Home
        NanoVue™ Plus Spectrophotometer
         Contents
1.   2  Sale  lease  license or other grant of rights to use this material or any material derived or produced from it   3  Use of this material to perform services for a fee for third parties  including contract research and drug screening     If you require a commercial license to use this material and do not have one  return this material unopened to  GE Healthcare Bio Sciences AB  Bjorkgatan 30  SE 751 84 Uppsala  Sweden and any money paid for the  material will be refunded    All third party trademarks are the property of their respective owners        2008   2010 General Electric Company   All rights reserved  First published Jan  2008    All goods and services are sold subject to the terms and conditions of sale of the company within  GE Healthcare which supplies them  A copy of these terms and conditions is available on request   Contact your local GE Healthcare representative for the most current information    GE Healthcare Europe GmbH   Munzinger Strasse 5  D 79111 Freiburg    Germany  GE Healthcare Bio Sciences Corp   800 Centennial Avenue  P O  Box 1327  Piscataway  NJ 08855 1327  USA  GE Healthcare  Japan Corporation   Sanken Bldg   3 25 1  Hyakunincho  Shinjuku ku  Tokyo 169 0073  Japan       GE Healthcare Bio Sciences AB  Bjorkgatan 30   751 84 Uppsala   Sweden    28 9301 69 AC 09 2010    
2.  chromatography     The instrument enables you to characterize oligonucleotide  primers by keying in the base sequence  up to 66 mer  to  obtain a conversion factor  ug ml   molecular weight   theoretical absorbance  AU umol   and theoretical Tm     DNA linearity    NanoVue Plus shows excellent linearity with double stranded  DNA  dsDNA  from 5 to 6000 ug ml  Fig 4   This large linear  dynamic range and high reproducibility means that you can  proceed to the next stage of your workflow with a high  degree of confidence in the data produced by NanoVue Plus  Spectrophotometer                 Calculated value  mg ml   N       0 1 2 3 4 5 6  DNA  mg ml     Fig 4  DNA concentration curve  0 5 to 6000 pg ml  generated from salmon  sperm dsDNA samples measured at 260 nm using NanoVue Plus  n   20  replicates  mean   1 standard deviation  SD       Proteins    NanoVue Plus Spectrophotometer can be used to determine  the concentration of protein samples by a variety of  methods including Bradford  BCA   Lowry  Biuret  and direct  UV methods with a choice of line fit and the ability to run up  to 27 standards  including replicates   These calibration  curves can be viewed on the graphical display  printed or  stored as a method for future use     Protein concentration can be determined in the near UV at  280 nm due to absorption by tyrosine  tryptophan  and  phenylalanine amino acids  The absorbance at 280 nm  varies greatly for different proteins because of their amino  acid content th
3.  contamination of the sample mechanism thus  improving data precision  Fig 2   The plate is compatible  with a wide range of chemicals including DMSO and DMF    e Low sample volumes  Reduce sample loss and eliminate  the need for dilution by using low volumes of 0 5 to 5 ul  for sample measurement  The appropriate pathlength for  your sample volume is selected automatically unless you  opt for manual selection    e Fast measurements  Read time is typically  lt  5 s per  sample and you save more time by not having to wash  cuvettes or dilute samples          Fig 1  NanoVue Plus Spectrophotometer uses a novel drop and measure  sample plate which eliminates the need for cuvettes     e Convenience and ease of use  Large  high resolution  graphical display enables a quick read of relevant results   Calibration curves  kinetics or ratio measurements are all  displayed at the touch of a button  Fig 3     e Flexible analytical performance  Full wavelength scan in  less than 5 s from 200 to 900 nm with zoom facility  peak  identification  and on peak confirmation  Visualization of  nucleic acid scans allows impurities to be detected and  this is especially useful with RNA samples    e Choice of data output  Print to an integrated printer   optional  or to any suitable PC via a USB  wireless   Bluetooth  connection or SD card    e Reliability and robust instrumentation  Press to read  feature reduces the amount of time the lamp stays  switched on  Optics with no moving parts minimizes 
4.  the incidence of optical misalignment       w           E oe ee            te ee    m  a    nN        pa        BSA carry over data  mg ml   w a          0 D  aaa OD    Sa  0 5 10 15 20 25    Sample number       O    water pre BSA sample       BSA sample           water post BSA sample    Fig 2  Sample carry over study using 25 mg ml BSA with successive additions  of a water protein water cycle of samples  which was repeated 25 times   The data showed no significant change in absorbance readings during the  study for water sample readings or BSA samples readings thus indicating  that there was no sample carry over     NanoVue Plus Life Science   Protein    oN oN  32    Life Science       o  amp   o  amp   oe    Protein UV Protein A280 BCA    of  eo ot    Bradford Lowry Biuret    Applications Favourites       o   o    Methods Utilities    A   B     Wavescan    ot DNA RE  0    A230 1 53A Sample   A260 2 39A A    A280 0 94 A  A320 0 18A                   Concentration    110 5    A260 A280 Units A  261  271  318             A260 A280  2 908                         1 637 pg ml Abs  0 171 0 204 0 104  Sample 4    C  D                                                  A 285nmA 0 036 A                   Fig 3  NanoVue Plus spectrophotometer screenshots showing   A  Simple  interface for easy selection of applications   B  Protein methods interface   C  DNA quantitation results and  D  An example of a wavelength scan        The key features of NanoVue Plus Spectrophotometer are  summ
5. 0 2116 30  illustra    plasmidPrep Mini Spin Kit 28 9042 69  illustra GenomiPhi    V2 DNA Amplification Kit 25 6600 30  illustra RNAspin Mini Kit 25 0500 71  illustra HotStart PCR Master Mix 25 1500 01  illustra GFX    PCR DNA and Gel Band Purification Kit 28 9034 70       GE  imagination at work  and GE monogram are trademarks of General Electric Company     Amersham  Cy  CyDye  GenomiPhi  GFX  HyPer  illustra  and NanoVue are trademarks of GE Healthcare  companies    CyDye  This product or portions thereof is manufactured under an exclusive license from Carnegie Mellon  University under US patent number 5 268 486 and equivalent patents in the US and other countries  Cy3 UTP  or Cy5 UTP  Cy3 5 dCTP or Cy5 5 dCTP  Cy3 CTP or Cy5 CTP  These products are manufactured for GE  Healthcare UK Limited by Perkin Elmer Life Sciences under US patent numbers 5047519 and 5151507   The  cyanine dyes in the product are manufactured under an exclusive license from Carnegie Mellon University  under US patent numbers 5 268 486 and equivalent patents in the US and other countries  The purchase of  CyDye products includes a limited license to use the CyDye products for internal research and development  but not for any commercial purposes  A license to use the CyDye products for commercial purposes is subject  to a separate license agreement with GE Healthcare  Commercial use shall include     1  Sale  lease  license or other transfer of the material or any material derived or produced from it 
6. GE Healthcare    Data file 28 9301 69 AC    UV Visible spectrophotometers    NanoVue    Plus Spectrophotometer    NanoVue Plus Spectrophotometer  Fig 1  is a stand alone  instrument with a novel drop and measure sample plate  that can be used with a wide range of chemicals  NanoVue  Plus can be used for the accurate and reliable quantitiation  of nucleic acids and protein samples  The ability to pipette  samples of 0 5 to 5 ul directly onto the sample plate  eliminates the need for cuvettes or other sample devices   After measurement  the sample can either be easily  recovered with a pipette or quickly discarded by wiping the  sample plate clean for the next measurement     NanoVue Plus is pre programmed with a range of preset  methods for the quantitation of nucleic acids and proteins   using UV or dye intercalation methods  Lowry  Bradford   BCA  and Biuret   In addition  wavelength scanning gives you  the flexibility to design your own methods  which can be  stored in a personal folder for easy access  The instrument  does not require a computer because it has a direct user  interface  Fast instrument start up with automatic self   calibration means NanoVue Plus Spectrophotometer is  always ready to use     NanoVue Plus Spectrophotometer offers     e Novel gold colored hydrophobic sample plate  Save  time and effort by directly pipetting your sample onto the  sample plate for measurement  The hydrophobic coating  facilitates sample recovery and minimizes the potential  for
7. arized in Table 1 below     Table 1  Features of NanoVue Plus Spectrophotometer    Feature Specification       200 to 1100 nm    Automatic upon switch on    Wavelength range  Wavelength calibration  Pathlength selection  Stored methods 90    Automatic with the option for manual selection       Spectral bandwidth 5nm   Wavelength accuracy  2nm   Wavelength reproducibility   0 5 nm   Absorbance accuracy Maximum   1   at 259 nm  at 0 7 to 0 8 A  using uracil    Light source Long life stabilized Xenon    Detector Twin CCD Array  Dimensions 260 x 390 x 100 mm  Weight 4 5 kg    Power requirements 100 to 240 VAC   10   50 60 Hz  50 VA       Note  Technical specifications were measured after the instrument had warmed up  at a constant ambient temperature     2 28 9301 69 AC    Assays    Nucleic acids    NanoVue Plus Spectrophotometer allows you to measure  the concentration and purity of nucleic acids in a variety of  units  ug ml  ng ul  ug ul  pmol ul  and pmol  and also  to  correct for dilution factors where necessary     NanoVue Plus displays both individual absorbance values  and absorbance ratios  260 280 and 260 230  on the screen   along with the sample concentration value  A graphical  display of the result is also available at the touch of a  button  The results can be printed immediately with the  on board printer option     You can examine the UV spectrum  220 to 330 nm  for  hybridization  PCR  and sequencing studies or for  quantitation of minipreps after isolation by 
8. erefore  the specific absorption value for a  particular protein must be determined     NanoVue Plus has in built extinction coefficients for BSA   IgG  and lysozyme thus allowing you to select the most  appropriate for your protein sample or you can input your  own extinction coefficients  Typical results for BSA and IgG  values using the absorbance at 280 nm program are shown  in Figures 5 and 6  respectively               ra0 9997    Calculated value  mg ml        0 10 20 30 40 50 60 70 80 90 100  BSA  mg ml     Fig 5  Protein concentration curve  0 01 to 100 mg ml  generated from BSA    protein samples measured at 280 nm using NanoVue Plus  n   20 replicates     mean   1 SD               Calculated value  mg ml           0 10 20 30 40 50  IgG  mg ml     Fig 6  Protein concentration curve  0 025 to 50 mg ml  generated from  bovine IgG samples measured at 280 nm using NanoVue Plus protein  absorbance 280 program with IgG preset option  n   20 replicates   mean   1 SD      Background correction    NanoVue Plus provides increased confidence in your results  by allowing you to use the background correction feature  at 320 nm to compensate for the effects of background  absorbance caused by turbidity  stray particulates  and  high absorbance buffer solutions  This feature of the  instrument is critical because very small samples are  susceptible to stray particulates  The results are displayed  on the screen with other absorbance results for quick  evaluation  You have the flex
9. ibility to turn the background  correction function on or off     CyDye       You can also measure purity  yields  and brightness of  fluorescently labeled in situ hybridization probes  CyDye  DNA function holds spectral data for 19 commonly used  dyes  including Cy   2  Cy3  Cy3B  Cy5  and Amersham     Hyper   5 and both sample concentration and frequency  of incorporation measurements are presented     Calibration    An accurate pathlength distance is important for the correct  functioning of all low volume spectrophotometers  We  recommend a regular pathlength check as a matter of good  laboratory practice  The pathlength calibration fluid supplied  with the instrument allows for simple checking of the  instrument pathlength whenever convenient for added  confidence in performance  Should re calibration be  necessary  you can easily and quickly recalibrate NanoVue  Plus in your laboratory without the inconvenience and cost  of sending the instrument away     28 9301 69AC 3    Drop  Measure  Done     NanoVue Plus provides intelligent performance across all  spectrophotometry applications  Fig 7      1  Just pipette your sample onto  the sample    2  Lower the sample plate head    3  Read your result                                                       st Protein UV     A260    Sample  2 33A 1  A280 Result    3 8 6  A320  0 000 A  Units  3 g ml          Fig 7  Operation of NanoVue Plus Spectrophotometer     Summary    NanoVue Plus Spectrophotometer is an easy to use and  re
10. liable instrument for measuring nucleic acid and protein  samples  Samples of 0 5 to 5 ul can be pipetted directly onto  a novel sample plate for measurement  and then simply  recovered using a pipette  In certain cases where sample  recovery is not necessary  the sample plate can be quickly  and easily wiped clean  This novel feature eliminates the  need for cuvettes  capillaries or other sample devices      just drop  measure  and you re done     For local office contact information  visit  www gelifesciences com spectros    GE Healthcare UK Limited  Amersham Place   Little Chalfont  Buckinghamshire  HP7 9NA  UK    www gelifesciences com    imagination at work       Ordering information       Product Code No    NanoVue Plus with integrated printer  US  amp  CA only  28 9569 63  anoVue Plus with integrated printer    Worldwide except US  amp  CA  28 9569 66   NanoVue Plus  print via computer  US  amp  CA only  28 9569 62   NanoVue Plus  print via computer    Worldwide except US  amp  CA  28 9569 65   NanoVue Plus with wireless connectivity  US  amp  CA only  28 9569 64   NanoVue Plus with wireless connectivity    Worldwide except US  amp  CA  28 9569 67  anoVue Plus with SD card  US  amp  CA only  28 9569 61  anoVue Plus with SD card  Worldwide except US  amp  CA  28 9569 60             Pathlength calibration kit  Worldwide  28 9244 05  Sample Plate Replacement Kit 28 9569 58  Lower Sample Plate Replacement Kit 28 9569 59  Related products   Ultrospec 10 Cell density meter 8
    