Home
        DataLyzer Spectrum Gage Management
         Contents
1.                           Serial Port to Test  The    Serial port to test    area includes a pull down list for specifying which RS 232 port will be used  for testing     Timeout  Use the    Timeout    field to enter the number of seconds the template will wait for an RS 232 string      ne   DataLyzer Spectrum Gage Management    from the instrument     Clear Button  The CLEAR button clears the window     Message   The    Message    field allows DataLyzer Spectrum Gage Management software to report messages  about the transmission being received  Messages such as    reading complete    or    time out error    may  display in this field     Analog to Digital  The    Analog to Digital    area allows you to high and low master the gage and set the software to do  an automatic interpolation     Use a large and small master gage block or other standard for this test   The procedure     1  Enter the dimensional size of your high master in the high master field    2  Enter the dimensional size of your low master in the low master field    3  Position your cursor in the    high gage reading    field  then put your high master into the gage and  transmit a value to the software using the MASTER HIGH button  The value will transmit from the  gage in terms of a number of digital segments as discussed earlier and will display in the    high  gage reading    field    4  Position your cursor in the low gage reading field  then put your low master into the gage and  transmit a value to the s
2.              Save  amp  Exit Exit                            Load Logo Button   To select a logo to be printed on your reports  press the LOAD LOGO button  A Windows   explorer  window will then appear where you can browse and select your logo graphic file  You must then  mark the    Include company logo on reports    checkbox for the logo to be printed  The logo graphic file  path is visible in the field below the buttons  If a file path is visible in this field  a file is loaded for use  in the reports  You cannot type in this field to change the path  It must be unloaded and reloaded to  be changed     Unload Button  To remove the logo that is currently loaded for use in the reports header  press the UNLOAD button     View Logo Button  Use the VIEW LOGO button to view the logo graphic you selected  This feature will only be  accessible if your logo has already been loaded     Include Company Logo on Reports Checkbox   If you have loaded a logo graphic file  you must check the    Include company logo on reports     checkbox to print the logo on all supported reports  This feature will only be accessible if your logo  has already been loaded     Include Company Name on Reports Checkbox   If this check box is marked  the text entered in the    company name    field on the    Identity    tab will  print as an additional title line at the top of each report  Checking this box will apply this feature to all  supported reports     Logo Width  mm   You can resize the logo graphic 
3.      OK Button    The OK button will close the window     System Information Button    The SYSTEM INFORMATION button will open the    Microsoft System Information    window for your  computer     Send Error Log Button   The SEND ERROR LOG button will appear on the    About DataLyzer      window if an error has  occurred and an error log was created by the system and exists in the Spectrum program folder   Press this button to send the error log via e mail to our technical support department  You must have  access to the internet to use this feature  but no e mail privileges are necessary          134   DataLyzer Spectrum Gage Management    9    9 1    9 2    Appendices    Appendix A   Computer Requirements    Computer Requirements    Item    CPU   RAM   Graphics   Operating System  RS 232 or USB ports   CD ROM Drive   Mass Storage    Uninterruptible power  supply  UPS     Minimum for use with  Microsoft   Access     Database    Pentium   233 MHz  128 MB   VGA  800x600   Windows   98  Required for Gage I F  Required    1 GB minimum    Minimum for use with Microsoft SQL  Server   or Oracle   Databases    Pentium 3   256 MB    VGA  800x600   Windows 98  Required for Gage l F  Required    1 GB minimum     Comport gage drivers are user configurable through the software  The operating system allows as    many as two comports to be used simultaneously     Appendix B   Installation    The DataLyzer Spectrum Gage Management software will operate alone or link to the DataLyzer    Spect
4.      Once you have set the lines per transmission correctly  you may proceed defining the template with  the        signs  see the following    Reading Description    fields      Reading Description Fields  Values for this area must be set through the CREATE TEMPLATE facility in the TEST menu  Entries  are visible from this screen for summary purposes only and cannot be changed using the keyboard     Reading   Field   In some cases a gage will transmit multiple measurements at one time  These fields indicate which  measurement is being referenced by the measurement locator fields  starting  ending  and sign  columns   These will be described later in detail  Refer also to the    Test Menu  Create Template     area for setting up a template to handle multiple readings per line  Up to eight readings can be  identified and used from one transmitted line       no   DataLyzer Spectrum Gage Management    7 12 1    Previous and Next Buttons   When a multiple reading template is being used  the PREVIOUS and NEXT buttons allow movement  between reading locations  When these buttons are used  you will note the    reading number   of       fields will change along the measurement locator fields to indicate which reading you are on and  where within the string that reading is located  A maximum of 64 individual measurements can be  identified and used from each transmission     Line Number of the Reading in the Gage Transmission   The    line number of the reading in the gage transmissio
5.     Master    Fist Master SN 112345 Tolerance   02         Nominal Bias Uncertainty   Tolerance    2 001 0 0047  3 98 0 0057 0 0092    6 03  0 0007 0 0067    8 02 0 0013 0 009    9 96 0 006 0 0149         The Nominal and Bias figures are copied from the spreadsheet  The Uncertainty column is a  calculation of the Combined Uncertainty  calculations shown above   In the graphic on the right the  minus and plus tolerances are shown as vertical lines  For each calibration point the bias  black line   and Combined Uncertainty  grey bar  are shown  If either of these are outside the tolerance limits the  calibration point has status  Fail      Calibration Label    This opens a window for printing calibration labels on A4 or similar sheets of peel off labels  Select a  study by clicking in its study number field in the Calibration History window  The row will highlight to  show it is selected  Select LABEL PRINT from the FILE menu to print a calibration label for the  selected study       is  Print calibration label  fx    Label Position Label Title  m Fields    Horizontal  fi  gt      Calibration IV Calibration Number F Description    Vertical 3  gt   E V Calibration Date I Model    IV Auto Increment Font   IV Calibration Status I Serial Number  IV Inspector TF Company Number             I    Gage Name IV Calibration Due    Font               Print   Close         If you have not previously printed labels using the type of label sheet  click the Print Setup tab and    Gage Studies 
6.     field     Barometer  Enter the barometric pressure when the study is performed in the    Barometer    field     Notes    You can use the    Notes    field to enter comments such as the condition or other aspects of the gage  being studied       62   DataLyzer Spectrum Gage Management    Instructions Button  When you press the Instructions button  the Measuring Instructions window will appear showing any  measuring instructions and attachments that were defined on the gage s    gage description    window     Calibration Datasets Frame   This frame controls the creation and removal of    Calibration Datasets     A calibration dataset is a  table like a spreadsheet where reference values and measurements can be entered for determining  the bias throughout a measurement mode of a gage  Because a cage can have more than one  measurement mode  e g  inside diameter  outside diameter  depth   you can have more than one  calibration dataset in a calibration study     When a study has one or more Calibration Datasets  the study status  pass  fail  or not set  is  automatically calculated from the measurements in the datasets  it cannot be set manually     When you add a calibration dataset  the Calibration Dataset Design window opens     The Dataset Table  The following shows the data entry area for a typical Calibration Dataset  The columns that appear  are determined by the Dataset window described above  Fields for user entry are white  other fields  show calculated results     
7.     for sign      Once created  use the TEST TEMPLATE item in the TEST menu so as not to disturb the template       signs  If the      sign template placement is incorrect  you will only get part of the measurement  value through to the software during data entry  A gage reading like 3 655 might come through as  655     The Up and Down Buttons  The UP and DOWN buttons allow you to scroll to the transmission lines that are off the screen     The Decimal Button   Pressing the DECIMAL button after you take a sample string from a gage will cause each character  in every line to be replaced with its ASCII decimal equivalent  The ASCII values will be in square  brackets  This function allows you to see unprintable characters that may be in the transmission     Example   1234567890  Gage transmission  X  5 42mm    Setup    Setup Menu  113    After pressing DECIMAL   10   88   58   32   53   46   52   50   109   109     Column  The    columr    field shows the position of the cursor within the    transmission    field     Advanced Concept Multiple Measurement Values per Transmitted Line   Some instruments transmit more than one measurement in a line  To accommodate these situations  without sending the same string twice  overwrite each character in the measurement  including the  decimal point  with Fs and save the template  At least one space should exist between the  measurements to separate them  When a       sign is found right in front of the measurement  i e   contiguous   over
8.    Data Menu    e      5 6    enter measurements for the type of label sheet you are using  After entering the measurements click  the Print Label tab     In the Label Position frame indicate the label on the sheet that you want to print on by clicking the  Horizontal and Vertical   or     buttons  Ifthe Auto Increment box is checked the printing position  will move after a label has been printed     The Label Title is printed centrally at the top of the label  The Font button allows you to change the  printed font for the label title     The Fields frame is for selecting the information that you want to appear on the label  The Font    button allows you to change the printed font for this information  You may need to change the font  size to ensure that there is room on the label for all the selected fields     Calibration Dataset Design    When you add a calibration dataset  the study design window opens          w  Calibration Study Design    Calibration Points  5  Measurements for each point  2    Use Note column E    Use Measurement Uncertainty column I    Use Master Uncertainty column E              Calibration Points   Typically  you will calibrate at more than one point within the gage   s range to check for bias  throughout its entire range called    linearity     Enter the desired number of calibration points  typically  3 8  in the Calibration Points    field     Measurements for each point   Enter the desired number of measurements that will be performed for each 
9.    How to use the Help System    While you are using the DataLyzer Spectrum Gage Management System you can press F1 on the  keyboard at any time to open a help screen  Often this will give specific help for the area of the  program you are currently using     When the help screens open use the four tabs located on the left to access general areas of Help or  move to individual help topics     Contents offers a list of topics from which to choose  Click on the topic to instantly access that  section of the document  You many need to click the small   beside an entry to open up a list of  sub topics     Index allows a fast search on specific key words or phrases     Search is a more general search for a word or phrase  It will return all occurrences of that word or  phrase in a list  You can select a particular citation by double clicking an item from that list  The  help system will jump to that location in the help text     You can broaden your search to  Match similar words  by checking that box below the tabs  You  can further refine a search by using the check box titled  Search previous results      Favorites allows you to save the location of a description found in the text  so you can access it  directly next time without searching or scrolling through the document     Purpose    The Purpose of the DataLyzer Spectrum Gage Management  System    The DataLyzer Spectrum Gage Management System was designed to assist supervisory personnel  determine the credibility of measure
10.    Other Supplies    field     User Fields 1 12  If custom user fields have been created in SETUP  SYSTEM DEFINITIONS  these fields will appear  beneath the    Other Supplies    field  Enter the specified information as necessary     GR amp R and Calibration Menu   The INSTRUCTIONS menu item opens the Instructions window in Set mode  This is used to setup  GR amp R and Calibration study instructions for the gage type  When assigning the Gage Type to a  Gage in the Gage Description window  the instructions for the Gage Type can be automatically  applied     File Menu   Save or  F2   The SAVE item  or pressing  F2  without opening the menu  saves the changes made in the    Gage  Type Definition    window     File Menu   Save  amp  Exit or  Ctrl   S   The SAVE  amp  EXIT item  or pressing  Ctrl   S  without opening the menu  saves the changes made  to the gage type definition and exits to the    Define Gage Types    catalog window     File Menu   Exit or  Ctrl   X   The EXIT item  or pressing  Ctrl   X  without opening the menu  closes the gage type definition  window without saving changes       96   DataLyzer Spectrum Gage Management    7 5 Define Masters    The DEFINE MASTERS item is provided to define master gage blocks for measuring and testing  purposes  When conducting accuracy or calibration studies  gage measurements are compared with  the    known    values of these standards to determine the accuracy of the gage     Define Masters pel    File    Name ID Number   MKROO
11.    The    Gage Suppliers    setup window will then appear  see here for a description of this window      Once a supplier  manufacturer  is selected  it appears in the field  You can view more specific  information about the supplier by clicking on the field  a window containing read only information  about the supplier will appear     Calibration Button and Field   No manual entry is allowed in the    Calibration    field  For entry in this field  use the CALIBRATION  button to open the    Suppliers    list  Note that the    gage calibration    option button is selected in this  window  allowing only predefined suppliers of that type to appear in the list for selection  To select a  predefined supplier s name from the list  double click on it or highlight it in the list and press the  SELECT  Press CANCEL to exit without selecting a supplier  To add a new supplier to the master  list  type a supplier name in the    Company    field and press SELECT  The program will notify you that  the supplier does not exist and ask if you want to create it  Press YES  The    Gage Suppliers    setup  window will then appear  see here for a description of this window      Once a supplier is selected  it appears in the field  You can view more specific information about the  supplier by clicking on the field  a window containing read only information about the supplier will  appear     Recalibration Interval  Enter the number of days between calibrations in the    Recalibration Interval    f
12.    To open an existing gage type  select it from the list or type its name in the    Gage Type    field  To  create a new gage  enter the name of the new gage type in the    gage type    field     OK Button  Press the OK button to open the    Gage Type Definition    window for the new or selected gage type     Delete Button  To delete a gage type  select it from the list or type its name in the    Gage Type    field and press the  DELETE button     Exit Button  The EXIT button closes the    Define Gage Types    catalog and returns to the main DataLyzer  Spectrum Gage Management software window        Gage Type Definition    Window    This section describes how a    Gage Type    is created  The    Gage Type    feature allows the user to      o    DataLyzer Spectrum Gage Management    group or flag similar gages under one label or    gage type     There is a twofold purpose     1  It makes gage entries and studies easier to find in a large database  2  If allowed by your quality manual and auditors  gage R  amp  R studies can be performed generalizing  results from one gage to all gages in the associated gage type group     To create or edit an existing gage type enter a new or existing gage type name and press OK in the     Define Gage Types    catalog  the    Gage Type Definition    window will appear  Any changes made in   the gage type s definition window after a gage type has been selected in other areas of the program  will not be applied in those areas  The    Gage Type 
13.    Trials  Enter the default number of trails used when conducting a variable GR amp R study in the    Trials    field     Parts  Enter the default number of parts tested when conducting a variable GR amp R study in the    Parts    field     Attribute Study Preferences  You can set the preference configuration to match the attribute Gage R amp R study configuration you  expect to use most frequently in a similar way to a variable study     Re Evaluation Time Units  The re evaluation period for Calibration studies and GR amp R studies can be set in days or months  Set  the Calibration re evaluation period for a gage in the Gage Description window  Set the re evaluation       2   DataLyzer Spectrum Gage Management    7 4    period for GR amp R studies when creating a GR amp R Study Definition     GR amp R calculations based on   Traditionally the full spread of variation due to components measured during GR amp R studies  EV  AV   R amp R etc  was considered to be 5 15 times Sigma  Standard Deviation or Estimated Standard  Deviation  of the component  5 15 times standard deviation represents 99  if the full spread  expected  assuming a    normal    distribution curve     When 5 15 Sigma is selected in this frame  deviation values calculated from the study are multiplied  by 5 15 for comparison with other sigma values or tolerance  When Process Variation if this is  selected as the Total Variation method  the standard deviation of characteristic measurements is  also multiplied 
14.    as determined by the software  Therefore  when  creating or editing gage supplier information  it is important to remember to select all possible     company types    that may apply for each supplier so these suppliers can be accessed in the  necessary lists     The    Gage Sales    company type indicates the supplier being defined sells gages     Gage Supplies     indicates the supplier being defined is a source for gage supplies  cleaners  solvents  tissues   chemicals  etc       Gage Calibration    indicates this company calibrates gages     Gage Technical  Support    signifies this company supplies technical support     Gage Manufacturer    denotes this  company manufactures gages     Certifying Agency    indicates the supplier does gage certifications     NOTE  If the supplier does certifications or calibrations  their standards should be traceable to a  national or international standard  Use the fields in the    Certification    area of this window to specify  the traceability information     Certification  Use the fields in the  certification  area to enter traceability information for    Gage calibration    and     Certifying agency    company types     Enter the agency that certified the supplier to do its certifications in the    Certifying Agency    field   Enter the certification number of the supplier authorizing it to do certifications in the    Cert       field   Enter the date of the certification   s expiration in the    Expiration Date    field  The    D
15.    measurement values  should have    R    or    R1    in their function fields  Characteristics looking for    Y    measurement values  should have    R2    in their function fields     If you are only interested in one of these measurement values  the whole string is processed but we  overwrite only the measurement value of interest with         signs  In this case  no    R    or    R1     references in the characteristic function areas are not necessary     The    reading description    area reports the line and column positions where measurements reside   Zeros will appear in the fields of this area until the template has been established through the testing  procedure       12    DataLyzer Spectrum Gage Management    7 12 7 Gage Template Example 3  Multiple Measurements Per Line    7 13       Example reading  _ _ _X 4 379 Y 3 465    This is an example  each time the gage transmits a one line ASCII string containing two  measurements  Elements of the transmission include  three spaces transmitted at the beginning of  the string  shown as underlines      X      then a measurement value followed by another space     Y       and another measurement value  Each space these characters occupy in the line is referred to as a  column   The decimal point is counted as a character   So  the    X    measurement starts in column    6  and ends in column    10     and the    Y    measurement starts in column    14    and ends in column    18      To get the    X    measurement out of 
16.   Fail    and    Not Set        Certificate Scan Frame   This frame provides a way to view a calibration certificate for the study  Make a file of the certificate  using a scanner  for example a jpeg or jpg file type and store it in a known location of the computer   Click the Browse button to locate the file  This location will be stored in the database  To view the  certificate click on the Show button     Advanced Mode    Clicking the Advanced button reveals a tabbed section which contains more fields for the Calibration  study    Gage Studies    Data Menu   61         File Edit Reports  Calibration number   000001 Date  0770472010 Time  fi 4 06  Performed by     First Inspector Manufacturer   Acme Calibration   E  FistGTFirstGaae     odel   Gage Type    First GT Serial Number    Serial No   Description   Desctiption 0000 Company Number   Eompany No    Status   gt  E        Certificate Scan m Calibration Datasets      Show   Browse      Add    Study Details                        Resolution    Temperature       Precision  VEA Humidity      Range  En Barometer  Tee  Units  VE  Instructions     Notes        Resolution  Precision  Range  Units    These fields are automatically filled in based on entries in the gage s    gage description    window   They cannot be changed in this window     Temperature  Enter the ambient temperature when the study is performed in the    Temperature    field     Humidity  Enter the ambient humidity when the study is performed in the    Humidity
17.   This area displays characters transmitted from the gage  To use this window     1  Set the com port to be used for this test using the option buttons   2  Access the READ GAGE button   A countdown will begin    3  Press the    transmit    or    print    button on your gage     The current template will be used and attempt to receive data from the gage   Messages such as    reading complete    or    time out error    may display in the message field     If the software receives anything from the gage through the template  it will show up in the filtered  readings area along with some additional comments supplied by the software related to the number  of channels received  the line number and item number in the line associated with each  measurement  etc   lt  lt  lt     Should have graphical eg gt  gt     Read Gage Button  The READ GAGE button attempts to take measurements from the gage for seven seconds  after  which    time out error    will be reported in the message field     Setup    Setup Menu  115    OK Button  The OK button saves and exits one level to the gage setup window     7 12 3 Special Menu    The SPECIAL menu allows selection of several types of special gaging situations  Once anything  other than    standard    has been selected  the associated window will appear for additional settings in  that area     Standard    Unless the gage is one of the special types listed on this window  the standard item should be  checked in this area     Single Channel A D  Ana
18.   Variable  Group  FirstCG    USL  foo  LSL  fi                     Study Parameters  Variable   Long  Parameters    3 Appraisers  10 Parts  3 Trials  Data Entry Method   Open Sheet   7 Re evaluation required every JU Days       Name  The Study Definition name must be unique    Gage frame   If you have defined gages click the Gage button to select a gage  If you have not defined any gages  you can just type in the name of the gage here  Ifthe gage was selected by clicking the Gage button  the category of the gage will be shown  variable or attribute  in the frame and the Gage Type  if the  gage is included in a Gage Type      If the gage is included in a Gage Type  then GR amp R studies performed using this Study Definition can  be considered as representative of all the gages in the gage type  If at least one GR amp R study has    Gage Details    File Menu    a7      been created using the Study Definition  the Gage Type will become    locked     This means that you  can only change the specified gage to another gage in the Type     Characteristic frame   If the program is being used in conjunction with Datalyzer Spectrum you can select a Datalyzer Part  and Characteristic using the buttons in this frame  Otherwise  just type in Part and Characteristic  names  If a Datalyzer characteristic is selected the Upper and Lower specification limits will be  displayed along with the category of the characteristic  variable or attribute   Ifthe characteristic is  included in a charac
19.   adjustment or repair     DataLyzer Spectrum Gage Management System calibration studies support up to 20 reference  points throughout the range of the gage  so bias can be detected at each end of the range or the  gage as well as numerous points in the middle     Gage Accuracy   Gage accuracy is a similar concept to calibration  The graphic below shows a gage s accuracy to be  the difference between the    true average    measurement and the gage s observed average  Unlike  calibration  multiple measurements are required  Again  this study is a way of determining bias     Reference  Value             Bias    Observed  Average  Value    Explanations  Gage Stability    Gage stability is the difference between a gage s observed measurement results as it measures the  same standard over time  Multiple measurements are taken  Stability determines bias over time       Stability      Time 2    Time 1    Gage Linearity   Gage linearity refers to accuracy of a gage throughout its measuring range  Often lower end  mid   range and high end masters are used to test for this condition  Linearity measures bias at more than  one location within a measuring system   s range  When a gage is used heavily in one area of its  measuring range  it can become worn and less accurate in that range in comparison with other  areas ofits range  Multiple measurements are taken of the several standard sizes     Reference Reference  Value Value  dl  Bias Larger Bias            Observed Observed  Average A
20.   transmission      This field is set from the    Test Create Template    area  This facility is only used when  the sign of the measurement is not contiguous with  right next to  the measurement  Also  it is also  not necessary if the measurements to be received will always be positive     File Menu    New  Use the NEW item from the FILE menu or  CTRL   N  from outside the menu to create a new gage  driver     Open  Use the OPEN item from the FILE menu or  CTRL   O  from outside the menu to open an existing  gage driver     Save  Use the SAVE item from the FILE menu or  CTRL   S  from outside the menu to save the current  gage driver  You must always save a template before you can use it to control the computer s RS     Setup    Setup Menu  u    232 com port s      Save As  Use the SAVE AS item from the FILE menu to save the selected gage driver with the name you have  supplied     Print    Use the PRINT item from the FILE menu to print the current gage driver s settings     Exit    Use the EXIT item from the FILE menu to exit without saving     7 12 2 Test Menu    Create Template   The CREATE TEMPLATE item allows data acquisition from a com port using the baud rate  parity  and stop bit settings of the selected gage template driver  This testing process is also used to set  the    reading description    fields of the    Gage Template    template window     Once communication parameters are set in the    Gage Template    window  the interface is tested  using the    Create T
21.  1  since there is only one range calculation   K3 is 5 15 d2 where d2 is dependent on the number of parts  m  and  g  is 1  since there is only one range calculation    ld2 is obtained from Table D3   Quality Control and Industrial Statistics     A J  Duncan                       Print the report using the FILE  PRINT menu item  Close the report preview window by using the  FILE  EXIT menu item     Report Header   The header of the report contains information about the gage used and the characteristic defining the  study    Statistics from the data sheet include     R double bar   Average of each range calculated from each part trial   all appraisers  all trials    X bar Diff   Average from each trial all averaged for each appraiser  The minimum of these averages  is subtracted from the maximum of these averages     Ro    Part Trial average maximum     Part Trial average minimum    range for the entire study    DataLyzer Spectrum Gage Management    Measurement Unit Analysis Area   Repeatability   Equipment Variation  EV   EV RxK  5  1 R from datasheet    K  is associated with the number of trials in the study    Reproducibility   Appraiser Variation  AV        EV    parts  x   trials        AV    Xpipp   K2  i  X DIFF from datasheet    K is associated with the number of appraisers in the study    Repeatability  amp  Reproducibility  R amp R     R amp R   JEV   AV   Part Variation  PV   Rp x K3    Rp from datasheet    K associated with the number of parts in the study    Tota
22.  10  or  100 subgroups with subgroup size 1  X  amp  Mr charts      Data Tab    The    data    tab produces a data table with parts measured as rows and each appraiser trial as a  column  This tab applies to both long variable and ANOVA studies       42   DataLyzer Spectrum Gage Management                                               GRER Average  amp  Range Method X   y Average Charts Individuals Charts Range  amp  Error Charts  Statistics i E Performance Curve  Part Appr  1 Appr  1 Appr  1 Appr 2   Appr 2   Appr 2   Appr 3   Appr 3   Appr  3  Trial  1 Trial  2 Trial  3 Trial  1 Trial  2 Trial  3 Trial  1 Trial  2 Trial  3  1 56 62 64 55 52 6 64 1 62  2 57 65 61 1 1 1 65 6  64  3 83 63 65 68 57 63 62 1 54  4 61 54 1 58 56 65 68 57 5   5 58 59 51 54 6 58 1 53    56  6 62 1 52 64 61 59 58 54 58  7 67 61 57 53  61 6 52 1 57  8 7 62 72 81 57 58 54 57 61  9 56 63 6 1 1 1 64 58 62  yo  9 53 54 54 58 61 61 61 68    Print m  Print Exit                Average Plot Tab  The    average plot    tab applies only to ANOVA studies     The purpose of the    averages plot    graph is to show potential differences in reproducibility between  operator averages for the same part  Option buttons designate which one of three graphical analyses  will be displayed  They include  runs chart  averages chart  and X Y plot           5 GREER ANOVA Method E    Statistics Average Plot          Appraiser 1 Appraiser 2 Appraiser 3    7 08    Appraiser Averages  Sr a   N o  N          an   gt     2 36      
23.  Define GR amp R Characteristic Groups    If you are using both the DataLyzer Spectrum Gage Management System and DataLyzer Spectrum  SPC software  the DEFINE GR amp R CHARACTERISTC GROUPS facility is used allows you to search  the shared database for characteristics with matching engineering specifications and methods of  being measured  These characteristics can then be grouped together so that the results from  studying one of these characteristics will be applied to all characteristics in the group  This type of  group application is usually only done when a quality department is relatively small so that the same  study appraisers would be involved in all GR amp R studies anyway  lt is not recommended where the  number of appraisers involved exceeds three     To use this facility  your DataLyzer Spectrum Gage Management System and DataLyzer Spectrum  SPC software must share a database  This allows the Gage Management system to read the SPC  characteristic information and search for matching characteristic criteria  specifications etc    The     database path  can be checked in both programs by going to HELP  ABOUT DATALYZER     After selecting the DEFINE GR amp R CHARACTERISTC GROUPS item  the following window will                      appear     Characteristic Group  Te  Exit  Delete    Name field    To create a new characteristic group  type the name of the new group in the    Name    field  To open  an existing characteristic group  select it from the list or type its 
24.  Enter the name or initials of the supervisor overseeing the test in the    Performed by    field  You can  type it in directly or select from a list of previously entered names by clicking the PERFORMED BY  button     Part Field and Button   Enter the part name or part number being studied in the    Part    field  You can type it in directly  or  if  sharing a DataLyzer Spectrum SPC software database  you can select a part from a catalog of all  existing DataLyzer Spectrum parts from the SPC system  complete with data group options  by  pressing the PART button  Using the part and characteristic entry the USL and LSL can be  referenced in the associated DataLyzer   SPC system to allow the calculation of  bias  If no  DataLyzer   SPC system is present then you can enter the part and characteristic information for  reference but you will also need to enter the USL and LSL or process variation range for a   bias  value to be calculated       e   DataLyzer Spectrum Gage Management    Characteristic Field and Button   Enter the characteristic being studied in the    Characteristic    field  You can type it in directly  or  if  sharing a DataLyzer Spectrum SPC software database  you can select it from a catalog of  characteristics from the selected part by pressing the CHARACTERISTIC button  If no part was  specified in the    Part    field  you will first be prompted to select a part  A different characteristic can  be assigned to each individual gage study     LSL and USL   If y
25.  Gage Description x    Fie GR amp R Calibration  Name ID     FirstGTFirstG aged    Gage Type   First GT y   Manufacturer                 Model     Model   Description   Descon      Category    E Variable    Attribute  GAR Re evaluation Interval   90 Days Precision     Calibration Re evaluation Interval    30 Days Range  EEN  Resolution    Units  P  Serial Number   Serial No  Company     Company Mo   Date of Purchase    PO    er  Cost   a  Status     First Status SS Location     FirstLocation        1 Gage Supplier        2 Gage Supplier    i   Documentation        Battery Type    1 Supplies Company        2 Supplies Company   ua   Other Supplies  N    Notes        NamellD     The    Name ID   is the name given to the individual gage  It will be used to identify the gage in the  gage catalog  The Name ID  is initially defined in the gage catalog window when creating a new  gage  however  it can be changed here     Gage Type    You can assign a Gage to a Gage Type by clicking on the arrow at the right hand end of the field    Gage Details    File Menu  31    and selecting from the dropdown list  When you select a gage type you will be offered the option to  automatically fill many of the fields from the corresponding entries for the Gage Type  Gage Types  are created in the SETUP menu of the main program window     Serial Number  The    Serial Number    is the unique serial number assigned to the gage by the manufacturer     Company    This field can be used to store a serial n
26.  Include Atmospheric Information    checkbox  the    Calibration Study    window  will include atmospheric fields such as temperature  humidity  and barometer     Define Gage Types    The DataLyzer Spectrum Gage Management System allows gage types to be defined as a typing  aid during gage creation  That means when you are creating a gage and declare it to be of a  particular type  many fields are automatically filled in from the information stored in the common  gage type area  Beyond that  optionally  GR amp R results from one gage can be applied to all gages of  the same type  check with your auditor before establishing this procedure     Setup    Setup Menu   s      To create a gage type  select the DEFINE GAGE TYPES item from the SETUP menu  The following     Define Gage Types    catalog window will appear          Define Gage Types       Gage Type        Caliper  Indicator  Master  Micrometer        Delete    Exit    ree                This window contains a catalog of all defined gage types  The    Master    gage type is automatically  created in the software  Master type gages are used as masters to calibrate other gages  Masters  typically do not undergo GR amp R studies  Since they are highly precise they are usually more  expensive than in process type gaging and therefore inventorying them more frequently is wise  management  Master gages are usually traceable back to an off site measuring standard  i e  a  national or international standard unit of measurement   
27.  Print                   Performance Curve Tab  The purpose of a gage performance curve is to determine the probability of a measurement system    Gage Studies    Data Menu    43      either accepting or rejecting parts improperly using data from Long Variable and Attribute gage  studies  This tab applies to long and short variable studies and long attribute studies     Once the amount of gage error has been determined via a gage study  it is possible to calculate the  probability of accepting a part of some reference value when using the gage  To accomplish this  the  assumption is made that repeatability and reproducibility are normally distributed with some  variance  s2    The gage performance curve provides a graphical method of communicating how well the  measurement system is discriminating good parts from bad parts as the measurements approach  the decision points at the upper and lower specs        GRER Average  amp  Range Method    Average Charts Individuals Charts Range  amp  Error Charts  Statistics   Data J    Performance Curve           Performance Curve  GRR   6  Bias      l 0  00 BERN  2 o 2 e    2 o e    Gi S 5   S a 5  ce Part Measurements    Probability ofAcceptance  A 15  es  vo B      4       5    o  h       o       a  re        5  Referen       FA Print                Pin   Ext            The graph contains x  and y  axes  The x axis along the bottom contains values as reflected by the  measurement system  The y axis relates to the percent confidence  Sp
28.  The average of all the measurements for the calibration point    Bias    calculated    Average minus Nominal    Meas U    calculated   Uncertainty of measurements   2 times the standard deviation of  measurements for the calibration point    Master U   This column is for any additional uncertainty you might know about  uncertainty of  Nominal value for example     Status    calculated   Not set  dark grey   some values for the calibration point are missing    Fail if Bias plus Total Uncertainty is greater than positive tolerance OR Bias minus Total  Uncertainty is lower than negative tolerance     Pass if Bias plus Total Uncertainty is less than positive tolerance AND Bias minus Total  Uncertainty is greater than negative tolerance     Total Uncertainty   Square root of  Meas U squared plus Master U squared   If the Meas U  or Master U column is not showing in the table  the relevant value is taken as zero     Note  Status for the whole study is Fail if any calibration point in any calibration dataset is  Fail  Status for the study is Pass if all calibration points in all calibration datasets are Pass  If  data is incomplete and no calibration points are Fail then the status for the study is Not Set     Graph  Click the Graph button with the mouse and hold down the left mouse key  This shows a simple  graphic of the results for each calibration point       e   DataLyzer Spectrum Gage Management    5 5    Study Details Outside Diameter      Title  Outside Diameter A O o 9  02
29.  When the DataLyzer Spectrum Gage Management System is run  the user can be warned if a gage  is due for calibration or a defined GRR study is due to be repeated  This feature is enabled in the  Preferences window     Basic Steps for Use    Basic steps for calibration studies     1  Define Gage  same definition as used for GR amp R studies    2  Define master  if master reference blocks are used    3  Enter calibration data   4  Print report s    governed and enhanced by certain preferences and options    Basic steps for gage repeatability and reproducibility  GR amp R  studies     1  Optional   Define Gage  assuming it was not already defined for calibration studies above    2  Optional     Define GR amp R Study Definition and define preferences for studies    3  Create a new GRR Study  choose a Study Definition or set parameters for the individual study   4  Enter gage R amp R data   5  Print reports     The Setup menu allows you to define optional conveniences for use  throughout the DataLyzer Spectrum Gage Management System     e The DEFINE GAGE TYPES item can be used to create a prototype gage of a given type     Introduction 11    Afterwards  when you create a new gage and indicate that it is a specific gage type  fields in the     Gage Description    window can be automatically completed based on the information entered for  the gage type    The DEFINE MASTERS item allows you to define master or    known    gage blocks to use as a  measurement reference in calibration 
30.  amp  Name  PartOne Gage Name  0 1 caliper Study Date  5 16 2008  Characteristic  varcharl Gage No   8876M09 Performed By  KRE  Specifications  0 10 Gage Type  caliper   User Field 1    From data sheet       2 0000 Ramer   0 2000 Rp   0 8889  Measurement Unit Analysis   Total Variation  TY   Repeatability   Equipment Variation  EV    Trias   Kaf 0 00 eve     EV RAxK   20000 305 2 ae   100  6 1000   6 2577     6 1000 f    4 2 50   97 32     Reproducibility   Appraiser Variation  AV    Appraisers   Ka   ZAV  100 AV   TV       0 0000 r   Number of Trials    N   UR  EV  2 3 65 ae  3 270 a     Y  0 2000 2702    6 100027 10x3  A  20   A  Repeatability  amp  Reproducibility  R amp A    Pans        R amp R   100  R amp R   TV   R amp R   Y EV    AV    6 1000    0 0000       6 1000   100  6 1000   6 2677      Part Variation  P      97 32   PV   RpxK   0 9989x 1 62    1 4400 ZPV  100 PV TV       100  1 4400   6 2677       22 98   Y 6 10007   1 44002      6 2677    Total Variation  TV     TV   Y R amp R    Py           All calculations are based upon predicting 5 15 sigma  99 0  of the area under the normal distribution curve   K1 is 5 15 d2  where d2 is dependent on the number of trials  m  and the number of parts times the number of appraisers  g  which is  lassumed to be greater than 15    AV   If a negative value is calculated under the square root sign  the appraiser variation  AV  defaults to zero  0    K2 is 5 15 d2 where d2 is dependent on the number of appraisers  m  and  g  is
31.  and sends you back to the gage setup window     Use fixed length lines   Length   To use the    use fixed length lines    option  mark the check box and set the number of characters  allowed in each line to be accepted from the buffer  This option is only used when the strings have  no discernible line terminator  This setting supersedes the line terminator designation on the  opening gage setup screen     Accept short input lines   The    accept short lines    option allows the program to attempt to process gage readings even when  the defined template extends beyond the end of the string received  If the template has   signs in  positions 1 through 8 and the transmission received is only 4 characters long  normally this  transmission would be discarded as an error  However  with this button selected  the program will    122 DataLyzer Spectrum Gage Management    attempt to evaluate the characters received     Test Reading Timeout  The    test reading time out    feature allows setting the number of seconds to accept data during use of  the communication test window  Seven seconds is the default     Seconds before timeout   The    Seconds before timeout    option is provided for gages that may take more than seven seconds  to run a test  where the gage is across the room  or any scenario where it is difficult to send a  sample reading in the default time     Delay between lines 1 10 second   When a gage sends multiple lines  the program assumes that they will be sent one aft
32.  any order     The    By Trial    method can be used with long or short variable and short attribute studies  Using this  method  all entries are made in one appraiser trial at a time  column by column   Data entry is  organized by trial and must be entered in the order prompted by the system     The    Randomized    method can be used with long or short variable and short attribute studies  For  valid results  this method must be used for long variable studies using the ANOVA calculation  method     Data Entry Method and Data Spreadsheet   All Gage R amp R studies should be done blind  i e  organize the samples but do not mark them  so that  on subsequent trials  the appraiser can not keep track of previous measurements on the same  samples  Keep samples organized out of sight of the appraiser  so a specific pattern of trials can be  followed     The selection of data entry method  study type  calculation method  for attribute gages   number of  appraisers  trials  and parts will affect how the data spreadsheet will appear     When using    Randomized        By Trial     or    Open sheet    methods  once data entry is complete  data  can be viewed in the    data    tab under ANALYZE  ANALYSIS TOOLS in the    Gage Repeatability  amp   Reproducibility Study    window or data can be viewed after calculations are complete by saving and  exiting the gage study and then reviewing the study by using the EDIT REVIEW STUDY button in  the    GR amp R History    window  If data needs t
33.  define the study     Location  The    Location    column shows the location where the study was performed     Master Value  The    Master Value    column shows master value of the standard being measured     Average  The    Average    column shows the average value of the readings taken by the gage     Bias      28   DataLyzer Spectrum Gage Management    The    Bias    column shows the difference between the master value and the observed average     New Study Button  The NEW STUDY button creates a new study  An empty form is displayed ready for data entry  See  the    Bias Data Window    description for more information regarding data entry into this form     Edit Review Study Button   The EDIT REVIEW button opens an existing study for editing or review  Select a study by clicking in  its study date field  Its row will highlight to show it is selected  Press the EDIT REVIEW button to  open the selected study  See the    Bias Data Window    description for more information regarding the  study window     Exit Button  The EXIT button closes the study history window and retums to the main DataLyzer Gages window               DataLyzer Spectrum Gage Management  4 Gage Details    File Menu     Select items from the File menu to create gages  delete gages and edit gage details  You can also  create  delete and edit GR amp R Study Definitions from the File Menu     4 1 Gage Description    If you select New Gage or Open Gage from the File menu the Gage Description window opens      
34.  display when typed  When using literals  enclose them in quotation  marks  e g     SEND         P    is a special character when used outside quotation marks that indicates a  one second pause  lt can be used between commands sent to the gage to allow the gage enough  time to process each command     Transmission to Prompt Gage for Reading   The    transmission to prompt gage for reading    field is used only when a gage interface requires a  particular set of characters to be transmitted to it before it will transmit each measurement  Enter  the necessary set of characters in this field  When a gage driver is assigned that contains a     transmission to prompt gage for reading     during data entry a small button will appear in the data  entry window  This button will contain a label of    Gage1        Gage 2     etc  These labels correspond to  the com port to which the associated gage driver is assigned  Press the screen button to cause a  measurement to be transmitted by the gage     You can use ASCII characters or literals as entries for this field  When using ASCII characters  enter  the associated ASCII number  separated by commas  for any characters such as carriage return and  line feed  e g  13  10  that do not display when typed  When using literals  enclose the word in  quotation marks  e g     SEND         P    is a special character that  when used outside quotation marks   indicates a one second pause  lt can be used between commands sent to the gage to allow 
35.  first create and set up your ODBC  connections  Look in the    SQL Docs    folder under the program folder for installation documents     CreatingSpectrumDBSQLSeneer or CreatingSpectrumDBSQLServer2K  SettingUpSpectrumWithSQLServer    CreatingSpectrumDB Oraclev8 or CreatingSpectrumDBOraclev9  SettingUpSpectrumWithOracle or 94SettingUpSpectrumWithOracle    Follow the instructions in these documents to create and set up your ODBC connections  then  install the Gage Management System software using the instructions below     Installation Instructions   Insert the DataLyzer Spectrum Gage Management CD into your CD drive  The    Setup    program will  run automatically after a few seconds  if not  run the setup exe program from the CD   The    Setup     program will prompt you through the installation procedure  Almost all users should use the    Typical     option for installation     Index   A     Additional Reports 85     B     Basic Concepts 14  Bias History 27  Bias studies 19  Bias Study 67     C     Calibration 18   Calibration  amp  GR amp R Due 82  Calibration Certificate 81  Calibration Dataset Design 65  Calibration History 23  Calibration Label 64  Calibration Study 59   Com Port Assignments 126     D     Define Gage Types 92  Define GR amp R Characteristic Groups 99  Define Masters 96    Ge    Gage Catalog 22   Gage Description 30   Gage Interface Template 105   Gage Locations 104   Gage RR 18   Gage R amp R   Performance Report 80  Gage Repeatability  amp  Reproducib
36.  for all gages   See the  header fields  description here for additional information     File Menu   Save or  F2   SAVE  or pressing  F2  from outside the menu  saves the changes made to the gage description  form     File Menu   Save  amp  Exit or  Ctrl   S   SAVE 8 EXIT  or pressing  Ctrl   S  from outside the menu  saves the changes made to the gage  description form and exits the gage description window     File Menu   Exit or  Ctrl   X    EXIT  or pressing  Ctrl   X  from outside the menu  exits the gage description window  If you have  made change to any of the fields in the window  you will be asked if you want to save the changes  before exiting     File Menu     Save as Another Gage  When you choose this menu option a new gage will be created with the current values in the fields   You must alter the Name ID  field to a unique name or number before selecting this option     GR amp R Menu   Instructions  Opens the Instructions Window for setting the instructions and attachments to be used for Gage  R amp R studies     GR amp R Menu   History  Opens the GR amp R History Window    Calibration Menu   Instructions  Opens the Instructions Window for setting the instructions and attachments to be used for  Calibration studies     Calibration Menu   History  Opens the Calibration History Window      a   DataLyzer Spectrum Gage Management    4 2 Measuring Instructions   Set    The Measuring Instructions windows is in Set mode if it is opened from the Gage Description window  
37.  in the list  To select a predefined supplier s  name from the list  double click on it or highlight it in the list and press the SELECT button  To  create a new supplier  type a supplier name in the    Company    field and press SELECT  The program  will notify you that the supplier does not exist and ask if you want to create it  Press YES  The     Gage Suppliers    setup window will then appear  see the description of this window   Once a supplier  is selected  it appears in the field  You can view more specific information about the supplier by  clicking on the field  a window containing read only information about the supplier will appear     Documentation     The    Documentation      field can be used to include the last invoice number used to purchase the  gage or other documentation such as a reference to the instrument user s manuals and their  location     Battery Type  The    Battery Type    field allows you to indicate what type of battery the gage requires  if any       1 and  2 Supplies Company Buttons   No manual entry is allowed in the     1 Supplies Company    or     2 Supplies Company    fields  For entry  in these fields  use the  1 SUPPLIES COMPANY and the  2 SUPPLIES COMPANY buttons to  open the    Suppliers    list to select or create a supplier  To select a predefined suppliers name from  the list  double click on it or highlight it in the list and press the SELECT button  Press CANCEL to  exit without selecting a supplier  To create a new supplier  ty
38.  is also available from the Reports menu          Calibration History  FirstGTFirstGage   Serial No    Study Date Performed By Calibration Due   Status      000001 07 04 2010 14 06 First Inspectorzz 07 05 2010 PASS             New Study   Edit Review Study   Exit         As studies are completed  results are stored in the system  Summaries are displayed in  spreadsheet form sorted by date time with the most recent on top  Each row across in the study  history window represents an individual study  Each column represents key information relating to  the study       24   DataLyzer Spectrum Gage Management    Number Column  This is a unique serial number automatically given to each study     Study Date Column    The    Study Date    column shows the date and time the study was performed     Performed By  The    Performed By    column shows the name or initials of the person who performed the study  This  information is taken directly from the study window as it entered the study     Calibration Due  This is the date when calibration must be performed again  The period between calibrations is set in  the Gage Description window     Status  The    Status    field shows if the study is Pass  Fail or Not Set     New Study Button and File Menu   New Study  not available from the Reports Menu   The NEW STUDY button or menu item creates a new study  An empty form is displayed ready for  data entry  See the    Calibration Study Window    description for more information regarding data e
39.  is an example of an ASCII character string that might be transmitted from a gage or gage  interface  In this example there are several elements  three spaces are transmitted at the beginning  of the string  shown as underlines      X      then a space  a          sign   another space  then the  measurement value  4 379   Each space these characters occupy in the line is referred to as a  column   The decimal point is counted as a character   So  the    X    measurement starts in column    9       Setup    Setup Menu  125    and ends in column    13     The sign is located in column    7     To get the    X measurement out of this  string automatically each time it is transmitted  overwrite the    X measurement value  4 379  with          signs and the sign with an    S     This way each time the transmission is sent by the gage  the  software will look through this template to pick up values from these column positions  Note  If the  sign is always transmitted right next to the measurement value with no spaces between  you can  use a         sign to overwrite the sign making it part of the measurement value itself     The    reading description    area reports column positions where measurement values and signs  reside  Zeros will appear in the fields of this area until the template has been established through the  testing procedure  As the         signs are entered in the template  the software picks up their column  positions and fills in the    column    fields automatic
40.  line  terminator  use the ASCII character 44 for the comma character     Using a line terminator is the most commonly used technique  however  there is an alternative that  can be used in certain situations  which is using fixed line length  described later      Lines Per Transmission Received from The Gage   The    lines per transmission received from the gage    field is used to enter the number of lines the  gage will send each time it transmits  See the gage s operating manual for this information  This  feature relates to the previously described line terminator item  Remember that the default line  termination character is carriage return     If it is not clear exactly how many lines are transmitted  it may be necessary to use the CREATE  TEMPLATE item in the TEST menu to take test transmission s  that displays the number of lines  being transmitted  To do this  set the number of lines to a large number such as    8    or    10    and read  the gage  You ll be able to see five lines at a time  By taking a test string and waiting for timeout  the  software will report how many lines were received from the gage  the maximum being the number of  lines set in this field   A field will appear next to the word    transmission    that will report how many  actual lines were received from the gage  then you can go back and adjust the number of lines per  transmission accordingly  Save the template  and then enter the actual number of lines received in  the gage template window
41.  menu  saves the changes made or data  entered to the study     File Menu   Save     Exit or  Ctrl   S   The SAVE  amp  EXIT item  or pressing  Ctrl   S  without opening the menu  saves the changes made  or data entered to the study and closes the study window     File Menu   Exit  The EXIT item closes the study window without saving any entered data or changes     Analyze Menu   Preview or  F5    The PREVIEW item  or pressing  F5  without opening the menu  will open a report preview window  for the current study  A full description of the report can be found in Reports Menu  You can PRINT  or EXIT the report from the FILE menu in this window       s   DataLyzer Spectrum Gage Management    Analyze Menu   Analysis Tools or  F6   This opens the Analysis Tools window     5 3  GR amp R Report Definition    This window appears if you attempt to print from the GR amp R Analysis Tools window and none of the  tabs have the Print checkbox checked        Gage R amp  amp R Report Definition            Data Table    Graphics     Average Plot Charts    Individuals Charts     Range  amp  Error Charts      Performance Curve    Print Select All Exit                Select the aspects of the report you wish to print  You will build a report by checking all items to be  included in the report  Each check box will cause an associated set of information up to 1 2 page in  size to be included in the printed report  Therefore  if all the sections are selected  the report will be  about three pages 
42.  of Trials     and    Number of Parts    fields  The  following list contains recommended values for the fields for each type of GR amp R study  Some values  are set and locked by the software  as indicated to force the study to conform to generally accepted  rules                      38   DataLyzer Spectrum Gage Management  a Variable Gage Study Attribute Gage Study    Copas  Tae   Fans  sopas  Trae   rm      Number of Appraisers   Enter the number of operators to be included in the study in the    Number of Appraisers    field  In  certain study types  the number of appraisers is set and locked by the software  See the table above  for the number of appraisers recommended or allowed for the different study types   Study validity  depends on these appraisers being the same individuals who ordinarily measure production parts         Number of Parts   Enter the number of sample parts to be measured in the    Number of Parts    field  In certain study  types  the number of parts is set and locked by the software  See the table above for the number of  parts recommended or allowed for the different study types     Number of Trials   Enter the number of trials to be included in the study in the    Number of Trials    field     Trial    refers to  the number of times a set of samples are measured  In certain study types  the number of trials is  set and locked by the software  See the table above for the number of trials recommended or allowed  for the different study types   Stu
43.  on initialization   If a gage sends initialization information on start up  it will need to be cleared from the buffer or it will  be treated as a reading transmission  and erroneous readings will be displayed on the screen   Select this option to clear this information before starting normal gage entries     Clear buffer after each reading   This facility was originally developed for Microvu optical comparators  It allows one reading or part of  a string to be accepted and processed then the buffer is cleared and another string can be accepted  without portions of the previous string remaining in the buffer     Turn debug mode on   This mode intercepts gage transmissions and reports back the trigger that caused the transmission  and the contents of the buffer   See also RS 232 options window and Appendix K for event codes  that may be returned      Setup    Setup Menu  123    Special Modes   The GagePort TM  Computer mode check box allows the DataLyzer Spectrum Gage Management  software to send special codes to the GagePort to pull data from it on demand  This mode causes a     gage    button to appear on the data entry window  The button will be numbered corresponding to the  computer s com port number where it is connected  This GagePort Computer Mode is usually used  in combination with the multichannel mode  This allows the gage button to access each channel of  the GagePort independently  E g  for GP channel    A     you set the multi channel prompt to send the  GP    
44.  or  mask mentioned earlier    4  Once the driver is created  you must assign it permanently to the com port where the instrument  is connected  Com port assignments during template tests are temporary     DataLyzer International  Inc  always recommends purchasing the RS 232 connecting cable directly  from the gage or interface manufacturer to eliminate any cable problems  The RS 232 cable between  the gage and the computer must be pinned correctly  If the individual conductors inside the RS   232cable are not connected to the correct pins at each end  communication will be impossible and  valuable time will be wasted during system implementation     Note about USB communication support    The DataLyzer system supports RS 232 communication for up to four com ports  com 1 and 2 along  with two virtual com ports for extended serial communication  The virtual com ports are often used for  supporting USB to RS 232 conversion cable systems where separate software is installed to make a  USB port emulate a com port or for metaframe serial communication under systems like Citrix or  Microsoft s Terminal Services  Gage interfaces that are already completely USB compatible require  no RS 232 conversion and therefore do not require the use of the gage template system  e g   Mitutoyo 264 012 USB single gage interface   In cases of USB compatible interfaces  the gage  assignments for the two virtual comports should be set to    no gage    because communication is  automatically built int
45.  results  or the gage can be re evaluated using a more rigorous long form attribute study  or the gage  should be repaired or replaced     Gage Studies    Data Menu  53    Number of Appraisers   Enter the number of operators to be included in the study in the    Number of Appraisers    field  In  certain study types  the number of appraisers is set and locked by the software  See the table above  for the number of appraisers recommended or allowed for the different study types   Study validity  depends on these appraisers being the same individuals who ordinarily measure production parts      Number of Parts   Enter the number of sample parts to be measured in the    Number of Parts    field  In certain study  types  the number of parts is set and locked by the software  See the table above for the number of  parts recommended or allowed for the different study types     Number of Trials   Enter the number of trials to be included in the study in the    Number of Trials    field     Trial    refers to  the number of times a set of samples are measured  In certain study types  the number of trials is  set and locked by the software  See the table above for the number of trials recommended or allowed  for the different study types   Study validity depends on these trials being conducted blindly so that  appraisers do not know results from previous associated trials      Data Entry Method  The    Open Sheet    method can be used with all studies  The user can enter study data in
46.  the changes made or data  entered to the study     File Menu   Save  amp  Exit or  Ctrl   S   The SAVE  amp  EXIT item  or pressing  Ctrl   S  without opening the menu  saves the changes made  or data entered to the study and closes the study window     File Menu   Exit  The EXIT item closes the study window without saving any entered data or changes     Reports Menu   Print or  F6   The PRINT item  or pressing  F6  without opening the menu  will print the report        DataLyzer Spectrum Gage Management    6    Reports    Reports Menu     The following procedure is used for printing Gage R amp R study  Gage Calibration  certificate of  calibration   and Performance reports     Select the type of report to be printed  GAGE R amp R STUDY  GAGE CALIBRATION   PERFORMANCE  from the REPORTS menu  The    Gage catalog    window will then appear  select  the desired gage from the gage catalog     After you have selected a gage  a    history    window will appear containing all studies performed for  the report type selected  The studies will be sorted by date and time with the most recent on top   If  no studies exist in the list  no studies have been performed for this gage      The history window that appears for printing GR amp R studies and performance reports looks like this                    GRER History  0 1 caliper   File  Study Date Type Performed By Part Characteristic Ev Ay TV Method  5 29 2008 13 22  Ecua Gage TOL PartOne varcharl 1 0 94 71 ANOVA  5 28 2008 15 41   Actu
47. 0 57 0 72 0 60  Average  0 61 0 61 0 70 0 72 0 56 0 71 0 62 0 68 0 60  Range  0 08 0 08 0 20 0 46 0 08 0 48 0 10 0 10 0 07  Appraiser  2   Trial    1 0 55 1 00 0 68 0 58 0 54 0 64 0 53 0 81 1 00  Trial    2 0 52 1 00 0 57 0 56 0 60 0 61 0 61 0 57 1 00  Trial    3 0 60 1 00 0 63 0 65 0 58 0 59 0 60 0 58 1 00  Average  0 56 1 00 0 63 0 60 0 57 0 61 0 58 0 65 1 00  Range  0 08 0 00 0 11 0 09 0 06 0 05 0 08 0 24 0 00  Appraiser  3   4          The data table can be printed using the PRINT option in the FILE menu  Close the window by using  the FILE  EXIT menu item  To copy from this window  highlight the desired cells and press  Ctrl   C   on the keyboard and then press  Ctrl   V  to paste into a spreadsheet or other document     DataLyzer Spectrum Gage Management    6 2    Gage R amp R   Long Variables   ANOVA    The    average plot    tab applies only to ANOVA studies     The purpose of the    averages plot    graph is to show potential differences in reproducibility between  operator averages for the same part  Option buttons designate which one of three graphical analyses  will be displayed  They include  runs chart  averages chart  and X Y plot        5 GREER ANOVA Method  Average Plot       Statistics          a a a  Appraiser 1 Appraiser 2 Appraiser 3    7 08    Appraiser Averages  f Ea a  N     N       w  a           J Print                   Reports    Reports Menu    77      6 3 Gage R amp R   Short Variables    The preview window for short variable GR amp R studies don
48. 0 8  0 7  0 6  2 0 5  0 4  0 3  0 2  0 1       0 0                0 427  0 434  0 455  0 462  0 469    0 441  0 448      oO Lo N D o  2       e P  in  DQ D 2 oa a  mm ise  tiz  m mm        o o     o o o ao g   mim          Eight parts must be selected each with a unique size  The largest and smallest can pass 100  and  fail 100  but the other parts must not pass the gage system 100  The gage must sometimes  accept and other times reject these parts during the 20 trials     The probability of accepting each size is calculated     a  number of accepted parts from the 20 measured  m  number of parts measured of that size    m     a  5     m      Probability of acceptance  if a m  lt   5 a  0  These are the values in the P column         Probability of acceptance  if a m  gt   5 a   20  These are the values in the P column     5   Probability of acceptance  if a m    5 These are the values in the P column    Reports    Reports Menu           6 5     X  atP    995     X  at P    005   Bias  Bias   Lower Limit     X  at P  5     Repeatability         To calculate whether the bias is significantly different than zero  _ 31 3 x Bias  Repeatability    Compare this result with the critical value in t distribution table for  025 or  975 and number of  samples you are testing minus 1  degrees of freedom   Ifthe calculated t value for this study is  greater than the critical t value found in the table  then bias in this study is significant     Gage R amp R   Short Attribute    The previe
49. 20001 length ref   Serial Number   25mm 2mm  Manufacturer    Mitutoyo Belgium NY  Calibration    Eaim       s   S  Recalibration Interval  fasso Days  Date of Purchase   11 09 1992 Cost   280 00  Status  Location    Documentation       Supplies Company      Other Supplies       Notes            Set bestaat uit volgende lengtes   2 5mm   890791  5 1mm   890204  7 7mm   900051    Last Calibration Date   27 09 2005 09   curent    Name ID Number  The    Name ID    field is automatically filled in with the name entered in the previous    Gage Catalog     window  The name can be edited or changed     Serial Number  In the    Serial Number    field  enter the serial number of the master  if it exists  for traceability  purposes     Manufacturer Button and Field  No manual entry is allowed in the    Manufacturer    field  For entry in this field  use the    Setup    Setup Menu    er      MANUFACTURER button to open the    Suppliers    list to select or create a manufacturer  Note that  the    manufacturer    option button is selected in this window  allowing predefined suppliers to be  selected from a list  To select a manufacturer from the list  double click it or highlight it in the list  and press the SELECT button  Press CANCEL to exit without selecting a supplier  To add a new  manufacturer to the master list  type a supplier name in the    Company    field and press SELECT  The  program will notify you that the supplier does not exist and ask if you want to create it  Press YES
50. A        A    in this command represents the channel number on the GagePort  Checking the  GagePort Computer Mode box causes the program to send the other command required in the  GagePort s sequence automatically  You do not need to enter the    As    anywhere     RS 232 Options       5 RS 232 Options    Handshaking       geonsnnsnnsunennennnunenn     M ATS CTS  MW DSR DTR   XON XOFF    Communication Events    Reject all hardware triggers  C Ignore all hardware triggers    C Allow all hardware triggers    Framing Errors      Reject all framing errors       Ignore all framing errors    C Allow all framing errors    Save Cancel                The SAVE button stores the data and sends you back to the gage setup window  The CANCEL  button cancels any changes made and sends you back to the gage setup window     Handshaking   The standard handshaking uses RTS CTS and DSR DTR  If the owners manual of your instrument  indicates XON XOFF protocol  that can be used by checking the XON XOFF box and unchecking the  others  Refer to the owner   s manual for your gage for the specific communication handshake     Communication Events  The    Communication Events    are typically related to conditions with cabling       124   DataLyzer Spectrum Gage Management    7 12 5    Reject all hardware triggers  default    The    Reject all hardware triggers    option away all data in the com buffer if other than trigger 2 com  event is received  Trigger 2 com event means the acceptable number of cha
51. Definition    window looks like this                Gage Type Definition BAX  Fie GR amp R Calibration  Category  ts G  GAR Re evaluation Interval  Days Precision   Calibration Re evaluation Interval  Days Range        Manufacturer             1 Gage Supplier     2 Gage Supplier    Documentation      Battery Type        1 Supplies Company     2 Supplies Company                              Other Supplies                          Category  Use the    Variable    or    Attribute    option buttons to select whether the gage type is for variable or  attribute  go no go  gages     GRR Re evaluation Interval  Days   Enter the number of days between GR amp R studies in the    GRR Re evaluation Interval    field  This is  often dictated by the organization   s quality manual or procedures     Calibration Re evaluation Interval  Days   Enter the number of days between calibration studies in the    Calibration Re evaluation Interval    field   This is often dictated by the organization   s quality manual or procedures     Precision  Enter the smallest interval the gage can measure in the    Precision    field     Range  Enter the range between the smallest and the largest sizes measurable by the gage in the    Range     field     Manufacturer   1 Gage Supplier   2 Gage Supplier   1 Supplies Company   2  Supplies Company Buttons and Fields  No manual entry is allowed in the    Manufacturer         1 Gage Supplier         2 Gage Supplier         1    Setup    Setup Menu    as      Suppli
52. EV     Equipment Variation  field  represent the variation in the measurement process associated with the gaging equipment itself   presented as a percentage of Total Variation  TV      AV  Appraiser Variation    Appraiser variation  AV    gage reproducibility  The values in the    AV     Appraiser Variation  field  represent the variation in the measurements between personnel using the gaging system  presented  as a percentage of Total Variation  TV      RR  Repeatability and Reproducibility    Repeatability and Reproducibility is the combined variation due to EV and AV  presented as a  percentage of Total Variation  TV   The RR field will turn green for pass  or red for fail after the study  is completed  Set the    pass      level in the Preferences window     The total variation  TV  method used in the above calculation is set in the GR amp R Study     Method  The    method    column displays the method used for the study e g  long variable  short attribute  etc     New Study Button and File Menu   New Study   The NEW STUDY button or menu item creates a new study  You are first prompted to select a  Study Definition  If you want to start a GR amp R study that is not based on a Study Definition  select      none      An empty form then is displayed ready for data entry  See    Gage R amp R Study    section for  more information regarding data entry into this form     Edit Button and File Menu   The EDIT button or menu item opens an existing study for editing or review  Sel
53. FILE  PRINT menu item  Close the history window  by using the EXIT button or FILE  EXIT menu item     To copy from this preview window  highlight the desired cells and press  Ctrl   C  on the keyboard  and then press  Ctrl   V  to paste into a spreadsheet or other document     6 10 Supplier Summary    A supplier summary report is a list of all suppliers in the database along with attending information   The following procedure is used for printing a supplier summary report     Select the SUPPLIER SUMMARY item from the REPORTS menu  A scrollable preview report will be  displayed     It will look similar to this               amp  Supplier Summary Report Jog  File  Company Name  1234 Gage Company Type  Gage Supplies    Address  999 Main Street  Contact 1  Tina Thompson Title  Sales Phone  555 888 4848  Contact 2  Title  Phone   Company Name  ABC Gage Company Type  Gage Sales  Supplies  Technical Support  Address  1111 2nd Street  Contact 1  Bob Smith Title  Sales Phone  555 555 0002  Contact 2  Julie Miller Title  Support Phone  555 555 0001  4 N om                The suppliers will be sorted alphanumerically  Print or exit the report using the options in the FILE  menu     Reports    Reports Menu    s    6 11 Additional Reports    Several additional reports are available via the analyze menu in the Gage R amp R Study windows  under  the main program window s data menu   Most of these graphical tools relate to long variable and    ANOVA studies with the exception of the perform
54. Study Definition can be considered as representative  of all the characteristics in the group  If one or more GR amp R studies has been stored using the Study  Definition  the characteristic group will become    locked     This means that you can only change the  specified characteristic to another characteristic in the group     Ifthe program is not being used in conjunction with Datalyzer Spectrum  just type in Part and  Characteristic names  You can enter USL and LSL manually     Tolerance  USL  LSL  is one of the Total Variation  TV  methods  Specification limits are not  necessary in short attribute studies  and only one specification limit is needed for long attribute  studies    Saved Calculations   This frame is for specifying the type of calculations  ANOVA or Average and Range  and total  variation method that will be used for displaying results in the GR amp R History screen and reports after  it is saved  However  you can always view analysis based on ANOVA or Average and Range  with  any available total variation method  by selecting items in the Analyze menu     Average  amp  Range Method  Long and Short Variable Studies    The    Average  amp  Range    method is the simple method used by AIAG in its SPC manual  It provides  results for equipment variation  repeatability  and appraiser variation  reproducibility  only  It does not  express interactions between appraiser variation and equipment variation     ANOVA  Long Variable Studies    The    ANOVA     analys
55. Study Details Outside Diameter      Title  Outside Diameter  Master   First Master    Tolerance    02  Tolerance     02             Nominal   Meas1   Meas2   Meas3   Average   Bias   MeasU   MasterU   Status  Ela 2 001 2 005 2 01 2 002 2 0057 0 0047 0 0081  001  Br 3 98 3 99 3 981 3 986 3 9857 0 0057 0 009  002  Eva 6 03 6 028 6 033 6 027 6 0293  0 0007 0 0064  002  _4   8 02 8 025 8 017 8 022 8 0213 0 0013 0 0081  004  5 9 96 9 97 9 965 9 963 9 966 0 006 0 0072 0 013          Title    Enter a title for the dataset  This title will also appear on the tab     Tolerance   and       Enter the permitted positive and negative tolerances for the measurement mode  Calculated fields in  the data entry spreadsheet are displayed to two more decimal places than the number of decimal  places shown in the tolerance fields  For example  if you set a tolerance to    0 20    calculated fields  will display to four decimal points     Gage Studies    Data Menu    e      Master Button and Fields   To enter a master block set in the    Master Set    field  press the MASTER button and select a master  from the list that appears  The masters in this list are defined under SETUP  DEFINE MASTER in  the main program window  Text included in the Notes field for the master will be printed on the  calibration certificate     Table columns   Nominal   The reference or known value ofthe master block or characteristic used for the calibration  point   Meas   The gage measurement s    Average    calculated 
56. Suppliers    setup window without saving  select EXIT from the FILE menu or  press  Ctrl   X  without opening the menu     Gage Locations    The GAGE LOCATIONS item allows you to set up a list of locations that you can expect gages to  be found  Setting a location for a gage is done in the Gage Description window  The gage locations  in the list are only suggested locations  it is not it is not compulsory to choose a location from the  list  When you select the GAGE LOCATIONS item  the following window will appear        Gage Locations Ea    Location Name     mee ooo        Blow Molding  Cycle  amp  Barst  Domestic Post Molding  Domestic Winding    OK    Engineering  Everpure  Industrial Filtration Delete      Industrial Post Molding   Industrial Winding   Lamineer Department   Maintenance EM    Pomp Assembly   Pool Post Molding   Poolfilter Assembly   Preforming   04 Office   Quality lab El          The OK button opens the selected gage location description window  The DELETE button deletes  the selected gage location  The EXIT button closes the    Gage Locations    catalog window     Location Name  Type the name of a new or existing storage location in the    Location Name    field  or select an  existing location name from the list to open     Setup    Setup Menu    os      7 10 Gage Statuses and Inspectors    Choosing Gage Status or Inspectors from the Setup menu opens a window like this  With this  window you can create a new Gage Status or Inspector or delete an existin
57. _ DATALYZER     Ahnternati    G    Partners in Continuous Improvement    DataLyzer Spectrum Gage Management       2010 Datalyzer International  Inc          Cotes   3      Table of Contents    Foreword  Part I Introduction 8  PUrPose  u    eu keller eher 8  Special Fe  t  reS          2  2 ns a A ii 9  Basie Steps for Use iii IR 10  Part Il Explanations 14  1     Basic CO DIS aa 14  LOCURA 18  A A O 18  4 Bias StUqICS ici a ia 19  Part Ill General Windows 22  tT Gage Catalonia 22  2 Calibration  History iia 23  3    GR amp R HISOLY id 25   amp  Bias ISON  in ode iaa nd 27  Part IV Gage Details    File Menu  30  T Gage Description sararan AE E aan ads 30  2 Measuring Instructions   Set    cioininrnccnonnccci rica 34  3 GREER Study  Definitions sssini 36  Part V Gage Studies    Data Menu  40  1T OGR amp R Analysis TOO Sii aa en in an is 40  2  Gage R amp R Study ii ar HE are Erna NARESE 48  3 GR amp R Report Definition                        4u4unsnnnnnnnnnannnnnnnnnnnnnnnannnnnnnnnnnnnnnanennnnnnn nenn nenn nenn 58  A    Calibration Study idas 59  5 Calibration Label    seien 64  6 Calibration Dataset Design             uusssnssnnsnnnennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnsnnnnnnnsnnnnnnnsnnnnnnn nenn 65  7 Measuring Instructions   Read             z 44sns4s4nn0n nenn nnn nn en nnnnnn sure nenn snnenn nen snnnnsnnennnnen nenn 66  8  Bias MUAY   n 67    Part VI Reports    Reports Menu  72    4 DataLyzer Spectrum Gage Management       1 Gage R amp R   Long Variables     Average  am
58. aders    User Field  1 poo User Field  2  User Field  3  User Field  4 AAA User Field  5     User Field  6  User Field  7 poo User Field  8 po User Field  9  User Field  10   User Field  11  User Field  12             Save  amp  Exit   Exit         In the    User Fields 1 12     enter a field label  Examples of field names may include such items as       line        operation     or    manager in charge     You can create user fields 1 12 depending on your  company   s specific needs     Reports Tab  The    Reports    tab is used to set report preferences  You can add a logo  change report titles and  include your company name to report titles     Setup    Setup Menu    ss         System definitions          Identity      Reports       Logo Options  Logo width  mm     25 4    View Logo   Load Logo       Unload        Header Fields       Report Options       Include company logo on reports           EASCSLOGO BMP    Reports    IV Include Company Name        Instructions    Due Date    Signature line    Report User Name  GRASA AVERAGE  amp  RANGE REPORT  GR amp R ANOVA REPORT GR amp R ANOVA REPORT  GRR ATTRIBUTE REPORT GR amp R ATTRIBUTE REPORT  GR amp R BIAS REPORT GR amp R BIAS REPORT  CALIBRATION HISTORY REPORT CALIBRATION HISTORY REPORT  Certificate Of Calibration Certificate Of Calibration    M Include   I Include  I Include  F Include    M Include  TF Include   I Include  F Include    V Include  I Include   I Include  I Include  I Include X    F Include E Include         
59. age Type Serial Number Active Location Storage Location Due Date LastDate Active  1 Presioe2 cae 836787 Lab1 Storage2 6 20 2008 5 6 2008 Yes  2 attgage caliper 53334 Lab Storage 6 27 2008 5 28 2008 Yes  3  0 1 caliper caliper 8876M09 Lab Storage  7 10 2008 6 10 2008 Yes  4 gage Unknown Type 988547 Labl Storage2 New Gage NewGage Yes  5  gagel Unknown Type 0977MT7 Lab Storagel NewGage NewGage Yes  6   1 2caliper Unknown Type 14443222 Lab Storagel NewGage NewGage Yes                      The gages in the list will appear in descending order based on the due date  Each gage listed will  include the number name  type  serial number  active and storage locations and active status as  defined in the    gage description    window  The    last date    column shows the date a study was last  performed for the gage  The number of days entered in the    re evaluation intervals    fields in the    gage  description    window are added to the    last date    to determine the    due date    for the next study  Print  the list by using the PRINT option in the FILE menu     To copy from this preview window  highlight the desired cells and press  Ctrl   C  on the keyboard  and then press  Ctrl   V  to paste into a spreadsheet or other document     GR amp R and Calibration Histories    The following procedure is used for printing GR amp R and Calibration History reports     Select the desired report type  GR amp R HISTORY  CALIBRATION HISTORY  from the REPORTS  menu  The    Gage catalog    wind
60. age transmission  1  Starting column of reading in line     Ending column of reading in line     iil    Sign column of reading in line             108   DataLyzer Spectrum Gage Management    Port Settings   Refer to the user s manual of your RS 232 gage or multiplexing interface for information regarding its  baud rate  parity  data bits  and stop bits  Set the    port settings    fields to match  Defaults are  9600  baud  no parity  eight data bits  one stop bit     Ordinarily  gage communication works on a    push    system where a transmit button is pressed and  the gage sends a measurement  however  as DataLyzer products have matured  features have been  developed to handle certain specific situations     Transmission to Initialize Gage   The    transmission to initialize gage    field is only used when a gage interface requires a particular set  of characters to    wake it up    or when the gage must receive a specific command to set it into a  mode for sending data  This transmission from the computer to the gage is typically only required  once per session  Refer to the gage manufacturer s operating manual and enter the necessary set of  characters in this field  Ifthe gage requires no initialization message  leave this field blank     You can use ASCII characters or literals as entries for this field  When using ASCII characters  enter  the associated ASCII number  separated by commas  for any characters such as carriage return and  line feed  e g  13  10  that do not
61. al Gage 1 0 95 73 Long  5 28 2008 15 13   Actual Gage widget2000 weight 1 0 81 64 Long  5 16 2008 11 54 Actual Gage KRE PartOne varcharl 1 D 97 30 Long  5 16 2008 10 59  Actual Gage TOL 1 0 1213 Long  5 16 2008 10 31   Actual Gage PYM widget thickness 0 970 0 000 96 84 Long  5 14 2008 15 09 Actual Gage KRE widget2 length 0 000 0 350 34 89 Short  5 14 2008 14 53 Actual Gage KRE widget2 length 0 9500 0 0000 95 07 ANOVA  5 6 2008 11 21 Actual Gage MTL widget width 092 000 91 71 Long  Preview   Print                        The history window that appears for printing calibration reports looks like this        g    Calibration History  0 1 caliper  File          Study Date Performed By Master Set Status ey  5 16 2008 13 34 Pass             Preview   Print                           The spreadsheet areas of these windows are described in Chapter 4     Data     Select the desired study by clicking on it  You can then print the study using the PRINT button or  FILE  PRINT menu item  You can preview the study using the PREVIEW button or FILE  PREVIEW  menu item  Close the history window by using the EXIT button or FILE  EXIT menu item     Reports    Reports Menu    73      6 1 Gage R amp R     Long Variables     Average  amp  Range    The preview window for long variable GR amp R studies done using the    average  amp  range    calculation  method looks like this        GRER Report  File Variables       Stephen Computer Services  Inc     GR amp R AVERAGE  amp  RANGE REPORT    Part No  
62. ally     You must use the TEST  CREATE TEMPLATE area to cause the column fields to be filled in     7 12 6 Gage Template Example 2  Multi Line       Example reading  _ _ _X 4 379  _ _Y 3 465    In this example  each time the gage transmits  it sends an ASCII character string with two lines   Elements of the transmission include  three spaces at the beginning of the string  shown as  underlines      X      then a measurement value  An invisible carriage return line feed also exists at the  end of the first line  The second line includes three spaces     Y      and another measurement value   Each space these characters occupy in their respective lines is referred to as a column   The  decimal point is counted as a character   So  the    X    measurement starts in column    6    and ends in  column    10    of line    1     and the    Y    measurement starts in column    6    and ends in column    10    of line     2     To get the    X measurement out of this string automatically each time it is transmitted  overwrite  the    X measurement value  4 379  with         signs  Similarly  the    Y    measurement should be  overwritten with         signs  This way each time the transmission is sent by the gage  the software will  use this template to look in these line and column positions for input values     When templates of this type are used  the reading number of the transmission must be referenced in  the FUNCTION area of each characteristic  Characteristics looking for    X 
63. ance curve which also applies to long attribute  studies     These facilities are not available from the reports menu but are informational reporting type  resources  so we have referenced them here  These analyses are intended to provide different views  of data and calculations to reveal measurement problems more clearly     The following graphic example shows one of the analyses available along with the tabs to access  others  The    print    checkbox on each tab allows you to select that analysis for inclusion on the                     report   GRER Average  amp  Range Method     Statistics 1 Data Ri Pestormence Curve  Average Charts 1 Individuals Charts Y Range  amp  Eno Charts  Chart Type    C Run Chart       C XY Plot    Averages Chart    za  Appraiser 3          DataLyzer Spectrum Gage Management    7    7 1    Setup    Setup Menu     System Definitions  The SYSTEM DEFINITIONS item brings up a window that contains three different tabs  All three tabs  use the following buttons     Save 8  Exit Button  The SAVE  amp  EXIT button saves the information and closes the    System Definitions    window     Exit Button  The EXIT button closes the    System Definitions    window     Header Fields Tab    The    Header Fields    tab allows you to define up to 12 custom fields for use elsewhere in the  program  They appear when creating gages  gage types  masters  and in certain reports     System definitions xi    Identity Reports i File Viewers              Gage Description He
64. and accuracy studies    The DEFINE GR amp R CHARACTERISTIC GROUPS item links the DataLyzer Spectrum Gage  Management System with the DataLyzer Spectrum SPC program and allows you to create groups  of characteristics with similar specifications for easy measuring characteristic selection    Using SYSTEM DEFINITIONS  you can create custom fields that will be included in the    Gage  Description    window and on reports  These fields may be used to store and display additional  information that is not already defined in the DataLyzer Spectrum Gage Management System   You can also use this item to modify report titles  add a logo and or your company name to  reports    e Use PREFERENCES to set default settings for GR amp R studies    The SUPPLIER items allow you to manage the gage supplier entries  Create  edit or delete various  types of gage supply companies  Maintain information such as shipping address  costs  service  records  etc  for several categories of supplier  These records allow quick reference in case the  measurement instrument needs service or replacement    The GAGE LOCATIONS item allows you to define and catalog locations for gages    e The GAGE STATUSES item allows you to define statuses to apply to your gages    The Inspectors item allows you to list the inspectors that will supervise GR amp R and calibration  studies    The GAGE INTERFACE TEMPLATE item is used to create and test gage templates for your  specific gages    Use the COM PORT ASSIGNMENTS item to e
65. as the difference  between the master value and the observed average  The percent bias is defined as 100     bias au  process variation     Bias studies do not draw conclusions regarding pass fail like calibration studies  Because bias  studies use an average of observed measurements rather than the one observed measurement at  each calibration point  they may be considered more thorough than calibration studies  Bias studies  also consider only one master  one calibration point   while calibration studies often use multiple  masters at multiple calibration points  So  if a calibration fails at one calibration point  the bias at  that point it can be studied more thoroughly using a bias study  Bias studies can only be done on  variable type gages     The bias value from the most recent bias study is used in calculations for the Performance Curve of  the gage        22 DataLyzer Spectrum Gage Management    3    3 1    General Windows    The Gage Catalog and Study History windows are used in many places in the DataLyzer Spectrum  Gage Management System     Gage Catalog    When you create a new gage  need to select a gage  or delete one  the Gage Catalog window  appears              Gage Catalog x  Gage  m Filter Gage List      Serial Number     FirstGTFirstGage  FirstGTSecondGage Gage Type   SecondGTFirstG age mM   ThirdGTFirstGage First GT                 C Master  C Second GT  C Third GT       Status  falL E      Location    faLL bd               Gage Field   Enter the name o
66. ation of a process     Gage Repeatability   Does a measuring system report the same result when the same master is    blindly    measured  several times by the same person  Can one operator be consistent in measuring the same part   characteristic using the same gage  Repeatability is also referred to as equipment variation   because it often reflects on equipment design or condition     Repeatability    Explanations 15    Gage Reproducibility   Does a measuring system report the same results when two or three operators measure the same  part characteristic    blindly    several times  Between operator reproducibility is the other dimension of  variation in a measuring process  Gage reproducibility is also referred to as appraiser variation   because it most significantly reflects on operators    consistency in using the measurement system   Solutions to problems of reproducibility often involve operator training       Operator B      AN Operator C    Operator A            Reproducibility    Gage R amp R studies are usually made up oftwo components  equipment variation and appraiser  variation  These components serve to focus any remedial attention on repairing or improving the gage  itself  EV equipment variation  or training operators to use the gage correctly or more consistently   AV appraiser variation   See report examples in the Reports descriptions for more information about  appraiser and equipment variation     Either total tolerance  USL LSL  or total process varia
67. by 5 15  When the Total Variation  TV  method is Tolerance  the study sigma values   after multiplication by 5 15  are compared with the total tolerance  USL LSL      6 times sigma represents 99 7  of the full spread of a    normal    curve  When 6 Sigma is selected in  this frame  deviation values based on ANOVA calculations are multiplied by 6 before comparison  with other sigma values  No multiplication factor is applied to values calculated from Average and  Range  Because no multiplication factor is applied  when Tolerance is selected as the Total  Variation method  the Tolerance value is divided by 6    USL LSL    6     Note that the AIAG MSA Manual shows calculations based on 5 15 Sigma in the first two editions   The Third Edition used 6 times Sigma for ANOVA and 1 times Sigma for Average and Range     Alert if calibration or GR amp R study is due in     days  Ifthis box is checked a message will appear when the program starts to warn if a calibration study  or GR amp R study is overdue or is due within the specified number of days     Passing   Error Limit  Enter the passing percent error limit for Gage R amp R studies in the    Passing   Error Limit    field     Use Appraiser Names   When you mark the    Use Appraiser Name as Labels    checkbox  the DataLyzer Spectrum Gage  Management software will use appraiser names rather than    appr      in the labels for the data entry  tables as well as on the reports     Include Atmospheric Information  When you check the   
68. calibration point  If this  number is greater than one they will be averaged to determine the gage s bias for the calibration  point     Use Note column  Ifyou check this box a column in the spreadsheet will be added for miscellaneous notes about the  calibration point  for example  the master blocks stacked to create the reference value      Use Measurement Uncertainty column  If the number of measurements for each calibration point is greater than 1  the Use Measurement      e   DataLyzer Spectrum Gage Management    Uncertainty column checkbox appears  If it is checked a column appears in the spreadsheet  showing the measurement uncertainty  This is a calculated column based on the variation of the  individual measurements  This uncertainty is taken into account when determining the status of the  calibration point  If there is no uncertainty column then pass or fail is calculated from a simple  average     Use Master Uncertainty column   If this checkbox is checked a Master Uncertainty column appears in the spreadsheet  The value in  this column is input by the user representing uncertainty of the reference value  This uncertainty is  taken into account when determining the status of the calibration point  If there is a Measurement  Uncertainty column and a Master Uncertainty column in the spreadsheet  then the values are  combined together to make a Total Uncertainty figure     5 7 Measuring Instructions   Read    The Measuring Instructions windows is in Read mode if it i
69. can be considered as representative of all the characteristics in the group  This mode is widely  employed  but should only be used with the approval of management and your auditors  This method  assumes that all gages of a particular type  Mitutoyo 0 1    micrometer for example  are identical  and  all characteristics in a characteristic group are identical  If the premise is acceptable  you can study  a specific gage and assume the study applies to all gages of that type and all characteristics in the  group  It is more cost effective because it requires much less extensive record keeping than dealing  with all gages and characteristics individually  However  this method is based on certain  assumptions that may not always be true and may not be acceptable to specific management and   or auditing teams  In any event  this method of generalizing results by gage type only applies to  gage R amp R studies  Calibrations continue to be done individually  so bias and inaccuracy can be  caught during calibration for individual gages     DataLyzer Spectrum Gage Management System Consolidation with  DataLyzer Spectrum SPC System   The DataLyzer Spectrum Gage Management System will operate alone or link to the DataLyzer  Spectrum SPC module  By default both programs are installed in the same folder and will use a  single database  therefore accessing common setup information so characteristic specifications and  gage R amp R results are shared between the applications     Study due  
70. cnnnncnnnnnnnerennnanerennnnnarennnanens 105  11   Gage  Interface Template iii nn been 105  12 Gage Template Window     ooccconnccnnncccnnncnonnnennnnnnnnnnnnnnnnnnrrnnnnnrnnrrrnnnernnnrrnnernnnnrrnnernnnnens 107  File 2a es ee here Heer p senken  Test Menu              Special Menu  Options Menu  Gage Template Example 1  Single Line  Single Measurement               nnusussunnnsannnnnnnnnnnnnnnnnnnnnnnnnnnnnnn 124  Gage Template Example 2  Multi Line                 u  z0us444044400nnn00nnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn 125  Gage Template Example 3  Multiple Measurements Per Line               usuusrss0nnnsannnnnnnnnnnnnnnnnnnnnnnnnnnnnn 126  13  Com Port Assignime ts  i           4   1 uc0 men a naar nee 126  14    Select Database         u      220 0 200 aa i   127  Part VIII Help    Help Menu  130  1  Help Documentation              n 2 u canada iaa andanada nenne neu nn nenne 130  2 DataLyzer Application Notes            ccccceece cece cece nese ee ee ee ee nese eee eeea nese essa eeae esse es nun nun ann ann 130    3 About GR amp R Spectrum              u r24u04404 RR 130    Contents 5    Part IX Appendices 134  1 Appendix A   Computer Requirement    oocoonccconnccconnconnncnonnncnnnrnnnnnnnnnnnnnnernnnncrnnarenanennns 134   2 Appendix B   Installation           ooococnconocnnononnnononnnoronnnoaronnnaronnnoronnnoronnnaronnraronnrarennnarenananes 134   137    Index         8   DataLyzer Spectrum Gage Management    1    1 1    Introduction 
71. dy Definitions can be set up  where a gage and DataLyzer characteristic are specified along with certain parameters of the study   If the specified gage in a GR amp R Study Definition is part of a Gage Type  and the Characteristic is  part of a Characteristic Group  the results can be considered as applying to all gages in the Type  and all characteristics in the Group  This will reduce the number of GR amp R studies required     Individual Gage History   This is the most thorough method  It involves creating a gage entry for each gage to be tracked and  for GR amp R studies associating that gage with one or more product characteristics it measures  This  method provides maximum traceability and the most accurate and specific information about gage  performance for each job  While this method offers the highest level of traceability  it is time  consuming and expensive to maintain     GR amp R using Study Definition   A GR amp R Study definition is a way to specify details for repeating Repeatability  amp  Reproducibility  tests of a gage and characteristic  part  combination on a regular baisi  If the gage in a Study  Definition is included in a gage type  then GR amp R studies performed using the Study Definition can      1   DataLyzer Spectrum Gage Management    1 3    be considered as representative of all the gages in the gage type  If the characteristic in a Study  Definition is part of a characteristic group  then GR amp R studies performed using the Study Definition  
72. dy validity depends on these trials being conducted blindly so that  appraisers do not know results from previous associated trials      Data Entry Method  The    Open Sheet    method can be used with all studies  Appraisers user can enter study data in any  order     The    By Trial    method can be used with long or short variable and short attribute studies  Using this  method  all entries are made in one appraiser trial at a time  column by column   Data entry is  organized by trial and must be entered in the order prompted by the system     The    Randomized    method can be used with long or short variable and short attribute studies  For  valid results  this method must be used for long variable studies using the ANOVA calculation  method          40   DataLyzer Spectrum Gage Management    5    5 1    Gage Studies    Data Menu     Selecting items in the Data menu allows you to create  delete or edit Calibration  GR amp R and Bias  studies    GR amp R Analysis Tools    Note  Most ofthe    Analyze    graphical tools relate to Long variable and ANOVA studies with the  exception ofthe Performance Curve which also applies to Long Attribute Studies  When you select  the ANALYSIS TOOLS item  or press  F5  without opening the menu   a window will appear  containing a set oftabs  These tabs contain different statistical or graphical analyses ofthe Gage  R amp R Study  You can preview information in these tabs and click the    Print    check box to include  them in a report  Som
73. e fields for each type of GR amp R study  Some values are set and  locked by the software  as indicated to force the study to conform to generally accepted rules                 Short Form Variable   The    Short Variable    study is the shortest and easiest variable GR amp R study  however  it only  provides an abbreviated picture of the measurement variability  It analyzes only appraiser variation   reproducibility  because the appraisers measure only one set of parts  one trial   then those results  are compared against each other  This study requires two appraisers to make one measurement    52 DataLyzer Spectrum Gage Management    each on the same characteristic for five parts  To begin the study  select five production parts at  random  or of sizes ranging through the specifications  and any two operator appraisers whose  regular job it is to measure the parts from the process  Each appraiser should measure each part  independently  so neither appraiser knows the results of the other s test  i e  a blind test   The  measurements are arranged in two columns so the results from each part are side by side     Short  Variable    studies are often preferred because they are quicker and easier than long variable studies   They save time for the operator and manager while still providing an indicator of the gage s  reproducibility  Short variable studies always use the    average 8 range    calculation method  Short  variable studies will only show appraiser variation  When th
74. e high and low values  measured in the trials  The average is represented by a dot on this whisker line  It is similar to the  normalized chart except the dot is located on the average and the connective line  whisker   emanates from there to the extreme values  high and low  for each appraiser s samples                 GRER Average  amp  Range Method E  Statistics Data Performance Curve  Average Charts Individuals Charts Range  amp  Error Charts  Chart Type  C Normalized G       C Scatter Plot    Part Averages   Whisker       Appraiser 1 Appraiser 2    Appraiser 3        Print             Print Exit             Scatter Plot    Using the    scatter plot    option  values of all readings for each part are plotted minus the grand  average  One might expect as many as three dots for each appraiser for each part  However  when  measurements are the same for a given appraiser on a given part  two or three measurements may    Gage Studies    Data Menu    ar      appear as one dot  This is a desirable condition showing repeatability of the appraiser                 GRER Average  amp  Range Method X   Statistics Data Performance Curve  Average Charts Individuals Charts Range  amp  Error Charts  Chart Type       C Normalized Whisker    Individuals Charts   Scatter Plot       Appraiser 1 Appraiser 2 Appraiser 3            Print       Print Exit                Range  amp  Error Charts Tab    Option buttons include  range chart and error chart  This tab only applies to long variable stud
75. e ofthese analysis tools are available in other forms in the reports menu but  this is the only place where you can select your own report components     For ANOVA variable studies the following tabbed sections are included  Statistics  Data  and  Average Plot     For Average and Range variable studies the following tabbed sections are included  Statistics  Data   Performance Curve  Average Charts  with run chart  averages chart  X Y Plot options   Individuals  Charts  with normalized  whisker  scatter plot options   Range  amp  Error Charts  with range chart  error  chart options      Each tab in the preview window is described below     On each of the tabbed sections  you will find these common items     Exit Button  The EXIT button closes the window and returns to the GR amp R study window     Print Checkbox    Mark the    print    checkbox to include the graph or information on the current tab in a report generated  by pressing the PRINT button     Print Button   Use the PRINT button to print a report  If you have marked the    print    checkbox on any of the tabbed  sections  any graphs and information on those marked tabs will be included on the report  If you  press the PRINT button without any checkboxes marked  the GR amp R Report Definition window will  appear which allows you to quickly select which report components you would like to include  This  unique area offers print options not available from the REPORTS menu in the main program window     Statistics Tab  T
76. e using the    average  amp  range    calculation  method looks like this        GRAR Report  Stephen Computer Services  Inc     GR amp R AVERAGE  amp  RANGE REPORT       Part No   amp  Name  wdget2 Gage Name  0 1 caliper Date  5 14 2008  Characteristic  length Gage No   3376M03 Performed By KRE  Specifications  5 000   0 100 Gage Type  caliper    Measurement Unit Analysis             Parts Appraiser A Appraiser B Range A   B   1 2 354 3 6857 1 332  2 3 654 4 974 1 32  3 1 688 3 7111 2 023  4 2 697 2 499 0 198  5 4 414 1 784 2 63  Average Range 1 501       GR amp R   6 49419                               Equipment variation  repeatability  is not allowed because only one trial per operator is involved   ANOVA analysis is not supported on short variables studies     Average Range   DR    2    GRER   Rid        WORER 100x  GR amp R    process standard deviation    DataLyzer Spectrum Gage Management    6 4    Gage R amp R   Long Attribute    The preview window for long attribute GR amp R studies looks like this              Gage Repeatability and Reproducibility Report X   File  Part No   amp  Name  PartOne Gage Name  alt gage Date  7722 2008  Characteristic  attcharl Gage No   53334 Performed by  JUL  Specifications  400   Infinite Gage Type     Appraiser 1  Status Pass  Part Size Accepts P  1 4 0 0 025  2  405 6 0 325  3 41 5 0 275  4  45 7 0 375  5 1 42 3 0 475  e  425 9 0 475  7 1 4 8 0 425  g   435 20 0 975          Performance Curve  Bias 0 022 GRR Val  0 053    1 0  0 9  
77. easurement system  whether total  process variation can absorb significant measurement variation or whether measurement variation  must be reduced and thereby reduce total process variation to an acceptable level  By performing  careful R amp R studies the user can draw conclusions about whether the measurement system can be  used as is  whether it must be replaced or can be improved sufficiently through a combination of  operator training and equipment repair     The program also offers an alternative graphical view relating GR amp R study results to a part   s  specification limits in the form of a performance curve  From the cur  the confidence interval for a  gage system can be determined  i e  the measurement range within which one may be 99   confident that a measurement system will be accepting or rejecting parts properly  The difference  between the confidence interval and the engineering specification limits should be applied as a  reduction in tolerance due to measurement error     Gage Inventory and Management Tools  Surveillance   The program stores hundreds of gages and provides several listing facilities to aid in inventory and  scheduling of regular measurement system evaluations  Supplier information containing shipping  address  product costs  service record  etc  can be stored and referenced later at the point of gage  replacement or purchase of supplies or parts     Special Features    For GR amp R analyses  gages can be studied individually or GR amp R Stu
78. easuring system s result at that size  For instance  in  the case above the gage being used needs to be replaced because at the lower specification limit   0 00  we can only be about 45  confident about making the right decision to accept or reject  product  worse than simply flipping a coin      Reports    Reports Menu  a      6 7    Discrimination required from the gaging system is a function of cost  of the gage itself  ofthe  products being measured and improperly scrapped  and of the personnel time required to do the  inspection     Calibration Certificate    There is no preview for the Calibration Certificate   Calculations used in the Calibration Certificate are   Bias   Average minus Nominal    Meas U   Uncertainty of measurements   2 times the standard deviation of measurements for the  calibration point    Status  Nothing  some values for the calibration point are missing    Fail if Bias plus Total Uncertainty is greater than positive tolerance OR Bias minus Total  Uncertainty is lower than negative tolerance     Pass if Bias plus Total Uncertainty is less than positive tolerance AND Bias minus Total  Uncertainty is greater than negative tolerance     Total Uncertainty   Square root of  Meas U squared plus Master U squared   If the Meas U  or Master U column is not showing in the table  the relevant value is taken as zero     Note  Status for the whole study is Fail if any calibration point in any calibration dataset is  Fail  Status for the study is Pass if all cal
79. ecification limits are shown  as vertical lines intersecting the x axis  Normal curve tails intersect these specs and represent  confidence indicators  At the intersection of either spec you can read the corresponding left or right  scale to see the confidence one has in the measuring system   s result at that size  For instance  in  the case above the gage being used needs to be replaced because at the lower specification limit   0 00  we can only be about 45  confident about making the right decision to accept or reject  product  worse than simply flipping a coin      Discrimination required from the gaging system is a function of cost  of the gage itself  ofthe  products being measured and improperly scrapped  and of the personnel time required to do the  inspection     The bias value from the most recent bias study is used in calculations for the performance curve     Average Charts Tab   The purpose of the    averages charts    analyses is to show potential differences in reproducibility  between operator averages for the same part  Option buttons designate which one of three graphical  analyses will be displayed printed  They include  runs chart  averages chart  and X Y plot     Run Chart  The    run chart    option displays data points that represent the average of each appraiser s trials for           DataLyzer Spectrum Gage Management    each part  Each appraiser s data points are displayed in a different color and shape  Lines of the  associated color connect each a
80. ect a study by  clicking in its study date field  Its row will highlight to show it is selected  Press the    Edit    button to  open the selected study  You can also select a study by double clicking it     3 4    General Windows  Exit Button and File Menu   Exit    The EXIT button closes the study history window and returns to the main program window     File Menu   Delete   The DELETE item will delete the selected study  Select a study by clicking in its study date field  Its  row will highlight to show it is selected  Select DELETE from the FILE menu to delete the selected  study  A window will appear to confirm that you want to delete the selected study     Bias History    The    Bias History    window looks like this        Bias History  0 1 caliper  Study Date Performed By Location Master Value Average Bias  6 10 2008 10 44 MTL 0 50       a   amp           CC EdivReview Study   Ext                           As studies are completed  results are stored in the system  Bias study summaries will be displayed  in spreadsheet form sorted by date time with the most recent on top  Each row represents an  individual study  Each column represents a key statistic or label relating to the study     Study Date Column  The    Study Date    column shows the date and time the study was performed     Performed By  The    Performed By    column shows the name or initials of the person who performed the study  This  information is taken directly from the study window as it was entered to
81. ed measurement and press OK  For attribute gages short form  enter     G    for Go or Accept and    N    for No Go or Not accept  The corresponding cell will turn green and the  actual value will be internally stored in the database  The value will be hidden so it will not cause    bias in later measurements     Open Sheet  The    Open Sheet    method can be used with long or short variable and short attribute studies  The       Open Sheet    entry spreadsheet looks something like this       s   DataLyzer Spectrum Gage Management      Gage Repeatability  amp  Reproducibility Study  x   File Edit Analyze Instructions  Performed By      Date  26 07 2010 Time   11 52    Dero o oo    r Characteristic    Part     LSL  ae  USL  p   Characteristic      r Saved calculations  Calculation Method   Average  amp  Range y  Total Variation Method   Study Variation y     Study Parameters   Variable   Long 3 Appraisers 10 Parts 3 Trials Data Entry Method   Open Sheet  Appr  1    Appr  2    Appr  3                                  Trial  3   Trial 1   Tral 2   Trial  3    Appr  1   Appr 1   Appr  1   Aen  2   Appr  2   Appr 2   Appr 3   Appr 3   Appr  3  Trial  1   Trial  2   Trial  3    Trial  1   Trial  2                   To use this method  enter values in the spreadsheet  in any order  until the form is complete  For  variable gages  enter the value for the specified measurement  For attribute gages short form  enter     G    for Go or Accept and    N    for No Go or Not accept     Open 
82. eld     Cancel Button      102   DataLyzer Spectrum Gage Management    The CANCEL button closes the    Suppliers    catalog and returns to the main DataLyzer Spectrum  Gage Management software window     7 8 Supplier Details    The following    Gage Suppliers    window is used to enter the details for a new supplier or edit existing  supplier information           Gage Suppliers  File   Company Information Company Type   Name    Gage Sales   eins   Gage Supplies   Address       Gage Calibration  IT   Gage Technical Support   City   SY   Gage Manufacturer   State   nn   Certifying Agency   oe l Certification   Country  OoOo O O Certifying Agency    Contact 1  IT   Title  Cert     A y  Phone  PA Espiration Date  ln  e Mail  A Document    o  Contact 2  no   Title  IT   Phone      e Mail  CY   Fax  fC   Website  r    Notes     Save Exit                   Company Information  Use the fields in the    Company Information    portion of the supplier window to enter general contact  information for the supplier     Company Type   If you selected a company type option button other than    all    in the previous    Suppliers    catalog when  creating a new supplier  this company type will be selected here  However  you can select as many  types that apply to the supplier being currently defined     Setup    Setup Menu    103      In certain areas of the program where the system accesses the    Suppliers    catalog  the catalog may  contain only suppliers of a particular    company type 
83. emplate    window  Ifthe parameters are set correctly and if the RS 232 cable  pinouts are correct  you should be able to receive a string of characters from the instrument  When  you click READ GAGE a seven second countdown starts  giving you time to press the button on  your instrument     send    or    print    or    transmit     that causes it to transmit a string of characters to the  computer  This string appears in the  transmission line      1    field  You will be asked to overwrite the  measurement number in the string with pound signs     to make a mask or template for future  transmissions from this instrument  Save the template and exit     The    Create Template    window looks like this              7         amp  Read sample data  Message  Serial port to test  dr Timeout   7 Clear  Transmission  Up   Dem   De   Column        H  HHEH    Read Gage Save Cancel                Serial Port to Test   Use the    serial port to test    area to temporarily designate a com port to use for the current test  This  selection will not be stored permanently  because the com port selection for ongoing use is made  from the COM PORT ASSIGNMENTS item in the SETUP menu     112 DataLyzer Spectrum Gage Management    Timeout  Use the    Timeout    field to enter the number of seconds the template will wait for an RS 232 string  from the instrument     Clear Button  The CLEAR button clears all transmission lines of previously set    Ps and extraneous characters that  may exist from p
84. emplate   The TEST TEMPLATE item allows you to test the template filter s  created in the    Create Template     window  If you test a template in the    Create Template    window  the test will overwrite the template   Ps and    S    signs  and you will have to replace them after each test  The    Test Template    window  not only confirms communications but also displays other information like the mode  reading  and  channel identifiers as well as any analog to digital  A D  conversions that are taking place     The    Test Template    window looks like this       114   DataLyzer Spectrum Gage Management             a Read sample data x   NER l Serial port to test  a    Timeout  E Clear     Transmission  Up   Doan   Deena   Column  BE       H  HHHHHHH    Read Gage Save Cancel             Serial Port   Use the    serial port to test    area to temporarily designate a com port to use for the current test  This  selection will not be stored permanently  because the com port selection for ongoing use is made  from the COM PORT ASSIGNMENTS item in the SETUP menu     Timeout  Use the    Timeout    field to enter the number of seconds the template will wait for an RS 232 string  from the instrument     Clear Button  The CLEAR button clears the test reading window     Message  The    Message    field will display operational comments that the software reports about the  transmission  Messages include  opening com port  send data now  communication timeout  etc     Filtered Readings
85. er another  immediately  However  some gages do some calculations between lines or have intermediate steps  that cause additional delays  If the program does not receive the complete transmission in a prompt  fashion it discards the entire transmission  By setting the delay higher the program will wait the  specified time between lines before giving up  Regardless of the delay specified  when the reading is  received the program will display it immediately  With a large value in this field  the program will  appear to be locked up once a transmission has started until either the transmission is completed or  until the specified delay has elapsed  The value entered is in tenths of seconds not seconds  When  using BAUD rates below 2400 it may be necessary to specify a small delay of up to 5 10 of a  second to account for the slow transfer rate  Another instance where this delay should be increased  from the default is when a whole report that is ordinarily printed is being redirected to the com port  for data acquisition     Buffer Size   The    buffer size    option boxes must also be set to control the size of each com port character buffer   Specific options are selectable as character buffer sizes  The default  1024  should be adequate for  almost all situations  Use larger buffer options only when pages of ASCII information are being  transferred through the com port with each transmission  If the buffer size is not large enough  data  will be lost     Special clear buffer
86. ers  i e  TAL Fowler software wedge   MicroRidge GageWay Il and direct USB interfaces   set the com port s  to    No Gage        Experience has proven that a high percentage of communication problems result from the use of an  incorrect RS 232 cable  We advise you to always buy RS 232 cables directly from the manufacturer  of the gage or interface     Advanced Concept  IRQ    Windows    by default  reserves two processor interrupts  IRQs   IRQ4 for com 1 and IRQ3 for com  2  With the advent of virtual com ports  different numbered com ports become available  usually  com5 and above   Separate software must usually be employed to establish the existence of these  additional com ports in the operating system  This allows the simultaneous use of more than two  RS 232 ports at a time     Select Database    This option is available only for the Microsoft   Access database platform  Use the windows browser  to locate and select the location of the desired database and press the    Open    button  You will be  prompted to restart the program so the different database can be accessed          130   DataLyzer Spectrum Gage Management    8    8 1    8 2    8 3    Help    Help Menu     Help Documentation    The HELP DOCUMENTATION item accesses the DataLyzer Gage Management help system  Use  the search facility to find all references to a topic you enter or move to the beginning of a general  topic by selecting it from the list of topics     Press F1 on the keyboard at any time to show t
87. ers do two things  they define the RS 232 communication environment in terms of baud  rate  parity  stop bits  etc   and they tell the DataLyzer Spectrum Gage Management package the  actual location of the measurement within the gage s transmission  The transmitted string may  contain information such as  a measurement value      sign  units of measure  serial number of  device  date and time stamp  etc  The order and contents of the transmitted string is different for  each instrument or interface  The purpose of the template is to define a mask to lay over the  transmitted string of characters  so that each time a string is received  the mask can pick out only a  specific set of characters  the measurement  from the string  Maybe the measurement number is  located in the 5th through the 12th space in the string of transmitted characters  That is where this  gage driver s template should be set to pick up valid values for data entry     Steps for gage driver creation and use     1  Configure the communication parameters to match the gage transmissions  This means the  software s baud rate  parity  data bits  and stop bits must all match the gage s settings  Refer to  the instrument s user s manual for this information    2  Enter any necessary commands that prompt the gage for readings  define line termination  characters and the number of lines to be transmitted in the string    3  Take a test transmission from the gage to verify communication and define the string template
88. es Company     or     2 Supplies Company    fields  For entry in these fields  use the associated  buttons to open the    Suppliers    list to select or create a gage supplier  Note that the applicable  option button  e g  manufacturer  gage supplies  is selected in this window  allowing only predefined  suppliers of that type to appear for selection  however  if you create a new supplier  you can specify  it to be any of the supplier types available in the system   To select a predefined suppliers name  from the list  double click it or highlight it in the list and press the SELECT button  Press CANCEL  to exit without selecting a supplier  To create a new supplier  type a supplier name in the    Company     field and press SELECT  The program will notify you that the supplier does not exist and ask if you  want to create it  Press YES  The    Gage Suppliers    setup window will then appear  see here for a  description of this window      Once a supplier is selected  it appears in the field  You can view more specific information about the  supplier by clicking on the field  a window containing read only information about the supplier will  appear     Documentation    The    Documentation  4    field can be used to locate the manual or  instructions for the gage    Battery Type  Enter the SKU or model number of the battery required by the gage  if any  in the    Battery Type     field     Other Supplies    Enter any other supplies required to maintain or use the gage in the 
89. es a one line per transmission gage setup with         signs in the gage  template  The multi channel facility is then accessed to tie a channel 1 D  to an individual part  characteristic     The window for setting up multi channel translation looks like this        Multi channel description       Channel information   64    Prompt     Identifier     Reading   ft Column        lt  lt Previous Save Cancel   Nexb gt     Channel Information  The    Channel information    area displays the current channel description number                 Prompt   Use the    Prompt    field to enter a message to be sent to the gage to request a reading from the  current channel  This field should not be used to request a reading from all channels  Any global  requests should be entered at the main gage setup screen     Identifier   Enter a set of characters in the    Identifier    field that uniquely identify the line containing the reading  for the current channel  This may be a label like     X axis     or    radius     or a channel number such as     01    or    A     It is important that you enter the characters in the exact column positions that they  occur in the gage transmission     Reading Number  The    Reading number    field allows for multiple readings per line to be handled  If only one  measurement is transmitted per line then enter    1    in this field     Column   The    Column    field does not allow input  It merely displays the cursor s position for the    identifier     fie
90. f the gage in the    Gage    field or select a gage from the list below  If you are creating  a new gage use a name that has not been used before  You cannot create a new gage with the  name of an existing gage     Filter Gage List frame  The frame on the right is for filtering the list of gages on the left  When creating a new gage this list  is for information only     Gage Type   Selecting a gage type in this catalog window will cause the window to refresh  showing only gages of  the specified type to appear  Marking the    ALL    check box causes the screen to refresh to display   all existing gages  Gage types are defined under SETUP in the main program window     General Windows 23    3 2    Status   If you select a status from the dropdown list  mouse click on the small arrow to reveal the list  only  gages with this status will show  Statuses are defined under SETUP in the main program window  A  gage is assigned a status in the Gage Description window under the FILE menu  To remove the filter  and show gages with all statuses  select ALL at the top of the list     Location  This filters the gage list in a similar way to Status     OK Button  Pressing the OK button selects the gage currently in the Gage field     Cancel Button  The CANCEL button exits to the DataLyzer Spectrum Gage Management main program window  without saving any changes     Calibration History    Select Enter Edit   Calibration from the Data menu to show the Calibration History window  This  window
91. for printing on the report by using the logo width field  Enter the  desired width in millimeters  100 mm max       Report Titles Table   The report titles table contains two columns     Report    column  default title  and    User Name    column   Use this area to re title any of the available reports  To rename a report  click or double click on the  name you want to change in the    User Name    column and enter the desired title     Identity Tab      90   DataLyzer Spectrum Gage Management    7 2    Company Name   Originally  the    Company Name    field will contain the name of the company to which DataLyzer  International  Inc  registered the software  The name in this field can be changed  Any changes  made to the company name will flow backward to change the official registration  software s owner of  record   The entry in this field is the one used for printing in the reports  if desired  see Reports tab      Note  In the event you wish to change the software s official assignment that appears in the HELP   ABOUT DATALYZER window  please contact DataLyzer International  Inc     File Viewers Tab       Identity   Reports   Header Fields afi File Viewers             Path to Microsoft Word Viewer Change Folder  Path to Microsoft Excel Viewer Change Folder  Path to Microsoft PowerPoint Viewer  Change Folder             Exit         Change Folder  Use the buttons to locate the executable   exe  files to use for viewing Microsoft Word documents   Excel spreadsheets and Powe
92. from the instrument     Message   The    Message    field allows DataLyzer Spectrum Gage Management software to report messages  about the transmission being received  Messages such as    reading complete    or    time out error    may  display in this field     Analog to Digital  The    Analog to Digital    area allows you to high and low master the gage and set the software to do  an automatic interpolation     The procedure     1  Enter the size measurement of your high master in the high master field    2  Enter the size measurement of your low master in the low master field    3  Position your cursor in the high gage reading field  then put your high master into the gage  Press  the MASTER HIGH button  If you are not using a gage prompt  you will need to press the  footswitch or send button on the gage  The value will transmit from the gage in terms of a number  of digital segments as discussed earlier and will display in the    high gage reading    field    4  Position your cursor in the low gage reading field  then put your low master into the gage  Press  the MASTER LOW button  If you are not using a gage prompt  you will need to press the  footswitch or send button on the gage  The value will transmit from the gage in terms of a number  of digital segments as discussed earlier and will display in the    low gage reading    field    5  Save the gage file template by using the SAVE button     This sets up an automatic interpolation relationship  The software does the f
93. g one     Create and Delete Gage Statuses    Ed          Active   Calibration externally Exit    Maintenance   New Gage   Rejected E            To create a new status or inspector  type in the name of the new status or inspector and click the  OK button     To delete a Status or Inspector click on the name in the list then click the DELETE button     The EXIT button closes the window     7 11 Gage Interface Template    Gage interface templates  or gage drivers  are created to allow direct communication between the  gage or gage interface and the DataLyzer Spectrum Gage Management software  They provide  specific information about the gage to ensure proper communication with DataLyzer Spectrum Gage  Management software  The template supports RS 232 communication for up to four com ports  com  1 and 2 along with two virtual com ports for extended serial communication  Gage interface  templates for the most commonly used gages and interfaces are provided with the DataLyzer  Spectrum Gage Management software  Other gage drivers can be created allowing the DataLyzer  Spectrum Gage Management software to operate with your specific gages  Once a gage driver is  setup  it can be used in the    Com Port Assignment    window     The information in this section describes the process of defining a gage driver that can then be used  by the DataLyzer Spectrum Gage Management software to control any ofthe computer s four com  ports       106   DataLyzer Spectrum Gage Management    Gage driv
94. he    statistics    tab contains a detailed listing of key statistics in the GR amp R study  This tab applies  to both Average   Range and ANOVA studies        ij  GR amp  amp R ANOYA Method    Statistics  SSS Average Plot    ANOVA Supporting Calculations  Source DF  s MS    Appraiser  2 jr fs   Parts B   fess   fasieo   ApprxPart fig foso fomss fosa  Error fo  27589  0 0460   Total fs   faen    m Total Variation    Study C Tolerance Process   when saved    Process   current 6 5364 6 Std Dev                       Results    Lower 90  CL 6Std  Dev  Upper 90  CL   of Variation   Contribution    Repeatability   Equipment Variation  EV  jf    1 2865 0 19   Reproducibility   Appraiser Variation  AV   5 4695  1 3698 fiseso9  Part x Appraiser Interaction foot  ooo  ooo  Repeatability te Reproducibility  R amp R   1 1997 fazas Jans  Part Variation  PY  faszo9  so   Ja    F Print                You can see the effects of different Total Variation methods by clicking on the options in the Total  Variation frame  The    Process    options are only available if a Datalyzer Characteristic was selected  for the study  With the    Process     when saved    option a calculation of the Standard Deviation of the  last one hundred data points was made when the GRR study was saved     Process     current    will  recalculate process variation with the most recent one hundred data points  One hundred data  points is irrespective of the subgroup size  for example 10 subgroups with subgroup size is
95. he help screen  If you have a window or dialogue box  open specific help for that window will be shown     DataLyzer Application Notes    Selecting this menu item takes you to the DataLyzer International website     About GR amp R Spectrum    ABOUT GR amp R SPECTRUM item provides some general information about the software and  installation  Information like version number  database type  ACC MS Access  SQL MS Sequel  Server and ORCL Oracle   program path and database path information is also included     The following window offers an example of what the    About DataLyzer   Spectrum    window looks  like     Help    Help Menu  131    About Datal yzer Spectrum Gaging Analysis xi      Datal yzer Gage  Version  0 30 12 DB Version  93 GRA DB Version  12  ACC                       Copyright 92001 by DataLyzer International  Inc     This program is licensed for       Serial    Demo Licensed copies  1    Program Path    TSCLIENTSD Work D atalyzer   sS ource D atalyzerG age a  Database Path     Tsclient d  Work D atalyzer   S ource D atalyzerG ages MarcD ata4ndLang22May       Warning  this computer program is protected by copyright law and international  treaties  Unauthorized reproduction or distribution of this program  or any portion of  it  may result in severe civil and criminal penalties  and will be prosecuted to the  maximum extent possible under the law        System Infol      Technical support     Phone   248  960 3535 Fax   248  960 3539 Email  support D atal yzer com  
96. her than the date and whether the  study is Pass or Fail     Performed By Field and Button  Enter the name or initials of the supervisor overseeing the test in the    Performed by    field  You can  type it in directly or select from a list by clicking the PERFORMED BY button     Gage Name  The    Gage Name    field displays the gage name  The contents of this field cannot be changed     Gage Type  The    Gage Type    field displays the gage type as specified in the gage   s    gage description    window  It  cannot be changed here     Date    Enter the date the study was initiated in the    Date    field  The current system date is automatically  entered into this field  but another date can be entered if desired     Time      e   DataLyzer Spectrum Gage Management    Enter the time the study was initiated in the    Time    field  The current system time is automatically  entered into this field  but another time can be entered if desired     Serial Number  The    Serial Number    field is automatically filled in based on the serial number entered in the gage s     gage description    window  lt cannot be changed in this window     Company Number  The    Company Number    field is automatically filled in based on the company number entered in the  gage s    gage description    window  lt cannot be changed in this window     Status Field and Set button    In simple mode the status Set button is available to manually set the status  Three statuses are  available     Pass      
97. ibration points in all calibration datasets are Pass  If  data is incomplete and no calibration points are Fail then the status for the study is Not Set       82   DataLyzer Spectrum Gage Management    6 8 Calibration 8 GR amp R Due    Select the desired report type  GAGE LISTS  CALIBRATION DUE  GR amp R DUE  MASTER DUE   from the REPORTS menu  The    Due List Setup    window will appear           Due List Setup  Reports Gage Type   v ALL  Calipers   C Due Next 7 Days indicators     Master   C Due Next 15 Days MASTER   C Due Next 30 Days None      Complete Chronological List    Status  C All      Active    C Inactive    OK Cancel             Reports    In the    Reports    section  select a time frame for the report list     Gage Status  In the    Status    section  select the status for gages to be included in the report  Status values are  originally set or edited in the    gage description    window     Gage Type  In the    Gage Type    list  select a gage type  if applicable  Gage Types are defined in setup and  selected in the    gage description    window     Cancel Button  Press the CANCEL button to close the    Due List Setup    window     OK Button  After pressing the OK button  a list of gages due will appear  If no gages appear  none were found to  be due under the conditions specified     The list of gages due will look similar to this     Reports    Reports Menu    s      6 9          TI GRER   GRER Due ListComplete  Sorted Chronologically Jon  Fie  Gage Number_G
98. ibute gages   short form  enter    G    for Go or Accept and    N    for No Go or Not accept  The corresponding cell will  turn green and the actual value will be internally stored in the database  The value will be hidden so it    will not cause bias in later measurements     By Trial Method  The    By Trial    method can be used with long or short variable and short attribute studies  Using this    method  all entries are made in one appraiser trial at a time  column by column   Data entry is  organized by trial and must be entered in the order prompted by the system     The    By Trial    entry spreadsheet looks something like this     Gage Studies    Data Menu  55    ij  Gage Repeatability  amp  Reproducibility Study xi    File Edit Analyze Instructions  Performed By    Date  26 07 2010 Time   11 52     E       Gage   A     r Characteristic  Part     LSL    USL     Characteristic        Saved calculations  Calculation Method  Average   Range  gt   Total Variation Method   Study Variation         Variable   Long 3 Appraisers 10 Parts 3 Trials Data Entry Method   By Trial                         Appr  1   Appr  1 Appr  2   Appr 2   Appr  3  Trial  1   Trial  2   Trial H2      Trial  3   Trial  1   Trial  2   Trial  3                         To begin  click the green ENTER button at the top of the first column  The system will display a data  entry window with an indicator as to which appraiser  trial  and part to measure  For variable gages   enter the value for the specifi
99. ield  This is often  dictated by the organization s quality manual or procedures     Date of Purchase  Enter the date the master was purchased in the    Date of Purchase    field     Cost  Enter the cost of the master in the    Cost    field     Status    Set the current status of the master by clicking on the Status button and selecting from the list  No  manual entry is allowed  you must pick from the list     Location  Set the current status of the master by clicking on the Location button and selecting from the list   You can also type in a location that is not in the list    Documentation    Enter for the location of the documentation or the documentation ID for the master in the     Documentation      field       8   DataLyzer Spectrum Gage Management    Supplies Company Button   No manual entry is allowed in the    Supplies Company    field  For entry in this field  use the  SUPPLIES COMPANY button to open the    Suppliers    list to select or create a supplier  Note that  the    gage supplies    option button is selected in this window  allowing only predefined suppliers of  that type to appear in the list for selection  however  if you create a new supplier  you can specify  them to be any of the supplier types available in the system   To add a new manufacturer to the  master list  type a supplier name in the    Company    field and press SELECT  The program will notify  you that the supplier does not exist and ask if you want to create it  Press YES  The    Gage  S
100. ies     Range Chart  The    range chart    option shows the range of each operator s readings for each sample measured  In    addition  a control limit is calculated and displayed  Note the control Llimit calculation in Appendix A                   GRER Average  amp  Range Method E  Statistics Data Performance Curve  Average Charts Individuals Charts   Range 4  Chart Type  Range Chart C Eror Chart  Range Chart       Appraiser 1 Appraiser 2 Appraiser 3       J Print                Print Exit             Error Chart      s   DataLyzer Spectrum Gage Management    On the    error chart     individual deviation  error  for each part is plotted  Error in this case is calculated  as observed value minus average measurement of the part  This analysis offers several good  applications      One appraiser s averages are higher than another     An appraiser s second reading is consistently lower than the others     A particular part is difficult to measure consistently by all appraisers       GRER Average  amp  Range Method X   Statistics Data Performance Curve  Average Charts Individuals Charts Range  amp  Error Charts   Chart Type     Range Chart             Error Chart  2 0    16  1 2  0 8  0 4  0 0   0 4   0 8   1 2  16        2 0    m a  Appraiser 1 Appraiser 2 Appraiser 3  J Print m             Print Exit                5 2 Gage R amp R Study    Once a new GR amp R or GR amp R by type study is created or an existing study is opened  a    Gage  Repeatability  amp  Reproducibility St
101. iety of RS 232 devices  The COM PORT ASSIGNMENTS and  GAGE INTERFACE TEMPLATE menu items work together to define this communication  Each  computer com port  1 2 and 2 virtual ports  can be controlled by a driver that defines how that com  port will communicate  The COM PORT ASSIGNMENTS menu item displays which drivers are  invoked on which com ports  The GAGE INTERFACE TEMPLATE menu item allows definition   editing and testing of the gage drivers themselves     7 14    Setup    Setup Menu  127                Com 2  Unavailable    No Gage v    Com   v  No Gage      Com       No Gage        OK Cancel             The DataLyzer Spectrum Gage Management package provides predefined drivers for most popular  interfaces and serial gages  You can assign these drivers to each com port field by selecting an  available gage driver     You can assign one gage driver to each com port  The sole purpose of the com ports window is to  assign existing gage drivers to control particular communication ports on your computer  If no RS   232 communication will be required  the    no gage    option should be used     No gage    is the default for  all com ports at installation  To activate a com port   s drop down menu  use  Alt  the respective  number  For example   Alt   1  activates the com port 1 menu and  Alt   2  activates the com port 2  menu     The program will show which com ports are available or unavailable for communication     When using external  non DataLyzer software TSR gage driv
102. ility Study  Gage Statuses 105    Index    Gage Template Window 107    GR amp R Analysis Tools    40    GR amp R and Calibration Histories 83    GR amp R History 25      l    Inspectors 105    L      Language 90    Measuring Instructions 34  66      P    Preferences 91     R     Reports    Reports Menu  72    ere    Supplier 102  Supplier Summary  Suppliers 101  System Definitions    84    88    
103. in length     Text Area    Statistics Checkbox  Marking the    statistics    checkbox will list all statistics and show calculations associated with the  study on the report     Data Table Checkbox  Marking the    data table    checkbox will include a complete data table in the report     Graphics Area   Use the checkboxes in the graphics area to select the items you would like to include in the report   The    average plot charts        individuals charts        range     error charts     and    performance curve     checkboxes apply to long variable studies  The    average plot charts    item applies to ANOVA studies     5 4    Gage Studies    Data Menu  NOS    Print Button  Press the PRINT button to print the report     Exit Button  The EXIT button closes the window without printing     Calibration Study    Use the ENTER EDIT CALIBRATION item in the DATA menu to define and enter  review  or edit  gage calibration study information for an individual gage     Simple Mode       File Edit Reports  Calibration number   000007 Date  07 04 2010 Time  has  Performed by     First Inspector SCS Manufacturer   Acme Calibration  Gage Name    Model  Model  Gage Type   Frer   Serial Number  E   Description   Desctiption    Company Number   Comman  Status  DAA se    Certificate Scan         Browse      Advanced               When a calibration study is started  the Calibration Study displays simple mode  Simple mode is  for situations where there is no need to record calibration data  ot
104. is of variance  method is a more accurate  more flexible method than the     Average 8 Range    method with no additional data entry requirements  ANOVA shows equipment  variation  repeatability  and appraiser variation  reproducibility  and adds discrimination for the  interactions between appraiser variation and equipment variation  The ANOVA method is similar to  the    long variable study average  amp  range    method and is the preferred method of GR amp R analysis  because it can handle any experimental setup  number or samples  operators  and trials   and it can  estimate variances with higher bias  but the computation  handled by DataLyzer Spectrum Gage  Management  and analyses are more complicated     When conducting an ANOVA study  data entry must be randomized to be valid  Use the     randomized    data entry method  Operators will be prompted for measurements of pieces in a  random order to ensure independent measurements  The draw back to this is that all operators must  be present at the time of the study where as you can enter data for one operator at time for Average  8 Range type studies     Total Variation Method  Variable Gages Only   The selection of the    Total Variation Method    is used along with the globally set    passing   error    Gage Studies    Data Menu  51    limit     located in SETUP  PREFERENCES  to determine if a gage passes or fails the R amp R study   There are three options for this item     Study Variation   Study variation is gage er
105. is study type is selected  the    Number of  Appraisers        Number of Trials     and    Number of Parts    fields will be set and locked by the software     Long Form Variable   The long study has flexible parameters for number of operators  number of parts measured  and  number of trials per part  We suggest using two or three appraisers operators  each making two or  three trial measurements on eight to ten sample products characteristics  The sample products are  chosen randomly from production parts  or organized by size ranging through the specifications    Appraisers are asked to make the desired number of measurements using their normal measuring  techniques  Appraisers should take measurements independent of each other so results between  appraisers have no influence on each other  The studies require significantly more sampling than the  short studies  but they provide more detailed information regarding the type and amount of  repeatability and reproducibility variation present in the measurement system  Two types of  calculation methods are available for long variable studies     average  amp  range    and    ANOVA     see the     Calculation Method    item for a description of these      Long Form Attribute   The long attribute gage study is the most in depth study and requires a substantial amount of time   The study consists of one operator making twenty measurements on 8 10 samples  Samples must  be selected carefully  This study may be used on both single a
106. l  Gage Name  0 1 caliper Performed By  KRE  Gage No   8876M09 Date  5 16 2008  Gage Type  caliper Serial Number  8876M09  100  confidence range  3 30 to 6 10  Probability of Acceptance at LSL  0 53 USL  0 53                The purpose of a gage performance curve is to determine the probability of a measurement system  either accepting or rejecting parts improperly using data from Long Variable and Attribute gage  studies  This tab applies to long and short variable studies and long attribute studies     Once the amount of gage error has been determined via a gage study  it is possible to calculate the  probability of accepting a part of some reference value when using the gage  To accomplish this  the  assumption is made that repeatability and reproducibility are normally distributed with some  variance  s2    The gage performance curve provides a graphical method of communicating how well the  measurement system is discriminating good parts from bad parts as the measurements approach  the decision points at the upper and lower specs     The graph contains x  and y  axes  The x axis along the bottom contains values as reflected by the  measurement system  The y axis relates to the percent confidence  Specification limits are shown  as vertical lines intersecting the x axis  Normal curve tails intersect these specs and represent  confidence indicators  At the intersection of either spec you can read the corresponding left or right  scale to see the confidence one has in the m
107. l Variation  TV   Total variation can be based on one of three different statistics    Study Variation    Ty   VR amp R  PV     Process Variation  TV   VR amp  R   PV  60 of the manufacturing process    Tolerance    Reports    Reports Menu    75      TV   USL   LSL      Total Variations  TV  Area   EV   100 EV TV  or 100 EV SV  or 100 EV TOL  depending on which was selected in study setup     EV   100  EV TV  where TV could be variation found in this R amp R Study or     3sigma of the most  recent SPC data or Percent of Tolerance depending on which one was selected when the study was  setup      AV   100 AV TV  or 100 AV SV  or 100 AV TOL  depending on which was selected in study setup   AV   100  AV TV  where TV could be Study variation      3sigma of the most recent SPC data or      of Tolerance depending on which one was selected when the study was setup      R amp R   100 R amp R TV  or 100 R amp R SV  or 100 R amp R TOL  depending on which was selected in study  setup    Variables Menu Item  When you access the VARIABLES menu item from the report preview window  a window will appear  containing a spreadsheet of the data used in the study     It looks like this   2 51 xi  File    Variables for GR  amp  R Average        Appraisers JPart  1 Part  2 Part  3 Part  4 Part  5 Part  6 Part  7 Part  8 Part  9    Appraiser  1   Trial    1 0 56 0 57 0 83 0 61 0 58 0 62 0 67 0 70 0 56  Trial    2 0 62 0 65 0 63 0 54 0 59 1 00 0 61 0 62 0 63  Trial    3 0 64 0 61 0 65 1 00 0 51 0 52 
108. ld  so you can locate the specific channel identifier in the exact position of the line s transmission  string  For example  ifthe gage interface transmits 02 as the channel identifier in position 10 and 11   you must put    0    in position 10 and    2    in position 11 of the    identifier    field  The column field shows  you where you are as you move the cursor       ns   DataLyzer Spectrum Gage Management    Previous and Next Buttons  The PREVIOUS and NEXT buttons allow you to scan and set up to 64 channel descriptions     Save Button  The SAVE button saves the template     Cancel Button  The CANCEL button exits without saving     Special Menu   Multi Channel A D   The    Multi channel    A D window is a combination of the two previously described windows  lt is used  in situations where you need to communicate with an analog to digital interface that has more than  one channel        Multi channel description       Channel information 1 64    Prompt     Identifier     Reading   hr Column       Serial port to test   1   Timeout   7    Message                   Analog to Digital  High Master  DD  Low Master  fo  High Gage Reading  fo Master High    Low Gage Reading  fo Master Low       lt  lt  Previous Save   Cancel Next  gt  gt       Channel Information  The    Channel information    area displays the current channel description number                 Prompt   Use the    Prompt    field to enter a message to be sent to the gage to request a reading from the  current chan
109. led locally  on drive C   for graphics  and framing objects support  Microsoft   PowerPoint Viewer must be installed locally  on drive C   to  view PowerPoint files  Adobe Reader must be installed locally  on drive C   to view Acrobat   pdf   files     To detach a file  press the desired SET button  When the    Open    browser window appears  press the  CANCEL button  The program will ask if you would like to remove the file  press YES     Specify Web Addresses   To enter or edit a web address as an attachment  mark the    specify web addresses    checkbox  before pressing the SET button  A dialog box window will appear where you can enter the web  address path required  To attach another file type  unmark the    specify web addresses    checkbox  before pressing another SET button     Attachment Buttons  After setting a file or website  click the Attachment to check that the link is working            DataLyzer Spectrum Gage Management    4 3    GR amp R Study Definition    A GR amp R Study definition is a way to specify details for Repeatability  amp  Reproducibility tests of a  gage and characteristic  part  combination that will be regularly repeated     If you select New GR amp R Study Definition or Open GR amp R Study Definition the Study Definition  window opens        ig  GR amp R Study Definition    Name  G1VarChart DK      Gage     G age    nn E xit      Category  Variable  Type  Typel               Characteristic    Part    Pa  Characteristic    Varchan    Category
110. log to Digital     This facility is used to master analog to digital transmissions  Typically A D converters break an  analog DC voltage range into digital segments  An example would be a 12 bit A D converter breaking  a given DC voltage range into 4096 digital segments  If the analog range is  2 volts DC to  2 volts  DC  this 4 volt range will be divided into 4096 segments or approximately 1000th of a volt per  segment  A D interfaces typically transmit a measurement in terms of the number of segments the  measurement represents out of the total voltage range  Therefore  if the gage can measure   010  inches to   010 inches  this range will be cut into 4096 segments with each segment approximately  equal to  000005 of an inch  This means that each 1000th of a DC volt increment is represented   by  000005 of an inch travel in the gage  Therefore  when the gage measures   005 inches it will  transmit the representative number of segments or 3000  This value is then converted into inches  with the result reported to the software as the measurement from the gage  in this case   005  inches  This process of interpolation can be setup automatically using the SINGLE CHANNEL A D  facility  It is sometimes referred to as mastering     The window looks like this                 a Sa   amp  Analog Gage  Message  Serial port to test  dr Timeout   7 Clear  Analog to Digital  High Master  0  Low Master  0 zava  High Gage Reading  0 Master High  _ I Cancel  Low Gage Reading  0 Master Low  
111. ments to be used in any statistical quality control effort and  follows the AIAG Gage Repeatability Reproducibility study procedures  The DataLyzer Spectrum  Gage Management System provides analytical tools for measurement assurance as well as gage  inventory and management     Measurement Assurance Tools   The DataLyzer Spectrum Gage Management System provides two basic types of measurement  system analysis  The first is a calibration analysis based on the accuracy of a measurement  system  i e  the extent to which bias exists in the system  The second is a set of statistical  techniques  gage repeatability and reproducibility studies  that quantifies the variation a  measurement system introduces into the total process variation when gages are put in the hands of  actual human appraisers     Gage Calibration  Accuracy and Extent of Bias     Introduction 9    1 2    Gage calibration allows you to find the difference between the observed average of measurements  and the reference or    true    value  Is this bias within an acceptable range  The DataLyzer Spectrum  Gage Management System further maintains a history of these accuracy analyses     Gage Repeatability and Reproducibility Studies  Measurement System  Variation    The DataLyzer Spectrum Gage Management System determines repeatability  equipment variation   and reproducibility  appraiser variation  for both variable and attribute gaging systems  These results  can then be used to evaluate the adequacy of a complete m
112. n    field is set from the    Test Create  Template    area  When multiple lines are transmitted from the gage  this feature allows you to tell the  software which line will contain the measurement you want  After setting this information in the     Test Create Template    window  it will flow through to be set in the gage template window as well     Starting Column of Reading in the Line   The    starting column of reading in the line    field is set from the    Test Create Template    area  lts  purpose is to locate the position of the first character in the transmission s actual measurement  value  No measuring units  channel identifiers  or other textual information in the transmission should  be confused with the location of the first character in the measurement  This first character of a  measurement is often a   or   sign and these should be included in the set of     when creating the  template in the test area     Ending Column of Reading in the Line   The    ending column of the reading in the line    field is set from the    Test Create Template    area  Its  purpose is to locate the position of the last character of the transmission s actual measurement  value  No measuring units  channel identifiers  or other miscellaneous information in the  transmission should be confused with the location of the last character in the measurement     Sign Column of Reading in Line   The    sign column of reading in line    is the location of the measurement s sign       in the
113. n you press the Instructions button  the Instructions window will appear providing any measuring  instructions and attachments that are defined on the gage s    gage description    window     Process Variation  In the    Process Variation    field  enter the value for either 6sigma of the process or tolerance   whichever you want to measure the bias against     Gage Studies    Data Menu    e      Observed Average  After the study is saved  the    Observed Average    field will display the average of the 10 samples  entered     Bias   After the study is saved  the    Bias    field will display the difference between the master value and the  observed average  The bias value from the most recent bias study is used in calculations for the  Performance Curve of the gage       Bias  After the study is saved  the      Bias    field will display the percent of the process being used by  gage bias   bias   process variation    100      Data Entry Spreadsheet   To enter data  in the    Master    cell  enter the value of the master being measured  The master is often  a part of known dimension  but can be a standard  like a gage block  that is similar in size to a  production part such that it is reasonable to assume similar results from measuring the standard or  a production part  Measure the master 10 times and enter the results into cells 1 10  Blind studies  are the best way to avoid bias     File Menu   Save or  F2   The SAVE item  or pressing  F2  without opening the menu  saves
114. name in the    Name    field     OK button    The OK opens the    Characteristic Group    window for the new or selected characteristic group     Exit Button      100   DataLyzer Spectrum Gage Management    The EXIT button closes the window     Delete Button  To delete an existing characteristic group  select a name from the list or type its name in the    Name     field and press the DELETE button     Characteristic Group Window    5 Characteristic Group   OD Jog    Part  Select Characteristic  30 clip v30 clip   b  40 clip 30 clip   bi                         Save Cancel   Find Similar Statistics Show Selected            Part List   The    Part    list  located in the left hand side of the window lists all parts in the shared DataLyzer  database  Select the part or parts whose characteristics you wish to group by double clicking on the  part or highlighting it and pressing enter     Select Characteristic   Once a part is selected  all of its characteristics appear in the    Select Characteristic    list on the  right hand side of the screen  Select the characteristics you wish to group together by marking their  checkboxes     Save Button  The SAVE button saves the characteristic group     Cancel Button  The CANCEL button closes the    Characteristic Group    window without saving changes     Find Similar Characteristics Button   The FIND SIMILAR CHARACTERISTICS button will search all parts in the database for  characteristics similar to the selected characteristics  at lea
115. nd double limit gages  however  in  both cases only one limit must be examined at a time so their actual sizes are distributed  symmetrically around the reference or true value  The smallest sample must have O accepts and at  least six other parts must have 1 19 accepts  At least one sample with 20 accepts is also required     The long attribute study analyzes around one spec  limit  The software will not allow data to be  stored unless it meets the criteria of the study  If you try to store data that does not comply with  what is required by the long attribute study definition  the software displays a window containing  directions about how the sampling needs to be corrected  The long attribute study shows equipment  variation only  It provides the following results on reports  gage performance curve plot  R probability  for repeatability at a given reference value  B bias     Short Form Attribute   The short attribute gage study evaluates go no go measurement systems  lt involves two appraisers  each taking two measurements on 20 sample parts  Some passing and some failing parts should be  included in the samples  Both appraisers should measure all the parts once  and then repeat to  measure all 20 parts again  If all four results for any sample do not match  the gage is not  acceptable  Conversely  the gage is acceptable if all four trials agree for every sample  Ifthe gage is  judged to be unacceptable  it can be re evaluated using this method if there is doubt about the 
116. nel  This field should not be used to request a reading from all channels  Any global  requests should be entered at the main gage setup screen     Identifier    Setup    Setup Menu    119      Enter a set of characters in the    Identifier    field that uniquely identify the line containing the reading  for the current channel  This may be a label like     X axis     or    radius     or a channel number such as     01    or    A     It is important that you enter the characters in the exact column positions that they  occur in the gage transmission     Reading Number  The    Reading number    field allows for multiple readings per line to be handled  If only one  measurement is transmitted per line then enter    1    in this field     Column   The    Column    field does not allow input  It merely displays the cursor s position for the    Identifier     field  so you can locate the specific channel identifier in the exact position of the line s transmission  string  For example  ifthe gage interface transmits 02 as the channel identifier in position 10 and 11   you must put    0    in position 10 and    2    in position 11 of the    Identifier    field  The column field shows  you where you are as you move the cursor     Serial Port to Test  The    Serial port to test    area includes a pull down list for specifying which RS 232 port will be used  for testing     Timeout  Use the    Timeout    field to enter the number of seconds the template will wait for an RS 232 string  
117. nstructions window  Clicking the attachment buttons at that point will  cause the system to follow the link you are establishing now and result in the instantaneous display  of the    attached    file linked to it  It may be a picture file like a jpg or tif  It may be an MS Word or  Excel file or it might even be a movie about how to do sampling for the study  These attachments  can be a powerful addition to assist in performing a study     Supported attachment file types include     Document Files Movie Files   Text Files    txt  AVI Move    avi    wmv    Rich Text Format Files    rtf  Audio File    wav   aif   au   snd   wma   aiff    Word Document Files    doc t MPEG Move    mpg   mp2   mpeg   m1v   mpe   mpv   Acrobat Files    pdf t Quick Time 2    qt   mov    Hypertext files    htm   html  Windows Media File    asf   wm   wma   wmv    Image Files Other   AutoCAD    dwg   dxf  Excel Files    xls   xlw t   CADKey    prt  PowerPoint Files    ppt t   BMP Images    bmp  Web addresses    GIF Images    gif    JPEG Images    jpg   jpeg   PCX Images    pcx    PNG Images    png    TIFF Images    tif    tiff   WMF Images    wmf     t Note  The DataLyzer Spectrum Gage Management software provides an internal utility for viewing    doc and  xls files  however  Microsoft   Excel Viewer must be installed locally  on drive C   for  embedded graphic and multiple worksheet support  otherwise only the cell contents of the first  worksheet are visible  Microsoft   Word Viewer must be instal
118. ntry  into this form     Edit Review Study Button and File Menu   Edit Review Study  not available from  the Reports Menu    The EDIT REVIEW button or menu item opens an existing study for editing or review  Select a study  by clicking in its study date field  lts row will highlight to show it is selected  Only the latest  calibration study of the selected gage is available for editing  other studies are read only  Press the  EDIT REVIEW button to open the selected study  See the    Calibration Study Window    description  for more information regarding the study window     Exit Button and File Menu   Exit  The EXIT button closes the study history window and retums to the main DataLyzer Gages window     File Menu   Delete   The DELETE item will delete the selected study  Select a study by clicking in its study number field   Its row will highlight to show it is selected  Select DELETE from the FILE menu to delete the  selected study  A window will appear to confirm that you want to delete the selected study     File Menu     Print Certificate    Select Select this menu item to print a calibration certificate for the currently selected calibration  study  Select a study by clicking in its study number field     File Menu   Print Label    This opens a window for printing calibration labels on A4 or similar sheets of peel off labels  Select a  study by clicking in its study number field     General Windows 25    3 3    File Menu   Print History  This prints the list of calibrati
119. o be edited because of user error or gage malfunction   save the study and open the study using the EDIT REVIEW STUDY button in the    GR amp R History     window  an open sheet is then available to edit all entries     Randomized Method  The    Randomized    method can be used with long or short variable and short attribute studies  For  valid results  this method must be used for long variable studies using the ANOVA calculation      s   DataLyzer Spectrum Gage Management    method     The    Randomized    entry spreadsheet looks something like this     ij  Gage Repeatability  amp  Reproducibility Study xi                         Fie Edit Analyze Instructions  Performed By     Date  26 07 2010 Time   1 52  Gage    Gage      r Characteristic  Part     LSL    USL     Characteristic      Saved calculations  pe Method    Average  amp  Range y   Total Variation Method   Study Variation X            Variable   Long 3 Appraisers 10 Parts 3 Trials Data Entry Method   Randomized    Appr  1  Appr  2  Appr  3           Appr  1   Appr 1   Appr 1   Appr 2   Appr 2   Appr 2   Appr 3   Appr 3   Appr  3  Trial  1 Tral 2   Trial  3   Trial  1 Trial  2   Trial  3   Trial  1 Tral 2   Trial  3 _                   To begin data entry  click the green START button in top left of the spreadsheet  The system will  display a data entry window with an indicator as to which appraiser  trial  and part to measure  For  variable gages  enter the value for the specified measurement and press OK  For attr
120. o the operating system and these interfaces should begin working as soon as  they are plugged in     From the SETUP menu  select the GAGE INTERFACE TEMPLATE item  The following window will  appear     7 12    Setup    Setup Menu  107             5 Open Create a Gage Interface Template File Y       Gage Name     Import Templates                  Enter a new gage name in the    Gage Name    field or select a gage file name from the list  Press OK  to open the    Gage Template    window     The CANCEL button closes the    Open Create a Gage Interface Template File    window   To delete a gage template  select the desired name from the list and click the DELETE button     When you are sharing a database with the DataLyzer Spectrum SPC package  an IMPORT  TEMPLATES button is available  Use it to import existing gage driver files that exist in the database     Gage Template Window    The  Gage Template  window looks like this      amp  Gage Template  C  Program Files DataLyzer Spectrum gages GAGEPORT                File Test Special Options    Port Settings  Baud rate   9600 y  Parity   None v Data bits  le y  Stop bits   1 z   Transmission to initialize gage  DOO O  Transmission to prompt gage for reading  A  Secondary gage prompt y   Reading Acknowledgement  y   Line terminator for lines received from gage  E  Lines per transmission received from gage  ET       Reading Description    Reading    fr of E  lt  lt Previous lt  lt   gt  gt     Next gt  gt   Line number of reading in g
121. o this field  but another date can be entered if desired     Time  Enter the time the study was initiated in the    Time    field  The current system time is automatically  entered into this field  but another time can be entered if desired     Gage frame   In the Gage frame  click the Gage button to select a gage  The category of the gage will be shown   variable or attribute  in the frame and the Gage Type  ifthe gage is included in a Type   If you have  not defined any gages you can just type in the name of the gage here     If Study is based on a Study Definition and the gage is included in a Gage Type  then this study can  be considered as representative of all the gages in the gage type  Ifone or more GR amp R studies has  been stored using the Study Definition  the Gage Type will become    locked     This means that you  can only change the specified gage to another gage in the Type       s   DataLyzer Spectrum Gage Management    Characteristic frame   Ifthe program is being used in conjunction with Datalyzer Spectrum you can select a Datalyzer Part  and Characteristic by clicking on the buttons  The Upper and Lower specification limits will be  displayed along with the category of the characteristic  variable or attribute   Ifthe characteristic is  included in a characteristic group  the group will be also be shown     If Study is based on a Study Definition and the selected characteristic is part of a characteristic  group  then GR amp R studies performed using this 
122. ocument      field  allows you to add intemal traceability information     Notes  Use the    Notes    field for any extra comments you wish to add about the supplier     Save Button  The SAVE button saves the changes made to the supplier s information and exits the window  The  SAVE button only appears when using the NEW SUPPLIER or OPEN SUPPLIER options     Delete Button    The DELETE button will delete the currently displayed supplier  The DELETE button will only appear  when using the DELETE SUPPLIER option     Exit Button    The EXIT button closes the supplier window without saving changes or without deleting     File Menu   Save or  F2     To save the information in the    Gage Suppliers    setup window  select SAVE from the FILE menu or  press  F2  without opening the menu     File Menu   Save  amp  Exit or  Ctrl   S   When using the NEW SUPPLIER or OPEN SUPPLIER options  select FILE  SAVE  amp  EXIT  or press      104   DataLyzer Spectrum Gage Management    7 9     Ctrl    S  without opening the menu  to save the information and close the    Gage Suppliers    setup  window  When using the DELETE SUPPLIER option  select FILE  SAVE  amp  EXIT  or press  Ctrl   S   without opening the menu  to delete the currently displayed supplier     File Menu   Print or  Ctrl   P     To print the information in the    Gage Suppliers    setup window  select PRINT from the FILE menu or  press  Ctrl   P  without opening the menu     File Menu   Exit or  Ctrl   X     To close the    Gage 
123. oftware using the MASTER LOW button  The value will transmit from the  gage in terms of a number of digital segments as discussed earlier and will display in the    low  gage reading    field    5  Save the gage file template by using the SAVE button     This sets up an automatic interpolation relationship  The software does the following calculation  automatically       High master size   Low master size     High   segments received   Low  segment received     This defines the    measurement size    corresponding to each digital segment received  The  multiplication of the number of segments by the measurement increment per segment is then done  automatically in the background so the actual measurement value is entered in the input window as    the current measurement value     Save Button  The SAVE button saves the template     Cancel Button  The CANCEL button exits without saving     Special Menu   Multi Channel    Setup    Setup Menu    17      Some gages transmit multiple measurements in a transmission in random order  When this is the  case  these measurements are typically labeled with a measurement identifier in the transmission   The purpose of the multi channel facility is to key on these identifying labels for input rather than on  the line number where it was transmitted  These multi channel gage files typically must be created  for each part file because of the specific nature of the ordered characteristics involved  Multi channel  gage operation typically utiliz
124. ollowing calculation      12   DataLyzer Spectrum Gage Management    automatically     High master size   Low master size     High   segments received   Low  segment received      This defines the    measurement size    corresponding to each digital segment received  The  multiplication of the number of segments by the measurement increment per segment is then done  automatically in the background so the actual measurement value is entered in the input window as  the current measurement value     Previous and Next Buttons  The PREVIOUS and NEXT buttons allow you to scan and set up to 64 channel descriptions     Save Button  The SAVE button saves the template     Cancel Button  CANCEL exits without saving     7 12 4 Options Menu    Communications Options  The    communications options    feature defines the RS 232 communication environment and sets  some general rules for communication     Setup    Setup Menu  121             amp  Communication Options  Line length    Length  f0    ru    Accept short input lines       Test reading timeout  Seconds before timeout  7  Delay between lines 1 10 second  3  Buffer size     e 1024    2048    4096    8192    Buffer usage    Special Clear Buffer on initialization         Clear buffer after each reading    Debug mode    Turn debug mode on    Special modes      Gageport TM   computer mode    Save   Cancel               The SAVE button stores the data and sends you back to the gage setup window  The CANCEL  button cancels any changes made
125. on studies     GR amp R History    The window will be empty if no studies have been performed  The NEW STUDY button will allow you  to create a new GR amp R study        File Show  Study Definition  an y  Gage   Al All      Study Date Study Definition   Characteristic   av   AR   Method  11 07 2010 12 38 i     Long  03 07 2010 20 13   10Daysb VarCharl  UA Sor  02 07 2010 20 09   10D ays VarCharl Long  01 07 2010 20 09   10Days VarCharl Long  07 07 2010 14 30   Long  07 07 2010 14 28   WarCharl Long  07 07 2010 14 27 VarCharl Long  07 07 201013 42  VarCharl Long  07 07 2010 10 34   VarCharl 20 0  BE  06 07 2010 17 26   VarCharl 20 1  Long  06 07 2010 17 19   21 0  ANOVA  06 07 2010 17 14 21 0  ANOVA  06 07 2010 17 05 20 1  ANOVA  06 07 2010 16 52 20 1  Long    New Study   Edit   Exit      As studies are completed  results are stored in the system  Summaries are displayed in  spreadsheet form sorted by date time with the most recent on top  Each row represents an individual  study  Each column represents a key statistic or label relating to the study  Each study type has  different statistics in the study history window  To delete a study  click anywhere on the study   s row  and press the delete key  It is important to note that certain studies do not provide all these  statistics and so certain fields will be blank  for example  long attribute studies only provide  equipment variation  EV  since they involve just one appraiser  Short attribute studies offer no AV   EV  or RR 
126. or Gage Type window     Example calibration gage   GR amp R Attachments xi    GRR Measuring Instructions        GRR Control Plan Reference     ee aaa    Attachment 1   Set  Attachment 2   Set  Attachment 3   Set  Attachment 4   Set     Attachment 5   Set    I Specify web addresses       Measuring Instructions Field   Enter instructions for how to run the GR amp R or Calibration study in the    Measuring Instructions    field   3000 characters max    This field can be left blank but it is not recommended  Different instructions  and attachments are stored for Calibration and GR amp R studies     Control Plan Reference  The    Control Plan Reference    field allows you to cite the applicable page or paragraph in your control  plan describing how Gage R amp R studies for this gage should be conducted  It can also be left blank     Set Buttons   By clicking any of the five SET buttons you can browse your computer and network resources to  select files such as images  documents  videos  or spreadsheets that will then be attached to the  selected button  Web addresses can also be attached  see    specify web address    below  These    Gage Details    File Menu  35    attachments are intended to augment measuring instructions for the gage at the time when it is  being studied  Up to five attachments of the types listed below can be selected  one per button   At  the point when you or your staffis actually performing the studies  attachment buttons will be  available in a measuring i
127. ou are sharing an SPC DataLyzer database and have assigned a specific characteristic to the  study  both the lower and upper specifications will be filled in automatically from the characteristic  you selected  Otherwise  type in the upper and lower specification limits for the characteristic into  the    LSL    and    USL    fields  Once data is saved in the study or a characteristic is assigned to the  study  the specification limits are locked and cannot be changed     Date  Enter the date the study was initiated in the    Date    field  The current system date is automatically  entered into this field  but another date can be entered if desired     Time  Enter the time the study was initiated in the    Time    field  The current system time is automatically  entered into this field  but another time can be entered if desired     Location  Enter where the bias study is performed in the    Location    field     Gage Name  The    Gage Name    field displays the gage name  The contents of this field cannot be changed     Gage Type  The    Gage Type    field displays the gage type as specified in the gage s    gage description    window  lt  cannot be changed here     Serial Number  The    Serial Number    field displays the serial number as specified in the gage s    gage description     window  It cannot be changed here     Notes  You can use the    Notes    field to enter comments such as the condition or other aspects of the gage  being studied     Instructions Button  Whe
128. ow will then appear  select the desired gage from the gage catalog     After you select a gage  the study    history    window will appear containing all studies performed for  the report type selected  The studies will be sorted by date and time with the most recent on top   If  no studies exist in the list  no studies have been performed for this gage       The    history    window will look similar to one of these                    GRER History  File   Study Date Type Performed By Part Characteristic EV   AY   TV   Method  5 29 2008 13 22  Estua Gage TOL PartOne varcharl 1 0 94 71 ANOVA  5 28 2008 15 41  Actual Gage 1 0 95 73 Long  5 28 2008 15 13   Actual Gage widget2000 weight 1 0 81 64 Long  5 16 2008 11 54   Actual Gage KRE PartOne varcharl 1 0 97 30 Long  5 16 2008 10 59 ActualGage TOL 1 0 1213 Long  5 16 2008 10 31 ActualGage PVM widget thickness 0 970 0 000 96 84 Long  5 14 2008 15 09   Actual Gage KRE widget2 length 0 000 0 350 34 89 Short  5 14 2008 14 53   Actual Gage KRE widget2 length 0 3500 0 0000 95 07 ANOVA  5 6 2008 11 21 Actual Gage MTL widget width 092 000 91 71 Long   Print                     s   DataLyzer Spectrum Gage Management    J    a T  Calibration History  0 1  indicator   12342134  File             Study Date Performed By Master Set Status Location  12 13 2005 14 16 Pass cib 2             Print                      The spreadsheet areas of these windows are described in Chapter 4     Data     Print the history report using the PRINT button or 
129. p  Range         unzzunsnnnennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nn 73  2 Gage R amp R   Long Variables     ANOVA     uuenenssnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnsnnnnnnn nennen 76  3 Gage R amp R  Short Variables               nee nenne 77  4  Gage  R amp R   Long Attribute              2222er anna 78  5 Gage R amp R   Short  Attribute            see nein 79  6 Gage R amp R     Performance Report         uuussseennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnsnnnnnnnsnnnnnnn nn 80  7   Galibration Certificate            2 2 ann oia danna da 81  8 Calibration   amp  GR amp R D  e  vicio ee 82  9 GR amp R and Calibration Histories                      440 44400nnnnnnnnnnnnnnnnnnnnannnnnnnnnnnnnnnannnnnnnnn nenn 83  10    Supplier SUMMARY  ciencia aaa a aan anne nenn EHRE nennen 84  11  Additional Reports                2   2  ar ii ie ea 85  Part VII Setup    Setup Menu  88  T System DEMOS di 88  EE A A A OO TO 90  3 Preferences icon ne einen 91  4  Define Gage Types ia 92  5    Define Masle S eea re ii i   Ain ti cascadas 96  6 Define GR amp R Characteristic Groups       ccconccconccnconccnnnnccnnnnnnnncnnnnnonnnernnnnnnnnrrnnnncrnnnrenanenes 99  7 Suppliers   New  Open 8 Delete        cccoonccoonccocnceconnconnnccnnnennnnnnnnnnnnnnnnnonarennnennnnnrnnenennns 101  8  Supplier Details sinirinin iinne iaa a akaa aeaaea aa a raaa aaia diia 102  9 Gage Locations zeuai ineeie detec cee di ia ii 104  10 Gage Statuses and Inspectors     ccoooncccccononcccnonancnnnnnncncnnnnnrnn
130. p R    The measuring process often contributes a significant amount of variation to the total variation of a  manufacturing process  There are several standardized ways to study this variation  Selecting one  over another depends on which type of measuring system is being used  variable or attribute gages   along with how much discrimination and precision is required vs  the amount of time available to do  the study  GR amp R studies can be entered for a specific gage or you can study one gage and  designate results to apply for all gages of a specified gage type     After conducting a study  a high repeatability value  also known as equipment variation or EV  can  point to the need for gage maintenance  repair or redesign  or the addition of fixturing  etc  A high  reproducibility value  also known as appraiser variation or AV  points to the need to increase  measurement consistency between operators  possibly better measuring procedures  or additional  appraiser training to develop uniform measuring techniques between operators     Repeatability and reproducibility values are compared to    total variation     TV  to determine a percent  R amp R i e  the percent of total variation being consumed by gage error  Total variation can be defined  three ways     1  Total    tolerance     USL LSL    2  Total     study    variation      3  Total    process    variation which uses a larger sample from the manufacturing data usually  collected in an associated SPC system     Sometime
131. pe a supplier name in the    Company     field and press SELECT  The program will notify you that the supplier does not exist and ask if you  want to create it  Press YES  The    Gage Suppliers    setup window will then appear  see the  description of this window   Once a supplier is selected  it appears in the field  You can view more  specific information about the supplier by clicking on the field  a window containing read only  information about the supplier will appear     Other Supplies   The    Other Supplies    field is an optional field used for noting external gage supplies like batteries   cables  chemicals  attachments  printer paper or other miscellaneous items consumed while using  the gage     Gage Details    File Menu  33    Notes   Use the    Notes    field to enter miscellaneous user notes about the gage  A maximum of 3000  characters are allowed in this field  Full calibration procedures  graphical attachments and work  instruction documents are available in the INSTRUCTIONS items under the GR amp R and  CALIBRATION menus     User Defined Fields 1 12   The    user    fields only appear if they have been previously setup in the    headers fields    tab under  SETUP  SYSTEM DEFINTIONS  These fields will appear at the bottom of the window beneath the     Notes    area not only in the in the Gage Definition window but also in the reports  They are available  for entering any information you wish  up to 24 characters per field     12 fields max  they are global 
132. ppraiser s trial avg points  In a perfect situation  all points would be  in identical locations on the graph                      GRER Average  amp  Range Method X   Statistics Data Performance Curve  Average Charts Individuals Charts Range  amp  Error Charts  Chart Type    oR    C Averages Chart C XY Plot  Run Chart         N m   An   m o a    10    Appraiser 1 Appraiser 2 Appraiser 3        Print LJ          Print Exit                Averages Chart  The    averages chart    option adds control limits that are calculated by using the average appraiser  trial points for each part as subgroups              GR amp R Average tt Range Method E  Statistics Data Performance Curve  Average Charts Individuals Charts Range  amp  Error Charts  Chart Type    C Run Chart C XY Plot       Averages Chart       r N Ly E Ll e  amp  p pP a    Appraiser 1 Appraiser 2 Appraiser 3     Print m             Print Exit                X Y Plot  The    X Y Plot    option ranks the parts in the study smallest to largest using a calculated grand    Gage Studies    Data Menu    45      average for each part  This ranking becomes the x axis and represents the order the part trial  averages are plotted left to right  Measured average is the y axis  A linear regression is plotted on  each operators points  Dramatic differences between operator performances can be addressed by  training or other remedial measures              GRER Average  amp  Range Method E  Statistics Data Performance Curve  Average Chart
133. r Point presentations     Language    Use the LANGUAGE item to select the desired operating language for the program  The default  language is English  The user defined language item can be used to enter your own localized  language  Contact DataLyzer International  Inc  for instructions regarding use of this option     Setup    Setup Menu   a      7 3 Preferences    The PREFERENCES item is used to set preferences for the various studies in the program                                            m Variable Study Preferences Attribute Study Preferences  Appraisers   3 2to4   Type Appraisers  fi   L  Parts  fo 5to10 oe Parts  fo 8to10  Trials   3 2to4 Lo Trials   20  r Re Evaluation Time Units GR amp A calculations based on    Days    Month C 5 15 Sigma   6 Sigma  IV Alert if calibration or GARA is due in le Days I    Use Appraiser Name  Passing   Error Limit  fo  Include Atmospheric Information  Save  amp  Exit   Exit            Variable Study Preferences   You can set the preference configuration to match the variable Gage R amp R study configuration you  expect to use most frequently  These will come up as defaults when doing a new study or creating a  GR amp R Study Definition  They can also be changed at that time as necessary     Type    Using the    Type    selection  choose whether the default GR amp R variable study will be long or short     Appraisers  Enter the default number of appraisers used when conducting a variable GR amp R study in the     Appraisers    field  
134. racteristics was received  by the com buffer     Ignore all hardware triggers  The    Ignore all hardware triggers    option leaves data in the com buffer time after time     Allow all hardware triggers  The    Allow all hardware triggers    option attempts to process data in the com buffer no matter what  garbage characters may exist     Framing Errors  These errors typically arise from software errors like incorrect baud rate or parity     Reject all framing errors  default   The    Reject all framing errors    option throws away all data with framing errors     Ignore all framing errors  The    Ignore all framing errors    option attempts to process data in the com buffer no matter what  garbage characters may exist     Allow all framing errors  The    Allow all framing errors    option attempts to process data in the com buffer no matter what  garbage characters may exist     Hexadecimal   In rare cases  instruments will transmit in Hexadecimal code  This option converts the hexadecimal  code into readable ASCII text before displaying the string  This item works like a toggle switch   When it is set  the message    HEX    appears next to the    Reading Description    heading  Select the  menu option again and    HEX    disappears     Clear  The CLEAR item resets all gage template fields to their default settings  This includes fields in the  SPECIAL and OPTIONS menus     Gage Template Example 1  Single Line  Single Measurement       Example string  _ _ _X    4 379    This
135. results  since only a pass or fail conclusion is drawn  Short variable studies require only  one trial from 2 operators so only AV can be measured                       Show Menu   Over time the number of studies stored can get very long  which can slow down the loading of the  list  As you are likely to be most interested in the most recent studies  the Show Menu allows you  to improve performance by limiting the number of studies listed     Study Definition  drop down list    The drop down list under the menu bar allows you to filter the table to show only studies which used  particular Study Definition  Click the small down arrow at the right hand end of the box to show the  list of Study Definitions  Select    All    to show all Study Definitions       26   DataLyzer Spectrum Gage Management    Gage Button  Click the button to select a Gage from the Gage Catalog  The table will then only show studies  which used the selected gage  Click the All button to remove the filter     Study Date  The    Study Date    field displays the date and time the study was performed     Study Definition    If the study was based on a Study Definition  the name will be shown in this column     Part  The    Part    field displays the name of the part being measured in the study     Characteristic  The    Characteristic    field displays the name of the characteristic being measured in the study     EV  Equipment Variation    Equipment variation  EV    gage repeatability  The values in the    
136. revious tests     Message  The    Message    field will display operational comments that the software reports about the  transmission  Messages include  opening com port  send data now  communication timeout  etc     Transmission Field s    The number of transmission fields displayed in the    Transmission    area is determined by the number  entered in the    lines per transmission received from gage    field in the    Gage Template    window   These fields are used to define the template that is used to mask and pull out measurement values  from the string  When a gage transmits one or more strings of information during testing  they will  appear in these fields     Use the arrow keys to access only the lines and columns that have characters received from the  gage  including spaces   Then overwrite the desired measurement number  including the decimal  point  in each line with pound signs     and save the template  Thereafter during standard use  each  time the gage transmits a string of characters  the software will pick up values located under the  positions designated by        s in the template they form and use these values as numeric input for  data entry  A maximum of 130 characters can be accommodated on each transmission line     When the       sign of the measurement value in the transmitted string is right in front of the  measurement  i e  they are contiguous   overwrite the sign with a      also  When they are not  contiguous  overwrite the sign with an    S 
137. ror expressed as a percent of total variation where total variation is  estimated by the part variation found in the parts used in the study  Total study variation TV is  similar to total process variation except that it uses part to part variation of parts used in the study to  estimate variation of the manufacturing process     Process Variation  Process variation refers to total variation in the manufacturing process  This is the Standard  Deviation of the most recent one hundred measurements    Tolerance   Tolerance is gage error expressed as a percentage of engineering tolerance for the selected  characteristic  i e  percentage of the spec range being used by error in the gaging system   The  engineering tolerance will be taken from values entered into the USL and LSL fields  If using a shared  DataLyzer Spectrum SPC software these values will be filled in automatically     Study Parameters  Clicking the Study Parameters button brings up the Study Parameters screen     Study Parameters x       Study Variation    GAR Attribute Study    Long Form C Short Form    Number of Appraisers  3 2to4  Number of Parts fi 0 5to 10  Number of Trials E 2t04    r Data Entry Method    Open Sheet    By Trial C Randomized      OK   Cancel      The study category  variable or attribute   and long or short form dictate what entries are acceptable  in the    Number of Appraisers        Number of Trials     and    Number of Parts    fields  The following list  contains recommended values for th
138. rum SPC module     Using the Gage Management System with the DataLyzer  Spectrum SPC Package with Microsoft Access  Microsoft SQL    Server  and Oracle Databases  If you are using both the DataLyzer Spectrum Gage Management System and the DataLyzer  Spectrum SPC packages  you should install the SPC packages first  if they are not already  installed  follow the instructions in the DataLyzer Spectrum SPC user s manual  After the DataLyzer  Spectrum SPC package s  has been installed  install the Gage Management System software     By default both programs are installed in the same folder  C  Program Files DataLyzer Spectrum    and will use a single database  in order to allow results to be shared by both programs     Note  When using a Microsoft Access database  ifthe SPC and Gage Management programs have  not been installed in the same folder or the dsdbase mdb database file has been installed to a    Appendices 135    location other than the default  such as a server   you will need to select the desired database using  SETUP  SELECT DATABASE in either package  You can see the path of the database currently  being used by the program by going to HELP  ABOUT DATALYZER     Using the Gage Management System Only   When using a Microsoft Access database  install the Gage Management System software following  the    installation instructions    below  The program is then ready to run with no additional steps  required     When using Microsoft SQL Server or Oracle databases  you must
139. s Individuals Charts Range  amp  Error Charts  Chart Type     Run Chart C Averages Chart       X Y Plot       3   a 6 3 5 Pat 4 8 1 2 10       Appraiser 1 Appraiser 2 Appraiser 3           Print      Print Exit                Individuals Charts Tab   On the    individuals charts    tab  part size is on the Y axis and parts measured follow the X axis   Option buttons include  normalized  whisker  and scatter plot  This tab only applies to long variable  studies     Normalized   Using the    normalized    option  scale is converted showing the grand average of all readings at 0  The  high and low measurements for each appraiser and each part are plotted as a deviation from the  grand average  The points for each operator are plotted in a different color and shape and connected  by a line  One might expect as many as two dots at the extremes  high and low  of each operator for  each part except when all measurements are the same for a given appraiser on a given part  In this   desirable  case only one dot will be plotted       45   DataLyzer Spectrum Gage Management                   GRER Average  amp  Range Method E  Statistics Data Performance Curve  Average Charts Individuals Charts Range  amp  Error Charts  Chart Type  oh ed    C Whisker   Scatter Plot    Data Values   Grand Average   Normalized       Appraiser 1 Appraiser 2    Appraiser 3            Print          Print   Exit             Whisker    Using the    whisker    option  for each part  a line  whisker  connects th
140. s customers set which TV criteria to use but most often the practitioner is left to decide   To compare variation values calculated during the study to total tolerance  the sigma values   standard deviation or estimated standard deviation  have traditionally been multiplied by 5 15  This  represent 99  of the expected full spread assuming a normal distribution curve  More recently a  multiplication factor of 6 has been used  this represents 99 7  of the full spread  DataLyzer  Spectrum Gage Management System can use either  choose the multiplication factor in the  Preferences screen     Guidelines for acceptance of gage repeatability and reproducibility    R amp R      Explanations        2 4    e Under 10  error  measurement is acceptable   e 10  to 30  error  measurement system may be acceptable depending on application  importance  cost of the gage  cost of repairs or improvement and existing process capability    e Over 30  error  measurement system needs improvement  depending on process capability    make every effort to identify problems and correct them     Source AIG   MSA Manual     Bias studies    Gage bias studies determine the difference between a measured value and a known value  To  determine bias in a gage  there must be    known    values to compare with readings taken by the gage   When the study is conducted  one operator measures one    known    standard or part ten times  and  the measured values are compared to the    known    value  The bias is defined 
141. s opened during a Calibration or R amp R  study     Example calibration gage   GR amp R Attachments Ea    GRR Measuring Instructions     GRA Control Plan Reference     lee a i       Attachment 1    Attachment 2    Attachment 3    Attachment 4    Attachment 5         Gage Studies    Data Menu    er      5 8    Measuring Instructions Field  The    Measuring Instructions    field shows instructions or procedures about conducting studies   Different instructions are stored for Calibration and GR amp R studies     Control Plan Reference  The    Control Plan Reference    field offers control plan cross reference     File Attachment Buttons   If attachments were included when the gage was defined  each of the five ATTACHMENT buttons  may link to a file or website to help describe procedures or work instructions for doing a study   Buttons with attachments will display the name of the linked file or website inside the label of the  button  clicking on the button will cause the system to follow the link and display the    attached    file  or website that is linked to it     Bias Study    The Bias Study window looks like this                                      Bias Data  File R E    Rd  Gage Name  D 1 caliper  Part   Gage Type   caliper y  Characteristic Serial Number   LSL  USL  Notes   Date   6 10 2008 Time   11 18  Location     Instructions    Process Variation    bserved Average  Bias    Bias     Master 1 2 3 4 5 6 2 8 9 10  Reading                Performed By Field and Button  
142. sheet    is the most flexible data entry    method     Long Attribute  The long attribute study uses a special spreadsheet for data entry     It looks like this        is  Gage Repeatability  amp  Reproducibility Study  File Edit Analyze Instructions    Performed By   Date  26 07 2010 Time  fiz07  Gage    Gage      r Characteristic  Part   LSL    USL     Characteristic                        GRA Attribute Study   Long 1 Appraisers 10 Parts 20 Trials Data Entry Method   Open Sheet    Appr  1  l Sort      Accepts                   Run each sample through the attribute gage 20 times and record the dimension  size  and enter the  number of accepts for each size in the corresponding    Accepts    field  The whole study should have  at least one sample within the gage   s limit with a result of    Go    for all 20 samples  and the study  should include at least one sample outside the gage   s limits resulting in 20    No Go    entries  Results  between the two extremes should have a variable number of acceptances  It is required that at least  six samples be measured where the number of acceptances is between 1 and 19  If the originally  selected pieces do not yield 6 samples with 1 19 acceptances  choose more samples from the  process with different measurements  taken by the variable gage   The    P    column is used to show  the estimated probability of acceptance and is calculated after data entry     File Menu   Save or  F2   The SAVE item  or pressing  F2  without opening the
143. smallest interval the gage can measure in the    Precision    field     Range of the Gage  Enter the range between the smallest and the largest sizes measurable by the gage in the    Range     field     32 DataLyzer Spectrum Gage Management    Manufacturer Button   No manual entry is allowed in the    Manufacturer    field  For entry in this field  use the  MANUFACTURER button to open the    Suppliers    list to select or create a manufacturer  To select a  predefined supplier s name from the list  double click on it or highlight it in the list and press the  SELECT button  To create a new supplier  enter a supplier name in the    Company    field and press  SELECT  The program will notify you that the supplier does not exist and ask if you want to create  it  Press YES  The    Gage Suppliers    setup window will then appear  see the description of this  window   Once a supplier is selected  it appears in the field  You can view more specific information  about the supplier by clicking on the field  a window containing read only information about the  supplier will appear      1 and  2 Gage Supplier Button   No manual entry is allowed in the     1 Gage Supplier    or     2 Gage Supplier    fields  For entry in these  fields  use the  1 GAGE SUPPLIER and the  2 GAGE SUPPLIER buttons to open the    Suppliers     list to select or create a supplier  Note that the    gage sales    option button is selected in this window   allowing only predefined suppliers of that type to appear
144. st one characteristic must be selected   checked for this button to work   All applicable characteristics will appear in the    Select  Characteristic    list     Show Selected  Pressing the SHOW SELECTED button hides all characteristics that are not selected as part of the    Setup    Setup Menu    or      characteristic group     7 7 Suppliers   New  Open  amp  Delete    The NEW SUPPLIER  OPEN SUPPLIER  and DELETE SUPPLIER items use nearly identical  windows to operate  After selecting one of these options from the SETUP menu  the following  window will appear         amp  Suppliers       Company Type  all  C Gage Sales    Gage Technical Support  C Gage Supplies C Gage Manufacturer     Gage Calibration    Certifying Agency    Company                       OK Cancel       Company Type   Use the    Company Type    option buttons to select a supplier type  If you select a type other than     all     when creating a new supplier  this will carry through to the    Gage Suppliers    setup window  The  option buttons determine what types of existing suppliers are contained in the list below the     Company    field     Company   Enter the new or existing supplier name in the    Company    field  Existing suppliers will appear in a  list below this field  To open or delete an existing supplier  you can type in the name or select it from  the list     OK Button    Press the OK button to open the    Gage Suppliers    setup window for the supplier name entered in the     Company    fi
145. stablish direct communication with electronic  gages via RS 232 protocol     Other program functions     e Scheduling for gage calibration  This facility requires the calibration interval to be entered for each  gage in its gage definition  then reports interpret and organize gages by calibration due dates    e Scheduling for GR amp R study  This facility requires the re evaluation interval to be entered in a  GR amp R Study Definition           pa DataLyzer Spectrum Gage Management    2    2 1    Explanations    Basic Concepts  Gage R amp R Calibration Basic Concepts    SPC analysis is largely affected by measurement system error  Measuring instruments  inspection  techniques of technicians and fixturing all make up a measuring system  Any of these components  can introduce variation in the measuring system  This variation in the measuring system is then  reflected as part of total process variation in the SPC analysis and negatively affects key quality  statistics like Cp  Ppl  Cpk  Ppk  Ppm  etc  The DataLyzer Spectrum Gage Management System  helps to quantify this measurement system variation so steps can be taken to reduce it     NX      Total Variation       Every process contains inherent variation  The sources of variation can be related to any of the  inputs from the mechanics of the process itself to raw material or personnel consistency  However   it is often true that the variation in the measurement process is the most significant  contributor to the overall vari
146. teristic group  the group will be shown     You can enter USL and LSL manually if the program is not being used in conjunction with Datalyzer   Tolerance  USL  LSL  is one of the Total Variation  TV  methods     If the selected characteristic is part of a characteristic group  then GR amp R studies performed using  this Study Definition can be considered as representative of all the characteristics in the group  If at  least one GR amp R study has been created using this Study Definition  the characteristic group will  become    locked     This means that you can only change the specified characteristic to another  characteristic in the group     Re evaluation required every    days   Ifthis box is checked a new GR amp R study with the Study Definition will be due after a specified time   Studies coming up for re evaluation can be seen in the Gage Lists   GRR Due report  The user can  also be alerted when the program is started if the time for a re evaluation is approaching     Parameters Button  Clicking the Parameters button brings up the Study Parameters screen     Study Parameters xi      Study Variation    GAR Attribute Study     Long Form C Short Form    Number of Appraisers  3 2to4  Number of Parts fi 0 5to 10  Number of Trials  3 2to4    r Data Entry Method     Open Sheet C By Trial C Randomized    OK   Cancel      The category of the study  variable or attribute   and long or short form dictate what entries are  acceptable in the    Number of Appraisers        Number
147. the  gage enough time to process each command     Secondary Gage Prompt   Sometimes instruments require a prompt to be sent causing them to take readings into their  memory  then another prompt to transmit those measurements from memory to the computer  As  one example  the GagePort s    computer mode    works in this manner  Enter into this field the  secondary prompt required by the gage for it to send from memory  As in previous cases  ASCII  characters and literals are supported     Reading Acknowledgment    Setup    Setup Menu    109      Sometimes instruments require the computer to return a    got it    message  confirming its  measurement was received before additional data can be collected  When this is true  put the string  expected by the instrument in the acknowledgment field  As in previous cases  ASCII characters  and literals are supported     Line Terminator for Lines Received by the Gage   The software must be told what character or set of characters will be at the end of each line sent by  the gage  The    line terminator for lines received by the gage    field is used to enter the line terminator  character s   In most cases  entries for this field will not be visible when typed  such as carriage  return and line feed  The default is carriage return  ASCII character 13      When more than one entry is made in this field  thereby establishing more than one character for  line termination  commas should be used between the characters  If the comma itself is a
148. this string automatically each time it is transmitted  overwrite the     X measurement value  4 379  with         signs  Since the    Y    measurement is in the same line  its  measurement should be overwritten with         signs  This way each time the transmission is sent by  the gage  the software will use this template to look in these column positions for input values            When templates of this type are used  the reading number of the transmission must be referenced in  the FUNCTION area of each characteristic  Characteristics looking for    X    measurement values  should have    R    or    R1    in their function fields  Characteristics looking for    Y    measurement values  should have    R2     in their function fields     The    reading description    area reports column positions where measurements reside  Zeros will  appear in the fields of this area until the template has been established through the testing  procedure     Com Port Assignments    The purpose of the com ports window is to assign existing gage interface template files  gage  drivers  to control particular communication ports on the computer  One gage driver can be assigned  to each com port  Data can then be sent directly from the gage to DataLyzer Spectrum Gage  Management software  Com port assignments can be changed at any time  If no gage is necessary   fields can be left at    No Gage    The DataLyzer Spectrum Gage Management software allows you to  communicate directly with a wide var
149. tion  6 sigma from long range SPC data  or  total study variation  6 sigma of data collected in the gage study only  can be used for calculating  the proportion used in measurement error  Customer requirements typically mandate which method  to use     Guidelines for acceptance of gage repeatability and reproducibility    R amp R      e Under 10  error  measurement is acceptable   e 10  to 30  error  measurement system may be acceptable depending on application importance  cost of the cage  cost of repairs or improvement and existing process capability    e Over 30  error  measurement system needs improvement  depending on process capability    make every effort to identify problems and correct them      Source AIG   MSA Manual   Gage Calibration  Bias    This is the difference between the    true     reference  value that the gage should be reporting and the     actual    measurement the gage is reporting       16   DataLyzer Spectrum Gage Management    Reference  Value    Bias    Observed  Average  Value    Calibration   Calibration is a one time look at a gage or measuring system under ideal conditions  temperature  controlled and clean environment   A trained and experienced technician carefully tests the  measuring system  By measuring a known standard once  the amount of bias for that standard size  is determined  Bias is the difference between the observed measurement of the gage and  the known standard being measured by the gage  Bias can often be reduced through gage
150. udy    window similar to the one below opens  the appearance of the  window may vary depending on whether it is a variable or attribute gage or due to other items such  as the study type   If you created a new study based on a Study Definition  some of the fields will  already be filled in     Gage Studies    Data Menu  IO    is  Gage Repeatability  amp  Reproducibility Study   Study Definition  10Days              lyze Instructions    H Date  23 07 2010 Time   17 45          Gage                   G 6111 Serial    12345  age   Category  Variable Type  Typel  Locked    Characteristic   Par    Parti is  fi ust  fico  Hin Kain e Category  Variable   _ Characteristics   VarCharl Group  FirstCG  Locked    Saved calculations  Cata  o Method   Average  amp  Range y   Total Variation Method   Study Variation y         Study Parameters   Variable   Long 3 Appraisers 10 Parts 3 Trials Data Entry Method   Open Sheet  Appr  1  Appr  2  Appr  3                   Appr  1   Appr  1   Appr 1   Appr  2   Appr 2   Appr  2   Appr 3   Appr 3   Appr  3  Trial  1 Trial  2 Trial  3 Trial  1 Trial  2 Trial  3 Trial  1 Trial  2 Trial  3                         Performed By Field and Button    Enter the name or initials of the supervisor overseeing the test in the    Performed by    field  You can  type it in directly or select from a list by clicking the PERFORMED BY button     Date  Enter the date the study was initiated in the    Date    field  The current system date is automatically  entered int
151. umber assigned to the gage by the company if this is  different from the entry in the Name ID  field     Date of Purchase  Enter the date the gage was invoiced by the supplier in the    Date of Purchase    field so in the future  you will know how old the gage is     PO     Enter the purchase order number used by your organization to purchase the gage in the    PO     field     Cost  Enter the price paid for this gage in the    cost    field     Status    Set the current status of the master by clicking on the Status button and selecting from the list  No  manual entry is allowed  you must pick from the list     Location  Set the current status of the master by clicking on the Location button and selecting from the list   You can also type in a location that is not in the list     Variable   Attribute   The    Variable    or    Attribute    items are used to further identify the gage type and the type of studies to  be performed  Variable gages enter numeric measurements  attribute gages give measurements  such as defect information or go no go information     Calibration Re evaluation Interval   Enter the interval between gage calibration analyses  Refer to your quality manual for this interval   Enter 0  zero  for gages that are only used for reference and do not need to be regularly studied  The  re evaluation period can be set in Days or Months  this is chosen using the in Preferences item in  the SETUP menu of the main program window     Precision of the Gage  Enter the 
152. uppliers    setup window will then appear  see here for a description of this window      Once a supplier is selected  it appears in the field  You can view more specific information about the  supplier by clicking on the field  a window containing read only information about the supplier will  appear     Other Supplies  Enter information about other supplies required that are specific to this master or master set in the     Other Supplies    field     Notes  Enter any further notes documenting this master or master set in the    Notes    field     Last Calibration Date  Enter the last calibration date for the master in the    Last Calibration Date    field     User Fields 1 12  If custom user fields have been created in SETUP  SYSTEM DEFINITIONS  these fields will appear  beneath the    Last Calibration Date    field  Enter the specified information as necessary     File Menu     Save or  F2   The SAVE item  or pressing  F2  without opening the menu  saves the changes made to the master     File Menu   Save  amp  Exit or  Ctrl   S   The SAVE 4 EXIT item  or pressing  Ctrl   S  without opening the menu  saves the changes made  to the master and returns to the main DataLyzer Spectrum Gage Management software window     File Menu   Exit or  Ctrl   X   The EXIT item  or pressing  Ctrl   X  without opening the menu  closes the    Define Masters    window  and exits to the main DataLyzer Spectrum Gage Management software window without saving     Setup    Setup Menu    ss      7 6
153. verage  Value Value   lower part of range   higher part of range     DataLyzer Spectrum Gage Management System calibration studies support up to 20 reference  points throughout the range ofthe gage  so bias can be detected at each end ofthe range or the  gage as well as numerous points in the middle       18   DataLyzer Spectrum Gage Management    2 2    2 3    Calibration    Calibration is the comparison of measurement and test equipment or measurement standard of  unknown bias in order to detect  report  or eliminate by adjustment any variation in the bias of the  instrument being studied  The process involves measuring several known standards up and down the  gage s range of measurement  then verifying any differences and average differences between the  gage s measurements and the true standard  If the difference is smaller than the allowed tolerance  for all calibration points  the gage passes the calibration study  Ifthe difference is larger than the  allowed tolerance for any calibration point  the gage fails the calibration study and must be adjusted   sent for repair  replaced or relegated to reference use only  Some gages have more than one  measurement mode  for example a calliper gage may have different jaws to measure outside  diameter  inside diameter and depth  The bias of each measurement mode will have to be  determined separately  Required bias is dependent on the application  along with the number of  points to be measured throughout its range     Gage R am
154. w window for short attribute GR amp R studies looks like this              Gage Repeatability and Reproducibility Report E  File  Part No   amp  Name  Gage Name  alt gage Date  772272008  Characteristic  Gage No   53334 Performed by  ALG  Specifications  Gage Type  User Field 1   Appraiser 1  Appraiser 2  Status Fail  Appr  1 Appr  1 Appr  2 Appr  2  Part Trial  1 Trial  2 Trial  1 Trial  2    D    D D  o  a                zzooonoooooooo zzonononm  05000000000000272000002z2z   D  oon2oooooooooooo DooDoo D    20  050000000000900027200000          In this short attribute study example  20 samples are measured by two appraisers in two trials each   Samples are selected in order to include examples of passing samples and failing products  Passing  products are represented on the report by    G    and failing samples in the list are represented by    NG      For the gage to pass  all trials must be the same across each row  In the example above you can  see entries on rows 1  2  7  8  19  and 20 are not the same  so this gage fails the R amp R study  If all  trials for each sample match  then the gage is deemed acceptable       s   DataLyzer Spectrum Gage Management    6 6    Gage R amp R   Performance Report    The preview window for a performance report looks like this        Performance Report E  File       Performance Curve  GRR 6 1 Bins 0     gt      gt     Probability of Acceptance     gt           Reference Part Measurements    Serial Number  Part  ne Serial Number  varchar
155. write the sign with a          When the       sign is not contiguous overwrite the sign  with      and add enough        s to the string to connect it with the measurement     Once created  use the TEST TEMPLATE item in the TEST menu so as not to disturb the template       signs  If the      sign template placement is incorrect  you will only get part of the measurement  value through to the software during data entry  A gage reading like 3 655 might come through as  655     There is a maximum of eight unique measurement values that the software can process per line   There is also a maximum of 64 values that can be processed for all lines of a full transmission  So   an eight line transmission could have a total of eight accessible measurements per line  Most of the  standard interfaces used  transmit only one measurement per line     The Read Gage Button   The READ GAGE button is used to set the software to receive a transmitted string  It will open the  designated com port and continue trying to receive a string from the gage for seven seconds  by  default   after which a    time out error    will be reported in the message field  This waiting interval can  be changed through the COMMUNICATION item in the OPTIONS menu     Save Button  The SAVE button stores all the template information  If no further changes are made to the main  gage template window  saving at the outer level would be redundant     Cancel Button  The CANCEL button exits without storing anything     Test T
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
User Manual (pdf 1002 KB)  Quelque chose de pourri dans le post-empire  Page 1 ` ) KDGBT61U KENWOOB L`autoradio Bluetooth Haut de  Ape50/100 テールランプキット 取扱説明書  Century rund (Bau-Anleitung)  Integrated multi-function meter User`s manual  Fiche technique  Insignia NS-B3112 User's Manual  C/CAR/DCA/10 – NE/17 Organización de Aviación Civil  manuel d`utilisation tronçonneuse cs-320tes cs-350tes    Copyright © All rights reserved. 
   Failed to retrieve file