1. Press RUN STOP button to invoke and display STAND BY mode the power unit will beep and therapy mode operating pressure and battery charge status will be displayed c The power unit will default to continuous therapy mode at 80mmHg To switch between continuous and intermittent therapy modes press and hold the THERAPY MODE button until power unit beeps to confirm change of mode Adjust vacuum level if required using the UP and DOWN arrow buttons Press RUN STOP button again to initialise and run the power unit NB Vacuum level and therapy mode can be adjusted when in stand by mode and for up to 1 minute after power unit is running Power unit will automatically lock 1 minute after last button operation when running to prevent inadvertent operation of button functions except MUTE as indicated by L on the display screen Press and hold the UNLOCK button until power unit beeps if further button operation is needed Display screen is only illuminated for a short period after button operation in battery mode 6 Once the power unit is running observe the wound site The dressing should contract noticeably becoming firm to the touch If the dressing fails to contract the dressing has not been completely sealed Reinforce the dressing seal and or adjust the drain and initiate suction again WARNING Particularly when used outside of a medical institution getimmediate A medical assistance from those responsible for the prescri
2. Talley User Guidelines for VENTURI AVANTI VENTURI COMPACT Negative Pressure Wound Therapy Systems Explanation of Label Symbols and Statements l Intertek I oO Note Abbreviation cs Caution Refer to instructions of use booklet Medical Devices Directive 93 42 EEC North America ETL listed Class II Equipment Double Insulated Do not dispose of with the normal household waste please refer to www talleygroup com for further details Manufacturer Date of Manufacture Fragile handle with care 4 fe P s iy QAas WARNING CAUTION IP22 Protect from heat and radioactive sources Temperature limitation Humidity limitation Atmospheric pressure limitation The Type 19 is suitable for connection to type BF applied parts This is a statement that alerts the user to the possibility of serious injury or other adverse reactions with the use or misuse of the device This is a statement that alerts the user to the possibility of a problem with the system associated with its use or misuse Operating Instructions IP Ingress Protection Pump only 2 Protection against fingers or other object not greater than 80mm in length and 12mm in diameter 2 Protection from vertically dripping water when tilted to 15 Negative Pressure Wound Therapy is abbreviated to NPWT throughout this document Contents Page Explanation of Symbols a
3. To stop the power unit press and hold the UNLOCK button until power unit beeps and L clears from display screen Then press and hold the RUN STOP button until power unit beeps three times to return to stand by mode if using the power adapter disconnect it from the power unit and unplug from the power source The power unit will then power off automatically after one minute of inactivity or may be forced off by pressing the MUTE key 10 Place the user manual in a safe place for future use User Information The VENTURI AVANTI VENTURI COMPACT power unit will automatically lock 1 minute after last button operation when running to prevent inadvertent operation of button functions except MUTE as indicated by L on the display screen Press and hold the UNLOCK button until power unit beeps if further button operation is needed i e change of therapy mode or vacuum level or returning power unit to stand by mode The power unit will lock again 1 minute after last button operation Please note that when running the VENTURI AVANTI VENTURI COMPACT power unit must be unlocked before it can be returned to stand by mode i e the power unit cannot be put into stand by mode when the L is displayed on the screen During operation the power unit should be placed upright on a horizontal surface with the display uppermost The VENTURI AVANTI power unit can be placed in the optional metal holder for use at the bedside or attached
4. 1 2 P 800 MHz to 2 5GHz d 2 3 VP 0 01 0 01 0 01 0 01 0 12 0 12 0 12 0 12 0 12 0 12 0 12 0 12 0 23 0 23 0 23 0 23 0 1 0 1 0 1 0 1 For transmitters rated at a maximum output power not listed above the recommended separation distance d in Meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in Watts W according to the transmitter manufacturer Note At 80MHz and 800MHz the separation distance for the higher frequency range applies Note These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people This medical device is compliant with IEC 60601 1 3rd edition Medical electrical equipment safety and essential performance IEC 60601 1 11 2010 Home healthcare environment USER MANUAL PART NUMBER 50 02 15 108 5 Talley Group Limited E k Premier Way Abbey Park Industrial Estate d 1 yv C y EW r wetaya Romsey Hampshire SO51 9DQ England E oe sa TILA DEMO TAMAR DATA UR Za ey TEL 44 0 1794 503500 FAX 44 0 1794 503555 AAAA Ye HIS Tell Aeolelan EMAIL sales talleygroup com 09 2013
5. the power unit will default to continuous therapy mode To switch between therapy modes press and hold the THERAPY MODE button until power unit beeps to confirm change of mode When in intermittent therapy mode the power unit will provide vacuum therapy for periods of 5 minutes followed by a 2 minute rest period The selected therapy mode is displayed on the power unit screen and can be altered in either stand by or unlocked run modes MUTE Press to silence the sounder and to clear the message from the display screen The MUTE button can also be used to force power off in stand by mode UNLOCK LIGHT The VENTURI AVANTI VENTURI COMPACT power unit will automatically lock 1 minute after last button operation when running to prevent inadvertent operation of button functions except MUTE as indicated by L on the display screen Press and hold the UNLOCK button until power unit beeps if further button operation is needed i e change of therapy mode or vacuum level or returning power unit to stand by mode The power unit will lock again 1 minute after last button operation Pressing this button will also illuminate the display screen for 10 seconds if power unit is in battery operation screen will always be illuminated in mains operation USER SELECTABLE FUNCTIONS VENTURI COMPACT only The following functions are selectable in a user menu e Language selection Tilt indicator switch on or off Internal contamination senso
6. Voltage 6V 3 7V Peak Current 850mA RMS 450mA RMS Max Load Wattage TW 12W Vacuum Application Continuous default therapy or Continuous default therapy or intermittent intermittent Pressure Range 10 to 180mmHg 10 to 180mmHg Fixed Internal Battery 6V 2 5Ah NiMH 3 7V 10Wh Lithium lon Rechargeable Cell CANISTER Construction ABS Trans smoked grey textured includes 1 x 8g dissolvable sachet of desiccant Capacity 600ml or 1200ml 300ml Dimensions 300ml n a 156mm x 163mm x 32mm 600ml 200mm x 190mm x 35mm n a 1200mI 200mm x 190mm x 7Omm n a POWER ADAPTER Mains Adapter Type FW 362M 15 supplied FW7556M 12 supplied Input 100 240V 50 60Hz 700mA 100 240V 50 GOHz 400mA Output 15V dc 2A 12V dc 1 5A Cable Length 4 metres 4 metres Part Number 34 26 05 103 34 26 05 104 The above mains adapters are considered part of the ME equipment 12V vehicle power adapter is available as an optional accessory The VENTURI AVANTI VENTURI COMPACT power units must only be used with the Specific external power adapters as supplied by Talley Talley products are manufactured to comply with International and National safety standards Talley design and manufacture products to conform to the requirements of ISO9001 2000 IS013485 2003 and Directive 93 42 EEC Talley reserves the right to modify the specification of any product without prior notice in line with a policy of continual product development Our standard te
7. competent electrical engineer to repair or service the system Talley s standard terms and conditions apply to all sales A copy is available on request For service maintenance and any questions regarding this or any other product please contact Talley It is the customer s responsibility to ensure the following prior to collection e the system is cleaned of any obvious contaminants e contamination status is documented e assistance is given to Talley personnel to bag the equipment if the device has been in a known or suspected infectious environment TRANSPORT AND STORAGE Handle with care Please report instances of damage or impact to Talley Service Department Transport 25 C without relative humidity control and 70 C ata relative humidity up to 93 non condensing An atmospheric pressure range of 700 hPa to 1 060 hPa Suitable for all standard modes of transport when in the correct packaging OPERATIONAL CONDITIONS A temperature range of 5 C to 40 C A relative humidity range of 15 to 93 non condensing Operational Atmospheric Pressure 00 hPa to 1060 hPa Suitable for pollution degree 2 Operational altitude lt 2 000 m IP Rating IP22 pump only MANUFACTURER S GUARANTEE All power units are covered by a 24 month manufacturer s guarantee The intended design life is 5 years if fully serviced Warning and Fault Indicators All sounders can be silenced and messages cleared by pressing the MUTE bu
8. observed e The VENTURI needs to be installed and put into service according to the EMC information below e The VENTURI is intended for use in the electromagnetic environment specified in the tables below The user of the VENTURI should assure that it is used in such environment e In general although the VENTURI complies too the EMC standards it can be affected by portable and mobile RF communications equipment Such as mobile telephones e The VENTURI should not be used adjacent to or stacked with other equipment In case adjacent or stacked use is necessary the VENTURI should be observed to verify normal operation Guidance and Manufacturer s Declaration Electromagnetic Emissions IEC 60601 1 2 Emissions Test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Class B The VENTURI systems are suitable for use in all establishments Harmonics emissions 61000 3 2 Class A including domestic establishments and those directly connected to O U S S qun the public low voltage pump supply network that supplies buildings Voltage fluctuations flicker emissions 61000 3 3 Complies used for domestic purposes 14 Guidance and Manufacturer s Declaration Electromagnetic Immunity IEC 60601 1 2 Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Electrostatic Floors should be wood concrete or ceramic tile If floors disch
9. reoccur after 10 minutes LOW VACUUM CHECK DRESSING alternating messages indicates vaccum pressure has fallen below minimum allowable levels Power unit will continue to run whilst these messages is displayed Check that wound dressing is completely sealed and that all tubing connections are secure Press the MUTE button to silence the sounder and clear the messages If fault re occurs contact Talley TUBE CONSTRICTION indicates that pump demand has abruptly ceased Note that a certain level of flow in the drain is required for the indicator to be active The message will occur for example in the case of a blocked tube kinked tube or if the dressing is obstructed Note that the tube constriction indicator can be selected on or off as required See User Selectable Functions on page 8 Default is on TILT VENTURI COMPACT only activates when the device is in active mode and is placed at an angle that could affect the canister full indication The sounder delays for 5 seconds and is self muting when the unit is returned to an upright position Note that the tilt indicator can be selected to be on or off as required see User Selectable Functions on page 8 Default is on FAULT INDICATORS EMI Fault indicates that the unit detects the pressure sensor amplifier is adversely affected by external RF fields The power unit will continue to run whilst this message is displayed Press the MUTE button to silence the sounder Thi
10. to an IV pole using the optional bracket The VENTURI COMPACT is supplied with a bracket for IV pole attachment To allow patient full mobility during therapy both power units can also be placed in the carry cases provided In all instances the power unit canister assembly should be kept upright during use to ensure correct operation The VENTURI COMPACT will display a warning if the unit is over tilted User Information continued BUTTON FUNCTIONS RUN STOP Press to invoke stand by mode prior to running power unit Press again to run power unit Press and hold whilst power unit is running and unlocked to cease operation and return to stand by mode In battery operation the power unit will then power off after 1 minute of inactivity If using mains power switch off by disconnecting power cable from power unit or turning off mains power VACUUM LEVEL wv Vacuum level can be adjusted when in stand by mode and for up to 1 minute after last button operation when power unit is running using the UP and DOWN arrow buttons Pressure can be adjusted in 5nmHg increments between 10mmHg and 180mmHg according to treatment requirements the most widely used being between 60 8OmmHg The power unit will begin operation at the default pressure of 80mmHg The selected vacuum level is shown on the display screen THERAPY MODE The VENTURI AVANTI VENTURI COMPACT power unit offers a choice of continuous or intermittent therapy modes On power on
11. RI COMPACT only Canister Supplied fitted to power unit capacity 600ml VENTURI AVANTI 300ml VENTURI COMPACT Carry bag V pole attachment bracket supplied with VENTURI COMPACT only WOUND CARE SETS supplied separately available in singles and packs of 10 Gauze Wound Care Set flat drain standard large abdominal Gauze Wound Care Set portal drain standard large Gauze Wound Care Set channel drain Foam Wound Care Set portal drain Small standard large Note Other Wound Care Sets may be available please contact Talley for latest information ALSO AVAILABLE e 300ml 600ml 1200ml canister with solidifier Spare battery e 12V Fused Vehicle adapter charger Y connectorx5 Adhesive Gel Patch x 5 Carry case Bedside holder VENTURI AVANTI only IV pole attachment bracket may not be included in all markets Cautions and Warnings There are no special skills required to operate the pump unit however Negative Pressure Wound Therapy should only be used under the advice recommendation and supervision of a licensed Physician and or a registered nurse The medical professional is responsible for applying his her best medical judgment when using this system Prior to use the medical professional s treating the wound must assess how to best use the system for an individual wound Select correct setting for therapy required Care should be taken not to accidently change pressure
12. arge ESD H A 7 ay iad are covered with synthetic material the relative humidity IEC 61000 4 2 should be at least 30 alae lm IEC 2 kV For mains supply lines 2 kV For mains supply lines Mains supply quality for the mains adapter should be 61000 4 4 1kV For input output lines 1kV For input output lines that of a typical commercial and or hospital environment Surge i i Mains supply quality for the mains adapter should be IEC61000 4 5 SAR MHS 1 line SRM line te Mg that of a typical commercial and or hospital environment Voltage dips short lt gt Ur gt 95 Ur for 0 5 cycle lt 5 Ur gt 95 Ur for 0 5 cycle Mains supply quality for the mains adapter should be interruptions and 40 Ur 60 dip in Ur for 5 cycles 40 Ur 60 dip in Ur for 5 cycles that of a typical commercial and or hospital environment voltage variations j _ jj _ j _ In the event of a mains interruption the VENTURI on mains supply 70 Ur 30 dip in Ur for 25 cycles 70 Ur 30 dip in Ur for 25 cycles system will automatically use internal battery power IEC 61000 4 11 gt 5 Ur gt 95 dip in Ur for 5 secs gt 5 Ur gt 95 dip in Ur for 5 secs _ Unless the battery is exhausted Mains frequency Mains frequency magnetic fields should be at levels 50 60Hz magnetic 3 A m 3 A m characte
13. ction N EB Figure shows VENTURI AVANTI The VENTURI power unit battery is installed inside VENTURI COMPACT is similar the unit and is not accessible by users Battery replacement should only be carried out by qualified Talley service personnel Wireless equipment such as mobile phones should be kept at least 10 feet or 3 3m away from this equipment The above warnings cautions and any safety considerations should be observed on a routine and regular basis not only upon installation How to Apply NPWT NB A Wound Care Set must be used with the VENTURI AVANTI VENTURI COMPACT system to carry out NPWT see page 3 for available options Please refer to the Wound Care Set Instructions for Use leaflet Supplied with Wound Care Set for dressing application and setup CAUTION The medical professional is responsible for using his her best medical judgment when using this system Prior to use the medical professional s treating the wound must assess how to best use the system for an individual wound 1 Remove all packaging from the power unit and mains adapter Note The power unit internal battery may be mostly discharged or in an inert state VENTURI COMPACT only on first use This will require the power unit to be used with the power adapter which will automatically wake an inert battery VENTURI COMPACT only operate the system and charge the internal battery If not already in place attach canister to
14. e VENTURI AVANTI Instructions for its use are included with the part Battery Information e A fully charged battery should operate the power unit continuously for at least 24 hours e Charge status is shown on the display of the power unit when it is in stand by and run mode e As the batteries are automatically charged as required when the system is operating on mains power the battery module should not require removing or changing in normal use Use only the mains adapter or optional vehicle adaptor supplied with the system When power unit operation times when running from the internal battery are noticeably shorter than normal it is time to replace the battery pack Contact Talley or authorised dealer for battery replacement service e Never use any battery pack that is damaged or worn out Use the battery pack only for its intended purpose e The battery pack is not serviceable and should be replaced if faulty indicated by applicable battery faults displayed in place of battery charge status icon Contact Talley or authorised dealer for battery replacement service For VENTURI COMPACT only The internal Lithium lon Battery may be received or become inert if left for extended periods in storage or a fully discharged condition Connection a power adapter will wake charge the battery and immediately allow the full use of the Battery Information continued vacuum power unit It is possible to force the vacuum power uni
15. flat face of power unit by matching up the rear location pegs and rotating locking knob 1 4 turn clockwise to secure Ensure canister is correctly located and secured otherwise NO CANISTER message will appear and power unit will not operate 3 Prepare and seal wound as described in Wound Care Set Instructions for Use How to Apply NPWT continued 4 Attach Wound Care Set to the VENTURI AVANTI VENTURI COMPACT power unit canister by lining up locator stud on the tubing connector with the notch on the canister tubing port located on top corner of canister pushing down gently and twisting clockwise to lock A tube guide is fitted to assist with the routing of the tubing 5 OPERATING THE VACUUM POWER UNIT a If using the mains or vehicle power adapter insert the smaller end DC outlet of the supplied power adapter cable into the side of the VENTURI AVANTI VENTURI COMPACT power unit and the other end into the appropriate power outlet The power adapter indicator should be illuminated NB The battery will charge when the unit is connected to the power source indicated by battery charge status icon on display screen scrolling from left to right and provides automatic power back up if the external power supply or adapter fails It is recommended to use the power adapter when convenient to do so as this will ensure the battery is fully charged when needed A fully discharged battery will take a number of hours to fully charge
16. ilising a choice of VENTURI Wound Care Sets supplied separately which include a choice of gauze or foam dressing a silicone drain and transparent adhesive film Foam Wound Care Sets are supplied with a portal drain and Gauze Wound Care Sets are available with a choice of portal flat or channel drain Both foam and gauze Wound Care Sets are available in a range of sizes according to the wound type being treated The VENTURI AVANTI VENTURI COMPACT NPWT system will benefit from careful installation and use providing a long and effective service life Please read and understand this document completely before applying NPWT CONTRAINDICATIONS Do not place NPWT dressings directly in contact with exposed blood vessels anastomotic sites organs or nerves NPWT is contraindicated for patients with Malignancy in the wound Untreated osteomyelitis Non enteric and unexplored fistulas Wounds with difficult haemostasis Necrotic tissue with eschar present NOTE After debridement of necrotic tissue and complete removal of eschar NPWT can be used te oN List of Components Your VENTURI AVANTI VENTURI COMPACT NPWT system should comprise the following items please ensure you have all of these before installation Note Wound Care Sets are supplied separately VENTURI AVANTI TG600 07 VENTURI COMPACT TG600 08 vacuum power unit e 15V Mains adapter FW7362M 15 VENTURI AVANTI only e 12V Mains adapter FW7556M 12 VENTU
17. medical devices to prevent damage to materials and the degradation of plastic surfaces after prolonged use The use of inappropriate cleaning and detergent materials on medical equipment could damage surfaces and may compromise the ability to decontaminate medical devices adequately or may interfere with device function Talley recommends the use of TECcare CONTROL antimicrobial wipes and fluid to clean and decontaminate all products it supplies to health and social care facilities TECcare CONTROL products provide class leading broad spectrum high level disinfection with an exceptional safety profile Being pH neutral TECcare CONTROL can be universally used on all hard and soft surfaces without any detrimental effect TECcare CONTROL is CE marked for cleaning medical equipment WOUND CARE SETS AND CANISTERS Wound Care Sets and canisters are disposable and intended for single use only After use please dispose of in an appropriate manner in accordance with local regulations and hospital best practice SERVICING Talley recommend that all power units should be serviced every two years or as indicated by the hours to service display The unit contains no user serviceable parts and should only be serviced by either Talley or an authorised dealer Talley will make available on request 10 11 Care and Maintenance continued service manuals component parts lists and other information necessary for Talley an authorised dealer or a
18. nd Statements 1 Introduction Intended Use 4 Contraindications List of Components 4 Cautions and Warnings Information 5 How to Apply NPWT Getting Started Operating the Vacuum Power Unit 6 User Information T Battery Information 9 Care and Maintenance 11 Warning and Fault Indicators 11 Specification t3 EMI EMC Statement 14 EMC Manufacturer s Declaration Back Cover Introduction Intended Use Thank you for choosing to use the VENTURI AVANTI VENTURI COMPACT Negative Pressure Wound Therapy NPWT system from Talley In doing so you have selected an efficient competitively priced product for the treatment of many wounds including pressure ulcers dehisced surgical wounds diabetic neuropathic ulcers venous leg ulcers post surgical wounds sinus drainage and management traumatic wounds and pre and post op flaps grafts The VENTURI AVANTI VENTURI COMPACT NPWT system offers a choice of continuous or intermittent therapy modes and features a lightweight versatile vacuum power unit which benefits from dual power technology offering a seamless choice of mains or battery operation The integral battery is charge optimised and provides long lasting power back up when needed The battery operation option allows the system to function away from a mains power supply for extended periods of time allowing the patient full mobility during therapy if required optional carry case with shoulder strap available NPWT is applied ut
19. nvironment Suitable for continuous or intermittent use Not suitable for sterilisation The materials used in the manufacture of all components of the system comply with the required fire safety regulations Talley advice against smoking whilst the system is in use to prevent the accidental secondary ignition of associated items which may be flammable Do not modify any of the medical devices or accessories in any way or use unspecified parts Do not use unspecified parts Choking may result from a child swallowing a small part that has become detached from the ME equipment A Wound Care Set must be used with the VENTURI AVANTI VENTURI COMPACT system to carry out NPWT Cautions and Warnings continued Please refer to the Wound Care Set Instructions for Use leaflet Supplied with Wound Care Set for dressing application and setup Note that the dressings in the set are sterile and for single patient single use only not intended for reuse Intended for home healthcare and professional healthcare facility environments The device is intended to be used with its carry bag or holder Do not connect to any other medical device or equipment When the canister is not fitted on the vacuum power unit it is essential that the aperture on the rear face of the unit Fig 1 is not blocked or covered e g with a label The power unit and power adapter should be cleaned between patient use refer to Care and Maintenance Se
20. ption and setting of the NPWT system should any of the following occur obvious bleeding or pain the wound site or exudate presents unexpected changes in its condition colour or odour the wound dressing becomes detached or ineffective the tubing becomes blocked 7 To change or remove dressing unlock power unit press and hold UNLOCK until power unit beeps then press and hold the RUN STOP button until power unit beeps three times to return to stand by mode Clamp Wound Care Set tubing and remove by turning anticlockwise and lifting out of tubing port on canister Dispose of used Wound Care Set according to local clinical waste policy If required apply new Wound Care Set and continue NPWT How to Apply NPWT continued 8 Canisters should be replaced as required or weekly To change canister make sure power unit is in stand by mode if still running press and hold the UNLOCK button followed by the RUN STOP button Clamp Wound Care Set tubing and remove by turning anticlockwise and lifting out of tubing port this can be reconnected to new canister and unclamped if wound dressing is not being changed Remove sealing plug from its location on top corner of canister and use to cap tubing port to seal in contents Rotate locking knob 1 4 turn anticlockwise and remove canister Dispose of used canister according to local clinical waste policy If continuing NPWT attach new canister and connect Wound Care Set tubing as previously described
21. r switch on or off Tube constriction indicator switch on or off Battery inert Note that it is essential that the battery is put to this condition before international transportation User Information continued The menu is accessed by a three second press of the unlock key The arrow key scrolls to the functions shown above and the mode select key allows the function to be toggled on or off or selected For language selection press and hold the mode key until the required language is displayed Releasing the mode key will then set the display for the language Pressing the unlock key returns the device to stand by mode IV POLE ATTACHMENT VENTURI COMPACT only The VENTURI COMPACT is supplied with an IV pole attachment bracket to provide safe deployment and additional mobility e To fit the IV pole attachment bracket engage the three screws on the bracket into the corresponding three holes on the underside of the VENTURI COMPACT The screws can be hand tightened and do not require high torque or a tool e The fit the bracket with VENTURI COMPACT attached use the black hand knob to secure it to the vertical IV pole The bracket will fit to IV poles between 19mm and 26mm diameter e When dismounting the IV pole bracket from VENTURI COMPACT note that the three screws are Captive and will remain on the bracket when unfastened to prevent their loss An IV pole attachment bracket is available as an optional accessory for th
22. ristic of a typical location in a typical commercial field IEC61000 4 8 and or hospital environment Note Ur is the A C mains voltage prior to application of the test level Guidance and Manufacturer s Declaration Electromagnetic Immunity IEC 60601 1 2 Immunity IEC 60601 Compliance Electromagnetic Environment Guidance Test Test Level Level Portable and mobile RF communications equipment should be used no closer to any part of the QUATTRO including cables than the recommended separation distance calculated from the equation appropriate to the frequency of the transmitter Recommend separation distance Conducted RF 3M MS g0 MHZ d 1 2 4P IEC 61000 4 6 3 V rms d 1 2 VP 80 MHz to 800 MHz 3 Vim d 2 3 VP 800 MHz to 2 5 GHz Radiated RF 3 V m where P is the maximum output power rating of the transmitter in watts W according to the 80 MHz f IEC 61000 4 3 2 5 GHz transmitter manufacturer and d is the recommended separation survey should be less than the compliance level in each frequency range distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site a Interference may occur in the vicinity of equipment marked with the following symbol A Note1 At 80 MHz and 800 MHz the higher frequency range applies Note2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from struc
23. rms and conditions apply EMI EMC Statement and Manufacturer s Declaration This equipment has been tested and found to comply with the limits of EN 60601 1 2 2001 These limits are designed to provide reasonable protection against harmful interference in both a medical and residential environment This equipment generates uses and can radiate radio frequency energy and if not used in accordance with manufacturer s instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception or other equipment which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one of the following measures e Reorient or relocate the receiving antenna e Increase the separation between the equipment e Connect the equipment to an outlet on a circuit different from that to which the receiver or equipment was connected The equipment having been tested to operate within the limits of electromagnetic compatibility Immunity to interference from nearby sources radiating radio frequency energy Sources exceeding these limits may give rise to operation faults Where possible the system will sense the interference and if it is of short duration transparently take countermeasures whilst operating near normally or failing
24. s indicator will clear when interference ceases CHECK MOISTURE VENTURI COMPACT only indicates that moisture or exudate has been detected in the power unit The power unit will cease to operate on this indication Refer to Talley for service assistance O C Batt or CCT Fail VENTURI AVANTI only will appear instead of battery charge status icon indicates that the battery is not functioning correctly and cannot be used The power unit will only operate using mains power whilst this message is displayed Contact Talley to order a replacement battery If any of the following faults are displayed the power unit will cease to operate PUMP OPEN NO CANISTER PUMP OFF FAULT PUMP SHORT RELEASE KEY BATTERY FAULT VENTURI COMPACT only Should any of these faults occur press the MUTE button to reset the power unit If fault remains re occurs contact Talley The VENTURI AVANTI VENTURI COMPACT power unit may also display the following information the power unit will continue to run whilst these messages are displayed e SERVICE DUE Contact Talley to arrange service e UNCALIBRATED Contact Talley for recalibration 12 Specification POWER UNIT VENTURI AVANTI VENTURI COMPACT Model Ref Venturi v ll T 600 07 Venturi v Il TG600 08 Construction Flame retardant ABS Flame retardant ABS Dimensions 210mm x 205mm x 105mm 161mm x 155mm x 90mm Weight 2 2 kg 1 0 kg DC Input Voltage 15V Nominal 12V Nominal Operating
25. s once set as the efficiency of the therapy may be reduced This could also be caused by pets pests or children The electricity supply is of the type indicated on the power adapter Check the power adapter lead is free from damage and is positioned so as not to cause an obstruction or injury e g strangulation Do not position the vacuum power unit or power adapter such that makes it difficult to disconnect the supply or drain plug Ensure the power adapter lead drain tube or vacuum power unit cannot become trapped or crushed e g via raising or lowering of bed or bed rails or any other moving object All tubes must be free of kinks twists properly connected and positioned so as not to cause an obstruction or injury Do not place the vacuum power unit or power adapter on or near a heat source Never use the power adapter whilst placed on top of or near to material which is flammable or can be damaged by heat The supplied power adapters plug directly into the power source partly mitigating this issue The VENTURI AVANTI TG600 7 vacuum power unit must only be used with the approved power adapter supplied by Talley see Specification on page 13 The VENTURI COMPACT TG600 8 vacuum power unit must only be used with the approved power adapter supplied by Talley see Specification on page 13 The VENTURI NPWT system Vacuum power unit and power adapter is not used in the presence of flammable anesthetics or in an oxygen enriched e
26. t into making the battery inert should this be required by some airline and other carriers See User Selectable Functions Menu for details Page 8 CHARGING THE BATTERY If not fully charged the battery will automatically charge when the power unit is plugged into a power source via the power adapter The battery will charge whilst the vacuum power unit is in standby or a run mode REPLACING THE BATTERY The battery is installed inside the unit and is not accessible by users Battery replacement should only be carried out by qualified service personnel Contact Talley or authorised dealer for battery replacement service Care and Maintenance AN POWER UNIT Always disconnect the VENTURI AVANTI VENTURI COMPACT power unit from the power adapter and the power adapter from the power source before carrying out maintenance repairs servicing or cleaning Check all electrical connections and power lead for signs of excessive wear The power unit power adapter can be wiped down with detergent or disinfectant solution or wipe Do not use solvents Unsuitable for sterilisation Dispose of the power unit power adapter in accordance with the local regulations including WEEE requirements The power unit power adapter should be cleaned between patient use as a minimum In line with the MHRA Medical Device Alert MDA 2013 019 Talley advises customers to use pH neutral high level disinfectant cleaning products to sanitise reusable
27. this will issue a warning and take measures for the continued safely of the user Further increased levels of energy may cause the system to stop operating continuously generate random faults or continuous resets Try to ascertain the source of the interference by turning nearby or Suspect equipment off and see if the interference effects stop In any such event the user is encouraged to try to correct the interference by one of the following measures e Have the interfering equipment repaired or replaced Reorient or relocate the interfering equipment Increase the separation between the equipment and the possible source of the interference Connect the equipment to an outlet on a circuit different from that to which the interfering equipment was connected Information regarding Electro Magnetic Compatibility EMC according to IEC60601 1 2 2001 clause 6 8 With the increased number of electronic devices such as PC s and mobile telephones medical devices in use may be susceptible to electromagnetic interference from other devices The EMC Electro Magnetic Compatibility standard IEC60601 1 2 defines the levels of immunity to these electromagnetic interferences From the other hand medical devices must not interfere with other devices IEC60601 1 2 also defines the maximum levels of emissions for these medical devices The VENTURI conforms to this IEC60601 1 2 standard for immunity and emission Nevertheless special precautions need to be
28. tton once Should any fault occur press the MUTE button to reset the power unit If fault remains re occurs contact Talley WARNING INDICATORS NO CANISTER indicates canister is missing or is not correctly fitted The power unit will fail to operate whilst this message is displayed Check that canister is correctly located and secured as detailed on page 6 NO BATTERY VENTURI AVANTI only indicates battery is not correctly fitted The power unit will only operate using the power adapter whilst this message is displayed Contact Talley or authorised dealer to check battery and installation LOW BATTERY only appears during battery operation the power unit will continue to run whilst this warning is displayed Press the MUTE button to silence the sounder and clear the message Note that the system will automatically shut down when the battery is fully discharged Plug into a power source to charge Warning and Fault Indicators continued CANISTER FULL indicates that the canister has reached its capacity and should be changed The power unit will cease to run and the message sounder will continue until the RUN STOP button is pressed or the canister is removed both of which cancels the message and sounder and returns the power unit to stand by mode Change canister as detailed on page 7 NB The Canister Full sounder can be silenced by pressing the MUTE button however if the canister is not removed changed the sounder will
29. tures objects and people 4 Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electro magnetic site survey should be considered If the measured field strength in the location in which the QUATTRO is used exceeds the applicable RF compliance level above the QUATTRO should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the QUATTRO gt Over the frequency range 150 kHz to 80MHz field strengths should be less than 3 V m Recommended Separation Distance Between Portable and Mobile RF Communications Equipment and the QUATTRO The QUATTRO is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The user of the QUATTRO can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the QUATTRO as recommended below according to the maximum output power of the communications equipment Output Power of Transmitter in Separation distance according to frequency of transmitter in Meters m Watts W 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d