image

Contents

1. S Yuna Zi Recommended user weight li Heel function gt lt 1 Chlorine X Do not iron lt 2 gt gt g nN XXXX XX XX Year Month Day CE Patient information category Da Do not wash Q Tumble dry Do not dry clean IP__ IP class Vv Foot placement ye Y a Read the user manual Group of function CE marked acc to Recycling MDD 93 42 EEC X XXX Category a Z Not in domestic refuse follow recycling instructions Directly on bed base Machine wash at 70 C Res Do not tumble dry Temperature range Double Insulated Ingress Protection Rating i The mattress should only be used lying in the lengthwise direction on the mattress 16 10 Spare parts Storlek Top part Olivia 80x200 cm CC 906911 CC 916210 80x210 cm CC 906912 CC 916211 85x200 cm CC 906913 CC 916212 85x210 cm CC 906914 CC 916213 90x200 cm CC 906916 CC 916214 90x210 cm CC 906917 CC 916215 105x200 cm CC 906910 CC 916216 120x200 cm CC 906920 CC 916218 Art no Name Size CC 901310 Cell PU 80x21 cm CC 901311 Cell PU 85x21 cm CC 901312 Cell PU 90x21 cm CC 901313 Cell PU 105x21 cm CC 901314 Cell PU 120x21 cm Please make a note on your order when ordering ventilating cells CC 901330 Heel cell PU 80x21 cm CC 901331 Heel cell PU 85x21 cm CC 901332 Heel cell PU 90x21 cm CC 901333 Heel cell PU 105x21 cm CC 901334 Heel cell PU 120x21 cm CC 8217 Fuse set of 2
2. CC 8201 Filter washable If other spare parts are needed please contact Care of Sweden or your local dealer for further assistance Top part CuroCell LF CuroCell Comfort fabric Bottom part CuroCell 17 CC 907410 CC 907411 CC 907412 CC 907413 CC 907414 CC 907415 CC 907416 CC 907418 CC 905511 CC 905512 CC 905513 CC 905616 CC 905617 CC 905510 CC 905521 11 Options and accessories 11 1 Options Comfort fabric For patients who are sensitive to fluctuations in the mattress a padded inner cover can be placed over the air cells to increase comfort Heel section There are two options available to relieve sensitive heels Emptying the air cells with check valves or lower air cells NOTE These options are only available with new orders Cable holder A pocket along one long side of the mattress that holds the power cord in place to minimise the risk of tripping over the cord or the cord becoming trapped under the wheels of the bed NOTE This option is only available with new orders Sheet holder Specialised clips to which the sheet can be attached Six required per mattress 11 2 Accessories Art no Description and size CC 900220 Heel function with valve 4 cells CC 900222 Heel function with valve 3 cells CC 9094 Transport bag 80 90 cm CC 9095 Transport bag 105 120 cm CC 9075 Sheet holder CuroCell one of Top part LF low friction Top part with increased sliding function can be used for heavy immobi
3. Care of sweden Supporting Life User manual 2015 03 05 CuroCell 4 Alternating pressure system Item number user manual 95 001044 f 33171345 013282 Contents Important safety precautions Introduction Product description Assembly Installation Operation Cleaning and reconditioning Storage Maintenance Troubleshooting Technical description Spare parts Options and accessories Warranty Other information Oo oOo O A W 11 12 13 14 15 17 18 19 19 IMPORTANT SAFETY PRECAUTIONS Read all instructions prior to use Warning To minimize the risk of fire electric shocks or bodily injury The product must only be used for its intended purpose as described in this manual The product may not be combined assembled or repaired with other parts e g pump and mattress accessories or spare parts other than those described in this manual or other documentation from Care of Sweden 2 Do not use the product in wet rooms or where there is a risk of the pump coming into contact with water or another liquid Never handle a pump that has come into contact with water liquid Pull the plug out of the electrical socket immediately and send the product to an authorized service technician for servicing 3 DO NOT USE in the vicinity of or in contact with fire sources hot surfaces such as fire burning cigarettes or hot lamps Although the product has undergone fire testing it could become damaged if it comes int
4. 1 Quick connector unit air 2 Alarm off 3 Off On 0 1 4 Static mode 5 Power cord 6 Fuses 2 7 Bed hooks 8 Air filter 2 3 Pump functions 1 Power on off 1 0 ON OFF Button to turn the power on or off NoRMAL pressure LOW PRESSURE 2 Lamp for normal pressure NORMAL PRESSURE gt At start up Lights when the mattress has reached the set pressure and is ready for use During operation Steady green light while pressure is normal 3 Comfort setting 1 8 Suitable air pressure is set with the comfort dial 1 8 Guidelines based on user weight is given in the table on the pump NOTE In all uses where the head of the bed is raised the pressure must be increased at least one 1 step 4 Static mode STATIC When the STATIC button is pressed the alternating process is terminated and the pressure between the cells evens out The function can be used in association with patient care repositioning transferring in and out of bed Turn off this function by pressing the button It will return automatically to cycling after approx 30 minutes 5 Low pressure LOW PRESSURE The pump has an acoustic and light alarm that is triggered when pressure in the mattress is too low The alarm continues to sound until the ALARM OFF button is pressed on the side of the pump In the event of an alarm check the following Disconnected pinched tubes open CPR valve leakage at air cell connector The CuroCel
5. are made regularly The time interval must be evaluated by the responsible personnel depending on the patient s status diagnosis and general condition The mattress should only be used lying in the lengthwise direction on the mattress with the feet at the end marked with the feet symbol The mattress must always be cleaned between patients NOTE e For certain patients such as amputees are the recommended minimum weight not valid These users may however have other requirements as the entire mattress is not being loaded Please refer to sections 4 2 and 4 3 for operating control e When used together with positioning pillows read the user s manual carefully to ensure correct positioning of the patient e If you are using this product with an evacuation sheet it is the user s responsibility to ensure that it is safe to use for evacuation e Be careful when using bed rails or other protection on the bed so that the mattress does not become trapped or damaged e The mattress must not be lifted with a person lying on it or used for any other type of transport other than a possible evacuation using an evacuation sheet e This mattress is unsuitable for use during x ray examinations because of the risk of blurred images or artefacts that may lead to diagnostic errors 1 National Pressure Ulcer Advisory Panel European Pressure Ulcer Advisory Panel and Pan Pacific Pressure Injury Alliance Prevention and Treatment of Pressure
6. or changing the comfort settings Insert your hand between the air cells directly underneath the user s pelvic region use a vertical hand There should be a clearance of 2 3 fingers between mattress base and the patient If the distance is not sufficient increase the comfort setting to the next highest step Wait 10 minutes and perform a new hand check 4 4 Alarms When the pressure in the mattress drops to an abnormally low level the yellow low pressure lamp lights and an acoustic alarm sounds First check that the CPR valve is closed and that all connections are intact and correctly installed in accordance with paragraph 3 The yellow LED blinks until the pressure is corrected and restored The alarm is deactivated by pressing the alarm off button If the problem is not resolved within 3 minutes the alarm will sound again The yellow lamp flashes until the pressure is corrected and reset See section 8 for troubleshooting WARNING If the air pressure remains low check the mattress and tubes for damage Repair if necessary or contact the manufacturer or your local dealer for help 4 5 CPR In the event of an emergency requiring CPR cardiopulmonary resuscitation Quickly pull the CPR strap remove the quick connectors at the side of the pump to deflate the mattress as fast as possible On this product there are two possible positions for the CPR valve at the head end 1a or the foot end 1b Position 1a At th
7. resolved by putting the pump on a flat steady surface or by placing a towel between the pump and bed If your questions are not answered in the above information please contact Care of Sweden or your local dealer 14 9 Technical description 9 1 Technical specification PUMP SPECIFICATION Model o Soe Voltage f 22OVe24OV 50 6062 S Power consumption JOVA Typeoffuse TIA SOV Cycle periods fF Somn S Dimension LxWxH 30x105x225em S Weist o e po Storage e50 oo o Shies S Po Humidity O Operation 20 80 non condensing a ee Storage 10 90 non condensing Atmospheric pressure 700hPa 1060hPa O Classification Class i Type BF P22 MATTRESS SPECIFICATION Model Ff Croce O Dimension LxWxH 80 85 90 105 120x200 210x21cem_ __ Weight Pf 10 5 kg 8OKZODEM o NOTE Care of Sweden retains the right to modify the product specification at any time WARNING Follow operating and maintenance instructions carefully to prevent injury as well as damage to the product or other equipment 9 2 Labelling The mattress are labelled with product name production date article number size care and wash instructions EAN bar code as well as space for own labelling The product also displays which side is up down and head foot 15 WO REF Article No Counteract shear force Machine wash at 95 C Class Il equipment 3 Symbol Definitions Manufacturer GROY LINNA x OLN 7
8. Disassembly and recycling Except for certain parts of the pumps energy recovery is possible for almost all material in CuroCell products through incineration in waste incineration facilities Pump A used CuroCell pump must not be disassembled It must instead be taken to a recycling centre and sorted as electronic waste Mattress A used CuroCell mattress must be taken to a recycling centre and sorted as combustible waste For more information please contact Care of Sweden or your local dealer 13 2 Returns and complaints Complaints are to be reported to the supplier Return of a product shall follow the supplier s instructions and can only be done after the complaint has been reported Damages due to the transportation should be reported to the transportation agent 19 Manufactured by Care of sweden Supporting Life Contact Phone 46 771106600 Fax 46 325 12840 Email info careofsweden se Internet www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Visit address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns vag 4A SE 514 33 Tranemo SWEDEN Supplied by Care of Sweden and CuroCell are trademarks of Care of Sweden AB Care of Sweden AB 2015 20
9. Ulcers Quick Reference Guide Emily Haesler Ed Cambridge Media Osborne Park Western Australia 2014 1 3 Other information Action plan Pressure ulcers To achieve the best possible result for this product it is recommended that a well planned and well structured working environment be used In our guideline Action plan you will find examples which can be applied and put into practice when using our products Mattress guide This practical guide gives you as a care giver help in choosing the correct mattress from the Care of Swedens range CuroCell 4 is a mattress in function group C This information can be found on our website www careofsweden com or you can order a copy from our Customer services by sending your details to info careofsweden se 2 Product description 2 1 Pump and mattress 1 CPR rapid deflation a Head b Foot 2 Pump 3 Mattress 4 Pump connector 1a Tubing system with CPR in head 4 static head cells 9 ventilated cells optional Static safety cells Cell in cell i ORE ae RO AAR RARAAARRARARARAA j j j A NESE EE SPREE EERE EEE EEE EES NJ 5 NNN C P R 1b Tubing system with CPR in foot 4 static head cells 9 ventilated cells optional Static safety cells Cell in cell A a a aS aa aa cls GES ARS As as Gs As aS aS a R With the optional heel section is the four cells in the foot end a bit lower to in order to relief the heels 2 2 Pump
10. an audible click sound when the connectors are properly attached 4 Check that the power switch on the side is set to O OFF Connect the plug to an approved electrical socket 220 V 4 Operation Note Always read the user manual prior to use 4 1 Start up 1 Start the pump by pressing 1 On 2 Air starts to inflate the mattress Note We recommend that the comfort dial be set to 8 max when inflating a mattress 4 2 Comfort setting 7 we i The air pressure in the mattress can be adjusted according to 0 30 1 2 3 the user s desire for comfort Always consult the prescriber 40 oe 4 or other authorized personnel for appropriate setting 50 3 4 5 1 Use the comfort dial to set a suitable air pressure based 60 70 4 5 6 on the settings table on the pump 80 90 5 6 f 100 110 6 7 8 2 When the green indicator lamp for normal pressure 120 140 7 8 8 lights the mattress has reached the set pressure 150 200 8 8 3 The mattress is now ready for use NOTE e lf the head of the bed is raised pressure must be increased at least one 1 level When only parts of the mattress are under load e g if the patient is an amputee the pressure should be increased 1 or 2 levels e For recommended user weights refer to section 1 2 4 3 Hand check Hand check is used to ensure that the mattress is properly inflated This should be done regularly once per shift and after changing the position in the bed
11. e head end of the mattress Position 1b On tubes between the on the right side user s right hand side pump and mattress NOTE This connector is held in place with the elastic strap 4 6 Transport function If the patient must be moved in bed the tubes can be connected together detached from the pump and the mattress put in transport mode We recommend using this mode only for short periods of time 10 4 7 Heel function with check valves option This feature allows pressure relief of the heels Disconnect the air cell which is directly under the heel This air cell will then be emptied for air Press the bottom 1 and disconnect the plug Let the cell remain in its place in the mattress The air cell is activated by assembling the connection again 5 Cleaning and reconditioning It is important to follow the instructions below before using the product to another user Consult the hygiene responsible or the supplier for help and responsible or the supplier for help and instructions when ensure 5 1 Cleaning PUMP WARNING Do not open the pump housing risk of electric shock Service and maintenance must be performed by authorised technicians Do not get the pump wet or submerge it in any liquid Wipe off the pump with a damp cloth and a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 If using a diffe
12. hat is well proven and been used for a long time Alternating pressure is when the pressure regularly alternates between the mattress cells to replicate body movement Variable pressure cycles are set up where there is little or no pressure This is optimal for skin recovery or to prevent skin damage Always read the user manual prior to use 1 1 General information This system is a medical device and is CE marked in accordance with MDD 93 42 EEC According to statutory regulations made by the authorities regarding medical devices the manufacturer is required to report all accidents or incidents involving there products We would be very grateful for all information involving accidents or incidents relating to our products by reporting them immediately to us here at Care of Sweden The system has been tested and approved according to the following standards IEC 600601 1 IEC 60601 1 2 SS EN ISO 12182 SS EN ISO 14971 SS EN ISO 10993 EN 597 1 EN 597 2 EN 14126 1 2 Intended use CuroCell 4 is an alternating pressure relief mattress replacement system The mattress can be used in all kinds of caring environments to prevent pressure ulcers as well as treatment up to Category 4 pressure ulcers The mattress is meant for use for persons of a recommended minimum length of 120 cm The recommended user weight is up to 200 kg CuroCell is a replacement mattress and shall be placed directly on the bed frame We recommend that position changes
13. l 4 mattress is made up of the patented cell in cell system Inside every air cell is an inner cell which is continuously filled with air This acts as a safety layer if a leak occurs 6 Alarm off ALARM OFF This turns off the acoustic alarm It will be activated again after about 3 minutes if the pressure does not return to normal In case of problems with the alarm see paragraph 8 for troubleshooting 3 Assembly Installation When unpacking the mattress check that none of the parts are damaged In the event of damage contact the manufacturer or your local dealer before using the product 1 Place the mattress on the bed base use the straps on the cover to securely fasten the mattress to the bed base Note e Make the bed with sheets for added comfort e Check that the CPR valve valves are properly closed e If the bed is height adjustable the mattress must only be attached to the bed s moving parts e Check the cells and snaps to make sure that they have been attached correctly 2 Hang the pump at the foot end of the bed or place it ona flat steady surface Route the power cord so that there is no risk of it causing someone to trip being run over by the bed wheels or getting pinched when the bed is raised or lowered If a cable holder is used optional open the snaps position the power cord in the pocket and close the snaps 3 Connect the air tubes from the mattress to the side of the pump There is
14. le patients 11 3 Other accessories To ensure the best level of comfort we recommend use of accessories from Care of Sweden since our products are designed to work well together Please contact Care of Sweden or your local dealer for accessories or more information on our products 18 12 Warranty 12 1 Scope This CuroCell system is covered by a 2 year warranty for manufacturing defects The warranty does not apply to normal wear and tear or product impairment or damage stemming from negligence or improper handling care 12 2 Service and maintenance This CuroCell mattress is a medical device in accordance with MDD 93 42 EEC that is used as an aid in preventing and treating pressure ulcers For this reason we recommend that the pump be regularly serviced and inspected to maintain functionality and performance Like other technical devices a pump that is properly cared for works better and lasts longer Service and maintenance must always be performed by Care of Sweden or one of its authorised technicians Only use spare parts that have been approved by Care of Sweden Use of non approved spare parts invalidates the warranty A warranty claim must be submitted to Care of Sweden before the product is returned Please contact Care of Sweden or your local dealer if you have questions regarding warranty and service 12 3 Life expectance This product s expected life is estimated to be 5 years 13 Other information 13 1
15. o contact with fire sources 4 Position and use the product in a manner that prevents it from becoming trapped or damaged NOTE Be particularly aware of trapping damage when using bed rails 5 Never use the product if the pump power cord or plug is defective the pump casing is damaged or it is not functioning properly Contact a service technician for examination and repair 6 Regularly check product functionality through a hand check see section 4 3 Also check that the comfort setting is correct as indicated in the table on the pump 7 Continual monitoring is required when the product is used for individuals needing special supervision such as children 8 The mattress are protected with a hygiene cover please try to avoid the simultaneous use of multiple hygiene cover with this mattress Because of the risk that the vapour permeability can be destroyed 9 The hygiene cover does not allow liquid or air to penetrate but is vapour permeable Make sure that the patient is positioned correctly to avoid the risk of suffocation 10 Prevent damage to the hygiene cover from sharp objects 11 If the cover are supplied with handles on the long sides theses handles are only to be used for lifting and moving the mattress itself All other use will be at your own responsibility and is not covered by the product warranty 1 Introduction CuroCell air mattresses operate according to the alternating pressure principle a technique t
16. rent agent choose one that does not affect the pump housing MATTRESS COVER Covers can be wiped off with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 Excrements and blood spots must be removed immediately with cold water Carefully follow local instructions and usage instructions for the cleaner 357 Q ll Bet RI Multi layer covers should be divided before washing Wash with like colors Chlorine and phenol based cleaners could have a negative impact on the PU surface A and should be avoided If chlorine is used we recommend a mixture of max 1 11 INNER COVER AND COMFORT FABRIC accessories ts CE ill BX BS Clean the affected area with a mild cleaning solution for this purpose such as alcohol with or without tensides Ilsopropanol 70 oxidating solutions or chlorine max 1 5 2 Reconditioning PUMP See section 5 1 for instructions MATTRESS 1 Disconnect the tubes and let the air out also open the CPR valve 2 Remove the cover See instructions in chapter 5 1 for cleaning covers If the product is equipped with a foam core this should be removed and cleaned separately n Move the mattress to a clean area which is suitable for cleaning Wipe off the cells all tubes and the CPR valve with a mild cleaning solution for this purpose such as alcohol with or without tensides lsopropanol 70 o
17. wards and the filter slides out of its holder 3 Place the new cleaned filter in the holder and refit to the cover plate If the pump is used in a dirty environment the filter should be checked regularly 13 8 Troubleshooting Problem Solution Pump does not start Check that the power cord is connected to the power socket Check the fuses Alarm is on Check that the quick connectors to the air tubes mattress is deflating on the side of the pump are properly connected Check that the CPR valve valves are closed Open the mattress and check for leaks from air cells or tubes Check that all tube connectors are properly fitted and connected No air in head cells When the mattress is inflated the 4 static head cells are filled last Patient is bottoming The comfort setting may be set incorrectly Increase the pressure and wait a few minutes Do a hand check Refer to section 4 3 Check that the CPR valve valves are properly closed Mattress moves around Check that the air mattress is fastened to the bed frame with the straps underneath two at the head and two at each side of the bed Some cells have less air This is normal since the air flow alternates between every other cell for a set cycle time one cycle 10 minutes Pump is noisy Check how the pump is hung on the bed Resonance can occur vibrations are felt in parts of the bed Remove the pump and listen to see whether this makes a difference Can be
18. xidating solutions or chlorine max 1 Let it dry 5 Wipe off and disinfect all the work surfaces with an alcohol based cleaning agent intended for this purpose 6 Put the mattress together If cells for any reason have loosened from the tubes these must be replaced according to drawing in chapter 2 1 NOTE Check the hygiene cover each time the product is cleaned If damaged it must be replaced or repaired Also check the pump tube connectors and power cord during cleaning Damaged parts must be replaced or repaired 6 Storage The mattress and pump are best stored in the product bag optional extra Handle the packed product with care Do not put anything heavy on top otf it 12 7 Maintenance 7 1 General 1 Check the power cord and plug for damage and scratches 2 Check that the mattress cover is intact and that the cover and air cells are correctly assembled 3 Check that the air flow alternates between each outlet on the side of the pump 4 Check for damaged or pinched tubes and connectors Contact the manufacturer or your local dealer for spare parts 7 2 Changing fuses 1 Pull the plug out of the electrical socket before replacing fuses 2 Remove the cover for the fuses with a screwdriver Pull out the fuse holder 3 Fit a new fuse of the same type T1AL 250 V and screw the cover back on There are two fuses 1 Unscrew the small round cover plate at the back of the pump 2 Lean the pump back

Download Pdf Manuals

image

Related Search

Related Contents

Package `coda`  FM/AM Compact Disc Player  Manuale di istruzioni Inverter 230V / 400V 0.25 kW 11  Registrador medidor de nivel de sonido Modelo 407764  Diesel Exhaust Fluid Transfer Pump - Fill-Rite  SPMA9584 AIRCRAfT TELEMETRy JETCAT  4-kanal-multifrequenz-empfänger für funkmikrofone    Operating Instructions and User Guide  Janvier 2014 - Neufchâteau  