and replaceable tip 630 0210 Replaceable tip Y 18mm 630 0406 Replaceable tip 19mm 630 0407 Replaceable tip 1 25mm 630 0408 Coupler 630 0421 Stepp
29. n set processing will remain uninterrupted until the STOP key is depressed If the START key is pressed and the time limit has been set processing will remain uninterrupted until the set time limit expires or the STOP key is pressed whichever occurs first If a footswitch is use and the time limit has not been set processing will remain uninterrupted as long as the footswitch is depressed If a footswitch is used and the time limit has been set processing will remain uninterrupted until the time limit expires or the footswitch is released whichever occurs first The START key and footswitch are mutually exclusive If the process is initiated by the START key the footswitch becomes inoperative If the process is initiated by the footswitch the STOP key becomes inoperative 1 Using the fittings on top of the converter circulate dry compressed air through the converter to cool the converter 2 Immerse the probe into the liquid Always immerse the probe deep enough below the surface of the sample to inhibit aerosoling or foaming Foaming substantially reduces cavitation Processing at a lower power setting without foam is more effective than processing at a higher power setting with foam Decreasing the power increasing processing time and lowering the temperature of the sample will usually prevent aerosoling and foaming Do not use any antifoaming agents or surfactants NOTE The probe is tuned to vibrate at a specific f
30. only used to lyse bacterial and mammalian cells The technique involves freezing a cell suspension for 10 minutes in a dry ice and isopropanol bath and then thawing the material at room temperature immediately prior to ultrasonic processing The freeze thaw cycle should be repeated three times prior to being subjected to the ultrasonics This method of lysis causes cells to swell and ultimately break as ice crystals form during the freezing process and then contract during thawing This pretreatment has been 24 shown to effectively release recombinant proteins located in the cytoplasm of bacteria and is recommended when lysing mammalian cells However this method for lysis is not recommended when working with nitrilases as it has been noticed that purified nitrilases suffer structural damages upon freezing Most animal tissues can be processed fresh unfrozen However it is important to keep them cold and to process them quickly within 30 minutes after dissection When disrupting fresh tissue the cells need to be sheared immediately at the time the GITC lysis solution is added This can be done by dispensing the lysing solution in the tube adding the tissue and immediately sonicating Samples should never be left sitting in lysis solution undisrupted Hard tissues should be first treated in a blender or a mechanical homogenizer Animal tissues that have been frozen after collection should be disrupted by grinding in liquid nitrogen with a
31. otoplasmic membrane and cytoplasm The cytoplasm is made up of nucleic acid protein carbohydrates lipids enzymes inorganic ions vitamins pigments inclusion bodies and about 80 water In order to isolate and extract any of these substances from inside the cell it is necessary to break the cell wall and protoplasmic membrane In some cases the cell may excrete the desired substance without assistance but in most cases the cells must be lysed in order for these intracellular substances to be released Breaking cell membranes and releasing the contents present significant challenges The process must be fast and thorough to maximize the protein yield Because the energy applied must be great enough to break the cell membranes or walls yet gentle enough to avoid physically or chemically damaging cell content the Vibra Cell with its variable intensity capability is ideally suited for this application Microorganisms differ greatly in their sensitivity to ultrasonic disintegration For example the most readily disintegrated are the rod like forms bacilli while the spherical organisms cocci are much more resistant The group Mycobacteria to which the tuberculosis organism belongs is particularly difficult to disrupt Yeast gram positive bacteria and to a lesser extent gram negative bacteria have considerably harder cell walls in comparison to animal cells and require relatively high power for cell disruption Bacteria are extremely
32. requency If the resonant frequency of the probe has changed due to cavitation erosion or fracturing a minimum reading will not be obtained If an overload condition exits or if minimum reading cannot be obtained less than 20 when the flow cell is empty or when using a probe without the flow cell and the probe is in air out of the sample check the instrument without the probe to determine if the probe has failed or is out of tune If minimum reading is obtained using the converter without the probe the probe is defective and should be changed A loose probe will usually generate a loud piercing sound Refer to Section Ill if an overload condition exists 3 Depress the START key 4 Using the A or Y keys increase or decrease the amplitude as required TIMER In the pulsed mode the processing time will be different from the elapsed time because the processing time function monitors and controls only the ON portion of the duty cycle For example for 1 hour processing time the elapsed time will be 2 hours if the ON and OFF cycle are set for 1 second To set the processing time press the TIMER key The screen will display Using the numeric keys set the processing time as required Press the ENTER REVIEW key The screen will display PULSER By inhibiting heat build up in the sample the pulse function enables safe treatment of temperature sensitive samples at high intensity In addition pulsing enhances processing b
33. t concentration of energy is beneath the probe it is imperative that the sample be kept as close to the tip as possible liquids are easily processed because the free moving cells circulate repeatedly below the probe Solid materials however have a tendency to be repelled by the ultrasonic and should be processed in a vessel large enough to accommodate the probe yet small enough to restrict sample movement For small samples conical shaped test tubes are recommended Although plastic tubes work well glass and stainless steel tubes usually work better than plastic ones Make sure that the probe is not touching the bottom of the vessel Allowing the probe to contact the vessel will decrease the power output and cause minute grey glass particles to migrate into the sample Although these glass particles will not adversely affect the chemical composition of the sample they will form a thin grey layer on centrifuging If the probe has to come in contact with a solid sample use a standard 20mm 3 4 diameter stainless steel centrifuge tube cut to 70mm 3 length Do not use a glass tube Microtips must never come in contact with anything but the liquid because the stress resulting at the point of contact with a hard surface will cause the microtip to fracture Although larger probes will not fracture if they come in contact with a glass vessel they may cause the vessel to fracture Before each application place the tip in 100 ethanol and energi
34. ure can be repeated as long as the power monitor reading is less than 20 with the probe out of the sample when the AMPLITUDE control is set at 100 If the reading exceeds 20 the probe or replaceable tip should be replaced with a new one SECTION lll SERVICE INFORMATION QA e iS Y 4 La pa A a a aes iy ii Are T A 144 p ines i a i P e E Y A 7 Your Ultrasonic Processor was designed to provide you with years of safe and dependable service Nevertheless because of component failure or improper usage the possibility does exist that it might not perform as it should shut down due to an overload condition or that it will stop working all together The most probable causes for malfunction are listed below and should be investigated The unit was plugged into an electrical outlet that provides a different voltage from the one required See Electrical Requirements The probe or booster is not secured properly A fuse s has failed 1 Set the AMPLITUDE to 50 and press the I and the START STOP key With the flow cell empty of any liquid or the probe in air out of sample the wattmeter should read below 10 watts If the reading exceeds 10 watts press the START STOP key and disconnect the probe from the booster 2 Press the START STOP key to restart the equipment If the wattmeter reads below 20 watts the probe has failed or is out of tune due to excessive erosion and should be repl
35. y allowing the material to settle back under the probe after each burst The ON and OFF pulse duration can be set independently from 1 second to 59 seconds During the OFF portion of the cycle the red indicator on the PULSER key will illuminate If the OFF portion of the cycle exceeds three seconds a cautionary message CAUTION PROBE ON STANDBY will warn the operator against touching the ultrasonic probe To set the pulser press the PULSER key The screen will display Using the numeric keys set the ON portion of the cycle then press the ENTER REVIEW key The screen will display e g Using the numeric keys set the OFF portion of the cycle The screen will display e g Press the ENTER REVIEW key The screen will display On Cycle Off Cycle TEMPERATURE The temperature function prevents overheating of the sample by continuously monitoring the sample temperature and terminating the ultrasonics when the temperature reaches a predetermined setpoint The ultrasonics is automatically reinstated when the temperature drops below the setpoint If the temperature of the sample must be monitored and or controlled insert the optional temperature probe forcefully info the small jack on the rear panel immerse the temperature probe in the sample and press the TEMP key The screen will display Using the numeric keys set the high temperature limit Setpoint The screen will display Press the ENTER REVIEW key
36. ze the power supply for a few seconds to remove any residual substances If still concerned about contamination from previous use clean the probe with a disinfectant such as 20 Virkon solution and rinse with distilled water Probes are autoclavable To inhibit sample loss in test tube due to sticking siliconize the test tube as follows Wash and dry the test tube thoroughly coat with silicone then air dry Sigmacote manufactured by Sigma Chemical Co 3050 Spruce Street St Louis Missouri 63103 USA phone 314 771 5765 is ideally suited for that purpose High viscosity and concentration are problematic 2 000 cps and 15 concentration by weight are maximum limits Ultrasonic processing propagates sound waves through the sample If the sample is so dense that it will not pour or circulate easily it will absorb the sound waves and be too thick for ultrasonic processing Use the Cup Horn for processing pathogenic radioactive and biohazardous materials in complete isolation without probe intrusion Because plastic tubes have a tendency to absorb vibrations it is preferable to contain the sample in a stainless steel tubes or glass tubes when working with a cup horn To expedite processing add glass beads to the sample If desired crushed ice can also be added to the water inside the cup horn in order to optimize cooling Processing samples in a Cup Horn will usually take 4 times longer than processing with direct probe intrusion V