Introduction to the Fusion TOC Analyzer
Contents
1. 0 9 LAN i ost El Ff uN Vv 0 7 E NH3 si Transm 0 6 e 5 CHI 0 1 t N20 S02 CO2 N20 A N 1 2 3 4 5 6 7 8 9 10 Wavelength um Figure 5 Absorption spectra for selected gases The tungsten light source sends a single beam through the CO2 gas This beam reflects off a gold plated mirror to the detector The Fabry Perot Interferometer filters the light beam at an absorbance wavelength of 4 3 microns which is specific for CO2 A second beam is then sent and filtered at an absorbance wavelength of 3 9 microns which is non specific for CO2 The difference between the absorbance readings is The Fabry Perot Interferometer is silicon based micro machine semi conductor They apply an electrical current to the semi conductor and are able to tune select different absorbance wavelengths The detector is a Thermopile a thermocouple which measures the heat potential and turns it into an electrical signal Introduction of the Fusion TOC Analyzer Sales Support 800 874 2004 Main 513 229 7000 4736 Socialville Foster Rd Mason OH 45040 www teledynetekmar com IR detector Hermetically sealed Cap Light source FPI detector Sample cell with reflector Tunable FPI filter lt gt An inherent advantage of this tec
2. Fusion uses takes place only after the gases reach equilibrium inside the detector cell This pressurization of the sample gas stream inside the detector a patent pending technique allows for increased sensitivity and precision compared to a flow through cell detector by measuring all of the oxidation products from the sample in one reading The output signal is proportional to the concentration of CO in the carrier gas from the oxidation of the sample Results and Discussion The Fusion enables the analyst to use fewer calibration curves and check standards to analyze a wide selection of samples The new TOC TekLink software allows for continual flow of analysis from auto calibration to analysis of samples in one continuous schedule TOC TekLink processes and selects the most recent standards and updates the calibration curve Acceptance criteria of the calibration curve such as r value y intercept and slope may be implemented in combination or individually All samples analyzed following the newly generated curve are then automatically applied to that curve Calibration and analyses of 0 5 1 5 and 10 mg C L content was performed The FUSION instrument parameters were set at the following levels e 4 0mL sample injection volume 0 5mL acid volume 0 6 mL reagent volume and a UV Reactor pre rinsing volume of 4 0mL e The auto calibration curve was made from deionized water 1 5 and 10 mg C L using a 20 mg C L potassium hydr
3. Grouping like compartments allows for easier maintenance as well as separation of the electronics and valves preventing components from possible liquid exposure which could cause damage in the unlikely event of valve failure There are three autosampler tray sizes 75 position 40mL vials 90 position 55mL test tubes 120 position 20mL test tubes All tray sizes have four 125ml bottle positions in the center of the rack Directly below the autosampler a uniquely designed and improved automated syringe delivery system allows for remarkable precision and accuracy The Fusion handles all liquid movement through a redesigned seven port valve that is capable of accurately delivering 0 5 25 mL volumes in 20 uL increments A dedicated IC chamber prepares samples for TOC analysis by adding phosphoric acid to the untreated sample and sparging away inorganic carbon Between the Syringe Pumper and the IC chamber is a highly efficient mist trap for removing moisture from the sample gas before it enters the NDIR Detector protecting the detector as well as eliminating possible water interference from the sample Sodium Persulfate is added to an enhanced UV reactor located left of the Syringe Pumper oxidizing all of the carbon to CO2 with the aid of an ultraviolet light This newly designed UV Reactor increases sample conservation and delivers better UV radiation interaction between samples and the oxidant The Fusion utilizes an enhanced halogen scrubb
4. e of the flow between the detectors there is a corresponding peak displayed usually in millivolts mV on the instrument Once carrier gas purges out carbon dioxide in the sample cell the detector output signal decreases to its previous level because the flow between the detectors has returned to equilibrium The mV seconds area of this peak is the raw data for the analysis Introduction of the Fusion TOC Analyzer Sales Support 800 874 2004 Main 513 229 7000 4736 Socialville Foster Rd Mason OH 45040 www teledynetekmar com Gas Out Gas In Infrared Energy Source F Front Detector Cell Rear Detector Cell Motor for Chopper Blade Mass Flow Sensor Sample Cell Cell Window Chopper Blade Figure 3 Phoenix 8000 flow through detector cell Sample Gas In Sapphire Gold Plated and a j Coated Mirror IR Source Infra Red IR IR Detector Absorbed by CO e ee a gt a x Vent Valve v Sample Gas Out Figure 4 TOC Fusion detector cell Anatomy of a SPC NDIR detector In contrast to an analyzer using a flow through cell detector the Fusion utilizes pressurized detection or static read figure 4 In this technique oxidation of the specific caroon component and its movement using a non interfering carrier gas into the detector is identical However the static read detector pressurizes the detector cell by closing the outlet valve and takes a single measure
5. er that removes chloride contained within the sample gas as it passes through on its way to the NDIR detector Introduction of the Fusion TOC Analyzer Sales Support 800 874 2004 Main 513 229 7000 4736 Socialville Foster Rd Mason OH 45040 www teledynetekmar com Seven valves direct the gas flow controlled by the MFC The combination of valves and the MFC allows for mode to mode changes to the flow rate and the user to optimize the sparge flow of the sample resulting in lower gas usage since flow rates can be set instantly Higher flow rates sweep the flow path between samples decreasing the time of analysis Along with COs gas an inert carrier gas from the UV oxidation chamber carries water vapor through the mist trap to the improved permeation dryer The permeation dryer has a smaller internal diameter that permits more surface area exposure to the carrier gas stream which improves its capacity to remove water vapor Inside the right partition of the Fusion is the detector designed to tolerate higher concentrations of S02 SO3 moisture and halogens This newly designed detector uses Static Pressure Concentration SPC combined with NDIR technology to achieve unrivaled performance at lower concentrations With traditional flow through cell detectors like the Phoenix 8000 UV Persulfate oxidation liberates carbon to create CO gas The detector measures the adsorption of the infrared light over time as the COs passes through figure 2 The res
6. hnique is that all of the CO is in the cell at the same time for the detector to measure This allows the sensitivity of the analysis to be significantly increased In addition static read detection does not have inherent error associated with integration inaccuracy figure 6 The elimination of this error allows for lower detection limits and increased precision Figure 6 Typical Static Pressure Concentration SPC read display Instrument Software Software is in a user friendly Windows format design From the main instrument setup screen access to nearly every function of TOC TekLink is available The main window appears initially as a grey box Five main dialogue boxes can toggle ON OFF into the main window according to user preference There is also the Unit Status window an Event Dialogue window and a Status bar at the bottom which shows the mode and the connection status On the Unit Status window are various details of the analytical mode instrument status and zones including flow rates pressures and error messages A Sample Information window a real time NDIR signal chart and a recent history of results are all displays within the Run screen Sample Setup allows the user to enter relevant information about the sample ID type of analysis sample calibration blank etc method of analysis TOC TC IC TC IC and the number of repetitions performed per sample The Tools menu contains many useful components From
7. ment to determine the amount of CO inside the detector cell This single reading correlates directly to the concentration of the carbon contribution from the sample The static read detector figure 6 uses no moving parts to measure carbon dioxide Inside the detector a miniature lamp produces an electronically pulsed infrared beam A mirror reflects the beam and re focuses it back to an IR detector Behind a silicon based Fabry Perot Interferometer FPI is the IR detector The tuning of the electronics of the miniature FPI is so that its measured wavelength converts between the absorption bandwidth of CO gas and a reference bandwidth When the FPI passband coincides with the absorption wavelength of the CO gas the IR detector experiences a decrease in the light transmission The measured wavelength of the FPI changes to the reference band that has no Introduction of the Fusion TOC Analyzer Sales Support 800 874 2004 Main 513 229 7000 4736 Socialville Foster Rd Mason OH 45040 www teledynetekmar com absorption lines and the IR detector experiences a full light transmission The degree of light absorption in the gas indicated by the ratio of these two signals is proportional to the gas concentration As stated earlier carbon dioxide shows a unique adsorption spectrum when infrared energy passes through it allowing any NDIR detector to distinguish it from other gases figure 5 CO2 CO CH4 NH3 H20 N20 SO2 1 7 U
8. ogen phthalate KHP stock solution shown in Figure 8 Introduction of the Fusion TOC Analyzer Sales Support 800 874 2004 Main 513 229 7000 4736 Socialville Foster Rd Mason OH 45040 www teledynetekmar com Calibration Graph 400 000 e Calitr ation Points elmer y mix 5 350 000 ere m 2 337 n n ani b 0 507 Coefficient of Correlation r 0 99991 250 000 200 000 NOIR Apsorbance Abs 150 000 1 100 000 50 000 Anshysse type TOC Re gr ssion Calculation Method Lirie TOC Comments Semple ID Y Row Vale x Expected Message 2007 10 54 AM O1 Waker 0 9727 15 2007 11 12 AM 11 000 pom 29 3493 3 15 2007 11 30 AM 15 000 pom 140 5493 315 2007 11 49 AM 10 000 pem 24 7427 3415 2007 12 07 PM Figure 8 Fusion calibration curve Sample Concentration Result n 3 ppm TOC RSD 100 ppm 103 2512 0 16 10 ppm 10 4552 0 34 100 ppm 102 6462 0 20 1 ppm 1 0048 1 60 Carryover calculated by using a dilution method that did not over range the detector Sample Cross Concentration Result Contamination n 3 ppm TOC RSD Carryover 1 ppm 0 9803 0 10 N A 50 ppm 34 5803 3 80 N A 1 ppm 1 1767 0 45 0 57 1 ppm 1 0691 0 59 0 26 1 ppm 1 0056 0 73 0 07 Table 2 Carryover calculated by using a non dilution method to over range the detector Two experiments demonstrate the ability of the Fusion to recover from analyzing high carbon concentration samples The first uses the automatic dilu
9. ples Tekmar Pittsburg Conference Presentation McHale W TOC Overview for High Temperature Combustion Tekmar Presentation The author thanks Eric Heggs and Stephen Proffitt for their invaluable input into this application note Introduction of the Fusion TOC Analyzer Sales Support 800 874 2004 Main 513 229 7000 4736 Socialville Foster Rd Mason OH 45040 www teledynetekmar com
10. rganic Non Purgeable Carbon POC Organic Carbon Dissolved DOC Figure 7 Carbon analysis diagram To determine TOC by the NPOC method the Fusion transfers the sample and reagents to its reactors by first removing inorganic carbon then oxidizing the carbon in the sample and finally releasing it in the form of carbon dioxide It then uses a carbon free carrier gas to transfer the CO from the sample to the NDIR detector in the following sequence 1 Removal and venting of IC and Purgeable Organic Carbon POC by acidification and sparging in the IC sparger HCO CO CO POC_ _ gt CO POC carrier gas acid Introduction of the Fusion TOC Analyzer Sales Support 800 874 2004 Main 513 229 7000 4736 Socialville Foster Rd Mason OH 45040 www teledynetekmar com 2 Following IC removal an aliquot of the sparged sample is transferred to the UV Reactor and persulfate reagent is added to oxidize the organic carbon based on the following chemical reactions Free radical oxidants formed S 0 gt 2S0 H O gt H OF SO H O gt SO OH H Excitation of organics k R Oxidation of organics R SO OH gt nCoO Once the oxidation products in Step 2 are in the NDIR detector via the inert carrier gas the pressure inside the detector rises to a predetermined set point by closure of the exit valve The unique single measurement methodology the
11. the Tools dropdown box the Maintenance Tasks Blank Editor Standard Editor Diagnostics Update Firmware System Properties User Manager and Change Password functions are available One of the most helpful maintenance tasks is the Leak Check function This permits pressurization of the flow path and ensures flow path integrity when the system is under pressure Unlike the Phoenix 8000 the Fusion performs a single system blank The Introduction of the Fusion TOC Analyzer Sales Support 800 874 2004 Main 513 229 7000 4736 Socialville Foster Rd Mason OH 45040 www teledynetekmar com instrument blank screen displays the latest blank version and a View History button of the previous system blanks The Standard Editor accesses the list of standard sets previously created and controls the addition of new standard sets or the editing of existing ones System suitability criteria can be set here as well When creating or modifying a standard set the user may choose to use the center autosampler positions for the stock standards or traditional 40mL vials Using the same calibration file for multiple ranges or modes with TOC TekLink software is allowable The Diagnostics submenu permits manual control of the instrument valves seven port valve and syringe drive autosampler diagnostics instrument benchmark software debug and Mass Flow Controller Chemistry of Oxidation and Analysis Flow Chart Total Carbon TC o Purgeable O
12. tion feature of TOC TekLink This feature allows the user to preset a method to dilute each 100 mg C L sample to a predetermined concentration within the active calibration range figure 8 and table 1 The major aspect of the carryover calculation is the effect of 100 mg C L on the first replicate of the subsequent solutions between a ten and one hundred fold difference The carryover for the 10 mg C L 1 mg C L samples is less than two percent In the final experiment the major aspect of the carryover calculation is the effect of 50 mg C L which will over range the linearity of the calibration curve by five fold on low level samples The Fusion again demonstrates its superb capability to rinse the sample pathway in between samples as carryover is less than 1 Introduction of the Fusion TOC Analyzer Sales Support 800 874 2004 Main 513 229 7000 4736 Socialville Foster Rd Mason OH 45040 www teledynetekmar com immediately following the out of linear range sample and two samples subsequently less than 0 1 carryover table 2 References Phoenix 8000 User Teledyne Tekmar Mason OH 2007 FUSION User Manual Teledyne Tekmar Mason OH 2007 Yang S S Miller M Martin J and Harris J Evaluation and Application of a New Total Organic Carbon Analyzer Tekmar Application Note Wallace B Application of an Innovative Infrared Detection Technology to Total Organic Carbon Analysis of Pharmaceutical and Environmental Water Sam
13. uling automatic dilution capabilities using stock standards or samples universal calibration and blank methods for all detector ranges Automated self diagnostic tools 21 CFR functionality Electronic Records and Electronic Signatures ERES exportable reports in various formats and simple sample scheduling with minimal method development are also new features standard with the Fusion Default analytical parameters are set for Total Carbon TC Total Organic Carbon TOC Inorganic Carbon IC and Total Organic Carbon by difference TC IC analyses The range of detection is from 0 2 ug C L 4 000 mg C L with carryover of less than one percent Sales Support 800 874 2004 Main 513 229 7000 4736 Socialville Foster Rd Mason OH 45040 Copyright 2007 Teledyne Tekmar Company All Rights Reserved www teledynetekmar com Hardware Instrument Hallmarks Integrated High Performance Autosampler Enhanced UV Oxidation Reactor Proficient Inorganic IC Carbon Removal Rapid and Precise Sample Delivery with Ultra low Carryover Figure 1 Hallmarks of the TOC Fusion The Fusion is divided into four main compartments an integrated autosampler and needle housing a visible wet chemistry section where all sample manipulations occur the Mass Flow Controller MFC the valves needed to route gas flow an internal water removal system Permeation Dryer and the NDIR detector and the electronics needed to control the instrument
14. ulting measurement correlates to a peak Integration of the peak height correlates to the concentration of carbon in the sample S E g Z Figure 2 Typical flow through cell detector gas peak Anatomy of a flow through cell NDIR detector Inside the detector are two detector cells one front and one rear Infrared energy traverses through the sample chamber then on to both detector cells A rotating chopper blade interrupts the infrared energy beam at regular intervals causing changes in the energy levels reaching the detector cell The pulsing of the infrared energy results a pulsed pressure flow through a mass flow sensor between the detector cells The mass flow sensor emits an electrical signal proportional to the magnitude of the flow When there is no carbon dioxide in the sample cell the maximum amount of infrared energy enters the detector cell Carbon dioxide displays a unique absorption spectrum when infrared energy passes through it which allows the NDIR detector to distinguish it from other gases see figure 5 As the sample cell receives the carbon dioxide gas it absorbs some of the infrared energy normally sent to the front detector The reduction of energy entering the forward detector cell causes a decline in its pressure This results in a flow change between the front and rear detector cells As the mass flow sensor located between the front and rear detectors emits an electrical signal proportional to the magnitud
15. we TELEDYNE INSTRUMENTS Introduction to the Fusion TOC Analyzer Tekmar A Teledyne Technologies Company Y Application Note By Stephen Lawson Abstract Building on nearly 40 years of design and manufacturing experience Teledyne Tekmar introduces its fifth generation Total Organic Carbon TOC UV Persulfate analyzer The Fusion The Fusion offers increased flexibility software intelligence and user friendliness making this analyzer ideally suited for demanding laboratory needs By implementing the patent pending Static Pressure Concentration SPC technology the Fusion is able to achieve unprecedented low end sensitivity from a Non Dispersive Infrared NDIR detector The Fusion demonstrates excellent precision at low ppb parts per billion levels along with the ability to minimize carryover Vt 7 SAYS KEY WORDS Fusion Total Organic Carbon TOC Total Carbon TC Inorganic Carbon IC TOC analyzers Non Dispersive Infrared Detection NDIR pharmaceutical drinking water waste water Introduction Total Organic Carbon TOC in water is an important analytical parameter in evaluating water quality Measurement of the TOC concentration plays a critical role in a variety of industrial and environmental applications The Fusion and TOC TekLink Software combine to provide a wide range of user friendly features and reliable data analysis The TOC TekLink software provides the user with unattended auto calibration sched