Home
        TGA Series Trace Gas Analyzers
         Contents
1.             fg Temperature  MS PreAmp            E Calculations Reference   Sample      Concentration  Analog Input Temp  40 00    40 00  C      Other  TGA Temperatures PID control gain 0 03  Data Output  Serial Numbers PID control tau 4 00  About TGA  Themistor Rzero  4 95  Thermistor gain 4369 00  Themnistor offset                         FIGURE 7 15  Controlling the detector temperature settings    Most TGAs have detector temperatures that are controlled by TE coolers  The  Temp parameter on the Reference and Sample tabs is the setpoint  Set the  detector temperatures to optimize the signal levels  as discussed in Appendix  D 3  Detector Temperature  Some TGAs have LN cooled detectors for  measuring absorption at longer wavelengths  The temperature of these  detectors is not measured or controlled  For LN   cooled detectors  uncheck the  Control detector temperatures box     The rest of the detector parameters should be set to the defaults shown in  FIGURE 7 15  The parameters for the sample detector are not shown  but are  the same as for the reference detector     PreAmp   The Settings  gt  Detectors  gt  PreAmp window and parameters are the same for  both versions of the software and are shown in FIGURE 7 16  See Appendix  D  Optimizing Detector Parameters  for details on optimizing the detector  preamp parameters     7 2 3 3 Calculations    TGA Series Trace Gas Analyzers       rameter Settings      Laser     Temperature     Current  t Line Lock      Other  V  Auto gain
2.          sssssseeeeeeenen nene 31   6 1 1 Plumbing Connections 0 0 0    ccceecesecesecseeereeeseeeeeeeeeeeeeeeeenseensees 33  6 1 2 Data Output Connections 20 0 0    cee ceccesececeereeeseeeeeeeseeeeeeeeenteeneees 35  NMAMMd uS                          M M   37  6 2 TGA Software Installation                   sese 39  6 2 1 Installation of PC Software              sssssssssseeeeee 39  6 2 2 Updating TGA Operating System                   sss 40  6 2 3 Configure Ethernet Connection                 sssssssseseeeee 42  6 2 4 Set TGA Serial Number and Identification String                       45  6 2 5  R  n the PC Software 4  neue eere ir eden 46  6 3 Detailed Setup Instructions              ssssssssseseeeeeeeneenenee 46    EG runt 46   7 1 Routine Operatii ea a E ener 46  Telt Startup Procedure enaa aa ce Gagne negem 46  7 1 2 Routine System Checks                  sse 47  7 1 3 Shutdown Procedure                     sse 48   7 2 Software User Interface            sssssssssssseeeeeeeeeeeeeneneneennn 49  7 2 1     Connect Window    esee ee eed ner en 49  1 2 2   Stat  s Window e eerie teens redet 51  17 2 3     Settings Window    esee ee eee eren 52   4 2 3  1    LASS iss  ecc de ete tete e Ore RH err dus 53  P2352 Detectors  eR er eH ere etes etre tns 59  7 2 3 3   Calculations    eene eee ree 61  TQ BA    Oth  r  iiie eret eee tis 66  7 2 4  Laser Window  eee reden 70  7 2 4 1 View Less View More              sse 71  7 2 4 2  Tabbed Expand                    sse 72  7 24 3  
3.          the N2O CHA laser is not available for the TGA200A    the ammonia laser is not available for the TGA200A    The concentration of the reference gas 1s used to calculate the concentration of    the sample gas  therefore  it must be entered into the TGA software  The  calculated sample concentration 1s scaled by this value  If it is not correct  the    measured concentration will have a corresponding scale error  For many trace    gas flux measurement applications  a measurement of the reference gas  accurate to 2  is adequate     A 5       Appendix A  Configuring TGAs for Specific Gas Species    NOTE    A 3 Detectors    For applications that require a more accurate concentration  measurement the TGA may be calibrated using two well known  calibration tanks as discussed in Appendix E  Calibration  For  these applications the reference gas concentration provides only a  preliminary estimate that is superseded by the measurements of  the calibration gases  This makes a highly accurate measurement  of the reference gas concentration unnecessary     The user must provide an appropriate regulator for the cylinder of reference  gas  The delivery pressure is normally set to approximately 0 psig  and the  flow is typically set to 10 ml min  so the regulator should be chosen for good  performance at low delivery pressure and flow     The reference gas flow is generally set by a needle valve and monitored with a  flow meter  A reference gas connection assembly  pn 15837  is avail
4.     Status   Settings   Laser   Find   Graph   Data   Files          FIGURE 7 4  Toolbar before  top  and after  bottom  connection    7 2 2 Status Window    The TGA Status window displays the TGA laser and detector temperatures   sample cell pressure  and line lock status  The toolbar will show one of two  icons for the TGA status  depending on whether or not an error is detected   The lower panel of FIGURE 7 5 shows the Status icon of the TGA toolbar  when an error has been detected     51    TGA Series Trace Gas Analyzers       ORUA      Status   Settings   Laser   Find   Graph      Data   Files    Connection X     A Settings n Find   Ies E1 kd    FIGURE 7 5  TGA Status with a detected error  bottom        Any detected errors will be highlighted in red  with error messages displayed at  the bottom as shown in the right panel of FIGURE 7 6     Laser Temp  Smp Det Temp   Smp Det Temp    Ref Det Temp   Ref Det Temp  TGA Pressure TGA Pressure  Lines Locked Lines Locked          FIGURE 7 6  TGA Status window without error  left  and with error   right  and line lock manually disabled     NOTE The TGA Windows is used for LN2 cooled lasers  and reports the  laser temperature in K  The 7GA TEC software is used for TE   cooled lasers  and reports the laser temperature in   C     7 2 3 Settings Window    The Settings window is used to display and change parameters associated with  the laser  detectors  calculations  and other miscellaneous information     In the Settings window  the
5.     The TGA200A Input Module  pn 7730  is shown in FIGURE H 2  The blue  label identifies this module as being updated for use with TE cooled lasers        FIGURE H 2  TGA input module    The TGA200A Output Module  pn 7726  is shown in FIGURE H 3  The blue  label identifies this module as being updated for use with TE cooled lasers     eee a a U      MEYN       FIGURE H 3  TGA output module    CAUTION Do not use unmodified input or output modules  without the  blue label  for TE cooled lasers  Do not use modified input  or output modules  with blue label  for cryogenic lasers  A  mismatch between electronics and laser will damage the  laser     The input and output modules shipped with a TGA200 may be returned to the  factory to be updated  The modules will be reconfigured for TE cooled laser  operation  tested  and identified by affixing a blue sticker     NOTE If the input and output modules are upgraded to use with TE   cooled lasers  they may not be reconfigured for cryogenic lasers   Users that may wish to go back to use a cryogenic laser at a later  date should purchase new input and output modules for TE cooled  laser operation     The input and output modules mounted in the electronics are shown in  FIGURE H 4     Appendix H  Upgrading Early Generation TGAs to TE cooled Laser       FIGURE H 4  Modules mounted into TGA200A electronics    The TGA TEC software  pn 30723  is an updated version of the TGA Windows  software which has been modified for use with TE cooled las
6.    H 1 3 Power Module    Optionally  the power supplies of a TGA200 may be upgraded to the pn 30981  TGA200A power module  see Section 4 1 1 1  TGA Power Module  for detail  on the TGA200A power module   This upgrade is not required to use the TE   cooled laser  but it is available in the event that the TGA200   s power supply  fails  This upgrade requires a special connector assembly to be mounted in the  TGA200  Contact Campbell Scientific for details     H 1 4 Purge Boot    H 2 TGA100A    The TGA200 was supplied with a purge boot between the laser dewar and the  optical assembly  The purge boot mounted to the front lens holder and pushed  up against the laser dewar to enclose the air gap between the laser dewar and  the lens  The TGA200 purge boot can be pushed up against the TE cooled  laser assembly in the same way as for the laser dewar  If the purge boot  becomes damaged it can be replaced with pn 21573     H 2 1 Basic Upgrade    Upgrading a TGA100A is similar to upgrading a TGA200  It requires the same  laser assembly  input and output modules  and software  See Section 6 2 2   Upgrading TGA Operating System  for details  In addition to this basic  upgrade  a TGA100A may require the following upgrades     Appendix H  Upgrading Early Generation TGAs to TE cooled Laser    H 2 2 Detectors    If the TGA100A was supplied with LN2 cooled detectors  these should be  replaced with TE cooled detector assemblies  Contact Campbell Scientific for  availability  Most TGA100As were
7.   Dimensions and Part Numbers for Swagelok Inserts   Tubing OD  in  Tubing ID  in  Swagelok  pn CSI pn  1 4 1 8 B 405 2 15834  1 4 0 17 B 405 170 15830  1 4 3 16 B 405 3 15713  3 8 1 4 B 605 4 9845  1 2 3 8 B 815 6 17380  5 8 1 2 B 1015 8 19495   Ferrules    Each Swagelok   fitting comes assembled with the front and back ferrules  included  These ferrules are permanently swaged onto the tubing at the first  assembly  so spare ferrules may be needed for replacing the ends of tubing      gt           Back ferrule    ON    M    w    Front ferrule    FIGURE G 2  Front and back Swagelok   ferrules                            TABLE G 3  Dimensions and Part Numbers for Swagelok   Ferrules  Tubing OD  in  Swagelok   pn  front back  CSI pn  1 set   1 8 B 203 1 B 204 1 N A  1 4 B 403 1 B 404 1 15890  3 8 B 603 1 B 604 1 15889  1 2 B 813 1 B 814 1 N A  5 8 B 1013 1 B 1014 1 N A             G 3    Appendix G  Using Swagelok   Fittings    Plugs   Swagelok   plugs are used to plug a fitting when its tube is disconnected  It is  strongly recommended to plug all fittings to keep them clean  Spare plugs may  be needed if they become lost or damaged        FIGURE G 3  Swagelok   plug                      TABLE G 4  Dimensions and Part Numbers for  Swagelok   Plugs   Tubing OD  in  Swagelok   pn CSI pn  1 8 B 200 P 26803  1 4 B 400 P 15891  3 8 B 600 P 13712  1 2 B 810 P 17381  5 8 B 1010 P N A                   Caps   Swagelok   caps are used to cap the end of tubes when they are disconnected  
8.   Example of using Laser Line Find function to determine  laser threshold current    Appendix C  Optimizing Laser Parameters    C 3 High Current    Laser  Current    Laser  Temperature    Wavenumber   temperature     Wavenumber   current     Wavenumber   combined     Detector  Response    High current    too low         Equilibrates slowly    The laser cools slightly at the start of the spectral scan when it is turned off by  reducing its current to the zero current value  as discussed in Section 4 2 5   Laser Scan Sequence  If the actual spectral scan started immediately thereafter   the laser temperature would rise during the entire spectral scan  The rise in  temperature would be more rapid at first  but slow near the end of the scan as  the temperature approached equilibrium  The change in temperature would  change the laser s emission frequency  adding an undesired spectral  modulation  as illustrated in the far left panel of FIGURE C 6  To minimize  this problem  the laser current is increased above the DC current by an amount  specified in the High current offset parameter  The duration of this high  current pulse is determined by the laser High current count parameter  When  these parameters are properly set  the heat from the increased current  compensates for the heat lost when the current is reduced  stabilizing the laser  temperature more quickly     High current  too high    High current  correct    Overshoots    onstant during ramp       Positive slope   Nearly 
9.   Modulation Current  and  initiate the line locking function     For multiple ramp operation  each of the ramps must be locked onto the proper  absorption line     If the expected absorption line is not found easily  use the laser Find tool   Section 7 2 5  Find      A 5 Air Gap Purge    For isotope ratio applications  the air gap between the laser and lens  and the  short sample cell  should be purged as shown in FIGURE A 1  TGA100 or  TGA100A  or FIGURE A 2  TGA200 or TGA200A   This is not required for  most trace gas applications where the ambient concentration is very low  and  there is very little absorption  The sample cell is at low pressure  making the  sample absorption very narrow compared to the pressure broadened ambient  absorption  Thus  the concentration measurement is relatively insensitive to  trace gases in the ambient pressure air gap     This is not the case  however  for CO   or water isotope measurements where  the ambient concentration is relatively high and can change rapidly  for  example  if the cover of the TGA is off and a person in the vicinity exhales    Because these types of applications require extremely high accuracy  the air  gap should be purged to prevent absorption  A tank of compressed nitrogen  should be connected to the purge inlet  A flow of approximately 10 ml min is  recommended  The regulator on the user s tank may be connected to the purge  using the flow meter  needle valve  and tubing included in the reference gas  connection  
10.   New Values written to FLASH  BMPSvr  port 3000   listening    Initialized TGA100A TGA200 TGA200A                                                 Pause  StertExport      SendFile                  FIGURE 6 20  Device configuration utility Terminal tab    6 2 4 Set TGA Serial Number and Identification String    The TGA serial number and an identification string are stored in nonvolatile  memory  The serial number can be displayed and edited on the Settings tab of  the Device Configuration Utility  Set the serial number to match the serial  number of your TGA  The serial number for a TGA100 or TGA100A is found  on the laser mounting plate  inside the analyzer enclosure  The serial number  for a TGA200 or TGA200A is found next to the plumbing connections on the  outside of the analyzer enclosure  This setting allows the TGA Windows or  TGA TEC software to verify it is connected to the proper TGA before sending    parameters  see Section 7 2 1  Connect Window      45    TGA Series Trace Gas Analyzers    46    The identification string is also displayed and edited on the Settings tab of the  Device Configuration Utility  The identification string is intended as a device  nickname and to be a more user friendly way than the serial number by which  to identify which TGA is connected to the PC software  The identification  string is displayed at the top of the toolbar when the TGA software is  connected to a TGA  The identification string can be any text  up to 19  characters     6 2
11.   TGA TOLAD OT eT oshisi eod e Re ret bere Re DER redes eds 22  TGA heated intake filter tete overhead 22  TGA filter element replacement                    sssssssssseeeeee 23  47 mm replacement filters                    essent 23  TGA100 and TGA100A optical configuration                       ess 24  TGA200 and TGA200A optical configuration                        esse 24  TGA laser scan sequence              ssssssssssesseeeeneeenenen enne 26  Basic components required for TGA100 operation                            30  Basic components required for TGA200 and TGA200A operation   31  TGA200A enclosure                sees eene 31  TGA100 and TGA100A transport locks                        sees 32  TGA200 and TGA200A shipping clamps                   seesss 33  Feedthrough cover of TGA200A             ssssssssseseseeeeeeee 33  Plumbing connections located under feedthrough cover of   TCEA200 Aic ie eve e ete ebat qe e POTE e ER ca 34  SDM connections of TGA               ssessssseeeeeeeeren eene 35  SDM cable connector on TGA CPU board                          esses 36  SDM cable tied to electronics box                    eee 36  DC power cable connected to TGA200A and secured on   electronics boX   1 aieo n p e tH e te EI Pes 37  Routing of SDM and power cable through TGA200A feedthrough   bracket  nente erem ree te eid ds 38  TGA200A power module with cables installed                                   39  PC shortcut icons for TGA Windows  left  and TGA TEC  right     40  S
12.   The TGA100 and TGA100A have an optical assembly that includes a long  sample cell  a short reference cell  and a short sample cell  The beamsplitter of  the early TGAs was located at the back of the instrument  near the detectors     TheTGA200 and TGA200A optical configuration has the beamsplitter at the  front  near the laser   with similar  long  sample and reference cells  The  optical alignment hardware and shipping clamps are also different     Temperature Control   Early TGA100s had fans inside the enclosure to ensure all parts of the optical  system  in particular  the sample cell and the reference cell  remained at the  same temperature  but had no heater or temperature controller  An optional  accessory  the TGAHEAT  was introduced in 2002  TGAHEAT controlled  power to a pair of heaters attached to fans which maintained the temperature of  each end of the optical bench  For TGA100As  the TGAHEAT was included  as a standard component     Temperature control was improved in the TGA200 and TGA200A by  controlling the temperature of the air at each end of the enclosure instead of the  optical bench and including temperature control in the TGA Windows  software  This allows the temperature setpoint to be changed through software  rather than opening the TGA enclosure to turn a potentiometer  It also allows  the user to monitor and record the fraction of power used for the heaters     TGA100s with the TGAHEAT option and TGA100As that are running the  TGA Windows or TGA T
13.   by telephoning  435  227 9000  USA   You are responsible for conformance with governing codes and regulations  including safety  regulations  and the integrity and location of structures or land to which towers  tripods  and any attachments are attached  Installation  sites should be evaluated and approved by a qualified engineer  If questions or concerns arise regarding installation  use  or  maintenance of tripods  towers  attachments  or electrical connections  consult with a licensed and qualified engineer or electrician     General   e Prior to performing site or installation work  obtain required approvals and permits  Comply  with all governing structure height regulations  such as those of the FAA in the USA    e Use only qualified personnel for installation  use  and maintenance of tripods and towers  and  any attachments to tripods and towers  The use of licensed and qualified contractors is highly  recommended    e Read all applicable instructions carefully and understand procedures thoroughly before  beginning work    e Wear a hardhat and eye protection  and take other appropriate safety precautions while  working on or around tripods and towers    e Do not climb tripods or towers at any time  and prohibit climbing by other persons  Take  reasonable precautions to secure tripod and tower sites from trespassers    e Use only manufacturer recommended parts  materials  and tools     Utility and Electrical   e You can be killed or sustain serious bodily injury if th
14.   concentration should equal the reference gas concentration  set to  1000 above   Sample detector nonlinearity will cause the measured  concentration to be underestimated  If the measured concentration is  too low  increase the value of the sample detector linearity coefficient  until the measured concentration is 1000  A typical value is 0 3     Repeat the previous step for each ramp  if in multiple ramp mode   Restore the plumbing to its normal configuration and set the    reference gas concentration and cell length parameters back to their  proper values     Appendix E  Calibration    The predominant sources of error in the TGA   s concentration measurement are  the offset error caused by Fabry Perot interference  and gain errors caused by  errors in reference gas analysis or by different pressure or temperature in the  reference and sample cells  For eddy covariance or gradient flux applications   the offset error cancels out and only the gain errors are significant     For measurements of absolute concentrations  the offset errors are also  significant  Therefore  the appropriate calibration procedure depends on the  application  All applications will benefit from the basic span calibration  described in the next paragraph  It should be performed after the TGA has  been set up as discussed in Section 6  Installation     The TGA calibration may be checked by switching the sample inlet between  two calibration tanks  Normally  one tank should have near ambient  concentrat
15.   that contains the laser mounted on a thermoelectric cooler equipped with a  thermistor to measure its temperature     The standard detectors used in the TGA are thermoelectrically  TE  cooled and  operate at wavelengths up to 5 microns  These detectors are used for most  gases of interest  including nitrous oxide  N20   methane  CH4   and carbon  dioxide  CO2   Some gases  such as ammonia  NH3   have the strongest  absorption lines at longer wavelengths  and require the optional long  wavelength  liquid nitrogen cooled detectors  These detectors can operate to  wavelengths beyond 10 microns  The TGA200A uses only ICLs  which cannot    25    TGA Series Trace Gas Analyzers    26    reach these long wavelengths  Therefore  the TGA200A always uses TE   cooled detectors     4 2 5 Laser Scan Sequence    The laser is operated using a scan sequence that includes three phases  the zero  phase  the high current phase  and the modulation phase  This is illustrated in  FIGURE 4 27  The modulation phase performs the actual spectral scan   During this phase  the laser current is increased linearly over a small range   typically   0 1 to 1 0 mA   The laser s emission wavenumber depends on its  current  Therefore  the laser   s emission is scanned over a small range of  frequencies  typically   0 03 to 0 06 cm             During the zero phase  the laser current is set to a value below the laser s  emission threshold   Zero  signifies the laser emits no optical power  it does  not mean the
16.  192 168  5   8 Port  3000    Parameter Synchronization       Receive Parameters from TGA      Send Parameters to TGA    TGA Connection  Update Interval  in ms   100    Backup Parameter Files  Synchronized from TGA            Load from Backup file        Save to Backup File                  FIGURE 7 2  Connect window of TGA software interface    To connect to a TGA  enter the TGA s IP address  To set the TGA s IP  address  use the Device Configuration software  as described in Section 6 2 3   Configure Ethernet Connection  Select Port  3000  which is always used for  TGA PC communication     Choose the Parameter Synchronization approach  Normally  this should be  set to Receive Parameters from TGA  This allows connecting to the TGA  without disrupting its current settings  The Send Parameters to TGA option  can be used to restore parameters that had previously been backed up  after  loading the backup file      If sending parameters to the TGA  the software will first check to make sure  the TGA serial number in the parameter file matches the serial number stored  in the TGA   s nonvolatile memory  If the serial numbers do not match  the  software will not send the parameters to the TGA  and the message shown in  FIGURE 7 3 will be displayed  If this happens  verify the parameters are  correct for the TGA  and edit the serial number in the parameter file  Section  7 2 3 4  Settings  gt  Other  or in the TGA  Section 6 2 4  Set TGA Serial Number  and Identification String  as 
17.  5 Run the PC Software    Start the TGA Windows  for LN2 cooled laser  or TGA TEC  for TE cooled  laser  software  which will bring up the toolbar  Descriptions of the toolbar  functions are given in greater detail in Section 7 2  Software User Interface     6 3 Detailed Setup Instructions    Operation    7 1    When the TGA is first installed  or if it is reconfigured  for example  with a  new laser  the operational parameters must be set for optimal performance   Appendices A through E give detailed instructions to configure the TGA for a  specific gas species  performing the optical alignment  optimizing the laser  parameters  optimizing the detector parameters  and calibration  If the TGA  has already been configured  see Section 7 1 1  Startup Procedure  for routine  startup instructions     Routine Operation    Once the TGA has been set up  it should be checked periodically to verify  proper operation and to perform routine maintenance  Consistently recording  operating parameters such as pressure and detector signals is strongly  encouraged  Significant changes in these values from one operational period to  the next  is often indicative of problems with the system     Startup Procedure    NOTE    NOTE    1  This section describes the routine procedure for starting the TGA  If  the TGA is equipped with a LN  cooled laser  cool the laser dewar     If the TGA is equipped with a TE cooled laser  it will be cooled  automatically     2  If the TGA is equipped with LN2 coole
18.  8 80 Laser                4 1 2 2 AC Mains Power Cord    Depending on the geographic location the system is to be used  detachable AC  mains power cords are available from Campbell Scientific to accommodate  local electrical requirements  For reference  the part numbers and associated  geographic regions are summarized in TABLE 4 2     11    TGA Series Trace Gas Analyzers    12                         TABLE 4 2  Available AC Mains Power Cords by Region  Part Number Geographic Compatibility  13999 North America  18652 Continental Europe  18653 United Kingdom and Ireland  18672 Australia and New Zealand  19295 China                4 1 2 3 Power Module Mounting Brackets    Various mounting brackets are available such that the TGA power module can  be mounted at a location that is compatible to the measurement site and  convenient for the user  The mounting brackets available are summarized in  TABLE 4 3        TABLE 4 3  Power Module Mounting Brackets                               Part Number Mounting Location  19002 Tripod Mast Mounting  18955 CMIXX Leg Mounting  19017 Tower Mounting  27390 Pole Mounting  10     25 cm  4 10 in  diameter pole  18520 Pole Mounting  8 cm  3 in  diameter pole       4 1 3 Common Accessories  4 1 3 1 TGA Reference Gas Connection    A TGA reference gas connection  pn 15837  is available for connecting a  reference gas source to the TGA  The assembly includes a flow meter  needle  valve  6 2 m  20 ft  tubing  and 1 4 in Swagelok  fittings at either en
19.  8 Ib   Power  140 W  Pumping speed  1 slpm at 50 mb  60Hz     Ambient temperature range for operation  10 to 40   C    Ultimate vacuum   lt 2 mb    For detailed installation and operating instructions  refer to the user manual  supplied with the pump     4 1 4 5 Bypass Vacuum Pump    Some TGA applications  such as atmospheric profiles and chamber  measurements  use a sampling system to switch between multiple inlets  These  sampling systems require a bypass pump to maintain air flow through the non   selected inlets  The actual flow rate and pressure required will depend on the  application  Two bypass vacuum pump options are available from Campbell  Scientific  The DOAV502 has a capacity of 50 L min   and is adequate for  most low flow applications  The DAAVOS L has a capacity of 100 L min    and is used for higher flow applications  The pumps are supplied with the  tubing and fittings needed to connect to the sampling system  A brief overview  of each of the pumps is given in the following descriptions     DOAVS502 Vacuum Pump   The DOAVS502  shown in FIGURE 4 15  is a single head diaphragm pump that  is often used with a trace gas analyzer system in low flow applications  such as  atmospheric profile or chamber measurements  The DOAV502 comes with 30  m  100 ft  of plastic tubing with an outer diameter of 1 27 cm  0 5 in  and an  inner diameter of 0 95 cm  0 375 1n         FIGURE 4 15  DOAV502 vacuum pump    DOAV 502 specifications     Length  22 7 cm  8 9 in   Width  16 
20.  9 pin male serial data cable  pn 20730  is  included  It is shown in FIGURE 4 7  The serial cable is used to connect the  RS 232 port on the TGA to an RS 232 port on a user supplied PC        f        lt  7 ume    FIGURE 4 7  Serial data cable    4 1 1 9 TGA TEC Support Software and Operating System    Included is a CD  pn 30723  that contains installation files for TGA TEC  Support Software and current TGA firmware used to support the TGA200A or  prior TGA models that have been upgraded to use a thermoelectrically cooled   TE cooled  laser  TGA TEC is a user interface software package that allows  users to connect to the analyzer  set settings  and monitor real time analyzer  performance  The TGA firmware includes the OS that must be uploaded to the  analyzer s CPU board before operating a TEC laser  The software 1s  compatible with Windows XP operating systems or newer     10    TGA Series Trace Gas Analyzers      TGA Windows   Connected to sn 22205     eek    ee Find Gah Dua Fhe          j rl  A IT Pi    n Wi N A EA      Xa NANI  VA OTN  ML y    Sta  WOWECXJ6O05H8    XS Hiro      C Inbo          4 1 2 Optional Components  4 1 2 1 Laser    Depending on the gas to be measured  the TGA200A can be configured with  any one of the five lasers described in TABLE 4 1                          TABLE 4 1  Part Numbers for Available  Gas Species Lasers  Part Number Description  30478 N20 Laser  30477 CH  Laser  31121 N20 and CO   Laser  31119 CO   and 6P C Laser  30877 CO    52C  and
21.  B 7  B 8  and B 9     To use the alignment tool  place it against the tip tilt screws  Back  the screws out as needed to allow the gage to fit against the screw  threads  Then screw the screws in until the heads of the screws just  touch the gage  Back the screws out just enough to remove the tool   The tip tilt mirrors will now be at their nominal positions     FIGURE B 7 shows the positioning of the alignment tool for the  screws for the mirror in the beamsplitter block  The view is from the  user s perspective  FIGURE B 8 shows the same scenario but from a  different perspective  The figure illustrates the way that the  alignment tool fits against the screws  FIGURE B 9 illustrates how  the alignment tool fits against the detector tip tilt screws  Refer back  to FIGURE B 5 and FIGURE B 6 for the positioning of the mirror  tip tilt screws     B 10    Appendix B  Optical Alignment       FIGURE B 7  Use of alignment tool for aligning mirror in TGA  beamsplitter block       FIGURE B 8  Use of alignment tool for aligning mirror in TGA  beamsplitter block  alternate angle    Appendix B  Optical Alignment       FIGURE B 9  Use of alignment tool to position tip tilt screws for aligning  detector side mirrors    7  Adjust the horizontal and vertical alignment knobs to look for the  sample detector signal  and then proceed to the next section  If no  response is observed  contact Campbell Scientific for assistance     B 2 3 Horizontal and Vertical Alignment    Once the system is al
22.  Det Signal CO2  Laser DC Current CO2  Smp Transmittance CO2  Ref Transmittance CO2  Ramp B  Ramp C  Isotope Ratios  Miscellaneous  Laser Temperature  Laser Heater  Pressure  TGA analog input  TGA  Temp 1  TGA Temp 2  Duty Cycle TGA Heater 1  Duty Cycle TGA Heater 2       Detectors       OO EH        Cancel            FIGURE 7 40  Options for graphical display of data in TGA Windows    TGA TEC   The steps to select data to be graphed are slightly different for TGA TEC   Click on a parameter  edit the options if necessary  and then press Add as  shown in FIGURE 7 41  Repeat these steps to add more parameters to the  graph  An example of the graphical output of the TGA TEC is shown in  FIGURE 7 42     85    TGA Series Trace Gas Analyzers          g5    Line Width 1 E     Num Decimal Points 3     E  Do Not Plot in Graph                               FIGURE 7 41  Adding parameters to a graph in TGA TEC          Ia uw OS ZC vtr OFS                Data Values U     N20 Conc StdDev 0 0011 ppm  N20 Conc 0 281 ppm  il  Graph Width    Auto 1m23s     Manual             8 58 52 8 59 02 8 59 12 8 59 22 8 59 32 8 59 42                FIGURE 7 42  Example graph showing N2O concentration and  standard deviation    Clicking Options will bring up a window to add  remove  or change the way a  trace is shown  or to change the axes or the other visual aspects of the graph   The Add button can also be used to add traces     86    7 2 7 Data    TGA Series Trace Gas Analyzers    Keeping the Graph Wi
23.  Ifthe laser is to be operated again in the near future  it is  recommended to keep the laser cold to avoid temperature cycling the  laser     e Ifthe TGA will not be used for an extended period  allow the laser to  warm up and evacuate the dewar     7 2 Software User Interface    When the TGA program is started as described in Section 6 2 5  Run the PC  Software  the toolbar is displayed as shown in FIGURE 7 1  The functions are  listed below and described in greater detail in the following sections            an         Graph      ki    Files             Status   Settings    m vy    Find    Data                      x   Laser         FIGURE 7 1  TGA tool bar functions    Connect  Connect or disconnect from TGA  set the data interval  and save  and load parameter files     Status  Display TGA status   Settings  Display and change parameters   Laser  Display reference and sample signals and modify laser parameters  Find  Find the absorption line s    Graph  Set up and display real time graphs of TGA measurements   Data  View TGA measurements in numerical form    Files  Collect TGA data to the PC hard disk    7 2 1 Connect Window    The Connect window is used to connect or disconnect from the TGA  set the IP  address of the TGA to communicate with  set the interval at which the software  updates data from the TGA  and to save and load parameter files  The Connect  window is shown in FIGURE 7 2     49    TGA Series Trace Gas Analyzers    50       Connection mx       Connect To 
24.  Laser Temperature Col 5  C02 Conc StdDey      Selection Col 6  Laser Heater Col 6  13C Conc  Conc Col 7  Laser DC Current CO2 Col 7  Mean 13C Conc  Detectors Col 8  Pressure Col 8  13C Conc StdDev  RAE Col 9  TGA Temp 1 Col 9  180 Conc  5 Col 10  TGA Temp 2 Col 10  Mean 180 Conc  Misc Col 11  TGA Status Flags Col 11  180 Conc StdDev     Advanced Col 12  Del13C  File Names Col 13  Mean Del13C  Sai sais bao     Other Col 16  Mean Del180  Event Log Col 17  Del180 StdDev    Col 18  Ref Det Temp   Col 19  Ref Det Cooler   Col 20  Ref Det Gain   Col 21  Ref Det Offset   Col 22  Ref Det Signal CO2  Col 23  Ref Transmittance CO2  Col 24  Ref Det Signal 13C  Col 25  Ref Transmittance 13C  Col 26  Ref Det Sianal 180        z                           FIGURE 7 44  Controlling PC recorded data options in the TGA    The Settings column in the Data Files window is used to select what is  displayed on the remainder of the window   Files     display the names of the values set up to be collected     File Lists     Show the values in both the Data File and Housekeep Data  File  Set ASCII or Binary output     Data File     List just the data file   Housekeep File     List just the Housekeeping file  The Housekeep file  contains all data   Selection     Chose the values to be stored in the data file   Conc     Chose the concentrations and isotope ratios to store   Detectors     Chose the detector values to store   Laser     Chose the Laser values to store     Misc     Chose the miscellaneous v
25.  Rate    eie ti epi t ed e RORR F 1  F2 Sample R  te sued te end Incem eme pet F 1  E3 Digital Filters a suos dat eo an ena ea ea tt ent F 2  Feds  Synchtronicity   recette en Oe CREE F 4  F 5 Sample Cell Residence Time                    sse F 5    Table of Contents    G  Using Swagelok   Fittings                                         G 1  Gil  General Notes  doin toii a a nte G 1  G2  Assembly  sedes ota sehe aie G 1  G 3 Common Replacement Parts                  ssssesssssseeeeeene G 2    H  Upgrading Early Generation TGAs to TE cooled    Laser H 1   FL   x TGA DOO bes rire ertet ee eig eee erede H 1  Hille    Basic  Upgrade  ueteres eR RSS H 1  H12 Detectot I                       a H 4   H  13  Power Module  tiere E EO R H 4  HAE  Purge Boot    onset tercie rene eerte ess H 4   H2  TGXAIO0A    nitet tete temet ette eee tete ettet H 4  H 2  T  Basic Upgrade    tieess ost e H 4  H22 Detects nan eed Ee ie tese mete RA HERE H 5  H 2 3 Holes in Enclosure for Cryocooler Refrigerant Tubes               H 5  H 2 4 Temperature Control Upgrade                      sese H 5  H 2 5     Power  Module    2 N eee bn OE H 6  H 2 6   Puree Boot  enedonisetabne aet ad H 6   H3  TGAL00    oon d etie a d tente mer e te et ire ets H 6  H 3 I   Basic Upgrade  s aseos v ausa eene ba ep H 6  IT 3 2   CPU  Module    on pen Aen o e pede  H 6  H 3 3 Input and Output Modules                     sse H 6  H 3 4 Detectors and Detector Cables                  sssssssssss H 7  H 3 5 Temperature Contro
26.  Status window  see Section 7 2 2  Status Window      Record the DC current and compare it to the expected value  An  abrupt change in the DC current may indicate the laser has shifted to    47    TGA Series Trace Gas Analyzers    48    another absorption line  A persistent trend over time in the DC  current may indicate premature aging of the laser     Record the reference transmittance and compare it to previously  recorded values  A change in the reference transmittance may  indicate a problem with the reference gas supply  It can also be  caused by a shift to a different absorption line  or a change in the  pressure or temperature in the sample cell     Record the sample pressure and compare it to previously recorded  values  If the TGA sample pressure is actively controlled by the  datalogger  a change in the pressure indicates a problem with the  flows and pressures in the sampling system  If the pressure is not  actively controlled it will be determined by the sample flow and the  pump capacity  In this case  the pressure will decrease over time as  the sample intake filter s  becomes plugged     If the TGA is equipped with a LN  cooled laser  record the laser  heater voltage and compare it to previously recorded values  The  vacuum inside the laser dewar will gradually degrade  This  degradation reduces the thermal isolation between the outer wall of  the laser dewar and the laser itself  Over time  as more heat is  transferred to the laser by the degraded vacuum  les
27.  Temperature  K        FIGURE C 3  Typical concentration noise as a function of laser  temperature    In some cases the minimum concentration noise may be at a different laser  temperature than the minimum reference transmittance  If the DC current is  near the laser threshold current  the laser s optical power output may be  reduced significantly at higher laser temperatures  lower DC current   This can  be verified by looking at the sample detector signal as a function of laser    Appendix C  Optimizing Laser Parameters    temperature  This is shown in FIGURE C 4 for the example described in the  figures above              Sample Signal  mV              102 5 103 103 5 104 104 5 105 105 5    Laser Temperature  K              FIGURE C 4  Typical sample detector signal as a function of laser  temperature    In this case  it may be possible to compensate for the reduced laser power by  reducing the detector temperatures  see Appendix D 3  Detector Temperature    If adjusting the detector temperatures results in low concentration noise at the  laser temperature which gives the minimum reference transmittance  this is the  optimal laser temperature     The other condition that can give a different optimum laser temperature for  reference transmittance and concentration noise is that in which the position of  a mode hop may also move with laser temperature  If the laser has a mode hop  near the absorption line  the concentration noise may increase as the mode hop  approaches the 
28.  also slightly different for  an initial installation than for subsequent reassembly     First time assembly  plastic tubing     1  Cut the tubing to length   2  Make sure the cut is square and free of burrs     3  Some types of plastic tubing have an aluminum layer  Take care not  to flatten the tube as you cut it     4  Push an insert into the end of the tubing     5  Do not remove the nuts and ferrules from the fitting  Simply insert  the tube into the assembled fitting until it bottoms out     6  Rotate the nut finger tight   7  While holding the fitting body steady  tighten the nut one and one     quarter turns   For 1 16 in or 1 8 in sized fittings  tighten the nut  three quarters turn      G 1    Appendix G  Using Swagelok   Fittings    First time assembly  metal tubing    Extra care is needed to avoid overtightening brass fittings when used with  metal tubing  These notes apply to reducers and port connectors as well as  metal tubing     NOTE No insert is required with metal tubing     1  Do not remove the nuts and ferrules from the fitting  Simply insert  the tube into the assembled fitting until it bottoms out     2  Rotate the nut finger tight     3  While holding the fitting body steady  tighten the nut until it feels  tight  This will normally be less than one full turn  Tightening a full  one and one quarter turns will damage the threads on the fitting and  nut     Reassembly  plastic or metal tubing   You may disassemble and reassemble Swagelok   tube fittings 
29.  are shown in TABLE  F 2  The processing lag for a 100 ms moving average is shown for comparison  in TABLE F 2     The lags are different for each ramp in two ramp  A B  or three     ramp mode  A B C   The filters are designed with these different  lags to correct for the fact that the three ramps are not measured at  the same time  Ramp B is measured 2 ms after ramp A  and ramp  C is measured 4 ms after ramp A     The concentration measurements for all three ramps are  synchronized when using the EC filter option     The moving average lag is half the moving average time and is the    same for all ramps  The concentration measurements for multiple  ramps are not synchronized when using the moving average     Appendix F  TGA Frequency Response                                  TABLE F 2  Processing Lags for EC filters  Ramp A Ramp B Ramp C  EC filter  1 ramp  372  EC filter  2 ramp  750 748  EC filter  3 ramp  746 744 742  100 ms moving average 50 50 50          TGA measurements are triggered by the TGA   s internal clock  asynchronous  of any SDM or analog data acquisition system  The synchronicity of  correlated measurements  for example  sonic anemometer data and TGA data  in an eddy covariance system  will be limited by how often the TGA updates  its measurement  This update time varies from 2 ms to 6 ms  depending on the  measurement mode  one  two  or three ramps   the digital filter  and the output  mode  The different times are summarized in TABLE F 3  The maximum  del
30.  as programming to customer specifications  electrical  connections to Products manufactured by CSI  and Product specific training  is  part of CSI s product warranty  CSI EXPRESSLY DISCLAIMS AND  EXCLUDES ANY IMPLIED WARRANTIES OF MERCHANTABILITY  OR FITNESS FOR A PARTICULAR PURPOSE  CSI hereby disclaims   to the fullest extent allowed by applicable law  any and all warranties and  conditions with respect to the Products  whether express  implied or  statutory  other than those expressly provided herein        Assistance    Products may not be returned without prior authorization  The following  contact information is for US and international customers residing in countries  served by Campbell Scientific  Inc  directly  Affiliate companies handle  repairs for customers within their territories  Please visit  www campbellsci com to determine which Campbell Scientific company serves  your country     To obtain a Returned Materials Authorization  RMA   contact CAMPBELL  SCIENTIFIC  INC   phone  435  227 9000  After an application engineer  determines the nature of the problem  an RMA number will be issued  Please  write this number clearly on the outside of the shipping container  Campbell  Scientific   s shipping address is     CAMPBELL SCIENTIFIC  INC   RMA    815 West 1800 North   Logan  Utah 84321 1784       For all returns  the customer must fill out a    Statement of Product Cleanliness  and Decontamination    form and comply with the requirements specified in it   The fo
31.  but other power options are available  The capacity of the RB0021 L is  adequate for most high flow TGA applications  such as eddy covariance  It  ships with 2 54 cm  1 0 in  ID suction hose  pn 7123   length specified per the  order  and 2 qts of pn 8143 oil  The pump is shown in FIGURE 4 13        FIGURE 4 13  RB0021 sample pump    15    TGA Series Trace Gas Analyzers    RB0021 L specifications     Length    Width    Height    Weight    Power    Pumping speed    Ambient temperature range for operation     Ultimate vacuum     44 2cm  17 4 in    29 2 em  11 5 in    26 9 cm  10 6 in    19 kg  42 Ib    950 W   18 slpm at 50 mb  60Hz   10 to 40   C   2 mb    For detailed installation and operating instructions  refer to the user manual    supplied with the pump     XDD1 Sample Pump    The XDD1 is a four head  three stage diaphragm pump  It is modified to  include a bleeder filter and needle valve  a 1 2 in Swagelok fitting for  connection to the TGA  and a 1 4 in Swagelok fitting for connection to a  pressure controller  shown in FIGURE 4 14  It ships with 15 2 m  50 ft  ofa  1 27 cm  0 5 in  OD LDPE tubing  pn 25539   The pump can be used on AC  mains supplies ranging from 100 to 120 Vac  50 or 60 Hz  or 200 to 230 Vac   50 or 60 Hz   selected by a voltage changeover switch        FIGURE 4 14  XDD1 sample pump    XDDI specifications     Length   Width   Height     16    31 3 cm  123 in   14 4 cm  5 7 in   21 5 cm  8 5 in     TGA Series Trace Gas Analyzers    Weight  6 7 kg  14
32.  cable ties evenly spaced  Route  the cable through the hole in the electronics mounting bracket and fasten with a  cable tie     Appendix I  Install Temperature Control Upgrade    Fasten the other thermistor probe to the sample cell  midway between the laser  dewar and the electronics assembly  using cable ties  See the final  configuration in FIGURE I 3        FIGURE 1 3  Location of second thermistor probe attachment    Fasten the cable to the sample cell with three cable ties evenly spaced  Route  the cable through the hole in the electronics mounting bracket and fasten with a  cable tie     Remove the RTD connectors  RTD 1A  RTD 1B  RTD 2B  RTD 2A  from the  temperature control module  Remove the jumper wires that tie these  connectors to the analog inputs     Wire the thermistors into the analog inputs as described below and referring to  FIGURE I 4     1  2  G  no connection   3  white of thermistor 1  detector end    4  green of thermistor 1  detector end    G  black and shield of thermistor 1  detector end   5  white of thermistor 2  laser end    6  green of thermistor 2  laser end    G  black and shield of thermistor 2  laser end    7  8  G  pressure sensor    Left pin of EXT  Red wires from both thermistors and the pressure sensor     Appendix I  Install Temperature Control Upgrade            xi 3   TGAN    ein fh    RTD IA RTD 18       FIGURE l 4  Thermistor cable wiring to analog inputs    1 2 Connect the Control Cable    Connect the control cable to the CSIO por
33.  can be found  perform the following checks     e If two lasers are installed  verify they are aligned to the correct  laser    e Verify the dewar cable is installed correctly  if two or more  lasers are installed  verify the correct cable is being used    e Verify the detector cables are correctly installed  e Verify the laser is enabled in the TGA program    e Recheck the laser temperature  and the settings for zero  DC   modulation  and high current    If a detector response is still not observed  Contact Campbell  Scientific for assistance     B 1 2 Horizontal and Vertical Alignment    NOTE    Once the system is aligned well enough to see a response in the sample  detector  follow these steps to optimize the horizontal and vertical alignment     Adjust the horizontal alignment screw  see FIGURE B 3  to maximize  the sample detector signal     The sample and reference signals may not reach their maxima  simultaneously  If so  ignore the reference detector signal and  adjust the alignment to maximize the sample detector signal   Adjust the horizontal position past the peak in each direction far  enough to make sure there is a single response peak  If there is a  single peak  leave it at the center of the peak  If there are multiple  peaks  leave the horizontal alignment at the center of the group of    Some older systems used a relatively coarse pitch screw for the horizontal  alignment at the laser end and a second horizontal adjustment screw at the  detector end  This scr
34.  cell containing a prepared reference gas having a  known concentration of the target gas  The reference signal provides a  template for the spectral shape of the absorption line  allowing the  concentration to be derived independent of the temperature or pressure of the  sample gas or the spectral positions of the scan samples  The reference signal  also provides feedback for a digital control algorithm to maintain the center of  the spectral scan at the center of the absorption line  The simple optical design  avoids the alignment and contamination problems associated with multiple   path absorption cells  The number of reflective surfaces is minimized to  reduce errors caused by Fabry Perot interference     TGA Series Trace Gas Analyzers    CPU   The TGA100 used a CPU module based on transputers  These CPU modules  had limited processing capability  so the TGA100 required a DOS PC  with  another transputer on an expansion card  to be connected in real time via a  fiber optic cable  The TGA100A used a new CPU module with sufficient  processing power to make the PC unnecessary  This transformed the TGA  from a PC based system to a synchronous device for measurement  SDM   sensor  Most of the TGA100s have been upgraded to this new CPU module   Contact Campbell Scientific if you have questions regarding this upgrade     Software   There have been four generations of TGA software  The TGA100 software  was distributed between three microprocessors  the transputer in the TGA100  
35.  current is zero  The zero phase is used to measure the detector s  dark response     The reduced current of the zero phase dissipates less heat in the laser  causing it  to cool slightly  The laser s emission frequency depends on its temperature as  well as its current  Therefore  the temperature perturbation caused by reduced  current during the zero phase  introduces a perturbation in the laser s emission  frequency  During the high current phase  the laser current is increased above  the current to be applied during the modulation phase to replace the heat    lost     during the zero phase  This stabilizes the laser temperature quickly   minimizing the effect of the temperature perturbation  The entire scan  sequence is repeated every 2 ms  Each scan is processed to give a  concentration measurement every 2 ms  500 Hz measurement rate      Modulation Phase  High Current Phase  Spectral Scan      Temperature  Stabilization       Laser  Current   Zero Phase    Laser Off       Omitted Used in Calculation      er   Detector  Response           2ms    FIGURE 4 27  TGA laser scan sequence    TGA Series Trace Gas Analyzers    TGAs can be configured to measure two or three gases simultaneously by  alternating the spectral scan wavelength between nearby absorption lines  This  technique requires that the absorption lines be very close together  within about  1 em     Given this  it can only be used in very specific cases  The multiple   ramp mode is used to measure isotope ratios 
36.  discontinuity was  caused by a mode hop  this indicates the end of the temperature tuning range  for the selected absorption line     It is generally not necessary to actually plot the data  For reference  FIGURE  C 1 shows a graph of a typical data set where the same absorption line is  scanned with a range of temperatures     Laser DC Current  mA           102 5 103 103 5 104 104 5 105 105 5    LaserTemperature  K           FIGURE C 1  Typical laser DC current as a function of temperature    Next  look at the reference detector transmittance as a function of temperature   The transmittance should have a minimum at the  optimum  laser temperature   It should be higher at temperatures above and below the optimum temperature   This increased transmittance results from an increased fraction of the laser s  energy at undesired frequencies  multimode operation   Again  it is usually not  necessary to plot the data  but FIGURE C 2 shows a typical example     Appendix C  Optimizing Laser Parameters          Reference Transmittance        102 5 103 103 5 104 104 5 105 105 5    Laser Temperature  K        FIGURE C 2  Typical reference transmittance as a function of laser  temperature    Optionally  look at the concentration noise as a function of temperature  The  concentration noise should generally have a minimum at the optimum laser  temperature  as determined by the minimum reference transmittance   This is  illustrated in FIGURE C 3     Concentration Noise  ppb     104    Laser
37.  factory and is stored in nonvolatile  memory  Ifthe memory storing the MAC address becomes corrupted  it will  be displayed as all zeros  0x000      To reenter the MAC address  execute the  following steps     1  Power down the TGA     43    TGA Series Trace Gas Analyzers    2  Remove the CPU module and note the serial number  Convert the  serial number from decimal to hexadecimal  base 16   Web based  converters are available  FIGURE 6 19 shows an example   converting serial number 1458 to it representation in hexadecimal   5B2     File Edit Viev  01  Decimal to Hexadecimal Converter         wow binaryhexconverter com decimal to hex converter     B  Most Visited     Getting Started     Suggested Sites   Web Slice Gallery    BinaryHex  Converter    Decimal to Hexadecimal Converter    To use this decimal to hex converter tool  you have to type a decimal value like 79 into the left field  below  and then hit the Convert button  Therefore  you can convert up to 10 decimal characters to    hex   DEI 3  109    Decimal Value  max  9 digits  Hexadecimal Value   1458   582                Converter    Decimal to hex conversion result in base numbers     1458 10    5B2 i6     v    v Firefox automatically sends some data to Mozilla so that we can improve your experience  Choose What I Share   Xx       FIGURE 6 19  Web based decimal to hexadecimal converter    3  Reinstall the CPU module and power up the TGA   4  Connect to the TGA with Device Configuration Utility     5  Select Termina
38.  far left side provides drop down menus which  allow the user to set the parameters or retrieve additional information for each  of four major categories     e Laser   e Detectors   e Calculations   e Other   The left panel is shown in FIGURE 7 7 with the submenu for each category  expanded  The submenus are explained in the following sections     52    7 2 3 1 Laser    CAUTION    TGA Series Trace Gas Analyzers          Laser  Temperature  Current  Line Lock  Other      Detectors  Temperature  PreAmp       Calculations  Concentration  Analog Input      Other  TGA Temperatures  Data Output  Serial Numbers  About TGA    FIGURE 7 7  Expanded view of the menu in the TGA s Settings  window    Temperature   The Settings  gt  Laser  gt  Temperature window is different for TGA Windows and  TG TEC  TGA Windows  FIGURE 7 8  supports LN2 cooled lasers that  measure temperature in K and have a heater to control their temperature above  the LN   tank temperature  TGA TEC  FIGURE 7 9  supports TEC lasers that  measure temperature in   C and have a TE cooler to control their temperature     The laser temperature setpoint should be set and optimized for the individual  laser  See Appendix C 1  Laser Temperature  for details on how to set the laser  temperature  The laser temperature may also be set from the Laser window   see Section 7 2 4  Laser Window      The maximum laser temperature  K or   C  provides a safety shutdown  If the  laser is warmer than this  the TGA will automatically turn
39.  filter  which is implemented as  a convolution of the data with a filter function  There are 5 sets of filter  coefficients  This allows the user to select the passband  1  2  3  4  or 5 Hz   The graph below shows the FIR filter coefficients  The five blue curves are for  the 1  2  3  4  and 5 Hz passband EC filters  The 5 Hz passband filter is the  tallest  narrowest one  A moving average can also be thought of as an FIR  filter  with uniform weighting  For comparison  FIGURE F 1 shows the red  curve as the filter coefficients for a 100 ms moving average                 Time  sec     FIGURE F 1  EC filter coefficients    FIGURE F 2 shows the frequency response of the filters  An ideal filter would  have a flat response in the passband and drop to zero at the passband cutoff   The EC filters  blue curves  are very flat to just beyond the passband cutoff   and then roll off to near zero by about 5 Hz above the passband cutoff  The  graph includes the 100 ms moving average frequency response  red   for  comparison  The moving average shows no truly flat passband and it rolls off  more slowly than the EC filters     No units    No units    o      r    0 001      0 0001      1e 005 4  0       A    Appendix F  TGA Frequency Response    Hertz    FIGURE F 2  EC filter frequency response  linear scale     FIGURE F 3 graphs the same frequency response curves as shown in FIGURE  F 2  but with a logarithmic Y axis  and an extended frequency range to show  the attenuation beyond the pas
40.  offset      Detectors    Temperature  D IPreAmp   Preamp gain 45 00       Automatically adjust detector gain offset         Calculations et     Concentration Reference   Sample       Analog Input   B  Other   TGA Temperatures      Data Output Offset 4     Serial Numbers      About TGA    CH4 linearity coeff 0 000         Gain 3    AI   iial                         FIGURE 7 16  PreAmp window default settings    Concentration  The Settings  gt  Calculations  gt  Concentration window and parameters are the  same for both versions of the software     The TGA measures the gas concentration every 2 ms and applies a digital filter  to these measurements before they are output  The top section of the Settings  gt   Calculations  gt  Concentration window  as shown in FIGURE 7 17  allows the  user to select the type of digital filter  either a moving average or a finite  impulse response  FIR  filter that is optimized for eddy covariance  applications  If Moving average is selected  enter the averaging time  in ms   If EC filter is selected  enter 1  2  3  4  or 5  depending on the amount of  filtering that is desired  See Appendix F  TGA Frequency Response  for  details     61    TGA Series Trace Gas Analyzers    62                   nIm       ES         TGA Parameter Settings x     Laser        Concentration Filter  Current  Line Lock    Moving average  ms  100   Other d  a Detectors y EC filter  Hz  1  Temperature  PreAmp Reference Gas Concentration     Caladations   CHA 500 00  Co
41.  so the temperature must be stabilized at the start of each spectral  scan  Overdriving the current  as discussed in Section 4 2 5  Laser Scan  Sequence  can help to stabilize the laser temperature more quickly  but setting  the zero current as high as possible minimizes the temperature perturbation     The TGA Windows software has an Auto button for setting the    Zero current  In some cases this function gives an incorrect result   The suggested procedure for setting the zero current for all lasers  is given below  This automatic function has been removed from  the TGA TEC software     Set the zero current to 0 mA  Run the Laser Line Find  refer to Section 7 2 5   Find  starting with the DC current at 0 mA  View the resulting graph to  determine the laser s threshold current  DC current at which the reference  voltage begins to rise   In the example shown in FIGURE C 5  the reference  signal is near zero up to a DC current of 50 mA  and rises steeply beyond 50  mA  This is the laser s threshold current  Set the Zero current to  approximately 90  of the threshold current  45 mA in this example      For multiple ramp mode  set the zero current to this value for each ramp  In  some cases the zero current may be reduce below this value for some ramps   see 0  High Current                   gt   E      o   a  m     o   gt   n  o  c  D  iz  D  2  D  a    5       10 15 20    25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100    Laser Find  8 00    DC Current  mA           FIGURE C 5
42.  sss B 1  Bihl Initial Alignment    einer B 4  B 1 2 Horizontal and Vertical Alignment                     sess B 5  B 1 3 Focus Adjustment          sse eie eer enda B 6  B 1 4 Reference Detector Coalignment                   sse B 7  B 2 Optical Alignment of TGA200 and TGA200A                sss B 7  B 2 1 Configure the TGA PC Software    B 9  B22 Initial Alignment    esee eie eda B 9  B 2 3 Horizontal and Vertical Alignment                    sssssssss B 12  B24 Focus Adjustment    tte etie tI RR TI NETUS B 13  C  Optimizing Laser Parameters                                  C 1  Colt Laser Temperatute     t retro eq ae qas C 1   42     Zero Currentses see E datei eee ae OP C 5   9  igh    Currents  sn tdeo ete ee eee C 7  C 3 1 High Current Count               sssssssssseeeeeeeeenenenee C 10  CA    Omitted Data Count    eee eene C 11  C 5 Modulation Current                   esee nennen C 11  C 6 Laser Maximum Temperature and Laser Maximum Current          C 12  C 7 Laser Multimode Correction                    seen C 12  D  Optimizing Detector Parameters                             D 1  D 1 Detector Gain and Offset                sssssssssssssseeeeeeene D 1  D  Detector Preamp  Gan   i   aree re E D 1  Di3 Detector Temp  rature    uie etre i beoe eere D 1  D 4 Detector Linearity Coefficients                sssssssssseeeeee D 2  E  Calibration uiii tice oor ERR ERE 132 SA pep sa ducens E 1  F  TGA Frequency Response                                       F 1  F 1 Measurement
43.  stored in two places  in the parameter file  and in the TGA s  nonvolatile memory  These serial numbers must match in order for the PC  software to be connected to the TGA  as discussed in Section 7 2 1  Connect  Window  The TGA serial number may be edited only when the TGA is  disconnected     Use extreme caution when sending parameters to the TGA  to make sure the correct parameters are being sent  If the  wrong parameters are sent for the laser in the TGA  the laser  may be damaged     The TGA description is intended as a device nickname and to be a more user  friendly way than the serial number by which to identify which TGA 1s  connected to the PC software  It may be set only through the Device  Configuration Utility  see Section 6 2 4  Set TGA Serial Number and  Identification String   The identification string is displayed at the top of the  toolbar when the TGA software is connected to a TGA     The serial numbers for the laser  dewar  and detectors provide a historical  record in case these components are replaced by the user     TGA Series Trace Gas Analyzers    The site name description allows the user to enter text to describe the physical  location of the TGA  This is stored in the parameter file as a historical record                     Laser     Temperature    Current   Line Lock   H t   Other   5  Detectors    i PreAmp  Calculations       Concentration    E  Other  i TGA Temperatures   Data Output    zx serial Numbers      About TGA    Temperature    Analog 
44.  supplied with TE cooled detectors and  require no changes to the detectors or detector cables     H 2 3 Holes in Enclosure for Cryocooler Refrigerant Tubes    If the TGA100A was supplied with a cryocooler system for the laser  the end  of the TGA enclosure has feedthrough holes for the refrigerant tubes  The  feedthrough grommets may be removed and the holes plugged with pn 15542  hole plug and pn 23 nut  two of each required   This is not required for the TE   cooled laser upgrade  but it is recommended to seal the holes previously used  for the cryocooler refrigerant tubes  Most TGA100As shipped with an LN     dewar for the laser and already have the hole plugs installed     FIGURE H 5 shows the feedthrough grommets for the cryocooler refrigerant  tubes and the holes plugged with the recommended hardware        FIGURE H 5  Cryocooler feedthrough holes with grommets  above   and plugs  below     H 2 4 Temperature Control Upgrade    TGA100As shipped with a temperature control module that provided power to  fans and heaters inside the enclosure to maintain temperature at each end of the  enclosure  The original design worked independently from the TGA  electronics and software  A temperature control upgrade is available to allow  the TGA temperature to be controlled through software     The temperature control upgrade includes thermistor probes and a control  cable  It gives more a more accurate temperature measurement  the  convenience of software control  and provides bette
45.  the laser current off  to protect the laser  This limit should be set for the individual laser     Setting the maximum laser temperature parameter above a    laser   s safe operation temperature greatly increases the risk  of damaging the laser     The parameters on the bottom part of the window should normally be set to the  factory defaults  as shown for each type of laser in FIGURE 7 8 and FIGURE  7 9     53    TGA Series Trace Gas Analyzers          i  TGA Temperatures   Data Output  j Serial Numbers          Laser Temperature  Laser temperature set point  K   Laser temperature measured    Maximum Laser Ratings  Maximum laser temperature  K     Laser Temperature Calculations    Laser temperature slope  K V   Laser temperature offset  K     Laser Heater Calculations  Laser heater update interval    Laser heater control gain  Laser heater control zero  Hz     Laser heater control pole  Hz        FIGURE 7 8  TGA Windows laser settings        Data Output     Serial Numbers          Laser Temperature  Laser temperature set point    C   Laser temperature measured    Maximum Laser Ratings  Maximum laser temperature    C     Laser Temperature Calculations  Thermistor Rzero  Themnistor gain  Thermistor offset    Laser Cooler Calculations  Laser cooler update interval  Laser cooler control PID gain  Laser cooler control PID tau       FIGURE 7 9  TGA TEC laser settings    54    CAUTION    TGA Series Trace Gas Analyzers    Current   The Settings  gt  Laser  gt  Current window is sli
46.  the relevant section below for  details     B 1 Optical Alignment of TGA100 and TGA100A    The TGA100 and TGA100A have a simple  robust optical design that makes it  easy to adjust and maintain its optical alignment  The optical system   illustrated in FIGURE B 1  includes the laser  a collimating lens in front of the  laser  a beamsplitter to reflect some of the laser   s energy onto the reference  detector  and two focusing lenses mounted in front of the sample and reference  detectors     Reference    To pump                Sample  amp     detector JU    N2 Beamsplitt    NOTE    defector     4           Reference gas in Dewar    Sample cell       EM laser    iw To pump Sample in    FIGURE B 1  TGA100 and TGA100A optical layout    To adjust the TGA100 and TGA100A optical alignment complete the  following steps     1  Loosen the transport lock screw and axial lock screw at the detector  end  refer to FIGURE B 2   These screws should both be loose  during all periods of operation to allow the detector end to move  slightly as the length of the long sample cell changes with  temperature     The transport lock screw and axial lock screw should be tightened    only for transporting the TGA     B 1    Appendix B  Optical Alignment    NOTE FIGURE B 2 shows the horizontal adjustment screw that was  included with older units  Newer units use a fine pitch horizontal  adjustment screw at the laser end which makes horizontal  adjustment at the detector end unnecessary     Horizontal  
47.  the transputer in the PC  and the PC itself  The PC provided the user interface  as well as storing data on the hard drive     The TGA100A  which included the new CPU module  separated the TGA  firmware  operating system of the TGA itself  from the user interface software  which ran on the user   s PC  With this upgrade  the TGA became a stand alone  sensor  and the PC became necessary only to check and adjust the operation of  the TGA  The second generation of the interface software ran only under DOS  PCs     The third generation of user interface software  TGA Windows  was developed  for the TGA200  TGA Windows greatly improves the ease of use  and allows  the use of a modern PC running a Windows operating system  This software is  available as an upgrade for both the TGA100  with CPU upgrade  and  TGAI00A     The fourth generation software  TGA TEC  is a minor update to the TGA  Windows software that is compatible with the thermoelectrically  TE  cooled  lasers used in the TGA200A  TGA Windows and TGA TEC are not mutually  compatible  TGA Windows is used only with LN2 cooled lasers and TGA TEC  is used only for TE cooled lasers  A representative screen of the TGA  Windows software is shown in FIGURE 4 1     TGA Series Trace Gas Analyzers       9 Laser Settings  TGA windows   Connected to sn 1110 miz  View Less   Ta  M ified v Colors   pr   lt           Fics    DE carent EZEK DE cieni EZE        CO2 Conc    180 Conc T    FRE Wa  2 Manual a                            FIGUR
48.  to access in the field  It allows the convenience of replacing the entire  heated filter holder in the field  and then opening the filter holder to replace the  filter element later  If there is convenient access to the sample intakes  a lower  cost option is to simply replace the filter element  pn 17575        FIGURE 4 22  TGA heated intake filter    The filter element  pn 17575  is a 7um pore size filter element used in  Swagelok   1 4 in inline filter holders  pn 17574   It is used as a replacement  element for the heated filter  pn 18058   heated filter  amp  orifice  pn 18072   and  the PD625 sample air dryer  The replacement filter element is shown in  FIGURE 4 23     TGA Series Trace Gas Analyzers       FIGURE 4 23  TGA filter element replacement    4 1 6 3 TGA Filter Elements  47mm  Qty 100     A box of 100 polypropylene filter elements  10 um pore size  is available as pn  9838 and shown in FIGURE 4 24  These are used with 47 mm filter holders  such as the pn 20553 TGA High Flow filter holder  The filter elements may be  purchased as a replacement item     FIGURE 4 24  47 mm replacement filters    4 2 Theory of Operation    Campbell Scientific TGAs measure trace gas concentration in an air sample  with tunable diode laser absorption spectroscopy  TDLAS   This technique  provides high sensitivity  speed  and selectivity  The optical configuration of  the TGA100 and TGAT100A is different than TGA200 and TGA200A  as  described in the following sections     4 2 1 Optic
49.  to measure its  concentration every 2 ms  500 Hz measurement rate   driven by its own  internal clock  If it is measuring multiple species  it alternates its measurement  scans between them  For example  if it is configured to measure N2O and CHa   each will be measured on alternate scans  every 4 ms  250 Hz measurement  rate   If it is configured to measure   O and PC isotope ratios in CO    each of  the three isotopologs will be measured every 6 ms  167 Hz measurement rate      F 2 Sample Rate    The TGA does not store data  so its measurements must be sampled and stored  by an external device  The sample rate is determined by the data storage  device  Three options are supported     e Campbell Scientific datalogger  CR6  CR1000  CR3000  or CR5000   e Windows PC    e Analog output  requires an optional interface module     Datalogger   When using a datalogger  the TGA is configured as an SDM sensor  The TGA  responds to the datalogger s data request by sending back the most current   digitally filtered sample of the high rate concentration measurements  Thus   the sample rate is determined by the datalogger program     PC   When TGA data are sampled by a PC  the user selects the update interval  in  ms  on the connection window  The TGA will send data to the PC at the  specified interval  which is based on the TGA   s internal clock  These data will  be the most current  digitally filtered sample of the high rate concentration  measurements  Thus the sample rate is determi
50.  value  This will disable  line locking if the signal level drops  during optical alignment adjustment  for  example     Set the Max transmittance well above the normal measured reference  transmittance  A typical setting would be half way between the measured  transmittance and 100   Setting the parameter this way will disable line  locking if there is a problem with the reference gas  such as forgetting to open  the valve on the reference gas cylinder               amp      m Y  TGA Parameter Settings e   B  Laser    Temperature  Current Maximum Pressure for line lock to be enabled   Other  B  Detectors  Temperature     PreAmp      Calculations Minimum Reference Detector Signal for line lock   Concentration  Analog Input Min signal strength 1 00  4 66 mV  B Other  i   TGA Temperatures  Data Output  Serial Numbers  About TGA    Max pressure 100 00 54 80 mb    Maximum Reference Transmittance for line lock     Max transmittance 85  73 91                  FIGURE 7 12  Typical Line Lock parameter settings    57    TGA Series Trace Gas Analyzers    58    Other  The Settings  gt  Laser  gt  Other parameter window is the same for both versions  of the software                                       Laser  Temperature  t  Current Number of Ramps  i Line Lock  i Number of ramps 15  B Detectors      Temperature  i oe PreAmp CH4  B  Calculations     Concentration  Analog Input Gas name CH4  B  Other Multimode power   0 00   TGA Temperatures  Data Output  Serial Numbers  About TGA L  Ram
51.  with the following steps   1  Open a Laser Settings window     2  Select the Magnified display mode     Appendix C  Optimizing Laser Parameters    Set the high current to an initial value  start at zero mA when in  doubt      Click the padlock icon next to DC current  or press  T  to start line  lock     Click the Auto button next to the Mod current to adjust the  modulation current     Click the padlock icon next to DC current  or click on the detector  graph and press I to stop line lock     Click on the detector graph to enable adjustment of the DC current by  using the  lt  and  gt  cursor keys  alternately you may click in DC  current box and use the A and V cursor keys  or simply click the up  and down adjuster boxes      Repeatedly press the   4 cursor key to decrement the DC current in 0 1  mA steps until the center of the absorption line is at the right edge of  the spectral scan  Count the number of steps required     Start line lock again  Watch the reference detector display as the lines  to come to the center of the spectral scan       Stop line lock and repeatedly press the    cursor key to increment the    DC current in 0 1 mA steps until the center of the absorption line is at  the left edge of the spectral scan  Count the number of steps required       Start line lock again  Watch the reference detector display as the lines    come to the center of the spectral scan       Evaluate the results and iterate as needed using the steps below and    referring to FI
52. 00 and TGA200A focus is adjusted at the factory  and normally no  further adjustment is needed  If a system is configured for multiple lasers that  have greatly different wavelengths  however  the focus may need to be adjusted  when switching between these lasers  To adjust the focus  loosen the locking  ring on the lens mounting tube and screw the lens mounting tube in or out to  maximize the detector signals  Retighten the locking ring when the optimum  focus is achieved     Appendix B  Optical Alignment    Appendix C  Optimizing Laser    Parameters    Normally the laser parameters are adjusted only when a new laser is installed   after transporting the system  or warming and recooling the laser  LN2 cooled  laser   These parameter settings optimize the performance for a specific  absorption line  so these steps should be performed after the correct absorption  line is chosen  see Appendix A 4  Finding the Absorption Line      C 1 Laser Temperature    NOTE    If the TGA   s laser were perfect  it would emit at only one frequency  single  mode   This emission frequency would depend only on the injection current   and the emission frequency could be tuned over a wide range  In fact  the real  laser   s emission frequency is dependent on both its current and temperature  it  may emit some of its optical energy at other frequencies  multimode   and its  emission frequency can be tuned over only a small range before it jumps to a  different frequency  mode hop   The multi mode p
53. 0877 Carbon Dioxide  6 C and 6 50  12160160  30160160  and 120180160       A 1 3 Dewar Cables    For LN   cooled lasers  the standard laser dewar can accommodate up to two  lasers  Each laser position  position 1 and position 2  requires a corresponding  dewar cable  Position 1 and 2 cables are nearly identical  but they are wired to  connect the Laser Current output from the TGA electronics to different pins on  the dewar s connector  Inside the dewar  these pins are connected to the lasers  mounted in position 1 and 2  Connecting the position 1 or 2 dewar cable  therefore determines which laser is active  See Appendix A 1 4  Changing  Lasers  for additional details on selecting the laser     An optional second laser mount assembly will allow up to two more lasers to  be installed in the dewar  The same position 1 and position 2 dewar cables are  used for the lasers on the second mount  If a replacement dewar cable is  needed  see TABLE A 3        TABLE A 3  Replacement Cables for TGA Dewars    Position 1 Position 2  TGA100A 17895 17896  TGA200 21954 22001             Dewar cables for the TGA100 are no longer available  However  in most cases  the TGA100 electronics may be upgraded to use the TGA100A cables   Contact Campbell Scientific for details     TE cooled lasers do not require a dewar or a separate dewar cable  The  appropriate cabling 1s supplied as part of the laser assembly     A 1 4 Changing Lasers    This section outlines the procedure to swap out the laser assem
54. 1    42       TGA Series Trace Gas Analyzers       Device Configuration Utility 2 10 Beta  File Backup Options Help         Deke Tipe Settings Editor   Send OS   Terminal    Logger Control            Logger Program    El       Current Program    CPU tga cr9    El          El    Last Compiled   01 08 13 15 26 52        ri     Last Compile Results     cpu tga cr9    Compiled OK     TGA100A TGA200    Communication Port       Baud Rate Program Send Status        Click  Send Program  to send a new program     Send Program       FIGURE 6 17  Device configuration utility Logger Control tab    6 2 3 Configure Ethernet Connection    There are two ways to set up the Ethernet connection between the TGA and a  PC  Ifa local area network  LAN  is available  connect the TGA and your PC  to the LAN  This option allows a PC to connect to the TGA from any location  in the world  as long as both are connected to the internet    If no LAN is available  connect your PC directly to the TGA with an Ethernet  crossover cable  pn 18148 is supplied with the TGA for this purpose   This  option requires the PC and the TGA to be physically close  but it requires no  additional infrastructure  Configure your PC for a peer to peer network with  the TGA  Consult your information technology support group as needed     With either type of Ethernet connection  the TGA s IP address  Netmask   Gateway  and IP port must be set  These settings are stored in the CPU  module s nonvolatile memory  They can be change
55. 100s had no temperature controller for the enclosure  The  TGAHEAT temperature controller was made available as an option for  TGA100s in 2002  and was always included in the TGA100A  This  temperature controller is not required for the TE cooled laser  but it is  recommended for all TGAs to help maintain a steady temperature within the  TGA enclosure  To install a temperature controller in an older TGA100   contact Campbell Scientific for details     A later upgrade for the TGA temperature controller allowed the temperature to  be controlled through software  See Appendix I  Install Temperature Control  Upgrade  for details     Appendix H  Upgrading Early Generation TGAs to TE cooled Laser    Appendix I     Install Temperature    Control Upgrade    This section shows how to install the TGA temperature control upgrade so that  the TGA temperature can be controlled through software  This upgrade is  available for TGA100As and TGA100s with the optional Temperature  Controller and the CPU upgrade  This upgrade requires a kit containing these  parts     e two thermistor probes pn 22209    e one special control cable  FIGURE I 1       FIGURE I 1  Special control cable    1 1 Install Thermistor Probes    Fasten one of the thermistor probes to the sample cell  midway between the  detector block and the electronics assembly  using cable ties as illustrated in  FIGURE I 2        FIGURE 1 2  Location of first thermistor probe attachment    Fasten the cable to the sample cell with three
56. 2 Fraction of full power applied to the TGA 5 1       enclosure  laser end  arb                Given in   C if TGA TEC    Voltage of cooler if TGA TEC    93       TGA Series Trace Gas Analyzers    94    Status Flags   The TGAStatus value gives an indication of the overall status of the TGA  A  value of zero indicates a normal condition  A nonzero value indicates one or  more of the bits are set  The meaning of each of the bits is given in TABLE                                     7 4   TABLE 7 4  Descriptions of TGAStatus Values  Bit Decimal Value Description  0 1 Line Lock for ramp A is OFF  1 2 Line Lock for ramp B is OFF  2 4 Line Lock for ramp C is OFF  3 8 Sample detector signal exceeded input range  4 16 Reference detector signal exceeded input range  5 32 Sample detector temperature is outside its specified range  6 64 Reference detector temperature is outside its specified  range  7 128 Laser temperature outside its specified range  8 256 Pressure is above its upper limit                   7 3 2 TGA Output to PC    The Ethernet connection from the TGA to the user s PC does allow data to be  collected via the TGA Windows or TGA TEC software  This is normally used  only for testing and troubleshooting the TGA     7 3 3 TGA Analog Outputs    For applications that require analog outputs instead of this SDM connection  it  is possible to add an analog output module that can be driven by the SDM  connection  Contact Campbell Scientific for details     Troubleshooting and Ma
57. 3  TGA200A power module with cables installed    6 2 TGA Software Installation    6 2 1    NOTE    This manual documents the TGA Windows and TGA TEC software  These two  software packages are very similar  but they are not interchangeable  TGA  Windows is used with LN   cooled lasers  while TGA TEC is used with TE   cooled lasers  These two packages are very similar  so most of the discussion  applies to both  Differences are noted as applicable     The TGA software runs on the user   s PC  which must be running Windows XP  or newer operating system  The software provides the user interface to the  TGA  allowing the user to view the operation of the TGA and set parameters   The software can also be used to collect TGA data although in most cases a  Campbell Scientific datalogger is used to collect data     Early DOS based versions of the TGA software are no longer    supported  Any TGA100A or a TGA100 with an upgraded CPU  module may be upgraded to the TGA Windows or TGA TEC  software  Contact Campbell Scientific for details     Installation of PC Software    To install the software  run the setup exe file on the installation disk  This will  install the TGA software file TGA_Windows exe or TGA TEC exe on your PC     39    TGA Series Trace Gas Analyzers    40    The default path is C  Program Files Campbellsci TGA   The installation will  install a shortcut to this file and put it on your desktop     For users with multiple TGAs and a combination of cryogenic and TE cooled  l
58. 3 2 TGA Insulated Enclosure Cover                        sss 13  4 1 4 Other Accessories seo ue eeann ar EERE RESER 14  4 1 4 1 TGA Air Sample Intake 0 0    eeeeeeeseesteeneeeneeeeeees 14  4 1 4 2 TGA Heated Intake Filter  amp  Orifice                  sseussse 14  4 1 4 3 TGA High Flow Filter Holder                          ssssss 14  4 1 4 4 Sample Vacuum Pump            cc ecceeceesceeeceseceteceeeeneeeneeeneeees 15  4 1 4 5 Bypass Vacuum Pump                 seen 17  4 1 46  Sample Air Dryers sete 18  A47   DIGASROmelet  sh eerte tds 21  4 1 5       Support Software s  5 soie Sorted waked eee 22  4 16    Replacement Partsi eb d is 22  4 1 6 1 TGA Heated Intake Filter                      sese 22  4 1 6 2 Filter Element                sess 22  4 1 6 3 TGA Filter Elements  47mm  Qty 100                              23  Theory of Operation    eene enne 23  4 2 1   Optical System    nennen 23  2 272   ESSGE o trit cod eret tere eset e E io EE 25  2 PPA E D TAE  CEDE E 25  4 2  Am  Detectors eee ee rette eee A reet ie ERR 25  4 2 5 Laser Scan Sequence nenea a aA ERE aa eT NSR 26  4 2 6 Concentration Calculation                  sssssssseseeeeeee 27    Table of Contents    5   SpecIfICcallolis      n nocere ee eee botcveevshennectaceeneeee 28  5 1 Measurement Specifications             essen 28  5 2 Physical Specifications    29  5 3 Power Requirements  ipse erener eren 29   6  InstallatiOno  iiis sicci ree audae p a cu rev R Rea nasa wees 30  6 1   Analyzer Installation       
59. 5 06 L s  TGA Model  L   ms   TGA100 0 480 95  TGA100A 0 480 95  TGA200 0 420 83  TGA200A 0 200 40                The residence time is less than 100 ms in every case  Therefore the  recommended EC filter passband setting is 5 Hz  see TABLE F 1      Appendix G  Using Swagelok   Fittings    G 1 General Notes    G 2 Assembly    This appendix gives a few tips on using Swagelok   tube fittings  For more  information  consult your local Swagelok   dealer or visit their web site at  www swagelok com     Do not use fitting components from other manufacturers     they are not  interchangeable with Swagelok   fittings     Do not attempt to use metric fittings  Six mm is very close to 1 4 in  but  they are not interchangeable  Metric fittings can be identified by the  stepped shoulder on the nut and on the body hex     Make sure that the tubing rests firmly on the shoulder of the tube fitting  body before tightening the nut     Never turn the fitting body  Instead  hold the fitting body and turn the nut     Keep tubing and fittings clean  Always use caps and plugs to keep dirt and  debris out     Do not overtighten fittings as it will damage the threads     If a nut cannot be easily tightened by hand  this indicates the threads have  been damaged  Replace any damaged nuts and fittings     The first time a Swagelok   fitting is assembled  the ferrules become  permanently swaged onto the tube  Assembly instructions vary depending on  plastic or metal tubing  The assembly instructions are
60. 5 cm  6 5 in   Height  25 4 cm  10 0 in   Weight  8 2 kg  18 0 Ib   Power  250 W    17    TGA Series Trace Gas Analyzers    18    Pumping speed  50 L min        Ultimate vacuum  180 mb    DAAV505 L Sample Pump   The DAAVS505  shown in FIGURE 4 16  is a double head diaphragm pump  used for applications that require more flow than the DOAV502 can supply   The   L  on a product model indicates that the tubing length is specified at the  time of order        FIGURE 4 16  DAAV505 vacuum pump    DAAV505 L specifications     Length  29 8 cm  11 7 in   Width  16 5 cm  6 5 in   Height  25 4 cm  10 0 in   Weight  11 4 kg  25 1 Ib   Power  370 W   Pumping speed  100 L min   Ultimate vacuum  180 mb    4 1 4 6 Sample Air Dryer    Accurate measurements of trace gas fluxes by eddy covariance or gradient  techniques require that variation in water vapor concentration be eliminated  either by drying the sample gas before it is measured or by correcting the trace  gas flux   Four sample air dryers are available  the PD200T  PD625  PDIT   and the PD1T 1 5  See TABLE 4 4 to compare specifications for these dryers           Webb  E K   Pearman  G I  and Leuning  R   1980   Correction of flux measurements  for density effects due to heat and water vapor transfer   Quart  J  Met  Soc  106  85   100     TGA Series Trace Gas Analyzers    The dryers work by forcing the humidity in the sample air through the walls of  the Nafion   tubing  where it is carried away by the purge flow  The sample air  flow
61. 9  Controlling the detector temperature settings                   sssssssss 60  PreAmp window default settings                 sssessseseeeee 61  Calculation concentrations settings    62  Default settings for Channel 1 in analog input screen                         63  Channel 2 settings for a TGA with thermistor probe                           64  Typical pressure value inputs  values are specific to each TGA        65  Typical values for calculating 613C on a TGA set with multiple   FAMPS i E 66    Table of Contents    7 22     7 23   7 24   7 25   7 26   7 27   7 28   7 29   7 30   7 31   7 32   7 33   7 34   7 35   7 36   7 37   7 38   7 39   7 40   7 41   7 42     7 43   7 44   7 45   7 46     A 1   A 2   B 1   B 2   B 3   B 4   B 5   B 6   B 7     B 8   B 9     C 1   C 2     C 3   C 4   C 5     C 6   C 7   C 8   C 9     F 2   F 3     TGA temperature control window for the two TGA enclosure    heaters a n eite ERU P RU ERE 67  Data output setting window for SDM                   see 68  Setting a TGA serial number and identification string                        69  About TGA  window  isse Pr REP nies Dar oe OE ERE det 70  The TGA TEC screen for setting laser parameters                            71  View Less View More function of the Laser Settings window           72  Tabbed Expand function of the Laser Settings window                      73  Laser display modes                      sse 74  Raw mode of laser display                    sesssssseseeeeee 75  Maximum V
62. CTS   THE CUSTOMER ASSUMES ALL RISK FROM ANY INJURY RESULTING FROM IMPROPER INSTALLATION  USE  OR  MAINTENANCE OF TRIPODS  TOWERS  OR ATTACHMENTS TO TRIPODS AND TOWERS SUCH AS SENSORS  CROSSARMS   ENCLOSURES  ANTENNAS  ETC     Table of Contents    PDF viewers  These page numbers refer to the printed version of this document  Use the  PDF reader bookmarks tab for links to specific sections     1     2  3   4    Overview    Introduction eet 1  Cautionary Statements                                                 2  Initial Inspection zu  oie i enc oro poison Eater ote 3  PEN 3  System Componenlts              eere eter t enne entered E ias 6  4 1 1 Standard Components                       sse 7  4 1 1 4 TGA Power Module                      sese 7  4 1 1 2 TGA Accessory  amp  Tool Kit                sss 8  4 1 1 3  TGA Test Int  ke         ecee aee tete rere 8  4 1 1 4 TGA Leak Check Nozzle                     sen 8  4 1 1 5 TGA SDM Cable    nennen 9  4161 6    Plastic Tubing xe ER abo eae eects 9  4 1 1 7 CATS Ethernet Crossover Cable    9  41 1 8     Serial Data Cable erred 10  4 1 1 9 TGA TEC Support Software and Operating System          10  4 1 2 Optional Components            c ceccesceeseeseeeceeceeeeeeeeeeeeeneceeenseenaes 11  ALT   LaSi ERE 11  4 1 2 2 AC Mains Power Cord                  sse 11  4 1 2 3 Power Module Mounting Brackets                         sss 12  4 1 3     Common Accessories eet eee ye e eee dene 12  4 1 3 1 TGA Reference Gas Connection    12  4 1 
63. Display Mode        isses 73  12 44     COTS  22 oh cee Recents tette thes eter a eee 81  PED Som nu C                              81  12 67 Graphen ee eR E 84  T2 Da  tdnssceiisesupeeni ne dotted i n rS 87  1 2 9  E seca  ettet epi era Vt er ets 88  7 3 Data Quip ut oeieo e indent ses RI O EE ORE 90  73 1 SOM QUEDUt  ter tit edere tite E t M Ee HESS 90  Peds    Suri q PC                          MH 91  1 3  1 2    Remarks hs ee ees A RER 91  732    TGA  Output to  PC  edipi eene etis 94  7 3 3     TGA Analog Outputs  oett retire ete sees 94   8  Troubleshooting and Maintenance                           94  8 1 Lasers and Detectors                essen 94  82   Reference C EEE ed edet e te go 95  83 Filtration and Sample Cell Cleaning                           esses 95  84 Sample Pumps    uie dccem reci 95    Table of Contents    Appendices  A  Configuring TGAs for Specific Gas Species          A 1  A  Laser Select    ice ete a Pe etes A 1  A11 LN cooled lasers                   eeeeeeeeneeeereen ene A 1  A 1 2  TE cooled Lasers irni ninen ite ree A 1  AN3     Dewar Gables  eere tee ettet A 2  ATA Changing Lasets 5  eerte arre Ee tete A 2  A2    Reference Gas ss iiie heec nr ha OPE aere ee A 4  A3    Detectors  een rient n she Eco ee iin ree iae A 6  A  Finding the Absorption Line    A 6  ASS Ai Gap Purge scun i kn Adana ae ere an edt A 7  B  Optical Alignment               cccceeeceeeeeeeeeeeeeeeeeeeeneeeeeees B 1  B 1 Optical Alignment of TGA100 and TGA100A                
64. E 4 1  Screen of TGA Windows software    Laser   The lead salt diode lasers used in the TGA100  TGA100A  and TGA200  required cryogenic cooling  These lasers were available at any wavelength  from 3 to 10 um and could be specified to detect many different gases  Most  early TGAs used liquid nitrogen to cool the laser  but some TGA100As used a  closed cycle refrigeration system  cryocooler   These lasers became  unavailable in 2012  when the only manufacturer discontinued production of  the laser  A new room temperature thermoelectrically  TE  cooled interband  cascade laser  ICL  became available in 2014  This enabled the release of the  TGA200A  These ICLs are available as an upgrade for some of Campbell  Scientific   s older TGAs  See Appendix H  Upgrading Early Generation TGAs  to TE cooled Laser  for details     Dewars   The TGA100 was introduced with a small dewar  1 5 L  that required refilling  with LN   once or twice per day  A much larger  10 4 L  dewar was introduced  in 2002 with the evolution of the TGA100  The larger dewar allowed LN   filling only twice per week  rather than daily  Along with this new larger  dewar a cryocooler system was introduced as an alternative  The 10 4 L dewar  and the cryocooler were also used for the TGA100A     The TGA200 used an even larger LN   dewar  14 5 L  to extend the refill  interval to once per week     TGA Series Trace Gas Analyzers    The TE cooled laser of the TGA200A operates without a LN2 dewar     Optical Configuration 
65. E Older units were supplied with a horizontal adjustment screw at  the detector end  but newer units use a fine pitch horizontal  adjustment screw at the laser end which makes horizontal  adjustment at the detector end unnecessary     2  Use the horizontal adjustment screw at the laser end to align the long  sample cell with the laser     3  Sight along the long sample cell to point it at the laser  which can be  viewed through the dewar window  hanging below the laser mount     CAUTION The TGA uses a Class 1M laser  Do not view the laser  directly with optical instruments     4  Use the vertical adjustment screw to align the long sample cell with  the laser  If no detector response is observed  set the vertical  adjustment near the center of its adjustment range     Appendix B  Optical Alignment    Alternately adjust the horizontal and vertical alignment screws  When  a response is observed in the sample detector  proceed to the next  section  If a response is not observed troubleshoot with the following  steps     If a detector response is not observed  it may be helpful to defocus the  optics intentionally  This will make the laser   s image on the detector  larger and easier to locate  Loosen the axial clamping screw at the  laser end  Slide the long sample cell back  away from the laser  about  5 mm from the center of its adjustment range and retighten the axial  clamping screw  Alternately adjust the horizontal and vertical  adjustment screws     If no detector response
66. EC software may be upgraded to control  temperatures through software  See Appendix I  Install Temperature Control  Upgrade  for installation details     4 1 System Components    The TGA200A consists of several components  some of which must be  supplied by the user  Some additional accessories are required to complete a  fully functioning TGA200A system and are described and illustrated in the  sections that follow     FIGURE 4 2 illustrates the main system components as well as additional  equipment needed to operate the TGA200A  The other TGA models are  similar     TGA Series Trace Gas Analyzers    TGA200 Analyzer       Datalogger    e EENAA ca                       ene axe d  Suction Hose  Sample Intake Ro M   B  _ oe al H    a  T  Reference Gas Sample Pump    FIGURE 4 2  TGA200A system components    4 1 1 Standard Components    The newest version of Campbell Scientific   s trace gas analyzers  the  TGA200A  comes with the components described in the following sections     4 1 1 1 TGA Power Module    The TGA200A power module  pn 30981  is a power module included with the  TGA200A that provides 12 Vdc and 24 Vdc to the TGA200A  It is shown in  FIGURE 4 3  The 12 Vdc is used to power the electronics of the trace gas  analyzer  while the 24 Vdc is used to power the heaters and fans that regulate  the TGA enclosure temperature  A 4 6 m  15 ft  power cable with keyed  connectors is included with the module  Corresponding keyed connectors are  found inside the power module enc
67. Flat Negative slope  Reduced Normal Increased  Mu    o  Q  2  v  z  2  T      o        Steep slope    y    FIGURE C 6    S    Symmetrical  about center    Falls off at  left edge      Effects of temperature perturbation    NEGET    C 7    Appendix C  Optimizing Laser Parameters     0 6 mA     0 3 mA    Line lock on     0 3 mA     0 6 mA    C 8              AA  JA    As shown in FIGURE C 6  improper high current adjustment makes the  absorption line appear asymmetrical  The goal of the high current adjustment  procedure is to make the absorption line symmetrical about its center     A more objective method to evaluate this symmetry is to adjust the DC current  to move the absorption line to the left and right of the center line  If the high  current is adjusted properly  the absorption line will move the same distance  left right of center for the same change     in DC current  This is illustrated in  FIGURE C 7  The left panel shows that with the high current too low   changing the DC current by   0 6 mA moves the absorption line only halfway  to the left edge of the scan  but changing the DC current by     0 6 mA moves it  all the way to the right edge of the scan  In the center panel  the absorption line  moves symmetrically  all the way to the left and right edge of the spectral scan  for the same     change in DC current           High current High current High current    correct too high    FIGURE C 7  High current adjustment procedure    To set the high current  proceed
68. GA200 and  TGA200A are located at the side of the enclosure     Other components  such as a sample air dryer  valves to switch between  multiple intakes  calibration gases  and others  may also be required  depending  on a given user   s application     TGA100 Analyzer        TGA100 PC    Fiber Optic  Cable        Reference Gas Connection Suction Hose    Sample Intake       Reference Gas Sample Pump    FIGURE 6 1  Basic components required for TGA100 operation    30    TGA Series Trace Gas Analyzers    TGA200 Analyzer       Datalogger             o          Computer        Ethernet Cable SDM Cable                               Een i aa Suction Hose  Sample Intake Bese ii  1     T  We  uem eR ERmEECMWMBE   Reference Gas Sample Pump    FIGURE 6 2  Basic components required for TGA200 and TGA200A  operation    6 1 Analyzer Installation    The TGA analyzer is housed in an insulated fiberglass enclosure  FIGURE  6 3  equipped with a temperature controller that allows the TGA to operate in a  wide variety of environmental conditions        FIGURE 6 3  TGA200A enclosure    The analyzer must be placed on a stable surface  If placed on uneven ground   wooden blocks or other supports can be used under the two pairs of rubber feet  near the ends of the enclosure     NOTE Early TGA100s had a third pair of rubber feet under the center of  the enclosure  These TGAs should be placed on blocks to lift the  TGA so these center feet are not used for support     Once the TGA is positioned  loos
69. GURE C 7     e Ifthe high current is too low  it will take more steps to move  the line to the left edge than to move the line to the right  edge  The absorption line may become narrower as it  approaches the left edge and wider as it approaches the right  edge  Additionally  if line lock is started with the absorption  line at the left or right edge  the absorption line may move  relatively slowly to the center  If this is observed  increase  the high current and repeat the steps above     e Ifthe high current is set correctly  it will take the same  change in DC current to move the absorption line to the left  or right edge  and this will be approximately equal to the  modulation current   the width of the absorption line will not  change noticeably as it is moved from left edge to right edge   and when line lock is started with the absorption line at the  left or right edge  the absorption line will jump quickly to the  center     e If it takes fewer steps to move the line to the left edge than to  move the line to the right edge  or if the absorption line  becomes wider as it approaches the left edge  or if starting    Appendix C  Optimizing Laser Parameters    C 10    line lock with the absorption line at either edge causes it to  overshoot the center  this means the high current is too high   Decrease the High current and repeat the test above     Usually these criteria will give a clear indication of the correct value for the  High current  However  sometimes other cri
70. Input       TGA Identification  TGA S N  TGA description    Serial Numbers    Laser S N  Dewar S N    Sample detector S N  Reference detector S N    Site name description    1113    Lab TGA200    1476 21 8    8745  8770A    CSI TGA Lab    FIGURE 7 24  Setting a TGA serial number and identification string    About TGA  The Settings  gt  Other  gt  About TGA window is shown in FIGURE 7 25  The   window gives the version of the PC software and also documents the version of  the firmware if a TGA is connected        69    TGA Series Trace Gas Analyzers         TGA TEC Parameter Settings  G   Laser    Tena bE CAMPBELL         SCIENTIFIC    Line Lock          Other     Detectors    mae     amugu     PreAmp  B  Calculations  Concentration       x   TGA TEC Interface Software    TGA Temperatures Version 1 0 0 37    Data Output Copyright     2014 Campbell Scientific  Inc   Serial Numbers    TGA TEC Firmware    Version  TGA TEC  1 0 21   OS Date  08 08 14 11 21 12   ID string  Lab TGA200   Copyright    2014 Campbell Scientific  Inc     http   www w campbellsci com                   FIGURE 7 25  About TGA window    7 2 4 Laser Window  The Laser window is used to display reference and sample detector signals and  to modify laser parameters     The value of a parameter may be adjusted by several methods     e Manually entering the new value in the box with the keyboard number  keys    e Clicking the up or down arrow next to the box    e Click in the box and use the up and down arrow keys 
71. K  Other     Detectors ro       Temperature CH4  PreAmp  B  Calculations s  1 DC current 400 00    Concentration     Analog Input Mod current 0 60   tn     Other Zero current 320 00  TGA Temperatures High current 30 00 2  Data Output  Serial Numbers  About TGA  Maximum Laser Ratings  Maximum laser current  mA  600 00                FIGURE 7 10  Laser current parameter settings for TGA Windows    55    TGA Series Trace Gas Analyzers    56          F  PEERS  TGA Parameter Settings mm       Laser    Temperature    Current  V  Laser on  Line Lock  p men  cHa    B  Detectors    Temperature  PreAmp  B Calculations  Concentration  Analog Input Zero current 50 00     Other  TGA Temperatures  Data Output  Serial Numbers       About TGA Maximum Laser Ratings       DC current 88 15  Mod current 1 30    HE    High current 0 00          Maximum laser current  mA  100 00                FIGURE 7 11  Laser current parameter settings for TGA TEC    Line Lock   The Settings  gt  Laser  gt  Line Lock window sets limits used to temporarily  disable the line lock function during an error condition  The line lock function  locks the absorption line to the center of the spectral scan  by automatically  adjusting the DC current  Line locking must be active for normal operation   see Section 7 2 4  Laser Window  for information on making line lock active   and see Section 7 2 2  Status Window  for details on viewing the TGA line lock  status   The TGA monitors three conditions     e sample cell press
72. New New New  CPU  upgradeable   Software Transputer DOS TGA Windows TGA TEC   Lead salt Lead salt Lead salt Interband  Laser   Cascade   Cooling options LN  LN  or Cryocooler LN Thermoelectric  Dewar capacity  L    1 5 10 4 14 5 None   Beamsplitter at Beamsplitter at Beamsplitter at Beamsplitter at  Optical detector end  detector end laser end  Long laser end  configuration sample and    reference cells        Absorption cells    Long sample  cell short reference  cell    Long sample  cell short reference  cell    Long sample and  reference cells    Long sample  and reference  cells           Temperature  control       Fans only   TGAHEAT optional  starting 2002        TGAHEAT  included       Software       Software          TGA Series Trace Gas Analyzers    Before proceeding  please study    e Section 2  Cautionary Statements  e Section 3  Initial Inspection  e Section 6  Installation    Operational instructions critical to preserving accurate measurements of the  system are found throughout this manual  Before proceeding  please study this  manual  Several other user manuals provide additional information and should  be consulted before using the TGA  These include     CR6 Measurement and Control Datalogger  CR1000 Measurement and Control Datalogger  CR3000 Micrologger   CR5000 Measurement and Control System    all available at www campbellsci com    2  Cautionary Statements  e DANGER     O    All Campbell Scientific TGAs use Class 1M lasers  These lasers  are safe under al
73. TGA Series  Trace Gas Analyzers    Revision  10 14          Copyright    1992 2014  Campbell Scientific  Inc     IVONVIN NOILONALSNI    Limited Warranty       Products manufactured by CSI are warranted by CSI to be free from defects in  materials and workmanship under normal use and service for twelve months  from the date of shipment unless otherwise specified in the corresponding  product manual   Product manuals are available for review online at  www campbellsci com   Products not manufactured by CSI  but that are resold  by CSI  are warranted only to the limits extended by the original manufacturer   Batteries  fine wire thermocouples  desiccant  and other consumables have no  warranty  CSI   s obligation under this warranty is limited to repairing or  replacing  at CSI   s option  defective Products  which shall be the sole and  exclusive remedy under this warranty  The Customer assumes all costs of  removing  reinstalling  and shipping defective Products to CSI  CSI will return  such Products by surface carrier prepaid within the continental United States of  America  To all other locations  CSI will return such Products best way CIP   port of entry  per Incoterms    2010  This warranty shall not apply to any  Products which have been subjected to modification  misuse  neglect  improper  service  accidents of nature  or shipping damage  This warranty is in lieu of all  other warranties  expressed or implied  The warranty for installation services  performed by CSI such
74. TGA TEC  with the exception  that the units for the laser temperature are different and the minimum current  step size is different     The Interactive mode  selected at the bottom of the Laser Line Find screen   maps the laser at one temperature  but then allows the user to select an  absorption line by first clicking on the key to select which ramp to set  and then  clicking on the absorption line in the graph  Alternately  the vertical colored  band that represents the spectral scan may simply be dragged to the desired  absorption line     It is helpful to display the Laser Settings window beside the Laser Line Find  window  FIGURE 7 37  while making these adjustments  If preferred  the  Laser Settings window  as described in Section 7 2 4  Laser Window  may be  used to make the adjustments and view the changes interactively in the Laser  Line Find window  An example is shown in FIGURE 7 38 for a CO   isotope  laser     81    TGA Series Trace Gas Analyzers    If the two windows do not agree well  try repeating the laser map with a  smaller step size for the DC Current        iLaser Line Find    Step Size     X ampbellScivT 35D ataFlles mmad yyyyhhrimmap     Advanced kile Options        FIGURE 7 37  Laser line find window    82    TGA Series Trace Gas Analyzers    9 Laser Find  350 00 to 450 00 ma E    vp sx   o  o1 sll       Laser Find  99 70    Co2        Reference Voltage  mv           FIGURE 7 38  Interactive Laser Find window for a CO2 isotope laser     The non interac
75. able from  Campbell Scientific for this purpose  see Section 4 1 3 1  TGA Reference Gas  Connection  for more details  This assembly includes a flow meter  needle  valve  and 6 2 m  20 ft  of tubing with Swagelok   fittings to connect to the  TGA  The outlet of the regulator must have a 1 4 in Swagelok   fitting to  attach this assembly     A tank of reference gas 5 7 m   200 f     will last approximately one year at a  continuous flow of 10 ml min     The most commonly measured trace gases  methane  nitrous oxide  and CO   isotopes  as well as many other gases   have strong absorption lines at  wavelengths from 3 to 5 um  and can be measured with the standard TE cooled  detectors  Some gases  such as ammonia  water isotopes  or the combination of  N20 and CH  must be measured at longer wavelengths and require the optional  LN   cooled detectors  The TGA100 and TGA100A were supplied with custom  LN   cooled detectors which required daily filling  The TGA200 had optional  LN   cooled detectors with a much larger dewar that required filling twice  weekly     The TGA200A always uses TE cooled detectors because its TE cooled lasers  cannot reach the longer wavelengths that require LN   cooled detectors     A 4 Finding the Absorption Line    A 6    The spectral scan of the TGA must be locked onto a selected absorption line   When the TGA is restarted  it will perform an automated sequence to  reestablish this line lock by waiting until the laser temperature has stabilized   and th
76. aded  to use the thermistor probes  see Appendix I   nstall Temperature Control  Upgrade   The settings for the Channel 2 tab are shown in FIGURE 7 19  The  Channel 3 tab is similar  but the Name is TGA Temp 2     63    TGA Series Trace Gas Analyzers    64          Channel 2   Channel 3   Pressure       Name TGA Temp 1    5 PRT sensor  TGA100A       Themistor  TGA200     Bie 222109 Input      Manual scale  offset       TGA Temperatures Scale 0 00  Data Output m   m    Serial Numbers ur  i  About TGA                      FIGURE 7 19  Channel 2 settings for a TGA with thermistor probe    Channel 4 measures sample cell pressure  Each TGA pressure sensor is  individually calibrated  to determine appropriate values for the Zero output   V  and the Full range output  V   These are recorded in documents supplied  with the TGA  Typical values are shown in FIGURE 7 20     TGA Series Trace Gas Analyzers          Zero output  V  0 230  Full ange output  V     47008  Full ange pressure  mb  1013 00                Measured pressure 54 14 mb                   FIGURE 7 20  Typical pressure value inputs  values are specific to  each TGA     Isotopes   The  sotopes parameters are the same for both versions of the software  This  window is hidden if the number of ramps is set to 1  An example of typical  values for calculating 5   C is shown in FIGURE 7 21     65    TGA Series Trace Gas Analyzers    66    7 2 3 4 Other       TGA TEC Parameter Settings      Laser  Temperature  Current   Line Lo
77. adjustment  screw    Tighten only when  transporting    Horizontal lock  screw    Axial lock screw       FIGURE B 2  Alignment hardware of detector end of TGA100 and  TGA100A    2  Loosen the horizontal and vertical clamping screws at the laser end   refer to FIGURE B 3      Horizontal  adjustment  screw    Horizontal  clamping  screws    Appendix B  Optical Alignment    Vertical Axial  adjustment clamping screw  screw    Vertical  clamping screw       Vertical  clamping screw    FIGURE B 3  Alignment hardware of laser end of TGA100 and    TGA100A    If the TGA is equipped with an iris in front of the focusing lens  open  it fully  this is recommended for normal operation      Start the TGA program and make sure the laser and detector  parameters are set appropriately for the laser     Display the reference detector signal and the sample detector signal in  a graph  The goal of the alignment procedure is to maximize these  signals     Edit the graph options to move one of the traces to the right axis  Set  the minimum value of each Y axis to zero and let the maximum of  each Y axis scale automatically     Set the reference and sample detector gains to zero  This will disable  automatic gain and offset adjustment which can cause confusion  during the alignment process when active     Set the detector temperatures as needed to avoid saturation  This  adjustment may need to be repeated during the alignment process if  the signal level increases too much  It is not important to h
78. after the zero and high current  phases of the scan  The Omitted count parameter specifies how many scan  points to omit  The parameter may have a set value from 4 to 20 counts  where  each count represents a 20 us interval  to give a total duration of omitted data  of 80 to 400 us  To set this parameter  look at the reference detector  transmittance in Magnified mode  The leftmost portion of the graph is at a  different color to mark the omitted data counts  meaning that those data are not  used to calculate concentration  This parameter is found on the Settings  gt   Laser  gt  Other window  Increase the omitted data counts to move this line to  the right  decrease it to move it to the left  The reference transmittance should  be flat  or rising slightly  as it approaches the left edge  as shown on the right  in FIGURE C 8  Set the omitted data counts to avoid the use of data that drop  into the    ghost    line  as shown on the left in FIGURE C 8  When in doubt it is  usually better to omit a few extra counts     V Too few omit counts Y Correct omit counts        o  D  D      a  O    Drops at  left edge  Flat at    FIGURE C 8  Adjustment of omitted data counts    Adjusting this parameter usually makes little difference in the TGA    s  performance  so when in doubt  this parameter should simply be left at the  default which is 20 counts  For multiple ramp mode  the omitted count  parameter is common to all ramps  and it should generally be set to its default     C 5 Modula
79. al System    TGA100 and TGA100A   The TGA100 and TGA100A optical configuration is shown schematically in  FIGURE 4 25  Infrared radiation from the laser is collimated and passed  through a 1 5 m  4 9 ft  long sample cell  where it is absorbed proportional to  the concentration of the target gas  A beamsplitter directs most of the energy  through a focusing lens and a short sample cell to the sample detector  and  reflects a portion of the beam through a second focusing lens and a short  reference cell to the reference detector  A reference gas of known  concentration flows through the reference cell  Reflective surfaces are  minimized to reduce errors caused by Fabry Perot interference     This simple optical design avoids the alignment problems associated with  multiple path absorption cells     23    TGA Series Trace Gas Analyzers       Reference  defector  H  To pump     i              Reference gas in Dewar     i ZA n    E Sample cell f CN  Sample a  mm Laser    detector            I l   ji      UPC To pump Sample in    FIGURE 4 25  TGA100 and TGA100A optical configuration    TGA200 and TGA200A   The TGA200 and TGA200A optical system is shown schematically in  FIGURE 4 26  The laser is partially collimated by a first lens  Most of the  energy then passes through the beamsplitter and a collimating lens  and into the  sample cell  The beamsplitter reflects a portion of the beam to an alignment  mirror and then through a second collimating lens into the reference cell  The  
80. alues to store     Advanced     Set file names and the location for the files     File Names     Set parameters for the file naming the folder in which to  store the files     Sample Rates     Set the rates at which data are collected and stored     89    TGA Series Trace Gas Analyzers    Other     look at the event log     Event Log     View or save a log of the status of the communication with    the TGA  An example of the messages in the log file is shown in  FIGURE 7 45           B  Files    i   File Lists     Data File    i    Housekeep File      Selection    i  Misc  B  Advanced    p File Names      Sample Rates      Other    Event Log                   Connected To  Connected to 192 168 76 31  Lab TGA200     Time J Messaae  09 17 14 08 29 26 Parameter File  C  CampbellSci  TGA ParameterFiles t gaConfiguration prm The  09 17 14 08 29 45 Connected to 192 168 76 31 3000  Lab TGA200   09 17 14 08 46 06 TGA starting up   Waiting for laser temperature to stabilize before establishing line  09 17 14 08 48 20 TGA laser stabilized  Line Lock tumed on  Line found   09 17 14 09 08 46 Data Collection Toggled On                  Clear All J   Save events to File               FIGURE 7 45  Example of log file messages    7 3 Data Output    The TGA may be configured to output data to the user s PC  or for analog  outputs  but normally data are sent to a datalogger via an SDM cable     7 3 1 SDM Output    90    SDM is a Campbell Scientific communication protocol that allows  synchroni
81. amp     C 1    Appendix C  Optimizing Laser Parameters    Open the Laser Settings window to view the absorption line  set the  laser temperature and view the DC current  reference detector  transmittance  and the sample detector signal     Open either a Data window or graph to view the concentration noise     Enable the line locking function and the automatic detector offset and  gain adjustment function     Record the laser operating temperature  the laser DC current  and the  reference detector   s percent transmittance at the center of the ramp   displayed at the top of the reference detector transmittance   The  sample detector signal and the concentration noise may also be useful  to note  although these are not absolutely necessary     10  Record these values in a notebook     11  TABLE C 1 gives an example of the process described in steps 6     8        TABLE C 1  Example Laser Temperature Optimization Data                               Laser Laser DC Reference Concentration Sample Signal   Temperature  K    Current  mA    Transmittance     Noise  ppb   mV   103 7 482 6 65 9 7 30 0  103 9 477 4 66 2 8 27 6  104 1 472 1 66 4 9 24 9  104 3 466 6 67 3 10 21 9  104 5 460 8 69 1 13 18 3  104 7 454 9 72 0 20 14 4  104 9 448 7 78 2 45 10 0  105 1 442 2 90 5 350 4 9       Transmittance and noise much worse   try going down                103 5 487 7 65 8 7 32 5  103 3 492 6 67 7 6 34 5  103 1 497 5 77 8 15 36 3  102 9 502 9 89 5 85 38 7       Transmittance and noise worse aga
82. ascade laser  TE cooled ICL  became available that is operable at ambient  temperature without additional cooling  Campbell Scientifics   s TGA200A  instrument was released shortly after the TE cooled laser became available   Some of the older TGAs can be upgraded with this laser  See Appendix H   Upgrading Early Generation TGAs to TE cooled Laser  for details     All lasers used in the TGA are infrared diode lasers with wavelengths longer  than 3 microns and a divergent exit beam  Optical power for lead salt lasers is  specified as  gt  0 1 mW  with typical powers up to approximately 1 mW  ICL  optical power is  gt  1 mW  typically 2 to 4 mW  For both types of lasers  the  power varies from one laser to another  Consult the vendor data sheet for the  individual laser for more information  The safety classification for both types  of lasers is Class 1M     The TGA uses a Class 1M laser  Do not view the laser    directly with optical instruments     The TGA100 came with a small LN   laser dewar  1 5 L  that required refilling  with LN2 once or twice per day  A much larger  10 4 L  LN laser dewar was  introduced in the evolution of the TGA100  in 2002  This dewar required LN   filling twice per week  Along with this new larger dewar  a cryocooler system  was offered as an alternative  The TGA200 used an even larger LN  dewar   14 5 L  which extended the refill interval to once per week  The TEC ICL of  the TGA200A requires no LN   dewar  This laser is housed in a sealed package
83. asers  it is possible to have both programs installed on a PC  The shortcut  icons to start these two programs are distinguished by the color and the text  as  shown in FIGURE 6 14  Installing the software also starts a communication  process that runs in the background and registers the software for later  installation or updates         Lm  T TG    uw MAC     TGA TGA TEC       FIGURE 6 14  PC shortcut icons for TGA Windows  left  and TGA TEC   right     6 2 2 Updating TGA Operating System    Run the installation program  for example TGA_Firmware2 1 exe  in the TGA  Operating System folder on the installation disk  This will extract the TGA OS  file  for example TGA_2   0bj  and the TGA program tga cr9 and place them  in the default directory C  Campbellsci Lib OperatingSystems      1  As shown in FIGURE 6 15  connect the PCs COM port to the RS232  port on the TGA s CPU module with a serial cable  pn 20730 is  supplied for this purpose      AlddNs HIMOA   SAMO  0106    YaMog       FIGURE 6 15  Serial cable connecting PC to RS232 port of TGA CPU  module    TGA Series Trace Gas Analyzers    2  Run the Campbell Scientific Device Configuration Utility  part of  LoggerNet  or available for free from Campbell Scientific      3  Select device type TGA100A TGA200 and follow the directions to  connect to the TGA     4  Select the Send OS tab and follow the directions  FIGURE 6 16  to  send the new operating system to the TGA        Device Configuration Utility 2 10 Beta    File Backu
84. at of  the PD200T  The PD625 s inlet filter  tubing connections and purge flow  meter range are also smaller than for the PD200T  The PD625 is normally  used in the two dryer configuration  so it does not include the sample flow  meter or sample needle valve included in the PD200T  The PD625 is shown in    FIGURE 4 20      gt   gt     FIGURE 4 20  PD625 air sample dryer        4 1 4 7 TGA Rotameter    The TGA rotameter  pn 19541  measures air sample flow in high flow  applications such as eddy covariance  It is temporarily attached to the sample  inlet while adjusting the needle valves that control the sample and bypass  flows  In FIGURE 4 21  the front of the rotameter is shown on the left and the  connectors at the back are shown on the right     21    TGA Series Trace Gas Analyzers    22    4 1 6 2 Filter Element            napi    nuez    TAE     ard b    o 3         FIGURE 4 21  TGA rotameter    4 1 5 Support Software    The PC support software and the TGA firmware are supplied on a CD with the  shipment  pn 30723     4 1 6 Replacement Parts    Replacement parts for the TGA200A are described in the following sections     4 1 6 1 TGA Heated Intake Filter    This heated filter  pn 18058  may be used to replace the sub assembly of the  TGA heated intake filter  pn 18072  see Section 4 1 4 2  TGA Heated Intake  Filter  amp  Orifice  when that filter becomes plugged  This spare heated intake  filter shown in FIGURE 4 22  is recommended when the air sample intake is  difficult
85. ave large    Appendix B  Optical Alignment    signals during alignment  When in doubt  set the detector  temperatures relatively high for relatively low signals     9  Ifno sample detector signal can be seen  perform the initial alignment   see Appendix B 2 2  Initial Alignment      10  Once a signal can be observed on the sample detector  adjust the  horizontal and vertical alignment  Appendix B 2 3  Horizontal and  Vertical Alignment      11  Adjust the focus  Appendix B 1 3  Focus Adjustment      12  When the focus and the horizontal and vertical alignment have been  optimized  tighten the horizontal  vertical  and axial clamping screws     NOTE Remember to leave the transport lock and axial lock screws loose     13  Make sure the reference detector is coaligned with the sample detector   Appendix B 1 4  Reference Detector Coalignment      B 1 1 Initial Alignment    If the optical system is significantly misaligned  there may be no observable  detector response  This initial alignment procedure will help to align the  system well enough to see a response  As soon as a detector response is  observed  the system is ready for the horizontal and vertical alignment  procedure described in Appendix B 1 2  Horizontal and Vertical Alignment     1  Ifthe TGA is equipped with a horizontal adjustment screw at the  detector end  loosen the horizontal lock screw  adjust the horizontal  position to near the center of its adjustment range  and retighten the  horizontal lock screw     NOT
86. ay represents a random lag  jitter  that cannot be removed by EC post   processing algorithms  However  the jitter is small enough  2 to 6 ms  to be  insignificant  even for EC applications        TABLE F 3  Summary of TGA Update Times                            Sampling Mode Filter Ramps Maximum Delay  ms   SDM Moving Avg 1  2  or 3 2  EC 1 6  2 4  3 6  Analog Moving Avg or EC 1 6  4  3 6  PC Moving Avg or EC 1  2  or 3 2             F 5 Sample Cell Residence Time    The frequency response of the TGA is ultimately determined by the time it  takes for a sample to flow through the sample cell  This residence time  depends on the volume of the sample cell and the actual flow rate as given by  the equation    t v q  where  t   residence time  s     v   sample cell volume  L     q     actual flow rate  L s     F 5       Appendix F  TGA Frequency Response    F 6    A typical example is using an RB0021 L sample pump  see Section 4 1 4 4   Sample Vacuum Pump  for eddy covariance  This pump has a capacity of 18  slpm at 50 mb  Assuming 3 slpm is used to purge a PD200T dryer  see Section  4 1 4 6  Sample Air Dryer   leaves 15 slpm for the TGA  Converting standard  flow to actual flow and converting units from minutes to seconds gives     50 A60    This gives a sample cell residence time for the various TGA models as shown  in TABLE F 4        TABLE F 4  Sample Cell Residence Time as a Function of Sample  Cell Volume and TGA Model                      Sample Cell Volume Residence time 
87. be 1 8  tube 1 4    Swagelok    1 4    Swagelok    e Purge 3 8    Swagelok     1 2    Swagelok     3 8    Swagelok     1 2    Swagelok    Sample pressure drop  at standard   mb per 1  T 4 13 10 5  temperature and pressure  min      0 5  zz o  l  Flow rate for    15   C dewpoint 1 min  0   C dewpoint  0 5 2 16  Sample volume  internal volume  af deyerbibine  ml 1 4 6 7 10 80  Purge flow meter range   min     NA NA 0to 5 Oto 10  PDIT    The PDIT dryer  FIGURE 4 17  uses a 6 ft length of 0 086 in ID Nafion   tubing  similar to Permapure MD110 72  www permapure com   It will dry 0 5  lpm to    15   C dewpoint     19       TGA Series Trace Gas Analyzers    20       FIGURE 4 17  PD1T air sample dryer    The sample inlet and outlet connections are 1 18 in OD stainless steel tubes   These SS tubes connect directly to the Nafion   tube inside the dryer shell  This  design completely eliminates dead volume in the sample flow     The purge connections are 1 2 in Swagelok    and the dryer shell is 0 5 in OD  Synflex 1300 tubing  This large size minimizes pressure drop in the purge  flow to allow the dryer to be purged with the output of the TGA  reflux mode      The dryer shell is flexible Synflex 1300 tubing to allow it to be integrated  easily into the sampling system  typically between a multiport sampling  manifold and the TGA inlet     PDIT 1 5   The PDIT 1 5  shown in FIGURE 4 18  is similar to the PDIT  but is designed  for lower capacity at lower cost  It is designed to remove en
88. ber of ramps parameter  the TGA will return zero for the ramp  B and or C values     See TABLE 7 3 for complete information        TABLE 7 3  TGA Instruction                                                    Name Description DataList   ScanMode   ConcA Trace gas concentration measured in Ramp 1 1  A  ppmv    ConcB Trace gas concentration measured in ramp B 1 2   ppmv    ConcC Trace gas concentration measured in ramp C 1 3   ppmv    TGAStatus Status flags  see TABLE 7 4  1 1   TGAPressure Sample cell pressure  mb  2 1   LaserTemp Laser temperature  K   2 1   DCCurrentA Laser DC current for ramp A  mA  2 1   DCCurrentB Laser DC current for ramp B  mA  2 2   DCCurrentC Laser DC current for ramp C  mA  2 3   TGAAnalogl TGA Analog Channel 1 voltage  V  3 1   TGATempl Mi AR inside TGA enclosure  detector 3 1  end    C    TGATemp2 Tepe inside TGA enclosure  laser 3 1  end    C   Voltage applied to the laser heater to   LaserHeater maintain the laser at the specified 4 1  temperature  V       Reference detector signal at the center of the  RerDer analy spectral scan for ramp A  mV  1                     TGA Series Trace Gas Analyzers       TABLE 7 3  TGA Instruction                                                                Name Description DataList   ScanMode  RefDetSignalB Reference detector signal at the center of the 4 2  spectral scan for ramp B  mV   RefDetSignalC Reference detector signal at the center of the 4 3  spectral scan for ramp C  mV   Reference detector transmit
89. bly  TE cooled  lasers  or to change which laser is active  multiple LN2 cooled lasers in a  dewar         CAUTION    DANGER    NOTE    WARNING    WARNING    WARNING    Appendix A  Configuring TGAs for Specific Gas Species    The TGA laser can be damaged by operating at a    temperature or current outside maximum limits  which are  unique to each laser  Follow the steps carefully to avoid  damaging the laser     The TGA uses a Class 1M laser  Do not view the laser    directly with optical instruments     1  Turn the analyzer electronics off   2  Disconnect the laser cable as appropriate for the laser     LN cooled laser  Disconnect the laser cable from the electronics and the dewar     TE cooled laser  Disconnect the laser cable from the electronics  the laser cable is  permanently attached to the TE cooled laser assembly      If switching between position 1 and position 2 lasers  skip step 3     3  Ifthe dewar must be rotated to select a laser in the optional second  laser mount  or if the dewar or TE cooled laser assembly is to be  exchanged  remove the four mounting screws  rotate exchange the  dewar or TE cooled laser assembly  and reinstall and tighten the  mounting screws     Do not connect the dewar cable at this time     4  Startthe TGA software on your PC  Read an appropriate parameter  file for the new laser     To avoid damaging the laser  ensure that the    parameters are valid for the new laser  In particular   verify the laser maximum temperature and laser  maxi
90. center the three clamping screws in their slots   Align the reference detector to the sample detector by loosening the three  screws that attach the reference detector holder to the beamsplitter block   turning the alignment cams to maximize the reference signal  and then  retightening the mounting screws     The optical alignment is now complete     B 2 Optical Alignment of TGA200 and TGA200A    Like its predecessors  the TGA200 and TGA200A also have a simple  robust  optical design making it similarly easy to adjust and maintain optical  alignment  The optical system  illustrated in FIGURE B 4  includes the laser   collimating lenses in front of the laser  a beamsplitter to reflect some of the  laser   s energy through the reference cell  and three mirrors for aligning the  beam through the reference cell and onto the sample and reference detectors     Appendix B  Optical Alignment    Sample  detector    Detector  alignment  mirrors    Reference  alignment    mirror Reference    detector       FIGURE B 4  TGA200 and TGA200A optical layout    The TGA optical alignment does not change during normal operation  but it  should be checked after transport  FIGURE B 5 illustrates the horizontal   vertical  and reference mirror tip tilt adjustment knobs  shown with the laser    removed for clarity         Sample  Horizontal         Sample  Vertical         Reference  Mirror Tip Tilt        FIGURE B 5  Alignment hardware of laser end of TGA200 variants    B 8    Appendix B  Optical A
91. ck  V  Calculate isotope ratios  Other     Detectors Isotope Ratio  Temperature  PreAmp   oc     Del 13C standard isotope ratio 0 01117970  Concentration  Analog Input  Isotopes  B  Other  TGA Temperatures  Data Output  Serial Numbers  About TGA          Isotope ratio definition 13C 12C Y                   FIGURE 7 21  Typical values for calculating 613C on a TGA set with  multiple ramps    TGA Temperatures   The Settings  gt  Other  gt  TGA Temperatures window is shown in FIGURE 7 22   To enable the heaters to maintain the TGA at a steady temperature  check the  Control TGA temperatures box and set Control Parameters as shown in  FIGURE 7 22  Set the two temperature setpoints as needed to maintain a  steady temperature  Usually  this will be 5 to 10   C above the highest likely  ambient temperature  Normally  the two setpoints are set to the same value   However  if one of the two heaters generally works harder than the other   higher duty cycle   adjust the setpoints to slightly different temperatures   usually within one  C  to balance the two heaters     TGA Series Trace Gas Analyzers             g Laser     Temperature  l   Current  V  Control TGA temperatures     Line Lock Control Parameters  i i  Other      Detectors Pulse period 0 10  Temperature  PreAmp Control coefficients  P  2 00     Calculations Control coefficients  I  0 01  Concentration Control coefficients  D    20 00  i  Analog Input     3  Other    TGA Temperature 1    Data Output e    23 36     Serial Numb
92. cm  83 in  211 cm  83 in  211 em  83 in  211 cm  83 in   Width 47 cm  18 5 in  47 cm  18 5 in  47 cm  18 5 in  47 cm  18 5 in   Height 55 cm  21 5 in  55 cm  21 5 in  55cm  21 5 in  55 cm  21 5 in   Weight  74 5 kg  164 Ib  88 9 kg  195 5 Ib  78 6 kg  173 Ib  62 8 kg  138 5 Ib     Sample path length 153 08 cm  60 27 in    153 08 cm  60 27 in    146 6 cm  57 72 in    146 4 cm  57 64 in   Reference path length   4 52 cm  1 78 in  4 52 cm  1 78 in  146 6 cm  57 72 in    146 4 cm  57 64 in   Sample cell volume 480 ml 480 ml 420 ml 200 ml  Operating temperature      20 to 45   C    20 to 45   C    20 to 45   C    20 to 45  C                      Weight of the TGA100A and TGA200 is shown for most common configuration  LN2  laser dewar and TE cooled detectors   PDoes not include the weight of the power module  pn 30981  which is 5 4 kg  12 0 Ib   with the accompanying power cable    5 3 Power Requirements  Analyzer  LN2 cooled laser      Analyzer  TE cooled laser      Heater     90 to 264 Vac  47 to 63 Hz  42 W  max     24 W  typical     90 to 264 Vac  47 to 63 Hz  34 W  max     22 W  typical     90 to 264 Vac  47 to 63 Hz  150 W  max     50 W  typical     29    TGA Series Trace Gas Analyzers    6  Installation    The basic components required to operate a TGA are shown in FIGURE 6 1  and FIGURE 6 2  The systems are very similar with the exception of the  connections  The connections are located at the end of the enclosure in the  TGA100 and TGA100A  whereas the connections of the T
93. d  The  assembly is shown in FIGURE 4 8  For applications requiring a purge gas  the  assembly may be ordered for delivering zero gas to the purge inlet on the TGA     TGA Series Trace Gas Analyzers       FIGURE 4 8  TGA reference gas connection    4 1 3 2 TGA Insulated Enclosure Cover    The insulated TGA enclosure cover  pn 16599  is recommended when the  TGA is operated in the field without additional shelter  The cover has a  rainproof  white exterior to reflect the sun   s heat  and additional insulation to  dampen diurnal temperature fluctuations  The cover is shown installed on a  TGA in FIGURE 4 9     The cover fits over the TGA attaching with integral hook and loop fasteners   The 16599 cover was introduced in 2002 for use with the TGA100A  These  early covers had one flap over the access hole in the top of the TGA enclosure  to allow easy refilling of the liquid nitrogen cooled laser dewar  A second flap  was added in 2008 with the introduction of the TGA200  to allow access to the  optional LN2 cooled detectors  Both flaps are omitted for the TGA200A     FIGURE 4 9  TGA with insulated cover enclosure    To install the TGA cover  place the bottom  uninsulated  piece under the TGA   The bottom is the loops portion of the hook and loop fastener strip around the  periphery  Orient the bottom piece so the periphery folds up  with the strips on  the outside  Place the top  insulated  piece over the analyzer  oriented with the  flap s  over the deck plate s  to allow the 
94. d detectors  used for long  wavelength operation   cool the detectors with liquid nitrogen     If the TGA is equipped with the standard TE cooled detectors   they will be cooled automatically     3  Start the sample vacuum pump     4  Turn on the reference gas  A regulator pressure of 0 psig and a flow  rate of approximately 10 ml min are recommended     10     11     12     TGA Series Trace Gas Analyzers    Turn on the purge gas if required  high accuracy applications   A  flow rate of approximately 10 ml min is recommended     Power up the TGA analyzer  It is supplied with universal input power  supplies  Connect the power supplies to AC power  90 to 264 Vac  47  to 63 Hz      On a user supplied PC  start the TGA program  and connect to the  analyzer and check its status  see Section 7 2 2  Status Window      Verify the TGA pressure is consistent with the previous operation of  the TGA  The sample pump capacity  the total flow at the pump  and  pressure control by the sampling system determine the pressure  If the  pressure has changed  it may indicate a problem in the plumbing     Verify the correct absorption line is being scanned  Usually it is  sufficient to simply verify the DC current  reference detector  transmittance  concentration and concentration noise are consistent  with normal operation  see Section 7 1 2  Routine System Checks   If  any of these values have changed significantly from their normal  values  see Appendix A 4  Finding the Absorption Line     Veri
95. d only through the Device  Configuration Utility  These settings are normally retained when the TGA  operating system is updated  but it is good practice to verify them  Connect to  the TGA with the Device Configuration Utility and select the Settings Editor    TGA Series Trace Gas Analyzers    tab to review or edit these settings  The IP address  subnet mask  and gateway  should be set as directed by your information technology support group  Set  the IP port for the TGA to 3000        Device Configuration Utility 2 10 Beta  File Backup Options Help  M  Settings Editor   Send OS   Terminal   Logger Control       Device Type  M       Current Setting   P Address       Phone Modem  Radio IP Address  h92 168 76 31    Gateway  192 168 2 19      Netmask    255 255 240 0  IP Port    soo       Serial Number  1113         Y    Identification string   Lab TGA200       MAC Address  Ox00D02C030562       TGA100A TGA200    Unknowr    Wireless Sensor  IP Address    Communication Port l    Specifies the IP address for the TGA100A TGA200     Jse IP Co    Baud Rate    Appl Cance Factory Defaults Read File       FIGURE 6 18  Device Configuration utility Settings Editor tab    NOTE The TGA requires a static IP address     Some local area networks are set up to require the network administrator to  know the MAC address of devices connected to the network  The Settings  Editor tab also displays the TGAs MAC address  but does not allow it to be  edited     The MAC address of each TGA is set at the
96. dard deviation of the concentration sampled at 10 Hz  and calculated over a relatively short time  10 s   The TGA multiple scan mode  can be used to measure suitable pairs of gases  Typical performance for isotope  ratio measurements is given in delta notation  For example  the 6  C for CO   is  given by        R  c     E  4   VPDB  where  R    ratio of the isotopolog concentrations measured by the    TGA   CO   CO    Ryppg   the standard isotope ratio  C 12C     8  C is reported in parts per thousand  per mil or  o        ii Allan  D W   1966     Statistics of atomic frequency standards   Proc  IEEE  54  221   231    TGA Series Trace Gas Analyzers                         TABLE 5 1  Typical Measurement Noise    Part Number Description Chemical Formula   Typical Noise    Units  30478 Nitrous Oxide N20 1 5 nmol mol   30477 Methane CH  7 0 nmol mol   31121 Nitrous Oxide and N20 1 8 nmol mol    Carbon Dioxide    CO  0 3 umol mol    31119 Carbon Dioxide CO   0 15 umol mol   and 81  C 83C 0 5  o   30877 Carbon Dioxide  CO  0 5 umol mol   DC  and 6  O LC 2 0  o  5180 2 0  o                          Preliminary  specifications are subject to change without notice  ballan deviation with 100 ms averaging time   Based on the  C 60 60 isotopolog                                     5 2 Physical Specifications  The physical specifications of all of the TGA variants are summarized in  TABLE 5 2   TABLE 5 2  Physical Specifications of TGA Variants  TGA100 TGA100A TGA200 TGA200A   Length 211 
97. ded by  folding about the center line  averaging the data with a reversed copy of the  data   This process forces the data to appear symmetrical about the center of  the spectral scan  the center vertical dashed line       amp   Laser Settings  View Less Expand Folded   Colors          17 00   17 00   C       FIGURE 7 34  Folded mode of laser display    79    TGA Series Trace Gas Analyzers    Absorbance  Absorbance mode displays the absorbance of the  folded  data instead of the  transmittance         Laser Settings mx  View Less Expand Absorbance v Col       17 00    17 00   C       FIGURE 7 35  Absorbance mode of laser display    80    7 2 4 4 Colors    7 2 5 Find    TGA Series Trace Gas Analyzers    The Colors view  shown in FIGURE 7 36  is used to customize the laser  settings window colors  The default colors for the laser can be reset from this  window     ETTTEMPBMERS X      Background Colors       Used in Calculations     Discarded Color       Text Back Color          Normal Color     Detector Saturation        Text Color       m i    Center Line Color                FIGURE 7 36  Setting options for laser window display    The Find tool is used to find and identify absorption line s   It tunes the laser s  output wavelength by varying the laser current and or temperature and displays  a graph of the reference detector signal versus current  This graph  also called  a laser map  is comparable to a transmittance spectrum     The Find tool is similar for TGA Windows and 
98. dth in auto mode will scale the graph to the native  resolution of the display  each datum is matched to one screen pixel   Resizing  the graph window width changes the amount of data  time  that will be  displayed  This mode is generally preferred because the data scroll more  smoothly  The screen width may also be set manually  in which case the traces  may not scroll as smoothly     The graph may be zoomed by drawing a box from left to right around the  portion of the graph to be displayed  This will temporarily rescale the axes and  freeze the display  To resume scrolling with the previous axis scaling  draw a  box from right to left or click the Resume button in the upper right corner of  the graph     Alternately  you may compress or expand the scales  or step up down or  left right by clicking one of the icons at the top of the graph  Most of these  functions have keyboard shortcuts  Click anywhere on the graph window to  select it and hover over an icon to bring up a help box that explains it function  and gives the keyboard shortcut     The Data tool on the TGA toolbar brings up the Watch Window  This window  is used to view the TGA measurements in numerical form  Expand the  headings to see the values  FIGURE 7 43 shows the Ramp A and Detectors  headings expanded     This tool is similar for TGA Windows and TGA TEC  except for the names and  units associated with measurements and control of the laser temperature     87    TGA Series Trace Gas Analyzers          Watc
99. e        Dest   The Dest parameter is an array where the results of the measurement can be  stored  The length of the input variable array depends on the values of  parameters DataList and ScanMode     SDMAddress   The SDMaddress parameter is a constant that defines the address of the TGA  with which to communicate  Valid addresses are 0 through 14  The SDM  address is entered as a base 10 number  unlike older  jumper settable SDM  instruments that used base 4     91    TGA Series Trace Gas Analyzers    92    DataList   The DataList parameter is a constant that specifies the data to be retrieved  from the sensor  If DataList   1  only concentration and status are returned  If  DataList   2  then sample cell pressure  laser temperature  and DC Current are  returned in addition to concentration and status  If DataList   3  then the TGA  analog signal 1 and TGA temperatures are also returned  If DataList   4  then    all data except DutyCycle1 and DutyCycle2 are returned  If DataList   5   then all data are returned     ScanMode  The ScanMode parameter is a constant that specifies the number of values to  be retrieved for scan specific data  Normally the ScanMode parameter  corresponds to the TGA Number of ramps parameter that specifies how many  absorption lines are being measured  If ScanMode is set to a lower number  than the TGA Number of ramps parameter  the data for ramp B and or C will  not be retrieved from the TGA  If ScanMode is set to a higher number than  the TGA Num
100. e  individual laser   s data sheet  This will help to protect the laser if the laser  warms up  or if the laser current parameters are inadvertently set for too much  current     C 7 Laser Multimode Correction    An ideal laser would emit at only one frequency  single mode   Unfortunately   some  multimode  lasers emit some of their power at frequencies other than the  desired frequency  side modes   This side mode power is not absorbed by the  selected absorption line  therefore it gives an error in the measured  concentration     The TGA software can correct for a laser   s multimode power  if it is known  what percentage of the laser   s power is in the undesired side modes  If known   this value may be entered into the Settings  gt  Laser  gt  Other window  There are  three different methods to estimate the multimode power  depending on the  type of laser and the TGA model     TE cooled lasers   The TE cooled lasers used in the TGA200A are manufactured with a  distributed feedback  DFB  feature to suppress any side modes  For these  lasers  set the Multimode power   to 0     LN cooled lasers   The LN   cooled lasers used in earlier TGAs do not have the DFB feature to  suppress side modes  These lasers are specified to have not more than 10  of  their output in undesired side modes  For these lasers  the user may determine  the multimode power experimentally  The experiment consists of increasing  the amount of the target gas in the absorption path to absorb virtually all 
101. e tripod  tower  or attachments you are  installing  constructing  using  or maintaining  or a tool  stake  or anchor  come in contact with  overhead or underground utility lines    e Maintain a distance of at least one and one half times structure height  20 feet  or the distance  required by applicable law  whichever is greater  between overhead utility lines and the  structure  tripod  tower  attachments  or tools     e Prior to performing site or installation work  inform all utility companies and have all  underground utilities marked    e Comply with all electrical codes  Electrical equipment and related grounding devices should  be installed by a licensed and qualified electrician     Elevated Work and Weather  e Exercise extreme caution when performing elevated work   e Use appropriate equipment and safety practices   e During installation and maintenance  keep tower and tripod sites clear of un trained or non   essential personnel  Take precautions to prevent elevated tools and objects from dropping   e Do not perform any work in inclement weather  including wind  rain  snow  lightning  etc     Maintenance  e Periodically  at least yearly  check for wear and damage  including corrosion  stress cracks   frayed cables  loose cable clamps  cable tightness  etc  and take necessary corrective actions   e Periodically  at least yearly  check electrical ground connections     WHILE EVERY ATTEMPT IS MADE TO EMBODY THE HIGHEST DEGREE OF SAFETY IN ALL CAMPBELL SCIENTIFIC PRODU
102. ease the concentration noise slightly  but it will improve accuracy by  reducing the detector nonlinearity  This is especially important for measuring  Isotope ratios     D 4 Detector Linearity Coefficients    An ideal detector would have linear response  such that any increase in the  incident optical power would increase its signal proportionally  In reality  real  detectors have nonlinear response  As the incident optical power is increased   the incremental response becomes gradually lower  Detector nonlinearity 1s  worse at lower detector temperatures and at higher flux density  large detector  signals      Appendix D  Optimizing Detector Parameters    The TGA software corrects detector nonlinearity using the quadratic  polynomial     r   r Cr     where  r   detector response  r     linearity corrected response  C   the linearity correction coefficient    The linearity correction coefficients are defined separately for the reference  and sample detector  and for each ramp  if using multiple ramp mode   These  coefficients are settable in the Settings  gt  Detectors  gt  Preamp window     The reference detector linearity coefficient should be set to zero based on the  assumption that the reference detector is perfectly linear  This is assumed  because it is difficult to quantify the nonlinearity in the reference detector  and  because it generally gives good results  Although the reference detector may  not be perfectly linear  it is much more linear than the sample detec
103. eda G 3  G 2  Front and back Swagelok   ferrules                      ssssssss G 3  G3  Swagelok   plug    esee ee eene IR Ee ee dee G 4  G4  OSwagelok amp  cap   aede ert eR UN te ee G 4  H 1  TE cooled laser assembly installed in a TGA 200A                         H 1  H 2  TGA input module                   esee nennen H 2  H 3  TGA output module                  sssssssssseeeeeenenene nennen H 2  H 4  Modules mounted into TGA200A electronics                           ss H 3  H 5  Cryocooler feedthrough holes with grommets  above  and plugs    below   toes tedio temi isst ies  foerit H 5  H 6  Older style TGA input module shipped with TGA1005s                   H 7  H 7  Older style TGA output module shipped with TGA100s                 H 7  H 8  Older style two piece detector holder and short cell                         H 8  H 9  Newer style combined detector holder short cell  shown with   newer style cable                  sse H 9  T 1  Special control cable                  seeneeeeens I 1  I 2    Location of first thermistor probe attachment                         ssss I 1  I 3  Location of second thermistor probe attachment                              I 2  I 4  Thermistor cable wiring to analog inputs                     sssseese I 3  I 5  Control cable connection                    seen I 3   Tables   1 1  Historical Summary of Campbell Scientific Trace Gas Analyzers        1  3 1  Parts Included with the TGA200A               sse 3  4    Part Numbers for Availab
104. edure will help to align the  system well enough to see a response  As soon as a detector response is    Appendix B  Optical Alignment    observed  the system is ready for the horizontal and vertical alignment  procedure described in Appendix B 2 3  Horizontal and Vertical Alignment     1  Make sure the clamp knobs are loose and that the optical assembly is  in position     2  Use the horizontal adjustment screw at the laser end to align the  sample cell with the laser     3  Use the vertical adjustment screw to align the sample cell with the  laser  If no detector response is observed  set the vertical adjustment  near the center of its adjustment range     4  Alternately adjust the horizontal and vertical alignment screws  When  a response is observed in the sample detector  proceed to the next  section     5  Ifno detector response can be found  perform the following checks     e If two lasers are installed  FGA200   verify you are aligning to  the correct laser     e Verify the dewar cable is installed correctly  If two or more  lasers are installed  TGA200   verify you are using the correct  cable     e Verify the detector cables are correctly installed   e Verify the laser is enabled in the TGA program     e Recheck the laser temperature  and the zero  DC  modulation   and high current settings     6  Ifnoresponse is observed  use the alignment tool  pn 25897  to put  the reference tip tilt mirror and the detector mirrors at their nominal  positions  as shown in FIGURES
105. en lock onto the strongest absorption line within its spectral scan   Normally  this will be the same absorption line it was locked onto previously   In some cases  however  the automatic startup sequence will not find any  absorption line  or it will lock onto the wrong line  It is recommended that the  user verify that the correct absorption line has been locked     The laser   s emission frequency can be changed manually by changing the DC  current and the modulation current  It may be helpful to think of the reference  detector display as a viewing window looking upon a portion of the  transmittance spectrum  Increase the DC current to move the window to the    Appendix A  Configuring TGAs for Specific Gas Species    right and decrease the DC current to move the window to the left  It may also  be helpful to temporarily increase the width of the spectral scan by increasing  the modulation current  See Section 7 2 4  Laser Window  for details on  displaying the absorption line and adjusting the laser DC current and  modulation current     After the absorption line is found  adjust the DC current up or down to find  nearby absorption lines  Compare the spacing and relative depth of the  observed absorption lines to the transmittance spectrum provided in the laser   s  user manual to verify it is the desired absorption line  Adjust the DC current to  position the selected absorption line near the center of the spectral scan   Readjust the modulation current  see Appendix C 5
106. en the shipping clamps     31    TGA Series Trace Gas Analyzers    32    NOTE    TGA100 and TGA100A   The TGA200 and TGA100A used nylon belts to secure the optical bench inside  the enclosure  Loosen these belts and then loosen the transport lock and the  axial lock screws as shown in FIGURE 6 4     Horizontal  adjustment  screw    Tighten only when  transporting    Horizontal lock  screw    Axial lock screw    AURI eme ULL       FIGURE 6 4  TGA100 and TGA100A transport locks    TGA200 and TGA200A   The TGA200 and TGA200A use a set of four cam clamps to lock the optical  bench down for transport  Flip the cam clamps to the loose position for  operation  Also loosen the four bolt clamps that lift the optical assembly off  the optical bench  and the additional bolt clamp that holds the alignment  mechanism down  The cam and bolt clamps are shown in FIGURE 6 5     The cam and bolt clamps should be tightened only for transport   They should be loose during operation to allow the optical bench  to move within its enclosure with changes in temperature     TGA Series Trace Gas Analyzers    Cam Clamps          FIGURE 6 5  TGA200 and TGA200A shipping clamps    Once the enclosure is properly positioned  the analyzer needs sample input  connections and data output connections  The analyzer should be connected to  other system components as follows     6 1 1 Plumbing Connections    A TGA requires only a few additional components to draw samples into the  system to be analyzed  The connec
107. ent  pn 9838 is a box of 100 replacement filter elements   This  element must be replaced when it plugs enough to cause a significant decrease  in the sample cell pressure  The replacement interval is typically monthly  but  could be longer or shorter depending on conditions     Atmospheric Profiles   Low flow applications  such as atmospheric profiles  often use the pn 18072  heated filter orifice assembly  This assembly uses a heated filter holder  pn  18058  which uses a replaceable filter element  pn 17575  This element will  typically require replacement annually  but this interval could be longer or  shorter depending on site conditions     See Section 4 1 6  Replacement Parts  for details on these filter holders and  replacement filter elements     The TGA sample cell and optics do not require routine cleaning  Historically   the only TGAs returned to CSI for cleaning were heavily contaminated by  inadvertently pulling large amounts of dirty water through the analyzer     8 4 Sample Pumps    The TGA requires a vacuum pump to pull the air sample through the analyzer   The sample pump is not included as part of the TGA  so the details of pump  maintenance depend on the choice of pump  Two sample pumps available  from Campbell Scientific are given as examples     RB0021 L   The RB0021 L sample pump is a large capacity pump used for eddy  covariance  It is an oil sealed rotary vane pump that requires periodic  replacement of the oil  typically monthly   Oil for the RB0021 
108. erial cable connecting PC to RS232 port of TGA CPU module       40  TGA100A TGA200 OS download instruction                            sess 41  Device configuration utility Logger Control tab                                 42  Device Configuration utility Settings Editor tab                            ssse 43  Web based decimal to hexadecimal converter                                   44  Device configuration utility Terminal tab                      esses 45  TGA tool bar functions        ee eeceseceecnceneesseceeeecaeeeeceaserceaeeaeeaeeneeees 49  Connect window of TGA software interface                   sss 50  TGA error message for incompatible serial numbers                          51  Toolbar before  top  and after  bottom  connection                            51  TGA Status with a detected error  bottom                         ssssssse 52  TGA Status window without error  left  and with error  right    and line lock manually disabled                              sss 52  Expanded view of the menu in the TGA   s Settings window               53  TGA Windows laser settings                sse 54  TGA TEC laser setting Sreser ren r A enne 54  Laser current parameter settings for TGA Windows                          55  Laser current parameter settings for TGA TEC seee 56  Typical Line Lock parameter settings                     een 57  Settings of the Settings  gt  Laser  gt  Other screen                      ssssse 58  Ramp synchronization prompt                 eese 5
109. ers     3000 E  About TGA Duty Cycle Heater 1 1 00    TGA Temperature 2    Temperature set point 3000 5 23 38  C  Duty Cycle Heater 2 1 00                   FIGURE 7 22  TGA temperature control window for the two TGA  enclosure heaters    Data Output   The Settings  gt  Other  gt  Data Output window is shown in FIGURE 7 23   Check the SDM output button for normal operation  This will configure the  TGA as an SDM sensor  allowing a datalogger to request and receive data via  an SDM cable  Set the SDM address for the TGA from 0 to 14  and make sure  the datalogger program uses the same address to request data  see Section  7 3 1  SDM Output      The TGA may be configured for analog output  but this option requires the use    of a separate analog output module  It is not possible to configure a TGA for  both SDM and analog output  Contact Campbell Scientific for details     67    TGA Series Trace Gas Analyzers    68    CAUTION       TGA TEC Parameter Settings X       Laser          SDM output  Line Lock SDM address for TGA OF  Other     Detectors eeuna  Temperature  PreAmp     Calculations  Concentration  Analog Input     Other N20 Conc  TGA Temperatures  Scaing  Serial Numbers 0 00  About TGA          Channel 1 l2  3    4    1 00                   FIGURE 7 23  Data output setting window for SDM    Serial Numbers  The Settings  gt  Other  gt  Serial Numbers window is shown in FIGURE 7 24     The TGA is identified by a serial number and by a description  The TGA serial  number is
110. ers     To install a TE cooled laser in a TGA200    1  Turn off power to the TGA200    2  Disconnect the dewar cable from the laser dewar and the electronics   and remove the cable from the TGA200  This cable will not be used  with the new TE cooled laser assembly  which has the cable  permanently attached    3  Remove the four screws that fasten the laser dewar to the TGA200  optical bench  Remove the dewar from the TGA200  but keep the    mounting screws     4  Mount the TE cooled laser assembly to the optical bench  using the  dewar mounting screws     H 3    Appendix H  Upgrading Early Generation TGAs to TE cooled Laser    NOTE    5  Replace the input and output modules with the ones that have been  modified for use with TE cooled lasers     6  Connect all of the cables to the input and output modules except the  Laser Current connector     Do not connect the Laser Current connector to the output module    at this time     7  Install the TGA TEC software on a Windows PC  See Section 6 2 1   Installation of PC Software     8  Update the TGA operating  See Section 6 2 2  Updating TGA  Operating System     9  Complete the configuration for a new laser  See Section 6 3  Detailed  Setup Instructions     H 1 2 Detectors    If the TGA200 was equipped with liquid nitrogen cooled detectors  the  detectors may be replaced with TE cooled detectors  This requires the  additional assembly  pn 21577  TGA200 TE cooled detectors  Contact  Campbell Scientific for details on this upgrade  
111. ers   Upgrading a TGA100 is similar to upgrading a TGA200 or TGA100A  It  requires the same laser assembly  input and output modules  and software   Most of the additional upgrade issues are also common to the TGA100A  See  the section on upgrading a TGA100A for details  Some TGA100s may require  additional upgrades  as detailed in the following sections     H 3 2 CPU Module    TGA100s shipped with a CPU module that required a real time connection to a  PC running DOS  This was replaced with the introduction of the TGA100A by  a new design that does not require a real time connection to a computer  The  newer style CPU module can be distinguished from the older style visually by  the color of the module cover  The older style has a black module cover  while  the newer style has a shiny nickel plated module cover  The newer style CPU  module is required for the TE cooled laser  Most early TGAs have already  been upgraded to the new CPU module  Contact Campbell Scientific for  details     H 3 3 Input and Output Modules    H 6    TGA100s shipped with an older style module cover and connectors for the  input and output modules  The older modules have black module covers  instead of the newer nickel plated covers  Compare the TGA100 modules  shown below to the TGA200A modules shown in Appendix H 3 1  Basic  Upgrade  Modules shipped with TGA100As and TGA200s  and TGA100  modules that have been upgraded  will look like the upgraded modules  but    Appendix H  Upgrading Early Generat
112. es are  approximately equal  within 0 2   If one duty cycle is significantly higher than  the other  adjust the corresponding temperature setpoint downward  Try a   0 5   C change  This will compensate for any average difference in thermistor  readings  etc  It is not necessary for the duty cycles to be perfectly matched     Appendix I  Install Temperature Control Upgrade    Campbell Scientific Companies    Campbell Scientific  Inc   CSD  815 West 1800 North  Logan  Utah 84321  UNITED STATES  www campbellsci com   info campbellsci com    Campbell Scientific Africa Pty  Ltd   CSAf   PO Box 2450  Somerset West 7129  SOUTH AFRICA  www csafrica co za   cleroux a csafrica co za    Campbell Scientific Australia Pty  Ltd   CSA   PO Box 8108  Garbutt Post Shop QLD 4814  AUSTRALIA  www campbellsci com au   info campbellsci com au    Campbell Scientific  Beijing  Co   Ltd   8B16  Floor 8 Tower B  Hanwei Plaza  7 Guanghua Road  Chaoyang  Beijing 100004  P R  CHINA  www campbellsci com   info campbellsci com cn    Campbell Scientific do Brasil Ltda   CSB   Rua Apinag  s  nbr  2018     Perdizes  CEP  01258 00     Sao Paulo     SP  BRASIL  www campbellsci com br   vendas campbellsci com  br    Campbell Scientific Canada Corp   CSC   14532     131 Avenue NW  Edmonton AB T5L 4X4  CANADA  www campbellsci ca   dataloggers campbellsci ca    Campbell Scientific Centro Caribe S A   CSCC   300 N Cementerio  Edificio Breller  Santo Domingo  Heredia 40305  COSTA RICA  www campbellsci cc   info a cam
113. ew provides a finer adjustment of the horizontal  alignment than the one at the laser end  allowing the signal to be more easily  maximized  Newer systems have a fine pitch horizontal adjustment screw at  the laser end and require no adjustment at the detector end     B 5    Appendix B  Optical Alignment    Adjust the vertical alignment screw  see FIGURE B 3  to maximize  the sample detector signal  in the same way as for the horizontal  alignment     Iterate the horizontal and vertical alignment until the sample detector  signal is maximized  If there is a single narrow peak horizontally and  vertically  the system is also in good focus  If the response peak is  broad or if it has multiple peaks  adjust the focus as outlined in the  following section  B 1 3  Focus Adjustment     B 1 3 Focus Adjustment    The optical system includes the long sample cell  with the lens holder at the  laser end  and the beamsplitter and detectors at the other end  To focus the  system  this entire assembly is moved closer or farther away from the laser     1     To adjust the focus  first note the sample detector signal at the current  focus position     Loosen the axial clamping screw  slide the optical assembly either  forward or back a short distance   2 mm   and retighten the axial  clamping screw     Readjust the horizontal and vertical alignment to find the maximum  sample detector signal at this new focus position     Compare the sample detector signal at this focus position to the signa
114. ference Gas  for more details on the reference gas     Connect the sample intake to the sample gas inlet  The sample intake  should be filtered to remove particulates  10 um maximum pore size   and should have an appropriate needle valve or fixed orifice to control  the sample gas flow  The test intake  pn 15838  is shipped with the  TGA to make the initial setup easier for the user  The test intake  includes a filter  needle valve  and tubing to allow the TGA to operate  without fully installing the eddy covariance  EC  intake and dryer  or  other intake assembly or sampling system that will be used in a field  installation     If the application requires very high accuracy  connect a purge gas  supply to the TGA purge inlet  The purge gas supply should have an  appropriate regulator  flow meter  and needle valve to supply  approximately 10 ml min  The reference gas assembly  pn 15837  is  available from Campbell Scientific for this connection  refer to  Section 4 1 3 1  TGA Reference Gas Connection     6 1 2 Data Output Connections    NOTE       Connect the SDM cable to the CPU module in the TGA and the datalogger  A  20 ft SDM cable  pn 22178  is included for this purpose  If a longer SDM  cable is needed  use pn CABLEACBL L     See Section 7 3 1  SDM Output  for details on SDM data output   If using the CABLE4CBL L  the colors of the wires are different     Remove the cable feedthrough cap from side of the TGA enclosure and insert  the end of the SDM cable  The connection
115. figuration  Both the TGA100 and TGA100A have a  long sample cell  153 08 cm  60 27 in  in front of the beamsplitter  and a pair  of short cells  reference and sample  each 4 52 cm  1 78 in  behind the  beamsplitter  see FIGURE 4 25   The normal configuration for the TGA100  and TGA100A is to use only the long sample cell and the short reference cell   In this case  Ls is set to zero  and the values of L4 and Le are entered  The  concentration calculation simplifies to      C XL4XD  _ 0 02959 C  XD     L    D     D        Q     27    TGA Series Trace Gas Analyzers    5     28    It is possible to measure very high concentrations of the target gas in the  TGA100 and TGA100A by changing the plumbing connections to flow the  sample air through the short sample cell instead of the long sample cell  In this  case  L 4is set to zero  and the actual values of Ls and Lp are entered  The  concentration calculation simplifies to     C    C   D   3     The TGA200 and TGA200A have only one sample cell and one reference cell   These are both long  146 6 cm  57 72  and located behind the beamsplitter   Because they are behind the beamsplitter  their lengths are entered as Ls and  Lg  and L  is set to zero  The concentration calculation simplifies to Equation   3   above     Specifications  Measurement Specifications    The typical concentration measurement noise is calculated as the square root of  the Allan variance   Allan deviation  with 100 ms averaging  which is  comparable to the stan
116. from the fitting  It is strongly recommended to cap all disconnected tubes to  keep them clean  Spare caps may be needed if they become lost or damaged        FIGURE G 4  Swagelok   cap                         TABLE G 5  Dimensions and Part Numbers for  Swagelok   Caps  Tubing OD  in  Swagelok   pn CSI pn  1 8 B 200 C 19219  1 4 B 400 C 15831  3 8 B 600 C 15547  1 2 B 810 C 17335  5 8 B 1010 C 19496                   Appendix H  Upgrading Early  Generation TGAs to TE cooled Laser    Most TGAs shipped since 2000 can be upgraded to use TE cooled lasers   Earlier TGAs will require more upgrades than the more recent TGA200s  The  following sections give details for each TGA model     H 1 TGA200    H 1 1 Basic Upgrade  Upgrading a TGA200 to use a TE cooled laser requires the following parts     TE cooled laser assembly  TGA200A Input Module  TGA200A Output Module  TGA TEC software    The TE cooled laser assembly includes the laser  mounting hardware  and  cable  The assembly is shown in FIGURE H 1 installed in a TGA200A  There  are five different laser assemblies available  to measure different gases     pn 30477 Methane  CH4    pn 30478 Nitrous Oxide  N20    pn 31121 Nitrous Oxide and Carbon Dioxide  N20 and CO    CO  is  based on the PC 60 60 isotopolog    pn 31119 Carbon Dioxide  CO2  and 613C   pn 30877 Carbon Dioxide  CO2   6180  and 613C       FIGURE H 1  TE cooled laser assembly installed in a TGA 200A    H 1    Appendix H  Upgrading Early Generation TGAs to TE cooled Laser
117. fy the detector signals are consistent with previous operation of  the TGA  If they have changed  check the operational parameters  see  Appendices B through D      Verify that the reference transmittance at the center of the absorption  line is consistent with previous operation of the TGA  This  transmittance is dependent on which absorption line is selected  the  concentration in the reference cell  the pressure in the reference cell   and the laser performance  A significant change indicates a problem     Check the concentration and the concentration standard deviation to  verify proper performance     The TGA is now fully functional     7 1 2 Routine System Checks    The TGA is often used for long term  continuous measurements  It is  necessary to periodically check the status of the system and perform routine  maintenance  The status can be checked either with the TGA software running  on the user   s PC  or through a datalogger connected to the TGA  Ifthe TGA is  in multiple ramp mode  check the values for each ramp  Parameters should be  tracked in a log book to establish normal values and variability  This will help  the user to recognize when a parameter changes beyond its normal variability     1     Verify the status flags  If using a datalogger  make sure the value of  the TGAStatus variable is zero  A nonzero value indicates a  problem  See Section 7 3 1  SDM Output  for details on the TGA  status flags  If using the PC software  check for error messages in the 
118. ghtly different for TGA Windows  and TGA TEC  TGA Windows  FIGURE 7 10  supports a rapid warm up  function to bring the laser more quickly to room temperature  This is helpful  for periodic  typically annual  evacuation of the LN   dewar  TE cooled lasers  do not require this feature  Therefore the Laser  gt  Current screen of the TGA  TEC does not show the option to heat the laser  FIGURE 7 11      Both versions have a box to turn the laser on off  Check this box for normal  operation  Uncheck this box to disable the laser current     The laser DC  Mod  modulation   Zero  and High currents determine the laser  scan sequence  They should be set and optimized for the individual laser  The  DC  Mod  Zero  and High current parameters may also be set from the Laser  window  See Appendix C  Optimizing Laser Parameters  for details on how to  set the laser current parameters  This section also discusses the Line Lock and  Auto buttons next to the DC  Mod  and Zero currents     The Maximum laser current  mA  provides a safety limit  The TGA  software will not allow the DC  Mod  Zero  or High current to be set to a value  that would apply a current through the laser greater than this limit  Set this  limit for the individual laser     Setting the maximum laser current parameter above a    laser s safe operation current greatly increases the risk of  damaging the laser           i  p    Y  TGA Parameter Settings al  E  Laser  Temperature E  V  Laser on  Line Lock Heat laser up to 310 
119. h Window Aaa       El                            Vj  EI bb LE                 Ramp A  C02 Conc 150 20323 ppm  Mean C02 Conc 164 219772 ppm  C02 Conc StdDev 48 20964 ppm  Smp Det Signal CO2 30 5060 mV  Ref Det Signal CO2 1 0053 mV     Laser DC Current CO2 403 44 mA   Smp Transmittance CO2 98 6258    Ref Transmittance CO2 97 7745     Ramp B   Ramp C   Isotope Ratios   Miscellaneous   Detectors  Smp Det Temp 45 000   C  Ref Det Temp 45 000   C     Smp Det Cooler 1392   Ref Det Cooler 1451  Smp Det Gain 3  Smp Det Offset 58  Ref Det Gain 5  Ref Det Offset 0    FIGURE 7 43  Data outputs of TGA TEC    7 2 8 Files    TGA data are usually collected using a datalogger  However  the data can also  be collected using the PC  The Files tool of the TGA toolbar is used to set up  the values to be collected and to turn data collection on and off  Clicking on  the Files icon of the TGA toolbar will bring up the Data Files screen  shown in    FIGURE 7 44     88    Data Files EE x     Turn Data Collection On      Settings                                      Files                                        Data Files       mm dd yyyy hhmm dat    File Format   Binary Y         Col 1  Time of Day       Housekeep Data File                  TGA Series Trace Gas Analyzers    mm dd yyvy hhmm  all    File Format  ascii Y      Day of Year    Col 2  C02 Conc Col 2  Time of Day  Data File Col 3  Ref Det Signal CO2 Col 3  C02 Conc  LH keep Fil Col 4  Smp Det Signal CO2 Col 4  Mean CO2 Conc  Quse Seine Col 5 
120. he ramp at lower DC current  Ifa  negative high current pulse is required  it may be helpful to also reduce the  zero current below its normal setting     C 3 1 High Current Count    The High current count parameter may be adjusted in conjunction with the  high current  The High current count parameter sets the duration of the high  current pulse  from 0 to 8 counts  where each count represents a 20 us interval   giving a total duration of 0 to 160 us  Generally  it is best to start with the high  current counts at its maximum value  which will give a low amplitude  long   duration pulse  However  if the high current is set to a small value  less than  20 mA for LN   cooled lasers or less than 2 mA for TEC lasers  it may be  helpful to reduce the high current counts and increase the high current  This  will allow more of the samples to be used in the concentration calculation   This parameter can be adjusted from the Settings  gt Laser  gt  Other window  see  Section 7 2 3 1  Laser     Adjusting this parameter usually makes little difference in the TGA   s  performance  so when in doubt  this parameter should simply be left at the  default which is 8 counts  For multiple ramp mode  the high current count is  common to all ramps  and should generally be left at the default  8 counts      Appendix C  Optimizing Laser Parameters    C 4 Omitted Data Count    Some additional data must be omitted from the concentration calculation to  allow the laser temperature to stabilize fully 
121. he tabs   Typical names for the more commonly measured  gases are given in TABLE 7 2  Text of the user s choice  up to 23 characters   may be entered as an alternative        TABLE 7 2  Suggested Naming for Gas Names Setting                Target Gas Suggested Naming  Methane CH4  Nitrous Oxide N20  Carbon Dioxide   C 50 60  12C       Carbon Dioxide   C 50 60  13C  Carbon Dioxide   C 50 60  180  Ammonia NH3                      The Multimode power   parameter is the percentage of the laser s power in  unwanted side modes  See Appendix C 7  Laser Multimode Correction  for  details on how to set this parameter     At the bottom of the window are two ramp timing parameters that are common  to all ramps  High current count is the duration of the high current phase of  the ramp  in counts  Omitted count is the number of counts to omit from  concentration calculations  The defaults are 8 and 20  For more information  on these parameters  see Appendix C 3 1  High Current Count  and Appendix  C 4  Omitted Data Count     Temperature   From the Settings  gt  Detectors  gt  Temperature window  parameters can be set to  control the temperature of the detectors  This window is the same for both  versions of the software  TGA Windows and TGA TEC  and is shown in  FIGURE 7 15     59    TGA Series Trace Gas Analyzers    60       arameter Se                E  Laser  Temperature    Ourent  V  Control detector temperatures  i  Line Lock  i  Other  B  Detectors Cooler control update interval 10 
122. iation  time parameter  displayed at the bottom of the window shown in FIGURE  7 17  determines the time over which these statistics are calculated     Analog Input  The Settings  gt  Calculations  gt  Analog Input parameters are the same for both  versions of the software     The TGA has four differential analog input channels  Channel 2 and 3 measure  temperatures inside the TGA enclosure for controlling the heaters  see  Appendix I  Install Temperature Control Upgrade   Channel 4 measures the    TGA Series Trace Gas Analyzers    sample cell pressure  Channel   is available for the user to add a sensor  Each  of these channels has a tab to configure it  Pressure is the label given to the tab  for channel 4     The default configuration for Channel 1 is shown in FIGURE 7 18                                 Laser     Temperature     Current Channel 1 Channel 2   Channel 3   Pressure   Line Lock  i   Other      Detectors    Name i  Es Ex TGA analog input  PreAmp  C  PRT sensor  TGA100A   B  Calculations    A200  Concentration V Mera    i f Analog Input   Manual scale offset  B  Other  TGA Temperatures Scale 1 00    Data Output  Serial Numbers ee      About TGA  Measured data 0 02                         FIGURE 7 18  Default settings for Channel 1 in analog input screen    Channels 2 and 3 measure temperatures inside the TGA enclosure  Early TGA  used platinum resistance thermometer  PRT  sensors  but the TGA200 and  TGA200A use thermistor probes  TGA100s and TGA100A may be upgr
123. iew mode of laser display                       sess 76  Magnified mode of laser display                       sss 77  Detrended mode of laser display              csscesseesseeeeeseeesceeeeeteeeeeenseenes 78  Folded mode of laser display                        sss 79  Absorbance mode of laser display                       ssssssssssssss 80  Setting options for laser window display                         sess 81  Laser line find window                      eese 82  Interactive Laser Find window for a CO2 isotope laser                      83  Noninteractive Laser Find window                       eee 84  Options for graphical display of data in TGA Windows                     85  Adding parameters to a graph in TGA TEC                    eee 86  Example graph showing N20 concentration and standard   deVIatlOTL   otn CR EC ROO E 86  Data outputs of TGA TEC              isses 88  Controlling PC recorded data options in the TGA                              89  Example of log file messages                  sseeeeeeenere 90  Setting TGA for SDM output from TGA Parameter Settings   Window  etae tpe REOR P eO Un 91  TGA100 or TGA100A optical layout with air gap purge                 A 8  TGA200 or TGA200A optical layout with air gap purge                 A 8  TGA100 and TGA100A optical layout                      esee B 1  Alignment hardware of detector end of TGA100 and TGA100A     B 2  Alignment hardware of laser end of TGA100 and TGA100A          B 3  TGA200 and TGA200A optical 
124. igned well enough to see a response in the sample  detector  follow these steps to optimize the horizontal and vertical alignment     1  Adjust the horizontal position  see FIGURE B 5  to maximize the  sample detector signal     NOTE The sample and reference signals may not reach their maxima  simultaneously  If so  ignore the reference detector signal and  adjust the alignment to maximize the sample detector signal   Adjust the horizontal position past the peak in each direction far  enough to make sure there is a single response peak  If there is a  single peak  leave it at the center of the peak  If there are multiple  peaks  leave the horizontal alignment at the center of the group of  peaks     2  Adjust the vertical position  see FIGURE B 5  to maximize the  sample detector signal  in the same way as for the horizontal  alignment     3  Adjust the sample detector mirror tip and tilt knobs to maximize the  sample detector signal     4  Adjust the beamsplitter mirror tip and tilt knobs to maximize the  reference detector signal     Appendix B  Optical Alignment    5  Adjust the reference detector mirror tip and tilt knobs to maximize the  reference detector signal     6  Iterate these steps until both detector signals are maximized  If there  is a single narrow peak horizontally and vertically  the system is also  in good alignment  If the response peak is broad or if it has multiple  peaks  contact Campbell Scientific for assistance     B 2 4 Focus Adjustment    The TGA2
125. in     go to optimum temperature          103 6                485 0 65 5 6 0 31 0       12  Increase the laser operating temperature by 0 1 or 0 2 K  Some lasers    will allow a larger temperature increment  but when in doubt  use 0 1  K  The line locking algorithm will decrease the DC current as needed  to keep the absorption line in the center of the ramp        Appendix C  Optimizing Laser Parameters    13  Wait until the laser temperature and DC current stabilize  and then  record the values     14  Iterate this process until the transmittance or concentration noise  increases noticeably     15  Return to the starting laser temperature  and then step the temperature  downward until the transmittance or concentration noise again  increases noticeably  Record the laser temperature  DC current   reference transmittance  concentration noise  and sample signal at  each step     To evaluate the results  first verify the DC current decreases approximately  linearly with increasing temperature  A discontinuity indicates the line locking  algorithm may have switched to a different absorption line  This can be caused  by other strong absorption lines nearby  or by a laser mode hop  If this  happens  repeat parts of this test near the discontinuity  using a smaller step in  laser temperature  Watch the reference detector transmittance carefully  If the  discontinuity was caused by another strong absorption line nearby  using a  smaller temperature step may solve the problem  If the
126. in carbon dioxide or water by  tuning each ramp to a different isotopolog     The multiple ramp mode may also be used to measure some other pairs of  gases  such as nitrous oxide and carbon dioxide or nitrous oxide and methane   The measurement noise  however  will generally be higher than if a single gas  is measured  For measurements of a single gas  the laser wavelength is chosen  for the strongest absorption lines of that gas     Choosing a laser that can measure two gases simultaneously involves a  compromise  Weaker absorption lines must be used in order to find a line for  each gas within the laser   s narrow tuning range     4 2 6 Concentration Calculation    The reference and sample detector signals are digitized  corrected for detector  offset and nonlinearity  and converted to absorbance  A linear regression of  sample absorbance vs  reference absorbance gives the ratio of sample  absorbance to reference absorbance  The assumption that temperature and  pressure are the same for the sample and reference gases is fundamental to the  design of the TGA  It allows the concentration of the sample  Cs  to be    calculated by   c  CDD    a Esso LL D   where  Cr   concentration of reference gas  ppm    Lr   length of the short reference cell  cm  Ls   length of the short sample cell  cm   L4  length of the long sample cell  cm   D   ratio of sample to reference absorbance    The terminology long short cell is historical  based on the original TGA100  and TGA100A optical con
127. intenance  8 1 Lasers and Detectors    TGAs equipped with cryogenic lasers and or detectors required extensive  routine maintenance  refilling the dewar with LN  daily to weekly  and  periodically  typically annually  evacuating the dewar to maintain its insulating  vacuum     TGAs equipped with TE cooled lasers and detectors require no routine  maintenance     Troubleshooting both types of lasers and detectors involves the routine system  checks described in Section 7 1  Routine Operation  and the detailed setup  described in Appendices A through E     TGA Series Trace Gas Analyzers    8 2 Reference Gas    All TGAs require a constant flow of reference gas from a high pressure  cylinder of prepared gas  A typical cylinder will last approximately one year  before it must be replaced  See Appendix A 2  Reference Gas  for details     8 3 Filtration and Sample Cell Cleaning    The air sample to be measured must be filtered to remove particulates  10 um  maximum pore size   The filter will eventually plug  requiring cleaning or  replacement  The filter is not included as part of the TGA  so the details will  depend on the choice of filter and the application  flow rate   Two popular  configurations are given as examples     Eddy Covariance   Most TGA eddy covariance applications use a PD200T dryer that includes a pn  20553 47 mm filter holder   Earlier TGAs used the PD1000  which had a  similar filter holder   These filter holders use a 47 mm diameter  10 um pore  size filter elem
128. ion TGAs to TE cooled Laser    without the blue sticker   Some of the older modules may have been upgraded  to the newer design to improve immunity to radio frequency  RF  interference   Modules with the nickel pated module cover require only the basic upgrade  described in Appendix H 3 1  Basic Upgrade  Modules with the older black  module cover may be upgraded at increased cost for the new module cover and  connectors          9058 TGA TRACE GAS INPUT       Ree    SA    a          FIGURE H 7  Older style TGA output module shipped with TGA100s    TGA100s that are still using the original black input and output modules will  also be using an older style cable for the detectors  See Appendix H 3 4   Detectors and Detector Cables  for details on upgrading the detector cable     H 3 4 Detectors and Detector Cables    Similar to the TGA100A  most TGA100s were supplied with TE cooled  detectors  However  if the TGA100 was supplied with LN2 cooled detectors   these should be replaced with TE cooled detector assemblies  Contact  Campbell Scientific for availability     TGA100s used detector cables with three pairs of wires within a gray cable  sheath  This cable terminated at the electronics end with green screw terminal  blocks  In 2005  with the introduction of the TGA 100A  this design was  replaced with a bundle of three separate blue cables  terminated with circular  metal connectors  The input and output modules were modified to use the    H 7    Appendix H  Upgrading Early Ge
129. ion and the other calibration tank should have zero concentration   For applications measuring very high concentrations  however  it is preferable  to bracket the expected measurement range  For example  if measuring isotope  ratios in ambient CO    calibration tanks with  300 ppm and  600 ppm may be  preferred  Configure the calibration tank connections to supply the same flow  rate  and to give the same sample cell pressure  as for the trace gas  measurements  The difference in the measured concentrations for the two  tanks should be equal to the true difference between the two calibration tanks   If it is not  adjust the Reference gas concentration  ppm  parameter on the  Concentration Calculation window proportional to the measured error     T    T   Cy  PA Corig C   1 2    where  Cue    the corrected reference gas concentration  Corig   the original reference gas concentration  T  and T     the true concentrations in the calibration tanks    Mi and M     the calibration tank concentrations measured  by the TGA    E 1    Appendix F  TGA Frequency Response    There are several issues related to the frequency response of the TGA including  measurement rate  sample rate  digital filtering  synchronicity  and volume   flow  and pressure of the sample cell  This discussion applies to all TGA  variants running either the TGA Windows  LN   cooled lasers  or TGA TEC   TE cooled lasers  software     F 1 Measurement Rate    The TGA scans a selected absorption line of a specific gas
130. is available  from CSI as pn 8143     See RB0021 user manual for troubleshooting suggestions   XDDI  The XDD1 sample pump is a multi stage diaphragm pump used for low flow    applications  such as atmospheric profiles  This pump requires no routine  maintenance     95    TGA Series Trace Gas Analyzers    See the XDD1 user manual for troubleshooting suggestions     See Section 4 1 4 4  Sample Vacuum Pump  for additional information about  both of these pumps     96    Appendix A     Configuring TGAs for    Specific Gas Species    Campbell Scientific TGAs can be configured to measure one of several gases  by selecting an appropriate laser  reference gas  and detectors  In some cases  multiple gases can be measured simultaneously  The following sections give  details on how to configure the TGA for the more commonly measured gases     A 1 Laser Selection    CAUTION    Each gas species has a unique set of absorption lines and tunable diode lasers  have limited tuning ranges  Therefore  in most cases a different laser is   required for each gas species to be measured  Two types of lasers have been  used in TGAs  LN   cooled lasers and thermoelectrically cooled  TEC  lasers     The TGA uses a Class 1M laser  Do not view the laser    directly with optical instruments     A 1 1 LN cooled lasers    The TGA100  TGA100A  and TGA200 used cryogenic lead salt lasers that  were available at wavelengths from 3 to 10 um  This wide range of  wavelengths allowed the TGA to measure a large numbe
131. is is the estimate of the laser s multimode power  Enter   this value as the Multimode power     parameter in the Settings  gt  Laser  gt    Other window  Verify the reference transmittance is near zero     C 13    Appendix C  Optimizing Laser Parameters    TGA200   The process for determining Multimode power   in the TGA200 is slightly  different than for the TGA100 and TGA100A because the TGA200 design has  the reference and sample cells of similar length  Therefore  the absorption  cannot be increased by putting the reference gas in the sample cell  Instead  the  reference gas must be replaced by a higher concentration of the target gas  If  the reference transmittance is 50  or lower with the normal reference gas  the  second gas concentration must be at least 10 times greater than the reference  gas  15 x for 50 to 60   or 20 x for 60 to 70    Otherwise the procedure is the  same as for the TGA100 and TGA100A     For multiple ramp operation the laser multimode power may be different for  each absorption line  Repeat the process described above for each ramp     Appendix D  Optimizing Detector  Parameters    Set the detector parameters after the laser parameters are adjusted     D 1 Detector Gain and Offset    The detector signals are processed in the TGA electronics  which include an  amplifier with programmable gain and offset in the input module  and a second  programmable gain amplifier in the analog module  The detector gains and  offsets are normally controlled auto
132. l  at the previous focus position  Step the focus again in the same  direction if the signal improved  or move it the other direction if the  signal decreased     Repeat this process until the sample detector signal has a single   narrow peak of maximum height  It may be helpful to record the  focus position and sample signal in a table  along with a qualitative  assessment of the focus  whether there is one or multiple peaks  and if  the peak seems broad or narrow     Generally the goal is to set the focus for a single narrow peak  giving  the maximum sample signal  In some cases it may be desirable to  intentionally defocus the system as in some of the following scenarios     e Isotope ratio measurement accuracy may be improved by  defocusing to reduce detector nonlinearity    e Ifthe laser signal is large enough that the detector temperatures  must be raised above   0   C to avoid detector saturation   defocusing will reduce the signal and it will also reduce  detector nonlinearity    e   If it is difficult to coalign the reference and sample detector   see Appendix B 1 4  Reference Detector Coalignment  it may  be helpful to defocus the system    In cases where the optics are to be intentionally defocused  start at the position  of best focus and move the optical system away from the laser by no more than    B 6    Appendix B  Optical Alignment    5 mm  It is generally best to defocus just enough to give a single relatively  broad  flat peak     B 1 4 Reference Detector C
133. l  shows the expanded window of the View More mode  View More is the default  mode     71    TGA Series Trace Gas Analyzers    72     amp  Laser Settings  EJ    Laser Settings  View More Expand Detrended   Co View Less Expand Detrended v Colors             DC current     88 14 5   amp           17 00    17 00   C       FIGURE 7 27  View Less View More function of the Laser Settings  window    7 2 4 2 Tabbed Expand    The Tabbed or Expand button 1s used to switch between individual displays for  each ramp or displaying only one ramp and having tabs to switch the displayed  ramp  The two display modes are shown in as shown in FIGURE 7 28  The  difference between the two modes is irrelevant for single  ramp mode     TGA Series Trace Gas Analyzers          r       View Less Tabbed Detrended   Colors    CH4   DCcurent     88 144 H   Mod curent 1 30 Ej  Zero current   50 00 2   Auto   High current 0 00 t    17 00 2  17 00 C    7 2 4 3 Display Mode       E       Laser Settings    View Less Expand Detrended     Colors    CH4  Gase            DC current     8814 HO  Mod current 1 30   futo   Zero current 50 00    Highcurent         0 005           FIGURE 7 28  Tabbed Expand function of the Laser Settings window    The Display Mode of the Laser Settings window 1s the third button from the  left  There are five display modes that can be set from here  The default mode  is Detrended     Raw  Maximum  Magnified  Detrended  Folded  Absorbance    A side by side comparison of the modes is gi
134. l conditions of normal use except when passed  through magnifying optics such as microscopes and telescopes   Do not view the laser directly with optical instruments     The power cords supplied with the TGA may not be of sufficient  length for a given site application  If longer AC power cords are  required  always have a qualified electrician perform the work     e WARNING     o    Before attempting to install or change a laser in the TGA  read all  documentation accompanying the laser and thoroughly review  Appendix A 1 4  Changing Lasers  before attempting to change  or use a new laser  Laser settings and proper parameters must be  set according to the laser s requirements and a defined order of  events must occur to initiate a new laser in the TGA  Failing to  do so could irretrievably damage the laser     The cam and bolt clamps of the TGA should be tightened during  transport  Failure to do so could damage the analyzer     e CAUTION     O    The cam and bolt clamps of the TGA should only be tightened  during transport  They should always be loose during operation  of the TGA  The optical bench of the TGA should be free to  move within its enclosure with changes in temperature  Failure to  do so could result in inaccurate measurements     TGA Series Trace Gas Analyzers    3  Initial Inspection    4  Overview    e Upon receipt of a Campbell Scientific TGA  inspect the packaging and  contents for damage  File damage claims with the shipping company   Contact Campbell Scie
135. l mode and then uncheck the All Caps box at the  bottom of the window as shown in in FIGURE 6 20     6  Enter C to view the Configuration Options    7  Enter 1  Ethernet Address   At the prompt  enter 0x followed by a  four digit hexadecimal value  If the serial number in hex is less than  four digits  use leading zeros to create a four digit hex number  In the    example shown in FIGURE 6 19  the serial number entered would be  0x05b2     NOTE The Ox preceding the four digit hex  must be a lowercase x     44    TGA Series Trace Gas Analyzers    8  Enter 6 to Save and Exit           EE  O Der oraa a umm WC iei    Camera  Datalogger  Datalogger  Other   Network Peripheral  Peripheral  Phone Modem     Radio  El Sensor  CRS450 Series  CS120  CS125  C5140  CS450 Series  CS451 Series  CS650 Series  EC100  OBS500    TGA 100A TGA200    E  Unknown     Wireless Sensor    Communication Port   COM1  Use IP Connection    Baud Rate  115200 v          x       TGA200A  C    Configuration Options   1  Ethernet Address  0x00D02C030022A   2  IP Port  3000   3  IP Address  192 168 28 91   IP Default Gateway  192 168 28 1   5  IP Network Mask  255 255 252 0   Save and Exit   Exit with no Save   Current Ethernet Address  0x00D02C03002A  Enter New Address  lower 2 bytes only    Ox05b2    Configuration Options   1  Ethernet Address  0x00D02C0305B2   2  IP Port  3000   IP Address  192 168 28 91   IP Default Gateway  192 168 28 1   IP Network Mask  255 255 252 0   Save and Exit   Exit with no Save   
136. laser dewar to be filled  Connect the  side flaps to the end flaps with the hook and loop fastener strips and then  attach the top to the bottom     13    TGA Series Trace Gas Analyzers    14    4 1 4 Other Accessories  4 1 4 1 TGA Air Sample Intake    The TGA air sample intake assembly  pn 17882  includes a rain diverter  insect  screen  stainless steel tubing  and mounting bracket  The assembly can be  mounted on a pipe  as shown in FIGURE 4 10  with an outer diameter from   1 3 cm  0 5 in  to 5 cm  2 in   This intake is suitable for low  or high flow  applications     ll     a    bi    n    FIGURE 4 10  TGA air sample intake    4 1 4 2 TGA Heated Intake Filter  amp  Orifice    The heated intake filter and orifice  pn 18072  is used with trace gas analyzer  systems in low flow applications  such as atmospheric profiles or chamber  measurements  The assembly consists of a plastic housing that covers and  insulates a heated filter and orifice  The filter and orifice is shown in FIGURE  4 11  The heater requires 12 Vdc  0 5 W of power and includes a connector to  attach a pn 18073 L cable from a 12 Vdc source  Nominal flow for the three  standard orifice sizes  0 007 in   0 010 in   and 0 013 in   is 0 26  0 56  and 0 90  slpm  at sea level             Ei    Una           FIGURE 4 11  TGA heated intake filter and orifice    4 1 4 3 TGA High Flow Filter Holder    The TGA high flow filter holder  pn 20553  uses a 47 mm filter membrane to  remove particulates from an air stream a
137. layout                     sess B 8  Alignment hardware of laser end of TGA200 variants                     B 8  Alignment hardware of detector end of TGA200 variants                B 9  Use of alignment tool for aligning mirror in TGA beamsplitter   DIOCK qp                          B 11  Use of alignment tool for aligning mirror in TGA beamsplitter   block  alternate angle    B 11  Use of alignment tool to position tip tilt screws for aligning   detector side mirrors                    esee B 12  Typical laser DC current as a function of temperature                      C 3  Typical reference transmittance as a function of laser   ie mperatute     uto etie eter testis leise e e C 4  Typical concentration noise as a function of laser temperature        C 4    Typical sample detector signal as a function of laser temperature    C 5  Example of using Laser Line Find function to determine laser    threshold c  rrent        tese Set eio UOS C 6  Effects of temperature perturbation                    eee C 7  High current adjustment procedure                     see C 8  Adjustment of omitted data counts                        sess C 11  Adjustment of modulation current                    eese C 12  ECfilt  r coefficients once ete etre rS F 2  EC filter frequency response  linear scale                                     F 3  EC filter frequency response  logarithmic scale                                F 3    vi    Table of Contents    G I   Swagelok   insert    saos ee RR t e r
138. le Gas Species Lasers                     sssss 11  4 2  Available AC Mains Power Cords by Region                     sss 12  4 3  Power Module Mounting Brackets                   sse 12  4 4  Sample Air Dryer Specifications                sse 19  5 1  Typical Measurement Noise              sss 29  5 2  Physical Specifications of TGA Variants                 esee 29  7 1  Appearance and Function of Line Lock Icons                           esss 57  7 2  Suggested Naming for Gas Names Setting                  ssssssssssss 59  Js  FEC ASMSHUCE OM 4 err eee eet deett 92  7 4  Descriptions of TGAStatus Values                esses 94  A 1  Discontinued Cryogenic Lead salt Lasers                        sess A 1  A2  TE codled Lasers a a a atea ORO o D Rat A 2  A 3  Replacement Cables for TGA Dewars                sese A 2  A 4  Suggested Reference Gas Concentrations                   sss A 5  C 1  Example Laser Temperature Optimization Data    eee C 2  F 1  Recommended Passband Settings                      sss F 4  F 2  Processing Lags for EC filters                  sse F 5  F 3  Summary of TGA Update Times                 sse F 5  F 4  Sample Cell Residence Time as a Function of Sample Cell   Volume and TGA Model                      senten F 6  G 1  Available Plastic Tubing Sizes  Construction  and Usage   Cuidelines   1iuo ehe tie e eee eet t G 2  G 2  Dimensions and Part Numbers for Swagelok   Inserts                     G 3  G 3  Dimensions and Part Numbers for Swagelok   Fe
139. lignment    FIGURE B 6 illustrates the alignment hardware at the detector end     NOTE This illustration is a view looking up from underneath  The  adjustment screws are not visible from above           Sample Detector  j Mirror Tip Tilt       Reference Detector  Mirror Tip Tilt    FIGURE B 6  Alignment hardware of detector end of TGA200 variants    B 2 1 Configure the TGA PC Software    1  Start the TGA program and make sure the laser and detector  parameters are set appropriately for the laser     2  Display the reference detector signal and the sample detector signal in  agraph  The goal of the alignment procedure is to maximize these  signals     3  Edit the graph options to move one of the traces to the right axis  set  the minimum value of each Y axis to zero and let the maximum of  each Y axis scale automatically     4  Setthe reference and sample detector gains to zero  This will disable  automatic gain and offset adjustment which can cause confusion  during the alignment process when enabled     5  Setthe detector temperatures as needed to avoid saturation  This  adjustment may need to be repeated during the alignment process 1f  the signal level increases too much  It is not important to have large  signals during alignment  When in doubt  set the detector  temperatures relatively high for relatively low signals     B 2 2 Initial Alignment    If the optical system is significantly misaligned  there may be no observable  detector response  This initial alignment proc
140. line  In this case  it may be necessary to choose a laser  temperature that gives a compromise between reference transmittance and  concentration noise  If there is no laser temperature that gives satisfactory  performance  it may be necessary to choose another absorption line     For multiple ramp operation  follow the process described above  but also  record the DC current  reference transmittance  and concentration noise for the  additional ramps  Ideally  all ramps will have the same optimum laser  temperature  In some cases  it may be necessary to set the laser temperature  between the optimum temperatures for the ramps to achieve acceptable  performance for all ramps     C 2 Zero Current    The laser current must be reduced below the lasing threshold briefly at the start  of each spectral scan  described in Section 4 2 5  Laser Scan Sequence  to  measure the detector response with no laser emission  If the zero current is set  too high  the laser will emit some energy when it should be off  and the TGA  will calculate the wrong transmittance  This will cause an error in the reported  concentration  This problem could be avoided by simply setting the zero  current to 0 mA to guarantee the laser is off  However  both current and  temperature affect the laser s emission frequency  and the laser s temperature is    C 5    Appendix C  Optimizing Laser Parameters    C 6    NOTE    affected by its current  The laser   s temperature falls slightly when the current  is reduced 
141. ller    H 9   l  Install Temperature Control Upgrade                      l 1  I 1 Install Thermistor Probes  0        cccccessessseesceesceeeeeeeceeceseceseenseeneeeeeenes I 1  L2 Connect the Control Cables iaieineea i I 3  L3 Enter Control Parameters                  sse I 4  L4 HEISE I 4  I 1 Operations sc so cnet est ea tli ure Ge eade ash wl I 5   Figures   4 1  Screen of TGA Windows software    5  4 2   TGA200A system components 0 0 0    ceeeeesteceeeeeeseeceeeeecsaeceeeeeenaeceenees 7  4 3   TGA200A power module                      sese 8  AA    OLGA testantake   s etd Oe DR RS GIN GU 8  4 5  TGA leak check nozzle                    seeeeeens 9  4 6  TGA CATS Ethernet Crossover Cable                     sssssssss 10  4 7  Serial data Cable sss ideae mes 10  4 8  TGA reference gas connection    13  4 9  TGA with insulated cover enclosure                      sss 13  4 10  TGA air sample intake                 ssssssssssseeeeneenee 14  4 11  TGA heated intake filter and orifice    14  4 12  TGA High Flow Filter holder                         see 15  4 13  RB0021 sample pump                sse 15  4 14  XDDI sample pump                sese ene 16  4  5  DOAV502 vacuum pump    nennen enne enne enne 17  4 16  DAAV505 vacuum pump    nennen nennen nennen 18  4 17  PDIT air sample dryer                  ssssssseeeeenee 20  4 18  PDIT 1 5 air sample dryer    20    Table of Contents    PD200T air sample dryer naan ken eene ROI erre 21  PD625 air sample diyet nenne e e E E aR 21
142. losure and in the TGA200A enclosure  AC  power input is required and the appropriate input cable type should be specified  at the time of order  The type of mounting hardware for the module is also  specified at the time of order     TGA Series Trace Gas Analyzers       FIGURE 4 3  TGA200A power module    4 1 1 2 TGA Accessory  amp  Tool Kit    The TGA200A comes with an accessory and tool pack  pn 15895  containing a  tube cutter  analyzer mirror position gauge  and screwdriver  as well as an  assortment of Swagelok  fittings  end wrenches  hex keys  and hose clamps   Many of these items are required during the installation of a trace gas analyzer     4 1 1 3 TGA Test Intake    The test intake for the TGA200A  pn 15838  includes a filter  needle valve   and 7 6 m  25 ft  of tubing for use as a sample intake  It is shipped with trace  gas systems and may also be purchased as a replacement part  The test intake  is shown in FIGURE 4 4        FIGURE 4 4  TGA test intake    4 1 1 4 TGA Leak Check Nozzle    A leak check nozzle for the TGA200A  pn 15836  includes 7 6 m  25 ft  of  tubing  and a needle valve and nozzle for detecting leaks around fittings and  tubing of a trace gas analyzer  Detecting leaks requires the user to supply a    TGA Series Trace Gas Analyzers    pressure regulator and gas tank  where the tank should contain a relatively high  concentration of the gas being measured  The leak check nozzle is shipped  with each trace gas analyzer and may be purchased as a re
143. many times  but  the assembly process is slightly different than the first assembly     1  Insert the tube with pre swaged ferrules into the fitting until the front  ferrule seats against the fitting body     2  Rotate the nut finger tight     3  While holding the fitting body steady  tighten the nut slightly with a  wrench     G 3 Common Replacement Parts    Tubing  Campbell Scientific can provide several types and sizes of plastic tubing as  shown in TABLE G 1  A tubing cutter  pn 7680  can be used to cut these                               tubes   TABLE G 1  Available Plastic Tubing Sizes  Construction  and Usage Guidelines  CSI pn   Tubing Type   OD  in    ID  in  Length  ft  Construction Notes  15702   Synflex 1300 1 4 0 17 500 Black HDPE Aluminum  jacket  overlapped   layer limits  15703 38 1   4 250 aluminum tape  diffusion  best  19164 1 2 3 8 250 ethylene for sample  copolymer liner tubes  26506   LLDPE 3 8 1 4 500 Black linear low  More flexible  density than HDPE  25539 1 2 3 8 500 polyethylene  19499   HDPE 5 8 1 2 100 Black High  Required for  density larger  polyethylene diameter                               Tubing inserts    Appendix G  Using Swagelok   Fittings    Inserts are recommended for use in plastic tubing  These inserts become  permanently attached to the tubing at the first assembly  so spare inserts may    be needed for replacing the ends of tubing     X  C          FIGURE G 1  Swagelok   insert                                              TABLE G 2
144. matically by the TGA software  To enable  this automatic control  click the Auto gain offset box on the Settings  gt   Detectors  gt  Preamp window  To set the offsets and gains manually  adjust  one of the values to disable the automatic function     The detector gains and offsets should usually be controlled automatically   However there are two exceptions  First  automatic control should be disabled  while performing some of the setup steps  such as optical alignment  Second   the automatic gain algorithm will not increase the sample gain beyond gain 7   Therefore  if the detector signals are extremely weak  it may be necessary to set  the detector gains and offsets manually     D 2 Detector Preamp Gain    The detector electronics have a fixed gain in the first preamplifier stage  Early  TGAs used a gain of 200  but this was changed to 45 for TGA200s and  TGA200As  This reduced gain allows the detector signals to be larger without  saturating the input range  This is an advantage for higher power lasers  and  for LN2 cooled detectors used for longer wavelengths  Some early input  modules have been modified to use a gain of 45 in the preamp  These  instruments are identified by a label on the printed circuit board  Gain 45   Input modules upgraded for use with TE cooled lasers have a gain of 45   These modules have a blue label on the top that designates them for use with  TE cooled lasers  see Appendix H 3 1  Basic Upgrade      The preamp gain  either 45 or 200  must be en
145. mum current     To avoid damaging the laser  make sure the laser cable    is not connected     5  Turn the analyzer electronics on     6  Sendthe new parameters to the TGA as you connect the TGA  software     7  Savethe new parameters in the TGA     8  Turn the analyzer electronics off     A 3    Appendix A  Configuring TGAs for Specific Gas Species    9  Connect the Laser Temperature  Laser Heater Cooler and the dewar   LN gt  cooled laser  connectors     To avoid damaging the laser  do not connect the Laser  Current connector at this time     10  Turn the analyzer on     11  Connect to the TGA with the TGA software  Receive the parameters    as you connect  This allows you to view the parameters stored in the  TGA     12  Verify all of the parameters are correct for the new laser  If any  parameters are incorrect  adjust the parameters as needed and repeat  steps 7 to 12     Do not proceed until you verify the correct parameters  are stored in the TGA s non volatile memory     13  Turn the analyzer off and connect the Laser Current connector   14  Turn the analyzer on     15  Connect to the TGA with the TGA software  Receive the parameters  as you connect     The new laser will now be active     To complete the installation  change the reference gas  Appendix A 2   Reference Gas  and detectors  Appendix A 3  Detectors  if necessary   and  then follow the steps in Appendix B  Optical Alignment  Appendix A 4   Finding the Absorption Line  Appendix C  Optimizing Laser Paramete
146. ncentration   oncentration  i  Analog Input  B  Other  TGA Temperatures   Data Output  Serial Numbers Absorption Cell Lengths  About TGA Length of long sample cell  cm  0 00  Length of short sample cell  cm  146 60  Length of reference cell  cm  146 60  Standard Deviation  Standard deviation time 5 00             FIGURE 7 17  Calculation concentrations settings    The center section of this window displays information about the Reference  Gas Concentration  Enter the concentration of the reference gas  in ppm   This parameter is used as detailed in Section 4 2 6  Concentration Calculation   to derive the concentration of the air in the sample cell  This parameter may be  adjusted as described in Appendix E  Calibration  to improve the accuracy of  the TGA  If the number of ramps is greater than one  each ramp will have a  reference gas concentration parameter     The Absorption cell lengths are also used to calculate the concentration of the  air in the sample cell  See Section 4 2 6  Concentration Calculation  for a  discussion of how these parameters are used in the calculation  and see Section  5 2  Physical Specifications  for the correct lengths to enter for each of the  TGA models     The TGA software calculates a mean and a standard deviation of the  concentration measurements  These statistics can be displayed either  graphically  Section 7 2 6  Graph  or numerically  Section 7 2 7  Data   or they  may be saved to the data files  Section 7 2 8  Files  The standard dev
147. nd has been designed with minimal  dead volume to maintain frequency response when performing eddy   covariance measurements  Itis included as part of the PD200T sample air  dryer  and may be used in any application that requires a large capacity filter  with minimal dead volume  The filter holder is shown in FIGURE 4 12  This  filter holder replaces pn 9839  which was used with the PD1000 sample dryer     TGA Series Trace Gas Analyzers       FIGURE 4 12  TGA High Flow Filter holder    4 1 4 4 Sample Vacuum Pump    The TGA requires a sample pump to pull the sample and reference gases  through the TGA at low pressure  The actual flow rate and pressure required  will depend on the application  Two sample pump options are available from  Campbell Scientific  The XDDI has a capacity of 1 slpm at 50 mb  0 8 slpm  with 50 Hz power   and is adequate for low flow applications  The RB0021 L  has a capacity of 18 slpm at 50 mb  15 slpm with 50 Hz power   and is used for  high flow applications  The pumps are supplied with the tubing and fittings  needed to connect to the TGA  A brief overview of each of the pumps is given  in the following descriptions     RB0021 L Sample Pump   The RB0021 sample pump is an air cooled  direct drive  oil sealed  rotary vane  pump  It is modified with a special oil return line that allows continuous  operation with minimal loss of pumping capacity  and a 2 54 cm  1 0 in  inlet  connection  Specifications are given for 115 Vac  60 Hz  single phase power  
148. ned by the settings in the PC  software  TGA Windows or TGA TEC      Analog   The TGA can be equipped with an optional SDM to analog converter to  provide an analog version of the digitally filtered concentration measurements  for analog data acquisition systems  These analog outputs will be updated at  different rates  depending on whether the TGA 1s measuring 1  2  or 3 gases   The analog outputs are updated every 6 ms  167 Hz   4 ms  250 Hz   and 6 ms   167 Hz  for single  double  and triple ramp operation  respectively  The  sample rate will be determined by the user supplied analog data acquisition  system     F 1    Appendix F  TGA Frequency Response    F 3 Digital Filters    No units       The TGA has two types of digital filter available  a moving average and an EC  filter  optimized for eddy covariance measurements      The moving average filter option is generally used for low flow rate  applications  such as atmospheric profiles or soil chambers  The user selects  the averaging time  from 2 to 2000 ms  The TGA   s actual measurement time is  2 ms  so setting the averaging time to 2 ms gives the original  unfiltered  data   The averaging time is normally set to the sampling interval  For example  if  data are collected by a datalogger at 10 Hz  100 ms sampling interval  the  moving average should be set to 100 ms  All of the original 500 Hz  measurements are thus represented in the averages saved by the datalogger     The EC filter is a finite impulse response  FIR 
149. needed     TGA Series Trace Gas Analyzers       Parameters could not be sent     l x  Connection Refused  Incompatible Serial Numbers    The TGA serial number in the parameter file  1112  does not match the serial  number of the TGA  1113  at IP address  192 168 76 31      Parameters are not sent           FIGURE 7 3  TGA error message for incompatible serial numbers    CAUTION Use extreme caution when sending parameters to the TGA  to make sure the correct parameters are being sent  If the  wrong parameters are sent for the laser in the TGA  the laser  may be damaged     After sending parameters  parameter synchronization will automatically default  back to Receive Parameters from TGA     Choose an update interval  which defines how often data are sent from the  TGA to the PC  The default is 100 ms  If the connection is slow the  connection may be unreliable  In this case  a longer interval  such as 1000 ms   may help to maintain the connection  The update interval may be changed  while the TGA is disconnected or connected     Once the TGA is connected  the Connect icon will change to reflect this as  shown in the bottom panel of FIGURE 7 4  If defined previously  see Section  6 2 3  Configure Ethernet Connection   the identification string will appear at          the top of the toolbar    eee  e  a   We 0X mw EE E ld   Connect   Status   Settings   Laser   Find   Graph   Dat Files           KO TGA TEC   Connected to Lab TGA200        Ea    WD  A   ov Ez  8  id    Connection
150. neration TGAs to TE cooled Laser    mating connector for these cables  see Appendix H 3 3  Input and Output  Modules   This newer cable provides better immunity to RF interference   Some TGA100s have already been upgraded to this style cable to improve RF  immunity  This newer style cable is required for TE cooled lasers  because it  has the proper connectors to mate with the updated input and output modules   The older  TGA100  cable may be replaced with the TGA100A Detector Cable  Set  pn 17897  Mounting this cable also requires new screws and washers  pn  18000  qty 6  and pn 2146  qty 6   This cable also requires that the detector  holder be of the later design     Early TGA100s had a two piece design for the short absorption cells and the  detector holders  This was replaced in 2002 by a one piece design that allowed  easier coalignment of the reference detector to the sample detector  The newer  style absorption cell detector holder is required for the TE cooled laser   because it is compatible with the TGA100A detector cable assembly  See  FIGURE H 8 and FIGURE H 9 to see the difference between these two  designs  Contact Campbell Scientific for availability of the newer design        FIGURE H 8  Older style two piece detector holder and short cell    H 8    Appendix H  Upgrading Early Generation TGAs to TE cooled Laser        gt      FIGURE H 9  Newer style combined detector holder short cell  shown  with newer style cable     H 3 5 Temperature Controller    Early TGA
151. nfiguration should result in both detectors  seeing the same amount of absorption     Appendix D  Optimizing Detector Parameters    TGA200 and TGA200A   Change the tubing connection outside the TGA enclosure to flow  the reference gas through both the sample cell and the reference  cell in parallel  Disconnect the sample and reference tubes   Insert a tee and two short tubes to connect the reference gas to  both the reference inlet and the sample inlet     This will split the reference flow to go through both absorption  cells in parallel  resulting in both detectors seeing the same  amount of absorption     Adjust the pressure in the TGA to give reference transmittance near  90   70 to 95  is also acceptable   Usually  this is accomplished by  changing the sample flow or by adjusting a bleeder valve at the pump     Adjust the reference flow to make sure there are no leaks  View the  reference and sample transmittance as you adjust the reference flow  from 10 to 50 ml min  A significant change indicates there may be a  leak     Temporarily change some parameter settings for this test     e Reference gas concentration  ppm   1000  each ramp if in  multiple mode     e Length of long sample cell  cm   0  e Length of short sample cell  cm   100  e Length of reference cell  cm   100    Start a Graph window to observe the mean concentration  The  reference and sample detectors are measuring absorption of the same  reference gas  through the same path length  Therefore  the measured
152. nsmittance measurement is affected by detector  nonlinearity  see Appendix D 4  Detector Linearity Coefficients    The reference detector is more linear because it has a smaller  signal  Therefore this measurement should be based on the  reference detector  not the sample detector     To achieve the optimum amount of absorption  first note the reference  transmittance with reference gas in the reference cell only  the normal  configuration   Normally  the reference gas concentration is chosen to give  approximately 50  absorption  For this test  however  it is best to have  between 70  and 80  absorption  Usually this can be accomplished by  reducing the sample flow to reduce the pressure in the analyzer  When the  proper absorption is achieved  record the sample cell pressure     It is acceptable for the transmittance to be less than 70   but it    must NOT be greater than 80      To configure the instrument for the multimode test  turn off the sample pump   swap the reference gas and sample connections  and turn the pump back on   This will put the reference gas in the long sample cell and the air sample in the  short reference cell  The total flow to the sample pump should be the same as  before  so the pressure should also be the same  Verify the pressure is within  10  of the previous pressure     Adjust the laser modulation current as needed     the increased absorption may  make the line wider  Record the reference transmittance  shown in the Laser  Settings window  Th
153. nt length of tubing inside the analyzer  enclosure  This brings both sample and reference gas to the temperature of the  inside of the enclosure     The absorbance of the reference gas depends primarily on the line strength of  the selected absorption line  the concentration of the reference gas  and the path  length  Pressure and temperature also affect the reference absorbance  The  reference gas concentration should be chosen to give an absorbance  in the  center of the absorption line  of 0 3 to 0 9  transmittance of 75  to 40    If  the absorbance is significantly more or less than this  the concentration noise  may increase  Suggested reference gas concentrations for the most commonly  measured gases are listed in TABLE A 4                                                        NOTE A higher reference gas concentration is required for the TGA100  and TGA100A because they have a shorter path length for the  reference cell   TABLE A 4  Suggested Reference Gas Concentrations  Gas Species TGA100 or TGA200 or Balance of  P TGA100A TGA200A Tank  Methane  CHa  15 000  1 5     500 N2  Nitrous Oxide  N20  2 000 60 Air or N2  N20 CO2 N20 2 000 90  Air or N2  CO2 300 000  30     15 000  1 5    N20 CH4 N20 10 000 350  m  2  CH4 20 000 850   Croon Dioxide  CO2  isotopic ratios  100 000  10     2500 Air   C only  Carbon Dioxide  CO    isotopic ratios  x d f  5180 and 83C 300 000  30     10 000  1   Air  Ammonia  NH3  5 000 160   Air or N2  Water or other Contact Campbell Scientific  
154. ntific to facilitate repair or replacement     e Immediately check package contents against shipping documentation   Thoroughly check all packaging material for product that may be trapped  inside  Contact Campbell Scientific about any discrepancies  Model  numbers are found on each product  On cables  the model number is often  found at the connection end of the cable  Check that correct lengths of  cables are received     e The TGA200A ships with the separate items listed in TABLE 3 1        TABLE 3 1  Parts Included with the TGA200A       Part Number Description  15895 TGA Accessory  amp  Tool Pack  15836 TGA Leak Check Nozzle  25ft tubing  15838 TGA Test Intake  5ft tubing  22178 TGA200A SDM Cable  20ft  Raw Plastic Tubing 1 4 in OD X  040 Wall                               pee Polyethylene Alum   18148 10Base T CATS Ethernet Crossover Cable  25ft  20730 9 Pin Female to 9 Pin Male Serial Data Cable  25ft  30723 TGA TEC Support Software  amp  OS   30981 TGA200A Power Module                The optical source of Campbell Scientific TGAs is a tunable diode laser that is  simultaneously temperature and current controlled to produce a linear  wavelength scan centered on a selected absorption line of the trace gas  A  beamsplitter allows most of the energy from the laser to pass through a 1 5 m   4 9 ft  sample cell  where it is absorbed proportional to the concentration of  the target gas  The portion of the beam that is reflected by the beamsplitter  passes through a reference
155. oalignment    Once the optical alignment has been optimized for the sample detector  check  the coalignment of the reference and sample detectors  Ideally  the sample and  reference detectors are optically coincident  and adjusting the horizontal and  vertical alignment gives a maximum response for both detectors at the same  position  Evaluate this by watching both detector signals while adjusting the  horizontal and vertical alignment  If they are not coincident  the reference  detector alignment must now be adjusted     The process is different for older and newer systems     Older systems    The beamsplitter mount can be rotated to adjust the vertical coalignment of the  reference detector to the sample detector  Loosen the three beamsplitter  clamping screws  rotate the beamsplitter mount to maximize the reference  detector signal  and retighten the beamsplitter clamping screws  It is  recommended that the system be at normal operating pressure  vacuum pump  on  for this step  If it is not possible to achieve adequate signal on the  reference detector signal by rotating the beamsplitter  it may be necessary to  make a small adjustment to the horizontal  vertical  and axial alignment to  reach a compromise between the reference detector signal and the sample  detector signal     Newer systems   These systems have a combined reference cell detector holder that includes  horizontal and vertical alignment cams  For these systems  the beamsplitter  mount should be rotated to 
156. of the  laser s power at the center of the absorption line  The side mode power at    NOTE    NOTE    Appendix C  Optimizing Laser Parameters    other frequencies will generally not be absorbed  The measured transmittance  at the center of the absorption line gives an estimate of the laser multimode  power  Two different methods are used depending on the TGA model    TGA100 and TGA100A   TGA100s and TGA100As have a long sample cell and a short reference cell   The laser   s multimode power can be estimated by temporarily putting reference  gas in the long sample cell  This increases the path length by a factor of almost  34  This test is best performed in conditions in which there is not an excessive  amount of absorption as this can lead to two possible problems     First  the absorption lines become broader  and absorption in the tails of the  absorption line can reduce the response at the edges of the spectral scan that are  assumed to be 100  transmittance  This will give an error in the estimate of  multimode power  especially if there is another absorption line nearby     Second  too much absorption will increase the chances of absorbing the  multimode power in some other absorption lines of the gas  The correct  absorption can usually be achieved by adjusting the pressure in the sample cell     To begin  go to the Settings  gt  Laser  gt  Other window and set the Multimode  power   to zero  Then note the reference transmittance in the Laser Settings  window     The tra
157. of the keyboard  to increase or decrease the value    The DC current and Mod current have automatic functions that can adjust  the value  Click the button to the right of the up down arrow buttons to enable  these automatic functions     NOTE The TGA Windows software also has an Auto button for Zero  current  The suggested procedure for setting the Zero current  does not use the function  see Appendix C 2  Zero Current   This  button has been removed from TGA TEC     70    TGA Series Trace Gas Analyzers    The DC current and modulation current can be adjusted by clicking inside the  graph in the window and then using the left and right arrow keys to adjust the  DC current  pan right left  and using the up down arrow keys to adjust the  modulation current  zoom in out      The Laser window for TGA Windows and TGA TEC are nearly identical  only  the units for the laser temperature are different  The TGA TEC window is  shown in FIGURE 7 26           Laser Settings  View Less Tabbed Detrended v Colors    CH4  DCcurent     8814   9y   Mod current       Lasertemp 17 00   17 00   C          FIGURE 7 26  The TGA TEC screen for setting laser parameters    7 2 4 1 View Less View More    The View Less View More toggle allows the expansion of what is shown in the  window  View Less shows only the DC current for each ramp  View More also  shows the modulation current  zero current  high current  and laser temperature  settings  The left panel of FIGURE 7 27 shows View Less  the right pane
158. or 2  on the laser side      4  Set both setpoints well below the measured values  Use a voltmeter to  measure the voltage across each heater  The voltage should be  approximately 0 V     NOTE The temperature control module works by switching the low side  to ground  When the heater is off  both terminals will be at  approximately 48 V with respect to ground     5  Set the TGA Temperature 1 setpoint well above the measured  temperature  Verify the corresponding duty cycle increases to 1     Appendix I  Install Temperature Control Upgrade    6  Measure the voltage across Heater 1  It should now be approximately    48 V   NOTE The heater does not greatly increase the temperature of the air  flowing through the fan  It is usually not possible to feel this  temperature increase by hand   7  Repeatthis test with TGA Temperature 2 setpoint and Heater 2   Operation    Set the temperature setpoints to the desired value  typically 35  C  They will  normally be set to the same value  The Duty Cycle Heater 1 and Duty Cycle  Heater 2 parameters will be automatically adjusted to control the temperatures   These duty cycles determine the fraction of the time the heaters will be turned  on for each pulse  The pulse repetition rate is determined by the Pulse Period  parameter     Put the enclosure lid on the TGA  The temperature should warm up to the  setpoint and stay there  This warm up typically takes approximately one hour     When the temperature is stable  verify that the two duty cycl
159. ough water vapor  at a sample intake to avoid condensation in the downstream tubing          raer    FIGURE 4 18  PD1T 1 5 air sample dryer    The Nafion   tube in the PD1T 1 5 is 1 5 ft long instead of the 6 ft length used  in the PDIT  It will dry 0 5 Ipm to 0   C dewpoint  compared to    15   C for the  PDIT     The sample inlet and outlet tubes are the same 1 8 in OD SS tubes as for the  PDIT  but are shorter than on the PDIT  The same design for the internal  connection to the Nafion   tube eliminates dead volume     The purge connections are 3 8 in Swagelok    and the dryer shell is 3 8 in OD  stainless steel tubing  The smaller shell reduces cost and the stainless steel  tube is more rugged for mounting at the sample intake     TGA Series Trace Gas Analyzers    PD200T   The PD200T consists of a 200 tube  48 in Nafion   dryer element manufactured  by Perma Pure  Inc   that is housed in a rugged dryer shell designed and  manufactured by Campbell Scientific  The PD200T includes a filter holder  a  flow meter to measure purge flow  needle valves to adjust the sample and  purge flow rates  and mounting hardware  Common accessories are spare filter  membranes  pn 9838  and a 4     40 lpm flow meter  pn19541  to measure the  sample flow  The PD200T is shown in FIGURE 4 19        FIGURE 4 19  PD200T air sample dryer    PD625   The PD625 is similar to the PD200T  but is designed for lower flow rates  Its  50 tube  24 in Nafion   dryer element has a drying capacity one eighth th
160. ower Module  Mounting Brackets     Remove the cable feedthrough cap from side of the TGA enclosure  and insert one end of the DC power cable  Plug the DC power cable  into the mating connector on the electronics box and secure the cable  to the corner of the electronics box as shown in FIGURE 6 11        FIGURE 6 11  DC power cable connected to TGA200A and secured on  electronics box    Replace the cap of the cable feedthrough by pushing it on  rotating it  to fit snugly against the cables  and tightening the thumbscrew     Secure the SDM and power cabling by routing along the bottom of the    feedthrough bracket and securing with cable ties as shown in FIGURE  6 12     37    TGA Series Trace Gas Analyzers       FIGURE 6 12  Routing of SDM and power cable through TGA200A  feedthrough bracket    5  Remove the feedthrough cap from the bottom of the power module   Insert the other end of the DC power cable into the power module and  connect it to its mating connector     6  Connect the power module to AC mains power  90 to 264 Vac  47 to  63 Hz  using the detachable power cord     NOTE If a long AC power cord is required  have a qualified electrician  connect the field wireable plug  pn 28771  to a user supplied cord     7  Replace the cap of the cable feedthrough by pushing it on  rotating it  to fit snugly against the cables  and tightening the thumbscrew  The  connected power module enclosure is shown in FIGURE 6 13     38       TGA Series Trace Gas Analyzers       FIGURE 6 1
161. ower and the mode hop  characteristics of a laser may change dramatically with temperature  Because  both temperature and current determine the emission frequency  changing the  current can compensate for a change in temperature  The goal in setting the  laser temperature is to find the combination of temperature and current that  minimizes multimode operation and avoids mode hops     The following discussion uses an LN2 cooled laser as an example   but the process is the same for all lasers  TE cooled laser  temperatures are given in units of   C instead of K     In principle  this is straightforward  but it is complicated by the iterative nature  of the process  All of the other laser parameters must be set to reasonably  appropriate values in order to evaluate the laser temperature  but the optimum  value of some of those parameters depend on temperature  To begin  set the  other laser parameters as follows     1  Set the zero current as described in Appendix C 2  Zero Current  but  then reduce it by approximately 20  before setting the other  parameters  This will help to avoid confusion caused by the laser   s  lower threshold current at lower temperature     2  Set the high current count and the omitted data count to their  maximum values     3  Set the high current as described in Appendix C 3  High Current     4  Set the modulation current as described in Appendix C 5  Modulation  Current     5  Ifthe laser is used in multiple ramp mode  set the parameters for each  r
162. p Options Help    Device Type    E E mE E    TGA100A TGA200    Communication Port    v    Settings Editor i Terminal      TGA100A TGA200 OS Download Instructions    This page is used to download an operating system to the TGA9032 CPU module in the  TGAIO0A TGA200 trace gas analyzer   1  Remove power from the TGAIOOA TGA2O0   2  Connect a cable between one of your computer s serial ports and the RS 232 port on the  TGA3032 CPU module     Make sure that the appropriate serial port is selected in the left panel     Click the Start button below     In the resulting file open dialogue box  select the   003 file that should be sent as an  operating system     After you have pressed the Open button in the file open dialogue  press and hold the white  card eject button on the TGA9032 module  see the image below  while applying power to    the TGATO0DA TGA200   Press and hold white button    While turning power on       Baud Rate    Start Print Instructions                FIGURE 6 16  TGA100A TGA200 OS download instruction    The TGA runs a program similar to a CR9000X datalogger program  This  program is installed at the factory and normally does not need to be updated  when the operating system is updated  However  if this file becomes corrupted  or needs to be updated  connect to the TGA with the Device Configuration  Utility which will bring up the screen shown in FIGURE 6 17  From there  select the Logger Control tab  Click the Send Program button and send the  file tga cr9     4
163. p Timing  High current count 8 5  Omitted count 20 2             FIGURE 7 13  Settings of the Settings    Laser    Other screen    Number of ramps may be set to 1  2  or 3  corresponding to how many gases  are to be measured  see Section 4 2 5  Laser Scan Sequence   If the number of  ramps is decreased  the laser current parameters  DC  Mod  Zero  and High  current  for the unused ramp remain in the parameter file  Increasing the  number of ramps gives the ramp synchronization prompt shown in FIGURE  7 14     To use the parameters that were previously stored in the parameter file for  RAMP B  click No  Click Yes to overwrite the new ramp s laser current  parameters with those of ramp A  This    synchronizes    the new ramp to  RAMP A  This is a good choice for initial set up of the laser for multiple   ramp mode  or if it is not clear to the user what parameters are in the file for  the new ramp     7 2 3 2 Detectors    TGA Series Trace Gas Analyzers       Increasing Number of Ramps WB    Would you like to syncrhonize RAMP B with RAMP A           FIGURE 7 14  Ramp synchronization prompt    The middle section of the Settings  gt  Laser  gt  Other window  FIGURE 7 13   will have a tab for each ramp  Click the tab to bring it to the foreground  Each  tab has two parameters  Gas name and Multimode power    Gas name is a  label that is used throughout the TGA software to designate what gas 1s  measured  for example  after defining these settings the TGA will  automatically label t
164. pbellsci cc    Campbell Scientific Ltd   CSL   Campbell Park  80 Hathern Road  Shepshed  Loughborough LE12 9GX  UNITED KINGDOM  www campbellsci co uk   sales campbellsci co uk    Campbell Scientific Ltd   CSL France   3 Avenue de la Division Leclerc  92160 ANTONY  FRANCE  www campbellsci fr   info campbellsci fr    Campbell Scientific Ltd   CSL Germany   FahrenheitstraBe 13  28359 Bremen  GERMANY  www campbellsci de   info campbellsci de    Campbell Scientific Spain  S  L   CSL Spain   Avda  Pompeu Fabra 7 9  local 1  08024 Barcelona  SPAIN  www campbellsci es   info campbellsci es    Please visit www campbellsci com to obtain contact information for your local US or international representative     
165. placement part as  well  The leak check nozzle assembly is shown in FIGURE 4 5        FIGURE 4 5  TGA leak check nozzle    4 1 1 5 TGA SDM Cable    The TG200A includes a 6 m  20 ft  cable  pn 22178  used to connect the SDM  ports on a trace gas analyzer to the SDM ports on a datalogger  The cable  includes three wires for SDM signals  one wire for ground  and a shield  One  end has untinned pigtail wires used to connect semi permanently to the trace  gas analyzer s SDM connector  The other end has tinned pigtail wires used to  connect to the ports on a datalogger or SDM hub     4 1 1 6 Plastic Tubing    The TGA200A includes 6 m  20 ft  of 1 4 in OD Synflex 1300 tubing  pn  15702  which has a high density polyethylene jacket  overlapped aluminum  tape  and ethylene copolymer liner  The aluminum layer limits diffusion of  gases through the wall of the tube  making it the best option for delivering  sample air from an intake to a gas analyzer     4 1 1 7 CAT5 Ethernet Crossover Cable    A 7 6 m  25 ft   unshielded CATSe crossover Ethernet cable  pn 18148  is  included  The Ethernet cable should be used when connecting two Ethernet   capable products directly together  or when connecting one directly to a PC  It  is not intended for use with hubs  switches  or routers  A CATSe crossover  Ethernet cable is shown in FIGURE 4 6     TGA Series Trace Gas Analyzers    cy     FIGURE 4 6  TGA CAT5 Ethernet Crossover Cable       4 1 1 8 Serial Data Cable    A 7 6 m  25 ft  9 pin female to
166. pn 15837  or similar hardware     In the TGA100A or TGA100A the purge connection is at the end of the  enclosure  The purge gas flows through the short sample cell and into a purge  boot around the air gap between the laser and lens holder  as shown in  FIGURE A 1     A 7    Appendix A  Configuring TGAs for Specific Gas Species          Reference  defector  m      No  To pump E i               Reference gas in Dewar  LE AN  EN   Sample cell J N f N  p s     Y  Sample gy   wm Laser    detector       J          S   j  Jj   N J  N2 CURT To pump Sample in    FIGURE A 1  TGA100 or TGA100A optical layout with air gap purge    In the TGA200 or TGA200A the purge connection is in the middle of the  enclosure  The purge gas splits  flowing to the detector block and the  beamsplitter block  The flow through the beamsplitter block purges the air gap  between the lens holder and the laser  This configuration is shown in FIGURE    A 2     detector    Detector  alignment  mirrors       Reference  alignment  mirror       Reference  detector    FIGURE A 2  TGA200 or TGA200A optical layout with air gap purge    A 8    Appendix B  Optical Alignment    Campbell Scientific TGAs have simple  robust optical designs that do not  require adjustments in normal use  The optical alignment may need minor  adjustments after transporting the system or if the laser is replaced  The optical  design of the earlier models  TGA100 and TG100A  is different than that of  the later models  TGA200 and TGA200A   See
167. r of different gases   Unfortunately  these lasers were discontinued by the manufacturer in 2012   For reference  the part numbers and gases measured are listed in TABLE A 1                                      TABLE A 1  Discontinued Cryogenic Lead salt Lasers  Part Number   pn  Target Gas es  Molecular Formula  7979 Methane CH4  7113 Nitrous Oxide N20  21400 Methane and Nitrous Oxide CH  and N20  21401 Nitrous Oxide and Carbon Dioxide   N20 and PC 60 90  17466 Carbon Dioxide and 8 C 2C160 60 and  C 60 60  17469 Carbon Dioxide  8C  and 5 8O 20160160  130160160  and 120180160  21398 Water Vapor  5D  and 8  O H H  O  7H H   O  and  H H 8O  21399 Ammonia NH3                A 1 2 TE cooled Lasers    The TGA200A uses TE cooled lasers that became available in 2014  These  lasers are available at wavelengths from 3 to 6 um  allowing the TGA200A to  measure the gases most commonly measured with earlier TGAs  In most cases  earlier TGAs may be upgraded with a new TE cooled laser to avoid the need    A 1       Appendix A  Configuring TGAs for Specific Gas Species    for cryogens  See Appendix H  Upgrading Early Generation TGAs to TE   cooled Laser  for details  Part numbers for these lasers are listed in TABLE  A 2                                   TABLE A 2  TE cooled Lasers  Part Number   pn  Target Gas es  Molecular Formula  30477 Methane CH4  30478 Nitrous Oxide N20  31121 Nitrous Oxide and Carbon Dioxide   N20 and PC 60 50  31119 Carbon Dioxide and 68 C 12C160 60 and PBCISO 60  3
168. r temperature stability   Some TGA100As have already been upgraded in conjunction with upgrades to  the TGA Windows software  This upgrade is not required for using TE cooled    H 5    Appendix H  Upgrading Early Generation TGAs to TE cooled Laser    lasers  but it is recommended for all TGA100As  See Appendix I   nstall  Temperature Control Upgrade  for details     H 2 5 Power Module    TGA100As shipped with two AC DC power adapters mounted under the  electronics mounting bracket  One of these power adapters supplies 12 Vdc  power to the electronics and the other supplies 48 Vdc power to the  temperature controller     In the event that a TGA100A power supply fails  an upgrade is available  This  upgrade includes a special version of the pn 30981 power module that has been  modified to supply 48 Vdc for the heaters instead of the 24 Vdc used be later  TGA models  TGA200 and TGA200A   Contact Campbell Scientific for  details     H 2 6 Purge Boot    H 3 TGA100    Some TGA100As included an optional purge boot between the laser dewar and  the optical assembly  The purge boot mounted to the laser dewar and sealed  around the front lens holder to enclose the air gap between the laser dewar and  thelens  The purge boot can be mounted to the TE cooled laser assembly in  the same way as for the laser dewar  If the purge boot becomes damaged  contact Campbell Scientific and request pn 15902     H 3 1 Basic Upgrade    Most TGA100s shipped since 2000 may be upgraded to use TE cooled las
169. reference and sample cells have the same length  see TABLE 5 2 for a  summary of these specifications for all systems   The back end of each cell has  a focusing lens  alignment mirror  and detector     Sample  detector    Detector  alignment    mirrors        Reference  alignment  mirror       Reference  detector       FIGURE 4 26  TGA200 and TGA200A optical configuration    24    4 2 2 Laser    DANGER    4 2 3 Dewars    4 2 4 Detectors    TGA Series Trace Gas Analyzers    The lead salt diode lasers used in the TGA100  TGA100A  and TGA200  required cryogenic cooling  These lasers were available at any wavelength  from 3 to 10 um  which could be specified to detect any one of many distinct  species of different gases  Lead salt tunable diode lasers have a limited tuning  range  typically 1 to 3 cm  within a continuous tuning mode  In some cases  more than one gas can be measured with the same laser  but usually each gas  requires its own laser  Most of these early TGAs used liquid nitrogen  LN   to  cool the laser  but some TGA100As used a cryocooler  The laser dewars and  cryocoolers have two laser positions available  four with an optional second  laser mount   allowing selection of up to four different species by rotating the  dewar  installing the corresponding cable  and performing a simple optical  realignment     In 2012  when the only manufacturer stopped production  lead salt diode lasers  became unavailable  In 2014  however  a thermoelectrically cooled interband  c
170. rm is available from our web site at www campbellsci com repair  A  completed form must be either emailed to repair campbellsci com or faxed to   435  227 9106  Campbell Scientific is unable to process any returns until we  receive this form  If the form is not received within three days of product  receipt or is incomplete  the product will be returned to the customer at the  customer   s expense  Campbell Scientific reserves the right to refuse service on  products that were exposed to contaminants that may cause health or safety  concerns for our employees     Precautions    DANGER     MANY HAZARDS ARE ASSOCIATED WITH INSTALLING  USING  MAINTAINING  AND WORKING ON OR AROUND  TRIPODS  TOWERS  AND ANY ATTACHMENTS TO TRIPODS AND TOWERS SUCH AS SENSORS  CROSSARMS  ENCLOSURES   ANTENNAS  ETC  FAILURE TO PROPERLY AND COMPLETELY ASSEMBLE  INSTALL  OPERATE  USE  AND MAINTAIN TRIPODS   TOWERS  AND ATTACHMENTS  AND FAILURE TO HEED WARNINGS  INCREASES THE RISK OF DEATH  ACCIDENT  SERIOUS  INJURY  PROPERTY DAMAGE  AND PRODUCT FAILURE  TAKE ALL REASONABLE PRECAUTIONS TO AVOID THESE HAZARDS   CHECK WITH YOUR ORGANIZATION S SAFETY COORDINATOR  OR POLICY  FOR PROCEDURES AND REQUIRED PROTECTIVE  EQUIPMENT PRIOR TO PERFORMING ANY WORK    Use tripods  towers  and attachments to tripods and towers only for purposes for which they are designed  Do not exceed design  limits  Be familiar and comply with all instructions provided in product manuals  Manuals are available at www campbellsci com or
171. rrules                   G 3  G 4  Dimensions and Part Numbers for Swagelok   Plugs                       G 4  G 5  Dimensions and Part Numbers for Swagelok  Caps                        G 4  I 1  Control Parameters for TGA Thermistors                    sseeeeess I 4    vii    Table of Contents    viii    TGA Series Trace Gas Analyzers    1  Introduction    Campbell Scientific has been manufacturing tunable diode laser absorption  spectrometer  TDLAS  trace gas analyzers  TGAs  since 1993  While the  TGAs have improved through a succession of four different generations  the  core technology remains the same  The TDLAS technique provides high  sensitivity  speed  and selectivity     All Campbell Scientific TGAs are rugged  portable instruments designed for  use in the field  Common applications include gradient or eddy covariance  flux measurements of methane or nitrous oxide  and isotope ratio  measurements of carbon dioxide  Many of the important improvements are  available as upgrades for older models  These substantial upgrades blur the  lines between the various models  This manual covers all of the TGA models  to some extent  with emphasis on the later variants  A brief summary of these  TGA models and the most relevant improvements are summarized in TABLE    1 1        TABLE 1 1  Historical Summary of Campbell Scientific Trace Gas Analyzers                            TGA100 TGA100A TGA200 TGA200A   Ship dates 1993     2004 2005     2009 2008     2012 2014    Transputer 
172. rs   Appendix D  Optimizing Detector Parameters  and Appendix E  Calibration     A 2 Reference Gas    A prepared reference gas having a known concentration of the gas to be  measured must flow through the reference cell  The beamsplitter directs a  fraction of the laser power through the reference cell to the reference detector   This gives a reference signal with the spectral absorption signature of the  reference gas  The reference signal provides a template for the spectral shape  and position of the absorption feature  This allows the concentration of the  sample gas to be derived without measuring the temperature or pressure of the  sample gas  or the spectral positions of the scan samples  The reference signal  provides feedback for a digital control algorithm to maintain the center of the  spectral scan at the center of the absorption line  line locking   The reference  signal also allows the user to identify the wavenumber of an absorption line by  comparing it to the theoretical absorption spectrum of the gas     The reference cell is kept at the same pressure as the sample cell by connecting  the outlets of both cells to a common vacuum manifold  A continuous flow of   reference gas must be maintained to avoid dilution of the reference gas with the  sample gas  A flow of 10 ml min is recommended     Appendix A  Configuring TGAs for Specific Gas Species    The reference gas and sample gas are brought to the same temperature by  flowing each of them through a sufficie
173. s at the TGA use a five pin green  connector as shown in FIGURE 6 8      12 not connected   G   SDM reference  black  and SDM shield  clear   C1   SDM data  green    C2   SDM clock  blue    C3   SDM enable  yellow     FIGURE 6 8  SDM connections of TGA    35    TGA Series Trace Gas Analyzers    The SDM cable connector attaches to the CPU board in the TGA as shown in  FIGURE 6 9     9010  FIDWER SUPPLY       FIGURE 6 9  SDM cable connector on TGA CPU board    Once the SDM cable is connected to the CPU board  secure the cable to the  outside of the electronics box at two places as shown in FIGURE 6 10               FIGURE 6 10  SDM cable tied to electronics box    Connect the other end of the SDM cable to a datalogger     To configure the TGA for SDM output  connect a PC running TGA  Windows TGA TEC  Open the window Settings  gt  Other  gt  Data Output     36    6 1 3 Power    TGA Series Trace Gas Analyzers    Click SDM Output as shown in FIGURE 7 46 of Section 7 3 1  SDM Output   and enter the SDM address for the TGA     The TGA100  TGA100A  and TGA200 used AC DC adapters inside the  analyzer enclosure to provide power for the analyzer and the temperature  controller  Connect these to AC mains power using the detachable power  cords supplied     The TGA200A includes a TGA power module  pn 30981  which is a separate  enclosure to house the AC DC adapters     1     Mount the TGA power module near the TGA  Various mounting  options are available as described in Section 4 1 2 3  P
174. s heat is needed to  maintain the laser at the set temperature  and the laser heater voltage  will gradually decrease  Therefore  monitoring the laser heater  voltage may give an indication of when it is time to evacuate the  dewar  If the TGA is equipped with a TE cooled laser  record the  laser cooler voltage     Verify that the concentration and concentration noise are as expected     If the TGA is equipped with a LN2 cooled laser  fill the laser dewar  with liquid nitrogen as needed  If the TGA is equipped with LN2   cooled detectors  fill the detector dewar as needed     Check the reference gas tank and regulator pressure  Check other  tanks  air gap purge  calibration  etc   as needed  Monitor the change  in pressure as these gases are consumed  to gauge when to order  replacement tanks and to identify possible leaks     7 1 3 Shutdown Procedure    This section describes the routine shutdown procedure for the TGA     l     If a PC is connected to the TGA  turn off data collection if data is  being collected  and quit the TGA program     Shut off the TGA sample pump  and bypass pump  if applicable    Shut off power to the TGA    Shut off reference gas supply    Shut off air gap purge supply  if applicable     Shut off calibration gas supplies  1f applicable     TGA Series Trace Gas Analyzers    If the TGA is equipped with a LN2 cooled laser  there are two recommended  shutdown states the TGA should be left in depending on how soon the TGA  will be put back into use     e
175. s through the insides of the tube  or bundle of tubes  and the purge air  flows outside of the tubes in the opposite direction  The water vapor is forced    through the wall of the tubes by a difference in vapor pressure  The sample  becomes progressively drier as it travels down the dryer  while the purge air  becomes progressively more humid  For best performance  the purge flow  should be very dry     40   C dewpoint   and should have an actual flow rate of  at least twice the sample flow  Although the purge flow could be supplied by  air from a compressed air tank or by ambient air dried with a chemical  desiccant  for most TGA applications it is provided by the dryer itself  A  portion of the sample flow is split off at the outlet of the dryer and pressure is  reduced by connecting the purge outlet to the TGA sample pump  Dropping  the pressure reduces the partial pressure of the water vapor and increases the  actual flow rate  allowing the purge requirements to be met with just a fraction  of the sample flow  More information on the dryer can be found at  www permapure com        TABLE 4 4  Sample Air Dryer Specifications                                                    Specification Units PDIT 1 5 PDIT PD625 PD200T  Campbell Scientific pn 21772 19206 16315 19200  Length cm  in  72  28  239  94  76  30  137  54   Weight kg  Ib  0 4  0 9  0 7  1 5  4 4 9 7  5 6  12 4   Connections    e Sample Inlet 1 8  tube 1 8  tube 1 4    Swagelok  3 8  tube  e Sample Outlet 1 8  tu
176. sband  The EC filters reduce noise by three to  four orders of magnitude  The 100 ms moving average reduces noise by    approximately one order of magnitude     CYWY            Yi AIL RANDE       i  40    i  60 80    Hertz    FIGURE F 3  EC filter frequency response  logarithmic scale     Appendix F  TGA Frequency Response    The EC filter passband should be chosen to preserve high frequencies that may  be in the data while attenuating noise  The optimum setting depends on the  TGA frequency response which  in turn  depends on the sample cell residence  time  see Appendix F 5  Sample Cell Residence Time   The recommended EC  passband setting as a function of sample cell residence time is given in TABLE  F 1        TABLE F 1  Recommended Passband Settings                      Residence Time  ms    Recommended Filter Bandwidth  Hz      100 5  100     150 4  150     200 3  200     300 2   gt  300 1                F 4 Synchronicity    NOTE    NOTE    For EC measurements it is very important to synchronize the wind vector data  from the sonic anemometer with the scalar data from the TGA  The FIR filters  introduce significant time delay  lag  into the concentration measurements   Because the lag is a constant  it can easily be removed by EC post processing  algorithms     In single ramp mode the EC filter processing lag is 372 ms  This lag is the  same regardless of the passband selected  but it is different for two ramp and  three ramp mode  The processing lags for the EC filters
177. served without absorption  Adjust the detector temperature  setpoints in the Settings  gt  Detectors  gt  Temperature window according to these  criteria     Sample Detector  For input modules with a preamp gain of 200     The maximum detector signal is approximately 52 mV  Set the sample  detector temperature as needed to give a sample signal of approximately  45 mV to make sure the input will not saturate     For input modules with a preamp gain of 45     The maximum detector signal is approximately 230 mV  but there is  generally little improvement in noise for signals above 50 mV  Set the  sample detector temperature for 50 to 100 mV sample signal     Reference Detector   Set the reference detector temperature to give a reference signal that is  approximately 10  of the sample signal  about 5 mV  This usually will be a  temperature close to that of the sample detector     Keep in mind that the TE cooled detectors can cool the detectors a maximum  of 80 to 90   C  Make sure the detectors can be maintained at the setpoint  even  with the TGA enclosure warmed up to its operating temperature     To first order  detector nonlinearity can be compensated using the detector  linearity coefficients  described in Appendix D 4  Detector Linearity  Coefficients  However  if concentration accuracy is more important than  precision  it is recommended to increase the sample detector temperature to  give a signal level of 35 to 40 mV  3 to 4 mV for the reference detector   This  may incr
178. splay    75    TGA Series Trace Gas Analyzers    Maximum View   Maximum View displays the reference and sample transmittance  scaled to the  maximum and minimum of all of the data  including the zero  high  and  omitted data      View Less Expand             17 00 r3 17 00   C       FIGURE 7 31  Maximum View mode of laser display    76    TGA Series Trace Gas Analyzers    Magnified   The Magnified view displays the reference and sample transmittance  scaled to  the maximum and minimum of the data used in the concentration calculation   including the zero  high  and omitted data   This mode is used to set High  current        Laser Settings     View Less Expand Magnified   Colo         CH4       17 00   17 00 C       FIGURE 7 32  Magnified mode of laser display    77    TGA Series Trace Gas Analyzers    78    Detrended   The Detrended display mode shows the data after they have been detrended by  fitting a line to the data and dividing by this line  The graph is scaled  horizontally to show only the data used for concentration calculations  This is  the default display mode         Laser Settings  View Less Expand Detrended x Colo       DC current 88 14 5  La  Mod current   1 30     Zero current   H    High current       17 00   17 00   C       FIGURE 7 33  Detrended mode of laser display    TGA Series Trace Gas Analyzers    Folded   Folded display mode is similar to the Detrended display mode  but the data  have been detrended differently  This mode shows data that are detren
179. t on the CPU module  not the RS232  port  and the RTD 1B and RTD 2A connectors on the temperature control  module  as shown in FIGURE I 5       a  caf 3 aya S MODULE     s    LH  di Te       FIGURE l 5  Control cable connection    Appendix I  Install Temperature Control Upgrade    1 3 Enter Control Parameters    1  Inthe TGA software  TGA Windows or TGA TEC   bring up the  Analog Input page in the Settings window     2  On the Channel 2 tab  select Thermistor  TGA200    3  On the Channel 3 tab  select Thermistor  TGA200      4  This will convert the measurements on analog inputs 2 and 3 to  temperature using the conversion equation for thermistor probes     5  Inthe TGA software  bring up the TGA Temperature page in the  Settings window     6  Verify the temperature readings next to the temperature setpoints are  approximately correct  These are not intended to be high accuracy    measurements  but they should be within 1 or 2   C     7  Setthe control parameters to the default values shown in TABLE I 1                       TABLE I 1  Control Parameters for TGA  Thermistors  Parameter Default Value  Pulse period 0 1  Control coefficient  P  2 0  Control coefficient  I  0 01  Control coefficient  D  20                l 4 Testing    1  Verify the TGA Temp 1 and TGA Temp 2 give reasonable values     2  Gently squeeze Thermistor 1  on the detector side  between finger and  thumb to warm the thermistor  Verify that TGA Temperature 1  increases     3  Repeat this test for Thermist
180. tance at the  Reese center of the spectral scan for ramp A     4    Reference detector transmittance at the  Beier ose center of the spectral scan for ramp B     i  2  Reference detector transmittance at the  Reme Trane center of the spectral scan for ramp C            RefDetTemp Reference detector temperature    C  4 1  Current applied to the thermoelectric cooler  RefDetCooler to maintain the reference detector at its 4 1  specified temperature  arb   RefDetGainOffset Gain and offset settings for the reference 4 1  detector preamplifier  arb     Sample detector signal at the center of the  penpDerwienel spectral scan for ramp A  mV  4      Sample detector signal at the center of the  PIGDDSUDIBFAIB spectral scan for ramp B  mV  2    Sample detector signal at the center of the  pppoe mena    spectral scan for ramp C  mV  1 2  Sample detector transmittance at the center  pip Deans of the spectral scan for ramp A     1 l  Sample detector transmittance at the center  Smp Dertigsh of the spectral scan for ramp B     5 2  Sample detector transmittance at the center  POPPEL ANE of the spectral scan for ramp C     1     SmpDetTemp Sample detector temperature    C  4 1  Current applied to the thermoelectric cooler  SmpDetCooler to maintain the sample detector at its 4 1  specified temperature  arb     Gain and offset settings for the sample  SEPP eatin detector preamplifier  arb  1 l  Fraction of full power applied to the TGA  Deae enclosure heater  detector end  arb       DutyCycle
181. tered in the parameters at the  Settings  gt  Detectors  gt  Preamp window  If it is unclear which gain is the  correct one for a specific input module  or if an input module upgrade is  desired  contact Campbell Scientific     D 3 Detector Temperature    Most TGAs are equipped with TE cooled detectors  but some are equipped  with LN   cooled detectors to work with longer wavelength lasers  LN2 cooled  detectors have no temperature measurement or control  They are designed to  operate near LN  s boiling point  77 K     TE cooled detectors include a thermistor to measure the detector s temperature   The TGA measures this temperature and adjusts the TE cooled current to cool  the detector to the setpoint  Generally  a lower detector temperature will  increase the detector signal and decrease the concentration noise  However   some lasers emit enough power to saturate the detectors 1f they are cooled to    D 1    Appendix D  Optimizing Detector Parameters    their lowest temperature  Cooling the laser can also increase detector  nonlinearity  Choosing the optimum temperature may involve a tradeoff  between precision  noise  and accuracy  nonlinearity      To choose the optimum detector temperatures  first set the reference and  sample linearity coefficients at the Settings  gt  Detectors  gt  PreAmp window to  zero  Observe the detector signals in the Laser Settings window  Increase the  DC current to move away from the absorption line  This allows the detector  signals to be ob
182. teria must be considered     e Ifthe high current is too high  the laser s frequency will  overshoot the absorption line at the beginning of the spectral  scan  quickly scan backwards through the absorption line   and then scan forward through the absorption line  This is  visible in the reference detector display as a second narrow  absorption line at the left edge of the spectral scan data  See  FIGURE C 6 and FIGURE C 7  This is acceptable as long  as this    ghost    line is in the omitted part of the scan  But if  the signal falls off at the left edge of the scan as illustrated in  the lower right of FIGURE C 6  the high current must be  reduced  even if the    counting steps    criterion indicates the  high current is already too low      e When incrementing the DC current  it may not be possible to  move the absorption line to the left edge because the two  absorption lines  the true line and its    ghost     may merge just  inside the left vertical dotted line  see FIGURE C 7  top  right   This also indicates the high current is too high     For multiple ramp mode  adjust the high current individually for each ramp  In  this case  the laser temperature is perturbed by the entire previous scan  not just  the zero pulse of the present scan  This means it will take a larger high current  pulse for the ramp with the higher DC current because the laser will have  cooled during the previous ramp at lower DC current  It may require a  negative value for the high current of t
183. tion Current    The laser modulation current parameter controls the width of the spectral scan   The edges of the spectral scan should extend slightly past the absorption lines  to measure the laser   s unabsorbed intensity  100  transmittance   Adjust the  modulation current until it occupies approximately one fourth of the spectral  scan  with a nearly flat portion on either side  as illustrated in FIGURE C 9   The TGA software includes an automatic algorithm to set the modulation  current  To use this algorithm  enable line locking and then click the Auto  button next to the Mod current on the Laser Settings window     Appendix C  Optimizing Laser Parameters          Modulation too low Modulation correct Modulation too high    FIGURE C 9  Adjustment of modulation current    For multiple ramp mode  the modulation current must be set individually for  each ramp  It may need to be set to a different value for each ramp to  compensate for residual temperature perturbation that cannot be completely  removed by the high current pulse     C 6 Laser Maximum Temperature and Laser    Maximum Current    The laser can be damaged by too much current or by operation at too high a  temperature  The TGA software will automatically disable the laser current if  the laser s temperature is above an upper limit or if the laser current parameters  are set to exceed an upper limit for any of the spectral scan points     Set the laser maximum temperature and current to the values specified on th
184. tions for the TGA200A are located at the  center of the TGA enclosure under the feedthrough cover  FIGURE 6 6   Lift  the cover following directions indicated on the cover to expose the  connections  Follow the steps below for this simple set up  Refer to FIGURE  6 7 as needed for the physical connections        TGA200A       FIGURE 6 6  Feedthrough cover of TGA200A    33    TGA Series Trace Gas Analyzers    34    NOTE       FIGURE 6 7  Plumbing connections located under feedthrough cover  of TGA200A    1  Connect the vacuum exhaust outlet of the analyzer to the sample  pump  The sample pump must be able to pull the required flow rate at  75 mb or less  The actual flow rate and pressure required will depend  on the application     Two sample pump options are available from Campbell Scientific     The RB0021 is used for high flow applications and the XDD1 is  adequate for low flow applications  See Section 4 1 4 4  Sample  Vacuum Pump  for more details on sample pumps     TGA Series Trace Gas Analyzers    Connect a reference gas supply to the reference gas inlet  The  reference gas supply should have an appropriate regulator  flow meter   and needle valve so that it will supply the reference gas at  approximately 10 ml min  The reference gas assembly  pn 15837  is  available from Campbell Scientific to provide the flow meter  needle  valve  and tubing for this connection  See Section 4 1 3 1  TGA  Reference Gas Connection  for specifics about the assembly and  Appendix A 2  Re
185. tive mode  Save to File button checked  will increment the  laser temperature and collect and store a laser current scan at each temperature   Each laser scan will be stored in a separate file  The data are displayed in the  laser map window as they are collected  At the completion of the data  collection  any of the laser scans may be selected by clicking the button next to  the temperature  A typical screen is shown in FIGURE 7 39     83    TGA Series Trace Gas Analyzers             gt   E  o  a  hi     o   gt   o  o  c  o     o     4  o  a       84    7 2 6 Graph      Laser Map  C 8 00  C 9 00    98 100 102 104 106 108  DC Current  mA       FIGURE 7 39  Noninteractive Laser Find window    The Graph tool is used to graphically display real time data from the TGA   The user interface is slightly different for TGA Windows and TGA TEC     TGA Windows   For TGA Windows  click the Graph icon in the toolbar to bring up the Select  Data window  FIGURE 7 40  Click the parameters  shown in bold  to  highlight which to include in the graph  Ifa selection is highlighted by  mistake  click it a second time to toggle it off  The sections for Ramp B   Ramp C  and Isotope Ratios are visible only if the TGA is in multiple ramp  mode  Once the choices are highlighted  click OK to bring up the Graph  window     TGA Series Trace Gas Analyzers    m   Select Data  o     Select Data you would like to graph          El  Ramp  amp   C02 Conc  Mean CO2 Conc  CO2 Conc StdDev  Smp Det Signal CO2  Ref
186. tor for two  reasons  First  the flux density on the reference detector is low due to the  beamsplitter transmitting most of the optical power to the sample detector and  reflecting less than 10  onto the reference detector  Second  the reference  detector is adjusted  see Appendix D 3  Detector Temperature  to give a  relatively low response  The reference detector may have a small amount of  nonlinearity  but this tends to be cancelled by setting the sample detector  linearity coefficient so that the sample detector matches the reference detector   as described below     1  Change the plumbing configuration to ensure the reference and  sample detectors see the same absorption  The step is different for the  different TGA models     TGA100 and TGA100A   Change the tubing connections inside the TGA enclosure such  that the reference gas flows through both short cells in parallel   This can be accomplished by disconnecting the tubing at the inlet  to the reference cell and inserting a tee and two short tubes  connected to the inlets of the reference cell and the short sample  cell     Similarly  disconnect the tubing at the outlet of the reference cell  and insert another tee and two more tubes connected to the outlets  of the reference cell and the short sample cell  This will split the  reference flow to go through the two short cells in parallel  The  sample inlet  long sample cell  should be connected  in the  normal way  to a source of air or nitrogen     This plumbing co
187. ure  e reference detector signal  e reference detector transmittance    The TGA will disable the line lock function if it detects a problem with any of  these values  This prevents the line lock algorithm from misadjusting the DC  current when the position of the absorption line cannot be determined reliably   In this case  the line locking algorithm is disabled temporarily  When the error  condition no longer exists  the line locking is automatically re enabled  The  state of the line lock function is shown by three different icons on the button on  the Settings  gt  Laser  gt  Current window and the Laser window  The  appearance and function of those icons is shown in TABLE 7 1     TGA Series Trace Gas Analyzers       TABLE 7 1  Appearance and Function of Line Lock Icons       Icon Function       e   Open yellow padlock indicates line locking has been turned off             Closed yellow padlock indicates line locking is active          Closed red padlock indicates line locking has been temporarily    disabled by an error condition                Set the Max pressure limit well above the typical operating pressure  refer to  FIGURE 7 12   A typical setting would be twice the operating pressure  This  will disable line locking when the sample pump is turned off     Set the Min signal strength well below the normal value for the reference  detector signal strength  The default is 1 mV  If the reference detector signal  is less than 2 mV  set this parameter to half the typical
188. ven in FIGURE 7 29 and a  description of each of the modes follows     73    TGA Series Trace Gas Analyzers              Laser Settings      View Less Expand Raw v Colors        Laser Settings      View Less Expand Maximum v Colo       Laser Settings  x  View Less Expand Magnified    Color                                         Laser Temp 17 00   17 00   C       Laser Temp 17 00     17 00   C       Laser Settings   i    Laser Settings    View Less Expand Detrended    Colo View Less Expand Folded    Colors View Less Expand Absorbance v Col          CH4       DC current   Mod current  Zero current  High current 0 00    Laser Temp 17 00 7  17 00 C 17 00                                             FIGURE 7 29  Laser display modes    74    TGA Series Trace Gas Analyzers    Raw    Raw displays the raw detector signals  with the Y axis scaled to match the input  range of the electronics  This mode is used to set the zero current and to verify  the detector signals are within the range of the electronics  not saturated   Two  colors are used on the right and left sides to show the range of data actually    used to calculate concentration  blue  on right side   The dashed vertical line  shows the center of the spectral scan        Laser Settings    View Less Expand Raw v Colors       CH4       DC current 88 4  gt  9   Mod current 1 30 H  Zero current 50 00     High current ook              Laser Temp 17 00 H    J gt   ee     17 00   C    1 1          FIGURE 7 30  Raw mode of laser di
189. zed measurement and rapid communication between a Campbell  Scientific datalogger and multiple devices including the TGA100A  TGA200   and TGA200A  The TGA100 requires the CPU upgrade for SDM  communication  Although nearly all Campbell Scientific dataloggers support  SDM  only the CR1000  CR3000  CR5000  and CR6 dataloggers support  communications with TGAs with the TGA   instruction     To configure the TGA for SDM output  connect a PC running TGA  Windows TGA TEC  Open the window Settings    Other    Data Output   Click the SDM output button as shown in FIGURE 7 46  and enter the SDM  address for the TGA        7 3 1 1 Syntax    7 3 1 2 Remarks       TGA Series Trace Gas Analyzers    See the applicable user manual for details on these dataloggers        TGA Parameter Settings x   E  Laser      Temperature     Current    SDM output    Line Lock SDM address for TGA 15  Other   E Detectors  Temperature  PreAmp      Calculations   Channel 1   2   3   4    Concentration  Analog Input   B  Other CH4 Conc            Analog output          TGA Temperatures  Scaling   Serial Numbers 0 00  About TGA                  FIGURE 7 46  Setting TGA for SDM output from TGA Parameter  Settings Window    The datalogger program must include the TGAQ  instruction to retrieve data  from the TGA     The TGAQ instruction is used to measure a TGA100A  TGA200  or    TGA200A trace gas analyzer system  It is also used for TGA100 with the  upgraded CPU module     TGA Dest  SDMA ddress  DataList  ScanMod
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
  HIV-1 p24 Elisa Assay  Samsung Samsung GT-B5722 DUOS Vartotojo vadovas  Sharp SJ-GC480VBK fridge-freezer  EN 60950-1 Standard Upgrade  ITINERARI  COPERTINA pressomat m  Observations sur le muscadier    Copyright © All rights reserved. 
   Failed to retrieve file