2. 44 Limitations of the Procedure Caution The TysonBio TB 100 System is designed for in vitro diagnostic use only and is not intended to test on newborns Any change or administer of medication based on the TysonBio TB100 blood glucose test results without the consent advice of a physician or healthcare professional is not recom mended The TysonBio TB 100 Test Strips are designed for use with fresh capillary whole blood samples obtained from the fingertip palm and forearm False results may occur when performing the test while severely dehydrated se verely hypotensive in shock or in a hyperglycemic hyperosmolar state If you believe you are suffering from any of the above symptoms consult a health care professional immediately Service and Warranty IMPORTANT Tyson Bioresearch Inc cannot endorse the performance of the TysonBio TB 100 System when used with Test Strips other than those de signed for the TysonBio TB 100 Meter The TysonBio TB 100 System manu facturer warranty is valid only when used properly within the guidelines of the provided User s Manual and is invalid when the TysonBio TB 100 System and TysonBio TB 100 Test Strip are used improperly Downloaded from www Manualslib com manuals search engine Manufacturer Warranty Tyson Bioresearch Inc guarantees that this device will be free of defects in materials and workmanship for a period of three years from the date of origi nal purchase During the stated three year pe
3. Downloaded from www Manualslib com manuals search engine Important Information EI Available Alternate Sites Testing Palm and Forearm TysonBio TB 100 Blood Glucose Monitoring System provides you alternate sites testing AST This system provides you to test on the palm and the forearm with the equivalent results to fingertip testing There are limitations for doing AST Please consult your healthcare professional before you do AST Forearm Downloaded from www Manualslib com manuals search engine Alternative site results differ from fingertip results when glucose levels are changing rapidly such as after a meal after taking insulin during exercise or when you are ill or under stress Use blood from a fingertip rather than an alternative site if e you think your blood glucose is low e you don t regularly have symptoms when your blood glucose is low or e how you feel doesn t match the results from the alternative site e Vhen you have just taken insulin or any time during or after exercise e Vhen there are unknown variables occurring in your day such as illness e Vhenever you are about to drive CAUTION 1 Physiologic differences in the circulation between the finger and other test sites like the forearm and palm may result in differences in blood glucose measurements from the other test sites and your fingertips Changes in blood glucose may be observed in finger blood samples sooner than blood
4. 20 Unit Selection 1 Select unit mg dL or mmol L on the back Note The mg dL is the standard unit in the United States Meters used in the United States must be set to read in mg dL TysonBio TB100 Blood Glucose Meter For in vitro diagnostic use only mmol L mg dL Downloaded from www Manualslib com manuals search engine The Control Solution Test The Control Solution is used to check and ensure that the TysonBio TB 100 Meter and TysonBio TB 100 Test Strip are working together properly and that you are performing the test correctly When to Perform a Control Solution Test 1 When you are using your Blood Glucose Monitoring System for the first time 2 When you are using a new batch of Test Strips 3 Anytime you question the performance of the system or on a regular basis to ensure accuracy e g once a week 4 When you adjust your diabetic medication plans 5 When your blood glucose test result is lower or higher than your normal level Note There are three Control Solutions at different desired range of aqueous glucose Low Normal or High For more information about the Control Solu tion please read your Control Solution package insert Two sets of control solution should be used a high level and a low level control solution of your choice For order information please contact Customer Service Downloaded from www Manualslib com manuals search engine 22 To Perform a Control Solution Test 1 Make
5. Test Port This is where you insert the Test Strip into the Meter Unit Switch Switch for mmol l or mg dL 1 mmol l 18 mg dL Battery Cover Label reset Button Press this button to re set the system Data Port Connect interface cable to transfer data OR od Downloaded from www Manualslib com manuals search engine About the TysonBio TB 100 Meter Display 1 Date O 2 Time 3 Control Solution Test x Appears when doing a control solution test and marking the result as a control solution test 4 Memory Symbol Indicates in memory mode 5 Units of Measurement Unit of mmol l or mg dL will ap pear with the test result Downloaded from www Manualslib com manuals search engine 6 Battery Symbol Appears when the battery is low or must be replaced 7 Blood Drop Symbol Flashes when it is ready to apply sample to Test Strip 8 Warning Symbol Appears when result is out of the range of your setting reference value 9 Alarm Function Symbol Appears when the clock alarm is set 10 Test Strip Symbol Appears to tell you when the meter is ready for test 11 Day Average Symbol Shows the average value of test results 12 Test Result Area Display test results 13 Temperature Message Appears if your meter is out of operating temperature range Downloaded from www Manualslib com manuals search engine About the TysonBio TB 100 Test Strip 13 The TysonBio TB 100 Test Strip is used to perform the blood
6. 26 Preparing the Test Strip 1 Wash hands using soap and warm water Rinse and dry thor oughly 2 Pull out a Test Strip from the vial and re cap the vial immedi ately 3 Insert the Test Strip white side face up into the test port 4 Abeep will sound and the sort line moving clockwise will ap pear followed by a flashing symbol Now you are ready to obtain a blood sample Downloaded from www Manualslib com manuals search engine HIN TN Obtaining a Blood Sample SAMPLE MAY BE OBTAINED FROM FINGER PALM or FOREARM For further instructions please see the insert provided with lancing device 1 Unscrew and remove the endcap of the lancing device 2 Insert the lancet into the lancet holder 3 Twist and remove the protective lancet cover to expose the sterile needle tip 4 Recapping and screwing the endcap to the body of the lancing device 5 Pull back on the cocking mechanism until it comes to a stop 6 Place the lancing device firmly against the finger tip and press the release button 7 After sampling remove the endcap containing the used lancet carefully 8 Remove the lancet from the lancing device Discard the used lancet prop erly in accordance with local state and federal law Downloaded from www Manualslib com manuals search engine 28 Applying Blood Sample to the TysonBio TB 100 Test Strip 1 After obtaining a blood sample place the drop onto the test wel
7. To set the third alarm press the M button y If you press S Fig 6 button n it will be moved on to health management alarm yf setting step 11 To change the hour press S button To c al enter the hour and move on to change the minutes press val M button To change the minutes press S button To enter the minutes and move on to choose the fourth alarm setting 330 press the M button Fig 7 o Fig 7 Downloaded from www Manualslib com manuals search engine 10 To set the fourth alarm press the M button y If you 11 press S button n it will be moved on to health manage ment alarm setting step 11 To change the hour press S button To enter the hour and move on to change the minutes press M button To change the minutes press S button To enter the minutes and move on to health man agement HI alarm setting press the M button Fig 8 Note When the alarm is ringing press S or M button to turn off or alarm will turn off automatically after 30 sec To change the HI alarm setting press the S button To enter the HI alarm setting and move on to LO alarm set ting press M button Fig 9 12 To change the LO alarm setting press the S button Fur ther press M button to enter the LO alarm setting and finish the meter setting turn off the meter Fig 9 Downloaded from www Manualslib com manuals search engine
8. When the 14m 113 memory is full the oldest result is removed and replaced with the newest result m 7 Press and release the S button to the end of the memory r ti t i 000 will show Then press the S button again or press M 7 7 7 button at anytime to turn off the meter Exiting the Memory Mode 1 Press the M button to turn off the meter 2 Meter will shut down automatically after 3 minutes with no further action Note The consecutive 7 14 30 day average is calculated from the blood glucose results obtained during the last consecutive 7 14 30 day To Record Your Results We have provided a log book for your recording convenience To learn more about how to record your results please refer to the log book for further instructions Downloaded from www Manualslib com manuals search engine Transfer Test Results to a Computer 33 You can use your meter with TysonBio Link Health Management Software to transfer test results to your personal computer 1 Obtain the required software and cable For order information please call Customer Service or visit website of www tysonbio com 2 Install the software on a computer Follow the instructions provided with software to install the software 3 Get ready to transfer test results Connect the interface cable to a serial port on your computer Then connect the other end of interface cable to the data port located on the side of the meter with the meter turned off P
9. sure the Control Solution is at room temperature 15 C to 35 C or 59 F to 95 F prior to testing 2 Insert a Test Strip white side face up Contact Bar s end first into the test port The Meter will automatically turn on Fig 10 3 All segments of the LCD display will appear a beep will sound and mov ing dotted lines will appear followed by a flashing symbol Fig 11 4 After the symbol appears press the S button for 3 sec to switch control solution mode The ctl symbol indicates that the Meter will mark your current test as a Control Solution test Now you are ready to apply the Control Solution Fig 12 Fig 11 Fig 12 Downloaded from www Manualslib com manuals search engine 5 Discard the first drop of Control Solution and squeeze a small drop of Control Solution on a clean nonabsorbent surface such as a clean piece of wax paper Do not apply Control Solution to the Test Strip directly from the bottle 6 Bring the tip of the Test Strip to lightly touch the drop of ca ong Control Solution Control Solution is automatically pulled Tad into the strip through the tip Hold until the meter beeps ic The meter will now being counting down from 5 to 1 and i y the Control Solution test result will appear Caution Please remember that in order to prevent contamination follow the above instructions for performing the Control Solution tes
10. C will appear on the display indicating that the meter is in the communication mode 4 Transfer data Follow the instructions provided in the software to download the results from the meter After finishing the data transfer press M button to turn off the meter Note While in the communication mode you will be unable to perform a blood glucose test If the meter is not in the PC link mode it will not respond to computer commands Downloaded from www Manualslib com manuals search engine 34 Care and Storage 1 Handle the Meter with care Dropping or throwing the Meter may cause damage to the device 2 Do not expose the Meter Test Strips and Control Solution to extreme con ditions such as high humidity heat freezing cold or dust 3 Always wash hands with soap and water and rinse and dry completely be fore handling the Meter and Test Strips 4 When cleaning the Meter gently wipe the exterior surface using a damp soft cloth DO NOT USE ANY ORGANIC SOLVENT for cleaning 5 The Meter should be stored at room temperature in a dry and clean space DO NOT STORE IN DIRECT SUNLIGHT OR AREAS WITH HIGH HU MIDITY AND OR DUST It is advised that you store the Meter and its ac cessories into the provided carrying case For healthcare professionals using this system on multiple patient please be aware that all items that come in contact with human blood should be handled as potential biohazards Users should follow the guidelin
11. TysonBio Blood Glucose Monitoring System User s Manual For in vitro diagnostic use only temi TB 100 tz B OO User s Manual Downloaded from www Manualslib com manuals search engine Table of Contents a important IM Ornato Ms ana 3 INtrOdUCUON aa 7 About the PIOdUGL Lillian 8 About the TysonBio TB 100 Meter i 9 About the TysonBio TB 100 Meter Display 11 About the TysonBio TB 100 Test Strip 13 About the TysonBio TB 100 Test Strip Vial i 14 Installing Replacing the Battery 15 sening Up the Meter reni as 16 Setting Up the Date TIME ii 16 Soning UPATI PRO RR O e CMe n Or Hane ye orn wanacr ee rsren Sener 17 URIFSCI ERO 20 Downloaded from www Manualslib com manuals search engine The Control Solution Test iii 21 Testing Your Blood Glucose Level xrrrrrrrrrriiiiie 25 Preparing the TEStStipalha rali ile ail 26 Obtaining a Blood Sample e 27 Applying Blood Sample to the TysonBio TB 100 Test Strip 28 Memory Feallufes a JJ 31 Transfer Test Results to a Computer iii 33 Careand St0rage nl aa 34 Display Messages 35 Specifications sicario ii 42 Limitations of the Procedure i 44 Service and Warrant 44 CUSIOME SEVICE sia dirai 46
12. ange Expiration Date Downloaded from www Manualslib com manuals search engine Installing Replacing the Battery When the battery is low the Meter will show a warning signal When this display appears it is time to replace a new battery When the low battery warning signal is shown the Meter will provide accurate results for ap proximately 50 more measurements How to Replace the Battery 1 From the back of the Meter gently slide and remove the battery cover 2 Turn over the battery cover and insert a 3 Volt Lithium battery CR 2032 with the side facing down 3 Slide battery cover back into place with the battery side facing down Downloaded from www Manualslib com manuals search engine 16 Setting Up the Meter To turn the Meter on first insert a battery then press the reset button located at the back of the Meter and proceed with the following instructions Setting Up the Date Time 1 Press the M button to enter this mode Press and release S button to choose 24hr or 12hr display With your favorite clock mode press M button and the year setting Fig 1 will be appeared Press and release the S Ma button until the correct year appears Fig 1 2 To enter the year and move on to choose the month press the M button To change Fig 2 the month press the S button Fig 2 3 To enter the month and move on to choose the day press the M button To change the day
13. cifications Meter Operating Conditions Temperature 10 C 40 C 50 F 104 F Humidity 10 90 RH Hematocrit 35 55 Test Sample Capillary Whole Blood from Fingertip Palm and Forearm Sample Volume gt 0 5 pl Measuring Unit mg dL or mmol L Measuring Range 20 600 mg dL 1 1 33 3 mmol L Test Time 5 sec Memory Capacity 300 most recent results Average 7 14 30 days average results External Output Standard RS232 PC interface Power supply 3 Volt lithium battery CR 2032 Battery Life Approximately 1000 tests Dimension and Weight 92 x58 x19 mm 60 g Downloaded from www Manualslib com manuals search engine m The TysonBio TB 100 Meter TysonBio TB 100 Test Strip and Control 43 Solution are in conformity with the IVDD 98 79 EC The Lancing device is in conformity with the MDD 93 42 EEC Manufacturer EU authorized representative Tyson Bioresearch Inc Medical Device Safety Service GmbH 5F 22 Ke E Road III Science Based Schiffgraben 41 D 30175 Hannover Industrial Park Chu Nan Germany Miao Li County Taiwan R O C TEL 49 511 62628630 TEL 886 37 585998 m The Lancets are in conformity with the MDD 93 42 EEC Manufacturer EU authorized representative Wuxi Xinda Medical Device Co Ltd EMDAR BV No 42 Xixin Road Zhangjing Town ljsselburecht 3 Postbus 5486 Wuxi Jiangsu 214194 China 6802 EL Arnhem Netherlands TEL 86 510 379 3149 TEL 31 263 653 375 Downloaded from www Manualslib com manuals search engine
14. es for prevention of blood borne transmittable diseases in a healthcare setting for potentially infectious human blood specimens as recommended in the National Committee for Clinical Laboratory Standards Protection of Laboratory Workers from Instrument Biohazards and Infectious Disease Transmitted by Blood Body Fluids and Tissue Approved Guideline NCCLS document M29 A ISBN 1 56238 339 6 Downloaded from www Manualslib com manuals search engine Display Messages DISPLAY WHAT IT MEANS ACTION gaggi ga gguw System check for verifying No action necessary o that all segments of the 0 ud Meter are functioning li nt Bn The system is ready to ac You may now apply a cept a blood sample blood sample The system is ready to You may now apply a accept a Control Solution drop of Control Solution sample sample Downloaded from www Manualslib com manuals search engine Indicate the result is HIGHER than the HI alarm setting of the health management Indicate the result is LOWER than the LO alarm setting of the health management Meter detects blood glu cose level is higher than 600 mg dL 33 3 mmol l Downloaded from www Manualslib com manuals search engine DISPLAY WHAT IT MEANS ACTION For your reference or you can change the default setting value 100 mg dL according to page 19 For your reference or you can change the default set ting va
15. eter will provide approximately 50 more measurements Battery power is too low for further usage Downloaded from www Manualslib com manuals search engine Press reset button and check the meter again with the test strip If the problem persists contact Customer Service for help Replace with a new 3 Volt lithium battery CR 2032 Replace with a new 3 Volt lithium battery CR 2032 immediately 5 30 All testing values recorded Press M or S button or wait in the memory have been for 3 minutes to turn off the shown meter The setting alarm is ring Press M or S button to shut ing to remind of doing the off or the alarm will turn off blood glucose test automatically after 30 sec Downloaded from www Manualslib com manuals search engine Other problems that may occur The Test Strip has not been inserted Review the instructions and re insert into the Meter properly the Test Strip correctly white side face up Defective Test Strip Replace with a new Test Strip Insufficient blood sample Repeat the test with new Test Strip Test Strip remains in the test port for Meter will automatically turn off Re more than 3 minutes prior to testing insert the Test Strip to the test port LCD display on the Meter is blank Contact Customer Service for help when trying to perform a test Downloaded from www Manualslib com manuals search engine 42 Spe
16. glucose test in conjunction with the Tyson Bio TB 100 Meter Each strip can be used only once The Test Strip consists of the following parts 1 Confirmation Window This is where you can confirm that enough blood has been applied to the target area 2 Target Area This is where blood is drawn into the Test Strip 3 Contact Bar This end of the Test Strip is inserted into the Test Port to activate the Meter white side face up Downloaded from www Manualslib com manuals search engine E About the TysonBio TB 100 Test Strip Vial VIAL LABEL TEST STRIP VIAL TysonBio 1B100 Blood Glucose Test Strips Store at 4 30 C 39 86 F Do not refrigerate Use within 90 days after first opening Do not store test strips outside this vial Re cap immediately WARNING Always close lid tightly For testing of glucose in capillary whole blood For in vitro diagnostic use Manufacturer Tyson Bioresearch Inc 5F 22 Ke E Road Ill Science Based Industrial Park Chu Nan Taiwan Tel 886 37 585998 50 TEST STRIPS Use TysonBio TB 100 meter only Lot H8T121205 EXP Aug 2011 Control Solution Range mg dL mmol l Low 31 61 1 7 3 4 Normal 84 144 4 7 8 0 High 348 438 19 3 24 3 39 86 F Do not refrige t strips outside this vial tely ys close lid tightly cose in capillary whole bl ostic use Inc Ill Science k Chu Nan 37 585998 Lot Number Control Solution R
17. l of the Test Strip Blood is automatically drawn into the Test Strip 2 As soon as enough blood has filled the Confirmation Window see picture of the Test Strip the Meter will beep and begin counting down from 5 to 1 Your blood glucose result will appear on the LCD display and will be stored into the Meter memory automatically pl Xi X Note To ensure accurate results make sure the Confirmation Window of the Test Strip is completely filled with your blood sample Downloaded from www Manualslib com manuals search engine 3 After the test result appears if no further tests are performed the Meter will automatically shut off after 3 minutes After finishing the test we rec ommend using a tissue to remove the used strip from the Meter for proper disposal Downloaded from www Manualslib com manuals search engine 30 HI and LO Readings 1 The test results range between 20 to 600 mg dL 1 1 to 33 3 mmol l If a HI message appears on the display your Meter has detected that your blood glucose level is higher than 600 mg dL 33 3 mmol l It is suggested that you review your testing procedure and test again with a Fig 13 new Test Strip to confirm the result If the same result oc curs consult your healthcare professional immediately Fig 13 2 If a LO message appears on the display your Meter has detected that your blood glucose level is lower
18. lue 70 mg dL accord ing to page 19 Indicates high blood glu cose level Repeat test If result still occurs consult your physician immediately Meter detects blood glu cose level is lower than 20 mg dL 1 1 mmol l Temperature is high during the test procedure Temperature is low during the test procedure Downloaded from www Manualslib com manuals search engine Indicates low blood glucose level Repeat test If result still occurs consult your physician immediately Result may be larger varia tion Move to a cooler envi ronment 10 40 C or 50 104 F and wait 15 minutes before re testing Result may be larger varia tion Move to a warmer environment 10 40 C or 50 104 F and wait 15 minutes before re testing DISPLAY WHAT IT MEANS ACTION Temperature is too high to Repeat the test in a cooler perform the test setting 10 40 C or 50 104 F Wait 15 minutes before re testing Temperature is too low to Repeat the test in a warmer perform the test setting 10 40 C or 50 104 F Wait 15 minutes before re testing An error message indicat Review the instructions ing a problem with the and try again with a new Test Strip Test Strip Downloaded from www Manualslib com manuals search engine DISPLAY WHAT IT MEANS ACTION An error message indicat ing a problem with the Meter Battery power is low M
19. nitor whole blood glucose levels obtained from the fingertip palm and forearm This is an over the counter OTC product Please read the entire User s Manual carefully before using this product Downloaded from www Manualslib com manuals search engine 3 About the Product Your TysonBio TB 100 System is made up of several components This booklet is designed to guide you through the operation of the system with detailed directions and pictures Your TysonBio TB 100 System contains TysonBio TB 100 Meter One vial of TysonBio TB 100 Test Strips 10 strips One Lancing Device option Ten Lancets option One Control Solution One carrying case User s Manual Log book option Quick Reference Guide Note 25 50 test strips vial can be purchased individually Downloaded from www Manualslib com manuals search engine About the TysonBio TB 100 Meter FRONT Downloaded from www Manualslib com manuals search engine BACK TysonBio TB100 Blood Glucose Meter For in vitro diagnostic use only 7 Display Your test results are displayed here The large easy to read screen guides you through the test by using numbers and symbols 2 M Button The M mode button is used to set time alarm enters the memory mode to show the consecutive 7 14 30 day average results and turns the Meter on and off 3 S Button The S select set button is used to select year month day time memory of results and control solution mode
20. press the S button Fig 3 Fig 3 Downloaded from www Manualslib com manuals search engine 4 To enter the day and move on to choose the hour setting eo press the M button Fig 3 5 To change the hour press the S button To enter the hour and move on to change the minutes press the M button Fig 4 6 To change the minutes press the S button To enter the minutes and move on to choose the alarm setting press JED 2 ru the M button Fig 4 od Setting Up Alarm t 4 7 To set the first alarm press the M button y If you press S button n it will be moved on to health manage 5 30 ment alarm setting step 11 To change the hour press o l S button To enter the hour and move on to change the minutes press M button To change the minutes press S button To enter the minutes and move on to choose the Fig 5 second alarm setting press the M button Fig 5 l Downloaded from www Manualslib com manuals search engine 8 To set the second alarm press the M button y If you 4E amp S fa press S button n it will be moved on to health manage a ment alarm setting step 11 To change the hour press S button To enter the hour and move on to change the min utes press M button To change the minutes press S button LIU To enter the minutes and move on to choose the third alarm c setting press the M button Fig 6 9
21. riod our company shall repair or replace any TysonBio TB 100 Meter found defective with a new TysonBio TB 100 Meter This warranty does not apply to the performance of a TysonBio TB 100 Me ter that has been accidentally damaged altered misused tampered with or abused in any way In no event shall our company be liable to the purchaser or any other person for any incidental consequential or punitive damages arising from or in any way connected with the purchase or operation of the TysonBio TB 100 Meter or its parts For manufacturer warranty services purchaser must contact Tyson Bioresearch Inc for help Downloaded from www Manualslib com manuals search engine 46 Customer Service If you have any questions or concerns regarding this product or its operation or any attempt to correct a problem fails please call our Customer Service Our trained specialists will be happy to assist you train you or even reaffirm your results Customer Service Hotline 1 800 948 3894 U S 24 hours a day 886 37 585998 or E Mail us at service tysonbio com website www tysonbio com When you call our Customer Service please have your TysonBio TB 100 Meter TysonBio TB 100 Test Strip and all other system supplies available This will allow us to answer any of your questions with speed and efficiency Downloaded from www Manualslib com manuals search engine CE 0597 B03Q106847 01 Downloaded from www Manualslib com manuals search engine
22. samples from the forearm and other alternate sites Rub the alternate test sites about 20 seconds before lancing If you are testing for hypoglycemia low blood glucose or if you suffer from hypoglycemia unawareness we recommend that you test on your fingertips Downloaded from www Manualslib com manuals search engine 2 Talk to your doctor to see if alternate site testing is right for you With a little bit of education you can give your fingertips a rest and maybe test more often than you do now For people with diabetes more frequent test ing is a good thing Just remember any time you want to be sure of an ac curate up to date blood glucose reading test on your fingertip 3 We strongly recommend you do AST ONLY in the following intervals e In a pre meal or fasting state more than 2 hours since the last meal e Two hours or more after taking insulin e Two hours or more after exercise DO NOT use AST if e You think your blood glucose is low e You are unaware of hypoglycemia e Your AST results do not match the way you feel e You are testing for hyperglycemia e Your routine glucose results are often fluctuating e f you are pregnant Downloaded from www Manualslib com manuals search engine e Low or high blood glucose readings can indicate a potentially serious med ical condition If your blood glucose reading is unusually low or high or if you do not feel the way your reading indicates repeat the test with a new tes
23. t Downloaded from www Manualslib com manuals search engine Comparing Control Solution Results The TysonBio TB 100 is functioning properly if the result falls inside the specified range printed on the Test Strip vial If the test result falls outside the specified range repeat the test Results falling outside the specified range may be caused by e Error in performing the test e Control solution temperature is lower than 15 C 59 F or higher than 35 C 95 F e Expired or contaminated Control Solution e Expired or contaminated Test Strips e Meter malfunction Note The result will not be incorporated into the average in the Control Solution test ctl mode DO NOT use the system if the problem persists Please contact Customer Service or website of www tysonbio com Downloaded from www Manualslib com manuals search engine Testing Your Blood Glucose Level 25 Before testing blood glucose you need the following items 1 TysonBio TB 100 Meter 2 TysonBio TB 100 Test Strip 3 Adjustable Automatic Lancing Device 4 Sterile Lancet Caution To reduce the chances of infection e Never share a lancing device and lancet with another person e Always use a new and sterile lancet Lancets are for single use only Always use a new Test Strip e Test Strips are for single use only e Do not get lotion oil dirt or debris in or on the lancet and lancing device Downloaded from www Manualslib com manuals search engine
24. t strip If your reading is not consistent with your symptoms or if your blood glucose result is less than 60 mg dL 3 3 mmol L or higher than 240 mg dL 13 3 mmol L you should contact your healthcare professional and follow his or her treatment advice e Any change or administer of medication based on the TysonBio TB 100 blood glucose test results without the consent and advice of a physician or healthcare professional is not recommended Severe dehydration and excessive water loss may cause false low results If you believe you are suffering from severe dehydration consult your phy sician immediately Downloaded from www Manualslib com manuals search engine Introduction Thank you for choosing the TysonBio TB 100 Blood Glucose Monitoring Sys tem the latest advancement in biosensor technology As you already know self monitoring of blood glucose SMBG is a necessary part of the treatment plan of people with diabetes mellitus The Diabetes Control and Complica tions Trial DCCT has confirmed the significant benefits from SMBG when practiced as part of a larger intensive strategy to tightly control blood glucose concentrations Adapted for its ease of use and quick response time the TysonBio TB 100 System has the ability to process accurate results utilizing only a small vol ume of blood for in vitro diagnostics The TysonBio TB 100 Blood Glucose Monitoring System is intended for use in the home and in professional set tings to mo
25. than 20 mg dL 1 1 mmol l It is suggested that you review your testing procedure and test again with a new Test Strip to confirm the result If the same result occurs con Fig 14 sult your healthcare professional immediately Fig 14 nd lt EN e aw Downloaded from www Manualslib com manuals search engine Memory Features The TysonBio TB 100 Meter automatically stores the 300 most recent blood glucose results with date and time in its memory It also provides you with a consecutive 7 14 30 day average of your blood glucose test results 1 Press M button to turn on the Meter All segments of the LCD display will appear The date and time will appear first 2 Press S button the result of control solution will appear on display 3 Press S button the consecutive 7 day average will be shown 4 Press S button the consecutive 14 day average will be shown Tac 5 Press S button the consecutive 30 day average will be shown 02 4 ng dl Note If there is no test results during this stage it will directly appear the most recent test result TAY Downloaded from www Manualslib com manuals search engine 32 6 Press S button the most recent test result will appear If it G53 gog has 14 day average result it will show under the screen at D the same time Each time you press the S button the meter ul will recall the last 300 test results accordingly