1. Light too bright Adjust light intensity control to a lower position Objective lens dirty Clean objective lens Eyepiece lens dirty Clean eyepiece lens Slide upside down Poor Resolution Image not sharp Turn specimen slide over cover slip facing up Cover slip on specimen slide too Use 0 17mm thick cover slip thick Too much light Adjust light intensity control to a lower position Condenser lens dirty Clean condenser lens Rack stop not set at a proper Adjust rack stop position Eyepiece dirty Clean eyepiece lens Spots in field of view Speciinensiidedirty Ciearsiide Condenser lens dirty Clean lens of condenser Nosepiece not properly indexed Revolve nosepiece into positive Uneven illumination of field index stop Diaphragm not properly indexed Adjust diaphragm to proper level MECHANICAL PROBLEM Reason for Problem Solution S Does not stay in focus Stage drops down Adjust tension adjustment knob
2. RICHTER PTICA Richter Optica info richter optica com Instructions for Models U1B U1T U1D UILCD Universty Student Biological Microscope RICHTER Eer ca Trinocular Port Interpupillary Adjustment Scale Eyepiece Diopter Adjustment Ring Head Locking Screw Revolving Nosepiece Objective o l Moveable Finger et Screws to secure specimen holder to stage Specimen Slide Holder Condenser locking Screw Stage Coarse Focus Abbe Condenser X Axis Control Fine Focus Knob Y Axis Control Condenser Control Knob Knob Iris diaphragm lever Swing out Filter Holder Lamp Housing RICHTER Err ca Thank you for your purchase of a Richter Optica microscope The information in this manual is provided to answer most questions that can arise when operating your microscope and to help you avoid unneccesary maintenance expenses in the future Please carefully read instructions before operating microscope Nomenclature used to describe components and controls are identified on opposite page UNPACKING Do not discard styrofoam container or packing materials until you are sure shipment is complete and undamaged retain styrofoam shipping container to store your microscope when it is not in use Remove alltape and packing material used to protect microscope during shipment Make certain lens surfaces do not come in contact with dirt fingerprints or oil Damage of lens surfaces occur when they come in contac
3. ly depress and rotate screwdriver to release the fuse cap Pull cap and fuse out of fuse holder Insert proper lamp electronic fuse into fuse cap Using screwdriver rotate fuse cap assembly in opposite direction of arrow until guide slot engages depress fuse cap and rotate 1 4 turn to lock into fuse TROUBLESHOOTING ELECTRICAL Problem Reason for Problem Solution Light Fails to Operate Outlet inoperative Have qualified service technician repair outlet AC Power cord not connected plug into outlet Replace with new LED bulb Bulb burned out Replace with new 1 amp fuse Fuse burned out Replace with proper 1 amp fuse Fuse burns out too soon Unit has short have qualified service technician repair Fuse blows instantly when replaced electrical short Problem Reason for Problem Solution Light burns out too soon The voltage is too high Adjust intensity control to the minimum position before turning the power switch on Light bulb burns out immediately Incorrect lamp used Have qualified service technician use proper bulb Plug unit into proper outlet 120v or 220v Light Flickers Lamp not properly inserted into Have qualified service technician make socket repairs Lamp about to burn out Fuse holder not locked into proper position Loose connection at AC outlet RICHTER rrica TROUBLESHOOTING IMAGE QUALITY No Image Nosepiece not indexed properly Move revolving nosepiece until objective lens clicks into position
4. position RICHTER Err ca 2 Place standard specimen slide cover slip up on top of stage surface a Swing movable finger on slide holder outward Place specimen slide against fixed side of slide holder Slowly release moveable finger until it makes contact with the specimen slide 3 Rotate coarse focusing controls until the specimen comes into focus 4 Adjust fine focus controls until soecimen is in sharp focus 5 Adjust diopter for difference in eyesight a Using right eye peer into the right eyepiece tube Adjust sharpness of image by turning diopter adjustment located on left eyepiece tube 6 Adjust the aperture opening of the iris diaphragm Iris diaphragm should not be used to control the brightness of illumination use light intensity control knob to adjust light level Iris diaphragms are designed to help achieve high resolution of soecimen and provide contrast in the image Smaller apertures will deliver contrast to the image However closing aperture too much will reduce resolution Experimentation is the best method of determining the correct opening of the diaphragm Some suggested openings for the iris diaphragm are Objective Diaphragm Opening 4x 1 8 open 10x 1 8 to 1 4 open 40x 1 4 to 1 2 open 100x 1 2 to 3 4 open 7 Changing magnification a Rotate revolving nosepiece to postition 10x objective into optical path b This microscope has been parfocalized which allows changes from one objective
5. s using a magnifying glass to ensure that the element is clean If immersion oil or specimen material of any kind is evident use a cotton swab dipped in a small amount of Windex to clean all foreign material from objective lens surface Such material will reduce or totally block image quality e Cleaning condenser lens Clean only the top lens surface visible when looking through hole in top of stage Use same procedure as used for eyepiece or objective lenses f Illuminator condenser lens Use same procedure as used for eyepiece or objective lenses 2 Electrical Maintenance WARNING FOR YOUR SAFETY TURN SWITCH OFF AND REMOVE PLUG FROM POWER SOURCE OUTLET BEFORE MAINTAINING YOUR MICROSCOPE 1 Replacement of Bulb a Carefully pull magnetic lamp housing up and off of the microscope to reveal the bulb b Using a tissue or cloth gently grasp the halogen bulb pull straight out of lamp socket C Your microscope requires an LED bulb available from your microscope dealer d Make certain that the new bulb is clean as fingerprints on the bulb can affect light transmission Grasping bulb gently with a tissue or cloth insert pins straight into the lamp socket e Replace magnetic lamp housing RICHTER Err ca 2 Replacement of Fuse The fuse is located on the back of the microscope on the right side of the power plug To remove the fuse from the holder insert a6mm screwdriver blade into slot located in rear of fuse cap Slight
6. t with such contaminants and image quality is reduced OPERATION A Illumination 1 Before operating microscope adjust intensity control located on side of base to the minimum position This should be done prior to each time the base is turned on or off This will extend bulb life 2 Insert power plug into base of microscope Then insert plug into outlet 3 Push rocker switch at back of the base to on position 4 Rotate intensity dial on illuminator base until the image is illuminated 5 Adjust intensity of light to match requirements of objective and specimen slide B Interpupillary Adjustment of Viewing Head 1 Looking through microscope adjust distance between the two eyepieces by grasping the eyetubes and moving them either closer or farther apart 2 When a full field of view is observed throught both tubes and images blend into one interpupillary distance is corrected for your eyes Check the interpupillary scale and note index reading for future reference in case other users will be changing this adjustment from time to time 3 Adjust the diopter scales located on each eyetube to the same numerical value as indicated on the interpupillary scale This must be done in order to maintain parfocality of objective lenses If interpupillary distance is changed adjust eyepiece diopters accordingly C Focusing the Microscope 1 Position the 4x objective lens into the optical path making sure that lens clicks into its proper
7. to another while requiring only a slight adjustment of the fine focus controls c When changing to the 40x and 100x objective lens care must be exercised in order to prevent damaging the front lens element and specimen slide d In order to obtain the maximum resolution of the 100x oil immersion lens it is necessary to apply immersion oil between the cover glass of slide and the front lens of the objective 1 Use a very small amount of immersion oil Only the tip of the lens should ever come in contact with the oil RICHTER Err ca 1 Optical Maintenance a Do not attempt to disassemble any lens component Consult an expert technical service company when repairs not covered by these instructions are required b Prior to cleaning any lens surface brush dust or dirt off lens surfaces using a camel hair brush Or use air to blow dust and lint off surfaces Compressed air available at any computer supply store is a good source of clean air C Cleaning eyepiece lenses Do not remove eyepiece from eyepiece tube Clean only the outer surface Breathe on lens to dampen surface then wipe with lens paper Do not wipe lens surface while dry as lenses are scratched very easily Wipe in a circular motion from center to outer edges d Cleaning objective lenses Do not remove objective lenses from the microscope Clean front lens element only Using a cotton swab saturated with distilled water clean front lens surface Inspect the len