Home

JEROME 431-X - Arizona Instrument LLC

image

Contents

1. In areas containing these highly volatile compounds the filter can quickly become saturated In such situations it is recommended that these gases be allowed to dissipate before sampling for the less volatile more persistent mercury vapor In addition a special filter designed to remove chlorine gas from the sample stream is available from Arizona Instrument and may be ordered as Chlorine Filter AZI P N Z2600 3940 Collection of air samples with Jerome gold coil dosimeters for analysis by the J erome mercury vapor analyzers will also eliminate interferences Ammonia in very high concentrations can cause an off gassing of accumulated acidic fumes from the internal acidic gas filter resulting in positive readings on the instrument In these cases the ammonia odors are very strong A special filter designed to remove ammonia gas from the sample stream is available from Arizona Instrument and may be ordered as Ammonia Filter AZI P N 990 0193 Again either allow the vapors to dissipate use the ammonia filter or use the dosimeters Filter replacement at regular intervals or when unexpectedly high readings are encountered in areas of these potential interferences may resolve these problems Visit the Tech Notes section at www azic com for more information concerning the chlorine and ammonia filters Volatile mercury compounds in general will cause the gold film to respond Alkyl organic mercuries such as methyl mercury and ot
2. 6000 1055 Jerome Communication Cable For current prices and delivery information call AZI Customer Service at 800 528 7411 or 602 470 1414 Service includes filter replacement component testing and instrument calibration to NIST traceable standards For scheduling and shipping authorization call AZI Customer Service at 800 528 7411 or 602 470 1414 AZI Customer Service 800 528 7411 or 602 470 1414 Page 36 of 67 11 APPENDIX A 431 X FUNCTIONAL TEST KIT If your application requires frequent verification of instrument functionality this test will benefit you If the test results fall within the expected range you may assume the instrument is functioning properly THIS IS A FIELD CHECK OF THE FUNCTIONALITY OF THE INSTRUMENT THIS TEST DOES NOT CALIBRATE THE INSTRUMENT NOTE Perform the functional test ONLY after a sensor regeneration The 431 X Functional Test Kit contains all accessories necessary to perform the functional test See the complete list on page 30 and verify that all the parts to the kit are present CAUTION The vial and thermometer contain liquid mercury and are possible sources of mercury contamination Follow the instructions for handling or transferring the mercury into the Functional Test Kit Vessel carefully For safety information see the supplier s Material Safety Data Sheets MSDS or call AZI Customer Service at 1 800 528 7411 or 1 602 470 1414 for assistance in obtai
3. Arizona Instrument LLO USER MANUAL JEROME 431 X MERCURY VAPOR ANALYZER OPERATION MANUAL January 2011 ARIZONA INSTRUMENT LLC 3375 N Delaware Street Chandler AZ 85225 USA 800 528 7411 602 470 1414 f 602 281 1745 www azic com Email General azi azic com International intl azic com Customer Service support azic com AZI P N 700 0046 D Last update January 2011 JEROME 431 X Mercury Vapor Analyzer Operation Manual PROPRIETARY RIGHTS NOTICE This manual contains valuable information and material developed by Arizona Instrument LLC for use with the Jerome 431 X Mercury Vapor Analyzer No part of this manual can be reproduced or transmitted in any form or by any means electronic mechanical or otherwise This includes photocopying and recording or in connection with any information storage or retrieval system without the express written permission of Arizona Instrument LLC ALL RIGHTS RESERVED Copyright 1990 2011 Arizona Instrument LLC Resisorb is a registered trademark of Avantor Performance Materials Tygon is a registered trademark of Norton AZI Customer Service 800 528 7411 or 602 470 1414 Page 2 of 67 Table of Contents 1 FOR THOSE WHO CAN T READ THE WHOLE MANUAL NOW cccccssssssssssssssees 5 Bix ANTRODUC LOIN ae nsec estcnceensecasescsessoutasscus costes ceensecunesssevsostsevcecseostssevessecstescscies 7 3 PRINCIPLE OF OPERA DION weiss scccesssncssccccscs
4. e Reinstall the cover e If the external 12volt lines are not powered power them now Connect them to the battery REPRESENTATIVE e Connect the instrument s AC power cord between the instrument and FRONT VIEW the front of the power inverter e Turn the instrument ON e Press the REGEN switch on the instrument Inverter operation can be verified in either of two ways gt Immediately after pressing REGEN the inverter will intermittently sing This tone slowly becomes nearly continuous and then ends after 64 seconds gt If the area is noisy use a voltmeter or test lamp to verify that approximately 115 volts is present for about 64 seconds starting when the REGEN switch is pressed e Allow the instrument to complete its regeneration before turning it off e With the instrument turned off complete the installation 1 e connect data logger communications cables or other devices and ensure that the DIP switches for the instrument and option board are set correctly AZI Customer Service 800 528 7411 or 602 470 1414 Page 65 of 67 16 WARRANTY Arizona Instrument LLC seller warrants to buyer that J erome products delivered pursuant to this agreement shall at the time of delivery and for a period of one 1 year Thereafter the Internal Battery Pack where applicable 1s warranted for a period of ninety 90 days only to be free from defects in material or workmanship and shall conform to s
5. Line voltage less than 100 VAC or less than 200 VAC for 220V units Cycles dipswitch set incorrectly Check power cord for connection Replace fuse Refer to page 22 Check line voltage settings Refer to page 21 Check input cycle settings Refer to 55 If fuse and line voltage are OK it may be circuit board adjustment or component failure Call AZI Customer Service at 800 528 7411 or 602 470 1414 Display reads E E E Very low battery Recharge battery Refer to page 17 Replace battery Refer to page 21 AZI Customer Service 800 528 7411 or 602 470 1414 Page 26 of 67 Range Sensitivity Precision Accuracy Response time sample mode Response time survey mode Flow rate Power requirements Fuse Internal battery pack Operating environment Case construction Dimensions standard model Dimensions XE model Weight standard model Weight XE model Digital meter display Certification 0 001 to 0 999 mg m 0 003 mg m Hg 5 relative standard deviation at 0 100 mg m Hg 5 at 0 100 mg m Hg 12 seconds 3 seconds 750 50ml min 0 75 05 liters min 100 120 VAC 50 60 Hz 1 A or 220 240 VAC 50 60 Hz 1 A FIA 250V 5mm X 20mm Rechargeable Nickel Cadmium 0 to 40 C non condensing non explosive Aluminum alloy 33 cm L x 15cm W x 10cm H 13 L x 6 W x 4 H 35cm Lx 18cm Wx 18 cmH 14 Lx 7 W x7 H 3 18 kilos 7 pounds 3 5 ki
6. e Press the Jerome 431 X power ON button e Inject 1 cc of mercury saturated vapor into the septum according to the Syringe Technique described on page 37 Appendix A e Make 3 additional 1 cc injections sample readings and record the displayed values include the decimal points e Average the results of the last 3 injections e Remove the septum assembly and zero air filter from the instrument e Connect the 10 1 dilution module to the instrument connect the septum assembly and zero air filter to the dilution module e Inject 1 cc of mercury vapor as above into the septum e Make three additional injections and average the three readings gt Divide this result into the average from the three direct injections e Use the result as the dilution module ratio in your dosimeter analysis Most Accurate Method Perform the above test however attach the dosimeter septum holder assembly and zero air filter to the sampling pump that will be used The technique is described in the next section Collection efficiencies should be approximately 100 AZI Customer Service 800 528 7411 or 602 470 1414 Page 49 of 67 e Connect your pump dosimeter septum holder assembly and zero air filter together with 1 8 and 3 16 Tygon tubing as shown to the right e Turn on the pump e Inject 10 cc s of Hg into the septum cc at a time gt A total of ten Icc injections one after another e Wait 30 seconds after t
7. 431 X has been factory calibrated using laboratory equipment containing NIST traceable permeation tubes In order to calibrate the Jerome 431 X a sophisticated calibration system is required that ensures stability of the calibration gas source eliminates any pressure in the calibration gas stream and controls the temperature of the calibration environment We strongly recommend you take advantage of our calibration and maintenance service at Arizona Instrument Call Customer Service at 800 528 7411 or 602 470 1414 to arrange re calibration A certificate of calibration is issued from AZI when your instrument 1s factory calibrated Verification of Functionality and Quality Control The Functional Test Kit FTK AZI P N Y431 0902 is used to determine if your instrument is functioning correctly between recommended annual factory calibrations It allows you to have complete confidence in the sample results This test verifies proper instrument operation through the introduction of a known concentration of Mercury Vapor into the J erome analyzer THIS IS A FIELD CHECK OF THE FUNCTIONALITY OF THE INSTRUMENT THIS TEST DOES NOT CALIBRATE THE INSTRUMENT If your application requires frequent verification of instrument function this test demonstrates the unit s operation and function Recording FTK results in an instrument log provides a quality control quality assurance record of instrument function between regular calibrations If test
8. This flashing signals the opening and closing of the solenoid sample bypass See the Principles of Operation on page 8 for details e At the end of the 12 second cycle read the digital meter gt The number shown on the digital meter is the mercury concentration in mg m or ng This value remains displayed until the next sample is taken The digital meter automatically zeroes at the start of each sample e When the sensor is completely saturated the digital meter displays 8 8 8 instead of a value No further operation is possible until a sensor regeneration is performed Refer to page 12 for the Sensor Regeneration procedure e Press the power OFF button when not in use Install the zero air filter in the instrument intake during storage Sampling Notes The Jerome 431 X is intended for vapor use only DO NOT allow the probe or the instrument s intake to come in contact with liquids dust or other foreign material Moisture or liquids drawn into the instrument can damage the sensor and flow system The Jerome 431 X Operates a minimum of 6 hours on a fully charged battery Use the probe AZI P N1400 2002 to locate mercury vapor in hard to reach places Plug the probe directly into the instrument s intake AZI Customer Service 800 528 7411 or 602 470 1414 Page 14 of 67 Survey Mode The survey mode takes samples every 3 seconds automatically Use this mode to locate mercury spills or to assess areas of potentially high merc
9. may become clogged and need replacement when sampling in a dusty or humid area p lt Open instrument and check for pinched crimped or disconnected internal tubing 2 In extreme conditions an additional particle filter may be installed on the intake High erratic results Readings vary more than 0 05 mg m3 when in survey mode Loose connections to gold film sensor Place a zero air filter into the intake Place the instrument in survey mode Move the unit as samples are being taken Call AZI Customer Service at 800 528 7411 or 602 470 1414 for assistance Low response or erratic readings after a long period of non use May need a second regeneration cycle 1 Wait 30 minutes and perform another sensor regeneration 2 Test with FTK Refer to page 37 3 If still unresponsive call AZI Customer Service at 800 528 7411 or 602 470 1414 for assistance AZI Customer Service 800 528 7411 or 602 470 1414 Page 25 of 67 False readings may go to 8 8 8 or LEL Extremely cold or extremely warm air sampled into unit If sampling under these conditions install zero air filter in intake Sample until display reads 0 003 mg m3 or less This equilibrates sensor temperature with the temperature of the sample air stream Remove filter and take samples Miscellaneous Problems Display reads P P P when regeneration is attempted Power cord not attached Blown fuse
10. about how to proceed call AZI Customer Service at 800 528 7411 or 602 470 1414 or send an e mail to support azic com for assistance AZI Customer Service 800 528 7411 or 602 470 1414 Page 56 of 67 JCS Kit Contents e Jerome Communication Software on CD ROM with security key e Jerome Communication Cable AZI P N 6000 1055 e User s manual System Requirements e Jerome 431 X with the Communications Option These Jerome instruments have a DB 25 connector and related internal hardware and firmware e Windows 98 Second Edition ME NT 2000 or XP At least one free serial port One free USB port Optional equipment e Jerome Data Logger AZI P N 6100 0010 to capture data without a computer nearby Data Logger Option The software can also program the Jerome Data Logger AZI P N 6100 0010 used with the Jerome analyzer The computer programs the data logger that then attaches to the DB 25 connector on the rear of the instrument The data logger initiates automatic sampling triggers alarms and stores data This optional accessory enables portable automatic sampling without a dedicated computer AZI Customer Service 800 528 7411 or 602 470 1414 Page 57 of 67 Proper use of this board requires that the base instrument be fully functional and set correctly for the intended operation ies 13 e NOTE If R101 is A Hot 3O moved from its P Dt v4 factory set position A Call customer service gt jii a at 8
11. average of the last three readings is still not within range refer to the section on Functional Test Troubleshooting below AZI Customer Service 800 528 7411 or 602 470 1414 Page 40 of 67 e Pull and hold the syringe plunger against the bar stop e Verify that the black mark on the syringe plunger aligns with the Icc mark on the syringe barrel gt If it does not the holder assembly must be adjusted Call AZI customer service at 602 28 1 1745 or 800 528 7411 for assistance e Insert the needle into the needle guide of the bottle stopper e Operate the plunger two or three times to pump mercury vapor into the syringe On the final stroke pull and hold the plunger against the bar stop e Holding the plunger against the bar stop remove the syringe from the bottle and move it to the septum attached to the instrument intake e Continue to hold the plunger against the bar stop and insert the syringe needle into the septum e Press SAMPLE on the instrument AZI Customer Service 800 528 7411 or 602 470 1414 Page 41 of 67 e When the display flashes release the plunger and allow gravity to feed the mercury vapor into the airstream If the plunger stops gently press it completely closed e Remove the syringe needle from the septum Temperature in C Digital Meter Response 091 to 123 100 to 135 108 to 146 129 to 174 138 to 187 151 to 204 164 to 222 177 to 240 118 to 159 AZI Cus
12. not turn the ZERO ADJUST potentiometer between samples Turn the ZERO ADJUST only after a sensor regeneration cycle otherwise invalid readings will result e Press the power OFF button and disconnect the power cord e The Jerome 431 X is ready for sampling CAUTION The Jerome 431 X is intended for vapor use only DO NOT allow the probe or the instrument s intake to be exposed to liquids dust or other foreign material Moisture or liquids drawn into the instrument can damage the sensor and flow system AZI Customer Service 800 528 7411 or 602 470 1414 Page 13 of 67 Sample Mode This mode used for standard operation produces optimum accuracy 5 at 0 100 mg m Hg with the Jerome 431 X e Press the power ON button gt The digital meter displays 000 If the unit is set to display in ng the digital meter displays 00 0 Disregard the digital meter s initial momentary readings Recharge or replace the battery pack if the LO BAT indicator REMAINS ON Refer to pages 17 and or 21 for the procedure e To ensure the instrument s electronics have stabilized allow a 1 minute warm up before beginning the next step e Press the SAMPLE button gt During the sampling cycle the bar or bars shown on the digital display indicate the current percentage of sensor saturation Refer to Meter Display Codes page 10 for code descriptions gt The bar or bars flash after 2 seconds and again after an additional 7 seconds
13. of 67 e In some cases the instrument cannot resume sampling after regeneration L L L appears on the display when the ZERO button is pressed and the error message manual bridge adjust needed is added to the notes column of the JCS text file when the JCS is used If this problem persists it may be necessary to re set the auto zero e When necessary to re adjust the auto zero point gt Turn the instrument off gt Make a note of the original DIP switch settings of SW100 on the option board gt On red DIP switch on the control board SW2 turn DIP switch 4 to ON gt Set the switches on the option board s blue DIP box SW100 to 1 2 6 OFF 3 4 5 ON gt Turn the instrument ON gt Press and hold the ZERO button and adjust the potentiometer on top of the instrument until the number 20 is displayed then release the ZERO button gt Switch option board DIP 1 from OFF to ON three times leaving it set to ON i e starting from OFF switch it ON OFF ON OFF ON gt Press and hold the ZERO button while turning the potentiometer on the option board until the number 20 is displayed Note the display may flicker one digit gt Return all switches to their original position NOTE The higher the auto zero number the lower the sensor capacity and the more sensor regenerations are needed Timed Regeneration If the unit is to be operated unattended for extended periods AZI recommends that the sensor be regenerated regular
14. on page 21 3 PRINCIPLE OF OPERATION A thin gold film in the presence of Mercury Vapor undergoes an increase in electrical resistance proportional to the mass of Mercury Vapor in the sample When the SAMPLE button is pressed an internal pump pulls ambient air through a scrubber filter and into the flow system Intake Scrubber Scrubber Gold Film Sensor Acidic Pump Gas Sample Filter Solenoid Closed After 2 seconds the sample solenoid bypass opens closing off the scrubber filter from the flow system Intake Scrubber Scrubber Gold Film Sensor Acidic Pump Gas Sample Filter en AZI Customer Service 800 528 7411 or 602 470 1414 Page 8 of 67 The sample air passes through a filter removing any acidic gases which interfere with the sensor s response to mercury and is drawn over the gold film sensor The sensor absorbs the Mercury Vapor Nine seconds after starting the sample solenoid bypass closes and the remainder of the sample is drawn through the scrubber filter and the flow system The instrument determines the amount absorbed and displays the measured concentration on the digital meter in milligrams per cubic meter mg m of mercury An internal DIP switch can be used to change the digital meter display from mg m to nanograms of mercury see page 22 The instrument s microprocessor automatically re zeroes the digital meter at the start of each sample cycle and free
15. on the entire range 0 999 mg m3 Note that neither analog output circuit is floating The negative terminals of both circuits are connected to the instrument s common ground buss SW101 Functions 0 2V analog output I 4 20 ma analog output AZI Customer Service 800 528 7411 or 602 470 1414 Page 60 of 67 Jerome 431 X instruments shipped after early 1995 are capable of providing 0 2 volts analog output Instruments shipped before that time can be upgraded by a firmware update and adjustment Instruments that are capable of 0 2 volt output can be upgraded to the 4 20 mA output with the addition of an option board upgrade This must be installed at the factory Connection and Setup e 0 2 volt devices connect as shown in Figure 1 If the instrument includes an option board be sure its analog jumper SW101 is set to the V position o O ogo oO o o opyo OO 8 CO 9O o0o0 0 ooo 90 REAR OF CONNECTOR CONNECT PINS 18 AND 23 FOR 0 2 VOLT OUTPUT JUMPER ON BOARD IS AT V Figure 1 e The 4 20 mA active receivers connect as shown in Figure 2 The active receiver contains a voltage source to power the loop current The receiver must have an isolated input circuit That is it must not be connected to ground or to a voltage source referenced to ground Be sure that jumper SW101 is set to the T position before power is applied e 431X reading Current 0 000mg m3 4mA 1 000mg m3 20mA e The 431X cur
16. read this manual before attempting operation If you have any questions about your application or operation please call AZI Customer Service at 800 528 7411 or 602 470 1414 or e mail support azic com for assistance 431 X Features e Automatic sensor regeneration when equipped with the communications option and used with the Jerome Communication Software JCS program and the J erome data logger e Regulated film heat voltage during sensor regeneration This allows the sensor to clean properly with voltages from 100 130 VAC or 200 260 VAC e Survey mode can be locked in e DIP switch setting can change the digital meter readings from mg m Hg to nanograms ng of Hg see page 22 e The Jerome 431 X can be operated from 100 130 or 200 260 VAC To change the default voltage range refer to Setting the Input Voltage page 21 Accessories and Maintenance Parts The Accessories and optional items available to support the 431 X are listed and pictured beginning on page 30 AZI Customer Service 800 528 7411 or 602 470 1414 Page 7 of 67 Applications Ambient air analysis Regulatory compliance Quality control Scrubber efficiency testing Accuracy check for other Mercury Vapor monitors and control systems Mercury Vapor source detection Leak detection Portable Mercury Vapor detection The Jerome 431 X can be operated from 100 120 or 200 240 VAC To change the default voltage range See Setting the Input Voltage
17. steps gt At least once a month wait until LO BAT appears on the digital meter before recharging the battery pack gt Charge the battery pack when the LO BAT indicator comes on Excessive discharge can damage the battery pack gt Before storing the instrument verify the power is OFF e When the batteries fail to hold a charge the battery pack should be replaced gt Battery life under normal usage is approximately 1 year depending on the number of charge and discharge cycles AZI Customer Service 800 528 7411 or 602 470 1414 Page 17 of 67 5 MAINTENANCE Preventive Maintenance Calendar To keep the J erome 431 X operating at peak performance follow the maintenance schedule below as a guide Since maintenance is more a function of application and amount of use rather than time your requirements may be different from the listed schedule Call AZI Customer Service at 800 528 7411 602 470 1414 or e mail support azic com for additional guidance for your environment and operation PART COMPONENT MAINTENANCE CYCLE a At least once per month after month s storage Charge batteries or when LO BAT appears Page 17 Change 25 inch fritware Weekly or as needed Page 19 rare miter era After 6 months of use or as needed Page 20 Replace zero air filter Annually Factory calibration Annually Page 23 _ Appendix Calibration check Monthly or as needed A Page 37 Annually or as needed Page 21 Replace batteries The ba
18. 00 528 7411 or 602 470 1414 I 5 i i t i YA DIP Switcha l SW 100 A ihia Jumper Sw101 With the option board installed the 431 X has a limited auto zero function This function cannot be disabled and is transparent to the user The instrument can be manually zeroed as described in Zero Adjust on page 13 However if the instrument is to be operated by personnel not familiar with the procedure or if it is operated unattended the auto zero function should satisfactorily zero the unit after each sensor regeneration instrument Zeroing The Jerome 431 X has essentially three ways to zero the sensor reading before samples are taken if the option board is installed e The instrument automatically re zeroes between samples so that each sample is a unique reading To take a sample simply press the SAMPLE button e The manually adjusted zero using the switch on the top of the 431 X is used to re establish a baseline between the reference and sensor gold films only after a sensor regeneration This zero is manually adjusted by pressing the ZERO button and turning the potentiometer on the top of the instrument until the display reads 0 Adjust only after sensor regeneration it is normal for H to be displayed after sampling e The 431 X option board provides an auto zero feature following regeneration that is invisible to the user AZI Customer Service 800 528 7411 or 602 470 1414 Page 58
19. 104 mg m 0 103 mg m average Loading the dosimeter 0 120 mg m 0 113 mg m 0 100 mg m 0 111 mg m Step 1 above 0 103 mg m x 10 1 030 mg m Step 2 above 1 030 mg m 0 111 mg m 9 4 Dilution module ratio 9 4 1 NOTE For normal applications a X8 to X12 ratio is recommended If your ratio is not within this range call Customer Service at 800 528 7411 for assistance AZI Customer Service 800 528 7411 or 602 470 1414 Page 51 of 67 NOTE Wait a minimum of 30 minutes after sensor regeneration before starting this procedure e Connect the system as shown in the figures below Jerome iluti JT LNSWNYLSNI YNOZINV e Attach the power cord to the 431 X and plug it into AC power gt AC power is required to heat the dosimeter e Press the Jerome 431 X power ON button and then press SAMPLE button The digital meter reading appears in 17 seconds e Record the digital meter reading include the decimal point Wait 30 seconds then press SAMPLE button again and record this reading gt Repeating the heating process ensures complete release of mercury from the dosimeter coil e Add the two digital meter readings together gt The sum of the two digital meter readings is the figure you will use in your calculations and is referred to as the meter response e The following calculations will provide the mercury concentration in mg m based on a time weighted average Dilution Mo
20. 32 2500 3001 1 of 1 8 Tygon tubing 2600 3039 25 inch fritware Z2600 3905 Zero Air Filter Z2600 3928 C M Filter Z2600 3930 Scrubber Filter Z4000 0907 Battery Pack Assembly Y990 0175 or Y990 0176 990 0177 990 0159 X412 0901 2500 3001 2500 3010 2100 6017 1300 0031 Z2600 3905 AZI Customer Service 800 528 7411 or 602 470 1414 Dosimeter Analysis Kit 110 VAC applications Dosimeter Analysis Kit 220 VAC applications Pocket Pump calibrated to 5 ml min Converter 220VAC to LIOVAC 200 WATT supplied with Y990 0176 for locations using 220 VAC power Personal Mercury Dosimeter 2 2 of 1 8 Tygon tubing 1 of 3 16 Tygon tubing Dosimeter Lead Set 3 16 to 1 8 Reducer Zero Air Filter Page 30 of 67 Jerome Data Logger an Y990 0169 Includes the Jerome Data Logger gt and JCS Software Kit ray yee sume DATA LOGGER J E ROME SYSTEM c MM Y990 0175 Personal Dosimeter Kit 110 VAC Includes the Pocket Pump two or mercury dosimeters zero air Y990 0176 filter regeneration cable 220 VAC connecting tubing and adaptor Y990 0168 PACATE J erome Communication 990 0170 Software Kit JCS without cable Hard Side Carry Case Includes a molded case with die Dens cut foam rubber inserts to hold the Jerome 431 X and accessories AZI Customer Service 800 528 7411 or 602 470 1414 Page 31 of 67 Soft Field Carrying Case 1400 0052 Hand shoulder case with pockets for acc
21. 528 7411 or 602 470 1414 Page 45 of 67 e Connect the system as shown gt Insert the dosimeter s large end in the 431 X s intake and gently tighten the intake tube nut to ensure an airtight seal gt Connect the Dosimeter Lead Set clips as shown Short red lead to the rear and long black lead to the far end e Attach the power cord to the 431 X and plug it into AC power AC power is required to heat the dosimeter e Attach the Dosimeter Lead Set 25 pin connector to the respective 25 pin communications port e Press the instrument s power ON button e Reading the dosimeter dosimeter desorption gt Press the SAMPLE button The digital meter reading appears in 17 seconds gt Record the digital meter reading include the decimal point gt Wait 30 seconds gt Press SAMPLE again and record this digital meter reading Repeating the heating reading process ensures complete release of mercury from the dosimeter coil e Add the two digital meter readings together The sum of the two digital meter readings is the figure you will use in your calculations and is referred to as the meter response MR NOTE A third dosimeter desorption pressing the SAMPLE button should give a reading of 005 mg m or less AZI Customer Service 800 528 7411 or 602 470 1414 Page 46 of 67 Perform the following calculation to obtain the mercury concentration in mg m based on a time weighted average or alternately DIP switch 2 can be set to OFF
22. GENTS OR REPRESENTATIVES AND EXCEPT FOR WARRANTY OF TITLE IS IN LIEU OF ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED INCLUDING ANY IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE AND MERCHANTABILITY SUCH OTHER WARRANTIES BEING SPECIFICALLY DISCLAIMED BY SELLER IN NO EVENT SHALL EITHER PARTY S LIABILITY FOR ANY BREACH OR ALLEGED BREACH OF THIS AGREEMENT EXCEED THE TOTAL EXTENDED PRICE OR PRICES SHOWN ON UNFILLED ORDERS NOR SHALL EITHER PARTY BE LIABLE FOR ANY SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM BREACH OR ALLEGED BREACH Notwithstanding the foregoing if any product covered by order s placed hereunder is designated as developmental prototype or experimental no warranty whatsoever except a warranty of title to component materials will be applicable thereto and buyer shall indemnify seller for any claims for liability asserted seller in connection therewith The foregoing state the entire liability of seller in connection with products supplied hereunder AZI Customer Service 800 528 7411 or 602 470 1414 Page 66 of 67 TRADEMARK AND COPYRIGHT PROTECTION Jerome Arizona Instrument AZI and the stylized AZI are all registered trademarks of Arizona Instrument LLC Instrument firmware is copyright protected All specifications subject to change without notice Copyright 1990 2011 Arizona Instrument LLC All Rights Reserved Resisorb is a registered trademark of Avantor Pe
23. alata a caiaas eicc ed estes iS ah scene A 22 6 CALIBRA TION orrai eara S EE T a E T EE 23 Verification of Functionality and Quality Control sssnnoneeeeeeesssssssssssssseseeeerressssssssssssssseeeeeeeee 23 Te 431 X TROUBLESHOOTING wae iscecscsicssceseicsecdasiesstidesvestecteniscecesedecensvessinssacosecsiusistsiesedecessutin 24 8 JEROME 431 X TECHNICAL SPECIFICATIONS cccccscscsssssssssssssssssssssssssssssssssssseseees 27 Optional Communications Capability cc csesssssssssseeecceeeeeeceeeeaaaeeseeseeeeeeeceeeeeeeeseesaaaaegseesess 21 listamenn VO MEIC E e N 28 Potential Interr erences arin na E E 29 9 ACCESSORIES amp MAINTENANCE PARTS eeeeeeeccecccccccccccccccccccccccccccccccccccccccccccccccccccccceee 30 PE a E REE NUN Ns Oe ee MER ONO 32 10 FACTORY CALIBRA TION SERVICE wccsccessvicsiciacercctsceccunscoceeatssssesesseavaatecssnscaticocessdassesessne 36 11 APPENDIX A 431 X FUNCTIONAL TEST KIT 1 0 0 ccccssssssssssssssssssssssssssssssssssssees 37 IPS AR ACO ee na stash eda a as aetna awn anets 3T Mercurs TRANS ie eseri a E 38 MW CSSCl DD ISASSEMADI y reina a a a a a A NA 38 Repic me Mern U arei E 39 Punctional Fest Proce ure aean E mtanie dadesasaonaninataneim ah antenieteanseanerauts 39 Sy Ge Pech a Qe xing cee ort eas dicta ett nee aie tu deli antes wkend ace aur nia acinus ashanti 4 A31 X Temiperature Conversion Cart sesices iii eto Oa ene ies 42 Funchonal Test Tronbleshooin 1c ccc ns Biel oR SN acta d
24. and the digital meter will display nanograms Hg directly refer to diagram page 55 Working Formula and Units of Measure Meter Response X ConversionFactor Pump FlowRate X Sampling Time Sample Concentraton Where Meter Response Total of the two digital meter readings in mg m 87 5 ng mg m a constant which changes the meter response to nanograms of Hg Conversion Factor Pump flow rate 5 0cc per minute calibrated value of the supplied SKC Pocket Pump Sampling time Duration of the sample in minutes Sample concentration Inng cc mg m EXAMPLE To calculate a time weighted average during an 8 hour period using the following values Meter response 0 600 mg m sum of the two meter response readings Conversion factor 87 5 ng mg m constant Pump flow rate Scc min Sampling time 8 hours 480 min e Convert the meter response the total of the two digital meter readings to nanograms of mercury gt 0 600 x 87 5 52 5 nanograms of Hg e Determine the total volume of air sampled gt 5 cc min x 60 min hr x 8 hr 2400 cc e Determine the Hg concentration time weighted average of the dosimeter 52 5ng 2400cc 0 022 ng cc of Hg 0 022 mgm of Hg AZI Customer Service 800 528 7411 or 602 470 1414 Page 47 of 67 Check the sensor status after each dosimeter analysis IMPORTANT To prevent the loss of a sample perform sensor regeneration as soon as the display shows f
25. ce the battery bracket and tighten the captive screws Connect the new battery connector to the board Close the case and replace the two 2 side screws Dispose of the old NiCad battery in accordance with state and federal regulations Instruments are factory set and calibrated to use the power setting requested on the order However the voltage setting is easily changed to use either 110 or 220 VAC e Ensure the instrument is turned OFF and eee cab olla ees de E unplugged i oa chiles mee ey e Locate the voltage selector on the rear of the gate Ps N instrument SE ENN AN i e Insert a small screwdriver in the voltage selector slot and turn the selector until the arrow points toward your setting choice and a click is heard AZI Customer Service 800 528 7411 or 602 470 1414 Page 21 of 67 e Remove the two screws toward the front of the instrument and open the lid e Locate the DIP Switch SW2 at the top of the main circuit board This is the red on illustrated at right e Set SW2 position 1 and 6 as follows a t 4 f ef e i o a TS FA 60Hz 50Hz Position 1 OFF OFF Position 6 OFF ON e CEOL Tooo yom A ies ACA AA FEFE PCEF BE m m me Pa de ot oe em ot 0209 Thy NOlDO WD Ly The instrument is factory set to display mg m milligrams per cubic meter of mercury XXX For some applications including dosimeter analysis the instrument s display can be converted to nano
26. dule Ratio l Meter Responseconverted to nanograms Ng of mercury X _ Sample Concentration Sample Volume e Alternately DIP switch 2 can be set to OFF and the digital meter will display nanograms Hg directly AZI Customer Service 800 528 7411 or 602 470 1414 Page 52 of 67 MR meter response total of the two digital meter readings in mg m conversion factor a constant which changes the meter response to nanograms of Hg DM dilution module ratio the ratio determined on page 51 87 5 ng mg m SV sample volume ae i rate in cc min multiplied by sample time in Sample concentration ng cc mg m EXAMPLE Assume the following values Meter Response MR 0 600 mg m sum of the two meter response readings Sampling time 8 hours 480 min Sample volume 5 cc min X 480 min 2400cc A time weighted average during an 8 hour period 1s calculated by 0 600 mg m X 87 5 ng mg m 0 021875 ng cc 2400 cc e Convert the meter response the total of the two digital meter readings to nanograms of mercury gt The MR 0 600 multiplied by the conversion factor 87 5 equals nanograms of mercury gt 0 600 x 87 5 52 5 nanograms of Hg e Determine the actual mass of Hg collected by the dosimeter gt Nanograms of mercury multiplied by the dilution module ratio gt 52 5 nanograms x 9 4 493 5 nanograms e Determine the total volume of air sampled gt The pump flow rate multiplied by 60
27. e on page 22 AZI Customer Service 800 528 7411 or 602 470 1414 Page 12 of 67 Zero Adjust e To ensure air entering the instrument is clean install the zero air filter in the instrument s intake and sample until the readings stabilize e While pressing the ZERO button turn the ZERO ADJUST potentiometer shown at right using the trimmer tool until the digital meter reads 0 gt If the LCD reads H turn the ZERO ADJUST counter clockwise gt If the LCD reads L turn the ZERO ADJUST clockwise NOTE A minimum 30 minute wait after the sensor regeneration cycle is complete ensures maximum sample accuracy The unit can be used immediately following the sensor regeneration if necessary When the sensor regeneration is complete press ZERO and adjust the ZERO ADJUST pot until 0 appears on the display Install the zero air filter in the intake and take several samples or lock the instrument into survey mode see page 15 After approximately one minute stop sampling and check the ZERO Adjust to 0 if necessary Repeat sampling through the zero air filter until sensor remains on 0 NOTE When ZERO is pressed depending upon internal configuration a number between 00 and 100 may appear on the display instead of H L or O See APPENDIX C INTERNAL DIP SWITCH SETTINGS on page 55 for details If the instrument is configured with an Option Board see APPENDIX E JEROME 431 X OPTION BOARD beginning on page 58 CAUTION Do
28. e filter once each week or as needed In dusty environments the fritware filter may need to be replaced as often as once a day Replacement 25 inch fritware filters are available from AZI Consumable Sales at 800 528 7411 or 602 470 1414 e Unscrew and remove the intake e Push the old fritware filter disc out of the intake with your trimmer tool Avoid touching the new fritware disc with fingers Use tweezers to insert the new fritware Use the blunt end of the trimmer tool to seat the fritware disc firmly against the inner ledge of the intake Screw the intake back on the Jerome 431 X A CAUTION oS The stem coming from the instrument onto which the outer intake housing is attached must be securely held in place If loose the tubing inside the instrument can become twisted when the intake housing is replaced It may be necessary to open the instrument and tighten the hold down nuts inside the instrument Call AZI Customer Service at 800 528 7411 601 470 1414 or e mail support azic com if you have any questions AZI Customer Service 800 528 7411 or 602 470 1414 Page 19 of 67 Internal Filters e Replace the internal filters after six 6 months of use or as needed e Press the power OFF button and unplug the power cord e Remove the two 2 side screws from the intake end of the instrument and open the case e Carefully disconnect the Tygon tubing from both ends of the filters and discard the old fil
29. e introduced into the Jerome 431 X Since this ratio can change slightly with use it is important to occasionally determine the current dilution module ratio to ensure accurate results For normal applications a X8 to X12 ratio is recommended The 431 X Functional Test Kit contains all accessories necessary to determine the current dilution module ratio Call Customer Service at 800 528 7411 or 602 470 1414 if you have questions about flow rates or applications The dilution module contains Resisorb mercury vapor adsorbent For safety information see the supplier s Material Safety Data Sheets MSDS or call AZI Customer Service at 1 800 528 7411 or 1 602 470 1414 for assistance in obtaining the MSDS AZI Customer Service 800 528 7411 or 602 470 1414 Page 48 of 67 Dilution Module Ratio Check The 10 1 dilution module was manufactured and calibrated to produce a 10 1 ratio in the sample delivered to the instrument Over time the ratio may change slightly To accurately determine the exact dilution offered by the 10 1 dilution module perform the following tests and calculate the exact dilution ratio The calculated ratio can then be used in all final calculations where the dilution module is used NOTE Wait a minimum of 30 minutes after sensor regeneration before starting this procedure Direct 431 X Readings e Connect the instrument septum holder assembly and zero air filter with arrow pointing to the instrument as shown
30. e pnenentan cece annan R E N 57 15 APPENDIX E JEROME 431 X OPTION BOARD c scsssssssssosssssscosssssecsssessceeeeeesecees 58 PENG ay Se 0 eee ne ee eee ORR E E ee rE nT oe ene ee ert een ee ee 58 TSN Cie GUNNS 0 7 E cts erectus eee eins cctools ee E ned E E E eases dec teem 58 Wiis GUN Ge MCF AIO I ic cocneseccectasseesseadadasneshaccscsasecesseaueesssuddade si eslue stone secdnoberseddonidadanueelnc usbesecdtabe ae 59 PMI a tesa seacoast aeesacataeeseeesaceegses4 nares aedaatawaanesoubeuanaccneansaewenedacees 60 Aa 2 OMIA PAMaAlOG OUT stata a E EE EEE EEEE 60 WACTER T ee 60 COMME CHO and SOC as stran cavers tacase te yrcsea sterespnuandasanatanasest in sccwtntenanausantadaatenas 61 oc dW a Was 0 2 016 6 Reema REO a er nen ne oe ME eA mR ne ne Se oot Oat ene on ERO ee 63 POS OP AOI ear a 64 DC Power Adaptor Kit AZI P N Y031 0902 snnsnnnnennnnnssessssssssssssseerrrresssssssssssssseseereees 64 NG VARRANT ec E 66 An up to date electronic copy of this manual can be found at http www azic com downloads aspx AZI Customer Service 800 528 7411 or 602 470 1414 Page 4 of 67 This manual contains details that will optimize the results and the life of your instrument Read and refer to the manual for complete details on operation maintenance and troubleshooting special voltage inputs and data output The Jerome 431 X is easy to operate and ready for use upon receipt from the factory e Remove the instrume
31. e the mercury e Carefully remove the stopper assembly from the calibration vessel A oT A BE SURE NEEDLE GUIDE IS FREE OF LIQUID MERCURY ae e Carefully pour the mercury into a disposal vessel Refer to Vessel Disassembly Instructions on page 38 gt Mercury can become trapped between the plastic calibration vessel and the glass inner liner e Replace the oxidized mercury with approximately 1 2 cc fresh mercury AZI P N A2600 0904 gt Do NOT use the syringe for measuring liquid mercury Dispose of oxidized mercury properly e Reassemble the calibration vessel e Reinstall the stopper assembly Functional Test Procedure NOTE Perform the functional test ONLY after sensor regeneration e Allow the calibration vessel to remain stable at room temperature for at least 2 hours gt The temperature range for the test is 16 24 C gt Temperature fluctuations during the test procedure will produce erratic results e Replace the 25mm fritware gt Refer to page 19 for instructions e Replace the septum on the septum holder assembly e Plug the tubing adapter end of the septum assembly into the instrument s intake and tighten the intake tube nut NOTE To check for a tight seal gently pull on the septum holder assembly If it comes out of the intake it may be necessary to remove the intake tube from the instrument and firmly press the tubing adapter through the intake Tighten the intake tube firmly to the i
32. ed to provide fresh air or conditioned air during sensor regeneration This may be necessary if the sample stream lacks molecular oxygen A low current six volt DC solenoid connected between pins 19 and 11 of the 25 pin rear panel connector will be energized during the regeneration cycle if the option board SW100 switch 6 is placed in the OFF position If needed the circuit may be built from the following components and configured as shown in the following diagram It will only function if the option board is installed in the 431 X instrument 3 1 8 tubing 1 25 pin male DB 25 connector Solder cup style AMP 747912 2 None AMP 749626 2 These are types not stocked by AZI but should be available overnight from many AMP stocking distributors such as Digi Key Corporation There are multiple suitable alternatives such as Radio Shack s 276 1547 and 276 1549 CE Se 2 16 TUBE TO INSTRUMENT TO SAMPLE j SOURCE i i TO FRESH AIR 1 16 TUBE SOURCE PIN 1 PIN 1 PAS PAS REAR OF CONNECTOR CONNECT TO PINS 11 AND 19 CONNECT TO PINS 19 AND 21 MOST INSTRUMENTS 431 X SHIPPED IN 1994 ONLY AZI Customer Service 800 528 7411 or 602 470 1414 Page 63 of 67 DC Power Operation Instruments with the 431 X option board modification can be used with any 12 VDC source for continuous operation if the AZI Power Inverter Kit P N Y031 0902 is installed along with the option board To preserve the life of the DC power source usua
33. eller s specifications or such other specifications as seller has agreed to in writing Seller s obligations with respect to claims under this warranty shall be limited at seller s option either to the replacement of defective or non conforming product or to an appropriate credit for the purchase price thereof subject to the provisions of seller s Warranty Policy as amended from time to time said Policy being incorporated herein by reference Returned products under warranty claims will be shipped to seller s plant by buyer at buyer s expense and shall be accompanied by a statement of the reason for the return and an approved Return Material Authorization Number issued by seller Buyer remains responsible for payment for products not accepted for warranty adjustment handling costs and freight costs associated therewith Notwithstanding the foregoing no warranty shall be enforceable in the event that product has been subjected to environmental or stress testing by buyer or any third party without written approval of seller prior to such testing Further no warranty shall be enforceable if the alleged defect is found to have occurred because of misuse neglect improper installation repair alteration accident or improper return handling procedure by buyer Discontinued product is warranted only for a credit or replacement at seller s option THE EXPRESS WARRANTIES GRANTED ABOVE SHALL EXTEND DIRECTLY TO BUYER AND NOT TO BUYER S CUSTOMERS A
34. en detected Refer to your safety policy for additional direction to confirm the concentrations DURING SAMPLING 0 25 sensor saturation 25 50 sensor saturation 50 75 sensor saturation 75 100 sensor saturation DURING SAMPLING USING SURVEY MODE Survey sampling minus sign flashes continuously WHEN ZERO IS Adjust to 0 only after sensor regeneration Itis normal for the DEPRESSED display to read H after sampling has started Zero ready to sample High turn Zero potentiometer counterclockwise Low turn Zero potentiometer clockwise AZI Customer Service 800 528 7411 or 602 470 1414 Page 10 of 67 Daily Operations Before each day s use of the Jerome 431 X perform the following steps to verify proper instrument operation e Press the power ON button gt The digital meter displays 000 Disregard the digital meter s initial momentary reading Recharge or replace the battery pack if the LO BAT indicator REMAINS ON Refer to Charging Batteries on page 17 and or Replacing the Battery Pack on page 21 To ensure the instrument s electronics have stabilized allow a 1 minute warm up before beginning the next Step e Use the Zero Air Filter to equilibrate the instrument to ambient air temperature gt Install the Zero Air Filter in the instrument s intake gt Sample continuously until the readings stabilize e Perform sensor regeneration Refer to page 12 for the procedure e Thirty minutes a
35. essories Personal X412 0901 Mercury Dosimeter Stopper A2600 0902 Assembly Z2600 3914 Septum Holder 3200 0011 Septa Syringe A2600 0903 Assembly Syringe Needles 2600 0022 22 Ga AZI Customer Service 800 528 7411 or 602 470 1414 Page 32 of 67 1400 2002 2300 0001 1300 0031 Z4000 0907 Z4000 0908 Z2600 3905 Z2600 3928 Z2600 3930 Z2600 3940 990 0193 1400 3010 Y2600 3945 PS 151 AZI Customer Service 800 528 7411 or 602 470 1414 Probe Trimmer 1 8 x 3 16 reducer Battery Pack Assembly 115V Battery Pack Assembly 230V Zero air filter C M filter Scrubber filter Chlorine Filter Ammonia Filter Tubing adapter Intake Kit Tube Nut Includes mounting hardware Page 33 of 67 2600 3039 25 inch fritware Tygon tubing S 1 8 I D 1 foot Tygon tubing 2500 3002 1 16 I D 1 foot 115 VAC IDC battery charger A Used to charge an uninstalled battery 230 VAC IDC battery charger ee Used to charge an uninstalled battery AZI Customer Service 800 528 7411 or 602 470 1414 Page 34 of 67 6000 4003 100 120 VAC Line Cord Alterate Tine Cord for 200 0003 land Alternate Tino Cord for 200 0008 Europe 990 0225 10 to 1 Dilution Module 311 LNSINYLSNI VNOZINV AZI Customer Service 800 528 7411 or 602 470 1414 Page 35 of 67 5100 1012 Spare Fuse
36. fter sensor regeneration is complete zero the instrument NOTE For maximum accuracy such as when testing with the Functional Test Kit wait 30 minutes after the sensor regeneration cycle to re zero the unit For immediate use the unit can be re zeroed immediately after sensor regeneration e Press the SAMPLE button gt During the sample cycle the digital meter displays bars or to indicate the amount of sensor saturation e At the end of the sampling cycle read the digital meter gt The number shown on the digital meter is the Mercury Vapor concentration in mg m3 gt This value remains on the display until the next sample is taken gt The digital meter automatically zeroes at the start of each sample e At the end of each day s use perform sensor regeneration as described in the next section DO NOT ALLOW MERCURY VAPOR TO STAY ON THE GOLD FILM SENSOR OVERNIGHT AZI Customer Service 800 528 7411 or 602 470 1414 Page 11 of 67 Sensor Regeneration Sensor regeneration is needed to clear the 431 X sensor of any accumulated Mercury Vapor and to prolong the life of the sensor This simple procedure should be done At the beginning of the day on which the instrument is to be used During the day when the sensor becomes saturated At the end of the day before storing the instrument At a minimum of 30 day intervals while the instrument is in storage CAUTION Ensure the voltage selector on the back of the
37. g these steps gt Connect the line cord between the connector on the back of the 431 X and an AC power outlet gt Press the ON switch and then press the REGEN button The instrument will begin a 10 minute regeneration cycle indicated by H H H flashing on the display Do not interrupt this cycle For a complete description of this process see page 12 If any error message such as P P P appears on the display see the Troubleshooting section beginning on page 24 e When regeneration is complete zero the sensor by pressing the ZERO button and turning the zero adjust screw located under the handle until the display reads 0 e The instrument is now ready to sample e To ensure the input to the instrument contains no Mercury Vapor or mercaptans use a Zero Air Filter AZI P N Z2600 3905 The Zero Air Filter cleans the air sample and should produce sample readings of less than 0 003 mg m Therefore use the filter to gt Equilibrate the instrument to temperatures that are higher or lower than the instrument Sample with filter installed until the reading is below 0 003 mg m gt Identify contamination within the unit gt Confirm the presence of Mercury Vapor when readings are elevated Install filter and verify that the readings go down with filter installed e When the instrument measures Mercury Vapor the zero display will be replaced with a value CAUTION Do not adjust the ZERO after the instrument has measu
38. grams e Ensure the instrument is turned OFF Remove the two screws near the front of the instrument and open the lid Locate DIP Switch SW2 at the top of the main circuit board Place Position 2 to OFF for nanograms display Return Position 2 to ON for the normal milligram display If the instrument display reads P P P when the instrument is connected to AC power or when REGEN is pressed or if the battery will not charge the fuse may need to be replaced The AC line power could also be less than 100 VAC 220 VAC Check the fuse with an ohmmeter and the AC line power with a voltage meter e Locate the power receptacle on the rear of the instrument e Insert a small screwdriver in the slot located in the power receptacle and gently slide the fuse compartment out e If the fuse in the open sided clip is open remove and discard it e Replace the discarded fuse with the spare fuse located in the slide out spare fuse compartment e Replace the fuse compartment in the power receptacle gt As soon as possible replace the spare fuse with another 1A 250V time delay fuse AZI P N 5100 1012 AZI Customer Service 800 528 7411 or 602 470 1414 Page 22 of 67 6 CALIBRATION The Jerome 431 X s gold film sensor is inherently stable and does not require frequent calibration The interval between calibrations depends upon the application and frequency of use however the recommended interval is every 12 months The Jerome
39. he last injection then turn off the pump e Remove the dosimeter septum assembly and zero air filter from the pump e Connect the instrument dilution module dosimeter zero air filter and dosimeter lead set as shown 371 LNSWNUYLSNI VNOZINV e Attach the power cord to the 431 X and plug it into AC power gt AC power is required to heat the dosimeter e Press the Jerome 431 X Power ON button and then press the SAMPLE button gt The digital meter reading appears in approximately 17 seconds e Record the digital meter reading include decimal point Wait 60 seconds then press SAMPLE again and record this reading e Repeating the heating process ensures complete release of mercury from the dosimeter coil e Add the two digital meter readings together gt The sum of the two digital meter readings is the figure you will use in your calculations and is referred to as the meter response MR e Repeat this procedure two more times e Average the three meter responses you obtained in this section AZI Customer Service 800 528 7411 or 602 470 1414 Page 50 of 67 Dilution Module Ratio Calculations e Multiply the average obtained for Direct 431 X Readings by 10 the number of 1 cc injections gt Divide this result by the average obtained in the section Loading the Dosimeter on page 50 e Use the result as the dilution module ratio in your dosimeter analysis EXAMPLE Direct 431 X readings 0 102 mg m 0 103 mg m 0
40. her straight chained compounds are typically extremely volatile and change the electrical resistance of the gold film sensor Any such responses should be considered qualitative not quantitative The instruments are designed and calibrated to elemental mercury vapor only Inorganic mercury salts such as mercuric chloride are not very volatile They may however generate some minute level of elemental mercury vapor to which the instruments will respond This response again should be considered a qualitative response only AZI Customer Service 800 528 7411 or 602 470 1414 Page 29 of 67 9 ACCESSORIES amp MAINTENANCE PARTS PART ITEM DESCRIPTION Y431 0901 431 Accessory Kit See pictures beginning on page 32 1400 2002 Probe 1400 3010 Tubing Adapter 1 4 to 1 8 2300 0001 Trimmer Tool 2600 3039 25 Fritware 6000 4003 Line Cord 115 VAC USA and Canada Alt 200 0003 Line Cord 220 240 VAC England Alt 200 0008 Line Cord 220 240 VAC Europe Z2600 3905 Zero Air Filter Y431 0902 431 Functional Test Kit See pictures beginning on page 32 A2600 0902 Stopper Assembly A2600 0903 Syringe Assembly A2600 0904 Mercury Vial 2600 0022 Syringe Needles 22 Ga Reusable 2600 0030 Calibration Vessel Thermos 3200 0011 Septa 20 Z2600 3905 Zero Air Filter Z2600 3914 Septum Holder Y431 0903 or 431 Maintenance Kit 110 VAC See pictures beginning on page 32 Y431 0904 431 Maintenance Kit 220 VAC See pictures beginning on page
41. ing with the Dosimeter The personal mercury dosimeter adsorbs mercury vapor over a set period of time Therefore before each day s use it is necessary to ensure the dosimeter is mercury free Perform the following steps to remove any accumulated mercury e Connect the system as shown gt Insert the dosimeter s large end in the 431 X s intake and gently tighten the intake tube nut to ensure an airtight seal gt Connect the Dosimeter Lead Set clips as shown Short red lead to the rear and long black lead to the far end e Attach the power cord to the 431 X and plug it into AC power AC power is required to heat the dosimeter e Attach the Dosimeter Lead Set 25 pin connector to the respective 25 pin communications port e Press the instrument s power ON button e Press the instrument s SAMPLE button gt The digital meter reading will appear in 17 seconds e Wait 60 seconds and press the SAMPLE button again gt The digital meter should display less than 0 005 verifying all mercury has been removed from the dosimeter coil e Wait 2 minutes to cool sensor to prevent false positive response e The dosimeter is ready for sample collection NOTE For best results dosimeter analysis should be performed immediately after collection If analysis cannot take place immediately place the red end caps on the dosimeter For accurate results perform dosimeter analysis no later than five days after sampling AZI Customer Service 800
42. instrument near the power cord inlet connector points to the local AC power value See Setting the Input Voltage on page 21 To clean and protect the sensor the supplied AC power must be 100 to 120 VAC or 220 to 240 VAC depending on the available power source Once sensor regeneration is initiated DO NOT interrupt the cycle e Attach the power cord to the 431 X and plug it into AC power AC power is required to thermally regenerate the sensor e Press the power ON button e Press the REGEN button gt The digital meter flashes H H H for the duration of the 10 minute cycle and displays 0 0 0 when the cycle is completed e A minimum 30 minute wait after the sensor regeneration cycle is complete ensures maximum sample accuracy However the unit can be used immediately following the sensor regeneration if necessary When the sensor regeneration is complete press ZERO and adjust the ZERO ADJUST pot until 0 appears on the display Install the zero air filter in the intake and take several samples or lock the instrument into survey mode See page 15 After approximately one minute stop sampling and check the ZERO Adjust to 0 Repeat sampling through the zero air filter until reading remains on 0 NOTE The digital meter will read P P P after REGEN is activated if the power cord is not plugged in or if the instrument s fuse needs to be replaced Connect the power cord or if necessary replace the fuse See Changing the Fus
43. istical analysis The software can also program the Jerome Data Logger AZI P N 6100 0010 This optional accessory enables data storage during manual sampling or portable automatic sampling without being attached to a computer The data logger initiates automatic sampling triggers alarms and stores data The logged data may then be downloaded to the computer when it is convenient The data logger stores up to 1 000 data points The JCS is menu driven and easy to use Each Ed AZI Jerome Communications Software display screen is designed for clarity with self Ee Soup Data Looe beets froment Aow be beg explanatory menu options such as Operate Py Instrument or Display Stored Data Select menu A a options using either a mouse or a track ball pointing 7 A device or a standard keyboard The user creates records or files for computer storage of collected data Data is easily retrieved for later viewing graphing printing or editing with spreadsheet or word processing software not provided Data can be used for ongoing record keeping or for fulfilling local regulatory requirements Arizona Instrument Ready f kas Nmf f 2 Before using this software familiarization with the operation of the Jerome Hydrogen Sulfide Analyzer or Mercury Vapor Analyzer is important Also prior to installation of this software you should be familiar with the personal computer and operating system you are using If you have any questions
44. lly a car or truck battery the power inverter will switch on automatically to supply the AC necessary for regeneration only The external switch on the inverter should always be OFF to preserve battery life during normal sampling To work with the power inverter kit place option board SW100 DIP Switch 6 to the ON position When the instrument starts a regeneration with option board SW100 DIP Switch 6 ON the instrument sends a signal to close the relay on the DC Power Adaptor AZI P N 1000 0089 mounted between the data logger and the instrument This switches the power inverter ON using the inverter s internal switch NOTE When this mode is enabled the instrument does NOT check for 115 VAC for the regeneration If there is no AC power to the instrument and a regeneration is initiated the instrument will flash H H H rather than P P P however the sensor will not heat nor will the sensor be cleaned DC Power Adaptor Kit AZI P N Y031 0902 e The DC power adaptor kit consists of gt DC Power Adaptor P N 1000 0089 gt DC Power Inverter P N 4000 1021 gt DC Power Cable Assembly P N 6000 1093 Installation e Ensure that the instrument s option board switches are set correctly for the intended operation with the option board s SW100 DIP Switch 6 set to ON for DC operation e Mount the interface board to the rear of the instrument Tighten the mounting screws e Place or mount the DC AC power inverter in a secure p
45. los 8 pounds Liquid crystal display LCD CE mark on 220 240 V 431 XE model only Optional Communications Capability Dosimeter power output Data output For dosimeter analysis and regeneration 1 RS 232 Serial Baud Rate 1200 for use with data logger and or Jerome communication program 2 RS 232 Serial data format with O amp 20mA current logic levels Baud Rate 1200 special industrial applications and Analog 20 mA output With Option Board See APPENDIX E JEROME 431 X OPTION BOARD on page 58 Analog output Auto sample interval Auto regeneration interval 0 to 2V or 4 to 20 mA 5 15 30 or 60 minutes 6 12 or 24 hours AZI Customer Service 800 528 7411 or 602 470 1414 Page 27 of 67 Instrument I O Interface The 431 X I O port 25 pin D sub provides the following functions e Serial data communication gt Interface type RS 232C full duplex DCE gt Parameters 1200 Baud 1 start bit 8 data bits 2 stop bits no parity gt Pin assignments Pin 1 Protective ground Pin 2 Data in Pin 3 Data out Pin 7 Data ground e Serial current loop gt Interface type 20mA current loop full duplex gt Parameters 1200 Baud start bit 8 data bits 2 stop bits no parity gt Pin assignments Pin 1 Pin 4 Pin 5 Pin 14 Pin 16 e Dosimeter Power gt Pin Assignments Pin 22 Pin 23 Protective ground Data out Data in Data out Data in Dosimeter Enable 24 t
46. ly Operation under JCS or data logger control automatically regenerates saturated sensors The option board can control regeneration on a regular basis every 6 12 or 24 hours The regeneration intervals are set through a combination of switch settings as shown in the following table REGENERATION Switch 1 Switch 2 Interval Hrs AZI Customer Service 800 528 7411 or 602 470 1414 Page 59 of 67 Auto Sample If a data logger is either not connected or is operating in the manual sampling mode the following automatic sampling rates may be selected with option board s SW100 dip switch settings Dip switch settings Sampling frequency No automatic sampling The switches have no effect if a data logger 1s connected and operating in automatic sampler mode as programmed through the JCS 4 20 mA Analog Output The analog output signal at pin 18 of the 25 pin connector can be configured to provide the instrument s native mode 0 2 Volt output or the optional 4 20 mA output by setting the option board jumper SW 101 to the V position for voltage or the I position for current Pin 23 is the ground pin for the analog output function Pin 18 is positive with respect to the ground pin e The 0 2 Volt output circuit can drive loads of 10 k ohms or higher e The 4 20 mA output is a passive transmitter and requires the connected receiver to supply between 10 and 28 volts of excitation potential The analog output signal is based
47. ly worn on a belt The dosimeter can also be used for multiple point area monitoring by placing a dosimeter with pump attached in various strategic locations We recommend a pump flow rate of 5 cc minute for the most accurate results when sampling in an atmosphere that for eight hours may contain an average of 025 mg m Hg If you are considering using any other flow rate see page 48 Non Standard Flow Rates and Dilution Modules After sample collection is completed the dosimeter is inserted in the J erome 431 X s intake A dosimeter lead set 1s connected between the dosimeter and the DB 25 connector on the back of specially equipped instruments The instrument supplies power to volatilize the accumulated mercury from the dosimeter to the gold film sensor The Jerome 431 X determines the mass of mercury collected by the dosimeter in a 17 second analysis The dosimeter is ready for immediate re use after a mercury measurement has been performed Dosimeter Technical Specifications Sensitivity Less than 0 5 nanograms of mercury Precision 15 RSD 0 100 mg m Hg Accuracy 15 0 100 mg m Hg Recommended flow rate 5 cc min 0 005 liters min for atmospheres up to 0 025 mg m Construction Nylon and glass housing a gold film coil Weight 1 5 ounces Dimensions 0 5 dia x 4 5 Capacity 1000 nanograms of mercury Analysis Time Less than two 2 minutes AZI Customer Service 800 528 7411 or 602 470 1414 Page 44 of 67 Before Sampl
48. min hr multiplied by 8 hours gt 5 cc min x 60 min hr x 8 hr 2400 cc e Determine the Hg concentration time weighted average of the dosimeter gt The mass of Hg collected by the dosimeter divided by the total volume of air sampled gt 493 5 nanograms divided by 2400 cc 0 205625 ng cc of Hg 0 205625 mg m of Hg e Check the sensor status after each dosimeter analysis e Seal the dosimeter with tubing after analysis to prevent excessive mercury contamination during storage IMPORTANT Perform a sensor regeneration as soon as the meter display shows four bars to prevent the loss of sample AZI Customer Service 800 528 7411 or 602 470 1414 Page 53 of 67 Dosimeter Reference Chart 431 X Expected concentration related to sample volume and meter response Indicates the optimum meter response for that concentration with the corresponding volume 431 X reading in mg m HL HL HL 0 857 HL HL HL 0 5 0 1 10 086 loi7 10 343 0 686 HL HL HL HL HL HL Estimated O95 10 043 10 086 lo171 10 343 0 686 HL HL HL HL HL concentration nmefm 0 025_ 0 021 0 043 0 086 0171 o343 0 686 HL HL HL HL 0 005 10 004 0 009 0 017 0 034 10 069 10 137 10 274 0 549 HL HL 0 001 ILOW LOW 0 003 0 007 0 014 0 027 10 055 10 110 10 219 Volume at 5ml min 75 150 300 600 1200 2400 4800 9600 19200 38400 Hours collected 0 25 0 5 1 2 4 8 16 32 64 128 Use the following formula for calculating the concent
49. nal is encountered An HL high mercury level is encountered The instrument is turned OFF Press the power OFF button when not in use AZI Customer Service 800 528 7411 or 602 470 1414 Page 15 of 67 Operating on AC Power or Generator e For stationary use the 431 X may be operated on AC power gt Operating the instrument on AC power at all times eliminates the need for the battery pack and its necessary maintenance gt The battery may be unplugged or removed completely whenever the instrument is operating on AC power e When a generator is used to power the Jerome 431 X a high quality line conditioner or voltage regulator is required to ensure a pure sine wave and regulated voltage is applied to the instrument The gold film sensor may be damaged by voltage that varies in amplitude or by surges spikes and or noise on the power line This is especially true during the sensor regeneration Operating on Internal Battery Power For portable use the 431 X may be operated on Battery power e When you operate the instrument on battery power please be aware of the following gt A fully charged battery pack AZI P N Z4000 0907 provides power for a minimum of six 6 hours of operation gt For operating more than six 6 hours an extra fully charged battery pack is needed gt Complete battery recharging takes 14 hours Refer to Charging Batteries on page 17 gt The 431 X uses a rechargeable Nickel Cadmium NiCad batte
50. ning the MSDS e Carefully unpack and inspect the parts of the kit e ENSURE that the mercury shipping container and mercury filled thermometer are not broken e Ina ventilated area preferably under a fume hood remove the mercury vial from its shipping container e Place the functional test kit vessel and the mercury vial close to each other and open the mercury vial CAUTION The edge between the plastic case and the glass inner vessel of the functional test kit vessel are not sealed tight enough to prevent mercury from entering the area between the inner and outer vessels ENSURE the mercury transferred in the next step does not contact the seal where the glass and plastic portions join NOTE The vessel may be disassembled to transfer the mercury and better prevent contamination of the outer portion of the vessel Instructions to disassemble the vessel can be found on page 38 AZI Customer Service 800 528 7411 or 602 470 1414 Page 37 of 67 Mercury Transfer e CAREFULLY pour the mercury into the center of the functional test kit vessel s opening gt ENSURE that no mercury residue is on the outside of the vessel See the supplier s Material Safety Data Sheets MSDS or call AZI Customer Service at 1 800 528 7411 or 1 602 470 1414 for clean up instructions e INSTALL the stopper assembly into the functional test kit vessel carefully to prevent breakage of the thermometer gt PRESS the stopper assembly into the ve
51. nt from the packing material Retain all packaging materials for any future shipment of the instrument If the instrument is returned to AZI for any reason it must be placed in the original packaging materials that have been tested and proven to be effective protection during shipment e Call AZI Customer Service at 800 528 7411 or 602 470 1414 for Return Material Authorization RMA information prior to returning a unit e For all shipments boxes and packing materials are available from AZI e Pack the Jerome instrument only inaJ erome shipping container AZI WILL NOT BE RESPONSIBLE FOR SHIPPING DAMAGE IF YOU RETURN THE INSTRUMENT IMPROPERLY PACKAGED OR SHIPPED YOU SHOULD INSURE IT FOR FULL VALUE e Check for any damage and confirm receipt of all parts on your packing list Contact Arizona Instrument Customer Service at 800 528 7411 or 602 470 1414 if you have any questions e Press the ON button The display should read 000 in less than one second gt ALO BAT message appears briefly in the upper left corner gt If the LO BAT message persists recharge the battery See page 17 AZI Customer Service 800 528 7411 or 602 470 1414 Page 5 of 67 e Check the voltage setting 110 or 220 VAC on the back of the instrument Ensure that it is set to the correct voltage If the pointer is not aligned to the local voltage turn the selector to point to the correct voltage e Perform a sensor regeneration by followin
52. ntake stem Attach a zero air filter to the septum assembly Press power ON Take 3 samples AZI Customer Service 800 528 7411 or 602 470 1414 Page 39 of 67 gt If the average meter reading is greater than 005 stop here The instrument may be contaminated See 431 X TROUBLESHOOTING on page 24 gt If the average meter reading is less than 005 continue to the next step e Note the temperature of the calibration vessel e Press the SAMPLE button wait 2 seconds and when the display flashes inject 1 cc of mercury vapor according to the syringe technique described on page 41 Be sure all mercury vapor has been injected before the solenoid closes second click and display flash e Record the meter reading e Repeat the instructions for mercury injection three more times gt The readings obtained for the last three 1cc injections should be within 5 of each other e Refer to the Temperature Conversion Chart page 42 for the acceptable range gt The average of the last three readings should fall within the range shown on the chart If the average is within range the JEROME 431 X is functioning properly e If the last three readings are not within 5 of each other gt Perform sensor regeneration Press ZERO and turn the ZERO ADJUST refer to page 12 and 13 for the complete sensor regeneration procedure gt Wait 1 hour before proceeding to the next step gt Repeat the mercury injection test procedure gt If the
53. o 28 Volts AC Battery Ground Pins 12 amp 24 Tied Dosimeter Power Pins 13 amp 25 Tied Dosimeter Power e Switched battery connection for data logger gt Pin assignments Pin 9 Battery Pin 7 Battery ground Pin 23 Battery ground e Unswitched battery connection for external battery pack pin assignments gt Pin assignments Pin 15 Battery Pin 19 Battery Pin 7 Battery ground Pin 23 Battery ground NOTE Pins 6 8 11 17 18 20 and 21 are non standard and should not be connected AZI Customer Service 800 528 7411 or 602 470 1414 Page 28 of 67 Potential Interferences ys Potential interferences to the Jerome mercury vapor analyzers are rare and most of these can be eliminated with proper maintenance procedures However erroneously high readings can sometimes occur Here are a few things to be aware of when using the instrument The gold film sensors used in the J erome mercury vapor analyzers do not respond to the following compounds e Hydrocarbons e CO COs and SO e Water vapor Note that water vapor condensation on the gold film can cause irreparable harm to the sensor and must be avoided The C M filter AZI P N Z2600 3928 an acidic gas filter contained in the internal filter system removes the following compounds that cause the gold film sensor to respond e Chlorine NO e Hydrogen Sulfide H2S e Most mercaptans organic sulfur compounds or thiols
54. o ambient air Continuous sampling with clean air will not cause saturation of the gold film sensor but will equalize temperatures faster to allow accurate analysis to begin sooner The Zero Air Filter can also be used to identify contamination within the instrument If the readings do not reduce to near zero with the filter installed contamination should be suspected If the readings do drop to near zero with the filter installed but elevate with the filter removed the presence of Mercury Vapor at the sampled location 1s confirmed For more information on the use of the Zero Air Filter contact customer service at 1 800 528 7411 1 602 470 1414 or visit our web site at http www azic com AZI Customer Service 800 528 7411 or 602 470 1414 Page 9 of 67 4 INSTRUMENT OPERATION LCD Codes LCD CODE i EXPLANATION Ready to sample No Mercury Vapor reading No Mercury Vapor reading display in nanograms 8 8 8 L L L Sensor saturated regeneration needed refer to page 12 Sensor regeneration in progress H H H flashes Re zero needed refer to page 13 Power cord required or low line power lt 100 VAC or 200 VAC see pages 16 and 17 Changing the Fuse 1f P P P remains on after the cord is connected H L P High line power greater than 130 VAC in 110 operation or 260 VAC in 220 operation LO BAT Recharge batteries refer to page 17 Same as LO BAT automatically shuts off ap m Very high concentration has be
55. osition near the instrument e Connect the cable from the DC AC power inverter to the matching connecter on the interface board Note that the connectors are keyed to prevent improper connection e Plug the instrument s AC power cord into the power inverter and connect it to the instrument AZI Customer Service 800 528 7411 or 602 470 1414 Page 64 of 67 Ensure that the inverter s power switch is in the OFF position LEAVE the power switch in the OFF position at all times The interface board will activate the inverter when necessary If the inverter power switch is placed in the ON position it will cause a continuous drain on the external 12 volt power system REAR VIEW SHOWN WITHOUT COVER FOR CLARITY w 4 5 w suz 3 Q oY Ea YELLOW VWVIRE BLACK WIRE TO BATTERY NEGATIVE SSS SSS RED WIRE TO BATTERY POSITIVE eee RED WIRE e Remove the screws from the rear cover of the inverter and remove the cover e Place the wires from the external DC source battery and the wires from the DC power cable through the holes in the end plate e Connect cables from the external 12 volt power source and the DC power cable assembly to the appropriate positive and negative terminals on the back of the inverter and tighten the hold down screws e Connect the yellow wire from the DC power cable to the REMOTE terminal on the power inverter and tighten the hold down screw
56. our bars This indicates that the sensor is 75 100 percent saturated e Seal the dosimeter with caps after analysis to prevent mercury contamination during Storage e If your readings exceed 75 nanograms or more try the recommendations described next or call AZI Customer Service at 800 528 7411 or 602 470 1414 for assistance Non Standard Flow Rates and Dilution Modules You may use a pump with a flow rate up to 50 cc min but be aware that there are certain limitations If your pump flow rate exceeds 5 cc min and your average dosimeter analysis produces nanogram levels of 75 or more it may be easy to collect enough mercury to saturate the 431 X sensor You thus risk over ranging your instrument and losing your collection data Higher flow rates may also impair the capture efficiency of the dosimeter We recommend that you drop your flow rate or use a dilution module AZI P N 990 0225 Lowering the flow rate to decrease the sample volume provides the greatest accuracy Using a dilution module introduces an additional 15 inaccuracy to your analysis As an alternative to the dilution module sample for shorter time periods Dilution Module Specifications Accuracy 15 of 10 1 ratio E O 6 1 cm w x 14 0 cm 1 x 2 5 cm h 2 4 w x 5 5 1x 1 h Weight 61 4 g 2 2 oz The dilution module is factory set to a 10 1 ratio The mass of mercury entering the dilution module is reduced by 90 leaving a 10 X10 dilution concentration to b
57. ration of mercury in air M eter Response X 87 5 i Concentration mg m Flow Rate of Sampling Pump X Time The following table shows the relation of flow rate and time to total volume collected cc min Total volume collected in cc 100 60 20 10 600 1200 2400 4800 7200 14400 28800 43200 5 f 300 60o 1200 2400 3600 7200 __14400 21600 l 2 4 8 12 24 48 72 HOURS AZI Customer Service 800 528 7411 or 602 470 1414 Page 54 of 67 Main circuit board RED DIP switches SW2 This is the red DIP switch box located at the top center of the instrument s main circuit board The 431 X provides regulated film heat at both 50 Hz and 60 Hz line frequencies This also provides two ranges of preset but unregulated film heat 100 120 200 240 volt 1 ie ie A 4 a 7 B ao k S OI a muy 4 m i Te waa st at ies L TEES a za ma se I and 110 130 220 260 volt ranges The two ranges are available to reduce the effects of chronic low or high line voltage Note The ranges are doubled when the AC line selector switch is set to the 220V position The DIP switch positions 1 and 6 must be properly set DIP 1 DIP 6 OFF OFF OFF ON ON OFF ON ON Function 60 Hz regulated film heat 100 130 200 260 VAC 50 Hz regulated film heat 102 130 205 260 VAC 50 60 Hz preset film heat 110 130 220 260 VAC 50 60 Hz preset film heat 100 120 200 240 VAC Regulated film heat should normally be u
58. red Mercury Vapor or before the next regeneration Occasionally the display may drop to L L L indicating low between the initial zeroing and the first sample It is OK to readjust the ZERO if the instrument has not measured Mercury Vapor e The instrument is designed for ambient air monitoring DO NOT allow the probe or the instrument s intake to be exposed to any liquid The instrument is not explosion proof Press the SAMPLE button to start a 12 second sampling cycle Perform sensor regeneration after each day s testing Perform another sensor regeneration and re zero the instrument before each day s use Perform sensor regeneration after 30 days of storage or inactivity Call AZI Customer Service at 800 528 7411 from the United States and Canada or 602 470 1414 if you have any questions If you prefer you may send e mail to support azic com AZI Customer Service 800 528 7411 or 602 470 1414 Page 6 of 67 2 INTRODUCTION The Jerome 431 X Mercury Vapor Analyzer is an ambient air analyzer with a range of 0 001 to 999 milligrams of mercury vapor per cubic meter mg m Hg CAUTION The Jerome 431 X is for vapor use only DO NOT allow the probe or the instrument s intake to be exposed to any liquid dust or other foreign material The 431 X is designed to be easy to operate for quick and accurate analysis of mercury vapor levels It has few maintenance requirements However please take a moment to
59. rent formula is soo o o oooo oooo REAR OF CONNECTOR CONNECT PINS 18 AND 23 FOR Example 5 4 20 mA OUTPUT A reading of 100mg m gt would produce an JUMPER ON BOARD ISAT I output current of Figure 2 16mA 0 1 4mA 5 6mA 16mA Display Reading 4mA Current AZI Customer Service 800 528 7411 or 602 470 1414 Page 61 of 67 e The 4 20 mA passive receivers do not contain a voltage source to power the loop current They require the addition of a separate isolated power supply Typically a supply that delivers 15 to 20 volts DC at 50 mA is sufficient Wire these as in Figure 3 Note that some 12 volt DC wall transformers as used on portable equipment may deliver 15 to 20 volts when they are lightly loaded The Digi Key T509 P1IP ND is a commonly available example of a 12 volt 200mA supply that will deliver around 18 volts nominal when loaded below 20 mA Be sure that both the power supply used and the passive receiver are floating not connected to earth ground If either is not floating the circuit will not work and damage may occur Ensure that jumper SW101 is set to the T position before power up AZI Customer Service 800 528 7411 or 602 470 1414 PIN 1 poco Oo oo oo o o oO foooca O O oococo REAR OF CONNECTOR CONNECT PINS 18 AND 23 FOR 4 20 mA OUTPUT JUMPER ON BOARD IS AT IT Figure 3 Page 62 of 67 Fresh Air Solenoid An external three way solenoid can be us
60. results fall within the expected range you may assume the instrument is functioning correctly See APPENDIX A 431 X FUNCTIONAL TEST KIT on page 37 for more information about the FTK procedures Complete instructions for use are supplied with the test kit AZI P N Y431 0902 To order the FTK contact your AZI Sales Representative at 800 528 7411 or 602 470 1414 AZI Customer Service 800 528 7411 or 602 470 1414 Page 23 of 67 7 431 X TROUBLESHOOTING Symptom Possible Cause Solution Power Problems Unit does not turn ON Unit Discharged battery or Recharge battery for a minimum of turns on when power cord is 14 hours Refer to page 17 plugged in LCD displays 000 when instrument is operating on Dead battery Replace battery Refer to page 21 AC power Unit does not turn on when Open fuse Replace fuse Refer to page 22 connected to AC power cord Insufficient power Use a voltmeter to verify there is power to the AC outlet Call AZI Customer Service for information at 800 528 7411 or 602 470 1414 Regeneration amp Zero Problems LCD displays 8 8 8 Sensor saturated Do not attempt to adjust zero pot The sensor must be regenerated Refer to page 12 for information LCD displays L L L when Changes in temperature Readjust zero pot See page 13 for taking first sample information LCD displays H at finish of Internal contamination may Remove and replace fritware filter sensor regeneration when zero is redeposi
61. rformance Materials Tygon is a registered trademark of Norton Arizona Instrument LLC Jerome 431 X Mercury Vapor Analyzer Operation Manual Part Number 700 0046 January 2011 If you have any questions regarding the operation of this instrument please call our toll free number 800 528 7411 Internationally call 602 470 1414 or fax 480 804 0656 AZI Customer Service 800 528 7411 or 602 470 1414 Page 67 of 67
62. ry Dispose of worn out batteries properly when you are replacing the battery pack External battery power A special version of the Jerome 431 X and a DC Power Kit are available to operate the instrument from a secondary DC source The source may be a car truck battery or a storage cell used in conjunction with solar panels Call AZI Customer Service at 800 528 7411 602 470 1414 or e mail support azic com for additional information AZI Customer Service 800 528 7411 or 602 470 1414 Page 16 of 67 Charging Batteries e Press the power OFF button e Connect the AC power cord between the 431 X power receptacle and an AC power source gt Complete battery recharging takes 14 hours gt The 431 X contains a trickle charger so it may be continually plugged into an AC power source without damaging the battery pack e The battery pack may be charged outside the instrument with an optional AZI IDC Battery Charger AZI P N 4000 1011 for 115 VAC P N 4000 1012 for 230 VAC Obtaining Maximum Battery Life There are certain inherent limitations to NiCad batteries The primary limitation is a memory effect that occurs when the batteries are partially discharged and then recharged repeatedly This memory leads to a drastic reduction in the usable battery life To prevent this memory effect periodically allow the battery pack to discharge completely and then recharge the battery pack e To obtain maximum battery life follow these three 3
63. sed DIP 1 OFF except in the few cases where extremely dirty line voltage conditions may exist These conditions might be found where large motors are being controlled or other situations may exist where the voltage may vary outside the 100 130 VAC range with regularity In those cases the two preset heat ranges will allow some degree of satisfactory operation Switch Normal Action Number Position 2 3 ON ON Nanograms mode Displays relative not true voltage during regen 0 255 4 OFF ON ON Display L O H when zero button pressed Display 00 99 when zero button pressed Locks into 0 10mg m range survey mode AZI Customer Service 800 528 7411 or 602 470 1414 Page 55 of 67 The Jerome Communications Software JCS is used with 431 X Mercury Vapor Analyzers that feature the communications configuration option e The JCS allows the user to program the instrument for unattended monitoring and to download recorded data stored in the Jerome data logger e Automatic sampling can be initiated every one 1 to sixty 60 minutes with programmable audible alarm levels The Jerome Communications Software JCS operates with the J erome 431 X Mercury Vapor and Jerome 631 X Hydrogen Sulfide Analyzers that have the Communications Configuration option installed The software can control instrument sampling for unattended continuous operation collect data graph this data in real time and perform stat
64. sscadecssncatccsceasiscsscdscsassevedestenastocsssasiadssedecsessevevane 8 FEA KOSS G1 tal gc Meee een NT Seon CT ere RARE Tee ont ORT Tay on TR Rene ter mee CNTR ene ee Ree eee gt 9 4 ANSERUMEN FT OPERA TION ciiin E E EEEE EEES 10 LEDE OUE sets restate tae E E EE A aie 10 PA ODC ON a E Tr ertry Nat ecer se mr arr Tra veneer ny eer meet tana een 11 Sensor ReceneraliOMmisicnianeisrcetiawinienieeelaecn laine ene E eee es 12 FAA COM SGN hs eee ee ee teen eM tea E ee Se Na MMe Rene Mee eae OR Ty oe ere ene er a 13 SPEM oye ca Neenentee eco ne a eee em ee MRE ER FR ee enT E Mente ot ete nen eee 14 UV Oy MIO a Montes ae anata chascasn ont eiae crave a ranma eaad 15 Operating on AC Power Or Generator seernes aa e E ai 16 Operating on Internal Battery POW er ainos a E a E OEN 16 ExXtetial battery DOWER cerauen a e a alee ti ee ans 16 Chno ma BAES an rataiest Puente E a Mees as aad 17 Obitaimime Maximum Battery Lifes mse ied en cenetsndstdartensenst ed hel a a a a a 17 De MAINTENANCE ici esaeo nna E E E 18 Preventive Maintenance Calendar acin A ingenious 18 EOW SyS ein a a eon aoe eras 19 PD SAIC ATAU are E e a N 19 Internal PIETS sareni a a a E a 20 Replacme the Battery Pae Keceri aaa eas Miran vie cadet eats saat eas ean tee iees tonite 21 Setting the Input Voltage and PrequenCy sessir n E EA 21 Display Nanograms or Milligrams Per Cubic Meterr cccssssssssssseeecccceeeceeeeeeaasseeesesseeeeeeeeees 22 Chans me the Purse acc ioeh iis se ose
65. ssel to achieve a good seal e USE the 431 X instrument to verify that the outside of the vessel is not contaminated and the mercury vapor emission level if any is below the OSHA TLV for mercury e ALLOW the kit to adjust to room temperature for at least two 2 hours before using gt The temperature range for the test is 16 24 C Avoid temperature fluctuations CAUTION Do not use the calibration vessel as a portable container If the calibration vessel is upset or greatly agitated mercury droplets will cling to the thermometer stem the rubber stopper the mouth of the calibration vessel and the needle guide CAUTION The inner portion of the vessel is made of glass Handle the vessel carefully to prevent breakage e LOOSEN BUT DO NOT REMOVE the base of the vessel The base unscrews from the body e SET the vessel on a firm surface e HOLD the base stationary and unscrew the body from the base e HOLD the base and the inner glass vessel with one hand while removing the body and gasket with the other hand e After the mercury is transferred into the glass inner vessel reassemble in the reverse order AZI Customer Service 800 528 7411 or 602 470 1414 Page 38 of 67 Replacing Mercury An oxide coating will form on the drop of mercury and will cause lower readings in your testing Gently swirl the vessel to disturb the outer oxidized surface of the droplet If this does not restore higher readings it may be necessary to replac
66. t Mercury Vapor from intake filter disk scrubber filters pressed flow system onto gold film and Tygon tubing Refer to Flow sensor System on Page 19 Internal component failure Check tubing for kinks or crimps Repeat regeneration cycle Refer to page 12 Zero adjust pot cannot be Pot not turned sufficiently 1 Turn zero adjust up to 20 times adjusted to 0 to reach the end Pot will click softly 2 If no 0 turn pot slowly in opposite direction until display reads 0 Sensor may be ruptured or pot may be broken 3 If still unchanged call AZI Customer Service at 800 528 7411 or 602 470 1414 AZI Customer Service 800 528 7411 or 602 470 1414 Page 24 of 67 Sampling Problems Airflow is restricted during the possible permanent damage sensor regeneration cycle causing Kinks and crimps in the Tygon tubing Periodically check the Tygon tubing inside the instrument Refer to page 20 High erratic results Internal Mercury Vapor contamination 1 Install zero air filter in intake and tighten intake nut Press SAMPLE button After three samples if readings are over 0 003 mg m replace fritware filter and Tygon tubing Refer to page 19 2 Perform sensor regeneration with the zero air filter in intake Refer to page 12 Retest if necessary Replace scrubber filters and Tygon tubing Refer to page 20 High erratic results Intake and internal filters
67. t Raa Shite Muntele 43 12 APPENDIX B PERSONAL MERCURY DOSIMETER csssssssssssssssssssssssssssssssees 44 Dosimeter TechimiCal Speci iC aOms saa a E E E 44 Berore Sampline withthe Dosimeter siri ciel nesses a E elesanaseneen mene 45 DOStimeter Analy S a ce cad sceladd nh an cea dg shatditnencaeban deta atedtaatesaecedes 46 Non Standard Flow Rates and Dilution Modulles cc ccccsssssssseseeeeeccceeeeeeeeeaaaeseeesseeeeeeeeeees 48 AZI Customer Service 800 528 7411 or 602 470 1414 Page 3 of 67 Dilution Module Ratio Check cc cece ccc ccc ecceccecccceccecceccscesceccscescscecescescescecescscectscesceseesesceseeses 49 Loading the D OSHC Er ct s 5 isssscsencazarereaeseiescnuisssgenssdasasensonsgsameehusanwasesrecesdasunsedoesasamoonuoaseuseanaaddans 50 Dilution Module Ratio Calculations oo cc ccecccsccccssccescceescccsscceeccuseseusceeseceueceuseseusceeeeses 51 Analysis with a Dilution Module c ccc ccsesesssseeeecccceeeceeceecaaaeesssseeeeeeceeceeeeeeeeeasaesssssseseseeeeeeees 52 13 APPENDIX C INTERNAL DIP SWITCH SETTINGS 00 ccsssccccssssccccssccccccccscceecees 55 14 APPENDIX D JEROME COMMUNICATIONS SOFTWARE j ccccccsssssssesscsesessesssceees 56 JOS KEC OMENS eee ee ete a me oe Ome ee at ene ere 57 System NCO IHC MIG ILS 44sd nsaacdeanntocddsneeacanseutadanieandecbbue odds ateimudanmnsbateninandcsbiuatardfatseadanieudentmecseiainnied 57 Dan Lo Se ODON aso etc ae sen sec atin set
68. ters CAUTION Filters contain either Sofnolime or Resisorb Used filters will contain trace amounts of Mercury also Use proper methods when disposing of used filters Call AZI Customer Service at 800 528 7411 601 470 1414 or e mail support azic com for a copy of the Resisorb or Sofnolime MSDS or for other questions e Connect the new filters to the Tygon tubing ensuring all straight hose barbs point toward the intake pump corner of the case and elbow hose barbs point toward the sensor housing as shown in the illustration gt Push the Tygon as far as it will go onto the filter fittings e Push the filters into the mounting clips e Remove any crimps or twists in the tubing and ensure that tubing connections are secure If the tubing is loose readings may not be accurate Replace any tubing that has deteriorated due to heat and or age e Close the case and replace the screws e Dispose of all filters in accordance with state and federal environmental regulations AZI Customer Service 800 528 7411 or 602 470 1414 Page 20 of 67 Battery Connector Hold Down Screws e Press the power OFF button e Unplug the power cord Remove the two 2 side screws from the intake end of the instrument and open the case lid Disconnect the battery connector from the board Loosen the two 2 captive screws holding the battery bracket and remove the bracket Remove the old battery pack and replace with a new battery pack Repla
69. tomer Service 800 528 7411 or 602 470 1414 Page 42 of 67 Functional Test Troubleshooting If you don t achieve good results with the functional test procedure check the following Typically too high Ensure the calibration vessel temperature is stable Be sure to inject the Hg vapor ONLY after the display flashes approximately 2 seconds after SAMPLE is pressed Ensure there is no oxidation on the mercury drop in the calibration vessel Gently swirl the mercury drop in the calibration vessel Too Low Replace if necessary Ensure the instrument s intake 1s not blocked with foreign matter Check flow with a flow meter Ensure syringe is calibrated to Icc Use a new syringe needle Straighten or replace crimped or blocked internal tubing If you find the above does not solve your problem please call AZI Customer Service at 800 528 7411 or 602 470 1414 AZI Customer Service 800 528 7411 or 602 470 1414 Page 43 of 67 The gold coil personal mercury dosimeter is a unique collection device for mercury vapor The Jerome 431 X Gold Film Mercury Vapor Analyzer and the Personal Mercury Dosimeter determine personal exposure levels and ambient air concentrations as well as low levels of mercury in natural and stack gases The dosimeters are not compatible with the 431 XE model For personal sample collection the dosimeter is worn as close to the wearer s breathing zone as possible and is connected by tubing to a pump usual
70. ttery pack contains NiCad batteries Dispose of properly NOTE Install the zero air filter into the instrument s intake during storage C M filters contain Sofnolime For safety information see the supplier s Material Safety Data Sheet MSDS or call AZI Customer Service at 800 528 7411 601 470 1414 or e mail support azic com for a copy Dispose of all filters properly Zero air filters and scrubber filters contain Resisorb For safety information see the supplier s Material Safety Data Sheet MSDS or call AZI Customer Service at 800 528 7411 601 470 1414 or e mail support azic com for a copy Dispose of all filters properly AZI Customer Service 800 528 7411 or 602 470 1414 Page 18 of 67 Flow System The Jerome 431 X s flow system is the crucial link between the sensor and the sample For the instrument to perform correctly the flow system must be properly maintained The user maintainable components of this system are the intake filter 25 inch fritware a C M Filter two scrubber filters and connecting tubing Check the Preventive Maintenance Calendar on page 18 for a suggested schedule for changing fritware and scrubber filters The Tygon tubing in the system must be free of crimps for proper flow Part Part Number Scrubber Filter Z 2600 3930 C M Filter Z2600 3928 25 inch Fritware Filter 2600 3039 Tygon Tubing 1 8 I D 1 2500 3001 25 inch Fritware Filter Replace the 25 inch fritwar
71. ury concentrations Sampling in the survey mode is not as accurate Due to the decreased sample volume the accuracy of the instrument is reduced to 20 100 mg m e Press the power ON button gt The digital meter displays 000 If the unit is set to display in ng the digital meter displays 00 0 Disregard the digital meter s initial momentary readings Recharge or replace the battery pack if the LO BAT indicator REMAINS ON Refer to pages 17 and or 21 for the procedure gt To ensure the instrument s electronics have stabilized allow a 1 minute warm up before beginning the next step e Press and hold the SAMPLE button gt The instrument takes a normal 12 second sample displays the concentration at the end of the cycle and then goes into the survey mode sampling every 3 seconds The display flashes the measured concentrations at the end of each 3 second sample cycle e When you are finished surveying release the SAMPLE button gt The final survey value remains displayed until the next sample is taken e To lock the instrument in a survey mode gt Hold the SAMPLE button down until the sensor status indicator bar s _ begins flashing on the display gt Press the ZERO button then release the SAMPLE button gt The pump should continue to run and the display should update every 3 seconds gt The instrument remains in the survey mode until one of the following occurs The sensor is saturated ALO BAT low battery sig
72. zes the meter reading until the next sample cycle is activated thus eliminating drift between samples During the sample mode cycle bars on the LCD represent the percentage of sensor saturation Depending on the concentrations approximately sixty five samples containing 0 1 mg m Hg may be taken before the sensor reaches saturation After absorbing approximately 500 nanograms of mercury the sensor becomes saturated and needs to be cleaned This is accomplished by a manually activated 10 minute heat cycle or sensor regeneration that burns the mercury from the sensor This mercury is absorbed on internal filters to prevent any external contamination The solenoid bypass closes during the sensor regeneration cycle causing the air to pass through the scrubber filter providing clean air for the regeneration process The flow system s final scrubber prevents contamination to the atmosphere from the desorbed mercury The heat generated during the regeneration may cause some low level thermal drift To ensure maximum sample accuracy wait 30 minutes after regeneration before zeroing and using the instrument Zero Air Filter The Zero Air Filter removes mercury vapor mercaptans and hydrogen sulfide from the air sample Readings with the filter installed should be near zero Because air that is cooler than the instrument will cause low readings and warmer air will cause higher readings the Zero Air Filter should be used to equilibrate the unit t

Download Pdf Manuals

image

Related Search

Related Contents

Haier HWM70-98 User's Manual  2. 消費者に対する製品の環境配慮情報の提供について 2  17730-244 - gruporacsa.net  Harbor Freight Tools Pencil Torch Product manual  質問及び回答書その2  MODE D`EMPLOI  取扱説明書    Polaroid m536 User's Manual  INSTALLATION & OPERATOR`S MANUAL  

Copyright © All rights reserved.
Failed to retrieve file