SeQual Equinox User Manual

image

Contents

1. 19 eQuinox User Manual 3 eQuinox Quick Start Guide Unpack Your eQuinox Getting to Know Your eQuinox Review all Warnings Cautions and additional device information in the rest of this manual Become familiar with the key features of the eQuinox and the User Control Panel Control E hee Handle External Power Receptacle Oxygen Outlet Port AC Power Supply FAA WITH NEMA POWER CORD Power Cartridge Latch Rating Label amp i eQuinox Cart Serial Number ee 27 Location Cartridge Battery BACK DC Cable Service Cabinet Exhaust Vent Filter Air inlet Filter Power Cartridge battery Service Port 4 4805 C3_01A The eQuinox User Control Panel displays important operating information amp ON OFF Button Green Indicator This button powers the device ON or OFF The Green Indicator is illuminated when the device is ON and functioning properly Increase or Decrease Flow Setting Buttons Use these buttons to set the flow to your prescribed setting Delivery Mode Button and Indicator The button toggles between Continuous Flow and Pulse Dose Mode The Pulse Dose Mode activates autoSAT Technology and allows a significant increase in the operating time while powered by the battery When the Pulse Dose Mode is activated the green Pulse Dose Mode Indicator ill
2. CONNECTION TO MAKE SURE IT REMAINS FREE OF DUST WATER AND PARTICLES 18 4805 C3_01A WARNING DO NOT USE LIQUID DIRECTLY ON THE UNIT A LIST OF UNDESIRABLE CHEMICAL AGENTS INCLUDES BUT IS NOT LIMITED TO THE FOLLOWING ALCOHOL AND ALCOHOL BASED PRODUCTS CONCENTRATED CHLORINE BASED PRODUCTS ETHYLENE CHLORIDE AND OIL BASED PRODUCTS PINE SOL LESTOIL THESE ARE NOT TO BE USED TO CLEAN THE PLASTIC HOUSING ON OXYGEN CONCENTRATOR AS THEY CAN DAM AGE THE UNIT S PLASTIC Optional Humidifier If used clean your humidifier daily to reduce the possibility of contamination Follow the cleaning recommendations provided by the home care provider or the manufacturer of the humidifier Replace the humidifier monthly or as recommended by the manufacturer or your home care provider Do not overfill the humidifier CAUTION DO NOT operate the eQuinox and ambulate while a humidifier is attached Remove the humidifier bottle before walking DO NOT lay the eQuinox down while attached to a humidifier bottle Universal Cart Before cleaning the Universal Cart remove it from the eQuinox Use a damp not wet cloth or sponge and mild detergent solution to clean the Universal Cart Power Cartridge Battery The Battery in the eQuinox requires special care to assure a longer life and the highest level of performance The SeQual Battery is the only approved Battery recommended for use with the eQuinox Use a damp
3. IN RAPID BURNING CAUSING PROPERTY DAMAGE BODILY INJURIES OR DEATH USE NO OIL GREASE OR PETROLEUM BASED OR OTHER FLAMMABLE PRODUCTS WITH THE OXYGEN CARRYING ACCESSORIES OR THE OXYGEN CONCENTRATOR ONLY WATER BASED OXYGEN COMPATIBLE LOTIONS OR SALVES SHOULD BE USED OXYGEN ACCELERATES THE COMBUSTION OF FLAMMABLE SUB STANCES WARNING THIS DEVICE SUPPLIES HIGH CONCENTRATION OXYGEN THAT PROMOTES RAPID BURNING DO NOT ALLOW SMOKING OR OPEN FLAMES WITHIN THE SAME ROOM OF 1 THIS DEVICE OR 2 ANY OXYGEN CARRYING ACCESSORY FAILURE TO OBSERVE THIS WARNING CAN RESULT IN SEVERE FIRE PROPERTY DAMAGE AND OR CAUSE PHYSICAL INJURY OR DEATH WARNING DO NOT USE YOUR OXYGEN CONCENTRATOR IN Note When the automobile in which you are using the Oxygen Con centrator unit is turned off disconnect and remove the device from the automobile Do not store the Oxygen Concentrator in a very hot or cold automobile or in other similar high or low temperature environments DO NOT leave the Oxygen Concentrator or the Power Supply plugged into the vehicle if the ignition is in the off position Doing so may drain the vehicle s battery CAUTION Do not operate unit in a restricted or confined space 1 a small case or handbag where ventilation can be limited This can cause the Oxygen Concentrator to overheat and impair performance Do not allow either the air intake or the air outlet vents to become blocked This can cause the Oxygen Concentr
4. S MAY NEED ASSISTANCE WITH MONITORING ALARMS WARNING IT IS VERY IMPORTANT TO SELECT ONLY THE Contraindications for Use WARNING IN CERTAIN CIRCUMSTANCES THE USE OF NON PRESCRIBED OXYGEN CAN BE HAZARDOUS THIS DEVICE SHOULD ONLY BE USED WHEN PRESCRIBED BY A PHYSICIAN WARNING NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANES THETICS WARNING AS WITH ANY ELECTRICALLY POWERED DEVICE THE USER MAY EXPERIENCE PERIODS OF NON OPERATION AS A RESULT OF ELECTRICAL POWER INTERRUPTION OR THE NEED TO HAVE THE EQUINOX SERVICED BY A QUALIFIED TECHNICIAN THE EQUINOX IS NOT APPROPRIATE FOR ANY PATIENT WHO WOULD EXPERIENCE ADVERSE HEALTH CONSEQUENCES AS THE RESULT OF SUCH TEMPO RARY INTERRUPTION eQuinox Safety Guidelines WARNING NO MODIFICATION OF THIS EQUIPMENT IS PERMITTED ij WARNING THE MANUFACTURER RECOMMENDS AN ALTER NATE SOURCE OF SUPPLEMENTAL OXYGEN IN THE EVENT OF A POWER OUTAGE ALARM CONDITION OR MECHANI CAL FAILURE CONSULT YOUR PHYSICIAN OR EQUIPMENT PROVIDER FOR THE TYPE OF RESERVE SYSTEM REQUIRED WARNING THIS DEVICE SUPPLIES HIGH CONCENTRATION OXYGEN THAT PROMOTES RAPID BURNING DO NOT ALLOW SMOKING OR OPEN FLAMES WITHIN THE SAME ROOM OF 1 THIS DEVICE OR 2 ANY OXYGEN CARRYING ACCESSORY FAILURE TO OBSERVE THIS WARNING CAN RESULT IN SEVERE FIRE PROPERTY DAMAGE AND OR CAUSE PHYSICAL INJURY OR DEATH WARNING DO NOT OPERATE UNIT IN A RESTRICTED OR CONFINED SPACE I E A SMALL CASE OR HANDBAG
5. and Indicator Press the Flow Mode button to select your desired flow mode Pressing this button repeatedly will toggle you back and forth between continuous flow and pulse dose modes In Continuous Flow Mode the oxygen is provided at a constant flow rate between 0 5 and 3 0 LPM In Pulse Dose Mode oxygen is supplied in a bolus at the beginning of each inspiration providing a selectable range setting of l6mL to 192mL and can be adjusted for sensitivity and breath bolus delivery Flow settings both modes should be prescribed for you by your physician Adjust Flow Setting to Prescribed Level A Increase or Decrease Flow Setting Button Using the or buttons adjust the setting to the prescription supplied by your physician Continuous Flow Mode Display Pulse Dose Mode Display SCRIBED LEVEL OF OXYGEN DO NOT CHANGE THE FLOW WARNING IT IS VERY IMPORTANT TO SELECT ONLY THE PRE SELECTION UNLESS YOU HAVE BEEN DIRECTED TO DO SO BY A LICENSED CLINICIAN THE PORTABLE OXYGEN CONCENTRATOR MAY BE USED DURING SLEEP UNDER THE RECOMMENDATION OF A QUALIFIED CLINICIAN 6 4805 C3 01A Begin Using the eQuinox Air Inlet Filter Air inlet Filter Your unit is equipped with an Air Inlet Filter extra filter provided Check daily and clean once a week Connect the oxygen supply tubing to the oxygen outlet and connect the oxygen cannula per the instructions provided with the cannula or per the instructions from yo
6. illuminate The unit will still operate on external power Rechargeable Power Cartridge Battery To remove the Battery pinch the black release lever While holding DOWN the release lever pull the Battery away from the eQuinox Battery Operating Times The battery operating time is displayed on the Control Panel in both symbol and alpha numeric formats A variety of factors such as flow setting Pulse or Continuous Flow Mode and breath rate will impact the operating time The table Typical New Power Cartridge Operating Times provides operating time estimates for the eQuinox using a new fully charged Battery based on flow settings and operating conditions as indicated BATTERIES STORE IN A COOL AND DRY LOCATION TO HELP ENSURE THE LONGEVITY OF YOUR BATTERY STORING YOUR OXYGEN CONCENTRATOR FOR EXTENDED PERIODS OF TIME AT HIGH WARNING FOR OXYGEN CONCENTRATORS EQUIPPED WITH TEMPERATURES OR WITH A FULLY CHARGED OR COMPLETELY DIS CHARGED BATTERY CAN DEGRADE ITS OVERALL BATTERY LIFE DO NOT ATTEMPT TO OPEN THE BATTERY THERE ARE NO SERVICEABLE PARTS INSIDE THE BATTERY KEEP BATTERIES AWAY FROM CHILDREN 12 4805 C3 01A Initial Battery Charging The new Battery supplied with your eQuinox 1 not fully charged when it 1s shipped from the factory Before using your eQuinox for the first time you must first fully charge the Battery With the AC Power Supply plugged in and the Battery correctly installed in the Power
7. not wet cloth or sponge to clean the Battery First spray the cloth or sponge with a mild detergent then clean the Battery case and the latch SeQual Recommended Monthly Battery Service Run the eQuinox on Battery completely down to power loss condition red light and alarm Plug eQuinox into AC power and recharge until full on indicator The device does not have to be running to charge which allows for shorter charge time SeQual recommends replacing the power cartridge after 2 years Battery Disposal Your Battery is rechargeable and can be recycled Always return it to your home care provider for proper disposal You can also contact your local city or town offices for instructions on proper disposal of the Battery General Airline Travel Information During taxi take off and landing the eQuinox must be turned off and stowed under the seat or in another approved stowage location so as to not block the aisle way or the entry way into the row if the eQuinox will not be used During taxi take off and landing the user must be in a seat location that does not restrict any other passenger s access to or use of any required emergency or regular exit or the aisle s in the passenger compartment of the aircraft if the eQuinox is used User s are not permitted to be seated in an exit row if using the eQuinox If the eQuinox is used when decompression of the cabin occurs and the oxygen cabin oxygen system deploys then the us
8. your eQuinox The device is capable of being operated directly from three 3 different power sources AC Power Supply with NEMA Power Cord e 2VDC Cable Rechargeable Power Cartridge Battery General Charging Information The eQuinox operates from external AC power DC power or a rechargeable battery The system includes a smart battery charger that recharges the internal battery whenever the eQuinox is connected to external power The system monitors and controls the power source and the battery charger The eQuinox charges the battery when using AC power and when the battery temperature is within safe charging temperature When external power is disconnected the device will automatically switch over to the battery as long as the battery is installed and charged When external power is restored by plugging into an AC outlet the device will accept power from the external power source and replenish the battery Proper connection of the AC Supply or 12VDC cable requires aligning the power plug with the recessed external power receptacle on the eQuinox The recessed external power receptacle is located on the right side of the device when facing the Control Panel Push the power plug into the recessed receptacle 10 4805 C3_01A AC POWER SUPPLY AC Power Supply with NEMA Power Cord The eQuinox includes a universal AC Power Supply for use at home or wherever standard AC power is available To connect to an A
9. ABLE CONCENTRATOR IS NOT IN USE WARNING SMOKING WHILE USING OXYGEN IS THE NUM CAUTION Always disconnect AC power supply from the wall before disconnection the AC power supply from the oxygen concentrator CAUTION Always place oxygen supply tubing and power cords in a manner that prevents a trip hazard CAUTION When using the Oxygen Concentrator in an automobile boat or on other DC sources with the 12VDC Cable make sure that the vehicle is started and running before connecting the Oxygen Concentrator If the 12VDC Cable does not illuminate and requires resetting disconnect the 12VDC Cable from the DC outlet restart your vehicle and then reconnect your 12VDC Cable into the DC outlet Failure to follow these instructions can result in the power supply not supplying power to the Oxygen Concentrator CAUTION When the automobile in which you are using the Oxygen Concentrator unit is turned off disconnect and remove the device from the automobile Do not store the Oxygen Concentrator in a very hot or cold automobile or in other similar high or low tem perature environments Note Portable and mobile RF communications equipment can effect medical electrical equipment Recommended Operating Environments For proper use of the device the following chart provides important information concerning the recommended operating environments or operating conditions Specifications Operating 32 F to 104 F 0 C to 40 C Tem
10. ARTRIDGES NEEDED Table accounts for the typical 150 duration requirement Pulse Dose Mode FLIGHT LENGTH in Hrs 10 2 0 3 0 40 5 0 60 70 8 mL SETTING Table is based on 20 Breaths Per Minute for Pulse Dose Mode only Table is based on 15 Breaths Per Minute for Pulse Dose Mode only eQuinox User Manual 19 eQuinox ul CAIRE Inc 2200 Airport Industrial Drive Ste 500 Ball Ground GA 30107 USA 1 800 482 2473 Worldwide www CAlREmedical com Medical Product Services GmbH Borngasse 20 35619 Braunfels Germany Worldwide www CAlREmedical com 0029 4277 CAIRE A Chart Industries Company AirSep SeQual SeQual and eQuinox and autoSAT are trademarks of Chart SeQual Technologies Inc Ball Ground GA 30107 Chart SeQual Technologies Inc reserves the right to discontinue its products or change the prices materials equipment Salter Labs registered mark of Salter Labs Arvin CA 92303 quality descriptions specifications and or processes to its products at any time without prior notice and with no further j obligation or consequence All rights not expressly stated herein are reserved by us as applicable Lysol Isa registered trademark of Reckitt Benckiser UK PN 4805 C3 01A
11. C power source be sure the AC Power Supply cord is securely inserted into the recessed power receptacle on the side of the device and that the cord from the AC Power Supply is plugged into a grounded AC outlet When the eQuinox is plugged in properly a green indicator on the AC Power Supply will light and the External Power Indicator 47 will appear on the user control panel Note Use only electrical voltage specified on the specification label affixed to the device CAUTION Use only accessories recommended by the manufacturer Use of any other may be hazardous cause serious damage to your oxygen concentrator A and will void the warranty CAUTION Do not use extension cords with this unit or connect too many plugs into the same electrical outlet The use of extension cords could adversely affect the performance of the device Too many plugs into one outlet can result in an overload to the electrical panel causing the breaker fuse to activate or fire if the breaker or fuse fails to operate CAUTION Use of cables and adapters other than those specified with the exception of cables and adapters sold by the manufac turer of the medical electrical equipment as replacement parts for internal components may result in increased emissions of decreased immunity of the Oxygen Concentrator eQuinox Note While operating the device from a 12VDC Cable the Battery may 12VDC Cable not charge Note When operating the eQ
12. Compartment of the device allow the Battery to completely charge The Battery is fully charged when the Power Cartridge Status Gauge on the Control Panel is no longer cascading The eQuinox can be used while either discharging or recharging the Battery It may take up to 4 hours dependent upon the flow setting to achieve 80 capacity from a fully discharged Battery It will charge when the eQuinox is plugged into AC power and turned off CAUTION When the automobile in which you are using the Oxygen Concentrator unit is turned off disconnect and remove the device from the automobile Do not store the Oxygen Concentrator in a very hot or cold automobile or in other similar high or low temperature environments DO NOT leave the Oxygen Concentrator or the Power Supply plugged into the vehicle if the ignition is in the OFF position Doing so may drain the vehicle s battery of the SeQual eQuinox CAUTION Only use the Manufacturer s provided batteries For prop er battery disposal contact your Equipment Provider or your local government agency for disposal requirements PS Batteries are recyclable In the event of an AC or DC power interruption the eQuinox will beep and automatically switch to Battery operation as long as the Battery is present and charged When AC power is restored the eQuinox will beep and the Battery will automatically start recharging if not fully charged or too warm If the Battery is not p
13. IS CAN CAUSE THE WARNING THE MANUFACTURER RECOMMENDS AN ALTER OXYGEN CONCENTRATOR TO OVERHEAT AND IMPAIR PERFORMANCE 8 4805 C3_01 A BER ONE CAUSE OF FIRE INJURIES AND RELATED DEATHS YOU MUST FOLLOW THESE SAFETY WARNINGS WARNING DO NOT ALLOW SMOKING CANDLES OR OPEN FLAMES IN THE SAME ROOM WITH THE DEVICE OR THE OXYGEN CARRYING ACCESSORIES WARNING SMOKING WHILE WEARING AN OXYGEN CANNULA MAY CAUSE FACIAL BURNS AND POSSIBLY DEATH WARNING IF YOU SMOKE THESE 3 STEPS MAY SAVE YOUR LIFE TURN OFF THE OXYGEN CONCENTRATOR TAKE OFF THE CANNULA AND LEAVE THE ROOM WHERE THIS DEVICE IS LOCATED WARNING NO SMOKING OXYGEN IN USE SIGNS MUST BE PROMINENTLY DISPLAYED IN THE HOME OR WHERE OXYGEN IS IN USE PATIENTS AND THEIR CAREGIVERS MUST BE INFORMED ABOUT THE DANGERS OF SMOKING IN THE PRESENCE OF OR WHILE USING MEDICAL OXYGEN WARNING DO NOT USE YOUR OXYGEN CONCENTRATOR IN THE PRESENCE OF FLAMMABLE GASES THIS CAN RESULT IN RAPID BURNING CAUSING PROPERTY DAMAGE BODILY INJURIES OR DEATH WARNING REMOVING THE CANNULA AND PUTTING IT ON CLOTH ING BEDDING SOFAS OR OTHER CUSHION MATERIAL WILL CAUSE A FLASH FIRE WHEN EXPOSED TO A CIGARETTE HEAT SOURCE OR FLAME WARNING DO NOT LEAVE A NASAL CANNULA ON OR UNDER CLOTH ING BED COVERINGS OR CHAIR CUSHIONS IF THE UNIT IS TURNED ON BUT NOT IN USE THE OXYGEN WILL MAKE THE MATERIAL FLAM MABLE SET THE I O POWER SWITCH TO THE 0 OFF POSITION WHEN THE PORT
14. K HAZARD TURN OFF THE UNIT AND DISCONNECT THE POWER CORD FROM THE ELECTRICAL OUTLET BEFORE YOU CLEAN THE UNIT TO PREVENT ACCIDENTAL ELECTRICAL SHOCK AND BURN HAZARD ONLY YOUR EQUIPMENT PROVIDER OR A QUALIFIED SERVICE TECHNICIAN SHOULD REMOVE THE COVERS OR SERVICE THE UNIT DO NOT USE LIQUID DIRECTLY ON THE UNIT A LIST OF UNDESIRABLE CHEMICAL AGENTS INCLUDES BUT IS NOT LIMITED TO THE FOLLOWING AL COHOL AND ALCOHOL BASED PRODUCTS CONCENTRATED CHLO RINE BASED PRODUCTS ETHYLENE CHLORIDE AND OIL BASED PRODUCTS PINE SOL LESTOIL THESE ARE NOT TO BE USED TO CLEAN THE PLASTIC HOUSING ON OXYGEN CONCENTRATOR AS THEY CAN DAMAGE THE UNIT S PLASTIC WARNING ELECTRICAL SHOCK HAZARD DISCONNECT THE POWER CORD FROM THE ELECTRICAL OUTLET BEFORE YOU CLEAN THE UNIT TO PREVENT ACCIDENTAL ELECTRICAL SHOCK AND BURN HAZARD ONLY YOUR EQUIPMENT PROVIDER OR A QUALIFIED SERVICE TECHNI CIAN SHOULD REMOVE THE COVERS OR SERVICE THE UNIT WARNING CARE SHOULD BE TAKEN TO PREVENT THE OXYGEN CON CENTRATOR AND THE BATTERY FROM GETTING WET OR ALLOWING FLUIDS TO ENTER THE UNIT THIS CAN CAUSE A MALFUNCTION OR SHUT DOWN AND CAUSE AN INCREASED RISK FOR ELECTRICAL SHOCK OR BURNS WARNING CLEAN THE CABINET CONTROL PANEL AND POWER CORD ONLY WITH A MILD HOUSEHOLD CLEANER APPLIED WITHA DAMP NOT WET CLOTH OR SPONGE AND THEN WIPE ALL SURFACES DRY DO NOT ALLOW ANY LIQUID TO GET INSIDE THE DEVICE PAY SPECIAL ATTENTION TO THE OXYGEN OUTLET FOR THE CANNULA
15. M on continuous flow while being operated with the DC Supply in a car The eQuinox will charge the battery while on 12VDC Cable in hil fl or Note Actual performance of the eQuinox may vary depending upon usage a car while running on continuos flow at or less If the eQuinox battery is being charged the battery symbol will waterfall If the eQuinox battery is not charging the battery symbol will be displayed solid AIR OUTLET VENTS TO BECOME BLOCKED THIS CAN CAUSE THE OXYGEN CONCENTRATOR TO OVERHEAT AND IMPAIR PERFORMANCE T WARNING DO NOT ALLOW EITHER THE AIR INTAKE OR THE WARNING IF THE VEHICLE S DC POWER SOURCE DROPS BELOW 11 5 VOLTS THE EQUINOX WILL REVERT TO THE BATTERY OPERATION IF THE BATTERY IS PRESENT AND CHARGED THE EXTERNAL POWER INDICATOR MAY REMAIN ON eQuinox User Manual 11 eQuinox Rechargeable Power Cartridge Battery The eQuinox can be powered by the rechargeable Power Cartridge Battery which is supplied with the device Installing the Battery Align the Battery so that it will slide into the empty Power Compartment Push the Battery into the empty Power Compartment until it locks clicks into position and is flush with the front of the device When the Battery is properly installed and the eQuinox is ON the Battery Status Gauge will appear on the Control Panel If you DO NOT have a Battery installed the Power Cartridge Status Gauge will not
16. User Manual eQuinox eQuinox A CAIRE A Chart Industries Company AirSep SeQual eQuinox User Controls System Status Indicators Definition Definition FAA Approved Symbol The U S Federal Aviation Administration FAA has approved this device for use on board commercial aircraft Read user manual before operation See user manual for instructions No Smoking Icon Do not smoke near unit Flow Setting Indicator Warnings ALERT Yellow Indicator ALARM Red Indicator External Power Indicator Power Cartridge battery Status Gauge ae Power Cartridge Symbol Pulse Mode Operation Device operating normally power button Increase Flow Setting Decrease Flow Setting No open or naked flames Alternating Current No serviceable parts inside Do not open cover iid us dd Type B Applied Part degree of protection against GR Certified for both the U S and Canadian markets to electric shock 00975 applicable U S and Canadian standards This device complies with the requirements of Directive 93 42 EEC concerning medical devices It therefore bears the CE marking as shown Drip Proof Equipment IPX1 The eQuinox Power Cartridge Battery AC and DC Power Supplies provide protection against the harmful effects of the ingress of liquids IPX1 per IEC 60529 poeeea oons Na Recycle Symbol This symbol is to remind the equipment owners to return it to a recycling facility a
17. VAL OF THE BATTERY FROM THE DEVICE FOR ALL INTERNATIONAL AIRLINE TRAV EL WHEN THE OXYGEN CONCENTRATOR IS CHECKED AS LUGGAGE WHEN SHIPPING THE OXYGEN CONCENTRATOR THE BATTERY MUST ALSO BE REMOVED FROM THE DEVICE AND PACKAGED PROPERLY WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS AND INSTRUCTIONS CONTACT YOUR EQUIPMENT PROVIDER BEFORE ATTEMPTING TO USE THIS EQUIPMENT OTHERWISE INJURY OR DAMAGE MAY RESULT eQuinox User Manual 7 eQuinox ING A MEDICAL EMERGENCY SEEK MEDICAL ASSISTANCE IMMEDIATELY WARNING OPERATING THE OXYGEN CONCENTRATOR OUTSIDE OF THE OPERATIONAL TEMPERATURE SPECIFICATIONS CAN LIMIT THE CONCENTRATOR S ABILITY TO MEET OXYGEN CONCENTRATION SPEC IFICATION REFER TO THE SPECIFICATION SECTION OF THIS MANUAL FOR TEMPERATURE LIMITS WARNING USE NO OIL GREASE OR PETROLEUM BASED OR OTHER FLAMMABLE PRODUCTS WITH THE OXYGEN CARRYING ACCESSORIES OR THE OXYGEN CONCENTRATOR ONLY WATER BASED OXYGEN COMPATIBLE LOTIONS OR SALVES SHOULD BE USED OXYGEN ACCEL ERATES THE COMBUSTION OF FLAMMABLE SUBSTANCES WARNING THE INCORRECT USE OF THE BATTERY CAN CAUSE THE BATTERY TO GET HOT IGNITE AND MAY CAUSE SERIOUS INJURY BE SURE NOT TO PIERCE STRIKE STEP ON DROP THE BATTERY OR OTH ERWISE SUBJECT THE BATTERY TO STRONG IMPACTS OR SHOCKS THE USE OF A DAMAGED BATTERY MAY CAUSE PERSONAL INJURY WARNING WHILE USING THE PORTABLE OX
18. WHERE VENTILATION CAN BE LIMITED THIS CAN CAUSE THE OXYGEN CONCENTRATOR TO OVER HEAT AND IMPAIR PERFORMANCE WARNING THE CONCENTRATOR SHOULD BE LOCATED AS TO AVOID SMOKE POLLUTANTS OR FUMES WARNING THE USE OF SOME OXYGEN ADMINISTRATING ACCESSO RIES NOT SPECIFIED FOR THIS OXYGEN CONCENTRATOR MAY IMPAIR ITS PERFORMANCE RECOMMENDED ACCESSORIES ARE REFERENCED WITHIN THIS MANUAL WARNING IF THE OXYGEN CONCENTRATOR HAS BEEN DROPPED DAMAGED OR EXPOSED TO WATER PLEASE CONTACT YOUR HOME CARE PROVIDER FOR INSPECTION OR POSSIBLE REPAIR OF THE DEVICE DO NOT USE THE OXYGEN CONCENTRATOR IF IT HAS A DAM AGED POWER CORD OR PLUG WARNING PROPERLY SECURED BELT OR OTHERWISE RESTRAIN THE OXYGEN CONCENTRATOR WHEN IN A VEHICLE DURING TRANSPORT TO PREVENT DAMAGE OR INJURY WARNING DO NOT ALLOW EITHER THE AIR INTAKE OR THE AIR OUTLET VENTS TO BECOME BLOCKED DO NOT DROP OR INSERT ANY OBJECTS INTO ANY OPENINGS ON THE DEVICE THIS CAN CAUSE THE OXYGEN CONCENTRATOR TO OVERHEAT AND IMPAIR PERFORMANCE WARNING DO NOT OVERFILL THE OPTIONAL HUMIDIFIER FILL THE OPTIONAL HUMIDIFIER WITH WATER ONLY TO THE LEVEL SHOWN BY THE MANUFACTURER OF THE HUMIDIFIER WARNING DO NOT OPERATE THE EQUINOX AND AMBULATE WHILE A HUMIDIFIER IS ATTACHED REMOVE THE HUMIDIFIER BOTTLE BEFORE WALKING DO NOT LAY THE EQUINOX DOWN WHILE ATTACHED TO A HUMIDIFIER BOTTLE WARNING THE US DEPARTMENT OF TRANSPORTATION DOT AND UNITED NATIONS UN REGULATIONS REQUIRE THE REMO
19. YGEN CONCENTRATOR OUTDOORS WITH THE AC POWER SUPPLY CONNECT THE POWER SUP PLY INTO A GROUND FAULT INTERRUPTED GFI OUTLET ONLY WARNING THE OXYGEN CONCENTRATOR SHOULD NOT BE USED AD JACENT TO OR STACKED WITH OTHER EQUIPMENT IF ADJACENT OR STACKED USE IS UNAVOIDABLE THE DEVICE SHOULD BE OBSERVED TO VERIFY NORMAL OPERATION CAUTION In the event of an alarm or you observe the Oxygen Concentrator is not working properly consult the Troubleshooting section in this manual If you cannot resolve the problem consult your Equipment Provider WARNING IF YOU FEEL DISCOMFORT ARE EXPERIENC CAUTION To prevent a void warranty follow all manufacturers instructions THAN THE POSSIBLE SOLUTIONS LISTED WITHIN THIS MANUAL DO NOT REMOVE COVERS ONLY YOUR EQUIP MENT PROVIDER OR A QUALIFIED SERVICE TECHNICIAN SHOULD REMOVE THE COVERS OR SERVICE THIS DEVICE WARNING DO NOT ATTEMPT ANY MAINTENANCE OTHER WARNING ONLY ACCESSORIES RECOMMENDED BY THE MANUFAC TURER USE OF ANY OTHER MAY BE HAZARDOUS CAUSE SERIOUS DAMAGE TO YOUR OXYGEN CONCENTRATOR AND WILL VOID THE WARRANTY NATE SOURCE OF SUPPLEMENTAL OXYGEN IN THE EVENT OF A POWER OUTAGE ALARM CONDITION OR MECHANI CAL FAILURE CONSULT YOUR PHYSICIAN OR EQUIPMENT PROVIDER FOR THE TYPE OF RESERVE SYSTEM REQUIRED WARNING DO NOT ALLOW EITHER THE AIR INTAKE OR THE AIR OUTLET VENTS TO BECOME BLOCKED DO NOT DROP OR INSERT ANY OBJECTS INTO ANY OPENINGS ON THE DEVICE TH
20. ail Filter able from your Equipment Provider CAUTION Do not operate the Oxygen Concentrator with out the air intake filter in place If a second filter is pro vided insert the replacement filter before you clean the dirty filter Clean the dirty filter in a warm soap and water solution then dry thoroughly prior to use Note The Manufacturer does not recommend the sterilization of this equipment Note Do not operate the Oxygen Concentrator without the air intake filter in place If a second filter is provided insert the replacement filter before you clean the dirty filter Clean the dirty filter in a warm soap and water solution then dry thoroughly prior to use The air inlet filter should be replace annually If the eQuinox 1 used in a dusty environment the filter may need to be replaced more frequently You should not operate the eQuinox without an air inlet filter installed for more than 5 minutes eQuinox User Manual 17 eQuinox AC Power Supply 12VDC Cable Power Cord and Outer Case Turn OFF the eQuinox and disconnect from AC or DC power before any cleaning or disinfection activity DO NOT spray the outer case directly Use a damp not wet cloth or sponge Spray the cloth or sponge with a mild detergent solution to clean the cabinet and power supplies To disinfect the eQuinox use Lysol Brand II disinfectant or equivalent Proceed as directed by the cleaner manufacturer WARNING ELECTRICAL SHOC
21. ator to overheat and impair performance eQuinox User Manual 9 eQuinox Operating Instructions Before Operating This users manual serves as your reference to help you operate and maintain the device If you have any questions or concerns please call your home care provider Important DO NOT attempt to operate the eQuinox without first reading the Safety Guidelines section of this manual Please follow all of the operating instructions Please observe all Warnings on the device and in the Users Manual In order to reduce the risk of fire personal injury and serious damage to the eQuinox please observe all of the safety precautions WARNING CLEAN THE CABINET CONTROL PANEL AND POWER CORD ONLY WITH A MILD HOUSEHOLD CLEANER APPLIED WITH A DAMP NOT WET CLOTH OR SPONGE AND THEN WIPE ALL SURFACES DRY DO NOT ALLOW ANY LIQUID TO GET INSIDE THE DEVICE PAY SPECIAL ATTENTION TO THE OXYGEN OUTLET FOR THE CANNULA CONNECTION TO MAKE SURE IT RE MAINS FREE OF DUST WATER AND PARTICLES CAUTION Do not allow either the air intake or the air outlet vents to become blocked This can cause the Oxygen Concentrator to overheat and impair performance Do not operate the Oxygen Concentrator without the air intake filter in place If a second filter is provided insert the replacement filter before you clean the dirty filter Clean the dirty filter in a warm soap and water solution then dry thoroughly prior to use Powering on
22. ease see the Accessories Catalog PN ML LOX0010 available on www CAIREmedical com Step 5 Adjust the Flow Setting to the Prescribed A Level Using the Flow Setting Buttons adjust flow setting to the prescribed setting supplied by your physician PRESCRIBED LEVEL OF OXYGEN DO NOT CHANGE THE T WARNING IT IS VERY IMPORTANT TO SELECT ONLY THE FLOW SELECTION UNLESS YOU HAVE BEEN DIRECTED TO DO SO BY A LICENSED CLINICIAN eQuinox Step 6 Begin Using Your eQuinox Breathe normally through the nasal cannula Step 7 eQuinox Cart Operation The eQuinox Cart is designed for use on pedestrian type surfaces Align the locater holes on the bottom of the device to the tabs on the Universal Cart Align the large threaded screw to the insert at the back of the eQuinox Then hand tighten the knob on the cart until the Universal Cart is tightly secured to the SeQual eQuinox Concentrator eee _ _ ade Depress the push button on the Universal _ Cart handle to adjust the height of the handle Step 8 Power Off the Device Press and hold the ON OFF Button for two 2 seconds to power OFF your eQuinox Store the eQuinox in a cool and dry location for a period no longer than 1 month The eQuinox must run for a minimum of 2 hours per month to ensure proper function of the unit Note DO NOT operate the eQuinox and ambulate while a humidifier is attached Remove the humidifier bottle befo
23. er 15 to discontinue use of the eQuinox and use the aircraft supplemental oxygen The eQuinox unit is to be turned off after securing the aircraft supplemental oxygen Prior to travel the user needs to inspect the eQuinox to ensure it is in good operational condition CAUTION The US Department of Transportation DOT and United Nations UN Regulations require the removal of the battery from the device for all international airline travel when the oxygen concentrator is checked as luggage When shipping the oxygen concentrator the battery must also be removed from the device and packaged properly eQuinox Power Cartridge 2 Airline Travel Table ar These tables calculate flight times only Please allow for ground time preflight check in security check in etc and layovers Most airlines require that you have an adequate number of fully charged batteries to power the device for at least 150 of the expected maximum trip duration flight times all ground time before and after flight and during connections and for unexpected delays It is suggested to always carry an extra battery with you on trips Each airline has their own requirements for traveling with oxygen Contact your airline at least 48 hours prior to departure to inform them you will be traveling with the eQuinox Continuous Flow FLIGHT LENGTH in Hrs 20 40 50 6 0 70 8 ESTIMATED OF POWER C
24. er will keep your eQuinox performing Preventative Maintenance PM to be completed on an annual properly Check this air inlet filter daily and clean it at least once or yearly basis Consult your provider to arrange PM schedule a week using the following procedure Your home care provider or qualified technician will perform inspections of the alarms internal 9 volt battery battery internal filters and other internal parts only as needed 2 Wash the filter in warm tap water using a mild soap detergent solution 1 Remove the air inlet filter Note If three beeps sound after the power ON self test this indicates the Qv battery is low in power Contact your home care provider if you hear 3 Rinse the filter thoroughly with clear tap water and squeeze three beeps out the excess water 4 Allow the filter to air dry User Care and Cleaning of the Device Cannula Replacement 5 Reinsert the filter in the cabinet Replace your supply tubing and cannula on a regular basis as Cleaning tip Keep a clean second filter as a replacement for use recommended by your home care provider Your physician or when one filter is drying home care provider will provide you with cleaning disinfection Ee RN and replacement information Note Always follow the cannula manufacturer s instructions for proper use Replace the disposable cannula as recommended by the cannula Air inlet manufacturer or your Equipment Provider Additional supplies are av
25. is between 2 and 5 hours depending upon the device flow setting When the battery is fully charged the battery gauge will cease cascading The battery can be charged in the unit when the eQuinox is turned off and plugged into the portable oxygen concentrator If the Battery becomes too warm during discharging recharging will not begin until the Battery sufficiently cools Removing the Battery and allowing it to cool may expedite this cooling process The Battery will not recharge when operating from the 12VDC Cable The Battery charges when in the device is operating on AC power only The Battery will also charge from the optional Desktop Charger eQuinox User Manual 13 eQuinox First Time eQuinox Operation Step 1 Positioning Your eQuinox for Use Place the eQuinox in a well ventilated well lit area Be sure the air inlet and exhaust vents are not obstructed Air inlet Filter Position the eQuinox so that all audible and visual indicators or alarms can be easily seen and heard Be sure the air inlet filter is in place before operating your eQuinox If the air inlet filter is missing contact your home care provider If the air inlet filter is dirty wash with warm soapy water rinse with clear water and allow it to dry before placing the air inlet filter back in the device If necessary replace it with a new clean air inlet filter Plug the device into a grounded Exhaust Vent AC Power outlet a DC Power sou
26. l not be illuminated or will flash on and off External Power Indicator When the eQuinox is properly plugged in and is using the AC or 12VDC Cable this indicator will appear on the User Control Panel Buzzer An audible alarm or buzzer alerts you to the operating condition of the device either a warning or failure and confirms a valid key press eQuinox Powering on the eQuinox To connect to AC power source Insert the AC Power Supply cord securely into the External Power Receptacle on the right side of the device Plug the AC power cord into a grounded AC outlet When the eQuinox is plugged in properly a green indicator on the AC Power Supply will light and the External Power Indicator will appear on the user control panel 497 Install the Battery Slide the battery into the empty compartment until it locks into position and is flush with the front of the device When the battery is properly installed and the eQuinox is ON the Power Cartridge Status Gauge B will appear ON OFF Button Green Indicator Press and hold the ON OFF button for two 2 seconds to power ON your eQuinox A brief audible and visual power on self test will occur Verify that all the indicators are illuminated and the buzzer sounds for three 3 seconds Press and hold the ON OFF button for two 2 seconds to power OFF your eQuinox eQuinox User Manual 5 eQuinox Select Flow Delivery Mode Flow Mode Button
27. ox to ensure adequate air flow Exceeding maximum breath Switch to Continuous Flow mode until respiratory rate reaches an acceptable rate rate in Pulse Dose mode See Pulse Dose mode Breath Rate Chart Contact your dealer provider Restriction in humidifier or Repair or replace humidifier or tubing tubing Low Oxygen Fow Filter Blocked Clean or replace Air Inlet Filter Contact your dealer provider No Oxygen Delivered in Pulse Tubing cannula longer than 7 Attach 7 foot 2 1m tubing cannula Flow Mode feet 2 1m Humidifier attached Remove humidifier No inspiration detected Contact your dealer provider 1 in Continuous Flow mode Switch to Pulse Dose mode Nasal cannula not in the nose Place nasal cannula in nostrils and breathe through nose or mouth breathing Power Cartridge Status Gauge Power Cartridge battery is Contact your dealer provider to replace Power cartridge battery never indicates fully charged aging Concentration 16 4805 C3_01A eQuinox Cleaning Care 2 Routine Maintenance Routine Maintenance Air Inlet Filter Servicing of the internal components inside the cabinet of the Ambient air 1 drawn into the device through the air inlet located eQuinox must be conducted by a CAIRE trained and qualified on the top rear of the device Cleaning the air inlet filter is the service technician as needed most important maintenance activity you will perform as a user Cleaning the air inlet filt
28. perature Operating 10 95 at an 82 4 F 28 C dew point Humidity Transport 4 F to 140 F 20 C to 60 C Storage H Non Temperature umidity Up to 95 Non Condensing Electrical Use no extension cords Use no electrical outlets con trolled by a switch Altitude 0 13 123 feet 0 4 000 meters Placement NOT block the air inlet top back cover or the exhaust vent back bottom left Place device a minimum of 3 inch es 7 5 cm away from walls draperies furniture etc Environment Must be smoke pollutant and fume free 24 hours a day when connected to an external AC or DC power source Operating Time Note If the Oxygen Concentrator has been stored for an extended period of time outside its normal operating temperature range the unit should be allowed to return to normal operating temperature before being turned on Refer to the Specifications section in this manual eQuinox Proper Placement of the eQuinox Select a location for the device that avoids the intake of smoke fumes and pollutants Correct placement of the device should allow intake of air through the air inlet filter at the top rear of the cabinet and allow exhaust air to freely leave the exhaust vent at the bottom right of the device Place the device such that the alarms can be heard Position the oxygen supply tubing in such a way that it does not kink or occlude THE PRESENCE OF FLAMMABLE GASES THIS CAN RESULT
29. rce or be sure there is a fully charged Battery installed Step 2 Power ON the Device and Allow it to Ox Warm Up Press and hold the ON OFF Button for one 1 second to power ON your eQuinox A brief visual power on self test will occur Note If there are three beeps at start up this indicates a low battery condition for the 9 volt battery inside the cabinet of the eQuinox Please contact your home care provider and request that the 9 volt battery be replaced The eQuinox has an oxygen concentration status indicator OCSI built in to the device The OCSI continually monitors the oxygen output of the device and incorporates a green light normal yellow light alert red light alarm and a buzzer These indicators will all illuminate during the power on self test at start up After initially powering ON the device please allow up to five 5 minutes for the device to reach its performance specifications When only the green indicator light is illuminated the device has reached its performance specifications 14 4805 C3_01A Step 3 Connect the Oxygen Supply Tubing or Nasal Cannula to the Oxygen Outlet Replace the cannula and oxygen supply tubing regularly as recommended by your home care provider Consult your home care provider for cannula and supply tubing cleaning and replacement information Note Ensure the cannula is fully inserted and secure This ensures that the Oxygen Concentrator can properly detec
30. re walking DO NOT lay the eQuinox down while attached to a humidifier bottle eQuinox User Manual 15 eQuinox User Troubleshooting I able Do Not Ignore Alarms AGE ALARM CONDITION OR MECHANICAL FAILURE CONSULT YOUR PHYSICIAN OR EQUIPMENT PROVIDER FOR THE TYPE OF RESERVE SYSTEM REQUIRED Symptom Possible Cause No battery installed Install battery or plug into external power Battery is discharged or no Plug into external power external power is present eQuinox does not power AC Power Supply or 12v Check connections at outlet Power Supply box and at the eQuinox to ensure secure on when ON OFF button is DC cable is not plugged in connections pressed correctly WARNING THE MANUFACTURER RECOMMENDS AN ALTERNATE SOURCE OF SUPPLEMENTAL OXYGEN IN THE EVENT OF A POWER OUT Insufficient power from the Ensure there is adequate power from the outlet Outlet may be controlled by a wall outlet switch or the circuit breaker may have engaged Contact your dealer provider Filter blocked Clean Air Inlet Filter Check humidifier attachment and tubing No Oxygen Power the SeQual eQuinox ON Tubing or cannula is kinked or Check tubing cannula and connections not properly connected Contact your dealer provider Restriction in tubing or Repair or replace tubing or humidifier Position your eQuinox to ensure adequate air humidifier flow Low Oxygen Filter restricted Clean or replace Air Inlet Filter Position your eQuin
31. resent or fully discharged during an AC power interruption the eQuinox will shut down and alarm The Battery is fully discharged when the Power Cartridge Status Gauge on the control panel is no longer illuminated The device will display a yellow light and sound a low Battery alert beep until the device completely shuts down with a red light and constant alarm This audible alert may continue for up to 5 minutes until the device shuts down Note Press and hold the eQuinox ON OFF button for a minimum of 5 seconds to silence the Loss of Power alarm Connecting the eQuinox to an external AC power supply for 2 minutes will also eliminate this alarm Typical New Power Cartridge Operating Times Operating Type Battery Type 12 cell 24 cell Continous Mode 1 LPM 1 96 hrs 4 4 hrs 2 LPM 1 1 hrs 2 5 hrs 3 LPM 0 73 44 min 1 67 hrs Pulse Mode 12 BPM Bolus 16 mL 2 38 hrs 5 4 hrs Bolus 48 mL 2 16 hrs 4 9 hrs Bolus 96 mL 1 54 hrs 3 5 hrs Note Operating the Oxygen Concentrator outside of its normal operating temperature range can affect performance and decrease battery run time and or increase battery charge time Refer to the Specifications section in this manual You will achieve longer operating time on the Battery if you operate your eQuinox in the Pulse Mode Physician approved mode eQuinox Typical Battery Recharge Time The typical time to recharge your Battery to achieve 80 capacity from a fully discharged Battery
32. rts to measured bolus delivery To exit Pulse Dose Mode and engage Continuous Flow Mode push the flow mode button Pulse Dose Mode and autoSat Features Your eQuinox has a feature called autoSAT that delivers a consistent sized bolus of oxygen up to 40 breaths per minute see Max Breath Rate table below As your breath rate increases 1e during ambulation the autoSAT feature servo controls the device delivering the set bolus size autoSAT Technology is proprietary to CAIRE and is utilized during Pulse Mode only Mas Breath Rate Bolus Size Pulse Dose Setting Max Breath Rate 15 mL 1 0 16 40 _ 2 0 32 40 5 3 0 48 40 LA E 4 0 64 40 5 0 80 37 6 0 96 31 128 23 160 18 3 192 15 o ca Note Bolus volume decreases as breath rate exceeds published range Note The pulse dose setting bolus is not equal to LPM MINED FOR EACH PATIENT INDIVIDUALLY FOR THEIR NEEDS AT REST DURING EXERCISE AND WHEN TRAVELING WARNING THE EQUINOX MAY NOT BE ABLE TO DETECT ALL IN SPIRATORY EFFORTS IN PULSE MODE IF THE EQUINOX DOES NOT RESPOND TO YOUR INSPIRATION EFFORTS ASK YOUR HOME CARE PROVIDER TO CHECK THE SENSITIVITY SETTING WARNING PULSE DOSE MODE SETTINGS MUST BE DETER Note To Equipment Provider The following oxygen administration acces sories are recommended for use with the Oxygen Concentrator e Nasal Cannula with 7 feet 2 1 m of tubing PN 5408 SEQ e For any additional recommended accessories pl
33. t inspiration for oxygen delivery During inhalation you should hear or feel oxygen flow to the prongs of the nasal cannula Step 4 Select the Flow Mode Press the Flow Mode button to select the desired mode Pressing this button repeatedly will toggle you back and forth between Pulse Dose Mode and Continuous Flow Mode The proper mode and proper setting should be prescribed for you by your physician Continuous Flow Mode Operation When operating in the Continuous Flow Mode a continuous supply of oxygen measured in liters per minute LPM will flow through your tubing and nasal cannula You may use a humidifier when in Continuous Flow Portable Concentrator Mode Pulse Dose Mode Operation When operating in Pulse Mode a bolus of oxygen measured in milliliters mL is delivered upon inspiration Remove the humidifier bottle before walking DO NOT use a humidifier in the Pulse Dose Mode DO NOT lay the eQuinox down while attached to a humidifier bottle To ensure proper saturations secure a prescription for pulse dose from your physician with consideration to your needs at rest during exercise and when traveling at altitude Contact your care provider for prescription and boarding forms When operating in pulse dose mode the eQuinox supplies a continuous flow of oxygen if a breath is not detected The device will continually check for an inspiratory effort every 15 seconds Once an inspiration is detected the device reve
34. t the end of its life per Waste Electrical and Electronic Equipment WEEE Directive Drip Proof Equipment IPX0 The eQuinox enclosure does not provide protection against the harmful effects of the ingress of liquids IPX0 per IEC 60529 Authorized representative in the European Community Name and address of manufacturer U off 2 4805 C3 01A eQuinox Contents User Controls amp System Status Indicators 2 Quick Start cll NUM 4 il 7 Contraindications for USe 7 VLIPIIVACHCIIIDD reete 7 Recommended Operating Environments 9 Instructions mM 10 First Time OCR AUN OWN mec M M 14 User Troubleshooting 16 Cleaning Care amp Routine 17 General Airline Travel Information 19 Power Cartridge amp Airline Travel Table
35. uinox from a 12VDC Cable it is recommend ed that the Pulse Dose Mode be used if prescribed to use a pulse dose device Note To ensure that the device is utilizing the DC power check that the External Power Indicator is illuminated If a beep is heard this could indi cate a loss of external power Check if the green Battery Power Verification LED is on or off If the battery is removed on DC power it will ensure no loss of charge while the unit is running If the indicator is not illuminated the battery is powering the device A 12VDC Cable allows the system to operate from DC outlets such as those found in motor vehicles Settings Available When Operating on DC for eQuinox 1 Start your vehicle Flow Setting Continuous Bolus Size mL Pulse 2 Insert the 12 Volt DC cord into the recessed power 0 5 Yes receptacle on the side of the device 1 0 Yes 16 Yes 1 5 Yes 24 Yes 3 Insert the 12VDC Cable plug into the DC power outlet in 2 0 Yes 32 Yes your motor vehicle 25 Yes 40 Yes m l Y 4 Y When the device is properly connected and receiving power from 28 x the DC power source a green indicator light on the Power Supply 64 vac and the External Power Indicator 47 on the control panel will T Y illuminate 80 88 Yes SeQual eQuinox DC Supply Connection 96 Y es When the eQuinox is connected to the 12VDC Cable in a vehicle the following applies 192 Yes The eQuinox will run at 3 0LP
36. uminates and a pulse of oxygen is delivered with each inspiratory effort ALERT Yellow Indicator Low and Medium Priority Alerts When illuminated this indicates a low priority awareness condition or Caution Continue to use your system and refer to the Troubleshooting Table for the proper response or contact your Home Care Provider A flashing yellow indicates a medium priority alert prompt response is necessary Typically low oxygen or low battery ALARM Red Indicator High Priority Alarms Indicates a high priority alarm condition An immediate response is necessary Refer to the Troubleshooting Table and contact your home care provider Typically loss of power low oxygen or malfunction Information Display e Cor P indicates Continuous Flow mode and P indicates Pulse Dose mode e Time on battery indicates estimated time remaining on battery e Setting continuous flow or pulse dose each time you power the device ON the previous mode and or setting has been saved and will be used at start up Power Cartridge battery Status Gauge This indicator displays the charge remaining in the battery Each of the five horizontal gray bars represents approximately 20 of the total battery charge When the battery is being charged the charge indicator bars will blink in a waterfall type fashion If the battery is not installed or if itis improperly installed the Power Cartridge battery Status Gauge wil
37. ur home care provider Breathe normally through the nasal cannula When operating on battery power utilizing the Pulse Dose Mode will achieve longer duration of use Universal Cart Operation Align the locater holes on the bottom of the device to the tabs on the Universal Cart Align the large threaded screw to the insert at the back of the eQuinox Then hand tighten the knob on the cart until the Universal Cart is tightly secured to the device Depress the push button on the telescoping Universal Cart handle to adjust the height of the handle Important Safety Instructions are defined as follows ARDS THAT MIGHT CAUSE SERIOUS INJURY WARNING IMPORTANT SAFETY INFORMATION FOR HAZ CAUTION Important information for preventing damage to the eQuinox Note Information needing special attention Indications for Use PRESCRIBED LEVEL OF OXYGEN DO NOT CHANGE THE FLOW SELECTION UNLESS YOU HAVE BEEN DIRECTED TO DO SO BY A LICENSED CLINICIAN THE PORTABLE OXYGEN CONCENTRA TOR MAY BE USED DURING SLEEP UNDER THE RECOMMENDATION OF A QUALIFIED CLINICIAN WARNING FEDERAL USA LAW RESTRICTS THIS DEVICE TO SALE OR RENTAL BY ORDER OF A PHYSICIAN OR OTHER LICENSED HEALTH CARE PROVIDER WARNING THIS UNIT IS NOT TO BE USED FOR LIFE SUPPORT GERIAT RIC PEDIATRIC OR ANY OTHER PATIENT UNABLE TO COMMUNICATE DISCOMFORT WHILE USING THIS DEVICE MAY REQUIRE ADDITIONAL MONITORING PATIENTS WITH HEARING AND OR SIGHT IMPAIR MENT