User Manual - Chart Industries
Contents
1. 8 Observe that the flowmeter on the oxygen concentrator now indicates the total combined amount of flow in use at the flow stations which are connected CAUTION It is desirable to establish a protocol to monitor and verify flows at all SureFlow outlets as well as from the oxygen concentrator on a periodic basis Units Equipped with Barb Fitting Oxygen Outlets When connecting oxygen concentrators with a barb fitting as the oxygen outlet such as AirSep s VisionAire Oxygen Concentrator the oxygen hosing will not be needed Instead securely connect standard oxygen tubing to the barb fitting on the oxygen concentrator and then using an oxygen adapter AirSep part number F0025 1 on the inlet of the SureFlow flow station connect the other end of the tubing See Figure 5 completely clockwise Figure 5 WARNING AIRSEP VISIONAIRE OXYGEN CONCENTRATOR MODELS REQUIRE A MINIMUM FLOWRATE OF 1 LPM TO ACHIEVE HIGH CONCENTRATION OXYGEN WHEN SUREFLOW IS USED WITH THE VISIONAIRE ENSURE THE COMBINED FLOWRATE IS AT LEAST 1 LPM IF COMBINED PATIENT USE IS LESS THAN 1 LPM IT WILL BE NECESSARY TO VENT SOME OF THE OXYGEN TO ATMOSPHERE THROUGH AN UNUSED OXYGEN OUTLET ON THE SUREFLOW IN ORDER TO ACHIEVE A TOTAL FLOWRATE OF AT LEAST 1 LPM Sure low Do not use abrasive powders or chemicals Clean only the exterior of the SureFlow unit which can be disinfected with either acommon chemical disinfectant or a diluted solut
2. Attach up to five 5 cannulas at the oxygen outlets of the flow station depending on how many patients will be receiving oxygen therapy Cannula length for each outlet can be a maximum of 50 feet 15 meters Figure 3 Note Before setting flowmeters ensure that the SureFlow station is level and that you are at the correct viewing angle to properly read and adjust the settings See Figure 3 Whether the flow stations are mounted on a wall or located on a desk or table top make sure that you are positioned in front of the flowmeters to be adjusted bending down to eye level as necessary so that when looking at the flowmeter each of its back and front black horizontal line markings are in exact alignment appear as just one horizontal line per increment Proper adjustment will appear to split the flowmeter ball horizontally across the center of the ball as just one line 7 Set the desired flow on each flowmeter used at the flow station The maximum flow will be dependant on the capacity of the oxygen concentrator used The sum of the flow settings on the flow station s flowmeters must not exceed the capacity of the oxygen concentrator WARNING IT IS VERY IMPORTANT TO RE ADJUST THE FLOWS AS NECESSARY ON ALL FLOWMETERS BEING USED ON THE FLOW STATION EACH TIME ADJUSTMENT OR DISCONTINUED USE TO ONE OR MORE OF THE FLOWMETERS IS MADE TO ENSURE ACCURATE PRESCRIPTIONS ARE ADMINISTERED THROUGH ALL OXYGEN OUTLETS
3. CAIRE Inc 2200 Airport Industrial Drive Ste 500 Ball Ground GA 30107 USA 1 800 482 2473 Worldwide www CAlREmedical com Medical Product Services GmbH Borngasse 20 35619 Braunfels Germany Worldwide www CAlREmedical com CE 0459 LF CAIRE A Chart Industries Company AirSep _ SeQual As of the date of this publication SureFlow AirSep and their related logos as applicable are trademarks registered by CAIRE Inc descriptions specifications and or processes to its products at any time without prior notice and with no further h g reserved by us a a H e en I III Il II H herein are reserved by us as applicable MN152 1 Rev B
4. II A Chart Industries Company AirSep SeQual User Manual SureFlow Sure er Contents User Controls amp System Status Indicators rsvrnnrnnvnnvnnvnnvrnnvnnvnnennen 3 Warnings Cautions Notes nsvranrnnvnnvnnvrnnnnnvnnennvnnnnnnnnnennennennnnnnennennenne 4 Introduction WE 4 Componenes scie osmotic aaa i uta ian anda 5 Procedure for Set Up greuss vas caiac aia aa lt ai 6 Using One SureFlow Station r rrnrnnvnnvnnvnnnvnnvnnvnnvnnnnnnvnnennennnnnnnnnennennen 6 Oxygen from Dual Flow Concentrators rsrnnvnnvnnnrnnvnnvnnennvnnnnnnennennen 7 Using Multiple SureFlow Stations s rnnrnnrnnvnnvnnvnnnnnnvnnvnnvnnvnnnnnnvnnennen 8 Units Equipped with Barb Fitting Oxygen Outlets mun 9 TT ns ias gat ot ast i ln niente nee 9 Sp cifications ac uni 10 2 MN152 1 B Sure er User Controls amp System Status Indicators Definition Warnings and Cautions No serviceable parts inside Do not open cover Do not expose to open flames No Smoking Icon Do not smoke near unit Complies with the 93 42 EEC directive drawn up by the approved organization No 0459 Use no oil or grease Read user manual before operation See user manual for instructions This symbol is to remind the equipment owners to return it to a recycling facility at the end of its life per Waste Electrical and Electronic Equipment WEEE Directive Proper disposal of waste of electrical and electronic equipment requir
5. ON THE UNIT IT MAY TAKE SEVERAL ADJUSTMENTS TO STABILIZE THE FLOW ON THE SUREFLOW FLOWMETERS 8 Observe that the flowmeter on the oxygen concentrator now indicates the total combined amount of flow in use at the SureFlow flow station CAUTION It is desirable to establish a protocol to monitor and verify flows at all SureFlow outlets as well as from the oxygen concentrator on a periodic basis WARNING AIRSEP NEWLIFE INTENSITY OXYGEN CONCENTRATOR MODELS REQUIRE A MINIMUM FLOWRATE OF 2 LPM TO ACHIEVE HIGH CONCENTRATION OXYGEN WHEN SUREFLOW IS USED WITH THE INTENSITY ENSURE THE COMBINED FLOWRATE IS AT LEAST 2 LPM IF COMBINED PATIENT USE IS BELOW 2 LPM IT WILL BE NECESSARY TO VENT SOME OF THE OXYGEN TO ATMOSPHERE THROUGH AN UNUSED OXYGEN OUTLET ON THE SUREFLOW IN ORDER TO ACHIEVE A TOTAL FLOWRATE OF AT LEAST 2 LPM Sure er Oxygen from Dual Flow Concentrators If the oxygen source for the SureFlow station s is made available from an oxygen concentrator with two flowmeters and two outlets such as from AirSep s NewLife Intensity or NewLife Intensity 10 it is possible to use one of the flowmeters on the concentrator for SureFlow and to use the second flowmeter for another application For example if the SureFlow is using a total combined flow of 4 LPM from a NewLife Intensity 10 which has been set for its maximum capacity of 10 LPM the flowmeter corresponding to the outlet connected to SureFlow refl
6. ects the 4 LPM flow in use This allows a maximum of 6 LPM to be set on the secondary flowmeter for another oxygen application Note The combined flows from the SureFlow flow station and the secondary flowmeter cannot exceed the total capacity of the oxygen concentrator If adjustments to any of the SureFlow flowmeters is made to increase or decrease the amount of flow the result will be that more or less flow will be available for the secondary flowmeter application WARNING AIRSEP NEWLIFE INTENSITY OXYGEN CONCENTRATOR MODELS REQUIRE A MINIMUM FLOWRATE OF 2 LPM TO ACHIEVE HIGH CONCENTRATION OXYGEN WHEN SUREFLOW IS USED WITH THE INTENSITY ENSURE THE COMBINED FLOWRATE IS AT LEAST 2 LPM IF COMBINED PATIENT USE IS BELOW 2 LPM IT WILL BE NECESSARY TO VENT SOME OF THE OXYGEN TO ATMOSPHERE THROUGH AN UNUSED OXYGEN OUTLET ON THE SUREFLOW IN ORDER TO ACHIEVE A TOTAL FLOWRATE OF AT LEAST 2 LPM SureFlow User Manual 7 Sure low Using Multiple SureFlow Oxygen Flow Stations Two or more SureFlow flow stations can be connected together to divert the flow from a single oxygen concentrator or other oxygen source to even more oxygen outlets to serve additional patients The maximum capacity of flow that can be set from all flowmeters combined from the flow stations must not exceed the capacity ofthe oxygen source used To supply oxygen to the SureFlow units Note During set up nothing is initially connected to the oxygen co
7. ed Name and address of manufacturer Authorized representative in the European Community AirSep Corporation is ISO 9001 2008 and ISO 13485 2003 w CMDCAS certified SureFlow User Manual 3 Sure low WARNING REFER TO THE INSTRUCTIONS OR MANUAL FROM THE OXYGEN SOURCE MANUFACTURER FOR ALL WARNINGS CAUTIONS AND NOTES THAT MAY BE APPLICABLE WARNING OXYGEN PROMOTES RAPID BURNING DO NOT ALLOW SMOKING OR OPEN FLAMES IN THE SAME ROOM WHERE OXYGEN IS BEING USED WARNING CERTAIN OXYGEN CONCENTRATORS MAY REQUIRE A MINIMUM FLOWRATE TO PRODUCE HIGH CONCENTRATION OXYGEN WHEN USING SUREFLOW WITH THESE MODELS THE MINIMUM FLOWRATE MUST BE ACHIEVED THROUGH COMBINED PATIENT USE OR WHEN NECESSARY BY ALLOWING SOME OF THE OXYGEN TO VENT OUT OF AN UNUSED OXYGEN OUTLET CAUTION The oxygen source must not exceed 20 psig 138 kPa Note This manual covers connecting the SureFlow to an oxygen concentrator source If you intend to use another oxygen source consult with the manufacturer or supplier of that oxygen equipment Introduction SureFlow is a new and uniquely engineered flow station developed by AirSep Corporation for economically administering medical grade oxygen to multiple patients from an oxygen concentrator or other oxygen source gas or liquid This product enables clinicians and medics around the world to manage and redirect the flow from a single oxygen source to a simple flow station to serve u
8. ending on how many patients will be receiving oxygen therapy Cannula length for each outlet can be a maximum of 50 feet 15 meters Note Before setting flowmeters ensure that the SureFlow station is level and that you are at the correct viewing angle to properly read and adjust the settings See Figure 3 Whether the flow stations are mounted on a wall or located on a desk or table top make sure that you are positioned in front of the flowmeters to be adjusted bending down to eye level as necessary so that when looking at the flowmeter each of its back and front black horizontal line markings are in perfect alignment appear as just one horizontal line per increment Proper adjustment will appear to split the flowmeter ball horizontally across the center of the ball as just one line 7 Set the desired flow on each flowmeter used at the flow stations The maximum flow will be dependant on the capacity of the oxygen concentrator used The sum of the flow settings on the flow station s flowmeters must not exceed the total capacity of the oxygen concentrator WARNING IT IS VERY IMPORTANT TO RE ADJUST THE FLOWS AS NECESSARY ON ALL FLOWMETERS BEING USED ON THE FLOW STATIONS EACH TIME ADJUSTMENT OR DISCONTINUED USE TO ONE OR MORE OF THE FLOWMETERS IS MADE TO ENSURE ACCURATE PRESCRIPTIONS ARE ADMINISTERED THROUGH ALL OXYGEN OUTLETS ON THE UNIT IT MAY TAKE SEVERAL ADJUSTMENTS TO STABILIZE THE FLOW ON THE SUREFLOW FLOWMETERS
9. ion of household bleach sodium hypochlorite 5 25 To use effectively mix a solution of 1 100 parts of bleach to water Wear eye and skin protection and allow the solution to remain on the surface for 10 minutes After using the disinfecting solution rinse with water and wipe dry Make sure unit is completely dry and then retest it before you return it to inventory Cleaning The manufacturers of sodium hypochlorite products recommend various strengths of a bleach solution for killing bacteria etc based on the type of germ to disinfect however a generally recommended solution is cups 237 ml of household bleach per gallon 3 79 L of water SureFlow User Manual 9 Sure low Specifications Dimensions Weight Maximum Inlet Pressure Flowmeters Flowmeter Accuracy Operating Temperature Storage Temperature 10 MN152 1 B 9 7 in W x 5 8 in H x 7 2 in D 24 6 cm W x 14 7 cm H x 18 3 cm D 3 28 Ib 1 49 kg 5 6 Ib 2 5 kg Shipping Weight 20 psig 138 kPa Low flow flowmeters can be used from 1 10 LPM to 1 LPM with 1 LPM increments with back lines for proper viewing angle 0 1 Ipm Flowmeter 3 of Full Scale 0 2 Ipm Flowmeter 5 of Full Scale 0 5 Ipm Flowmeter 5 of Full Scale 5 40 C 40 104 F 5 60 C 30 140 F Sure low CAIRE Inc reserves the right to discontinue its products or change the prices materials equipment quality i
10. ncentrator at this time This is important to prevent the concentrator from being overdrawn when using with the SureFlow units 1 Supply power to the oxygen concentrator and set the I O switch to the I position to turn it on 2 Set the flowmeter on the oxygen concentrator to the maximum capacity of the unit for example 5 LPM 8 LPM or 10 LPM depending on concentrator model 3 Use a second 7 ft 2 1 m oxygen hose AirSep Part No TU056 074 to connect the available end of the first flow station to an inlet fitting on the second flow station See Figure 4 Figure 4 4 If more than two flow stations are to be used repeat step 3 with additional oxygen hosing and flow stations wn As necessary test for leaks at all ends of the hose by applying soapy water to each of the connection points If bubbling occurs then leaks are present and the connection must be tightened properly until leaks are eliminated 8 MN152 1B t may be more apparent if there 1s a leak s by allowing the concentrator to run with all flowmeters turned off at the flow station If no leak is present the flowmeter on the concentrator should read zero Another aid in determining the source of a leak s is to put your thumb over each outlet at the SureFlow station one at a time The flowmeter ball must drop to zero on each test indicating no leak within the SureFlow station 6 Attach up to five 5 cannulas per flow station dep
11. p to five individuals per SureFlow unit Two or more SureFlow units can be connected together to serve even more individuals simultaneously Each flowmeter is adjusted separately to ensure precise control with a visual indication of flow for safety and comfort Dual flow models of NewLife Intensity 8 and 10 LPM concentrators provide greater flexibility in delivering oxygen by allowing SureFlow to operate connected to one of the concentrator s flowmeters while the secondary flowmeter can supply oxygen for a nebulizer treatment or a higher flow patient SureFlow units also meet the low flow oxygen needs of pediatric patients in a critical care nursery in developing countries around the globe The standard SureFlow is configured with 5 low flow flowmeters which display in 1 10 increments and can be used within a wide range of settings from 1 10 LPM to 1 0 LPM SureFlow can also be specially ordered with a combination of I LPM and 5 LPM flowmeters or with five 2 LPM flowmeters which display in 1 8 LPM increments 4 MN152 1 B Sure ze Components Flowmeter Station Inlet Outlet 5 Oxygen Outlet s 5 Oxygen Flowmeter s SureFlow User Manual 5 Sure low Procedures For Set Up Using One SureFlow Oxygen Flow Station The following step by step instructions apply to SureFlow flow stations used with oxygen concentrators although you may use any medical grade oxygen source that does not exceed 20 p
12. sig 138 kPa To supply oxygen to the SureFlow unit Note During set up nothing is initially connected to the oxygen concentrator at this time This is important to prevent the concentrator from being overdrawn when using with the SureFlow unit 1 Supply power to the oxygen concentrator and set the I O switch to the I position to turn it on 2 Set the flowmeter on the oxygen concentrator to the maximum capacity of the unit for example 5 LPM 8 LPM or 10 LPM depending on concentrator model 3 Connect the oxygen hosing supplied to the outlet of the oxygen concentrator as shown in Figure 1 Figure 1 6 MN152 1B 4 Connect the other end of the oxygen hosing to either inlet of the SureFlow unit See Figure 2 Wa AN Figure 2 s necessary test for leaks at both ends of the hose by applying soapy water to each of the connection points If bubbling occurs then leaks are present and the connection must be tightened properly until leaks are eliminated It may be more apparent if there is a leak s by allowing the concentrator to run with all flowmeters turned off at the flow station If no leak is present the flowmeter on the concentrator should read zero Another aid in determining the source of a leak s is to put your thumb over each outlet at the SureFlow station one at a time The flowmeter ball must drop to zero on each test indicating no leak within the SureFlow station