Home
BP OIL -- TOLEDO REFINERY
Contents
1. NOTE For the GasAlertMicro 5 hold the wand at the sample area for at least 2 second per foot of wand and hose This will give the tester pump adequate time to pull in the sample in order to give a proper response The standard 10 hose requires a five second test time The maximum sample draw for this detector is 65 feet Page 10 of 13 PAPER COPIES ARE UNCONTROLLED THIS COPY VALID ONLY AT THE TIME OF PRINTING THE CONTROL VERSION OF THIS DOCUMENT IS AT hittps wss0 bp com Toledo_ Refinery HSSE safety procedures aspx Procedure No SAF 023 Rev No 9 BW GasAlert Micro 5 PID Gas Detector Bump Checking and Calibration Guidelines The BW GasAlertMicro 5 PID provides the simultaneous gas testing for Oxygen O2 Combustibles LEL Hydrogen Sulfide H2S and Carbon Monoxide CO In addition this instrument includes a Photo lonization Detector PID A PID is a portable vapor and gas detector that detects a variety of organic compounds Photo ionization occurs when an atom or molecule absorbs light of sufficient energy to cause an electron to leave and create a positive ion The PID is comprised of an ultraviolet lamp that emits photons that are absorbed by the compound in an ionization chamber lons atoms or molecules that have gained or lost electrons and thus have a net positive or negative charge produced during this process are collected by electrodes The current generated provides a measure of the analyte concentration Th
2. 5 Radio Frequency RF transmittal keep your H2S meter away from your radio antenna Other Acceptable H S Meters Other H2S meters may be used however if they must meet these minimum expectations of performance Have audible visual and vibration alarms Display the current gas concentration on the screen Have 10 ppm low alarm 15 ppm high alarm settings Have peak reading which is cleared by the bump test Have a field audit ability for bump test or calibration such as flashing red light or sticker Be intrinsically safe Class 1 Div 2 minimum 44 45 4 Four gas or other multi gas meters may be worn in place of simple gas H2S meters as long as they also meet the above listed requirements Exemptions There are some times when an exemption from the personal H2S meter requirement is allowable The PPE policy SAF 026 allows traveling to a green zone area from a green zone area within an enclosed motor vehicle to be in regular clothes without the standard refinery PPE Therefore no HS meter is required Some work tasks may decrease the function of or damage the HS meter These and possibly other work tasks can be exempted High pressure water blasting Page 7 of 13 PAPER COPIES ARE UNCONTROLLED THIS COPY VALID ONLY AT THE TIME OF PRINTING THE CONTROL VERSION OF THIS DOCUMENT IS AT https wss0 bp com Toledo_ Refinery HSSE safety_procedures aspx Procedure No SAF 023 Rev No 9 Asbestos abatement Arc gou
3. e Select the appropriate sampling accessories based on the location to be sampled Common sampling accessories include rigid and collapsible wands floats and sintered filters Each is available in variety of lengths e Prior to drawing a sample through a pump wand or other accessory the entire sample train must be tested for leakage This should be accomplished by blocking the inlet at the point furthest from the instrument and ensuring a bump block alarm is obtained CAUTION Most refinery gases are heavier than air and settle or flow to the lowest level within a space At the same time some hazardous gasses are lighter than air Hydrogen and natural gas are both much lighter than air and will normally rise and diffuse into upper atmospheric levels They will be trapped in the top of the unvented spaces of vessels fireboxes roofs and platforms Make sure gas testing is performed at various elevations within every space to ensure proper characterization of the atmosphere e Place the probe inlet at the lowest level of the space being tested Continue to check each open manway or at various elevations in open spaces to obtain representative gas test results e Keep sampling probe inlet above flammable liquid or water levels Liquids drawn into the instrument will make it inoperative Also you want to measure the vapors not the liquid The use of a float accessory on the sample wand may be helpful Page 2 of 13 PAPER COPIE
4. This flushes the hose to avoid contamination for another test Return the slide switch to position 0 to eject the chip Chip Measuring Range Purge Time Ammonia 2 0 50 0 ppm 3 Minutes Benzene 0 20 10 0 ppm 2 Minutes Carbon Monoxide 5 0 150 ppm 1 Minute Petroleum Hydrocarbons 20 0 500 ppm 1 Minute Sulfur Dioxide 0 40 10 0 ppm 1 Minute Page 5 of 13 PAPER COPIES ARE UNCONTROLLED THIS COPY VALID ONLY AT THE TIME OF PRINTING THE CONTROL VERSION OF THIS DOCUMENT IS AT https wss0 bp com Toledo_Refinery HSSE safety_procedures aspx Procedure No SAF 023 Rev No 9 PERSONAL H2S METER SPERIAN BRAND The Personal H2S meter is designed to detect hydrogen sulfide H2S in the atmosphere by diffusion through the sensor port on the front of the instrument It serves as a personal alert device by alarming to the user that H2S concentrations are above 10 ppm This guideline will review the operation and bump testing calibration of this direct reading instrument Refer to SAF087 for more information on H3S THE SPERIAN H2S PERSONAL PORTABLE GAS DETECTORS HAVE BEEN DESIGNED FOR THE DETECTION AND MEASUREMENT OF POTENTIALLY HAZARDOUS ATMOSPHERIC CONDITIONS IN ORDER TO ASSURE THAT THE USER IS PROPERLY WARNED OF POTENTIALLY DANGEROUS H2S ATMOSPHERIC CONDITIONS ITIS ESSENTIAL THAT THE USER FULLY UNDERSTANDS AND FOLLOW THE OPERATING REQUIREMENTS OF THE METER General Operation of the Sperian Personal H2S Mete
5. both tubes Use the built in tube breaker located towards the inlet of the sample pump Confirm that the pump piston type handle is fully pushed in Insert the detector tube into the rubber inlet of the sampling pump making sure the arrow on the tube points toward the pump In the case of a twin tube connect the C marked ends of the two tubes by using the rubber connector provided Align the guide marks on the pump shaft and handle Pull out the handle fully until it locked Wait until the sample time has fully elapsed sample time is listed on the on the information sheet provided with the specific detector tube If multiple strokes are required repeat the previous three steps Remove the tube from the pump and read the indication at the end of the color change layer Unlock the handle by turning it more than 1 4 turn to restore it to the initial sampling position It is recommended to perform a field volume leak check before use as follows 1 2 3 4 Push the pump handle fully in and align the guide marks on the pump shaft and the handle Insert a fresh UNBROKEN detector tube into the rubber inlet Pull out the handle fully until it is locked and wait one minute Unlock the handle by turning it more than 4 turn and guide it back gradually by applying a little force Be certain to do this gently otherwise the handle will spring back due to the vacuum in the cylinder and could damage the pump stop Confirm the
6. handle returns within 1 8 inch from the initial position If this is not confirmed the sampling pump will require maintenance Draeger CMS with measuring chip cartridges Basic Operation NOTE The Drager CMS does not require calibration However the analyzer should be inspected by trained personnel in accordance with the factory calibration sticker located on the side of the instrument or after 500 measurements whichever comes first If the instrument reaches the 500 measurements a message will appear stating maintenance required This message will not affect the normal operation of the device but it is recommended that the instrument be sent out for inspection This can be coordinated through the onsite Fisher Safety Store The slide switch plate on the top of the instrument is its mechanism to operate When situated in the 0 position the analyzer is switched off To switch the analyzer on slide the switch plate to position number 1 The analyzer immediately performs a complete self test followed by the message function test ok displayed for three seconds Next the message insert chip prompts you to push the chip selected for measurement into the rear inlet flap and into the analyzer The analyzer will verify the chip type and measurement range Five seconds later you will be instructed to set the switch to position number 2 Page 4 of 13 PAPER COPIES ARE UNCONTROLLED THIS COPY VALID ONLY
7. AT THE TIME OF PRINTING THE CONTROL VERSION OF THIS DOCUMENT IS AT hittps wss0 bp com Toledo_ Refinery HSSE safety_procedures aspx Procedure No SAF 023 Rev No 9 3 The instrument conducts a leak test and will indicate that the chip is ok You are instructed to move to position number 3 4 In this position measurement begins as air is being pulled through the chip You will see a bar type indicator on the screen to inform you that sampling is actively progressing You may leave the instrument and come back to it without concern 5 Finally the result of the measurement appears on the display 6 You may move the switch to position number 2 in order to take another test This feature is only available on analyzers manufactured after November 2000 Alternately the switch can be moved back to position 0 and the analyzer will eject the chip You may repeat the process to take another test as described above Remember you have up to 10 tests on each chip cartridge 7 The analyzer may also be equipped with a remote sampling hose An additional high flow pump is added to the analyzer and a 9 foot extension hose fits into the pump inlet Prior to use the integrity of the sample train must be verified Instructions for testing the sample pump are as follows a Assemble the complete sample train including sampling hose and float if being used b Turn onthe Draeger CMS by sliding the switch to position 1 The analyzer p
8. BP OIL TOLEDO REFINERY Procedure No Certified Refinery Wide SAF 023 Effective Date Rev No 9 July 1 2010 Using Direct Reading Gas Testing Equipment Written By Auth By D C Durnwald Page 1 of 13 Brian Burre Signature on file This procedure is to serve as a guide for the use care calibration and or bump test of portable direct reading gas testing equipment Refer to the manufacturer s technical manual for complete instructions MOC 2010 548 001 HEALTH Insure that all calibration activities are performed in clean air and in a well ventilated room or fume hood Special PPE amp Special Hazards SAFETY Standard Refinery PPE Safe work practices using compressed gas cylinders REFERENCE Operating manuals for DOCUMENTS BW GasAlertMicro5e GasAlertMicro5 PID amp Microclip Detectors Gas Tec Tester with Detector Tubes Draeger CMS Sperian Instruments Personal H2S Meters SPECIAL MATERIALS amp EQUIPMENT Sample hoses and Probe Calibration Gases Regulators Tubing Detector Tubes and Chips QUALITY Verify the expiration date stamped on the gas cylinder detector tubes or chips or the last calibration of gas tester ENVIRONMENTAL PAPER COPIES ARE UNCONTROLLED THIS COPY VALID ONLY AT THE TIME OF PRINTING THE CONTROL VERSION OF THIS DOCUMENT IS AT hittps wss0 bp com Toledo_Refinery HSSE safety procedures aspx Procedure No SAF 023 Rev No 9 OVERVIEW Testing work atmospheres for potentially
9. S ARE UNCONTROLLED THIS COPY VALID ONLY AT THE TIME OF PRINTING THE CONTROL VERSION OF THIS DOCUMENT IS AT hittps wss0 bp com Toledo_ Refinery HSSE safety_procedures aspx Procedure No SAF 023 Rev No 9 e Do not take samples in a steam filled atmosphere Steam can damage gas testing equipment and or give false readings NOTE Consult the instrument user s manual for sample length limitations as well as recommended sampling times As a general rule samples taken with a pumped instrument should be taken for a minimum of 2 second per foot of wand and hose This will give the tester pump adequate time to pull in the sample in order to give a proper response e Standard 4 gas testers may be used to determine the oxygen concentration in a space or vessel down to zero percent e Testing of flammable gases can not be achieved in an oxygen deficient atmosphere If a space is oxygen deficient and flammability needs to be assessed a special gas tester Cannonball will be used Contact a safety advisor for this assessment WARNING Catalytic bead LEL sensors generally require a minimum of 10 Oxygen in the stream being sampled to function properly If it is necessary to determine the concentration of flammable gas in an oxygen deficient Less than 19 5 Oxygen atmosphere consult a Safety Advisor or Process Engineer for additional guidance Specialized equipment will be utilized to to obtain accurate gas test results 2 0 Trai
10. before using the meter again but after the peak reading has been recorded onto the traction report our supervisor NOTE DO NOT use your personal H2S meter to assess the gas levels in a work area Contact the BP operator for the risk assessment If there is H2S in the area leave and have the BP operator re evaluate the hazard These are personal alarm meters and are used to alert you if atmosphere conditions change They should not be used as gas testers Potential H S Sensor Interferences There are some recognized interferent materials that may give an inaccurate gas reading and or cause the meter to alarm NOTE It is important to understand these but the critical behavior is to not assume it is a false reading or alarm Respond as if it is an accurate reading and re assess the situation Some potential interferences 1 Isopropyl Alcohol component of some hand sanitizers surface cleansing wipes other cleaning products inspect spray fragrance cologne If you need to use a product that contains isopropyl alcohol you may place the H2S meter in a sealed plastic bag or store the meter remote from the location impacted by isopropyl alcohol 2 Hydrogen Gas while it is not a one to one ratio Hydrogen gas can give a false reading and further risk assessment is needed verify with a Draeger CMS using an H3S chip or Gas Tec tester 3 Hydrocarbon vent gases which do not contain H2S 4 Diesel Exhaust from the oxides of nitrogen NOx
11. bol will appear on the screen when lt 7 days of battery life remain If the detector still has time remaining the lithium battery can be replaced under warranty at the Safety Store 7 Wear the H2S meter within your breathing zone on a collar or pocket It must be on your outermost layer of clothing Do not wear it on a hardhat or belt 8 Ifthe meter alarms stop work assess wind direction leave immediately crosswind contact the operator or other BP supervisor complete a near miss report 9 Supplied air is required when H2S concentrations exceed 10 ppm PO Do Page 8 of 13 PAPER COPIES ARE UNCONTROLLED THIS COPY VALID ONLY AT THE TIME OF PRINTING THE CONTROL VERSION OF THIS DOCUMENT IS AT https wss0 bp com Toledo _Refinery HSSE safety procedures aspx Procedure No SAF 023 Rev No 9 BW GasAlert Micro 5 Gas Detector Bump Testing and Calibration Guidelines The GasAlertMicro 5 detector can provide simultaneous detection of up to five gases The most typical configuration of this instrument in the refinery is as a standard 4 Gas instrument testing for Oxygen Oz Combustibles LEL Hydrogen Sulfide H2S and Carbon Monoxide CO There are several instruments that are also equipped to test for Sulfur Dioxide The instrument s primary use is for hazard assessment It should NOT be used to estimate gas stream concentrations DO NOT use this instrument to measure gas streams under pressure The GasAlertMicro 5 req
12. e instrument s primary use is for hazard assessment It should NOT be used to estimate gas stream concentrations DO NOT use to measure gas streams under pressure NOTE The PID is far more sensitive than a traditional LEL sensor The PID identifies hydrocarbons in parts per million A 1 concentration represents 10 000 PPM Do not use a PID equipped detector to test atmospheres known to contain high levels of hydrocarbons This will contaminate the bulb and lead to unnecessary cleaning or repairs and instrument down time Use the BW GasAlert Micro 5 in areas suspected of having high hydrocarbon concentrations The GasAlertMicro 5 PID requires a bump check daily prior to use The instrument will prompt the user to perform a bump check the first time the detector is turned on after midnight The unit will not function without a bump check being performed Individuals are encouraged to personally bump check each instrument prior to use A bump check is performed using the MicroDock II Automated Instrument Docking Station All bump check records are stored on data cards included in both the instrument and in the docking station Additional recordkeeping is not required Data is also recorded while the instrument is in use The instrument holds over 500 000 lines of data and records the detector s readings every 5 seconds The oldest data is lost as new data is recorded How TO BUMP CHECK complete once a day in a clean atmosphere 1 Press on off bu
13. ensor fails bump check or calibration and is turned off before being successfully rechecked a warning screen will appear near the end of the start up sequence The instrument will indicate which sensor failed a previous check and ask you to accept You must select YES to accept this Selecting NO or failing to make a selection within six seconds will result in the power to the sensor that failed the check being shut off It will not pass a bump check or calibration until this message is acknowledged e If the meter does not pass the bump check the MicroDock II will indicate the failure with a red light The meter will attempt a full calibration NOTE If all four sensors fail there is likely an issue with the gas supply to the dock You should check the calibration gas supply to the MicroDock to ensure the bottle is not empty and is connected to the appropriate inlet Inlet 2 e If the instrument passes the calibration as indicated by a green light on the dock no further action is required The instrument can now be used If the instrument fails the calibration shut off and restart the instrument paying particular attention to the start up sequence Often issues can be identified during this process e The O2 sensor is calibrated to atmospheric oxygen each time the detector is started All other sensors require calibration every 180 days or upon a failed bump check The detector will automatically calibrate due sensors when placed in the dock
14. erforms a self test of its electronic functions c Wait for the load chip message d Insert a training chip with the arrow pointing toward the analyzer and barcode up Press chip in an upward direction in order to open the inlet port Slide chip evenly and completely into the Chip inlet e Slide switch to position 2 Wait for chip ok message f Slide switch to position 3 g Place your finger over the Gas inlet or pinch the hose at the point nearest the open end Ina few seconds you should hear a change in the low suction pump sound and see the message pump error appear on the screen h If your instrument fails to show this message it may not be used for remote sampling Please return it to the safety store for repair NOTE You must use the high flow accessory pump in order to use the remote sampling hose Otherwise there is not enough suction to get enough air through the 9 feet of hose and properly give a test result Also ensure that you follow the manufacturer s instructions for proper purge times between tests Times will vary among the different chips 8 When using a remote sampling hose you must also flush the hose with the environmental air you are testing according to the times below before moving to Position 2 This insures that there is adequate time to travel through the length of sample hose When finished remove the hose from the environment and draw in clean air for 1 to 2 minutes
15. ging Others identified by the Control of Work risk assessment and a discussion with the area safety advisor Visitors For one day visitors the Security reception area will have access to loaner HS meters The BP contact bringing in the visitor must notify Security that an HS meter is needed An instruction card will be provided and a sign off that he she has been issued a loaner personal H2S meter and they understand how to use it Short term Duration Contractor Employees For other contract workers who will be in the plant for 30 days or less the Fisher Safety Store will have a rental supply of H2S meters An instruction sheet will be given with the personal H2S meter to insure that the user understands its function A work order number is needed to account for these meters when issued The BP contact will be responsible to locate an accessible bump test calibration station for the rental meter SPERIAN BRAND PERSONAL H S METER QUICK CHECK SHEET NOTE The meter must be bump tested daily prior to use Obtain a personal H2S meter from the Safety Store Locate and use a dock station to perform bump tests and or calibration testing Bump test on a Dock Station daily prior to use The meters are programmed to perform a calibration every 30 days or whenever it fails a bump test The dock station performs both tests automatically If the instrument fails calibration bring it to the Safety Store for replacement or repair A battery icon sym
16. inging an instrument in for repair allow the instrument to run in a clean environment for at least 30 minutes to purge contaminants from the instrument Shut down and restart the instrument If the self test is successful bump check the instrument and proceed with your testing If the instrument continues to fail the self test return for repair 3 Display reads BLOCK PUMP INLET Place your finger over the pump inlet where the hose connects to perform a pump function test You must block the pump for approximately eight seconds NOTE If a pump fails immediately when it is blocked an electrical issue is likely the cause Return the instrument to the Safety Store for repair If a pump check continues beyond eight seconds and ultimately fails an air leak in the system is likely the cause Verify the internal filter and break away inlet are both properly seated and reattempt the pump check If it continues to fail return the instrument to the safety store for repair Page 9 of 13 PAPER COPIES ARE UNCONTROLLED THIS COPY VALID ONLY AT THE TIME OF PRINTING THE CONTROL VERSION OF THIS DOCUMENT IS AT hittps wss0 bp com Toledo_Refinery HSSE safety_procedures aspx Procedure No SAF 023 Rev No 9 4 Place the running instrument face down in the docking section of the MicroDock II and close the lid The bump check will begin automatically A successful bump check is indicated by a green pass light on the MicroDock II NOTE If any s
17. ning e Initial user training for the operation of portable gas testing equipment will be provided by a competent trainer This may include but is not limited to the area trainer the area safety representative the union safety representative or an asset supervisor This training alone is not sufficient for utilizing this equipment for gas testing associated with the control of work process Additional training as a Level 1 Level 2 or Level 3 Authorized Gas Tester is also necessary This training will be part of the required training for individuals expected to conduct gas testing as part of their job responsibilities Page 3 of 13 PAPER COPIES ARE UNCONTROLLED THIS COPY VALID ONLY AT THE TIME OF PRINTING THE CONTROL VERSION OF THIS DOCUMENT IS AT hittps wss0 bp com Toledo_Refinery HSSE safety_procedures aspx Procedure No SAF 023 Rev No 9 DETECTOR TUBE GAS TESTING EQUIPMENT Gas Tec Sampling Pump with Tubes and Draeger CMS with Measuring Chip Cartridges Direct reading detector tubes measure the reaction of a contaminant with a color producing agent inside the glass tube Gas Tec Sampling Pump with Gas Tec Tubes Basic Operation 1 2 oor 7 8 Select a detector tube to be used Verify that the tube is within the expiration date stamped on the tube box Various tubes are available at the onsite Fisher Safety Store Break off both ends of the detector tube in the case of a twin tube break off both ends of
18. nt in for repair allow the instrument to run in a clean environment for at least 30 minutes to purge contaminants from the instrument Shut down and restart the instrument If the self test is successful bump check the instrument and proceed with your testing If the instrument continues to fail the self test return for repair 3 Place the instrument face down in a MicroClip module on the MicroDock II and close the lid The bump check will begin automatically A successful bump check is indicated by a green pass light on the MicroDock II e If the meter does not pass the bump check the MicroDock II will indicate the failure with a red light The meter will attempt a full calibration e If the instrument passes the calibration as indicated by a green light on the dock no further action is required The instrument can now be used If the instrument fails the calibration shut off and restart the instrument paying particular attention to the start up sequence Often issues can be identified during this process e The O2 sensor is calibrated to atmospheric oxygen each time the detector is started All other sensors require calibration every 180 days or upon a failed bump check The detector will automatically calibrate due sensors when placed in the dock Page 13 of 13 PAPER COPIES ARE UNCONTROLLED THIS COPY VALID ONLY AT THE TIME OF PRINTING THE CONTROL VERSION OF THIS DOCUMENT IS AT hittps wss0 bp com Toledo_Refinery HSSE safet
19. o the appropriate inlet 4 Gas blend to Inlet 2 Isobutylene to Inlet 3 e If the instrument passes the calibration as indicated by a green light on the dock no further action is required The instrument can now be used If the instrument fails the calibration shut off and restart the instrument paying particular attention to the start up sequence Often issues can be identified during this process e The O2 sensor is calibrated to atmospheric oxygen each time the detector is started All other sensors require calibration every 180 days or upon a failed bump check The detector will automatically calibrate due sensors when placed in the dock NOTE For the GasAlertMicro 5 hold the wand at the sample area for at least 2 second per foot of wand and hose This will give the tester pump adequate time to pull in the sample in order to give a proper response The standard 10 hose requires a five second test time The maximum sample draw for this detector is 65 feet Page 12 of 13 PAPER COPIES ARE UNCONTROLLED THIS COPY VALID ONLY AT THE TIME OF PRINTING THE CONTROL VERSION OF THIS DOCUMENT IS AT hittps wss0 bp com Toledo_Refinery HSSE safety_procedures aspx Procedure No SAF 023 Rev No 9 BW MicroClip Gas Detector Bump Checking and Calibration Guidelines The BW MicroClip provides simultaneous gas testing for up to four gases The most typical configuration of this instrument in the refinery is as a standard 4 Gas instrumen
20. r e The meters are a two year disposable model and cannot be shut off once turned on e The meter has factory pre set alarm levels of 10 ppm low alarm and 15 ppm high alarm e Prior to initial use all H2S meters must be placed on the dock in the Fisher Safety Store to ensure it includes the most up to date configuration and to initially calibrate the instrument e Future bump testing is the responsibility of the user or assigned person There are numerous locations throughout the refinery with Dock Stations The mode operating button is located on the front of the instrument Press the mode button twice to view the max peak reading e Wear the meter on your outermost layer of clothing very near your breathing zone and within your line of sight DO NOT WEAR ON A BELT OR HARDHAT Bump Testing The H2S meter must be bump tested daily prior to use This action assures that the sensor is functioning properly If you exceed 24 hours since your last bump test a visual reminder light will begin to flash If the meter fails a bump test or when 30 days has elapsed since the last calibration the instrument will automatically perform a calibration when placed in the dock station NOTE Use calibration gas containing 25 ppm H2S can be used with the IQ Express dock stations e The IQ Express Dock Station is the preferred method to conduct bump or calibration testing on the Sperian Brand H2S meters e Prior to bump testing check the date on
21. s likely the cause Return the instrument to the Safety Store for repair If a pump check continues beyond eight seconds and ultimately fails an air leak in the system is likely the cause Verify the internal filter and break away inlet are both properly seated and reattempt the pump check If it continues to fail return the instrument to the safety store for repair 6 Place the running instrument face up in the docking section of the MicroDock II and close the lid The bump check will begin automatically A successful bump check is indicated by a green pass light on the MicroDock II NOTE If any sensor fails bump check or calibration and is turned off before being successfully rechecked a warning screen will appear near the end of the start up sequence The instrument will indicate which sensor failed a previous check and ask you to accept You must select YES to accept this Selecting NO or failing to make a selection within six seconds will result in the power to the sensor that failed the check being shut off It will not pass a bump check or calibration until this message is acknowledged e If the meter does not pass the bump check the MicroDock II will indicate the failure with a red light The meter will attempt a full calibration NOTE If multiple sensors fail there is likely an issue with the gas supply to the dock You should check the calibration gas supply to the MicroDock to ensure the bottle is not empty and is connected t
22. t testing for Oxygen Oz Combustibles LEL Hydrogen Sulfide H2S and Carbon Monoxide CO The instrument s primary use is for continuous monitoring It may not be used for permit gas testing without the addition of an auxiliary pump It should NOT be used to estimate gas stream concentrations DO NOT use this instrument to measure gas streams under pressure The BW MicroClip requires a bump check daily prior to use The instrument will prompt the user to perform a bump check the first time the detector is turned on after midnight The unit will not function without a bump check being performed Individuals are encouraged to personally bump check their instrument prior to use A bump check is performed using the MicroDock II Automated Instrument Docking Station All bump check records are stored on data cards included in both the instrument and in the docking station Additional recordkeeping is not required Data is also recorded while the instrument is in use The oldest data is automatically overwritten as new data is recorded How TO BUMP CHECK complete once a day in a clean atmosphere 1 Press on off button on the instrument 2 Ensure the instrument indicates a successful self test of all sensors NOTE If a sensor fails a self test during the start up procedure the power to the sensor is turned off The instrument will not successfully pass a bump check or calibration until all sensors pass the self test Prior to bringing an instrume
23. the gas bottle to insure that the gas is not expired e Place the H2S meter face down into the molded outline of the meter Insure that the D ring is pulled back so that it does not interfere with the sealing surface of the instrument and the Dock e Wait until a confirmation message appears All Tests Passed to acknowledge that the meter is ready for use If it fails the bump test the Dock Station will automatically perform a calibration NOTE If the H2S meter fails the bump test and calibration bring it to the Safety Store for replacement If the battery is low or dead bring it to the Safety Store for repair The meters use a lithium battery A small battery icon will be seen on the display when lt 7 days of battery life remains Page 6 of 13 PAPER COPIES ARE UNCONTROLLED THIS COPY VALID ONLY AT THE TIME OF PRINTING THE CONTROL VERSION OF THIS DOCUMENT IS AT https wss0 bp com Toledo_ Refinery HSSE safety_procedures aspx Procedure No SAF 023 Rev No 9 H2S Meter Alarms What to Do if you meter alarms If your personal HS meter alarms stop your work assess wind direction leave the area and travel crosswind Contact the BP operator or other BP Operations person to notify them of the event Contact your supervisor A traction report must be completed for an event that causes a personal HS meter to alarm The max or peak reading will be entered into the traction report Perform another bump test following an alarm event
24. toxic gases and vapors and flammability is an important function of safe refinery operations It provides an effective means for identifying leaking and residual flammable gases that supply fuel for fires and explosions or providing information from measured levels that could harm health Testing is particularly important for the protection of people and equipment involved in hot work confined space entry or any work near or around leaking equipment In these cases testing serves to detect flammable gases in a prospective hot work area determine the effectiveness of protective measures and reveal unsafe conditions that may arise during the course of the work Proper gas testing is very critical to safe work activities There is more to gas testing than simply reading the meter display of an instrument Proper gas testing must be based on user knowledge of the gas testing equipment functioning chemical properties of measured gases and vapors and understanding the scope of the work to be performed 1 0 Sampling Methods for e All direct reading portable gas detectors using catalytic bead LEL Reliable Results semiconductor H2S photoionization VOC or electrochemical sensors O2 must be bump checked daily prior to initial use Bump test results must be datalogged by the instrument or manually recorded e Prior to use verify that the instrument is within the manufacturers recommended calibration interval for the instrument and sensor type
25. tton on the instrument 2 Ensure the instrument indicates a successful self test of all sensors NOTE If a sensor fails a self test during the start up procedure the power to the sensor is turned off The instrument will not successfully pass a bump check or calibration until all sensors pass the self test Bump checking an instrument that failed a self test will not resolve the issue If the cause of the failed bump check is sensor saturation attempting to bump the instrument actually delays the recovery process since the sensors are exposed to additional gas Prior to shutting the instrument down or bringing an instrument in for repair allow the instrument to run in a clean environment for at least 30 minutes to purge contaminants from the instrument Shut down and restart the instrument If the self test is successful bump Page 11 of 13 PAPER COPIES ARE UNCONTROLLED THIS COPY VALID ONLY AT THE TIME OF PRINTING THE CONTROL VERSION OF THIS DOCUMENT IS AT hittps wss0 bp com Toledo_Refinery HSSE safety_procedures aspx Procedure No SAF 023 Rev No 9 check the instrument and proceed with your testing If the instrument continues to fail the self test return for repair 5 Display reads BLOCK PUMP INLET Place your finger over the pump inlet where the hose connects to perform a pump function test You must block the pump for approximately eight seconds NOTE If a pump fails immediately when it is blocked an electrical issue i
26. uires a bump check daily prior to use The instrument will prompt the user to perform a bump the first time the detector is turned on after midnight The unit will not function without a bump check being performed All Individuals are encouraged to personally bump check each instrument prior to use The bump check is performed using the MicroDock II Automated Instrument Docking Station All bump check records are stored on data cards included in both the instrument and in the docking station Additional recordkeeping is not required Data is also recorded while the instrument is in use The instrument holds approximately six months worth of data and records the detector s readings every 5 seconds The oldest data is automatically overwritten as new data is recorded How TO BUMP CHECK complete once a day in a clean atmosphere 1 Press on off button on the instrument 2 Ensure the instrument indicates a successful self test of all sensors NOTE If a sensor fails a self test during the start up procedure the power to the sensor is turned off The instrument will not successfully pass a bump check or calibration until all sensors pass the self test Bump checking an instrument that failed a self test will not resolve the issue If the cause of the failed bump check is sensor saturation attempting to bump the instrument actually delays the recovery process since the sensors are exposed to additional gas Prior to shutting the instrument down or br
27. y_procedures aspx
Download Pdf Manuals
Related Search
Related Contents
絶 縁 抵 抗 計 Orne Socialiste juin6.pub Samsung LN32C540 32" Full HD Black LCD TV USB1010 User`s Manual Untitled - Bonterra Home & Garden Référentiel des métiers-types des professionnels de l Philips myHomeOffice Floor lamp 69066/87/16 MANUAL CR-72 CR-73 CR-74 COM LOGO Manuel d`Utilisation 16 Copyright © All rights reserved.
Failed to retrieve file