1. 43 6232 553 50 Fax 43 6232 55354 Spain INVACARE S A C Areny S N Poligon Industrial de Celra ES 1 7460 Celr Girona Phone 34 972 49 32 00 Fax 34 972 49 32 20 Norway amp Iceland INVACARE AS Grensesvingen 9 Postbox 6230 Etterstad NO 0603 Oslo Phone 47 22 57 95 00 Fax 47 225795 0l Belgium amp Luxemburg INVACARE N V Autobaan 22 BE 82 10 Loppem Br gge Phone 32 50 83 10 IO Fax 325083 I0 II Portugal INVACARE Lda Rua Estrada Velha 949 PT 4465 784 Leca do Balio Phone 351 225 10 59 46 47 Fax 351 225 1057 39 Australia INVACARE Australia Pty Ltd Lenton Place North Rocks AU NSW 215 Phone 61 2 8839 5333 Fax 61 2 8839 5353 Manufacturer INVACARE REA AB Box 200 SE 343 75 DIO www invacare com
2. Invacare Octave User Manual EN CE Table of Contents General Information 3 2 Operating the Octave cece 7 3 Receiving the Octave ee eee n 4 Disassembly ofparts 14 5 WIDE s o 99PPPRRRRSPRPTRRRSPPPPRRPRPPPPRRONSESSSSRNSEN SES d 15 6 Part numbers for accessories l6 7 Cleaning and Maintenance 17 8 Trouble shooting the electrical system 20 9 Electrical dad s s sca 5 capes oe as 5s 308 a Gs a eh Oe 9 21 10 Technical specifications ccc ccc cc ccc cece reece cece 25 iis VVeIPt sa5429999929999499999999992 9294229 9 9 545 27 12 DISDOSOLa vo Oo 09090919 9 ee ew aia 0 90 239 18 08 a a a a wie ca S 27 General information This manual must be given to the user of the product Before using this product read this manual and save for future reference Please note that there may be sections in this User Manual which are not relevant for your bed All indications of right and left are based on a patient lying on his back in the bed Metal side rails and U profile bed ends may be disassembled without the use of any tools l l Intended use e The bed is designed for use especially for home care e The bed is designed for patients over the age of 12 years 150 cm a
3. the two locking pins together Lower the side rail Britt V wooden side rails Line Alu conforms to the European standard EN60601 2 52 For those however with a small body size there will be a risk of entrapment 2 4 Adjusting the height of the lifting pole Loosen the cord as shown in fig A The lifting handle can now be adjusted to the desired height Press the cord together as shown in fig B and check that the cord is locked in the cord lock by pulling the handle as shown in figure C 2 5 Emergency release of the mattress support parts An emergency release of the mattress support is necessary in the case of e g a power or motor failure An emergency release of the height adjustment is NOT possible e Remove the plug from the mains before the emergency release of the mattress support e n an emergency case the mattress sections are released by pulling out the cotter pin from the motor in question A minimum of 2 people are required to release a mattress section Both people should hold the mattress section in locked position One person pulls out the cotter pin Both slowly lower the mattress section until it is completely down If the mattres support is removed the mattress will not foll
4. 3 Use ordinary household cleaning agents Never use acids alkalines or solvents such as acetone or cellulose thinner 4 The backrest and the leg thigh sections can be lifted from the mattress support which facilitates cleaning and access to the control unit 5 The hand control motors and control unit may be washed with brush and water Important details for cleaning IPX6 and IPX4 hand controls control unit and motors Water may be pressurised but never use high pressure cleaning or steam cleaning directly on the electrical parts 6 Dry the bed after cleaning e i Make sure that all electric plugs are in place during washing 7 2 Maintenance A service contract can be made in the countries where Invacare has its own sales company In certain countries Invacare offers courses in service and maintenance of the Octave Spare parts lists and additional user manuals are available from the Invacare homepage Before use Ensure that all manual and electrical parts functions correctly and are in a secure state Check by raising and lowering the bed that the shear arms run smoothly in the guideways After 3 months Ensure that all manual and electrical parts are functioning and tighten bolts screws nuts etc Every year We recommend a safety test comprising the motors performance and mechanical state e Motors hand control and control unit These parts are serviced by exchanging the faulty part Every s
5. 4 Dismounting mattress support from the base The mattress support can be disassembled from the base To disassemble the mattress support from the base relevant tools and 2 persons are required a Dismount all accessories incl bed end b Lower the bed into the lowest position c Dismount the height adjustment motor cable from the control unit d Raise the back section to the highest position e Loosen and remove the screws nuts between mattress support and shear arms f Lift the mattress support free from the guideway g Lower the back section to the lowest position h Remove the power plug from the bed and lift the mattress support off the shear arm 4 2 Dismounting the BRITT V Line Alu side rail a Raise the bed about one third b Place the side rail in the highest position c Dismount the end stop brackets 4 pcs d Lower the side rail see section Operation of the BRITT wooden Line Alu side rail until it is completely below the bed end rail 5 Wiring Check the cabling by operating the motors of the bed to their outer positions When all wires have been correctly installed there is no squeezing risk of the cables It is normal for cables to loosen slightly after short usage The cables are equipped with plugs at both sections hence they can be replaced independently except for the height motor Lu The cables are attached to the top frame with cable clips e The hand con
6. load 200 kg extension E1074 D 484640 Mattress max load 350 kg extension PIX IUX IS oua env bi Seen ee eure sane 1451260 Cover for side rail height extender padded covering both the normal side rail and the height extender piece s does sc dae a E aci o 7 EROR ORAE EUR CUR du ied i iii 3 1558095 Cover for Line and Britt V side rail padded piece 1 eee 1449588 Cover for Line and Britt V side rail extra padded piece ceea 1449589 Cover for Line extended side rail 10 cm padded piece cee ee ce eee eens 144959 Cover for Line extended side rail 20 cm padded piece cece cee eee eens 1449592 Cover for Line and Britt V side rail net piece 0 ceea 1449590 Cover for Line extended side rail 10 cm net piece cece eee eee eens 1449593 Cover for Line extended side rail 20 cm net piece cee ccc eee eee eens 1449594 Scala Basic 2 and Basic plus 2 3 4 length collapsible steel side rail 168 cm x 40 cm mounted on the side tube Release button in the head sec tion Scala Medium 2 3 4 length collapsible steel side rail 165 cm x 46 cm mounted on the side tube Release button in the head sec tion Scala Decubi 2 3 4 length collapsible steel side rail 168 cm x 54 cm mounted on the side tube Release button in the head sec tion Line Standard Full length side rail 206 cm x 40 cm in aluminum or wood finish moun
7. upper position There will be an audible click when it locks Mount the end stop brackets C UAWN 3 5 Mounting support handle accessories a Mount the support handles on the tube of either the back or leg sections Note Do not mount the support handle on the seat or thigh section b Tighten the finger screws Note The distance between 2 support handles or support handle and bed end must be more than 32 cm A N Placing support handles in wrong positions can cause serious damage or suffocation P DY SRI if body parts are trapped between support handles or between a support handle and bed end i ES 3 6 Mounting the lifting pole accessories a Push the lifting pole bottom piece in the fitting under the back rest b Mount the enclosed screws and nuts and tighten them 3 7 Mounting the mattress support extension 10 cm Note t is not possible to use the wooden side rail in combination with the mattress support extension Therefore dismount the wooden side rails first The mattress support extension can be mounted in either the head or foot end or in both ends a Loosen screws from the bed end b Remove the bed end c Remove the ferrules in the tube ends for the mattress support and mount the extension d Tighten up with the enclosed screws e Mount the bed end in the outer position 4 screws per fitting Visible insert pipe is about 14 5 cm 4 Disassembly of parts
8. Declaration of conformity Labels symbols on the product Ocd A Maximum user weight O L Safe working load SWL Patient mattress side rails lifting pole other equipment INSTRUCTION Reference to the user manual Applied Part Type B complying with the specified requirements e for protection against electrical shock according to IEC60601 An applied part is a part of the medical equipment which is designed to or are likely to come into physical contact with the patient The product should be reused where possible To identify the product please look at the product label Type no of the bed Electrical information Serial no INVACARE REA AB Vhxj v gen 303 8E 343 15 00 SWEDEN l P Tel 48 476 63500 Fax 45 476 D3099 C lin Current in Type nr OCTAVE C1AB1 M1 Octave ES m di zi lin max 3A Int Intermitens Uin 230 VOL AL 50 Hz z I e Mex Mem IPX6 i Max Maximum Oe 2013 01 DE ot min minutes Code ENQOSOI 2 52 j oe 360K L sig a kg kilogram E A Ampere Date of production IPX4 or IPX6 depend Symbols ing on configuration 4 Description of the main parts in the mattress support Leg section Back rest Thigh section Seat section H LT ac 1 5 Safety and warnings The bed fulfils all requirements regarding maximum distances However if the bed is used for the care of patients with a small body dimensions it must be es
9. SD d TO GoNia ceramic tile If floors are covered aRar Sar with synthetic material the relative IEC 61000 4 2 humidity should be at least 30 2 kV for power 2 kV for power supply lines supply lines Mains power quality should be that of a typical commercial or hospital 1 kV for input output 1 kV for input output environment lines lines 1 kV differential 1 kV differential mode mode Electrostatic transient burst IEC 61000 4 4 Surge Mains power quality should be that of a typical commercial or hospital IEC 61000 4 5 2 kV common mode 2 kV common mode BENIKORIHOHE lt 5 96 Ur 5 U4 gt 95 dip in U gt 95 dip in Uz for 0 5 cycle for 0 5 cycle Mains power quality should be that 40 Ur 40 Uz of a typical commercial or hospital 60 dip in Uz 60 dip in Uz environment If the user of the for 5 cycles for 5 cycles Octave Bed requires continued operation during power mains TO Ur TO Ur interruptions it is recommended that 30 dip in Uz 30 96 dip in U7 the Octave Bed be powered from an for 25 cycles for 25 cycles uninterruptible power supply or a battery Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 lt 5 Ur lt 5 Ur gt 95 dip in U gt 95 dip in U for 5 sec for 5 sec Power frequency 50 60 Hz magnetic field Power frequency magnetic fields should be at levels characteristic of a typical l
10. and the Octave Bed The Octave Bed is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Octave Bed can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Octave Bed as recommended below according to the maximum output power of the communications equipment Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz Rated maximum output of 3 5 a transmitter d V WP d E WP For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 24 10 Technical specifications i 3 i Yoo eu R E 178 A T All measurements are stated in cm All angles are stated in degrees All measurements and angles are stated without tolerances Invacare reserv
11. cling facility at its end of life 27 Customer Sales and Service Sweden amp Finland INVACARE AB Fagerstagatan 9 Box 66 SE 163 9 Spanga Phone 46 8 761 70 90 Fax 468 761 81 08 Ireland INVACARE Ireland Ltd Unit 5 Seatown Business Campus Seatown Road Swords IE County Dublin Phone 353 810 70 84 Fax 353 18107085 Denmark INVACARE A S Sdr Ringvej 37 DK 2605 Brandby Phone 45 36 90 00 00 Fa 45 36900001 Switzerland INVACARE AG Benkenstrasse 260 CH 4108 Witterswil Phone 41 61 487 70 80 Fax 41 61 487 70 81 Italy INVACARE MECC SAN s r l Via dei Pini 62 IT 36016 Thiene VI Phone 39 0445 38 00 59 Fax 39 0445 38 00 34 Invacare REA AB Ident no 1567233 Version 06 Date 02 2013 Yes you can Netherlands INVACARE B V Celsiusstraat 46 INL 6716 BZ Ede Phone 31 318 69 57 57 Fax 31 318 69 57 58 United Kingdom INVACARE Ltd Pencoed Technology Park Pencoed GB Bridgend CF35 5AQ Phone 44 656 77 62 00 Fax 441 656 77 62 01 France INVACARE Poirier SAS Route de St Roch FR 37230 Fondettes Phone 33 2 47 62 64 66 Fax 332 47 42 1224 Germany INVACARE GmbH Alemannenstrasse 0 DE 883 16 Isny Phone 49 75 627 000 Fax 49 75 627 00 66 New Zealand INVACARE NZ 4 Westfield Place Mt Wellington NZ Auckland Phone 64 9 917 39 39 Fax 649917 3957 Austria INVACARE AUSTRIA GmbH Herzog Odilostrasse 10 AT 5310 Mondsee Phone
12. econd year Service should be performed after the maintenance chart See chapter 7 4 Maintenance chart When reconditioning the bed service must be carried out according to the service schedule 7 3 Lubrication plan We recommend to lubricate the Invacare Octave according to the following instructions Guideways lubricate with grease 2 Swivel on shear arms lubricate with oil 3 Swivel on mattress support accessories lubricate with oil 4 Swivel on motor lubricate with oil Lubricate with medically clean oil e g KEW WO 50 order no 813239 and grease order no 1497607 7 4 Maintenance chart Service and maintenance of the Invacare Octave must only be performed by personnel who have received the necessary instruction or training S N located on the mattress support 9 IL T common RR E DE LETS CE LESE DIE LE mem 0 ESTE eee TI Mee LETT Ree ff eee ff DEE LC E E RR DE wee RI E IRI RR Lubrication according to the lubrication plan chapter 12 The wooden side rails gliding system must not be lubricated with oil otherwise the wooden bars will move sluggishly DE RR E E RI Measurements can be done at internal threads after removal of finger screws 7 5 Motors control unit and hand control These parts are not serviceable they must be replaced A safety test comprising control of the motors performance and mechanical state is recom
13. ed ends O U dH C n2 3 2 Mounting the Invacare Octave Receiving the bed on the trolley 2 persons required 9 Applyall 4 braked castors on the trolley b Apply all 4 braked castors on the bed c Remove the delivered equipment from the trolley d Lower the bed and the trolley into horizontal position standing on the bed castors e Dismount the trolley from the bed The connection to the socket outlet must always be accessible in case an electric error should For ergonomic working position Raise the bed hi low to approx 60 cm above floor level See section 2 Operation of the Invacare Octave Assembling the fittings a Push bed end through the pipes on the top frame Visible insert pipe is about 4 5 cm b Tighten the bed end with enclosed screws 2 Mounting the wooden bed end a Slide the wooden bed end into the U profile 3 3 Mounting wooden adapters and wooden side rail accessories Mounting wooden trim a Remove the exposure on the bed end b Mount the wooden trim with the enclosed mounting screws c Make sure that all 4 wooden trims have been correctly mounted tightened 3 4 Mounting the Britt wooden side rail Line Alu accessories Raise the bed about one third Mount one end ata time Lift the upper bar and guide the slide bracket to the bed end rail A Press the locking pin of the side rail with a finger B Guide the side rail to the
14. es the right to change the stated measurements and angles without preceding warning Max patient weight 350 kg provided the equipment bed ends side rails support handles lifting pole and mat tress have a total weight of max 35 kg Safe Working Load Max 385 kg Patient accessories 10 1 Environmental conditions Storage and Operation Transportation 10 C 50 C WC 4C Relative humidity 1076 80 RH Atmospheric 800 hPa 1100 hPa pressure 25 10 1 Mattress dimensions Safety aspects regarding combination of side rails and mattresses In order to obtain the highest possible safety level when using the bed in combination with side rails it is important to respect the minimum and maximum measures for mattresses stated in the table To prevent the possibility of the user crawling over the top of the side rail with the risk of severe injury due to e g falling a minimum vertical distance of 22 cm between the top of the mattress and the top of the side rail is to be respected The maximum mattress height in combination with a given side rail is stated in the table To prevent head entrapment between the inside surface of the side rail and the mattress with the risk of e g suffocation a safety maximum space has to be respected It is especially important to focus on this risk when using very thick or soft mattresses low density or a combination of both The minimum width and length of mattresse
15. mended once a year Motors control unit and hand control must be regularly cleaned from dust and dirt and they must be inspected for mechanical damage or breakage Inspect anchor points cables piston rod casing and plugs and check the correct functioning of the motors 9 8 Trouble shooting the electrical system Note Make sure that the mains are connected Possible cause Mains not connected 2 Fuse in the control unit is blown 3 Control unit defective Mains indicator does not light up Motor plug not pushed fully into the control unit 2 Motor is defective 3 Motor cable is damaged 4 Control unit defective Mains indicator lights up but the motor is not running The relays in the control unit makes a clicking noise Hand control defective 2 Control unit defective Mains indicator lights up but the motor is not running No relay sound is heard from the control unit Hand control defective 2 Control unit defective Control unit is in order except one direction on one channel Motor is running but the piston rod does not move The motor cannot lift full load Motor damaged Motor noise but no movement of piston rod Piston rod operates inwards but not outwards Relay sound No motor sound Motor does not run No relay sound Motor sound not move 20 Piston rod does Connect mains 2 Replace the control unit 3 Replace the control uni
16. ment power supply cord should be kept clear of the acces sories or any other moving part of the bed Invacare accepts no liability for any use change or assembly of the product other than as stated in this User Manual If the bed shows any signs of damage upon delivery please see the terms of de livery Accessories not mentioned in this manual must not be used When using side rails and the according safety cover it is essential to ensure correct fitting otherwise there is danger of entrapment suffocation between mattress support side rail and bed end Do not place any objects underneath bed If the functions of the bed change send the bed immediately for a check up according to the maintenance chart 1 6 Warranty The warranty covers all material and production defects for 2 years from the date of delivery provided it can be demonstrated that such defects were present before delivery All manufacturing faults or defects must be promptly reported Invacare may repair the fault or replace the component The warranty provided by Invacare does not cover additional costs transport packaging labour sundry expenses etc are for the customer s account The warranty does not cover e Damage caused during transport that is not directly reported to the forwarder at the moment of delivery e Repairs performed by unauthorized centers and personnel e Parts subject to normal wear e Malicious damage or damaged caused by imprope
17. nd longer e he bed is a nursing bed with modular structure e Max patient weight 350 kg provided the equipment bed ends side rails support handles lifting pole and mattress have a total weight of max 35 kg Safe Working Load Max 385 kg Patient accessories e The Ocatve bed is defined to be used according to Application environment 3 Long term care in a medical area where medical supervision is required and monitoring is provided if necessary and medical electrical equipment used in medical procedures may be pro vided to help maintain or improve the condition of the patient Application environment 4 Care provided in a domestic area where medical electrical equipment is used to alleviate or compensate for an injury disability or disease 278 SUD d In order to optimise the patient s comfort Invacare recommends using a 15 cm mattress see order numbers in section 6 Part numbers for accessories Produces meer ed Laer atei 1 2 Certification e Octave is marked in accordance with directive 93 42 EEC concerning Medical Devices e Octave is tested and approved in accordance with EN 60601 2 52 of TUV Product Service in Germany e Octave has undergone a risk analysis according to EN 14971 The motors and control unit of the bed are approved according to IEC 60601 1 2005 3 rd edition The hand control control unit and motors are protected according to IPX6 or IPX4 A lock cam must be used on the cont
18. ocation in a typical IEC 61000 4 8 commercial or hospital environment U is the a c mains voltage prior to application of the test level 22 portable and mobile RF communications equipment and the Octave Bed The Octave Bed is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Octave Bed can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Octave Bed as recommended below according to the maximum output power of the communications equipment Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz Rated maximum output of 3 5 3 5 7 transmitter E P P d VP For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 23 portable and mobile RF communications equipment
19. ow the beds movements when adjusting the head leg sections It is also possible that the mattress slips sideways which could cause the user to fall to the ground or get trapped in the bed 3 Receiving Invacare Octave Due to condensation the bed may not be used until it has reached room temperature 10 50 C Check whether the bed shows any signs of damage If the bed is damaged see terms of delivery After each assembly check that all fittings are properly tightened and that all parts have the correc function The bed s mattress support and base frame will be delivered assembl By ordering a standard bed the following parts will be delivered e Assembled bed on trolley e Fitting 2 pcs incl mounting screws 8 pcs e Wooden bed ends 2 pcs The following can be delivered an option see chapter 8 for order no e BRITT wooden side rails Line Alu a pair only for L 200 cm standard in the UK e Wooden bar 4 pcs incl mounting screws 8 pcs e Support handle e Lifting pole incl mounting screws 2 pcs e Mattress support extension 10 cm Can be used in head and bed end 3 1 Parts of a standard bed Base with central braking system Inner shear arm Outer shear arm Motor hi low Motor for back rest Motor thigh section with control unit 7 Motor leg section 8 Top frame incl seat section 9 Backrest 10 Thigh section Leg section 2 Fitting 13 Wooden b
20. pe cially noted that there is a risk of such a patient slipping through the openings between the side rails or through the opening between the side rail and the mattress support The bed must not be used by patients under 12 years of age or by patients with body size equiva lent to an average 2 year old or smaller The bed in combination with side rails must not be used for persons under 45 kg or under 150 cm s height or for persons that are restless spasms or confused unless a professional risk assess ment has taken place and been accepted or a correctly installed safety cover on the side rails are used The bed must be placed so that the height adjustment up down is not obstructed by for example lifts or furniture Electromagnetic interference between the bed and other electrical products can occur To reduce or eliminate such electromagnetic interference increase the distance between the bed and the products or switch them off This medical bed can be used together with medical electrical equipment connected to the heart intracardially or the blood vessels intravascularly provided that following are respected e The medical bed should be equipped with means for potential equalization connection See the symbol in chapter Electrical data The medical electrical equipment should not be fixed on the bed s metallic accessories such as side rails lifting pole drip rod bed ends ect In addition the medical electrical equip
21. r use of the bed I 7 Service life The expected service life of this product is seven years when used daily and in accordance with the safety instructions maintenance intervals and correct use stated in this manual The effective service life can vary according to frequency and intensity of use 2 Operating the Octave Disconnect the plug from the mains before moving the bed The cable must be kept clear of the floor and the castors during transport Make sure that the mains cable cannot be jammed or in any other way damaged when the bed is used Always leave the bed in the lowest position Otherwise there is a risk of entrapment due to accidental lowering of the mattress support A person can be seriously injured under the bed during height adjustment There is a risk of entrapment of fingers on assembly and operation of the metal side rail The bed s castors must be locked whilst nursing a patient in bed or when using any positioning function The bed must be locked when the patient moving in and out of the bed The bed s electrical equipment must not be dismantled or combined with other electrical equip ment 2 1 Operating the Invacare Octave with the hand control To operate the mattress support the bed is equipped with one of the following types of hand controls e HB 80 or e HL 80 with integrated ACP function key UP DOWN lt Sitting position lt amp Backrest lt lt Leg section lt lt Thigh
22. rol unit if not Invacare can not guarantee the IP protection All service work described in chapter 7 may only be performed by personnel who have been trained or instructed by Invacare service course If the functions of the bed change please send the bed for a check up according to the maintenance chart in chapter 7 Invacare is certified according to EN ISO 9001 and ISO 13485 and ensures compliance by 3 rd party Throughout the entire production process our materials products are quality controlled by the operators Additionally a final test is carried out prior to packaging and shipping The operator carrying out the final test which includes a check of all moveable parts motors and castors issues a QA number on the product thereby confirming its quality If the product does not correspond to the quality demands of Invacare it will be discarded If contrary to our expectations a problem should arise in connection with the delivered product please contact your Invacare supplier An address list is Shown on the back of this manual 1 3 Labels and symbols Symbols in this user manual WARNING Indicates a hazardous situation that could result in serious injury or death if it is not avoided INFORMATION bd Gives useful tips recommendations and information for efficient trouble free use This product complies with Directive 93 42 EEC concerning medical devices The launch date of this product is stated in the CE
23. s in combination with a given side rail is stated in the table EEE ETE ER Mattress Mattress Mattress Mattress Density Dacapo heavy user Max 350 kg 5 cm Visco elastic foam 50 kg m3 5 cm Cold foam 50 kg m3 5 cm Cold foam 50 kg m3 Dacapo heavy user Max 250 kg 5 cm Visco elastic foam 50 kg m3 5 cm Cold foam 38 kg m3 5 cm Cold foam 50 kg m3 26 ll Weight Mattress SUDDOM tic 6224 art m n e n dt ae de aaa tea 72 0 kg Base incl shear arm 00 eee eens 71 0 kg Bed end fitting 24 465 esr rro Roh esac ae 5 0 kg pcs Wooden bed end 0 cee eee eee 8 5 kg pcs Complete bed excluding accessories 170 0 kg Wooden trims 2 pcs 00 cee eee eee ee 2 0 kg Wooden side rail BRITT V 7 5 kg pcs Lifting BOlG 10m etr dee da rib ERR ees 10 0 kg Support handle 4 2 packed arina pops 2 0 kg Mattress appFOX cse die ronde d ad e rer 8 0 kg 12 Disposal This product has been supplied from an environmentally aware manufacturer that complies with the Waste Electrical and Electronic Equipment WEEE Directive 2002 96 CE This product may contain substances that could be harmful to the environment if disposed of in places land fills that are not appropriate according to legislation ENG The crossed out wheelie bin symbol is placed on this product to encourage you to recycle wherever possible Please be environmentally responsible and recycle this product through your recy
24. section lt Height adjustment of the bed Key for locking and unlocking the lt electrical functions Insert the key below each button to respectively lock and unlock the hand control functions There is a risk of entrapment whilst operating the bed When entering or exiting the bed the height adjustment should be used Furthermore the back rest can be used as support yellow button on the hand control However the thigh and leg sec tion must be horizontal otherwise there is risk of overloading of the mattress support 2 2 Using the central brake Braking the bed If the brake is in neutral step on the red pedal Releasing the brake Step on the green pedal until the brake is in the neutral position The bed s castors must be locked whilst nursing a patient in bed or when using any positioning function The bed must be locked when the patient moving in and out of the bed The bed must not be transported when occupied The castors might under special conditions be difficult to manoeuvre due to the floor covering being untreated or in poor condition In matters of doubt Invacare recommends the placement of a suitable type of protection between the castors and the floor 2 3 Operation of the BRITT V wooden side rail Line Alu UP Pull up the top wooden bar until the Locking pins locking pin locks with an audible click DOWN Lift the top wooden bar and press W c
25. ssification of the hand control decides the overall classification of the bed 2l 9 1 Guidance and manufacturer s declaration electromagnetic emission Guidance and manufacturer s declaration electromagnetic emission The Octave Bed is intended for use in the electromagnetic environment specified below The customer or the user of the Octave Bed should assure that it is used in such an environment Emissionstest Compliance Electromagnetic environment guidance NE ie Group 1 CISPR 11 RF emissions Class B The Octave Bed uses RF energy only for its internal function CISPR 11 Therefore its RF emissions are very low and are not likely to Harmonic emissions cause any interference in nearby electronic equipment The Class A Octave Bed is suitable for use in all establishments including domestic establishments and those directly connected to the IEC 61000 3 2 public low voltage power supply network that supplies Voltage fluctuations buildings used for domestic purposes flicker emissions Complies IEC 61000 3 3 Guidance and manufacturer s declaration electromagnetic immunity The Octave Bed is intended for use in the electromagnetic environment specified below The customer or the user of the Octave Bed should assure that it is used in such an environment IEC 60601 i Electromagnetic environment Immunity test Compliance level test level guidance Electrostatic Floors should be wood concrete or discharge E
26. t Insert the motor plug properly into the control unit 2 Replace the motor 3 Replace the motor 4 Replace the control unit Replace the hand control 2 Replace the control unit Replace the hand control 2 Replace the control unit Replace the motor Motor plug not connected Defective motor Defective control unit Defective hand control Defective motor 9 Electrical data Voltage supply 230 Voltage AC 50 Hz Max current input 3A Intermittent Duty cycle of motors Int 10 2 min 18 min Degree of protection IPX6 or IPX4 Sound level 48 dB A Double insulated class II E Alternating current NJ Applied Part Type B complying with the specified requirements 9 for protection against electrical shock according to IEC60601 1 An applied part is a part of the medical equipment which is designed to or are likely to come into physical contact with the patient Potential equalization by The control unit has no contact breaker hence disconnecting the mains plug is the only means of disconnecting the power supply The IPX6 classification means that the electrical system is protected against water projected in powerful jets from any direction The IPX4 classification means that the electrical system is protected against water splashed against the compo nent from any direction The hand controls have different IP classes see label on back the IP cla
27. ted on the bed end rail Two centered release buttons for folding of the gate Line Extendable Full length side rail with telescopic extension 206 226 cm x 40 cm in aluminum or wood finish mounted on the bed end rail Two centered release buttons for folding of the gate Britt V Full length side rail 205 cm x 40 cm in wood mounted on the bed end rail Two centered release buttons for folding of the gate Verso Il 3 4 length collapsible steel side rail 156 cm x 40 cm mounted on the side tube Release button in the foot sec tion 7 Cleaning and maintenance The base frame must be supported during service inspections to prevent accidental lowering e Only personnel who have received the necessary instructions or training may perform service and mainte nance of the bed e With normal daily use service must be carried out according to the service schedule after 2 years use and thereafter every second year e For beds with electrical equipment a safety test comprising control of the motors performance and mechanical state is recommended once a year e When reconditioning the bed service must be carried out according to the service schedule 7 Cleaning The bed does not tolerate cleaning in automatic washing plants or the use of water jet based cleaning equipment Please make sure the power plug is removed from the wall socket outlet before cleaning 2 The bed should be washed down using a wet sponge cloth or brush
28. trol control unit and motors are protected according to IPX4 or IPX6 It can be dangerous to roll over the mains cable Do not bring mains cable into moving parts Disconnect the plug from the mains before moving the bed The cables must be mounted in such a way that they are kept clear of the floor and do not block the castors A lock cam must be used on the control unit if not Invacare cannot guarantee the IP protection 6 Part numbers for accessories Article Dimensions Order no Line side rail beech Complete a dara dct bain ae dusa cae to ca ge RP e a 7 cd ati eee de 522786 0125 Line side rail extendable beech complete ccm 522787 0125 BRITT V wooden siderail 2 pcs 3 et aceea ati apo ardea acit st ac pic ded br c aaa e d dl 538458 XXXX Wooden trim for Line and Britt V side rail ceea 1544208 0101 Side rail height extender for Line extendable beech piece c cc 1538345 0125 Side rail height extender for Britt IV and V piece ccc eee ee eee 1538420 0125 SUpport handle 1 PSC P e oaia hea dc t d td lei cee BE a eee phe dand 446353 0152 Eder cepe 50 57700 MO Mattress support extension I0 cM seaca macara d ace at qb Rede d gea d i ddd al ei 446498 0 52 Mattress max load 200 kg c cec Bio 404g PP atare d a a 484429 Mattress max load 350 Kg on ata atata dle eee da awed cea L16x209 S aca teeny eee ddd at n na 1451259 Mattress max
Download Pdf Manuals
Related Search
Related Contents
Serenity - Canin Renesas Synergy Development Kit DK Tecumseh AWA2457ZXTXC Performance Data Sheet User Manual ASUS CM6870 T7010 User's Manual Médiathèques du Val Maubuée : mode d`emploi - Université Paris-Est User Manual Samsung VC240 Bruksanvisning EIA引出し 取扱説明書 LK−FU33 Rapport d`activité 2004 Copyright © All rights reserved.