Home
        Cdiff - BD Molecular Diagnostics
         Contents
1.     e Check reagent strips for proper liquid fills  ensure that the liquids are at the bottom of the tubes   see  Figure 1        Do not remove desiccant from reagent pouches       Do not use reagents if desiccant is not present or is broken inside reagent pouches       Do not use reagents if the foil has been broken or damaged       Do not mix reagents from different pouches and or kits and or lots       Do not interchange or reuse caps  as contamination may occur and compromise test results    e Proceed with caution when using chemical solutions as Master Mix and Extraction tube barcode  readability may be altered       Do not use expired reagents and  or materials       To avoid contamination by amplicons  do not break apart the BD MAX    PCR Cartridges after use  The  seals of the BD MAX    PCR Cartridges are designed to prevent contamination    e Performing the BD MAX    Cdiff Assay outside the recommended time ranges can produce invalid  results    e Additional controls may be tested according to guidelines or requirements of local  state  provincial  and or federal regulations or accrediting organizations    e In cases where open tube PCR tests are conducted in the laboratory  care must be taken to ensure that  the BD MAX    Cdiff Assay  any additional reagents required for testing  and the BD MAX    System are  not contaminated  Gloves must be changed before manipulating reagents and cartridges    e Always handle specimens as if they are infectious and in accordance w
2.  or negative for toxigenic C  difficile    Note  It is recommended that bacterial strains be freshly prepared in saline to a turbidity of 0 5 McFarland   1 0   X 10    CFU mL  from isolated colonies and subsequently diluted with saline to obtain a final concentration  of  3 3 x 10   CFUML    2  One  1  External Positive Control and one  1  External Negative Control should be run daily until  adequate process validation is achieved on the BD MAX    System  Reduced frequency of control  testing should be based on adequate data and determined by the individual laboratory    3  An External Negative Control that yields a positive test result is indicative of a specimen handling  and or contamination problem  Review the specimen handling technique to avoid mix up and or  contamination  An External Positive Control that yields a negative result is indicative of a specimen  handling preparation problem  Review the specimen handling preparation technique     P0137 01  2013 04    4  An External Control that yields an Unresolved  Indeterminate or Incomplete test result is indicative of a  reagent ora BD MAX    System failure  Check the BD MAX    System monitor for any error messages   Refer to the  System Error Summary  section of the BD MAX    System User s Manual   for  interpretation of warning and error codes  If the problem persists  use reagents from an unopened  pouch or use a new BD MAX    Cdiff Assay kit    5  Each BD MAX    Cdiff Extraction Tube contains a Sample Processing Co
3. 5 distinct days  wherein each day 2 panels were  tested by 2 technologists  at 3 clinical sites using 1 lot of reagents  Site to Site   One of these clinical sites  participated in an extended study where 2 additional lots of reagents were tested  Lot to Lot   Results are  shown for each specimen category     For Site to Site Reproducibility  the overall percent agreement was 100  for MP  LP and Neg categories   with 92 2  and 50 0  negative agreement for HN1 100 and HN1 10 categories  respectively  Table 8      For Lot to Lot Reproducibility  the overall percent agreement was 100  for MP  LP and Neg categories   with 96 7  and 64 4  negative agreement for HN1 100 and HN1 10 categories  respectively  Table 9      second Derivative Peak Abscissa  SDPA   an internal criteria used to determine a final assay result  was  selected as an additional means of assessing assay reproducibility  Overall mean SDPA values with  variance components  SD and  CV  are shown in Tables 8 and 9     Table 8  Site To Site Reproducibility Study Results using One Lot of the BD MAX    Cdiff Assay    SITE  1  Site 1 Site 2 Site 3 Overall Percent SUES  Percent Percent Percent Agreement  Overall   as    Agreement Agreement Agreement Mean SP    CV    Category    1For the Negative and High Negative categories  SDPA values reported are for the SPC  For other categories  SDPA values reported are for the toxigenic C   difficile target   2 For the High Negative categories  the expected assay result was deemed 
4. CR  Cartridge are sealed by the system prior to initiating PCR to contain the amplification mixture  thus  preventing evaporation and contamination     The amplified DNA targets are detected using hydrolysis  TaqMan    probes  labeled at one end with a  fluorescent reporter dye  fluorophore   and at the other end  with a quencher moiety  Probes labeled with  different fluorophores are used to detect tcdB and SPC amplicons in two different optical channels of the  BD MAX    System  When the probes are in their native state  the fluorescence of the fluorophore is  quenched due to its proximity to the quencher  However  in the presence of target DNA  the probes  hybridize to their complementary sequences and are hydrolyzed by the 5  3 exonuclease activity of the  DNA polymerase as it synthesizes the nascent strand along the DNA template  As a result  the  fluorophores are separated from the quencher molecules and fluorescence is emitted  The amount of  fluorescence detected in the two optical channels used for the BD MAX    Cdiff Assay is directly  proportional to the quantity of the corresponding probe that is hydrolyzed  The BD MAX    System monitors  these signals at each cycle of the PCR and interprets the data at the end of the program to provide a final  result     REAGENTS AND MATERIALS    Contents Quantity    BD MAX    Cdiff Master Mix   Dried PCR Master Mix containing tcdB specific 24 tests  molecular probe and primers along with Sample   Processing Control specific mol
5. OCEDURE   A liquid or soft stool specimen is collected and transported to the laboratory  For testing  a disposable 10uL  inoculating loop is dipped into the stool material and the contents dispersed into a BD MAX    Cdiff Sample  Buffer Tube  The Sample Buffer Tube is closed with a septum cap and vortexed  A worklist is created and  the Sample Buffer Tube  the BD MAX    Cdiff unitized reagent strip and the BD MAX    PCR Cartridge are  loaded onto the BD MAX    System  The BD MAX    System automates sample preparation including target  lysis  DNA extraction and concentration  reagent rehydration  and target nucleic acid amplification and  detection using real time PCR  The BD MAX    System performs results interpretation automatically  The  assay also includes a Sample Processing Control  SPC  that is present in the Extraction Tube  The SPC  monitors DNA extraction steps  thermal cycling steps  reagent integrity and the presence of inhibitory  substances     Following enzymatic cell lysis  the released nucleic acids are captured on magnetic beads  The beads  with  the bound nucleic acids  are washed using Wash Buffer and the nucleic acids are eluted by heat in Elution  Buffer  Eluted DNA is neutralized using Neutralization Buffer and transferred to a Master Mix to rehydrate  PCR reagents  After reconstitution  the BD MAX    System dispenses a fixed volume of PCR ready solution  containing extracted nucleic acids into the BD MAX    PCR Cartridge  Microvalves in the BD MAX    P
6. Preparation    section     CULTURING OF CLINICAL SPECIMENS  To perform species identification directly from stools  clinical specimens may be cultured using hospital  procedures     LIMITATIONS OF THE PROCEDURE    This product is intended for use only with liquid or soft stools  performance characteristics of other  clinical specimen types have not been established    This product can only be used on the BD MAX    System    Incorrect test results may occur from improper specimen collection  handling or storage  technical  error  sample mix up or because the number of organisms in the specimen is below the analytical  sensitivity of the test  Careful compliance with the package insert instructions and the BD MAX     system User s Manual     are necessary to avoid erroneous results    Good laboratory technique is essential to the proper performance of this assay  Due to the high  analytical sensitivity of this test  extreme care should be taken to preserve the purity of all materials  and reagents    A BD MAX    Cdiff positive assay result does not necessarily indicate the presence of viable  organisms  It does however indicate the presence of the tcdB gene and allows for presumptive  detection of a C  difficile toxigenic organism  The BD MAX    Cdiff Assay cannot be used for species  identification as it does not contain primers and probes specific to C  difficile    As with all PCR based in vitro diagnostic tests  extremely low levels of target below the limit of  detection o
7. R at  29  C for a maximum of 5 hours after the end of the run  If an Indeterminate  IND   Unresolved  UNR   or  Incomplete  INC  result is obtained  or if an External Control failure occurs  a repeat test from the Sample  Buffer Tube must be performed within this timeframe  see    Repeat Test Procedure    section      QUALITY CONTROL   Quality control procedures monitor the performance of the assay  Laboratories must establish the number   type and frequency of testing control materials according to guidelines or requirements of local  provincial   state and country regulations or accreditation organizations  For general QC guidance  the user may wish  to refer to CLSI MM03  and C2415     1  The External Positive Control is intended to monitor for substantial reagent failure while the External  Negative Control is used to detect reagent or environmental contamination  or carry over  by tcdB  amplicons  External Control materials are not provided by BD  Various types of External Controls are  recommended to allow the user to select the most appropriate for their laboratory quality control  program     e Commercially available control materials  e g   ATCC   43255  a C  difficile strain bearing the tcdB  gene  and ATCC   700057  a non toxigenic C  difficile strain  can be used as positive and negative  controls  respectively     e Suspensions of bacterial strains characterized by the user as toxigenic or non toxigenic    e Previously characterized specimens known to be positive
8. SBD MAX  Cdiff       BD MAX  Cdiff Assay REF  443418    For In Vitro Diagnostic Use  For use with the BD MAX    System    ye 9  Sr  me d IVD  ae a    INTENDED USE   The BD MAX    Cdiff Assay performed on the BD MAX    System is an automated in vitro diagnostic test for  the direct  qualitative detection of the Clostridium difficile toxin B gene  tcdB  in human liquid or soft stool  specimens from patients suspected of having C  difficile infection  CDI   The test  performed directly on the  specimen  utilizes real time polymerase chain reaction  PCR  for the amplification of C  difficile toxin B  gene DNA and fluorogenic target specific hybridization probes for the detection of the amplified DNA  The  BD MAX    Cdiff Assay is intended to aid in the diagnosis of CDI     SUMMARY AND EXPLANATION OF THE PROCEDURE   Clostridium difficile is an anaerobic  gram positive bacillus that is the leading cause of antibiotic associated  diarrhea and pseudomembranous colitis in health care facilities     Incidence of CDI has been increasing   and severe cases are becoming more common      CDI disease symptoms range from mild diarrhea to  severe colitis  and even bowel perforation and death  The most common risk factor is exposure to  antibiotics       The diagnosis of C  difficile infection is based upon clinical signs and symptoms  such as diarrhea  as well  as laboratory tests or pathologic finding consistent with toxigenic C  difficile  It appears that Toxin B is  necessary for the devel
9. acterial interference  Toxigenic C  difficile negative specimens and toxigenic C  difficile  positive specimens at 2 3X LoD were tested with the highest amount of each compound likely to be found  in the specimens or with interfering organisms  1 X 108 CFU mL of each strain   Potentially interfering  substances include calcium carbonate  Tums    as well as magnesium and aluminum hydroxide  Maalox    liquid   Results demonstrated no reportable interference with any other tested substance except for  Mesalamine rectal suspension enema and Gynol II   that both showed slight inhibition  delay of Second  Derivative Peak Abscissa  in the BD MAX    Cdiff Assay  however  expected assay results were still  obtained  Table 7      Table 7  Endogenous and Commercial Exogenous Substances tested with the BD MAX    Cdiff Assay    Weight or Weight or  Brand Name or Description Volume Result Brand Name or Description Volume Result  Tested SBT Tested SBT  NI    Nystatin 10 uL Pepto Bismol     Hyderm    Hydrocortisone  cream  0 0246 g  Glycerin Suppositories 0 0129 g Metronidazole  lhle   s Paste 0 0388 g  id     Anusol   Plus 0 01239 Polysporin   0 0240 g       I     P me   10305   1   Trayeorsemmicoci   10m   N  Waal ga    ow  Mesalamine Rectal P      Sipentonenena   E   t  Fleet   Mineral Oil Enema  TE    Gynol ETE o s  ON    Palmitic Acid    Imodium AD   0 0062 g    E  coli   non toxigenic C  difficile 10 uL     Interference with the BD MAX    Cdiff Assay   NI  No reportable interference wi
10. aluation of a real time PCR for Clostridium difficile toxin  A and B genes  Eur J Clin Microbiol Infect Dis 31  2219 2225    11  Clinical and Laboratory Standards Institute  Protection of laboratory workers from occupationally  acquired infections  Approved Guideline  Document M29  Refer to the latest edition     12  Centers for Disease Control and Prevention  Biosafety in microbiological and biomedical laboratories   Richmond JY and McKinney RW  eds   1993   HHS Publication number  CDC  93 8395    13  BD MAX    System User s Manual  Refer to the latest version  BD Diagnostics  Sparks  MD  USA    14  Clinical and Laboratory Standards Institute  Molecular Diagnostic Methods for Infectious Diseases   Approved Guideline  document MMO3  Refer to the latest edition     15  Clinical and Laboratory Standards Institute Statistical Quality Control for Quantitative Measurements   Principles and Definitions  Approved Guideline  Document C24  Refer to the latest edition      P0137 01  2013 04    16  Cohen SH et al   1998  Isolation of a Toxin B deficient mutant strain of Clostridium difficile in a case of  recurrent C  difficile  Associated Diarrhea  Clin  Infect  Dis  26  410 412    17  Cohen SH et al   2000  Analysis of the pathogenicity locus in Clostridium difficile strains  J  Infect  Dis   181  659 63    18  MacCannell D  et al   2006  Characterization of a novel  tcdB deficient  NPA1 variant strain of  Clostridium difficile  46th Annual ICAAC  San Francisco  Sept  2006    19  McFa
11. any additional specimens for testing by repeating Steps 1 through 5  then  proceed immediately  to Step 7    7  Vortex all prepared samples simultaneously at maximum speed for one  1  minute with the Multi Tube  Vortexer  The BD MAX    Cdiff Assay must be performed immediately after the vortexing step     BD MAX    System Operation    Note  Refer to the BD MAX    System User   s Manual for detailed instructions  Operation section    Note  The BD MAX    Cdiff Assay must be performed immediately after the vortexing step above     Specimen  Preparation     Step 7    Note  It is recommended to use the reagent tubes removed from their protective pouch  unreconstituted  within 3  hours   1  Power the System on  if not already done  and log on by entering  lt user name gt  and  lt password gt    2  Remove the required number of BD MAX    Cdiff Reagent Strips from the BD MAX    Cdiff kit  Gently  tap each strip onto a hard surface to ensure that all the liquids are at the bottom of the tubes   3  Remove the required number of Cdiff Extraction Tube s  and Cdiff Master Mix tube s  from their  protective pouches  Remove excess air  and close pouches with the zip seal   4  For each sample to be tested  place one  1  BD MAX    Cdiff Reagent Strip on the BD MAX    System  Rack  starting with Position 1 of Rack A and continuing sequentially with no open spaces   5  Snap one  1  BD MAX    Cdiff Extraction Tube  white foil seal  into Position 1 of each of the BD MAX     Cdiff Strips as shown 
12. at steps 9 to 12 for Rack B    14  Place the BD MAX    Cdiff Sample Buffer Tube s  in the BD MAX    Rack s  following the same order  as entered in the worklist     Note  Place the tubes into the sample rack with the 1D barcode labels facing outward  this makes  scanning tubes easier during sample login      15  Place the required number of BD MAX    PCR Cartridge s  into the BD MAX    System  see Figure 2      e Each cartridge accommodates 2 runs of up to 12 samples for a total of 24 samples    e The BD MAX    System will automatically select the position and row on the PCR cartridge for each  run    e Cartridges are used on a per run AND rack basis  2 runs per cartridge and 1 cartridge per rack        Figure 2  Load PCR Cartridges    P0137 01  2013 04    16  Load rack s  onto the BD MAX    System  Figure 3   Ensure that the placement of rack s   left to right   corresponds to the worklist created  top to bottom         side B  Figure 3  Load Rack s  into the BD MAX    System     17  Close the BD MAX    System lid and click the  lt Start Run gt  button to begin processing    18  At the end of the run check the results immediately  or store the Sample Buffer Tubes at 2 8  C for a   maximum of 5 days OR at 25  C for a maximum of 5 hours  until the results are checked    Note  If a septum cap was damaged during the run  replace it with a new one before storing the specimen    Note  BD MAX    Cdiff Sample Buffer Tubes can be stored at 2 8  C for a maximum of 120 hours  5 days  O
13. ation of warning and  error codes        REPEAT TEST PROCEDURE   Note 1  Sufficient volume is available for one repeat test from the Sample Buffer Tube on the BD MAX     system  For Sample Buffer Tubes stored at room temperature  retesting must be performed within 5 hours  after the end of the run  Alternatively  for Sample Buffer Tubes stored at 2 8  C  retesting must be performed  within 120 hours  5 days   The remaining stool specimen may also be used for repeat testing within 5 days  of collection if stored at 2 8  C or within 48h if stored at 2 25  C    Note 2  New samples may be tested in the same run with repeat samples     UNRESOLVED RESULT   Unresolved results may be obtained in the event that specimen associated inhibition or reagent failure  prevents proper target or SPC amplification  Sample s  can be repeated from their corresponding  sample Buffer Tube s  within the timeframe defined above  Vortex the sample s  for one  1  minute and  restart from the    BD MAX    System Operation    section  The remaining stool specimen may also be  used for repeat testing within the timeframe defined above  Restart from the    Specimen Preparation     section     INDETERMINATE RESULT   Indeterminate results may be obtained in the event that a System failure occurs  Sample s  can be  repeated from their corresponding Sample Buffer Tube s  within the timeframe defined above  Vortex  the sample s  for one  1  minute and restart from the    BD MAX    Operation    section  The rema
14. ecular probe     BD MAX    Cdiff Strips  Reagent strips containing all liquid reagents and          24 tests  443418 disposable pipette tips necessary for specimen  processing and DNA extraction     BD MAX    Cdiff Extraction Tube   Freeze dried pellet containing DNA magnetic  affinity beads  Achromopeptidase and Sample  Processing Control   BD MAX    Cdiff Sample Buffer Tube    Septum Caps    24 tests       EQUIPMENT AND MATERIALS REQUIRED BUT NOT PROVIDED       BD MAX    PCR Cartridges  BD Diagnostic Systems catalog no  437519   e  VWR Multi Tube Vortexer  VWR catalog no  58816 115      Vortex Genie 2  VWR catalog no  58815 234  or equivalent  to vortex stool specimens only     P0137 01  2013 04    e NALGENE   Cryogenic Vial holder  VWR catalog no  66008 783    e Disposable inoculating loops  10 uL    e Disposable gloves  powderless      Dry  clean containers for the collection of liquid or soft stool specimens   e  f culture is performed for External Controls  Pre reduced agar plate for anaerobes  e g   Brucella Agar  with 5  sheep blood  hemin and vitamin K1 plate  BBL    catalog no  297716     WARNINGS AND PRECAUTIONS       The BD MAX    Cdiff Assay is for In Vitro Diagnostic use       Do not use the kit if the label that seals the outer box is broken       Do not use reagents if the protective pouches are open or broken upon arrival    e Close protective pouches of reagents promptly with the zip seal after each use  Remove any excess air  in the pouches prior to sealing
15. ed molecular assays  demonstrated excellent overall agreement     In comparison to one FDA cleared molecular test  Table 5a   the PPA and NPA of the BD MAX    Cdiff  Assay were 99 1   Cl  94 9  99 8   and 97 4   Cl  95 7  98 4    respectively     In comparison to a second FDA cleared molecular test  Table 5b   the PPA and NPA of the BD MAX     Cdiff Assay were 95 5   Cl  92 1  97 5   and 98 8   Cl  97 9  99 3    respectively     P0137 01  2013 04    Table 5a  Results Obtained with the BD MAX    Cdiff Assay in Comparison  to a Commercially Available FDA cleared Molecular Assay for C  difficile     FDA cleared Molecular Assay 1    T oa    All Sites    BD MAX    Cdiff  Assay    Positive Percent Agreement  99 1   95  Cl  94 9   99 8    Negative Percent Agreement  97 4   95  Cl  95 7   98 4         Table 5b  Results Obtained with the BD MAX    Cdiff Assay in Comparison to  a Second Commercially Available FDA cleared Molecular Assay for C  difficile     FDA cleared Molecular Assay 2    a Sea    All Sites    BD MAX    Cdiff  Assay    Positive Percent Agreement  95 5   95  Cl  92 1   97 5    Negative Percent Agreement  98 8   95  Cl  97 9   99 3         Analytical Sensitivity   The analytical sensitivity  Limit of Detection or LoD  for the BD MAX    Cdiff Assay was determined as  follows  individual inoculating loops were dipped into a wide range of C  difficile bacterial suspensions at  different concentrations  prepared and quantified from cultures of 4 C  difficile strains represe
16. f the assay may be detected  but results may not be reproducible    Mesalamine rectal suspension enema and Gynol II   may cause slight inhibition in the BD MAX    Cdiff  Assay  refer to    Interfering Substances    section for further details     Tums   and Maalox   liquid may inhibit the BD MAX    Cdiff Assay  refer to    Interfering Substances     section for further details     False negative results may occur due to loss of nucleic acid from inadequate collection  transport or  storage of specimens  or due to inadequate bacterial cell lysis  The Sample Processing Control has  been added to the test to aid in the identification of specimens that contain inhibitors to PCR  amplification  The Sample Processing Control does not indicate if nucleic acid has been lost due to  inadequate collection  transport or storage of specimens  or whether bacterial cells have been  adequately lysed    BD MAX    Cdiff Assay results may sometimes be Unresolved due to an invalid Sample Processing  Control  or be Indeterminate or Incomplete due to System failure  and require retesting that can lead to  a delay in obtaining final results     P0137 01  2013 04    e Mutations or polymorphisms in primer  or probe binding regions may affect detection of C  difficile tcdB  gene variants  resulting in a false negative result with the BD MAX    Cdiff Assay    e Variant toxigenic C  difficile without the tcdB gene or with a non functional Toxin B protein are very  rare    19  The BD MAX    Cdiff Assay 
17. gistered trademark of the American Type Culture Collection    BD  BD logo and all other trademarks are property of Becton  Dickinson and Company      2013 BD    P0137 01  2013 04    
18. in Figure 1   6  Snap one  1  BD MAX    Cdiff Master Mix Tube  green foil seal  into Position 2 of each of the BD  MAX    Cdiff Strips as shown in Figure 1     P0137 01  2013 04    Waste Pipette Tips    Extraction Tube    Master Mix tube       aa Of    Figure 1  Snap BD MAX    Cdiff Extraction tubes and Master Mix tubes into reagent strips    7  On the BD MAX    software  select the  lt Consumable info gt  tab under the Run screen    8  Enter the kit lot number for the BD MAX    Cdiff Assay  for lot traceability  by either scanning the 1D  barcode with the scanner or by manual entry   Note  Repeat Steps 7 and 8 for each new kit lot number    9  Select the  lt Work List gt  tab  click on the  lt Assay gt  field and using the pull down menu  select  lt BD MAX  Cdiff gt   This will automatically populate the remaining assay fields for Rack A with    BD MAX Cdiff       10  Enter the BD MAX    Cdiff Sample Buffer Tube ID  Patient ID and Accession Number  if applicable  for  Position 1 of Rack A either by scanning the 1 D barcode with the scanner or by manual entry    11  Click on the  lt Lot Numbers field and using the pull down menu  select the appropriate kit lot number   on the outer box   This will automatically populate the remaining lot number fields for Rack A with the  same lot number    Note  Step 11 must be repeated for each new kit lot number    12  Enter the information for Position 2 of Rack A and continue for all remaining Sample Buffer Tubes in the  rack    13  Repe
19. ining  stool specimen may also be used for repeat testing within the timeframe defined above  Restart from the    P0137 01  2013 04       Specimen Preparation    section  For the interpretation of warning or error code messages  refer to the  BD MAX    User   s Manual        Troubleshooting       section      INCOMPLETE RESULT   Incomplete results may be obtained in the event that the Sample Preparation or the PCR did not reach  its expected time points  Sample s  can be repeated from their corresponding Sample Buffer Tube s   within the allowed timeframe defined above  Vortex the sample s  for one  1  minute and restart from     BD MAX    Operation    section  The remaining stool specimen may also be used for repeat testing  within the timeframe defined above  Restart from the    Specimen Preparation    section  For the  interpretation of warning or error code messages  refer to the BD MAX    Users Manual         Troubleshooting    section      EXTERNAL CONTROL FAILURE   External Controls should yield expected results when tested  If specimens have to be repeated due to  an incorrect External Control result  the specimens should be repeated from their Sample Buffer Tube  along with freshly prepared External Controls within the timeframe defined above  Vortex the samples  for one  1  minute and restart from the    BD MAX    Operation    section  The remaining stool specimen  may also be used for repeat testing within the timeframe defined above  Restart from the    Specimen  
20. ith safe laboratory procedures  such as those described in the CLS  Document M2911 and in Biosafety in Microbiological and  Biomedical Laboratories     e Wear protective clothing and disposable gloves while handling all reagents    e Wash hands thoroughly after performing the test       Do not pipet by mouth       Do not smoke  drink  chew or eat in areas where specimens or kit reagents are being handled    e Dispose of unused reagents and waste in accordance with local  state  provincial and or federal  regulations    e Consult the BD MAX    System User s Manual     for additional warnings  precautions and procedures     STORAGE AND STABILITY  Collected specimens should be kept between 2  C and 25  C during transport  Protect against freezing or  exposure to excessive heat     Specimens can be stored at 2 25  C for a maximum of 48 hours or at 2 8  C for a maximum of 120 hours  5  days  before testing     BD MAX    Cdiff Assay components are stable at 2 25  C through the stated expiration date  Do not use  expired components     P0137 01  2013 04    BD MAX    Cdiff Master Mix and Extraction Tubes are provided in sealed pouches  To protect product from  humidity  immediately re seal after opening  Reagent tubes are stable for up to 7 days at 2 25  C after initial  opening and re sealing     INSTRUCTIONS FOR USE    Specimen Collection Transport   In order to obtain an adequate specimen  the procedure for specimen collection must be followed closely   Using a dry  clean containe
21. itive results due to carry over contamination     REFERENCES   1  Dubberke ER  Wertheimer Al  Review of current literature on the economic burden of Clostridium  difficile infection  Infect Control Hosp Epidemiol 2009  30 57 66    2  Poutanen SM  Simor AE  Clostridium difficile associated diarrhea in adults  Can Med Assoc J  2004 171 51 8    3  Redelings MD  Sorvillo F  Mascola L  Increase in Clostridium difficile related mortality rates  United  States  1999 2004  Emerg Infect Dis 2007 13 1417 9    4  McDonald LC  Owing M  Jernigan DB  Clostridium difficile Infection in Patients Discharged from  US Short Stay hospitals  1996 2003  Emerging Infectious Diseases  2006  12  3  409 415    5  Pepin J  Valiquettte L  Alary ME  Clostridium difficile associated diarrhea in a region of Quebec  from 1991 to 2003  a changing pattern of disease severity  CMAJ  2004 171 466 72    6  Bignardi GE  Risk factors for Clostridium difficile infection  J Hosp Infect 1998  40 1 15    7  Lyras D  O Connor JR  Howarth PM  et al  Toxin B is essential for virulence of Clostridium difficile   Nature 2009 458 1176 9    8  Peterson LR  Robicsek A  Does my patient have Clostridium difficile infection  Ann Intern Med  2009 151 176 9    9  Peterson L R  et al   2007  Detection of toxigenic Clostridium difficile in stool samples by real time  polymerase chain reaction for the diagnosis of C  difficile associated diarrhea  Clin  Infect  Dis    45 1152 60    10  de Jong et al   2012  Clinical and laboratory ev
22. nting 3  toxinotypes  0  III  VIII   Each loop was then transferred to a SBT  already containing fecal matrix negative  for toxigenic C  difficile  Each C  difficile strain was tested in replicates of 24 per concentration  by 2  different operators  using 3 different production lots of the BD MAX    Cdiff Assay  Analytical sensitivity   LoD   defined as the lowest concentration at which 95  of all replicates tested positive  ranged from 125  to 265 CFU per loop  Table 6      Table 6  Limit of Detection of the BD MAX    Cdiff Assay  C  difficile Strain Toxinotype LoD in CFU per loop    ATCC  43255 265  95  Cl  140  502     ATCC   9689 156  95  CI  82  298     ATCC   BAA 1805 205  95  Cl  102  412     ATCC   43598 125  95  Cl  66  235        Analytical Inclusivity   A variety of toxigenic Clostridium difficile strains were included in this study taking into account geographic  Origin  toxinotype  NAP1 outbreaks and temporal diversity  Sixty four  64  strains including 23  toxinotypes  22 and representing 21 countries were tested  including strains from public collections and  well characterized clinical isolates  The assay correctly identified 62 of the 64 toxigenic C  difficile strains  which were tested at  3xLoD  Two  2  strains  strains IS25  toxinotype XII  and R9385  toxinotype XV    produced low signal results and were false negative in 1 out of 5 replicates     P0137 01  2013 04    Analytical Specificity   The BD MAX    Cdiff Assay was performed on samples containing ph
23. ntrol  SPC  which is a  plasmid containing a synthetic target DNA sequence  The SPC monitors the efficiency of DNA capture   washing and elution during the sample processing steps  as well as the efficiency of DNA amplification  and detection during PCR analysis  If the SPC result fails to meet the acceptance criteria  the result of  the specimen will be reported as Unresolved  An Unresolved result is indicative of specimen   associated inhibition or reagent failure  Repeat any specimen reported as Unresolved according to the   Repeat Test Procedure  section below     RESULTS INTERPRETATION   Results are available on the    Results    tab in the    Results    window on the BD MAX    System monitor  The  BD MAX    System software automatically interprets test results  A test result may be called as NEG   negative   POS  positive  or UNR  unresolved  based on the amplification status of the target and of the  sample Processing Control  IND  indeterminate  or INC  incomplete  results are due to BD MAX    System  failure  Results are based on the following decisional algorithm     ASSAY RESULT REPORTED INTERPRETATION OF RESULT  tcdB gene DNA detected      NEG No tcdB gene DNA detected    UR Unresolved     inhibitory specimen or reagent failure    Indeterminate due to BD MAX    System failure   with Warning or Error Codes         INC Incomplete Run   with Warning or Error Codes           Refer to the    Troubleshooting    section of the BD MAX    System User s Manual for interpret
24. on for which diagnostic  tests were indicated and ordered  Only soft or liquid stools  and only one specimen per patient  were  included     The Comparative Reference Method consisted of direct culture complemented by enriched culture   Enriched culture analysis was completed for all specimens that were negative for toxigenic C  difficile by  direct culture  The anaerobic culture was used to isolate C  difficile  if present  This was followed by  confirmation of the isolate identification and a Tissue Culture Cytotoxicity Assay to determine the  toxigenicity of the isolate     Of the 2071 soft or liquid stool specimens compliant with the Reference Method  Direct and Enriched  Culture   1819 gave compliant and reportable results with the BD MAX    Cdiff Assay  In comparison to the  Reference Method  the BD MAX    Cdiff Assay identified 87 7  of the toxigenic C  difficile positive  specimens and 96 8  of the toxigenic C  difficile negative specimens  Tables 1 and 2      P0137 01  2013 04    Table 1  Results Obtained with the BD MAX    Cdiff Assay in Comparison to the Reference Method    Reference Method  All Sites    BD MAX    Cdiff  Assay    Sensitivity  87 7   265 302   95  Cl  83 6   91 0    Specificity  96 8   1469 1517   95  Cl  95 8   97 6     PPV  83 5   95  Cl  79 4   87 1     NPV  97 7   95  Cl  97 0   98 4    1 Further investigation was performed on specimens with discordant results between the Reference Method and the BD MAX     Cdiff Assay    e 2  of 48 False Positi
25. opment of CDI     Laboratory diagnosis of toxigenic C  difficile includes anaerobic  culture followed by detection of toxin by tissue culture cytotoxicity testing    or by detection of toxin gene s   by PCR testing  While culture for toxigenic C  difficile followed by tissue culture cytotoxicity is highly  sensitive and specific  it is also highly technical  time consuming and has very slow time to result  48 to 96  hours  making this method impractical in the management of patients  Enzyme immunoassay  EIA  used  for the detection of toxin A and toxin B and glutamate dehydrogenase  GDH  an enzyme found in all C   difficile strains   are currently used in many clinical laboratories because results are available the same day   are easy to perform  and are relatively inexpensive  However  the sensitivity is low  especially for the toxin  EIAs  which can lead to missed cases of CDI    PCR methods for the detection of toxin A and or toxin B  have been developed and have demonstrated high sensitivity and specificity as compared to toxigenic  culture  cell cytotoxicity and immunoassays   9  Additionally  these tests can be performed in approximately  2 hours  The combination of a highly sensitive and specific assay with rapidly available results may allow for  prompt targeted treatment of CDI patients and thus potential improvement in patient outcomes  reduced  recovery times  and improved infection control practices     tes 1  F   Ofpkt    P0137 01  2013 04    PRINCIPLES OF THE PR
26. r  liquid or soft stool specimens are collected according to the following  procedure     1  Transfer liquid or soft stool  but not urine  into the container  Avoid mixing toilet paper  water or soap  with the sample    2  Label the container    3  Ship the container to the laboratory according to hospital standard operating procedures  Refer to     Storage and Stability    section      Specimen Preparation   Note  One  1  Sample Buffer Tube  one  1  Septum Cap  one  1  Master Mix  one  1  Extraction Tube and  one  1  Strip are required for each specimen and each External Control to be tested  Remove the required  number of tubes strios from their protective pouches or boxes  Remove the excess air and close the  pouches with the zip seal     1  Label a Sample Buffer Tube  clear cap  with the appropriate specimen identification making sure not to  obscure  write or label over the barcodes on the Sample Buffer Tube    2  Vortex the specimen at high speed for 15 seconds and dip a 10 uL inoculating loop into the liquid or soft  stool for testing  For soft stool specimens  remove any excess stool present on the outside of the loop in  order to obtain approximately 10 uL    3  Remove the cap from the Sample Buffer Tube then place the loop into the liquid  Roll the shaft of the  inoculating loop between your fingers in order to release the specimen in the tube    4  Seal the tube with a Septum Cap    5  Place the Sample Buffer Tube in a NALGENE   Cryogenic Vial holder    6  Prepare 
27. rland  L V   et al   Implications of the changing face of Clostridium difficile disease for health care  practitioners  Am J Infect Control  2007  35 4   p  237 53    20  Rupnik M  et al  A novel toxinotyping scheme and correlation of toxinotypes with serogroups of  Clostridium difficile isolates  J Clin Microbiol  1998  36  8   2240 2247    21  Rupnik M  et al  Comparison of toxinotyping and PCR ribotyping of Clostridium difficile strains and  description of novel toxinotypes  Microbiology 2001  147  439 447    22  Rupnik M  et al  New types of Toxin A Negative  Toxin B Positive strains among Clostridium difficile  isolates from Asia  J Clin Microbiol  2003  41 3   1118 1125     Definition    Temperature limitations    In Vitro Diagnostic Medical Device    Authorized Representative    Reseal pouch after use    The purchase of this product allows the purchaser to use it for amplification and detection of nucleic acid  sequences for providing human in vitro diagnostics  No general patent or other license of any kind other  than this specific right of use from purchase is granted hereby        This product is sold under license  and purchase of this product does not include rights to use for certain  blood and tissue screening applications  nor for certain industrial applications        Mm  v BD  BD Diagnostics Technical Service  1 800 638 8663    GeneOhm Sciences Canada Inc  2555 Boul  du Parc Technologique  Qu  bec  QC   G1P 4S5   Canada       Made in Canada    ATCC   is a re
28. s tested with the BD MAX    Cdiff Assay  58  3 1   were initially  reported as unresolved  42 of those were repeated and 32 were resolved upon repeat testing  Overall  0 5   remained unresolved after repeat  Table 4      Table 4  Unresolved Rate    Initial Unresolved Rate Unresolved Rate After Repeat       3 1   58 1860    95  Cl  2 4   4 0   0 5   10 1844   95  Cl  0 3   1 0        Total number based on compliant specimens and BD MAX    Cdiff Assay results     Out of 1916 soft or liquid stool specimens tested with the BD MAX    Cdiff Assay  21  1 1   were initially  reported as indeterminate  Out of a total of 1844 compliant results  no result remained Indeterminate upon  repeat  considering valid results and compliant repeats      Out of 1916 soft or liquid stool specimens tested with the BD MAX    Cdiff Assay  28  1 5   were initially  reported as incomplete  Out of a total of 1844 compliant results  no result remained Incomplete upon valid  repeat  considering valid results and compliant repeats      Comparison Studies   In addition to the multi site investigational study  a comparison study was performed using the BD MAX     Cdiff assay and two  2  commercially available FDA cleared molecular assays for the detection of the C   difficile tcdB gene  Testing was performed on 2013 specimens at two external sites  The Positive Percent  Agreement  PPA  and Negative Percent Agreement  NPA  of the BD MAX    Cdiff assay  in comparison to  the two commercially available FDA clear
29. targets the tcdB gene and it is unknown whether it would detect  Toxin At  Toxin B  variant strains    e     An excess amount of stool may inhibit the BD MAX    Cdiff Assay    e     As with all in vitro diagnostic tests  positive and negative predictive values are highly dependent on  prevalence  BD MAX    Cdiff Assay performance may vary depending on the prevalence and  population tested     EXPECTED VALUES   In the BD MAX    Cdiff Assay clinical study a total of 1834 reportable results  from specimens compliant at  the specimen and PCR levels  were obtained from 6 geographically diverse sites  The study population was  grouped into in patient  out patient and unknown categories  The number and percentage of positive cases   as determined by the BD MAX    Cdiff Assay  are presented in the table below     BD MAX    Cdiff Assay                Group bee Number of Number Number Positive Percentage  pecimens Positive Negative  In patient 1249 184 1065 14 7   184 1249   Out patient 457 114 343 24 9   114 457   Unknown 128 17 111 13 3   17 128   Total    1834 315 1519 17 2   315 1834                          1 Total specimens based on compliant PCR results     PERFORMANCE CHARACTERISTICS   Clinical performance characteristics of the BD MAX    Cdiff Assay were determined in a multi site  prospective investigational study  Six  6  investigational centers participated in the study  To be enrolled in  the study  specimens had to be from patients suspected of having C  difficile infecti
30. th the BD Cdiff    Assay       Mesalamine rectal suspension enema and Gynol II    with nonoxynol 9  showed slight inhibition  delay of Second Derivative Peak Abscissa   in the BD MAX    Cdiff Assay  however  expected assay results were still obtained        P0137 01  2013 04    Precision  Within laboratory precision was evaluated for the BD MAX    Cdiff Assay at one  1  site  The Precision  panel consisted of 5 specimen categories as follows     Moderate Positive  MP   2 5 x LoD   Low Positive  LP   1  2 x LoD   High Negative 1 10  HN1 10   10 fold dilution of 1 x LoD   High Negative 1 100  HN1 100   100 fold dilution of 1 x LoD   True Negative  Neg    Escherichia coli ATCC 25922 was added to every tube in the panel  C  difficile strains were tested in each  of the specimen categories with the exception of the negative  Neg  specimen category     Testing was performed in duplicate  over 12 days  with 2 runs per day  by 2 technologists  Precision study  results for Neg  LP and MP samples demonstrated 100  agreement  Precision study results for HN1 100  and HN1 10 demonstrated agreement of 95 8  and 58 3   respectively     Reproducibility   The Reproducibility panel consisted of 5 specimen categories as follows    Moderate Positive  MP   2 5 x LoD   Low Positive  LP   1  2 x LoD   High Negative 1 10  HN1 10   10 fold dilution of 1 x LoD   High Negative 1 100  HN1 100   100 fold dilution of 1 x LoD   True Negative  Neg    Specimens in each category were tested in triplicate  on 
31. to be negative  Therefore  percent agreement was calculated for negative results        P0137 01  2013 04    Table 9  Lot to Lot Reproducibility Study Results using Three Lots of the BD MAX    Cdiff Assay  LOT  Lot 1 Lot 2 Lot 3 Overall Percent SDPA Values       Percent Percent Percent Agreement Overall  Agreement Agreement Agreement Mean    Category  SD  CV    1For the Negative and High Negative categories  SDPA values reported are for the SPC  For other categories  SDPA values  reported are for the toxigenic C  difficile target    2 For the High Negative categories  the expected assay result was deemed to be negative  Therefore  percent agreement was  calculated for negative results        Carryover   Cross Contamination   A study was conducted to investigate within run carryover and between run carryover while processing  specimens with high bacterial load of toxigenic C  difficile in the BD MAX    Cdiff assay  A panel made of  one high positive member and one negative member was used to prepare numerous samples  A  Clostridium difficile strain  Tox 0  ATCC 9689  was used for the high positive C  difficile panel member    3x108 CFU mL   The negative member did not contain any target analyte  Twelve  12  replicates of the  high positive panel member and 12 replicates of the negative panel member were tested in each run by  alternating negative and positive samples  Three  3  operators performed 3 consecutive runs for a total of 9  runs of 24 samples  There were no false pos
32. ve BD MAX    Cdiff specimens were also found to be positive using another commercially  available FDA cleared RT PCR assay targeting the C  difficile tcdB gene     e 32 of 37 False Negative BD MAX    Cdiff specimens  were also found to be negative using another commercially  available FDA cleared RT PCR assay targeting the C  difficile tcdB gene        Table 2  Results Obtained by Site using the BD MAX    Cdiff Assay in  Comparison to the Reference Method    Sensitivity Specificity  90 0   18 20  95 3   202 212    69 9   97 2       91 5   97 4    84 4   38 45  95 3   205 215    71 2   92 3    91 7   97 5    94 7   36 38  97 9   275 281    82 7   98 5    95 4   99      96 1   49 51  98 4   186 189    86 8   98 9    95 4   99 5    83 8   93 111  98 2   279 284    75 8   89 5    95 9   99 2         Numbers in parentheses express the 95  confidence interval boundaries    83 8   31 37  95 8   322 336    68 9   92 3    93 1   97 5         In comparison to direct culture  the BD MAX    Cdiff Assay identified 96 5  of the toxigenic C  difficile  positive specimens and 92 7  of the toxigenic C  difficile negative specimens  Table 3      P0137 01  2013 04    Table 3  Results Obtained with the BD MAX    Cdiff Assay in Comparison to Direct Culture    Direct Culture    All Sites    BD MAX    Cdiff  Assay    Positive Percent Agreement  96 5   194 201   95  Cl  93 0   98 3    Negative Percent Agreement  92 7   1507 1625   95  Cl  91 4   93 9         Out of 1860 soft or liquid stool specimen
33. ylogenetically related species   Clostridium other than toxigenic C  difficile  and other organisms  bacteria  viruses  likely to be found in  stool specimens     e Six  6  out of 6 C  difficile strains not bearing the tcdB gene  tested at a concentration 2 1 X 108  CFU mL  produced negative results with the BD MAX    Cdiff Assay    e   Thirty  30  out of 30 Clostridium strains other than C  difficile  including 4 strains of C  sordellii  tested at  a concentration 2 1 X 108 CFU mL  produced negative results with the BD MAX    Cdiff Assay    e Ninety five  95  out of 98 other bacterial strains  including 93 species and subspecies  were tested at a  concentration 2 1 X 108 CFU mL  or   1 X 108 genomic DNA cp mL or 1 X 108 elementary bodies mL   and produced negative results with the BD MAX    Cdiff Assay  An investigation was conducted and  determined the false positive results were due to contamination  New suspensions of 3 strains were  tested and generated the expected negative results    e Seven  7  out of 7 viruses  tested at a concentration 2 1 X 10   PFU mL  produced negative results with  the BD MAX    Cdiff Assay     Interfering Substances   Twenty five  25  biological and chemical substances occasionally used or found in perianal  rectal and or  stool specimens were evaluated for potential interference with the BD MAX    Cdiff Assay  Two  2   organisms  E  coli ATCC 25922 and non toxigenic C  difficile ATCC 700057  were also tested at high loads  in order to assess b
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
PICO Valve SV-060 Bedienungsanleitung  babe/ babe swing kalle lars 800/900/1100/1300 mats pelle tor vidar  T-Scan III Manual  SPC-0433 W - ミマキエンジニアリング  デジ例レフォースゲージ取扱説明書 DS2  Graco 6989 User's Manual  V60 - Daray Medical  Operating Instruction    Copyright © All rights reserved. 
   Failed to retrieve file