Home

SIARETRON 3000 ICU

image

Contents

1. 95mm e About 1 Kg Display area Measures parameters Messages Graphs Zoom Area 2 areas Functioning modes Measured parameters Messages Available graphs Zoom Area Pmax Pmean Poause PEEP VTi VTe ExpMV Rate FLOWi FLOWe I E Tinsp Tesp Tpause Cs Ri Operation mode Interruptions Functioning time Alarm Ventilation graphs Loop Trend Parameter zoom page 2 areas for ventilation parameter selection Function modes Zoom Area 2 areas Alarm settings default Loop Trend Help Monitor Alarm volume Demo display Configuration and gas type selection Siaretron 3000ICU_te_rev1 doc 4 6 siae Zoom Area Possibility to select and display any measured parameter Alarms Manual or automatic setting of minimum and maximum alarm limits Sound volume adjusting and visual flashing symbol on message display area alarm warning Available alarms Pmax Pmean Ppause PEEP VTi VTe ExpMV Rate Cs Ri Ventilation curves PAW Flow Volume Loop Volume Paw Volume Flow Trend Selection of trend range and period Available trends Pmax Pmean Ppause PEEP VTi VTe ExpMV Rate FLOWi Help Display of operation manual of the equipment optional Monitor Selection of the following display conditions LCD Remote 0 LCD Remote 1 Display of parameter zoom page Parameter Zoom Page Displayed parameters PEEP Flo
2. SIARETRON 3000 ICU Intensive care ventilator code 960136 AA S Jia rev 1 01 01 2009 ND S TARAP val lt MAIN CHARACTERISTICS Destination Resuscitation intensive care and emergency It can be equipped with TFT 6 5 colour spirometer monitor with curves and loops It is suitable for adults children and newborns weighing more than 3 5 Kg Dimensions W x H x D 360 x 245 x 300 mm without monitor 360 x 330 x 300 mm with monitor Weight 6 9 Kg 7 8 Kg with monitor Relative Humidity use 30 95 RH OPERATION DATA Operation principle e Time cycled at constant volume e Pressure cycled e Microprocessor controlled flow e Spontaneous breath with integrated valve Ventilation modalities IPPV IPPV AST SIMV CPAP PSV PCV TWINPAP BILEVEL PAP SIGH NEB APRV Apnoea BACK UP NIV Breathing rate From 5 to 150 bpm Tidal volume BTPS Adult from 100 to 3000 ml Paediatric Neonatal from 20 to 300 ml 1 E ratio From 1 4 to 3 1 Siaretron 3000ICU_te_rev1 doc 1 6 siere Inspiratory time From 20 to 75 Min 0 08sec Max 9 sec Inspiratory pause From 0 to 50 of inspiratory time Inspiratory pressure limit From 2 to 80 cmH20 PEEP From 0 to 20 cmH20 TWIN PAP BILEVEL PAP e Pressure low level 0 25 cmH2O0 e Time low level from 0 1 to 30 secs e Pressure high level 5 50 cmH2O e Time high level from 0 1 to 60 secs M
3. inute volume with 1 2 Max 40 l min ratio Accuracy on Minute 20 of real reading above 100ml Volume measurement 20ml of real reading below 100ml O2 concentration Adjustable from 21 to 99 with electronic integrated mixer Trigger Adjustable from off 1 to 9 cmH2O under set PEEP level Max inspiratory flow Adults from 6 to 120 l min Paediatric and Neonatal from 6 to 30 l min SIGH 1 breath every 100 with automatically changed alarm limit SIGH volume 150 of set Tidal Volume Supply pressure O2 Air pressure included between 280 kPa and 600 kPa 2 8 6 bar Max flow required by the ventilator 100 l min Patient circuit Double hose non rebreathing Other controls Button 100 O2 x 5 min INSP Hold and EXP Hold max 20 secs at a time Nebulizer Expandability Software upgradeable for future modalities Dead space compensation Automatic compensation of mechanical and patient circuit dead space Flow Sensor By magnetic disturbance st rilisable for several uses Atmospheric compensation Automatic Siaretron 3000ICU_te_rev1 doc 2 6 siae GAS CONSUMPTION Control About 2 l min Minute volume Same as set minute volume MONITORING Measured parameters PAW peak PEEP O2 Concentration Oxymeter Electronic with automatic calibration at start up ALARMS System alarms Test of electric and pneumatic devices Gas feeding Low l
4. t 2 7 bar Pressure Paw High Low limit Apnoea Alarm occurs in case of insufficient ventilation function of Apnoea BACK UP Activation time 25 sec FiO High Low limit Electric power supply Alarm occurs in case of failure of main power supply line Low battery Alarm occurs in case the battery voltage is lower than 11 Vdc POWER SUPPLY Electric power supply 100 230 Vac 50 60Hz 12 V de with external auxiliary battery Power consumption max Internal power supply 20 Watt 12Vdc battery 1 3 Ah Internal battery operation Max 1 5 hours Re charging time empty to full About 8 hours NORMS EN 1281 1 IEC 601 1 IEC 601 1 2 Dir ve 93 42 EEC IEC 601 1 4 EN 794 1 Class and type according to IEC 601 1 Class 1 Type B Class according to 93 42 EEC Directive Class Ilb Siaretron 3000ICU_te_rev1 doc 3 6 ACCESSORIES siae Supplied Accessories e User s Manual e O2 supply hose e AIR supply hose e O02 cell e Nebulizer set e Silicone patient circuit for adults e Antibacterial filter e Power cable SHUKO VDE Optional Accessories See on Export Price List TFT 6 5 colour spirometry monitor with curves and loops optional Code 713005 Operator Interface Key buttons function selection ESC enter alarm silencing and Encoder Technical Data e Box in white ABS RAL 9003 e 235x150x
5. w Insp Minute Vol I E Pmax Flow Exp Tidal Volume Rate Displayed curves PAW flow volume Configuration and gas type selection Measure gas Stop gas Set agent Set zero Set O2 Set Gas alarm Set CO2 unit CO2 N20 AG O2 Volume Adjustable from 1 to 3 Demo option Available Language option Available Flow sensor Available automatic setting of the OFF SET and GAIN parameters Power supply 12 Vdc through RS 232 port Power consumption 30 VA max External connections 9 poles connector RS 232 Siaretron 3000ICU_te_rev1 doc 5 6 siae a CAPNOMETRY optional G ns oe Capnometry connection Available O2 COs Description Ultra small infrared main stream multi gas probe comprising a multi channel IR bench barometric pressure sensor power regulator signal processor and a RS 232 interface Measuring mode Mainstream Dimensions and weight 37x 27 x 25 mm lt 30 g cable not included General specifications See on relative Multigas Analyzer Modules for ICU technical data sheet SIARE applies the UNI EN ISO 13485 2004 Quality System and the 93 42 EEC SIARE ENGINEERING INTERNATIONAL GROUP s r l Via Pastore 18 40056 Crespellano Bologna ITALY Tel 39 051 969802 Fax 39 051 969366 Email mail siare it http www siare it Siaretron 3000ICU_te_rev1 doc 6 6